HD 9212
INSTRUCTIONS MANUAL
– 15 –
HD 9212
ENGLISH
pHMETER THERMOMETER
GENERAL CHARACTERISTICS
The instrument allows measurements of pH, mV or temperature to be ta-
ken with a passive 4-wire Pt100 probe of the series TP9...
pH measurement
It is possible to take pH measurements by connecting a pH electrode to
the instrument (Fig. 1).
While measuring pH the instrument performs temperature compensa-
tion with the set temperature
value.
pH measurement is possible
in the range from -2.00 to
16.00 pH; the work range is
from 0.00 to 14.00 pH and
the value is indicated in
hundredths of pH.
If the measurement exceeds
the indicated limits or the
electrode is not connected,
the Err code appears on the
display.
mV measurement
When a pH electrode, an
electrode for measuring the
redox potential or a specific
ion electrode is connected,
the instrument performs mV
measurement (Fig. 1).
mV measurement is in the
range ±1999 mV. The mV
resolution of the instrument is
0.1 mV in the range ±199.9
mV and 1 mV in the remain-
ing range.
If the measurement exceeds
the indicated limits or the
electrode is not connected,
the Err code appears on the
display.
ENGLISH
– 16 –
EURO S7 CAP
Fig. 1
Temperature measurement
Temperature measurements may be taken by connecting to the instru-
ment a 4-wire Pt100 probe of the series TP9... (Fig. 2).
Temperature measurement is possible in the range from -50.0°C to
-199.9°C. The resolution of the instrument is 0.1°C.
The temperature
may be displayed
in either Centi-
grade or Fahren-
heit degrees.
If no temperature
probe is fitted
(and the pH or
mV probe is fit-
ted), the compen-
sation tempera-
ture of the pH
electrode is dis-
played, the set-
ting of which is
described below.
If the temperature
m e a s u r e m e n t
exceeds the indi-
cated limits or the
electrode is not
connected, and
there is no pH or
mV probe con-
nected, the Err
code appears on
the display.
Switching the instrument on and off
The instrument is switched on and off by pressing the On/Off button.
While switching on the instrument, all the symbols light up on the dis-
play for 2 seconds (test display), and the set calibration code of the
instrument (with the messages C01, C02 or C03), used for measure-
ment, and the pH measurement value are shown.
Auto power off
The instrument is provided with an auto power off function which switch-
es it off 8 minutes after the last button has been pressed.