User's Guide
High Resolution Benchtop MilliOhm Meter
Models 380560 and 380562
Introduction
Congratulations on your purchase of the Extech 380560 (117V) or 380562 (220V) High
Resolution MilliOhm Meter. This device offers seven resistance ranges with resolution as
low as 0.01mΩ. The 4-wire Kelvin clip connection ensures optimum accuracy. The built-
in comparator feature offers HI-LO-GO testing. Typical applications include transformer,
motor coil, and PC Board resistance measurements. Careful use of this meter will
provide years of reliable service.
380560, Version 1.6, 04/03
2
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one
year from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to
return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781)
890-7440 ext. 210 for authorization or visit our website at www.extech.com (click on ‘Contact Extech’ and go to ‘Service
Department’ to request an RA number). A Return Authorization (RA) number must be issued before any product is returned
to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in
transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation
outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any
implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental
or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth
above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Specifications
General Specifications
Circuit
Custom one-chip LSI microprocessor circuit
Display
0.7" (18 mm) 2000 Count LED Display
Connection type
4-Terminal Kelvin
Measurement ranges
Seven ranges (see listing below)
Test voltage
5V DC
Comparator
Built-in HI/LO/GO testing with audible beeper
Zero adjust
Automatic (no adjustment necessary)
Over-range indication
Display reads "1____" when the reading is out of range
Operating Temperature
32°F to 122°F (0°C to 50°C)
Operating Humidity
Max. 80% RH
Power Supply
110V (380560) or 220V (380562) ±15%, 50/60Hz
Weight
4.85 lbs (2.2kg)
Dimensions
11 x 8 x 3.5" (280 x 210 x 90mm)
Range Specifications
Range
Resolution
Test Current
Accuracy (%rdg)
Test Voltage
20 mΩ
0.01mΩ
1A
± (0.2% + 6 digits)
2.7V DC
200 mΩ
0.1mΩ
1A
3.3V DC
2 Ω
.001Ω
0.1A
3.5V DC
20 Ω
.01Ω
10mA
4.1V DC
200 Ω
0.1 Ω
1mA
2K Ω
.001kΩ
0.1mA
20K Ω
.01kΩ
10uA
± (0.2% + 4 digits)
4.5V DC
Note: Specifications based on RF Field Strength <3V/m and frequency <30MHz
International Symbols
380560, Version 1.6, 04/03
3
F+
S+
F-
S-
White Bands
Meter Description
1.
LED Display
2.
HI/LO/GO Status Indicators
3.
Power and Range pushbuttons
4.
Kelvin clip leads
5.
SET/TEST and HI/LO pushbuttons
6.
Comparator adjustment screws
7.
Kelvin clip lead input terminals
Note: The power cable input and audible alert button are on the rear of the instrument.
Measurement Precautions
Ensure that the meter is connected to the correct power source
(110V for model 380560 or 220V for model 380562).
Do not apply voltage to the meter input terminals. Meter damage may result.
Measurement Procedure
1. Connect the Kelvin test leads to the meter.
2. Press the POWER pushbutton to turn power on.
3. Position the SET/TEST pushbutton to the TEST position.
4. To check the meter zero, clip the test leads
together as shown at right.
5. Select the desired measuring range using
the labeled black pushbuttons. When the resistance of the device is unknown, start
with the highest range and work downward.
6. Clip the leads onto the device under test as shown below.
7. Observe the reading on the LED display.
1
2
3
4
5 6 7
Rx
F + S+ S - F -