Gait Up SA
EPFL Innovation Park, Bat C. CH-1015 Lausanne
www.gaitup.com
TVA CHE-145.136.760
PhysiGait user manual v1.3.1 Date: 29.04.2019
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
1
User manual and Outcome parameters
PhysiGait Live
Thank you for purchasing PhysiGait Live, the easy-to-use system to accurately measure
spatio-temporal gait performances in record time. Physilog®5 sensors measure the
movement in the field while a dedicated mobile application computes the relevant
information to offer immediate results with in-lab precision. This manual introduces how to
use the sensors and mobile application to perform gait analysis.
Gait Up SA
EPFL Innovation Park, Bat C. CH-1015 Lausanne
www.gaitup.com
TVA CHE-145.136.760
PhysiGait user manual v1.3.1 Date: 29.04.2019
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
2
Table of Contents
1.
PhysiGait Live elements ......................................................................................... 3
2.
Gait recording with Physilog®5 from tablet – Basics ............................................... 3
3.
Recording with Physilog®5 and the PhysiGait Live - details ................................... 4
3.1
Connection to mobile application ............................................................................ 4
3.2
Attachment ............................................................................................................. 5
3.3
Subject information, selection of options and comments ........................................ 5
3.4
Start the recording .................................................................................................. 6
3.5
Walk ........................................................................................................................ 6
3.6
Stop the recording .................................................................................................. 7
3.7
Show and compare gait reports .............................................................................. 7
4.
Gait Up Dashboard ................................................................................................. 9
5.
Practical Considerations ....................................................................................... 10
5.1
Charging Physilog®5 ............................................................................................ 10
5.2
Physilog®5 configuration for gait measurements ................................................. 10
5.3
Handling / Maintenance ........................................................................................ 11
5.4
Troubleshoot ......................................................................................................... 12
6.
Gait Parameter Overview ..................................................................................... 14
6.1
Essentials ............................................................................................................. 14
6.2
Temporal Analysis ................................................................................................ 14
6.3
Spatial Analysis .................................................................................................... 15
6.4
Turning Analysis ................................................................................................... 16
6.5
Clearance Analysis ............................................................................................... 16
7.
Limited Warranty & Support Policy ....................................................................... 17
8.
Certification claims for Physilog®5 sensors ........................................................... 18
Contact information ........................................................................................................ 19
Gait Up SA
EPFL Innovation Park, Bat C. CH-1015 Lausanne
www.gaitup.com
TVA CHE-145.136.760
PhysiGait user manual v1.3.1 Date: 29.04.2019
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
3
1. PhysiGait Live elements
2. Gait recording with Physilog®5 from tablet – Basics
Attach
Fix the sensors on the feet/
shoes using the rubber clips
or foot attachment straps.
Start
Select options and start
the recording from the
application. Sensors shine
green and start to blink.
Connect
Enter login information,
go to Gait analysis and
select “New measure”.
Connect the 2 Physilog® 5
to the application.
Walk
Wait until the two sensors
blink green synchronously
and ask the subject to walk
according to your protocol.
Stop
Stop the recording from
the application, the LED
turns orange and twinkles
while sending remaining
data to the tablet.
Display report
Once the files are
transferred to the
application and from there
to the server, the Gait report
is available.
Gait Up SA
EPFL Innovation Park, Bat C. CH-1015 Lausanne
www.gaitup.com
TVA CHE-145.136.760
PhysiGait user manual v1.3.1 Date: 29.04.2019
Copyright © 2018, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
4
3. Recording with Physilog®5 and the PhysiGait Live - details
3.1
Connection to mobile application
Each time you open the application from completely closed state you need to connect the
two Physilog®5 to the mobile device. Once they are connected you don't need to repeat this
procedure when switching between different parts of the application.
To connect Physilog®5 to the application for gait analysis, login with your user name (email
address) and password and enter the “Gait analysis” functions of the app. Allow Bluetooth
to be turned on the device and establish a connection to the internet. There is no need to
pair your Physilog® with the tablet prior to Bluetooth communication.
Click on the “Available Physilogs” refresh button below the picture of
the Physilog®5 on the left of the screen of the application to look for
available sensors. Chose two Physilog® from the list and click on the
“Connect” button beside the sensor name. The name of the sensor
contains the number which is written on the back of the Physilog®5.
Troubleshoot: If a sensor is not detected by the tablet, shortly press
the main button of the sensor to enter detection mode. If the sensor can still not be
connected, check that it has enough battery and is not connected to another mobile device
and if necessary ultimately do a reset of the sensor as explained in section 4.4 Troubleshoot.
If the sensor has incompatible firmware version for using with the app, the app displays an
error message and automatically disconnects from the sensor. If the sensor is blinking after
disconnection, you need to do a soft reset of the sensor (see section 4.4 Troubleshoot).
Physilog5 user manual explains how to update firmware.
When the sensors are connected the button changes to “Disconnect”
and a green dot appears besides the sensor name. You can see a
battery level estimation for each sensor below the battery symbol and
you can check if the sensor is measuring correctly by displaying the
raw data inside the Signals Visualizer part of the app.