Operating manual [PDF] (7 pages)
- Updated: 17.03.2026
Document Preview & Text Transcription
T14.0.0X.6C-02
Operating Manual for conductivity-transmitter
GLMU 200 MP
1 Specification
Measuring range 1 (one of the following can be configured):
Range (rAnG)=0
Range (rAnG)=1
Range (rAnG)=2
Range (rAnG)=3
Conductivity
0.0 ... 200.0 mS/cm
0.00 ... 20.00 mS/cm
0 ... 2000 µS/cm
0.0 ... 200.0 µS/cm
Resistivity
5.0 ... 100.0 Ohm·cm
50 ... 1000 Ohm·cm
0.50 ... 10.00 kOhm·cm
5.0 ... 100.0 kOhm·cm
TDS
-
-
0 ... 2000 mg/l
0.0 ... 200.0 mg/l
Salinity
0.0 ... 70.0
-
-
-
Meas. range 2 Temperature: -5.0...140.0 °C (resolution 0.1°C or 0.1°F, NTC10k)
Attention: Consider application range of measuring cell: 0..80°C
Accuracy: (at nominal temperature = 25°C)
Measuring:
conductivity:
±0,5% of measured value ±0,3 FS
temperature:
±0.1°C of measured value ±1 Digit
Add. output signal:
±0.2 % FS
Measuring cell:
4 electrode measuring cell with temperature sensor NTC10k,
7 pole connector
Cell constant K: 0.45..0.65, determined and adjusted ex works
Temperature compensation: non linear function, linear function (coefficient configurable) or none
Min-/Max-Value Memory:
Min and max measured values are stored
Output signal:
refer to type plate, freely scaleable
Scaling:
By entering display values for 4mA (or 0V) and 20mA (or 1V/10V) output
Connection:
4 - 20 mA (2-wire) – output signal electrically isolated
for option AV01, AV10:
0 - 1 (10) V (3-wire) – output signal electrically isolated
Auxiliary energy: (supply voltage)Uv = 12 - 30 V DC (4-20mA)
Uv = 12 - 30 V DC, max. 10mA (0-1V)
Uv = 18 - 30 V DC, max. 10mA (0-10V)
or refer to type plate
Reverse voltage protection: 50V permanent
Perm. impedance (at 4-20mA): RA(Ohm) < ( (Uv - 12V) / 0.02A )
Example: for Uv = 18V: RA < (18V - 12V) / 0.02A => RA < 300 Ohm
Permissible load (at 0-...V): RL(Ohm) > 3000 Ohm
Adjusting:
via keypress. Conductivity by editing cell constant K (CELL)
Temperature by offset and scale.
Display:
approx. 10 mm high, 4-digit LCD-display
Ambient conditions for electronics:
Nominal temperature:
25°C
Operating condition:
-25 to 50°C, 0 to 95 %RH (non-condensing)
Storage temperature:
-25 to 70°C
Ambient of standard measuring cell -5 to 80°C (short time up to 100°C)
Housing:
ABS (IP65)
Dimensions:
82 x 80 x 55 mm (without elbow-type plug and sensor tube)
Mounting:
With holes for wall mounting (in housing - accessible after cover has been removed).
Mounting distance:
50 x 70mm, max. shaft diameter of mounting screws is 4 mm.
Electrical connection:
elbow-type plug conforming to DIN 43650 (IP65),
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm
EMC:
In accordance with EN61326 +A1 +A2 (appendix A, class B), additional errors: < 1% FS.
When connecting long leads adequate measures against voltage surges have to be taken.
The electrode connections have to be protected sufficiently against ESD pulses, if the
device is used in areas with risk of ESD.
GREISINGER electronic GmbH
T14.0.0X.6C-02
User’s Manual GLMU 200 MP
page 2 of 8
2 Safety Requirements
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures
and special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to
any other climatic conditions than those stated under "Specification". If the device is transported from a
cold to a warm environment condensation may cause in a failure of the function. In such a case make
sure the device temperature has adjusted to the ambient temperature before trying a new start-up.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic
safety regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully.
Internal connection in third party devices (e.g. connection GND and earth) may result in not-permissible
voltages impairing or destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be
marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning:
Do not use these product as safety or emergency stop devices, or in any other application where failure of
the product could result in personal injury or material damage. Failure to comply with these instructions
could result in death or serious injury and material damage.
3 Installation
3.1 Design types, dimension
T14.0.0X.6C-02
User’s Manual GLMU 200 MP
page 3 of 8
3.2 General installation instructions
To mount the connection cable (2-, 3-, or 4-wire depending on type of device) the elbow-type plug screw has
to be loosened and the coupling insert has to be removed by means of a screw driver at the position
indicated (arrow).
Pull out connection cable through PG glanding and connect to the loose coupling insert as described in the
wiring diagram. Replace loose coupling insert onto the pins at the transducer housing and turn cover cap
with PG glanding in the direction desired till it snaps on (4 different starting positions at 90° intervals). Re-
tighten the screw at the angle plug.
3.3 Assignment of elbow-type plug: (depending on type)
4-20mA (2-wire connection)
voltage (3-wire connection)
1
= supply voltage +Uv
1
= signal +
2
= GND / signal
3
= supply voltage +Uv
(4) = supply voltage –Uv
The type current or voltage output is set by works and cannot be changed
3.4 Terminal assignment of meas. cell socket
contact no.
Standard-Cell (K ~ 0.55)
1
Temperature +
2
Temperature GND
3
Supply 1
4
Signal 1
5
Signal 2
6
Supply 2
7
Not busy
3.5 Usage of Unit-Labels
As the transmitter is a multiple purpose device, many different display units are
possible, eg. µS/cm, kOhm·cm.
Therefore unit-labels (within scope of supply) can be shoved between the case cover
and the front foil behind the transparent unit-window.
To replace a label, unscrew the cover, pull out the old label (if present) and shove in the
new one.
The unit depends on the settings of the Cell-constant, the function and the measuring range!
Please refer to table in chapter “Configuration of the device”
3.6 Calibration Services
Calibration certificates – DKD-certificates – other certificates:
If device should be certificated for its accuracy, it is the best solution to return it with the refering sensors to
the manufacturer.
Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy!