Hanna Instruments HI 9056 Manuals

1 pdf manual(s) Hanna Instruments HI 9056 for free download (Document Types: Instruction manual)

Document Preview & Text Transcription

Instruction Manual HI 9056 HI 9057 Portable Microprocessor Thermocouple Thermometers Manufacturers since 1978 These Instruments are in Compliance with the CE Directives 3 2 PRELIMINARY EXAMINATION GENERAL DESCRIPTION HI 9056 performs temperature measure- ments using interchangeable K-type ther- mocouple probes. HI 9057 performs temperature measure- ments with different types of thermocouples (K- J - T). This meter is equipped with a key that switches range from K-type to J-type or T-type thermocouples. The non-linearity of the temperature probe is linearized by the built-in microprocessor, in order to achieve a high degree of accu- racy and resolution. Also the HI9056 accu- rately compensates for temperature drifts in the measurement circuit and the reference junction. HI 9056 has two built-in K-type probe con- nectors. By simply pressing a button, the meter switches between the two readings on the display. A "∆T" button will display the difference between the two probe readings. TABLE OF CONTENTS Preliminary Examination............................ 3 General Description ................................... 3 Functional Description & Specifications of HI 9056 .......................... 5 Functional Description & Specifications of HI 9057 .......................... 6 Operational Guide...................................... 7 Display Codes Guide ............................... 11 Calibration ................................................ 12 Battery Replacement ............................... 12 Shockproof Rubber Boots........................ 13 Accessories .............................................. 14 Warranty .................................................. 18 CE Declaration of Conformity ................. 19 Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the neces- sary information for a correct use of the instrument, as well as a precise idea of its versatility. If you need more technical information, do not hesitate to e-mail us at [email protected]. These instruments are in compliance with directives. Remove the instrument from the packing ma- terial and examine it carefully to make sure that no damage has occurred during ship- ping. If there is any damage, notify your Dealer or the nearest Hanna Customer Ser- vice Center. Each meter comes supplied complete with a 9V battery and instruction manual: Note: Save all packing material until you are sure that the instrument functions cor- rectly. All defective item must be returned in the original packing with the supplied accessories. 5 4 1. Liquid Crystal Display 2. ON/OFF Key 3. Reading Memory/Hold Key 4. Measurement Resolution Key (0.1 or 1) 5. Input Probe Selection Key (T1 or T2) 6. Temperatures Difference Key 7. Memory Recalling Key. Specifications Range -200.0 to 1370°C Resolution 0.1°C (-200.0 to 999.9°C) 1°C (1000 to 1370°C) Accuracy ±0.2% F.S. (@20°C / 68°F) for one year, excluding probe error Typical EMC ±3°C Deviation ±6°F Probe K-thermocouple (optional) (see pages 14-17) Battery 9V/300 hours of continuous use Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz.) FUNCTIONAL DESCRIPTION AND SPECIFICATIONS FOR HI 9056 Standard features include: • Interchangeable probes • Clear Liquid Crystal Display • Two year Warranty • Possibility of switching between the Cen- tigrade and the Fahrenheit scale (for HI 9057 only) • HOLD function that freezes the reading on the display (for HI 9057 only) • HI-LO memory recall function (for HI 9057 only) • Low battery detection. All these functions are easily accessible through the keypad. Self-explanatory symbols are used to re- mind the user of the operating mode or condition. Hanna manufactures several K-type ther- mocouple probes for different applications. See pages 14-17 for more details. 7 6 OPERATIONAL GUIDE INITIAL PREPARATION Remove the battery cover on the rear of your thermom- eter, unwrap the battery and connect it to the clip. Insert the battery into the compart- ment and replace the cover. To switch on, press ON/OFF on the front of the unit. The thermometer performs a self diagnostic test. The LCD dispalys all the segments at once, to ensure none are missing. After the test, the thermometer will switch to the measurement mode. If a temperature probe is plugged in, the unit will display the temperature. If the probe is not plugged in, the display will show "----". To switch your thermometer off, press ON/ OFF again. MEASURING SCALE (°C/°F) (HI 9057 only) When the meter is turned on, it automati- cally defaults to the °C scale. Measurements can be performed on either the Centigrade or the Fahrenheit scale. BAT HOLD K J T ºC ºF HI LO ON OFF 1. Liquid Crystal Display 2. ON/OFF Key 3. Reading Memory/Hold Key 4. Thermocouple Type Selection Key (K or J or T) 5. Minimum Temperature Recalling Key 6. Maximum Temperature Recalling Key 7. Measuring Scale key (°C or °F). Specifications Range K -200.0 to 1370°C / -300.0 to 2500°F J -200.0 to 900.0°C / -300.0 to 1650°F T -200.0 to 400.0°C / -300.0 to 750.0°F Resolution 0.1°C (up to 999.9°C) 0.1°F(up to 999.9°F) 1°C (over 1000°C) 1°F (over 1000°F) Accuracy K= ±0.2% F.S. (@20°C / 68°F) J = ±0.2% F.S. T = ±0.5% F.S. for one year, excluding probe error Typical EMC ±3°C Deviation ±6°F Probe K, J and T thermocouple (optional) (for K-thermocouple see pages 14-17) Battery 9V/300 hours of continuous use Environment 0 to 50°C (32 to 122°F); 95% RH Dimensions 180 x 83 x 40 mm (7.1 x 3.3 x 1.6") Weight 350 g (12.4 oz.) FUNCTIONAL DESCRIPTION AND SPECIFICATIONS OF HI 9057