Instruction Manual
HI 93711
Free & Total Chlorine
ISM
w w w . h a n n a i n s t . c o m
This Instrument is in
Compliance with the CE Directives
Hanna Instruments reserves the right to modify the design, construction
and appearance of its products without advance notice.
WARRANTY
HI 93711 is warranted for two years against defects in workman-
ship and materials when used for its intended purpose and maintained
according to instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact the dealer from whom you purchased the
instrument. If under warranty, report the model number, date of pur-
chase, serial number and the nature of the failure. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If
the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service de-
partment and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protec-
tion.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner, Hanna Instru-
ments Inc., Woonsocket, Rhode Island, 02895 , USA.
Dear Customer,
Thank you for choosing a Hanna product. This manual will
provide you with the necessary information for the correct
operation of the meter. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with
directives.
Remove the instrument from the packing material and
examine it carefully to make sure that no damage has
occurred during shipment. If there is any damage, notify
your Dealer.
Each Ion Specific Meter is supplied complete with
• 9V Battery
• Two Sample Cuvets and Caps
• One Transport Cap
Note: Conserve all packing material until the instrument
has been observed to function correctly. Any defective
item must be returned in its original packing.
PRELIMINARY EXAMINATION
GENERAL DESCRIPTION
The HI 93711 meter measures the free and total chlorine
(Cl2) content in water and wastewater.
The meter uses an exclusive positive-locking system to ensure
that the cuvet is in the same place every time it is placed
into the measurement cell.
The reagents are in liquid and powder form and are
supplied in bottles and in packets. The amount of reagent
is precisely dosed to ensure maximum repeatability.
Display codes aid the user in routine operations.
The meters have an auto-shut off feature that will turn the
instrument off after 10 minutes of non-use.
SPECIFICATIONS
SPECIFICATIONS
Range
Free Cl2 0.00 to 2.50 mg/L
Total Cl2 0.00 to 3.50 mg/L
Resolution
0.01 mg/L
Accuracy
±0.03 mg/L ±3% of reading
Typical EMC
±0.01 mg/L
Deviation
Light Source
Light Emitting Diode @ 555 nm
Method
Adaptation of the EPA recommended
DPD method 330.5. The reaction
between the chlorine and the DPD
reagent causes a pink tint in the sample.
Light Detector
Silicon Photocell
Environment
0 to 50°C (32 to 122°F);
max 95% RH non-condensing
Battery Type/Life 1 x 9 volt/40 hours
Auto-Shut off
After 10' of non-use
Dimensions
80 x 83 x 46 mm (7.1 x 3.3 x 1.8")
Weight
290 g (10 oz.).
REQUIRED REAGENTS
POWDER:
Code
Unit
Description
Quantity
HI 93701-0 Free Cl2
DPD
1 packet
HI 93711-0 Total Cl2
DPD
1 packet
LIQUID:
Code
Unit
Description
Quantity
HI 93701A-T Free and Total Cl2 DPD1 indicator 3 drops
HI 93701B-T Free and Total Cl2 DPD1 buffer
3 drops
HI 93701C Total Cl2 only
DPD3 solution
1 drop
This indicates that the meter is in
ready state and zeroing can be pe
formed.
Sampling in Progress. This prompt ap
pears each time the meter is performin
a measurement.
This indicates that the meter is in
zeroed state and measurement can b
performed.
A zero reading was not taken. Insert
sample before adding reagent and pres
ZERO.
Under range. A blinking "0.00" indicate
that the sample absorbs less light tha
the zero reference. Check the procedu
and make sure you use the same cuv
for reference (zero) and measurement.
Over range. A flashing value higher tha
the maximum concentration readable (se
specifications) indicates that the samp
absorbs too much light, meaning tha
the concentration is too high. Dilute th
sample.
Light over range. The cuvet is not in
serted correctly and an excess ambien
light is reaching the detector. If the cove
is properly installed, then contact you
dealer or the nearest Hanna Custome
Service Center.
Light under range. The zero sample
too dark for proper zeroing. If this is no
the case, contact your dealer or th
nearest Hanna Customer Service Center
The "V" indicates that the battery vol
age is getting low and the battery need
to be replaced.
This indicates that the battery is dea
and must be replaced.
Note: once this indication is displayed
the meter will lockup. Change the ba
tery to restart.
DISPLAY CODE GUIDE
V
OPERATIONAL GUIDE
MEASUREMENT PROCEDURE
• Turn the meter on by pressing ON/OFF.
• When the LCD displays "- - -",
it is ready.
• Select free or total Chlorine by pressing FREE/TOTAL. An
"I" or a "C" will appear on the right corner to indicate
free or total chlorine, respectively.
• Fill the cuvet with 10 mL of
unreacted sample, up to the
mark, and replace the cap.
• Place the cuvet into the holder
and ensure that the notch on
the cap is positioned securely
into the groove.
• Press ZERO and "SIP" will appear on the display.
• Wait for a few seconds and
the display will show "-0.0-".
Now the meter is zeroed and
ready for measurement.
• Remove the cuvet
Powder reagents procedure
• Add the specific test reagent to the cuvet:
Free Chlorine: 1 packet of
Total Chlorine: 1 packet of
DPD Free Chlorine reagent
DPD Total Chlorine reagent.
ON
OFF
FREE
TOTAL
ZERO
• Replace the cap and shake
gently for 20 seconds (or 2
minutes in case of seawater
analysis).
• Replace the cuvet into the
holder and ensure that the
notch on the cap is positioned
securely into the groove.
• Wait for 1 minute in case of Free Chlorine or 2 minutes
and 30’’ seconds in case of Total Chlorine and then
press READ. The display will show “SIP” during
measurements.
• The instrument directly displays concentration in mg/L of
Free or Total Chlorine on the Liquid Crystal Display.
The instruction listed below should be carefully followed
during testing to ensure best accuracy.
• Do not touch the cuvet walls with hands.
• In order to maintain the same conditions during the
zeroing and the measuring phases, it is necessary to
close the cuvet to prevent any contamination.
• Do not let the test sample stand too long after reagent
is added or accuracy will be lost.
• Whenever the cuvet is placed into the measurement cell,
it must be completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth
prior to insertion.
• It is important that the sample does not contain any
debris. This would corrupt the readings.
• It is possible to take multiple readings in a row, but it
is recommended that a zero reading be taken for each
sample and that the same cuvet is used for zeroing and
measurement.
• It is important to discard the sample immediately after
the reading is taken because the glass might become
permanently stained.
TIPS FOR AN ACCURATE
MEASUREMENT
Recommendations for Users
Before using these products, make sure that they are entirely suitable for the enviro
ment in which they are used. Operation of these instruments in residential area cou
cause unacceptable interferences to radio and TV equipments, requiring the operator
take all necessary steps to correct interferences. Any variation introduced by the user
the supplied equipment may degrade the instruments' EMC performance.
To avoid damages or burns, do not perform any measurement in microwave ovens.
ACCESSORIES
REAGENT SETS
HI 93701-01 Reagents for 100 free chlorine tests (powde
HI 93701-03 Reagents for 300 free chlorine tests (powde
HI 93711-01 Reagents for 100 total chlorine tests (powde
HI 93711-03 Reagents for 300 total chlorine tests (powde
HI 93701-T Reagents for 300 chlorine tests (liquid)
OTHER ACCESSORIES
HI 710009
Blue rubber boot
HI 710010
Orange rubber boot
HI 721310
9V battery (10 pcs)
HI 731318
Tissue for wiping cuvets (4 pcs)
HI 731321
Glass cuvets (4 pcs)
HI 731325
Caps for cuvets (4 pcs)
HI 93703-50 Cuvets cleaning solution (230 mL).
CE DECLARATION
OF CONFORMITY
IST93711R4
09/03
10 mL
Liquid reagents procedure
• To an empty cuvet add
3 drops of HI 93701A-T
DPD1 indicator, 3 drops
of HI 93701B-T DPD1 buffer
and only in case of Total
Chlorine measurement add
also 1 drop of HI 93701C
DPD solution. Swirl gently
to mix and immediately
add 10 mL of unreacted sample. Replace
the cap and shake gently again.
• Reinsert the cuvet into the
instrument.
• Wait for 1 minute in case of Free Chlorine or 2 minutes
and 30’’ seconds in case of Total Chlorine and then
press READ. The display will show “SIP” during
measurements.
or
x 3
x 3
x 1
• The instrument directly displays concentration in mg/L of
Free or Total Chlorine on the Liquid Crystal Display.
Note: free and total chlorine have to be measured sepa-
rately with fresh unreacted samples following the above
procedure if both values are requested.
INTERFERENCES
Interference may be caused by:
Bromine
Iodine
Fluorine
Ozone
Oxidized manganese and Chromium
Alkalinity above 250 mg/L or acidity above 150 mg/L will
not reliably develop the full amount of color or it may
rapidly fade. To resolve this, neutralize the sample with
diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L
CaCO3, shake the sample for approximately 1 minute after
adding the powder reagent.
• Shaking the cuvet can generate bubbles in the sampl
causing higher readings. To obtain accurate measure
ments, remove such bubbles by swirling or by gent
tapping the vial.
• All the reaction times reported in this manual are referre
to 20°C (68°F). As a general rule of thumb, they should b
doubled at 10°C (50°F) and halved at 30°C (86°F).
or