Document Preview & Text Transcription
C6000910-01
issued October 2021
U Series Operator Instruction Manual
U-Series Operating Manual
2
C6000610
Contents
Contents ................................................................................................................. 2
Safety Instructions ................................................................................................ 3
Installation – U100, U100H, U300 & U300H ......................................................... 6
Installation – U500H, U1300H & U2500H ............................................................. 7
Quick Guide ........................................................................................................... 8
Manual I Control Panel Instructions (U100, U300) ............................................ 9
Manual II Control Panel Instructions (U100H) ................................................. 10
Manual III Control Panel Instructions (U300H, U500H, U1300H & U2500H) . 11
Technical Information ......................................................................................... 12
Ultrasonic cleaning detergents ......................................................................... 13
Specifications ...................................................................................................... 14
Maintenance ........................................................................................................ 15
Troubleshooting .................................................................................................. 15
Returning equipment to Ultrawave ................................................................... 16
Warranty ............................................................................................................... 17
WEEE Compliance .............................................................................................. 17
Compliance with the Control of Noise at Work regulations ........................... 18
U-Series Operating Manual
3
C6000610
Safety Instructions
Use of products
The following products are covered by this operating manual:
•
U100, U100H, U300, U300H, U500H, U1300H & U2500H
The products listed above are a series of ultrasonic cleaning baths designed
for indoor use by a professional user.
Electrical
Connect to a 230 VAC fully earthed supply via a 3pin plug.
It can be dangerous to operate an ultrasonic bath without an earth
connected.
The U Series machines are rated at up to 850W maximum.
The detachable mains lead is rated at 250V 6A. NEVER FIT A LEAD WITH
A LOWER RATING.
The mains plug is fitted with a 5A fuse. NEVER FIT A FUSE OF A HIGHER
RATING.
The detachable mains lead is the disconnect device and should remain
accessible while the machine is in use.
Ensure that excess mains cable is stored neatly.
Temperature: Care should be taken when operating the bath at
higher temperatures as external surfaces may become hot.
When the temperature is set to 54°C or above, the unit will become hot and
care should be taken when touching the case.
A risk assessment should be conducted – see BS EN ISO 13732-1:2008 for
full details
U-Series Operating Manual
4
C6000610
General use
Do not place fingers, hands, or other parts of the body in the liquid in the bath
when it is sonicating.
Always ensure the liquid is above the tank crease when in operation (page 7).
Do not place parts or containers directly on the bottom of the cleaning tank.
Always use a basket or instrument cassette to support items to be cleaned. Not
doing so may damage the bath and invalidate your warranty.
Never use toxic, flammable, acidic, caustic, or corrosive solutions in the bath.
Do not move the bath when it is full of water.
Before cleaning, drain the bath. Clean by wiping with a clean non-abrasive cloth.
The user should familiarize themselves with this user manual before operating
the equipment. Safety may be impaired if they are not followed.
Ultrawave will not be responsible for damage or injury caused by incorrect use
of the equipment.
The user should apply to Ultrawave or its agent for advice on cleaning techniques
or detergents.
IMPORTANT: Once the cleaning cycle has finished, remove the cleaned items,
and rinse them immediately in clean running water.
REMEMBER
Always ensure the liquid is at or above the crease in the tank when in operation.
Do not put hot water above 50°C into the bath.
Always use the basket to support items in the bath.
Never expose hands, fingers or other body parts to cleaning solutions.
Never use toxic, flammable, acidic, caustic or corrosive solutions.
Never breathe the fumes from strong solutions.
Rinse the items in clean water once the cycle is complete.
Subjecting the bath to improper treatment or misuse will invalidate the warranty.