MS2300 Baby Scale
OPERATOR’S MANUAL
Calibration Mode for Technical Maintenance
4
TABLE OF CONTENTS
Introduction …………………………………………………………2
Power supply ……………………………………………………… 2
Low battery ………………………………………………………… 2
Specifications ……………………………………………………… 2
Display and key function ……………………………………….. 3
Calibration ………………………………………………………….. 4
Trouble error measures ………………………………………..5 & 6
1
START
Press Test pin twice which at rear of the scale to
enter Test mode.
Press ON/OFF/ZERO key to zero the scale.
Display shows span count
(4000~~14000).
Display shows CAL
Press ON/OFF/ZERO key once.
Display shows calibration
weight in normal mode.
End
Load calibration weight
All scales come complete with plastic platform, plastic base, and two
soft steel handles. The keyboard has three easy to use function
keys: ON/ZERO/OFF, TARE, RECALL. With the standard DC
adaptor included and the 9V battery, the scale can be used in a
fixed location or as a portable scale.
zMODEL NUMBER: MS2300
zPOWER
The baby scale uses 9V battery or 9V 100mA DC adapter.
zLOW BATTERY
1.
(Low battery indication) will display on LCD display when
battery’s power is insufficient to power the scale.
2. When
is displayed, the operator has to replace battery or plug
in DC adaptor to instead of 9V battery.
zSPECIFICATIONS
2
ϥ
ϥ
ϥ
ϥ DISPLAY AND KEY FUNCTIONS
Key function:
1. ON /ZERO/OFF:
1) Power on the scale.
2) Power off the scale by pressing 3 seconds.
3) Zero the scale (±2% of full capacity).
2. TARE :
1) Before weighing baby, the weight of any protective paper or
napkin placed on the platform can be deducted by pressing
the start switch and obtaining zero. Baby can then be
weighed.
2) Press this key for 3 seconds to enter setting function mode.
3) In setting mode, this key means “select”.
3. RECALL:
1) Press this key to call the memory value and the arrow of
“MR” will appear. (Meanwhile, the memory value will be
cleared by software).
2) In setting mode, this key means “confirm”.
3) Disable the feature, please press ON/ZERO/OFF key.
3
Model# MS2300
Capacity
20 kg x 20g
Weight units
Kg
Power Supply
9V battery / 9V 100mA DC adaptor
Display
Easy to read 1.0” LCD display with 4 1/2 active digits
Operation Temperature
5к ~ 35к
Function Key
ON/ZERO/OFF,TARE, RECALL
Dimensions
Scale with handle: 24 x 14 x 2 in. (620 x 350 x 50 mm)
Platform surface: 20 x 14 x 2 in. (500 x 350 x 50 mm)
ϥ ERROR MESSAGE
ʳ
ˈˁ Power on zero count under calibration
zero –10%
6. Power on zero count over calibration
zero +10%
5
1. LOW BATTERY
If battery is discharged, the display
will show LOW. The operator should
replace the battery.
2. OVER ZERO
If zero count is higher than the setting
range, the display will show “Err”
message. The scale will turn off in 30
seconds if the status continues.
cocontinues.
4. OVERLOAD
3. UNDER ZERO
If zero count is below the setting range,
the display will show “Errl” message. The
scale will turn off in 30 seconds if
the status continues.
If the total load exceeds the maximum
capacity of the scale, the display will show
“Err” message. The weight must be removed
immediately from the surface.