A&D AD-4430B Instruction Manual
Read Online
Read the A&D AD-4430B Instruction Manual online using the built-in PDF viewer.
Text of the manual
This manual and Marks
All safety messages are identified by the following, “WARNING” or
“CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product
Safety Signs and Labels). The meanings are as follows:
WARNING
A potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
A potentially hazardous situation which, if not avoided,
may result in minor or moderate injury.
This is a hazard alert mark.
This manual is subject to change without notice, at any time, to improve the
product.
The contents of the product specifications and this manual are subject to
change without any obligation on the part of the manufacturer.
Under the copyright laws, the software (program) described in this manual
is copyrighted, with all rights reserved.
The software may be installed into one computer and may not be installed
into other computers without the prior written consent of A&D Company,
Limited. Copying includes translation into another language, reproduction,
conversion, photocopy and offer or loan to another person.
2013
All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or
translated into any language in any form by any means without the written
permission of A&D Company, Limited.
- 1 -
Contents
1.
Compliance.....................................................................................................................................4
1.1.
Compliance with FCC rules ....................................................................................................4
1.2.
Compliance with European Directives...................................................................................4
1.3.
Precautions For Safety Use....................................................................................................4
2.
Outline and Features......................................................................................................................5
3.
Specifications ..................................................................................................................................6
3.1.
Analog to Digital Unit (Load cell Input, A/D Converter)..........................................................6
3.2.
Display And Keys.....................................................................................................................6
3.3.
General.....................................................................................................................................6
3.3.1.
Interface..............................................................................................................................6
3.3.2.
Weighing Function.............................................................................................................7
3.3.3.
General...............................................................................................................................7
3.3.4.
Accessories........................................................................................................................7
3.3.5.
Dimensions ........................................................................................................................8
3.4.
Names (The Front Panel And Rear Panel )...........................................................................9
4.
Installing The Module....................................................................................................................10
4.1.
Conditions To Install The Module..........................................................................................10
4.2.
Power Supply.........................................................................................................................10
4.3.
Loadcell Cable .......................................................................................................................10
4.4.
Shield Cable........................................................................................................................... 11
4.5.
Connecting Loadcell Cable................................................................................................... 11
4.6.
Verifying Loadcell Cable........................................................................................................12
5.
Operations.....................................................................................................................................13
5.1.
General Functions .................................................................................................................13
5.1.1.
Zero Adjustment...............................................................................................................13
5.1.2.
Zero Tracking...................................................................................................................13
5.1.3.
The Tare Function............................................................................................................13
5.1.4.
Customizing The Function Of the Switch F ...................................................................14
5.1.5.
Customizing The Function Of the Display x...................................................................15
5.1.6.
Memory Backup...............................................................................................................15
5.1.7.
The Detection For The Zero Band..................................................................................15
5.1.8.
Upper Or Lower Limit Detection Function......................................................................15
5.1.9.
The Hold Function...........................................................................................................16
5.2.
Mode Map And Operation Switches.....................................................................................17
5.2.1.
Mode Map........................................................................................................................17
5.2.2.
Operation Switches .........................................................................................................18
5.3.
Calibration ..............................................................................................................................19
5.3.1.
Outline ..............................................................................................................................19
5.3.2.
The Calibration With An Actual Load (Using A Mass)...................................................20
5.3.3.
Gravity Acceleration Correction......................................................................................21
5.3.4.
Outline Of The Linearization Function............................................................................22
5.3.5.
The Actual Load Linearization Function.........................................................................23
- 2 -
AD-4430B
5.3.6.
The Function
Related To The Calibration.......................................................24
5.3.7.
The Function
Related To The Linearization Function....................................28
5.3.8.
Error Codes For The Calibration.....................................................................................29
5.3.9.
Adjustment Of The Loadcell Output...............................................................................29
5.4.
General Functions .................................................................................................................30
5.4.1.
The Procedure To Store New Parameters.....................................................................30
5.4.2.
The Adjustment Of The Digital Filter...............................................................................31
5.4.3.
The List Of The Basic Functions.....................................................................................32
5.4.4.
The List Of The Hold Function........................................................................................33
5.4.5.
The List Of The BCD Function........................................................................................34
6.
Interface.........................................................................................................................................35
6.1.
The Parallel BCD Output.......................................................................................................35
6.1.1.
Timing Chart Of The BCD Output ..................................................................................35
6.1.2.
Terminals Of The BCD Output........................................................................................35
6.1.3.
The State Of The BCD Output........................................................................................35
7.
Maintenance .................................................................................................................................36
7.1.
Error Messages......................................................................................................................36
7.2.
Verifying The Load Cell Connections ( DIAGNOS )............................................................36
7.2.1.
Guideline To Verify The Load Cell Connections............................................................36
7.2.2.
The Operation With The BCD Input...............................................................................37
7.2.3.
The Operation With The Switch......................................................................................37
7.2.4.
The Result Of Verifying The Connection........................................................................37
7.3.
Check Mode...........................................................................................................................38
7.3.1.
Entering The Check Mode..............................................................................................38
7.3.2.
Verifying The Switch Operation ......................................................................................38
7.3.3.
Monitoring The A/D Converter ( For Load cell Output)..................................................38
7.3.4.
Monitoring The Internal Value.........................................................................................39
7.3.5.
Monitoring The Program Version....................................................................................39
7.3.6.
Monitoring The Serial Number........................................................................................39
7.3.7.
Monitoring The Checksum Of The Program .................................................................39
7.3.8.
Monitoring The Checksum Of An Internal FRAM..........................................................39
7.3.9.
Displaying Function Parameters For Calibration...........................................................39
7.4.
Initializing Parameters ...........................................................................................................40
7.4.1.
Initializing Mode For RAM And Function Parameters...................................................40
7.4.2.
Initializing The Whole Data..............................................................................................40
7.5.
Verifying The Load cell Connections Using Multimeter.......................................................41
7.5.1.
Check List Of The Load cell Connections......................................................................41
7.6.
The Parameter List For The Function List............................................................................43
7.6.1.
Basic Functions................................................................................................................43
7.6.2.
The List Of The Hold Function........................................................................................44
7.6.3.
The List Of The BCD Function........................................................................................45
7.6.4.
The List Of The Calibration .............................................................................................46
7.6.5.
The List Of The Linearization Function ..........................................................................48
8.
Dimensions ...................................................................................................................................49
- 3 -
Illustrations
Illustration 1 Dimensions..............................................................................................................................8
Illustration 2 Front panel & rear panel.........................................................................................................9
Illustration 3 Mounting the module..............................................................................................................9
Illustration 4 Cable type..............................................................................................................................10
Illustration 5 Load cell connection.............................................................................................................11
Illustration 6 Peak hold / Averaging hold..................................................................................................16
Illustration 7 Mode Map (State Transition Diagram)...............................................................................17
Illustration 8 Gravity Acceleration Graph..................................................................................................22
Illustration 9 Digital linearization.................................................................................................................22
Illustration 10 Load cell output adjustment..............................................................................................29
Illustration 11 Wire Name of Load Cell....................................................................................................36
Illustration 12 Connection Check of Load Cell.......................................................................................41
- 4 -
AD-4430B
1.
Compliance
1.1.
Compliance with FCC rules
Please note that this equipment generates, uses and can radiate radio frequency energy.
This equipment has been tested and has been found to comply with the limits of a Class A
computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to
provide reasonable protection against interference when this equipment is operated in a
commercial environment. If this unit is operated in a residential area it may cause some
interference and under these circumstances the user would be required to take, at his own
expense, whatever measures are necessary to eliminate the interference.
(FCC = Federal Communications Commission in the U.S.A.)
1.2.
Compliance with European Directives
This appliance complies with the statutory EMC (Electromagnetic Compatibility) directive
2004/108/EC and the Low Voltage Directive 2006/95/EC for safety of electrical equipment
designed for certain voltages.
Note: The displayed value may be adversely affected under extreme electromagnetic influences.
1.3.
Precautions For Safety Use
Before use, confirm the following articles for safe operation.
Grounding The Module
Ground the module to the DIN rail with certainty. Separate this earth ground line from others,
like ground line of a motor, inverter or a power source. Unless the indicator is grounded, it
may result in receiving an electric shock, cause operation error or catch fire.
Proper Power Source And Power Cable
Confirm the AC voltage, frequency and power tolerance of the power cable. If the voltage
range of the cable is lower than the power line voltage, it may cause leakage or catching fire.
Use pole compression terminals to connect the power cable to the terminals.
Fuse
The fuse is installed to help prevent the module from catching fire. The module is equipped
with many safety circuits. Therefore, the fuse is not damaged in normal operation. If the fuse
is damaged, do not replace it, contact your local A&D dealer. This trouble may have been
caused by strong electric discharge.
Splashing Water
The module is not water-resistant type.
Flammable Gas
Do not install the module where flammable gas is present.
Heat Radiation Of The Module
Space out instruments to radiate heat sufficiently. Use a cooling fan to keep the operating
temperature of the module within specifications.
- 5 -
2.
Outline and Features
The AD-4430B has the following features.
The weighing module is a weighing indicator that amplifies signals from a load cell, converts it
to digital data and displays it as a mass value.
This indicator has the following performance :
Input sensitivity : ............................0.15 μV / d (d = minimum division)
Display resolution : ........................99,999 d in maximum
Sampling rate : ...............................1000 times/second
Input voltage range : ......................-35 mV to +35 mV (-7 mV/V to +7 mV/V)
The calibration using gravity acceleration correction :
The function compensates for weighing error due to the difference of gravity acceleration
between the calibration place and the measurement place.
The digital linearization function :
The digital linearization function can rectify and reduce the linearity deviation using weighing
points during the zero and capacity setting. Up to four weighing points can be specified. The
high-order correction curve is used between each points.
Digital span mode :
Keying in the load cell output voltage (mV/V) allows calibration to be performed without an
actual load.
- 6 -
AD-4430B
3.
Specifications
3.1.
Analog to Digital Unit (Load cell Input, A/D Converter)
Input sensitivity
0.15 μV/d or greater (d = minimum division)
Input voltage range
-35 mV to +35 mV (-7 to +7 mV/V)
Zero adjustment range
-35 mV to +35 mV (-7 to +7 mV/V)
Load cell excitation voltage
(Load cell drive capacity)
5 VDC ±5%, 60 mA with remote sensing capability
(Maximum 4 x 350 Ω load cells)
Zero
±0.02 μV/°C Typ. ±0.1 μV/°C max
Temperature coefficient
Span
±3 ppm/°C Typ. ±15 ppm/°C max
Non-Linearity
0.005% of full scale
A/D conversion method
Delta-sigma method
A/D resolution count
Approximately 16,000,000 counts
Display resolution
99,999 d max. (d = minimum division)
Sampling rate
1000 times/second
3.2.
Display And Keys
Display element Measurement display
5 - digit 7-segment green LED
Status indicators
6 red LEDs
Measurement
display
Numerical display
Switches between NET and GROSS
Decimal point
Selectable decimal places (101, 102, 103, 104)
Overflow display
All the digits turn OFF. (When the polarity is negative,
the minus sign appears at the highest-order digit.)
Status indicators
: GROSS,
: NET,
: HOLD,
: STABLE,
: ZERO,
: Preset function selected from the function list.
Key switches
/
,
(ZERO),
(TARE),
3.3.
General
3.3.1.
Interface
BCD Output
MDR connector 36pins female type
BCD Output And I/O Output
I/O Input
Output circuit
Open collector transistor
Input circuit
DC contact input
Isolation
Photo coupler
Open voltage
Approx. 5 V
Output current
50 mA max.
OFF current
0.1 mA max.
Saturation voltage 0.5V max. @50 mA
ON current
2 mA min.
Threshold voltage
2V
Chattering suppression time 10 msec.
- 7 -
3.3.2.
Weighing Function
Zero adjustment
Sets the gross weight to zero by pressing the
(ZERO) key.
Selection of disable or enable for the operation when unstable.
The zero value is backup by FRAM.
Zero adjustable range : Can be set optionally in the range of 1 to 100% of the
weighing capacity.
Zero tracking
Tracks the weight drift around the zero point to maintain zero.
Zero tracking time : 0.0 to 5.0 seconds
Can be set optionally within the range
Zero tracking band : 0.0 to 9.9 d
Can be set optionally within the range
Tare
Sets the net weight to zero by pressing the
(TARE) key.
The tare weight is backup by FRAM.
Tare range : Gross weight ≤ Weighing capacity
Stability detection
Turns ON the stabilization indicator when the variables of the weight values per
sampling are within the set band in the set time.
Stability detection time : 0.0 to seconds
Can be set optionally within the range
Stability detection band : 0 to 9d
Can be set optionally within the range
Digital filter
Cutoff frequency (-3 dB) range : 0.7 to 100 Hz
Zero detection
Detects to place nothing on the weighing pan.
Comparator
Compares the measurement with these limits and outputs the results.
Hold function
Displays the measurement value held.
Select from sample hold, peak hold, average hold.
3.3.3.
General
Data backup
Power failure countermeasure
Zero value, tare weight, calibration, function parameters backup by
FRAM. (Approx. 10 years)
Power source
DC 24 V, +10%, -15%
Power consumption
Approximately 6 W
Operating temperature
Operating humidity
-10 °C to +50 °C,
85 %RH or less (no condensation)
Installation method
DIN rail mount
Mass
Approximately 180 g
3.3.4.
Accessories
Item
Quantity
Model name
Power source connector
1
FMC1.5/2-ST-3.5
- 8 -
AD-4430B
110
35.3
(2.5)
100
1.3
34.5
39.4
28
20°
36.1
5.5
3.3.5.
Dimensions
Unit : mm
Illustration 1 Dimensions
- 9 -
3.4.
Names (The Front Panel And Rear Panel )
Illustration 2 Front panel & rear panel
Illustration 3 Mounting the module
Main display
Status indicators
Key switches
Vent
Sensor terminals
of load cell
Guide for rail
BCD connector
Power supply
connector
Hook for rail
Pull down it when removing.
Vent
DIN rail
Guide for rail
- 10 -
AD-4430B
Ex.: Shielded load cell cable
Ex. Power cable
Accessory connector
Press the button to remove & insert it.
Press button to remove & insert it.
4.
Installing The Module
This section is explained for the environment to install the module and for the terminals of the
power source and load cell cable. Refer to each section of options for I/O interface.
4.1.
Conditions To Install The Module
The module is a precision electronic instrument. Handle it carefully.
The operating temperature is -10°C to +50°C.
Do not install the module in direct sunlight.
4.2.
Power Supply
Earth ground the module to prevent electrical shock or indicator malfunction.
Before connecting the module to the power source, read the instruction manual thoroughly.
Do not connect the module to the power source before the installation is complete.
To avoid electrical shock, do not handle the power cable with a wet hand.
Earth ground the module. Do not share the ground line with other electrical power equipment.
The power requirement is 24 DCV, +10% to -15%.
Use a stable power source free from instantaneous power failure or noise.
To avoid a malfunction, do not share the power line with other devices.
The output voltage of a load cell is a very sensitive signal. Keep all electrical noise sources
away from the load cell and load cell cable.
Use shielded I/O cables. Connect the cable shield to the F.G. terminal or the module housing.
4.3.
Loadcell Cable
Load cell cables should have high insulation and shielding performance.
Use shielded cables with the insulator that is made of materials with high insulation resistance
such as Teflon and polyethylene. NOTE: Teflon is a registered trademark of DuPont. When
using the extension cable, we recommend using A&D's load cell cables.
AX-KO162-5M to 100M (5m to 100m)
Cross-sectional area of the conducting wire.....0.5mm2 6 wire cable equipped
Cable diameter......................................................φ9mm
Ex. Rod crimping terminal
Illustration 4 Cable type
- 11 -
4.4.
Shield Cable
Connect the load cell shield wire only to the shield terminal of the AD-4430B.
As a general rule, No ground is connected between the load cell and the AD-4430B. This is to
prevent the ground loop generated by multiple ground points. A ground loop can be created
along the wiring route, and interference such as high-frequency noise becomes more likely.
4.5.
Connecting Loadcell Cable
Shield Cable Connection
As a general rule, only connect the load cell cable's shield wire to the module's shield
terminal. This is because when there are multiple ground points, it becomes easier for noise
to occur as a result of a ground loop. However, when there are higher priorities than noise
prevention, such as explosion prevention systems or static electricity release routes, this rule
does not necessarily need to be followed.
Load Cell Terminal Connections
Two types of load cell connection are available: 6-wire connection and 4-wire connection.
For high precision and stable weighing, 6-wire connection is recommended.
Terminal No.
Terminal Name & Function
1
SIG-
Load cell input (-)
2
SIG+
Load cell input (+)
3
EXC- Load cell excitation voltage (-)
4
SEN- Sensing input (-)
5
SEN+ Sensing input (+)
6
EXC+ Load cell excitation voltage (+)
7
SHLD
Shield
6-wire connection to load cell (recommended)
4-wire connection to load cell (not recommended)
Illustration 5 Load cell connection
- 12 -
AD-4430B
Type
Advantages
Disadvantages
Description
6-wire
connection
(recommended)
The error is small even
when the load cell cable
is extended, a thin load
cell cable is used, or
multiple load cells are
used.
Complicated wiring
Use a 6-wire shielded
cable when a summing
box is used.
4-wire
connection
Simple wiring
The influence of the load
cell cable resistance
worsens the
temperature coefficient.
Prone to the influence of
the contact resistance of
the connector
The error increases
when the load cell cable
is extended or multiple
load cells are used.
Precautions to be taken when using the 4-wire connection:
If the 4-wire connection has to be used, please take the following measures.
Be sure to connect terminals EXC+(6) and SEN+( 5 ) and terminals SEN-(4) and EXC-(3).
When lengthening the load cell cable, try to use cable with a large cross-sectional area.
Also, keep the cable as short as possible.
4.6.
Verifying Loadcell Cable
When the load cell connection is complete, perform a connection check using the following
procedure.
Perform a visual check to ensure that the wiring is correct.
Turn the module on.
When calibration has not yet been carried out, the display value may be blank. However, as
long as there are no problems with the display, confirmation can still be carried out using
check mode.
Enter the check mode and check the load cell output value.
Refer to “7.3.Check Mode" to enter the A/D check mode.
Confirm that the displayed load cell output value matches the specified value. Normally the
displayed value will be the load cell rated output value or less.
If an error occurs, refer to "7.2. Verifying The Loadcell Connections ( DIAGNOS )" or "7.5.
Verifying The Loadcell Connections Using Multimeter".
- 13 -
5.
Operations
5.1.
General Functions
5.1.1.
Zero Adjustment
Zero adjustment is a function to set the gross weight to zero.
It is performed by pressing the
(ZERO) key.
The zero adjustment range is set in CAL05 ( Zero adjustment range ) and is expressed in
percent of the weighing capacity with the calibration zero point as the center.
Zero adjustment is disabled, even within the zero adjustment range, when the A/D converter
overflow occurs.
A ZERO error is output if zero adjustment is not performed.
The zero value is stored in the non-volatile memory and is maintained, even if the power is
disconnected.
Clearing the zero value is performed using the
key assigned to clear the zero value.
Functions Related To Zero Adjustment
CAL05( Zero adjustment range ) : A value between 0% and 100% can be specified.
CAL10( Tare and zero adjustment when unstable ) : The selection to enable or disable
tare and zero adjustment when unstable.
CAL16( Zero upon power-ON ) : The selection whether or not to perform zero upon
power-ON.
5.1.2.
Zero Tracking
The zero tracking is a function to track the gross weight drift around the zero point to maintain zero.
The zero tracking time is set in CAL06( Zero tracking time ) and the zero tracking band is set
in CAL07( Zero tracking band ). When the gross weight drift is within the specified ranges,
zero tracking is performed automatically.
A ZERO error is not output even if zero tracking is not performed.
Functions Related To Zero Tracking
CAL06( Zero tracking time ) : The value between 0.0 and 5.0 seconds can be specified.
CAL07( Zero tracking band ) : The value between 0.0 and 9.9 d can be specified.
(d = minimum division)
5.1.3.
The Tare Function
Tare is a function to store the gross weight as the tare value and set the net weight to zero.
The tare weight is stored in the non-volatile memory and is maintained, even if the power is
disconnected.
Clearing the tare weight is performed using the
key assigned to clear the tare weight.
Functions Related To The Tare Function
CAL10( Tare and zero adjustment when unstable ) : The selection to enable or disable
tare and zero adjustment when unstable.
CAL11( Tare when the gross weight is negative ) : The selection to enable or disable tare
when the gross weight is negative.
- 14 -
AD-4430B
Clearing The Tare Weight And Zero Adjustment
The way to clear the tare weight and zero adjustment : While pressing and holding the
(TARE) key, turn on the module. Another way : In turning off the module, while pressing
and holding the
(TARE) key, press the
key.
5.1.4.
Customizing The Function Of the Switch F
Assign a function to the
key in the general functions.
Functions Related To The
Key
Assigns a function to the
key from the functions of Fnc02(
key ) below :
0: None
1: Manual print command
2: Hold
3: Alternate switch
4: Momentary switch
5: Display selection
6: Clear the tare weight
7: Clear the zero value
CAL15( Clear the zero value ) : The selection to enable or disable clearing the zero value.
CAL20( Output when the gross weight is negative ) : The selection to enable or disable the
print command when the gross weight is negative.
Alternate switch and momentary switch
By assigning these switches to the
key, the ON/OFF status of the
key can be
transmitted to the master station. This is useful when building a network or performing
maintenance. The function of the display x can be specified by the
key and can be
monitored. These switches work as follows :
Alternate switch
When pressing and releasing the switch once, the state of the switch is maintained.
Press the switch again to turn off or on.
Momentary switch
Only while the switch is being pressed, the switch is ON. When it is released, it is OFF.
- 15 -
5.1.5.
Customizing The Function Of the Display x
Assign a function to the display x in the general functions.
Functions Related To The Display x
Assigns a function to the o display from the functions of Fnc04(Display x) below :
0: None
1: Zero tracking in progress
2: Alarm (Zero range error, over )
3:
key status
4: Zero band
5: HI output (Over the upper limit value)
6: OK output (Within upper and lower limit values)
7: LO output (Below the lower limit value)
8: User input 1
5.1.6.
Memory Backup
Zero value, tare weight, calibration data and function data are written into non-volatile
memory( FRAM). The FRAM does not require batteries. The data retention period is 10
years. Therefore, this module does not equipped the backup battery.
5.1.7.
The Detection For The Zero Band
Zero band is a function to detect whether an object has been placed on the weighing pan.
Zero band is defined as a state of the zero band when the weighing value is within the preset
value for the zero band.
Functions Related To The Zero Band
Fnc08(Set value of zero band) : The value of zero band can be specified.
Fnc09(Mass to be compared with zero band) : The selection of the gross weight or net
weight to compare the value of zero band.
5.1.8.
Upper Or Lower Limit Detection Function
This is a function to detect whether the weighed value is above an upper limit value or below a
lower limit value.
Functions Related To The Detection Function
Fnc10(Upper limit value) or Fnc11(Lower limit value) : A comparative upper or lower limit
value can be set.
Fnc12(Weight compared with upper and lower limits) : Gross weight or net weight to be
compared with the upper or lower limit value can be selected.
Fnc13(Logic output for upper and lower limits) : Positive logic or negative logic to output
the upper and lower limit can be selected.
key status
When alternate switch or momentary switch is selected at Fnc02, the
works.
The display turns ON when the
key is ON and turns OFF when the
key is OFF.
- 16 -
AD-4430B
5.1.9.
The Hold Function
There are three types of hold functions which can be used for different purposes.
Normal hold
Holds the value displayed at the time the hold command was received.
Peak hold
Holds the maximum value reached after the hold command was received.
The value will be refreshed if it increases again.
Averaging hold
This function averages weighing data over a certain period of time and then holds the result.
It is useful for measuring things that are difficult to weigh such as an animal that won’t settle
down, or for averaging out the weight of an object in an unstable state. In addition, it can
reduce the effects of breezes which the digital filter cannot eliminate.
Functions Related To The Hold Function
Fnc07(Hold) : The type of hold function can be selected.
To set operating conditions for the hold function : Set the averaging time length, standby
time, or start and stop conditions with HLd01 to 07. (Will not affect normal holds)
Peak hold
Averaging hold
Illustration 6 Peak hold / Averaging hold
Waiting time (Busy state)
(Start standby timer)
Averaging (Busy state)
(Averaging time)
Hold
Hold start command
Automatic start condition
Averaging start
Averaging finish
Hold cancellation
command
Automatic cancellation condition
Standard
state
Standard
state
Waiting time (Busy state)
(Start standby timer)
Hold
Hold start command
Automatic start condition
Hold cancellation command
Automatic cancellation condition
Standard
state
Standard
state
Refresh
(Busy state)
Hold
Judgment start
- 17 -
5.2.
Mode Map And Operation Switches
5.2.1.
Mode Map
The module has several modes to perform various operations. Perform mode switching by the key
operation shown below, only in the direction of the solid arrow. After setting a mode, the module
resets automatically. Alternatively, the module resets when the power is disconnected.
Illustration 7 Mode Map (State Transition Diagram)
Power on
Weighing mode
OFF mode (Standby)
key
In case of connecting the power without operation :
When the power is disconnected in the OFF mode and
connected again, the module will start in the OFF mode.
General function mode
+
key
Calibration mode
Calibration using actual load
Function mode related to
calibration
All data initialization mode
Automatic rest after setting
OFF state
ON state
Initialization mode
Check mode
Maintenance mode
+
key, next
(ZERO) +
key
Function mode related to
digital linearization mode
Digital linearization mode
+
key
Self-check mode
: Mode is switched with the
key.
: Mode is switched with the
key.
: Mode is switched automatically.
- 18 -
AD-4430B
5.2.2.
Operation Switches
Key
State
Function and Use
Weighing mode
The display switch between gross and net in factory setting.
Weighing mode
The zero key to perform the zero adjustment.
Setting mode
The key to change a selected item or move a flashed figure.
Weighing mode
The tare key.
Setting mode
The key to select parameter or increase number.
Weighing mode
When turning off the module, press and hold the key.
OFF mode (Standby) When turning on the module, press the key.
Setting mode
The key to store new settings.
Weighing mode
Function key can be selected the function and use.
Setting mode
The return key or escape key.
+
Weighing mode
The keys to proceed to the function mode.
+
Function mode
The keys to proceed to the check mode.
+
OFF mode (Standby) The keys to proceed to the calibration mode.
- 19 -
5.3.
Calibration
5.3.1.
Outline
In the calibration mode, the operation to relate the output voltage from a load cell to the mass
value and other operations directly related to weighing are performed.
Calibration with an
actual load
Calibration is performed using a calibration weight.
Zero calibration : ..................Press
key when no load is applied.
Span calibration : .................Enter the calibration weight value and place
the calibration weight.
When the module enters the mode of calibration with an actual load, the tare
weight and the zero value will be automatically cleared.
Digital span
Calibration is performed without an actual load, by numerical input of the load
cell output voltage (mV/V). Set the functions related to the calibration.
Input voltage at zero : ..........Numerical input of the load cell output at zero.
Input voltage of span : .........Numerical input of the load cell output of span.
(Load cell output at full capacity - load cell output at zero)
Calibration weight value of span : ..........Numerical input of the calibration
weight value corresponding to the input voltage of span.
( These value relate the input voltage of span and the weight value.)
Dual range
The dual range is the display feature that the weighing range is divided into two
depending on the value of gross or net and the minimum division is changed.
Gravity acceleration
correction
The correction calculates and corrects for gravity acceleration if the location
where the module was calibrated and the location where it is being used in are
different.
Digital linearization
Nonlinearity correction feature to correct weighing errors that occur halfway
between the zero point and weighing capacity. Up to 4 points can be input in
addition to the zero point, and the intervals between each point will be
calculated using curves.
Function related to
calibration
The function stores basic parameters of the module such as the minimum
division and weighing capacity and other data directly related to weighing is
performed. It includes the parameters for digital span calibration, dual range or
gravity acceleration correction is also performed.
All data initialization All the data such as zero value, tare weight, calibration data and function data
are initialized.
All the parameters in the calibration mode is stored in the FRAM.
- 20 -
AD-4430B
5.3.2.
The Calibration With An Actual Load (Using A Mass)
Calibration is performed using a calibration weight. Before performing calibration for the first
time, set the unit, decimal point position, minimum division and weighing capacity in the
calibration-related function mode ( C-Fnc ).
Note To avoid drift caused by changes in temperature, warm up the indicator for ten
minutes or more before performing calibration with an actual load.
Step 1
In the OFF mode (Standby), Press the
+
key to enter to
the calibration mode and display
.
Step 2
Press the
key to start the calibration and display
.
If returning to the weighing mode, press the
key.
Zero Calibration
Step 3
Press the
key to display
.
If zero calibration is not to be performed, press the
key and
proceed to step 5. If you want to check the current weighing value,
press the
key. When pressing the
key again,
is display.
Step 4
Placed nothing on the pan and wait for the stabilization ( LED ).
Press the
key.
is displayed for approximately
two seconds. If span calibration is not performed, press twice the
key for returning to the weighing mode.
Span Calibration
Step 5
Press the
key when
is displayed. The
calibration weight value (the current weighing capacity) is
displayed and the least digit of the value blinks. Correct the value
using the
and
key so as to be the value of the
calibration weight used. If span calibration is not performed, press
three times the
key for returning to the weighing mode.
Step 6
Place the calibration weight on the pan. Wait for the stabilization
( LED ). Press the
key.
is displayed for
approximately two seconds.
Step 7
is displayed.
Step 8
Press the
key.
is displayed and the
calibration data is stored in the memory.
Step 9
The current state is the same as that of step 2. If returning to the weighing mode, press
the
key.
If
X is displayed, an error has occurred. Refer to “5.3.8. Error Codes For The
Calibration” to take some measures. X : number.
The blinking decimal point means that the current value is not the weight value.
Example
- 21 -
5.3.3.
Gravity Acceleration Correction
When the scale (module) has been calibrated in the same place as it is being used, gravity
acceleration correction is not required.
A span error will appear if there is a gravity acceleration difference between the place where
the module has been calibrated and the place where it is being used. The gravity
acceleration correction calculates and corrects this span error by these gravity acceleration
correction values for both points (the place it was calibrated and the place it is being used in).
Note When the span is calibrated with actual load calibration, the gravity acceleration
correction settings are cleared, and the two gravity acceleration settings return to
their default values.
Functions Related To The Gravity Acceleration Correction
CAL26(Gravity acceleration of place of calibration) : The gravity acceleration where the
module has been calibrated.
CAL27(Gravity acceleration of place of use) : The gravity acceleration where the module
is being used.
Gravity Acceleration Table
Amsterdam
9.813 m/s2
Manila
9.784 m/s2
Athens
9.800 m/s2
Melbourne
9.800 m/s2
Auckland NZ
9.799 m/s2
Mexico City
9.779 m/s2
Bangkok
9.783 m/s2
Milan
9.806 m/s2
Birmingham
9.813 m/s2
New York
9.802 m/s2
Brussels
9.811 m/s2
Oslo
9.819 m/s2
Buenos Aires
9.797 m/s2
Ottawa
9.806 m/s2
Calcutta
9.788 m/s2
Paris
9.809 m/s2
Chicago
9.803 m/s2
Rio de Janeiro
9.788 m/s2
Copenhagen
9.815 m/s2
Rome
9.803 m/s2
Cyprus
9.797 m/s2
San Francisco
9.800 m/s2
Djakarta
9.781 m/s2
Singapore
9.781 m/s2
Frankfurt
9.810 m/s2
Stockholm
9.818 m/s2
Glasgow
9.816 m/s2
Sydney
9.797 m/s2
Havana
9.788 m/s2
Tainan
9.788 m/s2
Helsinki
9.819 m/s2
Taipei
9.790 m/s2
Kuwait
9.793 m/s2
Tokyo
9.798 m/s2
Lisbon
9.801 m/s2
Vancouver, BC
9.809 m/s2
London (Greenwich)
9.812 m/s2
Washington DC
9.801 m/s2
Los Angeles
9.796 m/s2
Wellington NZ
9.803 m/s2
Madrid
9.800 m/s2
Zurich
9.807 m/s2
- 22 -
AD-4430B
Illustration 8 Gravity Acceleration Graph
5.3.4.
Outline Of The Linearization Function
Even if zero and span points have been calibrated, weighing errors may occur between the
zero point and weighing capacity. This is a corrective function designed to non-linearly correct
weighing errors.
It is possible to input up to four points in addition to the zero point.
The zero point and each input point will be corrected to put them in a straight line.
Areas between input points that could not be corrected completely with straight line
correction or with secondary correction will be corrected using a curved line derived from
high-order equations.
When the actual load input for digital linearization is performed, the calibrated data will be
refreshed using zero point and final input point data. It is not necessary to calibrate again.
Illustration 9 Digital linearization
Zero
1st
4th
3rd
2nd
Mass value (Input point)
Weighing value
Before correction
Straight line correction
High-order equation correction
- 23 -
5.3.5.
The Actual Load Linearization Function
Set the digital linearization by loading/unloading masses.
Warm up the module for at least ten minutes to avoid the effects of temperature drift.
The input order should proceed from the smallest mass to the largest mass.
Step 1
Press the
+
key to enter to the calibration mode and
display
. Press the
key to start the calibration
and display
. Select
using the
key
and press the
key.
Step 2
is displayed.
If monitoring the current weighing value, press the
key.
When pressing the
key again,
is display.
Step 3
Placed nothing on the pan and wait for the stabilization ( LED ).
Press the
key.
is displayed for approximately
two seconds.
Step 4
is displayed.
If you want to check the current weighing value, press the
key.
When pressing the
key again,
is displayed.
Press the
key. The weight value (the current weighing
capacity) is displayed and the least digit of the value blinks.
Correct the value using the
and
key so as to be the
weight value used.
Step 5
Place the weight on the pan. Wait for the stabilization ( LED ).
Press the
key.
is displayed for approximately
two seconds.
Step 5
is displayed. Repeat step 4 and step 5. The procedure
proceeds in order of
→
→
.
Step 7
If finishing the input, proceed to step 8.
If you want to re-input the digital linearization, select the input point
using the
key. All data following the new input point will be
cleared.
Step 8
Press the
key.
is displayed and the inputted
data will be stored in the FRAM. At the same time, the calibrated
data is also refreshed. Press again the
key to return to
weighing mode.
If
X is displayed, an error has occurred. X : number.
Refer to “5.3.8. Error Codes For The Calibration” to take some measures.
The blinking decimal point means that the current value is not the weight value.
Example
- 24 -
AD-4430B
5.3.6.
The Function
Related To The Calibration
Step 1
Press the
+
key to enter to the calibration mode and display
.
Press the
key to start the calibration and display
.
Select
using the
key and press the
key.
Step 2
Select
using the
key and press the
key.
Step 3
Select a desired function number using the
key and press the
key.
The current data is displayed.
Step 4
When changing data, two methods of parameter selection and digital input depending on
the function are available.
Type
Description of method to change data
Parameter selection Only the available parameter is displayed and blinks.
Select a number using the
key.
Digital input
All the digits are displayed and a digit to be changed blinks.
Select a digit using the
key and
change the value using the
key.
After changing data, press the
key. The next function number is displayed.
When the value is not to be changed, press the
key to return to the function
number display.
Step 5
Press the
key to store new data in FRAM and
is displayed.
Press again the
key to return to the weighing mode.
The blinking decimal point means that the current value is not the weight value.
If a data exceeding the available range is inputted,
is displayed and data is
canceled.
- 25 -
Function No.
Parameter
and range
Function name
Description
Default
value
CAL02
0 to 0.0000
Decimal point
position
Decimal point position of the weight value
0
0.000
0.0
0.0000
0.00
0
CAL03
1 to 50
Minimum division
Minimum division (d) of the weight value
1
10
2
20
5
50
1
CAL04
1 to 99999
Weighing capacity
Weighing capacity of the module.
Weighing is possible up to the value of this setting plus
8 d. If the value exceeds this, overflow will occur and
will not be displayed. The decimal point position is the
same as the setting of CAL02.
70000
CAL05
0 to 100
Zero adjustment
range
Range to enable zero adjustment by the
(ZERO)
key Expressed as a percentage of the weighing
capacity with the calibration zero point as the center.
For example, if this is set to 2, the value in the range of
±2% of the weighing capacity with the calibration zero
point at the center will be to zero. When a power-ON
zero is performed, the initial zero point will be the
center.
2
CAL06
0.0 to 5.0
Zero tracking time
Performs zero tracking using this setting in
combination with the setting of the zero tracking band
CAL07. When 0.0, zero tracking will not be performed.
Unit : Second in 0.1 increments.
0.0
CAL07
0.0 to 9.9
Zero tracking band
Performs zero tracking using this setting in
combination with the setting of the zero tracking time
CAL06. When 0.0, zero tracking will not be performed.
Unit : d (minimum division) in 0.1 increments.
0.0
Displayed value
Displayed value
0 d
4.5 d
When CAL06 = 1.0, CAL07 = 4.5
1 second
Zero tracking follows the weight
value drifting a round the zero point
and adjusts to display as zero.
Zero tracking functions when the
weight value is drifting the range
shown in the graph.
1 second
2 second
5.0 d
4.5 d
4.0 d
3.5 d
3.0 d
2.5 d
2.0 d
1.5 d
1.0 d
0.5 d
0.0 d
When
CAL06 = 2.0,
CAL07 = 0.5
When
CAL06 = 1.0,
CAL07 = 4.5
- 26 -
AD-4430B
Function No.
Parameter
and range
Function name
Description
Default
value
CAL08
0.0 to 9.9
Stability detection
time
Performs stability detection using this setting in
combination with the setting of the stability detection
band CAL09. When 0.0, stability detection will not
be performed. (Stable all the time) Unit : Second in
0.1 increments.
1.0
CAL09
0 to 9
Stability detection
band
Performs stability detection using this setting in
combination with the setting of the stability detection
time CAL08. When 0, stability detection will not be
performed. (Stable all the time) Unit : d (minimum
division).
2
Stability detection outputs the STABLE
signal when changes in the weight
value are within a certain range during
a certain time.
CAL10
0 to 1
Tare and zero
adjustment when
unstable
Tare and zero adjustment when the weight value is
unstable.
0 : Disables both functions.
1 : Enables both functions.
1
CAL11
0 to 1
Tare when the gross
weight is negative
Tare when the gross weight is negative.
0 : Disables tare.
1 : Enables tare.
1
CAL12
0 to 1
Output when
overflow and
unstable
Standard serial output when the weight value
overflows and is unstable.
0 : Disables output.
1 : Enables output.
1
CAL13
1 to 3
Exceeding negative
gross weight
To judge when the negative gross weight is exceeded.
1 : Gross weight < -99999
2 : Gross weight < Negative weighing capacity
3 : Gross weight < -19d
1
CAL14
1 to 2
Exceeding negative
net weight
To judge when the negative net weight is exceeded.
1 : Net weight < -99999
2 : Net weight < Negative weighing capacity
1
CAL15
0 to 1
Clear the zero value
Select whether or not to clear the zero value.
0 : Disables.
1 : Enables.
1
CAL09
CAL08
Weight value
Time
CAL09
CAL08
STABLE signal
- 27 -
Function No.
Parameter
and range
Function name
Description
Default
value
CAL16
0 to 1
Zero setting when
power is turned on
Select whether or not to perform zero setting when
power is turned on.
0 : Disables.
1 : Enables.
0
CAL17
-7.0000
to 7.0000
Input voltage at zero
Input voltage from a load cell at zero. Unit : mV/V.
This value is determined in zero calibration during
the calibration with an actual load.
0.0000
CAL18
0.0001
to 9.9999
Input voltage at span
Input voltage from a load cell at span. Unit : mV/V.
This value and the value of CAL19 are determined
in span calibration during the calibration with an
actual load.
3.2000
CAL19
1 to 99999
Calibration weight
value corresponding
to Input voltage at
span
The calibration weight value corresponding to the
Input voltage at span of CAL18. When performing
digital span, CAL17, CAL18 and CAL19 are
required. The decimal point position is the same as
the setting of CAL02.
32000
NOTE:
*1 Record the setting values of CAL17, CAL18 and CAL19 in the “Setting List” at the end of the
manual to prepare against a failure.
*2 By changing the parameters of CAL17, CAL18 and CAL19, “zero calibration” and “span
calibration” can be adjusted optionally. (Digital span accuracy approximately 1/5000. The
accuracy varies depending on the load cell output accuracy and the conditions of calibration.)
Except for an emergency, perform calibration with an actual load.
CAL26
9.7500
to 9.8500
Gravity acceleration
of place of calibration
Gravity acceleration of the place of calibration.
Unit : in m/s2.
9.8000
CAL27
9.7500
to 9.8500
Gravity acceleration
of place of use
Gravity acceleration of the place where the scale is
being used. Unit : in m/s2.
9.8000
CAL28
0 to 1
Disable averaging
hold
0 : Disables.
1 : Enables.
0
CAL19
CAL18
CAL17
Input voltage
Displayed weight
0
- 28 -
AD-4430B
5.3.7.
The Function
Related To The Linearization Function
Confirm and change linearity settings.
To use this function, select
in the same way as calibration-related functions are
selected.
Function No.
Parameter
and range
Function name
Description
Default
value
Lnr01
1 to 4
Number of input
points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the set
value is between 1 and 2.
0
Lnr02
-7.0000
to 7.0000
Linear-zero
Voltage for linear-zero input. Unit : mV/V.
0.0000
Lnr03
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on CAL02.
0
Lnr04
0.0000
to 9.9999
Linear 1
Span
The span voltage between linear-zero and linear 1
input. Unit : mV/V.
0.0000
Lnr05
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on CAL02.
0
Lnr06
0.0000
to 9.9999
Linear 2
Span
The span voltage between linear-zero and linear 2
input. Unit : mV/V.
0.0000
Lnr07
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on CAL02.
0
Lnr08
0.0000
to 9.9999
Linear 3
Span
The span voltage between linear-zero and linear 3
input. Unit : mV/V.
0.0000
Lnr09
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on CAL02.
0
Lnr10
0.0000 to
9.9999
Linear 4
Span
The span voltage between linear-zero and linear 4
input. Unit : mV/V.
0.0000
- 29 -
5.3.8.
Error Codes For The Calibration
When an error occurs during calibration, the error number is displayed.
If calibration is finished without removing the error, the setting values will be restored to the state
before calibration.
Calibration errors and remedies
Error No.
Description
Treatment
C Er1
The display resolution (weighing
capacity / minimum division)
exceeds the specified value.
Make the minimum division greater or make the
weighing capacity smaller. The specified value
depends on specifications of the weighing system.
C Er2
Voltage at zero calibration
exceeds in the positive direction.
C Er3
Voltage at zero calibration
exceeds in the negative direction.
Check the load cell rating and connection. When
nothing is wrong with the rating and connection,
adjust the load cell output as described in the next
section. When the load cell or A/D converter may
be the cause of error, confirm this by using the
check mode.
C Er4
The value of the calibration weight
exceeds the weighing capacity.
C Er5
The value of the calibration weight
is less than the minimum division.
Use an appropriate calibration weight and
calibrate again.
C Er6
The load cell sensitivity is not
sufficient.
Use a load cell with higher sensitivity or make
the minimum division greater.
C Er7
Voltage at span calibration is less
than voltage at the zero point.
Check the load cell connection.
C Er8
The load cell output voltage is too
high when the load of the
weighing capacity is placed.
Use a load cell with a greater rating or make the
weighing capacity smaller.
5.3.9.
Adjustment Of The Loadcell Output
Add a resistor as shown below to adjust the load cell output.
Use a resistor with a high resistance value and a low temperature coefficient.
Illustration 10 Load cell output adjustment
Because the zero point of the module has a wide adjustable range, as long as there is no
problem with a load cell, correcting an output by the module is hardly ever required.
Before an output correction is carried out, confirm load cells (deformation, wiring mistakes,
contact with anything, or model selection etc.) and connections.
"C Er2"
When exceeding in the positive direction.
EXC+
SIG+
SIG-
EXC-
Resister added
"C Er3"
When exceeding in the negative direction.
EXC+
SIG+
SIG-
EXC-
Resister added
- 30 -
AD-4430B
5.4.
General Functions
In this section, the way of pre-setting and descriptions for the general functions is described.
General functions are divided into groups according to function and are indicated by the
group name with the function number.
NOTE: General functions determine the module performance and all of the settings are
stored in the FRAM.
5.4.1.
The Procedure To Store New Parameters
Step 1
Press the
+
key to enter to the function mode and display
.
Press the
key to start the function mode.
If returning to the weighing mode, press the
key.
Step 2
Press the
key to select the function group to be set.
Press the
key. The function group is as follows :
Display
Group Name
Basic functions
Hold functions
BCD Interface
Step 3
Press the
key to select the function number to be set.
Press the
key. The current setting value is displayed.
Step 4
When changing data, two methods of parameter selection and digital input depending on
the function are available.
Type
Description of method to change data
Parameter
selection
Only the available parameter is displayed and blinks.
Select a number using the
key.
Digital input
All the digits are displayed and a digit to be changed blinks.
Select a digit using the
key and
change the value using the
key.
After changing data, press the
key. The next function number is displayed.
When the value is not to be changed, press the
key to return to the function
number display.
Step 5
Press the
key to return to Step 2 .
Press the
key once to store new data in FRAM and to return to the weighing
mode.
The blinking decimal point means that the current value is not the weight value.
If a data exceeding the available range is inputted,
is displayed and data is
canceled.
- 31 -
5.4.2.
The Adjustment Of The Digital Filter
Adjust the digital filter using Fnc05 (digital filter).
The available settable range for the cutoff frequency is from 100 Hz to 0.7 Hz.
The cut off frequency is the frequency where vibrations declines to
times.
If the weighing value is unstable, set the cutoff frequency lower.
To make the response faster, set the cutoff frequency higher.
Cut off frequency is high.
Response rate is fast.
Weak against disturbances.
Cut off frequency is low.
Response rate is slow.
Strong against disturbances.
It is possible to make adjustments while watching the effects of the digital filter with your own eyes.
By pressing the
key while setting Fnc05 (digital filter), it is possible to check the weight
displayed.
key
Increase stability.
key
Return to the value setting display.
- 32 -
AD-4430B
5.4.3.
The List Of The Basic Functions
Function No.
Parameter
and range
Function name
Description
Default
value
Fnc01
0000 to 1111
Disable key
switch
Each digit of the setting corresponds to a key switch.
Only available in the weighing mode.
0 : Enables
Key assignment to each binary digit.
1 : Disables
4th
3rd
2nd
1st
0000
(2binary)
Fnc02
0 to 7
key
0 : None
1 : Manual print command
2 : Hold
3 : Alternate switch
4 : Momentary switch
5 : Changing between gross weight and net weight
6 : Clear the tare weight
7 : Clear the zero value
NOTE: Clearing the zero value can be enabled or
disabled by CAL15.
0
Fnc03
1 to 3
Display update
rate
1 : 20 times/second
2 : 10 times/second
3 : 5 times/second
1
Fnc04
0 to 9
x display
0 : None
1 : Zero tracking in progress
2 : Alarm (Zero range setting error, over)
3 :
key status
4: : Zero band
5 : HI output (Over the upper limit value)
6 : OK output (Between upper and lower limit values)
7 : LO output (Below the lower limit value)
8 : User input 1
9 : User output 1
0
Fnc05
0 to 16
Digital filter
Selects a cutoff frequency.
0: None
1: 100.0 Hz
9: 7.0 Hz
2: 70.0 Hz
10: 5.6 Hz
3: 56.0 Hz
11: 4.0 Hz
4: 40.0 Hz
12: 2.8 Hz
5: 28.0 Hz
13: 2.0 Hz
6: 20.0 Hz
14: 1.4 Hz
7: 14.0 Hz
15: 1.0 Hz
8: 10.0 Hz
16: 0.7 Hz
15
Fnc07
1 to 3
Hold
1 : Normal hold
2 : Peak hold
3 : Averaging hold
1
Fnc08
-99999 to
99999
Set value of
near-zero
The reference value for near-zero.
Decimal point position is linked to CAL02.
10
- 33 -
Function No.
Parameter
and range
Function name
Description
Default
value
Fnc09
1 to 2
Mass to be
compared with
near-zero
Item to be compared with near-zero.
1 : Gross weight
2 : Net weight
1
Fnc10
-99999 to 99999
Set value of
upper limit
Reference value for the upper limit.
Decimal point position is linked to CAL02.
10
Fnc11
-99999 to 99999
Set value of
lower limit
Reference value for the lower limit.
Decimal point position is linked to CAL02.
-10
Fnc12
1 to 2
Mass to be
compared with
upper and
lower limit
Item to be compared with the upper and lower limit.
1 : Gross weight
2 : Net weight
1
Fnc13
1 to 2
Output logic of
upper and
lower limit
Logic used when the result of the comparison with
the upper and lower limit is output
1 : Positive logic
2 : Negative logic
1
5.4.4.
The List Of The Hold Function
Function No.
Parameter
and range
Function name
Description
Default
value
HLd01
0.00 to 9.99
Averaging time Time to perform the averaging. Unit : second.
0.00 is not averaged.
0.00
HLd02
0.00 to 9.99
Start wait time Time to wait before commencing hold or averaging.
Unit : second.
0.00
HLd03
0 to 2
Condition of
automatic start
Condition to start the hold or averaging automatically.
0 : Do not use automatic start.
1 : Above the near-zero range, and stable.
2 : Above the near-zero range.
0
HLd04
0 to 1
Release due to
falling control
input
Release when control input is falling.
0 : Do not release
1 : Release
1
HLd05
0.00 to 9.99
Time release
Release after a set amount of time has passed.
Unit : second. 0.00 is not averaged.
0.00
HLd06
0 to 99999
Release due to
range of
fluctuation
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is the
same as the setting of CAL02.
0
HLd07
0 to 1
Release in
near-zero range
Release when the weighing value is in the near-zero
range.
0 : Do not release.
1 : Release.
0
This hold function only works when Fnc07 (hold) is set to 2 (peak hold) or 3 (averaging hold).
This hold function has no function when Fnc07 (hold) is set to 1 (normal hold).
HLd01 (averaging time) works only when Fnc07 (hold) is set to 3 (averaging hold).
- 34 -
AD-4430B
5.4.5.
The List Of The BCD Function
Function No.
Parameter
and range
Function name
Description
Default
value
bcd01
1 to 4
Output data
1 : Displayed weight
2 : Gross weight
3 : Net weight
4 : Data specified by BCD input
1
bcd02
1 to 3
Data transfer
mode
1 : Stream
2 : Automatic printing
3 : Manual printing
1
bcd03
5 to 1000
Data transfer rate
5 times/second
10 times/second
Display linked to Fnc03
20 times/second
100 times/second
1000 times/second
20
bcd04
1 to 2
Output logic of
data transfer
1 : Positive logic
2 : Negative logic
2
bcd05
1 to 2
Output logic of
negative sign
1 : Positive logic
2 : Negative logic
2
bcd06
1 to 2
Output logic of
status
1 : Positive logic
2 : Negative logic
2
bcd07
1 to 2
Output logic of
strobe
1 : Positive logic
2 : Negative logic
2
bcd08
0 to 5
Input selection
0 : None
1 : Clear zero
2 : Clear tare
3 : Changing between gross weight and net weight
4 : Print command
5 :
key
3
bcd09
0 to 12
Output selection
0 : None
1 : Stabilization
2 : During tare
3 : Zero band
4 : Hold busy state
5 : HI output (Over upper limit)
6 : OK output (Within upper and lower limits)
7 : LO output (Under lower limit)
8 : During operating weighing (On)
9 : During operating weighing (1 Hz)
10 : During operating weighing (50 Hz)
11 : Alarm (Zero correction error and tare error )
12 : Busy
key
1
- 35 -
6.
Interface
6.1.
The Parallel BCD Output
6.1.1.
Timing Chart Of The BCD Output
The strobe signal, will be turned on when starting data transmission, will be turned off after
the strobe time. The strobe time is a half of a period to transmit data.
Data output
Strobe signal
6.1.2.
Terminals Of The BCD Output
BCD Output And I/O Output
Output type
Open Collector
Isolation
Photo coupler
Output current
50mA max.
Saturation voltage 0.5V max. @50mA
I/O Input
Input type
DC input
Open circuit voltage Approx. 5V
Off current
0.1mA max.
On current
2mA min.
Saturation voltage
2V
Anti - chattering time 10msec.
6.1.3.
The State Of The BCD Output
Over Flow
The whole bits of the weighing data (1x100 to
8x104) are set to "1". Negative sign and
status bits are according to the situation.
Except Weighing Mode
The whole bits of the weighing data (1x100 to
8x104) are set to "1". Over bit is set to "1".
Other bits are set to "0". The strobe signal is
not output. A strobe signal is output once
when moving the mode from weighing mode
with stream mode.
1/2 Cycle
1/2 Cycle
1/2 Cycle
1
0
+30V max.
Output terminal
COM (Common)
Input terminal
+5V
Inside of the module
750Ω
- 36 -
AD-4430B
7.
Maintenance
7.1.
Error Messages
If an error message is displayed, refer to the chart below to take proper measures.
Error message
Cause
Treatment
Program checksum error.
Repair is required.
Data can not be acquired from
the A/D converter.
Repair is required.
Correct data can not be read
from the FRAM.
Perform initialization. If initialization does not clear
the error, repair is required.
Calibration data is not correct. Perform calibration.
X
Calibration error. X : number.
Refer to “5.3.8. Error Codes for the Calibration ”.
The setting value is out of the
settable range.
Check the setting value and set again.
7.2.
Verifying The Load Cell Connections ( DIAGNOS )
7.2.1.
Guideline To Verify The Load Cell Connections
The faulty wiring or disconnection of the load cell can be checked.
This verification is useful for new setting, pre-measurement inspection and periodical inspection.
No.
Check Item
Target
Judgment Criteria (Generally)
1 Load cell input voltage
Between SEN+
SEN-
3 V or more
2 SEN+ voltage
Between SEN+
AGND 4 V or more
3 SEN- voltage
Between SEN-
AGND 1 V or less
4 Load cell output voltage
Between SIG+
SIG-
Within ±35 mV
5 Load cell output rate
Between SIG+
SIG-
Within ±7 mV/V
6 SIG+ voltage
Between SIG+
AGND 1 V to 4 V
7 SIG- voltage
Between SIG-
AGND 1 V to 4 V
8 Internal temperature
-20 °C to +60 °C
AGND : Internal analog circuit ground
SIG- : Load cell output (-)
SIG+ : Load cell output (+)
EXC- : Load cell input (-)
SEN- : Sensing input (-)
SEN+ : Sensing input (+)
EXC+ : Load cell input (+)
SHLD : Shield. Frame ground.
Illustration 11 Wire Name of Load Cell
- 37 -
7.2.2.
The Operation With The BCD Input
When inputting "ON" level to the "DIAGNOS" terminal of the BCD input terminal for 1 second
or more, the self-check mode is started and each results are displayed and outputted. The
scan condition of the self-check mode can be specified by the following BCD input.
BCD Input
No.
DIAGNOS
GRS/NET
HOLD
TARE
ZERO
Scan
ON
OFF
OFF
OFF
OFF
1
ON
OFF
OFF
OFF
ON
2
ON
OFF
OFF
ON
OFF
3
ON
OFF
OFF
ON
ON
4
ON
OFF
ON
OFF
OFF
5
ON
OFF
ON
OFF
ON
6
ON
OFF
ON
ON
OFF
7
ON
OFF
ON
ON
ON
8
ON
ON
OFF
OFF
OFF
7.2.3.
The Operation With The Switch
Step 1
Press the
key while pressing and holding the
key (
+
) to
display
. Press the
key to enter to the function mode.
If returning to the weighing mode, press the
key.
Step 2
Press the
key while pressing and holding the
key (
+
) to
display
of the check mode. Press the
key to display the check item.
Step 3
Select the self-check mode
using the
key. Press the
key to enter
to it. The selected check is performed and displayed the results for approx. 16 seconds.
Each measurement value can be displayed using the
key.
7.2.4.
The Result Of Verifying The Connection
While doing the scan and display exchange,
is displayed and
is
outputted. The result of scan is displayed a value
added error codes.
When an error is nothing,
is displayed and
is outputted.
When detecting an error,
is displayed and
is displayed a value.
is added error codes. BCD output is not included decimal point. The resistance of the load
cell cable can be calculated as follows : (Resister value of the load cell) × (No.3) / (No.1)
No. Check item
Status LED
Display Range
Error Code
1 Load cell input voltage
0.001 V
1
2 SEN+ voltage
0.001 V
2
3 SEN- voltage
0.001 V
4
4 Load cell output voltage
0.001 mV
8
5 Load cell output rate
0.0001 mV/V
16
6 SIG+ voltage
0.001 V
32
7 SIG- voltage
0.001 V
64
8 Internal temperature
0.1 °C
128
: lighted
: not lighted
- 38 -
AD-4430B
7.3.
Check Mode
The check mode checks the performance of the display, key switches and external I/O.
7.3.1.
Entering The Check Mode
Step 1 Press the
key while pressing and holding the
key (
+
) to
display
.
If returning to the weighing mode, press the
key.
Step 2 Press the
key while pressing and holding the
key (
+
) to
display
of the check mode. Press the
key to display the check item
Step 3 Select the check item using the
key. Press the
key to enter to it.
Press the
key when exiting it
Display
Item To Be Checked
Key check
BCD check
A/D converter output check (Load cell)
Internal count
Program version
Serial number
Program checksum
Memory checksum (EEPROM)
CALF check (CAL01 to CAL28)
7.3.2.
Verifying The Switch Operation
When pressing the key, the corresponding segment moves.
If stopping the current check mode, press the
key twice.
7.3.3.
Monitoring The A/D Converter ( For Load cell Output)
The load cell output voltage rate is displayed in unit of mV/V .
ex. If internal count is 1.2345 mV/V ,
If value is above ±7 mV/V, a damage and connection error of the load cell may cause . Refer
to "7.5. Verifying The Load cell Connections Using Multimeter".
- 39 -
7.3.4.
Monitoring The Internal Value
Internal count ( 10 times of weighing value ) is displayed.
ex. If internal count is 123.
7.3.5.
Monitoring The Program Version
Program version is displayed.
ex. If version is 1.00 .
7.3.6.
Monitoring The Serial Number
ex. If low five digits of serial number is displayed.
7.3.7.
Monitoring The Checksum Of The Program
Checksum of the program is displayed.
ex. If checksum is EF.
7.3.8.
Monitoring The Checksum Of An Internal FRAM
Checksum of FRAM is displayed.
Memory of the general function is not checked
ex. If checksum is EF.
7.3.9.
Displaying Function Parameters For Calibration
Function parameters rated to calibration can be displayed.
Note
Concerning these operation and details, refer to "5.3.6. The Function
Related To
The Calibration".
Parameters can not be changed here.
- 40 -
AD-4430B
7.4.
Initializing Parameters
The initialization mode restores the parameters of the default values to the FRAM.
Three types of initialization mode are available as shown below.
Initialization mode
Display
Description
RAM
initialization
RAM memory is initialized. The center of zero and tare
value will be restored to 0.
General functions
initialization
Data of the general functions stored in the FRAM are
reset to factor settings.
All data
initialization
All data stored in the FRAM are initialized. So calibration
data is included, calibrate the module again.
7.4.1.
Initializing Mode For RAM And Function Parameters
Step 1
Press the
key while pressing and holding the
key (
+
) to
display
. If returning to the weighing mode, press the
key.
Step 2
Press the
key while pressing and holding the
key (
+
) to
display
of the check mode.
Step 3
Select the initialization mode
using the
key. Press the
key.
Step 4
Select an item to be initialized using the
key. Press the
key.
Step 5
Check that all status LED are blinking.
If performing the initialization, press the
key for 3 seconds or more.
After initialization, all segments light and return to the weighing mode.
If canceling the initialization, press the
key return to the weighing mode.
7.4.2.
Initializing The Whole Data
Step 1
In the standby mode (While turning off the module),
Press
+
key to display
.
If returning to the weighing mode, press the
key.
Step 2
Press the
key. to enter to calibration mode.
Step 3
Select all initialization mode using the
key. Press the
key.
Step 4
Check that all status LED are blinking.
If performing the initialization, press the
key for 3 seconds or more.
After initialization, all segments light and return to the weighing mode.
If canceling the initialization, press the
key return to the weighing mode.
- 41 -
7.5.
Verifying The Load cell Connections Using Multimeter
The load cell connection can be checked easily using a digital multimeter.
Illustration 12 Connection Check of Load Cell shows points to confirm the load cell connection.
When a summing box is used, the same measurement must be performed, even internally.
Illustration 12 Connection Check of Load Cell
7.5.1.
Check List Of The Load cell Connections
Measurement
Point
Item To Be Measured
Judging The Voltage
EXC+ SEN+ A drop voltage of the load
cell cable on EXC+ side.
Normally 100mV or less. However, it may exceed 1V
when an extremely long load cell cable is used. For the
4-wire connection, it must be 0V.
EXC+ EXC- Input voltage of load cell
Normal range is between 4.75 V to 5.25 V.
SEN- EXC- A drop voltage of the load
cell cable on EXC- side.
Normally 100mV or less. However, it may exceed 1V
when an extremely long load cell cable is used. For the
4-wire connection, it must be 0V.
SIG- EXC- Center point voltage
Approximately 2.5 V of a half of excitation voltage.
SIG+
SIG- Output voltage of load cell
Generally within 0 V to 15mV. Theoretical value is
calculated from the load cell rated capacity, actual load
and excitation voltage.
Output voltageof load cell
Center point voltage
Drop voltage between
EXC- and SEN-.
Input voltage of load cell
Drop voltage between
EXC+ and SEN+.
Load cell
Shield cable
SIG-
EXC-
SEN-
SIG+
SEN+
EXC+
- 42 -
AD-4430B
When the module does not operate properly, write the required items in the table below and contact
your local A&D dealer.
Item
User’s Usage Circumstances
Model Number, Rated, Measurement
Value etc.
Description
Connection method
4-wire connection
6-wire connection
When using the 4-wire connection,
connect between EXC+ and SEN+ and
between EXC- and SIG-.
Model number of load cells
Rated capacity of load cells
[Unit ]
Rated output of load cells
[mV/V]
Allowable overload of load
cells
[%]
Number of load cells used
[pieces]
Use of summing box
Length of the extension cable
[m] Length from the module to the summing
box.
Initial load on the module
[Unit ]
Weighing capacity of the
module
[Unit ] Write all digits including decimal figures.
Example : 0.002 kg
Weighing capacity of the
module
[Unit ] Write all digits including decimal figures.
Example : 10.000 kg
Load cell output during initial
load (when no load is applied)
[mV/V] From -0.1 mV/V to rated sensitivity value
(With the first load)
Relation between load cell
output and load of mass of
capacity (or chosen mass )
Load :
[Unit ]
Load cell output :
[mV/V]
When loaded to capacity, the output
value of the initial load + the rated output
value of the load cell. (It must be within
allowable overload.)
Measurement point
Measurement contents
Measurement result
EXC+
SEN+
A drop voltage of the load cell cable on EXC+ side.
[mV]
EXC+
EXC-
Input voltage of load cell
[V]
SEN-
EXC-
A drop voltage of the load cell cable on EXC- side.
[mV]
SIG-
EXC-
Center point voltage
[V]
SIG+
SIG-
Output voltage of load cell
[mV]
- 43 -
7.6.
The Parameter List For The Function List
When performing maintenance, use the following list as a memorandum.
When making inquiries about the product, inform your local A&D dealer of the user settings.
7.6.1.
Basic Functions
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
Fnc01
0000 to 1111
Disable key
switch
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0 : Enables
Key assignment to each binary digit.
1 : Disables
4th
3rd
2nd
1st
0000
(binary)
Fnc02
0 to 7
key
0 : None
1 : Manual print command
2 : Hold
3 : Alternate switch
4 : Momentary switch
5 : Changing between gross weight and net weight
6 : Clear the tare weight
7 : Clear the zero value
NOTE: Clearing the zero value can be enabled
or disabled by CAL15.
0
Fnc03
1 to 3
Display
update rate
1 : 20 times/second
2 : 10 times/second
3 : 5 times/second
1
Fnc04
0 to 9
x display
0 : None
1 : Zero tracking in progress
2 : Alarm (Zero range setting error, over)
3 :
key status
4: : Zero band
5 : HI output (Over the upper limit value)
6 : OK output (Between upper and lower limit values)
7 : LO output (Below the lower limit value)
8 : User input 1
9 : User output 1
0
Fnc05
0 to 16
Digital filter
Selects a cutoff frequency.
0: None
1: 100.0 Hz
9: 7.0 Hz
2: 70.0 Hz
10: 5.6 Hz
3: 56.0 Hz
11: 4.0 Hz
4: 40.0 Hz
12: 2.8 Hz
5: 28.0 Hz
13: 2.0 Hz
6: 20.0 Hz
14: 1.4 Hz
7: 14.0 Hz
15: 1.0 Hz
8: 10.0 Hz
16: 0.7 Hz
15
Fnc07
1 to 3
Hold
1 : Normal hold
2 : Peak hold
3 : Averaging hold
1
Fnc08
-99999 to
99999
Set value of
near-zero
The reference value for near-zero.
Decimal point position is linked to CAL02.
10
- 44 -
AD-4430B
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
Fnc09
1 to 2
Mass to be
compared
with
near-zero
Item to be compared with near-zero.
1 : Gross weight
2 : Net weight
1
Fnc10
-99999 to
99999
Set value of
upper limit
Reference value for the upper limit.
Decimal point position is linked to CAL02.
10
Fnc11
-99999 to
99999
Set value of
lower limit
Reference value for the lower limit.
Decimal point position is linked to CAL02.
-10
Fnc12
1 to 2
Mass to be
compared
with upper
and lower
limit
Item to be compared with the upper and lower limit.
1 : Gross weight
2 : Net weight
1
Fnc13
1 to 2
Output logic
of upper
and lower
limit
Logic used when the result of the comparison with
the upper and lower limit is output
1 : Positive logic
2 : Negative logic
1
7.6.2.
The List Of The Hold Function
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
HLd01
0.00 to 9.99
Averaging
time
Time to perform the averaging. Unit : second.
0.00 is not averaged.
0.00
HLd02
0.00 to 9.99
Start wait
time
Time to wait before commencing hold or averaging.
Unit : second.
0.00
HLd03
0 to 2
Condition of
automatic
start
Condition to start the hold or averaging automatically.
0 : Do not use automatic start.
1 : Above the near-zero range, and stable.
2 : Above the near-zero range.
0
HLd04
0 to 1
Release
due to
falling
control input
Release when control input is falling.
0 : Do not release
1 : Release
1
HLd05
0.00 to 9.99
Time
release
Release after a set amount of time has passed.
Unit : second. 0.00 is not averaged.
0.00
HLd06
0 to 99999
Release
due to
range of
fluctuation
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is
the same as the setting of CAL02.
0
HLd07
0 to 1
Release in
near-zero
range
Release when the weighing value is in the
near-zero range.
0 : Do not release.
1 : Release.
0
- 45 -
7.6.3.
The List Of The BCD Function
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
bcd01
1 to 4
Output
data
1 : Weighing value
2 : Gross weight
3 : Net weight
4 : Data specified at BCD input
1
bcd02
1 to 3
Data
transfer
mode
1 : Stream
2 : Automatic printing
3 : Manual printing
1
bcd03
5 to 1000
Data
transfer
rate
5 times/second
10 times/second
Display linked to Fnc03
20 times/second
100 times/second
1000 times/second
20
bcd04
1 to 2
Output
logic of
data
transfer
1 : Positive logic
2 : Negative logic
2
bcd05
1 to 2
Output
logic of
negative
sign
1 : Positive logic
2 : Negative logic
2
bcd06
1 to 2
Output
logic of
status
1 : Positive logic
2 : Negative logic
2
bcd07
1 to 2
Output
logic of
strobe
1 : Positive logic
2 : Negative logic
2
bcd08
0 to 5
Input
selection
0 : None
1 : Clear zero
2 : Clear tare
3 : Changing between gross weight and net
weight
4 : Print command
5 :
key
3
bcd09
0 to 12
Output
selection
0 : None
1 : Stabilization
2 : During tare
3 : Zero band
4 : Hold busy state
5 : HI output (Over upper limit)
6 : OK output (Within upper and lower limits)
7 : LO output (Under lower limit)
8 : During operating weighing (On)
9 : During operating weighing (1 Hz)
10 : During operating weighing (50 Hz)
11 : Alarm (Zero correction error and tare error )
12 : Busy
key
1
- 46 -
AD-4430B
7.6.4.
The List Of The Calibration
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
CAL02
0 to 0.0000
Decimal point
position
Decimal point position of the weight value
0
0.00
0.0000
0.0
0.000
0
CAL03
1 to 50
Minimum
division
Minimum division (d) of the weight value
1
2
5
10
20
50
1
CAL04
1 to 99999
Weighing
capacity
Weighing capacity of the module.
Weighing is possible up to the value of this
setting plus 8 d. If the value exceeds this,
overflow will occur and will not be displayed. The
decimal point position is the same as the setting
of CAL02.
70000
CAL05
0 to 100
Zero
adjustment
range
Range to enable zero adjustment by the
(ZERO) key Expressed as a percentage of
the weighing capacity with the calibration zero
point as the center. For example, if this is set to 2,
the value in the range of ±2% of the weighing
capacity with the calibration zero point at the
center will be to zero. When a power-ON zero is
performed, the initial zero point will be the center.
2
CAL06
0.0 to 5.0
Zero tracking
time
Performs zero tracking using this setting in
combination with the setting of the zero tracking
band CAL07. When 0.0, zero tracking will not be
performed. Unit : Second in 0.1 increments.
0.0
CAL07
0.0 to 9.9
Zero tracking
band
Performs zero tracking using this setting in
combination with the setting of the zero tracking
time CAL06. When 0.0, zero tracking will not be
performed. Unit : d (minimum division) in 0.1
increments.
0.0
CAL08
0.0 to 9.9
Stability
detection time
Performs stability detection using this setting in
combination with the setting of the stability
detection band CAL09. When 0.0, stability
detection will not be performed. (Stable all the
time) Unit : Second in 0.1 increments.
1.0
CAL09
0 to 9
Stability
detection
band
Performs stability detection using this setting in
combination with the setting of the stability
detection time CAL08. When 0, stability
detection will not be performed. (Stable all the
time) Unit : d (minimum division).
2
CAL10
0 to 1
Tare and zero
adjustment
when
unstable
Tare and zero adjustment when the weight
value is unstable.
0 : Disables both functions.
1 : Enables both functions.
1
CAL11
0 to 1
Tare when
the gross
weight is
negative
Tare when the gross weight is negative.
0 : Disables tare.
1 : Enables tare.
1
- 47 -
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
CAL12
0 to 1
Output when
overflow and
unstable
Standard serial output when the weight value
overflows and is unstable.
0 : Disables output.
1 : Enables output.
1
CAL13
1 to 3
Exceeding
negative
gross weight
To judge when the negative gross weight is
exceeded.
1 : Gross weight < -99999
2 : Gross weight < Negative weighing
capacity
3 : Gross weight < -19d
1
CAL14
1 to 2
Exceeding
negative net
weight
To judge when the negative net weight is
exceeded.
1 : Net weight < -99999
2 : Net weight < Negative weighing capacity
1
CAL15
0 to 1
Clear the zero
value
Select whether or not to clear the zero value.
0 : Disables.
1 : Enables.
1
CAL16
0 to 1
Zero setting
when power
is turned on
Select whether or not to perform zero setting
when power is turned on.
0 : Disables.
1 : Enables.
0
CAL17
-7.0000
to 7.0000
Input voltage
at zero
Input voltage from a load cell at zero. Unit :
mV/V.
This value is determined in zero calibration
during the calibration with an actual load.
0.0000
CAL18
0.0001
to 9.9999
Input voltage
at span
Input voltage from a load cell at span. Unit :
mV/V.
This value and the value of CAL19 are
determined in span calibration during the
calibration with an actual load.
3.2000
CAL19
1 to 99999
Calibration
weight value
correspondin
g to Input
voltage at
span
The calibration weight value corresponding to
the Input voltage at span of CAL18. When
performing digital span, CAL17, CAL18 and
CAL19 are required. The decimal point
position is the same as the setting of CAL02.
32000
CAL26
9.7500
to 9.8500
Gravity
acceleration
of place of
calibration
Gravity acceleration of the place of calibration.
Unit : in m/s2.
9.8000
CAL27
9.7500
to 9.8500
Gravity
acceleration
of place of
use
Gravity acceleration of the place where the
scale is being used. Unit : in m/s2.
9.8000
CAL28
0 to 1
Disable
averaging
hold
0 : Disables.
1 : Enables.
0
- 48 -
AD-4430B
7.6.5.
The List Of The Linearization Function
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting MEMO
Lnr01
1 to 4
Number of
input points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the
set value is between 1 and 2.
0
Lnr02
-7.0000
to 7.0000
Linear-zero Voltage for linear-zero input. Unit : mV/V.
0.0000
Lnr03
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on CAL02.
0
Lnr04
0.0000
to 9.9999
Linear 1
Span
The span voltage between linear-zero and linear
1 input. Unit : mV/V.
0.0000
Lnr05
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on CAL02.
0
Lnr06
0.0000
to 9.9999
Linear 2
Span
The span voltage between linear-zero and linear
2 input. Unit : mV/V.
0.0000
Lnr07
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on CAL02.
0
Lnr08
0.0000
to 9.9999
Linear 3
Span
The span voltage between linear-zero and linear
3 input. Unit : mV/V.
0.0000
Lnr09
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on CAL02.
0
Lnr10
0.0000 to
9.9999
Linear 4
Span
The span voltage between linear-zero and linear
4 input. Unit : mV/V.
0.0000
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN
Telephone: [81] (3) 5391-6132
Fax: [81] (3) 5391-6148
A&D ENGINEERING, INC.
1756 Automation Parkway, San Jose, California 95131, U.S.A.
Telephone: [1] (408) 263-5333
Fax: [1] (408)263-0119
A&D INSTRUMENTS LIMITED
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom
Telephone: [44] (1235) 550420
Fax: [44] (1235) 550485
A&D AUSTRALASIA PTY LTD
32 Dew Street, Thebarton, South Australia 5031, AUSTRALIA
Telephone: [61] (8) 8301-8100
Fax: [61] (8) 8352-7409
A&D KOREA Limited
한국에이.엔.디(주)
대한민국 서울시 영등포구 여의도동 36-2 맨하탄 빌딩 8층 우편 번호 150-749
( Manhattan Building 8th Floor, 36-2 Yoido-dong, Youngdeungpo-gu, Seoul, 150-749 Korea )
전화: [82] (2) 780-4101
팩스: [82] (2) 782-4280
OOO A&D RUS
OOO "ЭЙ энд ДИ РУС"
121357, Российская Федерация, г.Москва, ул. Верейская, дом 17
( Business-Center "Vereyskaya Plaza-2" 121357, Russian Federation, Moscow, Vereyskaya Street 17 )
тел.: [7] (495) 937-33-44
факс: [7] (495) 937-55-66
A&D INSTRUMENTS INDIA PRIVATE LIMITED
( 509, Udyog Vihar, Phase-
, Gurgaon - 122 016, Haryana, India )
: 91-124-4715555
: 91-124-4715599