Alert J5 Instruction Manual

Specifications

Brand: Alert
Category: Test Equipment
Model: J5 (Instruction Manual)
File Type: PDF
File Size: —
Pages: 20

Read Online

Read the Alert J5 Instruction Manual online using the built-in PDF viewer.

Text of the manual

Federal Communications Commission (FCC) Disclaimer

Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC rules. Operations are subject to the following 2 conditions: 1. This device may not cause harmful interference 2. This device must accept any interference received, including interference that may cause undesired operations Industry Canada This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: 1. This device may not cause interference, and 2. This device must accept any interference, including interference that may cause undesired operation of the device. Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts de licence. L’exploitation est autorisée aux deux conditions suivantes: 1. l’appareil ne doit pas produire de brouillage, et 2. l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement. Alcohol Countermeasure Systems Corp 60 International Boulevard Toronto, Ontario M9W 6J2 CANADA ALERT, Alcohol Countermeasure Systems, and the Molly are trademarks of Alcohol Countermeasure Systems (International) Inc. and are used under license. © Alcohol Countermeasure Systems Corp 2011 The information disclosed in this document is the valuable property of Alcohol Countermeasure Systems Corp and all copyright and other proprietary rights to this document are reserved. No reproduction of this document is permitted without the prior written consent of Alcohol Countermeasure Systems Corp.

TABLE OF CONTENTS

INTRODUCTION��������������������������������������������������������������v SAFETY AND PRECAUTIONS����������������������������������������v WARRANTY��������������������������������������������������������������������vi MAINTENANCE��������������������������������������������������������������vi CONTACT INFORMATION����������������������������������������������vi GETTING STARTED��������������������������������������������������������1 Buttons and Menus��������������������������������������������������2 BAC/BrAC Format���������������������������������������������������3 Power On/Shut Down����������������������������������������������3 Activating�����������������������������������������������������������������4 Battery���������������������������������������������������������������������4 TESTING��������������������������������������������������������������������������5 Taking a Breath Test������������������������������������������������5 Recalling a Breath Test��������������������������������������������7 PRINTING������������������������������������������������������������������������8 Serial Printing����������������������������������������������������������8 Inserting/Charging the Printer Battery�������������������10 Loading the Paper Roll������������������������������������������11 MAINTENANCE & TROUBLESHOOTING���������������������12 Troubleshooting Table��������������������������������������������12 INDEX����������������������������������������������������������������������������13

Thank you for purchasing the ALERT™ J5 – a breath

alcohol tester (BAT) that measures the alcohol level in human breath. Based on your jurisdiction, your ALERT J5 comes with preset levels that indicate whether the alcohol breath test result is a fail (red LED light), warning (amber LED light) or pass (green LED light). SAFETY AND PRECAUTIONS Take the following precautions to ensure smooth and uninterrupted use of your device: • Use the product for its intended purpose only • Use AA Alkaline batteries only • Do not store in extreme heat or cold and avoid exposing the device to direct sunlight for extended periods of time • Do not blow smoke or any fluids into the mouthpiece, as these can damage the sensor • Do not open the BAT case. Doing so will void the warranty and may also damage the components within • Keep all magnets and transmitting devices (such as mobile phones) away from the BAT to minimize RF interference • If you spill any liquid on the ALERT J5, remove all excess fluid with a cloth and allow the device to dry thoroughly before attempting to power on. Do not power on in case of excessive water damage

All ACS products are warranted to be free of defects in

workmanship and materials for a period of one year from the date of shipment. ACS agrees to replace or repair any defective device, provided the defect was not caused by misuse or mishandling. The ALERT J5 being returned for warranty repairs must be properly packaged and shipped prepaid to an authorized service provider. MAINTENANCE ALERT J5 requires periodic maintenance and calibration. After testing and calibrating at the factory, ALERT J5 is deactivated so it can be stored without losing any clock time from the calibration period. You need to activate it to start using it (see the “Activating” section). Once activated, ALERT J5 automatically increments the date for the next calibration by 365 days from the date of activation. NOTE: ALERT J5 is calibrated for optimal performance at the point of shipment. CONTACT INFORMATION [email protected] acs-corp.com Alcohol Countermeasure Systems Corp 60 International Boulevard Toronto, Ontario M9W 6J2 Canada T +1 416 619 3500 F +1 416 619 3501 Comp

The ALERT J5 is a professional series breath alcohol

tester (BAT) that is accurate and easy to use. It measures breath alcohol concentration (BrAC), which is directly proportional to blood alcohol concentration (BAC). This section familiarizes you with the basic features, such as the menus, buttons, and batteries of the device. Display Battery Compartment Intake Port Serial Port Indicator Light Two-Button Operation Mouthpiece

Buttons and Menus

The ALERT J5 has an interface similar to a cellular phone, which allows you to navigate all menus using the two buttons. Either button powers on the device. To navigate using the two buttons: • Use the right button to scroll down to your selection (scrolling up is not possible) • Use the left button to select (Enter) the option The device has the following menus: • Last 100 tests: Recalls and displays the last 100 test results. You have the option of printing using the right button. Refer to the “Printing” section for details • Calibration: Shows the date of last calibration. You can re-calibrate using the right button. Specialized equipment is required to accurately calibrate ALERT J5 and you should not attempt to calibrate it yourself • Deactivation/Activation: A service menu option for use by manufacturer or distributer. Units are supplied to customer in “Activated” state • Date/Time: Shows the current date and time of the device. Use the right button to scroll through the digits and the left button to change the values. You can also set the time to a 12 or 24 hour format • Settings: -- Change the units of the BrAC/BAC. Using the right button, scroll to the desired unit and press left button to select. The units available are: mg%, %BAC, g/L, mg/L, and μg/100mL -- Backlight: Set the brightness of the backlight and the time you would like the light to stay on

Shutdown timer: Set the time for auto shut off. This

time can be between 30 seconds and 900 seconds (15 minutes) • Language: Set the display language using the right button. Scroll through to view available options • Shut down: Use this menu to manually shut down the device without waiting for automatic shut off. This option does not deactivate the device; it only shuts off the device BAC/BrAC Format The blood alcohol concentration (BAC) or the breath alcohol concentration (BrAC) format is available in five formats; mg%, mg/L, %BAC, g/L and μg/100mL. You can select the desired unit from the Settings menu. Power On/Shut Down Press either of the two buttons to power on the device. A flashing amber LED light indicates that the device is warming up for a test, and you will see a WAIT message on the display (for approximately 3 seconds). A green flashing LED light with a READY message on display indicates that the device is ready and you can now do a breath test. You can manually shut down the device by selecting Shut down from the Main Menu. Alternatively, the device shuts down automatically after a specific period of time. You can set this time to suit your needs from the Settings menu option.

ALERT J5 is shipped deactivated after it is calibrated

and passes inspection at the factory. This allows the device to be stored without losing any time from the calibration period. You have to activate the device to start using it. Once activated, the date for the next calibration increments by 365 days from the date of activation. A Not Activated message when you initially power on the device indicates that it is not activated and no test can be performed until it is activated. NOTE: This feature is displayed on the Main Menu, only if your ALERT J5 has been previously calibrated. It is not possible for you to deactivate the device. To activate: 1. From the Not Activated screen, press the left button to activate it. The date the device was deactivated is shown. 2. Press the right button to activate the device. The calibration due date is now 365 days from the current date. Battery The ALERT J5 requires two AA Alkaline batteries to operate. To insert, remove the battery compartment rear cover (press the tab to release).

Taking a Breath Test

NOTE: Certain things like drinking, eating, and smoking may distort the results of a breath alcohol test. It is recommended that you rinse your mouth with water or wait for at least 20 minutes before taking a test. 1. Take a new mouthpiece and insert it on top of the ALERT J5 device. With the display facing you, connect the mouthpiece to the device by aligning the holes and having the rimmed (larger) end of the mouthpiece pointing to the left. NOTE: To ensure accurate results, use a new mouthpiece for each test. A used mouthpiece may be contaminated from the previous tests, which may result in inaccurate readings. Use part number 95-000130 to order more mouthpieces from ACS. 2. Power on the device by pressing either of the two buttons. You will see a flashing amber LED light and a WAIT message as the device goes through the boot up sequence. 3. After 2 to 3 seconds, once the flashing LED light turns green and the device displays a READY message, you can begin a breath test. ABOUTREADY MENU

4. Take a deep breath and blow into

the mouthpiece moderately, at a constant rate and pressure, until you hear a click sound (approximately 5 seconds). You should hear a continuous beep, indicating a proper breath flow. The click indicates the sampling process is complete. NOTE: If the breath sample is insufficient, the LED light flashes red and you will see a message to correct the problem. Refer to the “Troubleshooting” section for details. 5. Stop blowing when you hear the click and the tone stops. 6. The device analyzes the sample. Time taken (a few seconds) depends on the amount of alcohol present in the breath sample. If the breath test is being done outdoors in colder weather, the time to analyze may increase by a few seconds. 7. ALERT J5 analyzes and displays the results. NOTE: The colour of the LED light that accompanies the results is for reference only; results given in numbers should be considered at all times, especially for new and professional drivers that may have a different maximum limit as per local regulations. 0.020 % BAC PRINT MENU

Recalling a Breath Test

You can recall any of the 100 most recent tests to either view on screen or print. To recall a test: 1. Press the right button to access the menu. 2. Scroll down to the Last 100 Tests option, using the right button and select the option using the left button. This displays the details of the most recent breath test. NOTE: You can print a test result from here. Scroll down to the Print option using the right button and select the option by pressing the left button. Refer to the “Printing” section for more information. 3. To view other recent tests, use the right button to select the X/100 (such as 1/100) option and press the left button to scroll to the specific test you wish to view. Last 100 Tests Calibration Date/Time Enter Scroll BACK TO TEST Deactivation BACK TO MENU 3/100 0.049 %BAC MAR 20, 2008 12:49:13 PM Enter Print Scroll Scroll Through Breath Tests 8. You can: • Print the displayed result by pressing the left button. Refer to the “Printing” section for details • Start a new test by pressing the right button

An optional battery operated printer can be purchased to

print the BAC results on site. This rechargeable printer connects to ALERT J5 using a serial communication cable. The printer kit includes: • Rechargeable battery pack • 12V power adaptor • Paper roll • Serial communication cable (RS-232) NOTE: Fully charge the battery pack overnight before using the printer for the first time. The printer cannot print while the battery is charging (refer to the “Inserting/Charging the Printer Battery” section). Serial Printing Follow these steps to print the results of the current breath test using the provided communication cable: 1. Power on the printer by pressing the left power button. If the printer does not turn on, the battery may require charging. 2. Connect the ALERT J5 to the serial jack below the printer, using the serial communication cable. 3. With the breath test result showing on the ALERT J5 display, select the Print option by pressing the left button. ALERT J5 detects the cable, and uses the serial (RS-232) port. Line Feed Button Power Button Serial (RS-232) Port

NOTE: To print a recalled test, refer to

the “Recalling a Breath Test” section for details. 4. You will see Port: RS-232 message displayed to indicate serial printing. 5. Press the right button to start printing. This displays a Wait... message until the printing begins. 6. A Done! message is displayed when the printing is complete. 7. To return to the breath test result, press the left button to select the Back option. Port: RS-232 Back Print J5 Breath Alcohol Tester Serial Number: 000056 Unit of Measurement: %BAC Test Record No: 8 Subject Test August 28, 2007 0.051%BAC 09:57:06 Subject Name Licence Plate Operator Name Test Location Department Name Print Sample

Inserting/Charging the Printer Battery

1. Align the power terminals and push the battery pack in the compartment below the printer. 2. Plug the provided AC or DC car charger to the jack at the rear of the case. 3. The left power button on the printer also lights up red, indicating the battery is charging. Allow the battery to charge for 12-16 hours initially. 4. When the battery is fully charged both red lights turn off. Unplug the power to the case. NOTE: To prolong the battery life, remove the battery when the printer is not in use, and let the battery drain fully before recharging.

Loading the Paper Roll

1. Open the printer top cover by pushing up on either side of the printer. 2. Pull the tab to the right and insert the paper roll, so that the paper dispenses under. Release the tab to align the paper roll. 3. Feed the paper into the slot between the roller and ribbon cassette. 4. Turn the printer power on (left button), and press the right line feed button until the paper feeds smoothly through the printer. 5. Close the top cover. NOTE: Use part number 95-000171 to order more paper rolls from ACS. Insert Paper Under Line Feed Button

MAINTENANCE & TROUBLESHOOTING

Troubleshooting Table IF THE ALERT J5 THEN Does not power on Change the batteries Displays Not Activated Press the left button Cannot print serially (using communication cable) • Check that the printer is on • Check the cable connection • Check the cable Is displaying irregular results • Change the mouthpiece. Use part number 95-000130, to order more mouthpieces from ACS • Eating, drinking or smoking 20 minutes prior to taking the test can contaminate results. Wait for at least 20 minutes and repeat the test • Check if a calibration is due Has a clogged or dirty mouthpiece • Change the mouthpiece • Use part number 95-000130 to order more mouthpieces from ACS Is using a printer that is out of paper • Replace the paper roll • Use part number 95-000171 to order more rolls from ACS Is displaying PEAK ERROR • You are trying to calibrate the device with a solution less than 100 BAC • Do not attempt to calibrate yourself Requires service Contact ACS for return procedure