Olympus BX51 Instructions Manual
Text of the manual
A X 9 8 5 5
This instruction manual is for the Olympus System Microscopes Models BX51 and BX52. To ensure
the safety, obtain optimum performance and to familiarize yourself fully with the use of this micro-
scope, we recommend that you study this manual thoroughly before operating the microscope.
Retain this instruction manual in an easily accessible place near the work desk for future reference.
TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE
3-2 Focusing Block ...................................................................................................................................................................................... 11
1 Voltage Indication; 2 Using the Light Intensity Preset Switch; 3 Using the Filters
21-23
24-25
26
27-29
Correct assembly and adjustments are critical for the microscope to exhibit its full performance. If you are going to assemble
the microscope yourself, please read section 7, “ASSEMBLY” (pages 27 to 29) carefully.
IMPORTANT — Be sure to read this section for safe use of the equipment. —
1 NOMENCLATURE
2
3 USING THE CONTROLS
3-1 Base................................................................................................................................................................................................................. 8-10
3-3 Stage............................................................................................................................................................................................................ 12-14
3-4 Observation Tube...................................................................................................................................................................... 15-17
3-5 Condenser.......................................................................................................................................................................................... 18-19
1 Centering the Condenser; 2 Compatibility of Objectives and Condensers
3-6 Immersion Objectives................................................................................................................................................................. 20
1 Using Immersion Objectives
3-7 Objectives with Correction Collar ............................................................................................................................. 20
4 TROUBLESHOOTING GUIDE
5 SPECIFICATIONS
6 OPTICAL CHARACTERISTICS
7 ASSEMBLY
PROPER SELECTION OF THE POWER SUPPLY CORD ................................................................. 30-31
1 Replacing the Fine Adjustment Knob;
2 Adjusting the Coarse Adjustment Knob Tension; 3 Pre-focusing Lever
1 Placing the Specimen; 2 Adjusting the X- and Y-Axis Knob Tension;
3 Rotating the Stage; 4 Adjusting the Stage Height
1 Adjusting the Interpupillar Distance; 2 Adjusting the Diopter;
3 Using the Eye Shades; 4 Using Eyepiece Micrometer Disks;
5 Selecting the Light Path; 6 Adjusting the Tilt
1-3
— See this section for the replacement of the light bulb. —
12V100WHAL (PHILIPS 7724)
12V50WHAL-L (LIFE JC)
This microscope employs a UIS (Universal Infinity System) optical design, and should be used only with
UIS eyepieces, objectives and condensers for the BX2 series. (Some of the modules designed for the BX
series are also usable. For details, please consult Olympus or the catalogues.) Less than optimum
performance may result if inappropriate accessories are used.
1. Install the microscope on a sturdy, level table or bench so as not to block
the air vents on the underside of the base.
Do not place the microscope on a flexible surface, as this could result in
blocking the air vents and cause overheating or a fire.
2. The surfaces of the lamp housing on the rear of the microscope will
become extremely hot during operation. When installing the microscope,
make sure to allow ample free space (10 cm or more) around and in
particular above the lamp housing.
3. When installing the microscope, route the power cord away from the
lamp housing. Should the power cord come in contact with the hot lamp
housing, the power cord could melt and cause electric shock.
4. To avoid potential shock hazards and burns when replacing the light
bulb, set the main switch @ to “
” (OFF) then disconnect the power
cord from the wall outlet in advance. Whenever you replace the bulb
during use or right after use, allow the lamp socket ‚ and bulb to cool
before touching. (Fig. 1)
#The microscope also incorporate a fuse (this should be replaced by
the manufacturer or an authorized agent).
5. Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the section
“PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of
this instruction manual. If the proper power cord is not used, product
safety performance cannot be warranted.
6. Always ensure that the grounding terminal of the microscope and that
of the wall outlet are properly connected. If the equipment is not grounded,
Olympus can no longer warrant the electrical safety performance of the
equipment.
7. Never insert metallic objects into the air vents of the microscope frame
as this could result in electrical shock, personal injury and equipment
damage.
@
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to
the user and/or damage to the equipment.
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Warning engraving
position
Lamp housing
(Warning against high temperature)
Fig. 2
1 Getting Ready
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment
in the safest possible manner.
Warnings
Warning engraving is placed at parts where special precaution is required when handling and using the microscope.
Always heed the warnings.
1. A microscope is a precision instrument. Handle it with care and avoid
subjecting it to sudden or severe impact.
2. Do not use the microscope where it is subjected to direct sunlight, high
temperature and humidity, dust or vibrations. (For the operating condi-
tions, refer to section 5, “SPECIFICATIONS”.)
3. When moving the microscope, carefully carry it with the grasping part on
the rear of the arm and the base as shown in Fig. 2 (Weight: approx. 16
kg).
#If a module such as a photographic device is installed, the micro-
scope system will be very heavy. Remove the intermediate attach-
ment before carrying.
#Damage to the microscope will occur if you grasp it by the stage,
coarse/fine adjustment knob or binocular section of the observa-
tion tube.
4. The BX51/52 series can be used with up to two intermediate attach-
ments (e.g. a U-CA magnification changer, U-EPA2 eyepoint adjuster, etc.).
For restrictions when using two intermediate attachments, make sure to
read the instruction manual provided with the respective intermediate
attachments.
2 Maintenance and Storage
1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze slightly
moistened with a mixture of ether (70%) and alcohol (30%).
Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to keep
these chemicals away from open flames or potential sources of electrical sparks –– for example, electrical equip-
ment that is being switched on or off. Also remember to always use these chemicals only in a well-ventilated
room.
2. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean
them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.
4. When not using the microscope, keep it covered with a supplied dust cover.
3 Caution
If the microscope is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the
equipment may also be damaged. Always use the equipment as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).
adjustment scale (Page 15)
Diopter adjustment ring
(Page 15)
Slide holder (Page 12)
Aperture iris diaphragm ring
(Page 19)
X-axis knob (Page 13)
Y-axis knob (Page 13)
Pre-focusing Lever (Page 11)
Filter mount (Page 9)
Field iris diaphragm ring
(Page 18)
Allen screwdriver
Condenser height
adjustment knob (Page 18)
Voltage indicator LED
Main switch (Page 1)
Light intensity preset
switch (Page 8)
Light intensity preset
adjustment screw
(Page 8)
Fine adjustment knob (Page 11)
Coarse adjustment knob (Page 11)
Coarse adjustment tension adjustment ring (Page 11)
Brightness adjustment knob (Page 8)
(Lamp voltage adjustment knob)
(accommodation position)
(Detachable)
} If you have not yet assembled the microscope, read section 7, “ASSEMBLY” (pages 27 to 29).
Transmitted light specification models
BX51TF or BX52TF microscope frame.
Filter knobs (Page 9)
Optional filter knob (Page 9)
Light path selector knob (Page 16)
ND6, ND25 and LBD from the front to the rear
Transmitted light/reflected light specification models
BX51TRF and BX52TRF microscope frames
NOTE
* For the reflected illuminators and reflected lamp housings, please also refer to their instruction manuals.
The controls are identical to those of the transmitted light specification models except for the transmitted/
reflected light switch and the reflected light system (vertical illuminator, reflected lamp housing).
Reflected light illuminator*
· Universal illuminator
(BX-URA2)
· Fluorescence illuminator
(BX-RFA)
Reflected lamp housing*
· 100 W mercury burner housing
(U-LH100HG)
· 100 W mercury apo-burner housing
(U-LH100HGAPO)
· 75 W xenon lamp housing
(U-LH75XEAPO)
· 100 W halogen lamp housing
(U-LH100-3)
· System lamp housing
(U-ULH, halogen bulb specification)
(for exclusive use with
the halogen light source)
TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE
Set the main switch to “ I ” (ON) and adjust the
brightness.
@Main switch
²Brightness adjustment knob
Set the transmitted light
(BX51TRF/BX52TRF only)
³Transmitted/reflected light
switch
Engage only the LBD filter.
|LBD filter knob
Select the light path (trinocular tube).
ƒLight path selector knob
Place the specimen on the stage.
…Slide holder
†X-/Y-axis knobs
Engage the 10X objective in the light path.
‡Revolving nosepiece
Bring the specimen in focus.
ŠCoarse/fine adjustment knobs
Adjust the interpupillary distance.
Adjust the diopter.
Adjust the light axis.
‰Binocular tube
‹Diopter adjustment ring
ŒCondenser height adjustment
knob
™Condenser centering knob
Adjust the aperture iris and field iris diaphragms.
šAperture iris diaphragm ring
›Field iris diaphragm ring
Engage the objective to be used in the light path
and bring the specimen in focus.
‡Revolving nosepiece
Adjust the brightness.
²Brightness adjustment knob
Engage the required filters.
œFilters
(P. 5)
(P. 9)
(P. 8)
} Copy the observation procedure pages on separate sheets and post it near your microscope.
œ
›
Œ
Š
USING THE CONTROLS
3-1 Base
Fig. 3
@
Fig. 4
@
²
²
1 Voltage Indication
(Fig. 3)
1. Turn the brightness adjustment knob @ clockwise to increase the volt-
age and make illumination brighter.
2. The numerals to the right of the lamp voltage indicator LEDs ² indicate
the voltage.
2 Using the Light Intensity Preset Switch
(Fig. 4)
}The light intensity preset switch @ makes it possible to limit the light
intensity to a preselected level regardless of the position of the bright-
ness adjustment knob.
The light intensity preset has been preset to optimum brightness for
photography when the built-in LBD filter is engaged (approx. 9 V with
marking).
1. Press the light intensity preset switch @ to the ON position. (The face of
the switch lights when it is ON.)
2. Using a small flat-blade screwdriver, turn the preset adjustment screw ²
to obtain the required light intensity. Turning the screw clockwise increases
brightness.
3. When the light intensity preset switch is set to OFF, the brightness returns
to the level set by the brightness adjustment knob.
#While the light intensity preset switch is ON, turning the light bright-
ness adjustment knob does not affect brightness.
3 Using the Filters
(Figs. 5 - 10)
}You can place a filter in the light path with one of the following methods.
· Push in the knob of the desired built-in filter to engage it in the light path.
(Page 9)
· Place a filter on the filter mount on the base and engage in the light path.
(Page 9)
· Insert a filter in the U-FC filter cassette, attach it on the filter mount and
engage it in the light path by sliding the filter lever. (Page 9)
Using Built-In Filters (Fig. 5)
Fig. 5
@ ² ³ |
Each of filter knobs @ to | can be engaged into the light path by press-
ing the knob in. Pressing the button again disengages the filter from the
light path.
Filter Type/Purpose
@
ND6 (Neutral Density filter for light adjustment, transmittance 6%)
ND25 (Neutral Density filter for light adjustment, transmittance 25%)
LBD (for color balancing, daylight filter)
OP (optional) filter set*
* The set can be ordered at your local Olympus representative.
²
³
|
Mounting a Single Filter
(Fig. 6)
A filter with a diameter of 45 mm ƒ can be placed on the filter mount on
the base. If you need to place multiple filters, please purchase the U-FC
filter cassette.
# Even when a filter cassette is used, a filter with thickness of no more
than 3 mm can still be placed on the filter mount.
Using the Filter Cassette
(Figs. 7 - 10)
Loading Filters into Filter Cassette
}The filter cassette accommodates filters with a diameter of 45 mm and
thickness of 2.7 mm or less.
}The filter cassette has two filter levels on the right side and one on the left
side.
1. Move all filter levers to the OUT position except for the one belonging to
the slot into which the filter is to be inserted.
2. Slide the lever … to the IN position. Make sure that it clicks securely into
place. (Fig. 7)
3. Holding the lever in the position shown in Fig. 7, put the filter into the
cassette by inserting it in the direction indicated by the arrow.
4. Insert the other two filters in the same manner.
Neutral density filter
45G-530, 45G-533, 45IF550
Green
45Y-48
Yellow
45O-560
Orange
Fig. 8
Fig. 9
Fig. 10
Usable Filters
Applications
B&W contrast filter
‡
†
Š
‰
‹
Mounting the Filter Cassette
1. Fully loosen the filter cassette clamping screw †. (Fig. 8)
2. Align the key ‡ on the bottom surface of the filter cassette with the
positioning slot Š on the filter mount, then snap the filter cassette into
place from above.
3. Rotate the filter cassette to align its sides with the base. (Fig. 9)
4. Align the clamping screw † with the positioning hole ‰ on the filter
mount, then tighten the screw to fasten the filter cassette.
#When the filter cassette is installed, the stage may hit it when low-
ered. Therefore, exercise caution when lowering the stage with the
filter cassette installed.
Using the Filter Cassette
(Fig. 10)
Up to three of the above filters can be loaded in the filter cassette. Mov-
ing the levers ‹ on the left and right sides of the cassette to the IN
position brings the corresponding filter into the light path.
3-2 Focusing Block
Fig. 11
2
(Fig. 12)
Fig. 12
@
@
@
²
³
1 Replacing the Fine Adjustment Knob
(Fig. 11)
Fig. 13
}The fine adjustment knob is designed detachable to prevent interference
with hand during manipulation of the fine adjustment knob or X-/Y-axis
knobs.
Usually attach the knob on the opposite side to the X- and Y-axis knobs.
1. Using the Allen screwdriver, loosen the clamping screw @ and remove
the fine adjustment knob ².
2. Remove the seal from the fine adjustment knob screw hole on the other
side and attach the knob by reversing the removal procedure.
3. Attach a provided seal on the screw hole of the removed fine adjustment
knob ³.
}The fine adjustment knob can be operated with your fingertip or finger
surface at the same time as manipulating the X- or Y-axis knob.
Adjusting the Coarse Adjustment
Knob Tension
#Adjust the coarse adjustment knob tension using the tension ad-
justment ring @.
The coarse adjustment knob tension is preadjusted for easy use. How-
ever, if desired, you can change the tension using the tension adjust-
ment ring @. Turning the ring in the direction of the arrow increases
tension, and vice versa.
The tension is too low if the stage drops by itself of focus is quickly lost
after adjustment with the fine adjustment knob. In this case, turn the ring
in the direction of the arrow to increase tension.
3 Pre-focusing Lever
(Fig. 13)
}The pre-focusing lever ensures that the objective does not come in con-
tact with the specimen and simplifies focusing.
After focusing on the specimen with the coarse adjustment knob, turn
this lever @ in the direction of the arrow and lock; the upper limit on
coarse adjustment movement is set at the locked position.
After changing a specimen, refocusing is easily accomplished by rotat-
ing the coarse adjustment knob to reach the pre-focused position, then
making fine adjustments with the fine adjustment knob.
}Stage movement with the fine adjustment knob is not locked.
#If you want to lower the stage to the lower limit, release the lock in
advance.
1 Placing the Specimen
#The dimensions of the slide glass should be 26 x 76 mm with thick-
ness of 0.9 to 1.2 mm, and the cover glass should have thickness of
0.17 mm.
#When observing very large specimens, remove the slide holder and
place the specimen directly on the stage.
Microscopy with Double-Slide Holder (Fig. 14)
1. Turn the coarse adjustment knob @ to lower the stage.
2. Open the spring-loaded curved finger ² on the slide holder and place
one or two specimen slides on the stage from the front.
3. After placing the sides as far as they will go, gently release the curved
finger.
Microscopy with Single-Slide Holder (Fig. 15)
The specimen side can easily be placed by sliding it into the slide holder
from the front.
Using an Oil Immersion Objective
Adsorption of immersion oil can cause the specimen to drift. In such
cases, it is recommended to use the optional BH2-SCB-3 specimen clip
³ for oil immersion objectives. (Fig. 16)
Using an Oil Immersion Condenser
It may be effective to use the optional U-SVRO (right-hand knob) or U-
SVLO (left-hand knob) slotted stage which prevents close contact be-
tween the specimen and stage.
2 Adjusting the X- and Y-Axis Knob Tension
(Fig. 17)
Fig. 17
@
²
³
|
1. Hold the X-axis knob @ and slide up the Y-axis knob ² up to expose the
adjustment knobs.
2. Turning the X-axis adjustment knob ³ or Y-axis adjustment knob | clock-
wise (in the direction of the arrow) increases the tension and counter-
clockwise decreases it.
#If the tension is adjusted to tight, a creaking sound may be heard
during stage travel, and the stage stopping accuracy may be imper-
iled.
After long hours of use, the stage guide may be devi-
ated and the stage travel range may be decreased.
However, this is not malfunction and can be corrected
easily as described below.
[Treatment]
Horizontal direction: Hold the specimen holder and move the stage guide
to the left and right so that it hits the stoppers.
Vertical direction:
Hold the upper stage and move it to the front and rear
so that it hits the stoppers.
Stage Feed Knob Rubber Caps (Optional)
}When the X- and Y-axis knobs are fitted with the rubber caps, the knobs
can be adjusted without slipping and fine adjustment is possible by
holding the knobs with a very light force. The knob rubber caps also
reduce fatigue after long hours of operation.
The U-SHGT thick type (thickness 5 mm) and U-SHG thin type (thickness
2 mm) knob rubbers are available.
To attach the knob rubbers:
First fit the larger knob rubber to the Y-axis (upper) knob from below it,
then fit the smaller knob rubber to the X-axis (lower) knob from below it.
Angle of Rotation
Clockwise
Counterclockwise
Right hand knobs
230°
20°
Left hand knobs
20°
230°
Fig. 19
Fig. 20
@
@
²
³
3 Rotating the Stage
(Fig. 18)
1. Slightly loosen the stage clamping screw @.
2. The stage can be rotated both clockwise and counterclockwise by the
stage clamping screw.
#A click may be heard and felt during rotation. However, this is due to
the construction of the substage and does not indicate a malfunc-
tion.
}The angle of rotation varies depending on the positions of the X- and Y-
axis knobs.
4 Adjusting the Stage Height
(Figs. 19 & 20)
}By lowering the position of the substage, the microscope will accommo-
date specimens with maximum height of 35 mm. This is useful when
observing metallurgical specimens and other thick objects.
1 Lower the stage to the lower limit, then remove the stage from the micro-
scope.
2 Using the Allen screwdriver, loosen the substage bracket clamping screw
@ and remove the substage.
3. Turn the coarse adjustment knob and raise the focusing block ³ to
where the stopper screw ² on the arm becomes visible.
4. Using the Allen screwdriver, loosen and remove the upper stopper screw
².
5. Reattach substage bracket and stage.
3-4 Observation Tube
Fig. 21
Fig. 22
Fig. 23
Fig. 24
@
²
1 Adjusting the Interpupillar Distance
(Fig. 21)
While looking through the eyepieces, adjust for binocular vision until the
left and right fields of view coincide completely. The index dot · indicates
the interpupillary distance.
}Note your interpupillary distance so that it can be quickly duplicated.
2 Adjusting the Diopter
(Figs 22 & 23)
1. Looking through the eyepiece without the diopter adjustment ring, rotate
the coarse and fine adjustment knobs to bring the specimen into focus.
2. Looking through the eyepiece with the diopter adjustment ring, turn only
the diopter adjustment ring @ to focus on the specimen. (Fig. 22)
Using a Finder Eyepiece
1. Looking through the right eyepiece with your right eye, turn the top of the
eyepiece ² until a clearly defined double crossline can be seen in the
field of view. (Fig. 22 & 23)
2. Looking through the right eyepiece, rotate the coarse and fine adjust-
ment knobs to bring the specimen and double crossline into simulta-
neous focus.
3. Looking through the left eyepiece with your left eye, turn the diopter ad-
justment ring @ to focus on the specimen.
3 Using the Eye Shades
(Fig. 24)
When Wearing Eyeglasses
Use with the eye shades in the normal, folded-down position. This will
prevent the eyeglasses from being scratched.
When Not Wearing Eyeglasses
Extend the folded eye shades in the direction of the arrow to prevent
extraneous light from entering between the eyepieces and eyeglasses.
100% for binocu-
lar eyepieces
Observation of
dark specimens
Middle position
20% for binocular
eyepieces, 80%
for TV/photogra-
phy
Observation of
bright specimens,
photography,
TV observation
Pulled out
100% for TV/
photography
Photography,
TV observation
@
@
²
4 Using Eyepiece Micrometer Disks
(Fig. 25)
Eyepiece micrometer disks can be inserted into WH10X-H (or WH10X)
eyepieces.
Use 24 mm diam. x 1.5 mm thick micrometer disks.
Following Fig. 25, unscrew the micrometer mounting frame ² from the
eyepiece and place a micrometer disk @ into the mounting frame. The
engraving on the micrometer disk should face downward into the mi-
crometer mounting frame.
Screw the micrometer mounting frame back into the eyepiece sleeve.
Selecting the Light Path
(with the U-TR30-2 or U-SWTR-2)
Slide the light path selector knob @ to select the desired light path.
6 Adjusting the Tilt (with the U-TBI3)
(Fig. 27)
}Adjust the height and tilt of the observation tube to obtain the most com-
fortable viewing position.
Holding the binocular section with both hands, raise or lower it to the
desired position.
#Never attempt to force the binocular section past the upper or lower
stop position. Applying excessive force could destroy the limiting
mechanism.
}The U-TBI3 can be used in combination with only one intermediate at-
tachment.
}If you need photography using the U-TBI3, use the U-TRU trinocular tube.
With the U-ETBI
(Fig. 28)
The U-ETBI is an ergonomic observation tube with normal field capable
of tilting adjustment with an erect image and the eyepiece position can
be adjusted by 45 mm toward the front and rear.
1 Centering the Condenser
(Figs. 29 & 30)
1. Turn the condenser height adjustment knob @ to raise the condenser to
its upper limit.
2. Focus on the specimen using the 10X objective.
#When using the U-SC3 swing-out condenser, move the top lens into
the light path.
3. Rotate the field iris diaphragm ring ² in the direction of the arrow so that
the diaphragm image comes inside the field of view.
4. Manipulate the condenser height adjustment knob @ to focus on the
diaphragm image
5. Turn the two condenser centering screws ³ to move the iris diaphragm
image to the center of the field of view.
6. Gradually open the field iris diaphragm. The condenser is properly cen-
tered if the iris image is centered and inscribed in the field of view.
7. During actual use, open the field diaphragm slightly until its image cir-
cumscribes the field of view.
Effects of Field Iris Diaphragm (Fig. 30)
The field iris diaphragm restricts the diameter of the beam of light enter-
ing the objective and thus excludes extraneous light, improving image
contrast. The diameter of the field iris should e adjusted for objective
power to the extent that it just circumscribes the field of view. (See “Com-
patibility of Objectives and Condensers” on the next page.)
Fig. 30
Applicable to FN 22
2X
Applicable (FN 26.5) by
moving top lens out of
the light path.* (FN 26.5)
4X
Applicable to FN 22
10-60X
Applicable
(FN 26.5)
Applicable
(FN 26.5)
Engage top lens in
light path (FN 26.5)
100X
Applicable
(FN 26.5)
N.A. scale
@
²
Objective pupil
70-80%
30-20%
Aperture Iris Diaphragm (Figs. 31 & 32)
· The aperture iris diaphragm determines the numerical aperture of the
illumination system. Matching the numerical aperture of the illumination
system with that of the objective provides better image resolution and
contrast, and also increases the depth of focus.
· Since the contrast of microscope specimens is ordinarily low, setting the
condenser aperture iris diaphragm to between 70% and 80% of the N.A.
of the objective in use is usually recommended. If necessary, adjust the
ratio by removing the eyepiece and looking into the eyepiece sleeve
while adjusting the aperture iris diaphragm ring @ until the image shown
in Fig. 31 is seen.
}Using the numerical aperture scale:
Set the condenser numerical aperture scale to about 80% of the NA
value ² of the respective objective. (Fig. 32)
Example: With the UPlanFI40X (NA 0.75), set the scale to 0.75 x 0.8 = 0.6.
2 Compatibility of Objectives and Condensers
* When using the U-SC3 swing-out condenser together with a 1.25X to 4X objective, fully open the condenser aperture iris
diaphragm and use the field iris diaphragm in the base as aperture diaphragm. With the 1.25X to 2X objectives, the
surroundings of the field of view may become dark.
}To obtain better illumination in photomicrography using a 1.25X to 4X objective, use of the U-ULC-2 is recommended.
3-6 Immersion Objectives
Fig. 33
3-7 Objectives with Correction Collar
Fig. 34
@
1 Using Immersion Objectives
(Fig. 33)
1. Focus on the specimen with a lower-power objective.
2. Place a drop of provided immersion oil onto the specimen at the area to
be observed.
3. Turn the revolving nosepiece to engage the immersion objective, then
focus using the fine adjustment knob.
#Since air bubbles in the oil will affect the image quality, make sure
that the oil is free of bubbles.
a. To check for bubbles, remove the eyepiece and fully open the field and
aperture iris diaphragms, then look at the exit pupil of the objective inside
the observation tube. (The pupil should appear round and bright.)
b. To remove bubbles, turn the revolving nosepiece to repeatedly defocus
and refocus the oil immersion objective.
}f the condenser engraving shows a numerical aperture (NA) of 1.0 or
higher, the number applies only when oil is applied between the slide
glass and the top surface of the condenser. When oil is not present, the
NA is about 0.9.
4. After use, remove oil from the objective front lens by wiping with gauze
slightly moistened with an ether (70%)/alcohol (30%) mixture.
Caution in use of immersion oil
If immersion oil enters your eyes or contacts with your skin, immediately
take the following treatment.
Eyes: Rinse with fresh water (for 15 minutes or more).
Skin: Rinse with water and soap.
If the appearance of the eyes or skin is altered or pain persists, immedi-
ately see your doctor.
}If the cover glass thickness is not 0.17 mm, the objectives cannot mani-
fest their performances. If a correction collar equipped objective is used
in this case, the difference in thickness can be compensated for by ad-
justing the collar.
Adjustment Procedure
· If the cover glass thickness is known, set the correction collar @ to that
value. (Fig. 34)
· If the cover glass thickness is unknown, adjust the correction collar @
and fine adjustment knob alternately until the positioning with the high-
est resolution is obtained.
#Be careful not to touch the correction collar @ when turning the
revolving nosepiece.
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur,
please review the following list and take remedial action as needed. If you cannot solve the problem after checking the
entire list, please contact your local Olympus representative for assistance.
1. Optical System
28
3
Problem
Cause
Remedy
Page
a) Bulb does not light.
Bulb is burned out.
Replace bulb.
Power cord is unplugged.
Plug power cord into the power outlet.
Transmitted/reflected light switch is set
to
(reflected light). (TRF models only)
Set the switch to
(transmitted light
illumination).
b) Bulb operates, but field of view re-
mains dark.
Aperture and field iris diaphragms are not
opened wide enough.
Adjust them to proper sizes.
Condenser is lowered too much.
Adjust the condenser height position.
Light path selector knob is set to the
position.
Move the knob to the
or posi-
tion.
c) Field of view is obscured or not
evenly illuminated
Light path selector knob is in an inter-
mediate position.
Set the knob according to the obser-
vation method.
Revolving nosepiece is not correctly en-
gaged.
Make sure that the revolving nosepiece
clicks properly into place.
Condenser is not attached properly.
Re-attach it.
Revolving nosepiece is not attached
properly.
Push the side dovetail all the way until
it is stopped.
An objective that falls outside of the
condenser’s illumination range is used.
Use a condenser to match the pur-
pose.
Condenser is not properly centered.
Center the condenser.
Field iris diaphragm is stopped down too
far.
Open the field iris diaphragm until it
circumscribes the field.
Bulb is not mounted correctly.
Push the pins of halogen bulb all the
way until the stop position.
d) Dirt or dust is visible in the field of
view.
Dirt/dust on the eyepiece
Clean thoroughly.
Dirt or the to surface of the condenser
Dirt/dust on the specimen
e) Visibility is poor.
· Image is not poor.
· Contrast is poor.
· Details are indistinct.
· Image glares.
A non-UIS objective is used.
Use only UIS series objectives with this
microscope.
Condenser is lowered too far.
Adjust the condenser height position.
Aperture iris diaphragm is stopped down
too far.
Open aperture iris diaphragm.
Revolving nosepiece is not mounted
properly.
Push the slide dovetail all the way until
it is stopped.
Correction collar on correction collar
equipped objective is not properly ad-
justed.
While focusing, turn the correction
collar to find the best position.
Front lens of objective is dirty.
Clean objective.
Immersion oil is not being used with an
oil immersion objective.
Use immersion oil.
18/19
18
16
16
28
19
18
18
28
18
19
––
20
20
3
29
5
––
––
26
Immersion oil contains bubbles.
20
––
Remove the bubbles.
Recommended immersion oil is not
used.
Use the provided immersion oil.
Dirt/dust on specimen.
Clean it.
Dirt/dust on condenser
Inappropriate object side or cover glass
thickness.
Replace with glass of recommended
thickness.
f ) One side of image is blurred..
Objective is not correctly engaged in light
path.
Make sure that revolving nosepiece
clicks into place correctly.
Revolving nosepiece is not correctly
mounted.
Push slide dovetail all the way until it
is stopped.
Stage is not correctly mounted.
Re-attach it.
Specimen is not correctly mounted on
stage.
Place specimen correctly on top of
stage and secure it with slide holder.
g) Image appears to waver.
Revolving nosepiece is not corrected
mounted.
Push slide dovetail all the way until it
is stopped.
Objective is not correctly engaged in
light path.
Make sure that revolving nosepiece
clicks into place correctly.
Condenser is not properly centered.
Center the condenser.
h) Field of view becomes only slightly
brighter when the voltage is raised.
Condenser is not properly centered.
Center the condenser.
Condenser is lowered too far.
Adjust the condenser height position.
2. Electrical System
a) Bulb intermittently lights and goes
out.
Bulb is nearly burned out..
Replace bulb.
A connector is not properly connected.
Check all connectors.
b) Bulb burns out almost immediately.
Wrong type of bulb is being used.
Use correct bulb type.
c) Brightness does not change when
you turn light intensity adjustment
knob.
Light intensity preset button is set to ON.
Press button to OFF.
d) All voltage indicator LEDs light and
voltage cannot be varied with light
intensity adjustment knob.
Bulb is not installed.
Install bulb.
Bulb is burned out.
Replace bulb.
Lamp socket is not connected.
Connect lamp socket correctly.
e) Lamp voltage cannot be varied
with light intensity adjustment
knob.
Bulb is burned out.
Replace bulb.
3. Coarse/Fine Adjustment
a) Coarse adjustment knob is hard
to turn.
Tension adjustment ring is tightened ex-
cessively.
Loose ring.
You are trying to raise stage without
coarse adjustment knob while pre-focus-
ing lever is kept locked.
Unlock pre-focusing lever
b) Stage drifts down by itself or focus
is lost during observation.
Tension adjustment ring is too loose.
Tighten ring.
20
3
12
––
––
––
12
––
––
18
18
18
28
28
8
28
28
11
28
28
11
11
e) Visibility is poor.
· Image is not poor.
· Contrast is poor.
· Details are indistinct.
· Image glares.
c) Image cannot be focused.
When adjusting stage height, you forgot
to reattach upper stopper screw.
Reattach upper stopper screw.
d) Coarse adjustment will not go all
the way up.
Pre-focusing lever is locked at a too low
height.
Unlock pre-focusing lever.
e) Coarse adjustment will not go all
the way down.
Condenser holder is too low.
Raise condenser holder.
f ) Objective makes contact with
specimen before focus is obtained.
––
11
Specimen is mounted upside down.
Mount specimen correctly.
18
4. Observation Tube
a) Field of view of one eye does not
match that of the other.
Interpupillary distance is incorrect.
Adjust interpupillary distance.
Incorrect diopter adjustment.
Adjust diopter.
Different eyepieces are used on left and
right.
Change on eyepiece to match the
other so that both sides are the same
type.
Your view is not accustomed to micro-
scope observation.
Upon looking into eyepieces, try look-
ing at overall field before concentrat-
ing on specimen range. You may also
find it helpful to look up and into dis-
tance for a moment before looking
back into microscope.
5. Stage
a) Image shifts when you touch stage.
Stage is not properly mounted.
Clamp stage.
b) Specimen stops midway on the X-
axis traverse.
Specimen is not correctly positioned.
Place specimen correctly.
c) Horizontal and/or X- and Y-axis
knobs are too tight or too loose.
Tension of horizontal and/or X- and Y-
axis knobs is too high or too low.
Adjust tension.
d) Stroke has reduced.
Stage guide is deviated.
Use treatment method described on
page 13.
15
15
––
––
14
12
13
13
––
1. Optical system
UIS (Universal Infinity System) optical system
2. Illumination
Built-in transmitted Koehler illumination
12V, 100W long-life halogen bulb (pre-centered) 12V100WHAL-L (PHILIPS 5761) or 12 V, 50 W
long-life halogen bulb (pre-centered) 12V50WHAL-L (LIFE JC)
(Average life time: Approximately 2,000 hr. when used as directed)
Light intensity voltage range: 1.0 to 12.0 V DC (continuous)
Light intensity preset button (voltage adjustment range: 1.0 to 12.0 V DC)
Rated voltage: 100-120/220-240 V
, 1.8/0.8 A, 50/60 Hz
Power consumption: 140 W
3. Focusing
Stage height movement by roller guide (rack & pinion)
Stroke per rotation: 0.1 mm (fine), 17.8 mm (coarse)
Full stroke range: 25 mm
Upper limit stopper
Tension adjustment on coarse focus adjustment knob.
4. Revolving nosepiece
Type
U-5RE-2
U-D6RE
U-D7RE
U-P6RE
5-position revolv-
ing nosepiece
Universal 6-posi-
tion revolving
nosepiece
Universal 7-posi-
tion revolving
nosepiece
Universal 6-posi-
tion centered
revolving nose-
piece
Installable mod-
ules
None
Transmitted DIC prism,
transmitted light analyzer
5. Observation tube
Type
U-BI30-2
U-TBI3
U-TR30-2
U-SWTR-2
Widefield binocu-
lar tube
Widefield trinocu-
lar tube
Widefield tilting
binocular
Super widefield
trinocular
Field No.
22
26.5
Tube inclination
30°
5°-35°
continuous
30°
Interpupillary
distance adjust-
ment
50 mm to 76 mm
Light path selec-
tor
None
3 steps: @ BI 100%
² BI 20%, Photo 80%
³ Photo 100%
6. Stage
Type
U-SVRB-4
U-SVLB-4
Common-axis knobs on bottom right.
Common-axis knobs on bottom left.
Rectangular ceramic-coated, wire-driven stage
Size
156 mm (D) x 191 mm (W)
Movement
mechanism
X- and Y-axis knobs with adjustable tension.
Movement range: 52 mm in vertical (Y) direction, 76 mm in horizontal (X)
direction.
Specimen hold-
ers (single slide
holder)
U-HLS4
U-HLST4
U-HRS4
U-HRST4
Specimen hold-
ers (double slide
holder)
U-HLD4
U-HLDT4
U-HRD4
U-HRDT4
0.9 - 0.1
1.40
Aperture iris dia-
phragm
With numerical aperture scale
Objective range
4X (for FN 22 widefield),
10X - 100X
(for FN 26.5 super
widefield)
1.25X (for FN 22
widefield),
2X - 100X (for FN 26.5
super widefield)
10X - 100X (for FN 26.5
super widefield)
8. Operating environment
· Indoor use.
· Altitude: Max. 2000 meters
· Ambient temperature: 5° to 40°C (41° to 104° F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F), decreasing linearly
through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative humidity at 40°C (104°F).
· Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
· Pollution degree: 2 (in accordance with IEC664)
· Installation/Overvoltage category: II (in accordance with IEC664)
The following table shows the optical characteristics of com-
binations of eyepieces and objectives. The figure on the right
shows the performance data engraved on the objectives.
Name
Numerical aperture
NOTE
Magnification
Mechanical tube length
Cover glass thickness
Refer to the latest catalogue or consult your local Olympus
representative for the updated information on the eyepieces
and objectives that can be combined with this microscope.
Indicates the objective is provided with a correction collar.
Optical
character
Magni-
fication
N.A.
W.D.
(mm)
Cover
glass
thick-
ness
Reso-
lution
(µm)
Eyepiece
Remark
Objective
WH10X (FN22)
WH15X (FN14)
Total
mag.
Depth
of focus
(µm)
Field
of view
Total
mag.
Depth
of focus
(µm)
Field
of view
Ach/Ach-P
Achromat/Achromat
for Polarized Light
(FN 22)
· 60X and 100XOI
are only available
with Ach.
Iris
Plan/Plan-P
Plan Achromat/
Plan Achromat for
Polarized Light
(FN 22)
· Plan-P has 4X only.
UPlan-Fl/UPlan Fl-P
Universal Semi-
Apochromat/
Universal Plan
Semi-Apochromat
for Polarized Light
(FN26.5)
· 60X OI and 100X OI
are only available
with UPlan Fl.
UPlan Apo
Universal Plan
Apochromat
(FN26.5)
Plan Apo
Plan Apochromat
(FN26.5)
10X
0.25
6.1
—
1.34
100X
28.0
2.2
150X
20.9
1.4
20X
0.40
3.0
—
0.84
200X
6.09
1.1
300X
4.64
0.7
40X
0.65
0.45
0.17
0.52
400X
3.04
0.55
600X
2.35
0.35
60X
0.80
0.15
0.17
0.42
600X
1.76
0.37
900X
1.39
0.23
100XO
1.25
0.13
—
0.27
1000X
0.69
0.22
1500X
0.55
0.14
100XOI
0.60-1.25
0.13
-
0.27
1000X
0.69
0.22
1500X
0.55
0.14
4X
0.10
22.0
—
3.36
40X
175
5.5
60X
85.8
3.5
10X
0.25
10.5
—
1.34
100X
28.0
2.2
150X
20.9
1.4
10XCY
0.25
10.5
—
1.34
100X
28.0
2.2
150X
20.9
1.4
20X
0.40
1.2
0.17
0.84
200X
6.09
1.1
300X
4.64
0.7
40X
0.65
0.56
0.17
0.52
400X
3.04
0.55
600X
2.35
0.35
50XOI 0.50-0.90
0.20
—
0.37
500X
1.75
0.44
750X
1.30
0.28
100XO
1.25
0.15
—
0.27
1000X
0.69
0.22
1500X
0.55
0.14
4X
0.13
17.0
—
2.58
40X
127
5.5
60X
92.9
3.5
10X
0.30
10.0
—
1.12
100X
22.4
2.2
150X
16.5
1.4
20X
0.50
1.6
0.17
0.67
200X
7.00
1.1
300X
5.22
0.7
40X
0.75
0.51
0.17
0.45
400X
2.52
0.55
600X
1.93
0.35
60XOI
0.65-1.25
0.10
0.17
0.27
600X
0.98
0.37
900X
0.60
0.23
100XO
1.30
0.10
0.17
0.26
1000X
0.66
0.22
1500X
0.52
0.14
100XOI
0.60-1.30
0.10
0.17
0.26
1000X
0.66
0.22
1500X
0.52
0.14
4X
0.16
13.0
—
2.1
40X
99.5
5.5
60X
71.7
3.5
10X
0.40
3.1
0.17
0.84
100X
15.9
2.2
150X
11.5
1.4
20X
0.70
0.65
0.17
0.48
200X
4.65
1.1
300X
3.39
0.7
40X
0.85
0.2
0.11-0.23
0.39
400X
2.14
0.55
600X
1.62
0.35
40XOI
0.5-1.00
0.12
-
0.34
400X
1.70
0.55
600X
1.30
0.35
60X
0.90
0.20
0.11-0.23
0.37
600X
1.50
0.37
900X
1.17
0.23
100XOI
0.5-1.35
0.10
0.17
0.25
1000X
0.62
0.22
1500X
0.49
0.14
1.25X
0.04
5.1
—
8.38
12.5X
872
17.6
18.75X
639
11.2
2X
0.08
6.0
—
4.19
20X
398
11.0
30X
287
7.0
40X
0.95
0.14
0.11-0.23
0.35
400X
1.86
0.55
600X
1.40
0.35
60XO
1.40
0.10
0.17
0.24
600X
0.85
0.37
900X
0.64
0.23
100XO
1.40
0.10
0.17
0.24
1000X
0.59
0.22
1500X
0.47
0.14
UMPlan Fl
No Cover Universal
Plan Semi-Apochromat
(FN26.5)
40X
0.75
0.63
0
0.45
400X
1.66
0.55
600X
1.27
0.35
ND filter
Iris
Iris
Iris
Iris
Iris
Iris
ND filter
UApo
Universal
Apochromat
(FN22)
MPlan Apo
No Cover Plan
Apochromat
(FN26.5)
Plan Fl
Plan Semi-Apochromat
(FN26.5)
UPlan Apo-W
Universal Plan
Apochromat (Water-
Immersion)
(FN26.5)
100XO
1.40
0.10
0
0.24
1000X
0.59
0.22
1500X
0.47
0.14
10XCY
0.30
9.90
—
1.12
100X
22.0
2.2
150X
16.4
1.4
100X
0.95
0.20
0.14-0.20
0.35
1000X
1.0
0.22
1500X
0.83
0.14
20X/340
0.75
0.55
0.17
0.45
200X
4.3
1.1
300X
3.10
1.4
20XW/340
0.70
0.40
0.17
0.48
200X
4.08
1.1
300X
3.37
1.4
40X/340
0.90
0.20
0.11-0.23
0.37
400X
2.0
0.55
600X
1.50
0.35
40XW/340
1.15
0.26
0.13-0.25
0.29
400X
1.29
0.55
600X
1.08
0.35
40XOI/340 0.65-1.35
0.10
0.17
0.25
400X
1.2
0.55
600X
0.88
0.35
Iris
10XW
0.40
0.50
0.17
0.84
100X
14.0
2.2
150X
11.4
1.4
60XW
1.20
0.25
0.15-0.20
0.28
600X
0.90
0.37
900X
0.78
0.23
The diagram below shows the sequence of assembly of the various modules. The numbers indicate the order of
assembly.
The BX51TRF and BX52TRF microscope frames are assembled in the same procedures as the BX51TF and BX52TF
frames.
The module numbers shown in the following diagram are merely the typical examples. For the modules with which the
module numbers are not given, please consult your Olympus representative or the catalogues.
#When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any
parts or touching glass surfaces.
Assembly steps enclosed in will be detailed on the subsequent pages.
}Most assembly operations are possible by using the Allen screwdriver (
) provided with the microscope.
However, the assembly of the reflected light illuminator requires the use of the provided Allen wrench (
) for clamping
the internal screws (to ensure the performance, we recommend that you have your Olympus representative assemble or
remove this module).
ASSEMBLY
7-1 Assembly Diagram
Eyepiece
WH10X (FN 22)
35WH10X (FN 22)
SWH10X-H (FN 26.5)
35SWH10X (FN 26.5)
Observation tube
U-BI30-2 (FN 22)
U-TR30-2 (FN 22)
U-TBI3 (FN 22)
U-SWTR-2 (FN 26.5)
U-ETBI (FN 22)
Intermediate
attachment
U-EPA2
U-DO3
U-TRU, etc.
Tube clamping
screw
Revolving nosepiece
U-5RE-2
U-D6RE
U-D7RE
U-P6RE
Specimen holder
clamping knobs
Specimen holder
U-HLD-4
U-HRD-4
U-HLS-4
U-HRS-4
UIS
series
objective
Stage
U-SVRB-4
U-SVLB-4
U-SRG
Stage clamping
knob
Condenser
U-AC2
U-SC3
U-AAC
U-ULC-2
U-UCD8
Reflected light illuminator
BX-URA2
BX-RFA
Microscope frame
BX51TRF
BX52TRF
Power cord
100 W halogen bulb
12V100WHAL-L
12V50WHAL-L
100 W halogen
lamp housing
U-LH100-3
Lamp socket
clamping screw
Microscope frame
BX51TF
BX52TF
Condenser clamping
knob
Reflected
lamp
housing
Revolving nosepiece
clamping screw
Caution for Bulb Replacement During Use or Right After Use
Fig. 35
Fig. 36
Fig. 38
7-2 Detailed Assembly Procedures
@
²
³
³
@
²
1 Installing the Bulb
(Figs. 35 - 37)
}Use only the designated bulb 12V100WHAL-L (PHILIPS 7724) or
12V50WHAL-L (LIFE JC).
1. Fully loosen the lamp housing clamping screw @ on to of the lamp
housing cover with the provided Allen screwdriver.
2. Lift the lamp housing cover ²upward to remove it.
3. Turn the lamp socket ³ by 90° in the direction indicated by the arrow.
4. Holding the bulb ƒ with gloves or a piece of gauze, depress the bulb
clamping lever | and insert the bulb pins … fully into the pin holes † on
the lamp socket. Gently release the bulb clamping lever to the original
position to secure the bulb.
To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
5. Side the lamp housing cover onto the housing base from the above.
Tighten the clamping screw @ while pressing downward on the cover.
(Fig. 35)
The bulb and the lamp socket are areas near these will be extremely
hot during and right after use.
Set the main switch to “
” (OFF), disconnect the power cord from
the wall outlet, then allow the old bulb and lamp socket to cool
before replacing the bulb with a new of the designated type.
8 Attaching the Condenser
(Fig. 38)
1. Turn the coarse adjustment knob @ to raise the stage to its upper limit.
2. Turn the condenser height adjustment knob ² to lower the condenser
holder to its lowest position.
3. Fully loosen the condenser clamping knob ³.
4. Position the condenser so that the aperture scale is located in front.
Gently slide the condenser along the dovetail all the way until it is stopped.
}When attaching the U-SC3 swing-out condenser, align the positioning
pin at the back of the condenser with the groove in the condenser sleeve.
#When attaching a condenser with swing-out top lens, swing out the
top lens out of the way before inserting the condenser.
5. Tighten the condenser clamping knob and gently raise the condenser
holder to its highest position.
Fig. 37
|
…
ƒ
†
The power cord is vulnerable when bent or twisted. Never subject it
to excessive force.
Make sure that the main switch @ is set to “
” (OFF) before con-
necting the power cord.
Always use the power cord provided by Olympus. IF no power cord
is provided with the microscope, please select the proper power
cord by referring to section “PROPER SELECTION OF THE POWER
SUPPLY CORD” at the end of this instruction manual.
1. Connect the power cord plug ² to the AC receptacle ³.
The power cord should be connected to a grounded, 3-conductor
power outlet. If the power outlet is not grounded properly, Olympus
can no longer warrant the electrical safety performance of the equip-
ment.
2. Plug the power cord plug | into the wall outlet ƒ.
Should the power cord come in contact with the hot lamp housing,
the power cord could melt and cause electric shock. Be sure to
route the power cord away from the lamp housing.
Fig. 38
Fig. 39
Fig. 40
@
²
@
²
³
|
ƒ
9 Attaching the Eyepiece
(Fig. 39)
#When using the U-BI30-2 binocular tube, eyepieces with built-in mi-
crometer disk cannot be attached.
#When using a finder eyepiece or an eyepiece with focusing mirror,
attach it to the right-hand eyepiece sleeve.
When doing so, make sure that the eyepiece positioning pin @ fits
into the notch ² on the eyepiece sleeve.
#The super-widefield trinocular observation tube is equipped with a
positioning notch on both eyepiece sleeves. Make sure that the
positioning pins on both eyepieces fits into the respective notches.
Attaching the Power Cord
(Figs. 40 & 41)
10
PROPER SELECTION OF THE POWER SUPPLY CORD
If no power supply cord is provided, please select the proper power supply cord for the eqipment by referring to “ Specifications ”
and “ Certified Cord ” below:
CAUTION: In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the
electrical safety of the eqipment.
Specifications
Voltage Rating
Current Rating
Temperature Rating
Length
Fittings Configuration
125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area)
6A minimum
60°C minimum
3.05 m maximum
Grounding type attachment pulg cap. Opposite terminates in molded-on IEC con-
figuration appliance coupling.
Table 1 Certified Cord
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with an
agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in
Table 1. In case you are unable to buy locally in your country the power supply cord which is equivalent and authorized
agencies in your country.
Country
Agency
Certification
Mark
Country
Agency
Certification
Mark
Argentina
Australia
Austria
Belgium
Canada
Denmark
Finland
France
Germany
Ireland
IRAM
SAA
ÖVE
CEBEC
CSA
DEMKO
FEI
UTE
VDE
NSAI
Italy
Japan
Netherlands
Norway
Spain
Sweden
Switzerland
United
Kingdom
U.S.A.
IMQ
MITI
KEMA
NEMKO
AEE
SEMKO
SEV
ASTA
BSI
UL
Table 2 HAR Flexible Cord
APPROVAL ORGANIZATIONS AND CORDAGE HARMONIZATION MARKING METHODS
Approval Organization
Printed or embossed Harmoniza-
tion Marking (May be located on
jacket or insulation of internal wir-
ing)
Alternative Marking Utilizing
Black-Red-Yellow Thread (Length
of color section in mm)
Black
Red
Yellow
Comite Electrotechnique Belge
(CEBEC)
Verband Deutscher Elektrotechniker
(VDE) e.V. Prüfstelle
Union Technique de d´Electricite´
(UTE)
Instituto Italiano del Marchio di
Qualita´ (IMQ)
British Approvals Service for Electric
Cables (BASEC)
N.V. KEMA
SEMKO AB Svenska Elektriska
Materielkontorollanstalter
Österreichischer Verband für
Elektrotechnik (ÖVK)
Danmarks Elektriske Materielkontrol
(DEMKO)
National Standards Authority of Ireland
(NSAI)
Norges Elektriske Materiellkontroll
(NEMKO)
Asociacion Electrotecnica Y
Electronica Espanola (AEE)
Hellenic Organization for
Standardization (ELOT)
Instituto Portugues da Qualidade
(IPQ)
Schweizerischer Elektro
Technischer Verein (SEV)
Elektriska Inspektoratet
CEVEC
USE
IEMMEQU
BASEC
KEMA-KEUR
SEMKO
<ÖVE>
NEMKO
ELOT
l np l
SEV
SETI
10
30
10
30
10
10
30
30
10
10
30
50
10
10
30
10
30
30
10
10
50
30
10
50
30
10
30
30
30
50
10
10
70
30
10
70
30
30
70
10
10
90
10
30
90
10
30
90
Underwriters Laboratories Inc. (UL)
SV, SVT, SJ or SJT, 3 X 18AWG
Canadian Standards Association (CSA)
SV, SVT, SJ or SJT, 3 X 18AWG
This device complies with the requirements of both directive 89/336/EEC concerning electromagnetic compatibility
and directive 73/23/EEC concerning low voltage. The CE marking indicates compliance with the above directives.
This publication is printed on recycled paper.
Printed in Japan 2000 03 M 070–@
2-43-2,Hatagaya, Shibuya-ku, Tokyo, Japan
Postfach 10 49 08, 20034, Hamburg, Germany
2 Corporate Center Drive, Melville, NY 11747-3157, U.S.A.
491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373
2-8 Honduras Street, London EC1Y OTX, United Kingdom.
104 Ferntree Gully Road, Oakleigh, Victoria, 3166, Australia