Sartorius Basiclite Series
Sartorius Gemlite Series
Electronic Precision Balances
and Precious Metal Scales
Operating Instructions
98648-008-55
The Basiclite Series and Gemlite Series
from Sartorius offer precision balances/
scales with capacities ranging from
0.1 mg to 12 kg.
These balances/ scales meet the highest
requirements on the accuracy and
reliability of weighing results through the
following features:
Efficient filtering-out of unfavorable
ambient conditions such as vibrations,
drafts, etc.
Stable and reproducible results
Rugged, durable weighing system
Basiclite balances and Gemlite scales
save work and speed up simple routine
applications through:
Ultrafast response times
Easy operation
You can also choose from the following
extra functions for simple applications:
Toggling between weight units
Net-total formulation (tare memory)
Counting
Weighing in percent
Averaging
Contents
Practical Use
.................................. 2
Contents ........................................ 2
Warning and Safety Information ........ 2
Getting Started ................................ 3
Operating the Balance/ Scale
Basic Weighing Function .................... 8
Calibration/ Adjustment ...................... 9
Application Programs
Counting ........................................10
Weighing in Percent ..........................12
Weigh Averaging ............................14
Net-Total Formulation/
Second Tare Memory ........................16
Toggle Between Weighing Units ..........17
Configuring the Balance/ Scale
Setting the Parameters (Menu Codes) ....18
Balance/ Scale Operating Menu
(Overview) ......................................19
Error Codes ....................................20
Care and Maintenance
....................21
Instructions for Recycling ....................22
Overview
Specifications....................................23
Accessories (Options) ........................26
C Marking ......................................27
Warning and Safety Information
Read these operating instructions
thoroughly before using your balance/
scale to prevent damage to the equipment.
Keep these instructions in a safe place.
Follow the instructions below to ensure
safe and trouble-free operation of your
balance/ scale:
!
Make sure that the voltage rating
printed on the AC adapter is identical
to your local line voltage
!
Use only a commercially available
9-volt battery or rechargeable battery
!
Do not use this balance/ scale
in a hazardous area/ location
The only way to turn the power
off completely is to disconnect the
AC adapter
Connect only Sartorius accessories and
options, as these are optimally desi-
gned for use with your balance/ scale.
Protect the AC adapter from contact
with liquid.
Do not open the balance/ scale
housing. If the seal is broken, this will result
in forfeiture of all claims under
the manufacturer's warranty.
Do not miss out on the benefits of our
full warranty. Please contact your local
Sartorius office or dealer for further
information. Complete the warranty
registration card, if available indicating
the date of installation, and return the
card to your Sartorius office or dealer.
Storage and Shipping Conditions
Do not expose the balance/ scale to
shocks, vibration, moisture or extreme
temperatures.
Unpacking the Balance/ Scale
After unpacking the balance/
scale, check it immediately for
any visible damage as a result of
rough handling during shipment.
$
If you see any sign of damage,
proceed as directed in the chapter
entitled Care and Maintenance,
under the section on Safety
Inspection.
Save the box and all parts of the
packaging until you have successfully
installed your balance/ scale. Only
the original packaging provides the
best protection for shipment. Before
packing your balance/ scale, unplug all
connected cables to prevent damage.
Equipment Supplied
The equipment supplied includes the
components listed below:
BL210S, BL120S, BL60S, GC503
Balance/ scale
Weighing pan
Pan support
Shield ring
Shield plate
Dust cover
AC adapter
Gem tray (only with model GC503)
Calibration weight 100 g, F1
(only with model GC503)
BL150S
Balance/ scale
Weighing pan
Pan support
Glass draft shield with cover
AC adapter
Balances/ scales with a readability of
0.01g
Balance/ scale
Weighing pan
Pan support (only on models
with a round weighing pan)
AC adapter
Your balance/ scale is designed to
provide reliable weighing results under
normal ambient conditions. W hen
choosing a location to set up your
balance/ scale, observe the following
so that you will be able to work with
added speed and accuracy:
Set up the balance/ scale on
a stable, even surface
Avoid placing the balance/ scale
in close proximity to a heater or
otherwise exposing the balance/
scale to heat or direct sunlight
Protect the balance/ scale
from drafts that come from open
windows or doors
Avoid exposing the balance/ scale
to extreme vibrations during
weighing
Protect the balance/ scale from
aggressive chemical vapors
Do not expose the balance/
scale to extreme moisture over long
periods
Conditioning the Balance/ Scale:
Moisture in the air can condense
on the surfaces of a cold balance/
scale whenever it is brought into a
substantially warmer place. If you
transfer the balance/ scale to a
warmer area, make sure to condition it
for about 2 hours at room temperature,
leaving it unplugged from AC power.
Information
on Radio Frequency Interference
Warning!
This equipment generates, uses and
can radiate radio frequency energy
and, if not installed and used in
accordance with the instruction
manual, may cause interference to
radio communications. It has been
tested and found to comply with the
limits for a Class A computing device
pursuant to Subpart J of Part 15 of
FCC Rules, which are designed to
provide reasonable protection against
such interference, when operated in a
commercial environment. Operation of
this equipment in a residential area is
likely to cause interference, in which
case the user, at his own expense, will
be required to take whatever measures
may be required to correct the
interference.
Setting up the Balance/ Scale
Balances/ Scales with an Analytical Draft Shield
Chamber:
Place the components listed below inside
the chamber in the order given:
Shield plate
Shield ring
Pan support
Weighing pan
Gem tray (only with model GC503)
Balances/ Scales with a Glass Draft Shield:
Place the components listed below inside
the chamber in the order given:
Draft shield base place it on the balance so that
the edge for fitting the glass draft shield faces
upwards and turn it until it is firmly in place
Pan support
Weighing pan
Glass draft shield
Draft shield cover place it on the balance so
that the edge faces downwards
Balances/ Scales with a Round Weighing Pan
Place the components listed below inside
the chamber in the order given:
Pan support
Weighing pan
Balances/ Scales with a Rectangular Weighing Pan:
Place the weighing pan on the balance/ scale
Connecting the Balance/ Scale to AC Power/
Safety Precautions
Use only original Sartorius AC adapters:
for Europe: 6971948 for the RSA: 6971949
for the U.S.: 6971947 for Australia: 6971950
Insert the right-angle plug into the jack
The AC adapter rated to Class 2 can be plugged
into any wall outlet without requiring any
additional safety precautions
The ground is connected to the balance/ scale
housing, which can be additionally grounded
for operation.
Using a Non-Rechargeable/ Rechargeable Battery
(not for models BL210S, BL120S, BL60S, BL150S,
BL1500S, GC503, GM1502)
$
A non-rechargeable or rechargeable battery is not
included with the equipment supplied
!
Use only a commercially available non-
rechargeable or rechargeable 9-volt battery
!
W hen using a rechargeable battery, always use
an external charger to recharge the battery
Lay the balance/ scale on its side
Open the battery compartment:
Lift the compartment cover
Insert the 9-volt (rechargeable) battery
in the compartment
$
Make sure to connect the positive and negative
poles correctly
!
All used batteries are classified as waste
that requires special handling (not household
waste). Dispose of rechargeable batteries in
accordance with the applicable special waste
disposal regulations.
Close the battery compartment:
Press down on the cover until it clicks into place
Selecting the Line Voltage (Mains Voltage)
Use the following original AC adapters for selecting
the line voltage:
AC adapter TNG8, order no. 6971951
(universal)
or
AC adapter TNG8, order no. 6971952
(for the U.K.)
Use the switch to toggle between 230 V and 115 V
Leveling the Balance/ Scale
(only for models BL210S, BL120S, BL60S, BL150S,
BL1500S, GC503, GM1502)
Level the balance/ scale any time you set
it up in a new location. Use only the 2 front feet of
the balance/ scale for leveling.
Turn the 2 rear feet until they are in position
(only on models BL1500S and GM1502)
Turn the 2 front feet as shown here in the illustration
until the air bubble is centered in the level indicator
>
In most cases, this will require several
adjustment steps
Anti-theft Locking Device
To protect against theft, use the mounting lug on the
rear panel of the balance/ scale.
Secure the balance/ scale at the place of
installation, for example with a chain or a lock
Operating the Balance/ Scale
Basic Weighing Function
Available Features
Taring the balance/ scale
You can tare the balance/ scale within the entire weighing range.
Preparation
Turn on the balance/ scale: Press e
To change configurations: See the chapter entitled
Configuring the Balance/ Scale
To load factory-set configurations:
See Configuring the Balance/ Scale, parameter 9 1
$
To tare the balance/ scale: Press t
Additional Functions
To turn off the balance/ scale: Press e
Example
Basic weighing
Menu code settings: Factory-set codes
Step
Key (or instruction)Display/ Printout
1. Turn on the balance/ scale
e
Self-test is performed
2. Place container on balance/
+ 52.0 g
scale (here: 52 g)
3. Tare the balance/ scale
t
+ 0.0 g
4. Place sample in container on
+150.2 g
balance/ scale (here: 150.2 g)
Always calibrate/ adjust the balance/ scale after setting it up in a new location.
Available Features
Calibration/ adjustment can only be performed when
there is no load on the balance/ scale,
the balance/ scale is tared,
the internal signal is stable.
If these conditions are not met, an error message is displayed.
The weight required for calibration/ adjustment is displayed (see Accessories for
calibration weights; for model GC503 see Equipment Supplied).
On Basiclite balances, calibration/ adjustment can be performed
with any of three weight units: g, kg* * , or lb (menu code 1. 4. x)
Factory setting: Grams (menu code 1. 4. 1)
Example
Calibrate the balance/ scale
Menu code settings: Factory-set codes
Step
Key (or instruction)Display/ Printout
1. Turn on the balance/ scale
e
2. Tare the balance/ scale
t
0.0 g
3. Begin calibration
t
>2 sec.
+1000.0
Calibration weight is displayed
without weight unit (here: 1000 g)
4. Place the indicated calibration
1000.0
weight on the balance/ scale
After calibration, the calibration
+1000.0 g
weight is displayed with wt. unit
5. Remove the calibration weight
0.0 g
*
Calibration technically means to determine the difference between the balance/ scale readout and the
actual weight on the pan to determine the accuracy.
Adjustment means to bring a balance/ scale into the state of accuracy required for its use.
Therefore, calibration, as used in this manual, actually means adjustment.
* * = not on models with a readability of 0.1 mg
W ith the Counting program you can determine
the number of parts that each have approximately
equal weight.
Available Features
Store the current weight value to have it loaded
as the preset reference sample quantity next time
you initialize the Counting application
The reference sample quantity can be changed
in the operating menu:
See Configuring the Balance/ Scale
The average piece weight is automatically
output via the optional data interface port after
initialization, if the menu code for Printout
with data ID codes is set
Press v
to toggle the display between piece
count and weight
Factory Settings
Reference sample quantity: 10 (menu code 3. 3. 2)
Preparation
Configure the Counting application
in the operating menu:
See Configuring the Balance/ Scale
Menu code 2. 1. 4
Counting
Reference sample quantity:
Menu code 3. 3. 1
5 pcs
Menu code 3. 3. 2
10 pcs
Menu code 3. 3. 3
20 pcs
Menu code 3. 3. 4
50 pcs
Menu code 3. 3. 5
100 pcs
See also Configuring the Balance/ Scale
Determine an unknown piece count; weigh the preset reference sample quantity
Settings (changes in the factory settings required for this example):
Menu: Application program: Counting (menu code 2. 1. 4)
Reference sample quantity: 20 pcs (menu code 3. 3. 3)
Step
Key (or instruction)Display/ Printout
11. Turn on the balance/ scale
e
12. Tare the balance/ scale
t
0.0 g
13. Display the reference sample
v
>2 sec.
r ef 20 (briefly)
quantity (here: 20 pcs)
14. Place the reference sample
+
66.0 g
quantity (20 pcs) on the
balance/ scale (here: 66 g)
15. Start the application; if the
v
+
20 pcs
print format is set to include
data ID codes, the following
is printed:
wRef + 3.300 g
16. Weigh uncounted parts
+
174 pcs
(here: 174 pcs)
17. Display weight
v
+ 574.2 g
18. Display quantity
v
+
174 pcs
19. Unload the balance/ scale
0 pcs
10. Delete the reference value
c
11. Repeat the procedure starting
from step 6, if desired.
Weighing in Percent
Purpose
This application program allows you to obtain weight
readouts in percent which are in proportion to a
reference weight.
Available Features
Store the current weight value to have it loaded
as the preset reference percentage next time you
initialize the Weighing in Percent application
The reference percentage can be changed
in the operating menu:
See Configuring the Balance/ Scale
The reference percentage is automatically
output via the optional data interface port after
initialization, if the menu code for Printout
with data ID codes is set
Press v
to toggle the display between
percentage and weight
Factory Settings
Reference percentage: 10 (menu code 3. 3. 2)
Preparation
Configure the Weighing in Percent application
in the operating menu:
See Configuring the Balance/ Scale
Menu code 2. 1. 5
Weighing in percent
Reference percentage:
Menu code 3. 3. 1
5 %
Menu code 3. 3. 2
10 %
Menu code 3. 3. 3
20 %
Menu code 3. 3. 4
50 %
Menu code 3. 3. 5
100 %
See also Configuring the Balance/ Scale
Determine an unknown percentage; store the weight on the balance/ scale
as a reference percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (menu code 2. 1. 5)
Menu: Reference percentage 100 % (menu code 3. 3. 5)
Step
Key (or instruction)Display/ Printout
11. Turn on the balance/ scale
e
12. Tare the balance/ scale
t
0.0 g
13. Display the reference
percentage
v
> 2 sec.
r ef 100
14. Place the reference weight
+222.50 g
for 100% on the balance/ scale
(here: 222.5 g)
15. Start application; if the print
v
+100.00 %
format is set to include data
ID codes,
the following is printed:
Wxx% + 222.500 g
16. Place an unknown weight
+144.94 %
on the balance/ scale
(here: 322.5 g)
17. Display weight
v
+ 322.5 g
18. Display percentage
v
+144.94 %
19. Unload the balance/ scale
0.00 %
10. Delete the reference percentage c
11. Repeat the procedure starting
from step 6, if desired.
Use this program to determine weights under
unstable ambient conditions. In this program, the
balance/ scale calculates the weight as the average
value from a defined number of individual weighing
operations. These weighing operations are also
known as subweighing operations or subweighs.
Available Features
The measured result displayed is the arithmetic
mean shown in the selected weight unit;
a triangle indicates that this is a calculated value
You can set the number of subweighing
operations performed in the operating menu:
See Configuring the Balance/ Scale
Press v
for at least 2 sec. to display
the preset number of subweighing operations
Press v
to toggle the display between
the calculated result and the weight
Factory Settings
Number of subweighs for averaging: 10 (3. 3. 2)
Preparation
Configure the Weigh Averaging application
in the operating menu:
See Configuring the Balance/ Scale
Menu code 2. 1.12
Averaging
Number of subweighs for weigh averaging:
3. 3. 1
5 subweighs
3. 3. 2
10 subweighs
3. 3. 3
20 subweighs
3. 3. 4
50 subweighs
3. 3. 5
100 subweighs
See also Configuring the Balance/ Scale
Determine the weight of a sample in extremely unstable ambient conditions
by calculating the average of 10 subweighing operations.
Settings (changes in the factory settings required for this example):
Menu: Application program: Weigh Averaging (menu code 2. 1.12)
Step
Key (or instruction)Display/ Printout
1. Turn on the balance/ scale
e
2. Tare the balance/ scale
t
0.0 g
3. Display the number
of subweighs (here: 10)
v
>2 sec.
r ef 10 (briefly)
4. Place sample on the balance/
+ 8888
scale (weight readout fluctuates;
here: about 275 g)
5. Start measurement
v
+ 8888
10
9
8
...
1
After 10 subweighs
+ 275.5 g G
If the print format is set
to include data ID codes,
the following is printed:
RES + 275.5 g
6. Unload the balance/ scale
+ 275.5 g G
(stable display)
7. Delete the result
c
8. Repeat the procedure starting
from step 4, if desired.
Net-Total Formulation/ Second Tare Memory
W ith this application program you can weigh in components
for formulation of a mixture.
Preparation
Configure the Net-Total Formulation/ Second Tare Memory application
in the operating menu: See Configuring the Balance/ Scale
Menu code 2.1.3
Example
Step
Key (or instruction)Display/ Printout
1. Turn on the balance/ scale
e
2. Place an empty container on
+
65.0 g
the balance/ scale
3. Tare the balance/ scale
t
4. Add the first component
+ 120,5 g
5. Store the first component weight
v
0.0 gNET
If the print format is set
to include data ID codes,
the following is printed:
N1 + 120.5 g
6. Add the next component
+
70,5 g
7. Store the 2nd component weight v
0.0 gNET
8. Add further components,
As described
if desired
for steps 5 and 6
9. Display total weight
c
+ 191,0 g
and fill to desired final weight
Toggle Between Weight Units
W ith this application program you can toggle the display of a weight value back
and forth between two weight units.
Configure the Toggle Weight Units application in the operating menu:
See Configuring the Balance/ Scale, menu code 2.1. 2
Toggle weight units
(Factory setting on Gemlite models)
Menu code
Unit
Conversion factor
Abbr. on
printout
1. 7. 1
3. 1. 1
Taiwanese taels 4)
0.02666666000
o
(only selectable on Gemlite scales)
1. 7. 2 o
3. 1. 2 o
Grams
1.00000000000
g
1. 7. 3 1)
3. 1. 3 1)
Kilograms
0.00100000000
kg
1. 7. 4 2)
3. 1. 4
Carats
5.00000000000
ct
1. 7. 5
3. 1. 5
Pounds
0.00220462260
lb
1. 7. 6
3. 1. 6
Ounces
0.03527396200
oz
1. 7. 7
3. 1. 7 3)
Troy ounces
0.03215074700
ozt
1. 7. 8
3. 1. 8
Hong Kong taels
0.02671725000
tlh
1. 7. 9
3. 1. 9
Singapore taels
0.02645544638
tls
1. 7. 10
3. 1. 10
Taiwanese taels
0.02666666000
tlt
1. 7. 11
3. 1. 11
Grains
15.43235835000
GN
1. 7. 12
3. 1. 12
Pennyweights
0.64301493100
dwt
1. 7. 13
3. 1. 13
Milligrams
1000.00000000000
mg
1. 7. 14
3. 1. 14
Parts per pound
1.12876677120
/ lb
1. 7. 15
3. 1. 15
Chinese taels
0.02645547175
tlc
1. 7. 16
3. 1. 16
Mommes
0.26670000000
mom
1. 7. 17
3. 1. 17
Austrian carats
5.00000000000
K
1. 7. 18
3. 1. 18
Tola
0.08573333810
tol
1. 7. 19
3. 1. 19
Baht
0.06578947437
bat
1. 7. 20
3. 1. 20
Mesghal
0.21700000000
MS
1. 7. 22
3. 1. 22
lb/ oz Gemlite scales only
0.03527396200
o
= Factory setting
1) = not for models with a readability of 0.2 mg
2) = Factory setting only for model GC503
3) = Factory setting only for GM models
4) = Readability in increments of 1 digit
Function
To toggle the display between the 1st and 2nd weight units:
Press the v
key
Configuring the Balance/ Scale
Setting the Parameters (Menu Codes)
You can configure your Basiclite balance or Gemlite scale to meet individual requi-
rements by selecting from the parameters available in the menu.
Example: Adapt the balance/ scale to unstable ambient conditions:
Menu code 1 1 4
Step
Key (or instruction)Display
1. Turn off the balance/ scale
e
2. Turn the balance/ scale back on; e
while all segments
are displayed:
t
briefly
1.
$
To navigate within a menu
t
2.
level; the last menu option is
repeatedly
...
followed by the first option
9.
1.
3. Select the 2nd menu level
p
1. 1.
4. Select the 3rd menu level
p
1. 1. 2 o
5. In Menu Level 3:
t
Select the desired option
repeatedly
1. 1. 4
6. Confirm new setting;
the o indicates the currently set p
option
for 2 sec.
1. 1. 4 o
$
Select the next menu level
(here: move from the 3rd to
the 1st level)
p
1.
$
Set other menu codes, if desired
p
, t
7. Store parameter settings
t
and exit operating menu
for 2 sec.
or
$
Exit operating menu without
storing changes
e
>
Restart the application
0.0 g
Balance/ Scale Operating Menu (Overview)
Factory setting
User setting
Menu
1. Balance/
1.1. Adapt filter
1.1.1
Minimum vibration
scale functions
1.1.2
Normal vibration
1.1.3
Strong vibration
1.1.4
Extreme vibration
1.3. Stability range
1.3.1
1/ 4 digit
1.3.2
1/ 2 digit
1.3.3
1 digit
1.3.4
2 digits
1.3.5
4 digits
1.4. Weight unit for
1.4.1
Grams
calibration weight*
1.4.2
Kilograms* *
1.4.3
Pounds
1.6. Auto zero
1.6.1
On
1.6.2
Off
1.7. Weight unit 1
See Toggle between Weight Units
2. Application
2.1. Program selection
see the desired application program
programs
description
3. Application
3.1. Weight unit 2
see Toggle between Weight Units
parameters
3.3. Reference qty or %
see the desired application program
description
5. through 7.
applies only if an interface is installed; see the corresponding instructions
8. Extra functions
8.1. Menu
8.1.1
Parameter settings: alterable
8.1.2
Parameter settings: read only
8.2. External switch function
applies only if an interface is installed;
see the corresponding instructions
8.3. Power-on mode
8.3.1
Off/ on (battery shuts off
automatically after 3 min)
8.3.2
Standby/ on
9. Reset menu
9. Factory settings
9. 1
Reset
9. 2
Do not reset
* = only for BL models
* * = not for models with a readability of 0.1 mg
Menu level 1
Menu level 2
Menu level 3
Factory setting
Menu level 4
Error codes are shown on the main display for approx. 2 seconds, after which the program automati-
cally returns to the weighing mode.
Display
Cause
Solution
No segments appear on
No AC power is available
Check the AC power supply
the display
The AC adapter is not
Plug in the AC adapter
plugged in
Battery is dead
Replace the battery
Recharge the battery using
an external charger
h
The load exceeds the
Unload the balance/ scale
balance/ scale capacity
l
The weighing pan is not in place
Place the weighing pan on the
balance/ scale
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
e 01
Display capacity exceeded:
Decrease the weight on
Value to be output cannot be
the balance/ scale
shown on the display
e 02
Calibration parameter
Calibrate only when zero
requirement not met; e.g.:
is displayed
balance/ scale not zeroed
Press t
to tare the
balance/ scale
balance/ scale is loaded
Unload the balance/ scale
e 09
W hen gross value zero, no tare Tare the balance/ scale
e 10
The t
key is blocked when
Press c
to clear the
there is data in the second tare
tare memory and release the
memory (net-total) only 1 tare
tare key
function can be used at a time
e 11
Value input is not allowed for
Press t
second tare memory
e 22
Weight is too light or there is
Increase the reference quantity
no sample on the balance/ scale
or sample amount
e 30
Interface port for printer
Contact your local Sartorius
output is blocked
Service Center
Max. weighing capacity is less
The balance/ scale was
Place the weighing pan on the
than indicated under
turned on without the weighing
balance/ scale and press e
Specifications
pan in place
to turn the balance/ scale back on
The weight readout
The balance/ scale has not
Calibrate/ adjust the balance/
is obviously wrong
been calibrated/ adjusted
scale
The balance/ scale was not
Tare before weighing
tared before weighing
If any other errors occur, please contact your local Sartorius Service Center.
Care and Maintenance
Service
Regular servicing by a Sartorius technician will
extend the service life of your balance/ scale and
ensure its continued weighing accuracy. Sartorius
can offer you service contracts, with your choice of
regular maintenance intervals ranging from1 month to
2 years. The frequency of the maintenance intervals
depends on the user's operating conditions and the
tolerance limits.
Repairs
Repair work must be performed by trained service
technicians. Any attempt by untrained persons
to perform repairs may lead to hazards for the user.
Cleaning
!
Disconnect the balance/ scale from the AC
adapter and unplug any interface cables that are
connected to the balance/ scale
!
Make sure that no liquid enters the
balance/ scale housing
!
Do not use any aggressive cleaning agents
(solvents or similar agents)
Clean the balance/ scale using a piece of cloth
which has been wet with a mild detergent (soap)
After cleaning, wipe down the balance/ scale
with a soft, dry cloth
Removing and Cleaning the Weighing Pan:
Lift up and remove the weighing pan together
with the pan support by gripping them from under
the shield ring. Make sure that you do not
damage the weighing system in doing so.
If there is any indication that safe operation of the
balance/ scale with the AC adapter is no longer
warranted:
Turn off the power and disconnect the equipment
from AC power immediately
>
Lock the equipment in a secure place to ensure
that it cannot be used for the time being
Safe operation of the balance/ scale with the AC
adapter is no longer ensured when:
there is visible damage
to the AC adapter
the AC adapter no longer functions properly
The AC adapter has been stored for a relatively
long period under unfavorable conditions
In this case, notify your nearest Sartorius Service Center
or the International Technical Support Unit based in
Goettingen, Germany. Maintenance and repair work
may only be performed by service technicians who
are authorized by Sartorius and who
have access to the required maintenance manuals
have attended the relevant service training
courses
Instructions for Recycling the Packaging
To ensure adequate protection for safe shipment, your
balance/ scale has been packaged to the extent
necessary using environmentally friendly materials.
After successful installation of the balance/ scale, you
should return this packaging for recycling because it
is a valuable source of secondary raw material.
For information on recycling options, including
recycling of old weighing equipment, contact your
municipal waste disposal center or local recycling
depot.
Basiclite Series and Gemlite Series
Model
BL210S BL120S
BL60S
GC503
GM1205
Weighing capacity
g
210
120
60
505 ct/ 101
1200
Readability
mg
0.1
0.1
0.1
0.001 ct/ 0.2
50
Tare range (subtractive)
g
210
120
60
505 ct/ 101
1200
Repeatability
mg 0.1
0.1
0.1
0.001 ct/ 0.2
50
Linearity
mg 0.2
0.2
0.2
0.002 ct/ 0.4
100
Operating temperature range
+10 +30C
0 +40C
(50F86F)
(32F104F)
Sensitivity drift within
+10 +30C
/ K 2106
2106
2106
2106
2105
Response time (average)
s
3
3
3
3
2
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update (depends on
the filter level selected)
s
0.10.4
External calibration weight
g
200 (E2) 100 (E2) 50 (E2)
100 (F1)
1000 (F2)
(of at least accuracy class )
lb
0.4 (E2) 0.2 (E2) 0.1 (E2)
Net weight, approx.
kg/ lb 3.0/ 6.6 3.0/ 6.6 3.0/ 6.6 2.8/ 6.2
1.4/ 3.1
Pan size
mm
80
80
80
80
174 x 143
inches 3.3
3.3
3.3
3.3
6.9 x 5.6
Weighing chamber height
mm
200
200
200
133
inches 7.9
7.9
7.9
5.2
Dimensions (WxDxH)
mm
200x251x299
200x251x233 189x250x70
inches 7.9x9.9x11.8
7.9x9.9x9.2
7.4x9.8x2.8
AC power source/
power requirements
AC adapter, 230 V or 115 V, +15% 20%
Frequency
4860 Hz
Power requirements,
direct current
V
10 to 20
Power consumption (average)
W
1
1
1
1
0.75
Approx. hours of operation with:
9-volt battery, approx.
h
11
fully charged
rechargeable battery
h
2.5
Hours of operation
with the YRB07Z
rechargeable battery pack
h
20
20
20
20
25
Basiclite Series and Gemlite Series
Model
BL150S
BL1500S, BL610,
BL310,
GM212
BL150,
BL6100,
GM1502 GM612 GM312
GM152
GM6101
Weighing capacity
g
150
1500
610
310
210
150
6100
Readability
g
0.001
0.01
0.01
0.01
0.01
0.01
0.1
Tare range (subtractive) g
150
1500
610
310
210
150
6100
Repeatability
g
0.0015
0.015
0.01
0.01
0.01
0.01
0.1
Linearity
g
0.003
0.03
0.02
0.02
0.02
0.02
0.2
Operating temperature
range
0 +40C (263 313 K, 32F 104F)
Sensitivity drift within
+10 +30C
/ K
3.3106 3.3106 0.5105 1105
1105
2105
0.5105
Response time (average)s
2.5
2.5
2
2
2
1.5
2
Adaptation to ambient
conditions
By selection of 1 of 4 optimized filter levels
Display update (depends on
the filter level selected) s
0.20.8
External calibration
weight
g
100 (F1) 1000 (F1) 500 (F2) 100 (F2)100 (F2) 100 (M1) 5000 (F2)
(of at least
accuracy class )
lb*
0.2 (F1)
2 (F1)
1 (F2)
0.2 (F2) 0.2 (F2) 0.2 (M1) 10 (F2)
Net weight, approx.
kg
/ lb
1.6/ 3.5 1.4/ 3.1
1.1/ 2.4
1.1/ 2.4 1.1/ 2.4
1.1/ 2.4
1.4/ 3.1
Pan size
mm
100
174x143 116
116 116
116
174x143
inches3.9
6.9x5.6
4.6
4.6
4.6
4.6
6.9x5.6
AC power source/
power requirements
AC adapter, 230 V or 115 V, +15% 20%
Frequency
4860 Hz
Power requirements,
direct current
V
10 to 20
Power consumption
(average)
W
0.75
Approx. hours of operation with:
9-volt battery, approx. h
11
11
11
11
11
fully charged
rechargeable battery h
2.5
2.5
2.5
2.5
2.5
Hours of operation
with the YRB07Z
rechargeable battery pack
h
25
* = only for BL models
Basiclite und Gemlite Serie
Model
BL3100, BL1500
BL600,
BL12
BL6
BL3
GM3101
GM601
Weighing capacity
g
3100
1500
610
12000
6000
3000
Readability
g
0.1
0.1
0.1
1
1
1
Tare range (subtractive)
g
3100
1500
610
12
6
3
Repeatability
g
0.1
0.1
0.1
1
1
1
Linearity
g
0.2
0.2
0.2
2
2
2
Operating temperature range
0 +40C (263 313 K, 32F 104F)
Sensitivity drift within
+10 +30C
/ K 1105
2105
5105
2.5105 5105
10 105
Response time (average)
s
2
1.5
1.5
1.5
1.5
1.5
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update (depends on
the filter level selected)
s
0.20.8
External calibration weight
kg
1 (F2)
1 (M1)
0.5 (M1) 5 (M1)
5 (M1)
1 (M1)
(of at least accuracy class )
lb*
2 (F2)
2 (M1)
1 (M1)
10 (M1) 10 (M1) 2 (M1)
Net weight, approx.
kg/ lb 1.4/ 3.1
Pan size
mm
174x143 (6.9x5.6 inches)
AC power source/
power requirements
AC adapter, 230 V or 115 V, +15% 20%
Frequency
4860 Hz
Power requirements,
direct current
V
10 to 20
Power consumption (average)
W
0.75
Approx. hours of operation with:
9-volt battery, approx.
h
11
fully charged
rechargeable battery
h
2.5
Hours of operation
with the YRB07Z
rechargeable battery pack
h
25
* = only for BL models
Interface port, installation kit
for models with a
readability 1mg
YDO01BL
for models BL210S, BL120S,
BL60S, GC503
YDO02BL
Data printer
YDP04
Data printer
YDP03-0CE
with date/ time, statistics
evaluation and transaction
counter functions and LCD
(data interface required)
Paper (5 rolls)
6906937
Remote display
(data interface required)
reflective
YRD12Z
for overhead projectors,
transmissive
YRD13Z
External rechargeable battery packYRB08Z
with external battery charger
(hours of operation: 20 or 40.
depending on balance/ scale model)
SartoWedge data transfer softwareYSW01
for on-line storage of weighing data
Interface cable
7357312
for connecting a PC; 25-pin
Universal remote control switch:
(data interface required)
Foot switch with T-connector
YFS01
Hand switch with T-connector
YHS01
T-connector
YTC01
for connecting 2 peripheral
devices
(data interface required)
Carrying case
for models with a
readability 1mg
YDB01BL
for models BL210S, BL120S,
BL60S
YDB02BL
Product
Order No.
Dust cover, for use during weighing
for models with a rectangular
weighing pan
6960BC02
for models with a round
weighing pan
6960BC03
Attaching the dust cover to models
with a glass draft shield:
Remove adhesive strip from
balance/ scale housing
Place dust cover on balance/ scale
Stick adhesive strip on dust cover
Gem trays
300 ml, weight: 86 g,
stainless steel
6407
1000 ml, wt.: 240 g,
stainless steel
641211
300 ml, wt.: 22 g, aluminum
69641304
110 ml, 90 mm, aluminum
YWP02G
270 ml, wt.: 62 g,
137 mm stainless steel
YWP03G
Calibration weights
for BL210S (200 g; E2)
YCW5228
for BL120S, BL150S,
GC503* (100 g; E2)
YCW5128
for BL60S (50 g; E2)
YCW4528
for the BL150, BL310, GM512,
GM212, GM312 (100 g; F2)
YCW5148
for the BL610, BL600, GM612,
GM601, (500 g; F2)
YCW5548
for the BL1500S, GM1502
(1 kg; F1)
YCW6138
for the BL1500, BL3100, BL3,
GM3101, GM1205 (1 kg; F2) YCW6148
for the BL6100, BL6, BL12,
GM6101 (5 kg; F1)
YCW6538
or alternative
YSS653-00
* = F1 100-g weight supplied
with the GC503 scale
The C marking affixed to the equipment indicates that the equipment
meets the requirements of the following Directive(s):
Council Directive 89/ 336/ EEC Electromagnetic compatibility (EMC)
Applicable European Standards:
Limitation of emissions:
EN 50081-1
Residential, commercial and light industry
EN 50081-2
Industrial environment
Defined immunity to interference:
EN 50082-1
Residential, commercial and light industry
EN 50082-2
Industrial environment
Important Note:
The operator shall be responsible for any modifications to Sartorius equipment
and for any connections of cables or equipment not supplied by Sartorius
and must check and, if necessary, correct these modifications and connections.
On request, Sartorius will provide information on the minimum operating
specifications (in accordance with the Standards listed above for defined
immunity to interference).
Council Directive 73/ 23/ EEC
Electrical equipment designed for use within certain voltage limits
Applicable European Standards:
EN 60950
Safety of information technology equipment including
electrical business equipment
EN 61010
Safety requirements for electrical equipment for measurement,
control and laboratory use
Part 1: General requirements
If you use electrical equipment in installations and under ambient conditions
requiring higher safety standards, you must comply with the provisions
as specified in the applicable regulations for installation in your country.
37070 Goettingen, Germany
p Weender Landstrasse 94108, 37075 Goettingen, Germany
t
(+49/ 551) 308-0, f
(+49/ 551) 308-32 89
Internet: http:// www.sartorius.com
Copyright by Sartorius AG, Goettingen, Germany.
All rights reserved. No part of this publication
may be reprinted or translated in any form or by any means
without the prior written permission of Sartorius AG.
The status of the information, specifications and
illustrations in this manual is indicated by the date
given below. Sartorius AG reserves the right to
make changes to the technology, features,
specifications and design of the equipment
without notice.
Status: July 2000, Sartorius AG, Goettingen, Germany
Printed in Germany on paper that has been bleached without any use of chlorine W 199-A00 KT
Publication No.: W BL6005-e00072