GE Aespire 7100 Technical Reference Manual
Read Online
Read the GE Aespire 7100 Technical Reference Manual online using the built-in PDF viewer.
Text of the manual
Datex-Ohmeda products have unitserial numbers with coded logic which indicates a
...-...
product group code, the year of manufacture, and a sequential unit number for
_
identification. The serial number can be in one of two formats.
·胃F
AAAXllll1
AAAXX ll l1l1AA
The X represents an alpha character
The XX represents a number indicating
indicating the yearthe product was
the yearthe product was manufactured;
manufactured; H ~ 2004, J ~ 2005, etc. 04 ~ 2004, 05 ~ 2005, etc.
I and 0 are not used.
Aesplre, ProTlVA, SmartVent, and Unk-25 are registered trademarks of
Datex-Ohmeda Inc.
•
Other brand names or product names used in this manual are trademarks or registered
_
trademarks oftheir respective holders.
~
08/07
Mll10140
Technical Reference Manual
Covers the following:
Aespire 7100 anesthesia machlne
(with non-color or color dlsplay)
Aespire 100 anesthesla machine
515 ProTIVA anesthesia machine
50ftware Revisions 1.X and 2.X
This document is not to be reproduced in any manner, nor are the contents to be
disclosed to anyone, without the express authorization of the product service
department, Datex-Ohmeda, Ohmeda Drive, PO Box 7550, Madison, Wisconsin,
53707.
@ 2007 Datex-Ohmeda Inc.
The information contained in this Technical Reference manual pertains only ωthose
models of products which are marketed by Datex-Ohmeda as ofthe effective date of
this manual or the latest revision thereof. This Technical Reference manual was
prepared for exclusive use by Datex-Ohmeda service personnel in light of their training
and experience as well as the availability to them of parts, proper tools and test
equipment Consequently, Datex-Ohmeda provides this Technical Reference manual to
its customers purely as a business convenience and for the customer' s general
information only without warranty of the results with respectto any application of such
information. Furthermore, because ofthe wide variety of circumstances underwhich
maintenance and repair activities may be performed and the unique nature of each
individual's own experience, capacity, and qualifications, the fact that customer has
received such information from Datex-Ohmeda does not imply in anyway that Datex-
Ohmeda deems said individual to be qualified to perform any such maintenance or
repair service. Moreover, it should not be assumed that every acceptable test and
safety procedure or method, precaution, tool, equipmentor device is referred to within,
or that abnormal or unusual circumstances, may not warrant or suggest different or
additional procedures or requirements.
This manual is su国ectωperiodic review, update and revision. Customers are
cautioned to obtain and consult the latest revision before undertaking any service of
the equipment. Comments and suggestions on this manual are invited from our
customers. Send your comments and suggestions to the Manager ofTechnical
Communications, Datex-Ohmeda, Ohmeda Drive, PO Box 7550, Madison, Wisconsin
53707.
Lh. CAUTlON
Servicing ofthis product in accordance with this Technical Reference manual
should never be undertaken in the absence of proper tools, test equipment and
the most recent revision to this service manual which is clearly and thoroughly
understood.
Technical Competence
The procedures described in this Technical Reference manual should be performed by
trained and authorized personnel only. Maintenance should only be undertaken by
competent individuals who have a general knowledge of and experience with devices of
this nature. No repairs should ever be undertaken or a忧empted by anyone not having
such qualifications.
Datex-Ohmeda strongly recommends using only genuine replacement parts,
manufactured or sold by Datex-Ohmeda for all repair parts replacements.
Read completely through each step in every procedure before starting the procedure;
any exceptions may result in a failure to properly and safely complete the a忧:empted
procedure.
08;07 Ml1l0140
Table of Contents
Impo此ant..................................................................... .ii
Technical Competence . ...................... ............................... .... .ii
1.1 What this manual includes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Standard service procedures . . . . . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.1 User's Reference Manuals . . . . . .. . . .. .•. . . .. .. .. . . . . " . . .. •. .. . . .•. . .. .1-3
1.2.2 Software versions . . . .. . .. . . .. .. . . .. .. . . . . .. .. .. . .. . . . . .. .. .. . . . .. . .. .1-3
1.2.3 Ventilatortests ....................•...............................• .1-3
1.3 Overview .. . . . . . . . . .. .. . . . . . ... . . .. ... . . .. .. . . .. .. .. . . . . .. . . .. . . . . . ... .. .. .1-4
1.3.1 Aespire 100 ....................................................... .1-4
1.3.2 S/5 ProTIVA ....................................................... .1-4
1.4 Configuration options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.4.1 Standard con自guration .............................................. .1-5
1.4.20p回ons ........................................................... .1-5
1.5 Components .••....•.......•.....•••...........•.......•................• .1-5
1.6 Symbols used in the manual oron the equipment. . ..•. .. .. .... . .. . .•. .. .. .••.. 1-10
iii
2 Theory of Operatlon
Iv
2.1 Theoryoverview ................................... ........................ .2-2
2.2 Gas 何owthrough the anesthesia machine ..................................... .2-2
2.2.1 Overview .......................................................... .2-2
2.2.2 Physical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.2.3 Suction regulators •................................................. .2-7
2.2.4 System switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.2.5 Flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.3 Flow through the breathing system .......................................... 2-11
2.3.1 Overview offlow paths .. .... ...................... ....... ...... ..... 2-11
2.3.2 Manual ventilation.. .. . . .. . . . . . . .. . . . . .. . . . . . . .. . . . . .. . . .. .. .. . . . . .. 2-12
2.3.3 Mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2.3.4 Fresh gas and O2 flush flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
2.4 General Description ...................................................... 2-24
2.5 7100ventilatorfeatures......................... ........ ............. ..... 2-25
2.5.1 Safetyfeatures ............. ....... .................... ....... .... . 2-25
2.67100 ventilator components . . . . . . . . . . . . . . . . • . . . . . . . . . . . . . . • . . . . . . . . . . . . . .. 2-26
2.6.1 Control Module (color display) ....................................... 2-27
2.6.2 Control Module (non-color display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
2.6.3 Monitoring interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2.6.4 Serial interface .... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2.6.5 The Pneumatic Vent Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-30
2.7 Electronic and electrical components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.7.1 The Aespire 7100 ventilatorfunctional blocks . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-31
2.7.2 Power Supply ..................................................... 2-32
2.7.3 Sealed Lead Acid Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32
2.7.4 Control Board ..................................................... 2-33
2.7.5 Monitoring interface .. .. . . . . . .. . .. . . . . .. .. . . . . .. .. . . .. .. .. .. . . .. . . .. 2-37
2.7.6 Serial interface .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
08;07 Ml110140
3 Checkout Procedure
Table of Contents
3.1 Ventilator post-service checkout ............................................. .3-2
3.2lnspect伽e system ........................................................ .3-2
3.3 Pipeline and cylindertests ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.4 Flow control tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.1 With O2 monitoring....................... .................... ....... .3-4
3.4.2 Without O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 3-6
3.5 Pressure relieftests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.6 O2 supply alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.7 Flush FlowTest............................................................ .3-8
3.8 Vaporizer back pressure test ............................................... 3-10
3.9 Low-p陪ssure leaktest .................................................... 3-11
3.9.1 Negative low-pressure leak test ...................................... 3-11
3.9.2 ISO or BSI standard low-pressure leak test ............................. 3-12
3.10 Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 3-14
3.11 Breathingsystemtests................................................... 3-16
3.12 Auxiliary02 t1owmetertests ...............................................3-18
3.13lntegrated Suction Regulatortests ......................................... 3-18
3.14Powerfailuretest ....................................................... 3-19
3.15 Electrical safetytests .................................................... 3-19
4 Self Tests and Servlce Mode
Ml110140 08/07
4.1 Selftests ................................................................. 4-3
4.2 Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3 About Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
4.4 Alarm Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.5 Error Log . . .... . . ... . . . . . .. . . .. .. .. . . ..... . . . . .... . ... . . . . .. .. .. .. .. . . .. .. .4-7
4.6 Language ..... ........ ................................. .............. .... .4-8
4.7 User Settings .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.1 Screen Contrast .........
v
4.10 Diagnostic TestsjTools . . . .. . .. . . . .. . . . . .. . . . . . . .. . . . . .. . . . . . .. . . . ... . . . .. 4-24
4.10.1 Display A/D Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.10.2 Display Discrete 1/0 Signals . . .. . . . . .. . . . . . .. . . . . .. . . . . .. . . . . . . . . . . . 4-27
4.10.3 Display Battery Status .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
4.10.4 Test Panel Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
4.10.5 Valves -Test Tool ................................................. 4-30
4.10.6TestCPU and Memory ............................................. 4-31
4.10. 7Test EEPROM .................................................... 4-32
4.10.8 Test Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
4.10.9Test5VFaiIAlarm.................................................4-34
4. 10. 10Test InspiratoryValve ............................................. 4-35
4.10.11 Test PEEP Valve ................................................. 4-36
4.10.12TestPEEPSafetyValve ...........................................4-37
4.10.13 BreathingSystem LeakTest ....................................... 4-38
4.10.14 Test Pressure Limit Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
4.11 Upgrade Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
5.1 Primary Regulators ......................................................... 5-2
5. 1.1 Testsetup ......................................................... .5-2
5. 1.2 Testing Primary Regu lators …........................................ .5-3
5. 1.3 Adjusting Primary Regulators. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
5.2 Secondary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7
5.2.1 Testing/ Adjusting Secondary Regulators or Balance Regulators . . . . . . . . . . . . . .5-7
5.3 Flowmeter Needle Valve Calibration .......................................... .5-8
5.3.102 Needle Valve Calibration (Minimum Flow). . . . . . . . . . . . . . . . . . . . . . . . . . . .. 5-8
5.3.2 N20 Needle Valve Calibration (Minimum Flow). . . . . . . . . . . . . . . . . . . . . . . . . .. 5-10
5.3.3 Air Needle Valve Calibration (Minimum Flow). . . . . . . . . . . . . . . . . . . . . . . . . . .. 5-14
5.4 Link system calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 5-17
5.5 O2 Flush Regulator ....................................................... 5-22
5.6Ai阳ay pressure gauge .................................................... 5-23
5.6.1 Zero the pressure gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 5-23
5.6.2 Checking the pressure gauge accu阳cy ................................ 5-24
08/07 Mll10140
Table of Contents
6 Installatlon and Malntenance
7 Troubleshootlng
850甜ware Installatlon
Mlll0140 08/07
6.1 Aespire 7100 Ins姐lIation Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • . . . . . .6-2
6.2 Aespire 7100 Planned Maintenance. ... . .. .... . . .. .. . . .. .. .. .. . .•. . .. .. .. . . . . .6-4
6.2.1 Every twelve (12) months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
6.2.2 Every twenty-four (24) months ........................................ .6-5
6.3 Free breathingvalve maintenance . .. .. .. . . .•.. . . . .•. .. . . . . . ., . . • • •• •. . . .. .. . . .6-6
6.4 MOPV pressure reliefvalvetest .............................................. .6-7
6.4.1 Testsetup ......................................................... .6-7
6.4.2 Test procedure ...................................................... 6-7
6.5 Auxiliary 02 四owmetertests . . . .. . . . .•.. .. .. .. . . . . .. . .. . . .. .. . . .. . . . . . . .. .. . . .6-8
6.6lntegrated Suction Regulatortests ........................................... .6-9
7.1 GeneralTroubleshooting. .. . . .. .. . . . . ... . .. .. ... . .. ... . . . . ... . .. .... .. .. . .. . .7-2
7.2 BreathingSystem LeakTest Guide. .. .. .. . . . . . ... . . ..••. . . . . .. . . . . .•. . . . .. ... . .7-4
7.2.1 Breathingsystem leaktest....................... ............... ...... .7-5
7.2.2 BreathingSystem Troubleshooting Flowcharts ............................7-7
7.2.3 Leak lsola甘on Tests ........................•.............•......... 7-12
7.3 Troubleshooting guide ....................................................在28
7.4 Alarm and Error messages ......... . . . .. .. . . .. .. .. . . .. . . . . . . .. .. . ... .. . . . .. 7-30
7.5 Troubleshooting Flowcharts .. .. .... .. .... .. .... . . .. .... .. . . .. . . . . .. . . . .. .., 7-41
7.5.1 Ventilator assessment process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 7-41
7.5.2 No displaytroubleshooting. …………………………………… 7-42
7.5.3Inaccurate volume ventilation troubleshooting. . . .. . . . . . . .. . . . . .. . .. . . .., 7-43
7.5.4 VMB board evaluation .............................................. 7-44
7.5.5 No ventilation troubleshooting. ………………………………… 7-45
7.5.6 High intrinsic PEEP troubleshooting. …………………………….. 7-46
8.1 Before installing new software or replacing the Control Board •...................• .8-2
8. 1.1 KeyjBID label (Key Code) ............................................ .8-3
8.2 After replacingthe Control Board orthe Control Module .......................... .8-4
8.2.1 Calibration and Checkout ............................................ .8-5
8.3So阳are Installation .. . . . . . . . . . . . . . . . . . . • . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-6
vll
9 Repalr Procedures
viil
9.1 Howto bleed gas pressure from the machine . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . .. .. .9-4
9.2 Howto remove the rear panels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .9-4
9.2.1 To remove the rear upper panel . . . .. . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . .. .9-4
9.2.2 To remove the lower access panels.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .9-4
9.3 Howto remove thetabletop .. . . . . . . . . . .. . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . .. .9-5
9.4 Servicing the pipeline inlet manifold components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
9.4.1 Replace pipeline inlet filter ........................................... .9-6
9.4.2 Replace pipeline inletcheckvalve ..................................... .9-6
9.4.3 Replace the inlet manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. .. .9-7
9.5 Service the cylinder supply modules . . . . .. . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . .. .9-8
9.5.1 Tightening procedure for high-pressure tube 币ttings . . . . . . . . . . . . . . . . . . . . . .. .9-8
9.5.2 Replace primary regulator module (complete replacement) .................9-8
9.5.3 Replace cylinder inlet filter ........................................... .9-9
9.5.4 Replace cylinder check valve. . . . . . . . . .. . . . . . . . . . . . . . . . . . .. . . . . .. . . . . .. .9-9
9.5.5 Replace 3时-gas cylinder supply module .. .. . .. .. . .. .. .. . .. .. . .. .. .. .. . 9-10
9.6 Replace system switch assembly . . . . . . . . . . . . . . . . . .…………………….. 9-11
9.7 Service the flowmeter module .............................................. 9-13
9.7.1 Remove front flowmeter panel shield .................................. 9-13
9.7.2 Remove flowtubes for cleaning or replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
9.7.3 Remove complete flowmeter head .. .. .. .. . .. . .. .. .. .. .. . .. . . .. .. .. ... 9-15
9.7.4 Replaceflowmetermodules ......................................... 9-16
9.7.5 Replace flowmeterframe . . . . . . . .……………………………… 9-20
9.7.6 Replace O2 supply switch ........................................... 9-21
9.7.7 Checkout procedure for O2 supply switch .............................. 9-21
9.7.8 Replace secondary regulator manifold or balance regulator manifold ....... 9-22
9.7.9 Replace O2 or N20 needle valves (on machines with N20) . . . . . . . . . . . . . . . .. 9-23
9.7.10 Replace an Air needle valve on all machines or an O2 needle valve
on machines without N20 ........................................... 9-25
9.8 Service vaporizer manifold parts ........ .
08/07 MI110140
9.13 Replacethe bagsupportarm.............................................. 9-37
9.13.1 Servicing the bag support arm ...................................... 9-38
9.13.2 Replace 伽ction pad in lower bag arm assembly . . . . . . . . . . . . . . . . . . . . . . .. 9-39
9.13.3 Replace bag port housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 9-40
9.14 Replace a以iliary O2 flowmeter ............................ . . . . . . . . . . . . . . .. 9-41
9.15 Replace the suction control module ..................................... . .. 9-42
9.15.1 Front panel method ............................................... 9-42
9.15.2 Rear panel method ............................................... 9-43
9.16 Replace ABS breathing system components .............................. . .. 9-44
9.16.1 Replace Bag,lVentswitch assembly.................................. 9-44
9.16.2 Replace bellows base latch assembly ......... . .. . .. . . .. .. . . . . .. . . . .. 9-45
9.16.3 EZchange Canister spring replacement.………………………… .9-46
9.17 Replace casters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 9-47
9.18 Replace task light and switch ............................................. 9-48
9.18.1 To replace the task-light switch . . . . .. . . .. . .. . .. .. . . . . . . . . .. . . . . .. .. .. 9-48
9.18.2 To replace the task-light circuit board ................................ 9-48
9.19 Replace the display arm or display cables ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-49
9.19.1 Cabletieinstallation .............................................. 9-49
9.19.2 Removingthedisplayarm.......................................... 9-50
9.19.3 Replacing a display cable ...... . . .. . . . . .. .. . . .. .. . .. . .. . .. . .. . .. .. . 9-50
9.19 .4 lnstalling 由e long arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-51
9.19.5Installing the short arm ............................................ 9-52
9.20 Replace display and cables in ProTIVA machine .............................. 9-53
9.21 Change drive gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-54
9.22 Control module ............. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-55
9.22.1 Colorcontrol module .............................................. 9-56
9.22.2 Original control module . . .. .. .. . . .. .. . . .. .. . . .. .. . . . . .. . .. . . . . .. .. . 9-61
9.23 Servicing the Vent Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 9-67
9.23.1 To remove the Vent En
U
10 lIIustrated Pa此s
x
10.1 Service tools ........................................................... 10-3
10.1.1 Secondary regulator pilot pressure tool ..... . . . . . . . . . . . . . . . . . . . . . . . . .. 10-3
10.1.2 Test Devices ........................... . . . . . . . . . . . . . . . . . . . . . . . . " 10-4
10.1.3 Test Tools ...............................•....................... 10-5
10.2 Extemal components- 骨。ntview... . ... .. . ... . . . . .. . . . . . . .. . . . . ... .. .. ..... 10-6
10.3 Extemal compone阳- front view references ................................. 10-7
10.4 Extemal Components - rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-8
10.5 Control module mounting for a ProTIVA machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-9
10.6 Aespire 100 -exclusive components ...................................... 10-10
10.6.1 AC Inlet(Aespire 100) ............................................10-12
10.6.2 Display mount (Aespire 100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-13
10.7 Front panel, gauges and s:如tem switch .................................... 10-14
10.8 Rear panel components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-15
10.9 Tabletop components.. .. . . ... . . .. . . . . . .. . . . . " .. . . . . . . . . . . . . .. . . . .. .. .. 10-16
10.10 Right-side Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-17
10.11 臼ternal components -Iower assembly ................................... 10-18
10.12ACPowercords................................. ......................10-19
10.13 AC InletjOutlet Components . . . . . . . . . . . . . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . .. 10-20
10.14 Pipeline inlet 自创 ngs................................................... 10-22
10.15 Cylinder Gas Supplies ................................................. 10-23
10.15.1 Cylinder inlet 币四 ngs............................................. 10-24
10.16 Vaporizer manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-25
10.17 Flowmeter components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-26
10.17.1 Flowtube parts ................................................. 10-28
10.17.2 Secondary regulator components ................................. 10-30
10.18 ABS to machine Interface Components .................... . . . . . . . . . . . . . .. 10-32
10.18.1 Flush Regulator, Flush Valve, and ACGO Selector Switch . . . . . . . . . . . . . .. 10-33
10.19 Auxiliary O2 Flowmeter . . . . . . .
ω1/07 Ml110140
10.21 BreathingSystem ............ .................. ....................... 10-36
10.21.1 APL Valve ..................................................... 10-36
10.21.2 Bag/'VentSwitch .......••...................................... 10-37
10.21.3 Absorbercanister.. .. . . .. .. . . . . .. .. .. .... .. . .. .... . ... .. . . . ... .. 10-38
10.21.4 Row Sensor Module. …………………………………….. 10-39
10.21.5 Breathing Circuit Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-40
10.21.6 Exhalation valve . . . ... . .. .. .. . . .. .. .. .. . . .. . . ... . . . .. . . . . . . .. . .. 10-41
10.21. 7 Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-42
10.21.8 Bellow base..... ...... ............ ............•.....••... ..•... 10-43
10.21.9 BagArms ..................................................... 10-44
10.21.10 EZchange Canister system (C02 Bypass). . . . . . . . . . . . . . . . . . . . . . . . . .. 10-45
10.21.11 Condenser .......... ............................ ....... ...... 10-46
10.22 Display cables, serial board, AGSS flowtube, and sample return .............. 10-48
10.23 Vent Engine Housing .................................................. 10-49
10.24 Vent Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-50
10.24.1 Vent Engine - under side . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-51
10.24.2 Manifold plate. …………………………………………. 10-52
10.25 Display A部embly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-53
10.25.1 Display components (Color) .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-54
10.25.2 Display Keyboa刚 and LCD assembly (Color) . . . . . . . . . . . . . . . . . . . . . . . .. 10-55
10.25.3 Display components (non-color). ……………………………. 10-56
10.25.4 Display, front cover assembly (non-color) . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-57
s s10.25.5 Display Keyboard and LCD assembly (non-color) ........... . . . . . .. 10-58
10.26 Legris quick-release 币四ngs............................................. 10-59
10.27 Vent Drive and low-pressure tubing ...................................... 10-60
10.28 Tubing for use with Legris 币创 ngs......................................... 10-62
10.29 Cables and harnesses ................................................. 10-64
10.30 Cables and harnesses (Aespire 100) ........... . . . . . . . . . . . . . . . . . . . . . . . . .. 10-66
10.31 Anesthetic Gas Scavenging System - AGSS ... . . . . . . . . . . . . . . . . . . . . . . . . . . .. 10-68
10.31.1
11 Schematlcs and Dlagrams
Ml110140ω1/07
国
In this section
1his section provides a general overview ofthe Aespire 7100 line of anesthesia
machines.
M1110140
08/07
1.1 What this manual includes . . . . . . . .. . . . .. . .. . . . .. . .. . .. . . . . . .. . . . . . .. .. . . .. . . .1-2
1.2 Standard service procedures . . . . . . . . . . . .. . . . . . . . .. . . . . .. . . . . .. .. .. . . . . . . .. . . .1-3
1.2.1 User's Reference Manuals ...................... .......... .......... .. .1-3
1.2.2 Software versions . . .. .. . . .. . . . . . . .. . . . . .. . . . . ...…..….... ........ .1-3
1.2.3 Ventilatortests ..................................................... .1-3
1.3 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • . . . . . . • .1-4
1.3.1 Aespire 100 ....................................................... .1-4
1.3.2 S/5 ProTlVA ....................................................... .1-4
1.4Configur甜 on options ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • . . . . . . . . . . . . . . . . . . . • .1-5
1.4.1 Standard c。而guration ............................................... 1-5
1.4.20ptions ........................................................... .1-5
1.5 Components ..........................................•.................. .1-5
1.6 Symbols used in the manual oron the equipment .. . . .... . . . . .. . . .. ... . . .. . . ... 1-10
1-1
1.1 What this manual includes
Anesthesla Machlne
刊 is manual covers the service information for 由eA臼pire 7100 line of
anesthesia machines. It covers the following components:
• 7100 display components: color and non-color
• Vent Engine components
• gas delivery components,
• breathing system components,
• frame component (except those strictly associated with a specific
ventilator),
• optional suction regulator and auxiliary O2 flowmeter.
5/5 ProTlVA
The ProTIVA machine is configured with s'恼 ndard Aespire machine
components, with the exception ofthe vaporizer manifold and the Ventilator/
Monitoring Display (7100 Control Module) mounting solution
(referto Section 10.5).
B Braun equipment is not covered in this manual. Refer to B B 阳 un service
documen回tion.
Other equlpment
Other equipment may be attached to the system on the display mount, the
top shelf, or on the side dovetail rails. Consult separate documentation
relative ωthese items for details.
1-2
08/07 M1110140
1.2 Standard service procedures
1.2.1 User's Reference
Some sections ofthis manual referyouωthe User's Reference Manual forthe
Manuals
Aespire machine. To expedite repai邸, you must have, and be familiarwith,由e
User's Reference Manuals forthis product.
Referto the Aespire User's Reference Manual if you need further information
about the operation of the system.
1.2.2 Software verslons
The revision level is displayed on the ventilator sta阳 p menu. This manual
includes test and calibration procedures for Revision 1.X and 2.X software.
1.2.3 Ventllator tests
Service calibration functions let Datex-Ohmeda trained users and
Datex-Ohmeda service personnel perform ventilator setup functions, tests,
calibration and measurements from the front panel display.
Normal operational tests, calibration, and troubleshooting can be performed
on your 7100 ventilatorwithout removing components from the system. Repair
may require removing the ventilator components from the anesthesia
machine.
&. WARNING
Section 4, "SeNice Mode Tests and Calibration" must be performed
whenever you access any internal component of the ventilator to verify
that all critical parts ofthe ventilator are still operational and within
specification.
&. WARNING
Afterthe ventilator has been serviced, you must perform uposιSeNice
Checkout" to verify the entire anesthesia system is properly functioning
beforethes归tem can be retumed to clinical use.
Ml110140ω~
~
The Aespire 7100 machine is a compact, integrated and intuitive anesthesia
delivery system. The ventilator portion provides mechanical ventilation to a
patient during surgery as well as monitoring and displaying various patient
parameters.
刊 e system uses a microprocessor-controlled ventilatorwith internal
monitors, electronic PEEP, Volume Mode, and otheroptional features. A serial
interface permits communication to cardiovascular and respiratory gas
monitoring.
Note
Configurations available for this product depend on local market and
standards requirements. IlIustrations in this manual may not represent all
configurations ofthe product.
The Aespire machine is not suitable for use in an MRI environment.
1.3.1 Aesplre 100
The Aespire 100 machine is based on the standard Aespire 7100 machine
with the following exceptionsωavailable features or options:
• does not include the RS232 Seriallnterface
• does not include the Bi-Ievel LED light strip
• the 2 Vap manifold is standard (does not support 1 Vap manifold)
• not available with AC power outlets (area used by AC Inlet)
• uses 4-inch casters instead of 5-inch casters
1.3.25/5 ProTIVA
The S/5 ProTIVA is a special adaptation ofthe Aespire 7100 machine (does
not include vaporizer manifold) for use with B Braun intravenous drug delivery
components.
1-4
08/07 M1110140
1 I ntroduction
1.4Con由guration options
1.4.1 Standard
conflguratlon
刊e standard configuration includes the following items.
Items marked with an asterisk (叮 are not included in the Aespire 100
machine.
• 7100 Ventilator (with color or non-color display)
• Advanced Breathing System (ABS)
• Auxiliary Common Gas Outlet (ACGO)
• Seriallnterface - RS232*
• Bi-Ievel LED light strip*
• Two large drawe陪
1.4.2 Optlons
Options include the following items.
1.5 Components
Mll10140
08/07
Items marked with an asterisk(叮 are not available in the Aespire 100
machine).
• selected software features
• vaporizer manifold (1 Vap* or 2 Vap)
• pipeline configurations (02/N20, O2/ Air, or 02/N20/ Air)
• gas cylinder configurations (two inboard, one outboard)
• inboard configuration - 02/N20, 02/Air, or02/02
• outboard configuration - N20 only
• manual bag (on SUppo此 arm or on tube)
• EZchange Canister (C02 Bypass) and Condenser
• gas scavenging (active, adjustable, passive, orventuri)
• a suction regulator (pipeline vacuum or venturi vacuum)
• an auxiliary O2 flowmeter
• localized electrical poweroutlets* (isolated ornon-isolated)
• various display mounting solutions
The following figures show the front and rear views ofthe machine.
There are some differences between models.
Figure 1-1 • Aespire machine (frontview -Ieftside)
Figure 1-2 • Aespire machine (frontview - rightside)
Figure 1-3 • Aespire machine (rear view)
Figure 1-4 • S/5 ProTIVA with a typical B Braun fluid manager (fm) system
1町5
1. Auxlllary common gas outlet (ACGO) swltch
2. ACGO
3. Insplratory check valve
4. Insplratory flow sensor or 饵owpo此 adapter
5. Canlster (ca巾on dloxlde abω巾ent)
6. Canlster release
7. Explratory flow sensor or flow port adapter
8. Leak test plug
9. Explratory check valve
10. Breathlngsystem release
11. Manual bag port
12. APl(adJustable pre部ure-lImltlng) valve
13. Bag,tVent 棚Itch
14. Bello嗣 assembly
15. 脚,路ure gauge (allWay)
16. Sample gas retum port
17. Scavenglng flow Indlcator
18. R5-232 Serlal port (not avei灿 lein 由eh即ire 1∞ machlne)
Rgure 1-1 • Aespire machine (front view -Ieft side)
1-6
1
2
08/07 M1110140
1. AB5 (Advanced Breathlng 5ystem)
2. Row contrals
3.
Ve耐 lator Dlsplay/Co睛olMωule (color or non-color)
4. Dovetall ralls
5. Vaporlzer
6.
阴 pellne pressure gauge(s) (upper row)
7. System sw\tch
8.
句Ilnder pressure gauge(s) 刷刷rro呐
9.
02 阴山h
10. AuxlllalJ O2 匍跚meter
11. 5uc制,四 relÞllator
Figure 1.2 • Aespire machine (阳ntview -right side)
M1110140
08/07
4
s
e
y
llntroduction
1-7
AC Inlet for Aesplre 100 machlne
1
2
3
匠。,。啼,
1. Clrcult Breaker for Electrlcal Outlets.
2.
目ectrlcalOutlets.
3.
阴 pellne Connectlon(s)
4. Cyllnder 5uppllω
5.
S归lem Clrcult Breaker (AC Inlat fus笛 forA键plre 1∞ mac刷刷
6. Malns Inlet
7. Equlpotentlal 5tud
事 Items marked w灿 an asteñsk 俨) are not available in the Aespire 100 machine.
Rgure 1-3 • Aespire machine (rearview)
1-8
08/07 Mll10140
1. 刷刷latoνMonltorlng d比例町
2.
句rlnge pumps (3)
3.
fmωn阳11.,
4.
fm 创IIRputer
5.
阳 segment
1
Figure 1-4 • 5/5 Pro TIVA with a typi,臼I B Braun fluid manager(i加I)system
Mll10140
08/07
llntroduction
5
1-9
1.6 Symbols used in the manual or on the equipment
O
。
O
。
O
~
④
-L
l
专
1-10
On (power)
0仔 (power)
Standby
Symbols replace words on the equipment, on the display, or in Datex-Ohmeda
manuals. No one device or manual uses all of the symbols.
Warnings and Cautions tell you about dangerous conditions that can occur if
you do notfollow all instructions in this manual:
• Warnings tell about a condition that can cause injury to the operator or the
patient.
• Cautions tell about a condition that can cause damage to the equipment.
Read and follow all warnings and cautions.
因
Alarm silence
三只
Alarm silence
安
Type B equipment
Standby or preparatory state for part of
囚
Type BF equipment
the equipment
"ON" only for part ofthe equipment
圃
Type CF equipment
"OFF" only for part ofthe equipment
A
Caution, ISO 7000-0434
Direct current
&A
Attention, refer to product instructions,
IEC 60601-1
Alternating current
Dangerous voltage
Protective earth ground
G•
Electrical input
Earth ground
。争
Electrical output
Frame or chassis ground
E
Pneumatic input
Equipotential
曰争
Pneumatic output
08/07 M1110140
Plus, positive polarity
~ Movement in one direction
Minus, negative polarity
等--击'
Movement in 000 directions
4ζ二1
Variability
U
Read top offloat
..c::It3口
Variability in steps
e
Read to center offloat
-:C艾-
Lamp, lighting, illumination
。+
Vacuum inlet
II
Thiswayup
E
Suction bottle outlet
矿ICIÞ
Pipeline
Cylinder
自
Lock
]II[
Isolation transformer
自
Unlock
Jt'r
Linkage system
,.
Close drain
⑧
Risk of Explosion.
,.
Drain (remove condensate)
~~已
L川… l叫
•
⑧
Not autoclavable
因
Mechanical ventilation
1340 C
Autoclavable
~
Bag position/ manual ventilation
'&=cmH20
APL settings are approximate
02+
O2 Flush button
M1110140
08/07
1-11
Circle breathing circuit module
:•
Bain/Mapleson D breathing circuit
module
俨h
The primary regulator is set to pressure
俨
The primary regulaωr is setto pressure
less than 345 kPa (50 psi)
less than 414 kPa (60 psi)
9+
Absorberon
A
CO2 Bypass Option
C02
。毕
雷鸣2
EZchange Canister (C02 bypass)
Absorber off (C02 Bypass active)
00
No battery/ba忧:ery failure
-l
Ba忧ery in use. Bar indicates amount of
ba忧:ery power remaining.
REF
Stock Number
MAX
Maximum
SN
Serial Number
|政 ONLyl
Caution: federallaw prohibits dispensing
without prescription.
~叫
Authorized representa回ve in the European
CE
Systems with this mark agree with the
Community
European Council Directive (93/42/EEC)
E皿Z
for Medical Devices when they are used
as specified in their User's Reference
d
Date of Manufactur
manuals. The 以xx is the certification
number ofthe Noti币ed Body used by
Datex-Ohmeda's Quality Systems.
"
Manufacturer
e
GOST R Russian certification
1-12
08/07 MI110140
2 Theory of Operation
In this section
2.1 Theoryoverview ........................................................... .2-2
M1110140ω:/07
2.2 Gas flowthrough the anesthesia machine ..................................... .2-2
2.2.1 Overview .......................................................... .2-2
2.2.2 Physical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.2.3 Suction regulators .................................................. .2-7
2.2.4 System switch ................... .................. ............... .. .2-8
2.2.5 Flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.3 Flowthrough the breathingsystem .......................................... 2-11
2.3.1 Overviewofflowpaths .............................................. 2-11
2.3.2 Manual ventilation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-12
2.3.3 Mechanical ventilation . . . . . . . . • . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . " 2-15
2.3.4 Fresh gas and O2 flush flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-22
2.4 General Description ...................................................... 2-24
2.5 7100 ventilator features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • . . . " 2-25
2.5.1 Safety features .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-25
2.6 7100ventilatorcomponents ............................................... 2-26
2.6.1 Control Module (color display) ....................................... 2-27
2.6.2 Control Module (non-color display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-28
2.6.3 Monitoring interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-29
2.6.4 Serial interface ................ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-29
2.6.5 The Pneumatic Vent Engine . . . . • . . . . . . . . . . . • • . . . . . . . . . . . . . . . . . . . . . . " 2-30
2.7 Electronic and electrical components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31
2.7.1 刊e Aespire 7100 ventilatorfunctional blocks . . . . . . . . . . . . . . . . . . . . . . . . . .. 2-31
2.7.2 Power Supply ........•............................................ 2-32
2.7.3 Sealed Lead Acid Ba阳 ry. . .. .. .. . . .. . . . . .. . . . . . . .. . . . . . .. . . . . ... . '" 2-32
2.7.4 Control Boa时..................................................... 2-33
2.7.5 Monitoring interface .. . .. . . . .. .. . . .. . . . . .. . . . . . . .. . . . . . . .. . . .. . . . ... 2-37
2.7.6 Serial interface .......... . . . . . . . . . . . . . . . . . . . . .
2-1
2.1 Theory overview
刊 is section describes:
• The f10w of gas through the anesthesia machine.
• The f10w of gas through the breathing system.
• Electrical signals between the anesthesia machine, includingthe breathingsystem,
and the ventilator.
• Ventilator mechanical subsystems
2.2 Gas flow through the anesthesia machine
2-2
2.2.1 Overvlew
Refer to Figu re 2-1.
Gas supplles
Gas comes into the system through a pipeline (2) or cylinder (4) connection. AII
connections have indexed fittings, filters, and check valves (one-way valves). Gauges
show the pipeline (1) and cylinder (3) pre臼ures.
A primary regulator (5) decreases the cylinder pressuresωapproximately pipeline
levels. A pressure relief valve (6) helps protect the system from high pressures.
To help prevent problems with the gas supplies:
• Install yoke plugs on all empty cylinder connections.
• When a pipeline supply is adequate, keep the cylinder valve closed.
O2 flow
Pipeline or regulated cylinder pre臼ure supplies O2 directly to the ventilator (7a for O2
drive gas) and the venturi suction (21a for O2 drive gas) supply connection. An
additional regulator(13) decreases the pressure forthe flush valve (14a) and the
auxiliary flowmeter (25).
The flush valve supplies high flows of O2 to the fresh gas outlet (26 or 27) when you
push the flush button. The flush pressure switch (14b) monitors activation of the flush
valve.
When the system switch (8) is On, O2 flows to the rest ofthe system.
A secondary regulator (10) supplies a constant O2 pressureωthe O2 flow control valve
(11). There is a minimum flow of 25 to 75 mLjmin (for dual-tube flowmeters) or 175 to
225 mL/min (for single-tube flowmeters) through the O2 flowmeter (12).
The O2 pressure switch (9) monitors the O2 supply pressu陀. Ifthe pressure is too low,
an alarm appears on the ventilator display.
08/07 MI110140
Alrand N20 制。w
Note: Original
production machines
include a secondary
regu lator (18) atthe
inputto the Airflow
control valve (19).
Mlxedgas
MI110140
08/07
2 Theory of Qperation
Pipeline or regulated cylinder pressure supplies Air directly to the ventilator (7b for Air
drive gas) and the venturi suction (21b for Air drive gas) supply connection.
When the system switch (8) is On, airflows to the Airflow control valve (19). Because
there is no balance regulator, airflow continues at the set rate during an O2 supply
failure.
A balance regulator (15) controls the N20 supply pressureωthe N20 flow control
valve(16). The O2 secondary regulator pressure at a pilot port controls the output ofthe
balance regulator. The N20 output pressure drops with decreasing O2 supply pre臼ure
and shuts off hypoxic gas flow before the O2 supply pressure reaches zero.
A chain link system (Unk-25) on the N20 and O2 flow controls (16, 11) helps keep the
O2 concentration higherthan 21 % (approximate value) at the common gas outlet
The mixed gas goes from the flowmeter outlet, through the vaporizer manifold and
vaporizer (23) that is On , ωthe ACGO selector switch (曰. A pressure reliefvalve (24)
limits the maximum outlet pressure.
The ACGO selector switch directs the mixed gas to the selected circuit - to the
breathi ng system (26) orωthe ACGO (27).
2-3
Key to Numbered
Components
Note: Current production
machines do not include
a secondary regulator
(18) in the Air flowhead
module.
Key to Symbols
2-4
1. Pipeline pressure gauge
2. Pipeline inlet
3. Cylinder pressure gauge
4. Cylinder inlet (maximum of 3 cylinders)
5. Primary regulator (cylinder pressure)
6.
High-pre臼ure reliefvalve (758 kPa j 110 psi)*
7. Supply connections forthe ventilator
a. O2 drive gas
b. Air drive gas
8. System switch
9. Switch for low O2 supply pressure alarm (used with the ventilator)
10. O2 secondary regulaωr(207 kPa j 30 psi)*
11. O2 flow control valve
12. O2 flow tube(s)
13. O2 flush and auxiliary flowmeter 陀gulator(241 kPa j 35 psi)*
14. O2 Flush
a. Flush valve
b. Pressure switch (used with the ventilator)
15. N20 balance regulator
16. N20 flow control valve
17. N20 flow tube(s)
18. Air secondary regulator (207 kPa j 30 psi)*
19. Airflow control valve
20. Air flow tube(s)
21. Supply connection forVenturi suction
a. O2 drive gas
b. Air drive gas
22. Vaporizer port valve
23. Vaporizer
24. Low-pre臼ure reliefvalve (38 kPa j 5.5 psi)*
25. Auxiliary flowmeter (optional)
26. ToABS
27. ToACGO
28. Test port (primary regulator)
29. Test port (secondaryjbalance regulator)
* Approximate values
t唾)+(
Pneumatic Connection
obAV
Rlter
Direction of Flow
CheckValve
08/07 M1l10140
2 Theory of Operation
26
27
。2
4
IO()1
L旦」
148
日再「 1
l
1
W[D--τL
i)(|
02
4
2
Alr
4
28
2
N20
4
28
nu=ZE
守hdd
‘
-m
-Ul--
-n
主
-nrp
•
2
•
2
-2
啕,.
「|t|-
A. Cyllnder Supplles
B. 阳eumatlc Manlfold
C. 目øwmeter Head
D.Vaporlzer 翩翩lfold
E. ACGO Select SWltch
2-5
Figure 2-1 • Pneumatic circuit
08/07
Mll10140
Figure 2-2 shows the physical path that the gas takes.
connectlons
口问口守
h内田,吨
ABS
Rgure 2-2 • Typical tubing connections -picωrial
2-6
08/07 Ml110140
2 Theory of Operation
2.2.3 Suctlon
Plpellne vacuum
regulators
1he suction regulator (shown in Figure 2-2) uses an external vacuum source.
Venturl Drlve Gas
Alror02
Rgure 2-3 • Venturi suct;on
Mll10140ω1/07
Venturl Drlve vacuum
1he suction regulator (shown in Figure 2-3) uses an internal, venturi derived vacuum
source.
Drive gas (internally plumbed Alr or O2) enters the Venturi Module 阳M) atthe drive po同
(A). As the drive gas passes through the venturi module, a vacuum is created at po同 B.
1he drive gas exits the venturi module at port C and is exhausted outside the machine
through the muffler (D).
1he control port (E) on the venturi module r,臼ponds to pneumatic signals from the front
panel switch on the Suction Control Module (SCM) to turn the venturi vacuum drive gas
on or off. The check valve (CV) helps prevent pressur国tion ofthe suction circuitry ifthe
exhaust is occluded or the venturi unit fails.
E
-
c 圃
E 仁=:lIJ国
Suctlon
- --1
I D
E
t
2-7
1he system switch has two positions: On and Standby.
栅Itch
In the Standby posltlon
1he switch:
• Tums off the ventilator (electrical).
• Stops O2 and Air to the flowhead (pneumatic).
• Without O2 pressure,由e N20 balance regulaωrstops N20.
In the On posltlon
1he switch:
• Turns on the ventilator (electrical).
• Supplies O2 and Air to the flowhead.
• With adequate O2 pressure, the N20 balance regulator supplies N20.
2-8
句stem Swltch
( electrlcal)
tt::国D
(Rear阴阳}
句悦em S1回tch
(pneumatlc)
佣:;07 Ml110140
2 Theory of Operation
2.2.5 Flow control
Needle valves (one for each gas) adjust gas flows. Clockwise rotation decreases flow.
Countercloc阳ise increases flow. Mechanical stops set minimum flows for all gases.
The link system sets the maximum ratio of N20 to O2,
&. WARNING
The Link 25 Proportioning System sets a minimum O2 concentration in the fresh
gas stream when only O2 and N20 are used. Use of an absorber or another gas
can still cause a hypoxic mixture to be delivered to the patient, especially at low
O2 flow rates.
Mlnlmum flows
At minimum flow, two tabs prevent clockwise rotation of the valve stem. One tab is on
the stop collar; the other is on the valve body.
ValveStem
Stop
Collar
ValveBody
Llnk system
The chain link system helps assure an approximate minimum 1 to 3 ratio of flow
Mll10140
08/07
between O2 and N20. When engaged (minimum O2 concentration), a tab on the O2
knob is in contact with a tab on the O2 sprocket so that the O2 and N20 knobs tum
together:
• an increase in N20 flow causes an increase in O2 flow,
• a decrease in O2 flow causes a decrease in N20 flow.
UnkageTabs
O2 恼10b
2-9
Higher concentrations of O2 are possible when the linksystem is not engaged: either by
reducing the N20 flow below the point of engagement or by increasing O2 flow above
the point of engagement
When the N20 flow is below the point of engagement, increasing the N20 flow turns the
O2 sprocket without changing the O2 flow. At the point of engagement, the tab on the
O2 sprocket makes contact with the tab on the O2 knob. Once the linkage is engaged,
turning the N20 flow control counterclockwise (increase in N20 flow) also turns the O2
knob counterclocl例ise (increase in O2 flow) to maintain a nominal25% minimum O2
concentration.
Decreasing the N20 flow from the engagement point rotates the tab on the O2 sprocket
away from the tab on the O2 knob. Increasing the O2 flow rotates the knob tab away
from the sprocket tab. Either action increases the O2 concentration above 21 %.
Sufficiently decreasing O2 flow or increasingthe N20 flow brings the two tabs back into
contact and engages the linkage.
The kick-in point is defined as the N20 flow at which the N20 valve becomes engaged
with the O2 valve flowing at 200 mLjmin. This engagement point is an arbitrary
benchmark that assists in calibrating the proportioning system. The position of the
kick-in is set in the factory. During field calibration, you set the O2 flow to 200 mLjmin
and the N20 flow to the kick-in flow (usually in the range of 400 to 700 mLjmin) and
then install the sprockets with the O2 knobjsprocket engaged.
Maxlmum flows
When the Aespire machine was first produced, all gas flows in Canada required
maximum flow stops. Early production Aespire machines shipped to Canada include a
maximum stop collar on the body ofthe needle valve to set the maximum flow.
Canada no longer requires maximum flow stops forthe gasses that the Aespire
machine is designed to deliver. Current production Aespire machines do not include
the maximum stop collar.
Maxlmum
StopCollar
ValveStem
Valve Body
Mlnlmum Stop Collar
08;07 M1110140
2 Theory of Operation
2.3 Flow through the breathing system
2.3.1 Overvlewof
ßowpaths
Ml110140
佣1/07
1his section looks at four types of flow paths.
• Ventllatlon paths: How gas flows from the drive source (bag or bellows)ωandfrom
the patient.
• Fresh gas paths: Fresh gas can flow from the machine interface directly to the patient
through the inspiratory check valve, orthrough 由e absorber into the expiratory flow,
or directly to an extemal circuit through the optional auxiliary common gas outlet.
• Scavenged gas paths: APL or Pop-off.
• Flow 伽rough 伽e optlonal EZ抽ange Canlster and Condenser: EZchange ON and
EZchange OFF (C02 bypa臼).
2-11
2.3.2 Manual ventllatlon
Manuallnsplratlon
The Bag,lVent switch closes the ventilator path (8).
AP
B
FG
1
2
3
Gas flows from the bag (町, through the absorber (2), into the breathing circuit module,
and through a unidirectional valve (inspiraωry check valve)ωthe patient (3).
During inspiration, fresh gas (FG) flows from the machine into the inspiratory limb,
upstream of the inspiratory check valve.
得礼
~~ayPressure
BagNent switch to Bag
FreshGas
Flow to absorber
Flow from absorber
Inspiratory flow
Figure 2-4 • Gas flow during manual inspiration
2-12
ω:/01 M1110140
2 Theory of Operation
Manual 饵plratlon
The Bag,lVent switch keeps the ventilator path closed (8).
Gas f10ws from 由e patient (4), through a unidirectional valve (expiratory check valve),
and into the bag (5).
During exhalation, fresh gas flows bac阳ards through the absorber (FG) into 由e
expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system, the
retumed gas (SGR) enters the breathing system after the expiratory check valve.
费礼
AP Airway Pressure
B
BagNent switch to Bag
FG
FreshGas
SGR Sample Gas Retum
4
Expiratory flow
5
Flow to bag
Rgure 2-5 • Row during manual expiration
M1110140
08/07
2-13
1he APL valve sets a pre臼ure limit for manual ventilation.
⑦
As you turn the APL knob, it puts more or less force on the APL disc and seat (D/S). If
the circuit pressure is too high (6), the disc and seat inside the diaphragm opens and
vents gas to the scavenging system (7).
D/S APL disc and seat
6
APLflow
7
To scavenging
Rgure 2-6 • Row through the APL Valve
2-14
08/07 M1110140
2 Theory of Opera甘on
2.3.3 Mechanlcal ventllatlon
Mechanlcallnsplratlon
The Bag/Vent switch closes the manual path (V). Pilot pressure (P) closes the
exhalation valve.
Drive gas (0) pushes down on the bellows. Gas flows from the bellows (1), through the
abso巾er(2), and through a unidirectional valve (inspiratory check valve) to the patient
(3).
During inspiration, fresh gas flows into the inspiratory limb, upstream of the inspirato叩
check valve.
AP Airway Pressure
D
Drive gas
FG
Fresh Gas
P
Pi10t pressure
v
BagNent switch to Vent
1
Flow to absorber
2
Flow from absorber
3
Inspiratory ftow
Rgure 2-7 • Mechanical inspiration
M1110140
08;07
2-15
Drive-gas flow stops and the exhalation valve opens. Exhaled gas flows from the
⑤
AP
Airway Pressure
D
Drive gas
FG
FreshGas
patient (4), through a unidirectional valve (expiratory check valve) and into the bellows
(5). Residual drive gas (0) flows out of the bellows to the scavenging system (6).
If PEEP is selected, static pressure on the pilot port ofthe exhalation valve sets the
PEEPlevel.
During exhalation, fresh gas flows backwards through the absorber (FG) into the
expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system, the
returned gas (SGR) enters the breathing system after the expiraωry check valve.
SGR Sample Gas Retum
4
Expiratory flow
5
Flow to bellows
6
To scavenging
Rgure 2-8 • Mechanical expiration
2-16
08/07 Mll10140
( I f
⑨
AP Airway Pressure
D
Drive gas
FG Fresh Gas
P
Pilot pressure
2lheory of Operation
The Bag,'Vent switch closes the manual path (V). Pilot pressure (P) closes the
exhalation valve.
Drive gas (0) pushes down on the bellows. Gas flows from the bellows (白, through the
abso巾er(28), Condenser (2b), and through a unidirectional valve (inspiratory check
valve) to the patient (3).
During inspiration, fr,臼h gas flows into the inspiratory limb, upstream ofthe inspiratory
check valve.
v
BagIVent switch to Vent
1
Flow to absorber
2a
Flow 仕omabsorber
2b
Flow 仕om condenser
3
Inspiratory flow
Rgure 2-9 • Mechanical inspiration 伪rough Condenser with EZchange Canister and Condenser ON
Ml110140
08/07
2-17
AP Airway Pressure
D
Drive gas
FG
FreshGas
Drive-gas f10w sωpsandthe 创lalation valve opens. Exhaled gas f10ws from the
patient(剑, through a unidirectional valve (expiratory check valve) and into the bellows
(5). Residual drive gas (0) flows out of the bellows to the scavenging system (6).
If PEEP is selected, static pressure on the pilot po同时 the exhalation valve sets the
PEEPlevel.
During exhalation, fresh gas flows backwards through the Condenser and absorber
(FG) into the expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system, the
returned gas (SGR) enters 由e breathingsystem afterthe expiraωry check valve.
SGR Sample Gas Retum
~II I
4
Expiratory flow
5
Flow to bellows
6
To scavenging
Figure 2-10 • Mechanical expiration through Condenser with EZchange 臼nisterand Condenser ON
2-18
ω107 M1110140
( I f
⑨
AP
D
FG
P
Airway Pressure
Drive gas
Fresh Gas
Pilot pressure
2 Theory of Operation
The Bag,lVent switch closes the manual path (V). Pilot pressure (P) closes the
exhalation valve.
Drive gas (0) pushes down on the bellows. Gas flows from the b创 lows (1), through 由e
EZchange module bypassing the abso巾er (2), and through a unidirectional valve
(inspiratory check valve) to the patient (3).
During inspiration, fresh gas flows into the inspiratory limb, upstream ofthe inspiratory
check valve.
s
a
理
v
'
BagIVent switch to Vent
F10w to absorber
2
Bypass flow
3
Inspiratory ftow
Rgure 2-11 • Mechanical inspiration with EZchange Canister and Condenser OFF
M1110140 08/07
2-19
Drive-gas flow stops and the exhalation valve opens. Exhaled gas flows from the
patient (4), through a unidirectional valve (expiratory check valve) and into the bellows
(5). Residual drive gas (0) flows out of the bellows to the scavenging system (6).
Aespire 7100
Mechanlcal explraUon
(EZchange and
condenser
。仔}
If PEEP is selected, static pressure on the pilot port ofthe exhalation valve sets the
PEEPlevel.
During exhalation, fresh gas flows backwards thrOUgh the EZchange module (阳) into
the expiratory limb, downstream ofthe expiratory checkvalve.
For machines that are plumbed to return sample gas to the breathing system, the
returned gas (SGR) enters the breathing system after the expiratory check valve.
g
g
!:rì
AP
D
FG
SGR Sample Gas Retum
Expiratory flow
Flow to bellows
VA e
u
ca
gu
阳部二
m
vdgbE
WH@h
mvuvc
‘u
z·n
晦
ADR
4
5
6
To scavenging
08/07 Mll10140
Rgure 2-12 • Mechanical expiration with EZchange Canister and Condenser OFF
2-20
2 Theory of Operation
Pop-o何 valve
1he pop-offvalve limits the pressure inside 由e bellows to 2.5 cm H20 above the drive
gas pressure. 1his normally occurs when the bellows reaches the top of the housing at
the end of exhalation (5).
⑦
5
F10w to bellows
6
Pop-off flow
7
To scavenging
b归路 gas (6) ventsωthe scavenging system (7) through the pop-off valve and the
exhalation valve.
Figure2-13. Rowthroughthepop-offvalve
Mll10140
08/07
2-21
2.3.4 Fresh gas and O2 flush flow
To ABS breathlng
叼stem
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector Switch.
With the ACGO Selector Switch in the ABS position, fresh gas flow is channeled to the
breathingsystem through port3.
The output of the O2 Rush regulator (2) is channeled to the O2 Rush valve.
When activated, O2 flush flow joins the fresh gas flow in the ACGO Selector Switch.
l 翩
t翩翩
Absorber
一 ._._.1~
|2
SEer l
From
OzSupply
InspFlow
PawTrans
Rgure2-14 • 丹田hgas and O2 flush flow (to ABS)
2-22
08/07 Ml110140
Auxlllary Common Gas
Outlet
2 Theory of Operation
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector Switch.
With the ACGO Selector Switch in the ACGO position, fresh gas flow is channeled to the
ACGO outlet.
At the ACGO outlet, a small sample is dive民:ed to the O2 Sensor in the ABS for O2
monitoring.
The outputofthe O2 Aush regulator(2) is channeled to the O2 Aush valve.
When activated, O2 flush flow joins the fresh gas flow in the ACGO Selector Switch.
Absorber
i~
i~
nsar l
From
02 SUPP/Y
f-H Insp Flow
PawTrans
Rgure 2-15 • Fresh gas and O2 flush flow (to ACGO)
Mll10140
ω:;07
2-23
2.4 General Descrlption
、
?……、
;
队刷
卜-:嗣蛐n
"
lkIt
I~b由....,
r-.
---------一,
二‘….'
1田ideMac如lne
刊 e 7100 ventilaωr is a microprocessor based, electronically-controlled,
pneumaticall沪driven ventilaωr with a built-in monitoring system for inspired
oxygen, airway pre臼ure and exhaled volume. The ventilator is an integral
component of the Aespire anesthesia machine.
阳翩翩-国嗣
Note: For non-color displays, the
backlight is powered through the control
board, as shown. For color displays, the
control module includes a separate circuit
board to power the backlight.
:
爱
Figure 2-16 -1100 ventilator functional block diagram
2-24
08/07 Ml110140
2 Theory of Operation
2.5 7100 ventllator features
• Sensors in the breathing circuit are used to control and monitor patient
ventilation and measure inspired oxygen concentration. This lets 由e
ventilator compensate for compression losses, fresh gas contribution, valve
and regulator drift and smallleakages in the breathing absorber, bellows
andsystem.
• Positive End Expiratory Pressure (PEEP) is generated electronically. PEEP is
not active when mechanical ventilation is 0何.
• User settings and microprocessor calculations control breathing pa忧ems.
User interface se创 ngs are kept in non-volatile memory.
• Mechanical ventilation is started with the Bag/Vent switch on the breathing
sys恒m.
• The 7100 ventilator has minimum monitoring and alarms
managed on the ventilator panel (there is no other panel for safety
relevant alarm management, etc.).
• Ventilator hardware is regularly monitored by software tests.
• An RS在32 serial digi恼 I communications port connects to and
communicates with extemal devices.
• An exhalation valve modulates flow in the pressure mode.
• Pressure and volume modes are selectable by the operator.
• AII pneumatic components are located on one manifold.
• Exhausted drive gas and bellows pressure reliefvalve gases are mixed and
go through the ventilator exhalation valve.
• The exhalation valve block is autoclavable.
• Excess fresh gas released from the bellows and ventilator drive gas are
transferred from the exhalation valve to the Anesthesia Gas Scavenging
System (AGSS).
• Optimized for service with a low number of components.
2.5.1 Safety features
• Dual redundant airway overpressure protection, linked to Plimit setting.
• Volume over-delivery limits and protection.
• Proprietary hose connections and fixed manifolds.
• Proven mechanical components used.
• 10 VA electrical power limiting to potential oxygen enriched environment.
• 150 psi burstoverpressure protection.
Mll10140
08/07
2-25
2.6 7100 ventilator components
Major components of the 7100 ventilator are found in different locations of
the anesthesia machine.
The ventilator package consists of:
1. a Control Module (CM) - (color or non-color display) - the CM includes:
• a Control Board (CB)
• an LCD Display
• a Keyboard (wi由 rotary encoder switch)
• a Power Supply
• a backup battery
2. a Ventilator Monitoring Board (VMB).
3. a Seriallsolation Connector Board (SICB).
4. a Vent Engine (VE) with a Vent Engine Board (VEB).
3
1
2
4
Rgure 2-17 • Location of 71 00 ventilator components
2-26
08/07 Ml110140
21heory of Operation
2.6.1 Control Module
The control module with a color display consists oftwo enclosures.
7
{ωlor dlsplay)
The rear enclosure Includ阳
8
'
1. inline fuses
2. a powersupply
3. a cooling fan
4. a backup ba忧:ery
刊e power supply receives AC power from the anesthesia machine. AII the
power necessary to operate the ventilator comes from the power supply.
The front enclosure Includes:
5. a control board (controls operation ofthe ventilator)
6. a backlight inve由r
7. anLCDdisplay
8. a keyboard front panel
9. a rotary encoder
10.a speaker
2
'
10
3
4
5
6
Rgure 2-18 • 7100 ventilatorcontrol module
Ml110140
08/07
2-27
2.6.2 Control Module
The control module with a non-color display consists oftwo enclosures.
(non-color dlsplay)
The rear enclosure Includes:
1
8
'
1. inline fuses
2. a powersupply
3. a cooling fan
4. a backup battery
The power supply receives AC power from the anesthesia machine. AII the
power necessary to operate the ventilator comes from the power supply.
The 伽ont enclosure Includes:
5. a control board (controls operation ofthe ventilator)
6. a front panel assembly
The front panel assembly includes four submodules:
7. an LCD display
8. a keyboard front panel
9. a rotary encoder
10.a speaker
1
2
3
4
5
6
Figure 2-19 • 7100 ventilatorcontrol module
2-28
08/07 Ml110140
2 Theory of Operation
The Ventilator Monitoring Board (VMB) serves as the interface between 由e
ventilator's control board and the breathing system sensors and switches:
• the inspiratory and expiratory flow senso附
• the O2 sensor
• the BaglVent switch
• the VMB is hardwired to indicate a Circle module
• the canister release switch
• machines that include the canister release switch are hardwired to show
so伽are the presence of an EZchange (C02 Bypa臼) canister assembly
• the "control panel switch" signal in an Aespire machine indicates whether
the ABS breathing system is engaged or disengaged
• the Auxiliary Common Gas Outlet (ACGO) switch
• task light power
The VMB is located under the tabletop (below the worksurface). Machines
that do not include monitoring use a depopulated board (Task Light Interface
Board - TLlB) to power the task light and to interface the breathing system
switches.
2.6.4 Serial interface
The Seriallnterface and Connection Board (SICB) provides two functions.lt
Ml110140ω1/07
serves as the inte巾ce between the ventilator's control board and additional
switches located in the machine and channels serial communications signals
from the controller board to the RS232 connector.
The machine switches include:
• the System On/Standby switch
• the O2 supply pressure switch
• the O2 flush switch
The SICB also provides an on/off signal from the system switch ωa remote
monitorthrough the RS232 connector.
2-29
2.6.5 The Pneumatlc Vent
Englne
2-30
1he pneumatic engine enclosure is located in the back chamber of the
breathing system and is shielded to contain EMI emissions. 1he enclosure
includes the pneumatic Vent Engine (VE) and a pneumatic Vent Engine control
Board (VEB).
1he pneumatic Vent Engine consists of the hardware that drives the ventilator
bellows. It includes:
• a 2-micron inlet filter
• a pressure regulator
• a proportional inspiratory valve
• a mechanical overpressure reliefvalve
• a free-breathing check valve
• a PEEP safety valve
• a supply pressure sense switch
• a proportional PEEP valve
• a 200 mL reservoir
• a calibrated bleed orifice
1he pneumatic Vent Engine Board is an interface between the engine
components and the control board and includes:
• an ai阳ay pressure transducer
VE
VEB
08/07 Ml110140
2 Theory of Operation
2.7 Electronic and electrical components
The Aespire 7100 Ventilator electronicj electrical subassemblies or modules
include:
2.7.1 百1e Aesplre 7100
ventllator functlonal
blocks
• a Power Supply for operation under line power
and a backup ba忧ery for limited operation in case of powerfailure;
• a Control Board with digital, analog and power circuits to manage all
operations of the ventilaωr;
• a Front Panel Assembly that includes an LCD dlsplay for display of all
ventilation and monitoring parameters
and a k,叼board for operator input;
• a Ventllator Monltorlng Board to preprocess patientcircuit parameters and
to channel 由e breathing system switch states;
• a Serlallsolatlon Connectlon Board to channel machine switch states and
to provide a RS232 serial output for extemal communication.
面HD.-
守性.一田,‘
Control Board
MCF62060 COIdFlre Prac副嗣『
Møn盼ryer回 VO IleoodIr啕
FIa脯, 8陆M&EEPROM
SCR Clrcωtry
AIO- 口IA ConvM町
W曲白伺 S回m
阳晒阳"叩VOJve C刷刷 (tOVAllmltod)
PEEP Valve eontr剧 (tO VA linMod)
t2VDC8町>PIY (tOVA ImMod)
DC 阳PPIY-咱
剧团町Managemer鸭
r---------_____________________
τ
Insi由 Machlne
I
Note: For non-color displays, the backlight is
powered through the control board, as shown.
For color displays, the control module includes a
separate circuit boardωpower the backlight
'
BnI帽曲曲唱Sys幅m
2-31
Rgure2-20 • 日ectronic 仙nctionalblock diagram as usecl in an Aespire machine
ω1/07
Mll10140
The power supply receives AC input from the machine's AC Inlet Module. The
power supply is a universal40 watt switching supply that outputs two DC
voltages. The DC voltages are routed to the Control Board where they are
further regulated to produce the power requirements for the 7100 ventilator
system.
Aespire 7100
2.7.2 Power Supply
• Input:
Universal85-264 VAC 47 -63 Hz
• OutputVl:
6.0 VDC (:1:0.5%) at 0-5 A
• OutputV2:
9.0VDC(士5%) at 0-0.5 A
symbol.
High voltage in area of 与
&WARNING
6V return
w
9Vreturn
9V
O
···Eo
0000
。
口
E二3
Llne
Neutral
-o
一
A sealed lead acid battery supplies battery backup for the 7100 ventilator.
Since it only provides power in case of a powerfailure, the battery is in a float
charge state most of the time.
Figure 2-21 • 7100 Ventilator power supply
2.7.3 Sealed Lead Acld
Batte叩
The battery meets the following:
• capacity to operate ventilator system for 30 minutes (fully charged);
• long 目oat charge life;
• the ba忧ery is intemallyfused (auto-rese忧able).
Input: Nominally 6.8 VDC at 250C during float charge.
08/07 Ml110140
Output: +0.6ω+6 Amps during discharge
2-32
2.7.4 Control Board
The Control Board contains all of the major circuit functions nece臼aryto
control ventilator operation.
Note: The functions supported by
the color and non-color control
boards are identical. The circuit
descriptions that follow apply to
both control boards, except where
differences are noted to support
the color display.
The Control Board comprises three functional circuit types:
• power circuits,
• analog circui恼,
• di国恼 I circuits.
These circuits are detailed individually in the following sections.
Overall, the Control Board's functions include:
• Bus access control signals for all memory and peripheral devices
• Interrupt handling
• Clocks and timers for the system
• RS232C serial l/O
• Baud rate generator for serial po此
• Hard (power-up) and soft (watchdog error) reset generation
• Data bus buffers
• Memory and 1/0 decoding
• program memory with "memory stick" so仕ware upgrade
• Safety Relevant Computing (SRC)
• Watchdog system
• Data acquisition
• Aow valve control
• PEEP valve drive and PEEP safety valve drive
• Front panel interface
• Audio alarm
Cω忧rol Board - non-color dlspl8J
co四IroIBoard-ωlordlspl町
Mll10140
佣;/07
2-33
刊 e power section ofthe controller board receives the 6 VDC and the 9 VDC
outputs from the power supply.
Control Board
Power Clrcults
Aespire 7100
刊 e 9 VDC supply is used to charge the backup ba忧ery.
The 6 VDC supply is processed further to supply various power requirement
throughout the 7100 ventilator. In case of power failure, the battery is
switched in to supply power.
• powerωdrive the fan (5V)
• 5V supply for digital circuits
• 3.3V supply forthe CPU
• 5V supply for the LCD display backlight
• supply for the LCD display contrast adjustment
• -24V adjustable (non-color display)
• 1.3Vto 2.3V adjustable (color display)
• 1.5A supply for control of the Inspiratory and PEEP valves
• + 12V supply for analog circuits
• + 12V supply for the Monitoring Interface Assembly
• -12V supply for analog circuits
5V lDO (MIC2刷刷 5.0BU)
SUPPlY MONITOR 晶 SEl.E CTION
P.CH MOSFET SWITCHES
ON I STBY SWITCH
5V lDO (MIC2915 D-5.0町}
SHdz
‘
08/07 Ml110140
VOlTAGE 晶 CURRENT
MONITOR
日;gure 2-22 • Control board block d,阳gram -Power circuits
2-34
9.0 V 0 500mA MAX
BATTERY CIWIGER
uω906
2 Theory of Operation
The analog section of the controller board processes inputs from the
Monitoring Interface Board and the Pneumatic Engine Board. It multiplexes
由e inputs fordisplay bythe digital section.
• Inspiratory flow
• Expiratory flow
• Airway 02
• Airway pressure
Under the control of the digital section, the analog section includes drivers for
the pneumatic engine components:
• flow valve (inspira协 ry valve)
• PEEPvalve
• PEEP safety valve
The switch signals from the Monitoring Interface Board and the supply
pressure signal from the Pneumatic Engine Board are passed on as inputs to
the digital section.
HLPRESS
FLDW DAC
院EP_DAC
Figure 2-23. Control board block diagram -Analog circuits
Ml110140
08/07
2-35
The digital section ofthe controller board includes a MCF5206e ColdFire
microcontroller. The Aespire 7100 operatingsoftware isstored in 2MB of8-bit
(lMx16) Rash ROM. The board includes:
Control Board
Dlgltal Clrcults
Aespire 7100
• 1MB of 8-bit (512Kx16) static RAM (SRAM) for operation,
• and (*32Kx8 for non-color) or{* *64Kx16 forcolor) CMOS static RAM for
processing video information
The controller receives switch inputs from the front panel keyboard and the
system inputs from the Vent Monitoring Board, the Seriallsolation Connector
Board and the Vent Engine Board.
The patient circuit parameters are multiplexed through the analog section.
The LCD Display is driven through the Video Controller. It displays the
processed patient circuit paramete附 along with the derived alarm and
system condition messages.
:…
r
Additional outputs include an audio amplifier to drive the speaker and a
RS232 driver for external communication through the Seriallsolation
Connector Board.
l··················el--
••••••••••••••
,
飞咽
a「饵'''气,,
-dEEd
晕目zib--iwE-mu--······JJ
Elz
囊J固dRE‘"圃,
d··-
••
, ••
…
11Eli-…叫
EdE
二
it
…
!!
hh
口刊HMM二,,
EtdE
二
d"hH…
Hl
配川剧自丁噩「
-s二『
EU
HE--RB
……
ZEda2·tEM--2.
而叫旨-t··
吨wa--d
凰dz-ER目··…‘「盈
2··
-ad
温哥dgIE--aHEERgu-MIHE
P
‘2雹6··a!·建理hr-gaa
画画'da·-PSB」22
tpl'
-,fE···EBBEE--J''E·
,.
iiiij;;;;
..Ooseplra恤 bo.rd 阳coIord崎l町
-----_...…".
刷刷峭IIo8rd
conn幢幢
,
………………
,……..……………·、
SerlaIMI阳阳阳
‘ '
'.…..……---------_ ..
08/07 MI110140
Rgure 2-24 • Control board block diagram - Digi臼I circuits
2-36
2 Theory of Operation
The Vent Monitoring Board (VMB) is the interface between the patient circuit
sensors (the inspiratory and expiratory flow sensor, the O2 sensor) and the
ventilator control module. It also passes different switch functions through to
the ventilator control module and provides power to the task light
Respiratory gas flow, to and from the patient, is monitored by measuring the
differential pressure across a variable orifice in each flow sensor. The
pressure transducers for measuring the differential pressure are on the VMB.
Conditioning circuitry is supplied for these transducers and for the 伽ygen
sensor used in the breathing circuit
2.7.5 Monltorlng
Interface
Pre臼ure sense tubing and signal wiring is routed from the sensors and
switches in the breathingsystem to the VMB. A separate cable transfers
power and signals to and from the Control Board in the display module.
啤国」【
-m
吨圃,‘
撞a····0····za--u
吨却
目嗣 1-帽副d
ESD
缸...
回回归
B
酬
BEE-a
..
,a'a".
‘ •••
EzzaHe
2.37
Rgure 2.25 • Breathing Circuit Vent Monitoring Board (VMB)
08/07
Ml110140
The Seriallsolation Connector Board (SICB) provides three functions:
• It serves as an interface between the control board and switches that are
located in the machine itself (not in the breathing system).
• It processes serial communications signals from the control board to the
RS-232 Serial po此 on 由e breathing system side of the Aespire machine.
• Itprovides remote monitorOnjOffthrough an isolated relay.
2.7.6 Serlallnterface
The machine switches include:
• the System OnjStandby switch
• the O2 supply pressure switch
• the O2 flush switch
Machlne Swltches
The serial inte巾ce provides isolated RS232 serial communications. The lXD
(transmit) and RXD (receive) signals between the SICB and the control board
are at RS-232 levels.
Serlal Communlcatlons
Circuits on the board change the signals to digital5V levels; isolate them
through optocouplers; then, change them back to RS-232 levels before
sending them to the outside world.
• The extemal communications signals conform to standard
RS-232C signal standards.
• The extemal connector is a 15-pin female D connector.
• It is configured for Data Communications Equipment (DCE)
• Pin 6 - Receive da恒
• Pin 13 -Transmit data
• Pin 5 - Signal ground
• Pin 1 - Monitor OnjStandby
• Pin 9 - Monitor OnjStandby retum
』〔回回.向守.田,‘
Ea--··a···aa--
•••
,za
08/07 Ml110140
Rgure 2-26 • Seriallsolation Connector Board (SICB)
2-38
2 Theory of Operation
有le Vent Engine Board (VEB) provides two functions:
• Itserves as an interface between the control board and the pneumatic
engine.
• Itprocess臼 the output from the airway pressure transducer.
2.7.7 Vent Englne Board
The board provides a direct connection for the drive and retum lines for the
control valves on the pneumatic engine:
Pneumatlc 画191ne Interface
• Inspiraωry Flow Valve
• PEEPValve
• PEEP Safety Valve
The board routes the Supply Pressure Switch signals to 由 e control board.
刊e VEB includes a +5VA regulator to powerthe airway pressure transducer
circuitry. The circuits provide EMI filtering, signal amplification, and buffering.
Alrway Pressure Transducer
FLOW OR
FLOW_RET
EO-u
国zzau
az--adaz--EU--FH
PEEPVALVE
SAFETYVAl旺
LOW SUPPLY PRESSURE
SWITCH CONNECTOR
旦旦旦L四E
SUPPLY _LOW_RET
PEEP OR
丽#踹
SAFETY_VLV_R盯
2-39
g g g
+5VA
Rgure 2-27 • Vent Engine Board (VEB)
ω:/07
Ml110140
2.8 Ventilator mechanical subsystems
ReferωFigure 11-1, .System connection block diagram," in Section 11 forthe
complete pneumaticjmechanical subsystem.
The mechanical subsystem includes:
Pneumatic Vent Engine
• Drive gas inlet filter
• Supply gas pressure regulaω『
• Inspiratory flow control valve
• PEEP valve; PEEP safety valve; Pressure sense switch
• Mechanical Overpressure Valve (MOPV)
• Bleed resistor
• Free breathing valve
Exhalation valve
Bellows assembly
Breathing circuit flow sensors
2.8.1 Drlve gas
Drive gas (can be selected from O2 or Air) enters the Vent Engine (1) at a pressure of
241ω690 kPa(35ω100 psi) through a 2-micron filter (2) that is located under the
gas inlet fitting.
Smooth side up (undør platø)
Rgure 2-28 • Inlet filter
2-40
ω:/07 Ml110140
2 而eory of Operation
2.8.2 Pre钝ure Regulator
1he pressure regulator (3) is a non-relieving pressure regulator that regulates
high pressure filtered supply gas down ω172 kPa (25 psi).
Figure 2-29 • Pressure regulator
2.8.3 Insplratory flow
control valve
Figure 2-30 • Row control va伽e
Mll10140 08/07
1he inspiratory flow control valve (4) is cycled by the control board to supply
drive gas to the outer chamber of the bellows assembly at a rate determined
by ventilator set目ngs and sensor signals. 1he control valve modulates the
incoming 172 kPa (25 psi) drive gases to an output from 0 ω70 liters per
minute at pressures ranging from 0 to 100 cm H20.
2-41
The exhalation valve contains an elastomeric diaphragm that is used to control the
pre臼ures in the breathing circuit The exhalation valve includes two male ports on the
bottom for:
• Bellows drive gas (5)
• Exhalation valve pilot (6) - (manifold pressure)
The exhalation valve includes three ports on top that connect to the bellows base
manifold:
• Drive gas pass through (7)
• Drive gas return and pop-off valve flow (8)
• APL exhaust flow to scavenging (9)
A port at the back ofthe exhalation valve (10) connects to the down tube that directs all
the exhaust flows to the scavenging receiver.
The exhalation valve is normally open. When the exhalation port is open, gas flows from
the bellows housing to the scavenging port. Approximately
2 cm H20 of pilot pre臼ure is nece臼ary to close the valve. Pilot control ofthe
exhalation valve is done with PEEP Control Valve (A), Supply Pressure Switch (8), and
the PEEP Safety Valve (C).
Rgure 2-31 • Exhalation (PEE月阳Ive
2-42
08/07 MI110140
2 Theory of Operation
2.8.5 Reservolr and bleed
1he reservoir (11) is a 200 ml chamber that dampens the manifold (pilot)
Mll10140
08;07
reslstor
pressure pulses to the exhalation valve.
lE
1he bleed resistor(12) is a .controlled leak" from 0ω10 L/min in responseto
circuit pressures from 0 ω100 cm H20. 1he small quantity of pneumatic flow
exhaustingthrough the bleed resistor permits control ofthe exhalation valve' s
pilot pressure by modulation of the valve output 1he bleed resistor exhausts
only clean drive gas and must not be connectedωa waste gas scavenging
circuit 1he output is routed away from the electrical components to make
sure thatsystems usingoxygen drive gas meetthe 10VA limitation
requirement for oxygen enrichment.
@
Rgure 2-32 • ReseNoir and bleed resistor
2-43
The Bellows assembly is the interface between drive gas and the patient circuit in the
breathingsystem. The pre臼ure reliefvalve (or pop-offvalve) in the bellows
assembly (13) limits pre臼ure in the patient circuit. Excess fresh gas is discharged
through the exhalation valve into the gas scavenging system.
The pre臼ure relief valve is normally closed, maintaining approximately 1.5 cm H20 in
the breathing circuit in a no flow condition, enough to keep the bellows inflated. It is
piloted closed during inspiration and remains closed until the bellows is refilled during
exhalation. Ifthe pressure in the patient circuit exceeds 4 cm H20, the pop-off valve
opens to exhaust excess fresh gas flow at a rate up to 4 Lfmin.
lathlng
Val晴
Figure 2-33 • Bellows pressure reJiefvalve
2.8.7 Mechanlcal
Overpressure
Valve
The Mechanical Overpressure Valve (MOPV) is a mechanical valve (14) that operates
regardless of electrical power. It functions as a third level of redundancy to the
ventilator's pressure limit control functions, supplying pressure relief at approximately
110 cm H20.
Rgure 2-34 • Mechanical overpressure valve
2-44
08/07 Mll10140
Figure 2-35 • Free breathing valve
Ml110140
08107
2 Theory of Operation
The free breathing valve (15) helps assure the patient can spontaneously breathe. The
ventilator is programmed to supply a specified number of breaths per minute to the
patient If, in between one ofthese programmed cycles,由e patient needs a breath
(spontaneous), the free breathing valve permits the patient to inhale. The free
breathing valve is closed on mechanical inspiration.
2-45
Two flow sensors are used to monitor inspiratory and expiratory gas flow. The
inspiratory flow sensor is downstream ofthe gas system inspiratory check valve.
Feedback from the inspiratory transducer is used to supply tidal volumes that make
allowances forthe e仔ects offresh gas flow and circuit compressibility. The expiratory
flow sensor is located at the input to the gas system expiratory check valve. Feedback
from the expiratory flow sensor is used to supply signals for expiratory tidal volume
monitoring and the breath rate.
Airway
Pressure (p
Gauge
Rgure 2-36 • Row sensors
2-46
08/07 Mll10140
3 Checkout Procedure
In 由Issec目。n
3.1 Ventilator post-service checko川............................................. .3-2
3.2lnspectthesystem ........................................................ .3-2
3.3 Pipeline and cylindertests ...................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.4 Flow control tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.1 With O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.4.2Witho川 O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 3-6
3.5 Pressure relieftests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.6 O2 supply alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.7 Flush FlowTest... . . . . .. . . . ... . . . . .. .. . . .. .. . . . . .. . . . . .. .. . .. .. .. . . ... . .. .. .3-8
3.8 Vaporizer back pressure test ............................................... 3-10
3.9 Low-pressure leak test .................................................... 3-11
3.9.1 Negative low-pressure leaktest ...................................... 3-11
3.9.2IS0 or BSI standard low-pressure leaktest ............................. 3-12
3.10 Alarm tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.11 Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.12 Auxiliary 02 f10wmetertests .. . .. . . .... .. . '" .. . .. ... . .. . . .. . . .. . . . . .. . ... . 3-18
3.13lntegrated Suction Regulatortests ......................................... 3-18
3.14Powerfailuretest ...... .................................................3-19
3.15 日 ectrical safetyte筑s .................................................... 3-19
.&. WARNINGS
Afterany repairorservice ofthe Aespire system, complete all tests in thissection.
Mll10140ω:/07
Before you do the tests in 由 issection:
•
Complete all necessary calibrations and subassembly tests. Refer to the
individual procedures for a list of nece部ary calibrations.
•
Completely reassemble the system.
If a test failure occurs, make appropriate repairs and test for correct operation.
3-1
3.1 Ventilator post-service checkout
After servicing the 7100 ventilator, perfonn the following service mode Calibrations:
•
O2 Calibration
•
Zero Flow and Ai附ay Sensors
•
Adjust Drive Gas Regulator
•
Airway Sensor Span
•
PEEP Valve Calibration
•
Inspiratory Valve Calibration
•
Pressure Sensitivity
Then, complete the checkout procedure forthe entire machine in the following
sections.
3.2 Inspect the system
3-2
.&. CAunON
The upper shelfweight limit is 34 kg (75Ib) •
.&. WARNING
Do not leave gas cylinder valves open ifthe pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure.
Before testing the system, ensure that:
•
The equipment is not damaged.
•
Components are correctly attached.
•
The breathing circuit is correctly connected, not damaged.
•
Pipeline gas supplies are connected.
•
Cylindervalves are closed.
•
Models with cylinder supplies have a cylinder wrench attached to 由e
system.
•
Models with cylinder supplies have a reserve supply of O2 connected to
the machine during system checkout.
•
The casters are not loose and the brakes are set and prevent movement.
·刊e power cord is connected to a wall outlet. The mains indicator comes
on when AC Power is connected.
ω1/07 Ml110140
3 Checkout Procedure
3.3 Pipeline and cylinder tests
Mll10140
08/07
.&. CAunON
To prevent damage:
•
Open the cylinder valves slowly.
•
00 not force the flow controls.
If your system does not use cylinder supplies, do not do steps 2 and 3.
1. Disconnect the pipeline supplies and close all cylinder valves (if
equipped).lfthe pipeline and the cylinder pressure gauges are not atzero,
bleed all gasses from the sys'恒m.
a. Connect an 02supply.
b. Set the system switch to On.
c. Set the flow controls to mid range.
d. Make sure that all gauges but O2 are at zero.
e. Disconnect the O2 supply.
f. Make sure that the O2 gauge goes to zero. As pressure decreases,
alarms for low O2 supply pressure should occur.
2. Make sure thatthe cylinde隐 are full:
a. Open each cylindervalve.
b. Make sure that each cylinder has sufficient pressure. If not, close the
applicable cylindervalve and install a full cylinder.
3. Test one cylinder at a time for high pressure leaks:
a. Set the system switch to Standby, which stops 由e02flow.
b. If equipped, turn the auxiliary O2 flow cont刚 fullycloc阳ise (no flow).
c. If equipped, turn offventuri derived suction.
d. Open the cylinder.
e. Record the cylinder pressure.
f. Close the cylindervalve.
g. Record 由e cylinder pressure after one minute. If the pressure
decreases more than indicated below, there is a leak.
5000 kPa (725 pslg) forventilatordrive gas.
690 kPa (100 pslg) for non ventilator drive gas.
Install a new cylinder gasket and do this step again.
h. Repeat step 3 for all cylinde隐•
.&. WARNING
00 not leave gas cylinder valves open ifthe pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure.
4. Connect the pipeline supplies one at a time and ensure that the
corresponding gauge indicates pipeline pre臼ure.
3-3
3.4 Aowcontrol tests
3-4
3.4.1 Wlth O2 monltorlng
Ifthe system Includes O2 monltorl晤, complete the flow control tests
in Section 3.4.1, "With O2 monitoring".
Ifthe 甸事,tem does not Include O2 monltorlng, complete the flow control
tests in Section 3.4.2, "Without O2 moniωring飞
&. WARNING
Nitrous oxide (N20) f10ws through the system duringthis test. Use a safe
and approved procedure to collect and remove it.
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. Attach a patient circuit and plug the patient po比
c. At回ch a bagωthe bag port (or plug 由e bagpo民:).
d. Setthe Bag/VentswitchωBag.
e. Adjust the APL valve to minimum.
2. Connect the pipeline supplies or slowly open the cylinder valves.
3. Turn all flow controls fully clockwise (minimum flow).
4. Set the ACGO selector switch to ABS.
5. Turn on the system.
6. Confirm that the O2 sensor measures 21 % in room air and 100% in pure
O2, If not, calibrate the O2 sensor.
7. Make sure that:
• For a dual-tube O2 flowmeter,
the O2 flowtube shows 0.025ω0.075νmin.
• For a single-tube O2 flowmeter,
the O2 flowtube shows 0.175 to 0.225 Lfmin.
• The otherflowtubes show no gas flow.
8. Set the flow controls to mid range of each flowtube and make sure that the
flowtube floats rotate and move smoothly.
Note
Ifthe system does not include N20, skip steps 9 and 10.
9. Check the Link proportioningsystem concentration (increasing N 20 刊ow).
Observe the following precautions:
a. Start with all valves at the minimum se忧ing.
b. Adjust only the N20 flow control.
c. Increase the N20 flow as specified in the following table and make sure
the O2 concentration is in range.
08/07 Ml110140
3 Checkout Procedure
Note: Allow the O2 monitorto stabilize. Atthe lowerflows, the O2
monitor may take up to 90 seconds to stabilize.
d. If you overshoot a setting, turn the O2 f10w control clockwise until the
N20 f10w decreases to the previous setting before continuing the test.
Set 甜le N20 f10w (Ljmln)
Measured02
0.15 (dual flowtubes only)
21%minimum
0.5 (dual flowtubes only)
21%minimum
0.8
21%t030%
1.0
21%t030%
2.0
21%t030%
6.0
21%t030%
9.0
21%t030%
10.Check the proportioning system concentration (decreasing O2 f1ow).
Observe the following precautions:
a. Start with N20 valve at the maximum setting.
b. Adjust only the O2 flow control.
c. Decrease the O2 f10w as specified in the table and make sure the O2
concentration is in the allowed range.
Note: Allow the O2 monitor to stabilize. At the lower f1ows, the O2
monitor may take up to 90 seconds to stabilize.
d. If you overshoot a setting, turn the N20 f10w control counterclockwise
until the O2 flow increases to the previous setting before continuingthe
test.
Set the O2 f10w (Ljmln)
Measured O2
3.0
21%t030%
2.0
21%t030%
1.0
21%t030%
0.3
21%t030%
11. Check the linearity ofthe Airflow control.
• Turn the N20 f10w control fully clockwise to minimum stop.
Set the O2 f10w
Set 伽e A1r flow (Ljmln)
O2 monltor range
{νmln)
4.0
3.0
61%to 71%
3.5
6.0
45%t055%
1.5
8.0
28%t038%
3-5
3.4.2 Wlthout O2 monltorlng
3-6
& WARNING
1he following procedure will test for any significant malfunction ofthe
Link system but itwill notconfirm proper calibration. Periodic calibration
procedures using an accurate and properly calibrated O2 monitor must
be performed as recommended in the User's Reference Manual, Pa眈 2 ,
section 3 User Maintenance.
&
Nitrous oxide (N20) flows through the system during this test. Use a safe
and approved procedure to collect and remove it.
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. A'忧ach a patient circuit and plug the patient port.
c. Attach a bag to the bag po民 (or plug the bag po 此).
d. Setthe Bag,lVentswitchωBag.
e. Adjust 由e APL valve to minimum.
2. Connect the pipeline supplies or slowly open the cylinder valves.
3. Turn all flow controls fully clockwise (minimum flow).
4. Set the ACGO selector switch to ABS.
5. Turn on the system.
6. Make sure that:
• For a dual-tube O2 flowmeter,
the O2 flowtube shows 0.025ω0.075 Ljmin.
• For a single-tube O2 flowmeter,
the O2 flowtube shows 0.175ω0.225υmin.
.而e other flowtubes show no gas flow.
7. Set the flow controls to mid range of each flowtube and make sure that the
flowtube floats rotate and move smoothly.
Note
Ifthe system does not include N20, skip steps 8 and 9.
8. Check the Link proportioning system (increasing N20 flow).
Observe 由e following precautions:
a. Start with all valves at the minimum se忧ing.
b. Adjust only the N20 flow control.
c. Increase the N20 flow as specified in the following table and make sure
伽e O2 flow is as specified.
08/07 Ml110140
3 Checkout Procedure
d. If you overshoot a se剧唱, turn 由e O2 flow control clockwise until 由e
N20 flow decreases to the previous setting before continuing the test
SettheN20 帽owω阴阳I 协
The O2 flow must be
(L/mln)
greater 由an
(L/mln):
0.8
0.2
2
0.5
4
1.0
10
2.5
9. Check the Link proportioning system (decreasing O2 flow).
Observe the following precautions:
a. Setthe N20 flow to 9.0 Ljmin.
b. Setthe O2 flow to 3νmin or higher.
c. While reducing the 02 flow, set the N20 flow to the rates shown in the
table. The O2 flow must be greater than the minimum limits.
d. If you overshoot a setting, tum the N20 flow control counterclockwise
until the N20 flow increases to the previous setting before continuing
由etest
Set the N20 fl跚
The02 匍侧 mustbe
(using 伽e02 饵owcontrol)
greater 伽an
to(L/mln)
{νmln)
8.0
2.0
4.0
1.0
0.8
0.2
3-7
3.5 Pressure relief
tests
3.6 O2 supply alarm
test
3.7 Flush FlowTest
To check the pressure reliefvalve (vaporizer manifold outlet).
1. Turn all flow controls fully clockwise (minimum 币ow).
2. Setthe ACGO selectorswitch to ACGO.
3. Connect a gauge or a digital manometer to the
ACGO outlet using the positive pre臼ure leak
test adapter.
4. Adjust the O2 flow to 0.5 Lfmin.
5. Verify that the test device reading stabilizes
within the following range:
31-60 kPa, 230-450 mm Hg, 4.5-8.5 psl.
6. Remove the test device and the adapter.
1. Set all flow controlsω3νmin.
Test
Adapter
2. Stop the O2 supply. (Disconnect the pipeline supply or close the cylinder
valve.)
3. Make sure that:
a. The low O2 supply alarm occurs.
b. The N20 (if equipped) and O2 flows stop. The O2 flow stopslast.
c. Air (if equipped) flow continues.
d. Gas supply alarms occur on the ventilator ifthe ventilator uses O2 as
the drive gas.
4. Turn all of the flow controls fully clockwise (minimum flow).
5. Reconnectthe pipeline supplies.
Wlth Ventllator
1. Setthe BagjVentswitch to Vent.
3-8
2. Set the system switch to Standby.
3. Attach a patient circuit and plug the patient po比
4. Setthe ACGO selectorswitch to ABS (Circle).
5. Ensure that the bellows is completely collapsed.
6. Measure the amount of time it takes to fill the bellows when the O2 Flush
button is fully and continuously depressed.
7. Repeat the above measurement two more times (deflate bellows by
removing the plug from the patient port).
• The bellows should fill in 1.8 to 2.3 seconds.
08/07 Ml110140
3 Checkout Procedure
WlthoutVentllator:
1. Setthe Bag/VentswitchωBag.
2. Set the system switch to Standby.
3. At阳ch a patientcircuitand plugthe patientpo此.
4. At恼ch a 3-liter rebreathing bag on the bag arm or manual bag port
(if a 3-liter bag is notavailable, use the time specified for a 1-liter bagtimes
the volume of the bag used).
5. Set the ACGO selector switch to ABS (Circle).
6. Ensure that the rebreathing bag is completely collapsed.
7. Adjust 由e APL valve to maximum.
8. Measure the amount oftime ittakes to fill the rebreathing bagwhen the
O2 Aush button is fully and continuously depre部ed.
Note: When the airway pressure gauge exceeds 10 cm H20, the
rebreathing bag is full.
9. Repeat the above step two more times (deflate 伽e rebreathing bag by
removingthe plugfrom 由 epa甘entpo 民).
• A 3-liter bag should fill in 3.6 to 5.1 seconds.
• A 1-literbagshouldfill in 1.2ω 1. 7 seconds.
Posslble Causes of Fallure
• Large leak (if long filling time).
Ml110140ω;07
• Flush regulator se四ing (Section 5.5).
• Aush regulator cross-connection (if long filling time).
• ACGO selectorvalve inlet cross-connection (if short filling time).
From
~
To
F1ushValYe
(and optlonal Awdllary O2 F10wmete。
O2 F1ush Output
From
ACGO
S剑ector
剑IItch
Vaporlzer 翩翩Ifold
3-9
3.8 Vaporizer back pressure test
3-10
&.
WARNING
Anesthetic agent vapor comes out ofthe common gas outJet during this
test Use a safe, approved procedure to remove and collect the agent
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. Attach a patient circuit and plug the patient port.
c. Attach a bag to the bag port (or plug the bag port).
d. Set the Bag!Vent switchωBag.
e. Adjust the APL valve to minimum.
2. Set the ACGO selector switch to ABS.
3. Set the system switch to On.
4. Setthe O2 flow to 6 Ljmin.
5. Make sure that the O2 flow stays constant and the float moves freely.
6. Adjust the vaporizer concentration from 0 to 1% one click at a time. The O2
flow must not decrease more than 1 Ljmin through the full range. If the O2
flow decreases more than 1νmm:
a. Install a ditferent vaporizer and try this step again.
b. Ifthe O2 flow decreases less than 1 Ljmin with a ditferentvaporizer, the
malfunction is in the first vaporizer.
c. Ifthe O2 flow also decreases more than 1 Ljmin with a ditferent
vaporizer, the malfunction is in the Aespire system. Do not use the
system until it is serviced (repair vaporizer manifold po民 valve).
7. Complete steps 3 through 5 for each vaporizer and vaporizer position.
8. Setthe system switch to Standby.
08;07 Mll10140
3 Checkout Procedure
3.9 Low-pressure leak test
Note
.&. WARNING
3.9.1 Negatlve low-
pressure leak test
Perform either the MNega回ve low-pressure leak test" or 由eMISOorBSI
standard low-pressure leak test. It is not nece臼ary to perform both tests.
00 not use a system with a low-pressure leak. Anesthetic gas will go into
the atmosphere, not into the breathing circuit.
1. Test the leak test device:
a. Put your hand on the inlet of the leak test device. Push hard for a good
seal.
b. Squeeze the bulb to remove all air from the bulb.
c. Ifthe bulb completely inflates in less than 60 seconds, replace the leak
test device.
2. Setthe system switch to S恼 ndby.
3. Set the ACGO selector switch to ACGO.
4. Turn off all vaporize附.
5. Test the anesthesia machine for low-pressure lea悔:
a. Open the flow controls one and a half turns counterclockwise.
b. Con nect the test device to the ACGO outlet.
c. Compress and release the bulb until it is empty.
d. The vacuum causes the floats to move. This is usual. Ifthe bulb
completely inflates in 30 seconds or less, there is a leak in the low-
prl臼sure circuit.
6. Testeach vaporizerfor low-pressure leaks:
a. Setthevaporizerto 1%.
b. Repeat step 5.
c. Setthe vaporizerto OFF.
d. Test the remaining vaporizers.
7. Oisconnect the test device.
8. Turn all flow controls fully clockwise (minimum flow). 00 not over tighten.
& WARNING
Agent mixtures from the low仰ssure leak test stay in the system.
Mll10140 08/07
Always flush the system with O2 afterthe low-pressure leak test (1νmin
forone minute).
Tumo何 all vaporizers at the end of the low-pressure leak test
9.
F1ush the system wi由 O2 :
a. Setthe system switchωOn.
b. Set the O2 flow to 1νmin.
c.
Continue 世le O2 flow for one minute.
d. Tum the O2 flow control fully cloc阳ise (minimum flow).
e.
Setthes归tem switch to Standby.
3-11
3.9.2 150 or B51 standard
low-pressure leak test
3-12
& CAunON
Do the positive pre臼ure leak test at the ACGO outlet only.
1. Set the ACGO selector switch to ACGO.
2. Tum all flow controls fully clockwise (minimum flow).
3. Using 由e positive prl臼sure leak test adapter, connect the ISO or BSI
specific leak test device to 由e ACGO outlet. Push the adapter into the
ACGO outlet throughout the test to get a good seal.
T创t Devlce
σ'op Vl棚}
|咿|
Tωt Devlce
(FrontVlew)
4. Keep flowmeter of the test device vertical for accurate results.
5. Fully open the needle valve on the test device (counterclockwise).
& CAunON
If the needle valve is not fully open, this test can damage the pressure
gauge on the test device.
6. Setthe system switch to On.
ω/07 M1110140
3 Checkout Procedure
7. Open the O2 f10w control and set a to臼 1 f10w ofO.4 νminthrough 由e
f10wmeter on the test device.
8. Makesure 由at 由e pressu陪 gauge on the test device reads zero and that
all other f10w controls are fully closed.
9. CIose the needle valve on the 饱stdevice until 由 etestga咽.e reads:
I505358
3kPa
BSI4272.3
20 kPa
10.lfthe f10w through 由e test device is less than
。.35 L/mln (150) or
0.3νmln(B51) ,
there is a low pressure leak in the anesthesia machine.
11. Fully open the needle valve on the test device to decrease the back
pressure.
12. Test each vaporizer for low-pressu re lea阳:
a. Setthe vaporizer to 1%.
b. Repeat steps 7 由rough 10.
c. Turn the vaporizer OFF.
d. Test the remaining vaporizers.
13. Remove test tool and adapter.
& WARNING
Agent mixtures from the low-pressure leak test stay in the system.
Always 11 ush the system with 02 after the low-pressure lea k test ( 1νmin
for one minute).
Turn all vaporizers OFF atthe end ofthe low-pressure leak test.
14. Flush the system with O2:
a. Setthe system switch to On.
b. Set the O2 flowω1νmin.
c. Continue the O2 f10w for one minute.
d. Turn the O2 f10w control fully clockwise (minimum f10W).
e. Set the system switch to Standby.
3-13
3.10 Alann tes怡
3-14
1. Connect a test lung to the patient connection.
2. Setthe Bag,lVentswitch to Vent.
3. Set the system switch to On.
4. Setthe controls:
• Ventilation Mode: Volume control (select from main menu)
• Ventilator.
- 1idal Vol: 400 ml
- Rate: 12
- I:E Ratio:1:2
- Plimit:40 cm H20
-
PEEP:O阵
• Anesthesia Machine:
- O2 flow: minimum flow
- AII other gases: OFF
- ACGO selector switch to ABS
5. Push O2 Flush ωfill the bellows.
6. Set the Bag,lVent switch to Bag and back to Vent.
7. Make sure that:
a. Mechanical ventilation starts.
b. A subatmospheric pre臼ure alarrn does not occur.
Note: With active gas scavenging, too much scavenging flow can cause
subatmospheric alarrn.
c. The ventilator displays the correct data.
d. The bellows inflate and deflate during mechanical ventilation.
8. Set the O2 flow control ω5 lfmin.
9. Make sure that:
a. The end expiratory pressure is approximately 0 cm H20.
Note: Positive end expiratory pressure when PEEP is 0仔 may indicate
that the scavenging system is not removing enough gas.
b. The ventilator displays the correct data.
c. The bellows inflate and deflate during mechanical ventilation.
10. Test the low minute volume alarm:
a. Go to the alarrns menu.
b. Set the alarrn limit for low minute volume to 6.0νmin.
c. Make sure that a low minute volume alarrn occurs.
d. Go to the alarrns menu.
e. Set the low minute volume alarm ωOFF.
ω:/07 M1110140
3 Checkout Procedure
11. Testthe high ai附aypre臼urealarm:
a. Set Plimit to less than the peak airway pressure.
b. Make sure that 由ehigh ai附'aypre臼ure alarm occurs.
c. Set Plimit to correct level.
12. Testthe apnea and lowai阳aypre臼ure alarms:
a. Turn all flow controls fully cloc阳 ise.
b. Remove the test lung from the patient connection.
c. Other alarms such as low minute volume can occur.
d. Make sure that the low airway pressure and apnea alarms occur. The
apnea alarm occurs after 30 seconds.
13. Test the sustained airway pressure alarm:
a. Set the controls:
• APL valve - Closed (70)
• Bag,lVent switch - Bag
b. Mechanical ventilation stops when the Bag,lVent switch is set to Bag.
c. Occlude the bag port connector with a test plug.
d. Close the patient connection using the test plug located on the side of
the ABS and push the O2 Rush button.
e. Make sure that the sustained pressure alarm occurs after
approximately 15 seconds at the sustained pressure limit (6-30 cm
H20 varies with pressure limit).
14.Testthe O2 monitor and alarms:
a. Remove the O2 sensorfrom the circuit module.
b. Make surethe sensor measures approximately 21 % O2 in room air.
c. Set the low O2 alarm ω50%. Make sure a low O2 alarm occu附.
d. Set the low O2 alarm back to 21 % and make sure that alarm cancels.
e. Put the O2 sensor back in 由e circuit.
f. Remove the test lung from the patient connection.
g. Set the High O2 alarm to 50%.
h. Push the flush butωn to fill the breathingsystem.
i. Set the O2 flow control to 2υmin.
j. Make sure the high O2 alarm comes On.
k. Setthe high O2 alarm backto 100% and make sure that alarm cancels.
1. After 2 minutes in pure O2 • 由e O2 display reads approximately 100%.
m. Tum the O2 flow control fully clockwise (minimum flow).
15. Set the system switch to S恼 ndby.
3-15
3.11 Breathing system tests
3-16
..&.. WARNING
Objects in the breathingsystem can stop gas f10w to the patient. This can
cause injury or death:
•
00 not use a test plug that is small enough to fall into the breathing
system.
1. Verify thatAGSS is operating. For systems that have a flow indicator on the
side, make sure that the flow indicator shows a flow in 由e green (norrnal)
region.
2. Zero the pressure gauge (Section 5.6.1).
Check Valves
3. Make sure 由atthe check valves on the breathing circuit module work
correctly:
a. The Inspira协 ry check valve rises during inspiration and falls at the start
of expiration.
b. The Expiratory check valve rises during expiration and 旬 IIsatthes恼眈
of inspiration.
Ventlla协r Bellows
4. Ventilator bellows test:
a. Set the system switch to Standby.
b. Setthe Bag,tVentswitch to Ventilator.
c. Set all flow controls to minimum.
d. Close the breathing circuit at the patient connection. Use the test plug
located on the side of the ABS.
e. Push the O2 flush butωn to fill the bellows.
f. The pre臼ure must not increaseωmore than 15 cm H20 on the
pressure gauge.
g. Ifthe bellows falls more than 100 mLjmin (ωp of indicator), it has a
leak.
Servlce Mode Tests
5. Enterthe Service Mode: Push and hold the adjustment knob on the
ventilator's display and set由 e system switch to On.
a. Selectand confirrn .Service Modes."
b. Follow the menu structure outline belowto reach the adjustment for由e
inspiratory flow valve. Select and confirrn at each step.
• "Diagnostics TestsjTools"
• "Breathing System Leak Test"
c. Follow the instructions on the screen.
d. The leak rate should be less than 200 mLjmin.
For machines with a single-tube O2 flowmeter, the pressure reading
should reach 30 cm H20 atminimum flows greaterthan 200 mLjmin.
Note: If test fails, see Section 7.2, "Breathing System Leak Test Guide".
08/07 M1110140
3 Checkout Procedure
Bag Clrcult
6. Testthe Bag circuitfor lea悔:
a. Setthe system switch to On.
b. Set the Bag,tVentilator switch to Bag.
c. Plug the Bag port (use your hand or the approved test plug).
d. Close the APL valve (70 cm H20).
e. Set the O2 flow to 0.25νmin.
f.
Close 世le patientconnection (usinga hand ortest plugon the side of
the breathingsystem) and pressurize 由 e bag circuit with the O2 flush
bu忱。n to approximately 30 cm H20.
g. Release the flush button. The pressure must not decrease. A pressure
decrease la唱.e enough to see on the gauge indicates an unaccep切 ble
leak.
Note: Iftestfails, see Section 7.2, "Breathing System Leak Test Guide".
APL Valve
7. Testthe APL valve:
a. Fully close the APL valve (70 cm H20).
b. Set the total fresh gas flow to approximately 3 Ljmin and make sure
thatthe value on the inspiraωry pressure gauge does not exceed 85 cm
H20. Some pressure fluctuation is normal.
c. Fully open the APL valve (to 由 e MIN position).
d. Set O2 flow to 3 Ljmin. Tum any other gases 0仔.
e. Make sure thatthe value on the inspiratory pressure gauge is less than
approximately 5 cm H20.
f. Push the O2 flush button. Make sure that the value on the inspiratory
pressure gauge stays less than 10 cm H20.
g. Set the O2 flow to minimum and make sure that the value on the
inspiratory pressure gauge does notdecrease below 0 cm H20.
8.
Remove your hand or the test plug from 由e patient connection.
9. Setthe System switch ωS恼 ndby.
.&. WARNING
Make sure that there are no 恒st plugs or other obje也 caughtin
the breathing system.
Ml110140ωvm
~U
3.12 Auxiliary O2 flowmeter tests
1. Open the O2 cylindervalve or connect an O2 pipeline.
2. Rotate the flow cont刚 clockwise (decrease)ωshut off the flow. The ball
should rest at the bottom of the flow tube and not move.
3. Rotate the flow control counterclockwise (i ncrease)凰The ball should rise
immediately after rotation is begun. Itshould rise smoothly and steadily
with continued counterclockwise rotation. When a desired flow is set, the
ball should maintain in a steady position.
4. Occlude the auxiliary O2 outlet. The ball should rest at the bottom of the
flow tube and not move. A ball that does not rest at the bottom of the flow
tube indicates a leak and requires service.
5. Rotate the flow control clockwise to shut offthe flow.
3.13 Integrated Suction Regulator tests
3-18
The gauge needle should come to rest within the zero range bracket when no
suction is beingsupplied. Gauges which do notcomply may be outof
calibra甘 on.
1. Adjustthe regulator setting to minimum.
2. Turn the mode selectorto I (On).
3. Ensure the gauge remains less than 200 mmHg (26 kPa, 0.26 Bar).
4. Occlude the inlet.
5. Ensure the gauge remains less than 200 mmHg (26 kPa, 0.26 Bar).
6. Adjust the regulator in an increasing vacuum level.
7. The gauge should rise after rotation has begun. The gauge should rise with
continued rotation ofthe regulator adjustment.
8. Adjust the regulator setting to minimum.
9. Turn the Mode selectorto 0 (0的.
08/07 Mll10140
3.14 Power failure test
3 Checkout Procedure
1. Connect the power cord to a mains outlet. The mains indicator on the
display comes on when AC Power is connected.
If the indicator is not on, the display assembly is not receiving AC power.
• Verify AC powerωthe machine (reset circuit breaker).
• Check fuses in display assembly's inlet module.
• Check fuses in machine's AC inlet assembly.
2. Setthe system switch to On.
3. Unplugthe powerco时 with 由 e system turned on.
4. Make sure that the power failure alarm comes on.
5. Make sure the following message is on the ventilator display:
• .On Ba忧ery - Power OK?"
6. Connectthe powercable again.
7. Make sure the alarm cancels.
3.15 Electrical safety tests
Mll10140
08/07
Make sure the system is completely assembled and all accessory devices are
connected to electrical ou甘ets.
1. Connectan approved testdevice (e.g. UL, CSA, or AAMI) and verifythatthe
leakage current isless than:
Voltage
120jl00Vac
220j240Vac
Max. Leakage Current
300μAmps
500μAmps
2. Make sure thatthe resistance to ground isless than O.2Q between an
exposed metalsurface and the ground pin on the power cord.
3-19
4 Self Tests and Service Mode
.&. WARNING:
Post-Service Checkout is required after you complete 由 is 扭曲ion. You must
perform Section 3.1 Post-service checkoutafter performing any maintenance,
service or repair. Failure to do so may result in patient injury.
In this section
To ensure proper operation, the Aespire 7100 Ventilator includes several tests that run
automatically (selftests) and a series of menu pages that a qualified service person can
use to test, calibrate, or troubleshoot ventilator related components in the Aespire
anesthesia machine (Service Mode).
Mll10140 08/07
4.1 Selftests ................................................................ .4-3
4.2 Service Mode .. .. . . .. . . . . . .. . .. . . ... . .. . ... . . . ... . . . . . .. . . .. . ... . .. . .. .. . . .4-4
4.3 About Ventilator .. . .. . . . . . . .. . .. . . .. . . .. . .. . . . . . .. . . .. .. .. . . .. . . . . .. . .. .. . . .4-5
4.4 Alarrn Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.5 Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.6 Language . .. .. .. . . . . .. .. .. .. . . . . .. ... . ... . . . . . .. .. . .. . . . . . . .. . . . . . . .. .. .. .4-8
4.7 User Settings .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.1ScreenContrast .....................................................4-9
4.8 System Con币guration ..................................................... 4-10
4.9 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 4-12
4.9.1 02 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . " 4-13
4.9.2 Zero Flow and Airway Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . " 4-14
4.9.3 Adjust Drive Gas Regulator .......................................... 4-15
4.9.4AirwaySensorSpan ................................................4-16
4.9.5 PEEP Valve Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. 4-18
4.9.6 Inspiratory Valve Calibration ......................................... 4-19
4.9.7 Pressure Sensitivity ................................................ 4-21
4.9.8 Service Calibrations Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . " 4-23
4-1
4.10 Diagnostic Tests/Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
4.10.1 Display A/D Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.10.2 Display Discrete 1/0 Signals . . .. . . . . .. . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . 4-27
4.10.3 Display Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
4.10.4 Test Panel Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
4.10.5 Valves -Test Tool ................................…………… ..4-30
4.10.6 Test CPU and Memory ……………………………………… 4-31
4.10.7TestEEPROM .................................................... 4-32
4.10.8TestSeriaIPort ................................................... 4-33
4.10.9 Test 5V Fail Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.10.10 Test Inspiratory Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.10.11 Test PEEP Valve ................................................. 4-36
4.10.12TestPEEPSafetyValve ...........................................4-37
4.10.13 BreathingSystem LeakTest ................................… ....4-38
4.10.14 Test Pressure Limit Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
4.11 Upgrade Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
08/07 Ml110140
4.1 Self tests
4 Tests and Calibration
The Aespire 7100 Ventilator software includes self tests that determine whether or not
由e operating software is functioning properly and whether or not the electronic circuits
on the circuit boards are functional.
The self tests include:
•
powerup tests
•
continuous tests
•
periodic tests
Powerup tests
The following list of the tests run at powerup:
•
Sequential watchdog
•
Logical watchdog
•
Data RAM walking pa忧emtest
•
FLASH ROM CRC verification
•
PEEP Valve test
•
PEEP Safety Valve test
If one or more of these tests fail, the display provides a readout of the problem.
Contlnuous tests
These tests are run continuously during normal operation and alarms are associated
with each test. A failure causes an alarm ωdisplay on the screen in the alarm display
area.
•
Supply voltage chec恼
•
Battery voltage checks
•
Inspiratory control valve DAC and voltage feedback
•
PEEP control valve DAC and voltage feedback
Perlodlc tests
These tests are run every 30 seconds during normal operation. Alarms are associated
with each test. A failure causes an alarm to display on the screen in the alarm display
area.
•
CPUTest
•
Display RAM walking pattem test
•
Da恼 RAM walking pa忧.emtest
•
FLASH ROM CRC verification
Mll10140 08/07
4-3
4.2 Service Mode
刊e Servlce Mode is used to test, calibrate, or troubleshoot ventilator related
components in the anesthesia machine.
1here are two ways to enter the service mode:
•
Ifthe machine is tumed off, push and hold in the adjustment knob while setting
the system switch to On. Hold the adjustment knob pushed in until the .Service
Confirmation" menu appears. Use the adjustment knob to highlight .Service
Mode", then push the adjustment knob to confirm the selection.
•
Ifthe machine is already in normal operation, setthe Bag/Ventswitch to Bag.
1hen, press the VT/Pinsp, the PEEP, and the menu switches at the same time
to reset the so伽are (powerup). Push and hold the adjustment knob until the
.Service Confirmation" menu appears.
Note: Volume alarms must be in standby.
..- Language Speclflc
From this menu you can:
•
Go to the Service Modes
•
Return to Normal Operations
Servlce Modes Menu
The Servlce Modes main menu displays the servi四 te也 you can select.
4-4
SelVice Modes
圃'回-回..... ………
AI部mLog
Error Log
Lang崎ge
User Settlngs
Systell Contiguration
Calibrations
Diagn国际 Test矿Tools
Upgr~ Optlons
Go toN田旧 I ~rations
1he selectable service tests are displayed in categorical order. But you can select the
service test from this menu in any order. 1he following sections in this manual are
sequenced in the order that they appear on the screen.
08/07 Ml110140
4 Tests and Calibration
4.3 About Ventilator
Ml110140ω1/07
The About Ventllator menu identifies the current software loaded into the ventilator's
Rash-ROM memory and displays the total .on time" of the system.
•
Software Verslon
•
S。伽'areDate
•
Hardware ID
•
Total 句stemOn Tlme
•
Current Bootup Count - The bootup count is incremented each time the
machine is tumed on. The bootup count is not overwritten when new software
is loaded. However, if a new control board is installed, the bootup count will
reflect the value stored in the EEPROM of the new boar吼
旬flwi!le V阳 ion:
缸躏NareDate:
Har刷刷 ID:
About Ventilator . . .
1.1
A咽篇 2删节 :31:43
Total 句~阳伽Ti.e:
0 Vear( s] 0 阳~(s] 0 Hour( s] 0 MirUe(s]
CUrrent 勘ot叩 E咽d:
4-5
4.4 Alarm Log
The Alarm Log displays up to 20 of the most current alarrn messages that have been
logged. Each log entry shows:
•
Bootup Count number (the bootup count is incremented each time the
machine is turned on).
•
Time (ms) - the time in milliseconds since bootup when the latest alarm
condition occurred.
•
Alann message associated wi伽由 e particular alarm condition.
•
# Times - the number of times that the specific alarm condition has occurred
during the noted bootup count.
The bottom-Ieft comer ofthe screen displays additional information:
•
Bootup Count Last Cleared
•
Current Bootup Count
Clear Alarm Log
Bootup Count Last CI回red 1
Current Bootup Count:
55
Go 10 Service Mode$ Menu
Scroll Data
To view all the log entries, select "Scroll Data."
Then, use the control knob to scroll through the entries.
To exit Scroll Data, press the control knob.
Clear Alarm Log
To clearthe alarrn log, select "Clear Alarm Log." The system asks you to confirm thatyou
want to clearthe alarm log.
Remarks
Afteryou clearthe alarrn log:
•
the "Bootup Count Last Cleared" number will be reset to the "Current Bootup
Count" number.
•
the menu will show the message "No entries in log! ".
4-6
08/07 Ml110140
4.5 Error Log
4 Tests and Calibration
There are two special types of alarms:
•
Minimum monitoringalarms thatstop mechanical ventilation
•
Minimum shutdown alarms that stop mechanical ventilation and monitoring.
An alarm message that results from these special types of alarms is considered an error
alarm.
The Errol Log displays up to 20 ofthe most current Error messages that have been
logged.
Each log entry shows:
•
Bootup Count number (the bootup count is incremented each time the
machine is turned on).
•
T1me (ms) - the time in milliseconds since bootup when the latest alarm
condition occurred.
•
Error message associated with the particular alarm condition.
•
Address - the place in the software sequence where the last occurrence ofthe
errorωokplace.
The error address and software revision are important pieces of information to note if
technical suppo此 is required.
The bo忱。m-Ieft comer of the screen displays additional information:
•
Bootup Count Last Cleared
•
Current Bootup Count
口幅rEπOIL吨
自lOIUP C四ntL甜 CI剧创 1
0町田11 Boot叩 E国mt:
55
Go to SelVlce Modes M酬』
Scroll Data
To view all the log entries, select "Scroll Data."
Then, use the control knob ωscroll through the entries.
To exit Scroll Data, press the control knob.
Clear Error Log
To clearthe error log, select "Clear Error Log." 而esystem as陆 youωconfirm thatyou
want to clear the error log.
Ml110140 08/07
Remarks
Afteryou clearthe alarm log:
•
the "Boot Count Last Cleared" numberwill be resetto the "Current Boot Count"
number.
•
the menu will showthe message "No entries in log!".
4-7
1he text shown in the nonnal mode of operation is language sensitive. However, the
Service Confinnation menu (except for text MNonnal Operation") and all the Service
Modes menus are shown only in English.
刊e Language menu is used to set the specific language for nonnal operation.
Go to SelVlce Modea M回四
1he 7100 ventilator supports the following languages. 1he language selections appear
in language specific text.
. .. English
Español . . . . . . Spanish
Deutsch . . . . . . German
Polski. . . . . . . . Polish
p。此uguês . . . . portugue曾
Français . . . . . Fren由
Itallano. . . . . . . Italian
Ce剖na ...... Czech
Magyar . . . . . . Hungarian
Nederlands . . . Dutch
.. Greek
...... Russian
. . . Japanese
.. Chinese
1he following additionallanguages are available in Software revisions 1.4 and greater.
Suomi . . . . . . . Finnish
Tilrkçe. . . . . . . Turkish
Norsk. . . . . . . . Norweglan
1he language se创ng is stored in EEPROM with the default setting as English.
08/07 Ml110140