Leica DMI3000B Instructions Manual
Read Online
Read the Leica DMI3000B Instructions Manual online using the built-in PDF viewer.
Text of the manual
6.10 Installation and Replacement of the
transmitted Light Lamps:
107 or 107/2 Lamp Housing........................ 45
6.11 Installing the Lamp Housing Mount
and Mirror Housing) ................................... 46
6.12 Installation and Replacement of
Incident Light Lamps .................................. 48
6.13 Equipping the Incident Light
Turret Disk .................................................... 52
6.14 Inserting the Front Module Slider ............ 55
6.15 Installation of the Polarizer
and Analyzer................................................ 55
6.16 Optional Accessories................................. 57
6.17 Connection to the Electronics Box .......... 58
6.18 Connection to the Computer ..................... 59
6.19 Connection to the Power Supply ............. 59
7.
Start-up ........................................................ 60
7.1 Functional Principle .................................. 60
7.2 Switching on the Microscope .................. 64
7.3 The LeicaDisplay ........................................ 65
7.4 The Function Buttons on the Stand ......... 66
7.5 The SmartMove
Remote Control Module ............................ 69
7.6 Illumination .................................................. 69
7.6.1 Transmitted light .............................. 69
7.6.2 Incident Light - Fluorescence ........ 73
7.7 Checking Phase Contrast Rings ............... 74
7.8 Checking modulation contrast
slit diaphragms............................................ 77
7.9 Setting the Motorized Polarizer ............... 77
7.10 Adjusting the Light Sources ..................... 78
Contents
1.
Important Notes about this Manual ........
7
2.
Intended Purpose of the Microscope .....
8
3.
Safety Notes ................................................
9
3.1 General Safety Notes ................................
9
3.2 Electrical Safety.......................................... 10
3.3 Safety Instructions
for Handling the Light Sources ................ 12
3.4 Notes on handling laser devices ............. 12
3.5 Safety Instructions
for Handling Acids and Bases .................. 12
3.6 Disposal........................................................ 13
4.
Overview of the Leica DMI Series .......... 14
5.
Unpacking the Microscope ..................... 27
6.
Assembling the Microscope .................... 30
6.1 Assembly Tools ........................................... 30
6.2 Installation of the Transmitted Light
Illumination Carrier (TL) ............................. 31
6.3 Installation of the DIC Module
and DIC Objective Prisms ......................... 32
6.4 Installation of Specimen Stages .............. 33
6.5 Installation of Condensers ........................ 38
6.6 Installation of Eyepieces ........................... 43
6.7 Installation of Objectives .......................... 43
6.8 Installation of Filters
in the Illumination Arm............................... 44
6.9 Installing the transmitted Light
Lamp Housing ............................................. 44
Operation ..................................................... 81
8.1 Switching on................................................ 81
8.2 Contrast Methods ....................................... 83
8.2.1 Bright Field (TL) ................................ 83
8.2.2 Phase Contrast (TL) ....................... 85
8.2.3 Dark Field (TL) .................................. 86
8.2.4 Polarization (TL) ............................... 87
8.2.5 Differential
Interference Contrast (TL) ............. 88
8.2.6 Integrated Phase Contrast (TL) ..... 89
8.2.7 Integrated
Modulation Contrast (TL) ................ 90
8.3 Fluorescence............................................... 91
8.4 Combination Methods ............................... 93
8.5 Focusing ....................................................... 94
8.6 Tubes ............................................................ 96
8.7 Port selection ............................................. 96
8.8 Eyepieces..................................................... 97
8.9 Objectives .................................................... 98
8.10 Stages and Object Displacement ............ 101
8.11 Magnifi cation Changer .............................. 102
8.12 Light sources ............................................... 103
8.13 Aperture and Field Diaphragm ................ 104
9.
Troubleshooting .......................................... 105
10. Care of the Microscope ............................ 109
10.1 Dust Cover ................................................... 109
10.2 Cleaning ....................................................... 109
10.3 Handling Acids and Bases ........................ 110
11. Major Consumable
and Replacement Parts ............................. 111
12. Dimensions.................................................. 112
13. Abbreviations and Pictograms ................ 113
14. Index ............................................................ 115
15. EU Declaration of Conformity .................. 117
4. Overview of the Instruments
Fig. 1
Left side, Leica DMI4000 B and DMI6000 B
1
Eyepiece
2
Eyepiece tube
3 Top port
4 Intermediate pupil interface
5 LeicaScreen
6 Light intensity
7 Field diaphragm
8 TL/IL switching
9
Aperture diaphragm
10 Focus wheel (motorized Leica DMI6000 B,
manual (fi ne and coarse) Leica DMI4000 B)
11 Variable function buttons
12 Left side port
13 Booster lens
(Leica DMI6000 B fl uorescence microscopes only)
14 Lamp mount (fl uorescence microscopes only)
15 Condenser head
16 Condenser base
17 Field diaphragm
18 Transmitted light lamp housing
19 DIC objective prism disk
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
16
18
19
4. Overview of the Instruments
Fig. 2
Right side Leica DMI4000 B and DMI6000 B
1 E-Focus buttons (Leica DMI6000 B only)
2 Focus wheel (motorized Leica DMI6000 B,
manual (fi ne) Leica DMI4000 B)
3 Variable function buttons
4 Opener for drawer (fl uorescence microscopes only)
5 Drawer (fl uorescence microscopes only)
6 Right side port
7
Analyzer slot
8 Centering window (fl uorescence microscopes only)
9 Field diaphragm centering
(fl uorescence microscopes only)
10 Incident light lamp housing (fl uorescence microscopes
only)
11 Objective turret
12 Stage with attachable mechanical stage
1
2
3
4
5
6
7 8
12
9
10
11
4. Overview of the Instruments
Fig. 3
Front view Leica DMI4000 B and Leica DMI6000 B
1 LeicaScreen
2 Front control panel
3 Port switching
4 Top port
5 Manual transmitted light fi lters
6 Bertrand lens centering
1
2
3
4
6
5
4. Overview of the Instruments
Fig. 4
General view Leica DMI4000 B and Leica DMI6000 B with SmartMove remote control module
Fig. 3b
SmartMove remote control module
1
Travel in x
2
Travel in y
3
Focus
4 Variable function buttons
(pre assigned at factory)
1
2
3
4
Fig. 3a
Front control panel
1
Fluorescence cube
2
Shutter
3
100% light to all eyepieces
4 Port selection
5 Magnifi cation selection
6 1x tube lens
1
1
2
5
4
3
7.1 Functional Principle (Leica DMI4000 B and Leica DMI6000 B)
1 . Intelligent Automation
• Switching between contrast methods at the touch of a button. Light rings, DIC prisms, etc.
are automatically positioned in the beam path.
• The microscope recognizes the selected objective and associated contrast method.
The intensity (INT), aperture diaphragm (AP) and fi eld diaphragm (FD) are always set to suit-
able values.
• The INT, AP and FD values are always based on the currently activated illumination axis
(transmitted light or incident light).
• The INT, AP, and FD values can be adjusted individually. Manual adjustments overwrite the
previous settings. The current setting is stored and is retained from one session to the next
when power is switched off.
2. Controls
• SmartMove knobs
for stage and focus control
• Fixed function buttons on stand
for INT, AP, and FD, as well as for switching between transmitted light and incident light axis
• Variable function buttons on stand, SmartMove, STP6000
These function buttons have functions suitable to the confi guration of your microscope as-
signed to them at the factory. The functions can be reprogrammed and/or adapted to your
specifi c requirements, however.
• Complete control of microscope and camera via software
(Leica Application Suite (LAS/LAS AF))
Thanks to its intelligent automation, the Leica DMI4000 B and DMI6000 B can be controlled using a
variety of control elements.
(DMI4000 B and DMI6000 B)
Function
Function
Function
Rotary
and
buttons
buttons
buttons
knobs
STP6000
Stand
Stand
4000
6000
4000
6000
4000 6000 4000 6000 4000/6000
Select contrast method
-
-
+
+
+
+
-
-
+
Change transmitted light/incident light axis
+
+
-
-
-
-
-
-
+
Change to objective
-
-
-
+
-
+
-
-
+
Teach-in parfocality
-
-
-
-
-
-
-
-
+
Change operating mode (dry/imm)
-
-
+
+
+
+
-
-
+
Illumination Manager
+
+
+
+
+
+
-
-
+
Magnifi cation changer (motorized)
+
+
-
-
-
-
-
-
+
Focusing
-
+
-
-
-
-
-
+1)
+
Set stops
-
+
-
-
-
-
-
-
+
Go to stop
-
+
-
-
-
-
-
-
+
Change step increment (coarse/fi ne)
-
-
-
+
-
+
-
-
+
XY stage positioning
-
-
-
+
+
Change speed
-
-
-
-
-
-
-
-
+
Stage positions (store/go to)
-
-
-
-
-
-
-
-
+
Change to fi lter/refl ector cube
+
+
(+)
+
+
-
-
+
Side and bottom port (DMI6000 B only)
+
(+)
+
-
+
DIC fi ne adjustment
+
+
-
-
-
-
-
-
+
Adapted Focus Control (AFC)
-
+
-
+
-
+
-
-
+
+
always possible
(+) optional
-
not possible
1)
Focusing alternatively via wheels
Possible Assignments for Variable Function Buttons on Stand and SmartMove
For Leica DMI4000 B and Leica DMI6000 B:
Function button
Function
BF
Bright fi eld transmitted light
PH
Phase contrast transmitted light
ICT
Interference contrast, transmitted light
DF
Dark fi eld transmitted light
IMC
Integrated modulation contrast
POL
Polarization transmitted light
CHANGE TL
Cycle through all contrast methods
INT ↑
Increase intensity (transmitted light)
INT ↓
Reduce intensity (transmitted light)
AP ↑
Open aperture diaphragm (transmitted light)
AP ↓
Close aperture diaphragm (transmitted light)
FD ↑
Open fi eld diaphragm (transmitted light)
FD ↓
Close fi eld diaphragm (transmitted light)
SHUTTER TL
Open/close TL shutter
TL FLT 1
Enable/disable transmitted light fi lter at position 1
TL FLT 2
Enable/disable transmitted light fi lter at position 2
FLUO
Fluorescence (last fi lter cube)
CUBE 1-6
Select fi lter cube in position 1-6
CHANGE CUBE CW
Change cube clockwise (1 → 4)
CHANGE CUBE CCW
Change cube counterclockwise (4 → 1)
INT FLUO ↑
Increase intensity (fl uorescence)
INT FLUO ↓
Reduce intensity (fl uorescence)
FD FLUO ↑
Open fi eld diaphragm (fl uorescence)
FD FLUO ↓
Close fi eld diaphragm (fl uorescence)
CHG FW
Toggle fi lter functions
IFW
Activate external fi lter wheel
ExMan
Activate Excitation Manager
SHUTTER FL
Open/close fl uoshutter
COMBI
Combination method (PH fl uorescence or ICT fl uorescence)
CHANGE COMBI
Cycle through all combination methods
CHANGE OBJ CW
Cycle through objectives clockwise
CHANGE OBJ CCW
Cycle through objectives counterclockwise
Z FINE
Activate fi ne focusing (Leica DMI6000 B only)
Z COARSE
Activate coarse focusing (Leica DMI6000 B only)
XY PRECISE
Activate precise stage
XY FAST
Activate fast stage
BTP ON/OFF
Bottom port on/off (Leica DMI6000 B only)
DRY/IMM
Switch dry/immersion
CHANGE FLT
Switch TL fi lter
CHANGE CS
Switch to confocal application
OBJ 1-6
Select objective at position 1-6
MEM 1-6
Memory activated stored functions
AFC ON/OFF
Turns AFC on or off
AFC HOLD
Holds current position
If a component has not been installed correctly,
the LeicaScreen will display an error message.
See Troubleshooting chapter, → p. 105.
Components such as diaphragms, condensers,
light, and phase rings have been pre-centered
at the factory. It may be necessary to correct the
centering after the microscope has been trans-
ported and assembled.
Before performing the required steps, please fa-
miliarize yourself with the LeicaScreen and the
controls.
Caution!
After turning on the gas discharge lamp, the
burner must be immediately adjusted. There-
fore, do not turn on the power supply unit
yet. First, work in transmitted light in order
to familiarize yourself with the microscope’s
controls.
Fig. 81
LeicaScreen
Initialization
7.2 Switching on the Microscope
Leica DMI3000 B:
• Switch on the microscope’s power at the On/
Off switch. The signal lamp is lit when the in-
strument is ready. (For the Leica DMI3000 B
please continue at 7.4. Function Buttons on the
Stand)
Leica DMI4000 B and Leica DMI6000 B:
• Switch on the power of the electronics box at
the On/Off switch (80.1). The signal lamp (80.2)
is lit green when the unit is ready. All motor ized
microscope components will then run through
an initialization phase.
Note:
If a PC is connected, switch on the electronics
box fi rst, and then the computer.
After the initialization (Fig. 81) is complete, the
LeicaScreen will display the microscope’s cur-
rent settings (Fig. 82).
Fig. 80
front side
Leica CTR6000
1
On/Off switch
2
Signal lamp
1
2
Fig. 82
LeicaScreen
after
Initialization
7.3 The LeicaDisplay
(Leica DMI 4000 B and DMI 6000 B)
The screen displays the microscope’s current
settings. The content of the display depends on
the features of the individual microscope.
For information on the abbreviations used, please
turn to the table of abbreviations → p. 113.
The screen has a number of areas and lines.
Line 1:
contrast method
Line 2:
objective/ magnifi cation
Line 3:
illumination/ diaphragms
Line 4:
active ports
Line 5:
focus/ stops (DMI 6000 B only)
The content of the display changes according to
the active function.
Contrast method
Objective/
Magnifi cation
Focus/stops (DMI6000 B only)
Ports/Eyepiece
Illumination
Diaphragm
Fig. 83
Arrangement of the function buttons – overview
1
Four variable function buttons
2
Illumination Manager
3 Front control panel
4
Focus buttons (DMI6000 B only)
5
Three variable function buttons
6
SmartMove knobs
7
SmartMove function buttons
5
4
3
3
2
1
6
7
7
6
Pictograms
Changes to the light intensity as well as aperture
and fi eld diaphragm settings are stored for the in-
dividual objectives and contrast methods.
Fig. 84
Fixed function buttons (left side of stand)
1 Toggle transmitted light/incident light
2
Aperture diaphragm
3
light intensity
4
fi eld diaphragm
4
1
3
2
7.4 The Function Buttons on the Stand
Leica DMI3000 B:
• Focus wheels: the left-hand focus wheels can
be used for both coarse and fi ne focusing; the
right-hand focus wheel for fi ne focusing only
(a version of the Leica DMI3000 B with mir-
rored focus controls is also available)
• Light intensity: the transmitted light intensity
can be adjusted continuously from 0 to 12 V
using the potentiometer at the lower left of the
front of the microscope stand.
For the Leica DMI3000 B please continue at 7.6.
Illumination.
Leica DMI4000 B and Leica DMI6000 B:
A number of function buttons are located on both
sides of the stand. These can be broken down
into fi xed and variable buttons. The variable func-
tion buttons have different functions depending
on the features of the individual microscope.
Fixed function Buttons on the Left Side
The TL/IL button (84.1) toggles between the inci-
dent-light and transmitted light axis. The contrast
method last used with a given axis is restored
when switching.
The INT buttons (84.3) adjust the light intensity.
The adjustment can be made in coarse or fi ne
steps. Pressing both INT buttons at the same time
toggles between coarse and fi ne adjustment.
The AP buttons (84.2) for the aperture diaphragm
and FD (84.4) for the fi eld diaphragm open and
close their respective diaphragms.
The motorized DMI stands can be controlled
using the function buttons on the stand, the re-
mote control SmartMove or the STP6000.
Variable Function Buttons on the Stand
The variable function buttons are assigned func-
tions at the factory that are appropriate to the
features of your microscope. They are labeled
accordingly. For details on button assignments,
please refer to the included identifi cation sheet.
For information on the abbreviations used, please
refer to the list → p. 63.
Note:
The Leica Application Suite (LAS) software is re-
quired for changing the button assignments.
Fig. 85
Function buttons (left side of stand)
1
Variable function buttons
2
Open/close aperture diaphragm
3
TL/IL switching
4
Open/close fi eld diaphragm
5
Increase/decrease light intensity
Fig. 86
Function buttons (right side of stand)
1
Variable function buttons
BF
PH
ICT
DF
IMC
POL
CHANGE TL
INT ↑
INT ↓
AP ↑
AP ↓
FD ↑
FD ↓
SHUTTER TL
TL FLT 1
TL FLT 2
FLUO
CUBE 1
CUBE 2
CUBE 3
CUBE 4
CUBE 5
CUBE 6
AFC ON/OFF
AFC HOLD
CHANGE CUBE CW
CHANGE CUBE CCW
INT FLUO ↑
INT FLUO ↓
FD FLUO ↑
FD FLUO ↓
CHG FW
IFW
ExMan
COMBI
CHANGE COMBI
CHANGE OBJ CW (only DMI6000 B)
CHANGE OBJ CCW (only DMI6000 B)
Z FINE (only DMI6000 B)
Z COARSE (only DMI6000 B)
XY PRECISE
XY FAST
BTP ON/OFF (only DMI6000 B)
DRY/IMM
CHANGE FLT
CHANGE CS
OBJ 1-6
MEM 1-6
Turns AFC on or off
Holds current position
Possible functions*:
* See page 63 for abbreviations
2 3 4 5
1
1
Function Buttons on the Front Panel (Fig. 87)
100% of the light goes to the eye piece
(87.1).
Toggle function for the side ports
(87.2). This function depends on the
individual microscope confi guration.
Note:
Switching to the bottom port:
via the variable function buttons (Lei-
ca DMI6000 B only), switching to top
port: manually.
SHUTTER Opens and closes the shutter (87.3).
Switches
between
the
possible
mag nifi cations of the magnifi cation
changer (87.4).
The magnifi cation changer is set to
the magnifi cation 1x (87.5).
Fig. 87
Front control panel
1
100% light to eyepiece
2
Toggle ports
3
Shutter
4
Switch between subsequent magnifi cations
5
Subsequent magnifi cation 1x
6
Selecting fi lter cubes
Fig. 88
1
Focus control buttons
2
Open fi lter drawer
CUBE
The CUBE 1 to CUBE 6 (87.6) buttons
permit the direct selection of indi-
vidual fi lter cubes, provided the se-
lected cube is valid for the selected
method.
Press the CUBE 3 and CUBE 4 but-
tons at the same time to display the
assignments of the variable function
buttons. To reset the display, press
the buttons again or wait 3 seconds.
Focus buttons (Fig. 88) (DMI6000 B only)
Moves the Z drive in the indicated di-
rection.
SET + Z↑ Sets the upper focus stop.
SET + Z↓ Sets the lower stop.
1
Z↑
Z↓
2
1x
→
→
→
→
1
3
4
6
6
2
5
SmartMove remote control module
1
travel in x
2
Travel in y
3
Focus
4
Individual adjustment of button height
5 Variable function buttons (factory preset)
1
2
3
4
5
Shutter button + Cube buttons 1-6 (Leica
DMI4000 B and Leica DMI6000 B only)
The selected cube is moved to the
loading position for replacement.
„Load“ appears on the screen. After
pressing the button (88.2) the draw-
er is opened and the cube can be
changed. The next fi lter cube will be
moved to the loading position after
the drawer is closed.
7.5 The SmartMove Remote Control Module
SmartMove knobs
(Leica DMI4000 B and Leica DMI6000 B)
Use the knobs 89.1 and 89.2 to move the stage in
X and Y directions.
The image is focused using the knob 89.3 (Leica
DMI6000 B only).
The height of the knobs can be adjusted to a
comfortable working position by turning 89.4.
Variable function buttons on SmartMove
The variable function buttons are assigned func-
tions at the factory that are appropriate to the
features of your microscope. They are labeled
accordingly. For details on button assignments,
please refer to the included identifi cation sheet.
For information on the abbreviations used, please
refer to the list → p. 63.
Note:
The Leica Application Suite (LAS) software is re-
quired for changing the button assignments.
7.6 Illumination
7.6.1 Transmitted light
If your microscope has not yet been set up for
Koehler illumination, please continue with the
„Koehler Illumination“ section.
Leica DMI3000 B:
• If necessary, adjust the TL bright fi eld position
at the fi lter slider.
• Select an objective with moderate magnifi ca-
tion (10x–20x).
• Set the condenser to the bright fi eld position.
• Place a specimen on the stage.
• Focus on the specimen using the focus wheels.
• Adjust the light intensity.
• Close the fi eld diaphragm manually until the edge
of the diaphragm appears in the fi eld of view.
8.2 Contrast Methods
All of the contrast methods of the Leica DMI4000
B and Leica DMI6000 B can be selected and con-
trolled via the variable function buttons and the
Leica Application Suite (LAS). The only excep-
tions are methods that involve com ponents re-
quiring manual control (e.g. systems with manual
analyzers). The following section describes the
use of the function buttons on the stand. For in-
structions on the use of the software, please re-
fer to the separate manual.
Contrast methods for the Leica DMI3000 B are
controlled via the manual condenser, the manual
objective turret, as well as turning knobs and
sliders at the microscope.
8.2.1
Bright Field (TL)
Leica DMI3000 B:
• If necessary, adjust the TL bright fi eld position
at the fi lter slider.
• Set the condenser to the bright fi eld position.
• Remove all other optical components such
as analyzers, polarizers or IC prisms from the
beam path.
• Insert a transmitted light specimen.
• Select your objective
• Set the brightness at the light potentiometer
• Focus the image with the focus wheels.
Leica DMI4000 B and Leica DMI6000B:
Note:
If all positions of the fi lter turret are occupied, fi l-
ter cube „A“ can be swapped for fi lter cube „A-
TL“ using the Leica Application Suite (LAS/LAS
AF). TL contrast methods are possible with that
fi lter cube.
• Use the TL/IL function button to switch to
transmitted light (TL).
• Select the BF (bright fi eld) contrast method
by pressing the variable button BF.
Alternatively:
press
the
variable
button
CHANGE TL .
(For details on button assignments, please see
the identifi cation sheet.)
BF will appear on the LeicaScreen.
Motorized condensers will now move to the
bright fi eld position. Coded condensers must
be switched manually.
The fl uorescence fi lter turret will automati cally
go to an empty position or to the „A-TL“ fi lter
cube.
Fig. 108
Function buttons (left side of stand)
1
variable function buttons
2
Open/close aperture diaphragm
3
TL/IL switching
4
Open/close fi eld diaphragm
5
Increase/decrease light intensity
Fig. 109
Function buttons (right side of stand)
1
variable function buttons
5
3 4
2
1
1
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
• Focus the image with the knob on the Smart-
Move or the focusing wheel and adjust the in-
tensity with the INT function buttons.
Phase Contrast (TL)
(integrated phase contrast, see 8.2.6)
Leica DMI3000 B:
• If necessary, adjust the TL bright fi eld position
at the fi lter slider.
• Select a phase contrast objective.
• Select the suitable light ring on the condenser.
• Open the aperture of the condenser com-
pletely.
• Remove all other optical components such
as analyzers, polarizers or IC prisms from the
beam path.
• Insert a phase contrast specimen.
• Set the brightness at the light potentiometer
• Focus the image with the focus wheels.
Leica DMI4000 B and Leica DMI6000 B:
• Use the TL/IL function button to switch to
transmitted light (TL).
• Select the PH (phase contrast) contrast meth-
od
by pressing the variable button PH.
Alternatively:
press
the
variable
button
CHANGE TL .
(For details on button assignments, please see
the identifi cation sheet.)
PH will appear on the LeicaScreen.
Motorized condensers will now switch to the
correct light ring. Coded condensers must be
switched manually.
• Insert a transmitted light specimen.
• Rotate an appropriate objective into place.
Objectives that are suitable for phase contrast
are engraved with PH.
• Focus the image with the knob on the Smart-
Move or the focusing wheel and adjust the in-
tensity with the INT function buttons.
Note:
When selecting the phase contrast method, the
aperture diaphragm is opened fully and can not
be adjusted.
Leica DMI3000 B:
The fi lter slider (5a.9, S.24) is used to operate the
fl uorescence module.
• Pull the fi lter slider out completely to open the
beam path.
• Push the fi lter slider into the middle position
(1st detent) to bring the blue fi lter into the
beam path.
• Insert the fi lter slider fully in order to block the
beam path (shutter position).
• The fl uorescence illumination is controlled by
the rotary knob (5a.10, S.24).
• The fi lter cubes are swiveled manually into the
beam path by turning the incident light turret
disk.
Leica DMI4000 B and Leica DMI6000 B:
• Use the TL/IL function button to switch to fl uo-
rescence FLUO.
• Place a specimen on the stage and select a
suitable objective.
• The current fl uorescence fi lter cube will be
displayed on the LeicaScreen.
• You may protect your specimen from fading by
closing the incident light shutter.
To do so, press the SHUTTER button (87.3) on
the front panel.
The following pictogram will appear on th Lei-
caScreen:
�Changing the fl uorescence fi lter cube
�Fixed function buttons on the front panel:
CUBE 1 to CUBE 6 or Cube CCW
�Variable function buttons on the front panel
and SmartMove: CUBE CW or CUBE CCW
�Leica Application Suite (LAS) Software
• Focus the image with the knob on the Smart-
Move or the focusing wheel and adjust the in-
tensity with the INT function buttons.
• The intensity of the fl uorescence can be in-
creased by using the booster lens (Fig. 112) on
the left rear side of the stand (Fig. 113).
If bright fl uorescence is required in the center
of the fi eld of view, slide the booster lens into
the receptacle with the marking
facing the user. If a homogeneous distribution
over the entire fi eld of view is required, turn
the booster lens 180° so that the marking
is facing forward.
• For multiple fl uorescence, we recommend us-
ing the Excitation Manager and/or the ultrafast
internal fi lter wheel. Excitation wavelengths
can thus be changed in milliseconds. They are
controlled by the function buttons.
Fig. 112
Booster lens
Fig. 113
Booster lens in stand
1
•
1.4x
0.7x
8.4 Combination Methods
(Leica DMI4000 B and DMI6000 B)
Up to two combination methods are possible de-
pending on the features of the individual micro-
scope:
FLUO/PH and FLUO/DIC
• Select the combination method
by pressing the variable button COMBI .
Alternatively:
press
the
variable
button
CHANGE COMBI .
(For details on button assignments, please see
the identifi cation sheet.)
The content of the display changes accord-
ingly.
• Place a specimen on the stage and select a
suitable objective.
• Select the desired fi lter cube using the fi xed
function buttons on the front panel.
• The illumination settings for the fl uorescence
and transmitted light axes can be adjusted
separately.
• Toggle the illumination axes with the TL/IL
function button. The content of the LeicaS-
creen changes accordingly.
FLUO > DIC
The transmitted illumination is activated.
FLUO < DIC
The fl uorescence illumination is activated.
Note:
The manual analyzer (Fig. 110) must be used for
the FLUO/DIC method as described in Chapter
8.2.5, p. 88.
Leica DMI3000 B and Leica DMI4000 B:
The left-hand focus wheels can be used for both
coarse and fi ne focusing; the right-hand focus
wheel for fi ne focusing only (a version of the
Leica DMI3000 B with mirrored focus controls is
also available)
Leica DMI6000 B:
Note:
The parfocality teach-in has already been per-
formed at the factory. However, it may be neces-
sary to perform another teach-in after installing
the objectives when setting the microscope up.
We recommend checking parfocality before set-
ting the stops and performing a teach-in with the
Leica Application Suite (LAS) if necessary.
Focusing the image
The focusing is controlled by the knobs (116.3,
p. 101) on the SmartMove remote control module.
�
�
�
Alternatively, use the focus wheels on either side
of the stand.
The current Z position is shown on the LeicaS-
creen. In the case of motorized stages, the Z
drive will travel to its lowest position prior to the
stage initialization when switching the micro-
scope on.
The focus buttons Z↑ and Z↓ on the right side of
the stand (Fig. 114) permit fast focusing or lower-
ing of the objectives.
Setting stops
Set the lower focus stop by pressing and hold-
ing the SET button and pressing the Z↓ button as
well.
The display will show .
Pressing the button combination again will delete
the stop.
The display will show .
The lower focus stop can also be set using the
Leica Application Suite (LAS).
The lower stop is the same for all objectives and
can not be traversed.
In addition, a focus position that can not be tra-
versed can also be set.
To do so, press and hold the SET button and press
the Z button as well.
The display will show .
Pressing the button combination again will delete
the stop.
The display will show .
The focus position can also be set using the Leica
Application Suite (LAS).
Set the focus position for the dry objective at the
highest magnifi cation. The focus positions will
then be set automatically for all other objectives,
taking parfocality and working distances into ac-
count.
Abb. 114
1
Focus operating keys
2
Open fi lter drawer
1
2
�
�
�
�Set the stops via
�fi xed function buttons on stand
�Leica Application Suite (LAS) Software
Summary of pictograms
lower focus stop not set
lower focus stop set
focus position not set
focus position set
Going to the stops
Go to the lower stop by pressing and holding the
Z↓ button.
Go to the focus position by pressing and holding
the Z↑ button.
These functions can be assigned to variable
function buttons on the stand or SmartMove, or
they can be controlled via software.
�Go to stops via
�fi xed function buttons on stand
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
Note:
When going to the stops with the Z↑ and Z↓
buttons, hold the button until the stop has been
reached.
Setting the step increments
It is possible to toggle between Fine and Coarse
step increments.
The Fine value varies to suit the current objective. Suit-
able values have been predefi ned. The assignments
can be changed with the Leica Application Suite (LAS).
When selecting Coarse, the positioning speed is
the same for all objectives. Coarse corresponds
to the maximum speed.
Note:
The assignment of a specifi c step increment to
an objective not only applies to the Z drive, but
also to the step increments assigned to the stage
when Precise (→ p. 101) is selected.
�Switch between Fine and Coarse via
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
Only for DMI6000 B with AFC (Adaptive Focus
Control)
AFC actively holds a pre-defi ned focus position
over time. This feature is especially useful, if
e.g. during a time-lapse experiment at 37°C the
climate chamber has to be opened and a drop of
the temperature may occur.
Activate the AFC function, focus on your speci-
men either with the hand wheel on the micro-
scope stand or at the SmartMove and store the
current focus position as hold position.
AFC can be controlled either by.
• Variable function keys at the stand or SmartMove
• STP6000
• Software (Leica LASAF)
�
�
�
�
�
�
8.7 Port selection
Leica DMI 3000 B and Leica DMI4000 B:
Manual shifter rod activates and deactivates the
left-hand photo port.
VIS
LEFT
100%
0%
20%
80%
alternatively:
0%
100%
Leica DMI 6000 B:
The
button on the front control panel switches 100%
of the light to the eyepieces.
Use the
button, also on the front control panel, to select
the side ports.
8.6 Tubes
Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Adjusting the viewing distance
• Adjust the viewing distance of the eyepiec-
es so that a congruent total image is seen
(Fig. 115).
Adjusting the viewing angle
• Ergotubes feature a tilting binocular section for
a 30–45° viewing angle adjustment range.
Beam splitting in photo tubes
The beam splitting is set manually by pulling out
a control bar.
Observation
photo
100%
0%
0%
100%
alternatively
50%
50%
BL
activation of Bertrand lens*
�Light distribution via
�manual control bar
Fig. 115
Tube
setting
♦→
→
→
Depending on the confi guration, the screen will
now display
- the active port (right or left) and
- the percentage of light going to the port (100%,
80%, 50%).
Optional Leica DMI 6000 B:
The bottom port selection function can be as-
signed to one of the variable function buttons on
the stand or the SmartMove.
The top port can only be selected manually.
�Select ports via
�fi xed function buttons on stand (side ports)
�variable function buttons on stand and
SmartMove (bottom port)
�manual action (top port)
8.8 Eyepieces
Note:
The eyepiece’s aperture protector must be re-
moved or folded back, during microscopy while
wearing eyeglasses. We recommend removing
bifocals and spectacles with progressive-addi-
tion lenses when using the microscope.
• For the adjustable tubes with documentation
output, choose the 100% VIS position.
Eyepieces with inlaid reticle
• Focus the reticle by adjusting the eyelens.
• Focus on the object through this eyepiece.
• Then, close that eye and focus on the object
by adjusting only the second ocular.
Correction for Vision Problems
• With your right eye, look through the right eye-
piece and bring the specimen into sharp focus.
• Then, with your left eye, view the same speci-
men and rotate the left eyepiece tube until
the object is brought into sharp focus. Do not
change the Z position in the process!
Leica DMI3000 B and Leica DMI4000 B:
Select objectives manually with the objective tur-
ret.
The objective turret of the DMI4000 B is coded
so that the selected objective is shown on the
display.
Leica DMI6000 B:
The objectives can be selected with the func-
tion buttons on the stand or the SmartMove, or
by manually turning the objective turret. When
changing objectives manually, please ensure
that the turret clicks into position.
The positions of the objectives in the objective
turret have been specifi ed at the factory and
must be observed when installing the objectives.
(→ also see Objectives, p. 43).
When selecting an objective, the microscope au-
tomatically selects:
• the optimal setting for the fi eld diaphragm
• the optimal setting for the aperture diaphragm
• the light intensity for the current contrast
method
The objective magnifi cation and total magnifi ca-
tion are displayed on the LeicaScreen.
Note:
We recommend running a teach-in via the Leica
Application Suite (LAS) software when using
eyepieces not included in the scope of deliv-
ery. This will ensure that the total magnifi cation
shown in the LeicaScreen is correct.
• For immersion objectives use the appropriate
immersion medium.
OIL:
only use optical immersion oil
according to DIN/ISO standards.
Cleaning → p. 109.
W:
Water immersion.
IMM: Universal objective for water, glycerol,
oil immersion.
Caution!
Follow safety instructions for immersion oil!
Color coding of objectives
The magnifi cation of each objective is indicated by a color ring in accordance with DIN/ISO stand-
ards:
Immersion objectives are marked by an addition-
al, lower color ring.
black
oil or Imm (universal objective for
oil, water or glycerin)
white
water
orange
glycerin
The various engraved markings of the objectives
provide information on their applications:
black or
bright fi eld objectives,
dark blue
strain-free
green
phase contrast objectives,
strain-free
100x
63x
40x
25x
16x
10x
6.3x
4x
2.5x
1.6x
125x
50x
32x
20x
5x
150x
160x
white
dark-
light-
dark-
light-
yellow
orange
red
brown
gray
blue
blue
green
green
�Select objectives via
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
�Manual selection
Changing the operating modes
„dry“ ( DRY) and „immersion“ ( IMM)
Each objective is assigned to a specifi c objective
category:
1) Dry objectives (DRY)
2) Immersion objectives (IMM)
Note:
It is possible to use objectives for both operating
modes.
The mode can be assigned in the Leica Applica-
tion Suite (LAS).
Changing the operating mode
• First, select the operating mode (Imm or Dry)
using the function buttons.
The operating mode may also be selected in
the Leica Application Suite (LAS).
• The objective turret is lowered to its bottom
stop. This is to permit the application of the
immersion liquid when changing from a dry to
an immersion objective. It also permits the re-
moval of the liquid when changing to dry mode.
The current objective remains in the beam
path.
• Next, press the but on for the objective you in-
tend to use.
Note:
If the Imm or Dry operating mode buttons are
pressed accidentally, the original mode can be
restored by pressing the appropriate button.
�Change operating mode via
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
Note:
When replacing objectives, you must perform
a teach-in for the new objectives in the Leica
Application Suite (LAS). A parfocality teach-in
should also be performed.
Note:
For lockable immersion objectives lock these by
pushing the front part upwards until it stops (ap-
prox. 2 mm). Then, after a gentle turning motion
to the right, the objective is locked.
For objectives with corrective mounts turn the
knurl to adjust the objective to the thickness of
the cover glass.
8.10 Stages and Object Displacement
Leica DMI3000 B and Leica DMI4000 B:
The motorized stages are controlled via a sepa-
rate control unit.
Leica DMI6000 B:
Object displacement using SmartMove
The positioning of the stage is controlled by the
knobs (116.1, 116.2) on the SmartMove remote
control module.
Setting the step increments
The positioning speed of the stage can be varied
by switching between the Fast and Precise step
increments.
When selecting Fast, the positioning speed is the
same for all objectives.
The Precise speed varies to suit the current ob-
jective.
� Switch between Precise and Fast
via
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
Fig. 116
SmartMove remote control module
1
travel in x
2
Travel in y
3
Focus
4
Individual adjustment of button height
5 Variable function buttons (factory preset)
1
3
4
5
2
Storing and restoring stage positions
A variety of stage positions can be stored tempo-
rarily in the Leica Application Suite (LAS). The XY
position is stored, not the Z position.
In addition to a loading position (Load), 5 stage
positions can be set temporarily. When switch-
ing the microscope on, the stage will travel to a
previously-defi ned starting position.
�Temporarily store and
restore stage positions via
�Leica Application Suite (LAS) Software
8.11 Magnifi cation Changer
Leica DMI3000 B:
A mechanical magnifi cation changer can be
used optionally.
Magnifi cation factor: 1.6x.
A slider switches between 1x and the magnifi ca-
tion factor. The mechanical magnifi cation chang-
er affects the eyepieces and the top port.
Leica DMI4000 B and Leica DMI6000 B:
A mechanical magnifi cation changer can be
used optionally. The following magnifi cation fac-
tors are available: 1.5x, 1.6x and 2x
A slider switches between 1x and the magnifi ca-
tion factor.
The mechanical magnifi cation changer affects
the eyepieces and the top port.
The selected factor is shown in the LeicaDisplay
or the relevant window of the Leica Application
Suite (LAS) and taken into account when calcu-
lating the total magnifi cation.
Leica DMI4000 B and Leica DMI6000 B:
A motorized magnifi cation changer can be used
optionally. The following magnifi cation factors
may be selected: 1.5x, 1.6x, or 2x
The selected factor is displayed on the LeicaS-
creen and in the relevant fi eld of the Leica Ap-
plication Suite (LAS), and is taken into account
when calculating the total magnifi cation.
The motorized magnifi cation changer affects all
ports.
Pressing the left button (117.1) switches between
the possible magnifi cation factors; pressing the
right button selects the factor 1x.
Note:
a microscope can not have both types (manual
and motorized) of magnifi cation changers.
Fig. 117
Front control panel
1
Function buttons for magnifi cation changer
1
1
�Change magnifi cation via
�fi xed function buttons on stand
�Leica Application Suite (LAS) Software
�Adjust intensity via
�fi xed function buttons on stand
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS) Software
Fig. 118
Fixed function buttons, left side of stand
1
Aperture diaphragm
2 Transmitted light/incident light
3 Field diaphragm
4 Light intensity
1
2 3
4
8.12 Light sources
Leica DMI3000 B:
• Light intensity: the transmitted light intensity
can be adjusted continuously from 0 to 12V us-
ing the potentiometer at the lower left of the
front of the microscope stand.
• FLUO: The intensity can be adjusted in 5 fi xed
levels.
100% / 55% / 30% / 17% / 10%
( FIM=Fluorescence Intensity Manager)
If you continue turning to the left at 100% or to
the right after 10%, a 0% position (shutter posi-
tion) is set.
Leica DMI4000 B and Leica DMI6000 B:
• Adjust the intensity with the function buttons
(118.4). The INT function buttons are always
assigned to the currently active transmitted
light (TL) or incident light (IL) axis.
• For TL and IL:
The setting can be made in coarse and fi ne
steps. Pressing both INT (118.2) buttons as the
same time toggles between coarse and fi ne
adjustment. The light intensity displayed on the
LeicaScreen changes accordingly.
Coarse adjustment:
0–20
Fine adjustment:
0–255
• The intensity is individually adjusted and
stored for each objective and contrast method.
• FLUO: The intensity can be adjusted in 5 fi xed
levels.
100% / 55% / 30% / 17% / 10%
( FIM=Fluorescence Intensity Manager)
8.13 Aperture and Field Diaphragm
Leica DMI3000 B:
Transmitted light:
• The manual aperture diaphragm is adjusted on
the condenser.
• The manual fi eld diaphragm is adjusted on the
illumination arm.
Incident light:
• The fi eld diaphragm is set via the rotary knob
(5a.11, S. 24). Diaphragm apertures of different
sizes (round or angled) can be selected. (See
the labeling on the rotary knob.)
Round apertures are suitable for observations
through the eyepieces; angled apertures are
suitable for observations with CCD cameras.
Leica DMI4000 B and Leica DMI6000 B:
Both diaphragms have been set to suitable val-
ues for the current objective and contrast meth-
od at the factory.
The aperture diaphragm is controlled manually
when using the manual condenser.
The fi eld diaphragm is controlled manually when
using the manual illumination arm.
• The motorized diaphragms can be adjusted
at any time with the AP (aperture diaphragm)
(118.1) and FD ( fi eld diaphragm) (118.3) function
buttons. The values displayed on the Lei-
caScreen change accordingly.
The function buttons are assigned to the cur-
rently active transmitted light (TL) or incident
light (IL) axis.
Microscopes with structured illumination are not
equipped with a fi eld diaphragm.
Caution: The old values will be over-
written by the current ones!
Caution:
When using PH or DF, the aperture diaphragm is
fully open and can not be closed.
�Adjust diaphragms via
�fi xed function buttons on stand
�variable function buttons on stand and
SmartMove
�Leica Application Suite (LAS/LAS AF)
Software
9. Trouble Shooting
9. Trouble Shooting
Problem
Stand
The microscope does not respond.
Illumination
The image is completely dark.
The image is unevenly or not uniformly illumi-
nated.
The illumination fl ickers.
The lamp does not illuminate immediately upon
being switched on.
Cause/Remedy
�Ensure that the AC outlet has power.
�Ensure that the electronics box is connected to
an AC outlet.
�Check the cable connections.
�Inform Service and have the supply unit fuse
checked.
�Open the shutter (→ p. 68).
�Check the connections of the lamp housings
on the microscope (transmitted light/fl uores-
cence)
�Ensure that the lamps are connected to the
power supply and are not defective.
�Inform Service and have the ebq 100 supply
unit fuse checked.
�Remove all unneeded fi lters from the light
path.
�Center the lamp (→ p. 78ff)
�Replace the old lamp (→ p. 45, 49ff).
�Be sure that there is no loose connection at the
power supply.
�Replace the old lamp (→ p. 45, 49ff).
�The ebq 100 must be switched-on repeatedly.
�Hot Hg lamps should cool down before switch-
ing on again.
9. Trouble Shooting
Problem
Bright fi eld
The specimen can not be brought into focus.
Dark Field
No defi nite DF contrast is possible.
The image is unevenly or not uniformly illumi-
nated.
Undesirable stray light.
Cause/Remedy
� Use the correct immersion medium.
�Place the specimen on the stage with the cov-
erslip facing down.
�Make sure that the cover glass thickness is
correct and that it suits the indication on the
objective.
�Ensure that you are using an objective with
coverslip correction.
�Adjust the correction ring on the objective if
present.
�Be sure that a DF objective is being used.
�The objective aperture is too high:
maximum 0.7 for condenser S1
maximum 0.4 for condenser S23/28
If necessary, reduce the objective aperture us-
ing the iris diaphragm on the objective.
�Check the condenser centering.
�The magnifi cation is too weak.
Use a higher magnifi cation.
�Remove the condenser head or condenser
lenses.
�Clean the specimen and neighboring lenses
(→ p. 109f).
9. Trouble Shooting
Problem
Phase contrast
No phase contrast is possible.
Polarization
No polarization contrast is possible.
Transmitted light interference contrast
No transmitted light interference contrast is pos-
sible
Cause/Remedy
�The specimen is too thick.
�The cover glass is not placed evenly.
�Check the centering of the light rings
(→ p. 74).
�Bring the polarizer and analyzer into cross
position until they reach maximum darkness
(without specimen).
(→ p. 87).
�The specimen is too thick or too thin.
�Embedding medium or specimen are of bire-
fringent material. Rotate the specimen.
�The difference in the refractive indices of the
specimen and the embedding medium is too
small.
�The cover glass is too thick.
�Check the Koehler illumination (→ p. 70).
�Bring the polarizer and analyzer into cross
position until they reach maximum darkness
(without specimen).
(→ p. 87).
�Check whether the suitable condenser prism
and corresponding objective prism are select-
ed (manual alternative → p. 87).
�Ensure that the IC prisms are correctly seated
(→ p. 40).
9. Trouble Shooting
Problem
Fluorescence
The image is completely dark (no fl uorescence).
The fl uorescence is too weak.
LeicaScreen
Init Error!
Cause/Remedy
�Open the shutter (→ p. 68).
�Select the incident light axis (IL) (→ p. 66).
�Check your specimen, e.g. its antibody binding.
�Insert a new lamp (→ p. 45, p.49ff).
�Insert the booster (→ p. 92).
�Center the lamp (→ p. 78ff)
�Insert a new lamp (→ p. 45, p.49ff).
�Check the cable connections.
�Check whether the cover of the fi lter disk has
clicked into position.
�Check the installed objectives, fi lter cubes,
etc.
�Switch the microscope off and back on.
13. Abbreviations and Pictograms
Contrasting method
Magnifi cation
Illumination
Ports/Eyepiece
Focus
Lower focus stop not set
Lower focus stop set
Focus position not set
Focus position set
Shutter open
Shutter closed
Transmitted light fi lter
Field diaphragm, rectangular
Field diaphragm, round
Aperture diaphragm
Light distribution
13. Abbreviations and Pictograms
→
→
�
�
�
�
�
�
13. Abbreviations and Pictograms
AP
Aperture diaphragm
BF
Bright fi eld
COMBI
Combination method
CUBE
Fluo cube
DF
Dark fi eld incident/transmitted light
DIC
Differential Interference Contrast
FD
Field diaphragm
FLUO
Fluorescence axis (incident light)
ICR
Interference contrast, incident light
ICT
Interference contrast, transmitted light
IL
Incident light
INT
Intensity
IMC
Integrated modulation contrast
IPH
Integrated phase contrast
PH
Phase contrast
POL
Polarization, incident/transmitted light
TL
Transmitted light
Active ports 65
AFC Adaptive Focus Control 17, 62, 95
Allen key 30
Ambient temperature: 10
Analyzer 55, 56
Analyzer slot 21, 25
Aperture diaphragm 20, 66, 84, 104
Arc 80
Attachable mechanical stage 21, 25
Attachable mechanical stage for
fi xed micromanipul 34
Beam splitting 96
Bertrand lens 75, 96
Booster lens 20, 47, 92
Bright fi eld (TL) 106
C-mount 0.5x/0.63x 57
Camera 57
Centering key 30
Centering the mercury lamps 79
Centering window 21
Changing objectives 98
Cleaning glass surfaces 110
Cleaning objectives 110
Cleaning the stage 109
Coarse 95
Collector 50, 51
Color coding (objectives) 99
Combination Methods 93
Computer 64
Condenser 17, 38, 41, 55, 64
Condenser base 20
Condenser base S1-S28 38
Condenser cable 58
Condenser centering 72, 75
Condenser head 20, 42
Condenser head S1 38
Condenser head S28 38
Condenser prisms 39
Condenser tool 40
Consumable parts 111
Contrast settings 72
Contrasting method 14, 65, 83
Controls 18, 60
Correction for vision problems 97
Corrective mount 100
CUBE 68, 91
Dark Field (TL) 86, 106
Diaphragm 65
DIC module 32
DIC objective prism 32
DIC objective prism disk 20
DIC specimen 88
Differential Interference
Contrast (TL) 88
Digital camera 57
DIN VDE 8
Direct interface 58
Disposal 13
Drawer 21, 24
DRY 100
E-focus control buttons 21
ebq 100 supply unit* 59, 81
Electromagnetic compatibility 8
Electronics box 58, 64, 81
EU directive 8
EXT1-EXT4 sockets 30
Eyebase 96
Eyepiece tubes 20, 24
Eyepieces 20, 24, 43
Fast 101
Field diaphragm 20, 66, 73, 84, 104
Field diaphragm adjustment 73
Field diaphragm centering 21, 22, 26
Filter 44
Filter block 53
Filter cube 52, 68
Filter drawer 68, 94
Filter/refl ector cube 53
FIM 103
Fine 95
Fixed function buttons 62, 66, 103
Fixed micromanipulation stage 34
Fixed stage 35
Fixed stage (normal) 33
Fluorescence 91, 108
Fluorescence fi lter cubes 91
Fluorescence illumination 93
Focus 65
Focus control buttons 68, 94
Focus position 94
Focus stop 68
Focus wheel 20, 21, 24, 25
Focusing 94
Focusing telescope 74
Frequency 10
Front control panel 22, 23, 68, 102
Function button 63
Function button assignment 63, 67
Function buttons 67, 68
Fuses 10
Gas discharge lamps 81
General view 23
Glass insert 36
Graticule 97
Halogen lamp 12V 100W 45
Heating Insert P 36
Height compensation plate 112
Hg 100W lamp 48
Hg mercury burner 47
High-pressure mercury burner 49
High-pressure mercury burner 100W 49
High-pressure xenon burner 75 W 49
IC prisms 32
ICT condenser prisms 40
Illumination 65, 105
Image centering 72
IMM 100
Immersion objectives 99, 100
Immersion oil 111
Incident light - fl uorescence. 73
Incident light axis 14
Incident light lamp housing 21
Incident light turret disk 52
Inserting discharge lamps 49
Inserting the lamp 46
Inserts for attachable mechanical
stage 35
Installation location 28
Intelligent automation 60
Intensity control 103
Interfaces 18
Intermediate pupil interface 20
Knobs 62, 69
Koehler Illumination 70, 72
Lamp bases 49
Lamp housing 106 z 46, 48, 50, 78
Lamp housing 106 z L 48
Lamp housing 107 L 78
Lamp housing 107 or 107/2 45
Lamp housing mount 46
Lamp mount 20
Lamp power cable 58
Lamp replacement 44
Leica CTR6000 electronics box 10, 18
LeicaScreen
20, 22, 61, 64, 65, 82, 108
Light guide 12
Light intensity 20, 66
Light rings 38, 64
Light source adjustment 78
Light sources 103
Load 53, 69
Lower focus stop 68, 94
Magnifi cation 65
Magnifi cation changer 15, 68, 102
Manual method 87
Mechanical 3-plate stage 33
Medical instrument 8
Micromanipulation stage with
attachable mechanical 33
Motorized 3-plate or scanning
stages 37
Motorized method 87
Motorized polarizer 77
Multiple fl uorescence 92
Object displacement 101
Objective turret 16, 21, 24, 43
Objectives 65, 98
Observation ports 17, 18
Opener for drawer 21
Operating mode 100
Operating temperature 11
Overvoltage category: 10
Parfocality compensation 43
PCI card (PC) 57
Phase contrast 107
Phase contrast centering procedure 74
Phase contrast rings 74
Phase rings 39, 64
Phillips screwdriver 30
Pictograms 113
Polarization 107
Polarization (TL) 87
Polarizer 55, 75
Polarizer holder 55
Pollution degree 10
Port 96
Port selection 96
Port switching 22, 26
Power input 10
Power supply 59
Precise 101
Protective gloves 48
Protective mask 48
Refl ector cube 78
Relative humidity 10
Removing Immersion Oil 110
Replacement eyecup 111
Replacement lamp 111
Replacement of incident-light lamps 48
Reset function 61, 82
Right side port 20, 21, 24
RS232 ports 58
Safety class 9
Safety regulations 9
Screw lengths 34
Setting stops 94
Shutter 68, 91
Side-Port 21
SmartMove 23, 60, 62
SmartMove remote control module 23
Software 62
Software tools 19
Spare parts 111
Specifi cations 10
Stage 21, 25
Stage positions 101
Stages 16, 33, 101
Stand 105
Stand package 27
Step increments 95, 101
Stops 65
Supply unit 10, 50, 51, 59
Supply voltage: 10
System package 27
Three-plate micromanipulation
stage 33
TL/IL switching 20
Toggling transmitted light/incident
light 66
Top-Port 20, 22, 24, 26
Transmitted illumination 93
Transmitted Light Axis 14
Transmitted light
interference contrast 107
Transmitted light lamp housing 24
Transmitted-light 69
Transmitted-light fi lter 22, 26
Transmitted-light
illumination carrier 31
Transmitted-light lamp housing 20, 44
Transmitted-light specimen 84, 85
Transport 29
Tube 15
Tube setting 96
Turret disk 53
USB 58
Useful life 49
Variable function buttons
20, 21, 23, 62, 67
Variable function buttons on
SmartMove 69
Viewing angle 96
Xe 75 burner 50
XYZ-control 58
Z focus 17