Fluke 6062A Instruction Manual
Read Online
Read the Fluke 6062A Instruction Manual online using the built-in PDF viewer.
Text of the manual
Notwithstanding any provision of any agreement the following warranty
is exclusive:
The JOHN FLUKE MFG. CO., INC,, warrants each instrument
it manufactures to be free from defects in materia! and
workmanship under normal use and service for the period of
1 -yearfromdateof purchase, This warranty extends only
to the original purchaser. This warranty shall not apply to fuses, disposable batteries (rechargeable type batteries are
warranted for 90-days), or any product or parts which have been subject to misuse, neglect, accident, or abnormal
conditions of operations.
In the event of failure of a product covered by this warranty, John Fluke Mfg, Co., inc,, will repair and calibrate an
instrument returned to an authorized Service Facility within
1 year of the original purchase; provided the warrantor's
examination discloses to
its satisfaction that the product was defective. The warrantor may, at its optioh,- replace the
product in lieu of repair. With regard to any instrument returned within
1 year of the original purchase, said repairs or
replacement will be made without charge.
If the failure has been caused by misuse, neglect, accident, or abnormal
conditions of operations, repairs will be billed at a nominal cost. In such case, an estimate wilt be submitted before
work
is started,
if requested.
THE FOREGOING WARRANTY
IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS,
OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. JOHN FLUKE MFG, CO., INC., SHALL
NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN
CONTRACT, TORT, OR OTHERWISE,
It any failure occurs, the following steps should be taken;
1.
Notify the JOHN FLUKE MFG. CO., INC., or nearest Service facility, giving full details of the difficulty, and
include the model number, type number, and serial number. On receipt of this information, service data, or
shipping instructions will be forwarded to you,
2.
On receipt of the shipping
instructions, forward the instrument, transportation prepaid. Repairs will be
made at the Service Facility and the instrument returned, transportation prepaid.
SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT
All shipments of JOHN FLUKE MFG. CO.. INC., instruments should be made via United Parcel Serviceor "Best Way"'
prepaid. The instrument should be shipped
in the original packing carton; or
if
if
is not available, use any suitable
container that is rigid and of adequate size, if a substitute container is used, the instrument should be wrapped in paper
and surrounded with at least four inches of excelsior or similar shock-absorbing material.
CLAIM FOR DAMAGE IN SHIPMENT TO ORIGINAL PURCHASER
The instrument should be thoroughly inspected immediately upon original delivery to purchaser. Ail material in the
container should bechecked against the enclosed packing list. The manufacturer will not be responsible for shortages
against the packing sheet unless notified immediately.
If the instrument is damaged in any way, aciaim should be filed
with the carrier immediately, (To obtain a quotation to repair shipment damage, contact the nearest Fluke Technical
Center,) Final claim and negotiations with the carrier must be completed by the customer.
The JOHN FLUKE MFG. CO,, INC, will be happy to answer all applications or use questions, which will enhance your
use of this instrument, Please address your requests or correspondence to; JOHN FLUKE MFG. CO,, INC., P.O. BOX
C9090, EVERETT, WASHINGTON 98206, ATTN: Sales Dept, For European Customers: Fluke (Holland) B.V., P.O, Box
5053, 5004 EB, Tilburg, The Netherlands,
•For European customers, Air Freight prepaid,
John Fluke Mfg. Co., Inc., P.O. Box C9090, Everett, Washington 98206
Rev. 6/81
The 6062A Synthesized RF Signal Generator (referred to as the Generator or instrument) is a
fully-programmable, precision, synthesized signal generator. The Generator is designed for
applications that require good modulation, frequency accuracy, and output level performance
with moderate spectral purity. It is well suited for testing a wide variety of RF components and
systems including filters, amplifiers, mixers, and radios, particularly on-channel radio testing.
The shipping container should
include
a 6062A Synthesized RF
Signal
Generator, an
Operator Information Card, a Getting Started Manual, an Instruction Manual, a line power
cord and two BNC dust caps. Any accessories ordered for the Generator are shipped in a
separate container.
Section 2, Installation and Operation, gives instructions on inspecting the new Generator and
explains what to do if the instrument arrives with shipping damage. Reshipment information is
also included.
1-3.
SAFETY
This instruction manual contains information, warnings, and cautions that should be followed
to ensure safe operation and to maintain the Generator in a safe condition.
The Generator is designed primarily for indoor use, and it may be operated in temperatures
from 0 to 50°C without degradation of its safety.
TO AVOID ELECTRIC SHOCK, USE A POWER CORD THAT HAS A
THREE-PRONG PLUG. IF A PROPER POWER CORD IS NOT USED, THE
6062A CASE CAN DEVELOP AN ELECTRICAL POTENTIAL ABOVE EARTH
GROUND.
CAUTION
To avoid damage to the 6062A, check that the rear panel
line voltage
selection card and fuse are correct for the line voltage in your area. The
correct line voltage and fuse combinations are:
100/120V ac, ±10%, 47 to 63 Hz
220/240V ac, ±10%, 47 to 63 Hz
1.5 amp
.75 amp
1-1
The Operator Information Card has an adhesive backing so it can be affixed to the top of the
Generator in bench applications or to the operator console in remote applications. A copy of
the card
is located at the end of Section 8 of this manual as a convenient reference or for
duplication.
1-5.
GENERATOR DESCRIPTION
Fundamental features of the Generator are as follows:
®
0.10- to 1050-MHz frequency range in 10-Hz steps
•
1050- to 2100-MHz frequency range in 20-Hz steps
®
+16 to -137-dBm level range in 0.!-dB steps from 0.10 to 1050 MHz
•
+13 to -137-dBm level range in 0.1-dB steps from 1050 to 2100 MHz
•
AM, FM, 0M, and Pulse, internal or external
•
Internal 400- and 1000-Hz modulation oscillator
•
Relative frequency and amplitude
®
Volts/dBm conversion
®
Store/ recall memory
®
Instrument (or controller) prompted software compensation
•
Masler/slave
for
frequency,
amplitude,
and
modulation
functions
(IEEE-488
Interface controlled.)
®
Fluorescent display
®
5’/4-inches high, rack mountable
1~6.
Controller Functions
The Generator’s microprocessor controls
all operator interface functions, performs back-
ground operations such as status checks, and updates (strobes) the front panel displays.
Whether the operator is using local control with the front panel, or remote control with the
IEEE-488 Interface, the microprocessor provides self test and diagnostic capability. Economi-
cal instrument performance
is achieved by using software compensation and
accuracy-
enhancement circuitry.
1-7.
LOCAL CONTROL
The value of the basic output parameters of the Generator (i.e., amplitude, frequency, or
modulation) can be controlled in three ways:
®
Direct numeric entry
®
Incrementing or decrementing the bright digit
®
Step-up or step-down entry where the operator can program the step size via the front
Other controls provide selection of the modulation control, internai/external frequency
reference, instrument status, and special function modes. Refer to Tables 2-1 and 2-2 for a
more complete discussion.
1-8.
DISPLAY FIELD
The programmed values of modulation, frequency, and amplitude are displayed in the three
display fields. In addition, when the operator presses the [STATUS] key, uncal and reject entry
codes are displayed. The display also prompts the operator during the software compensation
procedures.
1-9.
REMOTE-CONTROL PROGRAMMING
The IEEE-488 Interface allows the Generator to be remotely controlled with any IEEE-488
bus controller. The operator can also use the instrument on the IEEE-488 bus without a
controller in a listen-only or talk-only mode by selecting the appropriate Generator rear panel
IEEE-488 switch settings.
All instrument controls can be controlled remotely except the POWER ON / OFF and the rear
panel REF INT/EXT switches. The IEEE-488 Interface provides additional commands not
available with local control, such as data transfer and individual control of internal I/O
control bits.
The IEEE-488 Interface allows two Generators to track amplitude, frequency, or modulation
in a master/ slave configuration when the operator uses the front panel step-up and step-down
entries on one of the instruments. For instance, frequency tracking is convenient for tests
involving mixers, and amplitude tracking is useful for two-tone intermodulation testing.
The specified frequency range
is 0.10 to 2100 MHz. The frequency
is synthesized from a
10-MHz reference and provides an output resolution of 10 Hz from 0.10 to 1050 MHz, and 20
Hz from 1050 to 2100 MHz. The relative frequency mode allows the operator to program the
frequency in relation to a center frequency or an offset frequency. This is convenient for testing
filters and mixers. The output frequency stability and accuracy depends on the reference,
whether that reference is internal or external.
1-11.
Reference
The internal frequency reference is a lO-MHz ambient crystal oscillator. If Option -130 High
Stability Reference or Option -132 Medium Stability Reference
is
installed,
the internal
oscillator is locked to the option oscillator.
With the REF INT/EXT switch set to EXT, the Generator can be operated from an external
I-, 2-, 2.5-, 5-, or 10-MHz, 0.3 to 4V peak-to-peak sine or square-wave reference applied to the
REF IN connector. In either position of the REF INT/EXT switch, the selected reference is
available as a 10-MHz signal at the rear panel 10 MHz OUT connector.
The Generator has two specified signal level ranges:
+ 16 to -137 dBm (I.4IV to 0.1 fjY) over a frequency range of 0.10 to 1050 MHz
+ 13 to -137 dBm (IV to 0.033 /jY) over a frequency range of 1050 to 2100 MHz.
The programming limits are +17 and -147.4 dBm, which correspond to limits of 1.58V to 0.01
fjY. The level entry can be in dBm or volts, or it can be converted from one to the other. The
relative amplitude mode allows the operator to account for cascaded gain or loss by displaying
the level relative to the initial setting.
1-13.
Modulation
The Generator has internal and external amplitude, frequency, phase, and pulse modulation
capabilities. The internal modulation oscillator
is selectable between 400 and lOOO Hz. AM
depths ofO to 99% are available in l% steps. FM deviation ranges of I, lO, I00,and 400 kHz are
available in steps of
i,
!0, lOOHz, and
I kHz, respectively. 0M deviation ranges ofO.lO,
l.O,
lO.O, and
40.0 radians (rad)
are
available
in
steps
of O.OOl,
O.OI,
O.l, and
O.l
radians,
respectively. The internal Pulse duty cycle is approximately 50%. External Pulse allows rates
from dc up to 16 M Hz, and duty cycles from 0 to 100%.
1-14.
OPTIONS AND ACCESSORIES
The following options are available for the Generator:
Option -130 High-Stability (Ovened) Reference
Option -132 Medium-Stability Reference
Option -651 Low-Rate FM (External only)
Option -830 Rear Panel RF Output, MOD Input, and Pulse MOD Input
Section 6 provides more detailed information on the options.
The following accessories are included with each Generator:
QUANTITY
Operator Information Card
Getting Started Manual
Instruction Manual
Line Power Cord
BNC Dust Cap
797928
1
794859
1
794842
I
284174
I
478982
2
The following accessories are available for the Generator:
DESCRIPTION
ACCESSORY
NO.
Rack Mount Kit. Includes M05-205-600 (5'/4-inch
Y6001
Rack Mount Ears) and MOO-280-610 (24-inch Rack Slides)
IEEE-488 Shielded Cable,
1 meter
Y8021
IEEE-488 Shielded Cable, 2 meters
Y8022
IEEE-488 Shielded Cable, 4 meters
Y8023
Coaxial Cable, 50 ohms, 3 feet, BNC (ra) both ends
Y9
!
1
Coaxial Cable, 50 ohms, 6 feet, BNC (m) both ends
Y91 12
1-15.
RECOMMENDED TEST EQUIPMENT
The test equipment recommended for the performance
tests, calibration adjustments, and
troubleshooting are listed in Table 4A-I. This equipment is assumed to be calibrated to the
manufacturer’s specifications. If the recommended test equipment is not available, equivalent
test equipment can be substituted.
NTRODUCTION AND SPECIFICATIONS
1-16.
MNEMONICS
The mnemonics used in the schematics, block diagrams, wiring diagrams, truth tables, and the
text are listed in Figure 8-1.
1-17.
SIGNAL GENERATOR SPECIFICATIONS
Unless
otherwise
noted,
the
following
performance
is
guaranteed
over
the
specified
environmental and ac power line conditions 20 minutes
after turn-on. Table
1-1
lists the
Generator specifications.
Table 1-1. 6062A Synthesized RF Signal Generator Specifications
Warranted performance^
20 minutes after
power-on within operating temperature
range.
FREQUENCY
(9 1/2-Digit
Display)
RANGE
0.1
to 2100.0
MHz
in 4
bands;
0.1
to 244.99999
MHz,
245
to 511.99999
MHz,
512
to 1049.99999 MHz,
1050
to 2100.0
MHz.
RESOLUTION
...
10 Hz
from
0.1
to 1050 MHz,
20 Hz
from 1050
to 2100
Itiz.
ACCURACY
Same as
reference
(See
REFERENCE).
REFERENCE
(Internal)
The
unit
operates
on an
internal
free-air 10-MHz crystal
oscillator
with
an ageing
rate
of
0 dBm,
terminated
in 50-ohms.
REFERENCE
(External)
Accepts
1-,
2-,
2.5-,
5-,
or 10-MHz
signal.
Level
of
0.3
to 4.0
Vpp
into 50-ohms termination.
AMPLITUDE
(3 1/2-Digit
Display)
RANGE
(Indicated)
+16
(+16 peak with
AM
enabled)
to -137
dBm,
from
0.1
to 1049.99999
MHz.
+13
(+13 peak with
AM enabled)
to -137
dBm,
from
1050
to 2100
MHz.
(Autoranging 6-dB
step attenuator).
RESOLUTION
0.1
dB
(<1%
or
1
nV
in
volts).
Annunciators
for
dB,
dBm,
V,
mV,
uV,
dB
mV,
and dB
uV.
ACCURACY
+1
dB
from +16
to -127 dBm
from
?
to 1049,99999
MHz.
*
+1.5
dB
from +13
to -127
dBm from
?050
to 2100 mz.
+2.0 dB
from +16
to -127 dBm
from
0.1
to .99999 MHz.
* +1.5
dB
at
temperatures
other
than
23°C +5°C
1-5
NTRODUCTION AND SPECIFICATION
Table 1 -1 . 6062A Synthesized RF Signal Generator Specifications (cont)
SOURCE
VSWR
<1,5:1
below
+1
dBm,
<2.0:1
elsewhere.
SPECTRAL
PURITY
(CW ONLY)
SPURIOUS
<~60 dBc
for offsets
greater
than 10
kHz
and
frequencies
fr«n
0.1
to
1049.99999 MHz.
<-54
dBc
for
offsets
greater
than
10 kHz
and
frequencies
from
1050
to 2100 Wz.
NOTE
Fixed frequency
spurs are <-60 dBc
or
<-140 dBm whichever
is
larger.
NOTE
dBc
refers
to decibels
relative to the
carrier frequency,
or
in this
case,
relative
to the
signal
level.
HARMONICS
<-30 dBc
for
levels
< +13 dBm and
frequencies
>1
MHz,
<-25 dBc
elsewhere.
SUBHARMONICS
<-50 dBc
for
output
frequencies
from
1050
to 2100 mz.
RESIDUAL
FM
(rms
in
0.3-
to 3-kHz band)
<12 Hz
for
0.1
to 244.99999
MHz,
<6 Hz
for
245
to 511 .99999
l^z,
<12 Hz
for
512 to 1049.99999 MHz,
<24 Hz
for
1050
to 2100
MHz.
RESIDUAL
FM
(rms
in
0.05-
to 15-kHz band)
<18 Hz
for
0.1
to 244.99999
Miz,
<9 Hz
for
245
to 511.99999
MHz,
<18 Hz
for 512 to 1049.99999
MHz,
<36 Hz
for
1050
to 2100 Miz.
RESIDUAL
AM
(in
0.05-
to 15-kHz
Band)
<-60 dBc.
AMPLITUDE MODULATION
(2-Digit
Display)
DEPTH
RANGE
0
to 99%.
RESOLUTION
U.
ACCURACY
±(2%
+ 4% of
setting)
to peak
amplitude of +13 dBm
for
frequencies
to 2100 MHz.
+(3%
+ 5% of
setting)
to peak
amplitude of +13 dBm
for
frequencies
of
0.1
to .99999 MHz.
1-1. 6062A Synthesized RF Signal Generator Specifications (con!)
<1,5%
total
harmonic distortion
(THD)
to 3CK
AM,
<3%
THD
to
70%
AM,
<5%
THD
to 90%
AM for frequencies
of
1
to
1049.99999 MHz
and peak
amplitude
<
+13
dBm.
<3%
THD to 70%
AM,
and <5% THD
to 90%
AM
for frequencies
of
1050
to 2100
MHz
and peak amplitude
< +13 dBm.
<3%
THD to 30%
AM,
<5% THD to 70%
AM,
and <7% THD to 90%
AM for
frequencies
of
0.1
to .99999 MHz
and
peak
amplitude
< +13
dBm.
BANDWIDTH
(3
dB)
20 Hz
to 50
kHz.
DC
to 50
kHz by
special
function.
(Note— valid
for
RF frequency - Mod
frequency >150
kHz)
INCIDENTAL
FM
<0.3
fm
for
internal
rates,
30%
or
less
AM,
and frequencies
from
0.1
to
1049.99999
MHz.
<0.6
fm
for
internal
rates,
30%
or
less
AM,
and
frequencies
from 1050
to
2100 mz.
FREQUENCY MODULATION
(3-Digit
Display)
DEVIATION
RANGES
MAXIMUM DEVIATION
Mod
Rate
fm
>,2
kHz
fm <,E
kHz
...
0
to 999 Hz,
1
to 9.99 kHz,
10 to 99.9
kHz,
and
100 to 400 kHz.
Max Dev
RF
Frequency
400
kHz
200
kHz
100
kHz
200
kHz
1050
to 2100
Mhz
512
to 1049.99999 MHz
245
to 511.99999 Mhz
0.1
to 244.99999 Mhz
lower
of
above
or;
2
fm
fo
245
to 2100 MHz
2
fm(fo+800)
0.1
to 244.99999 MHz
Where:
fm
=
mod
frequency
in
kHz
fo ®
RF frequency
in MHz
Deviation
is
in
kHz
Specs apply
where:
RF
Frequency
-
Dev >150
kHz
RF
Frequency
“
Mod Rate >150
kHz
RESOLUTION
3
digits.
ACCURACY
+
(7%
+
10 Hz)
for
rates
of 0,3
to
20
kHz.
Does
not
include effects
of
Residual
FM.
1-7
include effects of
Residual
FM.
BANDWIDTH
C3
dB)
20 Hz
to 100
kHz.
(Note--
valid
for
RF
Frequency
-
Mod
Frequency >150
kHz)
INCIDENTAL
AM
<1% AM at
1
kHz
rate,
for
the
maxiinum
deviation or
50 kHz,
whichever
is
less.
Valid
for
RF Frequency
>1
MHz.
PHASE MODULATION
(3 digit display)
DEVIATION
RANGES
MAXIMUM DEVIATION
RESOLUTION
ACCURACY
0
to .099
rad,
.100
to .999
rad,
1.00
to 9.99
rad,
and
10.0
to 40.0
rad.
20
rad from
0.1
to 244.99999
MHz.
10
rad
from
245
to 511 .99999
MHz.
20
rad
from 512
to 1049.99999 MHz
40
rad
from 1050
to 2100
MHz.
3 digits
+(7%
+
.01
rad)
at
1
kHz
rate.
Does
not
include effects
of
Residual
FM.
DISTORTION
BANDWIDTH
(3
dB)
....
........
INCIDENTAL
AM
<1%
THD
for
1
kHz
rate.
Does
not
include
effects
of
Residual
FM.
20 Hz
to 10 kHz.
(Note—
valid
for
RF
Frequency
-
Mod
Frequency
>150
kHz)
<1%
AM at
1
kHz
rate.
PULSE MODULATION
(RF
Frequencies
from 10
to 2100 MHz)
ON/OFF
RATIO
80
dB minimum
RISE
& FALL TIMES
<15
nsec
LEVEL ERROR
for
pulse widths >50 nsec,
power
in
the pulse within +0.5
dB
of measured
CW
level.
DUTY CYCLE
(ext mod)
0-100%
REP RATE
(ext mod)
DC-16 MHz
INTERNAL MODULATION
internal
rates,
approx
50%
duty
cycle.
EXTERNAL
PULSE MODULATION
The pulse
input
is TTL
compatible and
50 ohm terminated with
an
internal
active pull-up.
It
can be modeled as
1.2
V
in series with
50 ohms
at
the
pulse modulation
input
connector.
The
instrument
senses
input
terminal
voltage and turns the
RF off
when the
terminal
voltage
drops below
1
+
0.1
V.
Max allowable applied voltage, +10V.
PULSE MODULATION
(RF
Frequencies
<10 i^z)
RISE
& FALL
TIMES..
<2
X period of
RF
Frequency.
NTRODUCTION AND SPECIFICATIONS
Table 1-1. 6062A Synthesized RF Signal Generator Specifications (cent)
LEVEL
ERROR....
.....for
pulse
widths >10
X period
of
RF
Frequency,
power
in
the
pulse
will
be within +0.5 dB
of
the measured
CW
Level.
Other
specifications are
the
same as
for
the 10
to 2100 Wz
range.
NON-VOLATILE MEMORY
50
instrument
states
are
retained for
typicaUy
2
years,
even with
the power mains disconnected.
REVERSE
POWER PROTECTION
PROTECTION LEVEL
up to 25
watts
from
a 50
ohm
source.
up to 25
VDC.
Instrument
output
is
AC
coupled.
TRIP/RESET
Flashing
RF OFF annunciator
indicates
a
tripped condition.
Pushing
RF ON/OFF
button will
reset
instrument.
Protection
is provided when
instrument
is off.
IEEE-488
INTERFACE
FUNCTIONS
SH1,
AH1,
T5,
TEO,
L3,
LEO,
SR1,
RLl
,
PPO,
DC1,
0T1,
CO,
and
El
,
MODULATION
SOURCE
INTERNAL
0.4
or
1
kHz,
+3%
for
20
to 30°C;
add +0.1%/°C outside
this
range.
EXTERNAL
+5V max.;
IV peak provides
indicated
modulation
index.
Nominal
input
impedance
is 600 ohms.
TODES
Any combination
of
AM and
FM
or 0M,
internal
or
external,
may be
used-
If
external
AM and
FM
or 0M are
enabled,
the modulation
input
Z
will
drop
to approximately
560
ohms.
Pulse
modulation
is
completely
independent
and
can be used
in conjunction with
any
other
form(s)
of modulation.
GENERAL
TEMPERATURE
Operating
0
to 50®C
(32
to 122°F).
Non-Operating
-40
to 75°C
(-40
to 167°F).
HUMIDITY
RANGE
Operating
95%
to 30°C,
75%
to 40°C,
45%
to 50°C.
ALTITUDE
Operating
Up to 10,000
ft.
VIBRATION
Non-Operating
..............
5
to
15
Hz
at 0.06
inch,
15
to 25 Hz
at 0.04
inch,
and
25
to 55 Hz
at 0.02
inch,
double amplitude
(DA).
1-9
Table 1-1. 6062A Synthesized RF Signal Generator SpecHications (cont)
SHOCK
Non-Operating
Bench handling per
MIL
T 28^0C
Class
5,
Style
E.
ELECTROMAGNETIC
COMPATIBILITY
The
radiated emissions
induce <3 uV
(<1
uV
of
the Generator's
output
signal)
into
a 1-inch
diameter/
2-turn
Loop/
1-inch
from
any
surface as
measured
into
a 50-ohm
receiver.
Also
complies
with
the following
standards;
CE03
of MIL-STD-461B
(Power and
interconnecting
Leads)/
0.015
to 50
MHz.
RE02
of MIL-ST0-461B
(14
kHz
to 10 GHz).
FCC
Part
15
(j),
class
A.
CISPR
11.
SIZE
Width
Height
Depth
43
cm
13.3
cm
50.8
cm
17
in
5.25
in
20
in
POWER
» •
a a B
100/
120/
220,
240V
ac +10%, 47
to
63 HZ/
<180
VA
(<15
VA/
with Option
-130
installed/
and
the Generator
turned off
(standby).
WEIGHT
<15.7
kg
(35
lbs).
OPTION -130 HIGH-STABILITY
REFERENCE
AGING
RATE
<+5
x
10 -10/day,
after
21
days
continuous operation.
TEMPERATURE STABILITY
<+2
x
10 -10/°C
(Oven
remains powered
in
standby).
OPTION -132 MEDIUM-STABILITY
REFERENCE
AGING
RATE
<+1
x
10”^/month
after
5 days
continuous operation.
TEMPERATURE STABILITY
<+1
x
10"^/°C
(0 to 50°C)
(no
powered standby).
OPTION -651
LOW-RATE EXTERNAL
FH
MAXIMUM DEVIATION
9.99
kHz.
DROOP
<15% on
a
10-Hz
square wave.
BANDWIDTH
(3dB)
0.5
Hz
to 100 kHz
(typical).
MAX DC
INPUT
+10 mV.
.
<1%
AM at
1-kHz
rate
and deviation <10
kHz.
INCIDENTAL
AM
. .
.
Table 1-1. 6062A Synthesized RF Signal Generator Specifications (cent)
SUPPLEMENTAL
CHARACTERISTICS
The
following characteristics are provided to assist
in
the application
of
the Generator
and
to describe
the
typical
performance
that
can be
expected.
FREQUENCY
SWITCHING
SPEED
...
<100 mS
to be within 100 Hz
AMPLITUDE SWITCHING
SPEED
.... ...
<100 mS
to be
within
0.1
dB
AMPLITUDE RANGE
Programmable
from +17
to -147. A
dBm.
Fixed-range,
selected
by
special
function,
allows
for more
than 12 dB
of vernier without
switching
the
attenuator
.
AMPLITUDE ACCURACY
+2.0 dB
from -127.1
to -137 dBm
from
6.1
to 2100
MHz.
NOISE
(at 20-kHz
offset)
<--116 dBc/Hz
from 0.10
to
. 244.99999
f^z.
<"-122 dBc/Hz
from 245
to 511 .99999
iMHz.
<--116 dBc/Hz
from 512
to 1049.99999
MHz.
<--110 dBc/Hz
from 1050
to 2100 mz.
RESIDUAL
FM
(typical)
Freq
Range
.3
0.1
- 244.99999 MHz
8
245
- 511.99999 MHz
4
512
- 1049.99999 MHz
8
1050
- 2100 mz
16
3
kHz
.05
-
15
kHz
CCIT
Hz
12 Hz
7 Hz
Hz
6 Hz
3..5
Hz
Hz
12 Hz
7 Hz
Hz
24 Hz
14 Hz
EXTERNAL MODULATION
.
IEEE
.
FREQUENCY
DRIFT
PULSE MODULATION
Annunciators
indicate when
a IV peak
signal
is applied,
+2%,
over
a 0.02-
to 100-kHz band.
All
controls except
the power
switch
and
the
internal/external
reference
switch are
remotely programmable via
IEEE-488 Interface
(Std 488-1978).
All
status
including
the
option
complenent
are available remotely.
The
Store/Recall
memory data may be
transferred via
an
external
controller.
In talk-only,
the
appropriate
commands are generated
when the
front
panel
step-up and
step-down entries are made
to control
another 6062A,
6060A, 6060B ,6070A,
or
6071 A.
(6070/71 A only
have
FREQUENCY
STEP.)
<1
ppin/hr after
1
hour warmup at
constant
ambient
tanperature using
internal
free-air
crystal.
80
nsec
typ
65
nsec
typ
PULSE
DELAY
OFF/ON
section describes how
to
install and
operate
the Generator. This
section contains
information for an initial inspection, setting up the instrument, general operating information,
and local and remote operation.
2-2.
INITIAL INSPECTION
The Generator is shipped in a special protective container that should prevent damage during
shipment. Check the shipping order against the contents of the container, and report any
damage or short shipment to the place of purchase or the nearest Fluke Technical Service
Center. Instructions for inspection and claims are included on the shipping container.
If reshipment of the Generator is necessary, please use the original shipping container. If the
original container is not available, use a container that provides adequate protection during
shipment,
it
is recommended that the Generator be protected by
at least three inches of
shock-absorbing material on all sides of the container. Do not use loose fill to pad the shipping
container. Loose fill allows the Generator to settle to one corner of the shipping container,
which could result in the Generator being damaged during shipment.
2-3.
SETTING UP THE GENERATOR
The
following
paragraphs
describe how
to
set
up
the
Generator
for
operation.
This
information
includes;
line power
requirements,
line
voltage
selection
procedures,
fuse
replacement procedures, and rack mounting instructions.
2-4.
Line Power Requirements
The Generator uses a line voltage of 100 or 120V ac rms (± 10%) with a I.5A fuse; or 220V or
240V ac (± 10%) with a 0.75A fuse. The line frequency must be between 48 to 63 Hz. The power
consumption of the instrument is <180 VA with a full option complement.
2
Line Voltage and Fuse Selection
Verify that the intended line power source matches the line voltage setting of
the Generator before plugging in the line power cord.
Refer to Figure 2-1 to set the line voltage of the Generator to match the available source.
Figure 2-1 also shows how to replace the line fuse of the Generator. The correct fuse value for
each of the four line voltages is listed on a plate attached to the rear panel of the Generator.
NSTALLATION AND OPERATION
2-6,
IEEE-488 Address
The IEEE-488 address can be selected using the switches located next to the IEEE-488
connector on the rear panel. Talk-only and listen-only modes can also be selected on this
switch.
NOTE
To meet the specified radiated emissions, the IEEE-488 connector must
he terminated with a shielded IEEE-488 cable, such as a Fluke Y8021.
2-7.
RACK OR BENCH MOUNTING THE GENERATOR
Allow
at
least 3 inches
of clearance behind and on each side
of the
Generator to ensure proper air circulation.
The Generator may be placed directly on a work bench or mounted in a standard (24-inch
deep) equipment rack. Use the Fluke Y6001 Rack Mount Kit for mounting the Generator on
an equipment rack. Instructions for installing the Generator with the Rack Mount Kit are
provided in the kit. The outside dimensions of the Generator are shown in Figure 2-2. The
Rack Mount Kit is composed of the following parts:
5’4-inch Rack Adapter, P/N M05-205-600
24-inch Rack Slides, P/N MOO-280-610
2-8.
GENERAL OPERATING INFORMATION
The following paragraphs contain general information on the operation of the Generator. This
includes all the information required to familiarize the operator with the instrument and the
differences between local and remote operation.
2-9.
Familiarization
The Generator normally operates on an internal reference oscillator. However, if desired, the
Generator can be operated on an external reference by setting the rear panel REF INT/ EXT
switch to EXT and connecting the external reference to the REF IN connector.
When the Generator is operating on the internal reference, a 10-MHz signal
is present at the 10 MHz OUT connector on the rear panel. To meet the
specified radiated emissions, this connector must be terminated with a BNC
non-shorting
dust
cap. A dust cap, JF 478982,
is
supplied
with
the
Generator,
if a cable is connected,
it must be a double-shielded coaxial
cable such as RG-223 terminated in a SO-ohm load.
Output spectral degradation occurs if the Generator is operated on internal
reference with an external reference signal applied.
Figure 2-3 shows the front panel controls, indicator, and connectors and Table 2-I describes
the features.
Figure 2-4 shows the rear panel controls, connectors, and switches and Table 2-2 describes the
features.
2-10.
Local Versus Remote Operation
Two modes of controlling the output of the Generator are
local operation and remote
operation. In the local operation mode, the operator uses the keys on the front panel to control
the Generator. The remote operation mode
uses an IEEE-488 controller to control the
Generator through the IEEE-488 Interface. An overview of local operation is presented under
the heading Local and Remote Control Operations. The local and remote control operations
that have
similar entry methods
are
also described under Local and Remote
Control
Operations. The paragraphs under the heading Remote Operation contain information on
commands or descriptions that pertain only to remote operations.
2-11.
Power-On Sequence
When the Generator is turned on, a power-on sequence starts. During the power-on sequence,
the microprocessor tests the analog circuitry, the program ROM, the scratch-pad RAM, the
compensation memory, and the front panel displays. The front panel displays are tested by
lighting all segments for a brief period at the same time the rest of the self tests are performed.
If any of the self tests fail, an error code is displayed. If the operator initiates any front panel
entry before the power-on sequence is completed, the self test is aborted, and the Generator is
set to the state it was in when turned off. Table 2-3 lists the Instrument Preset State. Power-on
instrument
settings
that
relate
to
the IEEE-488
Interface
are described
in
the Remote
Operation paragraphs in this section. More detail on the power-on self tests are explained in
Section 4D.
2-12.
Changing Output Parameters
The six parameters of the Generator (i.e., frequency, amplitude, amplitude modulation (AM),
frequency
modulation (FM),
phase
modulation (0M),
and
pulse
modulation) may
be
individually changed by at least one of three methods;
Function Entry
Bright-Digit Edit Entry
Step Entry
and modulation frequency.
INT AM
Indicates
that
the
internal modulation oscillator
signal
is amplitude modulating
the Generator.
EXT AM
Indicates
that
the Generator
is amplitude modulated by
the
signal
connected to
the ^DD INPUT
connector.
INT
FM
Indicates
that
the
internal modulation oscillator
signal
is
frequency modulating
the Generator.
EXT
FM
Indicates
that
the Generator
is
frequency modulated
by
the
signal
connected to
the MOD INPUT
connector.
INT 0M
Irxiicates
that
the
internal modulation oscillator
signal
is
phase modulating
the Generator.
EXT 0M
Indicates
that
the Generator
is phase modulated by
the
signal
connected to the TOD INPUT
connector.
intJL
Indicates
that
the
internal
modulation oscillator
signal
is pulse modulating
the Generator.
EXT Jl.
Indicates
that
the Generator
is pulse modulated
by
the
signal
connected to the PULSE MODULATION
INPUT
connector.
STEP
Indicates
that
the
Step
or
Cv3
keys
(Step
Entry)
affect
the
current
Modulation display
value.
%
Indicates
that
the value displayed
is
the
AM Depth
in
percent.
kHz
DEV
Indicates
that
the value displayed
is the
FM Deviation
in kHz.
rad
Indicates
that
the value displayed
is the Phase Modulation
Deviation
in
radians.
the value displayed
is
the
target
Level
in dBm when performing
a
Level
compensation procedure.
AQO Hz
Indicates
that
the
internal modulating frequency
is 400 Hz.
1000 Hz
Indicates
that
the
internal modulating frequency
is 1000 Hz.
EXT HI
Indicates
that
.the external
modulation
signal
is more
than
2%
above
the nominal
IV peak
requirement
for
calibrated
operati on.
EXT LO
Indicates
that
the
external modulation signal
is more
than
2% below
the nominal
1V peak
input
requirement.
FREQUENCY
DISPLAY
FIELD
A 9-digit
display,
with six
indicators
used to display
the
output
frequency
of
the Generator.
Also
used to display
the
special
function
code,
status
error
codes,
or
the
memory
Location being
stored or
recalled,
as
well
as
relative and
actual
frequency,
when
REL
is
lit,
and
step
frequency.
STEP
Indicates
that
the
STEP
C'']
or
Cv3
keys
(Step
Entry)
affect
the
output
frequency.
REL
Indicates
that
the displayed frequency
is
relative
to
a
reference
frequency.
COMP
Indicates
that
the
compensation memory protection
switch
on the
A2
Controller
PCA
is set
to the enabled position.
FM
Indicates
that
an
FH compensation procedure
is
in progress
OUT
Indicates
that an Output
compensation procedure
is
in
progress
ATT
Indicates
that an
Attenuator
compensation procedure
is
in
progress.
®
AMPLITUDE
DISPLAY
FIELD
A 3 1/2-
(and
sign)
digit
display,
with
six
indicators,
used to display
the output
amplitude
of
the Generator
into
a 50-ohm
load.
STEP
Indicates
that
the
STEP
or
Cv3
keys
(Step
Entry)
affect
the output
amplitude.
REL
Indicates
that
the displayed amplitude
is relative
to
a
reference amplitude.
dBm
Indicates
that
the
output
amplitude
is
in decibels
relative to one milliwatt.
V
Indicates
that
the
output
amplitude
is
in volts.
uV
Indicates
that
the output
amplitude
is
in microvolts.
mV
Indicates
that
the output
amplitude
is
in millivolts.
®
STATUS
DISPLAY
FIELD
The
status display
field
is
composed of
11
indicators
used to denote
the
current
status
of
the Generator
or
instrument
entry.
EXT REF
Indicates
that
the
rear
panel
REF
switch
is
in
the
EXT
(external)
position.
Table 2-1. Front Panel Controls, Indicators, and Connectors (cont)
REJ
ENTRY
Lights when an
invalid entry
is made.
UNCAL
Lights when
a parameter entry
is outside
its
specified
range.
This
indicator
flashes
when any
of
the
internal
DAC's are
over
or
under-flow
or when any
abnormal
operation
is
detected.
RF OFF
Lights when the
RF OUTPUT
is disabled.
REMOTE
Lights when the Generator
is
in the
remote
ClEEE-488
interface)
mode
of
operation.
ADDR
Lights when the Generator
is addressed to
Listen
or
talk.
SRQ
Lights when
the Generator
has asserted the
IEEE-488
SRQ
signal.
©
MODULATION
ON/OFF
Used to
select
type^
source^
and
frequency
of modulation.
With
the exception
of
the
C400/1000T
key^
these
keys
operate
as
independent
push-on/push-off
switches
for
the
given
function.
Any combination
is allowed.
INT AM
Erables
internal amplitude modulation at
the frequency
annunciated by
the 400/1000 Hz
indicator.
INT
FM 0M
Enables
internal
frequency modulation
at
the frequency
annunciated by
the 400/1000 Hz
indicator.
EXT AM
Enables external
amplitude modulation using the
signal
applied to the
MOD INPUT
connector.
EXT
FM 0M
Enables external
frequency modulation using
the
signal
applied to the TOD
INPUT
connector.
ext_TL
Enables external
pulse modulation using
the
signal
applied to the
PULSE MODULATION
INPUT
connector.
400/1000
Alternately
sets
the
internal modulation oscillator's
frequency
to 400
or
1000 Hz,
Selected frequency
is
displayed only when
INT
AM^
FM^ 0M,
or
internal
pulse
is enabled.
©
FUNCTION
With
the exception
of
the
CSTEP]
and
CSPCL3
keys,
these
keys operate
as
interlocked
switches
that
select
the
parameter
to be entered or
edited.
For
the
CFREQ3,
[AMPLD,
CAM],
and
CFM]
FUNCTION
keys,
the bright
digit
appears
in the
corresponding display
of
the
selected
function.
FREQ
Selects
the frequency
parameter
of
the Generator
to be
programmed by using
the
DATA,
EDIT,
or
STEP entry
keys.
AMPL
Selects
the amplitude parameter
of
the Generator
to be
programmed by using
the
DATA,
EDIT,
or
STEP entry keys.
AM
Selects
the amplitude modulation
CAM)
parameter
of
the
Generator
to be programmed by
using
the
DATA,
EDIT,
or
STEP
entry
keys.
FMjBM
Selects
the frequency modulation
(FM),
or phase
modulation
(0M)
parameter
of
the Generator
to be
programmed by
using
the
DATA,
EDIT,
or
STEP entry keys.
NSTALLATION AND OPERATION
Table 2-1. Front Panel Controls, Indicators, and Connectors (cont)
SPCL
Enables
the
special
function mode.
Special
functions are
called up by
a
two-digit
code,
that
is entered by
using
the
DATA
keys.
Refer
to the paragraphs on Special
Function
in this
section
for
a detailed description and
a
list
of
the
special
functions.
STEP
After
one
of
the
four parameter
functions has
been
selected
for
programming,
pressing
this key
allows you
to
program
a step-wise
change
to
that
parameter.
The
step
increase
or
decrease
is
then performed every
time
the
STEP
or
Cv]
keys are pressed.
©
DATA
A
ten-digit
(plus
sign and decimal
key)
keypad used
for
entering
a parameter's value,
the
special
function
code,
or
a memory
recall/store
location.
©
Memory
STO
Used with
the DATA keys
to store
the
current
instrument
state
in
a memory
location.
Memory
locations
01
through
50 are available.
RCL
Used with
the
DATA
keys to recall
an
instrument
state
from
a memory
location.
Memory
locations
01
through 50
are available
for
operator-stored states; memory
location
98 contains the
Instrument
Preset
State
(see
Table 2-3.)
SEG
Sequentially
recalls,
in increasing
location
order,
the
instrument
states
stored
in memory.
While the
CSEQ3
key
is
pressed,
successive memory
locations are displayed.
When
the
key
is
released,
the
location
Last
displayed
is
recalled.
©
UNITS
These
keys,
with
the
exception
of
CCLR/LCL],
serve as the
terminating keystroke
of
a function entry,
thereby
causing
the Generator
to be progranmed.
The amplitude units
keys
are also used during Amplitude Units
Conversion entries.
MHz|V
Used with
the
CFREQ3,
CFM3,
and
CAMPLD
function keys
to
program
the
numerical
DATA entries
in terms
of
megahertz
(frequency
or
frequency modulation)
or
volts
(amplitude).
dB(tn)
Used with
the
CAMPL]
function
key
to program
the
numerical
DATA entries
in terms
of decibels per milliwatt.
kHz ImV
Used with
the
CFREQ],
EFM3,
and
CAMPL3
function keys
to
program
the
numerical
DATA
in
terms of
kilohertz
(frequency
or
frequency modulation)
or millivolts
(amplitude)
.
%
j rad
Used with
the
CAM]
function
key
to
program
the
numerical
DATA
entries
in terms of
percentage
AM depth.
Also used
with
the
CFMI0M3
function key
to program
the
numerical
DATA entries
in terms of
radians.
HzluV
Used with
the
CFREQ3,
CFM3,
and
CAMPL3
function
keys
to
program
the
numerical
DATA
in terms
of
hertz
(frequency
or
frequency modulation)
or microvolts
(amplitude).
CLRILCL
When
the Generator
is
in
local
operation,
this
key
is
used to
clear
the
current
entry
and
returns
the
Generator
to the previous
state.
When
the
instrument
is
in
remote
operation,
this key
is used to return
Local
control.
2-8
Table 2-1. Front Pane! Controls, Indicators, and Connectors (cont)
STEP
CvD
@
edit
t
number visible through
the window
on
the
selector
card
indicates
the
naninal
line
voltage
to which
the Generator
must
be connected.
The
line voltage
is
selected
by
orienting
the
selector
card appropriately.
A
1
1/2~ampere
fuse
is
required
for 100/120V operation and
a 3/4-ampere
fuse
is
required
for 220/240V operation.
REF
INT/EXT
Permits
selection
of
the Generator
frequency
reference.
When
set
to
INT,
the Generator
operates
on the
internal
reference,
which
is either
the
standard oscillator,
the
high“Stabil ity
oscillator
(if
the Option -130
is
installed)
or
the medium-stability
oscillator
(if
the
Option “132
is
installed).
In any
case,
the
internal
10-Miz
reference
signal
is available at
the
10 MHZ
OUT
connector.
When
set
to
EXT,
the
Generator
reference
is
the 10-MHz
signal
applied to the external
REF
IN
connector.
10 MHZ
IN/OUT
Connector
CBNC)
provides
the 10-Wz
reference
signal
Connector
(BNC)
is present
to accept
a
1~,
2-,
2.5~,
5-,
or 10~Wz,
0.3
to 4V p~p
sine
or
square wave
signal
into
nominally 50 ohms.
MOD
INPUT
Connector
(BNC)
is present
only
with
the Option
accept
a
IV peak external modulation
signal.
TOD
INPUT
Connector
(BNC)
is present
only
with
the Option -830
to
Table 2-3. Instrument Preset State
FUNCTION
SETTING
Frequency
Frequency
Step
Amplitude
Amplitude
Step
Modulation Rate
AM Depth
AM
Depth
Step
.....
FM Deviation without/with
Option -651
.
.
.
FM Deviation Step without/with Option -651
.
Modulation
Display
Bright-Digit
Location
Frequency Bright-Digit
Position
Amplitude Bright-Digit
Position
AM Bright-Digit
Position
FM Bright-Digit
Position without/with Option
Special
Functions
INT
AM
EXT
AM
INT
FM
EXT
FM
EXT PULSE
Step
Function
300.00000
MHz
1 ,00000
MHz
.
.
-10,0 dBm
1
dB
,
.
.
1000 Hz
30%
....
1
%
II*.!!’.".!!!
*5!o/. 5
kHz
100/10 Hz
AM Depth
.
.
.
.
Frequency
Bright
Digit
1
MH z
,
1
dBm
1
%
-651
10/1
Hz
.
.
.
.
20,30,40^50,60,70,80,90
Off
.
.
Off
. ...
Off
........
Off
Off
...
Frequency
Step
The command syntax for function entries
is:
Select Function — Enter Data — Select Unit
1.
Select one of the four parameters using the FUNCTION keys. The bright digit
appears in the corresponding display field. The presence of the bright digit in the
display
field
indicates
that
the value of the
selected parameter
is
ready
to be
programmed or changed.
2.
Enter the numeric data with the DATA keys. The numerics appear in the appropriate
display field.
3.
Select
a UNITS
key.
This
gives
the
data
its
absolute
value and
causes
the
microprocessor to internally program the Generator to the new state.
For the amplitude and frequency functions, the entered data programs the displayed value. If
the relative mode is enabled, the displayedvalue may be different from the actual output value.
Once
a
function
is
selected,
the
selected
parameter
or
feature
remains
in
the
active
programming mode until a new function is selected. Data for a selected parameter must be
followed by a unit value and must be within the range specified for the function. The display
field flashes, and the REJ ENTRY status indicator flashes if the entered data is not within the
specified range. A rejected entry does not affect the output of the Generator. The output of the
Generator remains at its previous values until a new value is accepted.
The function entry may be terminated at any time by the [CLRl LCL] key or by selecting
another function.
2-14.
BRIGHT-DIGIT EDIT ENTRY
Changing an instrument parameter by the bright-digit edit entry method is the fastest way to
make vernier (incremental) changes to one of the four parameters. The EDIT keys are used
with the four parameter FUNCTION keys to position the bright digit in the desired display
field and then increase or decrease the bright-digit value.
The command syntax for bright-digit edit entries is:
Select Display Field - Position Bright Digit — Change Bright-Digit Value
1.
Use one of the four FUNCTION keys to position the bright digit in the appropriate
display field.
2.
Use the [->] or [— ] EDIT keys to position the bright digit to the desired resolution,
and use the [t] or [|] EDIT keys to increase or decrease the value of the bright digit.
The position of the bright digit within a display field is maintained when the bright digit is
moved from one display field to another and then back to the original display field.
The repeat rate of the [|] or [i] EDIT keys may be changed to a faster or slower rate (a medium
repeat rate is the default) with a special function code (see Table 2-4). Refer to the paragraphs
on Special Function later in this section for the method and code.
2-15.
STEP ENTRY
Changing parameters by
the
step entry method
allows
the operator to
preset stepwise
increments of a parameter, then change that parameter (by the amount programmed in the
step function) with a single keystroke.
The command syntax for step entries is:
Select Step Function — Enter Data — Select Units — Change Parameter
1.
Select the parameter to be changed stepwise using one of the FUNCTION keys
followed by the [STEP] key to enable the step function.
2.
Program the numeric step amount using the DATA keys.
3.
Select a UNIT key to give the data its absolute value.
4.
The parameter value can now be changed by increments of the programmed step
amount by using the [f] or [|j STEP keys.
While the [STEP] key is pressed, the display field of the selected parameter shows the step
amount. The STEP indicator is
lit in the display field currently affected by the [STEP] key.
The repeat rate of the [t] or [|] STEP keys may be changed to a faster or slower rate (a medium
repeat rate is the default) with a Special Function code. Refer to the paragraphs on Special
Functions for the method and code.
A step entry is ignored when the result of that step entry would cause the value of the parameter
to exceed its programmable limit.
The [STATUS] key allows the operator to interrogate the Generator for an explanation of
uncalibrated or rejected entry operation (UNCAL or REJ ENTRY indicator is lit). Refer to
the paragraphs on Status and Clear Entry later in this section for a complete
list of status
codes.
See Section 3.
07
Set
SRQ.
08
Reset
SRQ.
09
Display
software revision
level.
Information appears
in the MODULATION
and
FREQUENCY
display
fields
for
5
seconds
or
until
another
key
is pressed.
10
Display
IEEE-488 mode and address
11
Display
self-test
results.
Zeros
in the display
fields
indicate
that
the
self
tests have passed.
See
Section 4.
12
Turn on displays.
13
Turn off
all
displays.
ALL
other
functions
still
operate.
14
Initialize memory
Locations
to instrument
Preset
State.
Sto
? appears
in
the
FREQUENCY
display
field
for
5
seconds.
If
during
this
time,
the
CST03
key
is
Pressed,
all memory
Locations are
initialized.
15
Latch
test.
See Section 4.
16
Display
option
loading.
See
Interrogate
Commands
in
this section.
17
Initiate
self
test
with
RF
on.
See
Section 4.
20
Disable
Relative Frequency.
See
Relative
Function
in this
section.
21
Enable
Relative
Frequency.
See
Relative
Function
in
this
section.
30
Disable
Relative Amplitude.
See Relative
Function
in this section.
31
Erable Relative Amplitude.
See
Relative
Function
in this
section.
40
Disable
Internal
Pulse
Modulation.
See Modulation Entry
in
this
section
41
Enable
Internal
Pulse
Modulation.
See
Modulation Entry
in this
section
50
Select
dBm
Amplitude Display
Mode.
See Amplitude Units Selection
in this
section.
51
Select
dBmV
Amplitude Display
Mode.
See Amplitude Units Selection
in
this section.
52
Select dBuV Amplitude Display
Mode.
See
Amplitude Units Selection
in
this section.
60
Disable
DC
AM mode.
See Modulation Entry
in this
section.
61
Enable
DC
AM mode.
See Modulation Entry
in this
section.
70
Set
repeat
rate
for
EDIT
and
STEP
keys
to medium.
71
Set
repeat
rate for
EDIT
and
STEP
keys
to
fast.
72
Set
repeat
rate
for
EDIT
and
STEP
keys
to
slow.
75
Display
Compensation Memory
status.
See
Section 4.
76
Repair
Compensation Memory
Errors.
See Section 4.
77
Load
FM MEC PROM
data
into Compensation Memory.
See
Section 4
78
Load Output
MEC
PROM data
into Compensation Memory.
See
section 4.
79
Load Attenuator
MEC PROM
data
into Compensation Memory.
See
section 4.
80
Enable
Amplitude correction.
Normal
operation.
81
Disable
Amplitude correction.
RF output
Level
may be up
to 12
dB
low.
82
Disable attenuator
correction.
Useful
as
a
troubleshooting
tool.
RF
input
to attenuator
is
flat.
83
Program
alternate 24
dB
attenuation.
See
Section 4.
84
Program
alternate 24 dB
attenuation.
See
Section 4.
85
Program
alternate
24
dB
attenuation.
See
Section 4.
86
Program
alternate
24
dB
attenuation.
See
Section 4.
90
Disable
Amplitude
Fixed Range.
See
Amplitude
Fixed Range
in this
section-
91
Enable Amplitude
Fixed Range.
See topLitude
Fixed Range
in this
section.
95
Initiate
FM Keyboard Compensation Procedure.
See Section 4E.
96
Initiate Output
Keyboard Compensation Procedure.
See Section 4E.
97
Initiate
Attenuator Keyboard Compensation Procedure.
See
Section 4E.
98
Initiate Output
with
Default
Attenuator
Data Keyboard
Compensation Procedure,
See Section 4E.
I LCL] key may be used to dear a partial DATA entry or clear the flashing REJ
ENTRY indicator.
2-17.
RF Output On/Off
The RF OUTPUT [ON/ OFF] key allows the operator to enable or disable the RF output of
the Generator. This feature
is useful in zeroing a power meter, finding the noise floor of a
system, or determining the presence or source of an unknown signal.
At power-on, the RF output of the Generator assumes the state it was in when the Generator
was turned
off.
Pressing
the RF OUTPUT [ON/ OFF] key
disables
the output
of the
Generator and causes the RF OFF indicator (in the STATUS display field) to light. If the RF
ON/ OFF was off at power-on, pressing the [ON/ OFF] key will enable output.
2-18.
Modulation On/Off and Rate
The MODULATION ON/ OFF keys allow the operator
to
select any combination
of
modulation or no modulation. The MODULATION display field indicates what combination
of modulation has been selected. Each modulation key is a push-on push-off type (except the
[400/ 1000] key).
The [400/ 1000] key toggles the internal modulation oscillator between 400 and 1000 Hz. The
400 Hz and 1000 Hz indicators are lit only when Internal AM, FM, <^M, or Pulse modulation is
enabled.
2”19.
Memory
Memory entry using the [STO] key allows the operator to save up to 50 complete front panel
settings for later recall. No memory location needs to be specified for the sequence operation.
The command syntax for memory entries
is:
Select Memory Function — Enter Memory Location
1.
Store the current front panel setting by pressing the [STO] key (located below the
DATA
keys). The
last memory location
stored
or
recalled
is displayed
in the
FREQUENCY display field.
2 .
Use the DATA keys to enter the two-digit memory location code. The location code
must contain both
digits
(e.g., 01 , 02 ,
... 50 ). The two-digit code appears in the
FREQUENCY display field as
it is entered.
3.
Recall a front panel setting by pressing the [RCL] key (located below the DATA
keys). The last memory location stored or recalled is displayed in the FREQUENCY
display field.
4.
Use the DATA keys to enter the memory location code of the desired front panel
setting. Remember, the location code must contain both
digits of the memory
location code.
Memory location 98 contains the Instrument Preset State that can be recalled at any time.
The [SEQ] key allows the front panel settings stored in memory to be sequentially recalled.
This process is activated by pressing the [SEQ] key at any time. When the [SEQ] key is pressed,
the memory location code of the currently recalled setting appears in the FREQUENCY
display field, and the location is recalled. When the last memory location is reached (50), the
[SEQ] key starts over at 01. The [SEQ] key repeats while pressed.
2-20.
Software Compensation Procedures
The software compensation procedures allow the operator to update the instrument-specific
compensation data after making related repairs. These procedures are easy to use and may be
performed
via the
front panel
(local)
or under computer control (remote) through the
IEEE-488 Interface. The local procedures enable secondary functions for many of the keys.
Each remote procedure enables a special set of compensation commands and disallows much
of the standard IEEE-488 command set. See Section 4E, Software Compensation Procedures,
for details.
2-21.
Special Function
Special function entries allow the operator to enable several special operating functions in the
Generator. For example, special functions allow the operator to change the repeat rate of the
STEP and EDIT keys, start the self tests, display the results of the power-on self tests, display
the IEEE-488 addresU enable relative and fixed-range features, and disable or enable special
attenuation features. A complete list of the special functions available is presented in Table 2-4.
The command syntax for special function entries
is:
Select Special Function — Enter Special Function Code
1.
Select the special function by pressing the [SPCL] key.
2.
The special function code is entered using the DATA keys.
2-22.
LOCAL AND REMOTE CONTROL OPERATIONS
The following paragraphs describe
local and remote operations that have similar entry
methods for each Generator function. The functions are arranged in alphabetical order. The
syntax of the command and allowable data ranges are for each function. Other information is
also presented.
2-23.
Amplitude and Frequency Entry
The following information describes how to control the carrier frequency and amplitude by the
FUNCTION-DATA-UNIT entry sequence. This method applies to both normal and relative
operations. The frequency display is a fixed-point display in MHz,. The amplitude display is
fixed point while displaying dBm but is Boating point when displaying voltage units.
The RF OUTPUT [ON/ OFF] must be enabled for the Generator to produce an output. (See
the reference material on RF OUTPUT ON/OFF Entry.)
Command Syntax:
Select Function — Enter Data
Select Unit
to 2100 MHz
10/20 Hz
1/2,3
Remote;
"FR"
float —
"GZ”
'MZ“
"KZ"
"HZ"
0.1
to 2100 MHz
10/20 Hz
1/2/4
Set
Amplitude
Local
:
campl: — DATA — IdBimn
CMHzjV]
CkHz|mV3
CHzjuV]
-137
to +13/16 dBm
0.032
uV to 1-41/1.58
0.1
dBm
V 3 digits
3,5,6
Remote:
"AP"
float —
"DB"
-V"
-137 to +13/16
dBm
0.032
uV
to 1.41/1.58
0.1
dBm
V 3 digits
4,5,6
”|viv''
"UV"
"NV"
Example:
Set
Frequency
to 10.7
MHz
and Amplitude
to -7.5
dBm.
Local:
[FREQ] U1
C03
L.l
Z71
LMHzjV]
CAMPL:
L-1
L71
L.I
Z51
LdBOn):
Remote:
"FR10.7MZ^AP-7.5DB"
Notes:
1.
Frequency ranging occurs at 245, 512, and 1050 MHz.
2.
Frequency may be programmed with 10 Hz resolution when frequency <1050 MHz,
and with 20 Hz resolution when frequency >1050 MHz.
3.
FUNCTION ([FREQ] or [AMPL]) remains selected until another FUNCTION or
[STEP], [STO], [RCL], or [SPCL] is pressed.
4.
Floating-point number is indicated by the word float.
5.
Amplitude uncaiibrated range from -147.4 to -137.1 dBm. Also uncalibraled from
+ 16.1 to +17 dBm when frequency <1050 MHz, and from +13.1 to +17 dBm when
frequency >1050 MHz.
6.
Amplitude ranging occurs at 1/2V, 1/4V, 1/8V,
... 1/2^^V with AM off and 1/4V,
1/8V, 1/16V,
... 1/2^'V with AM on.
Amplitude Fixed Range
Amplitude Units Conversion
Amplitude Units Selection
Bright-Digit Edit Entry
Relative Function
Step Entry
2-24.
Amplitude Fixed Range
The following information describes how to use the Fixed-Range special function. This special
function fixes the current amplitude range (holds the currently selected step of the Step
Attenuator). This function allows monotonic and nontransient level control over a limited
range around those levels where the Step Attenuator normally autoranges. This level control
may be accomplished with the Bright-Digit Edit Entry only.
The level vernier in fixed range has at least
1 8 dB of range (the specified accuracy range is
1
dB).
Command Syntax:
Select Fixed Range -
Enable or Disable
Summary:
COMMAND
NOTES
Enable
Fixed
Range
Local:
CSPCLD
C9: —
C13
1
Remote:
"SP"
"9"
--
”1”
Disable
Fixed Range
Local:
CSPCL:
1193
tOl
2
Remote:
"SP"
"9"
--
"0"
Example:
Set the Generator for monotonic and nontransient amplitude control (Bright-Digit Edit
only) over the range of the vernier level control below 0.25V.
Local:
[AMPLD
C.3
121
tSl
CMHzjVJ
CSPCL3
191
Cl]
Remote:
"AP.25V^ SP91
Notes:
1
The amplitude range is fixed only for Bright-Digit Edit operations. Other methods of
changing the amplitude cause the step attenuator to autorange if necessary.
2.
With amplitude fixed range disabled, amplitude ranging occurs at 1 / 2V,
1 / 4V,
1 / 8V,
... !/2'^V with AM off and 1/4V, 1/8V, 1/16V,... 1/2^'V with AM on.
Bright-Digit Edit Entry
Relative Function
2-25.
Amplitude Units Conversion
The following information describes how to convert the displayed amplitude level from dBm
to volts and from volts to dBm. The output of the Generator does not change during these
operations.
Command Syntax;
Select Amplitude Function — Select Unit
Summary:
COMMAND
Convert
dBm to volts:
Local:
CAMPL3
--
CMHz|V3
CkHzlmV]
CHzluV3
Remote;
"AP"
—
"V"
"MV"
"UV"
"NV"
Convert
volts
to dBm:
Local:
LAHPL: —
CdBCm)]
Remote:
"AP"
—
"DB"
NOTE
1
1
Example:
Change the displayed amplitude of -10.0 dBm to its voltage equivalent.
Local;
CAMPL3
CMHzjV]
Remote:
"APV"
Note:
1
.
Any voltage unit is accepted since the microprocessor automatically selects the units
appropriate for the value being displayed.
Related Operations:
Relative Function
Amplitude Units Selection
Amplitude Units Selection
The following information describes how to select the alternate level display units. When the
level is displayed as a dB quantity, units of dBm, dBmV or dB/iV may be selected by special
function. Entering the special function code converts the displayed quantity to the selected
units but does not change the output of the Generator.
Command Syntax:
Select Amplitude Units — Select Unit
Summary:
COMMAND
NOTE
Select
dBm
Display Units
Local:
CSPCL3
LSI —
IQl
Remote:
"SP"
"S'*
—
"0"
Select
dBmV Display Units
Local:
CSPCLJ
LSI —
111
Remote:
"SP"
"5"
—
"1"
Select
dBuV Display Units
Local:
:SPCL3
LSI —
L21
Remote:
"SP"
"5"
—
"2"
2-27.
Bright-Digit Edit Entry
The following information describes how to use a Bright-Digit Edit Entry to change an
instrument parameter. The output frequency, amplitude, and modulation indexes can be
modified with this entry method.
The RF OUTPUT [ON/OFF] must be enabled for the Generator to produce an output. (See
the reference material on RF OUTPUT [ON/ OFF] Entry.)
Command Syntax:
Select Display Field — Position Bright Digit — Change Bright-Digit Value
Summary:
COMMAND
NOTES
Edit
Frequency
Local:
EFREQD — EDIT
L<1/L>1
~~
EDIT
L^l/Lvl
1,2
Remote:
"FB"
—
float
"GZ"
—
"KF"
float
3,4,5
"MZ"
"KZ"
"HZ"
2-19
CAMPL3 — EDIT
C<3/C>3
•—
EDIT
:^]/Cv3
1,2
Remote:
"AB"
float
"D8"
"V"
"MV”
"UV"
"NV"
—
"KA" float
3,4,5
Edit
FM/0M Deviation
Local;
CFM10M3
-•-
EDIT
C<3/[>3
--
EDIT
:^3/Cv3
1,2
Remote:
“DB"
float
"GZ"
"MZ"
"KZ”
"HZ"
"RD"
—
"KD"
float
3,4,5
Edit
AM Depth
Local;
CAM3
EDIT
i:<3/C>3
-
EDIT
C''3/Cv3
1,2
Remote:
"PB"
float
"PC"
— "KP" float
3,4,5
Example
!:
Edit the displayed amplitude of 9.7 to 10.0 dBm.
Local;
Put
the
bright
digit
in the amplitude display by
pressing
EAMPLJ .
Select
the
least
significant
digit
in that display
by pressing EDIT 03
until
the
bright
digit
is on that
digit.
Increase
the
value
of
that
digit
by pressing EDIT
C'']
three
times.
Remote:
"AB.1DB,KA3"
Example 2:
Edit the displayed FM Deviation from 5.0 to 3.0 kHz.
Local;
Put
the
bright
digit
in the
FM
display by pressing
CFM3.
Select
the
1-kHz digit
by pressing
the
EDIT 03
or
EDIT
C<3
until
the
bright
digit
is on that
digit.
Decrease
the
value
of that
digit
by pressing EDIT
CvD
twi ce.
Remote:
"DB1KZ,KD-2"
Notes:
1.
The bright-digit field remains selected until another display field is selected.
2.
The bright-digit position is maintained for each of the four functions so that the
bright digit can be moved from one display to another and back without losing its
position in the previous display field.
2-20
Floating-point number is indicated by the word float.
In remote, the bright digit is positioned within a display field using a decade value and
associated unit. Minus signs are ignored.
5.
in remote, the bright digit is moved to the corresponding field and
is increased or
decreased by the signed integer following the “iCF,KA,KD,KP” messages. The
generic edit command “KB” may also be used to
edit up or down the current
bright-digit position. Positive integers do not require a sign.
Related Operations:
Relative Function
Amplitude Fixed Range
2-28.
Memory Entry
The following information describes how to use the memory function to store and recall front
panel settings. The Generator has 50 memory locations that are retained for 2 years with the
power off.
The sequence feature allows the operator to recall successive memory locations.
Command Syntax;
Select Memory Function — Enter Memory Location
Summary:
COMMAND
NOTES
Store
Local
:
[ST03
--
Cn3
Cn3
1,2,3
Remote
:
"ST"
--
int
1,4
Reca 1
Local
;
CRCL3
'-
En3
Cn3
1,2,3
Remote
:
"RC"
-—
int
1,4
Sequence
Local
:
[SEQ3
5,6
Remote; SQ"
5
Example:
Recall
the Instrument
Preset
State
(located
in memory location
98). Change
the
frequency parameter to 302 MHz, then store the new front panel setting in memory
location 06.
Local:
CRCU
[93
C83
EDIT
C^3
C^3
CST03
101
161
Remote:
''RC98,KF2,ST6"
The memory locations available for operator use are 01 through 50. Additionally, the
following special memory locations are available:
Memory location 00 contains a backup-memory location. After a recall {or sequence)
operation
it contains the last front panel setting. After a store operation,
it contains
the data in the stored memory location before the store operation. Thus, a recall
operation can be reversed by recalling location 00.
Memory location 98 contains the Instrument Preset State.
Memory location 99 contains the present instrument state.
2.
In local control, two data digits must be entered to specify the memory location. The
recall or store is performed when the second digit is released.
3.
The last memory location specified (used for sequence operations) is displayed while
the [STOJ or [RCL] button
is pressed.
4.
An unsigned integer is indicated with int.
5.
The sequence operation recalls the next higher memory location, starting from the
last memory location stored or recalled. No memory location needs to be specified.
When the highest location is reached, the sequence starts over again at location 01.
6.
While [SEQ]
is pressed, the next memory location number
is displayed and the
memory location is recalled. This key is repeating.
2-29.
Modulation Entry
The following information describes how to preset the modulation index (AM depth or
FM/^M deviation),
internal modulation
rate (400 or
1000 Hz), and how
to
select
the
modulation source (internal and/or external).
The FUNCTION-DATA-UNIT method of selecting the modulation index is summarized in
the following command syntax. The indices may also be modified using Bright-Digit Edit or
Step Entry. Since there is only one modulation display, the modulation index displayed
is
determined by the last modulation FUNCTION key pressed.
Command Syntax:
Select Function — Enter Data — Select Unit
Summary:
COMMAND
RANGE
RESOLUTION
NOTES
AM
Depth
Local
:
imi
--
DATA “
C%3
0 to 99%
1%
1,2
Remote
:
"AM" —
float — "PC"
0 to 99%
1%
1,2,3
Set
FM Deviation
rad 3 digits
Remote:
”FM" —
float —
”GZ"
0
to 400
kHz
3 digits
1^3,4,
5
"MZ”
"KZ"
"HZ"
"RD"
0 to 40.0
rad
3 digits
Summary:
COMMAND
NOTES
Select
Modulation Rate
Local:
[400/1000D
6
Remote:
"MR"
"0"
or
"1"
7
Summary:
COMMAND
NOTES
Enable
or
Disable
AM,
DCAM,
FM/0M
or
Pulse
Modulation
Local:
CINT
AM]
8,9
CINT
FM
0M]
CEXT AM]
CEXT
FM
0M]
CEXTJI.]
CSPCL]
C4]
--
CO]
or
Cl]
10
CSPCL]
C6] —
CO]
or
Cl]
11
Remote:
"Al"
"0"
or
"1"
12
"FI"
"0"
or
”1"
"AE" "0"
or
”1"
"FE"
"0"
or
"1"
"PE"
"0"
or
"1"
"PI"
"0"
or
"1"
"DA"
"0"
or
"1"
Example:
Set the FM deviation to 5 kHz, set the modulation rate to 400 Hz, and modulate the
carrier internally.
10,12
11,12
Local;
CFM|0M]
C5]
CkHz]
CINT
FM]
C400/1000]
Remote:
”FM5KZ,MR0,FI1
NSTALLATION AND OPERATION
Notes;
1.
This operation does
not change the Generator output unless the corresponding
modulation is enabled.
2.
Uncalibrated
if peak amplitude exceeds +16 dBm when frequency <1050 MHz, or
+13 dBm when frequency ^1050 MHz.
3.
Floating-point number is indicated with the word float.
4.
Uncalibrated if FM
is enabled and FM deviation is above (frequency -150 kHz).
5.
The maximum FM deviation is dependent on the output frequency. Deviations up to
400 kHz or 40.0 radians may be entered regardless of the output frequency; however,
the UNCAL indicator is flashed if the limits specified in Table 4D-14 are exceeded.
The maximum deviations are reduced to 40.0 kHz and 4.00 rad if the low-rate FM
option is installed.
6.
Toggles between 400 or 1000 Hz only. An indicator shows selected rate only if internal
modulation is on.
7.
"0” selects a modulation rate of 400 Hz; “1” selects 1000 Hz.
8.
These are ON/ OFF operations; any combination is allowed.
9.
Two indicators EXT HI and EXT LO are
lit when external AM or FM
is on to
indicate that the external modulation signal is 2% above or 2% below the nominal
1
peak input requirement.
10. Internal pulse modulation can be enabled by special -function or by the alternate
IEEE-488 command “PI”.
It can be enabled with any combination of the above
modulations.
11. The dc coupled AM mode can be enabled by special function or by the alternate
IEEE-488 command “DA”. The dc coupled AM mode works in conjunction with
External AM and only affects the output of the Generator if External AM
is also
enabled.
12. “0” turns the modulation source off; “I” turns
it on.
Related Operations:
Bright-Digit Edit
Entry Step Entry
2-30.
Relative Function
The following paragraphs describe how to change frequency and amplitude using the relative
mode. There are two steps;
1
Setting the reference
2
Changing the parameter relative to that reference
2-24
Setting the reference is done by setting the parameter to the desired value and then enabling the
relative mode for that parameter. This causes the REL indicator to light and the displayed
value to be zero in the corresponding display. The Generator output does not change during
these operations. In the relative mode, the usual means of changing the parameter may be used;
i.e,, Function, Step, or Bright-Digit Edit Entry.
In the relative frequency mode, the actual frequency
is the sum of the reference and the
displayed frequency. The actual frequency may be displayed by pressing the [FREQ] key. If
the 10/ 20 Hz frequency resolution boundary is crossed using the step or edit functions while in
the relative frequency mode, a 10-Hz rounding quantity may be applied to the displayed
frequency.
In the relative amplitude mode, the actual amplitude
is the sum of the reference and the
displayed amplitude when the reference and the displayed quantities have the same units.
However, with mixed units (volts and dB), the actual amplitude is the voltage value scaled by
the dB value. The actual amplitude may be displayed by pressing the [AMPL] key.
Command Syntax:
Select Relative Function — Enable or Disable
Summary:
COMMAND
NOTE
Frequency
Local:
CSPCL3
ZZl —
CO]
or
C13
1
Remote:
"SP”
”2"
—
"0"
or
"1”
1
Amplitude
Local:
CSPCL3
C33 —
C03
or
C13
1
Remote:
"SP"
”3"
—
"0"
or
"1"
1
Example:
Set the amplitude to -15 dBV; i.e., 15 dB below
1 volt.
Local:
CAMPL3
C13
CMHzlW
CSPCL3
C33
C13
CAMPL3
C“3
C13
C53
CdB(m)3
Remote;
"AP1V^SP31,AP“15D8"
Note:
1.
I enables the relative function; 0 disables the relative function.
Related Operations:
Amplitude and Frequency Entry
Bright-Digit Edit Entry
Step Entry
The following information describes how to enable the output of the Generator using the RF
OUTPUT [ON/OFF] key and the corresponding remote code.
Command Syntax:
RF Output On/ Off
Summary:
COMMAND
NOTE
RF Output
On
Local:
RF OUTPUT
CON/OFF3
when
RF OFF
is on
1
Remote:
"R01"
1
RF
Output
Off
Local:
RF OUTPUT
CON/OFF3
when
RF OFF
is off
Remote:
"ROD"
Note:
1.
Turning the RF Output on resets the RPP circuitry if it has tripped.
2-32.
Special Function Entry
The following information describes how to use the Special Function Entry to use the special
operating functions of the Signal Generator. Table 2-4 lists the special functions available.
The special function code is a two-digit number. The first digit indicates the classification of
the special function, and the second digit specifies the particular special function.
The special function is executed when the second special function code digit is entered. There
are ten classes of special functions. The special functions in the 0(n) and l(n) class cause an
action to be performed. Classes 2(n) through 9(n) cause an instrument state to change. The
status of classes 2{n) through 9(n) appears (left to right) in the frequency display field when the
[SPCL] key is pressed. Table 2-4 describes the special functions available with the Generator.
Command Syntax:
Select Special Function — Enter Special Function Code
COMMAND
NOTE
Local:
CSPCL3 —
Inl
Cn3
Remote:
"SP”
—
int
1
Example;
Change the repeat rate of the EDIT and STEP keys to slow.
Unsigned integer is indicated with int.
Related Operations:
Fixed Range
Relative Function
2-33.
Status and Clear Entry
The status entry allows the operator to interrogate the Generator for an explanation of either
uncalibrated operation (UNCAL indicator is lit) or rejected entry operation (the REJ ENTRY
indicator in
lit).
When either the UNCAL or REJ ENTRY indicator is lit, press and hold the [STATUS] k.ey to
display the Uncalibrated or Rejected Entry Error Code Message. These messages provide
detailed information on the nature of the uncalibrated or rejected entry condition. Table 2-5
contains
a
list and explanation of
all the Uncalibraled Error Code Messages. Table 2-6
contains a list and explanation of all the rejected entry error code messages.
The [CLR
I LCL] key may be used to clear a partial DATA entry or clear the flashing REJ
ENTRY indicator. Press the [STATUS] key while an UNCAL indication exists to display the
uncal error codes in three fields:
Flashing codes (denoted by *) indicate abnormal operation or aberrated output. Non-flashing
codes indicate operation outside the specified range.
Press the [STATUS] key while the REJ ENTRY indication exists to display the reject entry
error codes;
The following information describes how
to
use
the
step entry function
to change an
instrument parameter. The RF OUTPUT [ON/ OFF] must be enabled for the Generator to
produce an output. (See the RF OUTPUT [ON/ OFF] Entry information covered earlier in
this section.)
Command Syntax:
Select Step Function — Enter Data — Select Units — Change Parameter
Summary:
COMMAND
RANGE
RESOLUTION
NOTES
Frequency
CFREQ3i;STEP3--DATA — CMHz|V3 —STEP
C'^3/:v3
CkHzlmV3
CHz|uV3
0
to 2100
MHz
10 Hz
2
Local
:
— float— "GZ"
"MZ"
"KZ"
"HZ"
0 to 2100 MHz
10 Hz
1,2
Amplitude
Local:
CAMPL] CSTEPJ— DATA — CdB(m)3
—
STEP
C''3/Cv3
CMHzlV]
[kHz jfnVJ
CHzluvJ
0
to
164 dB
0.1
dB
0 nV to 1999
V
3 digits
Remote:
"LS"
— float— "DB"
—"LU'V'LD"
-V"
-MV'*
"UV"
"NV"
0 to 164
dB
0.1
dB
1,2
0
nV to 1999V
3 digits
1,2
FM/0M deviation
Local:
CFH 10MO CSTEP3
—DATA — CHHzjVJ —
STEP
C^T/Cv]
CkHzImV]
LHzjuV]
LX
i radj
0 to 400
kHz
3 digits
3
0
to 40.0
rad
3 digits
Remote;
"DS"
— float— "6Z"
—"DU'V'DD"
"MZ"
"KZ"
"HZ"
"RD"
0
to 400
kHz
3 digits
1,2,3
0
to 40-0
rad
3 digits
AM depth
Local:
CAM3CSTEP3
—DATA — U3
—
STEP
L^l/tyl
0
to 99%
1%
Remote;
"PS"
—float—"PC"
—''pu'V'pD"
0 to 99%
1%
1,2
Example:
Recall the Instrument Preset State: [RCL] [9] [8]. Step the displayed frequency of 300
MHz, in 10-MHz steps, to 270 MHz.
Local:
CFREQ]
[STEPJ
[13
C03
CMHzjVT
CvT
Cv3
Cv3STEP
Remote:
"FS1 OMZ, FO, FD, FD"
Notes:
1.
Floating-point number is indicated with the word float.
2-28
Table 2-5. UNCAL Error Codes
CODE
DESCRIPTION
000 000 000 =
No UNCAL
conditions
002 000 000
=:
FM Dev/0M >
(freq -150
kHz)
*004 000 000
= Excess
FM
Dev/0M,
main
or
reference
PLL unlocked
*010 000 000
=
FM
DAC
at
full
scale
200 000 000
Multiple COMP memory checksum
errors.
See
Compensation
Memory
Status
in Section
2.
*000 010 000 = Main or
reference
PLL unlocked
000 000 001
s Level
vernier
>
12 dB below bottom
of
range
000 000 002
- Peak amplitude
> +16 dBm
for
freq
<
1050
FWz
or
> +13 dBm
for
freq
>= 1050 MHz
*000 000 004
a Amplitude unleveled
*000 000 010
a
Fixed-range
level
vernier at
0
*000 000 020
s
Fixed-range
Level vernier
at
full
scale
*000 000 040
-
RPP tripped
000 000 100 a Level
< -137 dBm
000 000 200
= Level
correction disabled
*000 000 400
= RF
off
Flashing codes
(denoted by
*)
indicate abnormal
operation
or
aberrated
output .
Non--flashing codes
indicate operation outside
specified
range.
Table 2-6. REJect ENTRY Codes
CODE
DESCRIPTION
000 000 000
= No
rej ected entries
001
000 000
s:
FM Oev/0M not
between 0
and 400 kHz
(40.0
kHz
with Option -651)
002 000 000
=
FM Dev/0M Step
not
between
0
and 400 kHz
(40.0
kHz
with Option -651)
004 000 000
AM Depth
not
between
0
and 99%
010 000 000
AM Depth
Step
not between 0 and 99%
020 000 000
=
IEEE-488 command syntax
error
040 000 000
IEEE-488
input
value
out
of
range
100 000 000 =
MEC
compensation PROM error
200 000 000
=
IEEE
edit
or
step operation beyond
allowed
range
400 000 000
=
Invalid
frequency
in COMP memory
000
001
000
= Frequency
not
between 100
kHz
and
2100
Miz
000 002 000
=
COMP switch not enabled or
low-rate
FM option not disabled
000 004 000
= Frequency
Step
not
between 0 and
2100 MHz
000 010 000
= COMP
data may
not
be stored
if
procedure
incomplete
000 020 000
=
Invalid compensation command
000 040 OOQ
=
Invalid memory
location
000 100 000
=
Invalid data
in memory
000 200 000
8
Special
function not allowed
000 400 OQO =
COMP
data
range error
000 000
001
B
,Output amplitude not
between
10
nV and 1.58V
000 000 002
B
Insufficient
resolution
for
units
conversion
000 000
004
= Units
conversion
to volts not
allowed with
reference
in volts
000 000 010
s Units
conversion
to dB
not allowed with
reference
in volts
000 000 020
s Amplitude
Step
not
between 0 and
164
dB
or 0 and 1999V
000 000 040
= Units
conversion of
A^npLitude
or
FM/0M Step not allowed
000 000 100
= Amplitude or
FM/0M
step
and
current
display
not
in
same units
000 000 200
s
COMP
data
from
IEEE-488
out
of
range or
edit
beyond
COMP
limits
000 000 400
a
Internal
compensation data
transfer error
2.
Entering the step size from IEEE-488 does not select the step function. For example,
“FS 10MZ”does not select the step function; “FD” or “FU”must be used to select the
frequency step function. The generic step up/down commands “SU” and “SD”may
be used to step the current step function.
3.
If the Low-Rate FM option is installed, the range of FM deviation steps is 0 to 40.0
kHz, and the range of 0M deviation steps is 0 to 4.00 rad.
INSTALLATION AND OPERATION
Related Operations;
Relative Function
2-35.
REMOTE OPERATION (IEEE-488 INTERFACE)
The following paragraphs describe how to operate the Generator using the IEEE-488 Interface
(referred to as remote operation). The interface allows the operator to program the Generator
and operate instrument functions via the IEEE-488 bus (with the exception of the front panel
POWER switch and the rear panel REF INT/EXT switch). The IEEE-488 Interface also
provides additional programming features not accessible from the front panel.
The rest of this section is divided into two parts: the first part describes how to set up the
Generator for operation on the IEEE-488 bus and gives some ty pical programming examples.
The first part also includes a complete list of the programming commands recognized by the
Generator software.
The second part describes the implementation of the IEEE-488 Interface and programming
features that are accessible only from the IEEE-488 Interface. The second part also includes
typical timing data, provided as an aid to system programmers. This information can assist in
writing programs that have greater speed and efficiency.
The Generator can be used with any IEEE-488 controller in the normal addressed mode. The
following two additional modes are available for operation without a controller:
Listen-only mode
Talk-only mode
In the listen-only mode, the Generator responds to all data messages on the IEEE-488 bus. In
the talk-only mode, the Generator sends commands on the IEEE-488 bus to program another
606xA Generator (or a 607xA with some restrictions).
2-36.
Setting Up the IEEE-488 Interface
Figure 2-5 shows a 6062A Signal Generator connected to a 1722A via the IEEE-488 bus.
Use the following procedure to set up the Generator with the IEEE-488 Interface:
1
.
Connect a standard IEEE-488 cable between the Generator and the IEEE-488 device.
NOTE
The IEEE-488 Interface signal SHIELD (pin 12) can be disconnected
(when using an IEEE-488 cable without a metallic hood)from instrument
ground. To do this, use the left most address switch (as viewedfrom the
rear panel).
Figure 2-5. 6082A Synthesized RF Signal Generator Connected to a 1722A
2-30
Select the IEEE-488 address and mode as follows:
a.
Set both the LISTEN ONLY and TALK ONLY switches (located on the rear
panel of the Generator) to 0 (down). Set switches al through a5 to the desired
address 0 through 30. For example, for an address of 1
, set switches a2, a3, a4, and
a5 to 0 (down), and set switch al to
1 (up).
b.
For talk-only operation, set the TALK ONLY switch to
1 (up),
c.
For listen-only operation, set the LISTEN ONLY switch to
i (up) and set the
TALK ONLY switch to 0 (down).
3.
Verify the address and mode:
a.
Press the [SPCL] and the [1] [0] keys. Verify that the selected address appears in
decimal in the FREQUENCY display field.
b.
If the talk-only mode or listen-only mode has been selected, “to” or “lo” appears
to the left of the address in the FREQUENCY display field.
NOTE
The address switches are continuously monitored except when in remote.
The TALK ONLY and LISTEN ONLY svcUches are only read when the
Generator is powered on.
2-37.
Programming Commands
After the address and mode have been set, the Generator can be programmed by an IEEE-488
controller or from another generator. Tables 2-7 and 2-8 and the programming examples
following them provide' the basic information on how to program the Generator.
Table 2-7. Index of IEEE-488 Commands
FUNCT ION
COMMAND HEADERS
Amplitude Entry
AP,
SP3x,
RA ^
SP5x,
SP8> ,
SP9x
Binary
Learn Commands
LL
LM
Clear
Commands
ca.
CE,
CL
Compensation Mode Commands
CM
Canpensation Procedure Commands
CF,
CP,
ic.
IH
Edit
Entry
AB,
DB,
FB,
PB,
KB,
KA,
KD,
KF,
KP
Frequency
Entry
SP2> .
RF
Interface Mode Commands
EM,
RM,
TM,
VM,
UM,
a
Interrogate Commands
lA,
ID,
IE,
II,
IL,
10,
IR,
IS,
IT,
lU,
IV,
IZ
Hanory
Entry
RC,
ST,
SQ
Modulation Entry
AM,
AE,
Ai,
SP6x ,
DA ,
FM ,
FE,
FJ ,
MR ,
MF
PE,
SP4x ,
PI
Monitor Commands
IB,
OB,
OD,
RB,
RW,
DW,
WW,
XA,
X8,
XD,
XR
RF ON/OFF
Entry
RO
Special
Function
Entry
SP
SRQ Commands
IM,
SM,
XF
Step
Entry
FS,
LS,
PS,
DS,
SU,
SD,
FU,
FD,
LU,
LD,
PU,
PD,
DU,
DD
Trigger Commands
CT,
TR
INSTALLATION AND OPERATION
Table 2-8. IEEE-488 Commands
COMMAND
COMMAND
COMMENTS
USE
HEADER
NUMERIC
SUFFIX
AMPLITUDE ENTRY
Program
AP
f 1 oat
Program displayed amplitude
AmpLi tude
in
units
of:
V
volts
MV
millivolts
UV
mi crovolts
NV
nanovoLts
DB
dB,
dBm^
dBmV
or
dBuV
Convert
AP
none
Change amplitude
units
to;
Amplitude
V
volts
Units
MV
volts
UV
volts
NV
volts
DB
dB,
dBm^
dBmV
or
dBuV
Relative
SP
30/31
none
Disabie/enable relative
Amplitude
amplitude
operation
Relative
RA
0/1
Alternate
programming
command
Amplitude
for
disable/enable
relative
amplitude
operation.
Amplitude
SP
50
none
Select
dBm
units
Di splay
51
Select
dBmV units
Units
52
Select
dBuV units
Level
SP
80
none
Enable
all
Level
correction.
Correction
81
Disable all
level
correction.
82
Disable attenuator
correction.
Amplitude
SP
90/91
none
Disabie/enable amplitude
Fixed Range
fixed-range
operation.
BINARY LEARN
COMMANDS
Store
a
LI
int
string
The
Generator
stores
the
Front
Panel
string
into
the memory
Setup
Location specified
by
int.
See
the
Command
Description
paragraph
for
decoding
the
Learn
string.
Send
a
LM
int
none
The
Generator
responds
with
Front
Panel
the
contents
of
the memory
Setup
Location specified
by
int.
See
the
Ccxnmand
Descriptions
paragraph
for
decoding the
learn
string.
CLEAR
COMMANDS
Clear
IEEE-
CB
none
none
Clears
IEEE-488 output
488 Output
Buffer
buffer.
Clear
error
CE
none
none
Clears
the
IEEE-488
rejected
entry
status.
2-32
Dev i ce
CL
none
none
Clears the
instrument
state
Clear
and exits
all
compensation
procedures.
COMPENSATION
MODE
COMMANDS
Compensation
CM
none
AT
Begin Attenuator
compensation
Mode
FM
procedure.
Begin
FM
compensation
procedure.
OD
Begin Output
with
default
attenuator
compensation procedure.
OT
Begin Output
compensation
procedure.
sv
Save
compensation data
and exit
compensation procedure.
EX
Exit
compensation procedure
without
saving
the
data.
COMPENSATION
PROCEDURE COMMANDS
Compensa ti on
CF
float
G2
Accept
FM
deviation reading
FM
Entry
MZ
from modulation analyzer
KZ
during
FM
compensation
HZ
procedure.
Compensati on
CP
float
DB
Accept
amplitude
reading from
Ampl i tude
power
meter
during
level
Entry
compensati on
procedure.
Interrogate
IC
none
FR
Interrogate
the
frequency
of
Compensation
the
current
compensation
Step
Data
TG
procedure
step.
Interrogate
the target
level
or
FM
deviation of
the
current
compensation procedure
step.
Interrogate
IH
none
none
Interrogate
the
frequency
and
Het
Adjust-
level
offset
necessary
to make
ment
Data
the
het
compensation adjustment.
For example,
"+00001 20000 .,+00000000 .30"
indicates
the
adjustment
should
be made
at
120
kHz
and the
Level
needs
to
be
adjusted up
.3
dB.
EDIT ENTRY
Position
AB
float
V
Position the
bright
digit
in
Amplitude
MV
the AMPLITUDE
display with
Bright
Digit
uv
the
stated
resolution.
For
NV
example,
enter
"ABlOMV"
for
DB
a 10-mV
resolution.
Position
DB
float
GZ
Position the
bright
digit
in
FM
Bright
MZ
the
FM
display with
the
stated
Digit
KZ
resolution.
For
example,
enter
HZ
"DB1KZ"
for
a 1-kHz
resolution.
RD
ope rati on.
Relati ve
Frequency
RF
0/1
none
Alternate
programming
command
for
disable/enable
relative
frequency
operation.
j
INTERFACE
MODE
COMMANDS
Error
Mode
EM
0/1
none
Disable/enable
the
clear
error mode.
If
disabled,
the
IEEE-488 error
status
is
cleared only
when
inter-
rogated.
If
enabled,
the
error
status
is
cleared when
a
new message
is processed.
Record Mode
RM
0/1
none
Disable/enable
the
record
mode.
If
disabled,
the
message
unit
is
a
command.
If
enabled,
a message
unit
is
a
record.
The message
unit
is
the
smallest
group of
char-
acters
that
the
Generator
processes.
Table 2-8. IEEE-488 Commands (cont)
1
Record
Terminator
Mode
TM
0/1
none
Selects
the
LF/CR
character
as
the
record terminator.
The
record terminator
is used
on
input
in the
record mode
and
is
sent
following
all
output.
Output
Valid
Mode
VM
0/1
none
Oi sable/enable the output
valid mode.
In the
output
valid mode/
the Generator
waits to process
commands
until
the
RF output
has
become valid.
Unbuffered
Mode
UM
0/1
none
Di sa ble/enable the unbuffered
mode.
If
disabled/
all
input
is
buffered.
If enabled,
only
one
message
unit
is buffered.
"a"
Modes
a
int
none
The
"a"
command may
be
used
as
an alternate method of
programming
interface
modes.
INTERROGATE
COMMANDS
!
Interrogate
Attenuator
Counts
lA
none
none
The
Generator
responds
with
seven
counts.
Each
count
indicates
the
total
number
of
actuations
for
one
of
the
seven
attenuator
sections
in
the
Generator,
Instrument
Identi f i-
cati on
ID
none
none
The
Generator
responds
with
its
model
number,
for
example,
"6062A".
Elapsed
T ime
Indicator
IE
none
none
The
Generator
responds
with
the
total
operating
time
since
the
Generator
was manufactured.
Interface
Modes
II
none
none
Interrogate
the
interface
modes
selected.
The Generator
responds
with
an unsigned
integer.
Interrogate
Error
Log
IL
none
none
The
Generator
responds
with
ten error
Log entries.
Each
entry
is
an
uncaL
error
code
or
a
self
test
result
and
the
elapsed
time
of
when
the
error
was
Logged.
Opt i on
Loading
10
none
none
Interrogate
the
option
Load-
ing.
The
Generator
re-
sponds
with
the message:
d1,
d2, d3
d1
is
the
instrument
code.
d2
is
the
digital
and
synthesizer
options.
d3
is
the
output
options.
See
the
Interrogate
Commands
paragraphs
for
details.
Table 2-8. IEEE-488 Commands (cont)
Rej ected
Entry
IR
none
none
Interrogates
the
rejected en-
try error
codes.
The
Generator
responds
with
three
octal
fields:
"AAAAAA, BBBB88, CCCCCC".
See Table 2-6
for
a
list
of
rejected entry
error
codes.
Serial
Number
IS
none
none
Interrogates
the
instrument
serial
number.
Self Test
IT
none
none
Interrogates
the
results
of
the
self
tests.
The
Generator
responds
with
the
self-test
results.
See
paragra
graph 4D-20
for
self-test
codes.
UN CAL
lU
none
none
Interrogates
the
uncali-
brated output
error
codes.
The Generator
responds
with
three
octal
fields:
"AAAAAA, BBBBBB, CCCCCC”.
See
Table 2-5
for
a
list
of
uncal
error
codes.
Software
Version
IV
none
none
Interrogate
the
software
ver-
sion.
The
Generator
responds
with
the
status message;
"Vxx.x" where
x's are decimal
digits representing
the
current
software
revision
level.
Compensation
Memory
Status
11
none
none
Interrogates
the
compensation
memory
status.
The Generator
Responds
with
three
fields:
"AAAAAA, BBBBBB, CCCCCC".
See
Table 4D-7
for
a
list
of
status
codes.
MEMORY ENTRY
Recall
RC
int
none
Recall
the
front
panel
setup
stored
at
the memory
location
specified by
int.
Store
ST
int
none
Store
the
current
front
panel
setup
at
the memory
location
specified
by
int.
Sequence
SQ
none
none
Sequence
(recall)
to the
next
higher memory
location.
MODULATION
ENTRY
|
Program
AM
AM
float
PC
Program
AM
depth
in percent,
|
External
AM
AE
0/1
none
Di sable/enable external
AM
Internal
AH
A1
0/1
none
Disable/enable
internal
AM
INSTALLATION AND OPERATION
Table 2-8. IEEE-488 Commands (cont)
DC
coupled
AM
SP
60/61
none
Disable/enable
DC
coupled AM
DC
coupled
DA
0/1
none
Alternate programming
command
AM
for
disable/enable
DC
coupled
AM operation
Program
FM
FM
float
Program
FM
deviation
in units
of:
GZ
gigahertz
M2
megahertz
KZ
kilohertz
H2
hertz
RD
radians
External
FM
FE
0/1
none
Disable/enable external
FM
Internal
FH
FI
0/1
none
Disable/enable internal
FH
Program
MR
0/1
none
Program modulation frequency
Mod
Freq
to 400 Hz/1000 Hz.
Program
MF
float
Program modulation frequency
Mod
Freq
in
units
of:
GZ
gi gahertz
MZ
megahertz
KZ
ki 1 cfiertz
HZ
hertz
External
PE
0/1
Di sable/enable external
pulse
Pul se
modulation
Internal
SP
40/41
Disable/enable
internal
pulse
1
Pul se
modulati on
Interna 1
PI
0/1
Alternate
programming
command
Pulse
for
disable/enable internal
pulse modulation operation
MONITOR
COMMANDS
Input
Bit
IB
none
BIT
Respond with
the value
of
the
Desig-
designated
hardware
bit.
nator
Output
Bit
08
0/1
BIT
Set
the
designated hardware
Desig-
bit
to
0
or
1
.
nator
Output
Dac
OD
int
DAC
Set
the
value
of
the
desig-
Desig-
nated
hardware
DAC
to the
nator
value
specified by
int.
Read Byte
RB
int
none
Read the value
of
the ad-
dressed byte.
The Generator
responds
with
an unsigned
integer.
Read Word
RW
int
none
Read the
value
of
the
ad-
dressed word.
The Generator
responds
with
an
unsigned
integer.
2-37
into the address
specified with
the
define
write
address
command.
Write Word
WW
int
none
Write
int
into
the address
specified with
the
define
write
address
command.
Read
XA
none
none
Read the
current
attenua-
Attenuati on
tion.
The Generator
responds
with
an unsigned
integer.
Write
X8
none
none
Change attenuation to 6dB
Attenuation
times the unsigned
integer.
The
integer
can
be
0
to 23.
Set
XD
float
6Z
Set
the
frequency
hardware
Frequency
MZ
directly
to
the
specified
Di rect
KZ
synthesizer
frequency.
HZ
RF Output
1
XR
0/1
none
"XRO"
programs alt
attenuation.
|
"XR1”
restores attenuation to
!
its
previous
state.
|
RF ON/OFF
ENTRY
j
I
RF Output
RO
0/1
none
Turn
RF output
off/on.
j
SPECIAL
FUNCTION ENTRY
|
Special
SP
00
Clears
all
special
functions
Funct i ons
02
Initiates
self
test
03
Display check
04
Key
check
07/08
Set/reset
SRQ
09
Display
S/W
rev
and
instr
ID
10
Display IEEE-488 address
11
Display
self-test
results
12/13
Turn on/off
display
14
Initialize memory
15
Latch
test
16
Display
option
Loading
17
Initiates
self
test
with
RF
on
20/21
Disable/enable relative
freq
30/31
Di sable/enable relative
ampl
40/41
Disable/enable
internal
pulse
50
Select
dBm ampl
display
units
51
Select
dBmV »npl
display units
52
Select
dBuV
ampl
display units
60/61
Disable/enable
DC AM
70
Medium
key
repeat
rate
71
Fast
key
repeat
rate
72
Slow
key
repeat
rate
75
Display
COMP memory
status
I
76
Repair
COMP memory errors
I
__
77-79
Copy MEC
PROM
data
NSTALLATION AND OPERATION
Table 2-8. IEEE-488 Commands (cont)
80
Enable
all
level
correction
81
Disable all
level
correction
82
Disable attenuator
correction
83-86
Program
alternate 24dB attens
90/91
Di sable/enable ampL
fixed-rng
95-98
Local
COMP procedures
SRQ COMMANDS
Interrogate
IM
none
none
Interrogate
the
SRQ
mask.
The
SRQ Mask
Generator
responds
with
the
decimal
value
of
the
SRQ
mask.
Set
SRQ
SM
int
none
The
SRQ mask
is
set
to
int.
Local
XF
0/1
none
Disable/enable
a mode
to
set
Operation
SRQ
each
time
a
local
entry
Alert
Mode
is made.
This
SRQ
is enabled
by
setting
the
front
panel
bit
in
the
SRQ mask.
STEP ENTRY
Program
FS
float
Program
frequency
step size
FREQ
STEP
in
units
of:
Size
GZ
gigahertz
MZ
megahertz
KZ
ki 1 ohertz
HZ
hertz
Program
LS
float
Program
amplitude
step size
AMPL STEP
in
units
of:
Size
V
volts
MV
mil livolts
UV
microvolts
NV
nanovolts
DB
dB
Progran
PS
float
PC
Program
AM
step
size
in
AM
STEP
Size
pe rcent
.
Program
DS
float
Program
FM
step
size
in
units
FM
STEP
of;
Size
GZ
gigahertz
HZ
megahertz
KZ
ki 1 ohertz
HZ
hert 2
RD
radians
Step Up/Down
SU/SD
none
none
Step
the
currently
selected
step
function up/down one
step.
Step Up/Down
FU/FD
none
none
Change
the
current
step
Frequency
function
to frequency
and
step
frequency
up/down one
step.
Step Up/Down
Amplitude
LU/LD
none
none
Change
the
current
step
function to amplitude
and
step
amplitude up/down one
step.
INSTALLATION AND OPERATION
Table 2-8. IEEE-488 Commands (cont)
Step Up/Down
AM
PU/PD
none
none
Change
the
current
step
function
to AH and
step AM
up/down one
step.
Step Up/Down
FM
DU/DD
none
none
Change
the
current
step
function to
FM
and
step
FM
up/down one
step.
|
TRIGGER
COMMANDS
Configure
Trigger
CT
string
none
Configures
the
trigger.
Each
time
a
trigger
command
or
a
group execute
trigger
inter-
face message
is
received,
the
Generator
executes
the
string
of
commands.
The
string
record must
end
with
a record terminator.
T rigger
TR
none
none
Trigger
command.
Equivalent
to
the
group execute
interface
message.
Upon processing the
trigger
command,
the Generator
executes
the
string,
which
has
been
preprogrammed with
the
configure trigger
command.
1
More details about the commands can be found in two places. Commands that are available
from the front panel are described earlier in this section (Local and Remote Operation). Those
commands that are only available from the IEEE-488 Interface are described in the Command
Descriptions paragraphs later in this section.
Table 2-7 is an index for the IEEE-488 Commands used in Table 2-8. This index is a list of the
command headers according to function. Table 2-8
lists all the remote commands that are
recognized by the Generator. The commands are listed alphabetically by function.
The following three examples show how to use the IEEE-488 bus and how to use a variety of
controllers to program the Generator. In the first example, a Fluke 1722A Controller is used to
program the Generator. In the second example, two Generators are configured to track each
other in frequency. In the third example, a I722A is used to program the Generator with the
frequency step up controlled by the trigger command.
2-39.
PROGRAMMING EXAMPLE
1
Use the following procedure to program the Generator with a Fluke 1722A Instrument
Controller to this state:
Frequency
Amplitude
Modulation Freq
FM
Internal FM
AM
External AM
210 MHz
6 dBm
1000 Hz
5 kHz
ON
15%
ON
2-40
Connect the Generator to the controller with an IEEE-488 cable.
2.
Set the address switch of the Generator as follows (as viewed from the rear of the
instrument):
00000010
3.
Enter the following program into the controller:
1
f
Fluke
1722A BASIC program
to
control
a
6062A.
2
!
The
Addrees
of
the 6062A
is
2.
3
« 2X
10
!
Clear
the 6062A
so
that
it
is
in
a
known
state.
15
INIT PORT 0
20
REMOTE
©A7.
\
CLEAR SAX
iOO
!
SET THE 6062A.
no PRINT
SAX,
"FR2iOMZ,AP6DB,MRl,Fli5KZ,FIl, AlilSPC, AEl"
999 END
4.
Run the program by typing on the controller “RUN (RETURN)
The 6062A Synthesized RF Signal Generator can be connected to another signal generator in a
master-slave configuration. In the following example, two generators are configured to track
each other in frequency. This configuration may be used to track frequency, amplitude, AM,
or EM.
1.
Connect two signal generators together with an IEEE-488 cable.
2.
Set the rear panel address switch of the first generator (talker) as follows:
00100000
Cycle the power of the first generator so the software will read the talk only switch.
3.
Set the rear panel address switch of the second generator (listener) as follows:
01000000
Cycle the power of the second generator so the software will read the listen only
switch.
4.
Manually program the talker generator as follows:
FUNCTION
VALUE
KEY
SEQUENCE
Frequency
Step
Function
Frequency
Step
210 MHz
CFREQD
C2]
CID
COD
CMHzjVD
Frequency
CFREQD
CSTEP]
1.25
kHz
Cl]
C.]
C2]
C53
CkHzlmV]
5.
Manually program the listener generator as follows:
FUNCT ION
VALUE
KEY
SEQUENCE
195
MHz
CFREQ3
CID
C9]
C5D
CMHzjV]
Frequency
CFREQ3
CSTEP3
1.25
kHz
m
E.]
C23
£51
CkHzjmV]
Frequency
Step
Function
Frequency
Step
NSTALLATION AND OPERATION
6.
On the talker generator, press the [
t
] STEP or [
|
] STEP keys. Each lime the key is
pressed, the frequency of both generators increases or decreases by
i.25 kHz (the
Frequency Step) at frequencies 15 MHz apart.
Different functions on each generator can be programmed
to track
in the master-slave
configuration. In other words, while the master generator can be programmed to step increase
25 kHz FM, the slave generator can be programmed to step 25% AM.
NOTE
To use the stepfeaturefor otherfunctions, change the stepfunction on the
generators to the desiredfunctions.
2-41.
PROGRAMMING EXAMPLE 3
In
the following example, the Generator
is programmed by a Fluke 1722A Instrument
Controller
(via
the IEEE-488
bus)
to
the same
state
as
in Programming Example
1.
Additionally, the frequency step size is set to
1 .25 kHz, and the trigger buffer is programmed to
execute the step up command when the trigger command
is received. The SRQ mask of the
Generator is set to generate an SRQ when the RF output has settled and the Generator is ready
for more input from the bus.
The program then enters a loop where
it waits for the ready SRQ, sends the GET (group
execute trigger) interface message to step up the frequency, and waits again. At this time, do
the following;
1
.
Connect the Generator to the controller with an IEEE-488 cable.
2.
Set the rear panel addre.ss switch of the Generator as follows:
0000011
3.
Enter the following program into the controller:
20
100
no
120
121
130
140
150
160
170
IBO
190
300
310
320
800
810
820
830
850
860
870
999
!
Fluke
1722A BASIC program
to
control
a
6062A.
!
The address
of
the 6062A
is
7»
A% =
77.
!
Clear
the 6062A
so
that
it
is
in
a
known
state.
INIT PORT 0
REMOTE
SA7.
\ CLEAR
®A7.
'
Set
the 606SA.
PRINT @A%,
-FR210I1Z/ AP6DB.MR1, FM5KZ- FIl. AM15PC, AEl”
!
Set
the
freq^uency
step<
output valid
mode>
!
and
configure
the
trigger
buffer.
PRINT
®AX.
*’FS1.25K2. VMirCTSU"
!
Set
the SRQ mask
to enable
"output valid"
SRQ
PRINT
eA7..
"SM16"
I
Wait
for above
commands
to
finish
processing
WAIT
1000
\ S% « SPL
!
Trigger
the
first
step
up
TRIO
lA5i
!
Wait
for SRQ
ON SRQ GOTO BOO
WAIT FOR SRQ
OFF SRQ
!
Check
the
serial
poll
response
S7.
=
SPL(A7.)
IF
(S7. AND
647.+ i67.>
(
>
807, THEN PRINT
s7.;
*'
Bad Serial
Poll
!
Trigger
the next
step
up
TRIG
eA7.
i
Resume
operation— waiting
for
next SRQ
RESUME 300
END
iesponse’
4.
Run the program by typing on the controller “RUN (RETURN)”.
The Generator implements a subset of IEEE-488 Interface functions defined by the IEEE
Standard 488- 1978. Table 2-9 summarizes the IEEE-488 Interface functions implemented.
2-43.
Address Mode
In the address mode, the Generator may be operated from local (using the front panel keys) or
from remote (using the IEEE-488 Interface). The following paragraphs describe the operation
of the Generator in both states and transitions between the states.
The available IEEE-488 messages and their descriptions for the address mode of operation are
presented in Table 2-10.
2-44.
LOCAL OPERATION
The Generator powers up in the local mode. When in local mode, the following conditions are
present:
The front panel indicator REM, is not lit.
Device trigger (GET), device clear (DCL), and selected device clear (SDC) interface
messages are ignored.
All device-dependent messages are ignored.
If the data output was requested while the Generator was in the remote mode, the data
output of a talker may be sent.
2-45.
GOING FROM LOCAL TO REMOTE
The Generator switches from local to remote when the “my listen address message” (MLA) is
received, and the Remote Enable (REN) signal is true.
2-46.
REMOTE OPERATION
When in the remote mode, the following conditions are present:
The front panel REM indicator is
lit.
Device trigger (GET), device clear (DCL), and selected device clear (SDC) interface
messages are processed.
All device-dependent messages are processed during the remote mode.
Table 2-9. IEEE-488 Interface Function List
FUNCTION
DESCRIPTION
SH1
Complete
source handshake
capability
AH1
Complete acceptor
handshake
capability
T5
Basic talker>
Talk only.
Serial
poll.
Unaddressed
if
MLA
TEO
No extended talker
capability
L3
Basic
listener.
Listen only.
Unaddressed
if
MTA
LEO
No extended
listener
capability
SRI
Complete
service
request
capability
RL1
Complete
remote/Local
capability
PRO
No parallel
poll
capability
DC1
Complete device
clear
capability
DT1
Complete device trigger
capability
CO
No controller
capability
El
Open-coL lector
drivers
NSTALLATION AND OPERATION
Table 2-10. IEEE-488 Address Mode Message Descriptions
MESSAGE
DESCRIPTION
pon
Power-On
Talker
Operation
When powered up,
the Generator
generates
a
Power-
On message
(pon)
and
clears
its
output
buffer.
The
Generator
is not addressed
to talk when powered
up.
Li stener
Operation
The Generator
is
not addressed to
listen when
the power
is turned
on.
Service
Request
Operation
HTA
My
Talk
Address
The
state
of
the Service
Request
(SRQ)
signal
on
pon
is determined by
the
SRQ
mask.
The
SRQ mask
is the
same
as
when the
power was
removed-
Therefore,
if
the
SRQ mask
enables
the
power
on,
output
valid,
or
ready
SRQs,
the
SRQ
signal
will
be true
during pon.
Talker
Operation
The Generator
is addressed
to talk
upon
receipt
of
the MTA message.
The front
panel
ADDR
indicator
is
lit while the Generator
is addressed to
talk.
Listener
Operati on
The Generator
unlistens when
the
MTA message
is
received.
MLA
My
Listen
Address
Talker
Operation
The Generator
untalks when
the
MLA message
is
received.
Li stener
Operation
The Generator
is addressed
to
listen
when
the
MLA
message
is
received.
The
front
panel
ADDR
indicator
of
the
Generator
is
lit
while
the Generator
is addressed
to
listen.
Data
Talker
Ope ration
The Generator
sends
data
to the
IEEE-488 bus only when
requested by
a programming data message.
Message formats
are described
in
the
Command
Description paragraphs.
An
End
of
Record
(EOR)
character
is
sent
with
EOI
asserted
following
all
outputs.
The
EOR
character
is either
a
carriage
return
or
a
Line
feed,
depending on the
setting
of
the
terminator mode.
The parity bit
is always
zero.
Multiple output
requests are buffered until
the buffer
is
full.
Processing
of programming data messages
is
stopped until
the buffer
is
no
longer
full.
The buffer
can be cleared with
the Clear Buffer
command
("CB").
The
buffer
is also
cleared on power up
(pon),
with
a
Clear
Command
("CL"),
or
with
a Device
Clear
interface message
(OCL
or
SDC).
Li stener
Operation
Command
syntax,
error processing,
and
input
buffer
overflow
are described
in the paragraphs
on
Ccmimand
Processing.
Refer
to Table 2-8
for
a
list
of
IEEE-488
commands
that are recognized by
the
Generator.
IFC
Interface
Clear
Talker
Operation
The Generator
untalks and unlistens when
the
IFC message
is
received.
2-44
Tabie 2-10. IEEE-488 Address Mode Message Descriptions (cont)
1
MESSAGE
DESCRIPTION
Listener
The Generator
unlistens and
untalks when
the
IFC message
Operation
is
received.
OTA
Other
Talk
Address
Talker
Operation
The Generator
untalks
when
the OTA message
is
received.
SPE
Serial
Poll
Enable
Talker
After
receiving
the
SPE message,
the Generator
responds
Operation
with
the
serial
poll
status byte,
if
addressed
to talk.
SPD
Serial
Poll
Disable
Talker
After
receiving
the
SPD message,
the Generator
resumes
Operation
normal
talk operation.
ULA
Unlisten
Address
Listener
The Generator
unlistens when the ULA message
is
Operation
received.
RSV
Request
Service
Service
The
front
panel
SRQ
indicator
is
Lit
when
the
rsv
Operation
message
is
sent.
The Generator may
request
service
for
several
reasons.
Each
reason
for
service
request
can be
individually masked with
the
set mask
command
("SM”).
The
service
request
mask
can be
interrogated with
the
interrogate mask
command
("IM").
DCL
Device
Clear
Clear
The
DCL message
is
ignored when
in
local.
When
the
DCL
Operation
message
is received
(during
remote operation)
the
Generator
is
cleared.
Any characters
in the
input
buffer
are cleared followed by
the
same operation as
the
clear
command
("CL”).
The operation
of
the
DCL message
is
identical
to the
operation
of
the
selected device
clear
(SDC)
message.
The
cleared state
of
the Generator
is
described
in
the paragraphs on Power-On
Conditions.
SDC
Selected Device
Clear
Clear
The
SDC message
is
ignored during
Local
operation.
Operation
When
the
SDC message
is
received
(during rmote
operation),
the Generator
is
cleared.
Any
characters
in
the
input
buffer
are
cleared
followed
by
the
same
operation as
the
clear
command
("CL").
The operation
of
the
SDC message
is
identical
to the operation
of
the device
clear
(DCL)
message.
The
cleared state
of
the Generator
is described
in
the paragraphs on Power-On Conditions.
Table 2-10. IEEE-488 Address Mode Message Descriptions (cont)
MESSAGE
DESCRIPT ION
GET
Group
Execute
Trigger
Trigger
The GET message
is
ignored during
Local
operation.
When
Operation
the
GET message
is received
(during
remote operation)^
the Generator
executes
a
command
string
that
has
been
preprogrammed with
the Configure Trigger
command
("CT").
The operation
of
the GET message
is
identical
to the
operation
of
the
Trigger
("TR")
command.
Undefined
IEEE-488
Commands
All
undefined IEEE-488
commands
are acknowledged
by
the
Generator
handshake
sequence^
but
no action
is
taken.
2-47.
GOING FROM REMOTE TO LOCAL
The Generator switches from remote to local mode in one of the following ways: The IEEE-488
Go To Local (GTL) message is received, the remote enable signal REN is false, or a Return To
Local (rtl) message is generated by pressing the front panel [CLR
I LCL] key (if the Generator
is not in the local lockout mode).
The Generator enters the local lockout mode when the Local Lockout message (LLO)
is
received. The Generator exits the local lockout mode to the local mode when REN is false.
When
switching from remote
to
local, unprocessed commands
in
the
input
buffer
are
processed until the input buffer is cleared or a front panel entry is made. Switching to local has
no effect on the contents of the output buffer.
2-48.
Talk-Only Mode
Figure 2-6 shows two generators connected together with the IEEE-488 Bus.
To
select
the
talk-only mode,
set
the TALK ONLY
address
switch
to
1
(up),
if the
TALK-ONLY address switch and
the LISTEN-ONLY address switch
are
set
to
1, the
talk-only mode is selected.
In the talk-only mode, the listener, remote/local, service request, device clear, and device
trigger interface functions do not apply.
If the talk-only mode is selected, the Generator is always addressed to talk and the front panel
ADDR indicator is always lit. The Step Up (“SU”) or Step Down (‘‘SD”) message is sent when
the [f] STEP or [i] STEP front panel keys are pressed. This output is not buffered, and if no
listener is connected to the IEEE-488 Interface, no output will be sent. A carriage return
followed by line feed (with the EOI signal true) are always sent as the end of record.
2-49.
Listen-Only Mode
To select the listen-only mode, set the LISTEN ONLY address switch to
I (up). If the talk-only
address switch and the listen-only address switch are set to
1, the talk-only mode is selected.
Figure 2-6. 6062A IEEE-488 Bus Connected to a 6062A IEEE-488 Bus
If the listen-only mode
is selected, the Generator is always addressed to listen, and the front
panel ADDR indicator is always lit. The Generator listens and responds to all data messages
on the IEEE-488 Interface. The response to data messages is the same as in the addressed mode
of operation except that requests for talker output are ignored.
In the listen-only mode, the talker, remote/local, service request, device clear, and device
trigger interface functions do not apply.
2-50.
Command Syntax
The Generator IEEE-488 bus commands alphabet consists of the letters A through Z (upper
and lower case
letters are treated equally), digits 0 through 9, and the following special
characters:
@
.
,
; +
- CR LF
Spaces, tabs characters, and the parity bit are ignored.
The IEEE-488 commands for the Generator consist of the following three parts:
Header
Numeric
Suffix
The header is always required, but the numeric and suffix may be optional. This rule gives the
following four possible combinations:
(HEADER)
(HEADER) (NUMERIC)
(HEADER) (NUMERIC) SUFFIX)
(HEADER) (SUFFIX)
M ultiple commands may be separated with one of the end of string (EOS) characters
or
Use of EOS characters facilitates recovery in the event of a syntax error and will also enhance
readability.
2-51.
COMMAND HEADER SYNTAX
The command header is a two alpha-character string. A list of the IEEE-488 command headers
used on the Generator
is presented in Table 2-8. The header determines the syntax of the
numeric and suffix as listed in the table.
NUMERIC DATA SYNTAX
There are four types of numeric data: Boolean, unsigned integer, floating point, and trigger
string. The following paragraphs describes each of the four numeric data formats. A syntax
diagram
is included for each format.
1.
Boolean
Boolean numeric data must be either a “0” or a ‘T
All other characters will result in a
syntax error.
7
"
2.
Unsigned Integer
Unsigned integers may be specified in decimal or in hexadecimal. Any number of
decimal
digits
are
accepted.
However,
values
greater than 65,535
are
rejected.
Hexadecimal numbers are preceded by an “X”. Only four hexadecimal digits are
accepted. Specifying a number in hexadecimal for the read word and read byte
commands causes the response to be sent in hexadecimal. Decimal digits may be the
numerals 0 through 9. Hexadecimal digits may be the hexadecimal digits 0 through F.
3.
Floating Point
The floating-point numeric data format is the most flexible format. Digits may be the
numerals 0 through 9. Any number of digits is accepted for both the number and the
exponent. However, numbers greater than 2,147,483,629 are truncated, and expo-
nents greater than 32,749 are rejected.
2-48
The trigger string numeric data is a string of Generator commands terminated with an
EOR. The string may be up to 71 characters, not including the EOR. Commands in
the string are not checked for validity until the trigger string
is executed with the
trigger command.
EOR is the end of record character. This character
is selected with the terminator
mode command. “TMO” selects the linefeed character. “TMl” selects the carriage
return
character. The IEEE-488
interface
signal EOI
asserted
with any
other
character is also considered an end of record.
EOS is an end of string character, use either
or “»”•
x^--(TioRT)-
command
—
2-53.
SUFFIX SYNTAX
Suffixes are always one or two alpha-characters. Certain suffixes are used to scale the numeric
(the same as the front panel UNITS keys). Other suffixes mnemonically designate hardware
components. The five types of suffixes are described in Table 2-11.
2-54.
Command Descriptions
The following paragraphs describe the remote IEEE-488 Interface operating commands that
are not accessible from the front panel of the Generator. IEEE-488 Interface commands that
are accessible from the front panel of the Generator are described earlier in this section (Local
and Remote Operation).
2-55.
BINARY LEARN COMMANDS
Front panel setups are stored in the memory of the Generator in a packed binary format. The
binary learn commands are used to transfer this binary data between an IEEE-488 controller
and
the
Generator.
These commands
allow
the
operator
to
minimize
the amount
of
programming commands needed to program the entire instrument state. The binary learn
commands are:
“LM” Learn Memory
“LI” Learn Interface
The syntax for the Learn Memory (“LM”) command is as follows:
“LM” Memory Location Code
The Generator responds to the “LM”command with a string of 66 ASCII characters followed
by an
Table 2-11. Suffix Types
SUFFIX
TYPE
SUFFIX
MNEMONIC
EQUIVALENT EXPONENT
Frequency
GZ
gigahertz
9
MZ
mega hertz
6
KZ
kilohertz
3
HZ
hertz
0
Amplitude
V
volts
0
MV
mil livolts
-3
UV
microvolts
-6
NV
nanovolts
-9
DB
dB,
dBm,
dBmV,
or
dBuV
0
AM
PC
percent
0
FM
GZ
gigahertz
9
MZ
megahertz
6
KZ
kilohertz
3
HZ
hertz
0
RD
radians
0
DAC/BIT Designators
DAC
and BIT designators
are
two alpha-character
mnemonics
that
refer
to
hardware dacs
and
bits.
Refer
to the paragraphs
on Monitor
Commands
for
a
complete
list
of
designators.
Learn
Suffix
A
Learn suffix
is
a
string
of
ASCII
characters
that
contain
coded memory
location
contents.
Refer
to the
paragraphs on Binary
Learn
Commands
for
decoding
of
the
learn
string.
NOTE
The , end of record character,
is sent with EOl asserted. “TMO”
selects the linefeed character, and “TMI” selects the carriage return
character.
Example:
iEEE~488 Command: “LM98”
Response:
“BOABAAAAPPJMAAAKAAAAAAAABBOBKDAAAAAPECEAABPEAAABAACEBB
AGCLKKABAEAM” EOR
Refer to Figure 2-7 for information on how to decode this learn string.
The syntax for the Learn Interface “LI” command is as follows:
“LI” Memory Location Code: Learn String
CHAR.
POS.
AM
tepth
1
2
CHAR.
HEX
EQIV.
INSTRUCTION
VALUE
Reference Amplitude
CHAR.
POS.
5
6
7
8
60
CHAR.
A
A
A
A
K
HEX
EQIV.
0 X
0
INSTRUCTION
c
1
and
3
AM
Step
CHAR.
POS.
CHAR.
HEX
EQIV.
INSTRUCTION
VALUE
RESOLUTION
RESULT
Displayed Amplitude
9
10
11
12
59
!7i in n n
FM
Step
45
46
47
48
64
n
[
HEX
EQIV.
0
1
F
4
0
0
0
BINARY
EQUIV.
INSTRU CT ION
RESCLUTION/
UNITS
500
1
500C
Hz
1
10 ]
Hz
CHAR.
POS.
CHAR.
HEX
EQIV.
Data
Set #1
(miscellaneous
data)
49
50
51
52
A
A
C
E
0
0
2
4
lo b
0
0
b b b b
b b h b
b h b b
AHPL
Di splay
0
= dBm
1
= dBmV
2
= dBuV
3
“
invalid
DC
AM
*
0
=
oii,
1
= on
Phase
*
Modulation
0
= off,
1
= on
External
Pulse
*
0
a
off,
1
= on
Internal Pulse—*
0
“ off,
1
= on
External
FH —
0
= off,
1
=
on
Internal
FM
^.Relative
Amplitude
(special
function
3x).
0
= Off,
1
=
On
— Relative
Frequency
(special
function 2x)
0
=
off,
1
=
on
Step
Function
01
=
FREQ,
10
=
AMPL,
11
= MOD
L— Modulation
Rate
01
= 400
Hz,
10
= 1000 Hz
— Internal
AM
0
= off,
1
= on
External
AM
= 3rd digit
0100
= 4th
digit
0101
= 5th
digit
0110
= 6th
digit
0111
= ?th
digit
1000
=
8th
digit
1001
=
9th
digit
'Modulation display
0
=
1
FM
AMPL bright
digit'
00
=
1st
digit
01
=
2nd
digit
11
= 3rd digit
1—•AM bri ght
di
0
=
1 St
digi
»FM bright
digit
00
=
1 St
digit
01
= 2nd
digit
10
- 3rd digit
CHAR.
POS.
Amplitude
Special
Functions
63
CHAR.
HEX
EQIV.
BINARY
EQIV.
Amplitude
Fixed
Range
(Special
Function
9x)
(Special
Function
8x)
00
=
All
Correction on
01
= Attenuator
Correction
off
10
=
All
Correction off
Amplitude
Correction'
00
=
Fixed
Range
off
01
=
Fixed Range
on
(mi sceL laneous
data)
CHAR.
POS.
CHAR.
HEX
EQIV.
INSTRUCTION
BINARY EQUIV.
65
66
00
= valid
01
= EXT LO
10
= EXT HI
11
=
invalid
The Generator stores the learn string
in the memory location designated by the memory
location code. If the memory location specified is 99, the instrument is programmed to the data
sent in the learn string.
Example:
To program the Generator to the Instrument Preset State:
‘‘LI99BOABAAAAPPJMAAAICAAAAAAAABBOBKDAAAAAPECEAABPEAAABAAC
EBBAGCLKKABAEAM”
Note that the binary learn string in this example is the same as the learn string returned from
memory location 98, which contains a record of the Instrument Preset State.
Instructions:
1.
Convert the hexadecimal number to a signed decimal number as follows:
a.
Multiply the most significant hexadecimal digit by
16.
b.
Add the next significant digit to the value obtained in step a.
c.
Multiply the sum of step b by 16, and add in the next hexadecimal digit until the
least significant hexadecimal digit has been added.
NOTE
If the hexadecimal number started with an 8 through F, the number is
negative. Perform step dfor negative numbers.
d.
Subtract
16 from the number raised to the power of the number of
digits.
Example of Instruction
1:
To convert hexadecimal number IE:
(1 * 16}+ 14 = 30
(Since the most significant digit is
1, the number is not negative.)
To convert hexadecimal number FF9C:
{{{{(15 * 16) + 15) * 16) + 9) * 16) -M2 = 65436
(Since the most significant digit is F, the number is negative.)
Using Step d;
1
6“’ = 65536, 65436 - 65536 = -100. The signed decimal equivalent to FF9C is
- 100 .
2.
Amplitude quantities have a number and a resolution associated with them. This
applies to the Displayed Amplitude, Reference Amplitude, and Amplitude Step.
Use the following procedure to identify the resolution of an amplitude quantity;
a.
If the resolution is A or B (hexadecimal), the resolution of the number is 0. 1 dBm
or 0.1 dB units.
Example of Instruction 2:
The Displayed Amplitude (in Figure 2-7)
is -100 with a resolution of A.
a.
The actual displayed amplitude is -10.0 dBm.
b.
If the stored resolution is 0 through 9 (hexadecimal), the amplitude quantity is in
volts. To convert the number to the actual amplitude in nanovolts, multiply the
amplitude number by the power of ten represented by the resolution.
Example: An amplitude step of 12, with a resolution of 6 would be an actual
amplitude step of 1.2,000,000 nV or 12 mV.)
3.
If the Relative Amplitude mode
is off, the data stored in the reference amplitude
location is not used.
4.
If the Relative Frequency mode
is off, the data stored in the reference frequency
location is not used.
5.
FM and FM Step quantities have a number and a resolution associated with them.
The resolution nibble
is comprised of a 3-bit resolution value and a single bit units
value. If the units value is 0, the FM quantity is in Hertz. If the units value is 1, it is a
(f)M quantity in radians.
To convert the num.ber to the FM quantity in Hz, multiply the number by the power
of ten represented by the resolution.
Example: The FM number (in this figure) is 500 with a resolution of 10 Hz. The actual
FM deviation is 5000 Hz.
To convert the number to the (f)M quantity in radians, multiply the number by the
power of ten represented by the resolution, then divide the result by 1000.
Example: A deviation of 123 with a resolution of
1 and units of
i would be a <^M
deviation of 1.23 radians.
If the Low-Rate FM option is installed, the FM quantity adjusted by the resolution
has a resolution of 0.1 Hz. The 4>M quantity is derived exactly as
it is without the
option. The resolution remains the same, however, the range is reduced by a factor of
10 .
6.
The checksum data
is calculated by adding the data
in
the
learn
string, two
hexadecimal digits at a time. The total, including the checksum, should add up to a
number whose least significant two hexadecimal digits are 01.
7.
Only the current state of these parameters is relevant, so they are only valid when
learning memory location 99. Learn strings from all other memory locations have
characters
65
and
66
set
to “AM”.
These
parameters
are
provided
as
status
information only and cannot be set using the learn interface command.
NOTE
The memory location code must be an unsigned integer indicating the
memory location to he learned. Memory location 99 refers to the current
instrument settings. Memory location 98 refers to the Instrument Preset
State as listed in Table 2-3.
The characters in the learn string correspond to each Generator function. A description of how
to interpret the characters in the learn string
is given in Figure 2-7. Table 2-12 shows the
conversion from the learn string to the hexadecimal character.
2-56.
CLEAR COMMANDS
The following IEEE-488 clear commands are recognized by the Generator:
“CB” Clear IEEE-488 input buffer
“CE” Clear IEEE-488 rejected entry error status
“CL” Clear instrument
The “CB” command can be used to clear the Generator output buffer. The output buffer
should be cleared at the beginning of any remote program to synchronize data output. The
output buffer is also cleared at power-on with the “CL” clear Generator command, or the
buffer is cleared by the SDC and DCL clear interface messages.
The “CE” command can be used to explicitly clear the error status. The error status is also
cleared when
it
is
interrogated with
the “IR” command,
or
the “CL” clear Generator
command, or the SDC and DCL clear.
The “CL” command is used to clear the instrument state. The same actions are performed with
the SDC and DCL clear interface messages. (In addition, the input buffer is cleared with the
clear interface messages.) The following IEEE-488 commands are performed with the clear
Generatorcommand: “RC98, ROl, CE, CB, RMO, TMO, EMO, VMO, UMO, SM192, SP08,
XFO, DWO, CT’TEOR). If a Software Compensation Procedure is in progress,
it is exited
immediately and all measured data is discarded.
2-57.
INTERFACE MODE COMMANDS
Interface mode commands
are used
to
configure the Generator for
different modes
of
IEEE-488 interface operation. Since the Generator knows when its RF output has settled,
it
can be configured to synchronize itself with the controller. This eliminates WAIT statements in
the program, which are normally used to allow time for the output of the controlled device to
settle. Table 2-13 lists the interface mode commands.
Table 2-13. Interlace Mode Commands
COMMAND
DESCRIPTION
COMMAND
STATUS
"EM"
Error
Mode
1
~
on,
0
= off
"RM"
Record Mode
1
=
on, 0
= off
"TM"
Select
Terminator
1
=
CR,
0
= LF
"UM"
Unbuffered Mode
1
=
on,
0
s off
"VM"
Valid Mode
1
= on, 0
- off
The error mode selects when the IEEE-488 rejected entry status is cleared. If turned on, the
error status is cleared when a new message is processed. If turned off, the status is cleared only
when interrogated with the “IR” (interrogate rejected entry) command or when explicitly
cleared with the “CE” (clear error command).
The record mode selects whether the message unit is a record or a command. When turned on,
the message unit is a record. When turned off, the message unit is a command.
The terminator mode selects the character used as the record terminator. The terminator
character is not used for command processing unless the record mode is enabled. When turned
on, the record terminator CR (carriage return) is used. If turned off, the record terminator LF
(line feed)
is used. The record terminator character
is the last character in
all IEEE-488
messages sent from the Generator.
The unbuffered mode selects when messages from the IEEE-488 interface are processed. When
turned on, messages are read from the IEEE-488 interface only when the microprocessor is
ready to process them. In this mode, the input buffer will contain a maximum of one message.
(A message may be one command or one record, depending on the setting of the record mode.)
When turned off, messages are read from the IEEE-488 Interface to the input buffer of the
Generator at the fastest rate. In this mode, the input buffer may contain up to 80 characters.
The valid mode selects when messages are processed by the Generator microprocessor. When
the valid mode is turned on, processing of a new message is begun only after the RF output has
settled and become valid. When the valid mode
is turned off, a new message
is processed
immediately after the completion of the previous message.
The interface modes can also be programmed using the command, “(5)” n (where n
is an
integer). The interface modes are set to the value of n, where n is the sum of the codes for the
desired modes. The integer codes for the interface modes that can be programmed using the
n commands are as follows:
Terminator Mode=
I
Record Mode
=
2
Valid Mode
=
4
Unbuffered Mode=
8
Error Mode
= 16
For example, to select the record mode and valid mode, the command
is
6.
Interface Mode Example:
In this example, the RF output of the Generator is connected to a circuit that is being
measured by a voltmeter. The output of the Generator must be
settled before the
voltmeter is given the command to make a measurement.
A Fluke 1722A program might look something like:
PRINT
@1.
"CL, TMl. RMl, VMl, UMl "
!
PRINT
ei,
"FRIQOMZ, AP-25DB”
!
PRINT
@2.
"?”
!
INPUT
@2,
"R*"
!
initialize
the
6060,
select modes
program the 6060
trigger
the voltmeter
get
the reading
The entire record
is transferred into the Generator before processing begins. In this
example, processing begins when the record terminator CR is received. The following
character {LF in this case) will not be received into the Generator until the entire record is
processed and the output has settled. No wait statement is needed between setting up the
Generator and taking the measurement because the Generator will not handshake the LF
character until its output has settled.
NOTE
A record is a string of characters separated by . A message is the
smallest group of characters that the Generator can process when
it
is
programmedfrom the IEEE-488 Interface.
NOTE
The output valid state ofthe Generator occurs 45 ms after any hardware
has been changed.
2-58.
INTERROGATE COMMANDS
Interrogate commands allow the status of the Generator to be given over the IEEE-488
Interface. These commands consist of headers only. The interrogate commands available on
the IEEE-488 interface are:
“lA”
Attenuator Actuation Log
“ID”
Instrument ID
“IE”
Elapsed Time Indicator
“H”
Interface Mode Status
“IL”
Hardware UNCAL and Self-Test Error Log
“lO”
Option Loading
“IR”
IEEE-488 Rejected Entry Status
“IS”
SerialNumber
“IT”
Self-Test Results
“lU”
UNCAL (uncalibrated) Status
“IV”
Software Revision Level
“IZ”
Compensation Memory Checksum Status
The “lA” command interrogates the attenuator actuation log. Each time an attenuator pad is
removed from the RF path, the corresponding counter in non-volatile memory is incremented.
This means that if an attenuator pad is removed then inserted, the count will only increment
once. The Generator responds to the “IA” command with seven counts (one for each pad). The
counts are separated by commas, and the response
is terminated with the end of record
character ().
The response to the “lA” command will be:
0000000102,0000000099,0000001028,0000000009,0000019092,0000003908,
0000000412
When the “ID” command is sent, the Generator responds with its instrument model number
(and end of record character) as in “6062A”(EOR>
.
When the “lE”command is sent, the Generator responds with the time the instrument has been
in operation since it was manufactured. The time is 9 digits, a decimal point, a lOth digit and
the
indicates 3459.3 hours of operation since manufacture.
The “H” command interrogates the current selection of interface modes. A 5-digit integer
followed by the (EOR) character is the sum of the modes selected as follows:
Terminator mode =
I
Record mode
=
2
Valid mode
==
4
Unbuffered mode=
8
Error mode
= 16
The “IL” command interrogates the ten-entry error log in non-volatile memory. An error can
be either a failed self test or an uiicalibrated condition as a result of a hardware status line
(UNLOKL or UNLVLL). The elapsed time since the instrument was manufactured
is also
logged with the error.
Each of the ten error log entries consists of an uncalibrated (uncal) code or self-test code,
followed by a semicolon
followed by the elapsed time, followed by the end of record
character {
Self-test error (non-volatile memory
failure)
at
39012.2 hours since manufacture.
Seif-test error (non-volatile memory failure) and
the self-test sequence was aborted as indicated by the
This error occurred at 39012.1
hours since the Generator was manufactured.
000000,000000,000006;000020243.2(EOR)
Amplitude became
unieveled
at
20243.2
hours since manufacture (third number = 4). In addition, the peak amplitude was set to >
+ 13 dBm as indicated by the second uncal code (third number = 2).
000000,000000,000000;000000000.0 (EOR> This special code means that there are no
errors. One or more of the error log entries will be this code when less than ten errors have
been logged since the Generator was manufactured.
The number of characters in this response is much larger than the 80 characters in the output
buffer. Therefore,
if the entire response is not transferred to the IEEE-488 controller, the
Generator will not be able to respond to further programming commands. Any remaining
portion of the response that is not to be transferred can be cleared only by sending a device
clear to the Generator.
The ‘TO” command interrogates the Generator for
its option complement. The returned
record contains three integers, separated by commas, that indicate configuration of options.
For the 6062A,
the
first number
is always
4, which indicates that the Generator being
interrogated is in fact a 6062A. The two remaining numbers are the sum of the option related
numbers as follows:
2nd Number
4 = -130 High-Stability Reference Option
8 = -651 Low-Rate FM Option
16 = -132 Medium-Stability Reference Option
3rd Number
2 = -830 Rear Panel RF Output, MOD Input, and Pulse MOD Input Option
For example, “4,12,2” would indicate a 6062A with Options -651, -130, and -830.
The “IR” command interrogates the Generator for rejected entry error status. (See Table 2-6
for a list of rejected entry codes.) The returned record
is the sum of errors that have been
detected while processing IEEE-488 commands. The status is cleared when interrogated with
the “IR” command. The status can also be explicitly cleared with the “CE” command and is
also cleared on the “CL” command and the clear interface messages DCL and SDC.
The “IS” command interrogates the Generator for the serial number. The serial number is an
eight character ASCII string terminated with an
. The returned record is similar to the
following:
“01230000” (EOR)
The “IT” command interrogates the Generator for the self-test results. The self-test results are
reported in four fields. Any non-zero codes indicate that some tests have failed. Table 2-14
shows the self-test error code format. The self tests are performed at power-on and can also be
initiated with special function 02. Further details of the self-test results are listed in the Section
4 of this manual.
Table 2-14. Self-Test Error Codes
MOD
FREQ
AMPL
aaa
-bbb
ccc
ddd
aaa
= Modulation
(AM and
FM)
test
results
“
=
If
displayed;,
indicates the
self
test did not
complete
bbb
=
Frequency
test
results
ccc
= Digital
test
results
ddd
= Amplitude
test
results
The “lU” command interrogates the Generator for the UNCAL entry status. (See Table 2-5 for
a list of the UNCAL Error Codes.)
The “IV” command interrogates the Generator for its current software revision level. The
returned record is similar to the following:
“VI.O” (EOR>.
This means that the software revision level
is Version
l.O.
The “IZ” command interrogates the Generator for the compensation memory status. The
status is returned as three fields similar separated by commas, to the following:
“000010,100001,000000”
Table 2-15. Input/Output Monitor Commands
COMMAND
NAME
COMMAND
VALUE
DESIGNATOR
NAME
LOCATION
Read
Input
Bit
"IB"
"EX"
exref
A2,
U11-7
"HS"
hsopt
A2,
U11-17
"IE"
ieinl
A2,
U11-8
"LF"
1 rfm
A2,
U40-8
"MH"
mlevhi
A2,
U40-13
"ML"
m 1 ev 1 0
A2,
U40-7
"RO"
roptl
U40-17
"RP"
rppl
A2,
U11-11
"RT"
rptrpl
A2^
U11-14
"SR"
shre-f L
A2,
U40-4
"TQ"
trseqL
A2,
U40-3
"UK"
unLok
A2,
U11-3
"UV"
unlvL
A2^
U11-18
"XA"
extra-a
A2,
U40-14
"XB"
extra-b
kZ,
U40-18
Set Output
Bit
"OB"
0
or
1
"CL"
cLr
kZ,
U17-12
"DA"
dcam
kb,
U13-19
"EA"
extam
kb,
U13-12
"EF"
extf m
kb,
U13-6
"EP"
ext
pulse
kb,
U13-16
"FE"
fmen
A4,
U19-19
"FP"
front-panel
kZ,
U17-5
"HO"
haocth
kb,
U9-2
"HT"
het
kb,
U9-6
"lA"
intam
kb,
U13-9
"IF"
intfm
kb,
U13-5
"IP"
int
pulse
kb,
U13-15
"LP"
mid
kb,
U9-5
"MA"
rmuxO
kk,
U32-5
"MB"
rm uxl
A4,
U32-2
"MF"
mf400l
kb,
U13-2
"MG"
module-gate
kZ,
U17-15
"MS"
trmodl
A4,
U32-15
"NE"
nven
kZ,
U17-9
"PM"
phase mod
kb,
U23-12
"RI"
rinh
A4,
U32-9
"RR"
rprst
A2,
U17-6
"SE"
shen
kk,
U32-12
"SH"
shet
A4,
U32-16
"VH"
vcohl
kb,
U23-9
"VQ"
vcoq
kb,
U23-15
"XA"
x2l
kb,
U9-19
"XB"
x2fl1
kb,
U9-16
"XC"
x2fl2
A6^
U9-15
"XD"
x2fl3
A6,
U9-12
"XE"
x2fl4
A6,
U9-9
"XO"
xoenl
kU,
U32-6
"YA"
extra-a
kZ,
U17-19
"YB"
extra-b
A2^
U17-2
"YC"
extra-c
A2^
U17-16
Set Output
DAC
"OD"
(integer value)
(DAC
Location)
0-255
"AM"
AM dac
kb
U19
0-127
"AT"
attenuator
kZ,
U27-5,
b,
9,
12, 15,
16, 19
0-4095
"FM"
FM
dac
kb,
U25-4
through
15
0-7
"FR"
FM
range
A4,
U19-12,15,16
0-1023
"KN"
kn dac
A4,
U27-4 through 13
0-1023
"KV"
kv dac
A4,
U29-4 through 13
0-16383
"LE"
level dac
kb,
U21
0-255
"TC"
temp. comp. dac
A6,
U19
Table 2-16. Read/Write Monitor Commands
COMMAND
NAME
COMMAND
SYNTAX
NOTES
Read byte
Read word
Define write address
Write byte
Write word
"R8" memory
locati on
*'RW" memory
locati on
"DW" memory
locati on
"WB” value
W value
1
1
2
2
2
Notes
1
.
The Generator
or word
in the
be
an unsigned
If
the memory
is
returned
in
responds
to these
commands with
the value
of
the byte
memory
location addressed.
The memory
location must
integer.
The value
returned
is
fol lowed
by
an .
Location
is
specified
in hexadecimal,
then the value
hexadecimal
preceded
by
an
"X".
2.
The
Define Write
Address
command
specifies the write address used
with
the Write Byte and Write Word
commands.
When the Write Byte and
Write Word commands are
used,
the
specified data will
be written to
that write address.
Table 2-17. Hardware Control Monitor Commands
COMMAND NAME
COMMAND
SYNTAX
NOTES
Read attenuati on
"XA"
1
Set
attenuati on
"XB" integer
1
Set
synthesizer frequency
"XD"
f loati ng point decimal
"MZ"
2
RF on/off
"XR"
"0" or "1"
3
Notes
1
.
.The
current
settings
of
the Attenuator
can
be
read or
set.
The
attenuation
is
a
number from
0
to 23 where the number
specifies
the
multiple
of 6-dB
attenuation.
Zero indicates no attenuation,
and
23
is
the maximum attenuation.
Only
the attenuators are changed with
the
"XB"
command.
The value
of
the attenuation will
be output
on the "XA"
command
as
an unsigned
integer followed
by
.
2.
The
"XD"
command
can
be
used to program
the Generator
to the specified
frequency.
Only
the synthesizer
circuits on the Synthesizer
PCS
is
programmed.
No offset
is added,
no filters are programmed,
no VCO
compensation
is
calculated,
and no
level
correction is caUulated-
3 .
"XRO" programs all
attenuation,
and
"XR1" restores the attenuator
to
its previous
state.
The SRQ mask is set to the sum of the reason values listed in Table 2-18. These reason values
correspond to the allowable reasons that will be requiring service. The SRQ Mask is set by the
following command sequence;
SM” Sum of Reasons
The Generator asserts SRQ when one of the allowed reasons becomes true. The serial poll
response
is the sum of those values for reasons that are currently true, independent of the
seUing of the SRQ mask. For example, if the rejected entry SRQ is enabled with “SM2” and a
rejected entry occurs, the serial poll response will indicate that the Generator generated the
SRQ (value of 64) and that a rejected entry occurred (value of 2). In addition, other values may
be set. The default SRQ mask is 192.
The “IM” command interrogates the current SRQ mask, and an integer is returned.
The “XF” command enables a mode that causes an SRQ to be generated any time the
Generator processes an entry. In this mode, afront panel SRQ is generated (i.e., the serial poll
response indicates that afront panel SRQ is the cause of the SRQ). This mode is enabled and
disabled with its own command, not through the Set SRQ Mask commands (as are all other
SRQs). The Alert Mode is enabled/ disabled as follows:
“XFO” = Alert Mode off, “XFl” = Alert Mode on.
2-61.
TRIGGER COMMANDS
The Generator has the
ability to preprogram a command string of arbitrary Generator
programming commands. This command string is executed whenever the trigger command
“TR” orthe IEEE-488 group execute trigger interface message (GET) is received. This method
of programming the Generator can be used when a long string of commands is to be sent to the
Generator over and over. The programming time is shortened by the time required to transmit
the string of characters from the controller to the Generator.
The trigger commands are as follows;
“CT” Configure Trigger
“TR” Execute Trigger Buffer
The configure trigger command
is followed by
a string of any Generator programming
commands up to 71 characters in length. The validity of the programming commands
is not
checked until the trigger buffer is executed. The power-on value of the trigger string is null
(nothing).
The trigger command causes execution of the trigger buffer, which has been preprogrammed
with the configure trigger command “CT”. The trigger buffer can also be executed by sending
the IEEE-488 group execute trigger interface message (GET).
The following paragraphs describe how IEEE-488 commands are processed by the Generator.
Command processing is a term for how commands are executed and how errors are handled.
2-63.
COMMAND EXECUTION
The execution of the IEEE-488 commands depends on the selection of interface modes with
one exception;
if an IEEE-488 input
is buffered and
the
buffer becomes
full, command
execution starts and no further input is accepted until there is room in the input buffer. For
more details, refer to the paragraphs on Interface Modes.
2-64.
ERROR HANDLING
The Generator detects two types of errors while processing IEEE-488 commands; syntax
errors and processing errors. All errors are accumulated until the error status is interrogated or
is
explicitly
cleared. The IEEE-488
rejected
entry
status
is
interrogated
with
the “IR”
command. The error status is cleared with one of the following commands:
“CE” Clear Error Command
“CL” Clear Command
“DCL” or “SDC” Clear Interface Messages
The error status is also cleared at power-on.
The SRQ mask can be set to assert SRQ when an error is detected, The SRQ is unasserted
when the error status
is cleared.
Syntax errors are commands that do not have the correct syntax for the specified header. For
example, “FES”
is a syntax error because the external FM command requires a Boolean
numeric field. Unrecognized headers are also syntax errors. An IEEE-488 syntax error causes
all commands from the point of the error up to the next string terminator or record terminator
to be ignored.
Processing errors are commands that are syntactically valid, but the requested value is outside
the range of programmable values. For example, “FR99GZ” is syntactically correct, but the
Generator cannot be programmed to a frequency of 99 gigahertz. Command processing
continues with the next command.
2-65.
Timing Data
The programming time can be broken down into four-groups: transfer of commands to
Generator, command parsing time, software programming time, and instrument settling time.
The total programming time depends on the selection of the interface modes. In some modes,
programming steps are performed in parallel and can increase throughput. The following
paragraphs give some typical timing data for the above four programming steps and describes
how the interface modes affect their relative timing.
2-66.
TRANSFER OF COMMANDS TO GENERATOR
The maximum rate of transfer is 0.4 to 0.5 ms per character. With most IEEE-488 controllers,
all characters sent with a single output or print statement
is transferred together
at
the
maximum
rate. The
total time
to transfer commands
to
the Generator
is obtained by
multiplying the number of characters by the rate of transfer.
2-67.
COMMAND-PARSING TIME
Command-parsing time is the sum of the time required to process the header, the numeric, and
the suffix. Some commands do not have numerics or suffixes. Table 2- 1 9 gives the typical time
it takes to process the different components of a command.
2-68.
SOFTWARE PROGRAMMING TIME
The minimum time required to process a command is 25 ms. Most of the commands that do
not program the hardware (such as storing step values) are programmed in 25 ms. Table 2-20
gives the typical time value for programming the different functions in the Generator.
2-69.
INSTRUMENT-SETTLING TIME
The software-programming time
typically exceeds the
setting time
of the RF
circuitry,
therefore the instrument-settling time can generally be excluded from throughput calculations.
The exception
is when level correction
is disabled. Disabling level correction reduces the
software-programming time by 10 ms but does not affect the switching time of the frequency
programming circuitry. A 10 ms settling delay should be added if level correction is disabled
and a settled RF output is required.
2-70.
TIMING OPTIMIZATION
Timing depends upon the interface modes selected. Refer to the paragraphs on Interface Mode
Commands for a complete description of the interface modes.
The transfer of commands from the IEEE-488 controller to the Generator can never be
processed
in
parallel
with
anything
else. The
transfer
of commands
usually happens
simultaneously, regardless of which interface mode is selected.
The parsing of the command and programming the new instrument state is performed one
message unit at a time. The record mode selects a command or a record as the message unit.
The record mode off (“RMO”) is slower since there is extra processing between message units,
and the message unit is smaller.
If the valid mode is enabled, the processing of message units is delayed until the Generator has
settled from the previous message. While the RF output is typically settled before the software
finishes processing the command, enabling the valid mode adds an additional delay to ensure
that the output is settled before the next command
is processed.
If the output of the Generator does not need to be settled between programming strings, the
valid mode should be turned off to speed up processing.
If the output does not need to be
settled between commands, but needs to be settled between records, enable the valid mode and
the record mode. The instrument processes commands within the record as fast as possible and
waits for the output to settle only between records.
The instrument parameters at power-on are listed in Table 2-21
. The remote clear commands
can be used to reset all parameters except the last memory location and the remote/ local state.
Table 2-19. Command-Parsing Time
COMMAND
COMPONENT
TIME
Header
Boolean
Numeric
Unsigned
Integer Numeric
Floating-Point
Numeric
Trigger-String Numeric
Suffix
Learn-Interf ace
Suffix
2 ms
1
ms
2 ms
+
1
ms per
character
2 ms
+
1
ms per
character
10 ms +
0.5 ms per
character
1
to
1 .5 ms
40 ^ms
Table 2-20. Typical Programming Time of the Generator Functions
1
FUNCTION
TIME
(IN
ms)
NOTES
Frequency
70
1,
2,
3,
4
Amplitude
in Volts
60
1,
2,
4,
5
Amplitude
in dBm
95
1,
2,
4,
5
AM
Depth
50
2.
4
FM Deviation
35
4
Modulation
Frequency
25
Enable/Disable AM
55
2,
Enable/Di sable
FM
25
Recall 98
190
6,
5
RF Output
On
50
2
RF Output
Off
35
Notes
1.
May
take up
to
5 ms
longer
if
the
relative mode
is
enabled,
2.
Can
save 10 ms
if
all
Level correction
is disabled with
special
function
81
.
3.
Add
20 ms when
frequency
changes
from
greater
than
or
equal
to 245
MHz
to
less than 245
MHz.
4.
Edits
and
steps may
take up to
5 ms
longer than
the programming
function directly,
5.
Add
15 ms when
the
Attenuator
settings
change,
6.
Recalls
vary considerably depending on
the
stored data.
Maximum
is
approximately 250
ms.
Table 2-21. IEEE-488 Power-On State
INSTRUMENT PARAMETER
STATE
NOTES
Memory
location parameters
same
as
power
off
1
RF
on/off
same
as
power off
Last memory
Location
same
as
power
off
2
Remote/Local
state
local
IEEE
output
buffer
cleared
IEEE
input
buffer
cleared
Valid mode
same
as power
off
Record terminator
same
as power
off
Unbuffered mode
same
as power
off
Record mode
same
as
power off
SRQ mask
same
as
power
off
Trigger configuration
same
as power
off
SRQ
interface
signal
may be asserted
3
Notes
1.
The
contents of memory
location 98
(Instrcmient
Preset
State)
is
listed
in Table 2-3.
2.
The
last memory
location is used
for
sequence operations.
3,
If
the
SRQ mask
has
the power
on,
output
valid,
or
ready
1>RQ enabled.
the SRQ
interface
signal
is asserted on power-on.