Agilent Technologies 7820A Manual
Text of the manual
© Agilent Technologies, Inc. 2016
No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or transla-
tion into a foreign language) without prior
agreement and written consent from Agi-
lent Technologies, Inc. as governed by
United States and international copyright
laws.
Manual Part Number
G4350-90014
Edition
Fifth edition, August 2016
Fourth edition, June 2015
Third edition, November 2014
Second edition, May 2011
First edition, March 2009
Printed in China
Agilent Technologies (Shanghai) Co., Ltd.
412 Ying Lun Road
Waigaoqiao Free Trade Zone
Shanghai 200131 P.R. China
Warranty
The material contained in this docu-
ment is provided “as is,” and is sub-
ject to being changed, without notice,
in future editions. Further, to the max-
imum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a par-
ticular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or perfor-
mance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the sep-
arate agreement shall control.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Maintaining Your GC
3
Contents
1
About Maintaining the GC
Overview of Maintenance
10
Tools and Materials Required for Maintenance
11
Maintenance Methods for the 7820A GC
13
Safety Information
15
Finding a Replacement Part Number
16
2
Removing Covers
To Remove the Detector Top Cover
18
To Remove the Pneumatics Cover
19
To Remove the Side Cover
20
To Remove the Electronics Cover
21
3
Maintaining Capillary Columns
Consumables and Parts for Columns
24
To Install a Capillary Column Hanger
26
To Install Capillary Column Clips
27
To Condition a Capillary Column
28
To Cut a Loop from a Column
31
To Reverse a Column and Bakeout Contaminants
32
4
Maintaining the Split/Splitless Inlet
Consumables and Parts for the Split/Splitless Inlet
36
Exploded Parts View of the Split/Splitless Inlet
39
To Install a Capillary Column with the Split/Splitless Inlet
40
To Change the Septum on the Split/Splitless Inlet
44
To Clean the Septum Seat in the Insert Assembly of the Split/Splitless
Inlet
46
To Change the Liner and O-Ring on the Split/Splitless Inlet
48
To Replace the Gold Seal on the Split/Splitless Inlet 50
To Replace the Filter in the Split Vent Line
52
To Clean the Split/Splitless Inlet
55
To Bakeout Contaminants from the Split/Splitless Inlet
57
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
Consumables and Parts for the Purged Packed Inlet
60
Exploded Parts View of the Purged Packed Inlet
63
To Install a Capillary Column with the Purged Packed Inlet
64
To Change the Septum on the Purged Packed Inlet
68
To Clean the Septum Seat in the Purged Packed Inlet
70
To Install an Adapter on the Purged Packed Inlet
72
To Change the O-Ring on the Purged Packed Inlet
74
To Change the Glass Liner on the Purged Packed Inlet
75
To Install an Insulation Cup on the Purged Packed Inlet
77
To Clean the Purged Packed Inlet 78
To Bakeout Contaminants from the Purged Packed Inlet
80
To Install a Packed Metal Column
81
To Install a Packed Column Adapter on a Detector Fitting
83
To Install a Packed Glass Column
85
To Condition a Packed Column
88
To Install Ferrules on a Packed Metal Column
90
6
Maintaining the Packed Column Inlet
Consumables and Parts for the Packed Column Inlet
92
Exploded Parts View of the Packed Column Inlet
94
To Change the Septum on the Packed Column Inlet
95
To Clean the Septum Seat in the Packed Column Inlet
97
To Install an Adapter on the Packed Column Inlet
99
To Change the O-Ring on the Packed Column Inlet
101
To Change the Glass Liner on the Packed Column Inlet
102
To Install an Insulation Cup on the Packed Column Inlet
104
To Clean the Packed Column Inlet 105
To Bakeout Contaminants from the Packed Column Inlet
107
To Install a Packed Metal Column
108
To Install a Packed Column Adapter on a Detector Fitting
110
To Install a Packed Glass Column
112
To Condition a Packed Column
115
To Install Ferrules on a Packed Metal Column
117
Maintaining Your GC
5
7
Maintaining the COC Inlet
Consumables and Parts for the COC Inlet
120
Exploded Parts View of the COC Inlet
124
To Install a Capillary Column with the COC Inlet
125
To Check the Needle-to-Column Size on the COC Inlet
128
To Change a Septum on the COC Inlet
130
To Install an Insert on the COC Inlet
132
To Clean the COC Inlet
134
To Replace the 7693A Injector Needle Support Guide
136
To Replace the Needle Support Assembly in a 7683B Injector
137
To Replace a Needle in a Syringe
140
To Replace the Fused Silica Needle in a Syringe for the COC Inlet
141
To Bakeout Contaminants from the COC Inlet
143
8
Maintaining the FID
Consumables and Parts for the FID
146
Exploded Parts Views of the FID
149
Selecting an FID Jet
151
To Attach a Capillary Column Adapter on an Adaptable FID
153
To Install a Capillary Column in the FID
155
To Replace the FID Collector Assembly
158
To Replace an FID Jet
161
To Perform Maintenance on the FID Collector Assembly
165
To Check the FID Leakage Current
173
To Check the FID Baseline
174
To Install the FID Insulation Cup Assembly (Adaptable FID Only)
175
To Install the Optional FID PTFE Chimney Insert
177
To Bakeout the FID
178
9
Maintaining the TCD
Consumables and Parts for the TCD
182
To Install a Capillary Column in the TCD
184
To Install the Optional TCD Capillary Column Adapter
186
To Install a Capillary Column with the Optional TCD Capillary Column
Adapter
187
Maintaining Your GC
To Bakeout Contaminants from the TCD
190
10
Maintaining the µECD
Important Safety Information About the µECD
192
Consumables and Parts for the µECD
194
Exploded Parts View of the µECD
196
To Replace the µECD Fused Silica Indented Mixing Liner and Install the
Makeup Gas Adapter
197
To Install a Capillary Column in the µECD
200
To Install the Insulating Cup for the µECD
203
To Bakeout the µECD
205
11
Maintaining the NPD
Consumables and Parts for the NPD
208
Exploded Parts View of the NPD
211
Selecting an NPD Jet
212
To Attach a Capillary Column Adapter on an Adaptable NPD
214
To Install a Capillary Column in the NPD
216
To Replace the NPD Bead Assembly
219
To Maintain the NPD Collector, Ceramic Insulators, and Jet
224
To Check the NPD Leakage Current
230
To Bakeout the NPD
231
12
Maintaining the FPD+
Consumables and Parts for the FPD+ 234
Exploded Parts View of the FPD+ 236
To Install a Packed Column Adapter in the FPD+ 237
To Attach a Capillary Column to the FPD+ 239
To Change the FPD+ Wavelength Filter
241
To Remove the FPD+ Cover
244
To Replace the FPD+ Ignitor
246
To Install the FPD+ Cover
248
Cleaning the FPD+ Brazement
249
13
Maintaining the G3435A/G3436A FPD
Consumables and Parts for the FPD
252
Maintaining Your GC
7
Exploded Parts View of the FPD
254
To Install a Capillary Column Adapter in the FPD
255
To Attach a Capillary Column to the FPD
257
To Change the FPD Wavelength Filter
259
To Remove the FPD Vent Tube
262
To Replace the FPD Ignitor
264
To Install the FPD Vent Tube and Cover
266
14
Maintaining the PCM
Consumables and Parts for the PCM
268
Calibrating the PCM Interface
269
Installing or Replacing Frits in the PCM
270
15
Maintaining a Valve
Consumables and Parts for Valves
274
Exploded Parts View of GC Rotary Valves
275
To Replace a Gas Sampling Valve Loop
276
To Align a Rotary Valve Rotor
278
To Replace a Rotary Valve in the Valve Box
279
To Remove the Upper Valve Box
282
To Install the Upper Valve Box
284
A
Swagelok Connections
Making Swagelok Connections
288
Using a Swagelok Tee
292
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
1
About Maintaining the GC
Overview of Maintenance
10
Tools and Materials Required for Maintenance
11
Maintenance Methods for the 7820A GC
13
Safety Information
15
Finding a Replacement Part Number
16
This section provides an overview of the maintenance
procedures included in this document. It also lists the tools
needed for routine maintenance and the safety information one
should be aware of before performing a maintenance task.
Maintaining Your GC
1
About Maintaining the GC
Overview of Maintenance
This manual details the routine tasks needed to maintain the
7820A Gas Chromatograph (GC). The procedures assume a
basic knowledge of tool use and of GC operation. Readers are,
for example, expected to know how to:
• Safely turn devices on and off
• Load methods
• Change component temperatures, flows, and pressures
• Make typical pneumatic connections using Swagelok and
other standard fittings
• Reset GC service counters
Where to find a procedure
Included in this manual are chapters on maintaining the
following GC components:
• Capillary Columns
• Split/Splitless Inlet
• Purged Packed Inlet
• Packed Column Inlet
• COC Inlet
• FID
• TCD
• µECD
• NPD
• FPD+
• FPD
• PCM
• Valves
Each chapter includes:
• A list of the most commonly used consumables and parts for
the component
• An exploded parts view of the component
• Detailed procedures for routine maintenance tasks
associated with the component
About Maintaining the GC
1
Maintaining Your GC
11
Tools and Materials Required for Maintenance
Table
1 lists the tools needed for most GC maintenance
procedures. The specific tools required to perform a
maintenance procedure are listed in step 1 of the procedure.
Table 1
Tools and materials for GC maintenance
Common tools
Wrench, angled, septum nut (19251-00100)
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)*
Wrench, open-end, 9/16-inch and 7/16-inch (8710-0803)
Wrench, capillary inlet ( G3452-20512 )*
Flathead screwdriver
Column cutter, wafer (5181-8836, 4/pk)*
Driver, nut, 1/4-inch (8710-1561)*
T-20 Torx key (8710-1807) or screwdriver
T-10 Torx key (8710-2140) or screwdriver
3-mm hex key wrench (8710-2411)
Electronic flow meter(s) or bubble meter(s) capable of calibrated measurements
at 1, 10, and 100
mL/min flow ranges.
Electronic leak detector
Magnifying loupe, 20X (430-1020)
Metric ruler
Bench vise (for setting Swagelok fittings)
Razor or sharp knife
Tweezers (8710-0007) or thin needle-nose pliers (8710-0004)
Needle-nose pliers
ESD wrist strap (for installing new components)
Gloves, heat-resistant (for handling hot parts)
Wooden cotton swab (for removing FID filters)
Tools and materials for cleaning procedures
Cleaning brushes—The FID cleaning kit (9301-0985) contains appropriate
brushes for cleaning detectors and inlets
Cleaning brushes—(8710-1346) For cleaning split/splitless inlet split vent
fitting, FID and collectors
Maintaining Your GC
1
About Maintaining the GC
Jet cleaning wire (.010 inch)
Clean, lint-free cloth (to protect contamination-sensitive detector parts)
Small ultrasonic cleaning bath with aqueous detergent (for cleaning detector
and inlet parts)
Gloves, clean, lint-free, nylon (large: 8650-0030, small: 8650-0029) (for handling
contamination-sensitive parts)
Steel wool, 0- or 00-grade (for cleaning an inlet’s septum seating surfaces)
* Included with the GC ship kits
Table 1
Tools and materials for GC maintenance (continued)
About Maintaining the GC
1
Maintaining Your GC
13
Maintenance Methods for the 7820A GC
Before most maintenance procedures, the GC must be made
ready. Agilent recommends that you create and store the
following maintenance methods into the GC. The methods below
will:
• Prevent damage to the instrument (electronics, columns,
etc.)
• Avoid injury to the user (burns, shocks, etc.)
• Allow you to perform maintenance on specific areas while
leaving the rest of the GC components at operating
temperature
You can use the software keyboard or your Agilent data system
to create, save, and load these methods.
General GC Maintenance Method
Create this method for instrument column maintenance,
detector maintenance, and general GC maintenance tasks.
• Set the oven temperature to 35 °C. This allows the oven fan to
assist cooling.
• Set all inlet temperatures to 35 °C and set inlet gas pressures
to 0.0.
• If performing column maintenance, remember to wait for
the oven and column to cool down before turning off
column carrier gas flow at the source. Also remember to
cap both ends of the column to keep air out once it is
removed.
• If you are not performing column maintenance, keep inert
carrier gas (helium or nitrogen) flowing to protect the
column.
• Set all detector temperatures to 35 °C.
• If performing FPD+ maintenance, turn off the GC and
unplug the power cord.
• Some detectors (FID, NPD) use high voltages. For these
detectors, turn the electrometer Off to disable the high
voltage.
NOTE
Inlets and detectors at operating temperature may require 12 hours or
longer to reach the maintenance method setpoints below.
Maintaining Your GC
1
About Maintaining the GC
• The filament in the TCD will be damaged if exposed to air
while hot. To protect the filament, turn it Off.
• Set all detector flows to Off.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
Inlet Maintenance Method
This method prepares the inlet for maintenance while leaving
the detector at operating temperature.
• Set the oven temperature to 35 °C. This allows the oven fan to
assist cooling.
• Set all inlet temperatures to Off and set inlet gas pressures to
0.0.
• If performing column maintenance, remember to wait for
the oven and column to cool down before turning off
column carrier gas flow at the source. Also remember to
cap both ends of the column to keep air out once it is
removed.
• If you are not performing column maintenance, keep inert
carrier gas (helium or nitrogen) flowing to protect the
column.
• Maintain all temperature setpoints for installed detectors, if
desired.
• The filament in the TCD will be damaged if exposed to air
while hot. To protect the filament, turn it Off.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
About Maintaining the GC
1
Maintaining Your GC
15
Safety Information
Before performing a maintenance task, read the important
safety and regulatory information found in the Safety and
Regulatory Information book.
Maintaining Your GC
1
About Maintaining the GC
Finding a Replacement Part Number
Agilent now provides the Parts Finder tool to help you locate
replacement and consumable part numbers. If using an Agilent
data system, Parts Finder will be installed. If you want to install
the tool on another computer, Parts Finder is included on the
Agilent GC and GC/MS User Manuals & Tools DVD.
To find a consumable or replacement part using Parts Finder,
navigate to the part graphically, based on the part’s location in
the GC.
Figure 1
Quickly navigate to replacement parts by clicking on images of
GC components
Part numbers are also included in this manual.
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
2
Removing Covers
To Remove the Detector Top Cover
18
To Remove the Pneumatics Cover
19
To Remove the Side Cover
20
To Remove the Electronics Cover
21
This section describes how to remove covers as needed for
routine maintenance.
Only the covers listed in this chapter should be removed.
Removing other GC covers can compromise the safety features
of the GC, leading to personal injury or damage to the
instrument.
Maintaining Your GC
2
Removing Covers
To Remove the Detector Top Cover
This cover protects the detectors, valve box, and valve assembly.
To remove the detector top cover:
1
Raise it to a vertical position
2
Lift the right side and disengage the pin on the lower left
side.
CAUTION
Do not force the cover, either when installing it or closing it. This
could break the plastic parts.
Maintaining Your GC
19
To Remove the Pneumatics Cover
The pneumatics cover protects the flow manifolds in the back
top of the GC.
1
Disconnect any vent tubing connected to the split and
septum purge vents.
2
Raise and remove the detector top cover.
3
Remove the screw on the left side of the pneumatics cover
that secures the cover to the GC.
4
Loosen the one or two screws on the back of the GC that
secure the pneumatics cover to the GC. (Depending on the
date of manufacture, the number of screws securing the
pneumatics cover to the GC varies.)
5
Lift and remove the cover.
Maintaining Your GC
2
Removing Covers
To Remove the Side Cover
The side cover protects the GC main board and detector boards.
1
Remove the screw on the right side of the GC that secures the
side cover to the GC
2
Remove the one or two screws on the back of the GC that
secure the side cover to the GC. (Depending on the date of
manufacture, the number of screws securing the side cover
to the GC varies.)
3
Slide the side cover toward the back of the GC.
4
Tilt the top of the side cover away from the GC and then lift
the side cover off of the GC.
Maintaining Your GC
21
To Remove the Electronics Cover
You may need to remove the electronics cover to perform NPD
maintenance. The steps required depend on whether or not an
FPD+ is installed.
Without FPD+/FPD installed:
1
Raise or remove the detector top cover.
2
Remove the side cover.
3
Remove the screw on the left side of the electronics cover.
4
Loosen the screw on the back of the electronics cover.
5
Remove the cover.
With FPD+/FPD installed:
1
Raise or remove the detector top cover.
2
Remove the PMT tube. See To Change the FPD+ Wavelength Filter
or To Change the FPD Wavelength Filter as applicable.
3
Remove the heat shield (the metal “bracket” below the PMT
tube) by removing the four screws that secure it to the front
bracket/support.
4
Remove the side cover.
5
Remove the screw on the left side of the electronics cover.
6
Loosen the screw on the back of the electronics cover.
7
Remove the cover.
CAUTION
Raising the electronics cover exposes the GC electronics.
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
3
Maintaining Capillary Columns
Consumables and Parts for Columns
24
To Install a Capillary Column Hanger
26
To Install Capillary Column Clips
27
To Condition a Capillary Column
28
To Cut a Loop from a Column
31
To Reverse a Column and Bakeout Contaminants
32
Maintaining Your GC
3
Maintaining Capillary Columns
Consumables and Parts for Columns
See the Agilent catalog for consumables and supplies for a more
complete listing, or visit the Agilent Web site for the latest
information (www.agilent.com/chem/supplies).
Table 2
Nuts, ferrules, and hardware for capillary columns
Column id (mm)
Description
Typical use
Part number/quantity
.530
Ferrule, Vespel/graphite,
0.8-mm id
0.45-mm and 0.53-mm capillary
columns
5062-3512 (10/pk)
Ferrule, graphite, 1.0-mm id
0.53-mm capillary columns
5080-8773 (10/pk)
Ferrule, graphite, 0.8-mm id
0.53-mm capillary columns
500-2118 (10/pk)
Column nut, finger-tight (for
0.53-mm columns)
Connect column to inlet or detector
5020-8293
.320
Ferrule, Vespel/graphite,
0.5-mm id
0.32-mm capillary columns
5062-3514 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
.250
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
.100 and .200
Ferrule, Vespel/graphite,
0.37-mm id
0.1-mm and 0.2-mm capillary
columns
5062-3516 (10/pk)
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Ferrule, graphite, 0.4-mm id
500-2114 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
All
Ferrule, no-hole
Testing
5181-3308 (10/pk)
Capillary column blanking nut
Testing–use with any ferrule
5020-8294
Column nut, universal
Connect column to inlet or detector
5181-8830 (2/pk)
Maintaining Capillary Columns
3
Maintaining Your GC
25
Column cutter, ceramic wafer
Cutting capillary columns
5181-8836 (4/pk)
Pencil, diamond tipped
Cutting capillary columns
420-1000
Ferrule tool kit
Ferrule installation
440-1000
Table 2
Nuts, ferrules, and hardware for capillary columns (continued)
Column id (mm)
Description
Typical use
Part number/quantity
Table 3
Capillary column hangers
Description
Part Number
Column hanger
1460-1914
Capillary column clip kit, for 7-inch column basket
G1530-61580
Maintaining Your GC
3
Maintaining Capillary Columns
To Install a Capillary Column Hanger
1
Select either the front or back hanger position. (Hanger is
shown in back position.)
2
Insert the ends of the hanger into the slots in the selected
position.
WARNING
Be careful! The oven may be hot enough to cause burns. If the
oven is hot, wear heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Front position
Back position
Maintaining Capillary Columns
3
Maintaining Your GC
27
To Install Capillary Column Clips
1
Gather the following:
• Capillary column clip kit (see “Consumables and Parts for
Columns” on page
24.)
• T-20 Torx screwdriver
2
Loosen the four corner mounting screws, but do not remove.
3
Slip each corner screw through the large hole on the clip.
4
Slide the clip so that the screw is positioned in the slot.
5
Tighten the screws enough to hold the clips in place. Once
the column is installed, fully tighten the four corner screws
to secure the clips and column to the oven wall.
WARNING
Be careful! The oven may be hot enough to cause burns. If the
oven is hot, wear heat-resistant gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Column
clip
Mounting
screw
Maintaining Your GC
3
Maintaining Capillary Columns
To Condition a Capillary Column
1
Gather the following:
• One 7/16-inch, and 1/4-inch wrenches
• No-hole ferrule (See “Consumables and Parts for
Columns” on page
24.)
• Column nut
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Install the column into the inlet using the new ferrules. See:
• To Install a Capillary Column with the Split/Splitless Inlet
• To Install a Capillary Column with the Purged Packed
Inlet
• To Install a Capillary Column with the COC Inlet
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Maintaining Capillary Columns
3
Maintaining Your GC
29
4
Cap the detector column fitting.
5
Set a minimum velocity of 30 cm/s, or as recommended by
the column manufacturer. Let gas flow through the column at
room temperature for 15 to 30
minutes to remove air.
6
Program the oven from room temperature to the maximum
temperature limit for the column. Increase the temperature
at a rate of 10 to 15
°C/min. Hold at the maximum
temperature for 30
minutes.
7
Load the GC maintenance method and wait for the GC to
become ready.
8
Attach the column to the detector. For details, select your
specific detector from the following list:
• To Install a Capillary Column in the FID
• To Install a Capillary Column in the NPD
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the µECD
• To Attach a Capillary Column to the FPD+
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Maintaining Your GC
3
Maintaining Capillary Columns
9
Restore the analytical method.
• For FID or any FPD+, immediately turn off the flame.
• For NPD, immediately turn off the bead.
10 After the GC becomes ready, wait 10 minutes, then ignite the
detector flame or bead.
Maintaining Capillary Columns
3
Maintaining Your GC
31
To Cut a Loop from a Column
1
Gather the following:
• New ferrule(s) for the column inlet connection
• Column cutter
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Loosen the inlet column nut and remove the column from the
inlet.
4
Uncoil one loop of column from the column hanger.
5
Cut the unwanted loop from the column.
6
Install the column into the inlet using the new ferrules. See:
• To Install a Capillary Column with the Split/Splitless Inlet
• To Install a Capillary Column with the Purged Packed
Inlet
• To Install a Capillary Column with the COC Inlet
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Maintaining Your GC
3
Maintaining Capillary Columns
To Reverse a Column and Bakeout Contaminants
1
Gather the following:
• 1/4-inch wrench
• Column cutter
2
Load the maintenance method and wait for the GC to become
ready.
3
Disconnect the column from the inlet and detector.
4
If necessary, cut a loop from the column. (See “To Cut a Loop
from a Column” on page
31.) Do not attach the column to the
inlet.
5
Remove the column from the hanger and reverse its position
(inlet and detector ends) and place the column back on the
hanger.
6
Attach the column to the inlet.
Select your specific inlet from the following list:
• To Install a Capillary Column with the Split/Splitless Inlet
• To Install a Capillary Column with the Purged Packed
Inlet
• To Install a Capillary Column with the COC Inlet
7
Attach your column to the detector.
Select your specific detector from the following list:
• To Install a Capillary Column in the FID
• To Install a Capillary Column in the NPD
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the µECD
• To Attach a Capillary Column to the FPD+
WARNING
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Maintaining Capillary Columns
3
Maintaining Your GC
33
8
Set the column flow to the normal operating value, or set the
capillary column gas velocity to 30
cm/s.
For Split/Splitless inlets, select split mode and set the split
vent flow to 200
mL/min.
9
Purge the column with carrier flow for at least 10
minutes
before heating the oven.
10 Set the inlet temperature to 300
°C or 25
°C above the
normal operating temperature.
11 Set the column oven 25
°C above the GC method final oven
temperature to bake contaminants out of the inlet, mostly
through the split vent. Do not exceed the column
manufacturer’s maximum temperature limit.
12 Bakeout for 30
minutes.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
4
Maintaining the Split/Splitless Inlet
Consumables and Parts for the Split/Splitless Inlet
36
Exploded Parts View of the Split/Splitless Inlet
39
To Install a Capillary Column with the Split/Splitless Inlet
40
To Change the Septum on the Split/Splitless Inlet
44
To Clean the Septum Seat in the Insert Assembly of the Split/Splitless
Inlet
46
To Change the Liner and O-Ring on the Split/Splitless Inlet
48
To Replace the Gold Seal on the Split/Splitless Inlet 50
Check for leaks.
51
To Replace the Filter in the Split Vent Line
52
To Clean the Split/Splitless Inlet
55
To Bakeout Contaminants from the Split/Splitless Inlet
57
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
Consumables and Parts for the Split/Splitless Inlet
See the Agilent catalog for consumables and supplies for a more
complete listing, or visit the Agilent Web site for the latest
information (www.agilent.com/chem/supplies).
\Table 4
Split, splitless, direct, and direct connect inlet liners
Mode
Description
Deactivated
Part number
Split
Low-pressure drop, glass wool, single taper, 870 µL
Yes
5183-4647
Split
Glass wool, 990 µL
No
19251-60540
Split—Manual only
Empty pin and cup, 800 µL
No
18740-80190
Split—Manual only
Packed pin and cup, 800 µL
No
18740-60840
Splitless
Single taper, glass wool, 900 µL
Yes
5062-3587
Splitless
Single taper, no glass wool, 900 µL
Yes
5181-3316
Splitless
Dual taper, no glass wool, 800 µL
Yes
5181-3315
Splitless—Direct inject
2-mm id, quartz, 250 µL
No
18740-80220
Splitless—Direct inject
2-mm id, 250 µL
Yes
5181-8818
Direct inject —Headspace or
purge and trap
1.5-mm id, 140 µL
No
18740-80200
Direct column connect
Single taper, splitless 4-mm id
Yes
G1544-80730
Direct column connect
Dual taper, splitless 4-mm id
Yes
G1544-80700
Table 5
Nuts, ferrules, and hardware for capillary columns
Column id (mm)
Description
Typical use
Part number/quantity
.530
Ferrule, Vespel/graphite,
0.8-mm id
0.45-mm and 0.53-mm capillary
columns
5062-3512 (10/pk)
Ferrule, graphite, 1.0-mm id
0.53-mm capillary columns
5080-8773 (10/pk)
Ferrule, graphite, 0.8-mm id
0.53-mm capillary columns
500-2118 (10/pk)
Column nut, finger-tight (for
0.53-mm columns)
Connect column to inlet or detector
5020-8293
.320
Ferrule, Vespel/graphite,
0.5-mm id
0.32-mm capillary columns
5062-3514 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
37
.250
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
.100 and .200
Ferrule, Vespel/graphite,
0.37-mm id
0.1-mm and 0.2-mm capillary
columns
5062-3516 (10/pk)
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Ferrule, graphite, 0.4-mm id
500-2114 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
All
Ferrule, no-hole
Testing
5181-3308 (10/pk)
Capillary column blanking nut
Testing–use with any ferrule
5020-8294
Column nut, universal
Connect column to inlet or detector
5181-8830 (2/pk)
Column cutter, ceramic wafer
Cutting capillary columns
5181-8836 (4/pk)
Pencil, diamond tipped
Cutting capillary columns
420-1000
Ferrule tool kit
Ferrule installation
440-1000
Table 5
Nuts, ferrules, and hardware for capillary columns (continued)
Column id (mm)
Description
Typical use
Part number/quantity
Table 6
Other consumables and parts for the split/splitless inlet
Description/quantity
Part number
Septum retainer nut for headspace
18740-60830
Septum retainer nut
18740-60835
11-mm septum, high-temperature, low-bleed, 50/pk
5183-4757
11-mm septum, prepierced, long life, 50/pk
5183-4761
Merlin Microseal septum (high-pressure)
5182-3444
Merlin Microseal septum (30 psi)
5181-8815
Nonstick fluorocarbon liner O-ring (for temperatures up to
350
°C), 10/pk
5188-5365
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
Nonstick fluorocarbon liner O-ring for Flip Top Inlet Sealing
System, 10/pk
5188-5366
Graphite O-ring for split liner (for temperatures above
350
°C), 10/pk
5180-4168
Graphite O-ring for splitless liner (for temperatures above
350
°C), 10/pk
5180-4173
Split vent trap PM kit, single cartridge
5188-6495
Retaining nut
G1544-20590
Gold-plated seal (standard application)
5188-5367
Gold-plated seal with cross (high split flows) (includes SS
washer)
5182-9652
Stainless steel washer (0.375-inch od), 12/pk
5061-5869
Reducing nut
18740-20800
Column nut, blanking plug
5020-8294
Capillary inlet preventative maintenance kit, split
5188-6496
Capillary inlet preventative maintenance kit, splitless
5188-6497
Table 6
Other consumables and parts for the split/splitless inlet
Description/quantity
Part number
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
39
Exploded Parts View of the Split/Splitless Inlet
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Install a Capillary Column with the Split/Splitless Inlet
1
Gather the following (see “Consumables and Parts for the
Split/Splitless Inlet” on page
36):
• Column
• Ferrule(s)
• Column nut
• Septum
• Column cutter
• Isopropanol
• Lab tissue
• Metric ruler
• 1/4-inch open-end wrench
• Lint-free gloves
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Verify that the correct glass liner is installed. (See
“Consumables and Parts for the Split/Splitless Inlet” on
page
36.)
4
Place the column on the hanger with the ends pointing up
and the label to the front.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
41
5
Place a septum, capillary column nut, and ferrule on the
column.
6
Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
7
Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
8
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
9
Position the column so it extends 4 to 6 mm above the end of
the ferrule. Slide the septum up the column to hold the
column nut at this position.
Ferrule
Column nut
Septum
Good
Bad
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
10 Thread the column nut into the inlet but do not tighten.
11 Adjust the column position so that the septum contacts the
bottom of the column nut. Finger-tighten the column nut
until it begins to grip the column.
12 Tighten the column nut an additional 1/4 to 1/2
turn with a
wrench so that the column cannot be pulled from the fitting
with gentle pressure.
13 Configure the new column.
14 Condition the column per the manufacturer’s
recommendation. (See To Condition a Capillary Column.)
15 Install the column into the detector. See:
• To Install a Capillary Column in the FID
• To Install a Capillary Column in the NPD
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the µECD
• To Attach a Capillary Column to the FPD+
16 After the column is installed at both inlet and detector,
establish a flow of carrier gas and purge as recommended by
the column manufacturer.
17 Restore the analytical method.
• For FID or FPD+, immediately turn off the flame.
• For NPD, immediately turn off the bead.
4%6 mm
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
43
18 After the GC becomes ready, wait 10
minutes then ignite the
detector flame or bead.
19 Allow the oven, inlet, and detector to equilibrate at operating
temperature, then retighten the fittings.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If they are hot, wear heat-resistant gloves to protect
your hands.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Change the Septum on the Split/Splitless Inlet
1
Gather the following:
• Replacement septum. (See “Consumables and Parts for
the Split/Splitless Inlet” on page
36.)
• 0- or 00-grade steel wool (optional)
• Tweezers
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the septum retainer nut or Merlin cap.
4
Use tweezers to remove the septum or Merlin Microseal from
the retainer nut. Do not gouge or scratch the interior of the
septum head.
5
Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal should
face down (toward the oven).
6
Install the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1
mm above the nut.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Septum
Septum retainer
nut
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
45
7
Restore the analytical method.
8
Reset the septum counter.
CAUTION
Overtightening the septum nut can cause contamination.
1 mm
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Clean the Septum Seat in the Insert Assembly of the Split/Splitless Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Split/Splitless Inlet” on page
36.)
• 0- or 00-grade steel wool (optional)
• Tweezers
• Compressed, filtered, dry air or nitrogen
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Unscrew the insert nut from the inlet body. Lift the septum
assembly straight up and away from the inlet to avoid
chipping or breaking the liner.
4
Remove the septum retainer nut or Merlin cap.
5
Use tweezers to remove the septum or Merlin Microseal from
the retainer nut. (See “To Change the Septum on the
Split/Splitless Inlet” on page
44.)
6
Scrub the residue from the retainer nut and septum holder
with a small piece of rolled-up steel wool and tweezers. Do
not do this over the inlet.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Split/splitless insert nut
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
47
7
Use compressed air or nitrogen to blow away the pieces of
steel wool and septum.
8
Replace the insert retainer nut, tightening it to firm
finger-tightness. Do not overtighten.
9
Firmly press the new septum or Merlin Microseal into the
fitting. (See “To Change the Septum on the Split/Splitless
Inlet” on page
44.)
10 Replace the septum retainer nut or Merlin cap and
finger-tighten. (See “To Change the Septum on the
Split/Splitless Inlet” on page
44.)
11 Restore the analytical method.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Change the Liner and O-Ring on the Split/Splitless Inlet
1
Gather the following:
• Replacement O-ring (See “Consumables and Parts for the
Split/Splitless Inlet” on page
36.)
• Replacement liner
• Tweezers
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Unscrew the insert nut from the inlet body. Lift the septum
assembly straight up and away from the inlet to avoid
chipping or breaking the liner.
If using a Flip-Top inlet, lift the lever arm.
4
Loosen the O-ring from the sealing surface with tweezers.
5
Grasp the liner with tweezers and pull it out.
6
Inspect the surface of the gold seal for graphite or rubber
septum contamination. If required, replace the gold seal.
(See “To Replace the Gold Seal on the Split/Splitless Inlet” on
page
50.)
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Liner and O-ring in inlet
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
49
Clean the inlet if there is visible or suspected contamination.
(See “To Clean the Split/Splitless Inlet” on page
55.)
Clean O-ring residue from sealing surface.
7
Slide a new O-ring onto the replacement liner.
8
Return the liner to the inlet, pushing it all the way in until
the liner contacts the gold seal.
9
Replace the insert retainer nut, tightening it to firm
finger-tightness. Do not overtighten.
If using the Flip-Top inlet, close the lever arm.
10 Turn on the inlet. Allow the inlet and column to purge with
carrier gas for 15
minutes before heating the inlet or the
column oven.
11 Bakeout contaminants. (See “To Bakeout Contaminants from
the Split/Splitless Inlet” on page
57.)
12 Restore the analytical method.
13 Check for leaks.
14 Reset the liner counter.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Replace the Gold Seal on the Split/Splitless Inlet
1
Gather the following:
• Replacement gold seal (See “Consumables and Parts for
the Split/Splitless Inlet” on page
36.)
• Replacement washer
• 1/4-inch wrench (for column)
• 1/2-inch wrench
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the inlet liner.
4
Remove the column from the inlet. Cap the open end of the
column to prevent contamination. Remove the insulation cup
around the base of the inlet.
5
Loosen and remove the reducing nut. Remove the washer
and seal inside the reducing nut.
6
Put on gloves to protect the new gold seal and washer from
contamination. Put a new washer in the reducing nut and
place the new gold seal on top of it (raised portion facing
down).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
51
7
Replace the reducing nut and tighten securely with a wrench.
8
Replace the inlet liner.
9
Install the column and the insulation cup.
10 Bakeout contaminants. (See “To Bakeout Contaminants from
the Split/Splitless Inlet” on page
57.)
11 Restore the analytical method.
12 Check for leaks.
Inlet gold seal
Washer
Reducing nut
Gold seal side view
Make sure raised portion faces
down.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
To Replace the Filter in the Split Vent Line
1
Gather the following:
• New filter cartridge. (See “Consumables and Parts for the
Split/Splitless Inlet” on page
36.)
• T-20 Torx screwdriver.
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the pneumatics cover (top, back of GC).
4
Remove the retaining clip.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
WARNING
The split vent trap may contain residual amounts of any samples
or other chemicals you have injected into the GC. Follow your
company’s safety procedures for handling these types of
substances while replacing the trap filter cartridge.
Split vent valve
Split vent trap
Retaining clip
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
53
5
Completely loosen the two screws that secure the split vent
valve in place.
6
Lift the filter trap assembly and split vent valve from the
mounting bracket together and unscrew the split vent front
weldment on the filter trap assembly. Be careful not to stress
the tubing between the split vent valve and the trap.
7
Remove the old filter cartridge and two O-rings.
8
Verify the new O-rings are seated properly on the new filter
cartridge.
9
Install the new filter cartridge then reassemble the trap. Do
not fully tighten yet.
10 Place the filter trap assembly in the mounting bracket and
install the retaining clip.
11 Install the split vent valve.
12 Fully tighten the split vent front weldment onto the trap.
13 Check for leaks.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
14 Reset the EMF counter. See To reset an EMF counter.
15 Restore the analytical method.
16 Reset the split vent trap counter.
17 Install the pneumatics cover.
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
55
To Clean the Split/Splitless Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Split/Splitless Inlet” on page
36.)
• Replacement liner
• Replacement O-ring
• Replacement gold seal
• Replacement washer
• Solvent that will clean the type of deposits in your inlet
• Compressed, filtered, dry air or nitrogen
• Beaker
• Cleaning brushes—The FID cleaning kit (part number
9301-0985) contains appropriate brushes
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the inlet liner. (See “To Change the Liner and O-Ring
on the Split/Splitless Inlet” on page
48.)
4
Disconnect the column from the inlet.
5
Remove the reducing nut and gold seal. (See “To Replace the
Gold Seal on the Split/Splitless Inlet” on page
50.)
6
Place a beaker in the oven under the inlet to catch the
solvent.
7
Soak a cleaning brush in the solvent and scrub the inside of
the inlet weldment. Repeat 10
times.
8
Rinse the inlet with the solvent.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
4
Maintaining the Split/Splitless Inlet
9
Blow the inside of the inlet dry with compressed air or
nitrogen.
10 Install the gold seal and reducing nut.
11 Install the liner and O-ring.
12 Install the column. (See “To Install a Capillary Column with
the Split/Splitless Inlet” on page
40.)
13 Check for leaks.
14 Bakeout contaminants. (See “To Bakeout Contaminants from
the Split/Splitless Inlet” on page
57.)
15 Restore the analytical method.
Maintaining the Split/Splitless Inlet
4
Maintaining Your GC
57
To Bakeout Contaminants from the Split/Splitless Inlet
1
Put the inlet into split mode.
2
Set the column flow to the normal operating value, or set the
capillary column gas velocity to 30
cm/s.
3
Set the inlet split vent flow to 200
mL/min.
4
Purge the column with carrier flow for at least 10
minutes
before heating the oven.
5
If the column is attached to the detector, set the detector
25
°C above normal operating temperature.
If the column is not attached to the detector, cap the detector
fitting.
6
Set the inlet temperature to 300
°C or 25
°C above the
normal operating temperature to bakeout contaminants from
the inlet, mostly through the split vent.
7
Set the column oven 25
°C above the GC method final oven
temperature to bake contaminants from the column. Do not
exceed the column manufacturer’s maximum temperature
limit.
8
Bakeout for 30
minutes or until the detector baseline is free
of contamination peaks.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If they are hot, wear heat-resistant gloves to protect
your hands.
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
5
Maintaining the Purged Packed Inlet
Consumables and Parts for the Purged Packed Inlet
60
Exploded Parts View of the Purged Packed Inlet
63
To Install a Capillary Column with the Purged Packed Inlet
64
To Change the Septum on the Purged Packed Inlet
68
To Clean the Septum Seat in the Purged Packed Inlet
70
To Install an Adapter on the Purged Packed Inlet
72
To Change the O-Ring on the Purged Packed Inlet
74
To Change the Glass Liner on the Purged Packed Inlet
75
To Install an Insulation Cup on the Purged Packed Inlet
77
To Clean the Purged Packed Inlet 78
To Bakeout Contaminants from the Purged Packed Inlet
80
To Install a Packed Metal Column
81
To Install a Packed Column Adapter on a Detector Fitting
83
To Install a Packed Glass Column
85
To Condition a Packed Column
88
To Install Ferrules on a Packed Metal Column
90
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
Consumables and Parts for the Purged Packed Inlet
See the Agilent catalog for consumables and supplies for a more
complete listing, or visit the Agilent Web site for the latest
information (www.agilent.com/chem/supplies).
Table 7
Purged packed inlet parts
Description
Part number/quantity
Preventative maintenance kit
5188-6498
Purged packed glass liners and column adapters
Glass liner
5080-8732 (25/pack) or
5181-3382 deactivated
(5/pack)
0.53-mm column adapter
19244-80540
1/8-inch column adapter
19243-80530
1/4-inch column adapter
19243-80540
Recommended septa and O-rings for the purged packed inlet
11-mm solid septum, low-bleed, red
5181-1263 (50/pk)
11-mm septum with partial through-hole,
low-bleed, red
5181-3383 (50/pk)
11-mm septum, low-bleed, gray
5080-8896 (50/pk)
Merlin Microseal septum (30 psi)
5181-8815
11-mm high-temperature silicone septum
(350
°C and higher)
5182-0739 (50/pk)
Viton O-ring (Top insert weldment)
5080-8898 (12/pk)
Table 8
Nuts and ferrules for packed columns
Description
Typical use
Part number/quantity
1/8-inch id Swagelok
stainless steel nut, front
ferrule, back ferrule
1/8-inch column
5080-8751 (20 each/pk)
1/8-inch id Swagelok
brass nut, front ferrule,
back ferrule
1/8-inch column
5080-8750 (20 each/pk)
1/8-inch id Vespel/
graphite ferrule
1/8-inch column
0100-1332 (10/pk)
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
61
1/8-inch id
brass tubing nut
1/8-inch column
5180-4103 (10/pk)
1/4-inch id Swagelok
stainless steel nut, front
ferrule, back ferrule
1/4-inch column
5080-8753 (20 each/pk)
1/4-inch id Swagelok
brass nut, front ferrule,
back ferrule
1/4-inch column
5080-8752 (20 each/pk)
1/4-inch id Vespel/
graphite ferrule
Inlet/detector
liner/adapters
1/4-inch column
5080-8774 (10/pk)
1/4-inch id
brass tubing nut
1/4-inch column
5180-4105 (10/pk)
Table 8
Nuts and ferrules for packed columns (continued)
Description
Typical use
Part number/quantity
Table 9
Nuts, ferrules, and hardware for capillary columns
Column id (mm)
Description
Typical use
Part number/quantity
.530
Ferrule, Vespel/graphite,
0.8-mm id
0.45-mm and 0.53-mm capillary
columns
5062-3512 (10/pk)
Ferrule, graphite, 1.0-mm id
0.53-mm capillary columns
5080-8773 (10/pk)
Ferrule, graphite, 0.8-mm id
0.53-mm capillary columns
500-2118 (10/pk)
Column nut, finger-tight (for
0.53-mm columns)
Connect column to inlet or detector
5020-8293
.320
Ferrule, Vespel/graphite,
0.5-mm id
0.32-mm capillary columns
5062-3514 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
.250
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
.100 and .200
Ferrule, Vespel/graphite,
0.37-mm id
0.1-mm and 0.2-mm capillary
columns
5062-3516 (10/pk)
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
Ferrule, Vespel/graphite,
0.4-mm id
0.1-mm, 0.2-mm, and 0.25-mm
capillary columns
5181-3323 (10/pk)
Ferrule, graphite, 0.5-mm id
0.1-mm, 0.2-mm, 0.25-mm, and
0.32-mm capillary columns
5080-8853 (10/pk)
Ferrule, graphite, 0.4-mm id
500-2114 (10/pk)
Column nut, finger-tight (for
.100- to .320-mm columns)
Connect column to inlet or detector
5020-8292
All
Ferrule, no-hole
Testing
5181-3308 (10/pk)
Capillary column blanking nut
Testing–use with any ferrule
5020-8294
Column nut, universal
Connect column to inlet or detector
5181-8830 (2/pk)
Column cutter, ceramic wafer
Cutting capillary columns
5181-8836 (4/pk)
Pencil, diamond tipped
Cutting capillary columns
420-1000
Ferrule tool kit
Ferrule installation
440-1000
Table 9
Nuts, ferrules, and hardware for capillary columns (continued)
Column id (mm)
Description
Typical use
Part number/quantity
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
63
Exploded Parts View of the Purged Packed Inlet
Column nut
Ferrule
Insulation cup
Insulation
Adapter
Adapter nut
Ferrule
Glass liner
O-ring
Top insert weldment
Merlin Microseal
Merlin cap
Septum nut
Septum
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Install a Capillary Column with the Purged Packed Inlet
1
Gather the following:
• Column
• Ferrule (See “Consumables and Parts for the Purged
Packed Inlet” on page
60.)
• Column nut
• Glass liner
• Viton O-ring
• 0.53-mm column adapter
• Septum
• 1/4-inch wrench
• Metric ruler
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Install a 0.53-mm column adapter. (See “To Install an
Adapter on the Purged Packed Inlet” on page
72.)
4
Install a new Viton O-ring. (See “To Change the O-Ring on the
Purged Packed Inlet” on page
74.)
5
Place a septum, capillary column nut, and ferrule on the
column.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
WARNING
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
65
6
Score the column using a glass scribing tool. The score must
be square to ensure a clean break.
7
Break off the column end by supporting it against the column
cutter opposite the scribe. Inspect the end with a magnifying
loupe to make certain there are no burrs or jagged edges.
8
Wipe the column walls with a tissue dampened with
isopropanol to remove fingerprints and dust.
9
Position the column so it extends 1 to 2 mm above the end of
the ferrule. Slide the septum up the column to hold the
column nut at this fixed position.
Ferrule
Column nut
Septum
Good
Bad
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
10 Thread the column nut into the inlet adapter but do not
tighten.
11 Adjust the column position so that the septum is even with
the bottom of the column nut. Finger-tighten the column nut
until it begins to grip the column.
12 Tighten the column nut an additional 1/4 to 1/2
turn with a
wrench so that the column cannot be pulled from the fitting
with gentle pressure.
13 Configure the new column.
14 Condition the column per the manufacturer’s
recommendation. (See To Condition a Capillary Column.)
15 Install the column into the detector. See:
• To Install a Capillary Column in the FID
• To Install a Capillary Column in the NPD
• To Install a Capillary Column in the TCD
• To Install a Capillary Column in the µECD
• To Attach a Capillary Column to the FPD+
1%2 mm
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
67
16 After the column is installed at both inlet and detector,
establish a flow of carrier gas and purge as recommended by
the column manufacturer.
17 Restore the analytical method.
• For FPD+, immediately turn off the flame.
• For NPD, immediately set the bead voltage to 0.0.
18 After the GC becomes ready, wait 10 minutes then ignite the
detector flame or adjust offset on the NPD bead.
19 Allow the oven, inlet, and detector to equilibrate at operating
temperature, then retighten the fittings.
20 Reset the column EMF counters. See To Reset an EMF
Counter in the Operation Manual.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Change the Septum on the Purged Packed Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Purged Packed Inlet” on page
60.)
• Septum nut wrench
• 0- or 00-grade steel wool (optional)
• Tweezers
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the septum retainer nut or Merlin cap.
4
Use tweezers to remove the septum or Merlin Microseal from
the retainer nut. Do not gouge or scratch the interior of the
septum head.
5
Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal should
face down (toward the oven).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Septum retainer nut
Septum
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
69
6
Replace the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1
mm above the nut.
7
Restore the analytical method.
8
Reset the septum counter.
Merlin Microseal and cap
Standard septum
CAUTION
Overtightening the septum nut can cause contamination.
1 mm
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Clean the Septum Seat in the Purged Packed Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Purged Packed Inlet” on page
60.)
• Septum nut wrench
• 0- or 00-grade steel wool (optional)
• Tweezers
• Compressed, filtered, dry air or nitrogen
• Ultrasonic cleaning bath
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the septum retainer nut or Merlin cap.
4
Loosen the top insert weldment and remove.
5
Use tweezers to remove the septum or Merlin Microseal from
the top insert weldment. Do not gouge or scratch the interior
of the septum head.
6
Scrub the residue from the top insert weldment and septum
nut with a small piece of rolled-up steel wool and tweezers.
Ultrasonically clean the retainer nut and top insert
weldment.
7
Use compressed air or nitrogen to blow away the pieces of
steel wool and septum.
8
Wearing gloves, inspect the O-ring and replace, if necessary.
(See “To Change the O-Ring on the Purged Packed Inlet” on
page
74.)
9
Install the top insert weldment and hand-tighten firmly.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
71
10 Firmly press the new septum or Merlin Microseal into the
fitting.
11 Install the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1
mm above the nut.
12 Restore the analytical method.
13 Reset the septum counter.
CAUTION
Overtightening the septum nut can cause contamination.
1 mm
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Install an Adapter on the Purged Packed Inlet
1
Gather the following:
• Brass tubing nut (see “Consumables and Parts for the
Purged Packed Inlet” on page
60.)
• Adapter (0.53 mm, 1/8-inch packed, or 1/4-inch packed)
• 7/16-inch and 9/16-inch wrench
• Vespel/graphite ferrule
• Methanol
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Clean the end of the adapter with a lint-free cloth and
methanol to remove contamination such as fingerprints.
4
Place the tubing nut and Vespel/graphite ferrule on the
adapter.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
73
5
Insert the adapter straight into the inlet base as far as
possible.
6
Hold the adapter in this position and finger-tighten the nut.
7
Tighten an additional 1/4 turn with a wrench.
Inlet fitting
Ferrule
Tubing nut
Adapter
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Change the O-Ring on the Purged Packed Inlet
1
Gather the following:
• Replacement O-ring (See “Consumables and Parts for the
Purged Packed Inlet” on page
60.)
• Septum nut wrench
• Tweezers
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Loosen the top insert weldment to remove the top portion of
the inlet.
4
Use tweezers to remove the old O-ring.
5
Insert a new O-ring.
6
Install and tighten the top insert weldment.
7
Restore the analytical method.
8
Reset the EMF counter. See To Reset an EMF Counter in the
Operation Manual.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
75
To Change the Glass Liner on the Purged Packed Inlet
1
Gather the following:
• Replacement glass liner (See “Consumables and Parts for
the Purged Packed Inlet” on page
60.)
• 9/16-inch wrench
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Loosen the top insert weldment to remove the top portion of
the inlet.
4
Use a thin wire or wood splint to carefully lift and remove
the old glass liner.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Top insert weldment
Flared end
Glass liner
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
5
Wearing gloves, inspect the O-ring and replace, if necessary.
(See “To Change the O-Ring on the Purged Packed Inlet” on
page
74.)
6
Wearing gloves, grasp the flared end (top) of the replacement
glass liner with tweezers and install it in the inlet. If the glass
liner does not seat properly because a capillary column is
installed, remove the column, install the glass liner, and
replace the column. (See “To Install a Capillary Column with
the Purged Packed Inlet” on page
64.)
7
Install the top insert weldment and hand-tighten firmly.
8
Restore the analytical method.
9
Reset the EMF counter. See To Reset an EMF Counter in the
Operation Manual.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
77
To Install an Insulation Cup on the Purged Packed Inlet
1
Gather the following:
• No-hole ferrule
• Column nut
2
Install a plug (for example, a column nut with no-hole
ferrule) in the inlet capillary adapter.
3
Push the cup spring to the right. Slide the cup over the inlet
fitting so that the insulation at the top of the cup is flush
against the oven roof.
4
Place the spring into the groove in the inlet liner. Remove the
column nut and no-hole ferrule.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Clean the Purged Packed Inlet
1
Gather the following:
• Replacement O-ring (See “Consumables and Parts for the
Purged Packed Inlet” on page
60.)
• Replacement glass liner
• Replacement septum
• Solvent that will clean the type of deposits in your inlet
• Compressed, filtered, dry air or nitrogen
• Beaker
• Cleaning brushes—The FID cleaning kit (part number
9301-0985) contains appropriate brushes
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the column.
4
Remove the septum nut and septum.
5
Remove the top insert weldment.
6
Remove the glass liner and O-ring.
7
If used, remove the adapter.
8
Ultrasonically clean the septum nut, top insert weldment,
and adapter (if used) in a suitable solvent.
9
Place a beaker in the oven under the inlet to catch the
solvent.
10 Soak a cleaning brush with the solvent and vigorously scrub
the interior walls of the inlet.
11 Blow the inside of the inlet dry with compressed air or
nitrogen.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
79
12 Install the adapter, if used. (See “To Install an Adapter on the
Purged Packed Inlet” on page
72.)
13 Install the glass liner and O-ring. (See “To Change the Glass
Liner on the Purged Packed Inlet” on page
75.)
14 Install the top insert weldment and finger-tighten.
15 Install the septum and septum nut. (See “To Change the
Septum on the Purged Packed Inlet” on page
68.)
16 Attach the column. (See “To Install a Capillary Column with
the Purged Packed Inlet” on page
64.)
17 Check for leaks.
18 Restore the analytical method.
19 Set the septum and glass liner counters.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Bakeout Contaminants from the Purged Packed Inlet
1
Set the column flow to the normal operating value, or set the
capillary column gas velocity to 30
cm/s.
2
Purge the column with carrier flow for at least 10
minutes
before heating the oven.
3
If the column is attached to the detector, set the detector
25
°C above normal operating temperature.
If the column is not attached to the detector, cap the detector
fitting.
4
Set the inlet temperature to 300
°C or 25
°C above the
normal operating temperature.
5
Set the column oven 25
°C above the GC method final oven
temperature to bake contaminants out of the inlet. Do not
exceed the column manufacturer’s maximum temperature
limit.
6
Bakeout for 30
minutes or until the detector baseline is free
of contamination peaks.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
81
To Install a Packed Metal Column
1
Gather the following:
• 7/16-inch, 9/16-inch, and 1/2-inch wrenches
• Lint-free gloves
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Prepare the packed metal column. (See “To Install Ferrules
on a Packed Metal Column” on page
90.)
4
Install the 1/8-inch or 1/4-inch packed column inlet adapter,
if necessary. (See “To Install an Adapter on the Purged
Packed Inlet” on page
72.)
5
Attach the column to the inlet adapter. Finger-tighten the
nut.
6
Tighten the nut an additional 1/4 turn with a wrench (for a
1/8-inch column) or an additional 3/4 turn (for a 1/4-inch
column).
Use two wrenches, one on the column nut and the other on
the adapter, to prevent the adapter from rotating.
7
Configure the new packed column (make sure either column
length or diamter is zero).
8
Condition the column, if necessary. (See “To Condition a
Packed Column” on page
88.)
9
Load the GC maintenance method and wait for the GC to
become ready.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
10 If required, install the detector adapter. (See “To Install a
Packed Column Adapter on a Detector Fitting” on page
83.)
11 Attach the column to the detector or detector adapter.
Finger-tighten the nut.
12 Tighten the nut an additional 1/4 turn with a wrench (for a
1/8-inch column) or an additional 3/4 turn (for a 1/4-inch
column).
13 Establish a flow of carrier gas and purge as recommended by
the packing manufacturer. Generally:
• 20 to 30
mL/min for 2-mm id glass or 1/8-inch od metal
columns
• 50 to 60
mL/min for 4-mm id glass or 1/4-inch od metal
columns
14 Restore the analytical method.
• For NPD, immediately set the bead voltage to 0.0.
15 After the GC becomes ready, wait 10
minutes then ignite the
detector flame or adjust offset on the NPD bead.
16 Allow the oven, inlet, and detector to equilibrate at operating
temperature, then retighten the fittings.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
83
To Install a Packed Column Adapter on a Detector Fitting
1
Gather the following:
• 7/16-inch, 9/16-inch, and 1/2-inch wrenches
• Vespel/graphite ferrule (See “Consumables and Parts for
the Purged Packed Inlet” on page
60.)
• Brass column nut
• Lint-free gloves
• Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
• Consumables and Parts for the FID (Packed columns can
only be installed on an adaptable FID.)
• Consumables and Parts for the TCD
• Consumables and Parts for the NPD
• Consumables and Parts for the FPD+
• For µECD, 1/4-inch packed columns require no adapter.
For 1/8-inch packed columns, see Consumables and Parts
for the µECD.
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Assemble a nut and a ferrule onto the adapter.
WARNING
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
4
Insert the adapter straight into the detector base as far as
possible.
5
Hold the adapter in this position and finger-tighten the nut.
6
Tighten an additional 1/4 turn with a wrench (for a 1/8-inch
column) or an additional 3/4 turn (for a 1/4-inch column).
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
85
To Install a Packed Glass Column
1
Gather the following:
• 9/16-inch wrench
• Two 1/4-inch brass nuts (See “Consumables and Parts for
the Purged Packed Inlet” on page
60.)
• Two 1/4-inch Vespel/graphite ferrules
• Lint-free gloves
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Assemble a brass nut and Vespel/graphite ferrule on each
end of the column.
Glass columns must be simultaneously inserted into the inlet
and detector and installed parallel to the oven door. When
conditioning the column, do not attach the column to the
detector.
4
If conditioning the column, insert the column into the purged
packed inlet until it bottoms. Withdraw the column 1 to
2 mm. Finger-tighten the inlet column nut. (See “To
Condition a Packed Column” on page
88.)
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
CAUTION
Overtightening the column nut or forcing it to bottom in either the
inlet or detector may shatter the column.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
5
Tighten the inlet column nut 1/4
turn with a wrench.
6
After conditioning, remove the column from the inlet.
7
Simultaneously insert the column into the inlet and detector
fittings but do not force it. It may be necessary to start the long
end of the column in the inlet at an angle to clear the oven floor.
8
Withdraw the column 1 to 2
mm from both the inlet and
detector. Finger-tighten both column nuts.
9
Tighten both column nuts 1/4
turn with a wrench.
10 Configure the new packed column (make sure either column
length or diameter is zero).
11 Establish a flow of carrier gas and purge as recommended by
the packing manufacturer. Generally:
• 20 to 30
mL/min for 2-mm id glass or 1/8-inch od metal
columns
• 50 to 60
mL/min for 4-mm id glass or 1/4-inch od metal
columns
12 Restore the analytical method.
• For NPD, immediately set the bead voltage to 0.0.
13 After the GC becomes ready, wait 10
minutes then ignite the
detector flame or adjust offset on the NPD bead.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
CAUTION
Overtightening the column nut or forcing it to bottom in either the
inlet or detector may shatter the column.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
87
14 Allow the oven, inlet, and detector to equilibrate at operating
temperature, then retighten the fittings.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Condition a Packed Column
1
Gather the following:
• Capillary adapter, column nut, and no-hole ferrule (for
FID and NPD), or 1/8-inch Swagelok cap (for TCD)
• Two 7/16-inch wrenches
• 1/4-inch open-end wrench
• Lint-free gloves
2
Load the GC maintenance method and wait for the GC to
become ready.
3
Install the proper liner in the inlet and attach the column.
(See “To Install a Packed Metal Column” on page
81.)
4
Cap the detector(s) fittings with the capillary adapter,
no-hole ferrule and column nut (FID and NPD) or
1/8-inch
cap (TCD).
5
Enter a column flow as recommended by the packing
manufacturer or an appropriate flow as follows:
• 20 to 30
mL/min for 2-mm id glass or 1/8-inch od metal
columns
• 50 to 60
mL/min for 4-mm id glass or 1/4-inch od metal
columns
6
Raise the oven temperature slowly to the conditioning
temperature for the column. The conditioning temperature is
never higher than the maximum temperature limit for the
column; 30
°C less than the maximum is usually sufficient.
WARNING
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining the Purged Packed Inlet
5
Maintaining Your GC
89
7
Continue conditioning overnight at the final temperature.
Cool the oven to room temperature with carrier flow on.
8
Attach the column to the detector and maintain established
flow. (See “To Install a Packed Metal Column” on page
81.)
Maintaining Your GC
5
Maintaining the Purged Packed Inlet
To Install Ferrules on a Packed Metal Column
1
Gather the following:
• Wrenches
• Stainless steel male Swagelok fitting, 1/4- or 1/8-inch od
• Brass Swagelok nut and ferrule set (See “Consumables
and Parts for the Purged Packed Inlet” on page
60.)
• Lint-free gloves
2
Verify that the column end is cut square and is free of burns
and deformation.
3
Secure the fitting in a bench vise.
4
Assemble a Swagelok nut and ferrules onto the column.
5
Fully insert the column into the vise-held fitting, then
withdraw 1–2
mm. Finger-tighten the nut.
6
Tighten the nut an additional 3/4
turn with a wrench (for a
1/8-inch column) or an additional 1-1/4
turn (for a 1/4-inch
column).
7
Unscrew the column nut from the vise-held fitting and
remove the column. Ferrules should now be set in place on
the column with the column end correctly positioned.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Front ferrule
Back ferrule
Column nut
Agilent 7820A Gas Chromatograph
Maintaining Your GC
Agilent Technologies
6
Maintaining the Packed Column Inlet
Consumables and Parts for the Packed Column Inlet
92
Exploded Parts View of the Packed Column Inlet
94
To Change the Septum on the Packed Column Inlet
95
To Clean the Septum Seat in the Packed Column Inlet
97
To Install an Adapter on the Packed Column Inlet
99
To Change the O-Ring on the Packed Column Inlet
101
To Change the Glass Liner on the Packed Column Inlet
102
To Install an Insulation Cup on the Packed Column Inlet
104
To Clean the Packed Column Inlet 105
To Bakeout Contaminants from the Packed Column Inlet
107
To Install a Packed Metal Column
108
To Install a Packed Column Adapter on a Detector Fitting
110
To Install a Packed Glass Column
112
To Condition a Packed Column
115
To Install Ferrules on a Packed Metal Column
117
Maintaining Your GC
6
Maintaining the Packed Column Inlet
Consumables and Parts for the Packed Column Inlet
See the Agilent catalog for consumables and supplies for a more
complete listing, or visit the Agilent Web site for the latest
information (www.agilent.com/chem/supplies).
Table 10
Packed column inlet parts
Description
Part number/quantity
Preventative maintenance kit
5188-6498
Packed column glass liners and column adapters
Glass liner
5080-8732 (25/pack) or
5181-3382 deactivated
(5/pack)
1/8-inch column adapter
19243-80530
1/4-inch column adapter
19243-80540
Recommended septa and O-rings for the packed column inlet
11-mm solid septum, low-bleed, red
5181-1263 (50/pk)
11-mm septum with partial through-hole,
low-bleed, red
5181-3383 (50/pk)
11-mm septum, low-bleed, gray
5080-8896 (50/pk)
Merlin Microseal septum (30 psi)
5181-8815
11-mm high-temperature silicone septum
(350
°C and higher)
5182-0739 (50/pk)
Viton O-ring (Top insert weldment)
5080-8898 (12/pk)
Table 11
Nuts and ferrules for packed columns
Description
Typical use
Part number/quantity
1/8-inch id Swagelok
stainless steel nut, front
ferrule, back ferrule
1/8-inch column
5080-8751 (20 each/pk)
1/8-inch id Swagelok
brass nut, front ferrule,
back ferrule
1/8-inch column
5080-8750 (20 each/pk)
1/8-inch id Vespel/
graphite ferrule
1/8-inch column
0100-1332 (10/pk)
1/8-inch id
brass tubing nut
1/8-inch column
5180-4103 (10/pk)
Maintaining the Packed Column Inlet
6
Maintaining Your GC
93
1/4-inch id Swagelok
stainless steel nut, front
ferrule, back ferrule
1/4-inch column
5080-8753 (20 each/pk)
1/4-inch id Swagelok
brass nut, front ferrule,
back ferrule
1/4-inch column
5080-8752 (20 each/pk)
1/4-inch id Vespel/
graphite ferrule
Inlet/detector
liner/adapters
1/4-inch column
5080-8774 (10/pk)
1/4-inch id
brass tubing nut
1/4-inch column
5180-4105 (10/pk)
Table 11
Nuts and ferrules for packed columns (continued)
Description
Typical use
Part number/quantity
Maintaining Your GC
6
Maintaining the Packed Column Inlet
Exploded Parts View of the Packed Column Inlet
Column nut
Front ferrule
Insulation cup
Insulation
Adapter
Adapter nut
Ferrule
Glass liner
Top insert weldment
Merlin Microseal
Merlin cap
Septum nut
Septum
O-ring
Back ferrule
Maintaining the Packed Column Inlet
6
Maintaining Your GC
95
To Change the Septum on the Packed Column Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Packed Column Inlet” on page
92.)
• Septum nut wrench
• 0- or 00-grade steel wool (optional)
• Tweezers
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the septum retainer nut or Merlin cap.
4
Use tweezers to remove the septum or Merlin Microseal from
the retainer nut. Do not gouge or scratch the interior of the
septum head.
5
Firmly press the new septum or Merlin Microseal into the
fitting. The metal parts side of the Merlin Microseal should
face down (toward the oven).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
Septum retainer nut
Septum
Maintaining Your GC
6
Maintaining the Packed Column Inlet
6
Replace the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1
mm above the nut.
7
Restore the analytical method.
8
Reset the septum counter.
Merlin Microseal and cap
Standard septum
CAUTION
Overtightening the septum nut can cause contamination.
1 mm
Maintaining the Packed Column Inlet
6
Maintaining Your GC
97
To Clean the Septum Seat in the Packed Column Inlet
1
Gather the following:
• Replacement septum (See “Consumables and Parts for the
Packed Column Inlet” on page
92.)
• Septum nut wrench
• 0- or 00-grade steel wool (optional)
• Tweezers
• Compressed, filtered, dry air or nitrogen
• Ultrasonic cleaning bath
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Remove the septum retainer nut or Merlin cap.
4
Loosen the top insert weldment and remove.
5
Use tweezers to remove the septum or Merlin Microseal from
the top insert weldment. Do not gouge or scratch the interior
of the septum head.
6
Scrub the residue from the top insert weldment and septum
nut with a small piece of rolled-up steel wool and tweezers.
Ultrasonically clean the retainer nut and top insert
weldment.
7
Use compressed air or nitrogen to blow away the pieces of
steel wool and septum.
8
Wearing gloves, inspect the O-ring and replace, if necessary.
(See “To Change the O-Ring on the Packed Column Inlet” on
page
101.)
9
Install the top insert weldment and hand-tighten firmly.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
6
Maintaining the Packed Column Inlet
10 Firmly press the new septum or Merlin Microseal into the
fitting.
11 Install the septum retainer nut or Merlin cap and
finger-tighten. Tighten the septum retainer nut until the
C-ring is about 1
mm above the nut.
12 Restore the analytical method.
13 Reset the septum counter.
CAUTION
Overtightening the septum nut can cause contamination.
1 mm
Maintaining the Packed Column Inlet
6
Maintaining Your GC
99
To Install an Adapter on the Packed Column Inlet
1
Gather the following:
• Brass tubing nut (see “Consumables and Parts for the
Packed Column Inlet” on page
92.)
• Adapter (1/8-inch packed or 1/4-inch packed)
• 7/16-inch and 9/16-inch wrench
• Vespel/graphite ferrule
• Methanol
• Lint-free gloves
2
Load the inlet maintenance method and wait for the GC to
become ready.
3
Clean the end of the adapter with a lint-free cloth and
methanol to remove contamination such as fingerprints.
4
Place the tubing nut and Vespel/graphite ferrule on the
adapter.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause
burns. If the inlet is hot, wear heat-resistant gloves to protect your
hands.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Maintaining Your GC
6
Maintaining the Packed Column Inlet
5
Insert the adapter straight into the inlet base as far as
possible.
6
Hold the adapter in this position and finger-tighten the nut.
7
Tighten an additional 1/4 turn with a wrench.
Inlet fitting
Ferrule
Tubing nut
Adapter