Hioki 8861 Quick Start Manual
Text of the manual
QUICK START MANUAL
1
8860
8861
Read This Manual First
It describes preparations for use, basic
operating procedures and usage methods.
Confirming Package Contents....................................................2
Safety Information ......................................................................3
Operating Precautions................................................................5
Chapter 1
Overview__________________________________ 11
1.1
Product Overview and Features .......................................11
1.2
Measurement Workflow ....................................................12
Chapter 2
Names and Functions of Parts ________________ 19
2.1
Panel Names and Functions .............................................19
2.2
Screen Types and Contents .............................................23
2.3
Basic Operations ...............................................................32
2.3.1
To Change Settings ................................................................32
2.3.2
Entering Text and Numbers ....................................................33
2.3.3
Viewing Recorded Data ..........................................................35
Chapter 3
Measurement Preparations___________________ 37
3.1
Installation & Connection Procedures ...............................37
3.2
Connecting Cables and Probes ........................................38
3.3
Loading Recording Paper
(With a Printer Module Installed)........................................42
3.4
Connecting the Power Cord ..............................................45
3.5
Grounding the Instrument’s Functional Earth ...................45
3.6
Turning the Power On and Off ..........................................46
3.7
Setting the Clock ...............................................................48
Chapter 4
Basic Operations ___________________________ 49
4.1
Pre-Measurement Inspection ............................................49
4.2
Operation Workflow ..........................................................50
Recording Simple Waveforms ..........................................51
4.4
Actual Measurement and Analysis ...................................54
Chapter 5
Saving & Loading Data ______________________65
5.1
Recording (Storage) Media ...............................................66
5.2
Using PC Cards ................................................................67
5.3
Data that can be Saved & Loaded ....................................68
5.4
Saving Procedures and Contents .....................................70
5.5
Data Saving ......................................................................71
5.6
Data Loading ....................................................................74
Chapter 6
Printing ___________________________________75
6.1
Printing Methods and Contents ........................................76
6.2
Printing While Recording (Auto Print) ...............................77
6.3
Manual Printing (PRINT Key) ...........................................78
6.4
To Print a Detailed Settings List .......................................79
Chapter 7
Maintenance and Service ____________________81
7.1
Troubleshooting ................................................................81
7.2
Cleaning ............................................................................84
Appendix________________________________ A 1
Appendix 1 Waveform Screen Display & Icons..............................A 1
Appendix 2 Overview of Settings and Functions............................A 2
Appendix 3 Common Questions.....................................................A 7
Index ________________________________Index 1
Thank you for purchasing the HIOKI “Model 8860/ 8861 Memory HiCorder."
To obtain maximum performance from the instrument, please read this manual
carefully, and keep it handy for future reference.
In this document, the “instrument” means the Model 8860 or 8861 Memory
HiCorder.
The following documents are provided with this instrument. Refer to them as
appropriate for your application.
Registered trademarks
Windows is a registered trademark of Microsoft Corporation in the United States
and/or other countries.
If you encounter problems during operation
Introduction
Document
Description
1 Quick Start Manual
(This document)
Read this first.
It describes preparations for use, basic operating pro-
cedures and usage methods.
2 Input Module Guide
To connect input modules and measurement
cables, and when making input channel set-
tings;
this Guide describes the optional input modules, relat-
ed cable connection procedures, and their settings
and specifications.
3 Instruction Manual
To obtain setting details;
this Manual describes details of the functions and op-
erations of the instrument, and its specifications.
4 Analysis Supplement
The supplement describes usage of the cal-
culation functions to analyze measurement
data.
If you want to know a mea-
surement workflow
"1.2 Measurement Workflow" (⇒ p. 12)
For operation overviews:
"Chapter 3 Operation Overview" in the Instruction
Manual
For screen configurations
"2.2 Screen Types and Contents" (⇒ p. 23)
"2.2 Screen Organization" in the Instruction Manual
For input module connec-
tions and channel setting
procedures
Input Module Guide
When an error is displayed
"Appendix 1 Error Messages" in the Instruction
Manual
When measurement prob-
lems occur, or for general
measurement questions
"Appendix 3 Common Questions" (⇒ p. A7)
"Before returning for repair" (⇒ p. 82)
Confirming Package Contents
2
Confirming Package Contents
� 1 Quick Start Manual
� 2 Input Module Guide
� 3 Instruction Manual
� 4 Analysis Supplement
........................................One each
� Application CD
(Communications Manual)........... 1
� This instrument: Model 8860/ 8861 Memory
HiCorder (8860: 4-module, 8861: 8-module)
Confirm that these contents are provided.
Accessories
� Power Cord.............................................1
� Input Cable Labels...................................1
(If a printer module is installed)
For Model 8995 A4 Printer Unit:
� Model 9231 Recording Paper...........1 roll
� Paper Roll Holders ...........................1 pair
For Model 8995-01 A6 Printer Unit:
� Model 9234 Recording Paper............1 roll
� Paper Roll Axle........................................1
Other options specified when ordered
For information about options:
See "Appendix 5 Options" in the Instruction Manual
When you receive the instrument, inspect it carefully to ensure that no damage occurred
during shipping. In particular, check the accessories, panel switches, and connectors.
If damage is evident, or if it fails to operate according to the specifications, contact your
dealer or Hioki representative.
This manual contains information and warnings essential for safe operation of
the instrument and for maintaining it in safe operating condition. Before using it,
be sure to carefully read the following safety precautions.
The following symbols in this manual indicate the relative importance of cautions
and warnings.
Safety Information
This instrument is designed to comply with IEC 61010 Safety Standards,
and has been thoroughly tested for safety prior to shipment. However, mis-
handling during use could result in injury or death, as well as damage to
the instrument. Be certain that you understand the instructions and pre-
cautions in the manual before use. We disclaim any responsibility for acci-
dents or injuries not resulting directly from product defects.
Safety Symbols
In the manual, the
symbol indicates particularly important information that
the user should read before using the instrument.
The
symbol printed on the instrument indicates that the user should refer
to a corresponding topic in the manual (marked with the
symbol) before
using the relevant function.
Indicates a burn hazard if touched directly.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates both DC (Direct Current) and AC (Alternating Current).
Indicates a grounding terminal.
Indicates the ON side of the power switch.
Indicates the OFF side of the power switch.
Indicates that incorrect operation presents an extreme hazard that
could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard that
could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of injury to the
user or damage to the instrument.
Indicates advisory items related to performance or correct operation
of the instrument.
Symbols in text
Indicates the prohibited action.
(⇒ p. )
Indicates the location of reference informa-
tion.
Indicates quick references for operation
and remedies for troubleshooting.
*
Indicates that descriptive information is
provided below.
[ ]
Screen labels such as menu items, page
titles, setting items, dialog titles and but-
tons are indicated by square brackets [ ].
CURSOR
(Bold
characters)
Bold characters within the text indicate
operating key labels.
Unless otherwise specified, “Windows” represents Win-
dows 95, 98, Me, Widows NT4.0, Windows 2000, or Win-
dows XP.
Mouse operation terminology
Click:
Press and quickly release the left button of the
mouse.
Right-click: Press and quickly release the right button of
the mouse.
Double click: Quickly click the left button of the mouse
twice.
Drag:
While holding down the left button of the
mouse, move the mouse and then release the
left button to deposit the chosen item in the
desired position.
Activate: Click on a window on the screen to activate
that window.
We define measurement tolerances in terms of f.s. (full scale) values, with the following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or scale length. This is usually the name of the currently selected range.
In this instrument, the maximum displayable value is the range (V/div) times the number of divisions (20)
on the vertical axis. Example: For the 1 V/div range, f.s. = 20 V
Accuracy
This instrument complies with CAT II safety requirements.
This instrument’s input modules comply with CAT I or CAT II safety requirements.
To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for vari-
ous electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These
are defined as follows.
CAT I: Secondary electrical circuits connected to an AC electrical
outlet through a transformer or similar device.
CAT II: Primary electrical circuits in equipment connected to an
AC electrical outlet by a power cord (portable tools,
household appliances, etc.)
CAT III:Primary electrical circuits of heavy equipment (fixed
installations) connected directly to the distribution panel,
and feeders from the distribution panel to outlets.
CAT IV:The circuit from the service drop to the service entrance, and to the power meter and primary over-
current protection device (distribution panel).
Higher-numbered categories correspond to electrical environments with greater momentary energy. So a
measurement device designed for CAT III environments can endure greater momentary energy than a
device designed for CAT II.
Using a measurement instrument in an environment designated with a higher-numbered category than
that for which the instrument is rated could result in a severe accident, and must be carefully avoided.
Never use a CAT I measuring instrument in CAT II, III, or IV environments.
The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards.
Measurement categories (Overvoltage categories)
The applicable measurement category is determined by the input module being used. Refer to “Applicable Standards”
in "Chapter 5 Specifications" in the Input Module Guide for the applicable measurement categories.
Follow these precautions to ensure safe operation and to obtain the full benefits
of the various functions.
• Before using the instrument the first time, verify that it operates normally to
ensure that the no damage occurred during storage or shipping. If you find
any damage, contact your dealer or Hioki representative.
• Before using the instrument, make sure that the insulation on the probes and
connection cords is undamaged and that no bare conductors are improperly
exposed. Using the instrument in such conditions could cause an electric
shock, so contact your dealer or Hioki representative for replacements.
Operating temperature and humidity: 0 to 40°C at 20 to 80% RH (non-conden-
sating)
However, the operating temperature range may depend on the options used.
Temperature and humidity range for guaranteed accuracy:
23±5°C, 20 to 80%RH
Operating Precautions
Before Use
Instrument Installation
Operating Conditions
Operating Temperature
While changing the battery in the Model 9719 Memory Back-
up Unit
10 to 40°C
While using the Model 8995 A4 Printer Unit, 8995-01 A6
Printer Unit, 9716 FD Drive or 9718 HD Unit
5 to 40°C
Model 9717 MO Unit
5 to 35°C
Avoid the following locations that could cause an accident or damage
to the instrument.
Exposed to direct sun-
light
Exposed to high tem-
perature
In the presence of cor-
rosive or explosive
gases
Exposed to liquids
Exposed to high hu-
midity or condensa-
tion
Exposed to strong
electromagnetic fields
Near electromagnetic
radiators
Exposed to high lev-
els of particulate dust
Subject to vibration
• The instrument should be operated only with the bottom or rear side down-
wards.
• Vents must not be obstructed.
At least 50 mm
At least 50 mm spacing
on all sides
When using the optional printer, provide enough space for ejected recording
paper or use the Model 220H Paper Winder (optional recording paper auto
winder).
Handling the Instrument
• Do not allow the instrument to get wet, and do not take measurements
with wet hands. This may cause an electric shock.
• Never modify the instrument. Only Hioki service engineers should disas-
semble or repair the instrument. Failure to observe these precautions
may result in fire, electric shock, or injury.
• To avoid damage to the instrument, protect it from physical shock when trans-
porting and handling. Be especially careful to avoid physical shock from drop-
ping.
• Use the handle to carry the instrument.
See "Chapter 2 Names and Functions of Parts" (⇒ p. 19)
• When the Model 8958 16-Ch Scanner Unit is installed, do not lift up on the
connector section of the unit. The connector section could be damaged.
Handling the Cords and Probes
• Avoid stepping on or pinching cables, which could damage the cable insula-
tion.
• To avoid breaking the cables and probes, do not bend or pull them.
• To avoid damaging the power cord, grasp the plug, not the cord, when unplug-
ging it from the power outlet.
Use only the specified connection cables. Using a non-specified cable may
result in incorrect measurements due to poor connection or other reasons.
For additional information about input modules, their connections and
measurement precautions, refer to the Input Module Guide.
Before Connecting
Before turning power on
• Before turning the instrument on, make sure the supply voltage matches
that indicated on the its power connector. Connection to an improper
supply voltage may damage the instrument and present an electrical haz-
ard.
• To avoid electrical accidents and to maintain the safety specifications of
this instrument, connect the power cord only to a 3-contact (two-conduc-
tor + ground) outlet.
Installing and removing input modules
• To avoid electric shock accident, before removing or replacing an input
module, confirm that the instrument is turned off and that the connection
cords are disconnected.
• The mounting screws must be firmly tightened or the input module may
not perform to specifications, or may even fail.
• To avoid the danger of electric shock, never operate the instrument with
an input module removed. To use the instrument after removing an input
module, install a blank panel over the opening of the removed module.
Using an Optional Printer
When the instrument is not to be used for a long time, or when transport-
ing
To avoid straining some parts of the printer, and to prevent dirt adhering to the
print head, set the print head to its lowered position (so that the print head rests
against the print roller).
After storing the printer without use for a long time, inspect the following before
use:
1. Move the head raising lever up and down.
2. Perform test printing (printer check) three or four times.
When printing
• Avoid slow printing as much as possible in hot or humid environments. Other-
wise, printer life may be severely shortened.
• Please use only the specified recording paper. Using non-specified paper may
not only result in faulty printing, but printing may become impossible.
• If the recording paper is skewed on the roller, paper jams may occur. If the
recording paper is skewed on the roller, paper jams may result.
• Printing is not possible when the recording paper is loaded upside-down.
Refer to "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p.
42) for printing precautions.
Input and Measurement Precautions
The maximum input voltage and maximum rated voltage to ground
(between the grounds of the input terminals and the instrument, and
between inputs of other analog modules) of each input module and exter-
nal input terminals are shown below. To avoid risk of electric shock and
damage to the instrument, be careful to not exceed these ratings. The max-
imum rated voltage to ground is the same regardless of whether an input
attenuator or similar device is used.
Input Terminal (Input Module)
Maximum Input Voltage*1
Maximum Rated Voltage
to Ground*2
8936 Analog Unit
400 V DC max
370 V AC/DC
8956 Analog Unit
400 V DC max
300 V AC/DC
8946 4-Ch Analog Unit
30 V rms/ 60 V DC
30 V rms / 60 V DC
8938 FFT Analog Unit
400 V DC max
370 V AC/DC
8957 High Resolution Unit
400 V DC max
300 V AC/DC
8959 DC/RMS Unit
400 V DC max
370 V AC/DC
8937 Voltage/Temp Unit
30 V rms/ 60 V DC
30 V rms / 60 V DC
8939 Strain Unit
10 V DC max
30 V rms / 60 V DC
8960 Strain Unit
10 V DC max
33 V rms / 70 V DC
8940 F/V Unit
(BNC jacks and sensor re-
ceptacles)
30 V rms / 60 V DC
(BNC jack)
30 V rms / 60 V DC
(Sensor receptacle terminals)
Uninsulated
8947 Charge Unit*3
(BNC jack)
30 V rms / 60 V DC
(BNC jack)
30 V rms / 60 V DC
8958 16-Ch Scanner Unit
40 V DC max
33 V rms / 70 V DC
9322 Differential Probe
(CATII)
2000 V DC, 1000 V AC
(CAT III)
600 V AC/DC
With grabber clips
1500 V AC/DC (CAT II)
600 V AC/DC (CAT III)
With alligator clips
1000 V AC/DC (CAT II)
600 V AC/DC (CAT III)
I/O Terminals
Maximum input voltage
Maximum rated voltage
to ground
EXT.SMPL
EXT.TRIG
PRINT/EXT.IN3
STOP/EXT.IN2
START/EXT.IN1
-2 to 7 V DC
Uninsulated
TRIG OUT/CAL
NG/EXT OUT2
GO/EXT OUT1
-20 to 30 V DC,
500 mA max,
200 mW max
This Instrument
Input Module
GND
*1
*2
H
L
*2
“xxx V AC/DC” indicates an RMS
value, and “xxx V DC max” indi-
cates an instantaneous value.
(Input modules other than the
Model 8947 Charge Unit)
This Instrument
Input Module
GND
*3
*2
H
L
*2
*1
H
L
BNC
Miniature
Connector
Common
(8947 Charge Unit)
*1.Maximum input voltage
*2.Maximum rated voltage to
ground
*3.Maximum applied electric
charge
500 pC (high-sensitivity side, 6
ranges),
50,000 pC (low-sensitivity
side, 6 ranges)
Use the original packing materials when transporting the instrument, if possible.
The optional Model 9723 and 9724 Carrying Cases are available for shipping the
Models 8860 and 8861, respectively.
To avoid accidents in the event of a short circuit, when measuring com-
mercial mains voltage using the Model 8936, 8938, 8956, 8957 or 8959 input
modules, connection should be made only at the secondary side of a
breaker. Never make connections at the primary side of a breaker, as unre-
stricted current flow could cause a serious accident if a short circuit
occurs.
When measuring AC mains voltage, for example, using a power transformer, be
sure to connect the ground terminal of the power transformer to ground.
Using a power transformer (PT)
This Instrument
Input Module
GND
H
L
Input
Power transformer with ground terminal
Ground (Earth terminal of power source, or equivalent)
PT
Power transformer ground terminal
Power transformer without ground terminal
GND
H
L
Input
Ground (Earth terminal of power source, or equivalent)
PT
This Instrument
Input Module
Correct measurement may be impossible in the presence of strong magnetic
fields, such as near transformers and high-current conductors, or in the pres-
ence of strong electromagnetic fields such as near radio transmitters.
Shipping precautions
To avoid damage, observe the following when shipping the instrument:
• Remove the PC cards before shipping.
• If the optional printer module is installed, remove the paper. If the paper is left
in the instrument, the paper-handling components may be damaged by vibra-
tion.
1.1 Product Overview and Features
11
Overview
Chapter 1
1.1
Product Overview and Features
Recording various analog
signals
Measurements are
provided in combina-
tion with optional input
modules, cables and clamp
probes depending on the
application.
Voltage, Current, Tempera-
ture, Acceleration, Vibra-
tion, Frequency, Stress, etc.
Automatic and manual saving, loading
and file management of measurement
data (⇒ p. 65)
Data saving and loading are available with optional PC
Cards. Also, using optional proprietary drives, saving
and loading can be performed on floppy diskettes, MO
or hard disk. Data on storage media can be managed
with this instrument.
"Chapter 10 Saving/Loading Data & Managing Files" in
the Instruction Manual
Memory capacity can be expanded with optional mem-
ory boards. (This option must be specified when order-
ing.) 8860: 32 Megawords to 1 Gigaword, Model 8861:
64 Megawords to 2 Gigawords
Automatic and manual
printing of measurement
data (⇒ p. 75)
Measurement data can be printed
with the optional A4 or A6 internal
printer and special recording pa-
per. Printing is also available on
an external USB printer.
"Chapter 11 Printing" in the
Instruction Manual
Recording logic signals
Measures using optional logic
probes.
Mixed analog and
logic recording is
possible.
Remote control and data collection
The instrument can be connected to a PC for measure-
ment data analysis and remote control.
"Chapter 13 Communications Settings" in the
Instruction Manual
External Control
The instrument can be controlled by external signals.
"Chapter 14 External Control" in the Instruction Manual
Easy entry of text and numerical values
Screen operations and character input can be per-
formed by connecting a commonly available mouse
or keyboard.
"3.3 Common Operations" in the
Instruction Manual
Input Module Guide
X-Y Display
Zoom
Split-Screen Display
Numerical Calculations
Gauge and Level Monitor
Displays
In this case
• Viewing waveforms before and
after an anomaly.
• Capturing the waveform of an
intermittent anomaly.
• Recording multiple signal lines
over an extended period.
• Acquiring measurement data for
processing on a PC.
Overlay
Sheet Display
Analysis
Measures with functions
that support analytical
applications
Memory Function
Recorder Function
Real-Time Saving Function
FFT Function
Cursor Measurements
Many other functions are provided in addition to these. Refer to "Appendix 2 Overview of Settings and Functions"
(⇒ p. A2) and "Chapter 1 Overview" in the Instruction Manual.
1.2 Measurement Workflow
12
1.2
Measurement Workflow
1 Install & Connect
Connect the instrument to the measurement object, and
turn power on.
See"Chapter 3 Measurement Preparations" (⇒ p. 37) in this manual
See"Instrument Installation" (⇒ p. 5) in this manual
Details of input module connections and connection cables:
See"Chapter 2 Connections" in the Input Module Guide
See"3.6 Turning the Power On and Off" (⇒ p. 46) in this manual
To set the clock:
See"3.7 Setting the Clock" (⇒ p. 48)
2 Make Instrument Settings
Make appropriate instrument settings to acquire the mea-
surement data.
See"Chapter 4 Basic Operations" (⇒ p. 49) in this manual, (Auto-
matic setting is available (⇒ p. 51))
For details:
See"Chapter 4 Measurement Configuration Settings" in the
Instruction Manual
See"Chapter 3 Input Channel Settings" and "Chapter 4 Logic Input
Settings" in the Input Module Guide
See"Chapter 6 Trigger Settings" in the Instruction Manual
See"Chapter 7 Waveform Display Settings" in the Instruction Manual
See"Chapter 5 Saving & Loading Data" (⇒ p. 65) and "Chapter 6
Printing" (⇒ p. 75) in this manual,
"Chapter 10 Saving/Loading Data & Managing Files" and "Chap-
ter 11 Printing" in the Instruction Manual
Turn power on
Setting measurement
configuration
Set trigger criteria
3 Measure
Install the instrument
Select the input channels
Start recording
Analysis
Optionally save and print
Press the START key to start measuring.
Measurement data is recorded by the instrument with the selected set-
tings.
Measurement data recording stops automatically after the
preset period, or when you press STOP.
(Setting contents determine what causes measurement to stop.)
Analyze measurement data (⇒ p. 61).
Save and print measurement data as occasion demands.
See"Chapter 8 Waveform Screen Monitoring and Analysis" in the
Instruction Manual, and Analysis Supplement
5 Finish
Connect
Stop recording
4 Analyze, Save and Print
Select the Screen Layout
Turn the power off
Set saving and printing
conditions
Overall workflow and summary
If you don’t want to shut down completely (in order to measure
again immediately), press the STANDBY/ON key to activate the
standby state (⇒ p. 47).
(To record specified waveforms)
(To change the layout of the Waveform screen)
(To automatically save and print while
measuring)
See"3.3.6 Starting and Stopping Measurement" in the Instruction
Manual
(As occasion demands)
1.2 Measurement Workflow
13
Operation Overview: Recording Analog Waveforms
For more details, refer to "3.1 Measurement Workflow" in the Instruction Manual.
Also refer to "Appendix 2 Overview of Settings and Functions" (⇒ p. A2) and "Appendix 3 Common Ques-
tions" (⇒ p. A7).
Installation & Connections
Install the input modules and cables required for
measurement (⇒ p. 37).
Refer to the Input Module Guide for details about input mod-
ules and cables.
Power ON
Function Selection
Select the function according to your application.
See "Function Selection" (⇒ p. 29)
• Memory function
To record relatively fast signals from µs to minutes
• Recorder function
To record relatively slow signals at low speeds from ms to
hours
Record instantaneous
waveforms
Long-term measurements
Memory (MEM)
(Sampling point
recording)
Recorder (REC)
(Envelope recording)
Measurement Configuration
Settings
Make settings on the Status Settings screen.
Press the SET key to open the Settings screen
Press the SUB MENU keys to select the
menu
Press the SHEET/
PAGE keys to select
the [Use Ch] page
Verify the channel(s)
to be used
Press the SHEET/PAGE keys to select the
[Basic] page
1
Set the timebase (horizontal axis)
2
Set the recording length
1
2
See"4.2 Setting Measurement Configuration (Status Set-
tings Screen)" in the Instruction Manual
1.2 Measurement Workflow
14
Input Channel Settings
Make settings on the Channel Settings screen.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[One Ch] page
3
Select the Unit (module) and Channel
4
Select the measurement range (vertical axis)
Make input-module-related settings
5
Perform zero adjustment
6
(As occasion demands)
Set the scaling, waveform colors and
zero position
Trigger Settings
(If you want to record a specific waveform, such as an
anomaly)
Set on the Trigger Settings screen.
Press the SUB MENU keys to select the
menu
7
Set the trigger mode
8
Set the trigger criteria (AND/OR)
9
(Memory)
Set pre-trigger
(Recorder)
Set trigger timing
10
Set each trigger source
4
6
3
5
See"Chapter 5 Input Channel Settings" and "7.1 Making
Input Waveform Display Settings (Analog Waveforms)"
in the Instruction Manual
"Chapter 3 Input Channel Settings" in the Input Module
Guide
(after warm-up)
9
10
7
8
See"Chapter 6 Trigger Settings" in the Instruction Manual
Analog or Logic Trigger
External
Trigger
Timer Trigger
Default setting:
(Memory) [Auto]
(Recorder) [Single]
Default setting: [OR]
Default setting: All [Off]
1.2 Measurement Workflow
15
Display Sheet Settings
(If you want to change the layout of the waveform
screen to show any combination of channels)
Set on the Sheet Settings screen.
Press the SUB MENU keys to select the
menu
11
(As occasion demands)
Select the Screen Layout
Set the number screen divisions and the
split-screen layout
12
Select the channels to display
Saving Settings
(If you want to save data)
Set on the Save Settings screen.
Press the SUB MENUs key to select the
menu
13
Select automatic or manual saving
Specify the saving destination
14
Select what to save
12
See"Chapter 7 Waveform Display Settings" in the Instruction
Manual
Combination
of display
channels
11
Up to 32 channels can be
displayed per sheet.
When using multiple Model
8958 16-Ch Scanner
Units, channels are auto-
matically assigned to
sheets.
(Example: Auto Save)
See"Chapter 10 Saving/Loading Data & Managing Files" in
the Instruction Manual
13
14
Verify that storage media
has been inserted.
When saving manually,
settings can be changed
after measurement.
Default setting:
Auto Save [Off],
Manual Save [Selection
Save]
1.2 Measurement Workflow
16
Printing Settings
(If you want to print data)
Set on the Print Settings screen.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[Printer] page
15
Select automatic or manual printing
Press the SHEET/PAGE keys to select the
[Print Items] page
16
Select what you want to print
Start of Measurement
Press the START key (the green LED lights).
End of Measurement
Press the STOP key.
Recording stops after acquiring the specified length
(the green LED goes off).
Press twice to stop immediately.
If [Single] trigger is selected, recording stops automatically
after acquiring the specified data length.
Data Analysis
Analysis on the waveform screen.
See"Chapter 8 Waveform Screen Monitoring and Analysis"
in the Instruction Manual
Optionally save and print
Press the SAVE key to save. (Manual saving)
Press the PRINT key to print. (Manual printing)
Power OFF
Remove the cables from the measurement object,
and turn the power off.
16
See"Chapter 11 Printing" in the Instruction Manual
15
Verify that the paper is
loaded correctly.
When printing manually,
settings can be made after
measurement.
Default setting:
Auto Print [Off]
Manual Print [Selection
Print]
Data acquisition
Save & Print (when Auto enabled)
Waveform data is erased when power is turned off. However, measurement settings are retained.
To retain waveforms with power off, the optional Model 9719 Memory Backup Unit is required.
1.2 Measurement Workflow
17
Operation Overview: Recording Logic Waveforms
Installation & Connections
Connect the logic probes (⇒ p. 37).
Refer to the Input Module Guide for details about probes.
Power On
Measurement Configuration Settings
Make settings on the Status Settings screen.
Press the SET key.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[Use Ch] page
(With Memory function)
1
Select the channels to use
Press the SHEET/PAGE keys to select the
[Basic] page
2
Select the timebase
3
Set the recording length
Input Channel Settings
Make settings on the [Logic] page of the Channel
Settings screen.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[Logic] page
4
Enable (set to ON) the channels to be displayed
5
(As occasion demands)
Set the waveform display colors
1
See"4.2.1 Selecting Channels to Use" in the Instruction
Manual
(With Memory function)
2
3
See"4.2 Setting Measurement Configuration (Status Set-
tings Screen)" in the Instruction Manual
4
5
See"7.3.1 Setting the Waveform Display" in the Instruction
Manual
Default setting: [Off]
1.2 Measurement Workflow
18
Trigger Settings
(If you want to record a specific waveform, such as an
anomaly)
Make settings on the [Logic] page of the Trigger
Settings screen.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[Logic] page
6
Set the trigger mode
7
Set the trigger criteria (AND/ OR)
8
Set pre-trigger or trigger timing
9
Set the logic trigger
Display Sheet Settings
(If you want to change the display layout of the Wave-
form screen, to display any combination of waveforms)
Make settings on the [Logic] page of the Sheet Set-
tings screen.
Press the SUB MENU keys to select the
menu
Press the SHEET/PAGE keys to select the
[Logic] page
10
(As occasion demands)
Select the Screen Layout
Set the number screen divisions and
the split-screen layout
11
Select whether to display or not
12
Set display position
13
Set display height of recording
From here, proceed the same as for analog channels. Refer to "Saving Settings" (⇒ p. 15).
6
7
8
9
See"Chapter 6 Trigger Settings" and "6.8 Triggering by
Logic Signals (Logic Trigger)" in the Instruction Manual
Default setting: [Auto]
Default setting: [OR]
Default setting: [Off]
12
11
13
10
See"7.3.2 Setting the Display Position" and "7.3.3 Setting
the Display Height" in the Instruction Manual
Default setting: [On]
Default setting: [Normal]
2.1 Panel Names and Functions
19
Refer to "Chapter 2 Operating Keys and Screen Contents" in the Instruction Manual for details of operating
keys and screens.
Names and
Functions of PartsChapter 2
2.1
Panel Names and Functions
Do not apply heavy downward pressure with the
stand extended. The stand could be damaged.
Front Panel
Display (LCD)
Screen Display (⇒ p. 23)
(Example: 8860)
Top Panel
Printer
(When the optional Model 8995 A4 Printer
Unit is installed)
Recording Printer Installation (⇒ p. 42)
Vent
Bottom Panel
Ventilation Fan
Clean the filter periodically to avoid blockage (⇒ p. 84).
Protective Cover
Protects the connector section when carrying by the handle.
Foot
May be operated front-side-up.
The number of installable input modules and the position of the handle differs on the Models 8860 and 8861
Operating Keys (⇒ p. 21)
Stand
2.1 Panel Names and Functions
20
Left Side
Right Side
Power Inlet
Connect the supplied power cord here (⇒ p. 45).
GND Terminal
(Functional Earth)
Connect to Earth ground (⇒ p. 45).
Vent
PC CARD slots
Accepts PC Card storage media
(⇒ p. 67).
Eject Button
Ejects the PC Card.
MO Disk Slot
(With optional Model 9717 MO
Unit installed)
Accepts MO disk storage media.
("10.1.2 Using an MO Disk" in the
Instruction Manual)
LOGIC Connectors
(For logic probe connection)
Input connectors for proprietary
logic probes. (Channels A – D)
Various Connectors
(⇒ p. 22)
Input Module Compartment
Input modules install here.
Model 8860: Up to 4 modules,
Model 8861: Up to 8 modules
Logic Channels
Serial Number
MAC Address
This is required for network manage-
ment, so do not peel it off.
Handle
Use to carry the instrument.
(The Model 8861 has its handle on the
side)
POWER Switch
Turns the instrument on and off (⇒ p.
46).
( :Power On
:Power Off )
Module Unit and Analog Channel
Numbering
Module Unit Numbering
Channel Numbering
(From CH1 at the bottom)
UNIT1
CH1
CH2
CH3
CH4
UNIT3
UNIT2
UNIT4
CH1
CH2
8947
8946
CH1
CH4
CH16
8958
(8861)
(8860)
UNIT1 - UNIT8
GND
Terminal
(⇒ p. 45)
GND Terminal
(⇒ p. 45)
CHD
CHC
CHB
CHA
Vent
2.1 Panel Names and Functions
21
This section provides an overview of the operating keys. Refer to "Chapter 2 Operating Keys and Screen
Contents" in the Instruction Manual for functional details of each key.
Operating Keys
*1. FN Mode
Functions differ according to the type of display screen. Refer to "Chapter 2 Operating Keys and Screen Contents" in the
Instruction Manual. (For the Waveform Screen)
F1: Switches numerical display.
F5: Sets input channels.
F2: Switches displayed gauge.
F6: Forces (Manual) triggering.
F3: Displays level monitor.
F7: Searches the waveform.
F4: Executes auto setting.
F8: (supported by upgrade)
Select setting items
CURSOR keys: Moves among setting items
(Hold the left and right keys for three seconds to
activate the key-lock state. Hold again for three
seconds to cancel)
SELECT key: Accepts the selected character or
opens the pull-down menu
ENTER key: Accepts the setting
ESC key: Cancels an operation, closes a dialog
HELP/CONV key (supported by upgrade)
Set A/B cursors
TYPE key: Cursor type
SPEED key: Speed of cursor motion
A/B knobs: Inner knob A moves cursor A, Outer
knob B moves cursor B
Print (⇒ p. 78)
PRINT key
Feed recording paper
FEED key
Paper feeds while pressing
the key.
Save (⇒ p. 71)
SAVE key
Select the Sheet (Waveform screen) or
Page (Settings screen)
SHEET/PAGE keys
Scroll waveform
SCROLL control (Jog: inner, Shuttle: outer)
View waveforms
(Waveform Screen)
DISP key
Open Settings Screen
SET key
View a file on storage
media (File Screen)
FILE key
Toggle standby state (⇒ p. 47)
STANDBY/ON key
Change F key functions
FUNCTION MODE key
Select setting items
F1 to F8 keys
(Indicated as “F key” for ge-
neric reference)
Press the FUNCTION MODE
key to change the function
(FN mode)*1 labels that ap-
pear below F1 to F8.
Selects menus
SUB MENU keys
Stop measurement
STOP key
1 press: Measurement stops after the set recording length,
2 presses: Immediate stop
Start measurement
START key
The green LED at the left is lit during measurement.
Select the input channels
UNIT key: Module Unit selection
CH key: Channel selection
RANGE/POSN knobs:
Measurement range selection (RANGE knob: in-
ner)
Zero-position setting (POSN knob: outer)
Select the timebase
TIME/DIV key
(Input waveform acquisition rate)
2.1 Panel Names and Functions
22
SYNC. OUT Jack
Use to synchronize sampling with multiple instruments. ref
(Sample synchronizing signal output)
See "14.2.4 Synchronized Sampling Output (SYNC.OUT)" in the Instruction Manual
EXT SMPL Jack
Sampling can be synchronized by an external signal applied here. (External Sampling)
(Memory function only)
See "14.2.3 External Sampling (EXT.SMPL)" in the Instruction Manual
100Base-TX (RJ45) Jack
Accepts a LAN cable connection. For connection to a network.
See "Chapter 13 Communications Settings" in the Instruction Manual
PS/2 Mouse Jack
A PS/2 mouse can be connected here to control the instrument.
See "3.2.2 Using a Mouse" in the Instruction Manual
PS/2 Keyboard Jack
A PS/2 keyboard can be connected here to control the instrument.
See "3.2.3 Using a Keyboard" in the Instruction Manual
MONITOR Jack
The instrument’s display can be replicated on an external PC monitor connected here. (24-
pin analog RGB)
External I/O Terminal Strip
These terminals provide and accept external input and output control signals.
See "Chapter 14 External Control" in the Instruction Manual
USB Port
The optional Model 9716 FD Drive or a commonly available USB mouse, keyboard or
memory device can be connected here.
See Mouse and keyboard connection: "3.2 Before Operating" in the Instruction Manual
Model 9716 FD Drive connection: "10.1.3 Using a Floppy Disk" in the Instruction
Manual
Connectors
2.2 Screen Types and Contents
23
2.2
Screen Types and Contents
There are five general screen types. Refer to the Instruction Manual for details about each screen type.
Refer to the Analysis Supplement for details of the FFT function.
At power on, either the Opening screen or the Screen with the same settings that were in effect when power
was last turned off appears.
Opening Screen
System Configuration
List
Configuration can be verified
on the System screen
Setting Items
Functions can be selected with the F
keys.
"Function Selection" (⇒ p. 29)
The Waveform screen appears
when you select a function.
SET key
FILE key
To File Screen
(⇒ p. 28)
To Settings Screen
(⇒ p. 24)
To Waveform Screen
Function Menu
Select a function before mea-
suring (⇒ p. 29).
(Functions can be changed
from the Waveform or Settings
screens)
This screen appears first after power on. (When you turn the power off with the Waveform screen dis-
played, it reappears after this screen.) The boot process takes about 40 seconds.
Load existing data
Make settings
View waveforms
DISP key
For details of screen contents:"2.3 Opening Screen" in the Instruction Manual
To System Screen
(⇒ p. 27)
Press the F7 key, or
hold down the SET key
Waveform Screen
Press the DISP key.
DISP
Displays waveforms acquired by the instrument. Waveforms can be displayed on separate sheets, or as
the numerical values or an X-Y composite.
Record Data (⇒ p. 30)
Press the SHEET/PAGE keys
to change the displayed sheet.
Status bar (⇒ p. 30)
Function Menu
To change functions
(⇒ p. 51)
Scroll bar (⇒ p. 30)
For details of screen contents: "2.4 Waveform Screen" in the Instruction Manual
Setting items (⇒ p. 31)
Press the SUB MENU keys to
switch which setting items are dis-
played.
Press the F keys (F1 to F8) to select
a specific setting. (⇒ p. 32)
Clock (⇒ p. 48)
Shows the current time.
The
FUNCTION
MODE
key
changes the functions of the F1 to F8
keys.
[MACRO]
[SET]
[FN]
2.2 Screen Types and Contents
24
Settings Screen
Press the SET key.
SET
Displays screens for making various operation-related settings such as for measurement, data saving,
printing and calculations. Select the particular settings screen from the Settings menu. Settings menu
contents are function-dependent. Some Settings screens include multiple Settings pages.
Page
Press the SHEET/PAGE keys to select a page.
Settings Menu
Press the SUB MENU keys to select a Settings screen.
Function Menu
Status Settings Screen
([Basic]/[Use Ch]* Page)
Make basic settings here, such as the timebase
and recording length. You can specify which chan-
nels to use.* (* Memory function only)
About screen contents:
"2.5.1 Status Settings Screen" in the Instruction
Manual
Channel Settings Screen
([One Ch], [Comment], [All Ch], [Scaling],
[Variable] and [Logic] pages)
Set the measurement range, scaling and input
waveforms for input channels.
About screen contents:
"2.5.2 Channel Settings Screen" in the Instruction
Manual
Trigger Settings Screen
([Analog] and [Logic] pages)
Set trigger criteria when using a trigger event for
measuring.
About screen contents:
"2.5.3 Trigger Settings Screen" in the Instruction
Manual
Sheet Settings Screen
([Analog], [Logic], [X-Y Comp] and [Wcal]* pages)
Set the Waveform screen layout and Sheet set-
tings. (* Memory function only)
About screen contents:
"2.5.4 Sheet Settings Screen" in the Instruction
Manual
Measurement Configuration and Screen Display Settings
For details of screen contents: "2.5 Settings Screen" in the Instruction Manual
SUB MENU
2.2 Screen Types and Contents
25
Waveform Calculation (Wave Calc)
Settings Screen
([Formula], [Scale] and [Const] pages)
These are the display settings for waveform calcu-
lations.
About screen contents:
"2.5.7 Waveform Calculation (Wave Calc) Settings
Screen" in the Instruction Manual
Settings Menu
Press the SUB MENU keys to select a Settings screen. (function-dependent)
Memory Division (Mem Div)
Settings Screen
These settings control memory partitioning.
About screen contents:
"2.5.5 Memory Division Settings Screen" in the
Instruction Manual
Numerical Calculation (Num Calc)
Settings Screen
These are the display settings for numerical calcu-
lations.
About screen contents:
"2.5.6 Numerical Calculation (Num Calc) Settings
Screen" in the Instruction Manual
SUB MENU
Analysis Settings (Utility Functions) (Memory function only)
2.2 Screen Types and Contents
26
Settings Menu
Press the SUB MENU keys to select a Settings screen.
Save Settings Screen
([Auto Save] and [SAVE Key] pages)
Make saving-related settings such as what to save
and the saving method.
Automatic saving: [Auto Save] page
Manual saving: [SAVE Key] page
About screen contents:
"2.5.8 Save Settings Screen" in the Instruction
Manual
Print Settings Screen
([Printer] and [Print Items] pages)
Make printing-related settings such as what to print
and the printing method.
Print destination and printer settings: [Printer] page
Print item selection: [Print Items] page
About screen contents:
"2.5.9 Print Settings Screen" in the Instruction
Manual
Data Saving and Printing Settings
Page
Press the SHEET/PAGE keys to select a page.
SUB MENU
2.2 Screen Types and Contents
27
System Screen
Press the F7 [System] key on the Opening screen, or hold the SET key down.
Environment (Env) Settings Screen
Use this screen to configure the system environ-
ment, Waveform screen layout and operating key
functions.
Communication (Comm) Settings Screen
([Communication], [File], [Web] and [Command] pages)
Use this screen to make communication-related
settings.
Use this screen to make system-related settings. Select from the menu to display a settings screen.
Page
Press the SHEET/PAGE keys to select a page.
For screen details: "2.7 System Screen" in the Instruction Manual
Settings Menu
Press the SUB MENU keys to select a Settings screen.
SUB MENU
Setting Configuration (Setting) Screen
Use this screen to store and reload instrument set-
tings.
External Terminal (Ext Term)
Settings Screen
Use this screen to configure the external I/O termi-
nals.
2.2 Screen Types and Contents
28
Initialization (Init) Settings Screen
Use this screen to set the clock, initialize data, run
self-tests and set scanner module compensation.
Configuration (Config) List Screen
Displays the instrument’s system configuration.
Use this screen to make system-related settings. Select from the menu to display a settings screen.
For screen details: "2.7 System Screen" in the Instruction Manual
Settings Menu
Press the SUB MENU keys to select a Settings screen.
SUB MENU
File Screen
Press the FILE key
FILE
Use the SHEET/PAGE keys to
move the cursor.
Folder Tree
File List
Use this screen to load measurement data and manage files (copy, delete, etc.).
For screen details: "2.6 File Screen" in the Instruction Manual
File Info
Operating Procedures:
"10.4 Loading Data" in the
Instruction Manual
2.2 Screen Types and Contents
29
Function Selection
F1
F2
F7
F8
F3
F4
FFT Function (FFT)
Recommended for performing fre-
quency analysis of rotating ob-
jects, vibrations, sounds and etc.
Spectral analysis and transfer
functions are available.
Memory Function (MEM)
This function is most suitable for
oscilloscope-type measurements,
such as instantaneous waveforms
and transient phenomena.
Use to record relatively fast signals
with periods from µs to minutes.
(Sampling point recording)
System Settings
Sets the system environment.
Clock setting procedure (⇒ p. 48)
Select a function according to the desired recording data. Select from the Opening, Waveform or Set-
tings screens. For function details, refer to "4.1 Selecting the Function" in the Instruction Manual.
Select on the
Opening screen
Recorder Function (REC)
This function is suitable for use in-
stead of a pen recorder or pen os-
cilloscope, or to record fluctuations
and create records of slow phe-
nomena.
Use to record relatively slow sig-
nals with periods from ms to hours.
(Envelope recording)
File Operations
Displays the File screen. Use this
screen to load previously saved
files and for other file management.
Real-Time Saving Function
Recommended for long-term re-
cording instead of a data recorder.
Measurement data is saved to the
recording media in real time.
Move the cursor here, and select a function with the F1 to F8 keys.
Select on the
Waveform or
Settings screen
2.2 Screen Types and Contents
30
Viewing Recording Data (Waveform Screen)
Waveform
Data acquired with this in-
strument can be displayed
as waveforms or numerical
values.
To display as numerical
values:
"8.13 Viewing Waveform Data
as Numerical Values" in the
Instruction Manual
A/B Cursors
Move the cursors using the
A/B knobs (⇒ p. 63).
A/B Cursor details:
"8.8 Cursor Values" in the
Instruction Manual
A/B Cursor Values
Appear when the A/B cur-
sors are displayed. Numeri-
cal calculation results and
channel information can be
displayed.
To change the information
display:
Press the DISP key repeat-
edly to switch between
waveform, information and
gauge displays. (⇒ p. 63)
Time Value Display
To change the display type:
"12.1.3 Selecting the Time Value Display" in the
Instruction Manual
Trigger Mark
The trigger mark appears
at the point where a trigger
is applied.
Scroll Bar
Shows the range and position of the displayed
waveform.
Viewing the Scroll bar:
"8.1 Scrolling Waveforms" in the Instruction Manual
Status Bar (Waveform Screen)
Display Info
Shows the sheet number of the displayed waveform.
Internal Processing
Status
Shows the states of data ac-
quisition and internal process-
ing.
Trigger Info
Shows the trigger
source and the date and
time of triggering.
Setting Info
Indicates the setting states of Auto
Save, Auto Print and A/B Cursors.
Peripheral Device Setting
Info
Icons indicate when PC Cards
have been inserted, when con-
nected to peripheral devices,
and which internal devices are
present.
Number of Data Acquisitions
Shows the number of times data has been ac-
quired since the START key was pressed to
start measuring.
Sub Message
Shows the estimated time to
finish storing.
This bar indicates the current states of data acquisition, internal processing, settings and display in-
formation. For details about each indicator, refer to "Appendix 1 Waveform Screen Display & Icons"
(⇒ p. A1) in this manual and "2.4 Waveform Screen" in the Instruction Manual.
2.2 Screen Types and Contents
31
Setting Items and Contents (Waveform Screen)
Indicates current settings. These can be changed while measuring.
To change a setting, use the CURSOR keys to move the cursor to it, and press an F key to choose the
new value.
Press the SUB MENU keys to switch items. (Some items are function-dependent)
[Config] (Meas Config & Trigger Criteria)↔[Trigger](Analog Trigger)↔[Num Calc]*(Numerical Calculation) ↔
[Mem Div]*(Memory Division) (* Memory function only)
See "2.4 Waveform Screen";"Setting Items and Contents (Waveform Screen)", "4.4 Setting Measurement Configura-
tion on the Waveform Screen", and "6.12 Making Trigger Settings on the Waveform Screen" in the Instruction
Manual
Zoom
Sets the time (horizontal) axis
magnification ratio of the se-
lected section of the wave-
form.
(For Memory Function)
Timebase
Indicates the timebase (time per divi-
sion on the horizontal axis) and sam-
pling
rate
(sampling
interval).
(Settings can be made on the Status
Settings screen.)
Recording Length
Shows the recording length and time.
The recording length (time) for each
data acquisition is set as a number of
divisions. (The setting is made on the
Status Settings screen.)
Magnification
Sets the time (horizontal) axis
magnification ratio of the whole
waveform.
Selecting the [Mag] button dis-
plays the whole waveform.
Trigger Mode Setting
Sets subsequent triggering criteria after a
measurement
operation
is
finished.
(These settings are made on the Trigger
Settings screen.)
Pre-Trigger Setting
Set this to record data prior to a trigger
event, or for a specified period after-
wards.
Selectable when the Overlay function
is enabled.
Type of Set-
ting Item
Recording
Length
Recording
Time
Timebase
Sampling Rate
Zoom
x1
x5
Recording Length
(divisions)
Measurement
Range (e.g.,
Voltage Axis)
Timebase (time/div)
Acquired Waveform
x1
Timebase and Recording Length
Difference Between Magnification and Zoom
1div
Magnification
x2
2.3 Basic Operations
32
You have two ways to change settings.
This document describes how to change settings using the F keys.
In addition to the operating keys, operations can be performed using a mouse or
keyboard.
Refer to "3.2.2 Using a Mouse" and "3.2.3 Using a Keyboard" in the Instruction
Manual.
Initialization and Default Settings
Each item of waveform data and setting data can be initialized.
See "12.3.2 Initializing Waveform Data"
"12.3.3 Initializing System Settings (System Reset)"
"Appendix 2.1 List of Default Settings"
in the Instruction Manual
2.3
Basic Operations
2.3.1
To Change Settings
F Keys
Setting Items
Setting items appear at the right side of
the screen (GUI area).
Select using F1 to F8.
When [Page] appears at F8:
more selections are available.
Press the F8 key to see the additional
selections.
Selecting with an F key (F1 to F8)
Moving to a Selection
Select from a Pull-Down Menu
Move the cursor to setting
item.
Pull-Down Menu
ENTER key
SELECT key
CURSOR keys
Opens the pull-down menu.
Select a setting item.
Applies the setting.
2.3 Basic Operations
33
See "3.3.3 Entering Text and Numbers" in the Instruction Manual
Move the cursor to a setting item and select an input method with the F keys.
When using a mouse, double click on a setting item to display the virtual keypad.
Enter a numerical value using the settings (↑↑, ↓↓, ↑ and ↓) assigned to the F
keys.
Enter a numerical value using the virtual keypad.
Enter a numerical value using the virtual pushwheel switches.
2.3.2
Entering Text and Numbers
Entering Numbers
Direct entry by F keys
[↑↑] [↓↓] [↑] [↓]
Enter by virtual keypad
[Keypad]
Enter using the virtual pushwheel switches
[Pushwheel]
1. Select a digit ..........CURSOR keys
2. Accept ...................F1 [Set] or SELECT key
After entering all digits,
3. Accept entry............F6 [OK] or ENTER key
Cancel entry
F7 [Cancel] key or ESC key
1. Move to a digit .........
CURSOR keys
2. Set the digit value.....
CURSOR keys
After entering all digits,
3. Accept entry............F3 [OK] or ENTER key
Cancel entry ...........F4 [Cancel] or ESC key
2.3 Basic Operations
34
Move the cursor to a setting item and select an input method with the F keys.
When using a mouse, double click on a setting item to display the virtual key-
board.
See "5.2 Adding Comments"; "Comment Entry Example" in the Instruction Manual
Enter text using the virtual keyboard for character entry.
Connect a keyboard for direct keyboard entry.
Entering Text
Enter by virtual keyboard
[Edit]
Enter by external keyboard
[Direct]
1. Select a character....CURSOR keys
2. Enter it in the entry field
F1 [Set] or SELECT key
After entering all characters,
3. Accept entry ............F6 [OK] or ENTER key
Cancel entry............F7 [Cancel] key or ESC
key
Press F8 to switch character case.
1. Switch character case..F8 [Character Set] key
2. Select a character....CURSOR keys
3. Enter it in the entry field
F1 [Set] or SELECT key
4. Accept entry............F6 [OK] or ENTER key
5. Cancel entry............F7 [Cancel] key or ESC
key
[List/History]
Enter prestored text from the virtual keyboard
When entering a file name (to load a file on a PC)
Windows 2000 and XP cannot handle file names containing the following char-
acters, so they should not be used:
ASCII: + = [ ] \ / | : * ? “ < > ; ,
2.3 Basic Operations
35
Press the DISP key. The recorded waveform appears on the Waveform screen.
For analysis procedures, refer to "9. Analyze Data" (⇒ p. 61), or "Chapter 8
Waveform Screen Monitoring and Analysis" in the Instruction Manual.
See "8.1 Scrolling Waveforms" in the Instruction Manual
Turn the inner (Jog) or outer (Shuttle) knobs of the SCROLL control to scroll the
waveform display. (only when the recorded length is at least 25 divisions)
You can reverse the Jog and Shuttle scroll directions.
See "12.2.3 Specifying Jog & Shuttle Scroll Operations" in the Instruction Manual
The waveform scroll direction can be indicated vertically or continuously.
See "7.2.5 Setting Waveform Scrolling Orientation" in the Instruction Manual
To increase the scrolling speed:
The scroll speed is controlled by the degree of rotation of the outer Shuttle knob.
Turning the knob more increases the scrolling speed.
To scroll automatically (Auto-Scroll):
Turn the outer Shuttle knob in the direction desired to scroll the waveform, hold it
until “Auto-Scroll” appears on the screen, and release it.
When the Shuttle knob is released, the waveform continues to scroll automati-
cally.
To cancel, press any operating key.
See "8.8 Cursor Values" in the Instruction Manual
Press the TYPE key and set an A/B cursor using the Settings dialog. Turn inner
knob A or outer knob B to move the A/B cursors on the screen to confirm the
value at a cursor.
2.3.3
Viewing Recorded Data
Jog
Shuttle
Scrolling Waveforms
Turn CW
Scrolls later on the waveform
from the current point.
Turn CCW
Scrolls earlier on the waveform
from the current point.
Screen Display
Scroll Direction
Earlier
Later
Knob A
Knob B
Confirming Measured Values
3.1 Installation & Connection Procedures
37
Be sure to read the "Operating Precautions" (⇒ p. 5) before installing and connecting this
instrument.
Measurement
Preparations
Chapter 3
3.1
Installation & Connection Procedures
Install this instrument
(⇒ p. 5)
1
(When adding or changing an input module)
Install an input module
Refer to the Input Module Guide.
3
Which probes and cables to connect depends on
your measurement application (⇒ p. 38).
Refer to the Input Module Guide for details.
4
(When printing)
Load recording paper
(⇒ p. 42)
7
Connect the power cord
(⇒ p. 45)
(When measuring in an elec-
trically noisy environment)
Ground the Functional
Earth of this instru-
ment (⇒ p. 45)
Turn the power on
(⇒ p. 46)
When finished measuring, disconnect the
cable(s) from the measurement object(s),
and turn the power off.
Left Side
Right Side
Connect the input
cable(s) to the
input module
Connect a logic
probe to the
LOGIC receptacle
(When measuring
analog signals)
(When measuring
logic signals)
Top Panel
6
7
GND
2
POWER
5
6
Other Connections
To use a PS/2 mouse, connect it before turning the power on.
See "3.2.2 Using a Mouse" and "3.3.4 Mouse Operations" in the Instruction Manual
• If the clock is wrong after turning power on: "3.7 Setting the Clock" (⇒ p. 48)
• When using a PC Card: "5.2 Using PC Cards" (⇒ p. 67)
• When using other storage media or drives: "10.1 Storage Media" in the Instruction Manual
• When using a keyboard: "3.2.3 Using a Keyboard" in the Instruction Manual
3.2 Connecting Cables and Probes
38
When using an input module for measurement:
Connect the cables or sensors to the input module.
When measuring logic signals:
Connect the logic probe(s) to the LOGIC receptacle(s) on the instrument.
Refer to the Input Module Guide for precautions regarding connections and
details of connection procedures.
Connection Preparations
For voltage measurement
3.2
Connecting Cables and Probes
• Do not use cables other than those specified by Hioki. The specified
cables use insulated BNC connectors to avoid electric shock accidents.
An uninsulated BNC plug may cause electric shock or damage to the
BNC jack.
• Do not connect a cable to the instrument while it is connected to the
object to be measured. Otherwise, an electric shock accident may occur.
Connecting to the BNC jacks on input modules
When disconnecting the BNC connector, be sure to release the lock before pull-
ing off the connector. Forcibly pulling the connector without releasing the lock, or
pulling on the cable, can damage the connector.
Lock
2
Input Module Side
Guide Pins
BNC Plug Slots
1
Connecting
Align the two slots in the BNC plug with the
guide pins on the jack at the input module
side, then push and twist the plug clockwise.
Disconnecting
Push the BNC plug in, twist it counterclock-
wise, and pull it out.
Use only the specified connection cords. Using a non-specified cable may result
in incorrect measurements due to poor connection or other reasons.
3.2 Connecting Cables and Probes
39
Applicable Input Modules
• Model 8956 Analog Unit
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8946 4-Ch Analog Unit
• Model 8957 High Resolution Unit
• Model 8937 Voltage/Temp Unit
• Model 8940 F/V Unit
• Model 8959 DC/RMS Unit
• Model 8947 Charge Unit *1
• Model 8958 16-Ch Scanner Unit *2
(Maximum input voltage (⇒ p. 8))
Use to connect: Connection Cables
• Model 9197 Connection Cord
(Maximum input voltage: 500 V)
Large alligator clip type
• Model 9198 Connection Cord
(Maximum input voltage: 300 V)
Small alligator clip type
• Model 9217 Connection Cord
(Maximum input voltage: 300 V)
For measuring BNC output
If the voltage to be measured exceeds
the maximum input rating of the input
module being used:
• Model 9322 Differential Probe *3
• Model 9665 10:1 Probe
• Model 9666 100:1 Probe
Example: Model 9322
Measuring Voltage
Connect to the BNC jack on an input module.
Connection Procedure:
"2.2 Connecting the Cables" in the Input
Module Guide
*1. The Model 9198 Connection Cord is rec-
ommended for connection.
*2. Use an input cable to make connections.
Recommended cables:
0.14 to 1.5 mm2 single-strand, or 0.14 to 1
mm2 multi-strand AWG 16 to 26 stripped 5
mm
*3. A power cord option or AC adapter may be
required, depending on the input module
used.
Applicable Input Modules
• Model 8940 F/V Unit
• Voltage measuring input modules
Use to connect: Clamps
• Clamp-On Sensors:
Models 9270, 9271, and 9272
• Universal Clamp-On CTs
Models 9277, 9278, and 9279
• Clamp-On Probes
Models 3273, 3273-50,
3274, 3275, 3276,
9018-10, and 9132-10
Example: Models
9272 and 9318
Measuring Current
Connect to a sensor receptacle or BNC
jack on an input module.
Connection Procedure:
"2.2.5 Connecting to the Model 8940 F/V
Unit" and "2.3 Connecting Clamps" in the
Input Module Guide
Depending on the clamp, a conversion ca-
ble or power supply unit may be required
for connection. ("1.3 List of Input Mod-
ules, Cables, Probes and Clamp Combi-
nations" in the Input Module Guide)
Scaling is also required in some cases.
("5.4 Converting Input Values (Scaling
Function)" in the Instruction Manual)
3.2 Connecting Cables and Probes
40
Applicable Input Modules
• Model 8937 Voltage/Temp Unit
• Model 8958 16-Ch Scanner Unit
Use to connect: Thermocouple
Thermocouple
For the Model 8937 Voltage/Temp
Unit
(Compatible wire: AWG 16 to 26, 0.4
to 1.2 mm diameter)
For the Model 8958 16-Ch Scanner
Unit
(Compatible wire: AWG 16 to 26,
0.14 to 1.5 mm2)
Connect to terminal block
Connect to terminal block
Measuring Temperature
Connect to the terminal block on the input
module.
Connection Procedure:
"2.2.3 Connecting to the Model 8937 Volt-
age/Temp Unit" and "2.2.7 Connecting to
the Model 8958 16-Ch Scanner Unit" in the
Input Module Guide
Applicable Input Modules
Model 8940 F/V Unit
Use to connect: Connection Cable
Model 9198 Connection Cord
(Maximum input voltage: 300 V)
Small alligator clip type
Measuring Frequency, Count and Pulse Duty
Connect to the BNC jack on an input mod-
ule.
Connection Procedure:
"2.2.5 Connecting to the Model 8940 F/V
Unit" in the Input Module Guide
Applicable Input Modules
• Model 8939 Strain Unit
• Model 8960 Strain Unit
Use to connect: Sensor
Strain Gauge Transducer
(Not available from Hioki)
Example:
Connecting using a conver-
sion cable
Using a Strain Gauge to Measure Vibration or Displacement (Strain)
Connect to the BNC jack on an input mod-
ule.
Connection Procedure:
"2.2.4 Connecting to a Strain Unit (Models
8939 and 8960)" in the Input Module Guide
Use the conversion cable supplied with the
input module according to the sensor be-
ing used.
3.2 Connecting Cables and Probes
41
Applicable Input Modules
Model 8947 Charge Unit
Use to connect: Acceleration Sensor or Connection Cable
* Not available from Hioki
Preamp: Preamplified acceleration sensor*
(BNC plug, operating power 2 mA @ 15 V)
Charge: Charge-output type acceleration sensor*
(Miniature #10-32 connector)
Voltage Measurement: Model 9198 Connection Cord
(recommended)
Measuring Vibration using Piezoelectric (Charge- or Preamp-type) Sensors
Preamp
Connect to the BNC jack on an input mod-
ule.
Charge
Connect to the miniature receptacle on the
input module.
Connection Procedure:
"2.2.6 Connecting to the Model 8947
Charge Unit" in the Input Module Guide
Use to connect: Logic Probe
• Model 9320-01 Logic Probe
• Model 9321-01 Logic Probe
• Model 9327 Logic Probe
LOGIC Receptacles
Measuring Logic Signals
Connection Procedure:
"2.6 Connecting Logic Probes" in the Input
Module Guide
3.3 Loading Recording Paper (With a Printer Module Installed)
42
Printing is available using the following optional printer (specified when ordering
the instrument) and recording paper.
About the Printer
Recording Paper Usage Precautions
3.3
Loading Recording Paper (With a Printer
Module Installed)
Printer
Recording Paper (size)
Model 8995 A4 Printer Unit
Model 9231 Recording Paper (A4 size)
Model 8995-01 A6 Printer Unit
Model 9234 Recording Paper (A6 size)
The print head and surrounding metal parts of the Model 8995-01 A6
Printer Unit can become hot. Be careful to avoid touching these parts.
When the instrument is not to be used for a long time, or when transport-
ing
To avoid straining some parts of the printer, and to prevent dirt adhering to the
print head, set the print head to its lowered position (so that the print head rests
against the print roller).
After storing the printer without use for a long time, inspect the following before
use:
1. Move the head raising lever up and down.
2. Perform test printing (printer check) three or four times.
• Please use only the specified recording paper. Using non-specified paper may
not only result in faulty printing, but printing may become impossible.
• If the recording paper is skewed on the roller, paper jams may result.
• Printing is not possible if the front and back of the recording paper are
reversed.
• Always use the paper cutter on the printer’s cover to cut the printed paper.
Excessive paper dust can accumulate on the roller if the paper is cut by the
print head.
Handling and Storing Recording Paper
The recording paper is thermally and chemically sensitized. Observe the follow-
ing precautions to avoid paper discoloration and fading.
Avoid exposure to
direct sunlight.
Do not store thermal
paper above 40°C or
90% RH.
Avoid stacking with
wet Diazo copy paper.
Avoid exposure to vola-
tile organic solvents like
alcohol, ethers and ke-
tones.
Avoid contact with ad-
hesive tapes like soft vi-
nyl chloride and
cellophane tape.
3.3 Loading Recording Paper (With a Printer Module Installed)
43
Loading Recording Paper in the Model 8995 A4 Printer Unit
Retainer
Print Roller
Back Side of
Recording Paper
Paper Roll
Holders
Stock Cover
1
Pull the knob to open the stock cover.
2
Pull the head-raising lever forward.
The print head is separated from the print roller.
(Head-raised condition)
3
Put the paper roll holders in the ends of the
paper roll, and place it in the retainers.
4
Insert the paper through the slot under the print
roller, and pull the paper out so that the print
roller turns.
5
Align the edge of the paper that you pulled out
with the edge of the paper still on the roll.
If the paper is misaligned, it will crease as it feeds.
6
Return the head-raising lever to the lowered
position.
The print head should rest on the paper. (Head-low-
ered condition)
7
Insert the paper through the paper outlet in the
stock cover, and pull a little out.
8
Close the stock cover and cut the paper using
the cutter at the paper outlet to finish the proce-
dure.
Head-Raising Lever
Head-Raising
Lever
2
1
Print Roller
Knob
Paper Cutter
3.3 Loading Recording Paper (With a Printer Module Installed)
44
Retainer
Axle
2
1
Loading Recording Paper in the Model 8995-01 A6 Printer Unit
Print Roller
Back Side of
Recording Paper
Stock Cover
1
Push the stock cover to open the stock cover.
2
Raise the head-raising lever.
The print head is separated from the print roller.
(Head-raised condition)
3
Insert the paper through the slot under the print
roller, and pull the paper out so that the print
roller turns.
4
Insert the paper roll axle through the center of
the paper roll, and place it in the retainers.
5
Align the edge of the paper that you pulled out
with the edge of the paper still on the roll.
If the paper is misaligned, it will crease as it feeds.
6
Return the head-raising lever to the lowered
position.
The print head should rest on the paper. (Head-low-
ered condition)
7
Insert the paper through the paper outlet in the
stock cover, and pull a little out.
8
Close the stock cover and cut the paper using
the cutter at the paper outlet to finish the proce-
dure.
Print Roller
Paper Cutter
Head-Raising Lever
Head-Raising
Lever
Printer
Head
3.4 Connecting the Power Cord
45
When measuring in an electrically noisy environment, the effects of noise can be
minimized by grounding the Functional Earth terminals.
3.4
Connecting the Power Cord
• Before turning the instrument on, make sure the supply voltage matches
that indicated on the its power connector. Connection to an improper
supply voltage may damage the instrument and present an electrical haz-
ard.
Rated mains supply voltage (100 to 240 V AC), rated mains supply fre-
quency (50/60 Hz) (Voltage fluctuations of ±10% from the rated supply
voltage are taken into account.)
• To avoid electrical accidents and to maintain the safety specifications of
this instrument, connect the power cord only to a 3-contact (two-conduc-
tor + ground) outlet.
1
Verify that the instrument’s power switch is turned off.
2
Confirm that the mains supply voltage matches the
instrument, and connect the power cord to the power inlet
on the instrument.
3
Plug the power cord into the mains outlet.
Connection Procedure
Left Side
3.5
Grounding the Instrument’s Functional
Earth
Left Side
Functional Earth Terminals
Right Side
3.6 Turning the Power On and Off
46
When using the optional Model 9684 DC Power Unit, refer to "3.2.5 If the Model 9684 DC
Power Unit is Installed" in the Instruction Manual.
3.6
Turning the Power On and Off
Turn the POWER switch on ( | ).
After Power-On
After a short time, the STANDBY/ON key lights red and the Opening screen
appears.
The next time power is turned on, the Waveform screen for the function that
was displayed when power was turned off will appear.
Power On
Left Side
Turning Power On
Verify that the instrument and peripheral devices are correctly connected.
Before turning
power on
Before Starting Measurement
To obtain precise measurements, provide about 30 minutes warm-up (depend-
ing on installed input modules) after turning power on to allow the internal tem-
perature of the input modules to stabilize.
After that, perform zero adjustment before taking measurements.
Refer to the Input Module Guide for input module specifications and zero adjust-
ment.
Turn the POWER switch on (
).
When power is turned on again, the display appears with the settings that
existed when power was last turned off.
Turning Power Off
Power Off
Recording Data
When the POWER switch or STANDBY/ON key of the instrument is turned
off, internal recorded data is erased.
However, waveform data can be preserved when power is turned off if the
optional Model 9719 Memory Backup Unit is installed.
If you don’t want to lose recorded data, save it first to a PC Card or external
storage media.
Saving Procedure: "Chapter 5 Saving & Loading Data" (⇒ p. 65)
Instrument settings are preserved when power is turned off.
Before Turning
Power Off
3.6 Turning the Power On and Off
47
To Avoid the Startup Delay (Entering the Standby State)
Press the STANDBY/ON key.
“Enter Standby Mode?” appears.
Press the STANDBY/ON key again. (It slowly blinks red)
The screen turns off and the instrument enters the standby
state.
The fan continues to run for a while after entering standby. Also,
the fans run periodically during standby to avoid internal heating.
To resume operation from standby:
Press the STANDBY/ON key. (It lights red)
The initial or waveform screen reappears.
3.7 Setting the Clock
48
To adjust the clock, set the date and time on the Initialization screen of the Sys-
tem screen.
3.7
Setting the Clock
If you need to change both the time zone and the date and time, change the
time zone first.
See "12.3.1 Setting the Date and Time" in the Instruction Manual
5
Setting Procedure
Open the Initialization Settings Screen −−−−−−−−
1
When the Initialization screen appears:
Press the F7 [System] key.
If a screen other than the Initialization screen appears:
Use the CURSOR keys to move the cursor to the Function
menu at the upper left, and press the F7 [System] key.
The System screen appears.
2
Press the SUB MENU keys to select the [Init] menu item. The
Init (Initialization) Settings screen appears.
Set the time zone −−−−−−−−−−−−−−−−−−−−−−−−−−−−−−
3
Use the CURSOR keys to move the cursor to the locale set-
ting field as shown at the left, and select regions with the F
key.
4
Use the CURSOR keys to move the cursor to the [Execute]
key, and press the F1 [Execute] key.
Set the date−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−
Example: Set the date and time to May 5th 2005, 10:20:30.
5
Use the CURSOR keys to move the cursor to the date setting
field as shown at the left, and press the F1 [Set] key.
6
Use the
CURSOR keys to move to the year, month and
day, and change the numbers with the
CURSOR keys.
(Setting can be made with the F1 to F4 keys)
7
When finished changing the date, press F5 [OK] or the
ENTER key.
Set the time−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−
8
Press the
CURSOR key to move the cursor to the time set-
ting field, and press the F1 [Set] key.
Change the time setting like the date setting above.
9
When finished changing the time, press F5 [OK] or the
ENTER key.
Apply the new settings −−−−−−−−−−−−−−−−−−−−−−−−
10
Use the CURSOR keys to move the cursor to the [Execute]
key, and press the F1 [Execute] key.
F7
Function Menu
3
The date has been set.
Time Setting Field
System Screen
Init Menu Item
Date Setting Field
Time Zone Setting Field
6,7
8
9
2
4
4.1 Pre-Measurement Inspection
49
Before measuring, be sure to read the "Operating Precautions" (⇒ p. 5) and "Chapter 3
Measurement Preparations" (⇒ p. 37).
Basic Operations Chapter 4
4.1
Pre-Measurement Inspection
Do not use if damage is present, as
you could receive an electric shock.
Replace the damaged items.
Metal Exposed
Is the insulation of the probe or connection
cable to be used damaged, or is bare metal
exposed?
When using probes and connection cables
When using a clamp
Is the clamp cracked or damaged?
Yes
1
No Metal Exposed
2
To
No
2
Before using the instrument the first time, verify that it operates normally to ensure that the no
damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki
representative.
Peripheral Device Inspection
To
If damage is evident, request re-
pairs.
Is damage to the instrument or input mod- Yes
ules evident?
Instrument and Input Module Inspection
When turning power on
Do the fans rotate and the Hioki logo
appear on the screen?
No
2
The power cord may be damaged,
or the instrument may be damaged
internally. Request repairs.
Does the Initial or Waveform screen
appear?
No
Yes
The instrument may be damaged in-
ternally. Request repairs.
Nothing appears, or
the display is abnor-
mal
Yes
Inspection complete
4.2 Operation Workflow
50
4.2
Operation Workflow
Recording
Select with the SUB MENU keys
Settings
Set measurement configuration
Set up input channels
Load setting data (⇒ p. 74)
Set trigger criteria
Print (⇒ p. 75)
Save (⇒ p. 65)
Analyze
FILE key
START key
Waveform Screen
Select a Function
Settings Screen
File Screen
SET key
Waveform Screen
Range can be set auto-
matically. (⇒ p. 51) *1
Regular Recording
Record Specified Waveform (per trigger criteria)
Trigger criteria met
STOP key
Setting data can be loaded automatically. *2
Installation & Connections
"Chapter 3 Measurement Preparations" (⇒ p. 37)
"Chapter 10 Saving/Loading
Data & Managing Files" in the
Instruction Manual
"Chapter 11 Printing" in the
Instruction Manual
• "Chapter 8 Waveform Screen
Monitoring and Analysis" in the
Instruction Manual
• Analysis Supplement
Trigger mode: [Auto]
Settings for saving *3
Settings for printing *3
Set the screen layout of the
Waveform screen
Trigger criteria met
(Auto)
Trigger mode: [Single]
Trigger mode: [Repeat]
Record Once
Record Repeatedly
Turn Power On
(Awaiting trigger)
STOP key
*1. To enable automatic setting: "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual
*2. To load settings automatically: "10.5 Saving & Loading Auto Settings File (Auto Setup Function)" in the Instruction
Manual
*3. When saving or printing automatically, execution timing depends on the function.
When saving or printing manually, settings can be changed after measurement.
Finished Recording
Recording Starts
Recording Starts
Start Recording
Measure with New Settings
Measure with Existing Settings
Set as occasion demands.
Analysis, Saving &
Printing
4.3 Recording Simple Waveforms
51
First display the waveform for confirmation.
This description presumes that settings are at their factory default values. If the instrument has been used
before, or if you suspect that settings are not at their factory default values, reinitialize the instrument before
proceeding.
See "12.3.2 Initializing Waveform Data", "12.3.3 Initializing System Settings (System Reset)", "Appendix 2.1 List of
Default Settings" in the Instruction Manual
4.3
Recording Simple Waveforms
Example: Record a 120 V AC (60 Hz) waveform.
To display waveforms, set the horizontal axis (timebase), vertical axis (measurement range) and recording
length. In the measurement procedure described here, the range is set automatically.
Use Auto Setup to configure the lowest-numbered channel among those in use.
Auto Setup is recommended when the measurement range is undetermined, or when you want to verify
waveform input. However, depending on the measurement mode of the input module, Auto Setup may not
be available.
See "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual
Input modules and measurement modes not supported by Auto Setup:
• The [Temp] mode of the Model 8937 Voltage/Temp Unit
• The Model 8939 Strain Unit
• [Count], [Duty] and [50/60 Hz] (mains voltage measurement) modes of the Model 8940 F/V Unit
• [Charge] and [Preamp] modes of the Model 8947 Charge Unit
• The Model 8958 16-Ch Scanner Unit
• The Model 8960 Strain Unit
Turn Power On
Opening Screen
Select a Function
Press the F1 [MEM] key.
The Opening screen appears.
Make Necessary Connections
Refer to "Chapter 3 Measurement Preparations" (⇒ p. 37) and "Chapter 2 Connections" in the
Input Module Guide.
The Waveform screen appears.
4.3 Recording Simple Waveforms
52
Set Up the Instrument (Auto Setup)
1
Press the FUNCTION MODE key
to enable the FN mode.
2
Press the F4 [Auto Setup] key.
A confirmation message appears.
3
Press the F1 [OK] key.
The timebase, measurement range
and zero position are automatically
set, and recording starts.
Recording
continues
until
you
press the STOP key.
FN
Auto Setup makes the following settings:
While viewing a waveform, settings can be
changed as occasion demands.
See "To Change Settings" (⇒ p. 53)
Settings
Screen
Setting Item
Setting Value
Status
1 Timebase (Time)
1 ms/div
2 Shot (Recording
length)
(Unavailable in Auto
Setup)
Channel
Mode
Voltage
3 Range
20 V/div
4 Position
50 %
Trigger
5 Trigger Mode
Auto
Source (AND/OR)
OR
6 Pre-Trigger
20%
(Analog Trigger)
(Only lowest-
numbered module and
channel)
7
Type
Level
8
Level
Approx. 0 V
9
Filter
Off
1
2
4
5
1
3
1 div
7
9
8
6
SUB MENU keys
Finish Measurement
Press the STOP key.
4.3 Recording Simple Waveforms
53
To Change Settings
See "4.4 Setting Measurement Configuration on the Wave-
form Screen"
"5.7 Setting Input Channels from the Waveform Screen"
"6.12 Making Trigger Settings on the Waveform Screen"
in the Instruction Manual
Setting Items
Input Channel
Settings Dialog
After changing a setting, press the START key again.
However, if the change is made while measuring, the setting is reflected at the moment it is
changed.
Changing the timebase or recording length
Use the CURSOR keys to move the cursor to the set-
ting item, and press F1 to F8 to select the setting val-
ue. Pressing the SUB MENU keys changes available
setting items.
Changing the selected input channel
Press the UNIT or CH key to display the [CH SET] di-
alog, and select the channel to be changed.
• Changing the measurement range
Turn the inner RANGE knob.
• Changing the waveform display position (zero
position)
Turn the outer POSN knob.
• Erasing unwanted waveforms
Select the unit (module) and channel of the
waveform to be erased, and change [On] to [Off].
Press the ESC key to remove the dialog.
Verifying a Waveform
A waveform’s horizontal axis can be
expanded or compressed (⇒ p. 62)
The horizontal axis can be expanded or compressed
by setting the magnification in the [Mag] (Magnifica-
tion) field.
By selecting the [Zoom] button, a section of the wave-
form can be expanded.
See "8.9 Magnifying and Compressing Waveforms" in the
Instruction Manual
Reading values at cursors (⇒ p. 63)
Press the TYPE key to open the [A/B Cursor] dialog.
Select the cursor type, and move the A or B cursor by turning the
A or B knob, respectively.
The values at the A/B cursors appear at the top right.
See "8.8 Cursor Values" in the Instruction Manual
Shuttle
Scrolling a waveform
(⇒ p. 35)
Turn the inner Jog knob of the
SCROLL control to scroll slowly,
or the outer Shuttle knob to scroll
quickly.
See "8.1 Scrolling Waveforms" in the
Instruction Manual
Measurement
value at cursor
Cursor A
Cursor B
Jog
4.4 Actual Measurement and Analysis
54
This example uses an analog input module and connection cable to illustrate the basic measurement process
of recording and analyzing an anomalous waveform. Settings and measurement procedures differ according
to the specific application, as described in the Input Module Guide, the Instruction Manual and the Analysis
Supplement .
This description presumes that settings are at their factory default values. If the instrument has been used
before, or if you suspect that settings are not at their factory default values, reinitialize the instrument before
proceeding.
See "12.3.2 Initializing Waveform Data", "12.3.3 Initializing System Settings (System Reset)", "Appendix 2.1 List of
Default Settings" in the Instruction Manual
Measurement Example Overview
Record voltage sag on a 240 V, 60 Hz AC line. When the anomaly occurs, data is automati-
cally saved to a PC Card for analysis.
1 Make Basic Settings
Make the basic settings required for measurement such as measurement configuration (recording speed
[Timebase] or recording time [Shot]) and input channel selection.
2 Make settings to record the specified waveform anomaly (trigger criteria settings)
Make appropriate trigger settings to acquire the desired waveform data.
(Default settings are fine for continuously recording waveforms)
3 Make settings to save recording data to storage media
To save automatically while measuring, enable Auto Save before starting measurement. After measuring,
data remaining in the instrument's internal memory can be saved manually.
(Printing works the same way)
4 Start measurement, and analyze recorded data
After measuring, you can analyze the data, and save or print it as occasion demands.
4.4
Actual Measurement and Analysis
Setting required for this measurement:
*1. Unit (module) and channel number to use
*2. Trigger settings are not needed to only observe
fluctuations of the mains voltage.
Settings
Screen
Setting Item
Setting Value
(Function)
MEM (Memory)
Status
Timebase
5 ms/div
Shot (Recording
length)
50 div
Channel
Unit, Ch
Unit 1, Ch 1*1
Mode
Voltage
Range
20 V/div
Position
50 [%] (default setting)
Mag (Magnification)
x1/2
Trigger*2
Trigger Mode
Repeat
Pre-Trigger
10%
Type
Drop (Voltage Sag)
Level
280 V (RMS 197.99 V)
Freq (Frequency)
60 Hz
Save
Auto Save
On
Save in
(PC Card folder)
240 V AC
Load
Required equipment:
• This instrument
• Grounded power cord (2-conductor + ground)
• Model 8936 Analog Unit............1
• Model 9198 Connection Cord ...1
• PC Card
• Recording Paper
In this case, the Model 8936 Analog
Unit is installed as Unit 1 in the Model
8860, so the measurement proce-
dure describes connection of the
Model 9198 Connection Cord to Ch
1.
CH1
UNIT 1
CH2
2
3 4
8936
9198
H
L
Verifying Before Measurement
Input Module Compartment
4.4 Actual Measurement and Analysis
55
1
Connect the cable to a BNC jack
on the Model 8936 Analog Unit.
Connect the cable clips to the AC
power line.
1. Measurement Preparations
UNIT 1
See "Chapter 3 Measurement Preparations" (⇒ p. 37), and "Chapter 2 Connections" in the Input Module Guide
This example uses channel
1 (Ch 1).
Right Side
Left Side
Turn the power on.
2 3 4
1
Connect the power cord from
this instrument to an outlet.
Make sure the power switch is
turned off.
2
8936
CH1
4
9198 Connection
Cord
3
Twist clockwise to lock.
Insert a PC Card into PC CARD
Slot 1.
PC CARD Slot
No preparation is required when using the optional hard disk.
Verify that write protection on the storage media is disabled.
Preparations for Data Saving and Loading
Using a PC Card
Using an MO disk
(Using an internal MO unit)
USB Port
Model 9717 MO Unit
To save or print automatically, be sure to prepare the required devices before measurement.
In this case, insert a PC Card into PC CARD Slot 1.
Connect the drive to the USB
port, and insert storage media.
(Example: Model 9716 FD Drive)
Using a USB storage media
drive
Insert an MO disk into the MO
disk slot.
Verify that the paper is loaded correctly.
Using the internal printer: "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42)
Using an external printer: Connect to the USB port and load paper.
Printing Preparations
4.4 Actual Measurement and Analysis
56
2. Select the appropriate function
Function Menu
Opening Screen Case
Press the F1 [MEM] (Memory Function) key.
Using the CURSOR keys, move the cursor to the
Function menu, and press the F1 [MEM] key.
See "Function Selection" (⇒ p. 29)
Waveform or Setting Screen Case
3. Set measurement configuration
2 Press the SUB MENU keys to select the [Status] menu item.
3 Press the SHEET/PAGE keys to
select the [Basic] page.
4 Set the timebase to [5 ms/div].
This is the time per division on the hori-
zontal axis.
Sampling rate = timebase (s/div) / 100
(samples per division)
5 Set the recording length to [50 div].
Set the recording time in number of divi-
sions.
(Recording time = timebase (s/div) × re-
cording length (div))
Now use the CURSOR keys to move
among the setting items, and press the
F1 to F8 keys to select the desired set-
ting.
Make other settings as occasion demands.
Supplement
• When F8 [Page / ] appears, more selections are
available. Press the F8 key to display the
additional selections.
• In the memory function case, sampling speed is 1/
100th the timebase. Each division contains 100
data samples. More detailed analyses are
possible when the timebase is set smaller.
• Enable auto-ranging if you cannot determine the
appropriate range (⇒ p. 51).
1 Press the SET key.
• When using multiple channels, verify that
appears with the channel to be used on the [Use
Ch] page.
See: "Setting the Timebase (Horizontal Axis)" (⇒ p. A2) in
this manual
"4.2.1 Selecting Channels to Use",
"4.2.2 Setting the Timebase (Horizontal Axis) and
Sampling Rate",
"4.2.4 Setting the Recording Length (number of divi-
sions)"
in the Instruction Manual
See "Chapter 4 Measurement Configuration Settings" in the Instruction Manual
4.4 Actual Measurement and Analysis
57
4. Set up the input channel
Check This!
Indicates the
type of module
(unit) selected.
3 Select the module [Unit1] and chan-
nel number [Ch1] to be set.
4 Make appropriate input channel set-
tings such as the measurement
range.
Make other settings as occasion demands.
Setting Item
Setting Value
(Verify) [Module]:Analog, Mode: Voltage
Range(/div)
20 V (per division)
(Each vertical division repre-
sents 20 V)
Coupling
DC (default setting)
5 Make appropriate waveform display
settings.
Make other settings as occasion demands.
See "7.1 Making Input Waveform Display Set-
tings (Analog Waveforms)" in the Instruction
Manual
Setting Item
Setting Value
Wave Disp
On (default setting)
Mag
(Magnification)
x1/2
Position
50% (default setting)
Supplement
Determining the range:
Measurement range is denoted as value per division
on the vertical axis.
Setting a measurement range causes the measure-
ment area and display area of the level monitor (on
the left side of the screen) to be correspondingly
changed.
It is convenient to watch the indicated display area
on the level monitor while setting the range.
6 Select the [Zero-Adjust] button.
Execute zero-adjust 30 minutes after turning power on.
Proper warm-up time depends on the particular input mod-
ule. Verify the specifications in the Input Module Guide.
A message appears when zero-adjust is fin-
ished.
1 Press the SUB MENU keys to select the [Channel] menu item.
2 Press the SHEET/PAGE keys
to select the [One Ch] page.
See "Chapter 5 Input Channel Settings" in the Instruction Manual and "Chapter 3 Input Channel Settings" in the
Input Module Guide
Now use the CURSOR keys to move
among the setting items, and press the
F1 to F8 keys to select the desired set-
ting.
Level Monitor
To enter a comment for an individual channel:
A comment entered in the comment entry field ap-
pears on the waveform.
See "5.2 Adding Comments" in the Instruction Manual
To convert the displayed units for measurements
taken with a clamp or sensor, set the Scaling func-
tion.
See "5.4 Converting Input Values (Scaling Function)" on
the Instruction Manual
4.4 Actual Measurement and Analysis
58
5. Set Trigger Criteria
3 Press the SHEET/PAGE keys to
select the [Analog 1-4] page.
2 Set the trigger.
See "2.3.2 Entering Text and Numbers" (⇒ p. 33)
Setting Item
Setting Value
Trigger Mode *1
Repeat
Pre-Trigger (% setting) *2
10%
4 Set the analog trigger.*4
Make other settings as occasion demands.
See "Chapter 6 Trigger Settings" in the
Instruction Manual
Setting Item
Setting Value
[Normal]
(default setting)
(Applies one trigger to one
channel)
Type
Drop (trigger type)
Unit
1-1 (Unit1-Ch1)
Level *3
280 V (trigger level)
Parameter
Frequency: 60 Hz
Supplement
*1. About (trigger) modes
• Select [Auto] if you want to view the input
waveform continuously, even when no trigger is
applied.
• Select [Single] if you want to record just one event
and stop.
• Select [Repeat] to record continuously until you
press the STOP key.
See "6.3 Setting the Trigger Mode" in the Instruction
Manual
*2. About pre-triggering
To record the waveform prior to the trigger event, set
the trigger point at some percentage of the recording
length.
See "6.5 Pre-Trigger Settings" in the Instruction Manual
*3. About (trigger) level
The trigger level is set as an instantaneous value in-
stead of an RMS value.
1 Press the SUB MENU keys to select the [Trigger] menu item.
See "Chapter 6 Trigger Settings" in the Instruction Manual
Now use the CURSOR keys to move among
the setting items, and press the F1 to F8
keys to select the desired setting.
*4. About analog and logic triggering
• This selection is made from the Settings dialog.
Move the cursor to the [No.] field and press the F1
[All Settings] key to display the dialog. Set each
item in the dialog, and press the ENTER key or
[Close] button when done.
• [Normal] and [Expanded] analog triggers
Select [Normal] to apply one trigger type on one
channel, and select [Expanded] to apply multiple
triggers on one channel.
(The [Normal] setting is available for level, window
threshold and voltage sag trigger types.)
See "6.7 Triggering by Analog Signals" in the Instruction
Manual
4.4 Actual Measurement and Analysis
59
6. Enable Auto-Save
Check This! [On]
(default setting)
Trigger date and time are
automatically added as a
prefix to the specified sav-
ing name.
“Time_Date_Saving Name”
(default setting)
Select “Text” to make
the file readable on a
PC.
3 Set Auto Save [On].
4 Select the saving destination.
At [Save in 1], press the F1 key to open
the [Browse Folders] dialog.
The waveform can now be saved with other
settings in their default states. Change these
settings as occasion demands.
See "10.3 Saving Data"
"10.3.4 Setting Auto Save"
"10.3.7 Automatically Saving Waveforms"
in the Instruction Manual
8860
Supplement
• Using Auto Save, waveforms, numerical calculation
results and screen image can be saved at the same
time.
• When the [Save in 1] storage media becomes full, sub-
sequent recording data can be stored on [Save in 2]
storage media. (⇒ p. A10)
• If the [Save Method] setting is [Normal Save] (default
setting) and the storage media becomes full, an error
message appears. Stop measurement and change the
storage media.
1 Press the SUB MENU keys to select the [Save] menu item.
What about these situations?
To record over a long period
Use high-capacity storage media.
Specify a folder other than the root directory as the save destination. When
more than 5,000 files have been created in one folder, a new folder is automat-
ically created and saving continues.
To continue saving even when the storage media becomes full, select [Delete
Save] as the [Save Method]. When the storage media becomes full, the oldest
file is deleted and saving continues.
See "10.3.4 Setting Auto Save" in the Instruction Manual
To also print automatically
On the Print Settings screen, set Auto Print [On]. Every time a waveform is ac-
quired, it is saved and printed.
See "Chapter 11 Printing" in the Instruction Manual
Verify once again before starting measurement.
• Are connections made correctly?
• Is write-protection on the storage media disabled?
• Is storage media inserted?
See "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual
Now use the CURSOR keys to move
among the setting items, and press the
F1 to F8 keys to select the desired set-
ting.
CURSOR: Selects the storage
media.
Select [PC Card #1] (PC Card Slot 1).
To navigate the tree, press
CURSOR.
2 Press the SHEET/PAGE keys to
select the [Auto Save] page.
4.4 Actual Measurement and Analysis
60
7. Start Measurement
Press the START key.
The green LED lights, and the Waveform screen appears.
The instrument’s measurement status is displayed on the Status bar.
Recording does not occur until the specified trigger criteria are met.
The waveform is displayed.
(If the [Roll Mode] setting is [Off] on the Status setting screen, the
screen displays acquired data in quantities determined by the specified
by the recording length.)
Every time data is acquired, it is saved to the PC Card and the “Trigger
Wait” state is resumed.
Trigger Wait
Storing
Saving
Save to PC Card
Status Bar Display
In this measurement case, 50 divisions of data (including the pre-trigger setting of 10% of the 50-divi-
sion recording length) are recorded when the voltage is 280 V (197.99 Vrms) or less, at 5 ms per divi-
sion on the horizontal time axis.
Settings can be changed and the waveform can be scrolled while recording.
Recording restarts when setting are changed. (“Storage” is displayed)
See "3.3.6 Starting and Stopping Measurement" in the Instruction Manual
Press the STOP key.
8. Finish Measurement
See "3.3.6 Starting and Stopping Measurement" in the Instruction Manual
• Pressing Once: Recording stops when finished acquiring data for the recording length.
• Pressing Twice (Halt): Recording stops immediately.
If you halt while awaiting a trigger, no waveform is displayed. (However, with the Memory function,
the waveform does appear if the recording length is shorter than half of the maximum recording
length and a trigger event has occurred before the Halt command.)
What about these situations?
You can disable the keys to
prevent
inadvertent
opera-
tions
Hold both
CURSOR keys simultaneously for three seconds. “Key Lock”
appears under the clock display while the key-lock state is enabled. To cancel
key-lock, hold the keys again for three seconds.
When "Trigger Wait" contin-
ues to be displayed and no
waveform appears
If the trigger criteria are not met, no waveform is displayed (⇒ p. A10).
If you need to display the waveform, execute a forced (manual) trigger.
(FN Mode: F6 key)
To view part of the waveform
that is not visible while re-
cording
Use the SCROLL controls to scroll the waveform. When turning the inner Jog
or outer Shuttle knob, the [Scroll Trace] button appears at the left. Press the F1
[Scroll Trace] key to return to display of the currently recording waveform.
Refer to the scrolling methods: "2.3.3 Viewing Recorded Data" (⇒ p. 35)
To change settings while re-
cording
Timebase, recording length, input channel and trigger settings can be changed
from the Waveform screen.
See "To Change Settings" (⇒ p. 53),
"4.4 Setting Measurement Configuration on the Waveform Screen" in the
Instruction Manual
4.4 Actual Measurement and Analysis
61
9. Analyze Data
What about these situations?
To measure with changed set-
tings
After changing settings on the Waveform or Settings screen, press the START
key.
To load saved data from stor-
age media
Press the FILE key to display the File screen, and select the file to load.
See "10.4 Loading Data" in the Instruction Manual
To print recorded waveforms
or settings
Make the appropriate settings on the Print Settings screen, and press the
PRINT key.
See "6.3 Manual Printing (PRINT Key)" (⇒ p. 78) in this manual,
"11.4 Making Manual Print (PRINT Key Output) Settings" in the Instruction Manual
To save settings
Press the SAVE key and select [Settings] to choose the save type, and save.
By saving settings, you can reload them later when you want to repeat measure-
ment with the same settings. Saved settings can also be automatically reloaded
when power is turned on.
See "5.5 Data Saving" (⇒ p. 71) and "5.6 Data Loading" (⇒ p. 74) in this manual
To use recorded data on a PC
Select the waveform or display the image type, and save it.
For details about data saving types, refer to "5.3 Data that can be Saved &
Loaded" (⇒ p. 68)
Use the LAN interface to connect to a PC.
See "Chapter 13 Communications Settings" in the Instruction Manual
Time from Trigger Event
Number of Waveform Trigger Event Time
Acquisitions
Name of Currently Displayed Sheet
See "Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual
Displayed Channel
Trigger Event Position
4.4 Actual Measurement and Analysis
62
Viewing the Whole Waveform
Select the [Mag] (Magnification) button
The whole waveform over the specified recording length
is displayed.
Using the CURSOR keys, move the cursor to
[Mag], and press the F1 [Whole Wave] key.
Whole waveform
See "8.9 Magnifying and Compressing Waveforms" in the
Instruction Manual
Expanding Horizontally
See "8.9.1 Magnifying and Compressing Horizontally (Time
Axis)" in the Instruction Manual
Select the horizontal axis magnification
The waveform view is magnified as selected.
Use the CURSOR keys to move the cursor to the
magnification setting field, and press F1 to F8 to
select the desired magnification.
Zooming Part of a Waveform
See "8.9.3 Magnifying a Section of the Horizontal Axis (Time
Axis – Zoom Function)" in the Instruction Manual
Use the Zoom function
Use the SCROLL control to scroll the zoomed section
of the waveform.
To return to the previous view, move the cursor to
[Zoom], and press the F1 [Off] key.
1 Use the CURSOR keys to move the cursor to
[Zoom], and press the F2 [On] key.
The Zoom display is enabled.
2 Move the cursor to the Zoom setting field, and
select the desired magnification.
Zoomed Waveform Section
Waveform with Section to
be Zoomed
4.4 Actual Measurement and Analysis
63
Viewing Measurement Values
See "8.8 Cursor Values" in the Instruction Manual
Use the A/B cursors
Dialog
Press the TYPE key to select the cursor type.
The [A/B Cursor] dialog appears.
• To view the time and voltage (measurement) value:
[Trace]
• To see time and period: [Vertical]
• To see the voltage value: [Horizontal]
Make other settings as occasion demands.
Knob A
Knob B
Turn inner knob A or outer knob B to move the cur-
sor to the point to be read.
The value at the cursor is displayed.
Voltage value
Cursor A value
Cursor B value
Press the DISP key to display method. (Waveform/
Waveform and Information Display/ Gauge and Wave-
form Display/ Gauge, Waveform and Information Dis-
play)
The selection can be made in the dialog.
Press the ESC key to close the dialog.
See "8.4 Displaying Measured Values and Information" in the
Instruction Manual
To change the gauge of a displayed channel, press the
FUNCTION MODE key to enable the FN mode, and press
the F2 [Gauge] key. In the dialog, select the channel for
which to display the gauge.
To remove the displayed gauge, press the DISP key to switch
to Normal display.
See "8.5 Applying Gauges" in the Instruction Manual
Applying a Gauge
To remove the cursors, press the FUNCTION key to return
to the FN mode, and press the F1 [Switch Info] key.
Trace Cursor
Difference between A/B
cursors
Time from trigger event
Gauge Display
Channel
If Overlapping Display of Waveforms and
Numerical Values Obstructs Viewing
1
2
4.4 Actual Measurement and Analysis
64
1 Specify the section to print using the vertical or
trace cursors.
See"Viewing Measurement Values" (⇒ p. 63)
10. Printing a Selected Waveform Section
See "Chapter 11 Printing" in the Instruction Manual
2 Press the PRINT key.
The [Print] dialog appears.
3 Press the F2 [A-B Wave] key.
The specified section is printed.
This procedure describes Selection Printing (default setting).
Function, Timebase
Upper Limit of Display Area
Trigger Mark
Unit-Channel Number
Lower Limit of Display Area
Supplement
To print items other than waveforms such as a settings
list, gauge and upper and lower limits, select which items
you want to print on the [Print Items] page.
See "6.4 To Print a Detailed Settings List" (⇒ p. 79)
Set the number of cursors to [2].
Channel Settings
Saving & Loading
Data
Chapter 5
Compatible Recording Media
(⇒ p. 66)
Load into this instrument (File Screen)
• Settings Data
• Waveform Data
Compatible Data Types, and Procedures
(⇒ p. 71)
Settings Data
Waveform Data
• Save automatically while measuring (⇒ p. 59)
• Save waveform selection after measurement (SAVE
key)
Numerical Calculation Results
• Calculate and save automatically while measuring
• Calculate and save after measuring (SAVE key)
Screen Image (Display Screen)
• Save automatically while measuring
• Select and save a screen after measuring (SAVE
key)
The default at shipping and initialization of this setting is [Selection Save]. To save selections, press the
SAVE key, make the appropriate settings and save them. You can later change your saved settings from
the Save Settings screen as occasion demands.
For details, refer to "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual.
Data Types & Formats
(⇒ p. 68)
Saving Procedure Overview
(⇒ p. 71)
Loading Data
See "10.4 Loading Data" in the Instruction Manual
Viewing Data from this Instrument on a PC
See "Chapter 13 Communications Settings" in the
Instruction Manual
Handling Storage Media
See "10.1 Storage Media" in the Instruction Manual
This document describes saving and loading procedures
using PC Cards in the internal PC Card slots.
This document describes saving procedures commonly
used after measurement.
Save Procedure Details
See "10.3 Saving Data" in the Instruction Manual
File Size
See "Appendix 2.2 Waveform File Sizes" in the
Instruction Manual
File management on this instrument
(File Screen)
Initialize (format) storage media
Internal storage media file and folder operations
Formatting Storage Media
See "10.1.7 Initializing (Formatting) Storage Media" in
the Instruction Manual
File Management
See "10.7 Managing Files" in the Instruction Manual
• Copy
• Move
• Delete
• Rename
• Create folder
On-screen operations
• Sort
• Display only specified files (Filter Settings)
• Select display items from the File List (Display Set-
tings)
• Print the File List
5.1 Recording (Storage) Media
66
5.1
Recording (Storage) Media
Storage Media
Remarks and Hioki Options
Storage Media Designation
(when specifying the storage
media type)
PC Card
(Built-in slots provided)
• Model 9626 PC CARD 32M
• Model 9627 PC CARD 64M
• Model 9726 PC CARD 128M
• Model 9727 PC CARD 256M
• Model 9728 PC CARD 512M
• Model 9729 PC CARD 1G
Insert in a PC CARD slot on the right side of the instrument.
(⇒ p. 67) (Two slots are provided)
PC CARD #1 (PC CARD Slot 1) *2
PC CARD #2 (PC CARD Slot 2) *2
MO Disk
Model 9717 MO Unit *1
(Compatible MO disks: 3.5-inch, with capacity of 128, 230,
540 or 640 MB, and 1.3 or 2.3 GB)
See "10.1.2 Using an MO Disk" in the Instruction Manual
MO (with internal Model 9717 MO
Unit installed) *2
Floppy Diskette
Model 9716 FD Drive
(Compatible floppy diskettes: 3.5-inch, 2-HD or 2-DD, with
capacity of 720 KB or 1.44 MB)
See "10.1.3 Using a Floppy Disk" in the Instruction Manual
FDD #1, FDD #2, …, FDD #5
(using the Model 9716 FD Drive) *2
Hard Disk
Model 9718 HD Unit *1 (Capacity: 60 GB)
See "10.1.4 Using a Hard Disk" in the Instruction Manual
HDD
(USB drives)
Commonly available USB memory or other USB storage de-
vices can be connected to the USB port.*4
See "10.1.5 Using USB Memory Devices" in the
Instruction Manual
USB DISK #1, USB DISK #2, ... ,
USB DISK #5
(Networks)
Data can be transferred directly to and from PCs over a net-
work.
See "Chapter 13 Communications Settings" in the
Instruction Manual
NETWORK #1, NETWORK #2, ... ,
NETWORK #10 *3
For details about saving and loading: "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual
For details about inserting storage media and connecting drives:"Preparations for Data Saving and Loading" (⇒ p. 55)
*1. Either the Model 9717 MO Unit or 9718 HD Unit may be installed. Both cannot be installed together.
*2. Displayed when storage media is inserted.
*3. Displayed when connected to a network with sharing enabled.
*4. Not all commonly available media drives are supported.
When using a USB memory device:
Do not remove the USB memory device while it is being accessed (writing or reading).
Also, immediately after saving data and before removing it, verify on the File screen that the data has been saved.
No particular instrument operation is required to remove a USB memory device.
Be sure to read the above references regarding handling of each type of storage
media.
If damage occurs to the hard disk or internal memory, we cannot restore or ana-
lyze the lost data. We therefore recommend maintaining a backup of any impor-
tant data.
5.2 Using PC Cards
67
PC Card Insertion & Removal
5.2
Using PC Cards
Important
Use only PC Cards sold by Hioki (⇒ p. 66).
Compatibility and performance are not guaranteed for PC cards made by
other manufacturers. You may be unable to read from or save data to such
cards.
• Format new PC Cards before use.
To format with this instrument: "10.1.7 Initializing (Formatting) Storage Media"
in the Instruction Manual
• Be careful to avoid inserting a PC Card backwards or upside-down. The PC
Card or the instrument could be damaged.
• Never eject a PC Card while it is being accessed by the instrument. Data on
the PC Card could be lost.
• When storage media is inserted, a confirmation beep sounds, and an icon
appears in the Status bar at the lower right.
"Appendix 1 Waveform Screen Display & Icons" (⇒ p. A1)
• Be sure that storage media is inserted before selecting it for saving or loading.
It cannot be displayed in the File List if it has not been inserted.
• The Flash memory in a PC Card has a limited operating life. After long-term
usage, data storage and retrieval become difficult. In this case, replace the PC
Card with a new one.
• We cannot provide compensation for data loss in a memory card, regardless of
content or cause of the damage. Always maintain a backup of important data
stored on a memory card.
Right Side of Instrument
PC CARD Slots (two)
Eject Button
Inserting a PC Card
With the Front mark (�) pointing
toward the slot, insert the card in
the direction of the arrow all the way
in.
Removing a PC Card
Press the Eject button. When the
button pops out, press it again to
eject the PC Card.
5.3 Data that can be Saved & Loaded
68
The following data can be saved and loaded with this instrument. Auto Save saves the data automatically
after measurement.
5.3
Data that can be Saved & Loaded
“O” = Possible, “–” = Not Possible
File Type
File
Format
File Extension and Contents
Save
Load
PC-
Readable
Reference
Auto Manual
Settings Data*2
Settings made on the
Settings screen, such
as measurement
configurations
Binary
SET
Settings Data
(Measurement
Configuration)
—
O
O
—*5
Instruction Manual:
"10.3.6 Saving Settings Data"
"10.4.2 Loading Settings Data"*1
Waveform Data*2
All waveform data
acquired by the
instrument, or the
waveform section
demarcated by A/B
cursors.
Binary
MEM
Memory Function
Waveform Data
O
O
O
—*5,*6
Instruction Manual:
"10.3.7 Automatically Saving
Waveforms"
"10.3.8 Optionally Selecting
Waveforms & Saving (SAVE
Key)"
"10.6.2 Reading Waveform Data
on a PC"
"Chapter 9 Measuring with Real-
Time Saving"
REC
Recorder Function
Waveform Data
O
O
O
—*5,*6
RSM
Sampled waveform
data from the Real-
Time Saving function
O
—
O
—
RSR
Whole waveform data
from the Real-Time
Saving function
O
—
O
—
FFT
FFT Function Wave-
form Data
O
O
O
—*5
Text
TXT
Text Data
O
O
—
O
Waveform
Management
data (Divisions)*3
Management data for
dividing and saving
waveform data
(Index
file)
IDX
Index data for divided
saving
O
O
O
—*5
Instruction Manual:
"10.3.7 Automatically Saving
Waveforms"
"10.3.8 Optionally Selecting
Waveforms & Saving (SAVE
Key)"
(when [Binary] save type is
selected)
SEQ
Index data for memory
division (created
automatically for batch
saving)
O
O
O
—*5
Instruction Manual:
"4.3.3 Dividing Memory"
"8.12 Viewing Waveforms in
Every Display Block (Memory
Division)"
RSI
Index data for the Real-
Time Saving function
O
—
O
—
Instruction Manual:
"Chapter 9 Measuring with Real-
Time Saving"
Numerical
Calculation
Results
Text
TXT
Text Data
O
O
—
O
Analysis Supplement :
"1.4 Saving Numerical
Calculation Results"
"1.5 Reading Numerical
Calculation Results on a PC"
Captured Screen
Image *4
BMP
BMP
Image Data
O
O
—
O
Instruction Manual:
"10.3.9 Automatically Saving
Display Images"
"10.3.10 Optionally Selecting
Display Screens & Saving (SAVE
Key)"
PNG
PNG
Image Data
O
O
—
O
5.3 Data that can be Saved & Loaded
69
*1. Settings data can be loaded automatically at power-on. (Auto Setup Function)
*2. To reload data into this instrument:
Save it in binary format. Waveforms and some of the measurement settings are saved.
To load data into a PC: Save it in text format.
To save a waveform section, demarcate the section to save with the A/B cursors.
*3. To load all blocks at once when memory division is enabled:
Save using the [All Blocks] selection. A directory is created automatically, and files for the waveform data of each block
and an index file (SEQ) are created. This index file is used for reloading.
To reload waveform data saved with the Divided Saving function, load the index file (IDX).
To reload waveform data in the Real-time saving function, load the index file (RSI).
*4. BMP Format: This is a standard Windows graphics format. These files can be handled by many graphics programs.
PNG Format: This image file format has been internationally standardized as ISO/IEC15948.
*5. Loading is possible when using the optional Model 9725 Memory HiViewer.
*6. Loading is possible with the Waveform Viewer (Wv).
5.4 Saving Procedures and Contents
70
5.4
Saving Procedures and Contents
Basically, three methods are available for saving.
To save automatically while
measuring
To save immediately upon
pressing the SAVE key
To save selected contents
Auto Save
This method is convenient for saving
specific types of data, such as wave-
forms or images, or when you want
to save while monitoring. Specify
what you want to save before press-
ing the SAVE key.
(Default setting)
Press the SAVE key to select and
save the selected contents. No set-
tings need to be made on the Save
Settings screen, as settings can be
made on any screen.
Quick Save
Selection Save
Automatically saves all acquired
measurement data for the specified
recording length. Various types of
data can be saved at the same time.
Select the save destination and con-
tents before measuring (⇒ p. 59).
After installation and connection,
turn power on.
Insert storage media
Set measurement configuration,
etc.
Set the Auto Save setting [On]
Specify the saving destination
Select what to save
Measure
Saving proceeds automatically
From the Save Settings screen,
open the [Auto Save] page
The following can be saved
simultaneously:
• Waveform Data
• Numerical Calculation Results
• Screen Image
After installation and connection,
turn power on.
Insert storage media
Set the function of the SAVE key to
[Quick Save]
Specify the saving destination
Select what to save
Press the SAVE key*1
Save
From the Save Settings screen,
open the [SAVE key] page
Any of the following can be
saved:
• Waveform Data
• Settings Data
• Numerical Calculation Results
• Screen Image
After installation and connection,
turn power on.
Insert storage media
Set the function of the SAVE key to
[Selection Save]
Press the SAVE key.
Specify the saving destination
Select what to save
Select the [Execute] button*1
Save
From the Save Settings screen,
open the [SAVE key] page
Any of the following can be
saved:
• Waveform Data
• Settings Data
• Numerical Calculation Results
• Screen Image
Verify before saving
• Is storage media inserted?
• Is write-protection on the storage media disabled?
• Is the specified save destination correct?
• Is Auto Save set to [On] for automatic saving?
*1.To save only a section of a waveform, demarcate
the section with the A/B cursors. (Saving a wave-
form section is not available with Auto Save)
To save a screen image, display the desired
screen before saving.
5.5 Data Saving
71
The saving procedure for various data items using the default [Selection Save]
method is described here.
Available setting contents are the same for the [Quick Save] method.
Refer to "6. Enable Auto-Save" (⇒ p. 59) for more details about Auto Save.
Refer to "10.3 Saving Data" in the Instruction Manual for saving procedure
details.
5.5
Data Saving
Before setting, verify that write-protection is disabled on the destination
storage media, and insert it.
When entering a file name (to be read by a PC)
Do not use any of the following characters in a file name, as PCs running Win-
dows 2000 or XP may not be able to handle it:
ASCII: + = [ ] \ / | : * ? “ < > ; ,
Shared Settings in Selection Save
1 Press the SAVE key.
The [Save] dialog appears.
2 Press the F1 [Edit] key.
The [Browse Folders] dialog appears.
3 Use the
CURSOR keys to move the cursor to
the storage media, and press the F1 [OK] key.
To open a folder on the storage media, press the
CURSOR key with the
to the left of the folder select-
ed.
4 Press the F1 [Edit] key to enter the file name to
save. (Up to 40 characters)
See "2.3.2 Entering Text and Numbers" (⇒ p. 33)
If you have a keyboard connected, press the F2 [Direct]
key and enter from the keyboard directly.
5 If a file with the same name already exists in the
destination folder, select whether to overwrite or
append automatic numbering.
6 Select whether the trigger date and time should
be appended before or after the specified file
name.
Numbering
(Default setting)
If a file with the same name already ex-
ists, a number is automatically ap-
pended to the new file name to avoid
duplication. (Up to four digits)
Overwrite
Duplicate existing file names are over-
written.
Saving a Waveform
(⇒ p. 72)
Saving Settings Data
(⇒ p. 73)
Saving Captured Screen Images
(⇒ p. 73)
To enter the save name:
If a keyboard is connected,
you can press the F2 [Direct]
key and enter directly from the
keyboard.
5.5 Data Saving
72
Saving Waveforms
*2
See "10.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" in the Instruction Manual
1 Press the F2 [Waveform] key.
2 Select the data saving format.
3 Select the area to be saved.
4 Select the channels to save.
5 (when [Binary] saving format is selected)
Select to save by divisions depending on file size.
See "Appendix 2.2 Waveform File Sizes" in the
Instruction Manual
6 Press the F1 [Execute] key.
Binary
To be able to reload data into the instru-
ment (Default setting)
Text *1
To be able to load data into a PC
Whole
Save all measurement data.
(Default setting)
A-B
Save data in the area demarcated by
the A/B cursors.
Displayed Ch
Saves the channels on all sheets for
which waveform display is enabled
[On]. (Default setting)
All Ch
Saves all stored channels.
Off
Saves as one file without divisions.
(Divisions)
If the data is larger than the specified
recording length, it is divided and saved
in multiple files.
*1. When [Text] is selected as the save
type, data is saved with thinning.
Examples
2: Save every other data point.
(•x•x•x...)
5: Save one of every five data points.
(•xxxx•xxxx•xxxx..)
(•: saved, x: not saved)
"Shared Settings in Selection Save" (⇒ p. 71)
Use the CURSOR keys to move the cursor to the
setting item, and select the desired setting contents
with the F1 to F8 key.
*2: [Target Blocks]
This is displayed when using the Memo-
ry Division function.
When [All Blocks] is selected, the index
file (SEQ) and waveform data of all used
blocks are saved as a batch.
About Memory Division:
"4.3.3
Dividing Memory" in the
Instruction Manual
Note:
Data remaining in memory is saved by
manual saving with the Real-time sav-
ing function.
5.5 Data Saving
73
Saving Settings Data
See "10.3.6 Saving Settings Data" in the Instruction Manual
1 Press the F1 [Settings] key.
2 Press the F1 [Execute] key.
"Shared Settings in Selection Save" (⇒ p. 71)
Use the CURSOR keys to move the cursor to the
setting item, and select the desired setting contents
with the F1 to F8 key.
Saving Captured Screen Images
See "10.3.10 Optionally Selecting Display Screens & Saving (SAVE Key)" in the Instruction Manual
1 Press the F3 [Screen Image] key.
2 Select the file saving format.
3 Select whether to save the F key display (GUI
area).
4 Press the F1 [Execute] key.
BMP Color
Saves a color bitmap format file.
(Default setting)
Comp BMP
Saves a compressed color bitmap
format file.
BMP Gray
Saves a grayscale bitmap format file.
PNG
Saves a PNG format file
Without
The displayed area is not saved.
With
The displayed area is saved.
(Default setting)
"Shared Settings in Selection Save" (⇒ p. 71)
Use the CURSOR keys to move the cursor to the
setting item, and select the desired setting contents
with the F1 to F8 key.
5.6 Data Loading
74
5.6
Data Loading
Insert storage media
Select the data to load
(File extension: .SET)
Select F1 [Load]
Select F1 [Execute]
Load
Settings Data
Waveform Data
Insert storage media
Select the data to load
(File extension: .MEM, .REC)
Select F1 [Load]
Select F1 [Execute]
Load
Verify before loading
• Is storage media inserted?
• Is the loading source correct?
This instrument can load data saved
in binary format.
Loading Settings & Waveform Data
See "10.4.2 Loading Settings Data" and "10.4.3 Loading Waveform Data" in the Instruction Manual
1 Press the FILE key.
The File screen appears.
2 Use the CURSOR keys to select the storage
media, and press F1 [List].
The files on the selected storage media appear in the
File List.
3 Use the CURSOR keys to select the file to load,
and press the F1 [Load] key.
If [Load] does not appear, press the F8 key until F8
[Page 1/3] is displayed.
File List Operations
• To open a folder: ENTER key
• Return to previous folder: ESC key
A [Confirmation] dialog appears.
4 Press the F1 [Execute] key.
The data is loaded.
To cancel:
Press the F2 [Cancel] key.
Verify that storage media has been inserted.
Folder Tree
File List
Folder tree operations are also available.
To select a file or folder
CURSOR Keys
To move to or display
one deeper level
CURSOR Key
(Opens the level below the di-
rectory with the selected
symbol)
To move to or display
one higher level
CURSOR Key
(Closes levels below the di-
rectory with the
symbol)
Press the SHEET/PAGE
keys to move the cursor.
(⇒ p. 76)
• Automatic Printing
[Auto Print]: Memory Function and FFT Function
[Real Time Print]: Recorder Function
• Manual Printing
[Quick Print]: Prints when you press the PRINT
key
[Selection Print]: Press the PRINT key and
select what to print (Default setting)
Print Settings
• Printout Type (Waveform, Numerical Values and
Screen Link)
• Print Area (Whole Waveform, A-B)
• Display value of horizontal axis (Time Value Dis-
play)
Printer Settings
Internal Printer (Print Destination: [Printer])
• Print Density
• Paper Feed
• Print Quality
External Printer (Printing Destination: [USB])
• Paper Orientation, Margins
• Printing Colors
The shipping and initial default printing method is [Selection Print]. With this setting, printing occurs after
pressing the PRINT key and selecting what you want to print. You can change print settings as occasion
demands from the Print Settings screen. Refer to "Chapter 11 Printing" in the Instruction Manual for
details.
When using the optional printer unit, read the precautions in "3.3 Loading Recording Paper (With a Printer
Module Installed)" (⇒ p. 42).
Print Setting Details
(⇒ p. 79)
When printing waveforms *
• Grid Type
• Channel Markers
• List & Gauge
• Upper and Lower Limits
• Zero-Position Comments
• Counter
• Time Axis Expansion and Compression
• Gauge (when using external printer)
"11.6.2 Printing Waveforms" in the Instruction Manual
When printing numerical data
• Thinned numerical value data
"11.6.3 Printing Numerical Value Displays" in the
Instruction Manual
To print comments and settings data together *
• Print Comments (Analog, Logic)
• Print Title
• Settings Data
"11.6.5 Printing Comments and Setting Data" in the
Instruction Manual
* Some items can be set only when a particular function
is enabled.
Printable Items
(⇒ p. 76)
• Whole Waveform
• A-B Waveform
• Waveform before and after trigger event
• List (Settings List)
• Report
• Calculation results
• Screen image
Depending on the print method or function, some items
cannot be printed.
Printing Destination
• Internal Printer
• External Printer (on USB port, only HP printers
are supported)
"11.6.1 Common Settings" in the Instruction Manual
Description of printout types: (⇒ p. 78)
Printing examples: "11.7 Print Examples" in the
Instruction Manual
"11.5 Making Printer Settings" in the Instruction Manual
6.1 Printing Methods and Contents
76
6.1
Printing Methods and Contents
To print automatically while
measuring
To print immediately upon
pressing the PRINT key
To select and print specified
items after measuring
Auto Print
Specify what you want to print be-
fore pressing the PRINT key. It may
be convenient to print certain data
optionally and to print simultaneous-
ly while analyzing.
(Default setting)
After pressing the PRINT key, select
items to print before printing. This
method is convenient for printing dif-
ferent items each time.
Quick Print
Selection Print
With the Memory Function*, prints au-
tomatically after acquiring measure-
ment data for the specified recording
length.
With the Recorder Function (support-
ed only by the internal printer), prints
in real time while recording.
With the FFT Function, prints auto-
matically after FFT calculation.
In either case, before measuring, se-
lect the desired items to print.
* When the Roll Mode is enabled, waveforms can be printed while being dis-
played. (However, printing is delayed if the timebase is set faster than 500
ms/div.)
Basically, three printing methods are available.
Install recording paper (⇒ p. 42)
After installation and connection,
turn power on.
Set measurement configuration,
etc.
Set the Auto Print setting [On]
Select the print destination
Select items to print
Measure
Print
The following can be printed
simultaneously.
• Waveform Data
• Numerical Calculation Results
Install recording paper (⇒ p. 42)
After installation and connection,
turn power on.
Select [Quick Print] as the PRINT
key action
Select the print destination
Select items to print
Press the PRINT key
Print
Any of the following can be print-
ed:
• Whole Waveform
• A-B Waveform
• Waveform before and after
trigger event
• Report
• List (Settings List)
• Calculation results
• Screen Image
• (Screen Link)
Install recording paper (⇒ p. 42)
After installation and connection,
turn power on.
Select [Selection Print] as the
PRINT key action
Select the print destination
Press the PRINT key
Select items to print
Print
Any of the following can be
printed:
• Whole Waveform
• A-B Waveform
• Waveform before and after
trigger event
• Report
• List (Settings List)
• Calculation results
• Screen Image
Verify before printing
• Is the recording paper loaded
correctly?
• Is the printing destination set
correctly?
• For automatic printing: Is Auto
Print enabled [On]?
If both Auto Print and Auto Save are enabled, saving is performed first.
However, if the Roll Mode is used with the Memory function (default setting:
Auto), printing is performed first.
6.2 Printing While Recording (Auto Print)
77
6.2
Printing While Recording (Auto Print)
Enable Auto Print
Use the same settings to also
print numerical calculation re-
sults. (Memory function only)
1 Press the SUB MENU keys to select the [Print] menu.
2 Press the SHEET/PAGE keys
to select the [Printer] page.
Now use the CURSOR keys to move among the
setting items, and press the F1 to F8 keys to select
the desired setting.
3 Press the F2 [On] key.
Memory function and FFT function: Auto Print
Recorder function: Real-Time Printing
4 Select the printing destination.
(Memory function and FFT function only)
Make other printer settings as occasion demands.
See "11.5 Making Printer Settings" in the Instruction
Manual
Printer
Print using the internal printer.
USB
Print using an external printer. (only HP
printers are supported)
See "11.3 Making Auto Print Settings" in the Instruction Manual
Selecting Items to Print
You can select what infor-
mation prints on waveform
graphs (⇒ p. 79).
1 Press the SHEET/PAGE keys to
select the [Print Items] page.
2 Select the recording format for the printer.
3 Select the area to be printed.
4 Select the horizontal axis value (timebase
display).
Make other changes to waveform, numerical value, com-
ment settings as occasion demands (⇒ p. 79).
Waveform, Numeric, Screen Link (Default setting)
With [Screen Link], the printout corresponds to actual dis-
play appearance.
Whole (Default setting) or A-B
[A-B] prints the data between A/B cursors.
Time
Time from trigger event (units are
fixed) (Default setting)
Mod 60
Time from trigger event (units are mod-
ulo 60)
Scale
Number of divisions from trigger event
Date
Time of waveform acquisition
Samples
Number of samples from trigger event
See "11.6 Setting the Print Content" in the Instruction Manual
Start Measurement: Press the START key.
Data automatically prints after acquisition by the
instrument.
Comments and settings
data can be printed at
the same time (⇒ p.
Now use the CURSOR keys to move among the set-
ting items, and press the F1 to F8 keys to select the
desired setting.
6.3 Manual Printing (PRINT Key)
78
The [Selection Print] method is described here. To use the [Quick Print] method, refer to "11.4 Making
Manual Print (PRINT Key Output) Settings" in the Instruction Manual.
6.3
Manual Printing (PRINT Key)
Select when printing a
screen image.
3 Select the printing destination.
4 Press the F2 [Selection Print] key.
Change items to print as occasion demands on the [Print
Items] page (⇒ p. 77).
Printer
Print using the internal printer.
USB
Print using an external printer. (only HP
printers are supported)
1 Press the SUB MENU keys to select the [Print] menu.
5 Press the PRINT key.
The [Print] dialog appears.
Now use the CURSOR keys to move among the set-
ting items, and press the F1 to F8 keys to select the
desired setting.
2 Press the SHEET/PAGE keys to
select the [Printer] page.
6 Press the F1 to F8 keys to select what you want to print.
The selected items are printed.
*1. Memory function and Recorder function only.
*2. Memory function only.
Whole Wave
Prints all waveform data acquired by the instrument.
A-B Wave *1
Prints the section of the waveform demarcated by the A/B cur-
sors (vertical or trace cursors). When either of the A/B cursors
is off the screen, the section between A-B cursors is printed.
Before selecting: Specify the print area with the A/B knobs.
Starting and ending points can be represented by either cursor.
Trig Wave *1
Prints 10 divisions of the waveform before and after the central
trigger event.
Report
Prints the waveform data of the displayed area on the wave-
form screen, upper and lower limits and analog channel set-
tings.
List
Prints a list of the settings on the Settings screen.
Calc Results *2
Prints numerical calculation results.
Screen Image
Prints a captured image of the displayed screen.
Before selecting: Display the screen to be printed.
Feed
Feeds the paper.
Cancel
Cancels printing
Want to stop printing?
Press the STOP key.
To print numerical data:
On the [Print Items] page,
set the [Printout Type] to
[Numeric]
6.4 To Print a Detailed Settings List
79
6.4
To Print a Detailed Settings List
When printing waveforms
[Waveform Print Items]
1
2
3
4
5
6
7
(Default setting)
Setting Item
Selection
Description
1 Grid Type
Off, Normal, Fine,
Normal (Dark),
Fine (Dark),
Time Axis or
T-Axis (Dark)
Select the type and darkness of the
printed grid.
2
Channel Markers
Off, Ch No. or
Comments
Select whether to print channel numbers
or comments with waveforms. (Analog
channels only)
Marker Position
Inside or Outside
Select the channel marker position.
3 List & Gauge
Off, List, Gauge or
List & Gauge
Select whether to print a list or gauge.
4 Upper/Lower
Limits
Off or On
The upper and lower limits of each chan-
nel can be printed. When scaling is en-
abled, the limit values are in scaled
units.
5 Zero-Position
Comment
Off or On
Select comment printing at the zero po-
sition of each channel. (Analog channels
only)
6 Counter Printing
Off, Date or Name
Select whether to print the number of ac-
quired waveforms or counter name. This
is convenient for distinguishing similar
waveforms.
7 Mag/Comp
Screen Link or
No Screen Link
The time axis of waveforms to be printed
can be expanded or compressed.
(When [No Screen Link] is selected,
screen expansion and compression set-
tings do not affect printouts)
See "11.6.2 Printing Waveforms" in
the Instruction Manual
When printing numerical values
[Numerical Value Print Items]
(Default setting)
Setting Item
Selection
Description
Thinning
Screen Link or
No Screen Link
When data is displayed with numerical
values on the Waveform screen, select
whether the thinning setting is linked to
the screen.
See "11.6.3 Printing Numerical Value
Displays" in the Instruction
Manual
When printing comments
[Comment Printing Settings]
1
2
3
(Default setting)
Setting Item Selection
Description
1 Title
Off, Settings,
Comments,
Set& Com
Select whether to print a title and settings
data (Function, Timebase, Time Axis
Magnification Ratio, Trigger Time).
2 Analog
Off, Settings,
Comments,
Set& Com
Select whether to print comments for each
analog channel, or channel settings (mea-
surement range, vertical axis expansion
or compression, zero position, low-pass
filter, full-scale value of range [or upper
and lower limits when Variable is en-
abled]).
3 Logic
On or Off
Select whether to print comments for each
logic channel.
See "11.6.5 Printing Comments and
Setting Data" in the Instruction
Manual
Inspection and Repair
Refer to "Appendix 1 Error Messages" in the Instruction Manual for error mes-
sages.
Maintenance and
Service
Chapter 7
7.1
Troubleshooting
If damage occurs to the hard disk or MO disk, we cannot restore or analyze the
lost data. We therefore recommend maintaining a backup of any important data.
If damage is suspected, check the "Troubleshooting" section before contacting
your dealer or Hioki representative.
Transporting
• Pack the instrument so that it will not sustain damage during shipping, and
include a description of existing damage. We cannot accept responsibility for
damage incurred during shipping.
• To avoid damage to the instrument, be sure to remove the PC cards before
shipping.
Replaceable Parts
Certain parts require replacement periodically and at the end of their useful life:
(Useful life depends on the operating environment and frequency of use. Opera-
tion cannot be guaranteed beyond the following periods)
The fuse is housed in the power unit of the instrument. If the power does not
turn on, the fuse may be blown. If this occurs, a replacement or repair cannot be
performed by customers. Please contact your dealer or Hioki representative.
Part
Life
Part
Life
Fan Motor
Approx. 4 years
Model 9717 MO Unit
(Drive)
Approx. 5 years
Lithium Battery
Approx. 10 years
Model 9718 HD Unit
(Drive)
Approx. 20,000 hours
Printer
After using 1,000
rolls of the record-
ing paper
LCD
Approx. 74,000 hours
Backlight (to half
brightness)
Approx. 55,000 hours
Electrolytic
capacitors
Approx. 4 years
(The useful life of electrolytic capacitors varies greatly according to the
operating environment. In severe operating environments (40°C
ambient temperature), degradation occurs in about four years, so they
should be replaced periodically.)
Lithium battery
Approx. 10 years
(This instrument contains a built-in lithium battery to back up settings
and the real-time clock. Have the battery replaced if the date and time
are found to lag substantially or if settings are not retained when
power is turned off and back on. Contact your dealer or Hioki
representative.
Disposing of the Instrument
Before returning for repair
This instrument includes a lithium battery to retain settings. Remove this battery
before disposing of the instrument. Also remove the optional Model 9719 Mem-
ory Backup Unit, if installed.
See "Appendix 6 Disposing of the Instrument" in the Instruction Manual
If Power and Operating Keys Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
The
display
does
not
appear when you turn the
power on.
Is the power cord disconnected?
Are connections made correctly?
Verify that the power cord is connected prop-
erly.
"3.4 Connecting the Power Cord" (⇒ p. 45)
Keys do not work.
• Is any key being held down?
• Verify key operation.
• Is the key-lock state active (Key-Lock
message displayed)?
• Deactivate the key-lock state.
(Hold the
CURSOR keys for three
seconds)
• Is remote control from a PC over the
Web enabled?
• Press the SET key to release the key-
lock.
If the Display or Operations Malfunction
Symptom
Check Item, or Cause
Remedy and Reference
A waveform does not
appear when you press
the START key.
• Is the “Pre-Trig Wait” message dis-
played?
• Is the “Trigger Wait” message dis-
played?
When pre-triggering is enabled, triggering is
ignored until the pre-trigger portion of the
waveform has been acquired.
Recording starts when a trigger occurs.
No changes occur in the
displayed waveform.
• Is the clamp sensor or connection
cable connected correctly?
• Is the measurement range set prop-
erly?
• Is the low-pass filter enabled?
Verify that the clamp sensor or connection
cable is connected correctly.
"Chapter 2 Connections" in the Input
Module Guide
Verify the input channel settings.
While measuring with the
memory function, the dis-
played frequency is much
lower than the actual fre-
quency.
Aliasing may be occurring.
Change the timebase to use a faster sam-
pling rate.
"Appendix 4 Supplemental Technical Infor-
mation" in the Instruction Manual
Some channels are unus-
able.
Is the channel enabled for use?
"4.2.1 Selecting Channels to Use" in the
Instruction Manual
The displayed waveform
size does not change
when the input range is
changed.
Is the Variable function enabled?
Turn the Variable function off.
"8.9.4 Setting Arbitrary Waveform Height
and Position on the Vertical (Voltage) Axis
(Variable Function)" in the Instruction
Manual
If the cause is unknown
Try performing a system reset. All settings are returned to their factory defaults.
See "12.3.3 Initializing System Settings (System Reset)" in the Instruction Manual
For other problems, refer to the Appendix or the Index in the Instruction Manual.
Cannot Print, or Printing Malfunctions
Symptom
Check Item, or Cause
Remedy and Reference
Nothing
prints
on
the
paper.
Is the paper reversed (back to front)?
Verify that the recording paper is loaded cor-
rectly.
"3.3 Loading Recording Paper (With a
Printer Module Installed)" (⇒ p. 42)
Printout is too light.
• Is the specified recording paper being
used?
• Is the print density setting correct?
• Is the print head dirty?
Try changing the print density setting.
"11.5.1 Internal Printer Settings" in the
Instruction Manual
Clean the print head.
"Print Head Cleaning" (⇒ p. 85)
Recording traces are too
wide.
The input signal may have a ripple com-
ponent.
Enable the filter in the input module settings.
"3.10.3 Low-Pass Filter (LPF) Settings" in
the Input Module Guide
Recording traces are dou-
bled.
Is the waveform printing density set to
[Light]?
In this case, printed dots are spaced out
in the vertical direction. Therefore, a
slightly varying waveform prints some-
times as one line, and sometimes as two.
Change to waveform printing density setting
to something other than [Light]. (Print Set-
tings Screen)
"11.5.1 Internal Printer Settings" in the
Instruction Manual
Printing does not work.
• Is the printer’s head-raising lever in
the correct (head-lowered) position?
• Is the recording paper loaded cor-
rectly?
• Are the printer output settings appro-
priate?
Verify printer installation/connection and set-
tings, and try printing again.
"3.3 Loading Recording Paper (With a
Printer Module Installed)" (⇒ p. 42),
"Chapter 11 Printing" in the Instruction
Manual
If Saving is Not Possible
Symptom
Check Item, or Cause
Remedy and Reference
If saving is not possible to
PC Card or MO storage
media.
• Is write-protection on the storage media
disabled?
• Is the storage media inserted properly?
• Is the storage media formatted?
• Is the remaining capacity of the storage
media too low?
"5.1 Recording (Storage) Media" (⇒ p.
66)
"5.2 Using PC Cards" (⇒ p. 67)
Cleaning the Instrument and Input Modules
Cleaning the Air Filters
7.2
Cleaning
• To clean the instrument and input modules, wipe it gently with a soft cloth
moistened with water or mild detergent. Never use solvents such as benzene,
alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and
discolor the case.
• Wipe the LCD gently with a soft, dry cloth.
Air filters are installed in the bottom panel of the instrument. Clean them periodi-
cally before severe dust buildup occurs. If a filter becomes clogged, the instru-
ment’s internal cooling is impeded, and damage may result.
Also, be sure to thoroughly dry the filters before reinstalling. Operating with a wet
filter may cause damage.
1 Turn the instrument off.
2 Place the instrument with the front upwards, and
remove the filter covers from the bottom panel.
3 Remove the filters from the filter covers.
4 Use a vacuum cleaner to remove any dust or
grime from the filters.
If dust remains, wash with diluted detergent and
dry completely.
5 Replace the filters in their covers, and reinstall.
Filter Covers
Instrument’s Bottom Panel
Filter
Filter Cover
For instruments installed with filter covers
For instruments installed with metal filters
1 Turn the instrument off.
2 Place the instrument with the front upwards.
3 Use a vacuum cleaner to remove any dust or
grime from the filters.
Print Head Cleaning
Normally, no maintenance is required. However, depending on usage
conditions, dirt and paper dust may accumulate on the thermal head over the
long term, causing light or smeared printing. In this case, clean the head by the
following procedure.
1
Use the CURSOR keys to move the cursor
to the Function menu, and press the F7
[System] key.
The System screen appears.
2
Press the SUB MENU keys to select the
[Init] menu item.
The Init (Initialization) Settings screen appears.
3
Use the CURSOR keys to move the cursor
to the [Printer Check] button, and press the
F1 [Execute] key.
The [Printer Check] dialog appears.
4
Move the cursor to [Cleaning], and press the
F1 key.
The paper should print solid (100%) black.
Check for any smearing.
If sufficient improvement is not obtained
even after cleaning several times, wash
the print head (⇒ p. 86).
Self-Check Cleaning
Function Menu
Before Cleaning
Recording paper is required to clean the print head. Verify that the recording paper is
loaded correctly.
1
2
3
4
1 Moisten the back side of a piece of printing
paper with undiluted alcohol, and insert it into
the printer. If the front side is moistened, it
becomes discolored.
2 Lower the head-raising lever, and wash the
head by sliding the paper back and forth.
Back Side of
Printing Paper
Washing the Print Head
Be careful not to apply too much undiluted alco-
hol.
About the Print Head
To avoid discoloring or deforming the instrument, observe the following precau-
tions:
• Do not use organic solvents such as thinner or benzene
• After washing, allow the printer to dry completely before use.
About the Roller Surface
• White powder such as paper dust may accumulate on the roller surface after
long-term use. A small amount should have no effect on printing, but can be
removed with a commonly available camera blower brush if it causes concern.
• Always use the paper cutter on the printer’s cover to cut the printed paper.
Excessive paper dust can accumulate on the roller if the paper is cut by the
print head.
Appendix 1 Waveform Screen Display & Icons
A1
Appendix
Appendix 1 Waveform Screen Display & Icons
See "2.4 Waveform Screen" of the Instruction Manual
1
2
3
4
5
6
8
7
1 Internal Processing State
Pre-Trig Wait
Before acquiring data
Appears only when pre-trig-
ger recording is enabled.
Trigger Wait
Trigger wait state
Scanner Wait
Scanner preparation state
Storing
Data acquisition in progress
Storing Done
Finished acquiring data
Waveform
Waveform generation in
progress
Calculating (n/m)
Numerical value calculation
in progress
Calculating (Zn)
Waveform calculation in
progress
FFT (n/m)
FFT calculation in progress
Stopped
Operation stopped
Preparing
Preparation in progress
Ready
Idle state
Printing
Printing in progress
Saving
Saving in progress
Auto-Ranging
Automatic range detection in
progress
Complete
Finished automatic range
detection
(File Name)
Name of loaded file
2 Storage Events
(No. of Times)
Number of data acquisitions
3 Trigger Info
Trigger cause, date and time
4 Sub Message
Estimated time to storage completion
(Appears when 10s or longer)
Count to be Averaged(n/m)
Simple Averaging
(Time axis)
Simple Averaging
(Frequency)
Exponential Aver-
aging (Time axis)
Exponential Aver-
aging (Frequency)
Peak hold (Fre-
quency)
5 Setting Info
Auto Save (Waveforms)
Auto Save (Calculations)
Auto Save (Waveforms & Calcula-
tions)
Auto Save (Screen images)
Auto Save (Waveforms & Screen im-
ages)
Auto Save (Calculations & Screen im-
ages)
Auto Save (Waveforms & Calcula-
tions & Screen images)
Auto Print
Auto Print (External Printer)
Vertical Cursor
Horizontal Cursor
Trace Cursor
6 Display Sheet*1
Waveform
FFT
X-Y
Composite
Nyquist
Numerical
Values
FFT+Nyquist
Waveform + X-Y
Composite
Waveform +
FFT
Waveform +
Nyquist
7 Display Information
Block number when measuring with
Memory Division enabled*2
Displayed block number for Memory
Division*2
New
(FFT function) Use newly ac-
quired data for calculations.
MEM
(FFT function) Use pre-exist-
ing data for calculations.
Charging finishes about two hours
after power on.
8 Peripheral Device Info
PC Card
(Blank) No Card
Card Present
GP-IB Card
Internal Storage Media
Internal Hard Drive
Internal MO Drive
(Appears only when an MO
disk is present)
Model 9716 FD Drive
(Blank) No diskette present
Diskette present
LAN
Connection Status
(Blank) Disconnected
Power Supply
(Blank) AC power supply
DC power supply
Internal Printer
(Blank) Printer not installed
Printer Installed
Head-raised error *3
Out-of-Paper error *4
Battery Charge State
(When Model 9719 Memory Backup
Unit is installed)
(Blank) Disconnected
Rapid charging
Rapid charging finished
*3. Check the position the
head-raising lever.
*4. Replace with new recording
paper.
*1. Press the SHEET/PAGE keys to switch
sheets.
*2. Press the SHEET/PAGE keys to switch
blocks.
Appendix 2 Overview of Settings and Functions
A2
Appendix 2 Overview of Settings and Functions
Instrument Settings
About the Screen
The instrument’s LCD provides SVGA (800 × 600) resolution. The wave-
form display area consists of 625 horizontal dots and 500 vertical dots.
The waveform display area is divided into 25 divisions horizontally, and
20 divisions vertically, with each division composed of 25 dots horizon-
tally and vertically.
Each data frame (one division) represents 100 samples horizontally,
and 80 to 1600 data bits vertically. (depending on the input module)
Each displayed data frame (one division) changes in accordance with
expansion and compression of the time and voltage axes.
25 div
20 div
1 div = 100 samples
Setting the Timebase (Horizontal Axis)
See "4.2.2 Setting the Timebase (Horizontal Axis) and Sampling Rate", "8.9 Magnifying and Compressing
Waveforms", and "Appendix 4.3 Measurement Frequency Limit" in the Instruction Manual
The input signal acquisition rate corresponds to time per division on the
horizontal axis.
Determining the timebase: Calculate from the frequency and period.
f [Hz] = 1/t [s] (f: frequency, t: period)
Example: If the measurement frequency is 50 Hz: 50 [Hz] = 1/t [s]
t = 1/50 [s] = 0.02 [s] = 20 [ms]
So to display five cycles on the screen (25 divisions horizontally), 20
[ms] × 5/25 [div] = 4 ms/div
Select 5 ms/div for the timebase, which is the closest available setting
to the calculated value.
To measure phenomena with relatively fast signals such as
instantaneous waveforms, we suggest setting a small value (if the
frequency is 50 Hz, the timebase should be set faster than 5 ms/div).
During and after measurement, waveforms can be expanded and
compressed along the time axis.
1 div
Timebase:
with 1 div = 5 ms
• To expand and compress: Change the
expansion and compression ratio. (⇒ p.
62)
• To expand a specified section (Memory
function only): Use the Zoom function.
(⇒ p. 62)
Timebase:
with 1 div = 10 ms
Timebase and Sampling
See "Appendix 4.1 Sampling" and "Appendix 4.4 Recorder Function Values" in the Instruction Manual
For the Memory Function (Sampling point recording):
The sampling period is 1/100th of the timebase. When the timebase is set to 100 µs/
div, the sampling period is 1 µs. Changing the timebase setting also changes the
sampling period accordingly.
For the Recorder Function (Envelope recording):
One data point is recorded every 1/100th of the timebase as two amplitude values:
the maximum and minimum values measured during the specified sampling period.
With the 10 ms/div timebase, one data point is recorded every 100 µs. When the
sampling period is set to 1 µs, 100 samples are provided in order to record each data
point. The values recorded for each data point are maximum and minimum values
among these 100 samples.
Maximum
Minimum
1 div
1 2....
....100
Sampling Period
Sampling Points
Appendix 2 Overview of Settings and Functions
A3
Recording Length Setting
See "4.2.4 Setting the Recording Length (number of divisions)" in the Instruction Manual
Set the length (number of divisions) to record each time data is acquired. Each division of the recording length
consists of 100 data points. The number of data points of the whole recording length is the specified recording length
(divisions) × 100, + 1.
Example. The number of data points when the specified recording length is 50 divisions: 50 div × 100 data points +
1 = 5001 data points
Measurement Range Setting
See "Chapter 3 Input Channel Settings" in the Input Module Guide
Select the amplitude range for measurement. The setting value is
the number of units per division on the vertical axis.
About Resolution (per vertical division)
Measurement resolution depends on the particular input module.
(with x1 voltage axis magnification)
Input Modules
Resolution
(bits/div)
Screen Full
Scale (bits)
Minimum Resolution
8936 to 8940, 8946,
8947, 8959
80
1600
5 V/80 = 62.5 mV *
8956
100
2000
5 V/100 = 50 mV *
8957, 8960
8958 [Voltage] Mode
1600
32000
5 V/1600 = 3.125 mV*
8958 [Temp] Mode
1000
20000
10°C/DIV = 0.01°C
* When measurement range is 5 V/div
With 1 V/div range
(10 Vp-p input)
With 500 mV/div range
(10 Vp-p input)
0V
1 div
20 div
More about Measurement Range and Resolution:
"Chapter 5 Specifications" in the Input Module Guide
• To expand and compress: Change magnification of the
vertical axis. ("8.9 Magnifying and Compressing Wave-
forms" of the Instruction Manual)
• To shift the position on the vertical axis: Change the
zero position. ("7.1.2 Setting the Waveform Display Posi-
tion (Zero Position)" of the Instruction Manual)
• To display converted units: Use the Scaling function.
("5.4 Converting Input Values (Scaling Function)" of the
Instruction Manual)
• To freely set the display area of the vertical axis: Use
the Variable function. ("8.9 Magnifying and Compressing
Waveforms" of the Instruction Manual)
About Trigger and Pre-Trigger Recording
See "Chapter 6 Trigger Settings" in the Instruction Manual
When recording a specific signal with recording criteria (trigger criteria) set, the
waveform can be recorded only when the trigger event occurs. To view phenomena
prior to the trigger event, set the Pre-Trigger value to the amount of data your want
to record before the event occurs.
Example: When the recording length is set to 1000 div and Pre-Trigger is set to 5%
The pre-trigger recording length (data points) is 1000 div × 5% = 50 div (5000 data
points).
Pre-Trigger
T
Trigger Criteria Met
Appendix 2 Overview of Settings and Functions
A4
Convenience Functions
If the appropriate timebase or measurement range is unknown
Auto-Ranging Function
See "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the
Instruction Manual
Automatically selects the timebase and range settings to display 1 to 2.5
waveform cycles.
To record and search for waveform anomalies:
Trigger Function and Waveform Search Function
See "Chapter 6 Trigger Settings", "8.14 Searching a Waveform" in the
Instruction Manual
Trigger signals can be generally be obtained from the following sources: An-
alog trigger, logic trigger, external trigger, timer trigger and forced (manual)
trigger. A trigger can be applied as a result of one or a combination of these
criteria, except for the Manual trigger.
Also, after measurement, any waveform can be searched and displayed us-
ing the Search function.
T
T
T
T
T
T
To display with actual measurement units or to fix the display level of the waveform
Scaling Function
See "5.4 Converting Input Values (Scaling Function)" in the Instruction
Manual
Input values can be converted and displayed as physical values such as
speed, vibration, displacement and temperature. Two conversion methods
are available: setting by conversion ratio and setting by entering equivalent
values at two points.
Variable Function
See "8.9.4 Setting Arbitrary Waveform Height and Position on the Vertical
(Voltage) Axis (Variable Function)" in the Instruction Manual
The position and size of the displayed waveform can be freely set.
By entering the upper and lower limits of the waveform, it can be set to
display with full-screen amplitude.
The Scaling and Variable functions can be used in combination. The full span
of output from a sensor can be displayed.
mV
A
0%
100%
Appendix 2 Overview of Settings and Functions
A5
To view waveform data as numerical values/ To change the waveform display format
Numerical Values Display
See "8.13 Viewing Waveform Data as Numerical Values" of the Instruction
Manual
Split-Screen Display
See "7.2.4 Splitting the Display Screen (Split-Screen)" of the Instruction
Manual
Scroll Direction
See "7.2.5 Setting Waveform Scrolling Orientation" of the Instruction
Manual
The display format of measurement data can be freely selected.
0 ms
5 ms
10 ms
15 ms
20 ms
1-1
1-1
1-1
1-1
1-1
0.3V
71.5V
65.3V
-71V
0.2V
To view a waveform expanded or compressed, or to see the whole waveform
Zoom Display (Memory function only)
See "8.9 Magnifying and Compressing Waveforms" in the Instruction Manual
A normal waveform and an expanded section of it can be displayed together
on the screen.
Whole Waveform Display
All waveform fluctuations can be observed.
Zoom
Magnification
To determine the measurement value of a waveform, or to specify the waveform
range
Cursor Measurements
See "8.8 Cursor Values" in the Instruction Manual
Use the A/B cursors to display values at the cursor positions on a waveform,
and the time and number of cycles from the trigger position.
A specified data range can also be saved and printed.
A
B
A
B
A
B
Vertical
Cursors
Horizontal
Cursors
Trace
Cursors
To display a combination of measurement data
Sheet Display
See "7.2.1 Assigning Display Data to Sheets" in the Instruction Manual
Up to 32 channels of data can be assigned per screen, and freely displayed
on multiple sheets.
X-Y Composite Waveforms
See "7.4 Composite Waveforms (X-Y Waveforms)" in the Instruction Manual
Specify the channels for X and Y axes to form composite waveforms.
Sheet3
Sheet2
Sheet1
Appendix 2 Overview of Settings and Functions
A6
To compare with previous waveforms
Memory Division
See "4.3.3 Dividing Memory" of the Instruction Manual
By dividing the memory space into multiple blocks, waveform data can be re-
corded in any block, or overlaid with any other waveforms.
Overlay
See "4.3.2 Overlaying Waveforms" of the Instruction Manual
Displayed waveforms are retained on-screen and overlaid with new wave-
forms. Use this to compare new waveforms with those recorded immediately
before.
1 2 3 4 5 6
To apply calculations to measurement data
Numerical Calculations
See "Chapter 1 Numerical Calculation Functions" in the Analysis
Supplement
Nineteen types of numerical calculations are available, including RMS values,
maximum values and averages, and up to 16 calculations can be performed at
the same time. Waveforms and calculation results can be displayed separately
on one screen.
Waveform Calculations
See "Chapter 2 Waveform Calculation Functions" in the Analysis
Supplement
The waveform calculation functions allow display of the waveform calculated
by a previously specified formula.
FFT Calculations
See Analysis Supplement
Calculation
Results
Calculation
Results
Calculation
Results
CH(1,1)∗CH(1,2)
CH1,1
CH1,2
Time
Frequency
To perform remote control and data collection/ controlling the instrument through
the interface
LAN (Ethernet) Communications (ftp/Web server)
See "Chapter 13 Communications Settings" in the Instruction Manual
Use 100Base-TX compatible cable to connect the instrument to a network for
control by a PC.
GP-IB Interface
See Communications Operating Manual on the supplied applications disc
With the Model 9558 GP-IB Card installed, the instrument can be
controlled by a computer.
GP-IB card
Appendix 3 Common Questions
A7
Read this when you want to perform a certain task, or when the procedure for a
particular operation is unknown.
Refer to "7.1 Troubleshooting" (⇒ p. 81) if the display or operation seems
abnormal, or if damage is suspected.
Appendix 3 Common Questions
To Perform This Measurement
In this case
Description
Reference
Perform simultaneous mea-
surements with different sam-
pling rates
Set different sampling rates on Timebase 1 and Time-
base 2 from the Status screen of the Memory function,
where Timebase 1 or Timebase 2 can be selected for
each channel.
"4.2.3 Setting Different Sam-
pling Rates" in the Instruction
Manual
To send a warning when an
anomaly occurs
When a trigger is applied, a signal can be output from
the External I/O terminals.
"14.2.2 Trigger Output (TRIG
OUT/CAL)" in the Instruction
Manual
To use multiple recorders
Recording with multiple instruments can be performed
using the external control connectors.
"14.2.1 External Trigger Input
(EXT TRIG)" and "14.2.2 Trig-
ger Output (TRIG OUT/CAL)" in
the Instruction Manual
What happens to measure-
ment data when power to the
instrument is lost during mea-
surement?
Measurement data acquired prior to the power interrup-
tion is lost.
When [Auto-Resume] is enabled from the Environment
Settings Screen of the System Screen, recording
resumes automatically when power is restored. How-
ever, this action differs according to trigger settings.
(Default setting: [Off])
"12.2.2 Using the Auto-Resume
Function (Resume After Power
Restoration)" in the Instruction
Manual
To
measure
shock
when
dropping a test object, and to
search for the vibration reso-
nance point.
Use an acceleration sensor with the Model 8947 Charge
Unit to measure vibration and shock.
To record actual effective
(true RMS) values of mains
voltage
True RMS values can be recorded using the Model 8959
DC/RMS Unit.
To record anomalous waveforms, enable level triggering
to acquire waveforms when the RMS voltage drops.
Example. Trigger: Level, Level: 90 V, Slope: ↓ (Falling)
To monitor 120 V (50 or 60
Hz) power for a drop to 80%
or lower
Using a voltage measuring input module, enable trigger-
ing upon voltage sag.
To
observe
instantaneous
power dropouts on 120 V (50
or 60 Hz) commercial mains,
and the internal voltages of
affected devices.
Measurements can be made by the combination of a
voltage measuring input module and a clamp with con-
nection cable. Changes in commercial power, changes
in internal voltage of devices and changes in current flow
to devices (using a clamp-on sensor) at the instant of
power dropout are recorded on separate channels, and
analyzed. By splitting the screen, power and device-volt-
age waveforms can be displayed in separate traces.
Split-Screen Display:
"7.2.1 Assigning Display Data
to Sheets" in the Instruction
Manual
To observe surge noise on
240 V (60 Hz) AC mains.
AC mains and surge waveforms can be measured simul-
taneously. Surge noise can be detected by the slope trig-
ger. Acquired waveforms can be analyzed by expansion
and zooming.
Appendix 3 Common Questions
A8
To observe the voltage wave-
form on 600 V power lines
Use a voltage measuring input module and the Model
9322 Differential Probe or Model 9665 (10:1) or 9666
(100:1) attenuating probes.
"2.4 Connecting a Differential
Probe", "2.5 Connecting Atten-
uating Probes", and "3.10.15
Probe Attenuation Selection" in
the Input Module Guide
To perform frequency analy-
sis
Measure using the FFT function. We recommend using
an input module equipped with an anti-aliasing filter func-
tion (Models 8938, 8947, 8957 and 8960).
"Chapter 3 FFT Function" in the
Analysis Supplement
To record data without gaps,
or over a long term
Measure using the Real-Time
Saving function.
With the Memory and Recorder functions, data is saved
after each recording length, causing measurement to be
interrupted. With the Real-Time Saving function, data
can be stored continuously without gaps.
"Chapter 9 Measuring with
Real-Time Saving" in the In-
struction Manual
In this case
Description
Reference
Related Settings
In this case
Description
Reference
What is the basic setting pro-
cedure in order to view a
waveform?
To display a waveform, set the following: (1) timebase
(horizontal), (2) voltage axis range and zero position
(vertical), and (3) trigger [Off].
To use automatic setting for
measuring when the time-
base and the measurement
range are unknown
Press the FUNCTION MODE key to select the FN mode
on the Waveform screen (DISP key), and select F4
(Auto Setup).
The proper range is selected automatically, and record-
ing starts. Make other settings as occasion demands.
"3.3.5 Automatic Range Setting
(Auto-Ranging Function)" in the
Instruction Manual
To set up simply when the
measurement configuration is
already determined
Measurement configuration settings can be saved to
storage media as an auto-setup file (STARTUP.SET), so
the settings can be loaded automatically at power-on.
(Auto-Setup Function)
Also, the settings data can be save to the instrument
and it can be automatically loaded.
"10.3.6 Saving Settings Data",
"10.5 Saving & Loading Auto
Settings File (Auto Setup Func-
tion)" in the Instruction Manual
What happens to measure-
ment configuration settings
when power is turned off?
When power is turned off, measurement configuration
settings are preserved within the instrument. When
power is turned back on, those measurement configura-
tion settings are recalled.
"3.6 Turning the Power On and
Off" (⇒ p. 46) in this manual
What is the longest possible
recording length with the fast-
est timebase setting?
Recording time is determined by the [Shot]. The fastest
available timebase setting is [5 µs/div]. The recording
length can be increased by optional internal memory
expansion and by limiting which channels are used, so
the maximum time cannot be definitively specified.
A recording length of 10,000 divisions at 5 µs/div is
equivalent to 0.05 seconds recording time.
"4.2.4 Setting the Recording
Length (number of divisions)" in
the Instruction Manual
To observe phenomena such
as waveform anomalies prior
to a trigger event
From the Trigger Settings screen, enable Pre-Trigger
recording. (Set the position of the trigger point as a per-
centage of the recording length, where 0% represents
the recording start point and 100% represents the
recording end point)
"6.5 Pre-Trigger Settings" in the
Instruction Manual
If “Storage Off” appears
Verify that an input module is installed for the selected
channel.
If installed: On the Status Settings screen, verify that the
channel to be used for measurement is enabled [On].
If not installed: Select an installed module.
"4.2.1 Selecting Channels to
Use" in the Instruction Manual
Appendix 3 Common Questions
A9
How long will recording con-
tinue when the recording
length is set to [Cont] with the
Recorder function?
When trigger criteria are met, recording starts and con-
tinues until you press the STOP key. Up to 5,000 divi-
sions of data (with Model 9715 Memory Board installed)
from measurements prior to the stopping of recording is
retained in internal memory.
When Auto Print or Auto Save is enabled, the data ac-
quired by the instrument is printed or saved on each such
occasion.
About memory capacity and re-
cording length:
"4.2.4 Setting the Recording
Length (number of divisions)";
"Setting Continuous Recording
(Cont)" in the Instruction Manual
What
is
the
difference
between the Memory and
Recorder functions?
Which function is selected determines how the timebase
is set. The sampling method for each function is also dif-
ferent.
Because the purpose of the Memory function is acquisi-
tion of instantaneous waveforms, if you set a slow time-
base for long-term recording, the sampling period may
be unsatisfactorily lengthened. On the other hand, the
Recorder function samples and records maximum and
minimum values at the specified sampling rate regard-
less of the timebase setting. As a result, however, only
slow timebase settings are available.
"Function Selection" (⇒ p. 29)
in this manual,
"4.1 Selecting the Function" in
the Instruction Manual
In this case
Description
Reference
Modifying the Screen Display, and During Measurement
In this case
Description
Reference
To remove a Settings dialog
from the display
Move the cursor to the dialog, and press the ESC key.
If channel waveforms are
overlapping and hard to see
Channel waveforms can be displayed separately by
shifting the zero position of each waveform, or by sepa-
rating the waveform traces on the Sheet Settings
screen.
"7.2.4 Splitting the Display
Screen (Split-Screen)", "7.2.6
Assigning Display Channels to
Graphs (Analog Channels)",
and "7.1.2 Setting the Wave-
form Display Position (Zero Po-
sition)" in the Instruction Manual
The
grid
lines
(graticule)
causes waveforms to be hard
to see,
or you want to display the grid
To remove the grid, set the [Grid Type] on the Environ-
ment Settings screen of the System screen to [Off].
The grid is displayed when [Dotted Line] or [Solid Line]
is selected.
(Default setting: [Dotted Line])
"12.1.1 Selecting the Grid Type"
in the Instruction Manual
Waveform traces are too thick
Waveform traces appear thick when noise is present on
the signal. Enable the low-pass filter to suppress the
noise.
"3.10.3 Low-Pass Filter (LPF)
Settings" in the Input Module
Guide
Logic waveforms are not dis-
played
On the [Logic] page of the Sheet Settings and Channel
Settings screens, set the channels to be displayed to
[On].
If “Storage Off” is displayed:
Set the logic channel to [On] on the [Use Ch] page of the
Status Settings screen.
"4.2.1 Selecting Channels to
Use" and "7.3 Displaying Logic
Waveforms" in the Instruction
Manual
A waveform is continuously
acquired even when trigger
criteria have been set
Is the Trigger Mode set to [Auto]? (Continuously
acquires waveforms)
Select [Single] to record only once when trigger criteria
are satisfied, or select [Repeat] to record repeatedly.
"6.3 Setting the Trigger Mode"
in the Instruction Manual
Cursor values and waveforms
are overlapped and hard to
see
Press the DISP key to display waveforms and numerical
values separately.
"8.8 Cursor Values" in the
Instruction Manual
Appendix 3 Common Questions
A10
“Trigger Wait” is displayed
even though triggering is not
enabled
Is the time trigger enabled [On] or another trigger source
enabled?
Verify settings on the Trigger Settings screen.
“Storing” appears, but no
waveform is displayed
Is the Roll Mode function (on the Status Settings screen)
of the Memory function disabled (set to [Off])?
If set to [On] or [Auto], waveforms are displayed during
acquisition.
"4.3.1 Displaying Waveforms
During Recording (Roll Mode)"
in the Instruction Manual
When “Trigger Wait” is dis-
played and no waveform
appears
No waveform is displayed unless the trigger criteria are
satisfied.
To verify the current input waveform, change the trigger
mode of the Waveform screen to [Auto], and the wave-
form will appear.
However, when the trigger mode is [Auto], if no trigger
occurs within one second after pressing the START key,
recording starts anyway. (No trigger mark is applied)
"6.3 Setting the Trigger Mode"
in the Instruction Manual
To hide unwanted channel
waveforms
From the Waveform screen, press the FUNCTION
MODE key to change the FN mode, and press F5
[Channel Set]. (The Channel Settings dialog appears)
Move the cursor to [Col](color) for the channel to be hid-
den, and select F1 [Off].
"5.7 Setting Input Channels
from the Waveform Screen" in
the Instruction Manual
In this case
Description
Reference
Saving and Loading
In this case
Description
Reference
If you intended to save over a
long period using Auto Save,
and only saved a short period
New directories are created automatically when Auto
Save is [On], the [Normal Save] method is selected, and
Directory Creation is [On]. (Up to 5,000 files can be
saved in each directory)
If these settings are [Off], saving stops after writing the
maximum number of files that one directory can hold.
"10.3.4 Setting Auto Save" in
the Instruction Manual
Auto Save seems to be
enabled, but data is not saved
to the storage media
• Verify that write-protection on the storage media is
disabled. If write-protection is enabled, disable it.
• Verify that Auto Save is [On] before starting measure-
ment.
• When stopping measurement manually, note the dis-
played messages, and do not interrupt auto-saving
while writing to the storage media.
"10.3.4 Setting Auto Save" in
the Instruction Manual
Only short waveforms are
saved
Is the save area set to [A-B]?
To save the whole waveform, select [Whole].
"10.3.5 Setting Manual Save
(SAVE Key Output)" in the
Instruction Manual
When
storage
media
becomes full and you want to
continue saving to another
storage media
Specify as Media1 the storage media to which you want
to save first. For [Save on Error], set [Save in 2] and
specify the storage media on which to continue saving.
"10.3.4 Setting Auto Save" in
the Instruction Manual
To save to a shared folder on
a PC, or to load data
When a folder on a PC is registered for sharing over the
LAN, it can be accessed from this instrument.
"10.1.6 Using a Network
Shared Folder" and "Chapter 13
Communications Settings" in
the Instruction Manual