Hioki 8861 Quick Start Manual

Text of the manual

QUICK START MANUAL

1 8860 8861 Read This Manual First It describes preparations for use, basic operating procedures and usage methods.

Confirming Package Contents....................................................2

Safety Information ......................................................................3 Operating Precautions................................................................5 Chapter 1 Overview__________________________________ 11 1.1 Product Overview and Features .......................................11 1.2 Measurement Workflow ....................................................12 Chapter 2 Names and Functions of Parts ________________ 19 2.1 Panel Names and Functions .............................................19 2.2 Screen Types and Contents .............................................23 2.3 Basic Operations ...............................................................32 2.3.1 To Change Settings ................................................................32 2.3.2 Entering Text and Numbers ....................................................33 2.3.3 Viewing Recorded Data ..........................................................35 Chapter 3 Measurement Preparations___________________ 37 3.1 Installation & Connection Procedures ...............................37 3.2 Connecting Cables and Probes ........................................38 3.3 Loading Recording Paper (With a Printer Module Installed)........................................42 3.4 Connecting the Power Cord ..............................................45 3.5 Grounding the Instrument’s Functional Earth ...................45 3.6 Turning the Power On and Off ..........................................46 3.7 Setting the Clock ...............................................................48 Chapter 4 Basic Operations ___________________________ 49 4.1 Pre-Measurement Inspection ............................................49 4.2 Operation Workflow ..........................................................50

Recording Simple Waveforms ..........................................51

4.4 Actual Measurement and Analysis ...................................54 Chapter 5 Saving & Loading Data ______________________65 5.1 Recording (Storage) Media ...............................................66 5.2 Using PC Cards ................................................................67 5.3 Data that can be Saved & Loaded ....................................68 5.4 Saving Procedures and Contents .....................................70 5.5 Data Saving ......................................................................71 5.6 Data Loading ....................................................................74 Chapter 6 Printing ___________________________________75 6.1 Printing Methods and Contents ........................................76 6.2 Printing While Recording (Auto Print) ...............................77 6.3 Manual Printing (PRINT Key) ...........................................78 6.4 To Print a Detailed Settings List .......................................79 Chapter 7 Maintenance and Service ____________________81 7.1 Troubleshooting ................................................................81 7.2 Cleaning ............................................................................84 Appendix________________________________ A 1 Appendix 1 Waveform Screen Display & Icons..............................A 1 Appendix 2 Overview of Settings and Functions............................A 2 Appendix 3 Common Questions.....................................................A 7 Index ________________________________Index 1

Thank you for purchasing the HIOKI “Model 8860/ 8861 Memory HiCorder."

To obtain maximum performance from the instrument, please read this manual carefully, and keep it handy for future reference. In this document, the “instrument” means the Model 8860 or 8861 Memory HiCorder. The following documents are provided with this instrument. Refer to them as appropriate for your application. Registered trademarks Windows is a registered trademark of Microsoft Corporation in the United States and/or other countries. If you encounter problems during operation Introduction Document Description 1 Quick Start Manual (This document) Read this first. It describes preparations for use, basic operating pro- cedures and usage methods. 2 Input Module Guide To connect input modules and measurement cables, and when making input channel set- tings; this Guide describes the optional input modules, relat- ed cable connection procedures, and their settings and specifications. 3 Instruction Manual To obtain setting details; this Manual describes details of the functions and op- erations of the instrument, and its specifications. 4 Analysis Supplement The supplement describes usage of the cal- culation functions to analyze measurement data. If you want to know a mea- surement workflow "1.2 Measurement Workflow" (⇒ p. 12) For operation overviews: "Chapter 3 Operation Overview" in the Instruction Manual For screen configurations "2.2 Screen Types and Contents" (⇒ p. 23) "2.2 Screen Organization" in the Instruction Manual For input module connec- tions and channel setting procedures Input Module Guide When an error is displayed "Appendix 1 Error Messages" in the Instruction Manual When measurement prob- lems occur, or for general measurement questions "Appendix 3 Common Questions" (⇒ p. A7) "Before returning for repair" (⇒ p. 82)

Confirming Package Contents

2 Confirming Package Contents � 1 Quick Start Manual � 2 Input Module Guide � 3 Instruction Manual � 4 Analysis Supplement ........................................One each � Application CD (Communications Manual)........... 1 � This instrument: Model 8860/ 8861 Memory HiCorder (8860: 4-module, 8861: 8-module) Confirm that these contents are provided. Accessories � Power Cord.............................................1 � Input Cable Labels...................................1 (If a printer module is installed) For Model 8995 A4 Printer Unit: � Model 9231 Recording Paper...........1 roll � Paper Roll Holders ...........................1 pair For Model 8995-01 A6 Printer Unit: � Model 9234 Recording Paper............1 roll � Paper Roll Axle........................................1 Other options specified when ordered For information about options: See "Appendix 5 Options" in the Instruction Manual When you receive the instrument, inspect it carefully to ensure that no damage occurred during shipping. In particular, check the accessories, panel switches, and connectors. If damage is evident, or if it fails to operate according to the specifications, contact your dealer or Hioki representative.

This manual contains information and warnings essential for safe operation of

the instrument and for maintaining it in safe operating condition. Before using it, be sure to carefully read the following safety precautions. The following symbols in this manual indicate the relative importance of cautions and warnings. Safety Information This instrument is designed to comply with IEC 61010 Safety Standards, and has been thoroughly tested for safety prior to shipment. However, mis- handling during use could result in injury or death, as well as damage to the instrument. Be certain that you understand the instructions and pre- cautions in the manual before use. We disclaim any responsibility for acci- dents or injuries not resulting directly from product defects. Safety Symbols In the manual, the symbol indicates particularly important information that the user should read before using the instrument. The symbol printed on the instrument indicates that the user should refer to a corresponding topic in the manual (marked with the symbol) before using the relevant function. Indicates a burn hazard if touched directly. Indicates DC (Direct Current). Indicates AC (Alternating Current). Indicates both DC (Direct Current) and AC (Alternating Current). Indicates a grounding terminal. Indicates the ON side of the power switch. Indicates the OFF side of the power switch. Indicates that incorrect operation presents an extreme hazard that could result in serious injury or death to the user. Indicates that incorrect operation presents a significant hazard that could result in serious injury or death to the user. Indicates that incorrect operation presents a possibility of injury to the user or damage to the instrument. Indicates advisory items related to performance or correct operation of the instrument.

Symbols in text

Indicates the prohibited action. (⇒ p. ) Indicates the location of reference informa- tion. Indicates quick references for operation and remedies for troubleshooting. * Indicates that descriptive information is provided below. [ ] Screen labels such as menu items, page titles, setting items, dialog titles and but- tons are indicated by square brackets [ ]. CURSOR (Bold characters) Bold characters within the text indicate operating key labels. Unless otherwise specified, “Windows” represents Win- dows 95, 98, Me, Widows NT4.0, Windows 2000, or Win- dows XP. Mouse operation terminology Click: Press and quickly release the left button of the mouse. Right-click: Press and quickly release the right button of the mouse. Double click: Quickly click the left button of the mouse twice. Drag: While holding down the left button of the mouse, move the mouse and then release the left button to deposit the chosen item in the desired position. Activate: Click on a window on the screen to activate that window. We define measurement tolerances in terms of f.s. (full scale) values, with the following meanings: f.s. (maximum display value or scale length) The maximum displayable value or scale length. This is usually the name of the currently selected range. In this instrument, the maximum displayable value is the range (V/div) times the number of divisions (20) on the vertical axis. Example: For the 1 V/div range, f.s. = 20 V Accuracy This instrument complies with CAT II safety requirements. This instrument’s input modules comply with CAT I or CAT II safety requirements. To ensure safe operation of measurement instruments, IEC 61010 establishes safety standards for vari- ous electrical environments, categorized as CAT I to CAT IV, and called measurement categories. These are defined as follows. CAT I: Secondary electrical circuits connected to an AC electrical outlet through a transformer or similar device. CAT II: Primary electrical circuits in equipment connected to an AC electrical outlet by a power cord (portable tools, household appliances, etc.) CAT III:Primary electrical circuits of heavy equipment (fixed installations) connected directly to the distribution panel, and feeders from the distribution panel to outlets. CAT IV:The circuit from the service drop to the service entrance, and to the power meter and primary over- current protection device (distribution panel). Higher-numbered categories correspond to electrical environments with greater momentary energy. So a measurement device designed for CAT III environments can endure greater momentary energy than a device designed for CAT II. Using a measurement instrument in an environment designated with a higher-numbered category than that for which the instrument is rated could result in a severe accident, and must be carefully avoided. Never use a CAT I measuring instrument in CAT II, III, or IV environments. The measurement categories comply with the Overvoltage Categories of the IEC60664 Standards. Measurement categories (Overvoltage categories) The applicable measurement category is determined by the input module being used. Refer to “Applicable Standards” in "Chapter 5 Specifications" in the Input Module Guide for the applicable measurement categories.

Follow these precautions to ensure safe operation and to obtain the full benefits

of the various functions. • Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative. • Before using the instrument, make sure that the insulation on the probes and connection cords is undamaged and that no bare conductors are improperly exposed. Using the instrument in such conditions could cause an electric shock, so contact your dealer or Hioki representative for replacements. Operating temperature and humidity: 0 to 40°C at 20 to 80% RH (non-conden- sating) However, the operating temperature range may depend on the options used. Temperature and humidity range for guaranteed accuracy: 23±5°C, 20 to 80%RH Operating Precautions Before Use Instrument Installation Operating Conditions Operating Temperature While changing the battery in the Model 9719 Memory Back- up Unit 10 to 40°C While using the Model 8995 A4 Printer Unit, 8995-01 A6 Printer Unit, 9716 FD Drive or 9718 HD Unit 5 to 40°C Model 9717 MO Unit 5 to 35°C Avoid the following locations that could cause an accident or damage to the instrument. Exposed to direct sun- light Exposed to high tem- perature In the presence of cor- rosive or explosive gases Exposed to liquids Exposed to high hu- midity or condensa- tion Exposed to strong electromagnetic fields Near electromagnetic radiators Exposed to high lev- els of particulate dust Subject to vibration

• The instrument should be operated only with the bottom or rear side down-

wards. • Vents must not be obstructed. At least 50 mm At least 50 mm spacing on all sides When using the optional printer, provide enough space for ejected recording paper or use the Model 220H Paper Winder (optional recording paper auto winder). Handling the Instrument • Do not allow the instrument to get wet, and do not take measurements with wet hands. This may cause an electric shock. • Never modify the instrument. Only Hioki service engineers should disas- semble or repair the instrument. Failure to observe these precautions may result in fire, electric shock, or injury. • To avoid damage to the instrument, protect it from physical shock when trans- porting and handling. Be especially careful to avoid physical shock from drop- ping. • Use the handle to carry the instrument. See "Chapter 2 Names and Functions of Parts" (⇒ p. 19) • When the Model 8958 16-Ch Scanner Unit is installed, do not lift up on the connector section of the unit. The connector section could be damaged. Handling the Cords and Probes • Avoid stepping on or pinching cables, which could damage the cable insula- tion. • To avoid breaking the cables and probes, do not bend or pull them. • To avoid damaging the power cord, grasp the plug, not the cord, when unplug- ging it from the power outlet. Use only the specified connection cables. Using a non-specified cable may result in incorrect measurements due to poor connection or other reasons.

For additional information about input modules, their connections and

measurement precautions, refer to the Input Module Guide. Before Connecting Before turning power on • Before turning the instrument on, make sure the supply voltage matches that indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical haz- ard. • To avoid electrical accidents and to maintain the safety specifications of this instrument, connect the power cord only to a 3-contact (two-conduc- tor + ground) outlet. Installing and removing input modules • To avoid electric shock accident, before removing or replacing an input module, confirm that the instrument is turned off and that the connection cords are disconnected. • The mounting screws must be firmly tightened or the input module may not perform to specifications, or may even fail. • To avoid the danger of electric shock, never operate the instrument with an input module removed. To use the instrument after removing an input module, install a blank panel over the opening of the removed module. Using an Optional Printer When the instrument is not to be used for a long time, or when transport- ing To avoid straining some parts of the printer, and to prevent dirt adhering to the print head, set the print head to its lowered position (so that the print head rests against the print roller). After storing the printer without use for a long time, inspect the following before use: 1. Move the head raising lever up and down. 2. Perform test printing (printer check) three or four times. When printing • Avoid slow printing as much as possible in hot or humid environments. Other- wise, printer life may be severely shortened. • Please use only the specified recording paper. Using non-specified paper may not only result in faulty printing, but printing may become impossible. • If the recording paper is skewed on the roller, paper jams may occur. If the recording paper is skewed on the roller, paper jams may result. • Printing is not possible when the recording paper is loaded upside-down. Refer to "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42) for printing precautions.

Input and Measurement Precautions

The maximum input voltage and maximum rated voltage to ground (between the grounds of the input terminals and the instrument, and between inputs of other analog modules) of each input module and exter- nal input terminals are shown below. To avoid risk of electric shock and damage to the instrument, be careful to not exceed these ratings. The max- imum rated voltage to ground is the same regardless of whether an input attenuator or similar device is used. Input Terminal (Input Module) Maximum Input Voltage*1 Maximum Rated Voltage to Ground*2 8936 Analog Unit 400 V DC max 370 V AC/DC 8956 Analog Unit 400 V DC max 300 V AC/DC 8946 4-Ch Analog Unit 30 V rms/ 60 V DC 30 V rms / 60 V DC 8938 FFT Analog Unit 400 V DC max 370 V AC/DC 8957 High Resolution Unit 400 V DC max 300 V AC/DC 8959 DC/RMS Unit 400 V DC max 370 V AC/DC 8937 Voltage/Temp Unit 30 V rms/ 60 V DC 30 V rms / 60 V DC 8939 Strain Unit 10 V DC max 30 V rms / 60 V DC 8960 Strain Unit 10 V DC max 33 V rms / 70 V DC 8940 F/V Unit (BNC jacks and sensor re- ceptacles) 30 V rms / 60 V DC (BNC jack) 30 V rms / 60 V DC (Sensor receptacle terminals) Uninsulated 8947 Charge Unit*3 (BNC jack) 30 V rms / 60 V DC (BNC jack) 30 V rms / 60 V DC 8958 16-Ch Scanner Unit 40 V DC max 33 V rms / 70 V DC 9322 Differential Probe (CATII) 2000 V DC, 1000 V AC (CAT III) 600 V AC/DC With grabber clips 1500 V AC/DC (CAT II) 600 V AC/DC (CAT III) With alligator clips 1000 V AC/DC (CAT II) 600 V AC/DC (CAT III) I/O Terminals Maximum input voltage Maximum rated voltage to ground EXT.SMPL EXT.TRIG PRINT/EXT.IN3 STOP/EXT.IN2 START/EXT.IN1 -2 to 7 V DC Uninsulated TRIG OUT/CAL NG/EXT OUT2 GO/EXT OUT1 -20 to 30 V DC, 500 mA max, 200 mW max This Instrument Input Module GND *1 *2 H L *2 “xxx V AC/DC” indicates an RMS value, and “xxx V DC max” indi- cates an instantaneous value. (Input modules other than the Model 8947 Charge Unit) This Instrument Input Module GND *3 *2 H L *2 *1 H L BNC Miniature Connector Common (8947 Charge Unit) *1.Maximum input voltage *2.Maximum rated voltage to ground *3.Maximum applied electric charge 500 pC (high-sensitivity side, 6 ranges), 50,000 pC (low-sensitivity side, 6 ranges)

Use the original packing materials when transporting the instrument, if possible.

The optional Model 9723 and 9724 Carrying Cases are available for shipping the Models 8860 and 8861, respectively. To avoid accidents in the event of a short circuit, when measuring com- mercial mains voltage using the Model 8936, 8938, 8956, 8957 or 8959 input modules, connection should be made only at the secondary side of a breaker. Never make connections at the primary side of a breaker, as unre- stricted current flow could cause a serious accident if a short circuit occurs. When measuring AC mains voltage, for example, using a power transformer, be sure to connect the ground terminal of the power transformer to ground. Using a power transformer (PT) This Instrument Input Module GND H L Input Power transformer with ground terminal Ground (Earth terminal of power source, or equivalent) PT Power transformer ground terminal Power transformer without ground terminal GND H L Input Ground (Earth terminal of power source, or equivalent) PT This Instrument Input Module Correct measurement may be impossible in the presence of strong magnetic fields, such as near transformers and high-current conductors, or in the pres- ence of strong electromagnetic fields such as near radio transmitters. Shipping precautions To avoid damage, observe the following when shipping the instrument: • Remove the PC cards before shipping. • If the optional printer module is installed, remove the paper. If the paper is left in the instrument, the paper-handling components may be damaged by vibra- tion.

1.1 Product Overview and Features

11 Overview Chapter 1 1.1 Product Overview and Features Recording various analog signals Measurements are provided in combina- tion with optional input modules, cables and clamp probes depending on the application. Voltage, Current, Tempera- ture, Acceleration, Vibra- tion, Frequency, Stress, etc. Automatic and manual saving, loading and file management of measurement data (⇒ p. 65) Data saving and loading are available with optional PC Cards. Also, using optional proprietary drives, saving and loading can be performed on floppy diskettes, MO or hard disk. Data on storage media can be managed with this instrument. "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual Memory capacity can be expanded with optional mem- ory boards. (This option must be specified when order- ing.) 8860: 32 Megawords to 1 Gigaword, Model 8861: 64 Megawords to 2 Gigawords Automatic and manual printing of measurement data (⇒ p. 75) Measurement data can be printed with the optional A4 or A6 internal printer and special recording pa- per. Printing is also available on an external USB printer. "Chapter 11 Printing" in the Instruction Manual Recording logic signals Measures using optional logic probes. Mixed analog and logic recording is possible. Remote control and data collection The instrument can be connected to a PC for measure- ment data analysis and remote control. "Chapter 13 Communications Settings" in the Instruction Manual External Control The instrument can be controlled by external signals. "Chapter 14 External Control" in the Instruction Manual Easy entry of text and numerical values Screen operations and character input can be per- formed by connecting a commonly available mouse or keyboard. "3.3 Common Operations" in the Instruction Manual Input Module Guide X-Y Display Zoom Split-Screen Display Numerical Calculations Gauge and Level Monitor Displays In this case • Viewing waveforms before and after an anomaly. • Capturing the waveform of an intermittent anomaly. • Recording multiple signal lines over an extended period. • Acquiring measurement data for processing on a PC. Overlay Sheet Display Analysis Measures with functions that support analytical applications Memory Function Recorder Function Real-Time Saving Function FFT Function Cursor Measurements Many other functions are provided in addition to these. Refer to "Appendix 2 Overview of Settings and Functions" (⇒ p. A2) and "Chapter 1 Overview" in the Instruction Manual.

1.2 Measurement Workflow

12 1.2 Measurement Workflow 1 Install & Connect Connect the instrument to the measurement object, and turn power on. See"Chapter 3 Measurement Preparations" (⇒ p. 37) in this manual See"Instrument Installation" (⇒ p. 5) in this manual Details of input module connections and connection cables: See"Chapter 2 Connections" in the Input Module Guide See"3.6 Turning the Power On and Off" (⇒ p. 46) in this manual To set the clock: See"3.7 Setting the Clock" (⇒ p. 48) 2 Make Instrument Settings Make appropriate instrument settings to acquire the mea- surement data. See"Chapter 4 Basic Operations" (⇒ p. 49) in this manual, (Auto- matic setting is available (⇒ p. 51)) For details: See"Chapter 4 Measurement Configuration Settings" in the Instruction Manual See"Chapter 3 Input Channel Settings" and "Chapter 4 Logic Input Settings" in the Input Module Guide See"Chapter 6 Trigger Settings" in the Instruction Manual See"Chapter 7 Waveform Display Settings" in the Instruction Manual See"Chapter 5 Saving & Loading Data" (⇒ p. 65) and "Chapter 6 Printing" (⇒ p. 75) in this manual, "Chapter 10 Saving/Loading Data & Managing Files" and "Chap- ter 11 Printing" in the Instruction Manual Turn power on Setting measurement configuration Set trigger criteria 3 Measure Install the instrument Select the input channels Start recording Analysis Optionally save and print Press the START key to start measuring. Measurement data is recorded by the instrument with the selected set- tings. Measurement data recording stops automatically after the preset period, or when you press STOP. (Setting contents determine what causes measurement to stop.) Analyze measurement data (⇒ p. 61). Save and print measurement data as occasion demands. See"Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual, and Analysis Supplement 5 Finish Connect Stop recording 4 Analyze, Save and Print Select the Screen Layout Turn the power off Set saving and printing conditions Overall workflow and summary If you don’t want to shut down completely (in order to measure again immediately), press the STANDBY/ON key to activate the standby state (⇒ p. 47). (To record specified waveforms) (To change the layout of the Waveform screen) (To automatically save and print while measuring) See"3.3.6 Starting and Stopping Measurement" in the Instruction Manual (As occasion demands)

1.2 Measurement Workflow

13 Operation Overview: Recording Analog Waveforms For more details, refer to "3.1 Measurement Workflow" in the Instruction Manual. Also refer to "Appendix 2 Overview of Settings and Functions" (⇒ p. A2) and "Appendix 3 Common Ques- tions" (⇒ p. A7). Installation & Connections Install the input modules and cables required for measurement (⇒ p. 37). Refer to the Input Module Guide for details about input mod- ules and cables. Power ON Function Selection Select the function according to your application. See "Function Selection" (⇒ p. 29) • Memory function To record relatively fast signals from µs to minutes • Recorder function To record relatively slow signals at low speeds from ms to hours Record instantaneous waveforms Long-term measurements Memory (MEM) (Sampling point recording) Recorder (REC) (Envelope recording) Measurement Configuration Settings Make settings on the Status Settings screen. Press the SET key to open the Settings screen Press the SUB MENU keys to select the menu Press the SHEET/ PAGE keys to select the [Use Ch] page Verify the channel(s) to be used Press the SHEET/PAGE keys to select the [Basic] page 1 Set the timebase (horizontal axis) 2 Set the recording length 1 2 See"4.2 Setting Measurement Configuration (Status Set- tings Screen)" in the Instruction Manual

1.2 Measurement Workflow

14 Input Channel Settings Make settings on the Channel Settings screen. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [One Ch] page 3 Select the Unit (module) and Channel 4 Select the measurement range (vertical axis) Make input-module-related settings 5 Perform zero adjustment 6 (As occasion demands) Set the scaling, waveform colors and zero position Trigger Settings (If you want to record a specific waveform, such as an anomaly) Set on the Trigger Settings screen. Press the SUB MENU keys to select the menu 7 Set the trigger mode 8 Set the trigger criteria (AND/OR) 9 (Memory) Set pre-trigger (Recorder) Set trigger timing 10 Set each trigger source 4 6 3 5 See"Chapter 5 Input Channel Settings" and "7.1 Making Input Waveform Display Settings (Analog Waveforms)" in the Instruction Manual "Chapter 3 Input Channel Settings" in the Input Module Guide (after warm-up) 9 10 7 8 See"Chapter 6 Trigger Settings" in the Instruction Manual Analog or Logic Trigger External Trigger Timer Trigger Default setting: (Memory) [Auto] (Recorder) [Single] Default setting: [OR] Default setting: All [Off]

1.2 Measurement Workflow

15 Display Sheet Settings (If you want to change the layout of the waveform screen to show any combination of channels) Set on the Sheet Settings screen. Press the SUB MENU keys to select the menu 11 (As occasion demands) Select the Screen Layout Set the number screen divisions and the split-screen layout 12 Select the channels to display Saving Settings (If you want to save data) Set on the Save Settings screen. Press the SUB MENUs key to select the menu 13 Select automatic or manual saving Specify the saving destination 14 Select what to save 12 See"Chapter 7 Waveform Display Settings" in the Instruction Manual Combination of display channels 11 Up to 32 channels can be displayed per sheet. When using multiple Model 8958 16-Ch Scanner Units, channels are auto- matically assigned to sheets. (Example: Auto Save) See"Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual 13 14 Verify that storage media has been inserted. When saving manually, settings can be changed after measurement. Default setting: Auto Save [Off], Manual Save [Selection Save]

1.2 Measurement Workflow

16 Printing Settings (If you want to print data) Set on the Print Settings screen. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [Printer] page 15 Select automatic or manual printing Press the SHEET/PAGE keys to select the [Print Items] page 16 Select what you want to print Start of Measurement Press the START key (the green LED lights). End of Measurement Press the STOP key. Recording stops after acquiring the specified length (the green LED goes off). Press twice to stop immediately. If [Single] trigger is selected, recording stops automatically after acquiring the specified data length. Data Analysis Analysis on the waveform screen. See"Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual Optionally save and print Press the SAVE key to save. (Manual saving) Press the PRINT key to print. (Manual printing) Power OFF Remove the cables from the measurement object, and turn the power off. 16 See"Chapter 11 Printing" in the Instruction Manual 15 Verify that the paper is loaded correctly. When printing manually, settings can be made after measurement. Default setting: Auto Print [Off] Manual Print [Selection Print] Data acquisition Save & Print (when Auto enabled) Waveform data is erased when power is turned off. However, measurement settings are retained. To retain waveforms with power off, the optional Model 9719 Memory Backup Unit is required.

1.2 Measurement Workflow

17 Operation Overview: Recording Logic Waveforms Installation & Connections Connect the logic probes (⇒ p. 37). Refer to the Input Module Guide for details about probes. Power On Measurement Configuration Settings Make settings on the Status Settings screen. Press the SET key. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [Use Ch] page (With Memory function) 1 Select the channels to use Press the SHEET/PAGE keys to select the [Basic] page 2 Select the timebase 3 Set the recording length Input Channel Settings Make settings on the [Logic] page of the Channel Settings screen. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [Logic] page 4 Enable (set to ON) the channels to be displayed 5 (As occasion demands) Set the waveform display colors 1 See"4.2.1 Selecting Channels to Use" in the Instruction Manual (With Memory function) 2 3 See"4.2 Setting Measurement Configuration (Status Set- tings Screen)" in the Instruction Manual 4 5 See"7.3.1 Setting the Waveform Display" in the Instruction Manual Default setting: [Off]

1.2 Measurement Workflow

18 Trigger Settings (If you want to record a specific waveform, such as an anomaly) Make settings on the [Logic] page of the Trigger Settings screen. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [Logic] page 6 Set the trigger mode 7 Set the trigger criteria (AND/ OR) 8 Set pre-trigger or trigger timing 9 Set the logic trigger Display Sheet Settings (If you want to change the display layout of the Wave- form screen, to display any combination of waveforms) Make settings on the [Logic] page of the Sheet Set- tings screen. Press the SUB MENU keys to select the menu Press the SHEET/PAGE keys to select the [Logic] page 10 (As occasion demands) Select the Screen Layout Set the number screen divisions and the split-screen layout 11 Select whether to display or not 12 Set display position 13 Set display height of recording From here, proceed the same as for analog channels. Refer to "Saving Settings" (⇒ p. 15). 6 7 8 9 See"Chapter 6 Trigger Settings" and "6.8 Triggering by Logic Signals (Logic Trigger)" in the Instruction Manual Default setting: [Auto] Default setting: [OR] Default setting: [Off] 12 11 13 10 See"7.3.2 Setting the Display Position" and "7.3.3 Setting the Display Height" in the Instruction Manual Default setting: [On] Default setting: [Normal]

2.1 Panel Names and Functions

19 Refer to "Chapter 2 Operating Keys and Screen Contents" in the Instruction Manual for details of operating keys and screens. Names and Functions of PartsChapter 2 2.1 Panel Names and Functions Do not apply heavy downward pressure with the stand extended. The stand could be damaged. Front Panel Display (LCD) Screen Display (⇒ p. 23) (Example: 8860) Top Panel Printer (When the optional Model 8995 A4 Printer Unit is installed) Recording Printer Installation (⇒ p. 42) Vent Bottom Panel Ventilation Fan Clean the filter periodically to avoid blockage (⇒ p. 84). Protective Cover Protects the connector section when carrying by the handle. Foot May be operated front-side-up. The number of installable input modules and the position of the handle differs on the Models 8860 and 8861 Operating Keys (⇒ p. 21) Stand

2.1 Panel Names and Functions

20 Left Side Right Side Power Inlet Connect the supplied power cord here (⇒ p. 45). GND Terminal (Functional Earth) Connect to Earth ground (⇒ p. 45). Vent PC CARD slots Accepts PC Card storage media (⇒ p. 67). Eject Button Ejects the PC Card. MO Disk Slot (With optional Model 9717 MO Unit installed) Accepts MO disk storage media. ("10.1.2 Using an MO Disk" in the Instruction Manual) LOGIC Connectors (For logic probe connection) Input connectors for proprietary logic probes. (Channels A – D) Various Connectors (⇒ p. 22) Input Module Compartment Input modules install here. Model 8860: Up to 4 modules, Model 8861: Up to 8 modules Logic Channels Serial Number MAC Address This is required for network manage- ment, so do not peel it off. Handle Use to carry the instrument. (The Model 8861 has its handle on the side) POWER Switch Turns the instrument on and off (⇒ p. 46). ( :Power On :Power Off ) Module Unit and Analog Channel Numbering Module Unit Numbering Channel Numbering (From CH1 at the bottom) UNIT1 CH1 CH2 CH3 CH4 UNIT3 UNIT2 UNIT4 CH1 CH2 8947 8946 CH1 CH4 CH16 8958 (8861) (8860) UNIT1 - UNIT8 GND Terminal (⇒ p. 45) GND Terminal (⇒ p. 45) CHD CHC CHB CHA Vent

2.1 Panel Names and Functions

21 This section provides an overview of the operating keys. Refer to "Chapter 2 Operating Keys and Screen Contents" in the Instruction Manual for functional details of each key. Operating Keys *1. FN Mode Functions differ according to the type of display screen. Refer to "Chapter 2 Operating Keys and Screen Contents" in the Instruction Manual. (For the Waveform Screen) F1: Switches numerical display. F5: Sets input channels. F2: Switches displayed gauge. F6: Forces (Manual) triggering. F3: Displays level monitor. F7: Searches the waveform. F4: Executes auto setting. F8: (supported by upgrade) Select setting items CURSOR keys: Moves among setting items (Hold the left and right keys for three seconds to activate the key-lock state. Hold again for three seconds to cancel) SELECT key: Accepts the selected character or opens the pull-down menu ENTER key: Accepts the setting ESC key: Cancels an operation, closes a dialog HELP/CONV key (supported by upgrade) Set A/B cursors TYPE key: Cursor type SPEED key: Speed of cursor motion A/B knobs: Inner knob A moves cursor A, Outer knob B moves cursor B Print (⇒ p. 78) PRINT key Feed recording paper FEED key Paper feeds while pressing the key. Save (⇒ p. 71) SAVE key Select the Sheet (Waveform screen) or Page (Settings screen) SHEET/PAGE keys Scroll waveform SCROLL control (Jog: inner, Shuttle: outer) View waveforms (Waveform Screen) DISP key Open Settings Screen SET key View a file on storage media (File Screen) FILE key Toggle standby state (⇒ p. 47) STANDBY/ON key Change F key functions FUNCTION MODE key Select setting items F1 to F8 keys (Indicated as “F key” for ge- neric reference) Press the FUNCTION MODE key to change the function (FN mode)*1 labels that ap- pear below F1 to F8. Selects menus SUB MENU keys Stop measurement STOP key 1 press: Measurement stops after the set recording length, 2 presses: Immediate stop Start measurement START key The green LED at the left is lit during measurement. Select the input channels UNIT key: Module Unit selection CH key: Channel selection RANGE/POSN knobs: Measurement range selection (RANGE knob: in- ner) Zero-position setting (POSN knob: outer) Select the timebase TIME/DIV key (Input waveform acquisition rate)

2.1 Panel Names and Functions

22 SYNC. OUT Jack Use to synchronize sampling with multiple instruments. ref (Sample synchronizing signal output) See "14.2.4 Synchronized Sampling Output (SYNC.OUT)" in the Instruction Manual EXT SMPL Jack Sampling can be synchronized by an external signal applied here. (External Sampling) (Memory function only) See "14.2.3 External Sampling (EXT.SMPL)" in the Instruction Manual 100Base-TX (RJ45) Jack Accepts a LAN cable connection. For connection to a network. See "Chapter 13 Communications Settings" in the Instruction Manual PS/2 Mouse Jack A PS/2 mouse can be connected here to control the instrument. See "3.2.2 Using a Mouse" in the Instruction Manual PS/2 Keyboard Jack A PS/2 keyboard can be connected here to control the instrument. See "3.2.3 Using a Keyboard" in the Instruction Manual MONITOR Jack The instrument’s display can be replicated on an external PC monitor connected here. (24- pin analog RGB) External I/O Terminal Strip These terminals provide and accept external input and output control signals. See "Chapter 14 External Control" in the Instruction Manual USB Port The optional Model 9716 FD Drive or a commonly available USB mouse, keyboard or memory device can be connected here. See Mouse and keyboard connection: "3.2 Before Operating" in the Instruction Manual Model 9716 FD Drive connection: "10.1.3 Using a Floppy Disk" in the Instruction Manual Connectors

2.2 Screen Types and Contents

23 2.2 Screen Types and Contents There are five general screen types. Refer to the Instruction Manual for details about each screen type. Refer to the Analysis Supplement for details of the FFT function. At power on, either the Opening screen or the Screen with the same settings that were in effect when power was last turned off appears. Opening Screen System Configuration List Configuration can be verified on the System screen Setting Items Functions can be selected with the F keys. "Function Selection" (⇒ p. 29) The Waveform screen appears when you select a function. SET key FILE key To File Screen (⇒ p. 28) To Settings Screen (⇒ p. 24) To Waveform Screen Function Menu Select a function before mea- suring (⇒ p. 29). (Functions can be changed from the Waveform or Settings screens) This screen appears first after power on. (When you turn the power off with the Waveform screen dis- played, it reappears after this screen.) The boot process takes about 40 seconds. Load existing data Make settings View waveforms DISP key For details of screen contents:"2.3 Opening Screen" in the Instruction Manual To System Screen (⇒ p. 27) Press the F7 key, or hold down the SET key Waveform Screen Press the DISP key. DISP Displays waveforms acquired by the instrument. Waveforms can be displayed on separate sheets, or as the numerical values or an X-Y composite. Record Data (⇒ p. 30) Press the SHEET/PAGE keys to change the displayed sheet. Status bar (⇒ p. 30) Function Menu To change functions (⇒ p. 51) Scroll bar (⇒ p. 30) For details of screen contents: "2.4 Waveform Screen" in the Instruction Manual Setting items (⇒ p. 31) Press the SUB MENU keys to switch which setting items are dis- played. Press the F keys (F1 to F8) to select a specific setting. (⇒ p. 32) Clock (⇒ p. 48) Shows the current time. The FUNCTION MODE key changes the functions of the F1 to F8 keys. [MACRO] [SET] [FN]

2.2 Screen Types and Contents

24 Settings Screen Press the SET key. SET Displays screens for making various operation-related settings such as for measurement, data saving, printing and calculations. Select the particular settings screen from the Settings menu. Settings menu contents are function-dependent. Some Settings screens include multiple Settings pages. Page Press the SHEET/PAGE keys to select a page. Settings Menu Press the SUB MENU keys to select a Settings screen. Function Menu Status Settings Screen ([Basic]/[Use Ch]* Page) Make basic settings here, such as the timebase and recording length. You can specify which chan- nels to use.* (* Memory function only) About screen contents: "2.5.1 Status Settings Screen" in the Instruction Manual Channel Settings Screen ([One Ch], [Comment], [All Ch], [Scaling], [Variable] and [Logic] pages) Set the measurement range, scaling and input waveforms for input channels. About screen contents: "2.5.2 Channel Settings Screen" in the Instruction Manual Trigger Settings Screen ([Analog] and [Logic] pages) Set trigger criteria when using a trigger event for measuring. About screen contents: "2.5.3 Trigger Settings Screen" in the Instruction Manual Sheet Settings Screen ([Analog], [Logic], [X-Y Comp] and [Wcal]* pages) Set the Waveform screen layout and Sheet set- tings. (* Memory function only) About screen contents: "2.5.4 Sheet Settings Screen" in the Instruction Manual Measurement Configuration and Screen Display Settings For details of screen contents: "2.5 Settings Screen" in the Instruction Manual SUB MENU

2.2 Screen Types and Contents

25 Waveform Calculation (Wave Calc) Settings Screen ([Formula], [Scale] and [Const] pages) These are the display settings for waveform calcu- lations. About screen contents: "2.5.7 Waveform Calculation (Wave Calc) Settings Screen" in the Instruction Manual Settings Menu Press the SUB MENU keys to select a Settings screen. (function-dependent) Memory Division (Mem Div) Settings Screen These settings control memory partitioning. About screen contents: "2.5.5 Memory Division Settings Screen" in the Instruction Manual Numerical Calculation (Num Calc) Settings Screen These are the display settings for numerical calcu- lations. About screen contents: "2.5.6 Numerical Calculation (Num Calc) Settings Screen" in the Instruction Manual SUB MENU Analysis Settings (Utility Functions) (Memory function only)

2.2 Screen Types and Contents

26 Settings Menu Press the SUB MENU keys to select a Settings screen. Save Settings Screen ([Auto Save] and [SAVE Key] pages) Make saving-related settings such as what to save and the saving method. Automatic saving: [Auto Save] page Manual saving: [SAVE Key] page About screen contents: "2.5.8 Save Settings Screen" in the Instruction Manual Print Settings Screen ([Printer] and [Print Items] pages) Make printing-related settings such as what to print and the printing method. Print destination and printer settings: [Printer] page Print item selection: [Print Items] page About screen contents: "2.5.9 Print Settings Screen" in the Instruction Manual Data Saving and Printing Settings Page Press the SHEET/PAGE keys to select a page. SUB MENU

2.2 Screen Types and Contents

27 System Screen Press the F7 [System] key on the Opening screen, or hold the SET key down. Environment (Env) Settings Screen Use this screen to configure the system environ- ment, Waveform screen layout and operating key functions. Communication (Comm) Settings Screen ([Communication], [File], [Web] and [Command] pages) Use this screen to make communication-related settings. Use this screen to make system-related settings. Select from the menu to display a settings screen. Page Press the SHEET/PAGE keys to select a page. For screen details: "2.7 System Screen" in the Instruction Manual Settings Menu Press the SUB MENU keys to select a Settings screen. SUB MENU Setting Configuration (Setting) Screen Use this screen to store and reload instrument set- tings. External Terminal (Ext Term) Settings Screen Use this screen to configure the external I/O termi- nals.

2.2 Screen Types and Contents

28 Initialization (Init) Settings Screen Use this screen to set the clock, initialize data, run self-tests and set scanner module compensation. Configuration (Config) List Screen Displays the instrument’s system configuration. Use this screen to make system-related settings. Select from the menu to display a settings screen. For screen details: "2.7 System Screen" in the Instruction Manual Settings Menu Press the SUB MENU keys to select a Settings screen. SUB MENU File Screen Press the FILE key FILE Use the SHEET/PAGE keys to move the cursor. Folder Tree File List Use this screen to load measurement data and manage files (copy, delete, etc.). For screen details: "2.6 File Screen" in the Instruction Manual File Info Operating Procedures: "10.4 Loading Data" in the Instruction Manual

2.2 Screen Types and Contents

29 Function Selection F1 F2 F7 F8 F3 F4 FFT Function (FFT) Recommended for performing fre- quency analysis of rotating ob- jects, vibrations, sounds and etc. Spectral analysis and transfer functions are available. Memory Function (MEM) This function is most suitable for oscilloscope-type measurements, such as instantaneous waveforms and transient phenomena. Use to record relatively fast signals with periods from µs to minutes. (Sampling point recording) System Settings Sets the system environment. Clock setting procedure (⇒ p. 48) Select a function according to the desired recording data. Select from the Opening, Waveform or Set- tings screens. For function details, refer to "4.1 Selecting the Function" in the Instruction Manual. Select on the Opening screen Recorder Function (REC) This function is suitable for use in- stead of a pen recorder or pen os- cilloscope, or to record fluctuations and create records of slow phe- nomena. Use to record relatively slow sig- nals with periods from ms to hours. (Envelope recording) File Operations Displays the File screen. Use this screen to load previously saved files and for other file management. Real-Time Saving Function Recommended for long-term re- cording instead of a data recorder. Measurement data is saved to the recording media in real time. Move the cursor here, and select a function with the F1 to F8 keys. Select on the Waveform or Settings screen

2.2 Screen Types and Contents

30 Viewing Recording Data (Waveform Screen) Waveform Data acquired with this in- strument can be displayed as waveforms or numerical values. To display as numerical values: "8.13 Viewing Waveform Data as Numerical Values" in the Instruction Manual A/B Cursors Move the cursors using the A/B knobs (⇒ p. 63). A/B Cursor details: "8.8 Cursor Values" in the Instruction Manual A/B Cursor Values Appear when the A/B cur- sors are displayed. Numeri- cal calculation results and channel information can be displayed. To change the information display: Press the DISP key repeat- edly to switch between waveform, information and gauge displays. (⇒ p. 63) Time Value Display To change the display type: "12.1.3 Selecting the Time Value Display" in the Instruction Manual Trigger Mark The trigger mark appears at the point where a trigger is applied. Scroll Bar Shows the range and position of the displayed waveform. Viewing the Scroll bar: "8.1 Scrolling Waveforms" in the Instruction Manual Status Bar (Waveform Screen) Display Info Shows the sheet number of the displayed waveform. Internal Processing Status Shows the states of data ac- quisition and internal process- ing. Trigger Info Shows the trigger source and the date and time of triggering. Setting Info Indicates the setting states of Auto Save, Auto Print and A/B Cursors. Peripheral Device Setting Info Icons indicate when PC Cards have been inserted, when con- nected to peripheral devices, and which internal devices are present. Number of Data Acquisitions Shows the number of times data has been ac- quired since the START key was pressed to start measuring. Sub Message Shows the estimated time to finish storing. This bar indicates the current states of data acquisition, internal processing, settings and display in- formation. For details about each indicator, refer to "Appendix 1 Waveform Screen Display & Icons" (⇒ p. A1) in this manual and "2.4 Waveform Screen" in the Instruction Manual.

2.2 Screen Types and Contents

31 Setting Items and Contents (Waveform Screen) Indicates current settings. These can be changed while measuring. To change a setting, use the CURSOR keys to move the cursor to it, and press an F key to choose the new value. Press the SUB MENU keys to switch items. (Some items are function-dependent) [Config] (Meas Config & Trigger Criteria)↔[Trigger](Analog Trigger)↔[Num Calc]*(Numerical Calculation) ↔ [Mem Div]*(Memory Division) (* Memory function only) See "2.4 Waveform Screen";"Setting Items and Contents (Waveform Screen)", "4.4 Setting Measurement Configura- tion on the Waveform Screen", and "6.12 Making Trigger Settings on the Waveform Screen" in the Instruction Manual Zoom Sets the time (horizontal) axis magnification ratio of the se- lected section of the wave- form. (For Memory Function) Timebase Indicates the timebase (time per divi- sion on the horizontal axis) and sam- pling rate (sampling interval). (Settings can be made on the Status Settings screen.) Recording Length Shows the recording length and time. The recording length (time) for each data acquisition is set as a number of divisions. (The setting is made on the Status Settings screen.) Magnification Sets the time (horizontal) axis magnification ratio of the whole waveform. Selecting the [Mag] button dis- plays the whole waveform. Trigger Mode Setting Sets subsequent triggering criteria after a measurement operation is finished. (These settings are made on the Trigger Settings screen.) Pre-Trigger Setting Set this to record data prior to a trigger event, or for a specified period after- wards. Selectable when the Overlay function is enabled. Type of Set- ting Item Recording Length Recording Time Timebase Sampling Rate Zoom x1 x5 Recording Length (divisions) Measurement Range (e.g., Voltage Axis) Timebase (time/div) Acquired Waveform x1 Timebase and Recording Length Difference Between Magnification and Zoom 1div Magnification x2

2.3 Basic Operations

32 You have two ways to change settings. This document describes how to change settings using the F keys. In addition to the operating keys, operations can be performed using a mouse or keyboard. Refer to "3.2.2 Using a Mouse" and "3.2.3 Using a Keyboard" in the Instruction Manual. Initialization and Default Settings Each item of waveform data and setting data can be initialized. See "12.3.2 Initializing Waveform Data" "12.3.3 Initializing System Settings (System Reset)" "Appendix 2.1 List of Default Settings" in the Instruction Manual 2.3 Basic Operations 2.3.1 To Change Settings F Keys Setting Items Setting items appear at the right side of the screen (GUI area). Select using F1 to F8. When [Page] appears at F8: more selections are available. Press the F8 key to see the additional selections. Selecting with an F key (F1 to F8) Moving to a Selection Select from a Pull-Down Menu Move the cursor to setting item. Pull-Down Menu ENTER key SELECT key CURSOR keys Opens the pull-down menu. Select a setting item. Applies the setting.

2.3 Basic Operations

33 See "3.3.3 Entering Text and Numbers" in the Instruction Manual Move the cursor to a setting item and select an input method with the F keys. When using a mouse, double click on a setting item to display the virtual keypad. Enter a numerical value using the settings (↑↑, ↓↓, ↑ and ↓) assigned to the F keys. Enter a numerical value using the virtual keypad. Enter a numerical value using the virtual pushwheel switches. 2.3.2 Entering Text and Numbers Entering Numbers Direct entry by F keys [↑↑] [↓↓] [↑] [↓] Enter by virtual keypad [Keypad] Enter using the virtual pushwheel switches [Pushwheel] 1. Select a digit ..........CURSOR keys 2. Accept ...................F1 [Set] or SELECT key After entering all digits, 3. Accept entry............F6 [OK] or ENTER key Cancel entry F7 [Cancel] key or ESC key 1. Move to a digit ......... CURSOR keys 2. Set the digit value..... CURSOR keys After entering all digits, 3. Accept entry............F3 [OK] or ENTER key Cancel entry ...........F4 [Cancel] or ESC key

2.3 Basic Operations

34 Move the cursor to a setting item and select an input method with the F keys. When using a mouse, double click on a setting item to display the virtual key- board. See "5.2 Adding Comments"; "Comment Entry Example" in the Instruction Manual Enter text using the virtual keyboard for character entry. Connect a keyboard for direct keyboard entry. Entering Text Enter by virtual keyboard [Edit] Enter by external keyboard [Direct] 1. Select a character....CURSOR keys 2. Enter it in the entry field F1 [Set] or SELECT key After entering all characters, 3. Accept entry ............F6 [OK] or ENTER key Cancel entry............F7 [Cancel] key or ESC key Press F8 to switch character case. 1. Switch character case..F8 [Character Set] key 2. Select a character....CURSOR keys 3. Enter it in the entry field F1 [Set] or SELECT key 4. Accept entry............F6 [OK] or ENTER key 5. Cancel entry............F7 [Cancel] key or ESC key [List/History] Enter prestored text from the virtual keyboard When entering a file name (to load a file on a PC) Windows 2000 and XP cannot handle file names containing the following char- acters, so they should not be used: ASCII: + = [ ] \ / | : * ? “ < > ; ,

2.3 Basic Operations

35 Press the DISP key. The recorded waveform appears on the Waveform screen. For analysis procedures, refer to "9. Analyze Data" (⇒ p. 61), or "Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual. See "8.1 Scrolling Waveforms" in the Instruction Manual Turn the inner (Jog) or outer (Shuttle) knobs of the SCROLL control to scroll the waveform display. (only when the recorded length is at least 25 divisions) You can reverse the Jog and Shuttle scroll directions. See "12.2.3 Specifying Jog & Shuttle Scroll Operations" in the Instruction Manual The waveform scroll direction can be indicated vertically or continuously. See "7.2.5 Setting Waveform Scrolling Orientation" in the Instruction Manual To increase the scrolling speed: The scroll speed is controlled by the degree of rotation of the outer Shuttle knob. Turning the knob more increases the scrolling speed. To scroll automatically (Auto-Scroll): Turn the outer Shuttle knob in the direction desired to scroll the waveform, hold it until “Auto-Scroll” appears on the screen, and release it. When the Shuttle knob is released, the waveform continues to scroll automati- cally. To cancel, press any operating key. See "8.8 Cursor Values" in the Instruction Manual Press the TYPE key and set an A/B cursor using the Settings dialog. Turn inner knob A or outer knob B to move the A/B cursors on the screen to confirm the value at a cursor. 2.3.3 Viewing Recorded Data Jog Shuttle Scrolling Waveforms Turn CW Scrolls later on the waveform from the current point. Turn CCW Scrolls earlier on the waveform from the current point. Screen Display Scroll Direction Earlier Later Knob A Knob B Confirming Measured Values

3.1 Installation & Connection Procedures

37 Be sure to read the "Operating Precautions" (⇒ p. 5) before installing and connecting this instrument. Measurement Preparations Chapter 3 3.1 Installation & Connection Procedures Install this instrument (⇒ p. 5) 1 (When adding or changing an input module) Install an input module Refer to the Input Module Guide. 3 Which probes and cables to connect depends on your measurement application (⇒ p. 38). Refer to the Input Module Guide for details. 4 (When printing) Load recording paper (⇒ p. 42) 7 Connect the power cord (⇒ p. 45) (When measuring in an elec- trically noisy environment) Ground the Functional Earth of this instru- ment (⇒ p. 45) Turn the power on (⇒ p. 46) When finished measuring, disconnect the cable(s) from the measurement object(s), and turn the power off. Left Side Right Side Connect the input cable(s) to the input module Connect a logic probe to the LOGIC receptacle (When measuring analog signals) (When measuring logic signals) Top Panel 6 7 GND 2 POWER 5 6 Other Connections To use a PS/2 mouse, connect it before turning the power on. See "3.2.2 Using a Mouse" and "3.3.4 Mouse Operations" in the Instruction Manual • If the clock is wrong after turning power on: "3.7 Setting the Clock" (⇒ p. 48) • When using a PC Card: "5.2 Using PC Cards" (⇒ p. 67) • When using other storage media or drives: "10.1 Storage Media" in the Instruction Manual • When using a keyboard: "3.2.3 Using a Keyboard" in the Instruction Manual

3.2 Connecting Cables and Probes

38 When using an input module for measurement: Connect the cables or sensors to the input module. When measuring logic signals: Connect the logic probe(s) to the LOGIC receptacle(s) on the instrument. Refer to the Input Module Guide for precautions regarding connections and details of connection procedures. Connection Preparations For voltage measurement 3.2 Connecting Cables and Probes • Do not use cables other than those specified by Hioki. The specified cables use insulated BNC connectors to avoid electric shock accidents. An uninsulated BNC plug may cause electric shock or damage to the BNC jack. • Do not connect a cable to the instrument while it is connected to the object to be measured. Otherwise, an electric shock accident may occur. Connecting to the BNC jacks on input modules When disconnecting the BNC connector, be sure to release the lock before pull- ing off the connector. Forcibly pulling the connector without releasing the lock, or pulling on the cable, can damage the connector. Lock 2 Input Module Side Guide Pins BNC Plug Slots 1 Connecting Align the two slots in the BNC plug with the guide pins on the jack at the input module side, then push and twist the plug clockwise. Disconnecting Push the BNC plug in, twist it counterclock- wise, and pull it out. Use only the specified connection cords. Using a non-specified cable may result in incorrect measurements due to poor connection or other reasons.

3.2 Connecting Cables and Probes

39 Applicable Input Modules • Model 8956 Analog Unit • Model 8936 Analog Unit • Model 8938 FFT Analog Unit • Model 8946 4-Ch Analog Unit • Model 8957 High Resolution Unit • Model 8937 Voltage/Temp Unit • Model 8940 F/V Unit • Model 8959 DC/RMS Unit • Model 8947 Charge Unit *1 • Model 8958 16-Ch Scanner Unit *2 (Maximum input voltage (⇒ p. 8)) Use to connect: Connection Cables • Model 9197 Connection Cord (Maximum input voltage: 500 V) Large alligator clip type • Model 9198 Connection Cord (Maximum input voltage: 300 V) Small alligator clip type • Model 9217 Connection Cord (Maximum input voltage: 300 V) For measuring BNC output If the voltage to be measured exceeds the maximum input rating of the input module being used: • Model 9322 Differential Probe *3 • Model 9665 10:1 Probe • Model 9666 100:1 Probe Example: Model 9322 Measuring Voltage Connect to the BNC jack on an input module. Connection Procedure: "2.2 Connecting the Cables" in the Input Module Guide *1. The Model 9198 Connection Cord is rec- ommended for connection. *2. Use an input cable to make connections. Recommended cables: 0.14 to 1.5 mm2 single-strand, or 0.14 to 1 mm2 multi-strand AWG 16 to 26 stripped 5 mm *3. A power cord option or AC adapter may be required, depending on the input module used. Applicable Input Modules • Model 8940 F/V Unit • Voltage measuring input modules Use to connect: Clamps • Clamp-On Sensors: Models 9270, 9271, and 9272 • Universal Clamp-On CTs Models 9277, 9278, and 9279 • Clamp-On Probes Models 3273, 3273-50, 3274, 3275, 3276, 9018-10, and 9132-10 Example: Models 9272 and 9318 Measuring Current Connect to a sensor receptacle or BNC jack on an input module. Connection Procedure: "2.2.5 Connecting to the Model 8940 F/V Unit" and "2.3 Connecting Clamps" in the Input Module Guide Depending on the clamp, a conversion ca- ble or power supply unit may be required for connection. ("1.3 List of Input Mod- ules, Cables, Probes and Clamp Combi- nations" in the Input Module Guide) Scaling is also required in some cases. ("5.4 Converting Input Values (Scaling Function)" in the Instruction Manual)

3.2 Connecting Cables and Probes

40 Applicable Input Modules • Model 8937 Voltage/Temp Unit • Model 8958 16-Ch Scanner Unit Use to connect: Thermocouple Thermocouple For the Model 8937 Voltage/Temp Unit (Compatible wire: AWG 16 to 26, 0.4 to 1.2 mm diameter) For the Model 8958 16-Ch Scanner Unit (Compatible wire: AWG 16 to 26, 0.14 to 1.5 mm2) Connect to terminal block Connect to terminal block Measuring Temperature Connect to the terminal block on the input module. Connection Procedure: "2.2.3 Connecting to the Model 8937 Volt- age/Temp Unit" and "2.2.7 Connecting to the Model 8958 16-Ch Scanner Unit" in the Input Module Guide Applicable Input Modules Model 8940 F/V Unit Use to connect: Connection Cable Model 9198 Connection Cord (Maximum input voltage: 300 V) Small alligator clip type Measuring Frequency, Count and Pulse Duty Connect to the BNC jack on an input mod- ule. Connection Procedure: "2.2.5 Connecting to the Model 8940 F/V Unit" in the Input Module Guide Applicable Input Modules • Model 8939 Strain Unit • Model 8960 Strain Unit Use to connect: Sensor Strain Gauge Transducer (Not available from Hioki) Example: Connecting using a conver- sion cable Using a Strain Gauge to Measure Vibration or Displacement (Strain) Connect to the BNC jack on an input mod- ule. Connection Procedure: "2.2.4 Connecting to a Strain Unit (Models 8939 and 8960)" in the Input Module Guide Use the conversion cable supplied with the input module according to the sensor be- ing used.

3.2 Connecting Cables and Probes

41 Applicable Input Modules Model 8947 Charge Unit Use to connect: Acceleration Sensor or Connection Cable * Not available from Hioki Preamp: Preamplified acceleration sensor* (BNC plug, operating power 2 mA @ 15 V) Charge: Charge-output type acceleration sensor* (Miniature #10-32 connector) Voltage Measurement: Model 9198 Connection Cord (recommended) Measuring Vibration using Piezoelectric (Charge- or Preamp-type) Sensors Preamp Connect to the BNC jack on an input mod- ule. Charge Connect to the miniature receptacle on the input module. Connection Procedure: "2.2.6 Connecting to the Model 8947 Charge Unit" in the Input Module Guide Use to connect: Logic Probe • Model 9320-01 Logic Probe • Model 9321-01 Logic Probe • Model 9327 Logic Probe LOGIC Receptacles Measuring Logic Signals Connection Procedure: "2.6 Connecting Logic Probes" in the Input Module Guide

3.3 Loading Recording Paper (With a Printer Module Installed)

42 Printing is available using the following optional printer (specified when ordering the instrument) and recording paper. About the Printer Recording Paper Usage Precautions 3.3 Loading Recording Paper (With a Printer Module Installed) Printer Recording Paper (size) Model 8995 A4 Printer Unit Model 9231 Recording Paper (A4 size) Model 8995-01 A6 Printer Unit Model 9234 Recording Paper (A6 size) The print head and surrounding metal parts of the Model 8995-01 A6 Printer Unit can become hot. Be careful to avoid touching these parts. When the instrument is not to be used for a long time, or when transport- ing To avoid straining some parts of the printer, and to prevent dirt adhering to the print head, set the print head to its lowered position (so that the print head rests against the print roller). After storing the printer without use for a long time, inspect the following before use: 1. Move the head raising lever up and down. 2. Perform test printing (printer check) three or four times. • Please use only the specified recording paper. Using non-specified paper may not only result in faulty printing, but printing may become impossible. • If the recording paper is skewed on the roller, paper jams may result. • Printing is not possible if the front and back of the recording paper are reversed. • Always use the paper cutter on the printer’s cover to cut the printed paper. Excessive paper dust can accumulate on the roller if the paper is cut by the print head. Handling and Storing Recording Paper The recording paper is thermally and chemically sensitized. Observe the follow- ing precautions to avoid paper discoloration and fading. Avoid exposure to direct sunlight. Do not store thermal paper above 40°C or 90% RH. Avoid stacking with wet Diazo copy paper. Avoid exposure to vola- tile organic solvents like alcohol, ethers and ke- tones. Avoid contact with ad- hesive tapes like soft vi- nyl chloride and cellophane tape.

3.3 Loading Recording Paper (With a Printer Module Installed)

43 Loading Recording Paper in the Model 8995 A4 Printer Unit Retainer Print Roller Back Side of Recording Paper Paper Roll Holders Stock Cover 1 Pull the knob to open the stock cover. 2 Pull the head-raising lever forward. The print head is separated from the print roller. (Head-raised condition) 3 Put the paper roll holders in the ends of the paper roll, and place it in the retainers. 4 Insert the paper through the slot under the print roller, and pull the paper out so that the print roller turns. 5 Align the edge of the paper that you pulled out with the edge of the paper still on the roll. If the paper is misaligned, it will crease as it feeds. 6 Return the head-raising lever to the lowered position. The print head should rest on the paper. (Head-low- ered condition) 7 Insert the paper through the paper outlet in the stock cover, and pull a little out. 8 Close the stock cover and cut the paper using the cutter at the paper outlet to finish the proce- dure. Head-Raising Lever Head-Raising Lever 2 1 Print Roller Knob Paper Cutter

3.3 Loading Recording Paper (With a Printer Module Installed)

44 Retainer Axle 2 1 Loading Recording Paper in the Model 8995-01 A6 Printer Unit Print Roller Back Side of Recording Paper Stock Cover 1 Push the stock cover to open the stock cover. 2 Raise the head-raising lever. The print head is separated from the print roller. (Head-raised condition) 3 Insert the paper through the slot under the print roller, and pull the paper out so that the print roller turns. 4 Insert the paper roll axle through the center of the paper roll, and place it in the retainers. 5 Align the edge of the paper that you pulled out with the edge of the paper still on the roll. If the paper is misaligned, it will crease as it feeds. 6 Return the head-raising lever to the lowered position. The print head should rest on the paper. (Head-low- ered condition) 7 Insert the paper through the paper outlet in the stock cover, and pull a little out. 8 Close the stock cover and cut the paper using the cutter at the paper outlet to finish the proce- dure. Print Roller Paper Cutter Head-Raising Lever Head-Raising Lever Printer Head

3.4 Connecting the Power Cord

45 When measuring in an electrically noisy environment, the effects of noise can be minimized by grounding the Functional Earth terminals. 3.4 Connecting the Power Cord • Before turning the instrument on, make sure the supply voltage matches that indicated on the its power connector. Connection to an improper supply voltage may damage the instrument and present an electrical haz- ard. Rated mains supply voltage (100 to 240 V AC), rated mains supply fre- quency (50/60 Hz) (Voltage fluctuations of ±10% from the rated supply voltage are taken into account.) • To avoid electrical accidents and to maintain the safety specifications of this instrument, connect the power cord only to a 3-contact (two-conduc- tor + ground) outlet. 1 Verify that the instrument’s power switch is turned off. 2 Confirm that the mains supply voltage matches the instrument, and connect the power cord to the power inlet on the instrument. 3 Plug the power cord into the mains outlet. Connection Procedure Left Side 3.5 Grounding the Instrument’s Functional Earth Left Side Functional Earth Terminals Right Side

3.6 Turning the Power On and Off

46 When using the optional Model 9684 DC Power Unit, refer to "3.2.5 If the Model 9684 DC Power Unit is Installed" in the Instruction Manual. 3.6 Turning the Power On and Off Turn the POWER switch on ( | ). After Power-On After a short time, the STANDBY/ON key lights red and the Opening screen appears. The next time power is turned on, the Waveform screen for the function that was displayed when power was turned off will appear. Power On Left Side Turning Power On Verify that the instrument and peripheral devices are correctly connected. Before turning power on Before Starting Measurement To obtain precise measurements, provide about 30 minutes warm-up (depend- ing on installed input modules) after turning power on to allow the internal tem- perature of the input modules to stabilize. After that, perform zero adjustment before taking measurements. Refer to the Input Module Guide for input module specifications and zero adjust- ment. Turn the POWER switch on ( ). When power is turned on again, the display appears with the settings that existed when power was last turned off. Turning Power Off Power Off Recording Data When the POWER switch or STANDBY/ON key of the instrument is turned off, internal recorded data is erased. However, waveform data can be preserved when power is turned off if the optional Model 9719 Memory Backup Unit is installed. If you don’t want to lose recorded data, save it first to a PC Card or external storage media. Saving Procedure: "Chapter 5 Saving & Loading Data" (⇒ p. 65) Instrument settings are preserved when power is turned off. Before Turning Power Off

3.6 Turning the Power On and Off

47 To Avoid the Startup Delay (Entering the Standby State) Press the STANDBY/ON key. “Enter Standby Mode?” appears. Press the STANDBY/ON key again. (It slowly blinks red) The screen turns off and the instrument enters the standby state. The fan continues to run for a while after entering standby. Also, the fans run periodically during standby to avoid internal heating. To resume operation from standby: Press the STANDBY/ON key. (It lights red) The initial or waveform screen reappears.

3.7 Setting the Clock

48 To adjust the clock, set the date and time on the Initialization screen of the Sys- tem screen. 3.7 Setting the Clock If you need to change both the time zone and the date and time, change the time zone first. See "12.3.1 Setting the Date and Time" in the Instruction Manual 5 Setting Procedure Open the Initialization Settings Screen −−−−−−−− 1 When the Initialization screen appears: Press the F7 [System] key. If a screen other than the Initialization screen appears: Use the CURSOR keys to move the cursor to the Function menu at the upper left, and press the F7 [System] key. The System screen appears. 2 Press the SUB MENU keys to select the [Init] menu item. The Init (Initialization) Settings screen appears. Set the time zone −−−−−−−−−−−−−−−−−−−−−−−−−−−−−− 3 Use the CURSOR keys to move the cursor to the locale set- ting field as shown at the left, and select regions with the F key. 4 Use the CURSOR keys to move the cursor to the [Execute] key, and press the F1 [Execute] key. Set the date−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−− Example: Set the date and time to May 5th 2005, 10:20:30. 5 Use the CURSOR keys to move the cursor to the date setting field as shown at the left, and press the F1 [Set] key. 6 Use the CURSOR keys to move to the year, month and day, and change the numbers with the CURSOR keys. (Setting can be made with the F1 to F4 keys) 7 When finished changing the date, press F5 [OK] or the ENTER key. Set the time−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−−− 8 Press the CURSOR key to move the cursor to the time set- ting field, and press the F1 [Set] key. Change the time setting like the date setting above. 9 When finished changing the time, press F5 [OK] or the ENTER key. Apply the new settings −−−−−−−−−−−−−−−−−−−−−−−− 10 Use the CURSOR keys to move the cursor to the [Execute] key, and press the F1 [Execute] key. F7 Function Menu 3 The date has been set. Time Setting Field System Screen Init Menu Item Date Setting Field Time Zone Setting Field 6,7 8 9 2 4

4.1 Pre-Measurement Inspection

49 Before measuring, be sure to read the "Operating Precautions" (⇒ p. 5) and "Chapter 3 Measurement Preparations" (⇒ p. 37). Basic Operations Chapter 4 4.1 Pre-Measurement Inspection Do not use if damage is present, as you could receive an electric shock. Replace the damaged items. Metal Exposed Is the insulation of the probe or connection cable to be used damaged, or is bare metal exposed? When using probes and connection cables When using a clamp Is the clamp cracked or damaged? Yes 1 No Metal Exposed 2 To No 2 Before using the instrument the first time, verify that it operates normally to ensure that the no damage occurred during storage or shipping. If you find any damage, contact your dealer or Hioki representative. Peripheral Device Inspection To If damage is evident, request re- pairs. Is damage to the instrument or input mod- Yes ules evident? Instrument and Input Module Inspection When turning power on Do the fans rotate and the Hioki logo appear on the screen? No 2 The power cord may be damaged, or the instrument may be damaged internally. Request repairs. Does the Initial or Waveform screen appear? No Yes The instrument may be damaged in- ternally. Request repairs. Nothing appears, or the display is abnor- mal Yes Inspection complete

4.2 Operation Workflow

50 4.2 Operation Workflow Recording Select with the SUB MENU keys Settings Set measurement configuration Set up input channels Load setting data (⇒ p. 74) Set trigger criteria Print (⇒ p. 75) Save (⇒ p. 65) Analyze FILE key START key Waveform Screen Select a Function Settings Screen File Screen SET key Waveform Screen Range can be set auto- matically. (⇒ p. 51) *1 Regular Recording Record Specified Waveform (per trigger criteria) Trigger criteria met STOP key Setting data can be loaded automatically. *2 Installation & Connections "Chapter 3 Measurement Preparations" (⇒ p. 37) "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual "Chapter 11 Printing" in the Instruction Manual • "Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual • Analysis Supplement Trigger mode: [Auto] Settings for saving *3 Settings for printing *3 Set the screen layout of the Waveform screen Trigger criteria met (Auto) Trigger mode: [Single] Trigger mode: [Repeat] Record Once Record Repeatedly Turn Power On (Awaiting trigger) STOP key *1. To enable automatic setting: "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual *2. To load settings automatically: "10.5 Saving & Loading Auto Settings File (Auto Setup Function)" in the Instruction Manual *3. When saving or printing automatically, execution timing depends on the function. When saving or printing manually, settings can be changed after measurement. Finished Recording Recording Starts Recording Starts Start Recording Measure with New Settings Measure with Existing Settings Set as occasion demands. Analysis, Saving & Printing

4.3 Recording Simple Waveforms

51 First display the waveform for confirmation. This description presumes that settings are at their factory default values. If the instrument has been used before, or if you suspect that settings are not at their factory default values, reinitialize the instrument before proceeding. See "12.3.2 Initializing Waveform Data", "12.3.3 Initializing System Settings (System Reset)", "Appendix 2.1 List of Default Settings" in the Instruction Manual 4.3 Recording Simple Waveforms Example: Record a 120 V AC (60 Hz) waveform. To display waveforms, set the horizontal axis (timebase), vertical axis (measurement range) and recording length. In the measurement procedure described here, the range is set automatically. Use Auto Setup to configure the lowest-numbered channel among those in use. Auto Setup is recommended when the measurement range is undetermined, or when you want to verify waveform input. However, depending on the measurement mode of the input module, Auto Setup may not be available. See "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual Input modules and measurement modes not supported by Auto Setup: • The [Temp] mode of the Model 8937 Voltage/Temp Unit • The Model 8939 Strain Unit • [Count], [Duty] and [50/60 Hz] (mains voltage measurement) modes of the Model 8940 F/V Unit • [Charge] and [Preamp] modes of the Model 8947 Charge Unit • The Model 8958 16-Ch Scanner Unit • The Model 8960 Strain Unit Turn Power On Opening Screen Select a Function Press the F1 [MEM] key. The Opening screen appears. Make Necessary Connections Refer to "Chapter 3 Measurement Preparations" (⇒ p. 37) and "Chapter 2 Connections" in the Input Module Guide. The Waveform screen appears.

4.3 Recording Simple Waveforms

52 Set Up the Instrument (Auto Setup) 1 Press the FUNCTION MODE key to enable the FN mode. 2 Press the F4 [Auto Setup] key. A confirmation message appears. 3 Press the F1 [OK] key. The timebase, measurement range and zero position are automatically set, and recording starts. Recording continues until you press the STOP key. FN Auto Setup makes the following settings: While viewing a waveform, settings can be changed as occasion demands. See "To Change Settings" (⇒ p. 53) Settings Screen Setting Item Setting Value Status 1 Timebase (Time) 1 ms/div 2 Shot (Recording length) (Unavailable in Auto Setup) Channel Mode Voltage 3 Range 20 V/div 4 Position 50 % Trigger 5 Trigger Mode Auto Source (AND/OR) OR 6 Pre-Trigger 20% (Analog Trigger) (Only lowest- numbered module and channel) 7 Type Level 8 Level Approx. 0 V 9 Filter Off 1 2 4 5 1 3 1 div 7 9 8 6 SUB MENU keys Finish Measurement Press the STOP key.

4.3 Recording Simple Waveforms

53 To Change Settings See "4.4 Setting Measurement Configuration on the Wave- form Screen" "5.7 Setting Input Channels from the Waveform Screen" "6.12 Making Trigger Settings on the Waveform Screen" in the Instruction Manual Setting Items Input Channel Settings Dialog After changing a setting, press the START key again. However, if the change is made while measuring, the setting is reflected at the moment it is changed. Changing the timebase or recording length Use the CURSOR keys to move the cursor to the set- ting item, and press F1 to F8 to select the setting val- ue. Pressing the SUB MENU keys changes available setting items. Changing the selected input channel Press the UNIT or CH key to display the [CH SET] di- alog, and select the channel to be changed. • Changing the measurement range Turn the inner RANGE knob. • Changing the waveform display position (zero position) Turn the outer POSN knob. • Erasing unwanted waveforms Select the unit (module) and channel of the waveform to be erased, and change [On] to [Off]. Press the ESC key to remove the dialog. Verifying a Waveform A waveform’s horizontal axis can be expanded or compressed (⇒ p. 62) The horizontal axis can be expanded or compressed by setting the magnification in the [Mag] (Magnifica- tion) field. By selecting the [Zoom] button, a section of the wave- form can be expanded. See "8.9 Magnifying and Compressing Waveforms" in the Instruction Manual Reading values at cursors (⇒ p. 63) Press the TYPE key to open the [A/B Cursor] dialog. Select the cursor type, and move the A or B cursor by turning the A or B knob, respectively. The values at the A/B cursors appear at the top right. See "8.8 Cursor Values" in the Instruction Manual Shuttle Scrolling a waveform (⇒ p. 35) Turn the inner Jog knob of the SCROLL control to scroll slowly, or the outer Shuttle knob to scroll quickly. See "8.1 Scrolling Waveforms" in the Instruction Manual Measurement value at cursor Cursor A Cursor B Jog

4.4 Actual Measurement and Analysis

54 This example uses an analog input module and connection cable to illustrate the basic measurement process of recording and analyzing an anomalous waveform. Settings and measurement procedures differ according to the specific application, as described in the Input Module Guide, the Instruction Manual and the Analysis Supplement . This description presumes that settings are at their factory default values. If the instrument has been used before, or if you suspect that settings are not at their factory default values, reinitialize the instrument before proceeding. See "12.3.2 Initializing Waveform Data", "12.3.3 Initializing System Settings (System Reset)", "Appendix 2.1 List of Default Settings" in the Instruction Manual Measurement Example Overview Record voltage sag on a 240 V, 60 Hz AC line. When the anomaly occurs, data is automati- cally saved to a PC Card for analysis. 1 Make Basic Settings Make the basic settings required for measurement such as measurement configuration (recording speed [Timebase] or recording time [Shot]) and input channel selection. 2 Make settings to record the specified waveform anomaly (trigger criteria settings) Make appropriate trigger settings to acquire the desired waveform data. (Default settings are fine for continuously recording waveforms) 3 Make settings to save recording data to storage media To save automatically while measuring, enable Auto Save before starting measurement. After measuring, data remaining in the instrument's internal memory can be saved manually. (Printing works the same way) 4 Start measurement, and analyze recorded data After measuring, you can analyze the data, and save or print it as occasion demands. 4.4 Actual Measurement and Analysis Setting required for this measurement: *1. Unit (module) and channel number to use *2. Trigger settings are not needed to only observe fluctuations of the mains voltage. Settings Screen Setting Item Setting Value (Function) MEM (Memory) Status Timebase 5 ms/div Shot (Recording length) 50 div Channel Unit, Ch Unit 1, Ch 1*1 Mode Voltage Range 20 V/div Position 50 [%] (default setting) Mag (Magnification) x1/2 Trigger*2 Trigger Mode Repeat Pre-Trigger 10% Type Drop (Voltage Sag) Level 280 V (RMS 197.99 V) Freq (Frequency) 60 Hz Save Auto Save On Save in (PC Card folder) 240 V AC Load Required equipment: • This instrument • Grounded power cord (2-conductor + ground) • Model 8936 Analog Unit............1 • Model 9198 Connection Cord ...1 • PC Card • Recording Paper In this case, the Model 8936 Analog Unit is installed as Unit 1 in the Model 8860, so the measurement proce- dure describes connection of the Model 9198 Connection Cord to Ch 1. CH1 UNIT 1 CH2 2 3 4 8936 9198 H L Verifying Before Measurement Input Module Compartment

4.4 Actual Measurement and Analysis

55 1 Connect the cable to a BNC jack on the Model 8936 Analog Unit. Connect the cable clips to the AC power line. 1. Measurement Preparations UNIT 1 See "Chapter 3 Measurement Preparations" (⇒ p. 37), and "Chapter 2 Connections" in the Input Module Guide This example uses channel 1 (Ch 1). Right Side Left Side Turn the power on. 2 3 4 1 Connect the power cord from this instrument to an outlet. Make sure the power switch is turned off. 2 8936 CH1 4 9198 Connection Cord 3 Twist clockwise to lock. Insert a PC Card into PC CARD Slot 1. PC CARD Slot No preparation is required when using the optional hard disk. Verify that write protection on the storage media is disabled. Preparations for Data Saving and Loading Using a PC Card Using an MO disk (Using an internal MO unit) USB Port Model 9717 MO Unit To save or print automatically, be sure to prepare the required devices before measurement. In this case, insert a PC Card into PC CARD Slot 1. Connect the drive to the USB port, and insert storage media. (Example: Model 9716 FD Drive) Using a USB storage media drive Insert an MO disk into the MO disk slot. Verify that the paper is loaded correctly. Using the internal printer: "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42) Using an external printer: Connect to the USB port and load paper. Printing Preparations

4.4 Actual Measurement and Analysis

56 2. Select the appropriate function Function Menu Opening Screen Case Press the F1 [MEM] (Memory Function) key. Using the CURSOR keys, move the cursor to the Function menu, and press the F1 [MEM] key. See "Function Selection" (⇒ p. 29) Waveform or Setting Screen Case 3. Set measurement configuration 2 Press the SUB MENU keys to select the [Status] menu item. 3 Press the SHEET/PAGE keys to select the [Basic] page. 4 Set the timebase to [5 ms/div]. This is the time per division on the hori- zontal axis. Sampling rate = timebase (s/div) / 100 (samples per division) 5 Set the recording length to [50 div]. Set the recording time in number of divi- sions. (Recording time = timebase (s/div) × re- cording length (div)) Now use the CURSOR keys to move among the setting items, and press the F1 to F8 keys to select the desired set- ting. Make other settings as occasion demands. Supplement • When F8 [Page / ] appears, more selections are available. Press the F8 key to display the additional selections. • In the memory function case, sampling speed is 1/ 100th the timebase. Each division contains 100 data samples. More detailed analyses are possible when the timebase is set smaller. • Enable auto-ranging if you cannot determine the appropriate range (⇒ p. 51). 1 Press the SET key. • When using multiple channels, verify that appears with the channel to be used on the [Use Ch] page. See: "Setting the Timebase (Horizontal Axis)" (⇒ p. A2) in this manual "4.2.1 Selecting Channels to Use", "4.2.2 Setting the Timebase (Horizontal Axis) and Sampling Rate", "4.2.4 Setting the Recording Length (number of divi- sions)" in the Instruction Manual See "Chapter 4 Measurement Configuration Settings" in the Instruction Manual

4.4 Actual Measurement and Analysis

57 4. Set up the input channel Check This! Indicates the type of module (unit) selected. 3 Select the module [Unit1] and chan- nel number [Ch1] to be set. 4 Make appropriate input channel set- tings such as the measurement range. Make other settings as occasion demands. Setting Item Setting Value (Verify) [Module]:Analog, Mode: Voltage Range(/div) 20 V (per division) (Each vertical division repre- sents 20 V) Coupling DC (default setting) 5 Make appropriate waveform display settings. Make other settings as occasion demands. See "7.1 Making Input Waveform Display Set- tings (Analog Waveforms)" in the Instruction Manual Setting Item Setting Value Wave Disp On (default setting) Mag (Magnification) x1/2 Position 50% (default setting) Supplement Determining the range: Measurement range is denoted as value per division on the vertical axis. Setting a measurement range causes the measure- ment area and display area of the level monitor (on the left side of the screen) to be correspondingly changed. It is convenient to watch the indicated display area on the level monitor while setting the range. 6 Select the [Zero-Adjust] button. Execute zero-adjust 30 minutes after turning power on. Proper warm-up time depends on the particular input mod- ule. Verify the specifications in the Input Module Guide. A message appears when zero-adjust is fin- ished. 1 Press the SUB MENU keys to select the [Channel] menu item. 2 Press the SHEET/PAGE keys to select the [One Ch] page. See "Chapter 5 Input Channel Settings" in the Instruction Manual and "Chapter 3 Input Channel Settings" in the Input Module Guide Now use the CURSOR keys to move among the setting items, and press the F1 to F8 keys to select the desired set- ting. Level Monitor To enter a comment for an individual channel: A comment entered in the comment entry field ap- pears on the waveform. See "5.2 Adding Comments" in the Instruction Manual To convert the displayed units for measurements taken with a clamp or sensor, set the Scaling func- tion. See "5.4 Converting Input Values (Scaling Function)" on the Instruction Manual

4.4 Actual Measurement and Analysis

58 5. Set Trigger Criteria 3 Press the SHEET/PAGE keys to select the [Analog 1-4] page. 2 Set the trigger. See "2.3.2 Entering Text and Numbers" (⇒ p. 33) Setting Item Setting Value Trigger Mode *1 Repeat Pre-Trigger (% setting) *2 10% 4 Set the analog trigger.*4 Make other settings as occasion demands. See "Chapter 6 Trigger Settings" in the Instruction Manual Setting Item Setting Value [Normal] (default setting) (Applies one trigger to one channel) Type Drop (trigger type) Unit 1-1 (Unit1-Ch1) Level *3 280 V (trigger level) Parameter Frequency: 60 Hz Supplement *1. About (trigger) modes • Select [Auto] if you want to view the input waveform continuously, even when no trigger is applied. • Select [Single] if you want to record just one event and stop. • Select [Repeat] to record continuously until you press the STOP key. See "6.3 Setting the Trigger Mode" in the Instruction Manual *2. About pre-triggering To record the waveform prior to the trigger event, set the trigger point at some percentage of the recording length. See "6.5 Pre-Trigger Settings" in the Instruction Manual *3. About (trigger) level The trigger level is set as an instantaneous value in- stead of an RMS value. 1 Press the SUB MENU keys to select the [Trigger] menu item. See "Chapter 6 Trigger Settings" in the Instruction Manual Now use the CURSOR keys to move among the setting items, and press the F1 to F8 keys to select the desired setting. *4. About analog and logic triggering • This selection is made from the Settings dialog. Move the cursor to the [No.] field and press the F1 [All Settings] key to display the dialog. Set each item in the dialog, and press the ENTER key or [Close] button when done. • [Normal] and [Expanded] analog triggers Select [Normal] to apply one trigger type on one channel, and select [Expanded] to apply multiple triggers on one channel. (The [Normal] setting is available for level, window threshold and voltage sag trigger types.) See "6.7 Triggering by Analog Signals" in the Instruction Manual

4.4 Actual Measurement and Analysis

59 6. Enable Auto-Save Check This! [On] (default setting) Trigger date and time are automatically added as a prefix to the specified sav- ing name. “Time_Date_Saving Name” (default setting) Select “Text” to make the file readable on a PC. 3 Set Auto Save [On]. 4 Select the saving destination. At [Save in 1], press the F1 key to open the [Browse Folders] dialog. The waveform can now be saved with other settings in their default states. Change these settings as occasion demands. See "10.3 Saving Data" "10.3.4 Setting Auto Save" "10.3.7 Automatically Saving Waveforms" in the Instruction Manual 8860 Supplement • Using Auto Save, waveforms, numerical calculation results and screen image can be saved at the same time. • When the [Save in 1] storage media becomes full, sub- sequent recording data can be stored on [Save in 2] storage media. (⇒ p. A10) • If the [Save Method] setting is [Normal Save] (default setting) and the storage media becomes full, an error message appears. Stop measurement and change the storage media. 1 Press the SUB MENU keys to select the [Save] menu item. What about these situations? To record over a long period Use high-capacity storage media. Specify a folder other than the root directory as the save destination. When more than 5,000 files have been created in one folder, a new folder is automat- ically created and saving continues. To continue saving even when the storage media becomes full, select [Delete Save] as the [Save Method]. When the storage media becomes full, the oldest file is deleted and saving continues. See "10.3.4 Setting Auto Save" in the Instruction Manual To also print automatically On the Print Settings screen, set Auto Print [On]. Every time a waveform is ac- quired, it is saved and printed. See "Chapter 11 Printing" in the Instruction Manual Verify once again before starting measurement. • Are connections made correctly? • Is write-protection on the storage media disabled? • Is storage media inserted? See "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual Now use the CURSOR keys to move among the setting items, and press the F1 to F8 keys to select the desired set- ting. CURSOR: Selects the storage media. Select [PC Card #1] (PC Card Slot 1). To navigate the tree, press CURSOR. 2 Press the SHEET/PAGE keys to select the [Auto Save] page.

4.4 Actual Measurement and Analysis

60 7. Start Measurement Press the START key. The green LED lights, and the Waveform screen appears. The instrument’s measurement status is displayed on the Status bar. Recording does not occur until the specified trigger criteria are met. The waveform is displayed. (If the [Roll Mode] setting is [Off] on the Status setting screen, the screen displays acquired data in quantities determined by the specified by the recording length.) Every time data is acquired, it is saved to the PC Card and the “Trigger Wait” state is resumed. Trigger Wait Storing Saving Save to PC Card Status Bar Display In this measurement case, 50 divisions of data (including the pre-trigger setting of 10% of the 50-divi- sion recording length) are recorded when the voltage is 280 V (197.99 Vrms) or less, at 5 ms per divi- sion on the horizontal time axis. Settings can be changed and the waveform can be scrolled while recording. Recording restarts when setting are changed. (“Storage” is displayed) See "3.3.6 Starting and Stopping Measurement" in the Instruction Manual Press the STOP key. 8. Finish Measurement See "3.3.6 Starting and Stopping Measurement" in the Instruction Manual • Pressing Once: Recording stops when finished acquiring data for the recording length. • Pressing Twice (Halt): Recording stops immediately. If you halt while awaiting a trigger, no waveform is displayed. (However, with the Memory function, the waveform does appear if the recording length is shorter than half of the maximum recording length and a trigger event has occurred before the Halt command.) What about these situations? You can disable the keys to prevent inadvertent opera- tions Hold both CURSOR keys simultaneously for three seconds. “Key Lock” appears under the clock display while the key-lock state is enabled. To cancel key-lock, hold the keys again for three seconds. When "Trigger Wait" contin- ues to be displayed and no waveform appears If the trigger criteria are not met, no waveform is displayed (⇒ p. A10). If you need to display the waveform, execute a forced (manual) trigger. (FN Mode: F6 key) To view part of the waveform that is not visible while re- cording Use the SCROLL controls to scroll the waveform. When turning the inner Jog or outer Shuttle knob, the [Scroll Trace] button appears at the left. Press the F1 [Scroll Trace] key to return to display of the currently recording waveform. Refer to the scrolling methods: "2.3.3 Viewing Recorded Data" (⇒ p. 35) To change settings while re- cording Timebase, recording length, input channel and trigger settings can be changed from the Waveform screen. See "To Change Settings" (⇒ p. 53), "4.4 Setting Measurement Configuration on the Waveform Screen" in the Instruction Manual

4.4 Actual Measurement and Analysis

61 9. Analyze Data What about these situations? To measure with changed set- tings After changing settings on the Waveform or Settings screen, press the START key. To load saved data from stor- age media Press the FILE key to display the File screen, and select the file to load. See "10.4 Loading Data" in the Instruction Manual To print recorded waveforms or settings Make the appropriate settings on the Print Settings screen, and press the PRINT key. See "6.3 Manual Printing (PRINT Key)" (⇒ p. 78) in this manual, "11.4 Making Manual Print (PRINT Key Output) Settings" in the Instruction Manual To save settings Press the SAVE key and select [Settings] to choose the save type, and save. By saving settings, you can reload them later when you want to repeat measure- ment with the same settings. Saved settings can also be automatically reloaded when power is turned on. See "5.5 Data Saving" (⇒ p. 71) and "5.6 Data Loading" (⇒ p. 74) in this manual To use recorded data on a PC Select the waveform or display the image type, and save it. For details about data saving types, refer to "5.3 Data that can be Saved & Loaded" (⇒ p. 68) Use the LAN interface to connect to a PC. See "Chapter 13 Communications Settings" in the Instruction Manual Time from Trigger Event Number of Waveform Trigger Event Time Acquisitions Name of Currently Displayed Sheet See "Chapter 8 Waveform Screen Monitoring and Analysis" in the Instruction Manual Displayed Channel Trigger Event Position

4.4 Actual Measurement and Analysis

62 Viewing the Whole Waveform Select the [Mag] (Magnification) button The whole waveform over the specified recording length is displayed. Using the CURSOR keys, move the cursor to [Mag], and press the F1 [Whole Wave] key. Whole waveform See "8.9 Magnifying and Compressing Waveforms" in the Instruction Manual Expanding Horizontally See "8.9.1 Magnifying and Compressing Horizontally (Time Axis)" in the Instruction Manual Select the horizontal axis magnification The waveform view is magnified as selected. Use the CURSOR keys to move the cursor to the magnification setting field, and press F1 to F8 to select the desired magnification. Zooming Part of a Waveform See "8.9.3 Magnifying a Section of the Horizontal Axis (Time Axis – Zoom Function)" in the Instruction Manual Use the Zoom function Use the SCROLL control to scroll the zoomed section of the waveform. To return to the previous view, move the cursor to [Zoom], and press the F1 [Off] key. 1 Use the CURSOR keys to move the cursor to [Zoom], and press the F2 [On] key. The Zoom display is enabled. 2 Move the cursor to the Zoom setting field, and select the desired magnification. Zoomed Waveform Section Waveform with Section to be Zoomed

4.4 Actual Measurement and Analysis

63 Viewing Measurement Values See "8.8 Cursor Values" in the Instruction Manual Use the A/B cursors Dialog Press the TYPE key to select the cursor type. The [A/B Cursor] dialog appears. • To view the time and voltage (measurement) value: [Trace] • To see time and period: [Vertical] • To see the voltage value: [Horizontal] Make other settings as occasion demands. Knob A Knob B Turn inner knob A or outer knob B to move the cur- sor to the point to be read. The value at the cursor is displayed. Voltage value Cursor A value Cursor B value Press the DISP key to display method. (Waveform/ Waveform and Information Display/ Gauge and Wave- form Display/ Gauge, Waveform and Information Dis- play) The selection can be made in the dialog. Press the ESC key to close the dialog. See "8.4 Displaying Measured Values and Information" in the Instruction Manual To change the gauge of a displayed channel, press the FUNCTION MODE key to enable the FN mode, and press the F2 [Gauge] key. In the dialog, select the channel for which to display the gauge. To remove the displayed gauge, press the DISP key to switch to Normal display. See "8.5 Applying Gauges" in the Instruction Manual Applying a Gauge To remove the cursors, press the FUNCTION key to return to the FN mode, and press the F1 [Switch Info] key. Trace Cursor Difference between A/B cursors Time from trigger event Gauge Display Channel If Overlapping Display of Waveforms and Numerical Values Obstructs Viewing 1 2

4.4 Actual Measurement and Analysis

64 1 Specify the section to print using the vertical or trace cursors. See"Viewing Measurement Values" (⇒ p. 63) 10. Printing a Selected Waveform Section See "Chapter 11 Printing" in the Instruction Manual 2 Press the PRINT key. The [Print] dialog appears. 3 Press the F2 [A-B Wave] key. The specified section is printed. This procedure describes Selection Printing (default setting). Function, Timebase Upper Limit of Display Area Trigger Mark Unit-Channel Number Lower Limit of Display Area Supplement To print items other than waveforms such as a settings list, gauge and upper and lower limits, select which items you want to print on the [Print Items] page. See "6.4 To Print a Detailed Settings List" (⇒ p. 79) Set the number of cursors to [2]. Channel Settings

Saving & Loading

Data Chapter 5 Compatible Recording Media (⇒ p. 66) Load into this instrument (File Screen) • Settings Data • Waveform Data Compatible Data Types, and Procedures (⇒ p. 71) Settings Data Waveform Data • Save automatically while measuring (⇒ p. 59) • Save waveform selection after measurement (SAVE key) Numerical Calculation Results • Calculate and save automatically while measuring • Calculate and save after measuring (SAVE key) Screen Image (Display Screen) • Save automatically while measuring • Select and save a screen after measuring (SAVE key) The default at shipping and initialization of this setting is [Selection Save]. To save selections, press the SAVE key, make the appropriate settings and save them. You can later change your saved settings from the Save Settings screen as occasion demands. For details, refer to "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual. Data Types & Formats (⇒ p. 68) Saving Procedure Overview (⇒ p. 71) Loading Data See "10.4 Loading Data" in the Instruction Manual Viewing Data from this Instrument on a PC See "Chapter 13 Communications Settings" in the Instruction Manual Handling Storage Media See "10.1 Storage Media" in the Instruction Manual This document describes saving and loading procedures using PC Cards in the internal PC Card slots. This document describes saving procedures commonly used after measurement. Save Procedure Details See "10.3 Saving Data" in the Instruction Manual File Size See "Appendix 2.2 Waveform File Sizes" in the Instruction Manual File management on this instrument (File Screen) Initialize (format) storage media Internal storage media file and folder operations Formatting Storage Media See "10.1.7 Initializing (Formatting) Storage Media" in the Instruction Manual File Management See "10.7 Managing Files" in the Instruction Manual • Copy • Move • Delete • Rename • Create folder On-screen operations • Sort • Display only specified files (Filter Settings) • Select display items from the File List (Display Set- tings) • Print the File List

5.1 Recording (Storage) Media

66 5.1 Recording (Storage) Media Storage Media Remarks and Hioki Options Storage Media Designation (when specifying the storage media type) PC Card (Built-in slots provided) • Model 9626 PC CARD 32M • Model 9627 PC CARD 64M • Model 9726 PC CARD 128M • Model 9727 PC CARD 256M • Model 9728 PC CARD 512M • Model 9729 PC CARD 1G Insert in a PC CARD slot on the right side of the instrument. (⇒ p. 67) (Two slots are provided) PC CARD #1 (PC CARD Slot 1) *2 PC CARD #2 (PC CARD Slot 2) *2 MO Disk Model 9717 MO Unit *1 (Compatible MO disks: 3.5-inch, with capacity of 128, 230, 540 or 640 MB, and 1.3 or 2.3 GB) See "10.1.2 Using an MO Disk" in the Instruction Manual MO (with internal Model 9717 MO Unit installed) *2 Floppy Diskette Model 9716 FD Drive (Compatible floppy diskettes: 3.5-inch, 2-HD or 2-DD, with capacity of 720 KB or 1.44 MB) See "10.1.3 Using a Floppy Disk" in the Instruction Manual FDD #1, FDD #2, …, FDD #5 (using the Model 9716 FD Drive) *2 Hard Disk Model 9718 HD Unit *1 (Capacity: 60 GB) See "10.1.4 Using a Hard Disk" in the Instruction Manual HDD (USB drives) Commonly available USB memory or other USB storage de- vices can be connected to the USB port.*4 See "10.1.5 Using USB Memory Devices" in the Instruction Manual USB DISK #1, USB DISK #2, ... , USB DISK #5 (Networks) Data can be transferred directly to and from PCs over a net- work. See "Chapter 13 Communications Settings" in the Instruction Manual NETWORK #1, NETWORK #2, ... , NETWORK #10 *3 For details about saving and loading: "Chapter 10 Saving/Loading Data & Managing Files" in the Instruction Manual For details about inserting storage media and connecting drives:"Preparations for Data Saving and Loading" (⇒ p. 55) *1. Either the Model 9717 MO Unit or 9718 HD Unit may be installed. Both cannot be installed together. *2. Displayed when storage media is inserted. *3. Displayed when connected to a network with sharing enabled. *4. Not all commonly available media drives are supported. When using a USB memory device: Do not remove the USB memory device while it is being accessed (writing or reading). Also, immediately after saving data and before removing it, verify on the File screen that the data has been saved. No particular instrument operation is required to remove a USB memory device. Be sure to read the above references regarding handling of each type of storage media. If damage occurs to the hard disk or internal memory, we cannot restore or ana- lyze the lost data. We therefore recommend maintaining a backup of any impor- tant data.

5.2 Using PC Cards

67 PC Card Insertion & Removal 5.2 Using PC Cards Important Use only PC Cards sold by Hioki (⇒ p. 66). Compatibility and performance are not guaranteed for PC cards made by other manufacturers. You may be unable to read from or save data to such cards. • Format new PC Cards before use. To format with this instrument: "10.1.7 Initializing (Formatting) Storage Media" in the Instruction Manual • Be careful to avoid inserting a PC Card backwards or upside-down. The PC Card or the instrument could be damaged. • Never eject a PC Card while it is being accessed by the instrument. Data on the PC Card could be lost. • When storage media is inserted, a confirmation beep sounds, and an icon appears in the Status bar at the lower right. "Appendix 1 Waveform Screen Display & Icons" (⇒ p. A1) • Be sure that storage media is inserted before selecting it for saving or loading. It cannot be displayed in the File List if it has not been inserted. • The Flash memory in a PC Card has a limited operating life. After long-term usage, data storage and retrieval become difficult. In this case, replace the PC Card with a new one. • We cannot provide compensation for data loss in a memory card, regardless of content or cause of the damage. Always maintain a backup of important data stored on a memory card. Right Side of Instrument PC CARD Slots (two) Eject Button Inserting a PC Card With the Front mark (�) pointing toward the slot, insert the card in the direction of the arrow all the way in. Removing a PC Card Press the Eject button. When the button pops out, press it again to eject the PC Card.

5.3 Data that can be Saved & Loaded

68 The following data can be saved and loaded with this instrument. Auto Save saves the data automatically after measurement. 5.3 Data that can be Saved & Loaded “O” = Possible, “–” = Not Possible File Type File Format File Extension and Contents Save Load PC- Readable Reference Auto Manual Settings Data*2 Settings made on the Settings screen, such as measurement configurations Binary SET Settings Data (Measurement Configuration) — O O —*5 Instruction Manual: "10.3.6 Saving Settings Data" "10.4.2 Loading Settings Data"*1 Waveform Data*2 All waveform data acquired by the instrument, or the waveform section demarcated by A/B cursors. Binary MEM Memory Function Waveform Data O O O —*5,*6 Instruction Manual: "10.3.7 Automatically Saving Waveforms" "10.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" "10.6.2 Reading Waveform Data on a PC" "Chapter 9 Measuring with Real- Time Saving" REC Recorder Function Waveform Data O O O —*5,*6 RSM Sampled waveform data from the Real- Time Saving function O — O — RSR Whole waveform data from the Real-Time Saving function O — O — FFT FFT Function Wave- form Data O O O —*5 Text TXT Text Data O O — O Waveform Management data (Divisions)*3 Management data for dividing and saving waveform data (Index file) IDX Index data for divided saving O O O —*5 Instruction Manual: "10.3.7 Automatically Saving Waveforms" "10.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" (when [Binary] save type is selected) SEQ Index data for memory division (created automatically for batch saving) O O O —*5 Instruction Manual: "4.3.3 Dividing Memory" "8.12 Viewing Waveforms in Every Display Block (Memory Division)" RSI Index data for the Real- Time Saving function O — O — Instruction Manual: "Chapter 9 Measuring with Real- Time Saving" Numerical Calculation Results Text TXT Text Data O O — O Analysis Supplement : "1.4 Saving Numerical Calculation Results" "1.5 Reading Numerical Calculation Results on a PC" Captured Screen Image *4 BMP BMP Image Data O O — O Instruction Manual: "10.3.9 Automatically Saving Display Images" "10.3.10 Optionally Selecting Display Screens & Saving (SAVE Key)" PNG PNG Image Data O O — O

5.3 Data that can be Saved & Loaded

69 *1. Settings data can be loaded automatically at power-on. (Auto Setup Function) *2. To reload data into this instrument: Save it in binary format. Waveforms and some of the measurement settings are saved. To load data into a PC: Save it in text format. To save a waveform section, demarcate the section to save with the A/B cursors. *3. To load all blocks at once when memory division is enabled: Save using the [All Blocks] selection. A directory is created automatically, and files for the waveform data of each block and an index file (SEQ) are created. This index file is used for reloading. To reload waveform data saved with the Divided Saving function, load the index file (IDX). To reload waveform data in the Real-time saving function, load the index file (RSI). *4. BMP Format: This is a standard Windows graphics format. These files can be handled by many graphics programs. PNG Format: This image file format has been internationally standardized as ISO/IEC15948. *5. Loading is possible when using the optional Model 9725 Memory HiViewer. *6. Loading is possible with the Waveform Viewer (Wv).

5.4 Saving Procedures and Contents

70 5.4 Saving Procedures and Contents Basically, three methods are available for saving. To save automatically while measuring To save immediately upon pressing the SAVE key To save selected contents Auto Save This method is convenient for saving specific types of data, such as wave- forms or images, or when you want to save while monitoring. Specify what you want to save before press- ing the SAVE key. (Default setting) Press the SAVE key to select and save the selected contents. No set- tings need to be made on the Save Settings screen, as settings can be made on any screen. Quick Save Selection Save Automatically saves all acquired measurement data for the specified recording length. Various types of data can be saved at the same time. Select the save destination and con- tents before measuring (⇒ p. 59). After installation and connection, turn power on. Insert storage media Set measurement configuration, etc. Set the Auto Save setting [On] Specify the saving destination Select what to save Measure Saving proceeds automatically From the Save Settings screen, open the [Auto Save] page The following can be saved simultaneously: • Waveform Data • Numerical Calculation Results • Screen Image After installation and connection, turn power on. Insert storage media Set the function of the SAVE key to [Quick Save] Specify the saving destination Select what to save Press the SAVE key*1 Save From the Save Settings screen, open the [SAVE key] page Any of the following can be saved: • Waveform Data • Settings Data • Numerical Calculation Results • Screen Image After installation and connection, turn power on. Insert storage media Set the function of the SAVE key to [Selection Save] Press the SAVE key. Specify the saving destination Select what to save Select the [Execute] button*1 Save From the Save Settings screen, open the [SAVE key] page Any of the following can be saved: • Waveform Data • Settings Data • Numerical Calculation Results • Screen Image Verify before saving • Is storage media inserted? • Is write-protection on the storage media disabled? • Is the specified save destination correct? • Is Auto Save set to [On] for automatic saving? *1.To save only a section of a waveform, demarcate the section with the A/B cursors. (Saving a wave- form section is not available with Auto Save) To save a screen image, display the desired screen before saving.

5.5 Data Saving

71 The saving procedure for various data items using the default [Selection Save] method is described here. Available setting contents are the same for the [Quick Save] method. Refer to "6. Enable Auto-Save" (⇒ p. 59) for more details about Auto Save. Refer to "10.3 Saving Data" in the Instruction Manual for saving procedure details. 5.5 Data Saving Before setting, verify that write-protection is disabled on the destination storage media, and insert it. When entering a file name (to be read by a PC) Do not use any of the following characters in a file name, as PCs running Win- dows 2000 or XP may not be able to handle it: ASCII: + = [ ] \ / | : * ? “ < > ; , Shared Settings in Selection Save 1 Press the SAVE key. The [Save] dialog appears. 2 Press the F1 [Edit] key. The [Browse Folders] dialog appears. 3 Use the CURSOR keys to move the cursor to the storage media, and press the F1 [OK] key. To open a folder on the storage media, press the CURSOR key with the to the left of the folder select- ed. 4 Press the F1 [Edit] key to enter the file name to save. (Up to 40 characters) See "2.3.2 Entering Text and Numbers" (⇒ p. 33) If you have a keyboard connected, press the F2 [Direct] key and enter from the keyboard directly. 5 If a file with the same name already exists in the destination folder, select whether to overwrite or append automatic numbering. 6 Select whether the trigger date and time should be appended before or after the specified file name. Numbering (Default setting) If a file with the same name already ex- ists, a number is automatically ap- pended to the new file name to avoid duplication. (Up to four digits) Overwrite Duplicate existing file names are over- written. Saving a Waveform (⇒ p. 72) Saving Settings Data (⇒ p. 73) Saving Captured Screen Images (⇒ p. 73) To enter the save name: If a keyboard is connected, you can press the F2 [Direct] key and enter directly from the keyboard.

5.5 Data Saving

72 Saving Waveforms *2 See "10.3.8 Optionally Selecting Waveforms & Saving (SAVE Key)" in the Instruction Manual 1 Press the F2 [Waveform] key. 2 Select the data saving format. 3 Select the area to be saved. 4 Select the channels to save. 5 (when [Binary] saving format is selected) Select to save by divisions depending on file size. See "Appendix 2.2 Waveform File Sizes" in the Instruction Manual 6 Press the F1 [Execute] key. Binary To be able to reload data into the instru- ment (Default setting) Text *1 To be able to load data into a PC Whole Save all measurement data. (Default setting) A-B Save data in the area demarcated by the A/B cursors. Displayed Ch Saves the channels on all sheets for which waveform display is enabled [On]. (Default setting) All Ch Saves all stored channels. Off Saves as one file without divisions. (Divisions) If the data is larger than the specified recording length, it is divided and saved in multiple files. *1. When [Text] is selected as the save type, data is saved with thinning. Examples 2: Save every other data point. (•x•x•x...) 5: Save one of every five data points. (•xxxx•xxxx•xxxx..) (•: saved, x: not saved) "Shared Settings in Selection Save" (⇒ p. 71) Use the CURSOR keys to move the cursor to the setting item, and select the desired setting contents with the F1 to F8 key. *2: [Target Blocks] This is displayed when using the Memo- ry Division function. When [All Blocks] is selected, the index file (SEQ) and waveform data of all used blocks are saved as a batch. About Memory Division: "4.3.3 Dividing Memory" in the Instruction Manual Note: Data remaining in memory is saved by manual saving with the Real-time sav- ing function.

5.5 Data Saving

73 Saving Settings Data See "10.3.6 Saving Settings Data" in the Instruction Manual 1 Press the F1 [Settings] key. 2 Press the F1 [Execute] key. "Shared Settings in Selection Save" (⇒ p. 71) Use the CURSOR keys to move the cursor to the setting item, and select the desired setting contents with the F1 to F8 key. Saving Captured Screen Images See "10.3.10 Optionally Selecting Display Screens & Saving (SAVE Key)" in the Instruction Manual 1 Press the F3 [Screen Image] key. 2 Select the file saving format. 3 Select whether to save the F key display (GUI area). 4 Press the F1 [Execute] key. BMP Color Saves a color bitmap format file. (Default setting) Comp BMP Saves a compressed color bitmap format file. BMP Gray Saves a grayscale bitmap format file. PNG Saves a PNG format file Without The displayed area is not saved. With The displayed area is saved. (Default setting) "Shared Settings in Selection Save" (⇒ p. 71) Use the CURSOR keys to move the cursor to the setting item, and select the desired setting contents with the F1 to F8 key.

5.6 Data Loading

74 5.6 Data Loading Insert storage media Select the data to load (File extension: .SET) Select F1 [Load] Select F1 [Execute] Load Settings Data Waveform Data Insert storage media Select the data to load (File extension: .MEM, .REC) Select F1 [Load] Select F1 [Execute] Load Verify before loading • Is storage media inserted? • Is the loading source correct? This instrument can load data saved in binary format. Loading Settings & Waveform Data See "10.4.2 Loading Settings Data" and "10.4.3 Loading Waveform Data" in the Instruction Manual 1 Press the FILE key. The File screen appears. 2 Use the CURSOR keys to select the storage media, and press F1 [List]. The files on the selected storage media appear in the File List. 3 Use the CURSOR keys to select the file to load, and press the F1 [Load] key. If [Load] does not appear, press the F8 key until F8 [Page 1/3] is displayed. File List Operations • To open a folder: ENTER key • Return to previous folder: ESC key A [Confirmation] dialog appears. 4 Press the F1 [Execute] key. The data is loaded. To cancel: Press the F2 [Cancel] key. Verify that storage media has been inserted. Folder Tree File List Folder tree operations are also available. To select a file or folder CURSOR Keys To move to or display one deeper level CURSOR Key (Opens the level below the di- rectory with the selected symbol) To move to or display one higher level CURSOR Key (Closes levels below the di- rectory with the symbol) Press the SHEET/PAGE keys to move the cursor.

(⇒ p. 76)

• Automatic Printing [Auto Print]: Memory Function and FFT Function [Real Time Print]: Recorder Function • Manual Printing [Quick Print]: Prints when you press the PRINT key [Selection Print]: Press the PRINT key and select what to print (Default setting) Print Settings • Printout Type (Waveform, Numerical Values and Screen Link) • Print Area (Whole Waveform, A-B) • Display value of horizontal axis (Time Value Dis- play) Printer Settings Internal Printer (Print Destination: [Printer]) • Print Density • Paper Feed • Print Quality External Printer (Printing Destination: [USB]) • Paper Orientation, Margins • Printing Colors The shipping and initial default printing method is [Selection Print]. With this setting, printing occurs after pressing the PRINT key and selecting what you want to print. You can change print settings as occasion demands from the Print Settings screen. Refer to "Chapter 11 Printing" in the Instruction Manual for details. When using the optional printer unit, read the precautions in "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42). Print Setting Details (⇒ p. 79) When printing waveforms * • Grid Type • Channel Markers • List & Gauge • Upper and Lower Limits • Zero-Position Comments • Counter • Time Axis Expansion and Compression • Gauge (when using external printer) "11.6.2 Printing Waveforms" in the Instruction Manual When printing numerical data • Thinned numerical value data "11.6.3 Printing Numerical Value Displays" in the Instruction Manual To print comments and settings data together * • Print Comments (Analog, Logic) • Print Title • Settings Data "11.6.5 Printing Comments and Setting Data" in the Instruction Manual * Some items can be set only when a particular function is enabled. Printable Items (⇒ p. 76) • Whole Waveform • A-B Waveform • Waveform before and after trigger event • List (Settings List) • Report • Calculation results • Screen image Depending on the print method or function, some items cannot be printed. Printing Destination • Internal Printer • External Printer (on USB port, only HP printers are supported) "11.6.1 Common Settings" in the Instruction Manual Description of printout types: (⇒ p. 78) Printing examples: "11.7 Print Examples" in the Instruction Manual "11.5 Making Printer Settings" in the Instruction Manual

6.1 Printing Methods and Contents

76 6.1 Printing Methods and Contents To print automatically while measuring To print immediately upon pressing the PRINT key To select and print specified items after measuring Auto Print Specify what you want to print be- fore pressing the PRINT key. It may be convenient to print certain data optionally and to print simultaneous- ly while analyzing. (Default setting) After pressing the PRINT key, select items to print before printing. This method is convenient for printing dif- ferent items each time. Quick Print Selection Print With the Memory Function*, prints au- tomatically after acquiring measure- ment data for the specified recording length. With the Recorder Function (support- ed only by the internal printer), prints in real time while recording. With the FFT Function, prints auto- matically after FFT calculation. In either case, before measuring, se- lect the desired items to print. * When the Roll Mode is enabled, waveforms can be printed while being dis- played. (However, printing is delayed if the timebase is set faster than 500 ms/div.) Basically, three printing methods are available. Install recording paper (⇒ p. 42) After installation and connection, turn power on. Set measurement configuration, etc. Set the Auto Print setting [On] Select the print destination Select items to print Measure Print The following can be printed simultaneously. • Waveform Data • Numerical Calculation Results Install recording paper (⇒ p. 42) After installation and connection, turn power on. Select [Quick Print] as the PRINT key action Select the print destination Select items to print Press the PRINT key Print Any of the following can be print- ed: • Whole Waveform • A-B Waveform • Waveform before and after trigger event • Report • List (Settings List) • Calculation results • Screen Image • (Screen Link) Install recording paper (⇒ p. 42) After installation and connection, turn power on. Select [Selection Print] as the PRINT key action Select the print destination Press the PRINT key Select items to print Print Any of the following can be printed: • Whole Waveform • A-B Waveform • Waveform before and after trigger event • Report • List (Settings List) • Calculation results • Screen Image Verify before printing • Is the recording paper loaded correctly? • Is the printing destination set correctly? • For automatic printing: Is Auto Print enabled [On]? If both Auto Print and Auto Save are enabled, saving is performed first. However, if the Roll Mode is used with the Memory function (default setting: Auto), printing is performed first.

6.2 Printing While Recording (Auto Print)

77 6.2 Printing While Recording (Auto Print) Enable Auto Print Use the same settings to also print numerical calculation re- sults. (Memory function only) 1 Press the SUB MENU keys to select the [Print] menu. 2 Press the SHEET/PAGE keys to select the [Printer] page. Now use the CURSOR keys to move among the setting items, and press the F1 to F8 keys to select the desired setting. 3 Press the F2 [On] key. Memory function and FFT function: Auto Print Recorder function: Real-Time Printing 4 Select the printing destination. (Memory function and FFT function only) Make other printer settings as occasion demands. See "11.5 Making Printer Settings" in the Instruction Manual Printer Print using the internal printer. USB Print using an external printer. (only HP printers are supported) See "11.3 Making Auto Print Settings" in the Instruction Manual Selecting Items to Print You can select what infor- mation prints on waveform graphs (⇒ p. 79). 1 Press the SHEET/PAGE keys to select the [Print Items] page. 2 Select the recording format for the printer. 3 Select the area to be printed. 4 Select the horizontal axis value (timebase display). Make other changes to waveform, numerical value, com- ment settings as occasion demands (⇒ p. 79). Waveform, Numeric, Screen Link (Default setting) With [Screen Link], the printout corresponds to actual dis- play appearance. Whole (Default setting) or A-B [A-B] prints the data between A/B cursors. Time Time from trigger event (units are fixed) (Default setting) Mod 60 Time from trigger event (units are mod- ulo 60) Scale Number of divisions from trigger event Date Time of waveform acquisition Samples Number of samples from trigger event See "11.6 Setting the Print Content" in the Instruction Manual Start Measurement: Press the START key. Data automatically prints after acquisition by the instrument. Comments and settings data can be printed at the same time (⇒ p. Now use the CURSOR keys to move among the set- ting items, and press the F1 to F8 keys to select the desired setting.

6.3 Manual Printing (PRINT Key)

78 The [Selection Print] method is described here. To use the [Quick Print] method, refer to "11.4 Making Manual Print (PRINT Key Output) Settings" in the Instruction Manual. 6.3 Manual Printing (PRINT Key) Select when printing a screen image. 3 Select the printing destination. 4 Press the F2 [Selection Print] key. Change items to print as occasion demands on the [Print Items] page (⇒ p. 77). Printer Print using the internal printer. USB Print using an external printer. (only HP printers are supported) 1 Press the SUB MENU keys to select the [Print] menu. 5 Press the PRINT key. The [Print] dialog appears. Now use the CURSOR keys to move among the set- ting items, and press the F1 to F8 keys to select the desired setting. 2 Press the SHEET/PAGE keys to select the [Printer] page. 6 Press the F1 to F8 keys to select what you want to print. The selected items are printed. *1. Memory function and Recorder function only. *2. Memory function only. Whole Wave Prints all waveform data acquired by the instrument. A-B Wave *1 Prints the section of the waveform demarcated by the A/B cur- sors (vertical or trace cursors). When either of the A/B cursors is off the screen, the section between A-B cursors is printed. Before selecting: Specify the print area with the A/B knobs. Starting and ending points can be represented by either cursor. Trig Wave *1 Prints 10 divisions of the waveform before and after the central trigger event. Report Prints the waveform data of the displayed area on the wave- form screen, upper and lower limits and analog channel set- tings. List Prints a list of the settings on the Settings screen. Calc Results *2 Prints numerical calculation results. Screen Image Prints a captured image of the displayed screen. Before selecting: Display the screen to be printed. Feed Feeds the paper. Cancel Cancels printing Want to stop printing? Press the STOP key. To print numerical data: On the [Print Items] page, set the [Printout Type] to [Numeric]

6.4 To Print a Detailed Settings List

79 6.4 To Print a Detailed Settings List When printing waveforms [Waveform Print Items] 1 2 3 4 5 6 7 (Default setting) Setting Item Selection Description 1 Grid Type Off, Normal, Fine, Normal (Dark), Fine (Dark), Time Axis or T-Axis (Dark) Select the type and darkness of the printed grid. 2 Channel Markers Off, Ch No. or Comments Select whether to print channel numbers or comments with waveforms. (Analog channels only) Marker Position Inside or Outside Select the channel marker position. 3 List & Gauge Off, List, Gauge or List & Gauge Select whether to print a list or gauge. 4 Upper/Lower Limits Off or On The upper and lower limits of each chan- nel can be printed. When scaling is en- abled, the limit values are in scaled units. 5 Zero-Position Comment Off or On Select comment printing at the zero po- sition of each channel. (Analog channels only) 6 Counter Printing Off, Date or Name Select whether to print the number of ac- quired waveforms or counter name. This is convenient for distinguishing similar waveforms. 7 Mag/Comp Screen Link or No Screen Link The time axis of waveforms to be printed can be expanded or compressed. (When [No Screen Link] is selected, screen expansion and compression set- tings do not affect printouts) See "11.6.2 Printing Waveforms" in the Instruction Manual When printing numerical values [Numerical Value Print Items] (Default setting) Setting Item Selection Description Thinning Screen Link or No Screen Link When data is displayed with numerical values on the Waveform screen, select whether the thinning setting is linked to the screen. See "11.6.3 Printing Numerical Value Displays" in the Instruction Manual When printing comments [Comment Printing Settings] 1 2 3 (Default setting) Setting Item Selection Description 1 Title Off, Settings, Comments, Set& Com Select whether to print a title and settings data (Function, Timebase, Time Axis Magnification Ratio, Trigger Time). 2 Analog Off, Settings, Comments, Set& Com Select whether to print comments for each analog channel, or channel settings (mea- surement range, vertical axis expansion or compression, zero position, low-pass filter, full-scale value of range [or upper and lower limits when Variable is en- abled]). 3 Logic On or Off Select whether to print comments for each logic channel. See "11.6.5 Printing Comments and Setting Data" in the Instruction Manual

Inspection and Repair

Refer to "Appendix 1 Error Messages" in the Instruction Manual for error mes- sages. Maintenance and Service Chapter 7 7.1 Troubleshooting If damage occurs to the hard disk or MO disk, we cannot restore or analyze the lost data. We therefore recommend maintaining a backup of any important data. If damage is suspected, check the "Troubleshooting" section before contacting your dealer or Hioki representative. Transporting • Pack the instrument so that it will not sustain damage during shipping, and include a description of existing damage. We cannot accept responsibility for damage incurred during shipping. • To avoid damage to the instrument, be sure to remove the PC cards before shipping. Replaceable Parts Certain parts require replacement periodically and at the end of their useful life: (Useful life depends on the operating environment and frequency of use. Opera- tion cannot be guaranteed beyond the following periods) The fuse is housed in the power unit of the instrument. If the power does not turn on, the fuse may be blown. If this occurs, a replacement or repair cannot be performed by customers. Please contact your dealer or Hioki representative. Part Life Part Life Fan Motor Approx. 4 years Model 9717 MO Unit (Drive) Approx. 5 years Lithium Battery Approx. 10 years Model 9718 HD Unit (Drive) Approx. 20,000 hours Printer After using 1,000 rolls of the record- ing paper LCD Approx. 74,000 hours Backlight (to half brightness) Approx. 55,000 hours Electrolytic capacitors Approx. 4 years (The useful life of electrolytic capacitors varies greatly according to the operating environment. In severe operating environments (40°C ambient temperature), degradation occurs in about four years, so they should be replaced periodically.) Lithium battery Approx. 10 years (This instrument contains a built-in lithium battery to back up settings and the real-time clock. Have the battery replaced if the date and time are found to lag substantially or if settings are not retained when power is turned off and back on. Contact your dealer or Hioki representative.

Disposing of the Instrument

Before returning for repair This instrument includes a lithium battery to retain settings. Remove this battery before disposing of the instrument. Also remove the optional Model 9719 Mem- ory Backup Unit, if installed. See "Appendix 6 Disposing of the Instrument" in the Instruction Manual If Power and Operating Keys Malfunction Symptom Check Item, or Cause Remedy and Reference The display does not appear when you turn the power on. Is the power cord disconnected? Are connections made correctly? Verify that the power cord is connected prop- erly. "3.4 Connecting the Power Cord" (⇒ p. 45) Keys do not work. • Is any key being held down? • Verify key operation. • Is the key-lock state active (Key-Lock message displayed)? • Deactivate the key-lock state. (Hold the CURSOR keys for three seconds) • Is remote control from a PC over the Web enabled? • Press the SET key to release the key- lock. If the Display or Operations Malfunction Symptom Check Item, or Cause Remedy and Reference A waveform does not appear when you press the START key. • Is the “Pre-Trig Wait” message dis- played? • Is the “Trigger Wait” message dis- played? When pre-triggering is enabled, triggering is ignored until the pre-trigger portion of the waveform has been acquired. Recording starts when a trigger occurs. No changes occur in the displayed waveform. • Is the clamp sensor or connection cable connected correctly? • Is the measurement range set prop- erly? • Is the low-pass filter enabled? Verify that the clamp sensor or connection cable is connected correctly. "Chapter 2 Connections" in the Input Module Guide Verify the input channel settings. While measuring with the memory function, the dis- played frequency is much lower than the actual fre- quency. Aliasing may be occurring. Change the timebase to use a faster sam- pling rate. "Appendix 4 Supplemental Technical Infor- mation" in the Instruction Manual Some channels are unus- able. Is the channel enabled for use? "4.2.1 Selecting Channels to Use" in the Instruction Manual The displayed waveform size does not change when the input range is changed. Is the Variable function enabled? Turn the Variable function off. "8.9.4 Setting Arbitrary Waveform Height and Position on the Vertical (Voltage) Axis (Variable Function)" in the Instruction Manual

If the cause is unknown

Try performing a system reset. All settings are returned to their factory defaults. See "12.3.3 Initializing System Settings (System Reset)" in the Instruction Manual For other problems, refer to the Appendix or the Index in the Instruction Manual. Cannot Print, or Printing Malfunctions Symptom Check Item, or Cause Remedy and Reference Nothing prints on the paper. Is the paper reversed (back to front)? Verify that the recording paper is loaded cor- rectly. "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42) Printout is too light. • Is the specified recording paper being used? • Is the print density setting correct? • Is the print head dirty? Try changing the print density setting. "11.5.1 Internal Printer Settings" in the Instruction Manual Clean the print head. "Print Head Cleaning" (⇒ p. 85) Recording traces are too wide. The input signal may have a ripple com- ponent. Enable the filter in the input module settings. "3.10.3 Low-Pass Filter (LPF) Settings" in the Input Module Guide Recording traces are dou- bled. Is the waveform printing density set to [Light]? In this case, printed dots are spaced out in the vertical direction. Therefore, a slightly varying waveform prints some- times as one line, and sometimes as two. Change to waveform printing density setting to something other than [Light]. (Print Set- tings Screen) "11.5.1 Internal Printer Settings" in the Instruction Manual Printing does not work. • Is the printer’s head-raising lever in the correct (head-lowered) position? • Is the recording paper loaded cor- rectly? • Are the printer output settings appro- priate? Verify printer installation/connection and set- tings, and try printing again. "3.3 Loading Recording Paper (With a Printer Module Installed)" (⇒ p. 42), "Chapter 11 Printing" in the Instruction Manual If Saving is Not Possible Symptom Check Item, or Cause Remedy and Reference If saving is not possible to PC Card or MO storage media. • Is write-protection on the storage media disabled? • Is the storage media inserted properly? • Is the storage media formatted? • Is the remaining capacity of the storage media too low? "5.1 Recording (Storage) Media" (⇒ p. 66) "5.2 Using PC Cards" (⇒ p. 67)

Cleaning the Instrument and Input Modules

Cleaning the Air Filters 7.2 Cleaning • To clean the instrument and input modules, wipe it gently with a soft cloth moistened with water or mild detergent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as they can deform and discolor the case. • Wipe the LCD gently with a soft, dry cloth. Air filters are installed in the bottom panel of the instrument. Clean them periodi- cally before severe dust buildup occurs. If a filter becomes clogged, the instru- ment’s internal cooling is impeded, and damage may result. Also, be sure to thoroughly dry the filters before reinstalling. Operating with a wet filter may cause damage. 1 Turn the instrument off. 2 Place the instrument with the front upwards, and remove the filter covers from the bottom panel. 3 Remove the filters from the filter covers. 4 Use a vacuum cleaner to remove any dust or grime from the filters. If dust remains, wash with diluted detergent and dry completely. 5 Replace the filters in their covers, and reinstall. Filter Covers Instrument’s Bottom Panel Filter Filter Cover For instruments installed with filter covers For instruments installed with metal filters 1 Turn the instrument off. 2 Place the instrument with the front upwards. 3 Use a vacuum cleaner to remove any dust or grime from the filters.

Print Head Cleaning

Normally, no maintenance is required. However, depending on usage conditions, dirt and paper dust may accumulate on the thermal head over the long term, causing light or smeared printing. In this case, clean the head by the following procedure. 1 Use the CURSOR keys to move the cursor to the Function menu, and press the F7 [System] key. The System screen appears. 2 Press the SUB MENU keys to select the [Init] menu item. The Init (Initialization) Settings screen appears. 3 Use the CURSOR keys to move the cursor to the [Printer Check] button, and press the F1 [Execute] key. The [Printer Check] dialog appears. 4 Move the cursor to [Cleaning], and press the F1 key. The paper should print solid (100%) black. Check for any smearing. If sufficient improvement is not obtained even after cleaning several times, wash the print head (⇒ p. 86). Self-Check Cleaning Function Menu Before Cleaning Recording paper is required to clean the print head. Verify that the recording paper is loaded correctly. 1 2 3 4

1 Moisten the back side of a piece of printing

paper with undiluted alcohol, and insert it into the printer. If the front side is moistened, it becomes discolored. 2 Lower the head-raising lever, and wash the head by sliding the paper back and forth. Back Side of Printing Paper Washing the Print Head Be careful not to apply too much undiluted alco- hol. About the Print Head To avoid discoloring or deforming the instrument, observe the following precau- tions: • Do not use organic solvents such as thinner or benzene • After washing, allow the printer to dry completely before use. About the Roller Surface • White powder such as paper dust may accumulate on the roller surface after long-term use. A small amount should have no effect on printing, but can be removed with a commonly available camera blower brush if it causes concern. • Always use the paper cutter on the printer’s cover to cut the printed paper. Excessive paper dust can accumulate on the roller if the paper is cut by the print head.

Appendix 1 Waveform Screen Display & Icons

A1 Appendix Appendix 1 Waveform Screen Display & Icons See "2.4 Waveform Screen" of the Instruction Manual 1 2 3 4 5 6 8 7 1 Internal Processing State Pre-Trig Wait Before acquiring data Appears only when pre-trig- ger recording is enabled. Trigger Wait Trigger wait state Scanner Wait Scanner preparation state Storing Data acquisition in progress Storing Done Finished acquiring data Waveform Waveform generation in progress Calculating (n/m) Numerical value calculation in progress Calculating (Zn) Waveform calculation in progress FFT (n/m) FFT calculation in progress Stopped Operation stopped Preparing Preparation in progress Ready Idle state Printing Printing in progress Saving Saving in progress Auto-Ranging Automatic range detection in progress Complete Finished automatic range detection (File Name) Name of loaded file 2 Storage Events (No. of Times) Number of data acquisitions 3 Trigger Info Trigger cause, date and time 4 Sub Message Estimated time to storage completion (Appears when 10s or longer) Count to be Averaged(n/m) Simple Averaging (Time axis) Simple Averaging (Frequency) Exponential Aver- aging (Time axis) Exponential Aver- aging (Frequency) Peak hold (Fre- quency) 5 Setting Info Auto Save (Waveforms) Auto Save (Calculations) Auto Save (Waveforms & Calcula- tions) Auto Save (Screen images) Auto Save (Waveforms & Screen im- ages) Auto Save (Calculations & Screen im- ages) Auto Save (Waveforms & Calcula- tions & Screen images) Auto Print Auto Print (External Printer) Vertical Cursor Horizontal Cursor Trace Cursor 6 Display Sheet*1 Waveform FFT X-Y Composite Nyquist Numerical Values FFT+Nyquist Waveform + X-Y Composite Waveform + FFT Waveform + Nyquist 7 Display Information Block number when measuring with Memory Division enabled*2 Displayed block number for Memory Division*2 New (FFT function) Use newly ac- quired data for calculations. MEM (FFT function) Use pre-exist- ing data for calculations. Charging finishes about two hours after power on. 8 Peripheral Device Info PC Card (Blank) No Card Card Present GP-IB Card Internal Storage Media Internal Hard Drive Internal MO Drive (Appears only when an MO disk is present) Model 9716 FD Drive (Blank) No diskette present Diskette present LAN Connection Status (Blank) Disconnected Power Supply (Blank) AC power supply DC power supply Internal Printer (Blank) Printer not installed Printer Installed Head-raised error *3 Out-of-Paper error *4 Battery Charge State (When Model 9719 Memory Backup Unit is installed) (Blank) Disconnected Rapid charging Rapid charging finished *3. Check the position the head-raising lever. *4. Replace with new recording paper. *1. Press the SHEET/PAGE keys to switch sheets. *2. Press the SHEET/PAGE keys to switch blocks.

Appendix 2 Overview of Settings and Functions

A2 Appendix 2 Overview of Settings and Functions Instrument Settings About the Screen The instrument’s LCD provides SVGA (800 × 600) resolution. The wave- form display area consists of 625 horizontal dots and 500 vertical dots. The waveform display area is divided into 25 divisions horizontally, and 20 divisions vertically, with each division composed of 25 dots horizon- tally and vertically. Each data frame (one division) represents 100 samples horizontally, and 80 to 1600 data bits vertically. (depending on the input module) Each displayed data frame (one division) changes in accordance with expansion and compression of the time and voltage axes. 25 div 20 div 1 div = 100 samples Setting the Timebase (Horizontal Axis) See "4.2.2 Setting the Timebase (Horizontal Axis) and Sampling Rate", "8.9 Magnifying and Compressing Waveforms", and "Appendix 4.3 Measurement Frequency Limit" in the Instruction Manual The input signal acquisition rate corresponds to time per division on the horizontal axis. Determining the timebase: Calculate from the frequency and period. f [Hz] = 1/t [s] (f: frequency, t: period) Example: If the measurement frequency is 50 Hz: 50 [Hz] = 1/t [s] t = 1/50 [s] = 0.02 [s] = 20 [ms] So to display five cycles on the screen (25 divisions horizontally), 20 [ms] × 5/25 [div] = 4 ms/div Select 5 ms/div for the timebase, which is the closest available setting to the calculated value. To measure phenomena with relatively fast signals such as instantaneous waveforms, we suggest setting a small value (if the frequency is 50 Hz, the timebase should be set faster than 5 ms/div). During and after measurement, waveforms can be expanded and compressed along the time axis. 1 div Timebase: with 1 div = 5 ms • To expand and compress: Change the expansion and compression ratio. (⇒ p. 62) • To expand a specified section (Memory function only): Use the Zoom function. (⇒ p. 62) Timebase: with 1 div = 10 ms Timebase and Sampling See "Appendix 4.1 Sampling" and "Appendix 4.4 Recorder Function Values" in the Instruction Manual For the Memory Function (Sampling point recording): The sampling period is 1/100th of the timebase. When the timebase is set to 100 µs/ div, the sampling period is 1 µs. Changing the timebase setting also changes the sampling period accordingly. For the Recorder Function (Envelope recording): One data point is recorded every 1/100th of the timebase as two amplitude values: the maximum and minimum values measured during the specified sampling period. With the 10 ms/div timebase, one data point is recorded every 100 µs. When the sampling period is set to 1 µs, 100 samples are provided in order to record each data point. The values recorded for each data point are maximum and minimum values among these 100 samples. Maximum Minimum 1 div 1 2.... ....100 Sampling Period Sampling Points

Appendix 2 Overview of Settings and Functions

A3 Recording Length Setting See "4.2.4 Setting the Recording Length (number of divisions)" in the Instruction Manual Set the length (number of divisions) to record each time data is acquired. Each division of the recording length consists of 100 data points. The number of data points of the whole recording length is the specified recording length (divisions) × 100, + 1. Example. The number of data points when the specified recording length is 50 divisions: 50 div × 100 data points + 1 = 5001 data points Measurement Range Setting See "Chapter 3 Input Channel Settings" in the Input Module Guide Select the amplitude range for measurement. The setting value is the number of units per division on the vertical axis. About Resolution (per vertical division) Measurement resolution depends on the particular input module. (with x1 voltage axis magnification) Input Modules Resolution (bits/div) Screen Full Scale (bits) Minimum Resolution 8936 to 8940, 8946, 8947, 8959 80 1600 5 V/80 = 62.5 mV * 8956 100 2000 5 V/100 = 50 mV * 8957, 8960 8958 [Voltage] Mode 1600 32000 5 V/1600 = 3.125 mV* 8958 [Temp] Mode 1000 20000 10°C/DIV = 0.01°C * When measurement range is 5 V/div With 1 V/div range (10 Vp-p input) With 500 mV/div range (10 Vp-p input) 0V 1 div 20 div More about Measurement Range and Resolution: "Chapter 5 Specifications" in the Input Module Guide • To expand and compress: Change magnification of the vertical axis. ("8.9 Magnifying and Compressing Wave- forms" of the Instruction Manual) • To shift the position on the vertical axis: Change the zero position. ("7.1.2 Setting the Waveform Display Posi- tion (Zero Position)" of the Instruction Manual) • To display converted units: Use the Scaling function. ("5.4 Converting Input Values (Scaling Function)" of the Instruction Manual) • To freely set the display area of the vertical axis: Use the Variable function. ("8.9 Magnifying and Compressing Waveforms" of the Instruction Manual) About Trigger and Pre-Trigger Recording See "Chapter 6 Trigger Settings" in the Instruction Manual When recording a specific signal with recording criteria (trigger criteria) set, the waveform can be recorded only when the trigger event occurs. To view phenomena prior to the trigger event, set the Pre-Trigger value to the amount of data your want to record before the event occurs. Example: When the recording length is set to 1000 div and Pre-Trigger is set to 5% The pre-trigger recording length (data points) is 1000 div × 5% = 50 div (5000 data points). Pre-Trigger T Trigger Criteria Met

Appendix 2 Overview of Settings and Functions

A4 Convenience Functions If the appropriate timebase or measurement range is unknown Auto-Ranging Function See "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual Automatically selects the timebase and range settings to display 1 to 2.5 waveform cycles. To record and search for waveform anomalies: Trigger Function and Waveform Search Function See "Chapter 6 Trigger Settings", "8.14 Searching a Waveform" in the Instruction Manual Trigger signals can be generally be obtained from the following sources: An- alog trigger, logic trigger, external trigger, timer trigger and forced (manual) trigger. A trigger can be applied as a result of one or a combination of these criteria, except for the Manual trigger. Also, after measurement, any waveform can be searched and displayed us- ing the Search function. T T T T T T To display with actual measurement units or to fix the display level of the waveform Scaling Function See "5.4 Converting Input Values (Scaling Function)" in the Instruction Manual Input values can be converted and displayed as physical values such as speed, vibration, displacement and temperature. Two conversion methods are available: setting by conversion ratio and setting by entering equivalent values at two points. Variable Function See "8.9.4 Setting Arbitrary Waveform Height and Position on the Vertical (Voltage) Axis (Variable Function)" in the Instruction Manual The position and size of the displayed waveform can be freely set. By entering the upper and lower limits of the waveform, it can be set to display with full-screen amplitude. The Scaling and Variable functions can be used in combination. The full span of output from a sensor can be displayed. mV A 0% 100%

Appendix 2 Overview of Settings and Functions

A5 To view waveform data as numerical values/ To change the waveform display format Numerical Values Display See "8.13 Viewing Waveform Data as Numerical Values" of the Instruction Manual Split-Screen Display See "7.2.4 Splitting the Display Screen (Split-Screen)" of the Instruction Manual Scroll Direction See "7.2.5 Setting Waveform Scrolling Orientation" of the Instruction Manual The display format of measurement data can be freely selected. 0 ms 5 ms 10 ms 15 ms 20 ms 1-1 1-1 1-1 1-1 1-1 0.3V 71.5V 65.3V -71V 0.2V To view a waveform expanded or compressed, or to see the whole waveform Zoom Display (Memory function only) See "8.9 Magnifying and Compressing Waveforms" in the Instruction Manual A normal waveform and an expanded section of it can be displayed together on the screen. Whole Waveform Display All waveform fluctuations can be observed. Zoom Magnification To determine the measurement value of a waveform, or to specify the waveform range Cursor Measurements See "8.8 Cursor Values" in the Instruction Manual Use the A/B cursors to display values at the cursor positions on a waveform, and the time and number of cycles from the trigger position. A specified data range can also be saved and printed. A B A B A B Vertical Cursors Horizontal Cursors Trace Cursors To display a combination of measurement data Sheet Display See "7.2.1 Assigning Display Data to Sheets" in the Instruction Manual Up to 32 channels of data can be assigned per screen, and freely displayed on multiple sheets. X-Y Composite Waveforms See "7.4 Composite Waveforms (X-Y Waveforms)" in the Instruction Manual Specify the channels for X and Y axes to form composite waveforms. Sheet3 Sheet2 Sheet1

Appendix 2 Overview of Settings and Functions

A6 To compare with previous waveforms Memory Division See "4.3.3 Dividing Memory" of the Instruction Manual By dividing the memory space into multiple blocks, waveform data can be re- corded in any block, or overlaid with any other waveforms. Overlay See "4.3.2 Overlaying Waveforms" of the Instruction Manual Displayed waveforms are retained on-screen and overlaid with new wave- forms. Use this to compare new waveforms with those recorded immediately before. 1 2 3 4 5 6 To apply calculations to measurement data Numerical Calculations See "Chapter 1 Numerical Calculation Functions" in the Analysis Supplement Nineteen types of numerical calculations are available, including RMS values, maximum values and averages, and up to 16 calculations can be performed at the same time. Waveforms and calculation results can be displayed separately on one screen. Waveform Calculations See "Chapter 2 Waveform Calculation Functions" in the Analysis Supplement The waveform calculation functions allow display of the waveform calculated by a previously specified formula. FFT Calculations See Analysis Supplement Calculation Results Calculation Results Calculation Results CH(1,1)∗CH(1,2) CH1,1 CH1,2 Time Frequency To perform remote control and data collection/ controlling the instrument through the interface LAN (Ethernet) Communications (ftp/Web server) See "Chapter 13 Communications Settings" in the Instruction Manual Use 100Base-TX compatible cable to connect the instrument to a network for control by a PC. GP-IB Interface See Communications Operating Manual on the supplied applications disc With the Model 9558 GP-IB Card installed, the instrument can be controlled by a computer. GP-IB card

Appendix 3 Common Questions

A7 Read this when you want to perform a certain task, or when the procedure for a particular operation is unknown. Refer to "7.1 Troubleshooting" (⇒ p. 81) if the display or operation seems abnormal, or if damage is suspected. Appendix 3 Common Questions To Perform This Measurement In this case Description Reference Perform simultaneous mea- surements with different sam- pling rates Set different sampling rates on Timebase 1 and Time- base 2 from the Status screen of the Memory function, where Timebase 1 or Timebase 2 can be selected for each channel. "4.2.3 Setting Different Sam- pling Rates" in the Instruction Manual To send a warning when an anomaly occurs When a trigger is applied, a signal can be output from the External I/O terminals. "14.2.2 Trigger Output (TRIG OUT/CAL)" in the Instruction Manual To use multiple recorders Recording with multiple instruments can be performed using the external control connectors. "14.2.1 External Trigger Input (EXT TRIG)" and "14.2.2 Trig- ger Output (TRIG OUT/CAL)" in the Instruction Manual What happens to measure- ment data when power to the instrument is lost during mea- surement? Measurement data acquired prior to the power interrup- tion is lost. When [Auto-Resume] is enabled from the Environment Settings Screen of the System Screen, recording resumes automatically when power is restored. How- ever, this action differs according to trigger settings. (Default setting: [Off]) "12.2.2 Using the Auto-Resume Function (Resume After Power Restoration)" in the Instruction Manual To measure shock when dropping a test object, and to search for the vibration reso- nance point. Use an acceleration sensor with the Model 8947 Charge Unit to measure vibration and shock. To record actual effective (true RMS) values of mains voltage True RMS values can be recorded using the Model 8959 DC/RMS Unit. To record anomalous waveforms, enable level triggering to acquire waveforms when the RMS voltage drops. Example. Trigger: Level, Level: 90 V, Slope: ↓ (Falling) To monitor 120 V (50 or 60 Hz) power for a drop to 80% or lower Using a voltage measuring input module, enable trigger- ing upon voltage sag. To observe instantaneous power dropouts on 120 V (50 or 60 Hz) commercial mains, and the internal voltages of affected devices. Measurements can be made by the combination of a voltage measuring input module and a clamp with con- nection cable. Changes in commercial power, changes in internal voltage of devices and changes in current flow to devices (using a clamp-on sensor) at the instant of power dropout are recorded on separate channels, and analyzed. By splitting the screen, power and device-volt- age waveforms can be displayed in separate traces. Split-Screen Display: "7.2.1 Assigning Display Data to Sheets" in the Instruction Manual To observe surge noise on 240 V (60 Hz) AC mains. AC mains and surge waveforms can be measured simul- taneously. Surge noise can be detected by the slope trig- ger. Acquired waveforms can be analyzed by expansion and zooming.

Appendix 3 Common Questions

A8 To observe the voltage wave- form on 600 V power lines Use a voltage measuring input module and the Model 9322 Differential Probe or Model 9665 (10:1) or 9666 (100:1) attenuating probes. "2.4 Connecting a Differential Probe", "2.5 Connecting Atten- uating Probes", and "3.10.15 Probe Attenuation Selection" in the Input Module Guide To perform frequency analy- sis Measure using the FFT function. We recommend using an input module equipped with an anti-aliasing filter func- tion (Models 8938, 8947, 8957 and 8960). "Chapter 3 FFT Function" in the Analysis Supplement To record data without gaps, or over a long term Measure using the Real-Time Saving function. With the Memory and Recorder functions, data is saved after each recording length, causing measurement to be interrupted. With the Real-Time Saving function, data can be stored continuously without gaps. "Chapter 9 Measuring with Real-Time Saving" in the In- struction Manual In this case Description Reference Related Settings In this case Description Reference What is the basic setting pro- cedure in order to view a waveform? To display a waveform, set the following: (1) timebase (horizontal), (2) voltage axis range and zero position (vertical), and (3) trigger [Off]. To use automatic setting for measuring when the time- base and the measurement range are unknown Press the FUNCTION MODE key to select the FN mode on the Waveform screen (DISP key), and select F4 (Auto Setup). The proper range is selected automatically, and record- ing starts. Make other settings as occasion demands. "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the Instruction Manual To set up simply when the measurement configuration is already determined Measurement configuration settings can be saved to storage media as an auto-setup file (STARTUP.SET), so the settings can be loaded automatically at power-on. (Auto-Setup Function) Also, the settings data can be save to the instrument and it can be automatically loaded. "10.3.6 Saving Settings Data", "10.5 Saving & Loading Auto Settings File (Auto Setup Func- tion)" in the Instruction Manual What happens to measure- ment configuration settings when power is turned off? When power is turned off, measurement configuration settings are preserved within the instrument. When power is turned back on, those measurement configura- tion settings are recalled. "3.6 Turning the Power On and Off" (⇒ p. 46) in this manual What is the longest possible recording length with the fast- est timebase setting? Recording time is determined by the [Shot]. The fastest available timebase setting is [5 µs/div]. The recording length can be increased by optional internal memory expansion and by limiting which channels are used, so the maximum time cannot be definitively specified. A recording length of 10,000 divisions at 5 µs/div is equivalent to 0.05 seconds recording time. "4.2.4 Setting the Recording Length (number of divisions)" in the Instruction Manual To observe phenomena such as waveform anomalies prior to a trigger event From the Trigger Settings screen, enable Pre-Trigger recording. (Set the position of the trigger point as a per- centage of the recording length, where 0% represents the recording start point and 100% represents the recording end point) "6.5 Pre-Trigger Settings" in the Instruction Manual If “Storage Off” appears Verify that an input module is installed for the selected channel. If installed: On the Status Settings screen, verify that the channel to be used for measurement is enabled [On]. If not installed: Select an installed module. "4.2.1 Selecting Channels to Use" in the Instruction Manual

Appendix 3 Common Questions

A9 How long will recording con- tinue when the recording length is set to [Cont] with the Recorder function? When trigger criteria are met, recording starts and con- tinues until you press the STOP key. Up to 5,000 divi- sions of data (with Model 9715 Memory Board installed) from measurements prior to the stopping of recording is retained in internal memory. When Auto Print or Auto Save is enabled, the data ac- quired by the instrument is printed or saved on each such occasion. About memory capacity and re- cording length: "4.2.4 Setting the Recording Length (number of divisions)"; "Setting Continuous Recording (Cont)" in the Instruction Manual What is the difference between the Memory and Recorder functions? Which function is selected determines how the timebase is set. The sampling method for each function is also dif- ferent. Because the purpose of the Memory function is acquisi- tion of instantaneous waveforms, if you set a slow time- base for long-term recording, the sampling period may be unsatisfactorily lengthened. On the other hand, the Recorder function samples and records maximum and minimum values at the specified sampling rate regard- less of the timebase setting. As a result, however, only slow timebase settings are available. "Function Selection" (⇒ p. 29) in this manual, "4.1 Selecting the Function" in the Instruction Manual In this case Description Reference Modifying the Screen Display, and During Measurement In this case Description Reference To remove a Settings dialog from the display Move the cursor to the dialog, and press the ESC key. If channel waveforms are overlapping and hard to see Channel waveforms can be displayed separately by shifting the zero position of each waveform, or by sepa- rating the waveform traces on the Sheet Settings screen. "7.2.4 Splitting the Display Screen (Split-Screen)", "7.2.6 Assigning Display Channels to Graphs (Analog Channels)", and "7.1.2 Setting the Wave- form Display Position (Zero Po- sition)" in the Instruction Manual The grid lines (graticule) causes waveforms to be hard to see, or you want to display the grid To remove the grid, set the [Grid Type] on the Environ- ment Settings screen of the System screen to [Off]. The grid is displayed when [Dotted Line] or [Solid Line] is selected. (Default setting: [Dotted Line]) "12.1.1 Selecting the Grid Type" in the Instruction Manual Waveform traces are too thick Waveform traces appear thick when noise is present on the signal. Enable the low-pass filter to suppress the noise. "3.10.3 Low-Pass Filter (LPF) Settings" in the Input Module Guide Logic waveforms are not dis- played On the [Logic] page of the Sheet Settings and Channel Settings screens, set the channels to be displayed to [On]. If “Storage Off” is displayed: Set the logic channel to [On] on the [Use Ch] page of the Status Settings screen. "4.2.1 Selecting Channels to Use" and "7.3 Displaying Logic Waveforms" in the Instruction Manual A waveform is continuously acquired even when trigger criteria have been set Is the Trigger Mode set to [Auto]? (Continuously acquires waveforms) Select [Single] to record only once when trigger criteria are satisfied, or select [Repeat] to record repeatedly. "6.3 Setting the Trigger Mode" in the Instruction Manual Cursor values and waveforms are overlapped and hard to see Press the DISP key to display waveforms and numerical values separately. "8.8 Cursor Values" in the Instruction Manual

Appendix 3 Common Questions

A10 “Trigger Wait” is displayed even though triggering is not enabled Is the time trigger enabled [On] or another trigger source enabled? Verify settings on the Trigger Settings screen. “Storing” appears, but no waveform is displayed Is the Roll Mode function (on the Status Settings screen) of the Memory function disabled (set to [Off])? If set to [On] or [Auto], waveforms are displayed during acquisition. "4.3.1 Displaying Waveforms During Recording (Roll Mode)" in the Instruction Manual When “Trigger Wait” is dis- played and no waveform appears No waveform is displayed unless the trigger criteria are satisfied. To verify the current input waveform, change the trigger mode of the Waveform screen to [Auto], and the wave- form will appear. However, when the trigger mode is [Auto], if no trigger occurs within one second after pressing the START key, recording starts anyway. (No trigger mark is applied) "6.3 Setting the Trigger Mode" in the Instruction Manual To hide unwanted channel waveforms From the Waveform screen, press the FUNCTION MODE key to change the FN mode, and press F5 [Channel Set]. (The Channel Settings dialog appears) Move the cursor to [Col](color) for the channel to be hid- den, and select F1 [Off]. "5.7 Setting Input Channels from the Waveform Screen" in the Instruction Manual In this case Description Reference Saving and Loading In this case Description Reference If you intended to save over a long period using Auto Save, and only saved a short period New directories are created automatically when Auto Save is [On], the [Normal Save] method is selected, and Directory Creation is [On]. (Up to 5,000 files can be saved in each directory) If these settings are [Off], saving stops after writing the maximum number of files that one directory can hold. "10.3.4 Setting Auto Save" in the Instruction Manual Auto Save seems to be enabled, but data is not saved to the storage media • Verify that write-protection on the storage media is disabled. If write-protection is enabled, disable it. • Verify that Auto Save is [On] before starting measure- ment. • When stopping measurement manually, note the dis- played messages, and do not interrupt auto-saving while writing to the storage media. "10.3.4 Setting Auto Save" in the Instruction Manual Only short waveforms are saved Is the save area set to [A-B]? To save the whole waveform, select [Whole]. "10.3.5 Setting Manual Save (SAVE Key Output)" in the Instruction Manual When storage media becomes full and you want to continue saving to another storage media Specify as Media1 the storage media to which you want to save first. For [Save on Error], set [Save in 2] and specify the storage media on which to continue saving. "10.3.4 Setting Auto Save" in the Instruction Manual To save to a shared folder on a PC, or to load data When a folder on a PC is registered for sharing over the LAN, it can be accessed from this instrument. "10.1.6 Using a Network Shared Folder" and "Chapter 13 Communications Settings" in the Instruction Manual

Other Data Loggers Manuals