Hach POLYMETRON 9125 Operator's Manual
Read Online
Read the Hach POLYMETRON 9125 Operator's Manual online using the built-in PDF viewer.
Text of the manual
221=191=025 - Revision L - 23/02/2007
По вопросам продаж и поддержки обращайтесь:
Астана +7(77172)727-132, Волгоград (844)278-03-48, Воронеж (473)204-51-73, Екатеринбург (343)384-55-89,
Казань (843)206-01-48, Краснодар (861)203-40-90, Красноярск (391)204-63-61, Москва (495)268-04-70,
Нижний Новгород (831)429-08-12, Новосибирск (383)227-86-73, Ростов-на-Дону (863)308-18-15,
Самара (846)206-03-16, Санкт-Петербург (812)309-46-40, Саратов (845)249-38-78, Уфа (347)229-48-12
Единый адрес: [email protected]
Веб-сайт: hach.nt-rt.ru
Transmitter 9125 – Conductivity / resistivity measurement
621=191=025 Issue L
Hach Ultra Analytics
11/2005
This instrument conforms to the European Directives:
- 89/336/CEE modified by the directive 93/68/CEE
- 73/23/CEE modified by the directive 93/68/CEE
Warning !
There are no user-serviceable parts in either the transmitter or sensor. Only Hach
Ultra Analytics personnel or their authorized representative should attempt repair of
the system and only components expressly approved by the manufacturer should be
used. Any attempt to repair the instrument in contradiction of these guidelines may
result in damage to the instrument and injury to the person making the repair. It will
also void the warranty and may compromise the safe operation, electrical integrity or
CE compliance of the instrument.
Note:
This equipment has been tested and found to comply with the limits for Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his
own expense.
Precautionary Labels :
Read all labels and tags attached to the instrument. Personal injury or damage to
this instrument could occur if not observed.
This symbol, if noted on the instrument, references the instruction
manual for operation and / or safety information.
Electrical equipment marked with this symbol may not be disposed of in
European public disposal systems after 12 August of 2005. In conformity
with European local and national regulations (EU Directive 2002/96/EC),
European electrical equipment users must now return old or end-of life
equipment to the Producer for disposal at no charge to the user.
Note : For return for recycling, please contact the equipment producer
or supplier for instructions on how to return end-of-life equipment for
proper disposal.
Important document. Retain with product records.
!
Transmitter 9125 – Conductivity / resistivity measurement
Page 1
Table of contents
1. General presentation of the conductivity / resistivity measure
system ................................................................................................. 3
Presentation of the 9125 transmitter ................................................................................3
Introduction.......................................................................................................................4
Conductivity measurement...............................................................................................4
Characteristics..................................................................................................................6
2. Installation of the transmitter.......................................................... 11
Unpacking of the transmitter 9125 .................................................................................11
Advice for installation......................................................................................................11
Dimensions.....................................................................................................................12
Mounting types ...............................................................................................................13
3. Electric connections ........................................................................ 17
Electronic board lay-out in the 9125 transmitter enclosure ............................................17
Main connection .............................................................................................................20
Relays connections ........................................................................................................21
Sensor connections........................................................................................................22
4. Using the 9125 transmitter .............................................................. 23
Utilization rules for the menus ........................................................................................23
Modification of a value....................................................................................................24
Measures display............................................................................................................24
Main display....................................................................................................................25
Display options ...............................................................................................................25
Display options ...............................................................................................................26
Choice of the language.....................................................................................26
S/DISPLAY Menu .............................................................................................26
UNIT choice....................................................................................................................27
TDS choice.....................................................................................................................27
Temperature UNIT choice ..............................................................................................27
Concentration measure (TDS)........................................................................................28
Limits of the TDS measure.............................................................................................28
5. Programming the transmitter.......................................................... 29
Main menu......................................................................................................................29
Calibration Menu ............................................................................................................30
Conductivity calibration.....................................................................................30
Select to "Programming" and press enter. ...........................................30
Type of calibration ............................................................................................31
2 point calibration .................................................................................32
Transmitter 9125 – Conductivity / resistivity measurement
Page 2
1 point calibration .................................................................................32
Execution of a 2 point calibration .........................................................33
TDS measure calibration ..................................................................................34
Temperature electric calibration / Resistor adjust ................................35
Process calibration ...............................................................................35
PARAMETERS Menu.....................................................................................................36
HISTORIC Menu ............................................................................................................36
MAINTENANCE Menu ...................................................................................................37
PROGRAMMING Menu..................................................................................................38
S/MEASURE Menu ..........................................................................................39
PROBE.................................................................................................39
TEMPERATURE COMPENSATION ....................................................40
TDS coefficient adjusting......................................................................40
S/mA OUTPUTS Menu.....................................................................................41
OUTPUT 1/2.........................................................................................42
S/ALARMS menu .............................................................................................43
ALARM 1/2 (limit) .................................................................................44
ALARM 1/2 (USP) ................................................................................44
USP mode ............................................................................................45
ALARM 3 (Alarm system).....................................................................47
ALARM 4 (Timer)..................................................................................47
S/SPECIAL PROG. Menu ................................................................................48
TEST ....................................................................................................48
S/RS485 Menu .................................................................................................49
SERVICE Menu..............................................................................................................50
S/POLARIZATION Menu..................................................................................51
CABLE CAP. ........................................................................................51
TEST ....................................................................................................51
S/AVERAGE Menu...........................................................................................52
S/CODE Menu..................................................................................................53
S/SOFT ISSUE Menu.......................................................................................54
S/DEFAULT/VALUES Menu.............................................................................54
S/mA ADJUST Menu........................................................................................55
S/FACTORY Menu ...........................................................................................55
6. Polarization ....................................................................................... 57
Electric representation of the probe and its cable ..........................................................57
Frequency adjustment according to the conductivity measurement ..............................58
Automatic adjustment of the frequency..........................................................................59
7. Error messages ................................................................................ 61
Appendix A : Default values.................................................................A1
Appendix B : Spare parts list ...............................................................B1
Transmitter 9125 – Conductivity / resistivity measurement
Page 3
1.
General presentation of the conductivity /
resistivity measure system
Presentation of the 9125 transmitter
The 9125 transmitter and associated measuring
sensors has been designed for measuring and
continuous control of conductivity and resistivity (with
possibility of temperature measurement) in industrial
process.
Note :
The programming is displayed in 6 languages. To
modify this parameter see § 4 " Display Menu –
Choice of the language ".
Transmitter 9125 – Conductivity / resistivity measurement
Page 4
Introduction
The 9125 transmitter is a user-friendly instrument
(installation,
programming),
equipped
with
a
microprocessor it can be configured to correspond to
any application in the following sectors :
�
drinking water,
�
waste water,
�
process (chemistry, paper mills, sugar mills...),
�
measurement in pure/ultrapure water (energy
power plants, semiconductor industry, chemistry).
The 9125 transmitter should be connected to a probe
via a cable.
Conductivity measurement
The electric conductivity measures the transport of
electric charges in any field. In metal conductors, the
current flows by transport of electrons, whereas in
solutions, it flows by transport of ions such as Na+ and
Cl- which ensure the transport of charges. The higher
the transport of charges is, the greater is the
conductance of the solution.
Conductivity is the capacity a solution has to
conduct current.
In solution, conductivity is much more complicated
than in conductors because several species ensure
the transport of charges. For instance, in drinking
water the conductive species registered are sodium,
calcium,
magnesium,
ferrous
cations,
ferrites,
phosphates and nitrate ions. For slightly concentrated
solutions, the concentration of H+ protons and hydroxyl
OH- ions (stemming from the weak dissociation of the
water [H+] = [OH-] = 10-7 mol/l to 25°C) can no
longer be neglected in the presence of the product,
this
therefore
leads
to
a
non-linear
variation
Conductivity/Concentration.
Transmitter 9125 – Conductivity / resistivity measurement
Page 5
Mobility of these species in an electric field depends
naturally on their size, weight, transport charge,
viscosity of the field. The greater the concentration of
the species is, the greater the interactions between
these ionic species is.
Principle of electrolytic conductivity
Ohm’s law specifies that the current circulating in the
dipole is proportional to the difference in potential and
resistance of this dipole :
I = E / R
The resistance of a homogenous environment
depends
on
the
geometry
of
the
resitivity
(characteristic of the material) :
R = r . 1/K
Where "r" is the resistivity in Ohm.cm and "K", the cell
constant in cm-1.
Transmitter 9125 – Conductivity / resistivity measurement
Page 6
Characteristics
The 9125 is equipped with an input measurement
channel : a conductivity probe, 2 electrodes and
inductive may be connected as well as a temperature
Pt100 or Pt1000 probe.
The 9125 is also equipped with 2 analogue outputs (0
or 4-20 mA).
Options available on request
�
board with 4 relays
�
RS485 board
MAIN SPECIFICATIONS
Package
Delivered with instruction manual, 4 cable glands and
2 mounting screws and a specifications conformed
certificate
Maintenance
No particular maintenance required. Clean the
instrument with a soft tissue and without any
aggressive agent
OPERATING CONDITIONS
Ambient temperature
-20°C...+60°C
Relative humidity
10...90%
Power supply voltage
fluctuation
� 10 %
Over voltage category
2
Pollution degree
2 (as CEI 664)
Altitude
< 2000 m
Measurement category
I (overvoltage less than 1500 V)
ELECTRIC CHARACTERISTICS
Power supply voltage
� Standard version (� 10 %) :
- 100 V ... 240 VAC 50/60 Hz
� Low voltage version:
- 13...30 VAC 50/60 Hz
- 18...42 VDC
Transmitter 9125 – Conductivity / resistivity measurement
Page 7
Connections
2,5 mm2 screw terminals
Fuse
5 x 20 mm Cartridge
T2AL - 250 V
Consumption
25 VA
European standards
EN 61326-1997 and EN61326 A1-1998 (Industrial level
for immunity)
EN- 61010-1
UL and CSA agreement
File E226594
MECHANICAL CHARACTERISTICS
Dimensions
144 x 144 x 150 mm
Weight
2 Kg
Material
Housing : Polyester coated aluminium
Screws : stainless steel
Tightness
IP65
Mounting types
Wall
Pipe
Panel
Cable glands
2 x PG13.5
2 x PG11
PERFORMANCE
2 electrode probe
Cell constant
0,01 cm-1
0,1 cm-1
1 cm-1
Conductivity range
0,01 �S/cm...200 �S/cm
0,10 �S/cm...2 mS/cm
1 �S/cm....20 mS/cm
2 electrode probe
Cell constant
0,01 cm-1
0,1 cm-1
1 cm-1
Resistivity range
5 k�.cm...100 M�.cm
0,5 k�.cm...10 M�.cm
50 �.cm...1 M�.cm
Inductive probe
Cell constant
1 cm-1
2,35 cm-1
10 cm-1
Conductivity range
100 �S/cm...1 S/cm
200 �S/cm...2 S/cm
1 mS/cm...10 S/cm
Transmitter 9125 – Conductivity / resistivity measurement
Page 8
Inductive probe
Cell constant
1 cm-1
2,35 cm-1
10 cm-1
Resistivity range
1�.cm...10 k�.cm
0,5 �.cm...5 k�.cm
0,1 �.cm...1 k�.cm
Ambient temperature
range
-20°C...+200°C (-4…392 °F)
Display resolution
Conductivity/resistivity : automatic point drift (min.
resolution 0.001 �S/cm)
< 0.1 °C
Accuracy
Inductive :
K=1 :
� 2% of the displayed value or
��0,002 mS
K=2,35 :
� 2% of the displayed value or
��0,004 mS
K=10 :
� 2% of the displayed value or
��0,02 mS
2 electrode : � 1% of the displayed value
temperature < � 0.4 °C
Temperature sensor
Pt 100 / Pt 1000
Temperature
compensation
- No
- Automatic
- Manual
Automatic temperature
compensation range
-20 ... 200 °C
-4 ... 392 °F
Temperature compensation
range
Linear
Non linear :
- ultrapure water, HCl and NaCl
Sensor types
- 2 electrode sensor
- Inductive sensor
Cable length
100 m maximum
CALIBRATION
Conductivity calibration
type
- Electric
- 2 points
- 1 point
Slope matching
50 ... 150 %
Transmitter 9125 – Conductivity / resistivity measurement
Page 9
Temperature calibration
� 20 °C ( � 36 °F)
ANALOGUE OUTPUT
Output signals
2 galvanicly outputs insulated
Allocation
Conductivity/resistivity/temperature
Type
0 ... 20 mA
4 ... 20 mA
Mode
Linear
Dual
Logarithmic
Maximum load
800 �
Accuracy
0.1 mA
ALARMS
Alarm number
4
Function
- Standard limits
- Limits according to USP standard
- Alarm system
- Timer
Hysteresis
0 ... 10%
Temporisation
0 ... 999 s
Breaking power
(resistive load)
250 V AC, 3A max
100 V DC, 0,5A max
Use a cable (rated 105°C and AWG22 to 14). The
external cable insulation should be cut as close as
possible from the terminal block.
RS485
Baud rate
300 ... 19200 bauds
Insulation
Galvanic
Station number
32 max
Transmitter 9125 – Conductivity / resistivity measurement
Page 10
�
PROGRAMMING
Language
French
English
German
Italian
Spanish
Dutch
Display
Icones + graphic zone (80*64 pixels)
Protection codes
Calibration
Programming
Service
Transmitter 9125 – Conductivity / resistivity measurement
Page 11
2.
Installation of the transmitter
Unpacking of the transmitter 9125
Inspect the package at the reception to detect an
eventual damage due to the transport. Make sure the
package contents are not damaged.
Check if the package corresponds to your order :
�
quantity delivered,
�
type of instrument and version accordingly to the
instruction plates,
�
accessories : 4 cable glands and 2 mounting
screws + flanges,
�
instruction manual,
�
certificate of conformity to specifications.
Advice for installation
Choose a site where :
�
vibrations are not too excessive,
�
supply relays or commuters are away,
�
maintenance will be easy.
Note :
Information : It is preferable to mount the
instrument
above
eye
level,
allowing
an
unrestricted view of the front panel displays and
controls.
Transmitter 9125 – Conductivity / resistivity measurement
Page 12
Dimensions
(Dimensions are in mm [inches]).
Fig. 2.1 Transmitter
9125
dimension
Transmitter 9125 – Conductivity / resistivity measurement
Page 13
Mounting types
3 possibilities to mount the instrument (use of the red
clamping bow) :
The transmitter housing conforms to norm DIN 43700.
Panel mounting :
Panel cutting : 138 x 138 mm
Front panel dimensions : 144 x 144 mm
�
2 screws � 4 mm lg 16 flat head (provided) for
panel thickness 0 to 4 mm
�
2 screws � 4 mm lg 20 flat head (provided) for
panel thickness 4 to 8 mm
Fig 2.2 Panel mounting
Panel cutting
138 x 138 mm
(5.4 in. x 5.4 in.)
Front panel dimensions
144 x 144 mm
(5.8 in. x 5.8 in.)
Thickness panel
Inferior to 8 mm
Transmitter 9125 – Conductivity / resistivity measurement
Page 14
Wall mounting :
�
2 screws � 4 mm lg 60 flat head (not
provided)/ 80 mm center distance
Fig. 2.3 Wall mounting
Transmitter 9125 – Conductivity / resistivity measurement
Page 15
�
Pipe mounting :
�
� 2" maximum - 2 screws � 4 mm lg 60
(provided)
Fig. 2.4 Vertical mounting
Fig. 2.5 Horizontal mounting
Transmitter 9125 – Conductivity / resistivity measurement
Page 17
3.
Electric connections
Electronic board lay-out in the 9125 transmitter
enclosure
________________________________________________________________________
Fig. 3.1 Electronic board lay-out
1.
Terminal block 4-20 mA
2.
CPU board
3.
RS485 board (option)
4.
Relay board (option)
5.
Choice between an inductive (all 4 switches on position I) or a 2 electrode
(Kohlraush) probe (all 4 switches on position K)
6.
Programmed EEPROM
7.
Conductivity module
8.
Power supply board
9.
Program update connector
7�
1�
2�
3�
4�
5� 6�
8�
9�
Transmitter 9125 – Conductivity / resistivity measurement
Page 18
________________________________________________________________________
Fig. 3.2 9125 shielded plate
Electric connections are realised on the terminal inside
the housing. Put the cables into the appropriated
openings.
�
The main supply and relay cables should be
dispatched via the openings behind the shielded
plate. To remove the plate, unscrew the fixing
screw on the left side of the plate.
�
Sensor
and
mA
output
cables
should
be
dispatched via the openings provided on top of the
shielded plate.
�
Check the creeping of the cables when opening
the transmitter.
�
It is required to use shielded cables. The shielding
should be connected to the earth central shielding.
80°.
Transmitter 9125 – Conductivity / resistivity measurement
Page 19
________________________________________________________________________
Fig. 3.3 Power and relays connections
1. Terminal block connections
Analogue outputs
4-20 mA
2. Terminal block connections
"option RS 485"
3. Terminal block connections
of the conductivity sensors
4. Earth
1
3
2
4
4-20 mA
Sensor
Transmitter 9125 – Conductivity / resistivity measurement
Page 20
Main connection
Electrical connection should be performed only by
qualified personnel. For the base model, the power
supply accepts 100-240 VAC � 10 %, (50/60 Hz)
without changes in configuration. Before switching the
transmitter, make sure the site voltage corresponds to
the instrument voltage indicated on the identification
plate. The terminal block for power connections can be
lifted from its header for easier installation.
For safety reasons, it is required to observe the
precautions below :
�
Use a three wire mains supply cable (2 core + PE)
with a cross section between 0.35 and 2 mm²
(AWG 22 to 14) rated at 105°C minimum. The
external cable insulation should be cut as close as
possible from the terminal block.
�
The instrument should be connected to the power
supply by means of a breaker located close to the
instrument and be identified. The supply shall be
fitted with an overcurrent protection device rated at
20 Amp maximum.
�
This breaker should switch off phase and neutral in
case of electrical problems or when the user wish
to service the instrument. However the power
supply earth must always be connected.
�
Cabling should be specified for a minimum of 80°C
(176°F).
Note :
Before servicing the instrument, ensure that the
power supply is switched off.
Transmitter 9125 – Conductivity / resistivity measurement
Page 21
Relays connections
The 9125 is equipped with 4 relays. Relay S4may be
configured as a temporized relay. The nominal value
of the cutoff current of each relay is 2A for 250 Vca or
0,5 A for 100 Vcc. Cabling should be specified for a
minimum of 80°C (176°F). See figure 3.1 page 17 for
connection location.
�
S1 is located on the upper part of the terminal and
S2, S3 and S4 under. Each relay is connected to a
2 separated contact terminal, which may be
removed to facilitate the installation.
�
The relay operation is configured in the software
but the relay switches are always open when the
unit is switched off.
Output current connections (mA output)
The transmitter has two analogue outputs, which may
be set in 0-20 or 4-20 mA and which are galvanicly
insulated from the controller. Maximum load for each
output is 800 �. See Figure 3.1 page 17 for the
terminal location.
�
Use a signal cable with shielded twisted pair with
the earth shielding in the transmitter.
�
Connect the cable to the terminal according to the
drawing on the shielded plate.
Transmitter 9125 – Conductivity / resistivity measurement
Page 22
Sensor connections
The conductivity sensors have a double shielding, the
first one is connected the CPU board, the second one
(external) is connected to earth on the shielded plate.
Execute the connections according to the following
table :
Kohlrausch
Inductives
TEMP+
blue
green
TEMP-
black
yellow
IN
white
white
OUT
red
brown
GND
internal
shield
internal shield
(X2)
external
shield
external
shield
�
Use only Hach Ultra Analytics supplied cables.
Using other types of cables does not ensure the
conformity to the electromagnetic compatibility
standards.
Transmitter 9125 – Conductivity / resistivity measurement
Page 23
4.
Using the 9125 transmitter
Utilization rules for the menus
The user interface of the 9125 transmitter is made of a
display screen and 4 keys.
The (Esc) key is used to go back to the previous
menu.
The (Enter) key is used to validate the selections and
the data.
Both middle keys, right and left function keys, are
defined according to the words and symbols which are
displayed above each function key.
Modify a parameter
Select
Choose a menu
Main
Go back to the main display principal
Menu
Display the main menu
Disp2
Display screen 2
Disp3
Display screen 3
OK
Validate the measure during a calibration
-
Increase a value
+
Decrease a value
26.8
mS/cm
25.0°C
Disp2
Menu
S1
S2
S3
S4
Transmitter 9125 – Conductivity / resistivity measurement
Page 24
Modification of a value
The highlighted digit may be modified with the key
. Each digit can be validated by pressing
ENTER. Repeat both operations for each digit.
Note :
� If you do not use the display for at least 10
minutes,
the
instrument
returns
to
the
measuring mode except for the calibration and
maintenance mode.
� An access code may be required for the
calibration, programming and service menu (see
CODE menu).
Measures display
Measures display allow to display measures and state
of the device. There are three :
Reference
Description
�
Main display
�
Alarm state
�
Function keys
26.8
mS/cm
25.0°C
Disp2
Menu
S1
S2
S3
S4
�
�
�
�
�
Transmitter 9125 – Conductivity / resistivity measurement
Page 25
Main display
�
�
�
�
�
�
�
�
26.8 mS/cm : conductivity measurement
25.0 °C : temperature measurement
S1...S4 : alarm status (invisible if alarm inactive).
Display 2
S1...S4 : alarm status
S1 : activated by a temperature > 30.0°C
S2 : inactive
: S3 in alarm system
S3 closed by a temperature < - 20.0°C
S4 : not used
Display 3
Analogue outputs allocation:
measure or temperature.
Display of each output value with a bargraph + mA
indication.
26.8
mS/cm
25.0°C
Disp2
Menu
S1
S2
S3
S4
S1 :
S2 : OK
:
S4 : Not used
Disp3
Main
S1
S3
°C > 30.0
°C < - 20.0
�
I1 : Measure
I2 : Temp.
6.7 mA
Main
Transmitter 9125 – Conductivity / resistivity measurement
Page 26
Display options
Choice of the language
English is the default language. You can choose an
other language available (French, German, Italian,
Spanish or Dutch) by following the procedure below :
�
Use the right function key MENU.
�
Use the left function key (Select) to select the
menu SERVICE and press (Enter).
�
In the menu SERVICE, use the left function key
(Select) to select DISPLAY and press (Enter).
�
Select the language of your choice with the right
function key. ”Press (Enter)”.
S/DISPLAY Menu
� UNIT :
choice
of
the
display
of
conductivity/resistivity measurement.
� S/cm
� �.cm
� S/m
� �.m
� TEMP. :
choice
of
the
temperature
measurement.
� °C
� °F
� LANGUAGE : choice of the message language.
�
French,
�
English,
�
German,
�
Spanish,
�
Italian,
�
Dutch.
� Press Esc to go back to the DISPLAY menu.
�
CALIBRATION
MAINTENANCE
PROGRAMMING
SERVICE
MENU
Select
SERVICE
�
POLARIZATION
AVERAGE
CODE
SOFT ISSUE
DEFAULT VAL.
Ma ADJUST
FACTORY
SERVICE
Select
DISPLAY
�
UNIT
: S/cm
TDS
: NO
TEMP.
: °C
LANGUAGE :
DISPLAY
Select
GB
Transmitter 9125 – Conductivity / resistivity measurement
Page 27
UNIT choice
As the language used, it’s possible to choose the units
of the measures in which they will be display.
� Unit. allows to choose one of the units in
conductivity or resistivity in which values will be
done.
�
S/m
�
�.cm
�
S/cm
�
�.m
TDS choice
� TDS : yes/no.
When the option TDS is selected (Yes), the
measurements will be displayed in concentration
units : in ppt (pare per thousand), ppm (part per
million) ppb (part per billion) according to the
measurement level.
When the option TDS is not selected (No),
measurements are displayed in �.cm or S/cm
according to the unit selected ; Menus for
calibration of the TDS measurement are not
displayed anymore.
Temperature UNIT choice
Allow to choose a temperature display between
Celsius or Fahrenheit degree.
�
UNIT
: S/cm
TDS
: NO
TEMP.
: °C
LANGUAGE : GB
DISPLAY
Select
S/cm
�
UNIT
: S/cm
TDS
:
TEMP.
: °C
LANGUAGE : GB
DISPLAY
Select
YES
�
UNIT
: S/cm
TDS
: YES
TEMP.
: °C
LANGUAGE : GB
DISPLAY
Select
°C
Transmitter 9125 – Conductivity / resistivity measurement
Page 28
Concentration measure (TDS)
The Total Dissolved Solid (TDS) measure is a
concentration measure in the limits of measure where,
for a part given, the concentration of the solution is
proportional to its conductivity.
The next board show for the mains solutions, the
relations
existing
between
concentration
and
conductivity (in ppm).
The white part of the board show the limits where the
TDS display is possible.
The measure in concentration (TDS) is given by the
formula :
TDS = Cond x KTDS
The KTDS coefficient is calculated during the process
calibration of the device. It must be between 0.00 and
5.00. If it is not in this interval, an error message will
display during the calibration.
Limits of the TDS measure
For different substances, the limits of the measure
where the concentration is proportional to the
conductivity is given by the white part of the next
board:
The unit of the conductivity is μs/cm.
Weight % ppm
NaCl
NaOH
NH4OH
NH3
HCl
H2SO4
HNO2
HF
SO2
Acid
Acetic
0,0001
1
2,2
6,2
4,1
6,6
11,7
8,8
6,8
10
6,4
4,2
0,0003
3
6,5
8,3
8,3
12
50
61
20
30
18
7,4
0,001
10
21,4
61,1
17
27
116
85,6
67
99
54
15
0,003
30
64
182
31
49
340
251
199
290
150
30,6
0,01
100
210
603
58
84
1140
805
657
630
450
63
0,03
300
617
1780
102
150
3390
2180
1950
1490
1200
114
0,1
1000
1990
5820
189
275
11100
6350
6380
2420
3600
209
0,3
3000
5690
16900
329
465
32200
15800
18900
5100
7900
368
1
10000
17600
53200
490
810
103000
48500
60000
11700
17200
640
3
30000
48600
144000
790
1110
283000
141000
172000
34700
32700
1120
5
50000
78300
223000
958
1115
432000
237000
275000
62000
42000
1230
10
100000
140000
358000
1115
1120
709000
427000
498000
118000
61000
1530
20
200000
225000
414000
968
4251
850000
709000
763000
232300
1600
30
300000
292000
725
732000
828000
861000
390000
1405
40
400000
191000
460
770000
820000
1080
50
500000
150000
620000
717000
740
75
750000
182000
340000
168
100 1000000
10000
50000
1
Transmitter 9125 – Conductivity / resistivity measurement
Page 29
5.
Programming the transmitter
Main menu
The main menu gives access to 4 main functions of
the instrument :
�
The CALIBRATION menu enables to adjust the
instrument
measurement
according
to
the
reference measurements.
�
The MAINTENANCE menu enables to intervene
on the instrument.
�
The PROGRAMMING menu enables to program
the instrument according to the application.
�
The SERVICE menu is reserved to qualified
servicing personal.
�
CALIBRATION
MAINTENANCE
PROGRAMMING
SERVICE
MENU
Select
Transmitter 9125 – Conductivity / resistivity measurement
Page 30
Calibration Menu
Note :
Before any calibration launching, check the
parameters of the MEASURE menu (probe type,
temp. Comp.) Are correctly configured.
(See page 25).
This option allows to calibrate the conductivity
measure, the temperature measure and to display the
concentration measure (TDS).
Conductivity calibration
With the help of the « select » function touch, choose
COND. CALIB. option. Two options appears :
�
PROGRAMMING allows to choose the electric
conductivity calibration type :
� 2 points
� 1 point.
�
EXECUTION will allow to do the desired
calibration type.
Select to "Programming" and press enter.
Select the calibration type and press enter.
�
TDS CALIB.
COND. CALIB.
TEMP.CALIB.
PARAMETERS
HISTORIC
CALIBRATION
Select
�
EXECUTION
PROGRAMMING
COND. CALIB.
Select
�
EXECUTION
PROGRAMMATION
COND. CALIB.
Select
PROGRAMMING
�
ETALONNAGE
MAINTENANCE
PROGRAMMING
SERVICE
MENU
Select
CALIBRATION
Transmitter 9125 – Conductivity / resistivity measurement
Page 31
Type of calibration
�
With a 2 electrode probe
For the first point, remove the probe from liquid or
unscrew the connector from the probe.
For the second point connect a resistance to the
IN/OUT terminal of the conductivity module.
�
With an inductive probe
For the first point, remove the probe from liquid.
For the second point connect a resistance with a
looped wire through the probe.
�
TYPE : Elect.
R = 200 �
COND. CALIB.
Select
Transmitter 9125 – Conductivity / resistivity measurement
Page 32
2 point calibration
For the first point, remove the probe from liquid or
unscrew the connector from the probe (2 electrode
probe) in a known concentration solution. The user
enters this solution value when calibrating.
1 point calibration
Immerse the probe in a known concentration solution.
The user enters this solution value when calibrating.
Note :
Warning ! This calibration is only active on the
slope
The first point of the electric calibration and 2 point
calibration enables to realise an internal electronic
calibration and to measure the cable capacity with a
2 electrode probe.
It is mandatory to realise one of both calibrations
when starting the MONEC D9125 for the first time..
For the second calibration point use a resistance or a
solution with a significant difference from the first point.
�
TYPE :
COND. CALIB.
Select
1 pt
�
TYPE :
COND. CALIB.
Select
2 pts
Transmitter 9125 – Conductivity / resistivity measurement
Page 33
EXECUTION
Execution of a 2 point calibration
Program the manual 2 point calibration and execute it
as follows :
�
Remove the electrode from liquid.
�
When the measure is stable, the instrument goes
automatically to next step.
The symbol
flashes 10 to 20 seconds.
�
Press ENTER to accept the zero calibration.
�
Immerse the probe in the calibration solution
�
When the measure is stable, press OK.
�
Change the value displayed if you want.
�
Modify the date of the calibration if necessary by
pressing on the right function touch
.
To validate the calibration parameters press
ENTER. ESC does not validate the parameters
and the former calibration parameters are kept.
�
EXECUTION
PROGRAMMING
COND. CALIB.
Select
EXECUTION
�
PROBE OUT
Press
ENTER
COND. CALIB.
Transmitter 9125 – Conductivity / resistivity measurement
Page 34
TDS measure calibration
This option is available only if you have selected the
measurement TDS in the menu DISPLAY.
Select with the function key Choice CALIB TDS in the
menu CALIBRATION and press ENTER.
This calibration permits to adjust the coefficient K
TDS
between
the
concentration
value
and
conductivity.
The measurement TDS is displayed as well as the
conductivity value as secondary measurement.
Proceed to the following steps :
�
Press the function key OK.
�
Change the displayed value and enter the sample
value (TDS).
�
The instrument indicates the result of the
coefficient calculation TDS and press ENTER.
Temperature calibration
With the help of the right function touch choice, select
the programming line and press ENTER.
The right function touch
allow to select 2
types of calibration, electric or process.
�
6.98
ppm
9.90 K�
CALIBRATION
�
6.98
ppm
06.98 ppm
CALIBRATION
0
Transmitter 9125 – Conductivity / resistivity measurement
Page 35
Temperature electric calibration / Resistor adjust
This calibration is factory realised.
Used to realise an electric calibration for the Pt100
measurement. 2 resistances of known values should
be connected to the temp + and temp - of the module
measure. This resistors should have a precision about
0,1%.
Proceed as follows :
�
Launch the calibration when you have configured
an electric calibration. Lancer l'exécution après
avoir
configuré
l'étalonnage
en
étalonnage
électrique.
�
The transmitter requires the connecting of the first
resistance.
�
Press Enter.
�
Repeat the same procedure for the second
resistance.
Process calibration
Proceed as follows after you have configured a
process calibration :
�
Wait till the measurement is stable and press the
right function key OK.
�
You have the possibility to change the value
�
Press Enter.
�
The instrument executes a zero adjustment
required to display the value configured.
�
RESISTANCE 1
Press
ENTER
TEMP. CALIB.
�
189.9
189.9°C
TEMP. CALIB.
1
�
TYPE :
RES1 : 100 �
TEMP. CALIB
Elect.
Transmitter 9125 – Conductivity / resistivity measurement
Page 36
PARAMETERS Menu
This menu displays the calibration parameters of the
conductivity measurement.
Date of last calibration. This date is the date the user
has entered after a conductivity calibration ( electric, 2
points, 1 point).
The other parameters are those which are displayed
when the date is registered :
�
Slope of the last calibration which is a correction
factor of the conductivity probe slope.
�
Si la mesure est effectuée en TDS, le coefficient
de proportionnalité entre la conductivité et la
concentration.
�
Drift of the temperature measurement.
HISTORIC Menu
This
menu
displays
the
conductivity
slopes
corresponding to the last two conductivity calibrations
and allows to follow the probe clogging.
�
DATE
: 01/01/01
SLOPE : 100%
COEFF : 0.50
�T
: -0.0°C
PARAMETERS
�
DATE
: 01/01/01
SLOPE : 100%
HISTORIC
Transmitter 9125 – Conductivity / resistivity measurement
Page 37
MAINTENANCE Menu
When changing or cleaning a probe or servicing the
instrument, the transmitter continues to display
measures.
The analogue output value is the value programmed in
the mA menu. The relay status is not modified.
�
6.98
mS/cm
26.4°C
MAINTENANCE
Transmitter 9125 – Conductivity / resistivity measurement
Page 38
PROGRAMMING Menu
Note :
Warning ! An access code may be required if
programmed.
This menu enables the configuration of the instrument
according to its application.
In this operating mode, the measurements, the
analogue outputs and alarms remain active.
�
MESURE
SORTIES mA
ALARMES
RS485
PROGRAMMATION
Choix
MESURE
�
MESURE
mA OUTPUTS
ALARMS
RS485
PROGRAMMING
Select
MEASURE
Transmitter 9125 – Conductivity / resistivity measurement
Page 39
S/MEASURE Menu
The PROBE menu allows to configurate the utilized
probe
type,
TEMP.
COMP.
the
temperature
compensation, TDS the TDS measure coefficient
when it is actived.
PROBE
This menu allows to choose the utilized probe for the
conductivity measure.
TYPE :
- Induct.
- 2 elec.
choice of the electrode :
Inductive or 2 electrodes
Note :
Check both switches of the conductivity module are
correctly positioned :
- K : 2 electrodes
- I : inductive
K :
XXXX
Adjustment of the cell constant.
FREQ. :
- Auto
- 8000
- 4000
- 2000
- 1000
- 500
- 250
- 125
- 62,5
Choice between an automatic
adjustment
of
the
frequency
according to the measurement or
one
of
the
pre-programmed
frequencies. (see Chapter 5 for
further details).
Only if elect. 2 type has been
selected.
�
TEMP. COMP.
TDS :
�
MEASURE
Select
PROBE
�
TYPE : 2 élect or inductive
K : 001.00
FREQ. :
PROBE
2 elect.
0.50
Auto
Transmitter 9125 – Conductivity / resistivity measurement
Page 40
TEMPERATURE COMPENSATION
MEASURE :
- No
- Pt 100
- Pt1000
Choice
of
a
temperature
measurement
with
or
without
Pt100 / Pt1000.
TYPE :
- No
- Auto.
- Manual
Choice between no temperature
compensation or an automatic or a
manual temperature compensation
mode.
TEMP. :
XX
Possibility to enter the sample
temperature
in
a
manual
compensation.
TREF :
XX
Possibility to enter the reference
temperature.
COMP. :
- Coef.
- HCl
- NaCl
Possibility
to
choose
the
temperature compensation.
COEF :
XX
Possibility to enter the coefficient
value.
TDS coefficient adjusting
TDS coefficient adjusting is directly in the MEASURE
menu.
Utilize the left function touch to choose the TDS
adjusting.
TDS : Choose the TDS coefficient which is necessary
to calculate the solution concentration.
Utilize the right function touch to choose a value
between 0 and 5.
�
MEASURE :
TYPE : Manual
TEMP. : 15°C
TREF. : 025.0°C
COMP. : Coeff.
COEF. : 2.0%
TEMP. COMP.
Select
Pt100
�
PROBE
TEMP. COMP. :
TDS : 0.50
�
MEASURE
Select
0.50
Transmitter 9125 – Conductivity / resistivity measurement
Page 41
S/mA OUTPUTS Menu
This menu allows to adjust analogue outputs.
�
OUTPUT 1
OUTPUT 2
SPECIAL PROG.
TEST
mA OUPUTS
Select
Transmitter 9125 – Conductivity / resistivity measurement
Page 42
OUTPUT 1/2
� AFFECT. : choice
if
the
analogue
output
allocated to measure or temperature.
�
S/�
�
°C/°F
� TYPE :
choice of the analogue output type.
�
0/20 mA
�
4/20 mA
� MODE :
choice of operating mode : linear,
logarithmic or dual.
* In logarithmic mode , the beginning of range
should be different from 0.
�
Lin
�
Log
�
Dual
� LOWER :
limit value programming.
� MIDD. :
middle value only in dual mode.
� UPPER :
upper limit value programming.
�
AFFECT. :
TYPE : 0/20 mA
MODE : Dual
LOWER : 10.0 mS
MIDD : 15.0 mS
UPPER : 20.0 mS
OUTPUT 1/2
Select
S
Transmitter 9125 – Conductivity / resistivity measurement
Page 43
S/ALARMS menu
This sub-menu allows to reach the configuration of
alarms 1 to 4.
The MODE parameter allows to choose the operating
mode of the 4 alarms :
-
Limit : alarms 1...4
-
USP : only alarms 1 and 2
-
System : only alarm 3
-
Timer : only alarm 4
MODE :
�
No
�
Limit
�
USP
�
System
�
Timer
�
ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARMS
Select
�
MODE : Syst
ACCEPT : Manual
RELAY : NO
ALARM 3
Select
System
�
MODE : Timer
INTERV : 2440 mn
Impul. Nb : 1
Ton : 005 s
Toff : 005 s
TmA : 20 mn
ALARM 4
Select
Timer
Transmitter 9125 – Conductivity / resistivity measurement
Page 44
ALARM 1/2 (limit)
� AFFECT. :
choice of a limit on the measurement
or on the temperature.
�
No
�
S/�
�
°C/°F
� LIMIT :
value of the limit.
� DIR. :
choice of the direction :
�
Up
�
Down
� DELAY :
definition of the temporisation when
the relay is interlocking (in seconds).
� HYST :
definition of the hysteresis in % (10%
max).
� RELAY :
choice of the relays normally opened
or closed.
�
NO
�
NF
ALARM 1/2 (USP)
�
� The TEMP menu allows the temperature to be
entered in cases where no measurement is
provided by Pt100/Pt1000. If temperature is
measured
automatically,
this
menu
is
not
displayed.
� The LEVEL menu is used to set the safety margin
with respect to the USP standard: with a level of
50%, the 1.9 μS/cm threshold (50°C) drops to 0.95
μS/cm.
�
AFFECT. : Non
LIMIT : 20.3°C
DIR. : Up
DELAY : 30s
HYST. : 10%
RELAY : NO
ALARM 1
Select
°C/°F
�
MODE:
TEMP: 25.0 °C
LEVEL: 100%
TAB: P.W.
RELAY: NO
ALARM 2
Select
USP
Transmitter 9125 – Conductivity / resistivity measurement
Page 45
� The TAB menu is used to configure the stored
USP curve: W.F.I. (Water For Injection) or P.W.
(Pure Water).
USP mode
USP is a standard used in the pharmaceutical industry.
It recommends the use of a non temperature-
compensated conductivity measurements and sets the
upper limits of acceptable conductivity for USP-
compliant water according to temperature. The alarm
delivered therefore varies according to the measured
temperature. For this, the instrument possesses two
stored curves:
� a curve (W.F.I.) for conductivity measurements in
high purity water or for injectable preparations
(cf. table 1),
� a curve (P.W.) for conductivity measurements in
purified water (cf. table 2).
NB:
This function is only possible in 2 electrode
measurements
and
when
temperature
compensation has been deactivated.
T°C
Non-
compensated
conductivity
�S/cm
T°C
Non-
compensated
conductivity
�S/cm
T°C
Non-
compensated
conductivity
�S/cm
0
0.6
35
1.5
70
2.5
5
0.8
40
1.7
75
2.7
10
0.9
45
1.8
80
2.7
15
1.0
50
1.9
85
2.7
20
1.1
55
2.1
90
2.7
25
1.3
60
2.2
95
2.9
30
1.4
65
2.4
100
3.1
Table 1 - Temperature curve and conductivity requirements
for high purity water or for injectable preparations (non
temperature-compensated conductivity measurements).
Transmitter 9125 – Conductivity / resistivity measurement
Page 46
Example: if the measured temperature is equal to
22°C, the threshold is of 1.1 �S/cm and shall remain at
this value as long as the temperature stays at 22°C. If
the temperature varies, rising to 25°C, the threshold
automatically switches to 1.3 �S/cm.
NB: The delivered alarm corresponds to the value of
the threshold associated with the USP temperature
immediately below the measured temperature (i.e.
20°C and 25°C in the previous example).
T°C
Non-
compensated
conductivity
�S/cm
T°C
Non-
compensated
conductivity
�S/cm
T°C
Non-
compensated
conductivity
�S/cm
0
2.4
35
5.95
70
9.1
5
3.0
40
6.5
75
9.7
10
3.6
45
6.8
80
9.7
15
3.95
50
7.1
85
9.7
20
4.3
55
7.6
90
9.7
25
5.1
60
8.1
95
9.95
30
5.4
65
8.6
100
10.2
Table 2 - Temperature curve and conductivity requirements
for purified water (non temperature-compensated conductivity
measurements).
Transmitter 9125 – Conductivity / resistivity measurement
Page 47
ALARM 3 (Alarm system)
� In case of an alarm 3, choice between a limit
alarm or alarm system function.
MODE :
�
No
�
Limit
�
System
� In case of an alarm system, choice between an
automatic accept or a manual accept.
ACCEPT : �
Auto
�
Manu
� Choice between relays normally open or closed.
RELAY : �
NO
�
NC
ALARM 4 (Timer)
� In case of an alarm 4, choice between a limit or a
timer function.
MODE :
�
No
�
Limit
�
Timer
� INTERV : interval between two cleaning cycles in
minutes.
� Impul. Nb : number of pulses during a cleaning
cycle.
� Ton :
time
when
relay
is
activated,
in
seconds.
� Toff :
time when relay is desactivated, in
secondes.
� TmA :
hold time for the analogue outputs in
minutes.
�
MODE : Syst
ACCEPT : Manu
RELAY : NO
ALARM 3
Select
System
�
MODE : Timer
INTERV : 2440 mn
Impul. Nb : 1
Ton : 005 s
Toff : 005 s
TmA : 00 mn
ALARM 4
Select
Timer
Transmitter 9125 – Conductivity / resistivity measurement
Page 48
S/SPECIAL PROG. Menu
This screen allows to adjust 4..20 mA outputs states in
special events :
� MAINTENANCE
� CALIBRATION
� SYST. ALARM
� MODE :
choice
of
a
preset
value
during
calibration,
alarm
system
or
maintenance or timer.
�
last
�
preset
�
live
� VALUE :
indication of the preset value, 0 to
21 mA.
TEST
TEST menu test the analogue outputs by step of 1 mA
(0…21mA).
�
MODE :
VALUE : 00 mA�
SPECIAL PROG.
Select
Preset
�
CALIBRATION
SYST. ALARM
SPECIAL PROG.
Select
MAINTENANCE
�
VAL : �
TEST
Select
12 mA
Transmitter 9125 – Conductivity / resistivity measurement
Page 49
S/RS485 Menu
This option requires the RS485 kit.
N°
Monec number (0...32)
BAUD
300/600/1200/2400/4800/9600/19200
Transmission speed in bauds
PARIT.
- Without parity bit : No
- With odd parity bit : Odd
- With even parity bit : Even
BIT STOP
- 1 bit stop
- 2 bits stop
SWAP WORD Allow
to
reverse
the
« strong
weight », « light weight » size during
the manipulation of the real variable
(float type). Some equipment need
this reverse to read correctly the real
size data.
The communication protocol is MODBUS/JBUS.
The instrument may be equipped with a RS485 board
(optional) (see MODBUS 9100 manual).
�
N° : 4
BAUD : 9600
PARIT. : odd
BIT STOP : 1
SWAP WORD : NO
RS485
4
Transmitter 9125 – Conductivity / resistivity measurement
Page 50
SERVICE Menu
Note :
An access code may be required if it has been
programmed.
This screen allows to reach the 9125 transmitter
configuration screens.
The display options are detailed page 26.
�
POLARIZATION
AVERAGE
DISPLAY
CODE
SOFT ISSUE
DEFAULT VAL..
mA ADJUST
FACTORY
SERVICE
Select
Transmitter 9125 – Conductivity / resistivity measurement
Page 51
S/POLARIZATION Menu
This menu is displayed only if you use a 2 electrode
probe.
CABLE CAP.
CABLE measure the electrode cable capacity.
It is mandatory to disconnect the probe connector to
measure the capacity or to make an electric
calibration.
TEST
TEST measure the electrode polarization (see
Chap. 5). An error message is displayed if the
polarization is superior to 0,25.
�
CAP. CABLE
TEST
POLARIZATION
Select
�
CAPACITY
1.02 nF
CABLE CAP.
�
POLARIZATION
0.15
TEST
CABLE CAP.
Transmitter 9125 – Conductivity / resistivity measurement
Page 52
S/AVERAGE Menu
S/AVERAGE menu program a moving average on the
measurement.
AVERAGE : define the number of measures to
calculate the average.
TEST : visualize the difference between a measure
done with or without average.
Note :
The measurement cycle lasts 4 seconds if there is
a temperature measurement and only 2 seconds
without temperature measurement.
�
AVERAGE
28.3 mS/cm
REAL
27.9 mS/cm
TEST
�
AVERAGE : 9
TEST :
AVERAGE
Select
TEST :
Transmitter 9125 – Conductivity / resistivity measurement
Page 53
S/CODE Menu
� CALIB. :
access code for temperature and
conductivity calibrations menu.
� PROG. :
access code for "Programming" menu.
� SERVICE : access code for "service" menu.
Note :
If you have forgotten your access code, press
simultaneously ESC and ENTER to enter into the
menu.
�
CALIB. :
PROG. : 0000
SERVICE : 0000
CODE
Select
0000
Transmitter 9125 – Conductivity / resistivity measurement
Page 54
S/SOFT ISSUE Menu
The transmitter displays the type of instrument and the
software version installed.
S/DEFAULT/VALUES Menu
Note :
If you press YES, you load the default values and
you loose the programmed values and the
calibration parameters.
�
MONEC 9125
Cond X.XX
SOFT ISSUE
�
Loading
values
default ?
DEFAULT/VAL.
Yes
!
Transmitter 9125 – Conductivity / resistivity measurement
Page 55
S/mA ADJUST Menu
This menu adjustment of the analogue outputs to
20 mA
with
an
internal
coefficient
between
-9999...9999.
S/FACTORY Menu
Factory code necessary.
The user has no access to this menu.
�
OUTPUT 1
OUTPUT 2
mA ADJUST
Select
�
VALUE : 0000
OUTPUT 1/2
-
+
Transmitter 9125 – Conductivity / resistivity measurement
Page 57
6.
Polarization
Electric representation of the probe and its cable
In a conductivity measurement with a 2 electrode
probe the measurement current is transmitted via the
electrodes. The current is ensured by electrons in the
electrodes and by ionic migration in the solution
measured. An electron exchange process occurs
between the solution and the electrodes.
A significant example below :
2 H2O=H+ + O2 + 4e-
and
2 H2O +2e- = H2 +2 OH-
This reaction requires energy and causes a potential
difference which does not rely on conductivity.
This phenomenon is known as polarization and may
be represented by the scheme below :
Cp
Rx
Cc
V
Polarization
Conductivity
Cable capacity
Rp
electric equivalent of the conductivity probes
Transmitter 9125 – Conductivity / resistivity measurement
Page 58
where :
The dipole Rp, Cp represents the active and reactive
part of the energy necessary to the electron exchange
between the sample and the electrodes.
and :
Cc : the cable capacity
Frequency adjustment according to the conductivity
measurement
0,1
1
10
Cond (theo.) / Cond (meas.)
1E1
1E2
1E3
1E4
1E5
real conductivity (uS/cm)
Auto frequency
62.5 Hz
8000 Hz
Probe 8317 (K=1)
100 m of cable
In the example below 3 measuring ranges are
described :
�
measuring range 1 μS/cm to 100 μS/cm :
When the measuring frequency is high, the cable
capacity becomes important and induces an error
in the conductivity measurement.
If you use a long cable, use the lowest
frequency possible.
Transmitter 9125 – Conductivity / resistivity measurement
Page 59
�
measuring range 100 μS/cm to 1 mS/cm :
In this measuring range the cable capacity or the
polarization are negligible and do not perturb the
conductivity measurement.
�
measuring range 1 mS/cm to 20 mS/cm :
When the measuring frequency is low, the
polarization becomes important and induces an
error in the conductivity measurement.
It is required to use a high frequency.
Automatic adjustment of the frequency
The 9125 adjusts automatically the measurement
frequency according to the conductivity measured, the
cable capacity and the polarization.
The cable capacity is measured in the calibration of
the
first
calibration
point
or
in
the
menu
SERVICE\POLARIZATION\CABLE CAPA.
Polarization is regularly measured and an error
message is displayed if it is too high.
Transmitter 9125 – Conductivity / resistivity measurement
Page 61
7.
Error messages
Note :
In manual acquittal, in order to suppress an error
message press ENTER after correcting the default.
ERROR MESSAGE
DESCRIPTION/POSSIBLE CAUSE
�
Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
�
Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
� The resistivity value is inferior to the lower limit of the
measuring range.
� The conductivity value is superior to the upper limit of
the measuring range.
� The electrode polarization is too high.
10.3
mS/cm
Pt100/Pt1000
SHORT CIRCUIT
12.7
mS/cm
Pt100/Pt1000
OPEN CIRCUIT
60
�.cm
MEASURE
TOO LOW
15.2
mS/cm
MEASURE
TOO HIGH
15.2
mS/cm
POLARIZATION
TOO HIGH
Transmitter 9125 – Conductivity / resistivity measurement
Page 62
ERROR MESSAGES DURING A CALIBRATION
Note :
Press ESC to leave the menu and calibrate again.
�
The electronic zero shift is superior to the limit
programmed.
Calibrate again.
� The
slope
shift
is
superior
to
the
limit
programmed.
Limits : 50...150 %
� The temperature drift is superior to the limit
programmed.
Limits : � 20 °C
OFFSET
OUT OF LIMITS
SLOPE : 162 %
SLOPE
OUT OF LIMITS
�T : -30.0°C
�T
OUT OF LIMITS
COND. CALIB.
COND. CALIB.
TEMP. CALIB.
Transmitter 9125 – Conductivity / resistivity measurement
Page A1
Appendix A : Default values
CALIBRATION
COND. CALIB.
TYPE
: Electr.
RES.
: 1000 �
TEMP. CALIB.
TYPE
: Process
PARAMETERS
DATE
: 01/01/01
SLOPE
: 100.0%
�T
: 0.0°C
PROGRAMMING
MEASURE
PROBE
TYPE
: 2 elect.
K
: 001.00
FREQ.
: Auto
TEMP. COMP.
MEASURE : No
TEMP.
: 025.0°C
TREF.
: 025.0°C
COMP.
: Coef.
COEF.
: 2.0 %
ALARMS
ALARMS S1
AFFECT. : S
LIMIT
: 10.0 mS
DIR.
: Down
DELAY
: 000 s
HYST.
: 00%
RELAY
: NO
ALARMS S2
AFFECT. : S
LIMIT
: 10.0 mS
DIR.
: Down
DELAY
: 000 s
HYST.
: 00%
RELAY
: NO
ALARMS S3
AFFECT. : S
LIMIT
: 10.0 mS
DIR.
: Down
DELAY
: 000 s
HYST.
: 00%
RELAY
: NO
ALARMS S4
AFFECT. : S
LIMIT
: 10.0 mS
DIR.
: Down
DELAY
: 000 s
HYST.
: 00%
RELAY
: NO
mA OUTPUTS
OUTPUT 1
AFFECT.
: S
TYPE
: 4-20
MODE
: Lin
LOWER
: 1.0 �S
UPPER
: 10.0 �S
OUTPUT 2
AFFECT.
: °C
TYPE
: 4-20
LOWER
: 0 °C
UPPER
: 100 °C
Transmitter 9125 – Conductivity / resistivity measurement
Page A2
SPECIAL PROG.
MAINTENANCE
MODE
: Last
TIMER
MODE
: Last
CALIBRATION
MODE : Last
SYST. ALARM
MODE : Last
RS485
No
: 1
BAUD
: 19200
PARITY
: No
STOP BIT : 1
SWAP WORD : No
SERVICE
AVERAGE
AVERAGE : 1
DISPLAY
DISPLAY
UNIT
: S/cm
TEMP.
: °C
LANGUAGE : GB
CODE
CODE
CALIB.
: 0000
PROG.
: 0000
SERVICE : 0000
Transmitter 9125 – Conductivity / resistivity measurement
Page B1
Appendix B : Spare parts list
No other spare parts except those below in the table should be replaced in the
instrument.
Part number
Description
09125=A=1001
9125 equipped CPU board
09125=A=1500
9125 complete conductivity module
09125=A=2000
9125 power supply (standard version)
09125=A=2020
9125 power supply (low voltage version)
09125=A=4000
Relay board (option)
09125=A=1101
RS485 board (option)
09125=A=2485
RS485 kit (JBUS/MODBUS manual +board)
09125=C=3000
Mounted transmitter housing
425=110=221
Cable gland PG11
425=135=222
Cable gland PG13,5
351=007=001
Strap FLEXPAC 7 PTS
621=091=025
French instruction manual
621=191=025
English instruction manual
621=291=025
German instruction manual
621=491=025
Italian instruction manual
621=591=025
Spanish instruction manual
621=891=025
Dutch instruction manual
621=991=000
JBUS/MODBUS communication manual
По вопросам продаж и поддержки обращайтесь:
Астана +7(77172)727-132, Волгоград (844)278-03-48, Воронеж (473)204-51-73, Екатеринбург (343)384-55-89,
Казань (843)206-01-48, Краснодар (861)203-40-90, Красноярск (391)204-63-61, Москва (495)268-04-70,
Нижний Новгород (831)429-08-12, Новосибирск (383)227-86-73, Ростов-на-Дону (863)308-18-15,
Самара (846)206-03-16, Санкт-Петербург (812)309-46-40, Саратов (845)249-38-78, Уфа (347)229-48-12
Единый адрес: [email protected]
Веб-сайт: hach.nt-rt.ru