ATMOS® S 61
Servant vision
ENT treatment unit
GA1GB.110102.0
2017-10 Index: 23
Operating Instructions
English
Further information, accessories, consumables and spare
parts are available from:
ATMOS
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Germany
Phone +49 7653 / 689-0
Fax:
+49 7653 / 689-190
+49 7653 / 689-493 (Service Center)
[email protected]
www.atmosmed.de
Table of contents
1.0
Introduction .........................................................3
1.1
Notes on operating instructions ............................3
1.2
Intended use .........................................................3
1.3
Function ................................................................4
1.4
Explanation of pictures and symbols ....................4
2.0
For your safety ....................................................5
3.0
Assembly: Connecting conditions ....................6
3.1
Required connections for all units of
ATMOS® S 61 Servant .........................................6
3.2
Connection to electrical power line .......................6
4.0
Operation .............................................................7
4.1
Front view: Controls and options ..........................7
4.2
Connections ..........................................................8
4.4.3
On / off switch .......................................................8
4.4.4
Light modules and controls ...................................9
4.4.5
Endoscope management ....................................10
4.4.6
ATMOS® ENT camera .........................................12
4.4.7
ATMOS® LED stroboscope .................................13
5.0
Cleaning and care .............................................14
5.1
General information on cleaning and
disinfection ..........................................................14
5.2
Recommended instrument disinfectants .............15
5.3
Recommended surface disinfectants ..................16
5.4
Recommended endoscope disinfectants ............17
6.0
Hygienic plan .....................................................18
7.0
Maintenance and Service .................................20
7.1
Replacing the fuse ..............................................20
7.2
Sending in the device..........................................20
8.0
Troubleshooting ................................................21
9.0
Accessories .......................................................22
10.0
Technical data ...................................................23
11.0
Disposal .............................................................24
12.0
Notes on EMC ....................................................25
1.1 Notes on Operating instructions
These operating instructions contain important notes on how to operate the ATMOS® S 61 Servant
vision safely, correctly and effectively. Their reading helps to avoid risks, and also to reduce repair costs
and down-times. This increases, amongst other things, the reliability and service-life of the device.
These operating instructions serve not only for new operating personnel to be instructed in its use, but
also for use as a reference manual. Reprints (also in extracts) only with permission in written form by
ATMOS.
These operating instructions must always be kept available near the device.
Care and period tests in conjunction with professional execution provide for operational safety and
readiness for use of your ATMOS® S 61 Servant and are therefore a must besides regular cleaning.
Repair work and period tests may be carried out only by expert personnel authorised by ATMOS. By
applying only original spare parts you will have the guarantee that operational safety, readiness for work
and the value of your ATMOS® S 61 Servant vision will be preserved.
• The product ATMOS® S 61 Servant vision bears CE marking CE according to the EC Directive of
the council for medical products 93/42/EEC and meets the basic requirements of Appendix I of the
directive.
• The product ATMOS® S 61 Servant vision complies with all applicable requirements of the
Directive 2011/65/EC restricting the use of certain hazardous substances in electrical and electronic
equipment (“RoHS”).
• The declaration of conformity and our general standard terms and conditions can be obtained on our
website at www.atmosmed.com.
• The quality management system applied at ATMOS has been certified according to international
standards EN ISO 13485.
• Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible
dangerous situations.
1.2 Indended use
Name: ATMOS® S 61 Servant vision
Main functions:
ATMOS® S 61 Servant vision:
• Light sources
• Storage and heating of endoscopes
• Visualization
• Stroboscopy for larynx diagnostic
Medical indications / application:
Standard ENT examination and/or therapy
Specification of the main function:
ATMOS® S 61 Servant vision:
• Light source LED for light guide
• LED light source, 700 mA
• Camera (ATMOS® Cam)
• Stroboscope (ATMOS® Strobo 21 LED)
User profile:
Doctors and medical assistants
Patient groups:
Patients of all ages with and without restrictions
Application organ:
Mouth to pharynx, auditory canal to the ear drum and the
nasal cavities
Application time:
• ENT unit: Short term use (up to 30 days)
• Suction / Compressed air / ear rinsing / light source:
Temporary application on the patient (less than 60
minutes)
Application site:
Application sites are clinics and practices for ENT doctors
and phoniatrists. The examination and/or therapy with
the ENT unit may only be executed by medically trained
persons.
Contraindications:
May not be used for irrigation of the paranasal sinuses.
The ear irrigation should not be applied to an infected
auditory canal or a perforated eardrum.
The product is: active
Sterility: The ENT unit is no sterile product.
Single-use product / reprocessing:
The ENT unit is intended for multiple use. The device and
parts of the accessories are reusable. For information on
reprocessing, cleaning and disinfection, please see the
operating instructions.
1.4 Explanation of pictures and symbols
Short cuts / symbols on the unit
Foot switch
Signal input and output
Off (feed-in, power
connection)
On (feed-in, power
connection)
Alternating current
Ground wire connection
Warning, special diligent notice
Application part type BF
Fuse according to IEC 417/5016,
DIN 30600/0186
Potential equalization
Manufacturing date
Refer to the operating
instructions
This product complies with
the relevant requirements
of EU directives
Manufacturer
Professional disposal
1.0 Introduction
Short cuts / symbols contained in this operating instructions
Please press where
dot indicates
Subnumeration
Numeration
General information
Follow the arrows
Replace
Check
Please read,
important information
Move, plug... in this
direction
Engage, check
correct fit
Turn, shift ... in this
direction
Signal output
Signal input
0124
1.3 Function
• Coldlight channels (LED)
• Automatic activation over light barrier, headlight
hook, microscope arm
• Endoscope storage without/with heating
• Disinfectant timer
• Integration ATMOS® Cam
• Integration ATMOS® LED Strobo
• Instrument illumination
click
For your safety
2.0 For your safety
• The ATMOS® S 61 Servant vision is produced
according to IEC 601 / EN 60601 and listed in the
following classes:
-
VDE Class of protection 1
-
Class IIa (EEC 93/42).
• The unit should not be positioned directly next to a
wall, because of the ventilation openings on the rear
side! Do not restrict the air supply at the rear of the
unit!
• Caution! Mirror and endoscope heaters may generate
temperatures above 40°C!
• Attention with the cold-light source!
Because of the high energy of the light there is a
large amount of heat emission at the point of the
optical system. Avoid too small a distance between
the tissue and the field of light emission of the light
guide respective of the endoscope, as this may cause
coagulation of the patient's tissue. When using the
endoscope avoid the direct contact between area of
light emission and the tissue.
• Attention, Fire Hazard!
Do never place the area of light emission from
the light guide or from the endoscope onto heat-
absorbing surfaces (dark pieces of cloth, etc.),
because this will cause unacceptable high heating or
even ignition of the material. Switch the light off when
you do not require the light over a prolonged period
of time.
• Care is to be paid in respect to light sources when
working with endoscopes. The intensity of the light is
very high. Do not look directly into the light outlets! In
case of possible light failure remove the endoscope
from the working area.
• Always make sure that you do not blind patients with
the light source! Watch out that patients do not look
directly into the light source! You should always avoid
looking directly into the light source.
> Damages to the eyes due to blinding may be the
result.
• Exclusively connect ATMOS® HL 21 LED and
ATMOS® LS 21 LED to the connections for ATMOS®
HL 21 LED and ATMOS® LS 21 LED. Unsuitable
application parts may result in an electric shock or
damage. Cardiac arrhythmia and even death are
possible.
• The ATMOS® S 61 Servant vision may only be used
under the supervision of skilled staff who have been
authorised by ATMOS and trained in its operation
(IEC 601-1 / EN 60601-1).
• The mains voltage specified on the type plate must
correspond with the data of the power supply system.
• Make sure prior to every application of the equipment
that it is technically safe and in proper condition.
Damaged cables must be replaced immediately!
• Never leave the patient unattended at the treatment
unit.
• Correct configuration in assembly of country-specific
connections:
-
green/yellow: protective conductor (PE)
-
blue: neutral conductor (N)
-
black or brown: phase
• Please note:
A medical insulating transformer with earth leakage
monitor or any similar safety system acc. to EN 60
601-1 is required, if several devices are connected
over one common power supply. The transformer
must correspond to the power consumption of all the
devices to be connected.
• Do not place used contaminated instruments on the
ENT unit!
• The ambient conditions specified in the "Technical
data" must be strictly observed!
• Switch off main switch after finishing work in practice
and close water supply, if present.
• The ATMOS® S 61 Servant vision may be operated
only in rooms used for medical purposes, but not in
areas subject to explosion hazards and in oxygen rich
environments.
• The ATMOS® S 61 Servant vision meets the immunity
to interference requirements of IEC 601-1-2 / EN
60601-1-2 „Electromagnetic Compatibility – Medical
Electrical Devices“.
• The ATMOS® S 61 Servant vision may not be
operated with devices not complying with the
requirements of standard EN 60601-1 „Medical
Electrical Equipment“ and EN 60601-1-2
„Electromagnetic Compatibility“ (Medical Electrical
Equipment).
• ATMOS is not liable for personal injury and damage to
property if
-
no original ATMOS parts are being used,
-
the advice for use in these operating instructions
is not being observed,
-
assembly, new settings, alterations, extensions
and repairs have been carried out by personnel
not authorised by ATMOS.
• Never touch the device´s interfaces and the patient at
the same time!
• The unit may only be opened by a specialist
authorised by ATMOS!
• Please pay attention to the period tests in chapter 7.0
„Service and maintenance“ on page 20.
3.0 Assembly: Connecting conditions
3.1 Required connections for all units of ATMOS® S 61 Servant
Flexible multi-port distributors may not be used as a power supply for the ATMOS® S 61 Servant workstation, ATMOS® S
61 Servant vision or for an optional HF or radiofrequency surgical device.
Unit / Device
Maximum required connections
ATMOS® S 61 Servant ENT workstation
1 x earthing contact socket outlet
ATMOS® S 61 Servant vision
1 x earthing contact socket outlet (basic version) or
1 x fixed connection for the integrated camera or stroboscope LED
ATMOS® S 61 Servant instruments
3 x earthing contact socket outlet
Water separating system (WTA)
1 x earthing contact socket outlet
Monitor
1 x earthing contact socket outlet
Patient chair
1 x earthing contact socket outlet
An adequate number of socket outlets with earthing contact should be mounted for possible connection of further
electrically operated units which may be installed (e.g. installation of an electrically operated ATMOS® patient chair,
water separating system, camera, monitor, etc.)
3.2 Connection to electrical power line
Prerequisites
• Installation acc. to IEC 60346-7-710: earth leakage circuit breaker (FI-circuit breaker) with rated leakage current <0.03 A
• Connection of the power supply cable of the ATMOS® S 61 Servant vision to a safety socket outlet near the device,
max. 3 meters, preferably left (fig.). This may only be carried out by authorized qualified personnel.
• The supply circuit must be separated from other devices e.g. PCs etc.
• If isolating transformers are used then isolation monitoring must be integrated in the isolating transformer.
• Maximum power consumption:
-
ATMOS® S 61 Servant workstation 2.300 VA
-
ATMOS® S 61 Servant vision 200 VA
-
ATMOS® S 61 Servant instruments 250 VA
Flexible multi-port distributors may not be used as a power supply for the ATMOS@ S 61 Servant vision or ATMOS® S 61
Servant workstation.
ca.
250
<150
160
>350
250
640
979,5
1961
1550
740
1730
730
630
2940
445
These connection conditions refer to the overall unit with maximum equipment.
4.1 Front view: Controls and options at complete equipment
On / Off switch
Light port
Camera head
LED
stroboscope
ENT camera
Automatic light
switching
Instrument deposit
for 2 trays
Endoscope
management for
rigid, clean optics
Disinfectant quiver
Display of level of
brightness resp. of
time when switching
on the disinfectant
timer
Adjustment of
brightness resp.
disinfectant timer
Activation
disinfectant times
Display light module
Manual activation /
deactivation of light
ports
First start-up
Prior to shipment each ATMOS® S 61 Servant is being inspected by the manufacturer for function and safety. In order to
make sure that the appliance is working safely after transport and installation, the following points should be observed: The
user should put the appliance into operation only if
1. a functional test on the appliance at the place of operation has been carried out.
2. the operating instructions have been read and noticed.
Following transportation at low temperatures the appliance must be held for up to fours hours at ambient temperature
before first start-up. When the appliance has not been acclimatised the formation of condensation water is possible and a
malfunction might be the result.
4.3 On / off switch
4.0 Operation
After switching on the mains switch, all integrated and connected
devices are ready for operation.
→ green light at mains switch: Mains voltage switched on
When the device gets switched on, all LEDs are lit for a short while.
Please check them each time you switch on the device.
4.2 Connections
Foot switch ENT camera
Foot switch LED stroboscope
Microscope
Potential equalization
Connection supply network (figure: fixed connection)
Connections version 1:
1 , 2
Connections for light guide
3 , 4
Connections for ATMOS® HL
21 LED and ATMOS® LS 21
LED
The LED light module has different connections depending on the
configuration:
Version 1:
2 connections for light guide
2 connections for ATMOS® HL 21 LED and ATMOS® LS 21 LED
Version 2:
4 connections for light guide
Version 3:
4 connections for ATMOS® HL 21 LED and ATMOS® LS 21 LED
Exclusively connect ATMOS® HL 21 LED and ATMOS® LS 21 LED
to the connections for ATMOS® HL 21 LED and ATMOS® LS 21
LED. Unsuitable application parts may result in an electric
shock or damage. Cardiac arrhythmia and even death are
possible.
!
1
2
3
4
4.4 Light modules and controls
4.0 Operation
Hook for headlight
When the headlight is taken off the hook, it switches on automatically. The
same way, when you hang it up again it switches off.
LED headlight
LED light source
Light modules
Control panel
Display of level of
brightness resp. of
time when switching
on the disinfectant
timer
Adjustment of
brightness resp.
disinfectant timer
Activation
disinfectant times
Display light
module
Manual activation /
deactivation of light
ports
• After switching on, the light module is displayed amongst the corresponding light port (LED or no display [no display: no
light module integrated / intended]). The kind of display depends on the presetting from installing resp. your purchased
options.
• Activate light module by taking the cable out of the light port resp. take off the headlight.
• In the case that more than one light guide is taken out, the last removed one will be activated. As soon as this light
guide is put back, the light source of the other light guide is switched on again. When two light guides are taken out and
it should be impeded that the unused channel is activated, it can be switched off manually immediately after removal.
• Adjustment of brightness:
Choose port, press the up/down button on the right side (possible: from 100 % until 10 %). When switching off the
device the last adjustment will be kept.
4.5 Endoscope management
4.0 Operation
4.5.1 Overview
Metal quiver (removable) for storage of clean optics,
optionally heating
Plastics quiver (removable) for storage of used endoscopes
and filling with disinfectant dilution
4.5.2 Endoscope heating
10 - 15 mins
41 - 42 °C
4.5.3 Disinfection monitoring
Fire hazard!
Prior to each use on your patient check your instruments´ temperature (on the back of your hand or similar).
Only store and heat clean instruments!
Clean and disinfect the storage quivers regularly! Therefore please note the instructions in chapter “5.0 Cleaning”.
Please treat the instruments carefully, when inserting resp. removing.
Only fill quivers with admitted and nonhazardous disinfectants and always note application advices of the manufacturer.
Read more in chapter “5.0 Cleaning”.
Please note that the disinfecting quivers can be heated up to max. 42 °C!
!
!
Additionally please observe the maximal and minimal allowed residence time of instruments in the disinfectant dilution
Check the adjusted time prior to use!
An additional acoustic signal results after end of disinfection.
Perform disinfection
Stop disinfection
4.0 Operation
Switch off timer by pressing the button again when timer is active!
Adjustment of disinfection time
When switching on the unit the disinfection time is taken over from prior use.
keep pressed
Adjustment of
disinfection time
release key press-button
Disinfection time
adjusted
Displayed illuminated
segments
1
2
3
4
5
6
7
8
9
10
Time in mins
6
12
18
24
30
36
42
48
54
60
indicator light
changes to red
4.6 ATMOS® ENT camera – quick start
4.0 Operation
The integrated ATMOS® ENT camera ATMOS® Cam 21 / 31 features separate operating instructions.
Please note:
Read these separate operating instructions attentively and follow the stated notes for your safety to guarantee ideal and
safe use of all functions!
!
By switching on the unit ATMOS® S 61 Servant vision the ENT camera is automatically switched on also.
2
1
click
Choose light source
1.) at the treatment unit:
Take out the wanted light source. The adjustments on the camera (LED, Halogen) have been set ex works.
2.) directly at the camera:
Take out the wanted light source (see above). Choose desired optics by pressing the button at the camera ().
Choose the used light source ().
When using the camera next time, all adjustments will be taken over from prior use.
A
B
Connection for camera head
Connection for
LED light source
Buttons for selecting
application part and
light source
Mains on/off
switch
4.6.1 Controls and front view
4.6.2 Operation
1
2
4.7 ATMOS® LED stroboscope - quick start
4.0 Operation
The integrated ATMOS® LED Stroboscope ATMOS® Strobo 21 LED features separate operating instructions.
Please note:
Read these separate operating instructions attentively and follow the stated notes for your safety to guarantee ideal and
safe use of all functions!
!
4.7.1 Controls and front view
4.7.2 Operation
click
1
2
click
3
Phonation
read off measured
average values
Permanent light
Slow motion
Reflection of individual
phases/positions of vocal cord
oscillations
A
B
On / Off switch
Adjustments
Control knob
Connection for microphone
Connection for LED
light handle
Indication for permanent light
Indication for freeze image
Indication for slow motion
Display
4.7.3 Operate the foot switch
5.1 General information on cleaning and disinfection
5.0 Cleaning and care
Prior to cleaning
Medical devices like the ATMOS® S 61 Servant vision need to be fail safe at any time.
Therefore we recommend prior to every use:
if
required
5.1.1 Cleaning the unit surface
• The surfaces of the ATMOS® S 61 Servant vision are
resistant against all surface disinfectants listed in
chapter „Recommended surface disinfectants“. Please
note that long-term use of disinfectants with alcohol
might affect the flexibility and transparency of the
protective covers.
• Wipe the unit surface with a cloth moistened with a
cleaning or disinfecting solution.
• You may also use disinfectant sprays or disinfectant
tissues for cleaning and disinfection.
• Please take care that all surfaces are dry wiped.
Use a single-use cloth to absorb any liquid.
• In case that any disinfectant is spilled, please take care
to dry wipe the surface immediately in order that no
liquid may penetrate gaps and edges.
• Always observe the concentration specifications and
instructions by the respective manufacturer!
) Do not use
-
Disinfectants which contain organic or inorganic
acids or bases as they could cause corrosion
damage.
-
Disinfectants containing chloramides or phenol
derivatives, since these may cause stress cracks in
the material used for the housing of the unit.
• Only deposit clean instruments on the board!
• Clean and disinfect the instruments regularly!
5.1.3 Endoscope quivers
• The metal quivers of the endoscope holder are to be
used solely for holding the endoscopes, these first
having been cleaned and disinfected. The quivers are
to be cleaned daily and subsequently disinfected. For
doing this, the stopper at the lower end should be taken
off.
5.1.2 Instrument trays
• Before disinfection, thoroughly rinse the trays under
running water. A detergent or cleaning agent (surface
disinfectant) may also be used if required.
-
Use water to thoroughly rinse all residues of these
substances.
• Melamine and anodized aluminium trays cannot be
sterilised.
5.0 Cleaning and care
5.2 Recommended instrument disinfectants
Manual disinfection of instruments
Disinfectant
Ingredients
in 100 g Manufacturer
Korsolex® med AF
(Application concentrate)
N-dodecylpropane-1,3-diamine
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
surfactants, corrosion inhibitors, ph-value regulators, foam
inhibitors
15.6 g
5.1 g
Bode Chemie,
Hamburg
Korsolex® basic
(Application concentrate)
glutaral
(ethylenedioxy)dimethanol
surfactants, salts, corrosion inhibitors
15.2 g
19.7 g
Bode Chemie,
Hamburg
Korsolex® plus
(Application concentrate)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
didecyldimethylammonium chloride
surfactants, corrosion inhibitors, complexing agents, ph-value
regulators
9.2 g
13.0 g
Bode Chemie,
Hamburg
Korsolex® extra
(Application concentrate)
(ethylendioxy)dimethanol
glutaral
benzyl-C12-18-alkyldimethyl-ammonium chlorides
didecyldimethylammonium chloride
surfactants, foam inhibitors, corrosion inhibitors
15.3 g
7.5 g
1.0 g
1.0 g
Bode Chemie,
Hamburg
neodisher® Septo MED
(Application concentrate)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine
didecyldimethylammonium chloride
non-ionic surfactants, perfumes
9.2 g
13.0 g
Dr. Weigert,
Hamburg
neodisher® Septo 3000
(Application concentrate)
glutaral
(ethylenedioxy)dimethanol
15.2 g
19.7 g
Dr. Weigert,
Hamburg
Sekusept® PLUS
(Application concentrate)
glucoprotamin
25 g Ecolab,
Düsseldorf
Sekusept® aktiv
(Application concentrate)
sodiumpercarbonate, non-ionic surfactants, phosphonates
Ecolab,
Düsseldorf
Gigasept® Instru AF
(Application concentrate)
Cocospropylendiaminguanidindiacetate
Phenoxypropanols
Benzalkonium chloride
non-ionic surfactants, ph-value regulators, corrosion inhibitors
14 g
35 g
2.5 g
Schülke & Mayr,
Norderstedt
Gigasept® FF (new)
(Application concentrate)
succindialdehyde
dimethoxytetrahydrofuran
anionic and non-ionic surfactants, perfumes,
methylisothiazolinone
11.9 g
3.2 g
Schülke & Mayr,
Norderstedt
Gigazyme®
(Application concentrate)
non-ionic surfactants, enzymes, corrosion inhibitors
5 - 15 g Schülke & Mayr,
Norderstedt
Automatic disinfection of instruments
Disinfectant
Ingredients
in 100 g Manufacturer
Dismoclean® 24 Vario
(Application concentrate)
surfactants, micro-encapsulated enzymes, corrosion
inhibitors, complexing agents
Bode Chemie,
Hamburg
Dismoclean® 28 alka med
(Application concentrate)
alkali dispenser, complexing agents, corrosion inhibitors,
surface active materials
Bode Chemie,
Hamburg
Dismoclean® twin basic / twin
zyme
Bode Chemie,
Hamburg
Dismoclean® twin basic
alkali dispenser, complexing agents, corrosion inhibitors
Dismoclean® twin zyme
surface active materials, enzymes, stabilisers, corrosion
inhibitors
neodisher® FA
phosphates
15 - 30 g Dr. Weigert,
Hamburg
neodisher® MediClean forte
(Application concentrate)
non-ionic and anionic surfactants
enzymes
< 5 g Dr. Weigert,
Hamburg
Thermosept® alka clean forte
(Application concentrate)
non-ionic surfactants
anionic surfactants
NTA (nitrilotriacetic acid) and its salts
enzymes, poly carboxylates
corrosion inhibitors
< 5 g
< 5 g
< 5 g
< 5 g
Schülke & Mayr,
Norderstedt
Thermosept® RKN-zym
non-ionic surfactants, enzymes, corrosion inhibitors, glycols
5 - 15 g Schülke & Mayr,
Norderstedt
5.3 Recommended surface disinfectants
Coated surfaces
Disinfectant
Ingredients
in 100 g Manufacturer
Green & Clean SK
Di alkyl dimethyl ammonium chloride
Alkyl dimethyl ethyl benzyl ammonium chloride
Alkyl dimethyl benzyl ammonium chloride
< 1 g
< 1 g
< 1 g
Metasys, Rum
(Austria)
Dismozon® pur
(Granulate)
End of product 12/2014
magnesium monoperoxyphthalate hexahydrate
80 g Bode Chemie,
Hamburg
Dismozon® plus
(Granulate)
magnesium monoperoxyphthalate hexahydrate
95.8 g Bode Chemie,
Hamburg
Kohrsolin® FF
(Application concentrate)
glutaral
benzyl-C12-18-alkyldimethyl-ammonium chlorides
didecyldimethylammonium chloride
5 g
3 g
3 g
Bode Chemie,
Hamburg
Perform®
Potassium peroxymonosulfate
45 g Schülke & Mayr,
Norderstedt
Terralin® Protect
(Application concentrate)
benzyl-C12-16 alkyldimethyl, chloride
2-phenoxyethanol
aminoalkylglycine
non-ionic surfactants, perfumes
22 g
17 g
0.9 g
Schülke & Mayr,
Norderstedt
Other surfaces
Disinfectant
Ingredients
in 100 g Manufacturer
Dismozon® pur
(Granulate)
End of product 12/2014
magnesium monoperoxyphthalate hexahydrate
80 g Bode Chemie,
Hamburg
Dismozon® plus
(Granulate)
magnesium monoperoxyphthalate hexahydrate
95.8 g Bode Chemie,
Hamburg
Kohrsolin® FF
(Application concentrate)
glutaral
benzyl-C12-18-alkyldimethyl-ammonium chlorides
didecyldimethylammonium chloride
5 g
3 g
3 g
Bode Chemie,
Hamburg
Perform®
Potassium peroxymonosulfate
45 g Schülke & Mayr,
Norderstedt
Terralin® Protect
(Application concentrate)
benzyl-C12-16 alkyldimethyl, chloride
2-phenoxyethanol
aminoalkylglycine
non-ionic surfactants, perfumes
22 g
17 g
0.9 g
Schülke & Mayr,
Norderstedt
Surface disinfection F
312
alkyl-benzyl-dimethyl-ammonium chloride
non-ionic surfactants, complexing agents, hexyl cinnamal, butyl
phenyl methyl proionale, linalool
13 g Dürr Dental,
Bietigheim-
Bissingen
When using disinfectants containing aldehyde and amine at the same object. colour changes may occur.
5.0 Cleaning and care
5.0 Cleaning and care
5.4 Recommended endoscope disinfectants
Manual disinfection of endoscopes
Disinfectant
Ingredients
in 100 g Manufacturer
Helipur® H plus N
glutaral
2-propanol
ethyl hexanol
surfactants, complexing agents, corrosion inhibitors, colorants,
perfumes
12 g
7.5 g
0.5 g
BBraun,
Melsungen
Helix® Ultra
peracetic acid
BBraun,
Melsungen
Korsolex® basic
glutaral
(ethylendioxy) dimethanol
surfactants, salts, corrosion inhibitors
15.2 g
19.7 g
Bode Chemie,
Hamburg
neodisher® MediClean forte
(Application concentrate)
non-ionic and anionic surfactants
enzymes
< 5 g Dr. Weigert,
Hamburg
Sekusept® aktiv
(Application concentrate)
sodiumpercarbonate, non-ionic surfactants, phosphonates
Ecolab,
Düsseldorf
Automatic disinfection of endoscopes
Disinfectant
Ingredients
in 100 g Manufacturer
Korsolex® basic
glutaral
(ethylendioxy) dimethanol
surfactants, salts, corrosion inhibitors
15.2 g
19.7 g
Bode Chemie,
Hamburg
neodisher® MediClean
forte
(Application concentrate)
non-ionic and anionic surfactants
enzymes
< 5 g Dr. Weigert,
Hamburg
Gigasept® FF (new)
(Application concentrate)
succindialdehyde
dimethoxytetrahydrofuran
anionic and non-ionic surfactants, perfumes,
methylisothiazolinone
11.9 g
3.2 g
Schülke
& Mayr,
Norderstedt
Endozime® AW Plus
2-propanol
Ruhof,
Mineola
(USA)
AdaptacleanTM
Potassium hydroxide, surfactants
ASP,
Norderstedt
Qualifi ed and trained
staff who are familiar
with reprocessing.
(Please fi ll in the
responsible person
-> use a water-based
overhead marker)
Daily
Secretion canister
Hose connection (nozzle)
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Suction lid
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Gasket
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Bacterial fi lter
Daily exchange or when fi lter is blocked
X
Splash protection
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Float ball
X
X
Cleaning and disinfection (automatic or manual)
X
Suction hose in the
canister
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Secretion canister
X
X
Empty when the canister is full; at least daily
Cleaning and disinfection (automatic or manual)
X
Disposable canister
system
Exchange and disposal of full canister
X
Hose irrigation system
Suction nozzle
for hose irrigation
X
X3
Wipe cleaning and disinfection
X
Silicone attachment piece
X
X2,4,5,6
Cleaning and disinfection (automatic or manual)
X
Exchange of the Silicone attachment
X
Suction nipple
X
Manual cleaning after each patient
X
X2,4,5,6
Manual or automatic cleaning and disinfection
X
Secretion suction hose
X
Rinse the secretion hose with the hose irrigation
system after each procedure;
X
X2,4,5,6
Exchange or disinfection of the hose
X
Storage canister for
hose irrigation system
X
X2,4,5,6
Cleaning with a brush; Manual or automatic cleaning and
disinfection
X
Ear irrigation / Thermal nystagmus stimulation
Ear irrigation bowl
X
X2,4,5
Manual or automatic cleaning and disinfection
X
Handle
X
X3
Wipe cleaning and wipe disinfection
X
Jet connection
X
X2,4,5,6
Manual or automatic cleaning and disinfection
X
Splash protection
X
X2,4,5
Cleaning and disinfection (automatic or manual)
X
Hose tip (disposable)
Exchange after each application
X
Rinsing attachment
X
X2,4,5
Manual or automatic cleaning and disinfection
X
Hygiene fi lter
See operating instructions for hygiene fi lter
X
Rinsing lid with rinsing
hose
X
X2,4,5
Manual or automatic cleaning and disinfection
X
Rinsing bottle
X
X2,4,5,6
Cleaning and disinfection; cleaning in the dishwasher
with the glass care programme
X
Medication nebulisation / Politzer
Handle for
compressed air
X
X3
Manual cleaning and disinfection
X
Sprayer jet
X
Cleaning after each application
X
X2,4,5,6
Manual or automatic cleaning and disinfection
X
Sprayer head
X2,4,5
Multiple rinsing of the sprayer head with water
X
Hose at sprayer head
X
X
Weekly exchange of the hose or when changing the
medication
X
Sprayer bottle
X
X2,4,5,6
Cleaning in a cleaning and disinfection device; weekly or
when changing the medication
X
Politzer olive
X
X2,4,5,6
Exchange after each application.
Cleaning and disinfection
X
Politzer connection
X
X2,4,5,6
Exchange after each application.
Cleaning and disinfection
X
Endoscope management
Plastic quiver
X
X2,4,5
Cleaning with a brush; disinfection
X
Metal quiver
X
X2,4,5,6
Cleaning with a brush; disinfection (automatic or manual)
X
Fixation adapter for
plastic quiver
X
X2,4,5
Manual or automatic cleaning and disinfection
X
Protective sleeve (tefl on
element for metal quiver)
X
X2,4,5
Manual or automatic cleaning and disinfection
X
Cleaning and disinfection plan
ATMOS® S 61 Servant
Qualifi ed and trained
staff who are familiar
with reprocessing.
(Please fi ll in the re-
sponsible person ->
use a water-based
overhead marker)
Daily
Instrument management
ENT instruments
X
X2,4,5
X
Immerse instruments into solution immediately after use,
complete wetting is required, air must be removed from
any cavities, after the contact time instruments must be
rinsed with water, have to be dried and sterilised after-
wards. Please also observe the ATMOS operating
instructions for ENT instruments.
X
Instrument bowl
X
X4
Cleaning and disinfection (manual)
X
Instrument bowl with cover
X
X4
Cleaning with a brush; afterwards disinfection (manual)
X
Visualisation
ATMOS® Cam 21 / 31
X
X3
Wipe cleaning and wipe disinfection
X
ATMOS® Strobo 21 LED
X
X3
Wipe cleaning and wipe disinfection
X
Flexible scope
X
X1,7,8
X1)
Immediate pre-cleaning after the procedure
X
Rigid scope
X
X1,7,8
X1)
Immediate pre-cleaning after the procedure
X
Laryngoscope
X
X1,7,8
X1)
Immediate pre-cleaning after the procedure
X
Light conductor
X
X3
Wipe cleaning and wipe disinfection
X
Light grip
X
X3
Wipe cleaning and wipe disinfection
X
Microscope
X
X3
Wipe cleaning and wipe disinfection
X
Headlight
X
X3
Wipe cleaning and wipe disinfection
X
Radiofrequency surgery
ATMOS® RS 221 (surface)
X
X3
Wipe cleaning and wipe disinfection
X
Ergonomic handles
X
X1,2,4,5
X1)
Wipe cleaning and wipe disinfection
X
Bipolar tweezers
X
X1,2,4,5
X1)
Immediate pre-cleaning after the procedure
cleaning and disinfection (automatic or manual);
use of enzymatic detergents
X
Bipolar electrode
X
X1,2,4,5
X1)
X
Bipolar electrode cable
X
X1,2,4,5
X1)
Immediate pre-cleaning after the procedure
cleaning and disinfection (automatic or manual);
use of enzymatic detergents
X
Neutral electrode
X
X1,2,4,5
X1)
X
Neutral electrode cable
X
X1,2,4,5
X1)
Immediate pre-cleaning after the procedure
cleaning and disinfection (automatic or manual);
use of enzymatic detergents
X
ENT electrodes
X
X1,2,4,5
X1)
X
Surfaces
Housing
X
X3
Wipe cleaning and wipe disinfection
X
Roller shutter
X
X3
Wipe cleaning and wipe disinfection
X
System frame
X
X3
Wipe cleaning and wipe disinfection
X
Drawers
X
X3
Wipe cleaning and wipe disinfection
X
Writing surface
X
X3
Wipe cleaning and wipe disinfection
X
Instrument deposit
X
X3
Wipe cleaning and wipe disinfection
X
Mirror pre-heater
X
X3
Wipe cleaning and wipe disinfection
X
Tongue patches and
swab dispenser
X
X3
Wipe cleaning and wipe disinfection,
every day or when refi lling
X
Waste disposal
X
X3
Wipe cleaning and wipe disinfection,
every day or when emptying the canister
X
Instrument tray
X
X3
Wipe cleaning and wipe disinfection, daily when replacing
with new instruments
X
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Str. 16 79853 Lenzkirch/Germany
Phone +49 7653 689-0 Fax +49 7653 689-190
[email protected] www.atmosmed.de
The above stated hygiene requirements are based on the regulations
according to the Medical Devices Act, the Medical Devices Operator
Ordinance, §18 IfSG and the recommendations of the Robert Koch
Institute.
Defi nition of the required reprocessing steps result from the re-
commendations of the Robert Koch Institute: „Requirements for
the reprocessing of medical products“. The medical products were
categorised in the risk groups uncritical, semicritical and critical. The
reprocessing measures mentioned in this cleaning and disinfection
plan are a recommendation of ATMOS MedizinTechnik. Any additio-
nal reprocessing measures are at the operator‘s discretion.
All the recommended disinfectants which are stated herein are listed
disinfectants (VAH/RKI) and have been tested on their suitability of
use on the ATMOS® S 61 Servant. ATMOS MedizinTechnik cannot
be hold liable for any damage caused by wrong concentration of the
disinfectants or by the application of any other disinfectants.
Patients with suspicion of a clinical disease or who developed a
transmissible spongiform encephalopathy (CJK, vCJK, etc.) have to
be treated at facilities which are able to provide for the necessary
preventive measures against infection. The reprocessing of the reu-
sable instruments and material may only be performed at facilities
which have an externally certifi ed QM Management acc. to DIN EN
ISO 13485.
The Medical Devices Act, IfSG, the RKI directives, BGR 250 and
TRBA 250 always have to be considered.
Important information
Wipe cleaning and wipe disinfection: All sur-
faces have to be wiped with a clean (dispos-
able) wipe which is damped with disinfectant
solution; the entire surface has to be wiped tho-
roughly and may not be dried afterwards.
1) Please observe the manufacturer‘s operating instructions.
2) Alternative to manual cleaning und disinfection: Wash-
Disinfector 78°C / 172°F
6) Material dimensionally stable at 134°C
Wrong concentration of disinfectants
may lead to damage!
Recommended disinfectants
3) Surface disinfection
for coated surfaces:
Green & Clean SK (ATMOS)
Dismozon® plus (Bode Chemie)
Kohrsolin® FF (Bode Chemie)
Perform® (Schülke & Mayr)
Terralin® Protect (Schülke & Mayr)
Other surfaces:
Dismozon® plus (Bode Chemie)
Kohrsolin® FF (Bode Chemie)
Mikrobac® forte (Bode Chemie)
Perform® (Schülke & Mayr)
Terralin® Protect (Schülke & Mayr)
Surface disinfection FD 312 (Dürr Dental)
4) Instruments - manual disinfection:
Korsolex® med AF (Bode Chemie)
Korsolex® basic (Bode Chemie)
Korsolex® plus (Bode Chemie)
Korsolex® extra (Bode Chemie)
neodisher® Septo MED (Dr. Weigert)
neodisher® Septo 3000 (Dr. Weigert)
Sekusept® PLUS (Ecolab)
Sekusept® aktiv (Ecolab)
Gigasept® Instru AF (Schülke & Mayr)
Gigazyme® (Schülke & Mayr)
Gigasept FF neu (Schülke & Mayr)
5) Instruments - automatic disinfection:
Dismoclean® 24 Vario (Bode Chemie)
Dismoclean® 28 alka med (Bode Chemie)
Dismoclean® twin basic/twin zyme (Bode Chemie)
neodisher® FA (Dr. Weigert)
neodisher® MediClean forte (Dr. Weigert)
Thermosept® alka clean forte (Schülke & Mayr)
Thermosept® RKN-zym (Schülke & Mayr)
7) Endoscopes - manual disinfection:
Helipur® H plus N (BBraun)
Helix® Ultra (BBraun)
Korsolex® Basic (Bode Chemie)
neodisher® MediClean forte (Dr. Weigert)
Sekusept® aktiv (Ecolab)
8) Endoscopes - automatic disinfection:
Korsolex® Basic (Bode Chemie)
neodisher® MediClean forte (Dr. Weigert)
Gigasept® FF neu (Schülke & Mayr)
Endozime® AW Plus (Ruhof)
ADAPTACLEAN™ (ASP)
Please see the manufacturer‘s instructions for
concentration, contact time, temperature and
the compatibility of materials.
GA3GB.110100.0
2017-05 Index: 13
7.0 Maintenance and Service
7.1 Replacing the fuse
Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge
and are familiar with the product. To carry out these measures the person must have the necessary test devices and
original spare parts.
ATMOS recommends: Work should be carried out by an authorized ATMOS service partner. This ensures that repairs and
testing are carried out professionally, original spare parts are used and warranty claims remain unaffected.
At least every 12 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS
recommends an inspection according to the manufacturer‘s specifications.
7.2 Sending in the device
• Remove and properly dispose of consumables.
• Clean and disinfect the product and accessories according to the operating instructions.
• Place used accessories with the product.
• Fill in the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certificate.
) This form is enclosed to each delivery and can be found at www.atmosmed.com.
• The device must be well padded and packed in suitable packaging.
• Place the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certificate in an
envelope.
• Affix the envelope to the outside of the package.
• Send the product to ATMOS or to your dealer.
7.1.1 Earthing contact socket outlet
7.1.2 Fixed connection
> Defective electronics
Contact the ATMOS service!
Green control light at mains switch
does not glow, device does not
work
> AC power line not connected
Connect AC power line
> Device fuse is defective
Replace fuses
Green control light at mains switch
glows, but device does not work
> AC power line defective
Contact the ATMOS service!
> AC power line fuse defective
Contact the ATMOS service!
Microscope
No activation / deactivation
of microscope light port via
microscope column
> Cable of switching contact is not or
not correctly connected
Connect cable correctly to device
> Switching contact in microscope
column is defective / misaligned
Contact the ATMOS service!
Accessories for light package 2-channel LED for light conductor
531.1100.0
Adapter for light conductor with ATMOS®/Storz connection
530.6100.0
Adapter for light conductor with Olympus connection
530.6101.0
Adapter for light conductor with Pentax connection
530.6102.0
Adapter for light conductor with Wolf connection
530.6103.0
High-perf. light guide cable, Ø 4.8 mm, L = 1.8 m, Storz, straight
950.0152.0
Light conducting cable, Ø 3.5 mm, L = 1.7 m, Storz, straight
508.0663.0
Light conducting cable, Ø 3.5 mm, L = 1.8 m, Storz angled, 90°
508.0664.0
Accessories for light package 2-channel LED for ATMOS® HL 21 LED and ATMOS® LS 21 LED 531.1200.0
ATMOS® LS 21 LED
LED light handle for direct connection to endoscopes and laryngoscopes
507.4600.0
ATMOS® LS 21 LED, warm white
LED light handle for direct connection to endoscopes and laryngoscopes
507.4602.0
Lithium-ionic rechargeable battery
Battery for connection to ATMOS® LS 21 LED or ATMOS® HL 21 LED
507.4510.0
Universal battery quick charging power supply unit (100-240 V~)
011.1199.0
ATMOS® HL 21 LED
LED headlight with long-life and high-performance white light LED
530.4020.0
Headlight acc. to Binner with headband, 90°, with light conducting cable
502.0515.5
Headlight acc. to Binner without headband, 90°, with light conducting cable
502.0516.0
Accessories for endoscope management
Shock protection adapter
for endoscopes Ø 2.8 - 4 mm, teflon element for metal quiver
508.0777.5
Fixation adapter for plastics quiver
508.0782.0
Blind for light conductor support
531.0271.0
Accessories for LED stroboscope
Body sound stroboscope adapter
507.4775.0
10.0 Technical data at complete equipment
Voltage
100 - 230 V~ ± 10 %; 50/60 Hz
Power consumption
Max. 1.0 A
Power consumption
Max. 200 VA
Fuses
2 x T 3.15 A/H 250 V
Other safety equipment
Internal fuses on control circuit board
LED light module
Illuminance: min. 195 kLux (in 5 cm distance of a 4.7 mm high-performance light
guide)
Colour temperature: 5.500 K ± 10 %
Removable adaptor
Olympus, Wolf, Pentax
LED power supply for ATMOS® LS
21 LED
700 mA regulated
Endoscope management
For rigid and flexible optics (max. length 500 mm), one quiver per clean and
contaminated optics.
Endsocope heating
For 4 quivers, quivers temperature approx. 40 °C
Inspection of disinfection time
Timer adjustable in 10 steps from 6 to 60 minutes
Camera module
see ATMOS® Cam 21 / 31
LED stroboscope
see ATMOS® Strobo 21 LED
Operating time
Continuous operation
Protective earth conductor
resistance
max. 0.1 Ω
Earth leakage current
max. 0.5 mA
Enclosure leakage current
max. 0.1 mA
Patient leakage current
max. 0.1 mA
Ambient conditions
Transport / storage
-10...+50 °C
30...95 % humidity without condensing
air pressure 500...1060 hPa
Operation
+10...+35 °C
30...95 % humidity without condensing
air pressure 700...1060 hPa
Dimensions H x W x D
88.5 x 41.2 x 54.0 cm
Weight
15 - 30 kg, depending on configuration
Period tests
Repeat test of the electrical safety every 12 months.
Recommended: inspection according to the manufacturer‘s specifications.
Safety class (EN 60601-1)
I
Type of protection
Application parts type BF
Type of protection
IPX0
Classification in accordance with
Annex IX to EC Directive 93/42/
EEC
Class I, Standard 12
CE marking
CE
UMDNS code
11-585
GMDN code
11585
Ident-Nr.
531.0000.0
Please note the technical specifications of the single devices in the separate operating instructions!
Technical data unchanged since 21.01.2016
• The ATMOS® S 61 Servant vision does not contain any hazardous goods.
• The component parts of the ATMOS® S 61 Servant vision must be disposed of
correctly and the materials are to be separated carefully.
• The materials of the housing can be recycled completely.
11.0 Disposal
12.0 Notes on EMC
12.1 Guidelines and Manufacturer's Declaration - Emissions
The ATMOS® S 61 Servant vision is intended for use in the electromagnetic environment specified below. The customer or
user of the ATMOS® S 61 Servant vision should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions acc.to CISPR 11
Group 1
The ATMOS® S 61 Servant vision uses RF energy only
for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
HF transmission according to
CISPR 11
Class B
The ATMOS® S 61 Servant vision is suitable for use
in all establishments, including domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions according to
IEC 61000-3-2
Class A
Voltage fluctuations/flicker according
to IEC 61000-3-3
Corresponds
12.2 Guidelines and Manufacturer's Declaration - Immunity
The ATMOS® S 61 Servant vision is intended for use in the electromagnetic environment specified below. The customer or
user of the ATMOS® S 61 Servant vision should ensure that it is used in such an environment.
Immunity Test
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete, or
ceramics tile. If floors are synthetic, the
relative humidity should be at least 30
%.
Fast electrical transient/
burst IEC 61000-4-4
± 2 kV Mains
± 1 kV I/Os
± 2 kV Mains
Inapplicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surges IEC 61000-4-5
± 1 kV common-mode
± 1 kV differential mode
± 1 kV common-mode
± 1 kV differential mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Magnetic field at power
frequency 50/60 Hz acc.
to IEC 61000-4-8
3 A/m
Inapplicable
Power frequency magnetic fields should
be that of a typical commercial or
hospital environment.
12.0 Notes on EMC
Immunity Test
IEC 60601- Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Voltage Dips / Dropout
IEC 61000-4-11
< 5 % UT
(> 95 % Dip of the UT)
for 0.5 Cycle
40 % UT
( 60 % Dip of the UT)
For 5 cycles
70% UT
( 30 % Dip of the UT)
For 25 cycles
< 5 % UT
( 95 % Dip of the UT)
for 5 s
< 5 % UT
(> 95 % Dip of the UT)
for 0.5 Cycle
40 % UT
( 60% Dip of the UT)
For 5 cycles
70% UT
( 30 % Dip of the UT)
For 25 cycles
< 5 % UT
( 95 % Dip of the UT)
for 5 s
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the ATMOS®
S 61 Servant vision requires continued
function during interruptions of the
energy supply, it is recommended to
supply the ATMOS® S 61 Servant vision
from an uninterruptible power supply or
a battery.
NOTE UT is the alternating maims voltage prior to application of the test levels.
12.3 Guidelines and Manufacturer´s Declaration – electromagnetic immunity – for
devices and systems that are not life-sustaining.
The ATMOS® S 61 Servant vision is intended for use in the electromagnetic environment specified below. The customer or
user of the ATMOS® S 61 Servant vision should ensure that it is used in such an environment.
Immunity Test
IEC 60601- Test
Level
Compliance Level
Electromagnetic Environment - Guidance
Conducted RF IEC
61000-4-6
V1= 3 Veff
150 kHz to 80 MHz
3 V
Portable and mobile communications
equipment should be separated from the
ATMOS® S 61 Servant vision incl. the cables
by no less than the distances calculated/listed
below.
Recommended distances:
d = (3.5 / V1) * √(P)
d = (3.5 / E1) * √(P)
d = (7 / E1) * √(P)
where „P“ is the max. power in watts (W) and
D is the recommended separation distance in
meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site (a)
survey, should be less than the compliance
level (b).
Interference may occur in the vicinity of
equipment containing following symbol:
Radiated RF IEC
61000-4-3
E1 = 3 V/m
80 MHz to 2.5 GHz
3 V/m
12.0 Notes on EMC
12.4 Recommended safety distance between portable and mobile RF
Communications equipment and the ATMOS® S 61 Servant vision
NOTE 1 With 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in every case. The emanation of electromagnetic waves is affected by absorp-
tion and reflection of buildings, objects and people.
a
The field strength of stationary transmitters, such as base stations of cellular phones and mobile terrain radio equip-
ment, amateur radio transmitters, cbm broadcast and TV stations cannot be predestined exactly. To determine the elec-
tromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured
field strength at the location where the ATMOS® S 61 Servant vision is used exceeds the above compliance level, the
ATMOS® S 61 Servant vision is to be observed to verify the intended use. If abnormal performance characteristics are
noted, additional measures might be necessary, e. g. a changed arrangement or another location for the device.
b
Within the frequency range of 150 kHz to 80 MHz the field strength should be below 3 V/m.
The ATMOS® S 61 Servant vision is intended for use in electromagnetic environment in which radiated disturbances are
controlled. The customer or user of the ATMOS® S 61 Servant vision can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS® S 61
Servant vision as recommended below, according to the maximum output power of the communications equipment.
Safety distance, depending on transmit-frequency m
Nominal capacity
of the transmitter
W
150 kHz to 80 MHz
d = (3,5/V1) * √(P)
80 MHz to 800 MHz
d = (3,5/E1) * √(P)
800 MHz to 2.5 GHz
d = (7/E1) * √(P)
0.01
0.1167
0.1167
0.2333
0.1
0.3689
0.3689
0.7379
1.0
1.1667
1.1667
2.3333
10
3.6893
3.6893
7.3786
100
11.6667
11.6667
23.3333
For transmitters for which the maximum nominal output is not indicated in the above table, the recommended
safety distance d in meters (m) can be determined using the equation belonging to the respective column whereas
P is the maximum nominal output of the transmitter in watts (W) acc. to manufacturer´s specification.
NOTE 1 By 80 MHz and 800 MHz the higher frequency range applies.
NOTE 2
These guidelines may not be applicable in every case. The emanation of electromagnetic waves is affected by
absorption and reflection of buildings, objects and people.
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch / Germany
Phone: +49 7653 689-370
[email protected]
www.atmosmed.com