EXCELL PRECISION CO., LTD.

EX-2005C Check-weigh Indicator User Manual © EXCELL PRECISION CO., LTD. 2011. All rights reserved worldwide. The information contained herein is the property of EXCELL PRECISION CO., LTD. and is supplied without liability for errors or omissions. No part may be reproduced or used except as authorised by contract or other written permission. The copyright and the foregoing restriction on reproduction and use extend to all media in which the information may be embodied.

EXCELL PRECISION CO., LTD.

TABLE OF CONTENTS SAFETY PRECAUTION ...............................................................................................................2 CHAPTER 1 PANELS AND SPECIFICATION.............................................................................3 1-1 Front Panel.........................................................................................................................3 1-2 Rear Panel..........................................................................................................................4 1-3 Keypad Description...........................................................................................................5 1-4 A/D Conversion..................................................................................................................6 1-5 Power Supply.....................................................................................................................6 1-6 Dimensions (mm) ..............................................................................................................7 CHAPTER 2 GENERAL FUNCTION............................................................................................8 2-1 Function Setting Procedures ...........................................................................................8 2-2 Function Setting ................................................................................................................9 CHAPTER 3 CALIBRATION DESCRIPTION.............................................................................12 3-1 Load Cell Connection......................................................................................................12 3-2 Parameter Setting And Calibration Flowchart ..............................................................13 3-3 Specification Calibration ................................................................................................15 3-4 General Calibration .........................................................................................................17 3-5 Error Messages ...............................................................................................................18 CHAPTER 4 INTERFACE DESCRIPTION.................................................................................19 CHAPTER 5 MAINTENANCE ....................................................................................................26 5-1 Restore All Parameters To Default Factory Values.......................................................26 5-2 Test Mode.........................................................................................................................27 5-2-1 7-Segment Display Testing ......................................................................................28 5-2-2 Keypad And Calibration SW Testing .......................................................................28 5-2-3 Display A/D Internal Value Display..........................................................................28 5-2-4 RS-232 Serial Loop Back Testing............................................................................28 5-2-5 EEPROM Memory Testing........................................................................................28 5-2-6 IFC Factory Testing (reserved) ................................................................................28 5-2-7 RELAY_IO Interface Card Testing............................................................................28 APPENDIX 7-SEGMENT DISPLAY CHARACTERS .................................................................29

EXCELL PRECISION CO., LTD.

! SAFETY PRECAUTION � When the indicator is installed, connect an earth bonding conductor from FG to the earth connection marked “ ”. � Disconnect the mains power supply before opening the rear panel for whatever. � To install the optional interface cards, it is necessary to disconnect the mains power supply and fit a yellow/green earth bonding cable to the rear panel. (Fixed with another two earth bonding cables) � Before turning the power on ensure the supply voltage is within the acceptable range, AC 85 ~ 265 V. � The operating ambient temperature range is -10 ~ +40 °C (+14 ~ +104 °F). Thank you for purchasing EXCELL Check-weigh Indicator. To guide you to use our product correctly, please read the user manual carefully to avoid mis-operating this product.

EXCELL PRECISION CO., LTD.

CHAPTER 1 PANELS AND SPECIFICATION 1-1 Front Panel � � � � Display 6 digits, bright red, 7 segment LED display, character height 16mm (0.63”). � Indication icons “◄” HI ◄ : Weight over High Limit indication OK ◄ : Weight qualified indication LO ◄ : Weight below Lo Limit indication RUN ◄ : Check-weighing in operation indication � � � � Weight units Weighing Units “kg”,“g. Weight Unit Display Area Indication Weight Unit

EXCELL PRECISION CO., LTD.

1-2 Rear Panel Terminal Block Calibration Switch � Calibration Switch set to the left is “OFF” and to the right is “ON” � 13 Way Terminal Block 1st : : : : FG 2nd : : : : AC IN 3rd : : : : AC IN 4th : : : : NC 5th : : : : Current loop in 6th : : : : Current loop out 7th : : : : TXD 8th : : : : RXD 9th : : : : SG 10th : : : : EXC + 11th : : : : EXC - 12th : : : : SIG + 13th : : : : SIG -

EXCELL PRECISION CO., LTD.

1-3 Keypad Description : Exit key when setting parameters or calibration mode Generally, press this key to enter into or exit standby mode Enter into standby mode:all the indication (except HI indication “3”) and data output is closed Exit standby mode:restart the indicator : Press this key to recall the memorized HI times, LO times, OK times on the display : Press this key to clear all the memorized data : Press this key to clear all the memorized times Setting the qualified times of the indicator Mode 0: when reaching the qualified time, the signal 4 will be activated and output Mode 1: when reaching the qualified time, the operation process stops : Press this key to confirm all the setting are saved � Press and hold key and then press key to rezero the indicator � Switch on the indicator, press and hold key and then press key to enter into calibration mode. � Switch on the indicator, press and hold key and then press key to enter into function setting mode � When the indicator is switched on and counting down to zero, press and keys to restore all parameters to default value. � When the indicator is switched on and counting down to zero, press and keys to enter into testing mode. : Press this key to set the check-weighing High limit and Low limit Qty MR MC F Limit Enter Run Stop F Limit Enter Run Stop MC Enter Enter Enter Enter Enter

EXCELL PRECISION CO., LTD.

1-4 A/D Conversion Input Sensitivity Internal Resolution Max. Sampling Speed Application Range Load Cell Excitation Voltage : Over 0.12 µV/d : 1 / 1 000 000 : 120 times/s : - 0.1 ~ 4.0 mV/V : 5 V DC ± 5%, 120 mA (Up to eight 350 Ω load cells can be connected) 1-5 Power Supply ♦ AC 85 ~ 265 V 50 / 60 Hz ♦ Power consumption is about 10 W

EXCELL PRECISION CO., LTD.

1-6 Dimensions (mm) 154 140 92 45 48 96

EXCELL PRECISION CO., LTD.

First serial port interface Function setting CHAPTER 2 GENERAL FUNCTION 2-1 Function Setting Procedures Hold key and then press key after power on ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key Refer to 5-1 for details Qty MR F Limit Enter Run Stop MC F Limit Enter F Refer to 2-2 for details Enter Enter

EXCELL PRECISION CO., LTD.

2-2 Function Setting Press key Input desired parameter code Press key Display shows the previous parameter Input the desired new parameter, Press key To continue the next function setting Or press key to exit ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key *Function Parameter Code ⇒ Digital filter I ⇒ Digital filter II ⇒ Lock keypad function ⇒ Password setting function ⇒ Output signal (1) open time ⇒ Output signal (2) open time ⇒ Output signal (3) (4) open time ⇒ Input signal (1) reaction time ⇒ Input signal (2) reaction time ⇒ Sorting and falling internal time ⇒ Check-weighing time ⇒ Weight comparison parameter zero range ⇒ Weight sorting function setting ⇒ Display rewrite frequency ⇒ Turning on zero setting ⇒ Sorting mode setting Run Stop Enter Enter Enter Qty MR F Limit Enter Run Stop MC

EXCELL PRECISION CO., LTD.

� � � � FNC Group function setting Setting value Item Function Parameter Description Default 0 5 Hz 1 4.17 Hz 2 2.5 Hz 3 2.08 Hz 4 1.25 Hz 5 1.04 Hz 6 0.63 Hz 7 0.52 Hz 8 0.31 Hz FNC-01 Digital Filter I 9 0.26 Hz 4 0 Disable 1 2 3 4 FNC-02 Digital Filter II 5 weak strong 2 0 Normal (lock disable) FNC-03 Lock Keypad 000 000 ↓ 111 111 1 Close (lock enable) The bits and front panel key positions are related to each other 000 000 0 No password FNC-04 Password setting 1 Set password 0 FNC-05 Output signal (1) open time 0.0 ~ 25.5 s Set output signal (1) 0.2 FNC-06 Output signal (2) open time 0.0 ~ 25.5 s Set output signal (2) 0.2 FNC-07 Output signal (3) (4) open time 0.0 ~ 25.5 s Set output signal (3) (4) 0.3 FNC-08 Input signal (1) reaction time 0.0 ~ 25.5 s Set delay time of input signal (1) between reaction and operation 0.0

EXCELL PRECISION CO., LTD.

Setting value Item Function Parameter Description Default FNC-09 Input signal (2) reaction time 0.0 ~ 25.5 s Set delay time of input signal (2) between reaction and operation 0.2 FNC-10 Sorting and falling internal time 0.0 ~ 25.5 s Internal time of each sorting and falling time 0.1 FNC-11 Check-weighing time 0.0 ~ 25.5 s Set check-weighing time 0.1 FNC-12 Weight comparison parameter zero range 0 ~ 255 d Set zero range under check-weighing mode (the number below the range is set as 0) 2 0 Memory ordering function disable Single input and output enable FNC-13 Weight sorting function setting 1 Check-weighing and ordering function enable 12 memorized weights at most 0 0 No limit 1 20 times/s 2 10 times/s 3 5 times/s FNC-14 Display rewrite frequency 4 1 times/s 0 0 Disable FNC-15 Turning on zero setting 1 Enable 0 FNC-16 Check-weighing mode setting 0 - 1 Refer to Check-weighing mode flowchart description 0

EXCELL PRECISION CO., LTD.

CHAPTER 3 CALIBRATION DESCRIPTION 3-1 Load Cell Connection 2 2 2 2 When using a 4-wire cable to connect the load cell, the SEN+ and SEN- can be left unconnected (see below diagram) Shield SIG+ Load cell cable SIG- EXC- Load cell EXC+ CAL 8 7 7 8 10 10 9 11 13 12 ON

EXCELL PRECISION CO., LTD.

3-2 Parameter Setting And Calibration Flowchart Press key Specification calibration Calibration switch ON Calibration completed General Calibration Zero point voltage calibration Weight calibration Calibration switch ON Limit Enter Run Stop

EXCELL PRECISION CO., LTD.

Specification calibration General calibration Enter into general calibration mode Refer to 3-4 for details Enter into specification calibration mode Refer to 3-3 for details 2 2 2 2 Calibration Process Press key ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key Qty MR F Limit Enter Run Stop MC Limit Enter Enter Enter

EXCELL PRECISION CO., LTD.

3-3 Specification Calibration Press key Input desired parameter code Press key Display shows the previous parameter Input the desired new parameter, Press key To continue the next function setting Or press key to exit ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key *Calibration Parameter Code ⇒ Unit ⇒ Decimal point ⇒ Min. Division ⇒ Max. Capacity ⇒ Zero range ⇒ Time of zero tracking ⇒ Range of zero tracking ⇒ Investigation period of stability ⇒ Investigation range of stability ⇒ Zero and tare function when the weight is unstable ⇒ Tare function availability when gross weight is negative Qty MR F Limit Enter Run Stop MC Enter Run Stop Enter Enter

EXCELL PRECISION CO., LTD.

� � � � FNC Group function setting Setting value Item Function Parameter Description Default 0 None 1 g CSP-01 Unit 2 kg 1 0 None 1 1 decimal point 2 2 decimal point CSP-02 Decimal point 3 3 decimal point 1 1 2 5 10 20 CSP-03 Division 50 Division size 1 CSP-04 Max. Capacity 999 999 ↓ 000 000 Division size 999 999 CSP-05 Zero range 0 ~ 30 0 ⇒ full range 1 ~ 30 ⇒ ±1% ~ ±30% Zero range = calibration zero point ± (Max. capacity×setting value %) 0 CSP-06 Time of zero tracking 0.0 ~ 5.0 s Time and range of zero tracking should be use at the same time. If the time is set to 0.0, the zero tracking function is disabled 1.0 CSP-07 Range of zero tracking 0 ~ 9 Range of zero tracking = (setting value×½)D, D=division Range and time of zero tracking should be use at the same time. If the range is set to 0, the zero tracking function is disabled 9 CSP-08 Investigation period of stability 0.0 ~ 5.0 s Investigate time and range should be use at the same time. If the time is set to 0.0, the investigate time is disabled 2.0 CSP-09 Investigation range of stability 0 ~ 9 Investigate time and range should be use at the same time. If the range is set to 0, the investigate range is disabled 2 0 Enable CSP-10 Zero and tare function when weight is unstable 1 Disable 0 0 Enable CSP-11 Tare function available when gross weight is negative 1 Disable 0

EXCELL PRECISION CO., LTD.

3-4 General Calibration Press key to enter into general calibration mode Press key Zero Calibration No weight on the platform or in the hopper Press key After 5 seconds After 2 seconds Weight Calibration Use the front panel to key in the weight value Place the weight on the platform or inside the hopper After the weight is stable Press key After 5 seconds Press key General calibration completes � Zero calibration only, press key to escape after the display shows . � Span calibration only, press key entering directly to span calibration after the display shows . � Please refer to error message during calibration of the display show . X . ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key kg kg . . . . . kg . . . . . kg Enter Qty MR F Limit Enter Run Stop MC Run Stop Run Stop Run Stop Enter Enter Enter

EXCELL PRECISION CO., LTD.

3-5 Error Messages . ⇒ Load Cell output voltage is lower than - 0.1mV/V or higher than 4 mV/V . ⇒ Weight value ≤ previous weight value . ⇒ Actual measured weight value ≤ previous weight value . ⇒ Setting value is 0 . ⇒ mV/V value entered is higher than investigation range . ⇒ mV/V value entered is too small (SPAN – Zero is lower than 0 mV/V) . ⇒ Displayed resolution is less than 0.12 µµµµV/d When the above messages appear, please contact your supplier for professional help and service.

EXCELL PRECISION CO., LTD.

CHAPTER 4 INTERFACE DESCRIPTION Serial input / output interface (built-in) � � � � Pin location and setting 1. Built-in RS-232 and Current-loop � � � � Function setting First serial port interface setting procedures Press key Input desired parameter code Press key Display shows the last setting Input a new desired parameter Press key Continue to next function setting or press key to exit PIN Function 5 Current loop in 6 Current loop out 7 TXD 8 RXD 9 SG ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key Qty MR F Limit Enter Run Stop MC Enter Run Stop Enter Enter

EXCELL PRECISION CO., LTD.

Setting value Item Function Parameter Description Default 0 As display 1 Gross only 2 Net only 3 As display (simple) 4 Gross (simple) 5 Net (simple) 6 Comparison + As display (simple) 7 Comparison +Gross (simple) 8 Comparison +Net (simple) 9 Tare RS1- 01 Transmit format 10 Totalised (Accu.) Weight and transaction times 0 0 Transmit continuous + command mode 1 Auto transmit + command mode 2 Manual transmit + command mode RS1- 02 Transmit mode 3 Command mode 4 0 600 1 1 200 2 2 400 3 4 800 4 9 600 RS1- 03 Transmit speed 5 19 200 4 0 N,8,1 No parity,8 data bits,1 Stop bit 1 O,7,1 Odd parity,7 data bits,1 Stop bit RS1- 04 Parity Bit length Stop Bit 2 E,7,1 Even parity,7 data bits,1 Stop bit 2 0 Unlimited 1 1 time/s 2 2 times/s 3 5 times/s RS1- 05 Transmit times 4 10 times/s 0 RS1- 06 Transmission conditions 000 000 RS1- 07 Indicator poling address 00 ↓ 99 When set to 0, Indicator addressing is not used 0 Negative (net weight) Weight unstable Overload (OL) 0 ⇒ transmit continuously 1 ⇒ Stop transmit 0 0 0 0 0 0

EXCELL PRECISION CO., LTD.

� � � � Data format 1. General Format GROSS S T , G S , + 0 1 2 3 4 5 6 k g NET S T , N T , + 1 2 3 4 . 5 6 g TARE S T , T R , + 0 1 2 3 4 5 6 t + OL O L , G S , + SP SP SP SP SP SP SP SP SP - OL O L , G S , - SP SP SP SP SP SP SP SP SP UNSTABLE U S , G S , + 1 2 3 4 . 5 6 k g CR LF 2. Totalised (Accu.) Format Accu. Weight T W , + 1 2 3 4 5 6 . 7 8 9 k g Accu. Wt. Over+ T W , + SP SP SP SP SP SP SP SP SP SP SP SP Accu. Wt. Over - T W , - SP SP SP SP SP SP SP SP SP SP SP SP Accu. Count T N , + 0 1 2 3 4 5 6 7 8 9 SP SP Accu. Count over T N , + SP SP SP SP SP SP SP SP SP SP SP SP CR LF 3. Sample Format Gross/Net or as display + 1 2 3 4 5 6 Over load positive + SP SP SP SP SP SP Over load negative - SP SP SP SP SP SP CR LF 4. Comparison Condition (1) + Simple Format (Gross/Net or as display) + 1 2 3 4 5 6 CR LF bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0 bit 0 : Zero Band bit 1 : Over bit 2 : Under / HI bit 3 : SP1 / OK bit 4 : SP2 / LO bit 5 : SP3 bit 6 : Unloading bit 7 : Batch finish

EXCELL PRECISION CO., LTD.

5. Comparison condition (2) Byte 0 Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7 Byte 0 : Zero Band Byte 1 : Over Byte 2 : Under / Hi Byte 3 : SP1 / Go Byte 4 : SP2 / Lo Byte 5 : SP3 Byte 6 : Unloading Byte 7 : Batch finish Description Output ASCII Description OL 4FH、4CH Over load ST 53H、54H Weight stable Status 1 US 55H、53H Weight unstable GS 47H、53H Gross Weight NT 45H、54H Net Weight Status 2 TR 54H、52H TARE TW 54H、57H Totalised Weight TC 54H、43H Totalised Times 0 ~ 9 30H ~ 39H Figure of weight +、- 2BH、2DH Symbol (+ or -) of weight Space 20H Over load Data of Weight . 2EH Decimal Space, Space 20H、20H None Units kg 6BH、67H kg Ending code CR、LF 0DH、0AH Ending code Separating code , 2CH ON : 0 ( ASCII Code 30 H ) OFF : 1 ( ASCII Code 31 H )

EXCELL PRECISION CO., LTD.

� � � � Command mode 1. Command Format A Host Command Slave Command MZ Zero CZ Zero compensation On/OFF MT Tare CT Clear TARE value MG Gross Weight MN Net weight AT Accu. Current net weight and times plus 1. ST Deduct times of last accu. Value minus 1 DT Clear accu. Value and times BB Start batching (one time) HB Load stop BD Start unload SC Transmit continuous SA Auto transmit SM Manual transmit SO Command mode % Stop continuous transmission and enter the command mode 2. Command Format B Host Command < LF> Slave Data < LF> RW Read current weight RT Read TARE RG Read Gross Weight RN Read Net weight RB Read current display of wt (simple) RH Read Gross (simple) RI Read Net (simple) RJ Read comparison situation + current display of weight (simple) RK Read comparison situation + Gross (simple) RL Read comparison situation + Net (simple) RO Read comparison situation (2) RF Read prior completed weight RA Read accu. Value (incl. times) Note : : : : Prior command plus % Read Weight Compared value: RS□□ □□: Setting items FW Read target item of unload value S1 Read SP1 S2 Read SP2 S3 Read SP3 UD Read Under LO Read LO ZB Read Zero Band HI Read HI PR Reading Peak value OV Read Over Ex: Command : RSFW<CR><LF> EX2005 reply : RSFW□□□□□□ Finish 6 bytes

EXCELL PRECISION CO., LTD.

3. COMMAND FORMAT C Host Command + Data < LF> Slave Command + Data < LF> Write weight compared value WS□□XXXXXX XXXXXX : value (6 bytes) □□ : Setting items FW Write target item of unload value S1 Write SP1 S2 Write SP2 S3 Write SP3 UD Write Under LO Write LO ZB Write Zero Band HI Write HI PR Write Peak value OV Write Over � � � � Error messages E1: Format command fault E2: Setting parameters over range E3: Command not recognised � � � � Indicator poling address If the indicator has an address configured in RS1 – 07 as 0, it will directly respond to the command. If the indicator has an address configured in RS1-07 as 1~99, it will only respond to the messages prefixed the address. For example: The indicator poling address is set to 02, it would send the weight value only if it received the command: @02RW<CR><LF>

EXCELL PRECISION CO., LTD.

� � � � Equivalent Circuits � Open or short IN 0 ~ IN 7 and COM 1 to change input signal status Open ↔ OFF ,Short ↔ ON. � AC or DC connection input terminals. Input

EXCELL PRECISION CO., LTD.

CHAPTER 5 MAINTENANCE 5-1 Restore All Parameters To Default Factory Values (1) While the indicator is counting back to zero, adjust SW to ON and press (2) Display shows the flashing digits . (3) Confirm / abort (3-1) To confirm press key & don’t release it until the display shows , then release the key and return the calibration SW to OFF. (3-2) To abort, press key to exit. ⇒ Increase the flash value by one ⇒ Decrease the flash value by one ⇒ Move the cursor leftward ⇒ Move the cursor rightward ⇒ Confirm key ⇒ Exit key Qty MR F Limit Enter Run Stop MC Enter Run Stop F Enter

EXCELL PRECISION CO., LTD.

5-2 Test Mode While the indicator is counting down to zero Press . . . . . 7 segment display testing. Refer to 5-2-1 for details ⇒ Keypad and calibration SWITCH testing. Refer to 5-2-2 for details ⇒ Display A/D internal value Refer to 5-2-3 for details ⇒ RS-232 serial loop back testing. Refer to 5-2-4 for details ⇒ EEPROM memory testing Refer to 5-2-5 for details ⇒ Factory testing (reserved) ⇒ RELAY_IO interface card testing . ⇒ . Enter Run Stop F Limit F Limit F Limit F Limit F Limit F Limit

EXCELL PRECISION CO., LTD.

5-2-1 7 Segment Display Testing The display will show ~ , then display “.” and all of the icons. 5-2-2 Keypad And Calibration SW Testing Setting the calibration SW to “ON”, or pressing any key will cause the related display segment to change from → . 5-2-3 Display A/D Internal Value Display Display range is 0 ~ 520,000d (-0.1 ~ 4.0 mV/V). 5-2-4 RS-232 Serial Loop Back Testing Terminal pin 5 and pin 6 must be connected together at the rear of the indicator. If display shows , the interface is working normally. If display shows , the interface is not working normally. 5-2-5 EEPROM Memory Testing If the display shows , it means normal. If the display shows , the memory is not working correctly. 5-2-6 IFC Factory Testing (reserved) 5-2-7 RELAY_IO Interface Card Testing The display shows the indication light of corresponding input pins. When the input signal and IN_COM is short, the bit indication will change from → , press key to shift corresponding output pins and OUT_COM to short status. Limit

EXCELL PRECISION CO., LTD.

APPENDIX 7-SEGMENT DISPLAY CHARACTERS Number Display Letter Display Letter Display 0 A N 1 B O 2 C P 3 D Q 4 E R 5 F S 6 G T 7 H U 8 I V 9 J W K X L Y M Z