Anritsu MP1777A Operation Manual
Read Online
Read the Anritsu MP1777A Operation Manual online using the built-in PDF viewer.
Text of the manual
10 GHz Jitter Analyzer
Remote Control
Operation Manual
Vol. 2
Fourth Edition
To ensure that the equipment is used safely, read
the "For Safety" in the MP1777A 10 GHz Jitter Ana-
lyzer Operation Manual first.
Keep this manual with the equipment.
ANRITSU CORPORATION
Document No.: M-W1493AE-4.0
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the follow-
ing safety symbols to indicate safety-related information. Insure that you clearly understand the meanings of the
symbols BEFORE uslng the equipment.
Some or all of the symbols may not be used on this equipment. In addition, when drawings are included in this
manual, labels on the equipment may not be shown on them.
Safety Symbols Used in Manual
This indicates a very dangerous procedure that could result in death or serious
DANGER A injury if not performed properly.
This indicates a hazardous procedure that could result in death or serious injury if
A
not performed properly.
This indicates a hazardous procedure or danger that could result in light-to-severe
A
injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and/or in Manual
The followmg safety symbols are used inside or on the equipment near operation locations, andlor in manual to
provide information about safety items and operation precautions. Insure that you clearly understand the rnean-
ings of the symbols and take the necessary precautions BEFORE using the equipment.
This indicates a prohibited operation. The prohibited operation is indicated sym-
bolically in or near the barred circle.
0
This indicates an obligatory safety precaution. The obligatory operation is indi-
cated symbolically in or near the circle.
n
This indicates warning or caution. The contents are indicated symbolically in or
near the triangle.
n
U
This indicates a note. The contents are described in the box
These indicate that the marked part should be recycled.
MP1777A
10 GHz Jitter Analyzer Remote Control
Operation Manual Vol. 2
10 September
1998 (First Edition)
20 September
2002
(Fourth Edition)
Copyright O 1998-2002, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the
publisher.
The contents of this manual may be changed without prior notice.
Printed in Japan
ii
1. ALWAYS refer to the operation manual when working near locations at
which the alert mark shown on the left is attached. If the operation, etc.,
A
is performed without heeding the advice in the operation manual, there
is a risk of personal injury. In addition, the equipment performance may
be reduced.
Moreover, this alert mark is sometimes used with other marks and de-
scriptions indicating other dangers.
2. When supplying power to this equipment, connect the accessory 3-pin
power cord to a 3-pin grounded power outlet. If a grounded 3-pin outlet
is not available, before supplying power to the equipment, use a conver-
or
sion adapter and ground the green wire, or connect the frame ground on
the rear panel of the equipment to ground. If power is supplied withcut
grounding the equipment, there is a risk of receiving a severe or fatal
electric shock.
3. This equipment cannot be repaired by the user. DO NOT attempt to
open the cabinet or to disassemble internal parts. Only Anritsu-trained
Repair
service personnel or staff from your sales repr.esentative with a knowl-
edge of electrical fire and shock hazards should service this equipment.
There are high-voltage parts in this equipment presenting a risk of se-
vere injury or fatal electric shock to untrained personnel. In addition.
there is a risk of damage to precision parts.
4. This equipment should be used in the correct position. If the cabinet is
Falling Over
turned on its side, etc., it will be unstable and may be damaged if it falls
over as a result of receiving a slight mechanical shock.
5. DO NOT short the battery terminals and never attempt to disassemble it
Battery Fluid
or dispose of it in a fire. If the battery is damaged by any of these ac-
tions, the battery fluid may leak.
This fluid is poisonous.
DO NOT touch it, ingest it, or get in your eyes. If it is accidentally in-
gested, spit it out immediately, rinse your mouth with water and seek
medical help. If it enters your eyes accidentally, do not rub your eyes,
irrigate them with clean running water and seek medical help. If the
liquid gets on your skin or clothes, wash it off carefully and thoroughly.
LCD
6. This instrument uses a Liquid Crystal Display (LCD); DO NOT subject
the instrument to excessive force or drop it. If the LCD is subjected to
strong mechanical shock, it may break and liquid may leak.
This liquid is very caustic and poisonous.
DO NOT touch it, ingest it, or get in your eyes. If it is ingested acciden-
tally, spit it out immediately, rinse your mouth with water and seek medi-
cal help. If it enters your eyes accidentally, do not rub your eyes, irrigate
them with clean running water and seek medical help. If the liquid gets
on your skin or clothes, wash it off carefully and thoroughly.
Disposing the batteries
1. Before changing the fuses, ALWAYS remove the power cord from the
poweroutlet and replace the blown fuses. ALWAYS use new fuses of
the type and rating specified on the fuse marking on the rear panel of the
cabinet.
T _-A indicates a time-lag fuse
There is risk of receiving a fatal electric s h o c ~
if the fuses are replaced
with the power cord connected.
2. Keep the power supply and cooling fan free of dust.
Clean the power inlet regularly. If dust accumulates around the power
pins, there is a risk of fire.
Keep the cooling fan clean so that the ventilation holes are not ob-
structed. If the ventilation is obstructed. the cabinet may overheat
and catch fire.
3. Use two or more people to lift and move this equipment. or use a trolley.
There is a risk of back injury, if this equipment is lified by one person.
4. This equipment uses a lithium battery to back-up the memory. This bat-
tery must be replaced by a servlce engineer when it has reached the end
of its useful life; contact the Anritsu sales section or your nearest repre-
sentative.
NOTE: The battery used in this equ~pment has a maximum useful life
of 7 years. It should be changed before this period has
elapsed.
5. The main unit of the MP1777A uses lithium batteries. When d~sposing of
the batteries, make sure to conform with the local regulation.
Anritsu Corporation Contact
If [hi\ t.ql!rpmtmt ilc\elop\ :I t;tt~it. coiitirit .-\r:r~t\u Coqmrar~w or- ~ t \
repsc\i.nt:\:r\c\
itt thz aJdrc\> In tills iilarirul.
Notes On Export Management
-
-
Th~s product and its manuals may require an Export LicenseiApproval by
the Government of the product's country of origin for re-export from your
country.
Before re-exporting the product or manuals, please contact us to confirm
whether they are export-controlled items or not.
When you dispose of export-controlled items, the products/manuals are
needed to be brokenlshredded so as not to be unlawfully used for military
purpose.
Trademark and Registered Trademark
I
Quick Basic is a registered trademark of Microsoft Corporation.
I
viii
Composition of the MP1777A Operation Manual
VOL 1
-
- -
--
--
---
I
i
Operat~on manual for the maln un~t
MP1777A
L - - - -
-
-2
--
4
I Operat~on Manual
, -
.
- -
-
-_
--
_ __ _ - - -
L
I VOL 2
- - -1
I
Operat~on manual for remote control
Table of Contents
For Safety ............................................................
iii
Composition of the MP1777A Operation Manual ..
I
Section 1 Outline ................................................ 1-1
1 . 1 Interface Function .................................................................... 1-2
...............................................................
.
1 2 System Setup Example
1-3
....................................
Section 2 GPlB Interface
2-1
............................................................
2.1 GPlB Interface Functions
2-2
2.2 Device Message List ..................................................................
2-3
...........................................................................
2.3 Bus Commands
2-4
........................................................
2.4 Connecting the GPlB Cable
2-5
2.5 Setting GPlB ................................................................................ 2-6
2.6 Initializing the Device ................................................................... 2-8
Section 3 Listener Input Format ........................ 3-1
3.1 Listener Input Program Message Format .................................... 3-2
.................................
3.2 Funct~onal Elements of Program Message
3-4
3.3 Program Data Format ................................................................
3-14
Section 4 Talker Output Format ........................ 4-1
4.1 Talker Output Response Message Format ................................. 4-2
4.2 Functional Elements of Response Message ............................... 4-3
Section 5 Common IEEE488.2 Commands ...... 5-1
....................................................................
5.1 Common IEEE488.2
5-2
Section 6 Status Report ..................................... 6-1
...............................
6.1 Conficjurat;on of hlP1 777A Status Registers
6-2
6.2 Status Registers Spec~fied by IEEE438.2 ................................... 6-4
6.3 Status Register Specifled by SCPl .............................................. 6-6
6.4 Status Register Specific to MP1777A ......................................... 6-7
6.5 Reading. Writing and Clearing Status Reg~sters
......................... 6-8
Section 7 SCPl Outline ....................................... 7-1
.........................................................................................
7 1 Outline
7-2
7.2 Command Structure ...................................................................
7-3
7.3 Command Description Method ...............................................
7-4
-
- -
/ . 4 Compounding Commands ...................... ...
.............................. i -.J
.............................................................................
7.5 Parameter
.-- 7-6
Section 8 Device Message Details .................... 8-1
......................................................................
8.1 Response Format
6-2
8.2 Buffer Size Stipulations ............................................................
8-5
8.3 Devics-Specific Comma~ds
................................ ..
.................. 6-5
Appendix A SCPl Error Messages ................... A-1
.......................................................................
A 1 Command Errors
A-2
..........................................................................
A.2 Fiun Time Errors
A-4
A.3 Device-spec if:^ Errors ......................... ...
.......................... A-5
.............................................
A.4 Query Errors ............................... ..
A-6
Appendix B
Details of initialization Command
Functions ...................................... B-1
. .
B . l I r ~ ~ t ~ a z ~ n g
Bus ..............................................................................
B-2
...................................................................
B.2 Iriitiahz~r.
g Messqge
B-3
......................................................................
E.3 Icir~alizin~
Device
B-4
G.4 Device Status at P.2 wer Act~vatian ...........................................
2-5
n 1 Outline
1 . I In!s:face Function ............................ ..
..........................................
1-2
1 .2 Syskn; Setup Example ................... ..
...........................................
1-3
Section 1 Outline
Interface Function
The MP1777A has a connector for remote control on the back face.
Remote control of devices is enabled by fitting GPIB.
The GPIB setting is performed on the Secondary system screen in the Main setup screen.
GPIB interface: The GPIB interface for this device conforms to the IEEE (Institute of Electrical and Electronic Engi-
neers) standards, 488.1-1987. The software conforms to the standards IEEE488.2 and SCPI (Standard
Commands for Programmable Instruments). (See the Section 7 for more infom~ation.)
This device incorporates the following interface functions.
Control of functions except for certain functions such as power source switch and Local key.
Reading of all the setup conditions and screen display.
1.2 System Setup Example
Control from a host computer
\i'hec conrlec,tecl to :I hn\t C O I I I ~ L I ~ L , ~ .
~ t l r dc\.ice ptsforrllh ;iutoril:ttlc mc~~\i~rt.rllc:~il.
Host compllter
Fig. 1-1
Block Diagram
Section 2 GPlB Interface
2.1 GPlB Interface Functions .................................................................
2.2 Device Message List ........................................................................
................................................................................
2.3 Bus Commands
2.4 Connect~ng the GPlB Cable .............................................................
2.5 Sett~ng GPlB ...................................................................................
. .
2.6 lnitial~z~ng
the Device .......................................................................
. .
2.6.1
ln~t~alizing
the Bus ..............................................................
....................................................
2.6 2
lnltializing the Message
2.6.3
Initializing the Device .........................................................
2.6.4
Device Status at Power Activation .......................................
Section 2 GPIB lnterface
2.1
GPIB lnterface Functions
This device only functions as a device and does not function as a controller. Accordingly, the GPIB interface has the
following functions.
Table 2-1 GPIB lnterface Functions
Cord
PPO
DC I
Interface Function
All the source handshake functions is available.
IEEE488.2 Standards
Incomoration of all the functions as the standard features
All the acceptor handshake functions is available.
Basic talker functions is available.
Serial port function is available.
Talk only mode function is available.
Talker cancellation function is bv MLA available.
Incorporation of all the functions as the standard features
The device shall incorporate one of the following sub-
sets: TS, T6, TES or TE6.
Basic listener functions is available.
No listen only mode function is available.
Listener cancellation function is bv MTA available.
The device shall incorporate one of the following sub-
sets: L3, L4, LE3 or LE3.
All the servlce request functions is available.
All the remote and local functions is available.
All the device trigger functions is available.
/ DTO (no functions) or DT1 (all functions)
Incorporation of all the functions as the standard features
RLO (no functions) or RLI (all functions)
No parallel port function is available.
All the device clear functions is available.
PPO (no functions) or PPI (all functions)
Incorporation of all the functions as the standard features
No system controller functions is available.
CO (no functions). C4 and C5, or one of the following
sub-sets: C7, C9 or C 1 1.
Device Message List
2.2
Device Message List
I)<\ icc Ilrc\\ase\ Lire d,lta nlc.;;i_rc\
exclinrlgcd l~,et\iezr-i tt~e coii11-01lcr ,111d ~ l i c dev~ce vi;~ the a! \ten1 interl'acc \\ her1 the
hu\ i 4 wt I(> tht, (I:II:I ri~oile ( ~ i
liv~l the :\I U Iirii' i \ \et 31 "H" ). :md art. ci:tsaified Into t\vo t!peh. the proyrmi rrlt.sqe\ arid
r-c\po~i\e me\\age\.
1 ' 1 1 ~ p r o p r n rnt.i.;'tgei .IK ;\SC'II d,ita rne\\;ige\ tr-an\fcrreil fron~ the controller ti: the tic\ lcc, \+I~llc
the re\;>oriw r n ~ ~ v
\LI;C\
:Ire ~1:il.i I I I C \ \ : I _ ~ ' ~ \ t~-;~ri\t~r~ecl
from the tlevlce to the conlrollcr.
I':.cyrxn x ~ i l
re\ponw lile\\:ige\ :ire further cla\\lfied Into the fc~llo\i 111s
ille\hapc typcs.
Table 2-2 Device Message List
Program Message
(See the Section 4 for more information.)
~'I-illll lll\tructl~~ll\
Pro,
C'~>rnl~i;iriil\
~lniqut' 10 the ile\ i ~ . e
I See the Stx,t~on 7 !'or 111c)re ~nft?rrilatior;.
1
(~'oriinion IEEE48X.2 C L ) I I I I I ~ ; I I ~ ~ \
(See the Sectii>r~
5 "Si:\tu\ Repor\" for iticvc intor
r l 1 ~ l l l t ~ I l .
1
1'1-ocrirni clueric\
Response Message
(See the Section 5 for more information.)
Statu\ rrle\\agc\
(See the Sectic~n
(3 "SCPI Outlinc' for 111111.2 irifor7n;1[1o1i
)
Kc\ponw n!t.\sagc\
I
I
I
Program m e s s a c e
I
I
n
-
- tt?===d
-
<--PI
Responce message
r
y
r.3
I
Fi6. 2-1 Device Message
i'he mei\;tgcL; rllt:~it~oried :iho\e are t ' \ ~ h a l i ~ ~ ~ l
thro~lg.h thc 110 hut'kr- i:f tht. de\,ice. The follc)\+ing xx.[lori pro\ i..!c\ .I
h~net' de\crip~~uri
ot ~iie IT) b ~ i f t ~ r .
Table 2-3 I:0 Buffer
r
p
u
t
suffer
I
Output Queue --
.\ f I F 0 I ~ - I S \ [ (11 ~.'II-\I OLI!) type ri;eri~rr> AYC;I 1 1 1 [ C I I ~ ~ O -
I .,\ f . i l . 0 !ypt, siie!~~(.>i.>
:ir.~:t for [lie C \ : . ~ L I ~ ~ .
:\\I
ill? L>-\L3',
Section 2 GPIB Interface
2.3 Bus Commands
Bus commands refer to internal interface communications exchanged when the bus is set to the command mode (when the
ATM line is set at "L"). The table below shows a list of bus commands.
Table 2-4 Details of Bus Commands
Bus Command
Operations
DCL (Device clear)
Initializes the exchange of messages among all the devices connected to the GPIB
bus.
SDC (Selected Device Clear)
Initializes the exchange of messages among addressed devices. The operations are
the same as those for DCL.
GET (Group Execute Trigger)
Performs the same operations as those triggered when the Start/Stop key is pressed
IFC (Interface Clear)
Init~allzes the ~nterface.
Section 2 GPlB Interface
2.5 Setting GPlB
To use GPIB as a remote interface, set the device to the local mode and set the address by performing key input on the
front panel.
Table 2-5 Details of Setting when GPlB lnterface is Used
Details of Setting
I
Setting Item
Perform the setting shown above on the Secondary system screen in the hhin setup screen.
Setting Range
Address setting
(1) (a) Pressing the Setup key causes the Main setup screen to open.
The Main setup screen has "Setup" indicated on the left top section of the screen.
S e t u ~
I Interface
1 ITx&Rxl
Time XX : XX : XX XWXXWXX
Address
Bit Rate
[ 2,488 M ]
Jitter Setting: Equalizer
[
OFF
]
Clock
[ Internal
]
0 to 30
Fig. 2-3 Main setup Screen
(b) Moving the cursor to the shaded section of the figure above and pressing the Setup key causes a window as
shown below to open. To select the Secondary system screen, move the cursor onto "System" using the cursor
key and press the Set key.
Fig. 2-4 Screen Selection Window
2.5 Setting GPlB
Buzzer
[
OFF
1
Date & Time adjust
[ 00 : 11 : 17 01iJan195 ]
Setup
System
. -
-- - Address
1
Fig. 2-5 Secondary System Screen (When GPlB Interface is Used)
Section 2 GPIB Interface
2.6
Initializing the Device
IEEE488.2 st~pulates three levels of system initialization: initialization of the bus, message and device.
Table 2-6 Types of System lnitialization
Level
Type of Initialization
Outline
1
Initialization of the bus
Initializes the functions of all the interfaces connected to the bus with the IFC
message from the controller.
2
Initialization of the message
Disables the function to report to the controller that the initialization of the
exchange of messages among all the devices on GPIB and relevant operations
are completed by DCL, one of the GPIB bus commands, and that the initial-
ization of the exchange of messages among specified devices on GPIB and
relevant operations are completed by SCL, also one of the GPIB bus com-
mands.
3
Initialization of the device
Returns the device to the original state unique to the device by *RST, regard-
less of the status during past usage.
2.6.1
lnitializing the Bus
IFC:
Initialization of the bus by the IFC statement
Function:
Switches the IFC line into an active state for about 100 ps and initializes the interface function of all the
devices connected to the GPIB bus line. IFC can be transmitted only by the system controller.
2.6.2
lnitializing the Message
DC'L and SDC: Initialization of the exchange of messages by the DCL and SDC bus commands
Initializes the exchange of messages for all the devices on GPIB or specified devices on GPIR.
Functions:
Initialization of the message exchange is aimed at setting preparations to enable the controller to send
new instructions in the event that sections related to the message exchange in devices are set in a state
inappropriate for the control to be executed from the controller because, for example, other programs
were executed earlier, although changing the panel setting is not required.
DCL:
Initializes the message exchange among all the devices on GPIB
SDC:
Initializes the message exchange among specified devices.
2.6 Initializing the Device
2.6.3
Initializing the Device
,KS'I':
Inrti,il~zat~m
of the de\ ici. b> tlic .KS1 cornril,inci
2.6.4
Device Status at Power Activation
ection 3 Listener Input Format
3.1 Listener Input Program Message Format ........................................
3-2
3.2 Functional Elements of Program Message ......................................
3-4
3.2.1
...........................
3-4
3.2.2 iPFiOGRAM MESSAGE TERMINATOR> ..........................
3-5
3.2.3 ......................... ..............
.............................. 3-5
3.2.4 .................................................
3-6
3.2.5 .................. 3-7
3.2.6
..........................................
3-7
3.2.7
and .... 3-8
3.2.8 .................................. 3-10
3.2.9
.......................................
3-12
3.2.1 0 ...............................
3-13
3.2.1 1 ..................................
3-13
3.3 Program Data Format .................... ..
..... ..
..................................... 3-14
3.3.1
................................... 2-15
3.3.2
......................... 3-16
3.3.3
........................................ 3-17
3.3.4 .................................. ....
.
.
3-17
Section 3 Listener lnput Format
3.1
Listener lnput Program Message Format
The following section shows program messages when, for example, 4 UI is selected for the reception jitter and the
reception signal is set to 9953M.
A
v
A
,
: SENS : TEL : RANG U14
S P
T
I /
'
\
\
'
~whlte space,
: SENS :TEL: RANG
\
'\
Listener
(device)
SP
; SP
: SENS : TEL : BRAT ~9953-
,r
i
,
;
I
.r ----
7
SENS TEL RANG
t
i
7
Talker
(controller)
I
I
'\\
'boolean program data>
U14
The program message format comprises a sequence of functional elements divided into the units of minimum level to
express functions. The uppercase letters in angled brackets ( < and > ) in the figure above show examples of functional
elements. The functional elements are further divided into elements called the coding elements. The lowercase letters in
angled brackets ( < and > ) in the figure above show examples of coding elements.
The following pages provide explanations of the program message format using the functional syntax diagram and coding
syntax diagram.
Functional syntax diagram: Graphic representation of selection of functional elements along specific routes
- Coding syntax diagram:
Graphic representation of selection of coding elements along specific routes
3.1 Listener Input Program Message Format
1 hi. URI I F* ~r:d Kf .\I> io11111i~nt1~
t ~ k c
the tolloutng forrn,~tj
pp
-.
pppp-
-
I
WRITE @
Outputs data to the devlce
MFormat
-- -
WRITE @
Dev~ce number Data
Data-+; Arithmetic expression
1 Charxter string expression
~
..
~ .---p---------pp--.-.pp
--
The same as the example given on the previous page.
WRITE @03
- -.--
: ": SENS : TEL : RANG U14 : : SENS : TEL : BRAT M9953"
A
Listener address
iin case that the GPlB address of the 10 GHz J~tter Analyzer is set at 3)
READ @
Subst~tutes the data Input froin the dev~ce to the var~dble
IFormat -
-- --
- - -
-
p
p
-
-
p
--
READ @Dev~ce number Vanable
Substitutes the data input from the device (input tangent~al line setting)
to the variable, A$.
WRITE @03 : ": SENS : TEL : RANG?" t Inquires about the input
READ 0 0 3 : A$
tangential line setting.
i
Listxer address
(irl case that the GPlB address of the 10 G!iz
Jitter Analyzer is set at 3)
Section 3 Listener Input Format
3.2 Functional Elements of Program Message
This measuring instrument accepts the program message by detecting the terminator located at the end of the program
message. The following section explains each functional element of the program message.
3.2.1
The is defined as follows.
The constitutes a data message that incorporates all the functional elements
Refer to 3.2.4
necessary for the controller to send data. to this measuring instrument.
To complete the transfer of the , the is added to
MESSAGE
the end of the .
Refer to 3.2.2
-
Example: to send two instructions
3.2 Functional Elements of Program Message
Section 3 Listener Input Format
The is defined as follows.
SEPARATOR>
Refer to 3.2.5
I I
*
Refer to 3.2.6
The is composed of a sequence of zero, one or multiple
elements.
The elements signify programming instructions or data to be sent from the controller to
this measuring instrument. The is used as the separator for delimitting
multiple elements.
Example: Selects the reception jitter range of4 CTZ and sets the reception signal to 9953M.
3.2 Functional Elements of Program Message
3.2.5
The i n h~te \pace> I\ defined a\ tollou s.
. --
. -
--C?
3.2.6
/-*
Refer to 3 2 7
I
--
7)
- - - _ I
\
C
Section 3 Listener Input Format
3.2.7 and
(1) The is defined as follows.
Example: Sets the data.
\\
I
3.2 Functional Elements of Program Message
----- .
.
/ < F ~ O G R A M '
DATA
'* -
SEPARATOR>
I
I
- -
-
' Refer to 3 2 11 ,
,
.
. --
-----
HEADER>
I
SEPARATOR>
L
- A
\
Refer to 3 2 9,
,Refer to 3 2 lo,,
.
,
.
-
CALC : TGR : DATA? 1995, 3,27,12.54
-- 4
-
-
A
h
b
L
4
Section 3 Listener Input Format
3.2.8
The is defined as follows.
The can be placed before each header.
I
/
\ -
Refer to 3.2.3
\--I program headers I--
I Refer to (2) of 3.2.8 /
Refer to (1) of 3.2.8
( 1 ) The
(2) The is defined as follows.
( 3 ) The is defined as follows.
Refer to (3) of 3.2.8
1
w
Section 3 Listener Input Format
3.2.9
The is defined as follows.
The can be placed before each header.
1
1 6
Refer to 3.2.3
1 Refer to (2) of 3.2.9 1
-
Refer to (1) of 3.2.9
\ -
(1) The are defined as follows.
---,
(2) The is defined as follows.
3.2.9.1
The is a query command, that the controller preliminarily sends to this measuring
instrument so that the controller will be able to receive the response message from this measuring instrument.
It is always characterized by the addition of the query indicator, or a question mark ( ? ), to the end of the header.
Refer to (3) of 3.2.8
The format mentioned above comes with the query indicator, or a questlon mark ( ? ),
added to the end of the header
As other features are the same as those for the , see the Section 3.2.8 "" for more information.
*
Functional Elements of Program Message
3.2.1 1
--
- - 7
- Refer to 3.2 3
---C
- 7---
'
Refer to 3.2.3 I
(
: SYST : DATE
1993,7,15
4
A
b
- r
PAL4 L - A
4 - - ,
3,"- '7
Section 3 Listener Input Format
3.3
Program Data Format
This section provides explanations of the format of illustrated by the functional syntax diagram of
the Section 3.2.7 " and " from among formats of pro-
gram messages terminated as mentioned in the earlier sections.
The functional elements of are used to transfer parameters of various types related to the program
header. The sections shaded in the figure below are the program data used by this measuring instrument.
\I
PROGRAM DATA>
\
Refer to 3.3.1
PROGRAM DATA>
Refer to 3.3.3
BLOCK
PROGRAM DATA>
3.3 Program Data Format
-
Section 3 Listener Input Format
3.3.2
The expresses a numeric value of a decimal digit and is defined as follows.
( I ) The is defined as follows.
(2) The is defined as follows.
This measuring instrument uses the integer type of decimal digits.
Integer type
Expresses the integer values of decimal digits. The underline ( - ) stands for a space.
Zero (0) enabled to be input at the header .............................. 005
Space between a sign and digit disabled ................................ +5 (good), +-5 (no good)
Space enabled to be inserted to follow a numeric value ........ +S- -
The plus sign can be either added or omitted ......................... +5, 5
....................
A comma disabled to be used to separate digits
1.234 (no good)
3.3 Program Data Format
3.3.3
i'lic .;BOOI~E.\N PRC)(iII!\%I l>.\.l',A-. reiri-\ ti) t
!
~
progrnin d:it:~ \prc~l'icd h! SCPJ and cupreswi :t tlicci~.et~cal
\,~liic.
A\ L:IILIC\
t o i ~ ) i ~ - t ' \ p ~ i ~ i
10 t h ~
iriitii :tnd tal\~t\. 0%
aritf OFF ol'the i(~kI.4II;\C"I'i:K l'KOCiR.\L~I [):\.!'A>. :lnd 1 criiil 0 of
[he 1)!X'I%l;\I.
NL'MERIC 1'kOCiK.-\\I I).\T.\.. arc defiried.
I
SOURce TELeconi Jltter GN
SOURce TELecom Jitter
1
ection 4 Talker Output Format
......................................
4.1 Talker Output Response Message Format
4-2
.................................
4.2 Functional Elements of Response Message
4-3
.........................
4.2.1
4-3
.........................
4.2.2
4-3
..............................
...........
4.2.3
...
4-4
.................
4.2.4
4-4
....................
.............
4.2.5
..
4-5
..................................
4.2.6
4-5
4.2.7
........................................................
4-6
Section 4 Talker Output Format
4.1 Talker Output Response Message Format
The figure below shows responses to the query for the selection of the receiving jitter range, : SENS : TEL : RANG?, and
to the query for the sign speed of the received signal, : SENS : TEL : BRAT?.
As the SCPI response has no header, the response only consists of data.
U14
9
M9953
t
) in the figure above show
examples of functional elements. The functional elements are further divided into elements called the coding elements.
The lowercase letters in angled brackets ( < and > ) in the same figure above show examples of coding elements. Accord-
ingly, the same syntactical notational convention is used for the talker and the listener.
I
-
-
-
-
1.. 1
U14 ; M9953
r~---
I
Talker
(controller)
4.2 Functional Elements of Response Message
4.2
Functional Elements of Response Message
4.2.1
l l i c --TI K h I I h 11 P I ) KL 4POh\I kI1-54 IGF> 1s dztlned ,i\ folio\\\
I
4
. -
- --
----+
MESSAGE
Refer to 4.2 3
TERMINATOH>
-- -
Ltsteier '
Talker
U14 ; M9953 cNL>
idevlce) \
1
-
1
-
I
-
-
_ _ --
-d
(coi7troller)
L - -
_ _ A
.
\
I
-
--
\
I
\
-
-
-
_ _ - 7
I Functional elements I
Esiirnple: Itldicates 1111 exi~mple of tltc program to reed the' state of rrceiving,jitter range currently set.
Section 4 Talker Output Format
The is defined as follows.
MESSAGE UNlT
/
SEPARATOR>
\
The is composed of a sequence of one or multiple ele-
ments
The element signifies a single message to be sent from this measuring instrument to the
controller. The element is used as the separator to delimit multiple
elements.
Refer to 4.2.5
Example:
Indicates responses to queries for the selection of receiving jitter range and for the setting of sign speed of received signal.
-
1
4.2.4
The is defined as follows.
The separates with a semicolon ( ; ) that constitutes the , the elements when a sequence of multiple elements is output as one .
4.2 Functional Elements of Response Message
4.2.5
,,I
--.
'
., Refer to 4 2.6 ,
I
--,
--/I
I
I
I
--
- .
d
1
Refer to 3 2 7 1
4.2.6
Section 4 Talker Output Format
4.2.7
The shaded sections in the figure below refer to the used by this measuring instrument
The response data to be returned depends on the query message.
\
RESPONSE DATA>
\
RESPONSE DATA>
\
\
\
\
\
5 Common1 EE488.2 Commands
5.1 Common IEEE488.2 ......................................................................
5-2
Section 5 Common IEEE488.2 Commands
5.1
Common IEEE488.2
The table below lists common IEEE488.2 commands supported by this device.
Table 5-1 List of Common IEEE488.2 Commands
Mnemonic
Description
*IDN?
Identification Query
*OPC
Operation Con~plete Command
*OPC?
Operation Complete Query
*WAI
Wait Continue Command
*CLS
Clear Status Command
*ESE
Standard Event Status Enable Command
*ESE?
Standard Event Status Enable Query
*ESR?
Standard Event Status Register Querq
- -
-- -
-
-SRE
Servlce Request Enable Command
*SRE?
Service Request Enable Query
*STB?
Read Status Byte Query
*PSC
Power On Status Clear Command
*PSC?
Power On Status Clear Query
* SAV
Save Command
*RCL
Recall Command
-- --
-
*OPT?
Option Ident~ficat~on
Query
5.1 Common IEEE488.2
. IDN?
Identification Query
<:'IKBITRAKL' .-'iSC'II RESl'ONStr. L):\?.A>
< N m e of product ni:tri~~f:~cturt'r>.
4 l o d C l name>. .
111 t h ~ \ d e \ ~ c e : ANKITSV. MP1777.4. 0. 01
f-urict~o~l
Kcports data iricl~~ding
the name of product rnariufxt~rrer 2nd ~nc~dcl
~i;inie
.OPC
Operation Complete Command
I~~u~icti(jn
Set\ the h ~ r 0 thlt to end o p e r ~ ~ t i o ~ i \ )
c)t the \r:~nd:ird e\ent .;t;trus seg~\tzl arid
itcht.\ SRQ
O N \\ hen the e\ecutlcili of tlir pre\ I O U \ r~l\t~xct~ori
I \ i~ornplcteii.
Ii\:\rllpI~' ot llS?
-> . l<(-.I. I . . OP('
OPC?
Operation Complete Query
--
Kr\pori\e
cYKI YL 2ILRIC KE5PO\\I I) ' i l \>
WAI
Wait Continue Command
Keep\ the ~.urilrlleiiccrricr1t 01- the ekecution of thc 11t'kt ~ i i \ t r t ~ i l ~ ~ ~ i
on hold ~ ~ r l t i l
i l l ? cxreutlc'ri
ot the pre\ l o u ~
rnrr-uctiun I \ conipleted.
I t:\ccu!r\
the o\crl:q7 co~iin1;11xl
;I\ thc e~~uer~t1;11
c o ~ i i ~ ~ i ; ~ r i ~ i .
j
Till\ f ~ ~ n c t ~ o ~ i
I eflc
Section 5 Common IEEE488.2 Commands
*CLS
Clear Status Command
Parameter
None
Function
Clears all the status composition excluding the output queue and MAV summary message
However, the enable register and Transition filter are not cleared.
Clears both the output queue and MAV bit when *CLS is sent immediately after the , and at the same time, before the element.
Example of use
> *CLS
-ESE
Standard Event Status Enable Command
Parameter
Integer value between 0 and 255.
Set as the parameter the sum total of the bits desired to be enabled among the standard event
status enable register.
Set 1 for enable and 0 for disable.
See the Section for the "Status Byte" for the composition of the register of this device.
Function
Sets and clears the standard event status enable register.
Example of use
Sets bits 2 and 4 of the standard event status enable register.
> *ESE 20
*ESR?
Standard Event Status Register Query
Response
Function
Example of use
Integer value between 0 and 255.
Set as the response the sum total of the bita of the standard event status register.
See the Section for the "Status Byte" for the composition of the register of this device.
Inquires about the current value of the standard event status registel
When a command error is found.
> *ESR?
< 32
5.1 Common IEEE488.2
sSRE
Service Request Enable Command
fJ,~rmetcr
ti11 CIhl \I \,LI"LII. IKIC I'KO(iK4M L14 1A>
enable regi\tcr.
Set 1 for enable anti 0 for di\ahls.
See the S r c t ~ o r ~
for the "Statu\ Bg te" for the cori:po\itiori ot the reyi\ter ot't111\ de\ ice
Service Request Enable Query
Inte~cr \:iiu~ b c t ~ r e i ~
0 ; I I I C ~ 255,
Set ;I\ tl~e ~t'~,pcvl\c
the k u n l total ot'thr b ~ t \ of tht.
ice er~;~ble
reglst'r.
See the St'ctioi: for the " S ~ a ~ u s
H ~ r c '
for the c o r n p o \ ~ t i o ~ ~
ot the regI\tt'r ot [hi, dr:\ ice
STB?
Read Status Byte Query
Inquires nhout the curl-ent v:~lue of the \t:ttus b> tc regi\ti.r ir~clud~ng
the 1lSS (M,~\ter S u n -
r r i q Statu\ I hit.
Section 5 Common IEEE488.2 Commands
* PSC
Power On Status Clear Command
Parameter
Function
O ................... Sets the power ON status clear flag to false.
1 ................... Sets the power ON status clear flag to true.
Determines whether or not to clear each enable register of the service request, standard event
status and parallel poll of the status report module when the power is switched ON.
When the mode is set to 0. the enable register will not be cleared and the device is enabled to
generate SRQ after the power source is turned ON.
When the mode is set to 1, the enable register will be cleared and the device is disabled to
generate SRQ after the power source is turned ON.
Example of use
Generates SRQ without clearing the power ON status flag.
> vPSC 0 ; *SRE 32 ; *ESE 128
*PSC?
Power On Status Clear Query
Response
0 ................... Sets the power ON status clear flag to false.
1 ................... Sets the power ON status clear flag to true.
Function
Inquires about the truth or falsity of the power ON status clear flag.
Example of use
2 *PSC?
< 0
* SAV
S a v e Command
Parameter
Function
1 to 10 ......... 4-4
Writes the current device setting into the memory of specified number.
There is no limit to the state to be saved. See the Table 7- 1 in the Section 7.5 "Parameter".
As an SCPI command that has the same function, : SYSTem : MEMory : STORe is avail-
able.
Example of use
Writes the current setting into the memory number 1.
z -SAV 1
5.1 Common IEEE488.2
*RCL
Recall Command
Parameter
0 to 10
Function
Calls the memory of a specified number and, by doing so, returns the device to the previous
state.
There is no limit to the state to be loaded. See the Table 7-1 in the Section 7.5 "Parameter".
As an SCPI command which has the same function, : SYSTem : MEMory : RECall is avail-
able.
Example of use
Calls the information stored in memory number 1 and performs setting.
> *RCL 1
*OPT?
Option Identification Query
Response
Function
Characters to correspond to the option or unit
O ~ t i o n
Interface (Bit Rate) selection condition option
Option 04
03,062 M, 6,125 M, 12,249 M
Character
Interface (Bit Rate) selection condition option
Standard
l 2,488 M, 4,977 M. 9,953 M
Interface (Bit Rate) selection condition option
Option 0 1
l 2,494 M, 3.988 M, 9,977 M
Interface (Bit Rate) selection condition option
Option 02
l 2,666 M, 5.332 M, 10,644 M
OPT0
OPT 1
OPT2
Reports on the list of options and units mounted.
Reports on options and units mounted all delimited by a comma.
Interface (Bit Rate) selection condition option
Option 05
l 3,069 M. 6,138 M, 12,276 M
Interface (Bit Rate) selection condition option
Option 06
02,677 M, 5,355 M, 10.709 M
Interface (Bit Rate) selection condition option
Option 07
l 2,578 M, 5,156 M, lO,3 13 M
Example of use
Options 01 and 02 are mounted.
> .OPT?
< OPT1 , OPT2
OPT5
OPT6
OPT7
Section 5 Common IEEE488.2 Commands
Element
( I ) CHARACTER
RESPONSE DATA
Example:
ABC
DEFG
(2) NR 1 NUMENIC
RESPONSE DATA
Example:
123
+I23
-1233
(3) NR2 NUMENIC
RESPONSE DATA
Example:
12.3
+12.34
-12.345
(4) NR3 NUMENIC
RESPONSE DATA
Example:
1.23E+35
-12.3E+45
Function
Expresses short mriemonic data.
Expresses integer values of decimal digits.
Expresses fixed point numerical values.
Expresses real numbers of declmdl dlg~ts w ~ t h
exponent.
( 5 ) STRING
RESPONSE DATA
Example:
"1233"
"ABCD"
" 1234.5"
(6) ARBITARY ASCII
RESPONSE DATA
Example:
NLAEND
5.1 Common IEEE488.2
--
Function
Expresses a character string in double quotations (").
/
quote char>
1
I
Sends ASCII data bytes excluding the NL character without delimiting them. For this rea-
son, NLAEND (or NL only) is placed next to the last data, and the data i, accordingly ter-
minated without an exit point.
An example of using the GPIB interface is shown below.
u
data byte>
Section 6 Status Report
The configuration of the status registers of the MP1777A conforms to the SCPI stipulations. (SCPI: Standard Command
for Programmable Instruments, see the Section 7 for more information.)
This section provides descriptions of the configuration of status registers and definition of the status register bit specific
to the device.
6.1 Configuration of MP1777A Status Registers ...................................
6-2
........................................
6.2 Status Registers Specified by IEEE488.2
6-4
6.3 Status Register Specified by SCPl ................................................. 6-6
6.4 Status Register Specific to MP1777A ..........................................
6-7
6.5 Reading, Writing and Clearing Status Registers .............................
6-8
Section 6 Status Report
6.1
Configuration of MP1777A Status Registers
SCPI stipulates that the status register configuration must conform to the configuration specified by IEEE488.2 and
incorporate the SCPI OPERation status register and QUEStionable status register specific to SCPI.
The figure below shows a simple block diagram of status registers mounted on this device. (Explanations of the bit
position and width to be provided later)
status register
OPERation
status register
7
-
-
-
-
4
Standard even1
:
status register
i : *
. . . . . . . . . . . . . . . . . .
Errorlevent aueue
Output queue Ll
Status byte
IEEE 488.2 stipulations
SCPI stipulations
1
. . . . . . . . . . . . . . . . .
Fig. 6-1 Block Diagram of Status Registers
Registers specified by IEEE488.2 are event register and status byte register.
Register specified by SCPI is OPERation status register.
Device-specific register is INSTrument status register.
6.1 Configuration of MP1777A Status Registers
Status registers excluding the registers specified by IEEE488.2 are configured as follows.
Condition
Transition
Event
Event enable
register
filter
register
register
.
.
.
.
- Summary Message
Fig. 6-2 Block Diagram of each Status Register
Table 6-1 Definitions of Register and Filter
Register and Filter
Definition
Condition register
Monitors the device status and performs real-time change in accordance with the device
Transition filter
status.
For this reason, this register does not store the status.
Sets the condition register details to the event register
The transition filter comes in three modes, depending on the changes of the condition regis-
ter to be evaluated.
Positive direction change: The event becomes true only when the corresponding condition
changes from false to true.
Negative direction change: The event becomes true only when the corresponding condition
changes from true to false
Both direction change:
The event becomes true when a change into the positive or
negative direction takes place.
Event register
Stores the output of the transition filter.
Event enable register
Selects the bit of corresponding Event Reg~ster to trigger a sh~ft of summary message to true.
Section 6 Status Report
6.2
Status Registers Specified by IEEE488.2
IEEE488.2 specifies the two status registers shown below.
Table 6-2 Definitions of Status Registers Specified by IEEE488.2
Status Register
Definition
Status byte register
A register to set RQS and seven summary message bits.
Being used in combination with the service request enable register, this register sets
SQR ON when the logical OR of the two is not zero. RQS is system reserved in bit 6
and this bit reports to the external controller the presence of service request.
Standard event status register
Sets eight types of events the device will encounter as standard events.
The log~cal OR output bit is summarized and displayed in bit 5 of the status byte
register as' the ESB (Event Status Bit) summary message.
Standard event status
Standard event
enable register
status register
- -@+-
OPC
1
---- 1
NOT USED
-
2
. 2 QYE
DDE
EXE
CME
NOT
PON
USED
Service request
Status byte
1
enable register
-
-
register
-
jO/
NOTUSED
1
Errorlevent queue I
Output queue F
1 -
2
2
QUE
-
3
-
3
QUES
-
4 -
5 .
---5
ESB
-
status register
OPERation
status register
. .
Logical OR
Service Request
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.. ....
Generation
6.2 Status Registers
The tables below show the definition of register bit specified by IEEE488.2.
Table 6-3 Definition of Status Byte Register Bit
Specified by
Bit
Status Byte Register
Definition
DB2
QUE
(ErrorEvent QUEue)
Indicates that the error and event queues are not empty.
DB4
MAV
(Massege Available)
Indicates that the output queue is not empty.
DB5
ESB
(Event Summary Bit)
Standard event status register summary
DB6
RQS
(Request Service)
RQS message
MSS
(Master Summary Status)
Indicates that the device has a cause to request at least one
service.
DB7
OPER (OPERation status register summary)
OPERation status register summary
Table 6-4 Definition of Standard Event Status Register Bit
Bit
Standard Event Status Register
Definition
DBO
OPC
(Operation Complate)
Indicates that all the specified operations are completed.
DB2
QYE
(Query Error)
Indicates that a query error has taken place.
DB3
DDE
(Device-dependent Error)
Indicates that an error other than a command, query or run
time error has taken place.
DB4
EXE
(Execution Error)
Ind~cates that a run time error has taken place.
DB5
CME
(Command Error)
Indicates that a command error has taken place.
DB6
URQ
(User Request)
Indicates that a local control error has taken place.
DB7
PON
(Power on)
Indicates that the power source has switched from OFF to
ON.
Section 6 Status Report
6.3
Status Register Specified by SCPI
The following section shows the definition of the register bit specified by SCPI.
OPERation
Status register
NOT USED
NOT USED
NOT USED
NOT USED
MEAS
NOT USED
To the status
NOT USED
byte register (Bit 7)
NOT USED
*
NOT USED
NOT USED
NOT USED
NOT USED
NOT USED
INST
NOT USED
NOT USED
Fig. 6-3 Definition of Register Bit Specified by SCPI
-
Table 6-5 Definition of OPERation Status Register Bit
-
Bit
OPERation Status Register
Definition
Indicates that the measurement is being performed.
-
DB 13
INST
(INSTrument status register summary) INSTrument status register summary
6.4 Status Register Specific to MP1777A
6.4
Status Register Specific to MP1777A
The following section shows the definition of the device-specific register bit.
INSTrument
Status register
NOT USED
NOT USED
EOT
UNL
ALC
NOT USED
NOT USED
To the OPERation
NOT USED
status register (Bit 13)
*
NOT USED
NOT USED
NOT USED
NOT USED
NOT USED
NOT USED
NOT USED
NOT USED
Fig. 6-4 Definition of INSTrument Register Bit
Table 6-6 Definition of INSTrument Status Register Bit
Bit
INSTrument Status Register
Definition
DB2
EOT
(End Of Test period)
Indicates that the test (measurement) has finished.
DB3
UNL
(UNLock)
Indicates that Unlock has taken place.
DB4
ALC
(Alarm Change)
Indicates that the alarm has changed
Section 6 Status Report
6.5 Reading, Writing and Clearing Status Registers
(1) Reading and writing methods
The table below shows the methods of reading and writing the details of status registers.
Table 6-7 Reading and Writing Methods of Status Registers
- - -
Reading Methods
Writing Methods
Disabled
Register
Serial poll
Seven-bit status byte and RQS message
bit are retumed. In this case, the status
byte value does not change.
Common *STB? query
Status byte register
The details of the status byte register
and one numerical value from MSS
(Master Summary Status) are returned.
Common *SRE? command
Disabled
Comn~on *ESE? command
Service request enable register
Common *SRE? query
Common *ESR? query
Standard event status register
In this case, the details of the register is
cleared after it is read.
Standard event status enable
register
Common *ESE? query
In this case, the detail of the register
will not change.
SCPI command (STATus subsystem)
: STATus : ... : EVENt?
SCPI event register
Disabled
In this case, the details of the register
will be cleared.
SCPI command (STATus subsystem)
: STATus : ... : ENABle
SCPI command (STATus subsystem)
: STATus : ... : ENABle?
SCPI enable register
In this case, the details of the register
will not change.
SCPI command (STATus subsystem)
: STATus : ... : PTRansition?
: STATus : ... : NTRansition?
-
SCPI Transition filter
SCPI command (STATus subsystem)
: STATus : ... : PTRansition
: STATus : ... : NTRansition
In this case, the details of the register
will not change.
ErrorEvent Queues
SCPI command
Disabled
: SYSTern : ERRor?
NOTES :
1. The SCPI event register, SCPI enable register and SCPI Transition filter correspond to the event register,
enable register and Transition filter in the status registers specified by SCPI and the device-specific status
registers.
2. See the Section 5 "Common IEEE4XS.Z Commands" for more information on the common command\ and
queries and the Section 8 "Device Message Details" for more infonnation on the SCPI commands.
Reading, Writing and Clearing Status Registers
(2) Clearing and resetting methods
The table below shows clearing and resetting methods of status registers and the range of their effects.
Table 6-8
Clearing and Resetting Methods of Status Registers
1
*RST
Register
Standard event status
No change
No change
Clear
No change
Execution of *ESE 0
enable register
Status byte register
1 No change
Service request
1 No change
Reading of events by
SCPI event register
No change
Clear
Clear
No change
: STATUS : ... : EVENt? to
clear the status register
*CLS
Clear
No change
Power
Source ON #
I
SCPI enable register ! No change
I
1 Reading of all the events
Enorievent queue
No change 1
Clear
1
Clear
No change
1 bv : SYSTeni : ERRor?
Clear
Clear
NOTES :
STATUS: PRESet
No change
Execution of the
,
1. The SCPI event register, SCPI enable register and SCPI transition filter correspond to the event register, enable
register and transition filter in the status registers specified by SCPI and device-specific status registers.
Other Clearing
Methods
No change
No change
No change
SCPI Transition filter
2. Cleared (or reset) when the power source is switched ON with the *PSC (Power-ON Status Clear) flag set as
true by the common PSC command.
Execution of *SRE 0
Reset
No change
The table below shows the reset values of registers influenced by the : STATus : PRESet command.
Reset
Reset
1 : ST.ATus : ... : PTRansition O
Reset
I ; : STATus : ... : NTRansition 0
Execution of
: STATus : ... : ENABle 0
Reaister 0
EnableIFilter
I
Reset Value
1
OPERational status register
QUEStionable status register
ALL 1
ALL 1
ALL 0
INSTrument status register
Enable register
PTRansition filter
NTRansition filter
Enable register
PTRansition filter
NTRansition filter
ALL 0
ALL 1
ALI* 0
Other status registers
Enable register
PTRansitinn filter
NTRansition filter
ALL l
ALL 1
ALL 1
Section 7 SCPI Outline
The MP1777A adopts SCPI (Standard Commands for Programmable Instruments) as commands to perform remote con-
trol.
This section provides the outline of SCPI and explanations of the command system.
This section and subsequent Sectlons describe the examples of command use and response as follows.
> Program message (program command. query command)
< Response
..............................................................................................
7.1 Outline
7-2
.........................................................................
7.2 Command Structure
7-3
7.3 Command Description Method ........................................................
7-4
...............................................................
7.4 Compounding Commands
7-5
7.5 Parameter .........................................................................................
7-6
Section 7 SCPI Outline
7.1
Outline
SCPI is a device command language defined by the SCPI consortium and is independent from the hardware.
SCPI is designed to shorten the period for development of automatic measuring instrument (ATE). For this reason, the
programming environments such as the device control and data handling are made consistent.
Moreover, it is characterized by its ability to perform totally even control of devices of the same model as well as devices
of different models compatible with each other that are equipped with the same functions.
7.2 Command Structure
7.2
Command Structure
The SCPI command form a hierarchical structure. Commands are divided into groups, each of which consists of com-
mands of similar functions, and each group forms a hierarchical structure called the subsystem.
This specification expresses these subsystems by the command tree as shown below.
SCPI allows the same headers to be present in a tree and the difference in the location of the header corresponds to the
difference in the function. Therefore, a command must be described in a full path up to the header to be used.
SOURce -
: TELecom
: BRATe
: BRATe?
: EQUAlizer
: EQUAlizer?
Fig. 7-1 Example of SCPI Command Tree
Section 7 SCPl Outline
Command Description Met hod
: SOURce : TELecom : BRATe ~brate>
: SOURce : TELecom : BRATe?
: SOURce : TELecom : EQUAlizer
: SOURce : TELecom : EQUAlizer?
Fig. 7-2 Example of SCPl Commands
The command tree shown in the previous section consists of the commands shown above.
The section below provides explanations of the stipulations concerning descriptions of commands.
A command begins with a colon ( : ).
Or, a command is structured by concatenating headers n ith a colon ( : ).
7.4 Compounding Commands
7.4 Compounding Commands
Commands can be compounded by using semi-colon ( ; ) as shown by the example below.
The second command is also referenced as the command located at the same level as the lowest hierarchical level of the
first command.
For this reason, the second command can be described in a full path as shown by Example 1. However, as Example 2
shows, headers located in the layer above Type can also be abbreviated.
Example I :
> : SENSe : TELecom
Example 2:
> : SENSe : TELecom
: PATTem : TYPE UWORd16 ; : SENS : TELecom : PATTern : UWORd " 1 1001 1001 1001 100"
: PATTem :TYPE UWORdl6 ; UWORd "1 10011001 1001100"
Section 7 SCPI Outline
Parameter
The table below shows types of parameters used by this measuring instrument
This operation manual expresses parameter types in lowercase characters enclosed by angled brackets ( < and > ) in the
table below.
Moreover, the type corresponding to the parameter type, specified by IEEE488.2 (or SCPI), is
expressed in uppercase characters.
The correspondence between each parameter type and specified by IEEE488.2 (or SCPI) is de-
scribed for each command.
Table 7-1 Details of Parameter Type
Parameter T V D ~
(Defined by SCPI)
, ..etc
Expresses a decimal digit. such as MINirnum and
MAXimum are included as special numerical value type data.
The numerical values used by this device are mainly integer type.
Therefore, the fractional part is rounded.
Expresses a theoretical value.
OFF or 0 corresponds to false and ON or 1 corresponds to true. The setting can be done
by O and 1, or OFF and ON. However, the setting for the response to a query must be
done by O or I.
Expresses a character string, made of ASCII characters enclosed by single quotation
marks ( ' ) or double quotation marks ( " ) .
Either long or short form can be used.
Example: 'LOF : M 139' or "LOF : M 139"
Expresses a character data. Expressed in a short character string that corresponds to the
setting details
Either the long or short form can be used.
Section 8 Device Message Details
This section provides detailed explanations of the device messages supported by the MP1777A
Examples of command use and responses are described as follows in this Section.
> Program message (program command and query command)
< Response
8.1 Response Format .................................. ................ ........................... 8-2
8.2 Buffer Size Stipulations .................................................................
8-4
8.3 Device-Specific Commands .................................................... ........
8-5
INSTrument Subsystem .....................................................
8-7
SOURce Subsystem . .. . . ... .. .. .... .. . . ... ... ... . . . . .... ... . . . . . . . . .
8-8
SENSe Subsystem .............................................................. 8-1 4
DlSPlay Subsystem ............................................................
8-21
CALCulate Subsystem .........................................................
8-24
SYSTem Subsystem ......................................................... 8-25
STATUS Subsystem ........ .................................... ........... 8-29
Section 8 Device Message Details
8.1
Response Format
This section explains the format of the rehponse to the query command
The format is shown in the table below.
Table 8-1 Response Format (By Response Type)
Response Type
Format
,
I A string enclosed by the quotation marks (") is returned
Makes the number of digits of the response variable and sets the maximum number of
digits within the range of numerical values as the maximum number of digits. No space
is inserted between the sign and numerical value.
> : SYSTem : DATE?
i
1993.7, 14
, ... etc
A short form of a character is returned.
> : INSTrument : COUPle?
< ALL
/ See Table 8-2 for the response format, Form 6.
, ... etc
When short and long forms are both found in the string of the corresponding command,
the short form is returned.
> : DISPiay : DSELect : NAME?
< "SET" (short form of "SETUP")
8.1 Response Format
Table 8-2 Details of Response Format
Format
"XXXX"
"XXXX"
"XXX.XN
"XX.XXW
"XX.XX"
"XX.XX"
"X.XXXM
When 0.000 I Value _< 3,232 In the 3,200 U1 Range UIPP
When 0.000 I Value I 1,6 16 in the 1,600 UI Range UIPP
When 0.000 I Value 1808.0 in the 800 UI Range UIPP
When 0.000 I Value < 80.80 In the 80 UI Range UIPP
When 0.000 I Value 140.10 In the 40 UI Range UIPP
When 0.000 I Value I 20.20 In the 20 UI Range UIPP
When 0.000 I Value 10.505 In the 0.5 UI Range UIPP
F ~ v e characters al~gned to the right among slx characters
> : SOURce : JITTer : AMPLltude'?
< "2000"
When Value > 3,232 (1 Ul Range, UIPP)
When Value > 1,6 16 (1 UI Range, UIPP)
When Value > 808.0 (1 UI Range, UTPP)
When Value > 80.80 ( 1 UI Range, UIPP)
When Value > 40.40 (1 UI Range, UIPP)
When Value > 20.20 (1 UI Range. UIPP)
When Value > 0.505 ( I UI Range. UIPP)
When 0.000 I Value I 1.010 in the 1 UI Range UIPP
When 0.000 I Value 10.505 in the 1 UI Range UI+P, UI-P
When 0.000 1 Value I 0.375 in the 1 UI Range UIrms
Five characters aligned to the right among six characters
> : CALCulate : DATA? "JAMPLitude : PTPeak"
< " 1.234"
When 0.00 I Value 5 4.04 in the 4 UI Range UIPP
When 0.00 I Value 5 2.02 In the 4 UI Range UI+P, UI P
When 0.00 I Value 5 1.43 In the 4 UI Range UInnc
Four characters al~gned to the r~ght among SIX characters
> : CALCulate : DATA? "JAMPLltude : RMS"
< "7.00"
When Value > 1.010 (1 UI Range, U~PP)
When Value > 4.04 (4 UI Range, UIPP)
When Value > 0.505 (1 UI Range, UI-P/UI-P)
When Value > 2.02 (4 UI Range, UI-P/UI P)
When Value > 0.357 (1 UI Range, UInni)
When Value > 1.43 (4 UI Range, UInn\)
In the absence of data to corre\pond to the query
Section 8 Device Message Details
8.2
Buffer Size Stipulations
A response in the format explained earlier is sent from the device to the controller. Therefore, the controller must have a
buffer of a size enough to accept the response.
When the commands of this device are used, a maximum of 255 bytes of buffer space will suffice.
8.3 Device-Specific Commands
8.3
Device-Specif ic Commands
The following section provides details of the device-specific commands.
As the command descriptions are classified by the subsystem, see the Appendix D correspondence between commands
and screens for more information on the correspondence with screens.
Incidentally, the device-specific commands supported by this device are sequential commands with some exceptions
An explanation is glven for commands other than sequential commands every time they come up.
When a set value by a program command causes set values for other items to become unacceptable, they will be changed
to allowed values.
Moreover, the measurement starts again when settings are modified during measurement.
See the details of the operation method in this operation manual for conditions to change set values for other items and to
execute restart.
The following section shows examples of descriptions of commands.
When a restriction on command use exists, an item called the restriction is added to the example below.
Program command
Parameter type (SCPI, device-specific)
I
1
Parameter type (IEEE488.2, SCPI)
I
1
/ I i
: SENS : TEL : RANG
-.
Parameter
Function
Example of use
=
+ ;;:
Jitter receiving range 1 UI
Jitter receiving range 4 UI
Selects the range of jitter reception.
Sets the range of jitter reception at UI 4.
> : SENS : TEL : RANG U14
4
I
i
\% \
Parameter details
Example of command use
The right angled blacket ( > )
at the header shows that the
segment is a Program mes-
sage and left angled blacket
( < ) shows it is a Response.
Section 8 Device Message Details
8.3 Device-Specific Commands
8.3.1 INSTrument Subsystem
The INSTrument subsystem is used to specify the same setting for the transmission and reception or independent setting
for each.
: INSTrurnent
: COUPle
7
: COUPle?
: INSTrument : COUPle
Command type
Sequential command
Parameter
Function
Restriction
=
ALL ............. Tx&Rx (Same)
NONE .......... Tx/Rx (Independent)
Specifies the selection of whether to establish the same setting for the transmission and
reception modules or independent setting for each.
When the mode is switched from Tx/Rx to Tx&Rx, the set value for Rx is set for the Bit
Rate.
Example of use
Sets the same setting for the transmission and reception.
> : INSTrument : COUPle ALL
: INSTrument : COUPle?
Command type
Sequential command
Response
=
Same as : INSTrument : COUPle
Function
Makes a query of whether the setting is the same for the transmission and reception
Example of use
> : INSTrument : COUPle?
< ALL
Section 8 Device Message Details
8.3.2 SOURce Subsystem
The SOURce subsystem controls (setting and display) the transmission side.
: BRATe?
: EQUAlizer
: EQUAlizer?
: SOURce
: SOURce?
: JITTer
: JITTer?
: OFFSet
-
: OFFSet?
:
: RANGe
: RANGe?
: AMPLitude?
8.3 Device-Specific Commands
: SOURce : TELecom : BRATe
-
-
Command type
Sequential command
Parameter
=
(Standard)
(Option 05)
.........
M2,488 ........ 2,488 Mbit/s
M3069
3,069 Mbit/s
.........
M4,977 ........ 4,977 Mbit/s
M6138
6,138 Mbitls
M9,953 ........ 9, 953 Mbit/s
M12276 ....... 12,276 Mbitis
(Option 01)
(Option 06)
.........
M2494 ......... 2,494 Mbitls
M2677
2,677 Mbitls
M4988 ......... 4,988 Mbit/s
M5355 ......... 5,355 Mbitls
.......
M9977 ......... 9,977 Mbit/s
M 10709
10,709 Mbitls
(Option 02)
(Option 07)
M2666 ......... '2,666 Mbit/s
M2578 ......... 2,578 Mbit/s
M5332 ......... 5,332 Mbitis
M5156 ......... 5,156 Mbit/s
MI0664 ....... 10,664 Mbit/s
MI0313 ....... 10,313 Mbit/s
(Option 03)
M3062 ......... 3,062 Mbit/s
M6125 ......... 6,125 Mbit/s
M12249 ....... 12,249 Mbit/s
Function
Sets the sign speed of transmission signal.
Restriction
Sending the command while the option above is not set results in a run time error.
When the same setting is established for the transmission and reception modules (Tx&Rx),
the reception Bit Rate will be set.
Example of use
Sets the sign speed of the transmission signal to 9,953 Mbitls.
> : SOURce : TELecom : BRATe M9953
: SOURce : TELecom : BRATe?
Command type
Sequential command
Response
=
Function
Makes a query of the sign speed of the transmission signal.
Example of use
> : SOURce : TELecom : BRATe?
< M9953
Section 8 Device Message Details
: SOURce : TELecom : EQUAlizer
Command type
Sequential command
Parameter
=
OFF or 0 ...... Equalizer OFF
ON or 1 ........ Equalizer ON
Function
Sets ON/OFF of the Jitter Setting Equalizer.
Example of use
Sets the Jitter Setting Equalizer ON.
> ; SOURce : TELecom : EQUAlizer ON
- --
: SOURce : TELecom : EQUAlizer?
Command type
Sequential command
Response
...................
0
Equalizer OFF
1 ................... Equalizer ON
Function
Makes a query of the setting status of the Jitter Setting Equalizer.
Restriction
None
Example of use
> : SOURce : TELecom : EQUAlizer?
< 1
: SOURce : TELecom : CLOCk : SOURce
Command type
Sequential command
Parameter
=
INTernal ................. Internal
........
LOCK-2MHB
2 MHz
(Balanced)
.......
LOCK-2MHU
2 MHz
(Unbalanced)
........
LOCK-2MBB
2 Mbit/s
(Balanced)
LOCK-2MBU ........ 2 Mbit/s
(Unbalanced)
LOCK-I SMH ........ 1.5 MHz
(Unbalanced)
LOCK-1 SMB ........ 1.5 Mhit/s (Balanced)
EXTernal ................ External
...........
LOCK-1 OM
Lock 1 OM (Balanced)
Function
Sets the clock source of the transmission signal.
Example of use
Sets the clock source of the transmission signal to internal.
> : SOURce : TELeconi : CLOCk : SOURce INTemal
8.3 Device-Specific Commands
: SOURce : TELecom : CLOCk : SOURce?
Command type
Sequential command
Response
=
INT .............. Internal
EXT ............. External
LOCK .......... Lock
REC ............. Receive
Function
Makes a query of the clock source of the transmission signal.
Example of use
> : SOURce : TELecom : CLOCk : SOURce?
< INT
: SOURce : TELecom : JlTTer
Command type
Sequential command
Parameter
=
OFF or 0 ...... No jitter to be generated
........
ON or 1
Jitter to be generated
Function
Establishes setting of whether or not to generate jitter.
Restriction
When no jitter generation is set, attempting to set the frequency offset and jitter generation
range results in an error, disabling the setting.
Example of use
Sets the jitter generation ON.
> : SOURce : TELecom : JITTer ON
: SOURce : TELecom : JITTer?
Cornmand type
Sequential command
Response
...................
O
No jitter to be generated
...................
1
Jitter to be generated
Function
Makes a query of whether or not jitter will be generated.
Example of use
> : SOURce : TELecom : JITTer?
< 1
Section 8 Device Message Details
: SOURce : TELecom : OFFSet
Command type
Sequential command
Parameter
=
-50 to +50 ... step 0. I
Function
Performs the frequency offset setting.
Example of use
Sets the value of frequency offset at +25.
> : SOURce : TELecom : OFFSet 25
: SOURce : TELecom : OFFSet?
Command type
Sequential command
Response
-50 to +50 ... step 0.1
Function
Makes a query of the value of frequency offset.
Example of use
> : SOURce : TELecorn : OFFSet?
< -25
: SOURce : JITTer : RANGe
Command type
Sequential command
Parameter
Function
Restriction
= CHARACTER PROGRAM DATA>
UI3200 ........ 3,200 UIPP
UI1600 ........ 1,600 UIPP
UI800 .......... 800 UIPP
U180 ............ 80 UIPP
U140 ............ 40 UIPP
UI20 ............ 20 UIPP
UI05 ............ 0.5 UIPP
Performs setting of the transmission jitter generation range.
Effective when transmission jitter is generated.
Restrictions on setting apply depending on the Bit Rate.
Example of use
Sets the transmission jitter generation range at 0.5 UIPP.
> : SOURce : JITTer : RAN& UI05
Bit Rate
10 G
UI Range
3.200. 80, 0.5 UI
8.3 Device-Specific Commands
: SUORce : JlTTer : RANGe?
Command type
Response
Function
Restriction
Transmission sequential command
=
Makes a query of the transmission jitter generation range
Effective when jitter is generated.
Restrictions on setting apply depending on the Bit Rate.
Example of use
I
Bit Rate
/
UI Ranae
I
I
SG
1
1.600. SO, 0.5 UI
I
> : SOURce : JITTer : RANGe?
< UI05
: SUORce : JlTTer : AMPLitude?
Command type
Sequential command
Response
=
Function
Performs reading of the value of Tx JITTer AMPLitude.
Example of use
< : SOURce : JITTer : AMPLitude?
> 1.51
Section 8 Device Message Details
8.3.3 SENSe Subsystem
The SENSe subsystem performs control (setting and display) of measurement conditions and reception side.
: SENSe
- : TELecom
: BRATe
: BRATe?
: RANGe
: RANGe?
: FILTer
t
: FILTer?
: MEASure -
- : MEASure
: STATe?
: STAR1
: STOP
: STIMe?
I
:TYPE
:TYPE?
: PERiod
,
: PERiod?
8.3 Device-Specific Commands
p~
: SENSe : TELecom : BRATe
Command type
Sequential command
Parameter
=
(Standard)
(Option 05)
M2488 ......... 2,488 Mbitls
M3069 ......... 3,069 Mbit/s
M4977 ......... 4,977 Mbit/s
M6138 ......... 6.138 Mbit/s
M9953 ......... 9,953 Mbit/s
M 12276 ....... 12,276 Mbit/s
(Option 01)
(Option 06)
M2494 ......... 2,494 Mbit/s
M2677 ......... 2,677 Mbit/s
M4988 ......... 4.988 Mbit/s
M5355 ......... 5,355 Mbit/s
M9977 ......... 9,977 Mbit/s
M 10709 ....... 10,709 Mbitls
(Option 02)
(Option 07)
M2666 ......... 2,666 Mbit/s
M2578 ......... 2,578 Mbit/s
M5332 ......... 5,332 Mbit/s
M5156 ......... 5,156 Mbit/s
MI0664 ....... 10,664 Mbit/s
MI0313 ....... 10,313 Mbit/s
(Option 04)
M3062 ......... 3.062 Mbit/s
M6125 ......... 6,125 Mbit/s
Mi2249 ....... 12,249 Mbit/s
Function
Sets the sign speed of receiving signal.
Restriction
Sending the command while the option above is not set results in a run time error.
Example of use
Sets the sign speed of the reception signal to 9,953 Mbit/s.
> : SENSe : TELecom : BRATe M9953
-
SENSe : TELecom : BRATe?
-
Command type
Sequential command
Response
=
Function
Makes a query of the sign speed of the reception signal.
Example of use
> : SENSe : TELecom : BRATe?
< M9953
Section 8 Device Message Details
: SENSe : TELecom : RANGe
Command type
Sequential command
Parameter
=
U14 .............. 4 UIPP
UI 1 .............. 1 UIPP
Function
Sets the measurement range of receiving jitter.
Example of use
Sets the measurement range of receiving jitter at 1 UIPP,
> : SENSe : TELecom : RANGe UIl
: SENSe : TELecom : RANGe?
Command type
Sequential command
Response
=
Function
Makes a query of receiving jitter measurement.
Example of use
> : SENSe : TELecom : RANGe?
< UII
: SENSe : TELecom : FlLTer
Command type
Sequential command
Parameter
=
LP ................ LP
(80 M)
HPlLP ......... HPl+LP (10 K to 80 M)
HPl SLP ....... HPI1+LP (20 K to 80 M)
HP2LP ......... HP2+LP (4 M to 80 M)
HP3LP ......... HP3+LP (6 M to 80 M)
HPLP ........... HP+LP
( 12 K to 80 M)
HPSLP ......... HP'+LP (50 K to 80 M)
Function
Performs filter setting.
Restriction
HP1+LP, HPl'+LP setting is disabled when the Bit Rate is set at 2.5 G or 5 G.
HP3+LP can be set only for Option 07.
Example of use
Sets the filter at HP'+LP, HPll+LP.
> : SENSe : TELecom : FILTer HPSLP
8.3 Device-Specific Commands
: SENSe : TELecom : FILTer?
Command type
Sequential command
Response
=
Function
Makes a query of the filter setting.
Restriction
HP'+LP, HPlq+LP setting is disabled when the Bit Rate is set at 2.5 G or 5 G
Example of use
> : SENSe : TELecom : FILTer?
< HPlSLP
: SENSe : TELecom : MEASure : TYPE
Command type
Sequential command
Parameter
=
MANual ...... Manual measurement
SINGle ........ Single measurement
REPeat ......... Repeat measurement
Function
Performs the measurement mode setting.
Example of use
Sets the measurement mode to the repeat measurement.
> : SENSe : MEASure : TYPE REPeat
: SENSe : TELecom : MEASure : TYPE?
Command type
Sequential command
Response
=
MAN ........... Manual measurement
SING ........... Single measurement
REP ............. Repeat measurement
Function
Makes a query of the measurement mode.
Example of use
> : SENSe : MEASure : TYPE'?
< REP
Section 8 Device Message Details
: SENSe : TELecom : MEASure : PERiod ,
Command type
Sequential command
Parameter
=
1 to 99
=
EI .................. hour
M ................. minute
S ................... second
Function
Performs the measurement time setting.
Restriction
The setting is disabled when the measurement mode is set to manual measurement.
Example of use
Sets the measurement time at one hour.
> : SENSe : TELecom : MEASure : PERiod 1 , H
: SENSe : TELecom : MEASure : PERiod?
Command type
Sequential command
Response
,
=
=
Function
Makes a query of the measurement time.
Example of use
> : SENSe : TELecom : MEASure : PERiod?
< 1, H
8.3 Device-Specific Commands
: SENSe : MEASure : STATe?
Command type
Sequential command
Response
,
=
MAN ........... Manual measurement
NON ............ No target measurement found
Function
Restriction
=
0 ................... Measurement completed
1 ................... Measurement current1 y being performed
Makes a query of the measurement status.
Measurement of only one item enabled to he executed.
When no measurement is performed, the follow~ng indication is output.
< NON, 0
Example of use
> : SENSe : MEASure : STATe?
: SENSe : MEASure : STARt
: SENSe : MEASure : STOP
Command type
Sequential command
Parameter
None
Function
Halts measurement currently being executed.
Restriction
None
Example of use
> : SENSe : MEASure : STOP
Section 8 Device Message Details
: SENSe : MEASure : STIMe?
Command type
Sequential command
Response
, , , , ,
Function
Makes a query of the time to start the measurement.
Restriction
The clock can be set within the range between the year 1993 and 2093. When a query is
made while the measurement has yet to be started, the time to start measurement previously
set will be applied.
Example of use
> : SENSe : MEASure : STIMe?
< 1996,2, 5, 12, 12. 13
8.3 Device-Specific Commands
8.3.4 DlSPlay Subsystem
The DISPlay subsystem performs control (setting and display) of screens.
: DISPlay
: DSELect
[:NAME]
7
[:NAME]
: SETup
[:NAME]
[:NAME]
: RESult
Section 8 Device Message Details
: DlSPlay : SETup [:NAME]
Command type
Sequential command
Parameter
=
"INTerface" ....... Interface screen
........
"MEMory"
Memory screen
"SYSTem" ......... System screen
Function
Selects display item on the Setup screen
Example of use
Selects "INTerface" from the display items on the Setup screen.
> : DISPlay : SETup : NAME "INTerface" or : DISPlay : SETup "INTerface"
: DlSPlay : SETup [:NAME]?
Command type
Sequential command
Response
=
"INT" ........... Interface screen
"MEM" ........ Memory screen
"SYST" ........ System screen
Function
Makes a query of the display item on the Setup screen.
Example of use
> : DISPlay : SETup : NAME? or : DISPlay : SETupc?
< "INT"
: DlSPlay : RESult : JITTer : MODE
Command type
Sequential command
Parameter
=
CURRent ..... Current display
LAST ........... LAST display
Function
Switches the display mode of the Result screen.
Example of use
Sets the display mode to the Current display.
> : DISPlay : RESult : JITTer : Mode CURRent
8.3 Device-Specific Commands
-
--
: DlSPlay : RESult : JlTTer : MODE?
Command type
Sequential command
Response
= CHARACTER RESPONSE DATA>
Function
Makes a query of the display mode of the Result screen.
Example of use
> : DISPlay : RESult : JITTer : MODE'?
< CURR
: DlSPlay : RESult : JlTTer : UNIT
Command type
Sequential command
Parameter
=
PEAK .......... PEAK detection
RMS ............ RhlS detection
Function
Sets the format of data to be displayed on the Result screen.
Example of use
Sets the format of data to be displayed on the Result screen at Peak
> : DlSPlay : RESult : JITTer : UNIT PEAK
: DlSPlay : RESult : JlTTer : UNIT?
Command type
Sequential command
Response
=
Function
Makes a query of the format of data to be displayed on the Result screen.
Example of use
> : DISPlay : RESult : JITTer : UNIT?
< PEAK
Section 8 Device Message Details
8.3.5 CALCulate Subsystem
The CALCulate subsystem performs the performance measurement setting and measured result display
: CALCulate -
: DATA?
: CALCulate : DATA?
Command type
Sequential command
Parameter
Response
=
=
See the Table 8-1 for .
Function
Reads the measured results.
Example of use
> : CALCulate : DATA? "CURRent : JAMPlitude : PTPeak"
< 1.00
Table 8-1 String
Item
Jitter manual measurement
CURRent
Peak to Peak
+Peak
-Peak
RMS
LAST
Peak to Peak
+Peak
-Peak
RMS
"[CURRent:] JAMPIitude : PTPeak"
"[CURRent:l JAMPlitude : PPEak"
"[CURRent:] JAMPlitude : MPEak"
"[CURRent:] JAMPlitude : RhlS"
"LAST : JAMPlitude : PTPeak"
"LAST : JAMPlitude : PPEak"
"LAST : JAMPlitude : MPEak"
"LAST : JAMPlitude : RMS"
Response Format
Form 6
Form 6
Form 6
Form 6
Fom16
Form 6
Form 6
Form 6