This document is an integral part of the following Operating Instructions:

Safety Instructions XA00126D/06/EN/13.12 71195714 Ex documentation Proline Promag 53 HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, Modbus RS485 NEPSI Zone 1, Zone 21 • BA00047D, Proline Promag 53 HART • BA00051D, Proline Promag 53 FOUNDATION Fieldbus • BA00053D, Proline Promag 53 PROFIBBUS DP/PA • BA00117D, Proline Promag 53 Modbus RS485 Contents General warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Special conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Installation instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 COC certificates of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Description of measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Nameplates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Type code. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Temperature table compact version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Temperature table remote version. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Gas and dust explosion protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Design of measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Cable entries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Connection of remote version connecting cable. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Electrical connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Terminal assignment and connection data, power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Terminal assignment and connection data for signal circuits (intrinsically safe circuits) . . . . . . . . . . . . . . . . . . . 12 Terminal assignment and connection data for signal circuits (non-intrinsically safe circuits) . . . . . . . . . . . . . . . 14 Service adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Device fuse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

0NEPSI Zone 1, Zone 21

Proline Promag 53 2 Endress+Hauser General warnings • For installation, use and maintenance of the flow meter, the instruction manual and the following standards shall be observed: – GB50257-1996 "Code for construction and acceptance of electric device for explosive atmospheres and fire hazard electrical equipment installation engineering" – GB3836.13-1997 "Electrical apparatus for explosive gas atmospheres – Part 13: Repair and overhaul for apparatus used in explosive gas atmospheres" – GB3836.15-2000 "Electrical apparatus for explosive gas atmospheres – Part 15: Electrical installations in hazardous area (other than mines)" – GB3836.16-2006 "Electrical apparatus for explosive gas atmospheres – Part 16: Inspection and maintenance of electrical installation (other than mines)" – GB15577-2007: Safety regulations for dust explosion prevention and protection". (Only if installed in dust hazardous areas.) – GB12476.2-2006 "Electrical apparatus for use in the presence combustible dust – Part 1-2: Electrical apparatus protected by enclosures and surface temperature limitation – selection, installation and maintenance" (only if installed in dust hazardous areas). • The flow meter shall not be modified in order to ensure the explosion protection performance of the equipment. Any change may impair safety. • Installation, connection to the electricity supply, commissioning and maintenance of the devices must be carried out by qualified specialists trained to work on Ex-rated devices. • Compliance with all of the technical data of the device (see nameplate) is mandatory. • Open the device only when it is de-energized (and after a delay of at least 10 minutes following shutdown of the power supply) or in non-hazardous (classified) locations. • It is not permissible to connect the service adapter whilst the atmosphere is considered to be explosive. • Opening the transmitter housing and the connection housing of the remote version is only permitted for a brief time. During this time, ensure that no dust or water enters the housing. • To guarantee resistance to dust and water, the transmitter housing, the connection housing of the remote version and the cable entries must be tightly sealed. • The suitability of the device in the event of simultaneous occurrence of gas-air and dust-air mixtures requires an additional assessment. Special conditions • The device must be integrated into the potential equalization system. Potential must be equalized along the intrinsically safe sensor circuits. Further information is provided in the "Potential equalization" section on Page 9. • The sensors may only be used for those media, for which the wetted parts are known to be suitable. • For the application of the transmitters in an ambient temperature of less than –20 °C suitable cables and cable entries certified for this condition shall be used. Entry holes which are not needed, shall be closed by stopping plugs separately certified for this purpose. Installation instructions • For terminals No. 20 to No. 27 of the transmitter, only devices with ratings Um  260 V and Im  500 mA are allowed to be connected (does not apply to intrinsically safe circuits). • The measuring device must only be used in the permitted temperature class. The values of the individual temperature classes can be found in the temperature tables on Page 6. For Zone 21: The surface temperature of the measuring device must not exceed 2/3 of the ignition temperature of a dust cloud. The maximum surface temperature must maintain a safe distance of 75 °C to the smolder temperature of a dust layer of 5 mm. Example: Operation in temperature class T4 (135 °C) is, therefore, suitable for dust with an ignition temperature of 202.5 °C (1.5 × 135 °C or 135 °C = 2/3 of 202.5 °C) and a smolder temperature of 210 °C (135 °C + 75 °C). • The following applies when using the terminal compartment in type of protection "flameproof/Ex d":  Only cable entries and cable glands, which are approved by NEPSI in accordance with GB3836.1-2010 and GB3836.2-2010 and which are suitable for an operating temperature of up to 80 °C, they shall be used. • The following applies when using the terminal compartment in type of protection "increased safety/Ex e":  Only cable entries, cable glands and blanking plugs, which are approved by NEPSI in accordance with GB3836.1-2010 and GB3836.3-2010 and which are suitable for an operating temperature of up to 80 °C and for an ingress protection of IP67, they shall be used. Alternatively Ex e cable glands specified or provided by Endress+Hauser Flowtec AG can be used. The cables must be installed in such a way, that they are fixed in place in order to ensure adequate strain relief. • Suitable cables and suitable, certified cable glands, cable entries and blanking plugs must be used for measuring devices operated at temperatures below –20 °C. • The cable entries and openings not used must be sealed tight with suitable components.

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 3 Installation instructions (continued) • Turning the local display: the screw cap has to be removed before the local display can be turned, and this must be done with the device de-energized (and after a delay of at least 10 minutes following shutdown of the power supply) or when the area is known to be non-hazardous. " Caution! The explosion group for the measuring device can be IIC. However, it is reduced to IIB if the permitted, external capacitance/inductance for the intrinsically safe communication circuits is increased (see Page 13). # Warning! For entity concept, the criteria for interconnection between the I/O circuits and the associated circuits is as follow: • Uo  Ui • Io  Ii • Po  Pi • Co  Ci + Cc • Lo  Li + Lc Cc and Lc stand for the distributed capacitance and distributed inductance of the cable. Turning the transmitter housing 1. Unscrew the grub screw. 2. Rotate the transmitter housing cautiously clockwise until the end stop (end of the thread). 3. Rotate the transmitter housing counter-clockwise (max. 360°) in the wanted position. 4. Tighten the grub screw properly after each manipulation # Warning! Ensure that the grub screw is tightened properly. A0006944 Fig. 1: Turning the transmitter housing COC certificates of conformity COC certificates of conformity By affixing the certification number the product conforms with the following standards: Certification numbers: Inspection body NEPSI, National Supervision and Inspection Centre for Explosion Protection and Safety of Instrumentation Description of measuring system The measuring system consists of transmitters and sensors. Two versions are available: • Compact version: transmitters and sensors form a mechanical unit. • Remote version: transmitters and sensors are installed separately and connected to each other via connecting cables. N ic ht u nt e r S p a n n un g öf f ne n K e e p c o v er ti g ht w h il e ci r c u it s a re al i v e N e p a s o u v ri r l’ ap p ar ei l s o u s t e n si o n K e e p co v e r t ig ht w hi l e ci r c u it s a re al i v e max. 360° Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschützt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans sécurité intrinsèque protégées par Ip40 1 2 3 4 • GB3836.1-4 – 2010 • GB3836.20 – 2010 • GB12476.1 – 2000 • GYJ12.1047X

0NEPSI Zone 1, Zone 21

Proline Promag 53 4 Endress+Hauser Nameplates The nameplates, which are mounted in a clearly visible position on the transmitter and sensor, contain all of the relevant information about the measuring system. A0008261 Fig. 2: Example for nameplates of a transmitter and of a sensor A Transmitter nameplate B Sensor nameplate 1 Manufacturer / Certificate recipient 2 Transmitter or sensor type 3 Order code and serial number 4 Power supply, frequency and power consumption 5 Additional specification, e.g. ECC = Electrode cleaning circuitry, EPD = empty pipe detection,  R/B = reference electrode, etc. (only if present) 6 Available inputs/outputs 7 Space for additional information on special products 8 Year of manufacture 9 Type of enclosure protection 10 Type of protection 11 Number of the NEPSI certificate of conformity 12 Space for notes, e.g. delays, etc. (only if necessary) 13 Ambient temperature range 14 C-Tick symbol 15 NEPSI Symbol 16 Associated Ex documentation 17 Space for other approval specifications and certificates, e.g. PROFIBUS, etc. (only if present) 18 Calibration factor/zero point 19 Nominal diameter/nominal pressure 20 Fluid temperature range 21 Materials in contact with the medium 22 Measuring electrode material 23 Calibration tolerance 24 Direction of flow A B Order Code: Ser.No.: TAG No.: XXXXX-XXXXXXXXXXXX 12345678901 ABCDEFGHJKLMNPQRST 20-55VAC/16-62VDC 50-60Hz 15VA/W IP67 XA...D/06/../.... 4153 Reinach Switzerland 2007 i Promag 5X I-OUT (HART), f-OUT RELAY, STATUS-IN N12895 1 3 13 12 9 10 11 16 14 17 15 2 4 6 7 8 18 20 21 2 3 1 8 14 16 15 -20°C(-4°F)

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 5 Type code The type code describes the exact design and the equipment of the measuring system. It can be read on the nameplate of the transmitter and sensor and is structured as follows: Electronics (Item No. 2 in type code) Sensor (Item No. 3 in type code) Approvals (Item No. 12 in type code) P R O M A G 5 * * * * – * * * * * * * * * * * * Pos. no.: 1 Instrument family 2 Electronics 3 Sensor 4 to 5 Nominal diameter 6 Hyphen 7 Lining/process connection 8 Process connection/seal 9 Electrodes / material 10 Calibration 11 Factory certificates 12 Approvals 13 Version (compact/remote) 14 Remote version cable 15 Type of cable entry 16 Display, operation, power 17 Software 18 Outputs/inputs * Transmitter Type of explosion protection 3 Promag 53 • Transmitter electronics: [Ex ia] IIC • Transmitter housing: Ex d or Ex de IIC * Sensor Lining W,  P,  H Promag W • Hard rubber • Polyurethane Promag P • PFA • PTFE Promag H • PFA * Type of explosion protection Transmitter Sensor Remote version Compact version with Ex ia input/output Compact version other K 1) Ex de[ia Ga] IIC/IIB T6 Gb DIP A21 Ta, T6 IP6X Ex deia[ia Ga] IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X Ex deia IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X Ex eia IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X S 2) Ex de[ia Ga] IIC/IIB T6 Gb DIP A21 Ta, T6 IP6X Ex deia[ia Ga] IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X Ex deia IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X Ex eia IIC/IIB T1~T6 Gb DIP A21 Ta, T6 IP6X 1) Terminal compartment in type of protection Ex d 2) Terminal compartment in type of protection Ex e

0NEPSI Zone 1, Zone 21

Proline Promag 53 6 Endress+Hauser Type (compact/remote; Item No. 13 in type code) Type of cable entry (Item No. 15 in type code) Display, operation, power supply (Item No. 16 in type code) Outputs/inputs (Item No. 18 in type code) !Note! A detailed explanation of these values, regarding the available outputs and inputs, as well as a description of the associated terminal assignments and connection data can be found from Page 11 onwards. Temperature table compact version Max. medium temperature [°C] for T1 ~ T6 in relation to the maximum ambient temperature Ta 1) with polyurethane lining; 2) with hard rubber lining 1) with PFA lining; 2) with PTFE lining The minimum ambient temperature is –20 °C. A version for ambient temperatures up to –40 °C is optionally available * Type Min. ambient temperature Ta min A, P, U Compact –20 °C V, 6 –40 °C G, N, T, W Remote –20 °C 7, 8 –40 °C * Thread form A M20 × 1.5 B NPT ½" C G ½" * Thread form A, C, E, G, P, R, 0, 2, 4, 7 85 – 260 V AC B, D, F, H, Q, S, 1, 3, 5, 8 20 – 55 V AC/16 – 62 V DC * Type of protection A, B, C, D, J, K, L, M, N, P, Q, V, 2, 4, 5, 7 non Ex ia (intrinsically safe) input /output F, G, S, T Ex ia (intrinsically safe) input /output DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*W**-************ 25 to 1200 1) +50 °C 50 50 50 50 50 50 50 to 2000 2) +50 °C 80 80 80 80 80 80 DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*P**-************ 25 to 200 1) +40 °C 80 95 130 150 150 150 15 to 600 2) 130 130 130 25 to 200 1) +45 °C 80 95 130 130 130 130 15 to 600 2) 25 to 200 1) +50 °C 80 95 95 95 95 95 15 to 600 2) DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*H**-************ 2 to 150 +40 °C 80 95 130 150 150 150 +45 °C 80 95 130 130 130 130 +50 °C 80 95 95 95 95 95

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 7 Temperature table remote version Sensor Max. medium temperature [°C] for T1 ~ T6 in relation to the maximum ambient temperature Ta 1) with polyurethane lining; 2) with hard rubber lining 1) with PFA lining; 2) with PTFE lining; 3) versions in IP 68 are limited to 130 °C. The minimum ambient temperature is –20 °C. A version for ambient temperatures up to –40 °C is optionally available. Transmitter The transmitter of the remote version has the temperature class T6 when installing into the Ex d housing up to an ambient temperature of Ta = 60 °C. The maximum ambient temperature range is –20 to +60 °C. A version for ambient temperatures up to –40 °C is optionally available. DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*W**-************ 25 to 1200 1) +50 °C 50 50 50 50 50 50 50 to 2000 2) +60 °C 80 80 80 80 80 80 DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*P**-************ 25 to 200 1) +50 °C 80 95 130 150 3) 150 3) 150 3) 15 to 600 2) 130 130 130 25 to 200 1) +60 °C 80 95 130 130 130 130 15 to 600 2) DN Ta T6 (85 °C) T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) Promag 5*H**-************ 2 to 150 +50 °C 80 95 130 150 150 150 2 to 25 +60 °C 80 95 130 130 130 130 40 to 150 80 95 130 150 150 150

0NEPSI Zone 1, Zone 21

Proline Promag 53 8 Endress+Hauser Gas and dust explosion protection Determine the temperature class for gas in relation to the ambient Ta and medium temperature TM Determine the maximum surface temperature for dust in relation to the max. ambient temperature Ta and max. medium temperature TM Example Device: compact version, Promag 53 , DN 25, PTFE Max. ambient temperature: Ta = 45°C Max. medium temperature: TM = 110 °C A0005216 Fig. 3: Procedure for calculating the max. surface temperature 1. In the associated temperature table (compact version), the selection of the measuring device (Promag 53P), the nominal diameter (DN 25), the lining material (PTFE) and the ambient temperature Ta (45 °C) determine the line in which the max. medium temperature can be found. 2. The fluid temperature TM (110 °C), which is smaller or equal to the max. fluid temperature, determines the column, i.e. the temperature class, for gas (110 °C  130 °C T4). 3. The maximum temperature of the calculated temperature class corresponds to the maximum surface temperature (T4 = 135 °C = maximum surface temperature for dust). Design of measuring system A0005610 Fig. 4: Design of the measuring system, compact/remote version A Transmitter housing (compact version) B Transmitter housing on connection housing, remote version C Sensor connection housing, remote version, Promag W/P D Sensor connection housing, remote version, Promag H, DN  E Sensor connection housing, remote version, Promag H, DN  40 a Screw terminal for connecting to the potential equalization b Connection compartment cover c Connecting cable, coil current cable d Connecting cable, signal cable m to n see following chapter "Cable entries" !Note! Connection of remote version connecting cable Page 10 Promag 5*P**-************ DN T T6 (85 °C) a T5 (100 °C) T4 (135 °C) T3 (200 °C) T2 (300 °C) T1 (450 °C) 25...200 1) 2) 15...600 25...200 1) 15...600 2) 25...200 1) 2) 15...600 +40 °C +45 °C +50 °C 80 80 80 80 80 80 95 95 95 95 95 95 130 130 130 130 95 95 150 130 130 130 95 95 150 130 130 130 95 95 150 130 130 130 95 95 1) PFA; PTFE 2) DN 25 PTFE Ta= 45 °C TM = 110 °C � 130 °C) ( A B C b D E a b b a a a c d c d c d N ic h t u nt e r S p a n n u n g öf f n e n K e e p c o v e r ti g h t w hi le ci r cu it s a re a li v e N e p a s o uv ri r l ’a p p ar e il so u s t e n s io n K e e p c o v er ti gh t w h il e ci rc ui ts ar e a li v e N ic h t u nt e r S p a n n u n g ö ff n e n K e e p c o v er ti g ht w hi le ci rc ui ts a re a li v e N e p a s o uv ri r l ’a p p ar ei l s o u s t e n s io n K e e p c o v e r ti g ht w h il e ci r cu it s ar e a li v e m m n m n n n

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 9 Cable entries • m for connection compartment (Ex d version): power supply cable and cable of the communication circuit  Choice of thread for cable entries M20 × 1.5, ½" NPT or G ½". Make sure that the Ex d cable glands/entries are secured to prevent working loose and that the seals are installed immediately adjacent to the housing. • m for connection compartment (Ex e version): power supply cable and cable of the communication circuit  Choice of cable gland M20 × 1.5 or thread for cable entries ½" NPT or G ½".  The cables must be installed such that they are fixed in place. Adequate strain relief must be ensured. • n for remote version connecting cable:  Choice of cable gland M20 × 1.5 or thread for cable entries ½" NPT or G ½" # Warning! The leak-tight of the cable glands and cable entries is to ensure. Cable specification You can find information about the cable specification in the associated Operating Instructions. Potential equalization • The transmitter (compact and remote version) is to be securely connected to the potential equalization system using the screw terminal on the outside of the transmitter housing. Alternatively, the transmitter of the compact version as of serial number 4Axxxxxx000 can be connected to the potential equalization system via the pipeline if a ground connection via the pipeline according to regulations can be assured. • When using the remote version, the connection housing of the sensor must be grounded via the external screw terminal. Alternatively, the sensor can be integrated into the potential equalization via the pipeline as long as the pipeline provides a ground connection conforming to regulations. !Note! Further information about potential equalization, shielding and grounding can be found in the associated Operating Instructions. Potential equalization with shield grounded at both sides for fieldbus version A0005215 Fig. 5: Example for connecting potential equalization lines 1 Distributor/T-Box 2 Bus devices for potentially explosive atmospheres 3.1 Bus terminator PROFIBUS PA and FOUNDATION Fieldbus 3.2 Bus terminator PROFIBUS DP and Modbus 4 Bus supply unit or automation system 4a Potential equalization line is fed out into the safe area !Note! The length of the spur must be observed. 1 1 2 2 3.1 4 4a EX EX R C 3.2 R GND +5 V R R

0NEPSI Zone 1, Zone 21

Proline Promag 53 10 Endress+Hauser Connection of remote version connecting cable A0005614 Fig. 6: Connection of remote version connecting cable A Transmitter housing on connection housing, remote version B Sensor connection housing, remote version, Promag W/P C Sensor connection housing, remote version, Promag H, DN  D Sensor connection housing, remote version, Promag H, DN  40 a Ground terminals (are provided for connecting a potential equalization connection) b Connecting cable, coil circuit c Signal circuit connecting cable (electrodes) n.c.  not connected, insulated cable screening Wire colors  terminal number: 5/6 = brown, 7/8 = white, 4 = green, 37/38 = yellow Terminal assignment and connection data The connection of the remote version, between sensor and transmitter, is carried out in the explosion protections Ex i (signal circuit) and Ex e (coil circuit). When using the connecting cable supplied by Endress+Hauser, the intrinsic safety of the circuit is ensured at a maximum cable length of 90 m (for IIC) and 200 m (for IIB). " Caution! Only preterminated connecting cables supplied by Endress+Hauser must be used. Terminal 6 5 7 8 4 37 36 42 41 S1 E1 E2 S2 GND E S 2 1 Designation Measured variable Pipe EPD Signal circuit Coil circuit Functional values U = 60 V P = 2.5 W Electric circuit [Ex ia] IIC/IIB Ex e UO 37 V IO 25 mA PO 138 mW LO LO IIC = 50 mH / LO IIB = 200 mH CO CO IIC = 39 nF / CO IIB = 353 nF A B b C D c b c a b b c 6 5 5 7 7 8 4 4 37 37 36 42 42 41 41 E1 E2 GNDE 2 1 S1 E1 E2 S2GND E S n.c. n.c. n.c. n.c. n.c. n.c. a N ic h t u nt e r S p a n n u n g öf f n e n K e e p c o v e r ti g h t w hi le ci r cu it s a re a li v e N e p a s o uv ri r l ’a p p ar e il so u s t e n s io n K e e p c o v er ti gh t w h il e ci rc ui ts ar e a li v e

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 11 Electrical connection Connection compartment Transmitter housing compact/remote version (terminal assignment, connection data  Page 11 ff.) Fig. 7: Electrical connections *) Fixed communication boards (permanent assignment) **) Flexible communication boards a Power supply cable (terminal assignment, connection data Page 11) b Signal cable (terminal assignment, connection data Page 12ff.) c Ground terminal for signal cable shield / fieldbus cable / RS485 line d Ground terminal for protective ground e Service adapter for connecting service interface FXA 193 (Fieldcheck, FieldCare) f Further connections: – PROFIBUS DP *: Cable for external termination, optional (terminal assignment, connection data Page 11) – PROFIBUS DP **/ Modbus RS485 */**: Signal cable (terminal assignment, connection data Page 12) Terminal assignment and connection data, power supply Terminal assignment and connection data 4 to 20 mA HART A0005611 PROFIBUS PA Ex ia * FOUNDATION Fieldbus * A0005613 PROFIBUS DP */** Modbus RS485 */** A0005617 1 2 L1 (L+) N (L–) 20 21 22 23 24 25 26 27 e b c a + – + – + – + – d 1 2 L1 (L+) N (L–) 20 21 22 23 24 25 26 27 e 3 b c a + – + – + – + – d 1 2 L1 (L+) N (L–) 20 21 22 23 24 25 26 27 e 3 b c a + – + – + – + – f d All transmitters 1 L (+) 2 N (–) * Designation Supply voltage Protective earth Functional values AC: U = 85 to 260 V; AC: U = 20 to 55 V DC: U = 16 to 62 V Power consumption: 15 VA / 15 W Caution! Observe the grounding plans of the system! Intrinsically safe circuit no Um 260 V AC

0NEPSI Zone 1, Zone 21

Proline Promag 53 12 Endress+Hauser Terminal assignment and connection data for signal circuits (intrinsically safe circuits) !Note! The following tables contain values/specifications, which are dependent on the type code (type of measuring device). Please compare the following type code to the one shown on the nameplate of your measuring device. A graphic representation of the electrical connections can be found on Page 11. The communication circuits, option F and G, they meet all requirements for a FISCO Field Device  (IEC 60079-27). Terminal assignment of transmitter 53***-***********F Terminal assignment of transmitter 53***-***********G Transmitter Terminal no. (inputs/outputs) 20 (+) 21 (–) 22 (+) 23 (–) 24 (+) 25 (–) 26 (+) 27 (–) Assignment – – – PROFIBUS PA passive PA + PA – Electric circuit – – – Ex ia Safety-related values – – – Ui Ii Pi Li Ci FISCO 30 V DC 600 mA 8.5 W  10 μH  5 nF Field device Functional values – – – galvanically isolated, UBus 9 to 32 V DC IBus 11 mA IEC 61158-2 (MBP) Transmitter Terminal no. (inputs/outputs) 20 (+) 21 (–) 22 (+) 23 (–) 24 (+) 25 (–) 26 (+) 27 (–) Assignment – – – FOUNDATION Fieldbus passive FF + FF – Electric circuit – – – Ex ia Safety-related values – – – Ui Ii Pi Li Ci FISCO 30 V DC 600 mA 8.5 W 10 μH 5 nF Field device Functional values – – – galvanically isolated, UBus 9 to 32 V DC IBus 12 mA IEC 61158-2 (MBP)

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 13 Terminal assignment of transmitter 53***-***********S Terminal assignment of transmitter 53***-***********T Transmitter Terminal no. (inputs/outputs) 20 (+) 21 (–) 22 (+) 23 (–) 24 (+) 25 (–) 26 (+) 27 (–) Assignment – – Pulse/frequency output, passive Current output HART, active Electric circuit – – Ex ia Ex ia Safety-related values – – Ui Ii Pi Li Ci 30 V DC 500 mA 600 mW negligible 6 nF Uo Io Po Lo IIC/IIB Co IIC/IIB 1) Lo IIC/IIB 1) Co IIC/IIB Ui Ii Pi Li Ci 21.8 V DC 90 mA 491 mW 4.1 mH/15 mH 160 nF/1160 nF 2 mH/10 mH 80 nF/300 nF 30 V DC 2) 10 mA 2) 0.3 W 2) negligible 6 nF Functional values – – galvanically isolated, passive: 30 V DC / 250 mA Open Collector Full scale frequency  2 to 5000 Hz galvanically isolated, active: 0/4 to 20 mA RL < 400  RL HART  250  1) Permitted values in the event of simultaneous occurrence of concentrated inductances and capacitances. 2) The interconnection must be assessed according to the valid construction provisions. Transmitter Terminal no. (inputs/outputs) 20 (+) 21 (–) 22 (+) 23 (–) 24 (+) 25 (–) 26 (+) 27 (–) Assignment – – Pulse/frequency output, passive Current output HART, passive Electric circuit – – Ex ia Ex ia Safety-related values – – Ui Ii Pi Li Ci 30 V DC 500 mA 600 mW negligible 6 nF Ui Ii Pi Li Ci 30 V DC 100 mA 1.25 W negligible 6 nF Functional values – – galvanically isolated, passive: 30 V DC / 250 mA Open Collector Full scale frequency  2 to 5000 Hz galvanically isolated, passive: 4 to 20 mA voltage drop 9 V RL < [(Vp. supply – 9 V) ÷ 25 mA]

0NEPSI Zone 1, Zone 21

Proline Promag 53 14 Endress+Hauser Terminal assignment and connection data for signal circuits (non-intrinsically safe circuits) !Note! The following tables contain values/specifications, which are dependent on the type code (type of measuring device). Please compare the following type code to the one shown on the nameplate of your measuring device. A graphic representation of the electrical connections can be found on Page 11. Terminal assignment Transmitter Terminal no. (inputs/outputs) 20 (+) 21 (–) 22 (+) 23 (–) 24 (+) 25 (–) 26 (+) 27 (–) Non-convertible communication boards (fixed assignment) 53***-*…*A – – Pulse/frequency output Current output HART 53***-*…*B Relay output 2 Relay output 1 Pulse/frequency output Current output HART 53***-*…*J – – External termination PROFIBUS DP * +5 V DGND B A 53***-*…*K – – – FOUNDATION Fieldbus FF + FF – 53***-*…*Q – – Status input Modbus RS485 * B A Convertible communication boards 53***-*…*C Relay output 2 Relay output 1 Pulse/frequency output Current output HART 53***-*…*D Status input Relay output Pulse/frequency output Current output HART 53***-*…*L Status input Relay output 2 Relay output 1 Current output HART 53***-*…*M Status input Pulse/frequency output Pulse/frequency output Current output HART 53***-*…*N Current output Pulse/frequency output Status input Modbus RS485 * B A 53***-*…*P Current output Pulse/frequency output Status input PROFIBUS DP * B A 53***-*…*V Relay output 2 Relay output 1 Status input PROFIBUS DP * B A 53***-*…*2 Relay output Current output Pulse/frequency output Current output HART 53***- *…*4 Current input Relay output Pulse/frequency output Current output HART 53***-*…*5 Status input Current input Pulse/frequency output Current output HART 53***-*…*7 Relay output 2 Relay output 1 Status input Modbus RS485 * B A Safety-related and functional values of signal circuits Page 15 * PROFIBUS DP, Modbus RS485: – terminal 26 (+) B (RxD/TxD-P) – terminal 27 (–) A (RxD/TxD-N)

Proline Promag 53

NEPSI Zone 1, Zone 21 0 Endress+Hauser 15 Safety-related and functional values of signal circuits Signal circuits Functional values Safety-related values Current output HART galvanically isolated, active/passive can be selected: • active: 0/4 to 20 mA RL < 700 , RL HART  250  • passive: 4 to 20 mA Vs = 18 to 30 V DC, Ri  150  intrinsically safe Um Im = no = 260 V = 500 mA Current output galvanically isolated, active/passive can be selected: • active: 0/4 to 20 mA RL < 700  • passive: 4 to 20 mA Vs = 18 to 30 V DC, Ri  150  Pulse/frequency output galvanically isolated, active/passive can be selected: • active: 24 V DC / 25 mA (max. 250 mA during 20 ms) RL > 100  • passive: 30 V DC / 250 mA Open Collector Full scale frequency 2 to 10000 Hz (fmax = 12500 Hz) Status output galvanically isolated, Open Collector max. 30 V AC / 250 mA Relay output galvanically isolated, max. 30 V AC / 500 mA max. 60 V DC / 100 mA Current input galvanically isolated, active/passive can be selected: • active: 4 to 20 mA Ri  150  Uout = 24 V DC, short-circuit proof • passive: 0/4 to 20 mA Ri  150  Umax = 30 V DC Status input (53***-*...*D, L, M) galvanically isolated, 3 to 30 V DC Ri = 5 k Status input (53***-*...*N, P, Q, V, 7) galvanically isolated, Independent of polarity, 3 to 30 V DC Ri = 3 k PROFIBUS DP galvanically isolated, RS485 as per Standard EIA/TIA-485 FOUNDATION Fieldbus galvanically isolated, UBus = 9 to 32 V DC IBus = 12 mA IEC 61158-2 (MBP) Modbus RS485 galvanically isolated, RS485 as per Standard EIA/TIA-485

0NEPSI Zone 1, Zone 21

Proline Promag 53 16 Endress+Hauser Service adapter The service adapter is only used for connecting service interfaces approved by Endress+Hauser. # Warning! It is not permissible to connect the service adapter whilst the atmosphere is considered to be explosive. Device fuse # Warning! Use only fuses of the following types; the fuses are installed on the power supply board: • Voltage 20 to 55 V AC / 16 to 62 V DC: fuse 2.0 A slow-blow, disconnect capacity 1500 A (Schurter, 0001.2503 or Wickmann, Standard Type 181 2.0 A) • Voltage 85 to 260 V AC: fuse 0.8 A slow-blow, disconnect capacity 1500 A (Schurter, 0001.2507 or Wickmann, Standard Type 181 0.8 A) Technical Data Dimensions Please refer to the respective Technical Information for these dimensions: • Promag 50W, 53W TI00046D • Promag 50P, 53P TI00047D • Promag 50H, 53H TI00048D Weight The weight of the Ex d version is approx. 2 kg greater than that of the standard version.