Mastech MS2109A Operation & User’s Manual
Read Online
Read the Mastech MS2109A Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
AC/DC Clamp Meter
V
Ω
Hz%
TEMP
MS2109A
AC/DC CLAMP METER
NCV
SEL
REL
Hz%
MAX/MIN
/
60
1. Safety Information….......................1
1.1 Preparation.......................................1
CONTENTS
1.2 Usage................................................2
1.3 Signs and Labels................................3
1.4 Maintenance......................................4
2. Description….................................4
2.1 Part Name..........................................5
2.2 Switch and Button Description..............7
2.3 LCD Display........................................8
3. Specification…...............................9
3.1 General..............................................9
3.2 Technique Data...................................9
4. Operating guidance.......................15
4.1 Reading Hold....................................15
4.2 Measuring Range Switch....................15
4.3 Frequency/Duty Ratio Switch.............15
4.4 Maximum/Minimum
Measurement Choice....................16
4.5 Function Choice..................................17
4.6 Relative Measurement and
Surge Measurement..........................17
4.7 Back Light and Clamp Head Light..........17
4.8 Automatic Power-Off
4.9 Measurement Preparation....................18
4.10 Current Measurement........................19
4.11 Voltage Measurement........................20
4.12 Frequency and
Duty Ratio Measurement
4.13 Resistance Measurement..................23
4.14 Diode Test
4.15 Circuit Continuity test
4.16 Capacitance Measurement................25
5. Maintenance...................................27
5.1 Replace Battery...................................27
5.2 Replace Probe
6. Accessories....................................27
........................24
........................................24
..................21
............................18
4.17 NCV Measurement............................26
4.18 Temperature Measurement................26
....................................27
1. Safety Information
1.2 Usage
Please particularly note that inappropriate use
may cause shock or damage to the meter. When
using the meter, comply with common safety
procedures and completely follow the safety
measures in this operation manual.
To make full use of the meter's functions and
ensure safe operation, please carefully read and
follow the procedures in the operation manual.
The meter meets GB/T 13978-92 digital multimeter
general technology conditions, GB4793.1-1995 (IEC
61010-1, IEC 61010-2-032) electronic measurement
instrument safety requirements with secondary pollution
and over-voltage standard of CAT III 600V.
Please follow the safety operation guidelines to ensure
the safe use of meter.
The meter will provide satisfactory service to you if you
use and protect it appropriately.
Warning
1.1 Preparation
1.1.1 When using the meter, the user should comply with
standard safety rules:
- General shock protection
- Prevent misusing the meter
1.1.2 After receiving the meter, please check for damage
that may have occurred during transportation.
1.1.3 If the meter is stored and shipped under hard
conditions, please confirm that the meter is
undamaged.
1.1.4 Probe should be in good condition. Before use,
please check whether the probe insulation is
damaged, whether metal wire is bare.
1.1.5 Use the probe table provided with the meter to
ensure safety. If necessary, replace the probe
with another identical probe or one with the same
specification.
1.2.1 When using the meter, select the right function
and measuring range.
1.2.2 Don't measure when exceeding the maximum
value in each measuring range.
1.2.3 When measuring a circuit with the meter
connected, do not touch the probe tip (metal part).
1.2.4 When measuring, if the voltage to be measured
is more than 60 V DC or 30 V AC (RMS), always
keep your fingers behind the finger protection
device.
1.2.5 Do not measure voltage greater than AC 600V.
1.2.6 In the manual measuring range, when the value
to be measured is not known in advance, choose
the highest measuring range to begin, and then
gradually select lower ranges until the correct
range is identified.
1.2.7 Before rotating conversion switch to change
measuring function, remove probe from the
circuit to be measured.
1.2.8 Don't measure resistors, capacitors, diodes and
circuits that are energized.
1.2.9 During the test of current, resistors, capacitors,
diodes and circuit connections, do not allow the
meter to connect with voltage source.
1.3 Part name
Note (Important safety information. Refer to the
operation manual).
Dangerous electric conductor.
Double insulation protection (class II).
CAT III According to pulse voltage tolerance protection
level provided by IEC 61010-1 standard
overvoltage (installation) level III and grade
2 for pollution.
The meter complies with EU standard
Ground (earth)
1.4 Maintenance
1.4.1 Don't try to open the meter bottom case to adjust or
repair. Such operations only can be made by
qualified technicians who fully understand the
meter and electrical shock hazard.
1.4.2 Before opening the meter bottom case or battery
cover, remove probe from the circuit to be measured.
1.4.3 To avoid wrong readings and potential electric
shock, when “ ”appears on the meter display,
replace the battery immediately.
1.4.4 Clean the meter with damp cloth and mild
detergent. Do not use abrasives or solvents.
1.4.5 Power off the meter when it is not used. Switch the
transfer switch to OFF position.
1.4.6 If the meter is not used for long time, please take
the battery out to prevent the meter being damaged.
1.2.10 Don't measure capacitance unless the capacitor
is discharged completely.
1.2.11 Don't use the meter in explosive gas, vapor or
dusty environments.
1.2.12 If you find any abnormal condition or failure on
the meter, stop using the meter and have it
serviced by a qualified technician.
1.2.13 Unless the meter bottom case and the battery
cover are completely fastened in their original
places, do not use the meter.
1.2.14 Don't store or use the meter in direct sunlight,
high temperature or high humidity.
2. Description
- The meter is a portable, professional measuring
instrument with LCD display and back light for easy
reading by users. Measuring range switch is operated
by single hand for easy operation. The meter has
overload protection and low battery indicator. It is an
ideal multifunction meter for professionals, factories,
schools, fans and family use.
- The meter is used to measure AC current, DC current,
voltage, DC voltage, frequency, duty ratio, resistance,
capacitance, circuit connections, diodes, non-contact
voltages and temperatures.
- The meter has an auto measuring range function.
- The meter has a reading hold function.
- The meter has a max. measuring function.
- The meter has a min. measuring function.
(1) Current clamp head: used for current measurement.
(2) NCV indicator
(3) Key
(4) LCD display
(5) Common end jack
(6) Resistance, capacitance, voltage, frequency, diode,
continuity and temperature input jack
(7) Knob
(8) Clamp head center position
(9) NCV sense position
2.1 Part name
- The meter has a clamp head frequency measurement
function.
- The meter has an auto power-off function.
- The meter has a relative measuring function.
1
2
3
4
9
6
7
8
5
V
Ω
Hz%
TEMP
MS2109A
AC/DC CLAMP METER
NCV
SEL
REL
Hz%
MAX/MIN
/
60
2.2 Switch, Button and Input Jack Description
B.L/HOLD button: used for reading hold or
back light control
MAX/MIN button: used for maximum/minimum
measurement function switch.
SEL button: used for measuring function switch.
NCV button: Non-contact voltage detection switch.
REL button: used for entering relative measurement state.
Hz/% button: used for frequency and duty ratio
measurement function switch.
OFF position: used for shutting off the power.
INPUT jack: voltage, resistance, frequency, duty ratio,
capacitance, diode, circuit connection,
temperature measurement input wire
connecting terminal.
COM jack: voltage, resistance, frequency, duty ratio,
capacitance, diode, circuit connection,
temperature measurement common wire
connecting terminal.
Transfer switch: used for selecting function and
measuring range.
2.3 LCD Display
ALTERNATING CURRENT, direct current
Diode, continuity
Automatic measuring range mode
Maximum measurement state
AUTO
Minimum measurement state
Relative measurement mode
MAX
MIN
Automatic power-off state
REL
LOW BATTERY
Reading hold state
Percentage (duty ratio)
%
Millivolt, Volt (voltage)
mV, V
A
Amperes (Current)
nF, μF
Nanofarad, Microfarad
Ohm, Kilohm, Megohm (resistance)
Ω,kΩ, MΩ
H
Hz,kHz,MHZ Hertz, Kilohertz, Megahertz (frequency)
Celsius degree, Fahrenheit degree
(temperature)
°C,°F
At least once a year, the meter should be recalibrated
in temperatures of 18°C ~ 28°C and relative humidity
less than 75%.
3.1 General
Auto measuring range.
Full measuring range overload protection.
The maximum allowable voltage between measurement
end and ground: 600V DC or 600V AC
Operational height: maximum 2000m
Display: LCD
Displayed maximum value: 5999 digit.
Polarity indication: automatic indication, '-' means
negative polarity.
Exceeding measuring range display: '0L' or '-0L'.
Sampling rate: about 3 times/s, simulation bar 30 times/s.
Unit display: has function and power unit display.
Auto off time: 15 min
Power supply: 1.5V 3AAA battery.
Battery undervoltage indication: LCD displays symbol.
Temperature coefficient: less than 0.1x accuracy°C
Operating temperature: 18°C~28°C.
Storage temperature: -10°C~50°C
Dimension: 220×81×41mm
Weight: about 286g (include battery)
3.2 Technical Indicators
Environment temperature: 23±5°C, relative humidity
(RH): <75%
3.2.1 AC Current
Measuring
range
Resolution
Accuracy
±(2.5% of reading + 6 digits)
60A
600A
0.01A
0.1A
- Maximum input current: 600A AC
- Frequency range: 40~400Hz
3.2.3 DC Voltage
600mV
0.01V
0.1V
±(0.7% of reading + 2digits)
6V
60V
600V
0.1mV
0.001V
- Input impedance: 10MΩ
- Maximum input voltage: 600V AC (RMS) or 600V DC
3.2.2 DC Current
±(3.0% of reading + 6 digits)
60A
600A
0.01A
0.1A
- Maximum input current: 600A DC
Measuring
range
Resolution
Accuracy
Measuring
range
Resolution
Accuracy
3.2.4 AC Voltage
0.01V
0.1V
±(0.8% of reading + 3 digits)
6V
60V
600V
600V
0.001V
1V
±(1.0% of reading + 4 digits)
- Measuring scope: 40Hz ~ 10kHz
- The input voltage range: > 600mV AC (RMS) (input
voltage will increase when the frequency to be
measured increases)
- Input impedance: 10MΩ
- Maximum input voltage: 600V AC (RMS)
- Input impedance: 10MΩ
- Maximum input voltage: 600V AC (RMS) or 600V DC
- Frequency range: 40 ~ 400Hz
3.2.5 Frequency
3.2.5.1Clamp head measuring frequency
(through grade A):
± (1.5% of reading + 5 digits)
99.99Hz
999.9Hz
0.1KHz
0.01KHz
- Measuring scope: 40Hz ~ 1kHz
- The input signal range: > 6A AC (RMS) (input current
will increase when the frequency to be measured
increases)
- Maximum input current: AC 600A (RMS)
3.2.5.3 Through grade HZ/DUTY:
9.999Hz
±(0.5% of reading + 3 digits)
99.99Hz
0.001KHz
0.01Hz
0.001Hz
0.1Hz
999.9Hz
9.999KHz
99.99KHz
999.9KHz
9.999MHz
0.001MHz
0.1KHz
0.01KHz
- Overload protection: 250V DC or AC (RMS)
- The input voltage range: > 2V (input voltage will
increase when the frequency to be measured
increases)
Note:
In the small voltage measuring range, the probe is not
connected with the circuit to be tested, and the meter
may have fluctuating readings. This is normal and
caused by the meter's high sensitivity. It does not
affect actual measurement results.
Measuring
range
Resolution
Accuracy
Measuring
range
Resolution
Accuracy
± (1.5% of reading + 5 digits)
99.99Hz
999.9Hz
0.1Hz
0.01Hz
Measuring
range
Resolution
Accuracy
9.999kHz
0.001KHz
3.2.5.2 Through grade V:
Measuring
range
Resolution
Accuracy
If the resistance of circuit to be
measured is less than 70Ω, the
meter's built-in buzzer may sound.
3.2.6.2 Through grade V:
- Frequency response: 40 ~ 10kHz
- Input voltage range: > 600mV AC
- Input impedance: 10MΩ
- Maximum input voltage: 600V AC (RMS)
3.2.6.2 Through grade HZ/DUTY:
- Frequency response: 1 ~ 10MHz
- The input voltage range: >2V AC (RMS) (input
voltage will increase when the frequency to be
measured increases)
- Maximum input voltage: 250V AC (RMS)
3.2.7 Resistance
60kΩ
600kΩ
6MΩ
60MΩ
±(1.2% of reading + 3 digits)
- Open circuit voltage: about 0.4V
- Overload protection: 250V DC or AC (RMS)
3.2.8 Circuit Continuity Test
- Overload protection: 250V DC or AC (RMS)
0.1Ω
Display approximate diode
forward voltage value
3.2.10 Diode Test
- Forward DC current is about 1mA
- Backward DC voltage is about 3.3V
- Overload protection: 250V DC or AC (RMS)
Accuracy
40nF
±(4.0% of reading + 5 digits)
0.01μF
0.1nF
0.01nF
0.001μF
0.001mF
0.1μF
3.2.9 Capacitance
4μF
- Overload protection: 250V DC or AC (RMS)
400nF
40μF
400μF
4000μF
3.2.6.1Through grade A (from clamp head):
- Frequency response: 40 ~ 1kHz
- Input current range: >6A AC (RMS)
- Maximum input current: AC 600A
3.2.6 Duty Ratio
Measuring range
Resolution
Accuracy
0.1%-99.9%
0.1%
±(3%+5)
Measuring range
±(0.8% of reading + 3 digits)
Measuring
range
Resolution
Measuring
range
Resolution
Measuring
range
Resolution
Function
4. Operating Guidance
1) In the process of measurement, if reading hold is
required, press “HOLD/B.L” key. The display value
in the display will be locked. Press “HOLD/B.L”
key again to release reading hold.
4.1 Reading Hold
4.2 NCV Key
NCV key is used for non-contact voltage detection.
Press this key and hold it. Place the NCV sensor
zone close the conductor to be detected. It voltage
is more than AC 90V, NCV indicator will flash and
the buzzer will sound at regular intervals.
4.3 Frequency/Duty Ratio Switch
1) When the meter is in AC voltage, AC current position,
if “Hz/%” button is pressed, the meter will measure
AC voltage and AC current signal frequency. Click
“Hz/%” button again, and the meter will measure
voltage and current signal duty ratio. If it is in
HZ/DUTY position, pressing HZ % key will switch
between HZ and DUTY.
4.4 Maximum/Minimum Measurement Choice
1) Press “MAX/MIN” key to enter MAX mode, and
always store measurement maximum value. Press
“MAX/MIN” key again, and the meter will enter
minimum value measurement state. Press
“MAX/MIN” key for the third time, and the meter
will display the difference of maximum and minimum
value. Press “MAX/MIN” key to cycle through the
above operations.
2) After entering MAX or MIN mode, the meter will
automatically save the measured maximum or
minimum value.
3) If the user presses “MAX/MIN” key more than 2 sec,
the meter will restore normal measuring range.
1) When the meter is in the maximum/minimum value
measurement state, it is manual measuring range
mode.
2) When the meter is in the frequency, duty ratio
measurement state, it can't be switched to
maximum/minimum value measurement mode.
3) When the meter is in the maximum/minimum value
measurement state, SEL, REL key is invalid.
Note:
3.2.11 Temperature
Measuring
range
Resolution
Accuracy
-20°C~0°C
±(3.0% of reading + +4 digits)
1°C~400°C
401°C~1000°C
1°C
±(1.0% of reading + +3 digits)
±2.0% of reading
Note:
When the meter is in the maximum/minimum value
measurement state, it can't be switched to frequency,
duty ratio measurement mode.
2) If “Hz/%” button is pressed again, the meter will
revert to voltage, current measurement state.
- Temperature indicator does not include thermocouple error.
- Overload protection: 250V DC or AC (RMS)
4.5 Function Switch
1) In the resistance mode, pressing “SEL” button will
switch among resistance, diode, continuity and
capacitance detection.
2) In the voltage and current mode, pressing “SEL” key
will switch between AC and DC.
3) In the temperature mode, pressing “SEL” key will
switch between Celsius degree and Fahrenheit
degree.
17
18
1) REL key is relative value measurement button. When
the user taps this button, it will enter relative value
measurement mode. The current display value can
be stored in the memory as reference value. When
the user measures later, the display value is the
difference for input value minus reference value. ie.
REL∆(current reading)= Input value - Reference value.
2) The relative value measurement can only be
performed in the manual measuring mode.
4.6 REL Measurement
4.7 Back Light And Clamp Head Light
1) In the process of measurement, if ambient light is too
dark to read, press “B.L/HOLD” key for more than 2
seconds to turn on the backlight. The backlight will
automatically turn off in about 10 seconds.
2) During this period, pressing “B.L/ HOLD” key more
than two seconds will turn off backlight.
3) In the current measuring mode, the meter will turn
on backlight. At the same time it will turn on clamp
head light. Backlight luminophor is LED with high
current draw. If backlight is used often, it will shorten
battery life. Therefore, use the backlight only when
necessary.
4.8 Automatic Power-Off
1) If there is no operation for 15 minutes after turning
the machine on, the meter will automatically power
off to save electricity. Within 1 minute before
shutdown, the buzzer will sound for five times. Just
before powering off, the buzzer will make a long sound.
2) After automatic power-off, press any key to restore
the meter to ready status.
3) Holding the “SEL” key when powering on will cancel
automatic power-off function.
Note:
When battery voltage is <3.6V, LCD displays “ ”
(undervoltage) symbol.
But when the user uses the backlight, when the battery
voltage frequency is >3.6V, battery voltage drops due
to its high working current, “ ” symbol may display.
(When “ ” symbol shows, it does not guarantee the
accuracy of measurement). At this time, don't replace
the battery. Use the meter normally without using
back light, and wait to replace the battery when the
“ ” symbol is displayed.
4.9 Measurement Preparation
1) Turn the transfer switch to turn on the power. When
battery voltage is low (about <3.6V), LCD displays
“ ” symbol, replace the battery.
2) “ ” symbol means that input voltage or current
should not be more than instruction value, which is
to protect the internal line from damage.
3) Place transfer switch to required measuring function
and range.
4) When connecting line, please connect the common
test line first, then connect charged test line.
When removing line, remove charged test line first.
1) Clamping two or more leads of circuit to be tested
simultaneously will give invalid readings.
2) To get accurate readings, place the lead to be
tested at the center of clamp head.
3) “ ”indicates that maximum input AC current is 600 A.
4) To improve measurement precision, in the DC
current measurement state, because the earth and
other magnetic fields have an impact on DC current
measurement precision, if the LCD display is not
zero, place the clamp head vertical to conductor to
be measured, then press REL back to zero, then
measure.
4.11 Voltage Measurement
Note:
Electric shock hazard.
Pay special attention to avoid shock when
measuring high voltage.
Don't input voltage more than AC 600 V RMS.
Warning
1) Insert black probe to COM jack, insert red probe to
INPUT jack, choose appropriate measuring range.
2) Place transfer switch in voltage position At
this time, the meter is DC voltage measurement
state. To measure AC voltage, press SEL key to
enter AC voltage measurement state.
3) Connect the probe with voltage source or both
ends of load in parallel.
4) Read the voltage on the LCD.
Note:
1. In the small voltage measuring range, the probe is
not connected with the circuit to be tested, and the
meter may have fluctuating readings. This is normal
and is caused by the meter's high sensitivity. When
the meter is connected with the circuit to be tested,
you will get actual measured value.
2. “ ”indicates that maximum input voltage is 600V
AC or 600V DC.
3. If the reading measured by the meter is more than
750V AC or 1000V DC, it will send out “beep” alarm.
4.10 Current Measurement
Electric shock hazard.
Remove the probe from the meter before
measuring with current clamp.
1) When measuring switch is placed to position A, the
meter is in AC current measurement state. Choose
the appropriate measuring range. If you can't
determine the size of current to be tested, choose
the maximum measuring range, then choose lower
ranges until the correct range is displayed. If you
want to measure DC current, press SEL key to enter
DC current measurement state.
2) Hold the trigger, open clamp head, and clip one lead
of measurement circuit to be tested into the clamp.
3) Read the duty ratio value on the LCD display.
Warning
V
4.12 Frequency And Duty Ratio Measurement
Electric shock hazard.
Remove the probe from the meter before
measuring with current clamp.
Warning
(1) Place masuring switch nposition A. Choose
appropriate measuring range.
(2) Hold the trigger, open clamp head, clip one lead of
measurement circuit to be tested in the clamp.
(3) Press “Hz/%” key to switch to frequency
meauring state.
(4) Read the current value on the LCD display.
(5) Pressing “Hz/%” again can enter duty ratio
measuring state.
(6) Read the current value on the LCD display.
Note:
(1) Clamping two or more leads of circuit to be tested
simultaneously will give invalid readings
(2) Frequency measurement range is 40Hz~1kHz
If the frequency to be tested is less than 40Hz,
measuring frequency higher than 10 kHz is possible,
but accuracy is not guarantee
(3) Duty ratio measuring range is 10~95%.
(4) “ ” means that maximum input current is 600A
AC (RMS).
1) Clamp head measuring frequency
(through grade current):
2) Through grade voltage:
Electric shock hazard.
Pay special attention to avoid shock when
measuring high voltage.
Don't input voltage more than AC 600V RMS.
Warning
(1) Insert black probe to COM jack, insert red probe
to INPUT jack.
(2) Place transfer switch to position, press SEL to
enter AC voltage measurement state.
(3) Press “Hz/%” key to switch to frequency
measuring state.
(4) Connect the probe with signal or both ends of
load in parallel
(5) Read on the LCD.
(6) Pressing “Hz/%”again to enter duty ratio
measuring state and read data on the LCD.
V
(1) Frequency measurement range is 10Hz ~ 1kHz If
the frequency to be tested is less than 10Hz,LCD
will show “00.0”measuring frequency higher than
10 kHz is possible, but it accuracy is not guarantee
(2) Duty ratio measuring range is 10 ~ 95%.
(3) “ ”means that maximum input voltage is 600V
AC (RMS).
Note:
3) Through grade HZ/DUTY:
Electric shock hazard.
Pay special attention to avoid shock when
measuring high voltage.
Don't input voltage more than AC 250V RMS.
(1) Insert black probe to COM jack, insert red probe to
INPUT jack.
(2) Place transfer switch in position HZ.
(3) Connect the probe with signal or both ends of load
in parallel.
(4) Read on the LCD.
(5) Press “Hz/%” again to enter duty ratio measuring
state.
Warning
Note:
Frequency measurement range is 1 Hz~10 MHz. If the
frequency to be measured is less than 1 Hz, the LCD
will display “00.0”.
4.13 Resistance Test
Electric shock hazard.
When measuring circuit impedance, determine
that the power supply is disconnected and the
capacitor in the circuit is completely discharged.
Warning
1) Insert black probe to COM jack, insert red probe to
INPUT jack.
2) Place measuring range switch to position. At
time, the meter is in the measurement state.
3) Connect the probe to the both ends of resistor or
circuit to be tested..
4) LCD will show readings.
23
Ω
Note:
1) When the input end is open, LCD shows “0L”
over-range state.
2) When the resistance to be tested is >1MΩ, the meter
reading will stabliize after a few seconds, which is
normal for high resistance readings.
4.14 Diode Test
1) Insert black probe to COM jack, insert red probe to
INPUT jack.
2) Place measuring switch in position.
3) Press “SEL” button to switch to measuring state.
4) Connect the red probe to diode anode and connect
the black probe to diode cathode to make test.
5) Readings will display on the LCD.
Ω
Note:
1) What the meter shows is approximation of diode
forward voltage drop.
2) If the probe has reverse connection or the probe
is open, the LCD will show “0L”.
4.15 Circuit Continuity Test
Electric shock hazard.
When measuring circuit continuity, determine
that the power supply is disconnected and the
capacitor in the circuit is completely discharged.
Warning
1) Insert black probe to COM jack, insert red probe to
INPUT jack.
2) Measuring switch is placed to position.
3) Press “SEL” key to switch to circuit continuity
measuring state.
Ω
If the probe is open or circuits resistance to be tested
is more than 600Ω, the display will show “0L”.
4.16 Capacitance Measurement
4) Connect the probe to the both ends of circuit to
be tested.
5) If the resistance of circuit to be measured is less
than 70Ω, the meter's built-in buzzer may sound.
6) Read the circuit resistance value on the LCD.
Electric shock hazard.
To avoid electric shock, before measuring
capacitance, discharge capacitance completely.
Warning
1) Insert black probe to COM jack, insert red probe to
INPUT jack.
2) Measuring switch is placed to position
3) Press “SEL” button three times to switch to
capacitance measuring state.
3) After discharging capacitance completely, connect
the probe to the both ends of capacitor to be tested
for measurement.
4) Read the capacitance on the LCD.
Ω
Note:
1) When measuring bulk capacitor, readings will
stabilize after a few seconds (400μF and 4000μF).
2) To improve the accuracy below 40nF measuring
value, subtract the distributed capacitance of meter
and cable.
4.17 NCV Measurement
1) Press NCV key. .
2) Place the NCV sensor zone close to the conductor.
When test voltage is greater than AC 110V (RMS),
when the meter is close to the conductor, the meter
induction voltage indicator will turn on and buzzer
will sound.
Note:
1: Even there is no indication, voltage may exist still.
Don't use non-contact voltage detector to judge
whether there is voltage in the wire. Detection
operation could be affected by socket design,
insulation thickness, type and other factors.
2: When inputting voltage on the meter input terminal,
due to the existence of the induced voltage, voltage
induction indicator also may light.
3: Interference sources of external environment (such
as flashlight, motor, etc.) may trigger invalid
non-contact voltage detection.
4.18 Temperature Measurement
1) Insert black probe to COM jack, insert red probe to
INPUT jack.
2) Rotate switch to “TEMP” position. Normal
temperature will show on the LCD display
simultaneously.
3) Insert K-type thermocouple to the input jack of the
meter with correct polarity (insert red one to INPUT
jack, and insert black one to COM jack).
4) Measure the inner and outer surfaces of object to
be tested with the measuring end of thermocouple.
5) Read the measuring value from LCD display.
Before opening the meter battery cover,
remove probe from the circuit to be measured
to avoid electric shock.
Warning
1) When the battery indicator “ ” appears, the battery
should be replaced immediately.
2) Unscrew the fastening screw and remove the meter
battery cover.
3) Replace battery.
4) Put the battery cover back as before.
5. Maintenance
5.1 Replace Battery
Note:
Do not reverse the battery connections. Do not replace
battery when the meter is turned on.
5.2 Replace Probe
When replacing probe, use an identical probe
or one with the same specification. The probe
should be in good condition, with a capacity
of 1000V, 10A.
Warning
If the probe is damaged, such as having a bare metal
wire, replace the probe.
6. Accessories
1) probe Level:1000V 10A One pair
2) User Manual 1 pc
3) Battery 6F22 9VOLTS
HYS007092