Icare PRO Manual 

Text of the manual

Icare PRO tonometer Quick Guide

ENGLISH TURNING THE TONOMETER ON AND LOADING THE PROBE 1.�Press�the�main�button. 2.�Go�to�Measure�and�press�the�main�button. 3.�Partially�open�the�probe�package. 4.�Insert�the�probe�into�the�tonometer � from�partially�opened�package� � without�touching�the�probe�(4a).� Lightly press (4b) the�probe�into�the� � probe�base�to�lock�the�probe.� � Be�careful�not�to�bend�the�probe. 5.�Press�the�main�button�once�to� � activate�the�inserted�probe.� MEASUREMENT 1. Select�the�eye�you�want�to�measure�using�the�left�and�right�navigation�buttons. 2. Bring�the�tonometer�near�the�patient’s�eye.�**Note:�the�angle�must�be�90°�in�both�horizontal�and� � vertical�measurement.�** 3. The�distance�from�the�tip�of�the�probe�to�the�cornea�must�be�3-7mm�(1/8�-�9/32�inch).�If�necessary,� � adjust�the�distance�using�the�forehead�support. 4. Press�the�main�button�lightly�to�perform�one�individual�measurement,�taking�care�not�to�shake�the� � tonometer.�The�tip�of�the�probe�should�make�contact�with�the�central�cornea. 5. Make�six�single�measurements.�Once�the�six�measurements�are�complete,�the�final�IOP�is�displayed.� � Save�the�measurement�result�by�pressing�the�main�button�after�results! ONE SHORT BEEP Indicates a correct single measurement. DOUBLE BEEP Indicates an incorrect single measurement. (See error messages from the instruction manual). LONG BEEP Indicates that six succesful measurements are done and IOP result is available on the display. 4a. 4b. 3. From the main menu, select “Turn off” and confirm by pressing the main button twice. If not used for three minutes, the tonometer will automatically switch off. 1. 5. 3–7 mm (1/8 - 9/32’’) 90˚ Correct measuring positionings with Icare PRO are horizontal and vertical.

PARTS OF THE TONOMETER

CLEANING THE PROBE BASE The�probe�base�can�be�reused�after�careful�cleaning.�Make�sure� that�the�probe�base�is�completely�dry�before�using�it. 1. Unscrew�the�probe�base�collar�and�carefully�remove�the�probe��� � base�from�the�tonometer. 2.��Invert�the�probe�base�over�a�container�filled�with�70-100%�� � � isopropyl�alcohol�solution,�allowing�the�probe�base�to�drop�into�� � the�container�and�let�soak�for�5�-�30�minutes. 3.��Dry�the�probe�base�by�blowing�some�compressed�air�into�it. 4.��Carefully�reinsert�the�completely�dry�probe�base,�screw�the� � collar�back�into�the�tonometer. Please�refer�to�the�manual�for�more�detailed�instructions�of� changing�and�cleaning�the�probe�base. CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID UNNECESSARY SERVICE. www.icaretonometer.com Display� Navigation�buttons: Left,�Down,�Up�&�Right� Main�button� �Forehead�support�adjusting�wheel �Forehead�support �Collar �Probe�base � � Please�refer�to�the�manual� for�additional�information.�

Other Measuring Instruments Manuals