Endress+Hauser iTHERM TM411 Technical Information
Read Online
Read the Endress+Hauser iTHERM TM411 Technical Information online using the built-in PDF viewer.
Text of the manual
Easy-to-use imperial version with outstanding
sensor technology
Applications
• Specially designed for use in hygienic and aseptic applications in the Food &
Beverages and Life Sciences industries
• Measuring range: –200 to +600 °C (–328 to +1 112 °F)
• Pressure range up to 40 bar (580 psi)
• Protection class: up to IP69K
Head transmitter
All Endress+Hauser transmitters are available with enhanced accuracy and reliability
compared to directly wired sensors. Easy customizing by choosing one of the
following outputs and communication protocols:
• Analog output 4 to 20 mA, HART®
• PROFIBUS® PA, FOUNDATION Fieldbus™
Your benefits
• User-friendly and reliable from product selection to maintenance
• iTHERM inserts: globally unique, automated production. Full traceability and
consistently high product quality for reliable measured values
• iTHERM QuickSens: fastest response times (t90s: 1.5 s) for optimum process control
• iTHERM StrongSens: unsurpassed vibration resistance (> 60g) for ultimate plant
safety
• iTHERM QuickNeck – cost and time savings thanks to simple, tool-free
recalibration
• iTHERM TA30R: 316L terminal head for easier handling and lower installation
and maintenance costs, and with highest IP69K rating
• International certification: explosion protection e.g. ATEX/IECEx/FM/CSA and in
compliance with hygiene standards according to 3-A®, EHEDG, ASME BPE, FDA,
TSE Certificate of Suitability
Products
Solutions
Services
Technical Information
iTHERM TM412
Trend-setting, modular resistance thermometer
for hygienic and aseptic applications
TI01348T/09/EN/01.17
71362828
Table of contents
Function and system design . . . . . . . . . . . . . . . . . . .
3
iTHERM Hygiene line . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4
Modular design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6
Family of temperature transmitters . . . . . . . . . . . . . . . . .
6
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7
Wiring diagrams for RTD . . . . . . . . . . . . . . . . . . . . . . . .
7
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8
Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . .
11
Performance characteristics . . . . . . . . . . . . . . . . . . 11
Reference conditions . . . . . . . . . . . . . . . . . . . . . . . . . .
11
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . 11
Influence of ambient temperature . . . . . . . . . . . . . . . . .
12
Self heating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13
Insulation resistance . . . . . . . . . . . . . . . . . . . . . . . . . .
15
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . .
15
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Ambient temperature range . . . . . . . . . . . . . . . . . . . . .
18
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . 18
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . .
18
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Process temperature range . . . . . . . . . . . . . . . . . . . . . .
18
Thermal shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
18
Process pressure range . . . . . . . . . . . . . . . . . . . . . . . . .
18
Medium - state of aggregation . . . . . . . . . . . . . . . . . . . . 19
Mechanical construction . . . . . . . . . . . . . . . . . . . .
20
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . .
20
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24
Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24
Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . .
24
Terminal heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Extension neck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Tip shape . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
31
Certificates and approvals . . . . . . . . . . . . . . . . . . .
33
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Hygiene standard . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Other standards and guidelines . . . . . . . . . . . . . . . . . . .
33
Parts in contact with the medium . . . . . . . . . . . . . . . . . .
33
Surface purity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Material resistance . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Material certification . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
33
Protection tube testing and load capacity calculation . . . . .
34
Ordering information . . . . . . . . . . . . . . . . . . . . . . .
34
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . .
35
Communication-specific accessories . . . . . . . . . . . . . . . .
37
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . .
38
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Function and system design
iTHERM Hygiene line
This thermometer is part of the product line of modular thermometers for hygienic and aseptic
applications.
Differentiating factors when selecting a suitable thermometer
TM4x1
TM4x2
Metric version
Imperial version
↓
↓
TM41x characterizes the device that uses cutting-edge technology, with features such as a replaceable insert,
quick-fastening extension neck (iTHERM QuickNeck), vibration-resistant and fast-response sensor technology
(iTHERM StrongSens and QuickSens) and approval for use in hazardous areas
TM411
mm
A0018239
TM412
inch
A0018691
↓
↓
TM40x characterizes the device that uses basic technology, with features such as a fixed, non-replaceable
insert, application in non-hazardous areas, standard extension neck, low-cost unit
TM401
mm
A0018692
TM402
inch
A0018693
Measuring principle
Resistance thermometer (RTD)
These resistance thermometers use a Pt100 temperature sensor according to IEC 60751. The
temperature sensor is a temperature-sensitive platinum resistor with a resistance of 100 Ω at
0 °C (32 °F) and a temperature coefficient α = 0.003851 °C-1.
There are generally two different kinds of platinum resistance thermometers:
• Wire wound (WW): Here, a double coil of fine, high-purity platinum wire is located in a ceramic
support. This is then sealed top and bottom with a ceramic protective layer. Such resistance
thermometers not only facilitate very reproducible measurements but also offer good long-term
stability of the resistance/temperature characteristic within temperature ranges up to
600 °C (1 112 °F). This type of sensor is relatively large in size and it is comparatively sensitive to
vibrations.
• Thin film platinum resistance thermometers (TF): A very thin, ultrapure platinum layer,
approx. 1 μm thick, is vaporized in a vacuum on a ceramic substrate and then structured
photolithographically. The platinum conductor paths formed in this way create the measuring
resistance. Additional covering and passivation layers are applied and reliably protect the thin
platinum layer from contamination and oxidation, even at high temperatures.
The primary advantages of thin film temperature sensors over wire wound versions are their smaller
sizes and better vibration resistance. A relatively low principle-based deviation of the resistance/
temperature characteristic from the standard characteristic of IEC 60751 can frequently be observed
among TF sensors at high temperatures. As a result, the tight limit values of tolerance category A as
per IEC 60751 can only be observed with TF sensors at temperatures up to approx. 300 °C (572 °F).
Measuring system
Endress+Hauser offers a complete portfolio of optimized components for the temperature measuring
point – everything you need for the seamless integration of the measuring point into the overall
facility. This includes:
• Power supply unit/barrier
• Display units
• Data managers
• Overvoltage protection
For more information, see the brochure 'System Products and Data Managers - Solutions for the
loop' (FA00016K)
FieldCare
Memograph M
RSG45
RIA15
TM4xx
PLC
A0033768
1
Example of application, measuring point layout with additional Endress+Hauser components
• iTHERM TM4x2: Installed RTD thermometer with integrated HART® head transmitter
• Display unit RIA15:
- Display of 4 to 20 mA measured values or HART® process variables
- Loop-powered
- Voltage drop ≤1 V (HART® ≤1.9 V)
• Data Management Memograph M RSG45:
- Tamper-proof data storage and access (FDA 21 CFR 11)
- HART® gateway functionality; Up to 40 HART® devices connected at a time
- Communication capabilities: Modbus, Profibus DP, PROFINET, EtherNet/IP
• PLC / FieldCare: Field Data Manager Software MS20 - Automatic service for report generation,
printing reports, read out of data, storing of data, secure export, pdf generation Read out
measured data via online interface or from mass storage Online visualization of instantaneous
values (“live data”) . More information on this can be found in the Technical Information, see
"Documentation".
Modular design
Design
Options
7
U
5
6
1
2
3
4
A0034469
1: Terminal
head → 24
• 316L, high or low head, optionally with display window
• Aluminum, high or low head, with or without display window
• Polypropylene, low head
• Polyamide, high head, without display window
Your benefits:
• Optimum terminal access thanks to low housing edge of bottom
section:
– Easier to use
– Lower installation and maintenance costs
• Optional display: local process display unit for added reliability
• IP69K protection: optimum protection even with high-pressure
cleaning
2: Wiring,
electrical
connection,
output signal
→ 6
• Ceramic terminal block
• Flying leads
• Head transmitter (4 to 20 mA, HART®, PROFIBUS® PA,
FOUNDATION™ Fieldbus), single-channel or two-channel
• Attachable display (optional)
3: Connector or
cable gland
→ 28
• PROFIBUS® PA / FOUNDATION™ Fieldbus connector, 4-pin
• 8-pin connector
• Polyamide cable glands
4: Extension
neck → 28
Quick fastener iTHERM QuickNeck or NPT 1/2"
Your benefits:
• iTHERM QuickNeck: tool-free removal of the insert:
– Saves time/costs on frequently calibrated measuring points
– Wiring mistakes avoided
• IP69K protection: safety under extreme process conditions
5: Process
connection
→ 29
Wide range of different versions.
• Various diameters
• Various tip shapes (straight or reduced)
7: Insert
→ 23
Sensor models: wire wound (WW) or thin-film sensor (TF).
Your benefits:
• iTHERM QuickSens - insert with the world's fastest response
time:
– Insert: 3 mm (0.12 in) or 6 mm (0.24 in)
– Fast, highly accurate measurements, delivering maximum
process safety and control
– Quality and cost optimization
– Minimization of necessary immersion length: better product
protection thanks to improved process flow
• iTHERM StrongSens - insert with unbeatable durability:
– Vibration resistance > 60g: lower life cycle costs thanks to
longer operating life and high plant availability
– Automated, traceable production: top quality and maximum
process safety
– High long-term stability: reliable measured values and high
level of system safety
Input
Measured variable
Temperature (temperature-linear transmission behavior)
Measuring range
Depends on the type of sensor used
Sensor type
Measuring range
Pt100 thin-film
–50 to +200 °C (–58 to +392 °F)
Pt100 thin-film, iTHERM
StrongSens, vibration-
resistant > 60g
–50 to +500 °C (–58 to +932 °F)
Pt100 thin-film, iTHERM
QuickSens, fast-response
–50 to +200 °C (–58 to +392 °F)
Pt100 wire wound, extended
measuring range
–200 to +600 °C (–328 to +1 112 °F)
Output
Output signal
Generally, the measured value can be transmitted in one of two ways:
• Directly-wired sensors - sensor measured values forwarded without a transmitter.
• Via all common protocols by selecting an appropriate Endress+Hauser iTEMP temperature
transmitter. All the transmitters listed below are mounted directly in the terminal head and wired
with the sensory mechanism.
Family of temperature
transmitters
Thermometers fitted with iTEMP transmitters are an installation-ready complete solution to
improve temperature measurement by significantly increasing accuracy and reliability, when
compared to direct wired sensors, as well as reducing both wiring and maintenance costs.
PC programmable head transmitters
They offer a high degree of flexibility, thereby supporting universal application with low inventory
storage. The iTEMP transmitters can be configured quickly and easily at a PC. Endress+Hauser offers
free configuration software which can be downloaded from the Endress+Hauser Website. More
information can be found in the Technical Information. → 39
HART® programmable head transmitters
The transmitter is a 2-wire device with one or two measuring inputs and one analog output. The
device not only transfers converted signals from resistance thermometers and thermocouples, it also
transfers resistance and voltage signals using HART® communication. It can be installed as an
intrinsically safe apparatus in Zone 1 hazardous areas and is used for instrumentation in the
terminal head (flat face) as per DIN EN 50446. Swift and easy operation, visualization and
maintenance by using universal device configuration tools like FieldCare, DeviceCare or
FieldCommunicator 375/475. For more information, see the Technical Information. → 39
PROFIBUS® PA head transmitters
Universally programmable head transmitter with PROFIBUS® PA communication. Conversion of
various input signals into digital output signals. High accuracy over the complete ambient
temperature range. The configuration of PROFIBUS PA functions and of device-specific parameters is
performed via fieldbus communication. For more information, see the Technical Information.
→ 39
FOUNDATION Fieldbus™ head transmitters
Universally programmable head transmitter with FOUNDATION Fieldbus™ communication.
Conversion of various input signals into digital output signals. High accuracy over the complete
ambient temperature range. All transmitters are released for use in all important process control
systems. The integration tests are performed in Endress+Hauser's "System World". For more
information, see the Technical Information. → 39
Advantages of the iTEMP transmitters:
• Dual or single sensor input (optionally for certain transmitters)
• Unsurpassed reliability, accuracy and long-term stability in critical processes
• Mathematical functions
• Monitoring of the thermometer drift, sensor backup functionality, sensor diagnostic functions
• Sensor-transmitter matching for dual sensor input transmitter, based on Callendar/Van Dusen
coefficients
Wiring
• According to the 3-A® Standard electrical connecting cables must be smooth, corrosion-
resistant and easy to clean.
• Grounding or shield connections are possible via special ground terminals on the terminal
head.
Wiring diagrams for RTD
Type of sensor connection
Head mounted transmitter TMT18x (single input)
3
5
6
RTD
3
4
5
6
RTD
1
2
3-wire
4-wire
Power supply
head transmitter and
analog output 4 to 20 mA,
or bus connection
(red)
(red)
(red)
(red)
(white)
(white)
(white)
mA
A0016433-EN
Head mounted transmitter TMT8x (dual input)
-
+
+
1
-
2
7
6
5
4
3
1
2
7
6
5
4
3
Sensor input 2
Sensor input 1
RTD 4- and 3-wire
:
RTD 3-wire
:
Bus connection
and supply voltage
Display connection
white
red
red
white
green
black
black
A0034449-EN
Terminal block mounted
3-wire single
2 x 3-wire single
4-wire single
Outside screw
Red1
Red3
White5
White3
Red1
Red2
Black5
Black6
Green4
Red1
Red3
White4
White6
A0034451-EN
Head mounted field transmitter TMT162 (dual input)
+
-
1
2
3
4
5
6
+
-
3
4
S1
3
S2
-
+
+
-
#
(white)
(white)
(red)
(red)
RTD
RTD
RTD
Sensor 1
Sensor 2
3
3
1
1
4
2
2
(black) 5
(black) 6
(green) 4
Sensor 1
Sensor 2
3-wire
4-wire
Power supply
field transmitter and
analog output
4 to 20 mA,
or bus connection
A0034450-EN
Cable entries
See the 'Terminal head' section
Endress+Hauser offers a wide variety of connectors for the simple and fast integration of the
thermometer into a process control system. The following tables show the PIN assignments of the
various plug connector combinations.
Abbreviations
#1
Order: first transmitter/insert
#2
Order: second
transmitter/insert
i
Insulated. Wires marked 'i' are not connected and are
insulated with heat shrink tubes.
YE
Yellow
GND
Grounded. Wires marked 'GND' are connected to the
internal grounding screw in the terminal head.
RD
Red
BN
Brown
WH
White
GNYE
Green-yellow
PK
Pink
BU
Blue
GN
Green
GY
Gray
BK
Black
Terminal head with one cable entry
Plug
1x PROFIBUS PA
1x FOUNDATION™
Fieldbus (FF)
8-pin
Plug thread
M12
7/8"
7/8"
M12
PIN number
1
2
3
4
1
2
3
4
1
2
3
4
1
2
3
4
5
6
7
8
Electrical connection (terminal head)
Flying leads
Not connected (not insulated)
3-wire terminal
block (1x Pt100)
RD
RD
WH
RD
RD
WH
RD
RD
WH
RD
RD
WH
i
4-wire terminal
block (1x Pt100)
WH
WH
WH
WH
WH
WH
WH
WH
6-wire terminal
block (2x Pt100)
RD
(#1)
1)
RD
(#1)
1)
WH (#1) 1)
RD
(#1)
1)
RD
(#1)
1)
WH (#1) 1)
RD
(#1)
1)
RD
(#1)
1)
WH (#1) 1)
WH
BK
BK
YE
1x TMT 4 to 20
mA or HART®
+
i
-
i
+
i
-
i
+
i
-
i
+
(#1)
i
-
(#1)
i
i
2x TMT 4 to 20
mA or HART® in
the terminal
head with a high
cover
+
(#1)
+
(#2)
-
(#1)
-
(#2)
+
(#1)
+
(#2)
-
(#1)
-
(#2)
+
(#1)
+
(#2)
-
(#1)
-
(#2)
+
(#2)
i
-
(#2)
i
1x TMT
PROFIBUS® PA
+
i
-
GND
2)
+
i
-
GND
2)
Cannot be combined
Cannot be combined
2x TMT
PROFIBUS® PA
+
(#1)
-
(#1)
+
-
1x TMT FF
Cannot be combined
Cannot be combined
-
+
GND
i
Cannot be combined
2x TMT FF
-
(#1)
+
(#1)
PIN position and
color code
1 BN
2 GNYE
3 BU
4 GY
1
4
3
2
A0018929
1 BN
2 GNYE
3 BU
4 GY
1
4
3
2
A0018930
1 BU
2 BN
3 GY
4 GNYE
1
4
3
2
A0018931
1 WH
2 BN
3 GN
4 YE
5 GY
6 PK
7 BU
8 RD
A0018927
1)
Second Pt100 is not connected
2)
If a plastic housing TA30S or TA30P is used, insulated 'i' instead of grounded GND
Terminal head with two cable entries
Plug
2x PROFIBUS® PA
2x FOUNDATION™ Fieldbus (FF)
Plug thread
#1
#2
A0021706
M12(#1) / M12(#2)
7/8"(#1) / 7/8"(#2)
7/8"(#1) / 7/8"(#2)
PIN number
1
2
3
4
1
2
3
4
1
2
3
4
Electrical connection (terminal head)
Flying leads
Not connected (not insulated)
3-wire terminal block (1x Pt100)
RD/i
RD/i
WH/i
RD/i
RD/i
WH/i
RD/i
RD/i
WH/i
4-wire terminal block (1x Pt100)
WH/i
WH/i
WH/i
WH/i
WH/i
WH/i
6-wire terminal block (2x Pt100)
RD/BK
RD/BK
WH/YE
RD/BK
RD/BK
WH/YE
RD/BK
RD/BK
WH/YE
1x TMT 4 to 20 mA or HART®
+/i
i/i
-/i
i/i
+/i
i/i
-/i
i/i
+/i
i/i
-/i
i/i
2x TMT 4 to 20 mA or HART® in the
terminal head with a high cover
+(#1)/
+(#2)
-(#1)/
-(#2)
+(#1)/
+(#2)
-(#1)/
-(#2)
+(#1)/
+(#2)
-(#1)/
-(#2)
1x TMT PROFIBUS® PA
+/i
-/i
GND/
GND
+/i
-/i
GND/
GND
Cannot be combined
2x TMT PROFIBUS® PA
+(#1)/
+(#2)
-(#1)/
-(#2)
+(#1)/
+(#2)
-(#1)/
-(#2)
1x TMT FF
Cannot be combined
Cannot be combined
-/i
+/i
i/i
GND/
GND
2x TMT FF
-(#1)/
-(#2)
+(#1)/
+(#2)
PIN position and color code
1 BN
2 GNYE
3 BU
4 GY
1
4
3
2
A0018929
1 BN
2 GNYE
3 BU
4 GY
1
4
3
2
A0018930
1 BU
2 BN
3 GY
4 GNYE
1
4
3
2
A0018931
Connection combination: insert - transmitter
Insert
Transmitter connection 1)
1x 1-channel
2x 1-channel 2)
1x 2-channel
2x 2-channel 2)
1x Pt100, flying
leads
Pt100 (#1) : transmitter (#1)
Pt100 (#1) : transmitter (#1)
(Transmitter (#2) not
connected)
Pt100 (#1) : transmitter (#1)
Pt100 (#1) : transmitter (#1)
Transmitter (#2) not connected
2x Pt100, flying
leads
Pt100 (#1) : transmitter (#1)
Pt100 (#2) insulated
Pt100 (#1) : transmitter (#1)
Pt100 (#2): transmitter (#2)
Pt100 (#1) : transmitter (#1)
Pt100 (#2) : transmitter (#1)
Pt100 (#1) : transmitter (#1)
Pt100 (#2) : transmitter (#1)
(Transmitter (#2) not
connected)
1x Pt100 with
terminal block 2)
Pt100 (#1) : transmitter in
cover
Cannot be combined
Pt100 (#1) : transmitter in
cover
Cannot be combined
2x Pt100 with
terminal block 2)
Pt100 (#1) : transmitter in
cover
Pt100 (#2) not connected
Pt100 (#1) : transmitter in
cover
Pt100 (#2) : transmitter in
cover
1)
If 2 transmitters are selected in a terminal head, transmitter #1 is installed in the high cover. A TAG cannot be ordered for the 2nd transmitter as
standard. The bus address is set to the default value and, if necessary, must be changed manually before commissioning.
2)
Only in the terminal head with a high cover, only 1 transmitter possible. A ceramic terminal block is automatically fitted on the insert.
To protect against overvoltage in the power supply and signal/communication cables for the
thermometer electronics, Endress+Hauser offers the HAW562 surge arrester for DIN rail mounting
and the HAW569 for field housing installation.
For more information see the Technical Information 'HAW562 Surge arrester' TI01012K and
'HAW569 Surge arrester' TI01013K.
Performance characteristics
Reference conditions
These data are relevant for determining the accuracy of the temperature transmitters used. More
information on this can be found in the Technical Information of the iTEMP temperature
transmitters.
Maximum measured error
RTD resistance thermometer corresponding to IEC 60751
Class
max. Tolerances (°C)
Characteristics
RTD maximum error type TF
Cl. A
± (0.15 + 0.002 · |t| 1))
A
AA
-200 -100 0 100 200 300 400 500 600°C
0.5
1.0
1.5
2.0
B
2.5
3.0
- 0.5
- 1.0
- 1.5
- 2.0
- 2.5
- 3.0
B
A
AA
Max. deviation (°C)
Max. deviation (°C)
A0008588-EN
Cl. AA,
former 1/3
Cl. B
± (0.1 + 0.0017 · |t|) 1)
Cl. B
± (0.3 + 0.005 · |t| 1))
1)
|t| = absolute value °C
For measurement errors in °F, calculate using equations in °C, then multiply the outcome by 1.8.
Temperature ranges
Sensor type
Operating
temperature range
Class B
Class A
Class AA
iTHERM
StrongSens
–50 to 500 °C
(–58 to 932 °F)
-
–30 to 300 °C
(–22 to 572 °F)
0 to 200 °C
(32 to 392 °F)
iTHERM
QuickSens
–50 to 200 °C
(–58 to 392 °F)
-
–30 to 200 °C
(–22 to 392 °F)
0 to 200 °C
(32 to 392 °F)
Pt100 thin film
sensor (TF)
–50 to 200 °C
(–58 to 392 °F)
–50 to 200 °C
(–58 to 392 °F)
–30 to 200 °C
(–22 to 392 °F)
-
Pt100 wire-
wound sensor
(WW)
–200 to 600 °C
(–328 to 1 112 °F)
-
–100 to 450 °C
(–148 to 842 °F)
–50 to 250 °C
(–58 to 482 °F)
Influence of ambient
temperature
Depends on the head transmitter used. For details, see Technical Information. → 39
Self heating
RTD elements are passive resistances that are measured using an external current. This
measurement current causes a self-heating effect in the RTD element itself which in turn creates an
additional measurement error. In addition to the measurement current, the size of the measurement
error is also affected by the temperature conductivity and flow velocity of the process. This self-
heating error is negligible when an Endress+Hauser iTEMP temperature transmitter (very small
measurement current) is connected.
Response time
Tests in water at 0.4 m/s (1.3 ft/s), according to IEC 60751; 10 K temperature step change.
Response time without heat transfer paste 1)
Protection tube
Shape of tip
Insert
1x Pt100
iTHERM
QuickSens,
TF
1x Pt100
iTHERM
StrongSens
, TF
1x Pt100
wire
wound
WW
2x Pt100
wire
wound
WW
1x Pt100
standard
thin-film
TF
2x Pt100
standard
thin-film
TF
t50
t90
t50
t90
t50
t90
t50
t90
t50
t90
t50
t90
6.35 mm (¹⁄₄ in)
Reduced 4.76 mm (³⁄₁₆ in)
x 19.05 mm (0.75 in)
3 mm (0.12 in)
1.6 s
5.9 s
-
-
7.8 s 21.8 s 7.8 s 21.8 s
-
-
-
-
9.53 mm (³⁄₈ in)
Straight
6 mm (0.24 in)
8.5 s
47 s 25.9 s 80.9 s
-
-
-
-
-
-
-
-
Straight
6.35 mm (¹⁄₄ in)
-
-
-
-
23.6 s 67 s 21.8 s 65.2 s 18.4 s 55.8 s 18.4 s 55.8 s
Reduced 4.76 mm (³⁄₁₆ in)
x 19.05 mm (0.75 in)
3 mm (0.12 in)
1.5 s
5.5 s
-
-
8.4 s
23 s
8.4 s
23 s
-
-
-
-
12.7 mm (¹⁄₂ in)
Straight
6 mm (0.24 in)
8.2 s 34.8 s 23.4 s 70.6 s
-
-
-
-
-
-
-
-
Straight
6.35 mm (¹⁄₄ in)
-
-
-
-
20.1 s 55.4 s 21.3 s 61.8 s 17.9 s 51.5 s 17.9 s 51.5 s
Reduced 4.76 mm (³⁄₁₆ in)
x 19.05 mm (0.75 in)
3 mm (0.12 in)
1.8 s
6.2 s
-
-
8.8 s 24.1 s 8.8 s 24.1 s
-
-
-
-
1)
If using a protection tube.
Response time for directly wired insert without transmitter.
Calibration of thermometers
Calibration involves comparing the measured values of a device under test (DUT) with those of a
more precise calibration standard using a defined and reproducible measurement method. The aim is
to determine the deviation of the DUT's measured values from the true value of the measured
variable. Two different methods are used for thermometers:
• Calibration at fixed-point temperatures, e.g. at the freezing point of water at 0 °C,
• Calibration compared against a precise reference thermometer.
The thermometer to be calibrated must display the fixed point temperature or the temperature of
the reference thermometer as accurately as possible. Temperature-controlled calibration baths with
very homogeneous thermal values, or special calibration furnaces into which the DUT and the
reference thermometer, where necessary, can project to a sufficient degree, are typically used for
thermometer calibrations.
The measurement uncertainty can increase due to heat dissipation errors and short immersion
lengths. The existing measurement uncertainty is listed on the individual calibration certificate.
For accredited calibrations according to ISO17025, the measurement uncertainty shouldn’t be twice
as high as the accredited measurement uncertainty. If this is exceeded, only a factory calibration can
be performed.
Evaluation of thermometers
If a calibration with an acceptable uncertainty of measurement and transferable measurement
results is not possible, Endress+Hauser offers customers a thermometer evaluation measurement
service, if technically feasible. This is the case when:
• The process connections/flanges are too big or the immersion length (IL) is too short to allow the
DUT to be immersed sufficiently in the calibration bath or furnace (see the following table), or
• Due to heat conduction along the thermometer tube, the resulting sensor temperature generally
deviates significantly from the actual bath/furnace temperature.
The measured value of the DUT is determined using the maximum possible immersion depth and the
specific measuring conditions and measurement results are documented on an evaluation certificate.
Sensor transmitter matching
The resistance/temperature curve of platinum resistance thermometers is standardized but in
practice it is rarely possible to keep to the values precisely over the entire operating temperature
range. For this reason, platinum resistance sensors are divided into tolerance classes, such as Class
A, AA or B as per IEC 60751. These tolerance classes describe the maximum permissible deviation of
the specific sensor characteristic curve from the standard curve, i.e. the maximum temperature-
dependent characteristic error that is permitted. The conversion of measured sensor resistance
values to temperatures in temperature transmitters or other meter electronics is often susceptible to
considerable errors as the conversion is generally based on the standard characteristic curve.
When using temperature transmitters from Endress+Hauser, this conversion error can be reduced
significantly by sensor-transmitter matching:
• Calibration at three temperatures at least and determination of the actual temperature sensor
characteristic curve,
• Adjustment of the sensor-specific polynomial function using Calendar-van Dusen (CvD)
coefficients,
• Configuration of the temperature transmitter with the sensor-specific CvD coefficients for
resistance/temperature conversion, and
• another calibration of the reconfigured temperature transmitter with connected resistance
thermometer.
Endress+Hauser offers this kind of sensor-transmitter matching as a separate service. Furthermore,
the sensor-specific polynomial coefficients of platinum resistance thermometers are always provided
on every Endress+Hauser calibration certificate where possible, e.g. at least three calibration points,
so that users themselves can also appropriately configure suitable temperature transmitters.
For the device, Endress+Hauser offers standard calibrations at a reference temperature of
–20 to +500 °C (–4 to +932 °F) based on the ITS90 (International Temperature Scale). Calibrations in
other temperature ranges are available from your Endress+Hauser sales center on request.
Calibrations are traceable to national and international standards. The calibration certificate is
referenced to the serial number of the device. Only the insert is calibrated.
Minimum insertion length (IL) for inserts required to perform a correct calibration
The insert length IL is automatically calculated for every thermometer configuration in the
Endress+Hauser Configurator+Temperature software application. The system also automatically
checks whether the selected insert length suffices to perform a factory calibration. For further
information, see the 'Accessories' section → 38
Due to restrictions of the bath geometries, minimum immersion lengths must be maintained at high
temperatures in order to be able to perform a calibration with acceptable measurement uncertainty.
The same applies when a temperature head transmitter is used. Due to the heat dissipation,
minimum immersion lengths must be maintained in order to ensure the functionality of the
transmitter –40 to +85 °C (–40 to +185 °F).
Minimum insertion length (IL):
Calibration temperature
Minimum immersion length (IL)
–196 °C (–320.8 °F)
120 mm (4.72 in) 1)
+80 to +250 °C (+112 to +482 °F)
No minimum immersion length needed
+251 to +550 °C (+483 to +1 022 °F)
300 mm (11.8 in)
+551 to +600 °C (+1 023 to +1 112 °F)
400 mm (15.8 in)
1)
with temperature head transmitter min. 150 mm (5.91 in) needed
IL
IL*
ΔL
A0033648
2
Insertion lengths for sensor calibration
IL
Insertion length for factory calibration or recalibration onsite without the iTHERM QuickNeck extension neck
IL*
Insertion length for recalibration onsite with the iTHERM QuickNeck extension neck
ΔL
Additional length, depending on the calibration unit, if the insert cannot be fully immersed
• To check the actual accuracy rating of the thermometers installed, a cyclic calibration of the
installed sensor needs to be performed frequently. The insert is normally removed for comparison
with a precise reference thermometer in the calibration bath (see graphic, left part).
A reproducible calibration requires the insert to have a minimum insertion length IL. If the insert
is shorter than this minimum length, this reproducibility cannot be guaranteed.
• The iTHERM QuickNeck enables quick, tool-free removal of the insert for calibration purposes. The
entire upper part of the thermometer is released by turning the terminal head. The insert is
removed from the protection tube and directly immersed into the calibration bath (see graphic,
right part). Make sure that the cable is long enough to be able to reach the mobile calibration bath
with the cable connected. If this is not possible for the calibration, it is advisable to use a
connector. → 28
Advantages of iTHERM QuickNeck:
• Considerable time savings when recalibrating the device (up to 20 minutes per measuring point)
• Wiring mistakes avoided when re-installing
• Minimum plant downtime, thereby saving costs
The minimum immersion length is the length of the insert that is fully immersed in the
calibration bath. For a valid recalibration, the value selected for the length IL* must be at least
the value of the previously defined minimum insertion lengths (IL) of the specific types of
insert. For more detailed values, see the previous tables, values without head transmitter.
If the calibration unit used does not allow the insert to be fully immersed as far as the bottom
edge of the top part of the iTHERM QuickNeck, it might be necessary to add an additional
length (ΔL) to IL*. → 13
Formulas for calculating the IL* when recalibrating onsite with iTHERM QuickNeck
Thermowell version
Formula
Protection tube diameter 6.35 mm (¹⁄₄ in)
IL* = U + T + 19.05 mm (0.75 in)
Protection tube diameter 9.53 mm (³⁄₈ in)
Protection tube diameter 12.7 mm (¹⁄₂ in)
Insulation resistance
Insulation resistance ≥ 100 MΩ at ambient temperature.
Insulation resistance between the terminals and the outer jacket is measured with a minimum
voltage of 100 V DC.
Installation
Orientation
No restrictions. However, self-draining in the process must be guaranteed. If there is an opening to
detect leaks at the process connection, this opening must be at the lowest possible point.
Installation instructions
The immersion length of the thermometer can influence the accuracy. If the immersion length is too
small then errors in the measurement are caused by heat conduction via the process connection and
the container wall. If installing into a pipe then the immersion length should ideally be half of the
pipe diameter.
• Installation possibilities: Pipes, tanks or other plant components
• To minimize the error caused by heat conduction, a minimum immersion length is recommended
depending on the type of sensor used and the design of the insert. This immersion depth
corresponds to the minimum insertion length for the calibration. → 13
• ATEX certification: Observe the installation instructions in the Ex documentation! → 39
1, 2 Perpendicular to flow direction, installed at a minimum angle of 3° to ensure self-draining
3
On elbows
4
Inclined installation in pipes with a small nominal diameter
U
Immersion length
In the case of pipes with a small nominal diameter, it is advisable for the tip of the thermometer
to project well into the process so that it extends past the pipe axis. Installation at an angle (4)
could be another solution. When determining the immersion length or installation depth all the
parameters of the thermometer and of the medium to be measured must be taken into account
(e.g. flow velocity, process pressure).
The use of iTHERM QuickSens inserts is recommended for immersion lengths U <
70 mm (2.75 in).
1
2
A0008947
4
Process connections for thermometer installation in pipes with small nominal diameters
1
Varivent® process connection type N for DN40
2
Tee or elbow thermowell (illustrated) for weld-in as per DIN 11865 / ASME BPE
Process connections for thermometer installation in pipes with small nominal diameters
1
Varivent® - process connection D = 50 mm for DN25 pipes
2
Clamp or micro-clamp
1
2
3
Sensor
with Varivent
connection
Shaped
gasket
Companion
connection
O-ring
Companion
connection
Gasket
(O-ring)
Welding boss
Leak detection hole
Vessel wall
A0033600-EN
6
Detailed installation instructions for hygiene-compliant installation
1
Varivent® process connection for VARINLINE® housing
2
Clamp according to ISO 2852
3
Liquiphant-M G1" process connection, horizontal installation
The counterpieces for the process connections and the seals or sealing rings are not included in
the scope of supply for the thermometer. Liquiphant M weld-in adapters with associated seal
kits are available as accessories. → 35
Procedure in case of seal failure indicated by leak detection port:
• Disassembling of the thermometer, validated cleaning procedure of thread and sealing ring
groove
• Replacement of the seal or sealing ring
• CIP after re-assembly
In the case of weld-in connections, exercise the necessary degree of care when performing the
welding work on the process side:
• Suitable welding material
• Flush-welded or with welding radius > 3.2 mm (0.13 in)
• No recesses, folds or gaps
• Honed and polished surface, Ra ≤ 0.76 µm (30 µin)
As a general rule, the thermometers should be installed in such a way that does not impact
their ability to be cleaned (the requirements of the 3-A Sanitary Standard must be observed).
The Varivent® and Liquiphant-M weld-in adapter enable flush-mounted installation.
Ambient temperature range
Terminal head
Temperature in °C (°F)
Without mounted head transmitter
Depends on the terminal head used and the cable gland or fieldbus
connector, see 'Terminal heads' section → 24
With mounted head transmitter
–40 to 85 °C (–40 to 185 °F)
With mounted head transmitter and
display
–20 to 70 °C (–4 to 158 °F)
Extension neck
Temperature in °C (°F)
iTHERM QuickNeck
–50 to +140 °C (–58 to +284 °F)
Storage temperature
For information, see the ambient temperature.
Humidity
Depends on the transmitter used. If Endress+Hauser iTEMP head transmitters are used:
• Condensation permitted as per IEC 60 068-2-33
• Maximum relative humidity: 95% as per IEC 60068-2-30
Climate class
As per EN 60654-1, Class C
Degree of protection
Max. IP69K, depending on the design (terminal head, connector, etc.)
Shock and vibration
resistance
The Endress+Hauser inserts meet the requirements of IEC 60751 which specify shock and vibration
resistance of 3g in the range from 10 to 500 Hz. The vibration resistance at the measuring point
depends on the sensor type and design, see the following table:
Version
Vibration resistance for the sensor tip
Pt100 (WW or TF)
30 m/s² (3g) 1)
iTHERM StrongSens Pt100 (TF)
iTHERM QuickSens Pt100 (TF), version: ⌀6 mm (0.24 in)
> 600 m/s² (60g)
1)
Vibration resistance also applies for the quick-fastening iTHERM QuickNeck.
Electromagnetic
compatibility (EMC)
Depends on the head transmitter used. For details see the Technical Information. → 39
Process
Process temperature range
Depends on the type of sensor used, maximum –200 to +600 °C (–328 to +1 112 °F).
Thermal shock
Thermal shock resistance in CIP/SIP process with a temperature increase from
+5 to +130 °C (+41 to +266 °F) within 2 seconds.
Process pressure range
The maximum possible process pressure depends on various influencing factors, such as the design,
process connection and process temperature. For information on the maximum possible process
pressures for the individual process connections, see the 'Process connection' section. → 29
It is possible to check the mechanical loading capacity as a function of the installation and
process conditions online in the Thermowell (TW) Sizing Module for protection tubes in the
Endress+Hauser Applicator software. See 'Accessories' section. → 38
Example of the permitted flow velocity depending on the immersion length and process
medium
The highest flow velocity tolerated by the thermometer diminishes with increasing insert immersion
length exposed to the stream of the fluid. In addition it is dependent on the diameter of the
thermometer tip, on the kind of measuring medium, on the process temperature and on the process
pressure. The following figures exemplify the maximum permitted flow velocities in water and
superheated steam at a process pressure of 40 bar (580 PSI).
L (mm)
A
0
10
20
30
40
50
60
70
80
90
100
200
300
400
500
v (m/s)
B
0
4
8
12
16
20
L (in)
0
30
65
100
130
160
200
230
260
290
v (ft/s)
A0008967
7
Permitted flow velocities, protection tube diameter 9.53 mm (3/8 in)
A
Medium water at T = 50 °C (122 °F)
B
Medium superheated steam at T = 400 °C (752 °F)
L
Immersion length exposed to flow
v
Flow velocity
Medium - state of
aggregation
Gaseous or liquid (also with high viscosity, e.g. yogurt).
All dimensions in mm (in). The design of the thermometer depends on the protection tube version
used:
• Thermometer without protection tube - for installation in an existing protection tube
• Diameter 6.35 mm (¹⁄₄ in)
• Diameter 9.53 mm (³⁄₈ in)
• Diameter 12.7 mm (¹⁄₂ in)
Various dimensions, such as the immersion length U for example, are variable values and are
therefore indicated as items in the following dimensional drawings.
Variable dimensions:
Item
Description
E
Extension neck length, variable depending on the configuration or predefined for the version with
iTHERM QuickNeck
IL
Insert length
L
Protection tube length (U+T)
B
Protection tube base thickness: predefined, depends on protection tube version (see also the
individual table data)
T
Length of protection tube shaft: variable or predefined, depends on protection tube version (see
also the individual table data)
U
Immersion length: variable, depending on the configuration
ØID
Insert diameter 6 mm (0.24 in), 3 mm (0.12 in) or 6.35 mm (¹⁄₄ in)
Without protection tube
For installation in an existing protection tube (all dimensions in inch)
1
3
4
5
2
IL
ILa
IL
IL
ILa
ILa
ILa
ILa
IL
IL
E=1
E=2.5
E=1.5
E=3.5
E=
0.75
0.32
0.32
0.32
A0034489
1
Thermometer with extension hex nipple, for connection thread ½" NPT
2
Thermometer with extension nipple union nipple (NUN), for connection thread ½" NPT
3
Thermometer with quick-fastening iTHERM QuickNeck, upper part
4
Thermometer with quick-fastening iTHERM QuickNeck complete, for connection thread ½" NPT
5
Thermometer with 1¼" x 18 UNEF thread
IL
Insert length
ILa
Insert length (without nipple)
E
Length of the extension neck available at point of installation (provided one is available)
Pay attention to the following equations when calculating the insert length ILa for immersion into a
protection tube TT412 already available:
Version 1, 2, 4 and 5
ILa = U + T +38.1 mm (1.5 in) 1)
Version 3
ILa = U + T +19.05 mm (0.75 in) 1)
1)
ILa = Insert length (without nipple); U = Protection tube immersion length; T = Protection tube shaft
length
With protection tube (1/4", 3/8", 1/2")
1
2
3
4.76
(3/16)
4.76
(3/16)
4.76
(3/16)
B=3.2
(0.125)
B=3.2
(0.125)
B=3.2
(0.125)
8.1
(0.32)
8.1
(0.32)
8.1
(0.32)
E=1.5"
E=2.5"
E=1"
8.1
(0.32)
8.1
(0.32)
4
5
6
7
IL
IL
IL
U
U
U
1.75"+T
1.75"+T
1.75"+T
U
U
U
U
1.75"+T
1.75"+T
1.75"+T
1.75"+T
E
E
E
E
A0034495
8
Protection tube with neck connection NPT ½" and various process connection versions:
1
Thermometer with extension hex nipple and Tri-clamp process connection
2
Thermometer with extension nipple union nipple (NUN) and Tri-clamp process connection
3
Thermometer with quick-fastening iTHERM QuickNeck and Tri-clamp process connection
4
Thermometer with extension hex nipple and Tri-clamp process connection
5
Thermometer with extension hex nipple and Varivent® process connection
6
Thermometer with extension hex nipple and Liquiphant adapter
7
Thermometer with cylindrical weld-in adapter
IL
Insert length
U
Protection tube immersion length
Length of the extension neck available at point of installation (provided one is available)
T
Protection tube shaft length
B
Base thickness
For diameter ½" not available: Tri-clamp 3/4"
Item
Version
Length
Length of protection
tube shaft T 1)
Version 1, 2, 4: Tri-clamp with NPT
Version 3: Tri-clamp with QuickNeck
Version 5: Varivent® with NPT
Version 5: Varivent® with QuickNeck
Version 6: Thread ISO 228 for Liquiphant, with NPT
Version 6: Thread ISO 228 for Liquiphant, with QuickNeck
Version 7: Cylindrical weld-in with NPT
Version 7: Cylindrical weld-in with QuickNeck
0-6"
1-6"
1-6"
1.5-6"
2-6"
2-6"
2-6"
2-6"
Immersion length U
Independent of the version
Variable, depending on
the configuration
Base thickness B
6.35 mm (¹⁄₄ in) Protection tube:
Reduced tip 4.76 mm (³⁄₁₆ in)
3.2 mm (0.125 in)
9.53 mm (³⁄₈ in) Protection tube:
Reduced tip 4.76 mm (³⁄₁₆ in)
Straight tip
3.2 mm (0.125 in)
3 mm (0.12 in)
12.7 mm (¹⁄₂ in) Protection tube:
Reduced tip 4.76 mm (³⁄₁₆ in)
Straight tip
3 mm (0.12 in)
6.3 mm (0.25 in)
Extension neck length E
Version 1:
Thermometer with extension hex nipple and Tri-clamp
process connection
E = 25.4 mm (1 in)
Version 2:
Thermometer with extension nipple union nipple (NUN)
and Tri-clamp process connection
E = 63.5 mm (2.5 in)
Version 3:
Thermometer with quick-fastening iTHERM QuickNeck
and Tri-clamp process connection
E = 38.1 mm (1.5 in)
1)
Depends on the process connection
Insert
Depending on the application, iTHERM TS212 inserts with different RTD sensors are available for
the thermometer:
Sensor
Standard thin-film
iTHERM
StrongSens
iTHERM QuickSens 1)
Wire wound
Sensor
design;
connection
method
1x Pt100, 3- or 4-
wire, PTFE
insulated
2x Pt100, 2x3-
wire, PTFE
insulated
1x Pt100, 3- or
4-wire, mineral
insulated
1x Pt100, 3- or
4-wire, mineral
insulated
1x Pt100, 3- or 4-
wire, PTFE
insulated
1x Pt100, 3-
or 4-wire,
mineral
insulated
2x Pt100,
2x3-wire,
mineral
insulated
Vibration
resistance of
the insert tip
Up to 3g
Enhanced
vibration
resistance > 60g
> 60g
3g
Up to 3g
Measuring
range;
accuracy
class
–50 to +200 °C (–58 to +392 °F), Class A
or B
–50 to +500 °C
(–58 to +932 °F),
Class A or AA
–50 to +200 °C (–58 to +392 °F), Class
A or AA
–200 to +600 °C
(–328 to +1 112 °F), Class A
or AA
Diameter
6.35 mm (¹⁄₄ in)
6 mm (0.24 in)
6 mm (0.24 in)
3 mm (0.12 in)
6.35 mm (¹⁄₄ in);
3 mm (0.12 in)
1)
Recommended for immersion lengths U < 70 mm (2.75 in)
The iTHERM TS212 insert is available as a spare part. The insert length (IL) depends on the
immersion length of the protection tube (U), the length of the extension neck (E) and the length of
the protection tube shaft (T). The insertion length (IL) must be taken into consideration when
replacing the unit.
This can be calculated by means of the following formula: IL = U + T + E + 38.1 mm (1.5 in)
For more information on the deployed insert iTHERM TS212 with enhanced vibration
resistance and fast-response sensor, see the Technical Information.
Spare parts currently available for your product can be found online at:
http://www.products.endress.com/spareparts_consumables, product root: TM412. Always
quote the serial number of the device when ordering spare parts! The insertion length IL is
automatically calculated using the serial number.
Weight
Depending on the configuration
Material
Extension neck and thermowell, insert, process connection.
The temperatures for continuous operation specified in the following table are only intended as
reference values for use of the various materials in air and without any significant compressive load.
The maximum operating temperatures can be reduced considerably in cases where abnormal
conditions such as high mechanical load occur or in aggressive media.
Designation
Recommended max.
temperature for
continuous use in air
Properties
AISI 316L
650 °C (1 202 °F) 1)
• Austenitic, stainless steel
• High corrosion resistance in general
• Particularly high corrosion resistance in chlorine-
based and acidic, non-oxidizing atmospheres
through the addition of molybdenum (e.g.
phosphoric and sulfuric acids, acetic and tartaric
acids with a low concentration)
• Increased resistance to intergranular corrosion and
pitting
• The wetted part from a 316L or 1.4435+316L
thermowell withstand a passivation process with a
3% sulphuric acid
1)
Can be used to a limited extent up to 800 °C (1472 °F) for low compressive loads and in non-corrosive
media. Contact your Endress+Hauser sales team for further information.
Surface roughness
Values for wetted surfaces: 1)
Standard surface
Ra ≤ 0.76 µm (30 µin)
Finely honed surface 2)
Ra ≤ 0.38 µm (15 µin)
Finely honed surface and electropolished
Ra ≤ 0.38 µm (15 µin)+ electropolished
1)
Exception: Inside welding seams of the tee and elbow thermowells
2)
Not compliant with ASME BPE
Terminal heads
All terminal heads have an internal shape and size in accordance with DIN EN 50446, flat face and a
thermometer connection with a ½" NPT or M24x1.5 thread. All dimensions in mm (in). The sample
cable glands in the diagrams correspond to M20x1.5 connections with non-Ex polyamide cable
glands. Specifications without head transmitter installed. For ambient temperatures with head
transmitter installed, see the 'Environment' section. → 18
As a special feature, Endress+Hauser offers terminal heads with optimized terminal accessibility for
easy installation and maintenance.
(1.1) 78 (3.1)
15.5 (0.6)
A0009820
• Protection class:
– IP66/68 (NEMA Type 4x encl.)
– For ATEX: IP66/67
• Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
• Material: aluminum, polyester powder coated
Seals: silicone
• Threaded cable entry: G ½", ½" NPT and M20x1.5;
• Protection armature connection: ½" NPT, M24x1.5
• Head color: blue, RAL 5012
• Cap color: gray, RAL 7035
• Weight: 330 g (11.64 oz)
• Ground terminal, internal and external
• Available in conjunction with 3-A marked sensors
TA30A
Specification
107.5 (4.23)
91.6 (3.61)
28
(1.1)78 (3.1)
15.5 (0.6)
A0009821
• Protection class:
– IP66/68 (NEMA Type 4x encl.)
– For ATEX: IP66/67
• Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
• Material: aluminum, polyester powder coated
Seals: silicone
• Threaded cable entry: G ½", ½" NPT and M20x1.5
• Protection armature connection: ½" NPT, M24x1.5
• Head color: blue, RAL 5012
Cap color: gray, RAL 7035
• Weight: 420 g (14.81 oz)
• With TID10 display
• Ground terminal, internal and external
• Available in conjunction with 3-A marked sensors
TA30D
Specification
107.5 (4.23)
110 (4.3)
28
(1.1) 78 (3.1)
15.5 (0.6)
A0009822
• Protection class:
– IP66/68 (NEMA Type 4x encl.)
– For ATEX: IP66/67
• Temperature: –50 to +150 °C (–58 to +302 °F) without cable
gland
• Material: aluminum, polyester powder coated
Seals: silicone
• Threaded cable entry: G ½", ½" NPT and M20x1.5
• Protection armature connection: ½" NPT, M24x1.5
• Two head transmitters can be mounted. In the standard
version, one transmitter is mounted in the terminal head
cover and an additional terminal block is installed directly
on the insert.
• Head color: blue, RAL 5012
• Cap color: gray, RAL 7035
• Weight: 390 g (13.75 oz)
• Ground terminal, internal and external
• Available in conjunction with 3-A marked sensors
• Protection class: IP65
• Max. temperature: –40 to +120 °C (–40 to +248 °F)
• Material: polyamide (PA), antistatic
Seals: silicone
• Threaded cable entry: ½" NPT and M20x1.5
• Protection armature connection: M24x1.5 or ½" NPT
• Two head transmitters can be mounted. In the standard
version, one transmitter is mounted in the terminal head
cover and an additional terminal block is installed directly on
the insert.
• Head and cap color: black
• Weight: 135 g (4.8 oz)
• Types of protection for use in hazardous locations: Intrinsic
Safety (G Ex ia)
• Ground terminal: only internal via auxiliary clamp
TA30R (optionally with display window in
cover)
Specification
64 (2.52)
96 (3.8)
64 (2.52)
71 (2.8)
25 (1)
96 (3.8)*
A0017145
* Dimensions of version with display window
in cover
• Degree of protection - standard version: IP69K (NEMA Type
4x encl.)
Degree of protection - version with display window: IP66/68
(NEMA Type 4x encl.)
• Temperature: –50 to +130 °C (–58 to +266 °F) without cable
gland
• Material: stainless steel 316L, abrasive-blasted or polished
Seals: EPDM
Display window: polycarbonate (PC)
• Cable entry thread ½" NPT and M20x1.5
• Weight
– Standard version: 360 g (12.7 oz)
– Version with display window: 460 g (16.23 oz)
• Display window in cover optionally for head transmitter with
display TID10
• Protection armature connection: M24x1.5 or ½" NPT
• Ground terminal: internal in standard version; external
terminal optionally available
• Available in conjunction with 3-A marked sensors
TA30R (high version for two transmitters)
Specification
116 (4.57)
64 (2.52)
64 (2.52)
96 (3.8)
25
(1)
A0034644
• Degree of protection: IP69K (NEMA Type 4x encl.)
• Temperature: –50 to +130 °C (–58 to +266 °F) without cable
gland
• Material: stainless steel 316L, abrasive-blasted or polished
Seals: EPDM
• Cable entry thread ½" NPT and M20x1.5
• Weight: 460 g (16.23 oz)
• For two head transmitter
• Protection armature connection: M24x1.5 or ½" NPT
• Ground terminal: internal in standard version; external
terminal optionally available
• Available in conjunction with 3-A marked sensors
TA30S
Specification
122 (4.8)
28
(1.1)
78 (3.1)
50 (1.97)
87 (3.43)
81.7 (3.22)
A0017146
• Degree of protection: IP65 (NEMA Type 4x encl.)
• Temperature: –40 to +85 °C (–40 to +185 °F) without cable
gland
• Material: polypropylene (PP), FDA-compliant, seals: O-ring
EPDM
• Cable entry thread: ¾" NPT (with adapter for ½" NPT),
M20x1.5
• Protective assembly connection: ½" NPT
• Color: white
• Weight: approx. 100 g (3.5 oz)
• Ground terminal: only internal via auxiliary terminal
• Available in conjunction with 3-A marked sensors
Field housing T17 for TMT162
Specification
°C
10
0
20
30
40
50
114 (4.49)
114 (4.49)
94 (3.7)
A0034506
• Protection class:
– IP66
– IP68
• Temperature:
– Without display: –40 to +85 °C (–40 to +185 °F)
– With display: –40 to +80 °C (–40 to +80 °F)
• Material: Stainless steel 1.4404/316L for hygienic
applications
• Threaded cable entry: 1x or 2x M20x1.5, G½" or ½" NPT
• Protection armature connection: ½" NPT
• Display rotatable in 90° increments
• Brilliant blue backlit display with ease of visibility in
bright sunshine or pitch darkness
• Separate electronics compartment and connection
compartment
• Gold plated terminals to avoid corrosion and additional
measurement errors
• Weight: 1.25 kg (2.76 lb)
• Available in conjunction with 3-A marked sensors
Cable glands and fieldbus connectors
Type
Suitable for cable entry
Degree of protection
Temperature range
Cable gland, polyamide
½" NPT, ¾" NPT,
M20x1.5 (optionally 2x
cable entry)
IP68
–40 to +100 °C (–40 to +212 °F)
½" NPT, M20x1.5
IP69K
–20 to +95 °C (–4 to +203 °F)
Cable gland for dust ignition-proof area, polyamide
½" NPT, M20x1.5
(optionally 2x cable
entry)
IP68
–20 to +95 °C (–4 to +203 °F)
Fieldbus connector (M12x1 PA, 7/8" PA, FF)
½" NPT, M20x1.5
IP67, NEMA Type 6
–40 to +105 °C (–40 to +221 °F)
Fieldbus connector (M12, 8-pin)
M20x1.5
IP67
–30 to +90 °C (–22 to +194 °F)
Extension neck
Standard version of extension neck, or optionally with the quick-fastening iTHERM QuickNeck.
• Tool-free removal of the insert:
– Saves time/costs on frequently calibrated measuring points
– Wiring mistakes avoided
• IP69K protection class
Dimensions of extension neck, different versions, each with NPT ½" thread to the terminal head
1
Extension hex nipple
2
Extension nipple union nipple (NUN)
3
Quick-fastening iTHERM QuickNeck
4
Quick-fastening iTHERM QuickNeck - upper part, for installation in an existing thermowell with iTHERM
QuickNeck
5
Quick-fastening iTHERM QuickNeck complete, for mounting in existing thermowell with NPT ½" connection
6
1¼" x 18 UNEF thread, for mounting in existing protection tube
IL
Insert length
ILa
Insert length (without nipple)
E
Length of the extension neck available at point of installation (provided one is available)
Process connections
All dimensions in mm (in).
For welding in
Type
Version
Dimensions
Technical properties
Weld-in adapter
U
�d
h
44.45 + T
(1.75")
A0033743
Cylindrical ½" NPS
d = ½" NPS, h = 38.1 mm (1.5 in), U =
immersion length from lower edge of thread,
T = min. 50.8 mm (2 in)
• Pmax. depends on the weld-in
process
• With 3-A symbol and
EHEDG certification
• ASME BPE compliance
Cylindrical ¾" NPS
d = ¾" NPS, h = 38.1 mm (1.5 in), U =
immersion length from lower edge of thread,
T = min. 50.8 mm (2 in)
Cylindrical 1" NPS
d = 1" NPS, h = 38.1 mm (1.5 in), U =
immersion length from lower edge of thread,
T = min. 50.8 mm (2 in)
Releasable process connection
Type
Version
Dimensions
Technical properties
d: 1)
D
a
Clamp according to ASME BPE or ISO 2852
�D
U
�d
�a
�d
U
�D
R1.2
R0.8
0.4
1.6
A
A: Tri-clamp
A0033985
A Different seal geometries for Tri-clamp
Tri-clamp ¾"
(DN18) 2)
25 mm (0.98 in)
-
• Pmax. = 16 bar (232 psi),
depends on clamp ring and
suitable seal
• With 3-A symbol
Clamp ISO 2852 ½"
(DN12 - 21.3)
34 mm (1.34 in)
16 to 25.3 mm
(0.63 to 0.99 in)
• Pmax. = 16 bar (232 psi),
depends on clamp ring and
suitable seal
• With 3-A symbol and
EHEDG certification
(combined with Hyjoin
PEEK/stainless steel seal or
Dupont de Nemours Kalrez/
stainless steel seal)
• ASME BPE compliance 3)
Tri-clamp 1" - 1½"
(DN25 - 38)
50.5 mm (1.99 in)
29 to 42.4 mm
(1.14 to 1.67 in)
Tri-clamp 2" (DN40 -
51)
64 mm (2.52 in)
44.8 to 55.8 mm
(1.76 to 2.2 in)
Tri-clamp 2½"
(DN63.5)
77.5 mm (3.05 in)
68.9 to 75.8 mm
(2.71 to 2.98 in)
Tri-clamp 3"
(DN70-76.5)
91 mm (3.58 in)
> 75.8 mm (2.98 in)
1)
Pipes in accordance with ISO 2037 and BS 4825 Part 1
2)
Tri-clamp ¾" only possible with protection tube diameter 6.35 mm (¹⁄₄ in) or 9.53 mm (³⁄₈ in)
3)
Not for DN12-21.3
Type
Version G
Dimensions
Technical properties
L1 thread
length
A
1 (SW/AF)
Thread according to ISO 228 (for
Liquiphant weld-in adapter)
U
L1
1
G
A
A0009572
G¾" for
FTL31/33/20
adapter
16 mm
(0.63 in)
25.5 mm (1 in)
32
• Pmax. = 25 bar (362 psi) at
max. 150 °C (302 °F)
• Pmax. = 40 bar (580 psi) at
max. 100 °C (212 °F)
• With 3-A symbol and
EHEDG certification
• ASME BPE compliance
• Minimum extension neck
lengths: ≥ 76.2 mm (3 in)
G¾" for FTL50
adapter
G1" for FTL50
adapter
18.6 mm
(0.73 in)
29.5 mm
(1.16 in)
41
Type
Version
Dimensions
Technical properties
D
A
B
h
Pmax.
Varivent®
Type B
31 mm
(1.22 in)
105 mm
(4.13 in)
-
22 mm
(0.87 in)
10 bar
(145 psi)
• With 3-A symbol and
EHEDG certification
• ASME BPE compliance
Type F
50 mm
(1.97 in)
145 mm
(5.71 in)
135 mm
(5.31 in)
24 mm
(0.95 in)
The VARINLINE® housing connection flange is suitable for weld-in into the conical or torispherical head in tanks or containers with a small
diameter (≤ 1.6 m (5.25 ft)) and up to a wall thickness of 8 mm (0.31 in).
Type
Technical properties
Varivent® for VARINLINE® housing for installation in pipes
U
�D
�i
�a
A0009564
• With 3-A symbol and
EHEDG certification
• ASME BPE compliance
Version
Dimensions
Pmax.
D
i
a
Type N, according to DIN
11866, series C
68 mm (2.67 in)
OD 1½": 34.9 mm (1.37 in)
OD 1½": 38.1 mm (1.5 in)
OD 1½" to OD 2½":
16 bar (232 psi)
OD 2": 47.2 mm (1.86 in)
OD 2": 50.8 mm (2 in)
OD 2½": 60.2 mm (2.37 in)
OD 2½": 63.5 mm (2.5 in)
Type N, according to DIN
11866, series C
68 mm (2.67 in)
OD 3": 73 mm (2.87 in)
OD 3": 76.2 mm (3 in)
OD 3" to OD 4": 10 bar (145 psi)
OD 4": 97.6 mm (3.84 in)
OD 4": 101.6 mm (4 in)
Type F, according to DIN
11866, series C
50 mm (1.97 in)
OD 1": 22.2 mm (0.87 in)
OD 1": 25.4 mm (1 in)
16 bar (232 psi)
Due to the small immersion length U, the use of iTHERM QuickSens inserts is recommended.
Tip shape
The thermal response time, the reduction of the flow cross-section and the mechanical load that
occurs in the process are the criteria that matter when selecting the shape of the tip. Advantages of
using reduced thermometer tips:
• A smaller tip shape has less impact on the flow characteristics of the pipe carrying the medium.
• The flow characteristics are optimized, thereby increasing the stability of the protection tube.
• Endress+Hauser offers users a range of protection tube tips to meet every requirement:
– Straight tip
– Reduced tip with 4.76 mm (³⁄₁₆ in): walls of lower thickness significantly reduce the response
times of the overall measuring point
Protection tube tips available (reduced or straight)
Item No.
Protection tube (D1)
Tip (D2)
Insert (ID)
1
6.35 mm (¹⁄₄ in)
Reduced tip with 4.76 mm (³⁄₁₆ in)
3 mm (0.12 in)
2
9.53 mm (³⁄₈ in)
• Reduced tip with 4.76 mm (³⁄₁₆ in)
• Straight tip
• 3 mm (0.12 in)
• 6.35 mm (¹⁄₄ in) or
6 mm (0.24 in)
3
12.7 mm (¹⁄₂ in)
• Reduced tip with 4.76 mm (³⁄₁₆ in)
• Straight tip
• 3 mm (0.12 in)
• 6.35 mm (¹⁄₄ in) or
6 mm (0.24 in)
It is possible to check the mechanical loading capacity as a function of the installation and
process conditions online in the TW Sizing Module for protection tubes in the Endress+Hauser
Applicator software. See 'Accessories' section. → 35
Certificates and approvals
CE mark
The product meets the requirements of the harmonized European standards. As such, it complies
with the legal specifications of the EC directives. The manufacturer confirms successful testing of the
product by affixing to it the CE-mark.
Hygiene standard
• EHEDG certification, type EL - CLASS I. Permitted process connections in accordance with EHEDG,
see 'Process connections' section → 29
• 3-A authorization no. 1144, 3-A sanitary standard 74-06. Permitted process connections in
accordance with 3-A, see 'Process connections' section → 29
• ASME BPE 2009, certificate of conformity can be ordered for indicated options
• FDA-compliant
• All product contact surfaces are produced without animal fats (TSE Certificate of Suitability)
Ex approval
Information about currently available Ex versions (ATEX, FM, CSA, etc.) can be supplied by your E+H
Sales Center on request. All explosion protection data are given in separate documentation which is
available upon request.
Other standards and
guidelines
• EN 60079: ATEX certification for hazardous areas
• IEC 60529: Degrees of protection provided by enclosures (IP code)
• IEC 61010-1: Safety requirements for electrical equipment for measurement, control and
laboratory use
• IEC 60751: Industrial platinum resistance thermometers
• ASTM E 1137/E1137M-2008: Standard Specification for Industrial Platinum Resistance
Thermometers
• EN 50281-1-1: Electrical apparatus protected by enclosures
• DIN 43772: Protection tubes
• DIN EN 50446: Terminal heads
• IEC 61326-1: Electromagnetic compatibility (electrical equipment for measurement, control and
laboratory use - EMC requirements)
Parts in contact with the
medium
Parts of the thermometer in contact with the medium comply with the following European
regulations:
• (EC) No. 1935/2004, Article 3, paragraph 1, Articles 5 and 17 on materials and articles intended
to come into contact with food.
• (EC) No. 2023/2006 on good manufacturing practice for materials and articles intended to come
into contact with food.
• (EC) No. 10/2011 on plastic materials and articles intended to come into contact with food.
• Canadian Registration Number (CRN):
The CRN approval is only available for certain options of protection tubes. These will be marked
and shown during the configuration of this device.
Surface purity
Free from oil and grease, optional
Material resistance
Material resistance - including terminal heads - to the following cleaning agents/disinfectants from
the company Ecolab: P3-topax 66, P3-topactive 200, P3-topactive 500 and P3-topactive OKTO as
well as demineralized water.
Material certification
The material certificate 3.1 (according to standard EN 10204) can be requested separately.
Calibration
The "Factory calibration" is carried out according to an internal procedure in a laboratory of Endress
+Hauser accredited by the European Accreditation Organization (EA) to ISO/IEC 17025. A
calibration which is performed according to EA guidelines (SIT/Accredia) or (DKD/DAkkS) may be
requested separately. The calibration is performed on the replaceable insert of the thermometer. In
the case of thermometers without a replaceable insert, the entire thermometer - from the process
connection to the tip of the thermometer - is calibrated.
Protection tube testing and
load capacity calculation
• Protection tube pressure tests are carried out in accordance with the specifications in DIN 43772.
With regard to protection tubes with reduced tip that do not comply with this standard, these are
tested using the pressure of corresponding straight protection tubes. Tests according to other
specifications can be carried out on request. The liquid penetration test verifies that there are no
cracks in the welded seams of the protection tube.
• PMI test, dye penetration test, TW welding, internal hydrostatic pressure, etc. each with inspection
certificate
• Load capacity calculation for the protection tube as per DIN43772
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Device-specific accessories
Weld-in adapter for FTL31/33/20
ø26 (1.0)
25 (1.0)
ø29 (1.1)
30 (1.2)
G
ISO 228
¾"
ø32 (1.3)
A0008265
G¾", d=29 mm, without flange
Material: 316L
Roughness in μm (μin): 1.5 (59.1)
Order number: 52028295 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52021717 1), FDA-
compliant
1)
A seal is included in the delivery.
Weld-in adapter for FTL31/33/20
ø50 (2.0)
26 (1.0)
21 (0.8)
ø32 (1.3)
G ¾
ISO 228
"
A0008810
G¾", d=50 mm, with flange
Material: 316L
Roughness in μm (μin): 0.8 (31.5)
Order number: 52018765 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52021717 1), FDA-
compliant
With EHEDG certification and 3-A symbol
1)
A seal is included in the delivery.
Weld-in adapter for FTL50
26 (1.0)
ø55 (2.2)
21 (0.8)
ø32 (1.3)
G ¾
ISO 228
"
A0008274
G¾", d=55 mm, with flange
Material: 316L
Roughness in μm (μin): 0.8 (31.5)
Order number: 52001052 (without inspection certificate EN10204-3.1
material)
Order number: 52011897 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52014473 1), FDA-
compliant
With EHEDG certification and 3-A symbol
1)
A seal is included in the delivery.
Weld-in adapter for FTL50
ø50 (2.0)
ø53 (2.1)
47.8 (1.9)
24.6 (1.0)
30°
21 (0.8)
ø41 (1.6)
G 1"
ISO 228
A0011927
G1", d=53 mm, without flange
Material: 316L
Roughness in μm (μin): 0.8 (31.5)
Order number: 71093129 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52014472 1), FDA-
compliant
1)
A seal is included in the delivery.
Weld-in adapter for FTL50
ø41 (1.6)
G 1"
ISO 228
ø60 (2.4)
29.6 (1.2)
24.6 (1.0)
A0008267
G1", d=60 mm, with flange
Material: 316L
Roughness in μm (μin): 0.8 (31.5)
Order number: 52001051 (without inspection certificate EN10204-3.1
material)
Order number: 52011896 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52014472 1), FDA-
compliant
With EHEDG certification and 3-A symbol
1)
A seal is included in the delivery.
Weld-in adapter for FTL50
26 (1.02)
5 (0.2)
ø65 (2.56)
A0008272
G1", can be aligned
Material: 316L
Roughness in μm (μin): 0.8 (31.5)
Order number: 52001221 (without inspection certificate EN10204-3.1
material)
Order number: 52011898 (with inspection certificate EN10204-3.1
material)
Order number seal (5-pc. set): silicone O-ring 52014424 1), FDA-
compliant
1)
A seal is included in the delivery.
Flexible handle cap to cover the
QuickNeck bottom part
�D
20 (0.79)
35 (1.38)
A0027201
Diameter ⌀D: 24 to 26 mm (0.94 to 1.02 in)
Material: Thermoplastic polyolefin - elastomer (TPE), free from
plasticizers
Maximum temperature: +150 °C (+302 °F)
Order number: 71275424
Maximum process pressure for the weld-in adapters:
• 25 bar (362 PSI) at maximum 150 °C (302 °F)
• 40 bar (580 PSI) at maximum 100 °C (212 °F)
For more information on the weld-in adapters, see Technical Information (TI00426F).
Communication-specific
accessories
Configuration kit TXU10
Configuration kit for PC-programmable transmitter with setup software and
interface cable for PC with USB port
Order code: TXU10-xx
Commubox FXA195
HART
For intrinsically safe HART communication with FieldCare via the USB interface.
For details, see Technical Information TI00404F
Commubox FXA291
Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
Common Data Interface) and the USB port of a computer or laptop.
For details, see Technical Information TI00405C
HART Loop Converter
HMX50
Is used to evaluate and convert dynamic HART process variables to analog current
signals or limit values.
For details, see Technical Information TI00429F and Operating Instructions
BA00371F
Wireless HART adapter
SWA70
Is used for the wireless connection of field devices.
The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
For details, see Operating Instructions BA00061S
Fieldgate FXA42
Gateway for remote monitoring of connected 4 to 20 mA analog as well as digital
field devices.
For details, see Technical Information TI01297S and Brief Operating
Instructions KA01246S
Field Xpert SFX370
Field Xpert is an industrial PDA with integrated touchscreen for commissioning
and maintaining field devices in explosion hazardous and safe areas. It allows
efficient configuration of FOUNDATION fieldbus, HART 5/6/7 and WirelessHART
devices via Bluetooth and/or Wifi interfaces.
For details, see Technical Information TI01114S and Operating Instructions
BA01202S
Field Xpert SMT70
The Field Xpert SMT70 is an out-of-the box tablet PC for universal device
configuration that has pre-installed devices libraries. It supports devices with
protocols such as HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, Modbus, and
Endress+Hauser service protocols.
For details, see Technical Information TI01342S
Software for selecting and sizing Endress+Hauser measuring devices:
• Calculation of all the necessary data for identifying the optimum measuring
device: e.g. pressure loss, accuracy or process connections.
• Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters over the entire life cycle of a project.
Applicator is available:
• Via the Internet: https://wapps.endress.com/applicator
• On CD-ROM for local PC installation.
Konfigurator+Temperatur
Software for selecting and configuring the product depending on the measuring
task, supported by graphics. Includes a comprehensive knowledge database and
calculation tools:
• For temperature competence
• Quick and easy design and sizing of temperature measuring points
• Ideal measuring point design and sizing to suit the processes and needs of a wide
range of industries
The Konfigurator is available:
On request from your Endress+Hauser sales office on a CD-ROM for local PC
installation.
W@M
Life cycle management for your plant
W@M supports you with a wide range of software applications over the entire
process: from planning and procurement, to the installation, commissioning and
operation of the measuring devices. All the relevant device information, such as
the device status, spare parts and device-specific documentation, is available for
every device over the entire life cycle.
The application already contains the data of your Endress+Hauser device. Endress
+Hauser also takes care of maintaining and updating the data records.
W@M is available:
• Via the Internet: www.endress.com/lifecyclemanagement
• On CD-ROM for local PC installation.
DeviceCare SFE100
The fastest way to configure Endress+Hauser field devices is by using the dedicated
tool DeviceCare. User-centered design allows for transparent and intuitive device
connection and configuration.
For details, see Technical Information TI01134S and Operating Instructions
KA01303S and BA00027S
FieldCare SFE500
FieldCare is Endress+Hauser's universal tool for configuring field devices that
provides a range of functionality from device parameterization to engineered
Condition Monitoring solutions.
For details, see Technical Information TI00028S and Operating Instructions
KA01303S, BA00027S and BA00065S
System components
Accessories
Description
Display RIA15
It is integrated in the 4 to 20 mA or HART® loop and transmits the measuring
signal or HART® process variables in digital form. The process display unit does not
require an external power supply. It is powered directly from the current loop.
For details, see the "Technical Information" document TI01043K
Memograph M, RSG45
Advanced Data Manager with tamper-proof data storage and access (FDA 21 CFR
11) HART® gateway functionality; Up to 40 HART® devices connected at a time
Communication capabilities: Modbus, Profibus DP, PROFINET, EtherNet/IP.
For details, see the "Technical Information" document TI01180R
Resistance thermometer for hygienic and aseptic applications iTHERM TM402: TI01349T
Modular resistance thermometer for hygienic and aseptic applications iTHERM TM412:
TI01348T
Insert for thermometer installation iTHERM TS212: TI01336T
Thermometer protection tube for hygienic and aseptic applications iTHERM TT412: TI01350T
Process Indicator RIA15: TI01043K
Advanced Data Manager Memograph M RSG45: TI01180R
Field Data Manager Software MS20, MS21: TI01022R
Hazardous area supplementary documentation:
iTHERM TM412 Intrinsically safe Ex ia IIC: XA01024T
iTHERM TM412 Dust-explosion protection Ex ta/tb: XA01023T
iTEMP temperature head transmitter:
TMT80, PC-programmable, single-channel, RTD and TC: TI00153R
TMT180, PC-programmable, single-channel, Pt100: TI00088R
TMT181, PC-programmable, single-channel, RTD, TC, Ω, mV: TI00070R
HART® TMT182, single-channel, RTD, TC, Ω, mV: TI00078R
HART® TMT82, two-channel, RTD, TC, Ω, mV: TI01010T
PROFIBUS® PA TMT84, two-channel, RTD, TC, Ω, mV: TI00138R
FOUNDATION FieldbusTM TMT85, two-channel, RTD, TC, Ω, mV: TI00134R
iTEMP temperature field transmitter:
TMT162, dual input temperature field transmitter with HART®, FOUNDATION Fieldbus™ or
PROFIBUS® PA protocol: TI00086R