Endress+Hauser Micropilot FMR51 Technical Information
Text of the manual
Level measurement in liquids
Application
• Continuous, non-contact level measurement of liquids, pastes and slurries
• Horn antenna (FMR51); flush mounted, completely filled PTFE horn antenna
(FMR52)
• Maximum measuring range: 70 m (230 ft)
• Temperature: –196 to +450 °C (–321 to +842 °F)
• Pressure: –1 to +160 bar (–14.5 to +2 320 psi)
• Accuracy: ± 2 mm
• International explosion protection certificates; WHG; marine approvals
• Linearity protocol (3-point, 5-point)
Your benefits
• Reliable measurement even for changing product and process conditions
• HistoROM data management for easy commissioning, maintenance and
diagnostics
• Highest reliability due to Multi-Echo Tracking
• SIL2 according to IEC 61508, SIL3 in case of homogeneous or heterogeneous
redundancy
• Seamless integration into control or asset management systems
• Intuitive user interface in national languages
• Easy proof test for SIL and WHG
• Heartbeat Technology™
Products
Solutions
Services
Technical Information
Micropilot FMR51, FMR52
Free space radar
TI01040F/00/EN/08.17
71375707
Micropilot FMR51, FMR52
2
Endress+Hauser
Table of contents
Important document information . . . . . . . . . . . . . . . 4
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Function and system design . . . . . . . . . . . . . . . . . . .
6
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operating frequency . . . . . . . . . . . . . . . . . . . . . . . . . .
13
Transmitting power . . . . . . . . . . . . . . . . . . . . . . . . . . .
13
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . .
20
Device plug connectors . . . . . . . . . . . . . . . . . . . . . . . . .
27
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
28
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . .
30
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . .
30
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . .
31
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . .
31
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
31
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
31
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . .
32
Performance characteristics . . . . . . . . . . . . . . . . . . 33
Reference operating conditions . . . . . . . . . . . . . . . . . . .
33
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . 33
Measured value resolution . . . . . . . . . . . . . . . . . . . . . .
34
Reaction time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
34
Influence of ambient temperature . . . . . . . . . . . . . . . . .
34
Influence of gas layer . . . . . . . . . . . . . . . . . . . . . . . . . .
34
Gas phase compensation with external pressure sensor
(PROFIBUS PA, FOUNDATION Fieldbus) . . . . . . . . . . . . .
35
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . .
36
Measuring conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Mounting cladded flanges . . . . . . . . . . . . . . . . . . . . . . . 41
Installation in vessel (free space) . . . . . . . . . . . . . . . . . .
41
Installation in stilling well . . . . . . . . . . . . . . . . . . . . . . . 45
Installation in bypass . . . . . . . . . . . . . . . . . . . . . . . . . .
47
Vessels with heat insulation . . . . . . . . . . . . . . . . . . . . .
49
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Ambient temperature range . . . . . . . . . . . . . . . . . . . . .
50
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . .
50
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . .
58
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
58
Altitude according to IEC61010-1 Ed.3 . . . . . . . . . . . . . .
58
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . .
58
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Cleaning the antenna . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . .
59
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
60
Process temperature, Process pressure . . . . . . . . . . . . . .
60
Dielectric constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Mechanical construction . . . . . . . . . . . . . . . . . . . .
63
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
63
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
73
Materials: GT18 housing (stainless steel, corrosion-
resistant) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . . 75
Materials: GT20 housing (die-cast aluminum, powder-
coated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
76
Materials: Antenna and process connection . . . . . . . . . . .
77
Materials: Weather protection cover . . . . . . . . . . . . . . . .
79
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
80
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . .
80
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
81
Operation with remote display and operating module
FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
81
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
82
Integration in tank gauging system . . . . . . . . . . . . . . . . . 85
SupplyCare inventory management software . . . . . . . . . .
86
Certificates and approvals . . . . . . . . . . . . . . . . . . .
89
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
RCM-Tick marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
89
Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . .
89
Functional Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . .
89
WHG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . .
89
NACE MR 0175 / ISO 15156 . . . . . . . . . . . . . . . . . . . . . 90
NACE MR 0103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
90
Pressure equipment with allowable pressure
≤ 200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . .
90
Marine certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . .
90
Radio standard EN302729-1/2 . . . . . . . . . . . . . . . . . . .
90
Radio standard EN302372-1/2 . . . . . . . . . . . . . . . . . . .
91
FCC / Industry Canada . . . . . . . . . . . . . . . . . . . . . . . . .
91
Japanese radio approval . . . . . . . . . . . . . . . . . . . . . . . .
92
CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
92
Track record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
93
Test, Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Product documentation on paper . . . . . . . . . . . . . . . . . .
94
Other standards and guidelines . . . . . . . . . . . . . . . . . . .
95
Ordering information . . . . . . . . . . . . . . . . . . . . . . .
96
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 96
3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . .
97
5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . .
98
Customized parametrization . . . . . . . . . . . . . . . . . . . . .
99
Micropilot FMR51, FMR52
Endress+Hauser
3
Tagging (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
99
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
99
Application Packages . . . . . . . . . . . . . . . . . . . . . .
100
Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . .
100
Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . .
101
Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 102
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 103
Communication-specific accessories . . . . . . . . . . . . . . .
108
Service-specific accessories . . . . . . . . . . . . . . . . . . . . .
109
System components . . . . . . . . . . . . . . . . . . . . . . . . . .
109
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Standard documentation . . . . . . . . . . . . . . . . . . . . . . . 110
Supplementary documentation . . . . . . . . . . . . . . . . . . . 110
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . .
110
Registered trademarks . . . . . . . . . . . . . . . . . . . . . 113
Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
113
Micropilot FMR51, FMR52
4
Endress+Hauser
Important document information
Symbols
Safety symbols
Symbol
Meaning
DANGER
DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.
WARNING
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.
CAUTION
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.
NOTICE
NOTE!
This symbol contains information on procedures and other facts which do not result in
personal injury.
Electrical symbols
Symbol
Meaning
Direct current
Alternating current
Direct current and alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
Symbols for certain types of information
Symbol
Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
A
Reference to page.
Reference to graphic.
Visual inspection.
Micropilot FMR51, FMR52
Endress+Hauser
5
Symbols in graphics
Symbol
Meaning
1, 2, 3 ...
Item numbers
1., 2. , 3.…
Series of steps
A, B, C, ...
Views
A-A, B-B, C-C, ...
Sections
-
Hazardous area
Indicates a hazardous area.
.
Safe area (non-hazardous area)
Indicates the non-hazardous area.
Symbols at the device
Symbol
Meaning
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
Temperature resistance of the connection cables
Specifies the minimum value of the temperature resistance of the connection cables.
Micropilot FMR51, FMR52
6
Endress+Hauser
Function and system design
Measuring principle
The Micropilot is a "downward-looking" measuring system, operating based on the time-of-flight
method (ToF). It measures the distance from the reference point (process connection) to the product
surface. Radar impulses are emitted by an antenna, reflected off the product surface and received
again by the radar system.
F
L
D
E
R
0%
100%
A0017871
1
Setup parameters of the Micropilot
R
Reference point of the measurement (lower edge of the flange or threaded connection)
E
Empty calibration ( = zero)
F
Full calibration (= span)
D
Measured distance
L
Level (L = E - D)
Input
The reflected radar impulses are received by the antenna and transmitted into the electronics. A
microprocessor evaluates the signal and identifies the level echo caused by the reflection of the radar
impulse at the product surface. The unambiguous signal identification is accomplished by the
PulseMaster® eXact software together with the Multi-echo tracking algorithms, based on many
years of experience with time-of-flight technology.
The distance D to the product surface is proportional to the time of flight t of the impulse:
D = c · t/2,
with c being the speed of light.
Based on the known empty distance E, the level L is calculated:
L = E – D
The reference point R of the measurement is located at the process connection. For details see the
dimensional drawing:
• FMR51: → 64
• FMR52: → 70
The Micropilot is equipped with functions to suppress interference echoes. The user can activate
these functions. Together with the multi-echo tracking algorithms they ensure that interference
echoes (i.e. from edges and weld seams) are not interpreted as level echo.
Micropilot FMR51, FMR52
Endress+Hauser
7
Output
The Micropilot is commissioned by entering an empty distance "E" (=zero), a full distance "F" (=span)
and application parameters which automatically adapt the instrument to the process conditions. For
models with a current output, the factory adjustment for zero point "E" and span "F" is 4 mA and 20
mA. For digital outputs and the display module, the factory adjustment for zero point "E" and span "F"
is 0 % and 100 %.
A linearization with max. 32 points, based on a table entered either manually or semi-automatically,
can be activated locally or remotely. This function provides a measurement in engineering units and
a linear output signal for spheres, horizontal cylindrical tanks and vessels with conical outlet.
Life cycle of the product
Engineering
• Universal measuring principle
• Measurement unaffected by medium properties
• Hardware and software developed according to SIL IEC 61508
Procurement
• Endress+Hauser being the world market leader in level measurement guarantees asset protection
• Worldwide support and service
Installation
• Special tools are not required
• Reverse polarity protection
• Modern, detachable terminals
• Main electronics protected by a separate connection compartment
Commissioning
• Fast, menu-guided commissioning in only a few steps on site or from the control room
• Plain text display in national languages reduces the risk of error or confusion
• Direct local access of all parameters
• Short instruction manual at the device
Operation
• Multi-echo tracking: Reliable measurement through self-learning echo-search algorithms taking
into account the short-term and long-term history in order to check the found echoes for
plausibility and to suppress interference echoes.
• Diagnostics in accordance with NAMUR NE107
Maintenance
• HistoROM: Data backup for instrument settings and measured values
• Exact instrument and process diagnosis to assist fast decisions with clear details concerning
remedies
• Intuitive, menu-guided operating concept in national languages saves costs for training,
maintenance and operation
• Cover of the electronics compartment can be opened in hazardous areas
Retirement
• Order code translation for subsequent models
• RoHS-conforming (Restriction of certain Hazardous Substances), unleaded soldering of electronic
components
• Environmentally sound recycling concept
Micropilot FMR51, FMR52
8
Endress+Hauser
Input
Measured variable
The measured variable is the distance between the reference point and the product surface.
The level is calculated from this distance, taking into account the empty distance "E" entered by the
user.
If required, the level can be converted into other variables (volume, mass) by means of a linearization
(up to 32 points).
Measuring range
Maximum measuring range
Device
Maximum measuring range
FMR51 - standard version
40 m (131 ft)
FMR51 - with "Advanced dynamics" application package
70 m (230 ft)
FMR52 - standard version
40 m (131 ft)
FMR52 - with "Advanced dynamics" application package
60 m (197 ft)
Usable measuring range
The usable measuring range depends on the size of the antenna, the reflectivity of the medium, the
mounting location and eventual interference reflections.
The following tables describe the groups of media as well as the achievable measuring range as a
function of application and media group. If the dielectric constant of a medium is unknown, it is
recommended to assume media group B to ensure a reliable measurement.
Media groups
Media groups
DC (εr)
Example
A
1.4 to 1.9
non-conducting liquids, e.g. liquefied gas 1)
B
1.9 to 4
non-conducting liquids, e.g. benzene, oil, toluene, …
C
4 to 10
e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, …
D
> 10
conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
1)
Treat Ammonia NH3 as a medium of group A.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)
Micropilot FMR51, FMR52
Endress+Hauser
9
Device
Storage tank
A0018833
Calm product surface (e.g. intermittent filling, filling from bottom, immersion tubes)
Antenna size
FMR51
40 mm (1½ in)
50 mm (2 in)
80 mm (3 in)
100 mm (4 in)
FMR52
-
50 mm (2 in)
80 mm (3 in)
-
A
B
25
(82)
15
(49)
C
D
3
(9.9)
10
(32)
15
(49)
5
(16)
8
(26)
5
(16)
A0018858
A
B
C
D
4
(13)
8
(26)
8
(26)
15
(49)
12
(39)
25
(82)
35
(110)
40
(131)
A0018859
A
B
C
D
8
(26)
15
(49)
30
(98)
40
(131)
10
(32)
20
(65)
40
(131)
60
(197)
A0018860
A
B
C
D
10
(32)
25
(82)
40
(131)
40
(131)
15
(49)
30
(99)
45
(148)
70
(229)
A0018862
Measuring range [m (ft)]
Legend
Measuring range of the standard version
Measuring range for the "Advanced dynamics" application package (product structure: feature 540:
"Application Package", Option EM: "Advanced dynamics")
Micropilot FMR51, FMR52
10
Endress+Hauser
Device
Buffer tank
A0018835
Moving surfaces (e.g. continuous filling, from above, mixing jets)
Antenna size
FMR51
40 mm (1½ in)
50 mm (2 in)
80 mm (3 in)
100 mm (4 in)
FMR52
-
50 mm (2 in)
80 mm (3 in)
-
B
C
D
(6.6)
4
(13)
7.5
(24)
10
(32)
(16)
5
2
A0018863
B
C
D
3
(9.9)
7.5
(24)
10
(32)
5
(16)
10
(32)
15
(49)
A0018864
A
B
C
D
2.5
(8)
5
(16)
10
(32)
15
(49)
5
(16)
10
(32)
15
(49)
25
(82)
A0018865
A
B
C
D
5
(16)
10
(32)
15
(49)
25
(82)
7.5
(24)
15
(49)
25
(82)
35
(110)
A0018866
Measuring range [m (ft)]
Legend
Measuring range of the standard version
Measuring range for the "Advanced dynamics" application package (product structure: feature 540:
"Application Package", Option EM: "Advanced dynamics")
Micropilot FMR51, FMR52
Endress+Hauser
11
Device
Process tank with agitator
A0018837
Turbulent surface (e.g. filling from above, agitators, baffles)
Antenna size
FMR51
40 mm (1½ in)
50 mm (2 in)
80 mm (3 in)
100 mm (4 in)
FMR52
-
50 mm (2 in)
80 mm (3 in)
-
B
C
D
1
(3.2)
2
(6.6)
3
(9.8)
5
(16)
A0018867
B
C
D
2
(6.6)
3
(9.8)
5
(16)
7.5
(25)
10
(32)
A0018868
B
C
D
2.5
(8.2)
5
(16)
12
(39)
15
(49)
8
(26)
A0018869
B
C
D
4
(13)
(16)
5
8
(26)
15
(49)
20
(65)
10
(32)
A0018870
Measuring range [m (ft)]
Legend
Measuring range of the standard version
Measuring range for the "Advanced dynamics" application package (product structure: feature 540:
"Application Package", Option EM: "Advanced dynamics")
Micropilot FMR51, FMR52
12
Endress+Hauser
Device
Stilling well
Bypass
A0018842
A0018840
Antenna size
Antenna size
FMR51
40 ... 100 mm ( 1½ ... 4 in)
40 ... 100 mm ( 1½ ... 4 in)
FMR52
50 ... 80 mm ( 2 ... 3 in)
50 ... 80 mm ( 2 ... 3 in)
20
(66)
A, B, C, D
A0018851
20
(66)
C, D
A0018852
Measuring range [m (ft)]
Micropilot FMR51, FMR52
Endress+Hauser
13
Operating frequency
K-band (~ 26 GHz)
Up to 8 Micropilot transmitters can be installed in the same tank because the transmitter pulses are
statistically coded.
Transmitting power
Distance
Average energy density in beam direction
Standard version
With "Advanced dynamics" application
package 1)
1 m (3.3 ft)
< 12 nW/cm2
< 64 nW/cm2
5 m (16 ft)
< 0.4 nW/cm2
< 2.5 nW/cm2
1)
Product structure, feature 540: "Application package", option EM: "Advanced dynamics"
Micropilot FMR51, FMR52
14
Endress+Hauser
Output
Output signal
HART
Signal coding
FSK ±0.5 mA over current signal
Data transmission rate
1 200 Bit/s
Galvanic isolation
Yes
PROFIBUS PA
Signal coding
Manchester Bus Powered (MBP)
Data transmission rate
31.25 kBit/s, voltage mode
Galvanic isolation
Yes
FOUNDATION Fieldbus
Signal coding
Manchester Bus Powered (MBP)
Data transmission rate
31.25 kBit/s, voltage mode
Galvanic isolation
Yes
Switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.
Switch output
Function
Open collector switching output
Switching behavior
Binary (conductive or non-conductive), switches when the programmable switch
point is reached
Failure mode
non-conductive
Electrical connection values U = 16 to 35 VDC, I = 0 to 40 mA
Internal resistance
RI < 880 Ω
The voltage drop at this internal resistance has to be taken into account on
planning the configuration. For example, the resulting voltage at a connected
relay must be sufficient to switch the relay.
Insulation voltage
floating, Insulation voltage 1 350 VDC to power supply aund 500 VAC to ground
Switch point
freely programmable, separately for switch-on and switch-off point
Switching delay
freely programmable from 0 to 100 s, separately for switch-on and switch-off
point
Number of switching cycles
corresponds to the measuring cycle
Signal source
device variables
• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Relative echo amplitude
• Diagnostic values, Advanced diagnostics
Number of switching cycles
unlimited
Micropilot FMR51, FMR52
Endress+Hauser
15
Signal on alarm
Depending on the interface, failure information is displayed as follows:
• Current output (for HART devices)
– Failsafe mode selectable (in accordance with NAMUR Recommendation NE 43):
Minimum alarm: 3.6 mA
Maximum alarm (= factory setting): 22 mA
– Failsafe mode with user-selectable value: 3.59 to 22.5 mA
• Local display
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
• Operating tool via digital communication (HART, PROFIBUS PA, FOUNDATION Fieldbus) or
service interface (CDI)
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
Linearization
The linearization function of the device allows the conversion of the measured value into any unit of
length or volume. Linearization tables for calculating the volume in cylindrical tanks are pre-
programmed. Other linearization tables of up to 32 value pairs can be entered manually or semi-
automatically.
Galvanic isolation
All circuits for the outputs are galvanically isolated from each other.
Protocol-specific data
HART
Manufacturer ID
17 (0x11)
Device type ID
0x1128
HART specification
7.0
Device description files (DTM, DD)
Information and files under:
• www.endress.com
• www.fieldcommgroup.org
HART load
min. 250 Ω
HART device variables
The measured values can be freely assigned to the device variables.
Measured values for PV (primary variable)
• Level linearized
• Distance
• Electronic temperature
• Relative echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Measured values for SV, TV, FV (second, third and fourth variable)
• Level linearized
• Distance
• Electronic temperature
• Terminal voltage
• Relative echo amplitude
• Absolute echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Supported functions
• Burst mode
• Additional transmitter status
Wireless HART data
Minimum start-up voltage
16 V
Start-up current
3.6 mA
Start-up time
65 s
Minimum operating voltage
14.0 V
Micropilot FMR51, FMR52
16
Endress+Hauser
Multidrop current
4.0 mA
Set-up time
15 s
PROFIBUS PA
Manufacturer ID
17 (0x11)
Ident number
0x1559
Profile version
3.02
GSD file
Information and files under:
• www.endress.com
• www.profibus.org
GSD file version
Output values
Analog Input:
• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Digital Input:
• Digital output AD 1
• Digital output AD 2
• Switch output
Input values
Analog Output:
• Analog value from PLC (for sensor block external pressure to compensate gas phase
effects)
• Analog value from PLC to be indicated on the display
Digital Output:
• Extended diagnostic block
• Level limiter
• Sensor block measurement on
• Sensor block save history on
• Status output
Supported functions • Identification & Maintenance
Einfachste Geräteidentifizierung seitens des Leitsystems und des Typenschildes
• Automatic Ident Number Adoption
GSD compatibility mode with respect to the preceding product Micropilot M FMR2xx
• Physical Layer Diagnostics
Installation check of the PRFIBUS segment and the Micropilot FMR5x via the terminal
voltage and telegram surveillance.
• PROFIBUS Up-/Download
Up to 10 times faster writing and reading of parameters via PROFIBUS up-/download
• Condensed Status
Simple and self-explanatory diagnostic information by categorization of occurring
diagnostic messages.
FOUNDATION Fieldbus
Manufacturer ID
0x452B48
Device type
0x1028
Device Revision
0x01
DD Revision
Information and files can be found:
• www.endress.com
• www.fieldcommgroup.org
CFF Revision
Device Tester Version (ITK Version)
6.0.1
ITK Test Campaign Number
IT085300
Link Master (LAS) capable
yes
Micropilot FMR51, FMR52
Endress+Hauser
17
Link Master / Basic Device selectable
yes; default: Basic Device
Node address
Default: 247 (0xF7)
Features supported
Following methods are supported:
• Restart
• ENP Restart
• Setup
• Linearization
• Self Check
Virtual Communication Relationships (VCRs)
Number of VCRs
44
Number of Link Objects in VFD
50
Permanent entries
1
Client VCRs
0
Server VCRs
10
Source VCRs
43
Sink VCRs
0
Subscriber VCRs
43
Publisher VCRs
43
Device Link Capabilities
Slot time
4
Min. inter PDU delay
8
Max. response delay
20
Transducer Blocks
Block
Content
Output values
Setup Transducer Block
Contains all parameters for a standard commissioning
procedure
• Level or volume 1)
(Channel 1)
• Distance (Channel 2)
Advanced Setup
Transducer Block
Contains all parameters for a more detailed
configuration of the device
no output values
Display Transducer Block
Contains all parameters for the configuration of the
display module
no output values
Diagnostic Transducer
Block
Contains diagnostic information
no output values
Advanced Diagnostic
Transducer Block
Contains parameters for the Advanced Diagnostic
no output values
Expert Configuration
Transducer Block
Contains parameters which require detailed knowledge
of the functionalities of the device
no output values
Expert Information
Transducer Block
Contains information about the state of the device
no output values
Service Sensor Transducer
Block
Contains parameters which can only be operated by
Endress+Hauser service personnel
no output values
Service Information
Transducer Block
Contains information on the state of device which is
relevant for service operations
no output values
Data Transfer Transducer
Block
Contains parameters which allow to backup the device
configuration in the display module and to restore it
into the device. Access to these parameters is restricted
to the Endress+Hauser service.
no output values
1)
depending on the configuration of the block
Micropilot FMR51, FMR52
18
Endress+Hauser
Function Blocks
Block
Content
Number of
permanent
blocks
Number of
instantiable
blocks
Execution
time
Functionality
Resource Block
The Resource Block contains all the
data that uniquely identifies the
field device. It is an electronic
version of a nameplate of the
device.
1
0
-
enhanced
Analog Input
Block
The AI block takes the
manufacturer's input data, selected
by channel number, and makes it
available to other function blocks at
its output.
2
3
25 ms
enhanced
Discrete Input
Block
The DI block takes a discrete input
value (e.g. indication of an level
limit), and makes it available to
other function blocks at its output.
1
2
20 ms
standard
Mutiple
Analog Output
Block
This block is used to transfer analog
data from the bus into the device
1
0
20 ms
standard
Mutiple
Discrete
Output Block
This block is used to transfer
discrete data from the bus to the
device.
1
0
20 ms
standard
PID Block
The PID block serves as
proportional-integralderivative
controller and is used almost
universally to do closed-loop-
control in the field including
cascade and feedforward.
1
1
25 ms
standard
Arithmetic
Block
This block is designed to permit
simple use of popular measurement
math functions. The user does not
have to know how to write
equations. The math algorithm is
selected by name, chosen by the
user for the function to be done.
1
1
25 ms
standard
Signal
Characterizer
Block
The signal characterizer block has
two sections, each with an output
that is a non-linear function of the
respective input. The non-linear
function is determined by a single
look-up table with 21 arbitrary x-y
pairs.
1
1
25 ms
standard
Input Selector
Block
The input selector block provides
selection of up to four inputs and
generates an output based on the
configured action. This block
normally receives its inputs from AI
blocks. The block performs
maximum, minimum, middle,
average and ‘first good’ signal
selection.
1
1
25 ms
standard
Micropilot FMR51, FMR52
Endress+Hauser
19
Block
Content
Number of
permanent
blocks
Number of
instantiable
blocks
Execution
time
Functionality
Integrator
Block
The Integrator Function Block
integrates a variable as a function
of the time or accumulates the
counts from a Pulse Input block.
The block may be used as a
totalizer that counts up until reset
or as a batch totalizer that has a
setpoint, where the integrated or
accumulated value is compared to
pre-trip and trip settings,
generating discrete signals when
these settings are reached.
1
1
25 ms
standard
Analog Alarm
Block
1
1
25 ms
standard
Up to 20 blocks can be instantiated in the device altogether, including the blocks already
instantiated on delivery.
Micropilot FMR51, FMR52
20
Endress+Hauser
Power supply
Terminal assignment
2-wire: 4-20mA HART
8
9
1
+
2
4...20mA
HART
10 mm
Spare part
71108xxx
2- wire level
4-20 mA
4-20 mA
HART
[21]
open
-
1
+
2
4-20mA
1-channel overvoltage protection
-
[16]
A
7
B
�
4...20 mA
5
4
1
2
3
+
–
+
–
�
+
–
4...20 mA
5
4
1
2
3
+
–
6
�
A0011294
2
Terminal assignment 2-wire; 4-20mA HART
A
Without integrated overvoltage protection
B
With integrated overvoltage protection
1
Active barrier with power supply (e.g. RN221N): Observe terminal voltage
2
HART communication resistor (≥ 250 Ω): Observe maximum load
3
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4
Analog display device: Observe maximum load
5
Cable screen; observe cable specification
6
4-20mA HART (passive): Terminals 1 and 2
7
Overvoltage protection module
8
Terminal for potential equalization line
9
Cable entry
Micropilot FMR51, FMR52
Endress+Hauser
21
2-wire: 4-20mA HART, switch output
1
+
2
4...20 mA
HART
10 mm
Spare part
71108xxx
2- wire
4-20 mA
PFS
HART
[02/03]
open
-
A
1
+
2
-
3
+
4
-
1
3
+
+
2
4
4-20mA/
FIELDBUS
4-20mA/
2-channel overvoltage protection
-
-
[17]
B
10
9
8
7
11
+
+
-
-
2
2
3
3
4
4
6
5
5
1
1
4...20 mA
≥250 Ω
3+
3+
4-
4-
+
+
–
–
�
�
�
4...20 mA
≥250 Ω
A0013759
3
Terminal assignment 2-wire; 4-20mA HART, switch output
A
Without integrated overvoltage protection
B
With integrated overvoltage protection
1
Active barrier with power supply (e.g. RN221N): Observe terminal voltage
2
HART communication resistor (≥ 250 Ω): Observe maximum load
3
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4
Analog display device: Observe maximum load
5
Cable screen; observe cable specification
6
4-20mA HART (passive): Terminals 1 and 2
7
Switch output (open collector): Terminals 3 and 4
8
Terminal for potential equalization line
9
Cable entry for 4-20mA HART line
10
Cable entry for switch output line
11
Overvoltage protection module
Micropilot FMR51, FMR52
22
Endress+Hauser
2-wire: 4-20mA HART, 4-20mA
1
3
+
+
2
4
4...20mA
HART
4...20mA
10 mm
Spare part
71108xxx
2- wire level
4-20 mA
4-20 mA
HART
[04/05]
open
-
-
11
A
1
3
+
+
2
4
4-20mA/
FIELDBUS
4-20mA/
2-channel overvoltage protection
-
-
[17]
14
13
12
+
+
+
+
-
-
-
-
1
1
2
2
3
3
9
9
5
5
8
8
6
6
7
7
4
4
4
4
+
+
–
–
+
+
–
–
4...20 mA
4...20 mA
10
B
4...20 mA
4...20 mA
�
�
�
�
�
A0013923
4
Terminal assignment 2-wire, 4-20 mA HART, 4...20mA
A
Without integrated overvoltage protection
B
With integrated overvoltage protection
1
Connection current output 2
2
Connection current output 1
3
Supply voltage for current output 1 (e.g. RN221N); Observe terminal voltage
4
Cable screen; observe cable specification
5
HART communication resistor (≥ 250 Ω): Observe maximum load
6
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
7
Analog display device ; observe maximum load
8
Analog display device ; observe maximum load
9
Supply voltage for current output 2 (e.g. RN221N); Obeserve terminal voltage
10
Overvoltage protection module
11
Current output 2: Terminals 3 and 4
12
Terminal for the potential equalization line
13
Cable entry for current output 1
14
Cable entry for current output 2
This version is also suited for single-channel operation. In this case, current output 1 (terminals
1 and 2) must be used.
Micropilot FMR51, FMR52
Endress+Hauser
23
4-wire: 4-20mA HART (10.4 to 48 VDC)
3
1
+
L+
4
2
4...20 mA
HART
10.4...48 V=
10 mm
Spare part
71108xxx
2- wire
4-20 mA
HART
[08]
open
-
L-
A
13
12
11
9
10
2
3
4
7
8
5
1
�
�
+
-
6
�
4...20 mA
≥250 Ω
A0011340
5
Terminal assignment 4-wire; 4-20mA HART (10.4 to 48 VDC)
1
Evaluation unit, e.g. PLC
2
HART communication resistor (≥ 250 Ω): Observe maximum load
3
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4
Analog display device: Observe maximum load
5
Signal cable including screening (if required), observe cable specification
6
Protective connection; do not disconnect!
7
Protective earth, observe cable specification
8
4...20mA HART (active): Terminals 3 and 4
9
Supply voltage: Terminals 1 and 2
10
Supply voltage: Observe terminal voltage, observe cable specification
11
Terminal for potential equalization
12
Cable entry for signal line
13
Cable entry for power supply
Micropilot FMR51, FMR52
24
Endress+Hauser
4-wire: 4-20mA HART (90 to 253 VAC)
3
1
+
L
4
2
4...20 mA
HART
90...253 V~
10 mm
Spare part
71108xxx
2- wire
4-20 mA
HART
[09]
open
-
N
A
13
12
11
9
10
6
7
8
�
�
2
3
4
5
1
+
-
4...20 mA
≥250 Ω
A0018965
6
Terminal assignment 4-wire; 4-20mA HART (90 to 253 VAC)
1
Evaluation unit, e.g. PLC
2
HART communication resistor (≥ 250 Ω): Observe maximum load
3
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4
Analog display device: Observe maximum load
5
Signal cable including screening (if required), observe cable specification
6
Protective connection; do not disconnect!
7
Protective earth, observe cable specification
8
4...20mA HART (active): Terminals 3 and 4
9
Supply voltage: Terminals 1 and 2
10
Supply voltage: Observe terminal voltage, observe cable specification
11
Terminal for potential equalization
12
Cable entry for signal line
13
Cable entry for power supply
LCAUTION
To ensure electrical safety:
‣
Do not disconnect the protective connection (6).
‣
Disconnect the supply voltage before disconnecting the protective earth (7).
Connect protective earth to the internal ground terminal (7) before connecting the supply
voltage. If necessary, connect the potential matching line to the external ground terminal (11).
In order to ensure electromagnetic compatibility (EMC): Do not only ground the device via the
protective earth conductor of the supply cable. Instead, the functional grounding must also be
connected to the process connection (flange or threaded connection) or to the external ground
terminal.
An easily accessible power switch must be installed in the proximity of the device. The power
switch must be marked as a disconnector for the device (IEC/EN61010).
Micropilot FMR51, FMR52
Endress+Hauser
25
PROFIBUS PA / FOUNDATION Fieldbus
1
3
+
+
2
4
FIELDBUS
2-channel overvoltage protection
-
-
[17]
4-20mA/
4-20mA/
B
1
1
+
+
2
2
FIELDBUS
Spare part
71023457
PA/FF
[06/07]
FIELDBUS
-
-
1
3
+
+
2
4
PA/FF
10 mm
Spare part
71108xxx
2- wire level
4-20 mA
PFS
FIELDBUS
[26/27]
open
-
-
A
4
1
1
2
3
6
5
3+
3+
4-
4-
�
A0011341
7
Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus
A
Without integrated overvoltage protection
B
With integrated overvoltage protection
1
Cable screen: Observe cable specifications
2
Switch output (open collector): Terminals 3 and 4
3
PROFIBUS PA / FOUNDATION Fieldbus: Terminals 1 and 2
4
Terminal for potential equalization line
5
Cable entries
6
Overvoltage protection module
Micropilot FMR51, FMR52
26
Endress+Hauser
Connection examples for the switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.
3+
+-
4-
Ri
A0015909
8
Connection of a relay
Suitable relays (examples):
• Solid-state relay: Phoenix Contact OV-24DC/480AC/5 with mounting
rail connector UMK-1 OM-R/AMS
• Electromechanical relay: Phoenix Contact PLC-RSC-12DC/21
3+
2
1
+
4-
Ri
A0015910
9
Connection of a digital input
1
Pull-up resistor
2
Digital input
For optimum interference immunity we recommend to connect an external resistor (internal
resistance of the relay or Pull-up resistor) of < 1 000 Ω.
Micropilot FMR51, FMR52
Endress+Hauser
27
Device plug connectors
For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected
without opening the housing.
Pin assignment of the M12 plug connector
2
1
3
4
A0011175
Pin
Meaning
1
Signal +
2
not connected
3
Signal -
4
Ground
Pin assignment of the 7/8" plug connector
4
2
3
1
A0011176
Pin
Meaning
1
Signal -
2
Signal +
3
Not connected
4
Screen
Micropilot FMR51, FMR52
28
Endress+Hauser
Supply voltage
An external power supply is required.
Various supply units can be ordered from Endress+Hauser: see "Accessories" section→ 109
2-wire, 4-20mA HART, passive
"Power Supply, Output" 1) "Approval" 2)
Terminal voltage U at the
device
Maximum load R, depending on the supply voltage U0 at the
supply unit
A: 2-wire; 4-20mA HART
• Non-Ex
• Ex nA
• Ex ic
• CSA GP
10.4 to 35 V 3) 4)
R [ ]
�
U0 [V]
10
10.4
21.4
20
30
35
0
500
A0017140
Ex ia / IS
10.4 to 30 V 3) 4)
• Ex d(ia) / XP
• Ex ic(ia)
• Ex nA(ia)
• Ex ta / DIP
13 to 35 V 5)
U0 [V]
10
13
24
20
30
35
0
500
R [ ]
Ω
A0034771
Ex ia + Ex d(ia) / IS + XP 13 to 30 V 5)
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
For ambient temperatures Ta≤ -20 °C (-4 °F) a minimum voltage of 15 V is required for the sartup of the device at the MIN error current (3,6
mA). The startup current can be parametrized. If the device is operated with a fixed current I ≥ 5,5 mA (HART multidrop mode), a voltage of U ≥
10,4 V is sufficient throughout the entire range of ambient temperatures.
4)
In the current simulation mode a voltage U ≥ 12.5 V is required.
5)
For ambient temperatures Ta≤ -20 °C (-4 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
"Power Supply, Output" 1)
"Approval" 2)
Terminal voltage U
at the device
Maximum load R, depending on the supply voltage U0 at the
supply unit
B: 2-wire; 4-20 mA HART,
switch output
• Non-Ex
• Ex nA
• Ex nA(ia)
• Ex ic
• Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
13 to 35 V 3)
U0 [V]
10
13
24
20
30
35
0
500
R [ ]
Ω
A0034771
• Ex ia / IS
• Ex ia + Ex d(ia) / IS + XP
13 to 30 V 3)
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
Micropilot FMR51, FMR52
Endress+Hauser
29
"Power Supply, Output" 1)
"Approval" 2) Terminal voltage U at the
device
Maximum load R, depending on the supply voltage U0 at the
supply unit
C: 2-wire; 4-20mA HART,
4-20mA
any
13 to 28 V 3)
R [ ]
�
U0 [V]
10
13
24
20
28
0
500
A0034841
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
Polarity reversal
protection
Yes
Admissible residual ripple
at f = 0 to 100 Hz
USS < 1 V
Admissible residual ripple
at f = 100 to 10000 Hz
USS < 10 mV
Micropilot FMR51, FMR52
30
Endress+Hauser
4-wire, 4-20mA HART, active
"Power supply; Output" 1)
Terminal voltage
Maximum load
Rmax
K: 4-wire 90-253VAC; 4-20mA HART
90 to 253 VAC (50 to 60 Hz),
overvoltage category II
500 Ω
L: 4-wire 10,4-48VDC; 4-20mA HART
10.4 to 48 VDC
1)
Feature 020 of the product structure
PROFIBUS PA, FOUNDATION Fieldbus
"Power supply; Output" 1)
"Approval" 2)
Terminal voltage
E: 2-wire; FOUNDATION Fieldbus, switch output
G: 2-wire; PROFIBUS PA, switch output
• Non-Ex
• Ex nA
• Ex nA(ia)
• Ex ic
• Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
9 to 32 V 3)
• Ex ia / IS
• Ex ia + Ex d(ia) / IS + XP
9 to 30 V 3)
1)
Feature 020 of the product structure
2)
Feature 010 of the product structure
3)
Input voltages up to 35 V will not spoil the device.
Polarity sensitive
No
FISCO/FNICO compliant
according to IEC 60079-27
Yes
Power consumption
"Power supply; Output" 1)
Power consumption
A: 2-wire; 4-20mA HART
< 0.9 W
B: 2-wire; 4-20mA HART, switch output
< 0.9 W
C: 2-wire; 4-20mA HART, 4-20mA
< 2 x 0.7 W
K: 4-wire 90-253VAC; 4-20mA HART
6 VA
L: 4-wire 10,4-48VDC; 4-20mA HART
1.3 W
1)
Feature 020 of the product structure
Current consumption
HART
Nominal current
3.6 to 22 mA, the start-up current for multidrop mode can be parametrized (is
set to 3.6 mA on delivery)
Breakdown signal
(NAMUR NE43)
adjustable: 3.59 to 22.5 mA
PROFIBUS PA
Nominal current
14 mA
Error current FDE (Fault
Disconnection Electronic)
0 mA
Micropilot FMR51, FMR52
Endress+Hauser
31
FOUNDATION Fieldbus
Device basic current
15 mA
Error current FDE (Fault
Disconnection Electronic)
0 mA
FISCO
Ui
17.5 V
Ii
550 mA
Pi
5.5 W
Ci
5 nF
Li
10 μH
Power supply failure
• Configuration is retained in the HistoROM (EEPROM).
• Error messages (incl. value of operated hours counter) are stored.
Potential equalization
No special measures for potential equalization are required.
If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).
Terminals
• Without integrated overvoltage protection
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• With integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
Cable entries
Connection of power supply and signal line
To be selected in feature 050 "Electrical connection"
• Gland M20; Material dependent on the approval:
– For Non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastics M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
– For Dust-Ex, FM IS, CSA IS, CSA GP, Ex nA:
– For Ex d:
No gland available
• Thread
– ½" NPT
– G ½"
– M20 × 1.5
• Plug M12 / Plug 7/8"
Only available for Non-Ex, Ex ic, Ex ia
Connection of remote display FHX50
Dependent on feature 030: "Display, Operation":
• "Prepared for display FHX50 + M12 connection":
M12 socket
• "Prepared for display FHX50 + custom connection":
M16 cable gland
Cable specification
• Devices without integrated overvoltage protection
Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
• For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.
Micropilot FMR51, FMR52
32
Endress+Hauser
HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
• For 4-wire devices: Standard device cable is sufficient for the power line.
PROFIBUS
Use a twisted, screened two-wire cable, preferably cable type A.
For further information on the cable specifications, see Operating Instructions BA00034S
"PROFIBUS DP/PA: Guidelines for planning and commissioning", PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).
FOUNDATION Fieldbus
Endress+Hauser recommends using twisted, shielded two-wire cables.
For further information on the cable specifications, see Operating Instructions BA00013S
"FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP).
Overvoltage protection
If the measuring device is used for level measurement in flammable liquids which requires the use of
overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), overvoltage protection has to be ensured by an
integrated or external overvoltage protection module.
Integrated overvoltage protection
An integrated overvoltage protection module is available for 2-wire HART as well as PROFIBUS PA
and FOUNDATION Fieldbus devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".
Technical data
Resistance per channel
2 × 0.5 Ω max.
Threshold DC voltage
400 to 700 V
Threshold impulse voltage
< 800 V
Capacitance at 1 MHz
< 1.5 pF
Nominal arrest impulse voltage (8/20 μs)
10 kA
External overvoltage protection
HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.
For detailed information please refer to the following documents:
• HAW562: TI01012K
• HAW569: TI01013K
Micropilot FMR51, FMR52
Endress+Hauser
33
Performance characteristics
Reference operating
conditions
• Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
• Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
• Humidity = 60 % ±15 %
• Reflector: metal plate with a minimum diameter of 1 m (40 in)
• No major interference reflections inside the signal beam
Maximum measured error
Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1;
percentage values in relation to the span.
Device
Value
Output
digital
analog 1)
FMR51/FMR52
Standard version
Sum of non-
linearity,
nonrepeatability and
hysteresis
± 2 mm (0.08 in)
± 0.02 %
Offset/Zero
± 4 mm (0.2 in)
± 0.03 %
FMR51/FMR52
Version with application package
"Advanced dynamics" 2)
Sum of non-
linearity,
nonrepeatability and
hysteresis
± 3 mm (0.12 in)
± 0.02 %
Offset/Zero
± 4 mm (0.2 in)
± 0.03 %
1)
Only relevant for 4-20mA current output; add error of the analog value to the digital value.
2)
Product structure: Feature 540 "Application Package", Option EM "Advanced dynamics"
Differing values in near-range applications
2 (0.08)
0
-2 (-0.08)
A
1 (3.3)
4 (0.16)
-4 (-0.16)
R
� [mm] ([in])
D [m] ([ft])
A0019035
10
Maximum measured error in near-range applications; values for standard version
Δ
Maximum measured error
A
Lower edge of the antenna
D
Distance from the lower edge A of the antenna
R
Reference point of the distance measurement
Micropilot FMR51, FMR52
34
Endress+Hauser
3 (0.12)
0
-3 (-12)
A
2 (6.6)
D [m] ([ft])
20 (0.79)
-20 (-0.79)
R
� [mm] ([in])
A0019034
11
Maximum measured error in near-range applications; values for version with the "Advanced dynamics"
application package
Δ
Maximum measured error
A
Lower edge of the antenna
D
Distance from the lower edge A of the antenna
R
Reference point of the distance measurement
Measured value resolution
Dead band according to DIN EN IEC 61298-2 / DIN EN IEC 60770-1:
• digital: 1 mm
• analog: 1 μA
Reaction time
The reaction time can be parametrized. The following step response times (as per DIN EN IEC
61298-2 / DIN EN IEC 60770-1) 1)are valid if the damping is switched off:
Tank height
Sampling rate
Step response time
< 10 m (33 ft)
≥ 3.6 s–1
< 0.8 s
< 70 m (230 ft)
≥ 2.2 s–1
< 1 s
Influence of ambient
temperature
The measurements are carried out in accordance with DIN EN IEC 61298-3 / DIN EN IEC
60770-1
• Digital (HART, PROFIBUS PA, FOUNDATION Fieldbus):
– Standard version: average TK = 2 mm/10 K
– Version with advanced dynamics 2): average TK = 3 mm/10 K
• Analog (current output):
– zero point (4 mA): average TK = 0.02 %/10 K
– span (20 mA): average TK = 0.05 %/10 K
Influence of gas layer
High pressures reduce the propagation velocity of the measuring signals in the gas/vapor above the
fluid. This effect depends on the kind of gas/vapor and of its temperature. This results in a
systematic measuring error that gets bigger as the distance increases between the reference point of
the measurement (flange) and the product surface. The following table illustrates this measured
1)
According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1 the response time is the time which passes after a sudden change of the input signal
until the output signal for the first time assumes 90% of the steady-state value.
2)
Feature 540 "Application Package", option EM
Micropilot FMR51, FMR52
Endress+Hauser
35
error for a few typical gases/vapors (with regard to distance; a positive value means that too large a
distance is being measured):
Gas layer
Temperature
Pressure
°C
°F
1 bar (14,5 psi)
10 bar (145 psi)
50 bar (725 psi)
100 bar (1450 psi)
160 bar (2320 psi)
Air/
Nitrogen
20
68
0.00 %
0.22 %
1.2 %
2.4 %
3.89 %
200
392
–0.01 %
0.13 %
0.74 %
1.5 %
2.42 %
400
752
–0.02 %
0.08 %
0.52 %
1.1 %
1.70 %
Hydrogen
20
68
–0.01 %
0.10 %
0.61 %
1.2 %
2.00 %
200
392
–0.02 %
0.05 %
0.37 %
0.76 %
1.23 %
400
752
–0.02 %
0.03 %
0.25 %
0.53 %
0.86 %
Water
(saturated
steam)
100
212
0.02 %
-
-
-
-
180
356
-
2.1 %
-
-
-
263
505,4
-
-
8.6 %
-
-
310
590
-
-
-
22 %
-
364
687
-
-
-
-
41.8 %
When the pressure is known and constant, this measured error can, for example, be
compensated by means of a linearization.
Gas phase compensation
with external pressure
sensor (PROFIBUS PA,
FOUNDATION Fieldbus)
PROFIBUS devices can receive the signal of an external pressure sensor through the bus and use it to
perform a pressure dependent time-of-flight correction. In the case of saturated steam in the
temperature range from 100 to 350 °C (212 to 662 °f), for example, the measuring error of the
distance measurement can be reduced by this method from up to 29 % (without compensation) to
less than 3 % (with compensation).
Micropilot FMR51, FMR52
36
Endress+Hauser
Installation
Installation conditions
Mounting position
A
1
2
3
A0016882
• Recommended distance A from wall to outer
edge of nozzle: ~ 1/6 of tank diameter.
Nevertheless the device should not be
installed closer than 15 cm (5.91 in) to the
tank wall.
• Not in the center (2), as interference can
cause signal loss.
• Not above the fill stream (3).
• It is recommended to us a weather protection
cover (1) in order to protect the device from
direct sun or rain.
Vessel installations
A0018944
Avoid any installations (point level switches,
temperature sensors, braces, vacuum rings,
heating coils, baffles etc.) inside the signal
beam. Take into account the beam angle
→ 38.
Micropilot FMR51, FMR52
Endress+Hauser
37
Reduction of interference echoes
A0016890
Metallic screens mounted at a slope spread the
radar signal and can, therefore, reduce
interference echoes.
Measurement in a plastic vessel
If the outer wall of the vessel is made of a non-conductive material (e.g. GRP), microwaves can also
be reflected off interfering installations outside the vessel (e.g. metallic pipes (1), ladders (2), grates
(3), ...). Therefore, there should be no such interfering installations in the signal beam. Please
contact Endress+Hauser for further information.
2
3
1
A0017123
Micropilot FMR51, FMR52
38
Endress+Hauser
Optimization options
• Antenna size
The bigger the antenna, the smaller the beam angle α and the fewer interference echoes
→ 38.
• Mapping
The measurement can be optimized by means of electronic suppression of interference echoes.
• Antenna alignment
Take into account the marker on the flange or threaded connection → 41 → 44.
• Stilling well
A stilling well can be applied to avoid interferences → 45.
• Metallic screens mounted at a slope
They spread the radar signals and can, therefore, reduce interference echoes.
Beam angle
�
D
W
�
D
_
=
2
2.
. tan
W
A0016891
12
Relationship between beam angle α, distance D and beamwidth diameter W
The beam angle is defined as the angle α where the energy density of the radar waves reaches half
the value of the maximum energy density (3-dB-width). Microwaves are also emitted outside the
signal beam and can be reflected off interfering installations.
Micropilot FMR51, FMR52
Endress+Hauser
39
Beam diameter W as a function of beam angle α and measuring distance D:
FMR51
Antenna size
40 mm (1½ in)
50 mm (2 in)
80 mm (3 in)
100 mm (4 in)
Beam angle α
23°
18°
10°
8°
Measuring
distance (D)
Beamwidth diameter W
3 m (9.8 ft)
1.22 m (4 ft)
0.95 m (3.1 ft)
0.53 m (1.7 ft)
0.42 m (1.4 ft)
6 m (20 ft)
2.44 m (8 ft)
1.9 m (6.2 ft)
1.05 m (3.4 ft)
0.84 m (2.8 ft)
9 m (30 ft)
3.66 m (12 ft)
2.85 m (9.4 ft)
1.58 m (5.2 ft)
1.26 m (4.1 ft)
12 m (39 ft)
4.88 m (16 ft)
3.80 m (12 ft)
2.1 m (6.9 ft)
1.68 m (5.5 ft)
15 m (49 ft)
6.1 m (20 ft)
4.75 m (16 ft)
2.63 m (8.6 ft)
2.10 m (6.9 ft)
20 m (66 ft)
8.14 m (27 ft)
6.34 m (21 ft)
3.50 m (11 ft)
2.80 m (9.2 ft)
25 m (82 ft)
10.17 m (33 ft)
7.92 m (26 ft)
4.37 m (14 ft)
3.50 m (11 ft)
30 m (98 ft)
-
9.50 m (31 ft)
5.25 m (17 ft)
4.20 m (14 ft)
35 m (115 ft)
-
11.09 m (36 ft)
6.12 m (20 ft)
4.89 m (16 ft)
40 m (131 ft)
-
12.67 m (42 ft)
7.00 m (23 ft)
5.59 m (18 ft)
45 m (148 ft)
-
-
7.87 m (26 ft)
6.29 m (21 ft)
60 m (197 ft)
-
-
10.50 m (34 ft)
8.39 m (28 ft)
70 m (230 ft)
-
-
-
9.79 m (32 ft)
FMR52
Antenna size
50 mm (2 in)
80 mm (3 in)
Beam angle α
18°
10°
Measuring distance (D)
Beamwidth diameter W
3 m (9.8 ft)
0.95 m (3.1 ft)
0.53 m (1.7 ft)
6 m (20 ft)
1.9 m (6.2 ft)
1.05 m (3.4 ft)
9 m (30 ft)
2.85 m (9.4 ft)
1.58 m (5.2 ft)
12 m (39 ft)
3.80 m (12 ft)
2.1 m (6.9 ft)
15 m (49 ft)
4.75 m (16 ft)
2.63 m (8.6 ft)
20 m (66 ft)
6.34 m (21 ft)
3.50 m (11 ft)
25 m (82 ft)
7.92 m (26 ft)
4.37 m (14 ft)
30 m (98 ft)
9.50 m (31 ft)
5.25 m (17 ft)
35 m (115 ft)
11.09 m (36 ft)
6.12 m (20 ft)
40 m (131 ft)
12.67 m (42 ft)
7.00 m (23 ft)
45 m (148 ft)
-
7.87 m (26 ft)
60 m (197 ft)
-
10.50 m (34 ft)
Micropilot FMR51, FMR52
40
Endress+Hauser
Measuring conditions
• In case of boiling surfaces, bubbling or tendency for foaming use FMR53 or FMR54. Depending
on its consistence, foam can either absorb microwaves or reflect them off the foam surface.
Measurement is possible under certain conditions. For FMR50, FMR51 and FMR52, the additional
option "Advanced dynamics" is recommended in these cases (feature 540: "Application Package",
option EM).
• In case of heavy steam development or condensate, the maximum measuring range of FMR50,
FMR51 and FMR52 may decrease depending on density, temperature and composition of the
steam → use FMR53 or FMR54.
• For the measurement of absorbing gases such as ammonia NH3 or some fluorocarbons 3), please
use Levelflex or Micropilot FMR54 in a stilling well.
• The measuring range begins, where the beam hits the tank bottom. Particularly with dish bottoms
or conical outlets the level cannot be detected below this point.
• In stilling well applications, the electromagnetic waves do not propagate completely outside the
tube. It must be taken into account that the accuracy may be reduced in the area C. In order to
guarantee the required accuracy in these cases, it is recommended to position the zero-point at a
distance C above the end of the tube (see figure).
• In case of media with a low dielectric constant (εr = 1.5 to 4) 4) the tank bottom can be visible
through the medium at low levels (low height C). Reduced accuracy has to be expected in this
range. If this is not acceptable, we recommend positioning the zero point at a distance C (see
figure) above the tank bottom in these applications.
• In principle it is possible to measure up to the tip of the antenna with FMR51, FMR53 and FMR54.
However, due to considerations regarding corrosion and build-up, the end of the measuring range
should not be chosen any closer than A (see figure) to the tip of the antenna.
• When using FMR54 with planar antenna, especially for media with low dielectric constants, the
end of the measuring range should not be closer than A: 1 m (3.28 ft) to the flange.
• The smallest possible measuring range B depends on the antenna version (see figure).
• The tank height should be at least H (see table).
100%
0%
B
A
C
H
øD
A0018872
Device
A [mm (in)]
B [m (ft)]
C [mm (in)]
H [m (ft)]
FMR51
50(1.97)
> 0.2 (0.7)
50 to 250 (1.97 to 9.84)
> 0.3 (1.0)
FMR52
200(7.87)
3)
Affected compounds are e.g. R134a, R227, Dymel 152a.
4)
Dielectric constants of important media commonly used in various industries are summarized in the DC manual (CP01076F) and in the Endress
+Hauser "DC Values App" (available for Android and iOS).
Micropilot FMR51, FMR52
Endress+Hauser
41
Mounting cladded flanges
• Use flange screws according to the number of flange holes.
• Tighten the screws with the required torque (see table).
• Retighten the screws after 24 hours or after the first temperature cycle.
• Depending on process pressure and process temperature check and retighten the screws at
regular intervals.
Usually, the PTFE flange cladding also serves as a seal between the nozzle and the device
flange.
Flange size
Number of screws
Recommended torque [Nm]
minimum
maximum
EN
DN50/PN16
4
45
65
DN80/PN16
8
40
55
DN100/PN16
8
40
60
DN150/PN16
8
75
115
ASME
2"/150lbs
4
40
55
3"/150lbs
4
65
95
4"/150lbs
8
45
70
6"/150lbs
8
85
125
JIS
10K 50A
4
40
60
10K 80A
8
25
35
10K 100A
8
35
55
10K 100A
8
75
115
Installation in vessel (free
space)
Horn antenna (FMR51)
Alignment
• Align the antenna vertically to the product surface.
The maximum range may be reduced if the horn antenna is not vertically aligned.
• A marking at the flange (somewhere between the flange holes) or the boss enables alignment of
the antenna. This marking must be aligned towards the tank wall as well as possible.
Micropilot FMR51, FMR52
42
Endress+Hauser
90°
90°
90°
90°
90°
90°
90°
90°
90°
90°
A
B
A0018974
Depending on the device version the marking may be a circle or two short parallel lines.
Nozzle mounting
For optimum measurement, the tip of the antenna should extend below the nozzle. Depending on
the antenna size this is achieved by the following maximum nozzle heights:
Hmax
A0016820
13
Nozzle height for horn antenna (FMR51)
Antenna 1)
Maximum nozzle height Hmax
BA: Horn 40mm/1-1/2"
86 mm (3.39 in)
BB: Horn 50mm/2"
115 mm (4.53 in)
BC: Horn 80mm/3"
211 mm (8.31 in)
BD Horn 100mm/4"
282 mm (11.1 in)
1)
Feature 070 of the product structure
Conditions for longer nozzles
If the medium has good reflective properties, higher nozzles can be accepted. In this case the
maximum nozzle height, Hmax, is dependent on the nozzle diameter, D:
Micropilot FMR51, FMR52
Endress+Hauser
43
Hmax
øD
A0023611
Nozzle diameter D
Maximum nozzle height
Hmax
Recommended antenna 1)
40 mm (1.5 in)
100 mm (3.9 in)
BA: Horn 40mm/1-1/2"
50 mm (2 in)
150 mm (5.9 in)
BB: Horn 50mm/2"
80 mm (3 in)
250 mm (9.8 in)
BC: Horn 80mm/3"
100 mm (4 in)
500 mm (19.7 in)
BD: Horn 100mm/4"
150 mm (6 in)
800 mm (31.5 in)
BD: Horn 100mm/4"
1)
Feature 070 of the product structure
If the antenna doesn't extend below the nozzle, observe the following:
• The nozzle end must be smooth and free of burrs. If possible its edge should be rounded.
• An interference echo suppression must be performed.
• Please contact Endress+Hauser for applications with higher nozzles than those indicated in
the table.
• For mounting in heigh nozzles the device is available in a version with an antenna extension
of up to 1 000 mm (39.4 in) 5)
• The antenna extension may cause interference echoes in the near range. In this case it may
occur that the maximum measurable level is reduced.
Threaded connection
For devices with a threaded connection it may be necessary - depending on the antenna size -
to unmount the horn before fastening the device and to mount it again afterwards.
• Tighten with the hexagonal nut only.
• Tool : 55 mm hexagonal wrench
• Maximum permissible torque: 60 Nm (44 lbf ft)
Mesurement from the outside through plastic walls (FMR50/FMR51)
• Dielectric constant of the medium: εr > 10
• If possible, use an antenna 100 mm (4 in).
• The distance from the lower edge of the antenna to the tank ceiling should be about
100 mm (4 in).
• If possible, avoid mounting locations where condensation or build-up might occur.
• In case of outdoor mounting, the space between antenna and vessel has to be protected from the
elements.
• Do not mount any potential reflectors (e.g. pipes) outside the tank in the signal beam.
5)
Feature 610 "Accessory mounted" of the product structure.
Micropilot FMR51, FMR52
44
Endress+Hauser
Suitable thickness of the tank ceiling:
Penetrated material
PE
PTFE
PP
Perspex
DK / εr
2.3
2.1
2.3
3.1
Optimum thickness 1)
3.8 mm (0.15 in)
4.0 mm (0.16 in)
3.8 mm (0.15 in)
3.3 mm (0.13 in)
1)
Other possible values for the thickness are multiples of the values listed (e.g. for PE: 7,6 mm (0.3 in), 11,4
mm (0.45 in)
Horn antenna, flush mount (FMR52)
Alignment
• Align the antenna vertically to the product surface.
The maximum range may be reduced if the horn antenna is not vertically aligned.
• A marking at the flange (somwhere between the flange holes) or the boss enables alignment of
the antenna. This marking must be aligned towards the tank wall as well as possible.
90°
90°
90°
90°
90°
90°
90°
90°
90°
90°
A
B
A0018974
Depending on the device version the marking may be a circle or two short parallel lines.
Nozzle mounting
Hmax
A0016819
14
Nozzle height for horn antenna, flush mount (FMR52)
Micropilot FMR51, FMR52
Endress+Hauser
45
Antenna 1)
Maximum nozzle height Hmax
BO: Horn 50mm/2"
500 mm (19.7 in)
BP: Horn 80mm/3"
500 mm (19.7 in)
1)
Feature 070 of the product structure
Please contact Endress+Hauser for applications with higher nozzle.
• For flanges with PTFE cladding: Observe the notes on the mounting of cladded flanges
→ 41.
• Usually, the PTFE flange cladding also serves as a seal between the nozzle and the device
flange.
Installation in stilling well
1
90°
A0016841
15
Installation in stilling well
1
Marking for antenna alignment
• For horn antenna: Align the marking towards the slots of the stilling well.
• Measurements can be performed through an open full bore ball valve without any problems.
• After mounting, the housing can be turned 350° in order to facilitate access to the display and the
terminal compartment.
Recommendations for the stilling well
• Metal (no enamel coating; plastic on request).
• Constant diameter.
• Diameter of stilling well not larger than antenna diameter.
• Diameter difference between horn antenna and inner diameter of the stilling well as small as
possible.
• Weld seam as smooth as possible and on the same axis as the slots.
• Slots offset 180° (not 90°).
• Slot width or diameter of holes max. 1/10 of pipe diameter, de-burred. Length and number do not
have any influence on the measurement.
• Select horn antenna as big as possible. For intermedaite sizes (e.g. 180 mm (7 in)) select next
larger antenna and adapt it mechanically (for horn antennas)
• At any transition (i.e. when using a ball valve or mending pipe segments), no gap may be left
exceeding 1 mm (0.04 in).
• The stilling well must be smooth on the inside (average roughness Rz ≤ 6.3 μm (248 μin)). Use
extruded or parallel welded metal pipe. An extension of the pipe is possible with welded flanges or
pipe sleeves. Flange and pipe have to be properly aligned at the inside.
• Do not weld through the pipe wall. The inside of the stilling well must remain smooth. In case of
unintentional welding through the pipe, the weld seam and any unevenness on the inside need to
be carefully removed and smoothened. Otherwise, strong interference echoes will be generated
and material build-up will be promoted.
• In the case of smaller nominal widths flanges must be welded to the pipe such that they allow for a
correct orientation (marker aligned toward slots).
Micropilot FMR51, FMR52
46
Endress+Hauser
Examples for the construction of stilling wells
100%
0%
100%
0%
A
B
A
A
A-A
150…500
(5.91...19.7)
150…500
(5.91...19.7)
mm (in)
2
1
4
4
C
D
5
6
6
7
1
3
1 (0.04)
≤
1 (0.04)
≤
A0019009
A
Micropilot FMR50/FMR51: Horn 40mm(1½")
B
Micropilot FMR50/FMR51/FMR52/FMR54: Horn 80mm(3")
C
Stilling well with slots
D
Full bore ball valve
1
Marking for axial alignment
2
Threaded connection
3
e.g. welding neck flange DIN2633
4
hole max. 1/10 stilling well
5
hole max. 1/10 stilling well; single sided or drilled through
6
Inside of holes deburred
7
Diameter of opening of ball valve must always be equivalent to pipe diameter; avoid edges and constrictions.
Micropilot FMR51, FMR52
Endress+Hauser
47
Installation in bypass
1
2
2
90°
90°
A0019446
16
Installation in bypass
1
Marking for antenna alignment
2
Tank connectors
• Alighn the marker perpendicular (90°) to the tank connectors.
• Measurements can be performed through an open full bore ball valve without any problems.
• After mounting, the housing can be turned 350° in order to simplify access to the display and the
terminal compartment.
Recommendations for the bypass pipe
• Metal (no plastic or enamel coating).
• Constant diameter.
• Select horn antenna as big as possible. For intermediate sizes (e.g. 95 mm (3.5 in)) select next
larger antenna and adapt it mechanically (for horn antennas).
• Diameter difference between horn antenna and inner diameter of the bypass as small as possible.
• At any transition (i.e. when using a ball valve or mending pipe segments), no gap may be created
exceeding 1 mm (0.04 in).
• In the area of the tank connections (~ ±20 cm (7.87 in)) a reduced accuracy of the measurement
has to be expected.
Micropilot FMR51, FMR52
48
Endress+Hauser
Example for the construction of a bypass
100%
0%
A
A
A
A-A
mm (in)
200
(7.87)
200
(7.87)
200
(7.87)
200
(7.87)
B
5
4
4
1
2
C
3
3
≤ 1 (0.04)
A0019010
A
Micropilot FMR50/FMR51/FMR52/FMR54: Horn 80mm(3")
B
Full bore ball valve
C
Minimum distance to upper connection pipe: 400 mm (15,7 in)
1
Marking for axial alignment
2
e.g. welding neck flange DIN2633
3
Diameter of the connection pipes as small as possible
4
Do not weld through the pipe wall; the inside of the bypass must remain smooth.
5
Diameter of opening of ball valve must always be equivalent to pipe diameter. Avoid edges and constrictions.
Micropilot FMR51, FMR52
Endress+Hauser
49
Vessels with heat insulation
1
2
A0019142
If process temperatures are high, the device must be included in normal tank insulation to prevent
the electronics heating up as a result of heat radiation or convection. The insulation may not exceed
beyond the neck of the housing.
Micropilot FMR51, FMR52
50
Endress+Hauser
Environment
Ambient temperature range
Measuring device
–40 to +80 °C (–40 to +176 °F); –50 °C (–58 °F) with manufacturer declaration on
request
Local display
–20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at
temperatures outside the temperature range.
When operating the device in the open with strong sunlight:
• Mount the device in a shady position.
• Avoid direct sunlight, especially in warmer regions.
• Use a weather protection cover (see accessories).
Ambient temperature limits
The following diagrams take into account only functional aspects. There may be further
restrictions for certified device versions. Please refere to the separate Safety Instructions
→ 110.
With a temperature (Tp) at the process connection the admissible ambient temperature (Ta) is
reduced according to the following diagram (temperature derating):
Information concerning the derating tables
Option
Meaning
A
2-wire; 4-20 mA HART
B
2-wire; 4-20 mA HART, switch output
C
2-wire; 4-20 mA HART, 4-20 mA
E
2-wire; FF, switch output
G
2-wire; PA, switch output
K
4-wire 90-253VAC; 4-20 mA HART
L
4-wire 10, 4-48VDC; 4-20 mA HART
FMR51
Seal:
• Viton GLT, –40 to 150 °C (–40 to 302 °F)
• Kalrez, –20 to 150 °C (–4 to 302 °F)
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
150
(302)
66
(151)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
67
(153)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
61
(142)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
67
(153)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
150
(302)
63
(145)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
150
(302)
68
(154)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
Endress+Hauser
51
FMR51
Seal:
• Viton GLT, –40 to 150 °C (–40 to 302 °F)
• Kalrez, –20 to 150 °C (–4 to 302 °F)
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
150
(302)
63
(145)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
62
(144)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR51
Seal:
• Viton GLT, –40 to 150 °C (–40 to 302 °F)
• Kalrez, –20 to 150 °C (–4 to 302 °F)
Housing: GT19 (Plastics PBT)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
80
(176)
80
(176)
80
(176)
150
(302)
56
(133)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
76
(169)
76
(169)
76
(169)
150
(302)
56
(133)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
60
(140)
60
(140)
60
(140)
150
(302)
38
(100)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
56
(133)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
74
(165)
74
(165)
74
(165)
150
(302)
55
(131)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
79
(174)
79
(174)
79
(174)
150
(302)
56
(133)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
63
(145)
63
(145)
63
(145)
150
(302)
41
(106)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR51
Seal:
• Viton GLT, –40 to 150 °C (–40 to 302 °F)
• Kalrez, –20 to 150 °C (–4 to 302 °F)
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
150
(302)
69
(156)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
70
(158)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
64
(147)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
52
Endress+Hauser
FMR51
Seal:
• Viton GLT, –40 to 150 °C (–40 to 302 °F)
• Kalrez, –20 to 150 °C (–4 to 302 °F)
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
150
(302)
70
(158)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
150
(302)
66
(151)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
150
(302)
71
(160)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
150
(302)
65
(149)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
150
(302)
65
(149)
150
(302)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR51
Seal: Graphite, –40 to 250 °C (–40 to 482 °F)
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
250
(482)
63
(145)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
250
(482)
64
(147)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
250
(482)
58
(136)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
250
(482)
64
(147)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
250
(482)
61
(142)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
250
(482)
65
(149)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
250
(482)
60
(140)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
250
(482)
59
(138)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
Endress+Hauser
53
FMR51
Seal: Graphite, –40 to 250 °C (–40 to 482 °F)
Housing: GT19 (Plastics PBT)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
80
(176)
80
(176)
80
(176)
250
(482)
44
(111)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
76
(169)
76
(169)
76
(169)
250
(482)
44
(111)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
60
(140)
60
(140)
60
(140)
250
(482)
32
(90)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
250
(482)
44
(111)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
74
(165)
74
(165)
74
(165)
250
(482)
44
(111)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
79
(174)
79
(174)
79
(174)
250
(482)
44
(111)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
63
(145)
63
(145)
63
(145)
250
(482)
35
(95)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR51
Seal: Graphite, –40 to 250 °C (–40 to 482 °F)
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
250
(482)
67
(153)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
250
(482)
68
(154)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
250
(482)
62
(144)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
250
(482)
68
(154)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
250
(482)
64
(147)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
250
(482)
69
(156)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
250
(482)
64
(147)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
250
(482)
63
(154)
250
(482)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
54
Endress+Hauser
FMR51
Seal: Graphite, –196 to 450 °C (–321 to 842 °F)
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P6
P5
P4
P3
P2
P1
Ta
A0019344
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-196
(-321)
81
(178)
81
(178)
81
(178)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
B
Switch output not used
-196
(-321)
82
(180)
82
(180)
82
(180)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
B
Switch output used
-196
(-321)
77
(171)
77
(171)
77
(171)
450
(842)
25
(77)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
C
Channel 2 not used
-196
(-321)
82
(180)
82
(180)
82
(180)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
C
Channel 2 used
-196
(-321)
79
(174)
79
(174)
79
(174)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
E, G
Switch output not used
-196
(-321)
83
(181)
83
(181)
83
(181)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
E, G
Switch output used
-196
(-321)
78
(172)
78
(172)
78
(172)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
K, L
-196
(-321)
77
(171)
77
(171)
77
(171)
450
(842)
26
(79)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-15
(+5)
FMR51
Seal: Graphite, –196 to 450 °C (–321 to 842 °F)
Housing: GT19 (Plastics PBT)
Temperature unit: °C (°F)
Tp
P6
P5
P4
P3
P2
P1
Ta
A0019344
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-196
(-321)
80
(176)
80
(176)
80
(176)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
B
Switch output not used
-196
(-321)
76
(169)
76
(169)
76
(169)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
B
Switch output used
-196
(-321)
60
(140)
60
(140)
60
(140)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
C
Channel 2 not used
-196
(-321)
82
(180)
82
(180)
82
(180)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
C
Channel 2 used
-196
(-321)
74
(165)
74
(165)
74
(165)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
E, G
Switch output not used
-196
(-321)
79
(174)
79
(174)
79
(174)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
E, G
Switch output used
-196
(-321)
63
(145)
63
(145)
63
(145)
450
(842)
-29
(-20)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
9
(48)
Micropilot FMR51, FMR52
Endress+Hauser
55
FMR51
Seal: Graphite, –196 to 450 °C (–321 to 842 °F)
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P6
P5
P4
P3
P2
P1
Ta
A0019344
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-196
(-321)
81
(178)
81
(178)
81
(178)
450
(842)
39
(102)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
B
Switch output not used
-196
(-321)
82
(180)
82
(180)
82
(180)
450
(842)
39
(102)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
B
Switch output used
-196
(-321)
77
(171)
77
(171)
77
(171)
450
(842)
36
(97)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
C
Channel 2 not used
-196
(-321)
82
(180)
82
(180)
82
(180)
450
(842)
39
(102)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
C
Channel 2 used
-196
(-321)
79
(174)
79
(174)
79
(174)
450
(842)
38
(100)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
E, G
Switch output not used
-196
(-321)
83
(181)
83
(181)
83
(181)
450
(842)
39
(102)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
E, G
Switch output used
-196
(-321)
78
(172)
78
(172)
78
(172)
450
(842)
38
(100)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
K, L
-196
(-321)
77
(171
77
(171)
77
(171)
450
(842)
37
(99)
450
(842)
-40
(-40)
-40
(-40)
-40
(-40)
-196
(-321)
-20
(-4)
FMR52
Antenna: Horn 50mm/2"
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
200
(392)
61
(142)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
61
(142)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
55
(131)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
58
(136)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
56
(133)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
56
Endress+Hauser
FMR52
Antenna: Horn 50mm/2"
Housing: GT19 (Plastics PBT)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
80
(176)
80
(176)
80
(176)
200
(392)
42
(108)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
76
(169)
76
(169)
76
(169)
200
(392)
42
(108)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
60
(140)
60
(140)
60
(140)
200
(392)
31
(88)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
42
(108)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
74
(165)
74
(165)
74
(165)
200
(392)
42
(108)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
42
(108)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
63
(145)
63
(145)
63
(145)
200
(392)
33
(91)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR52
Antenna: Horn 50mm/2"
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
200
(392)
65
(149)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
65
(149
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
59
(138)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
66
(151)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
200
(392)
66
(1512)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
200
(392)
61
(142)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
60
(140)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
Endress+Hauser
57
FMR52
Antenna: Horn 80mm/2"
Housing: GT18 (316 L)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
51
(124)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
54
(129)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
200
(392)
53
(127)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
52
(126)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
FMR52
Antenna: Horn 80mm/2"
Housing: GT19 (Plastics PBT)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
80
(176)
80
(176)
80
(176)
200
(392)
36
(97)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
76
(169)
76
(169)
76
(169)
200
(392)
36
(97)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
60
(140)
60
(140)
60
(140)
200
(392)
27
(81)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
36
(97)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
74
(165)
74
(165)
74
(165)
200
(392)
36
(97)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
36
(97)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
63
(145)
63
(145)
63
(145)
200
(392)
30
(86)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Micropilot FMR51, FMR52
58
Endress+Hauser
FMR52
Antenna: Horn 80mm/2"
Housing: GT20 (Alu, coated)
Temperature unit: °C (°F)
Tp
P5
P4
P3
P2
P1
Ta
A0019351
Power Supply; Output (Pos. 2
of the product structure)
P1
P2
P3
P4
P5
P6
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
Tp
Ta
A
-40
(-40)
81
(178)
81
(178)
81
(178)
200
(392)
61
(142)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
B
Switch output used
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
56
(133)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 not used
-40
(-40)
82
(180)
82
(180)
82
(180)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
C
Channel 2 used
-40
(-40)
79
(174)
79
(174)
79
(174)
200
(392)
58
(136)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output not used
-40
(-40)
83
(181)
83
(181)
83
(181)
200
(392)
62
(144)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
E, G
Switch output used
-40
(-40)
78
(172)
78
(172)
78
(172)
200
(392)
58
(136)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
K, L
-40
(-40)
77
(171)
77
(171)
77
(171)
200
(392)
57
(135)
200
(392)
-40
(-40)
-40
(-40)
-40
(-40)
-
-
Storage temperature
–40 to +80 °C (–40 to +176 °F)
–50 °C (–58 °F) with manufacturer declaration on request
Climate class
DIN EN 60068-2-38 (test Z/AD)
Altitude according to
IEC61010-1 Ed.3
• Generally up to 2 000 m (6 600 ft) above MSL.
• Above 2 000 m (6 600 ft) if the following conditions are met:
– Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions)
– Supply voltage U < 35 V
– Supply voltage of overvoltage category 1
Degree of protection
• With closed housing tested according to:
– IP68, NEMA6P (24 h at 1.83 m under water surface)
– For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water
surface) 6)
– IP66, NEMA4X
• With open housing: IP20, NEMA1
• Display module: IP22, NEMA2
Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the
PROFIBUS cable is plugged in and is also rated IP68 NEMA6P.
Vibration resistance
DIN EN 60068-2-64 / IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s2)2/Hz
Cleaning the antenna
The antenna can get contaminated, depending on the application. The emission and reception of
microwaves can thus eventually be hindered. The degree of contamination leading to an error
depends on the medium and the reflectivity, mainly determined by the dielectric constant εr.
If the medium tends to cause contamination and deposits, cleaning on a regular basis is
recommended. Care has to be taken not to damage the antenna in the process of a mechanical or
6)
This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C
("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").
Micropilot FMR51, FMR52
Endress+Hauser
59
hose-down cleaning. The material compatibility has to be considered if cleaning agents are used! The
maximum permitted temperature at the flange should not be exceeded.
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility to all relevant requirements of the EN 61326- series and NAMUR
recommendation EMC (NE21). For details see declaration of conformity. 7).
If only the analogue signal is used, unshielded interconnection lines are sufficient for the
installation. In case of using the digital signal (HART/ PA/ FF) use shielded interconnection lines.
Max. fluctuations during EMC- tests: < 0.5 % of the span. As an exception to this, the maximum
fluctuations may amount to 2 % of the span for devices with plastic housing and see-through lid
(integrated display SD02 or SD03) if strong electromagnetic interferences in the frequenency range
of 1 to 2 GHz are present.
7)
Can be downloaded from www.endress.com.
Micropilot FMR51, FMR52
60
Endress+Hauser
Process
Process temperature, Process
pressure
The specified pressure range may be reduced due to the selected process connection. The
pressure rating (PN) specified on the flanges refers to a reference temperature of 20 °C, for
ASME flanges 100 °F. Pay attention to pressure-temperature dependencies.
Please refer to the following standards for the pressure values permitted for higher
temperatures:
• EN 1092-1: 2001 Tab. 18
With regard to their temperature stability properties, the materials 1.4435 and 1.4404 are
grouped under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials
can be identical.
• ASME B 16.5a - 1998 Tab. 2-2.2 F316
• ASME B 16.5a - 1998 Tab. 2.3.8 N10276
• JIS B 2220
FMR51
p
p
-40
(-40)
-40
(-40)
-20
(-4)
+150
(+302)
+250
(+482)
0
(+32)
0
(+32)
Tp
[°C]
([°F])
Tp
[°C]
([°F])
1
2
3
[bar] ([psi])
[bar] ([psi])
40 (580)
40 (580)
-1 (-14.5)
-1 (-14.5)
0 (0)
0 (0)
A0022656-EN
17
FMR51: Admissible range of process temperature and process pressure
1
Seal: Graphite (Feature 090 "Seal", Option D3)
2
Seal: Viton (Feature 090 "Seal", Option A5)
3
Seal: Kalrez (Feature 090 "Seal", Option C1)
Micropilot FMR51, FMR52
Endress+Hauser
61
p
-196
(-321)
-100
(-148)
+310
(+590)
+350
(+662)
+400
(+752)
+450
(+842)
0
(+32)
Tp
[°C]
([°F])
160 (2320)
154 (2233)
148 (2146)
142 (2059)
[bar] ([psi])
A0020555-EN
18
FMR51: Admissible range of process temperature and process pressure of the HT version (Feature 090
"Seal", Option D2)
Feature 090 "Seal"
Admissible process
temperature
Admissible process pressure
A5: Viton GLT
–40 to +150 °C
(–40 to 302 °F)
prel = –1 to 40 bar (–14.5 to 580 psi)
C1: Kalrez
–20 to +150 °C
(–4 to 302 °F)
D3: Graphite,
40...250°C/-40...482°F
–40 to +250 °C
(–40 to 482 °F)
D2: Graphite,
196...450°C/-321...842°F (HT)
–196 to +450 °C
(–321 to 842 °F)
prel = –1 to 160 bar (–14.5 to 2 320 psi)
Observe the temperature-dependent
restrictions indicated in the diagram
above 1).
1)
For devices with CRN approval the pressure range may be reduced → 92
FMR52
p
-40
(-40)
0
(+32)
Tp
[°C]
([°F])
+200
(+392)
+100
(+212)
2
1
[bar] ([psi])
25 (362.5)
-1 (-14.5)
0 (0)
A0023250-EN
19
FMR52: Admissible range of process temperature and process pressure
1
Valid for the following process connections: 6" (ANSI), DN150 (EN), 10K150A (JIS)
2
Valid for all other types of process connection
Micropilot FMR51, FMR52
62
Endress+Hauser
Feature 100 "Process Connection" Admissible process
temperature
Admissible process pressure
• AFK: 2" 150lbs (ANSI)
• AGK: 3" 150lbs (ANSI)
• AHK: 4" 150lbs (ANSI)
• CFK: DN50 PN10/16 (EN)
• CGK: DN80 PN10/16 (EN)
• CHK: DN100 PN10/16 (EN)
• KFK: 10K 50A (JIS)
• KGK: 10K 80A (JIS)
• KHK: 10K 100A (JIS)
• MRK: DIN11851 DN50
• MTK: DIN11851 DN80
• TDK: Tri-Clamp 2"
• TFK: Tri-Clamp 3"
• THK: Tri-Clamp 4"
–40 to +200 °C
(–40 to 392 °F)
prel = –1 to 25 bar (–14.5 to 363 psi) 1) 2)
• AJK: 6" 150lbs (ANSI)
• CJK: DN150 PN10/16 (EN)
• KJK: 10K 150A (JIS)
• For T = –40 to +100 °C (–40 to 212 °F):
prel = –1 to 25 bar (–14.5 to 363 psi)
• For T = 100 to 200 °C (212 to 392 °F):
prel = 0 to 25 bar (0 to 363 psi)
1)
For devices with CRN approval the pressure range may be reduced → 92
2)
Process connections up to 25 bar (363 psi) on request
Dielectric constant
• For liquids
– εr ≥ 1.9 in free-field applications
– εr ≥ 1.4 in stilling well
• For bulk solids
εr ≥ 1.6
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)
Micropilot FMR51, FMR52
Endress+Hauser
63
Mechanical construction
Dimensions
Dimensions of the electronics housing
146 (5.75)
142 (5.59)
115.25 (4.54)
ø108.5 (4.27)
78 (3.07)
90 (3.54)
98 (3.86)*
ø103.5 (4.07)
R110
A0011666
20
Housing GT18 (316L); Dimensions in mm (in)
*for devices with integrated overvoltage protection.
144 (5.67)
113.5 (4.47)
ø106 (4.17)
78 (3.07)
90 (3.54)
99 (3.9)*
ø106 (4.17)
R110
A0011346
21
Housing GT19 (Plastics PBT); Dimensions in mm (in)
*for devices with integrated overvoltage protection.
145.5 (5.73)
141.5 (5.57)
ø112 (4.41)
ø108.5 (4.27)
78 (3.07)
90 (3.54)
98 (3.86)*
ø103.5 (4.07)
R110
61 (2.4)
A0020751
22
Housing GT20 (Alu coated); Dimensions in mm (in)
*for devices with integrated overvoltage protection.
Micropilot FMR51, FMR52
64
Endress+Hauser
FMR51: Version T ≤ 150 °C (302 °F); without antenna extension
A
B
C
R
L
L
68
(2.68)
63
(2.48)
5 (0.2)
26 (1.02)
b
5 (0.2)
ød
55
ød
øD
øA
D
L
65
(2.56)
b
2 (0.08)
ød
øD
A0023383
23
FMR51: Version T < 150 °C (302 °F); without antenna extension; dimensions: mm (in)
A
Process connection: Thread
B
Process connection: Tri-Clamp ISO2852
C
Process connection: Flange 316L
D
Process connection: Flange AlloyC>316L
R
Reference point of the measurement
Valid for the following device versions
Feature 090 "Seal":
– A5: Viton GLT
– C1: Kalrez
• Dimensions of the horn antenna (L, ⌀d): → 68
• Dimensions of the process connection (⌀A, ⌀D, b): → 68
Micropilot FMR51, FMR52
Endress+Hauser
65
FMR51: Version T ≤ 150 °C (302 °F); with antenna extension
A
R
L
68 (2.68)
26 (1.02)
5 (0.2)
100(3.94)
ød
55
B
C
L
63 (2.48)
b
5 (0.2)
100(3.94)
ød
øD
øA
D
E
63
(2.48)
68
(2.68)
b
47 (1.85)
ød
øD
101...1000
(3.98...39.3)
L2
ø27(1.06)
ø42.5(1.7)
55
26 (1.02)
A0023384
24
MR51: Version T < 150 °C (302 °F); with antenna extension; dimensions: mm (in)
A
Process connections: Thread; Accessory mounted: 100mm/4" antenna extension
B
Prozessanschluss: Tri-Clamp ISO2852; Accessory mounted: 100mm/4" antenna extension
C
Process connection: Flange 316L; Accessory mounted: 100mm/4" antenna extension
D
Process connection: Thread; Accessory mounted: ..... mm/inch antenna extension
E
Process connection: Flange 316L;Accessory mounted: ..... mm/inch antenna extension
R
Reference point of the measurement
Valid for the following device versions
• Feature 090 "Seal":
– A5: Viton GLT
– C1: Kalrez
• Feature 610 "Accessory mounted"
– OM: 100mm/4" antenna extension (diagrams A, B, C)
– OU: ..... mm antenna extension (101 ... 1000 mm) (diagrams D, E)
– OV: ..... inch antenna extension (3,98 ... 39,3 in) (diagrams D, E)
• Dimensions of the horn antenna (L, L2, ⌀d): → 68
• Dimensions of the process connections (⌀A, ⌀D, b): → 68
Micropilot FMR51, FMR52
66
Endress+Hauser
FMR51: Version T ≤ 250 °C (482 °F)
A
B
R
L
122 (4.8)
116(4.57)
26 (1.02)
b
ød
55
øD
C
D
122 (4.8)
116(4.57)
26 (1.02)
b
55
øD
47 (1.85)
ød
101...1000
(3.98...39.3)
L2
ø27 (1.06)
ø42.5(1.7)
A0023385
25
FMR51: Version T < 250 °C (482 °F); dimensions: mm (in)
A
Process connection: Thread
B
Process connection: Flange 316L
C
Process connection: Thread; Accessory mounted: ..... mm/inch antenna extension
D
Process connection: Flange 316L; Accessory mounted: ..... mm/inch antenna extension
R
Reference point of the measurement
Valid for the following device versions
Feature 090 "Seal":
D3: Graphite, -40...250°C/-40...482°F
• Dimensions of the horn antenna (L, ⌀d): → 68
• Dimensions of the process connection (⌀D, b): → 68
Micropilot FMR51, FMR52
Endress+Hauser
67
FMR51: Version T ≤ 450 °C (842 °F)
L
141 (5.55)
b
ød
øD
R
A0023386
26
FMR51: Version T < 450 °C (842 °F); dimensions: mm (in)
Valid for the following device versions
Feature 090 "Seal":
D2: Graphite, -196...450°C/-321...842°F
• Dimensions of the horn antenna (L, ⌀d): → 68
• Dimensions of the process connection (⌀D, b): → 68
Micropilot FMR51, FMR52
68
Endress+Hauser
FMR51: Dimensions of the horn antenna
Feature 070 "Antenna"
Dimension
BA: Horn 40mm/1-1/2"
BB: Horn 50mm/2"
BC: Horn 80mm/3"
BD: Horn 100mm/4"
L
86 mm (3.39 in)
115 mm (4.53 in)
211 mm (8.31 in)
282 mm (11.1 in)
L2
51 mm (2.01 in)
80 mm (3.15 in)
176 mm (6.93 in)
247 mm (9.72 in)
d
40 mm (1.57 in)
48 mm (1.89 in)
75 mm (2.95 in)
95 mm (3.74 in)
FMR51: Dimensions of the process connections
Flanges according to EN1092-1 (suitable for DIN2527) 1)
Pressure rating Dimension
Nominal diameter 2)
DN50
DN80
DN100
DN150
PN10/16
b
18 mm (0.71 in)
20 mm (0.79 in)
20 mm (0.79 in)
22 mm (0.87 in)
⌀D
165 mm (6.5 in) 200 mm
(7.87 in)
220 mm (8.66 in) 285 mm
(11.2 in)
PN25/40
b
20 mm (0.79 in)
24 mm (0.94 in)
24 mm (0.94 in)
-
⌀D
165 mm (6.5 in) 200 mm
(7.87 in)
235 mm (9.25 in) -
PN63
b
-
-
30 mm (1.18 in)
-
⌀D
-
-
250 mm (9.84 in) -
PN100
b
-
-
36 mm (1.42 in)
-
⌀D
-
-
265 mm (10.4 in) -
1)
For flanges with a higher pressure rating than those listed in the product structure please contact Endress
+Hauser.
2)
see feature 100 "Process connection"
Flanges according to JIS B2220 1)
Nominal diameter 2)
Pressure
rating 2)
Dimension 50A
80A
100A
150A
10K
b
16 mm (0.63 in)
18 mm (0.71 in)
18 mm (0.71 in)
22 mm (0.87 in)
D
155 mm (6.1 in) 185 mm (7.28 in) 210 mm (8.27 in) 280 mm (11 in)
63K
b
-
-
44 mm (1.73 in)
-
D
-
-
270 mm (10.6 in) -
1)
For flanges with a higher pressure rating than those listed in the product structure please contact Endress
+Hauser.
2)
see feature 100 "Process connection"
Micropilot FMR51, FMR52
Endress+Hauser
69
Flanges according to ANSI B16.5 1)
Nominal diameter 2)
Pressure
rating 2)
Dimension 2"
3"
4"
6"
150 lbs
b
17.5 mm
(0.69 in)
22.3 mm
(0.88 in)
22.3 mm
(0.88 in)
23.9 mm
(0.94 in)
D
150 mm (6 in)
190 mm (7.5 in)
230 mm (9 in)
280 mm (11 in)
300 lbs
b
20.7 mm
(0.81 in)
27 mm (1.1 in)
30.2 mm (1.2 in)
35 mm (1.4 in)
D
165 mm (6.5 in)
210 mm
(8.25 in)
255 mm (10 in)
320 mm
(12.5 in)
1)
For flanges with a higher pressure rating than those listed in the product structure please contact Endress
+Hauser.
2)
see feature 100 "Process connection"
Tri-Clamp according to ISO 2852
Nominal diameter 1)
Dimension
DN40-51 (2")
DN70-76.1 (3")
A
64 mm (2.52 in)
91 mm (3.58 in)
1)
see feature 100 "Process connection"
Micropilot FMR51, FMR52
70
Endress+Hauser
FMR52 with dairy coupling DIN11851
R
A
B
ø100 (3.94)
ø49 (1.93)
ø68.5 (2.7)
ø81 (3.19)
14 (0.55)
29 (1.14)
5 (0.2)
6 (0.24)
90 (3.54)
122 (4.8)
A0023387
27
FMR52 with dairy coupling DIN11851; dimensions: mm (in)
A
Antenna: Horn 50mm/2"; Process connection DIN11851 DN50 PN25
B
Antenna: Horn 80mm/3"; Process connection DIN11851 DN80 PN25
R
Reference point of the measurement
Valid for the following device versions:
Feature 100 "Process connection"
– MRK: DIN11851 DN50 PN25 slotted nut, PTFE>316L (diagram A)
– MTK: DIN11851 DN80 PN25 slotted nut, PTFE>316L (diagram B)
Micropilot FMR51, FMR52
Endress+Hauser
71
FMR52 with Tri-Clamp ISO2852
R
A
B
øa
øa
ø81 (3.19)
ø49 (1.93)
29 (1.14)
14 (0.55)
2.6 (0.1)
2.6 (0.1)
120 (4.72)
93(3.66)
A0023388
28
FMR52 with Tri-Clamp ISO2852; dimensions: mm (in)
A
Antenna: Horn 50mm/2"
B
Antenna: Horn 80mm/3"
R
Reference point of the measurement
Antenna 1)
BO: "Horn 50mm/2"
BP: Horn 80mm/3"
Process
connection 2)
TDK: Tri-Clamp DN40-51
(2")
TFK: Tri-Clamp DN70-76.1
(3")
THK: Tri-Clamp DN101.6
(4")
a
64 mm (2.52 in)
91 mm (3.58 in)
119 mm (4.69 in)
1)
Feature 070 of the product structure
2)
Feature 100 of the product structure
Micropilot FMR51, FMR52
72
Endress+Hauser
FMR52 with flange
R
A
B
øb
øb
øc
øc
28.6 (1.13)
13.8 (0.54)
4 (0.16)
a
a
4 (0.16)
94(3.7)
121(4.76)
ø48 (1.89)
ø75 (2.95)
A0023389
29
Dimensions FMR52 with flange; dimensions: mm (in)
A
Antenna: Horn 50mm/2"
B
Antenna: Horn 80mm/3"
R
Reference point of the measurement
Flanges according to EN1092-1 (suitable for DIN2527)
Antenna 1)
BO: Horn 50mm/2"
BP: Horn 80mm/3"
Process connection 2)
CFK: Flange DN50
CGK: Flange DN80
CHK: Flange DN100
CJK: Flange DN150
a
20 mm (0.79 in)
20 mm (0.79 in)
20 mm (0.79 in)
22 mm (0.87 in)
b
102 mm (4.02 in)
138 mm (5.43 in)
158 mm (6.22 in)
212 mm (8.35 in)
c
165 mm (6.5 in)
200 mm (7.87 in)
220 mm (8.66 in)
285 mm (11.2 in)
1)
Feature 70 of the product structure
2)
Feature 100 of the product structure
Flanges according to ANSI B16.5
Antenna 1)
BO: Horn 50mm/2"
BP: Horn 80mm/3"
Process connection 2)
FK: Flange 2"
AGK: Flange 3"
AHK: Flange 4"
AJK: Flange 6"
a
17.5 mm (0.69 in)
22.3 mm (0.88 in)
22.3 mm (0.88 in)
23.9 mm (0.94 in)
b
92 mm (3.62 in)
127 mm (5 in)
158 mm (6.22 in)
212 mm (8.35 in)
c
150 mm (6 in)
190 mm (7.5 in)
230 mm (9 in)
280 mm (11 in)
1)
Feature 70 of the product structure
2)
Feature 100 of the product structure
Flanges according to JIS B2220
Antenna 1)
BO: Horn 50mm/2"
BP: Horn 80mm/3"
Process
connection 2)
KFK: Flange 10K
50A
KGK: Flange 10K
80A
KHK: Flange 10K
100A
KJK: Flange 10K
150A
a
16 mm (0.63 in)
18 mm (0.71 in)
18 mm (0.71 in)
22 mm (0.87 in)
b
96 mm (3.78 in)
127 mm (5 in)
151 mm (5.94 in)
212 mm (8.35 in)
c
155 mm (6.1 in)
185 mm (7.28 in)
210 mm (8.27 in)
280 mm (11 in)
1)
Feature 70 of the product structure
2)
Feature 100 of the product structure
Micropilot FMR51, FMR52
Endress+Hauser
73
Weight
Housing
Part
Weight
Housing GT18 - stainless steel
approx.
4.5 kg (9.9 lb)
Housing GT19 - plastic
approx.
1.2 kg (2.7 lb)
Housing GT20 - aluminium
approx.
1.9 kg (4.2 lb)
Antenna and process connection
Device
Weight of antenna and process connection
FMR51
max. 3.0 kg (6.6 lb) + weight of flange 1)
FMR52
max. 4 kg (8.8 lb) + weight of flange 1)
1)
For the weight of the flange refer to Technical Information TI00426F.
Micropilot FMR51, FMR52
74
Endress+Hauser
Materials: GT18 housing
(stainless steel, corrosion-
resistant)
2.1
1
2.2
5.2
5.1
6
7
8
9
4
3
A0013788
No. Part
Material
1
Housing
CF3M similar to 316L/1.4404
2.1
Cover of the electronics compartment
• Cover: CF3M (similar to 316L/1.4404)
• Window: glass
• Cover seal: NBR
• Seal of the window: NBR
• Thread-coating: Graphite-based lubricant varnish
2.2
Cover of the terminal compartment
• Cover: CF3M (similar to 316L/1.4404)
• Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish
3
Cover lock
• Screw: A4
• Clamp: 316L (1.4404)
4
Lock at the housing neck
• Screw: A4-70
• Clamp: 316L (1.4404)
5.1
Dummy plug, cable gland, adapter or plug
(depending on the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
• Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2
Dummy plug, cable gland or adapter (depending on
the device version)
• Dummy plug: 316L (1.4404)
• Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6
Dummy plug or M12 socket (depending on the
device version)
• Dummy plug: 316L (1.4404)
• M12 socket: 316L (1.4404)
7
Pressure relief stopper
316L (1.4404)
8
Ground terminal
• Screw: A4
• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)
9
Nameplate
• Plate: 316L (1.4404)
• Groove pin: A4 (1.4571)
1)
For the version with M12 plug the sealing material is Viton.
2)
For the version with 7/8" plug, the sealing material is NBR.
Micropilot FMR51, FMR52
Endress+Hauser
75
Materials: GT19 housing
(plastic)
2.1
1
2.2
5.2
5.1
6
7
8
9
4
A0013788
No. Part
Material
1
Housing
PBT
2.1
Cover of the electronics compartment
• Cover glass: PC
• Cover frame: PBT-PC
• Cover seal: EPDM
• Thread-coating: Graphite-based lubricant varnish
2.2
Cover of the terminal compartment
• Cover: PBT
• Cover seal: EPDM
• Thread-coating: Graphite-based lubricant varnish
4
Lock at the housing neck
• Screw: A4-70
• Clamp: 316L (1.4404)
5.1
Dummy plug, cable gland, adapter or plug
(depending on the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2
Dummy plug, cable gland or adapter (depending on
the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6
Dummy plug or M12 socket (depending on the
device version)
• Dummy plug: Nickel-plated brass (CuZn)
• M12 socket: Nickel-plated GD-Zn
7
Pressure relief stopper
Nickel-plated brass (CuZn)
8
Ground terminal
• Screw: A2
• Spring washer: A4
• Clamp: 304 (1.4301)
• Holder: 304 (1.4301)
9
Adhesive nameplate
Plastic
1)
For the version with M12 plug the sealing material is Viton.
2)
For the version with 7/8" plug, the sealing material is NBR.
Micropilot FMR51, FMR52
76
Endress+Hauser
Materials: GT20 housing
(die-cast aluminum, powder-
coated)
2.1
1
2.2
5.2
5.1
6
7
8
9
4
3
A0013788
Nr. Part
Material
1
Housing, RAL 5012 (blue)
• Housing: AlSi10Mg(<0,1% Cu)
• Coating: Polyester
2.1 Cover of the electronics compartment; RAL 7035
(gray)
• Cover: AlSi10Mg(<0,1% Cu)
• Window: Glass
• Cover seal: NBR
• Seal of the window: NBR
• Thread-coating: Graphite-based lubricant varnish
2.2 Cover of the terminal compartment; RAL 7035
(gray)
• Cover: AlSi10Mg(<0,1% Cu)
• Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish
3
Cover lock
• Screw: A4
• Clamp: 316L (1.4404)
4
Lock at the housing neck
• Screw: A4-70
• Clamp: 316L (1.4404)
5.1 Dummy plug, cable gland, adapter or plug
(depending on the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on
the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6
Dummy plug or M12 socket (depending on the
device version)
• Dummy plug : Nickel-plated brass (CuZn)
• M12 socket: Nickel-plated GD-Zn
7
Pressure relief stopper
Nickel-plated brass (CuZn)
Micropilot FMR51, FMR52
Endress+Hauser
77
Nr. Part
Material
8
Ground terminal
• Screw: A2
• Spring washer: A2
• Clamp: 304 (1.4301)
• Holder: 304 (1.4301)
9
Adhesive nameplate
Plastic
1)
For the version with M12 plug the sealing material is Viton.
2)
For the version with 7/8" plug, the sealing material is NBR.
Materials: Antenna and
process connection
FMR51
A
E
F
G
H
D
B
C
1
7
7
3
2
2
5
2
4
3
6
3
3
6
2
2
7
A0018961
A
Version with screw-in adapter
B
XT version with threaded connection
C
XT version with flange
D
HT version with flange
E
Version with flange and antenna extension > 100 mm (3.94 in) to 500 mm (19.7 in)
F
Version with threaded connection and antenna extension > 100 mm (3.94 in) to 500 mm (19.7 in)
G
Version with clamp adapter and antenna extension 100 mm (3.94 in)
H
Version with flange and antenna extension 100 mm (3.94 in)
Pos.
Part
Material
1
Process adapter
316L (1.4404)
Locking washer
2
Horn
316L (1.4404)
AlloyC22 (2.4602)
Screws
A4
AlloyC22 (2.4602)
Spring-lock washer (compact version)
A4
without spring-lock
washer
Micropilot FMR51, FMR52
78
Endress+Hauser
Pos.
Part
Material
Nordlock washer
(extension >100 mm, XT and HT
version)
A4
without Nordlock
washer
Process separation cone
Standard:
PTFE
XT:
PEEK
HT:
Al2O3
Seal
Standard:
• Viton: FKM
• Kalrez: FFKM
(K6375)
XT:
Graphite
HT:
Graphite
3
Flange
316L (1.4404/1.4435) optionally AlloyC22 (2.4602) clad
4
Tube extension + adapter
316L (1.4404)
5
Extension process adapter
316L (1.4404)
Locking washer
6
Process adapter
316L (1.4404)
7
Housing adapter
316L (1.4404)
Temperature reduction (XT)
316L (1.4404)
Intermediate adapter (HT)
316L (1.4404)
Process adapter
316L (1.4404)
FMR52
A
B
C
1
2
4
4
4
2
3
1
1
A0018951
A
Dairy coupling
B
Clamp adapter
C
Flange version
Pos.
Part
Material
1
Adapter
304 (1.4301)
2
Process connection
316L (1.4435)
3
Flange
316L (1.4404/1.4435)
4
Cladding
PTFE
(in accordance with FDA 21 CFR 177.1550 and USP <88>
Class VI)
Micropilot FMR51, FMR52
Endress+Hauser
79
Materials: Weather
protection cover
1
2
3
4
5
7
6
A0015473
No
Part: Material
1
Protection cover: 316L (1.4404)
2
Molded rubber part (4x): EPDM
3
Clamping screw: 316L (1.4404) + carbon fibre
4
Bracket: 316L (1.4404)
5
• Cheese head screw: A4-70
• Nut: A4
• Spring washer: A4
6
Ground terminal
• Screw: A4
• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)
7
• Washer: A4
• Cheese head screw: A4-70
Micropilot FMR51, FMR52
80
Endress+Hauser
Operability
Operating concept
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Operating languages
• English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vietnamese)
• čeština (Czech)
Feature 500 of the product structure determines which of these languages is preset on delivery.
Quick and save commissioning
• Interactive wizard with graphical interface for easy commissioning via FieldCare/DeviceCare
• Menu guidance with brief explanations of the individual parameter functions
• Standardized operation at the device and in the operating tools
Integrated data storage device (HistoROM)
• Enables transfer of configuration when changing electronic modules
• Records up to 100 event messages in the device
• Records up to 1000 measured values in the device
• Saves the signal curve on commissioning which can later be used as a reference.
Efficient diagnostics increase measurement reliability
• Remedy information is integrated in plain text
• Diverse simulation options and line recorder functions
Micropilot FMR51, FMR52
Endress+Hauser
81
Local operation
Operation
with
Pushbuttons
Touch Control
Order code for
"Display;
Operation"
Option C "SD02"
Option E "SD03"
A0032219
A0032221
Display
elements
4-line display
4-line display
white background lighting; switches to red in
event of device error
Format for displaying measured variables and status variables can be individually configured
Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
Operating
elements
local operation with 3 push buttons (, ,
)
external operation via touch control; 3 optical
keys: , ,
Operating elements also accessible in various hazardous areas
Additional
functionality
Data backup function
The device configuration can be saved in the display module.
Data comparison function
The device configuration saved in the display module can be compared to the current device
configuration.
Data transfer function
The transmitter configuration can be transmitted to another device using the display module.
Operation with remote
display and operating
module FHX50
1
2
A0032215
30
FHX50 operating options
1
Housing of the remote display and operating module FHX50
2
Display and operating module SD02, push buttons; cover must be removed
3
Display and operating module SD03, optical keys; can be operated through the glass of the cover
Micropilot FMR51, FMR52
82
Endress+Hauser
Remote operation
Via HART protocol
1
4
5
7
9
6
8
2
3
A0028746
31
Options for remote operation via HART protocol
1
PLC (programmable logic controller)
2
Transmitter power supply unit, e.g. RN221N (with communication resistor)
3
Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475
4
Field Communicator 475
5
Computer with operating tool (e.g. DeviceCare/FieldCare , AMS Device Manager, SIMATIC PDM)
6
Commubox FXA191 (RS232) or FXA195 (USB)
7
Field Xpert SFX350/SFX370
8
VIATOR Bluetooth modem with connecting cable
9
Transmitter
Via PROFIBUS PA protocol
ENDRESS + HAUSER
T
PROFIBUS DP
PROFIBUS PA
1
2
3
5
4
4
4
A0015775
1
Segment coupler
2
Computer with Profiboard/Proficard and operating tool (e.g. DeviceCare/FieldCare
3
PLC (Progrommable Logic Controller)
4
Transmitter
5
Additional functions (valves etc.)
Micropilot FMR51, FMR52
Endress+Hauser
83
Via FOUNDATION Fieldbus
FF-H1
FF-H1
BT
BT
PS
3
4
PS
BT
SB
BT
FF-HSE
IN
LD
ENDRESS + HAUSER
E
+
–
%
1
1
2
2
A0017188
32
FOUNDATION Fieldbus system architecture with associated components
1
FFblue Bluetooth modem
2
Field Xpert SFX350/SFX370
3
DeviceCare/FieldCare
4
NI-FF interface card
IN
Industrial network
FF-HSE
High Speed Ethernet
FF-H1
FOUNDATION Fieldbus-H1
LD
Linking Device FF-HSE/FF-H1
PS
Bus Power Supply
SB
Safety Barrier
BT
Bus Terminator
Micropilot FMR51, FMR52
84
Endress+Hauser
Via service interface (CDI)
1
2
3
A0032466
1
Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface)
2
Commubox FXA291
3
Computer with DeviceCare/FieldCare operating tool
Micropilot FMR51, FMR52
Endress+Hauser
85
Integration in tank gauging
system
The Endress+Hauser Tank Side Monitor NRF590 provides integrated communications for sites with
multiple tanks, each with one or more sensors on the tank, such as radar, spot or average
temperature, capacitive probe for water detection and/or pressure sensors. Multiple protocols out of
the Tank Side Monitor guarantee connectivity to nearly any of the existing industry standard tank
gauging protocols. Optional connectivity of analog 4...20 mA sensors, digital I/O and analog output
simplify full tank sensor integration. Use of the proven concept of the intrinsically safe HART bus for
all on-tank sensors yields extremely low wiring costs, while at the same time providing maximum
safety, reliability and data availability.
NXA820
16...253 VAC
1
2
3
4
6
7
8
5
A0017982
33
The complete measuring system consists of:
1
Tankvision workstation
2
Commubox FXA195 (USB) - optional
3
Computer with operating tool (ControlCare) - optional
4
Level measuring device
5
Temperature measuring device
6
Tank Side Monitor NRF590
7
Pressure measuring device
8
Tankvision Tank Scanner NXA820
Micropilot FMR51, FMR52
86
Endress+Hauser
SupplyCare inventory
management software
SupplyCare is a web-based operating program for coordinating the flow of material and information
along the supply chain. SupplyCare provides a comprehensive overview of the levels of
geographically distributed tanks and silos, for instance, providing complete transparency over the
current inventory situation, regardless of time and location.
Based on the measuring and transmission technology installed onsite, the current inventory data are
collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide
additional security for material requirements planning.
The main functions of SupplyCare:
Inventory visualization
SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current
and historical inventory data and calculated forecasts of future demand. The overview page can be
configured to suit the user's preferences.
Master data management
With SupplyCare you can create and manage the master data for locations, companies, tanks,
products and users, as well as user authorization.
Report Configurator
The Report Configurator can be used to create personalized reports quickly and easily. The reports
can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be
transmitted in many ways, such as by http, ftp or e-mail.
Event management
Events, such as when levels drop below the safety stock level or plan points, are indicated by the
software. In addition, SupplyCare can also notify pre-defined users by e-mail.
Alarms
If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to
the System Administrator and the Local System Administrator.
Delivery planning
The integrated delivery planning function automatically generates an order proposal if a pre-set
minimum inventory level is undershot. Scheduled deliveries and disposals are monitored
continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not
going to be met as planned.
Analysis
In the Analysis module, the most important indicators for the inflow and outflow of the individual
tanks are calculated and displayed as data and charts. Key indicators of material management are
automatically calculated and form the basis for optimizing the delivery and storage process.
Geographical visualization
All the tanks and the tank inventories are represented graphically on a map (based on Google Maps).
The tanks and inventory situations can be filtered by tank group, product, supplier or location.
Multi-language support
The multi-language user interface supports 9 languages, thereby enabling global collaboration on a
single platform. The language and settings are recognized automatically using the browser settings.
SupplyCare Enterprise
SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server
in an Apache Tomcat environment. The operators and administrators operate the application via a
Web browser from their workstations.
Micropilot FMR51, FMR52
Endress+Hauser
87
6
.
10
9
1
3
4
6
5
8
7
2
A0034288
34
Example of inventory management platform with SupplyCare Enterprise SCE30B
1
SupplyCare Enterprise (via Web browser)
2
SupplyCare Enterprise installation
3
SupplyCare Enterprise on mobile devices (via Web browser)
4
Ethernet/WLAN/UMTS
5
Fieldgate FXA42
6
Power supply 24 V DC
7
Modbus TCP via Ethernet as server/client
8
Converter from Modbus to HART Multidrop
9
HART Multidrop
10
4 x 4 to 20 mA analog input (2-wire/4-wire)
Micropilot FMR51, FMR52
88
Endress+Hauser
Cloud-based application: SupplyCare Hosting
SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is
installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress
+Hauser portal.
6
.
10
9
1
3
4
6
5
8
7
2
WWW
A0034289
35
Example of inventory management platform with SupplyCare Hosting SCH30
1
SupplyCare Hosting installation in Endress+Hauser data center
2
PC workstation with Internet connection
3
Warehouse locations with Internet connection via 2G/3G with FXA42 or FXA30
4
Warehouse locations with Internet connection with FXA42
5
Fieldgate FXA42
6
Power supply 24 V DC
7
Modbus TCP via Ethernet as server/client
8
Converter from Modbus to HART Multidrop
9
HART Multidrop
10
4 x 4 to 20 mA analog input (2-wire/4-wire)
With SupplyCare Hosting, users do not need to make the initial software purchase or install and run
the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance
the capability of the software in conjunction with the customer. The hosted version of SupplyCare is
thus always up-to-date and can be customized to meet different customer requirements. Other
services are also offered in addition to the IT infrastructure and the software that is installed in a
secure, redundant Endress+Hauser data center. These services include defined availability of the
global Endress+Hauser Service and Support Organization and defined response times in a service
event.
Micropilot FMR51, FMR52
Endress+Hauser
89
Certificates and approvals
Currently available certificates and approvals can be called up via the product configurator.
CE mark
The measuring system meets the legal requirements of the applicable EC guidelines. These are listed
in the corresponding EC Declaration of Conformity together with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
RoHS
The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).
RCM-Tick marking
The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.
A0029561
Ex approval
• ATEX
• IEC Ex
• CSA
• FM
• NEPSI
• KC
• INMETRO
• TIIS (in preparation)
Additional safety instructions must be observed for applications in hazardous areas. They are
contained in the separate "Safety Instructions" (XA) document, which is included in the scope of
delivery. Reference is made to the XA on the nameplate of the device.
Details on the available certificats as well as the associated XAs can be found in the Associated
documentation chapter in the Safety instructions section: → 110.
Dual seal according to
ANSI/ISA 12.27.01
The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the
user to waive the use and save the cost of installing external secondary process seals in the conduit
as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These
instruments comply with the North-American installation practice and provide a very safe and cost-
saving installation for pressurized applications with hazardous fluids.
Further information can be found in the Safety Instructions (XA) of the relevant devices.
Functional Safety
Used for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous or inhomogeneous
redundancy), independently assessed by TÜV Rhineland as per IEC 61508. Other information see
documentation SD01087F: "Functional Safety Manual".
WHG
WHG approval: Z-65.16-524
Sanitary compatibility
FMR52 with flange cladding made of PTFE conforms to FDA 21 CFR 177.1550 and USP <88> Class
VI
Micropilot FMR51, FMR52
90
Endress+Hauser
3A and EHEDG approval with Tri-Clamp and DIN11851 process connection.
To avoid risk of contamination, install according to the “Hygienic Equipment Design Criteria
(HDC)” as stated in the Subgroup Design Principles of the EHEDG, Doc. 8 from April 2004.
Suitable fittings and seals must be used to ensure hygiene-compliant design according to the
specifications of 3-A SSI and EHEDG.
The leak-tight connections can be cleaned with the cleaning methods usually used in this
industry without leaving residues.
The 316L housing (GT18), option B in the order attribute 40, is not suitable for applications
which need a 3A approval.
NACE MR 0175 / ISO 15156
For FMR51 and FMR54:
• The wetted, metallic materials comply with the requirements of NACE MR 0175 / ISO 15156.
• Declaration of conformity: see product structure, feature 580, option JB.
NACE MR 0103
For FMR51 and FMR54:
• The wetted, metallic materials comply with the requirements of NACE MR 0103.
• The certificate of compliance is based on NACE MR 0175.
The hardness and the intercrystalline corrision have been tested, as well as the heat treatment
(solution annealed) has been carried out. The used materials comply therefore with the
requirements of NACE MR 0103.
• Declaration of conformity: see product structure, feature 580, option JE.
Pressure equipment with
allowable pressure
≤ 200 bar (2 900 psi)
Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
pressure.
Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.
Marine certificate
Device
Marine certificate 1)
GL
ABS
LR
BV
DNV
FMR51
FMR52
1)
see oredering feature 590 "Additional approval"
Radio standard
EN302729-1/2
The devices Micropilot FMR50, FMR51, FMR52, FMR56 and FMR57 are conform with the LPR
(Level Probing Radar) standard EN302729-1/2. The devices are allowed to be used inside or outside
of closed bins or tanks in countries of th EU and the EFTA. Precondition is that the country itself
already has implemented the directive.
At present, the following countries have already implemented the directive:
Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, Great Britain, Ireland, Iceland, Italy,
Liechtenstein, Lithuania, Latvia, Malta, Netherland, Norway, Austria, Poland, Romania, Sweden,
Switzerland, Slovakia, Spain, Czech Republic and Cyprus.
All countries not mentioned above are at present in the implementation phase.
For the use of the devices outside of closed bins or tanks, the following has to be observed:
1.
The installation has to be done by trained personnel.
2.
The antenna of the device has to be mounted on a fixed place and vertically to the bottom.
3.
The mounting place has to be in a distance of at least 4 km to the mentioned astronomical
stations or an appropriate permission has to be issued by the national authority. If the device is
mounted at a distance of 4 to 40 km from the stations mentioned, the maximum mounting
height is restricted to 15 m (49 ft).
Micropilot FMR51, FMR52
Endress+Hauser
91
Astronomical stations
Country
Name of the station
Geographical latitude
Geographical longitude
Germany
Effelsberg
50°31'32" N
06°53'00" E
Finland
Metsähovi
60°13'04" N
24°23'37" E
Tuorla
60°24'56" N
24°26'31" E
France
Plateau de Bure
44°38'01" N
05°54'26" E
Floirac
44°50'10" N
00°31'37" W
United Kingdom
Cambridge
52°09'59" N
00°02'20" E
Damhall
53°09'22" N
02°32'03" W
Jodrell Bank
53°14'10" N
02°18'26" W
Knockin
52°47'24" N
02°59'45" W
Pickmere
53°17'18" N
02°26'38" W
Italy
Medicina
44°31'14" N
11°38'49" E
Noto
36°52'34" N
14°59'21" E
Sardinia
39°29'50" N
09°14'40" E
Poland
Krakow Fort Skala
50°03'18" N
19°49'36" E
Russia
Dmitrov
56°26'00" N
37°27'00" E
Kalyazin
57°13'22" N
37°54'01" E
Pushchino
54°49'00" N
37°40'00" E
Zelenchukskaya
43°49'53" N
41°35'32" E
Sweden
Onsala
57°23'45" N
11°55'35" E
Switzerland
Bleien
47°20'26" N
08°06'44" E
Spain
Yebes
40°31'27" N
03°05'22" W
Robledo
40°25'38" N
04°14'57" W
Hungary
Penc
47°47'22" N
19°16'53" E
In general the requirements of EN 302729-1/2 have to be taken in consideration.
Radio standard
EN302372-1/2
The devices Micropilot FMR50, FMR51, FMR52, FMR53, FMR54, FMR56 and FMR57 are conform
with the TLPR (Tanks Level Probing Radar) standard EN302372-1/2 and can always be used in
closed tanks or bins. For installation, points a to f in Annex B of EN302372-1 have to be taken into
account.
FCC / Industry Canada
This device complies with Part 15 of the FCC rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
Canada CNR-Gen Section 7.1.3
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to
the following two conditions: (1) This device may not interference, and (2) this device must accept
any interference, including interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts
de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas
produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique
subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
[Any] changes or modifications not expressly approved by the party responsible for compliance could
void the user's authority to operate the equipment.
Micropilot FMR51, FMR52
92
Endress+Hauser
In addition, the devices FMR50 8), FMR51 9), FMR52 10), FMR56 and FMR57 are compliant with the
LPR (Level probe radar) regulation also for free space applicactions according to the FCC Code of
Federal Regulations, CFR 47, Part 15, Sections 15.205, 15.207, 15.209, 15.256 for antenna sizes
bigger than 50 mm (2.0 in) 11). For these applications the devices must be professionally installed in
a downward operating position. In addition, the devices are not allowed to be mounted in a zone of
4 km around RAS stations and within a radius of 40 km around RAS stations the maxium operation
height of devices is 15 m (49 ft) above ground.
Japanese radio approval
The devices FMR50, FMR51, FMR52, FMR54 and FMR57 comply with the Japanese Radio Law,
Article 6, Section 1(1).
CRN approval
Some device versions have a CRN approval. Devices are CRN approved if the folowing two conditions
are met:
• The device has a CSA approval (Product structure: Feature 010 "Approval")
• The device has a CRN approved process connection according to the following table.
Feature 100 of the product structure
Process connection
AFJ
NPS 2" Cl.150 RF, 316/316L
AFK
NPS 2" Cl.150, PTFE >316/316L
AFM
NPS 2" Cl.150, AlloyC >316/316L
AGJ
NPS 3" Cl.150 RF, 316/316L
AGK
NPS 3" Cl.150, PTFE >316/316L
AGM
NPS 3" Cl.150, AlloyC >316/316L
AHJ
NPS 4" Cl.150 RF, 316/316L
AHK
NPS 4" Cl.150, PTFE >316/316L
AHM
NPS 4" Cl.150, AlloyC >316/316L
AJJ
NPS 6" Cl.150 RF, 316/316L
AJK
NPS 6" Cl.150, PTFE >316/316L
AJM
NPS 6" Cl.150, AlloyC >316/316L
ARJ
NPS 2" Cl.300 RF, 316/316L
ARM
NPS 2" Cl.300, AlloyC >316/316L
ASJ
NPS 3" Cl.300 RF, 316/316L
ASM
NPS 3" Cl.300, AlloyC >316/316L
ATJ
NPS 4" Cl.300 RF, 316/316L
ATM
NPS 4" Cl.300, AlloyC >316/316L
MRK
DIN11851 DN50 PN25 slotted nut, PTFE>316L
MTK
DIN11851 DN80 PN25 slotted nut, PTFE>316L
RGJ
Thread ANSI MNPT1-1/2, 316L
RVJ
Thread EN10226 R1-1/2, 316L
TDJ
Tri-Clamp ISO2852 DN40-51 (2"), 316L
TDK
Tri-Clamp ISO2852 DN40-51 (2"), PTFE>316L, 3A, EHEDG
TFJ
Tri-Clamp ISO2852 DN70-76.1 (3"), 316L
8)
Except of FMR50-######BM* (Horn 40mm/1-1/2", PVDF encapsulated)
9)
Except for FMR51-######BA* (Horn 40mm/1-1/2") and FMR51-######BB* (Horn 50mm/2")
10)
Except for FMR52-######BO* (Horn 50mm/2", flush mount)
11)
Micropilot FMR51, FMR52
Endress+Hauser
93
Feature 100 of the product structure
Process connection
TFK
Tri-Clamp ISO2852 DN70-76.1 (3"), PTFE>316L, 3A, EHEDG
THK
Tri-Clamp ISO2852 DN101.6 (4"), PTFE>316L, 3A, EHEDG
• Process connections without CRN approval are not included in this table.
• Refer to the product structure to see which process connections are available for a specific
device type.
• For some process connections not listed in the product structure a CRN approval is available
on request.
• CRN approved devices are marked with the registration number OF15872.5C on the
nameplate.
For device versions listed in the table below, the maximum allowed pressure is reduced if they
have a CRN approval. For device versions not listed in the table below the pressure range as
indicated in the "Process" chapter → 60 is not affected by the CRN approval.
Product
Antenna 1)
Process connection 2)
Seal 3)
max. pressure
FMR52
TFK: Tri-Clamp 3"
10 bar (145 psi)
THK: Tri-Clamp 4"
10 bar (145 psi)
FMR51
D2: Graphite (HT)
100 bar (1 450 psi)
1)
Feature 070 of the product structure
2)
Feature 100 of the product structure
3)
Feature 090 of the product structure
Track record
FMR5x is the upgrade model of the corresponding FMR2xx series.
Micropilot FMR51, FMR52
94
Endress+Hauser
Test, Certificate
Feature 580
"Test,
Certificate"
Designation
Available for
JA
3.1 Material certificate, wetted metallic parts, EN10204-3.1 inspection
certificate
FMR51
JB
Conformity to NACE MR0175, wetted metallic parts
FMR51
JD
3.1 Material certificate, pressurized parts, EN10204-3.1 inspection
certificate
FMR52
JE
Conformity to NACE MR0103, wetted metallic parts
FMR51
JF
Conformity to AD2000, wetted metallic parts:
Material of all wetted/pressurized parts conform to AD2000 (Technical
rules W2, W9, W10)
• FMR51
• FMR52
KD
Helium leak test, internal procedure, inspection certificate
FMR51
KE
Pressure test, internal procedure, inspection certificate
• FMR51
• FMR52
KG
3.1 Material certificate+PMI test (XRF), internal procedure, wetted metallic
parts, EN10204-3.1 inspection certificate
• FMR51
• FMR52
KP
Liquid penetrant test AD2000-HP5-3(PT), wetted/pressurized metallic
parts, inspection certificate
FMR51
KQ
Liquid penetrant test ISO23277-1 (PT), wetted/pressurized metallic parts,
inspection certificate
FMR51
KR
Liquid penetrant test ASME VIII-1 (PT), wetted/pressurized metallic parts,
inspection certificate
FMR51
KS
Welding documentation, wetted/pressurized seams
consisting of
• Drawing
• WPQR (Welding Procedure Qualification Report)
• WPS (Welding Process Specification)
• Manufacturer Declaration
FMR51
KV
Conformity to ASME B31.3:
The dimensions, materials of construction, pressure / temperature ratings
and identification markings of the device comply with the requirements of
ASME B31.3
• FMR51
• FMR52
Test reports, declarations and material test certificates can be obtained electronically from the
W@M Device Viewer :
Enter the serial number from the nameplate (www.endress.com/deviceviewer)
This applies to the options of the following ordering features:
• 550 "Calibration"
• 580 "Test, Certificate"
• 590 "Additional Approval", option LW: "CoC-ASME BPE"
Product documentation on
paper
Test reports, declarations and material certificates can be ordered as hardcopies via ordering feature
570 "Service", option I7 „Product documentation on paper“. The printouts will be enclosed in the
product delivery.
Micropilot FMR51, FMR52
Endress+Hauser
95
Other standards and
guidelines
• EN 60529
Degrees of protection by housing (IP code)
• EN 61010-1
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures.
• IEC/EN 61326
"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC
requirements)
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 107
Status classification as per NE107
• NAMUR NE 131
Requirements for field devices for standard applications
• IEC61508
Functional safety of electrical/electronic/programmable electronic safety-related systems
Micropilot FMR51, FMR52
96
Endress+Hauser
Ordering information
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Micropilot FMR51, FMR52
Endress+Hauser
97
3-point linearity protocol
The following notes must be taken into account if option F3 ("3 point linearity protocol") has
been selected in feature 550 ("Calibration").
The 3 points of the linearity protocol are defined as follows:
R
A
500 (19.7)
6000 (236)
1
3
2
A0023272
36
Points of the 3-point linearity protocol; Dimensions: mm (in)
A
Distance from reference point R to first measuring point
R
Reference point of the measurement
1
First measuring point
2
Second measuring point (centrally between first and third measuring point)
3
Third measuring point
Measuring point
Position
1st measuring point
• At the distance A from the reference point
• A = antenne length + antenna extension (if present) + 500 mm (19.7 in)
• Minimum distance: Amin = 1 000 mm (39.4 in)
2nd measuring point
centrally between 1st and 3rd measuring point
3rd measuring point
6 000 mm (236 in) below the reference point, R
The position of the measuring points may vary by ±1 cm (±0.04 in).
The linearity is checked under reference conditions.
Micropilot FMR51, FMR52
98
Endress+Hauser
5-point linearity protocol
The following notes must be taken into account if option F4 ("5 point linearity protocol") has
been selected in feature 550 ("Calibration").
The five points of the linearity protocol are evenly distributetd across the measuring range (0% to
100%). In order to define the measuring range, Empty calibration (E) and Full calibration (F) have
to be specified 12).
The following restrictions have to be taken into account when defining E and F:
R
A
F
E
A0017983
Device version
Minimum distance
between reference point
(R) and 100% level
Minimum span
Maximum value for
"Empty calibration"
FMR51/FMR52
Without antenna extension
A ≥ antenna length +
200 mm (8 in) 1)
F ≥ 400 mm (16 in)
E ≤ 24 m (79 ft)
FMR51 with 100 mm/4" antenna extension 2)
A ≥ antenna length +
antenna extension
100 mm (4 in) +
200 mm (8 in)
FMR51 with varaible antenna extension 3)
A ≥ antenna length +
maximum antenna
extension
1 000 mm (40 in) 4) +
200 mm (8 in)
1)
Minimum value: A ≥ 400 mm (16 in)
2)
Feature 610 "Accessory Mounted", option OM
3)
Feature 610 "Accessory Mounted", option OU or OV
4)
This value is independent of the actual size of the antenna extension.
The linearity is checked under reference conditions.
The selected values of Empty calibration and Full calibration are only used to record the
linearity protocol and are reset to their probe specific default values thereafter. If values
different from the default are required, they must be ordered as a customized parametrization
→ 99.
12)
If E and F are not specified, probe dependent default values will be used instead.
Micropilot FMR51, FMR52
Endress+Hauser
99
Customized parametrization
If the option IJ "Customized parametrization HART", IK "Customized parametrization PA" or IL
"Customized parametrization FF" has been selected in feature 570 "Service", customer specific
presettings can be selected for the following parameters:
Parameter
Communication
Selection list / range of
values
Setup → Distance unit
• HART
• PA
• FF
• in
• ft
• mm
• m
Setup → Empty calibration
• HART
• PA
• FF
max. 70 m (230 ft)
Setup → Full calibration
• HART
• PA
• FF
max. < 70 m (230 ft)
Setup → Adv. Setup → Current output 1/2 → Damping
HART
0 to 999.9 s
Setup → Adv. Setup → Current output 1/2 → Failure mode
HART
• Min
• Max
• Last valid value
Setup → Adv. Setup → Current output 1/2 → Burst mode
HART
• Off
• On
Tagging (TAG)
Ordering feature
895: Marking
Option
Z1: Tagging (TAG), see additional spec.
Position of the measuring
point marking
To be selected in the additional specifications:
• Tag plate Stainless Steel
• Self-adhesive paper label
• Supplied label/plate
• RFID TAG
• RFID TAG + Tag plate Stainless Steel
• RFID TAG + Self-adhesive paper label
• RFID TAG + Supplied label/plate
Definition of the measuring
point designation
To be defined in the additional specifications:
3 lines containing up to 18 characters each
The measuring point designation appears on the selected label and/or the
RFID TAG.
Designation in the Electronic
Name Plate (ENP)
The first 32 characters of the measuring point designation
Designation on the display
module
The first 12 characters of the measuring point designation
Services
The following services can be selected via the product structure in the Product Configurator 13):
• PWIS free (PWIS: paint wetting impairment substances)
• Customized parametrization HART → 99
• Customized parametrization PA→ 99
• Customized parametrization FF → 99
• Without tooling DVD (FieldCare)
13)
Feature 570 of the product structure
Micropilot FMR51, FMR52
100
Endress+Hauser
Application Packages
Heartbeat Diagnostics
Availability
Available in all device versions.
Function
• Continuous self-monitoring of the device.
• Diagnostic messages output to
– the local display.
– an asset management system (e.g. FieldCare/DeviceCare).
– an automation system (e.g. PLC).
Advantages
• Device condition information is available immediately and processed in time.
• The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation
NE 107 and contain information about the cause of the error and remedial action.
Detailed description