Endress+Hauser Micropilot FMR51 Technical Information

Text of the manual

Level measurement in liquids

Application • Continuous, non-contact level measurement of liquids, pastes and slurries • Horn antenna (FMR51); flush mounted, completely filled PTFE horn antenna (FMR52) • Maximum measuring range: 70 m (230 ft) • Temperature: –196 to +450 °C (–321 to +842 °F) • Pressure: –1 to +160 bar (–14.5 to +2 320 psi) • Accuracy: ± 2 mm • International explosion protection certificates; WHG; marine approvals • Linearity protocol (3-point, 5-point) Your benefits • Reliable measurement even for changing product and process conditions • HistoROM data management for easy commissioning, maintenance and diagnostics • Highest reliability due to Multi-Echo Tracking • SIL2 according to IEC 61508, SIL3 in case of homogeneous or heterogeneous redundancy • Seamless integration into control or asset management systems • Intuitive user interface in national languages • Easy proof test for SIL and WHG • Heartbeat Technology™ Products Solutions Services Technical Information Micropilot FMR51, FMR52 Free space radar TI01040F/00/EN/08.17 71375707

Micropilot FMR51, FMR52

2 Endress+Hauser Table of contents Important document information . . . . . . . . . . . . . . . 4 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Function and system design . . . . . . . . . . . . . . . . . . . 6 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Operating frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Transmitting power . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Device plug connectors . . . . . . . . . . . . . . . . . . . . . . . . . 27 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . 31 Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 32 Performance characteristics . . . . . . . . . . . . . . . . . . 33 Reference operating conditions . . . . . . . . . . . . . . . . . . . 33 Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . 33 Measured value resolution . . . . . . . . . . . . . . . . . . . . . . 34 Reaction time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Influence of ambient temperature . . . . . . . . . . . . . . . . . 34 Influence of gas layer . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Gas phase compensation with external pressure sensor (PROFIBUS PA, FOUNDATION Fieldbus) . . . . . . . . . . . . . 35 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . 36 Measuring conditions . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Mounting cladded flanges . . . . . . . . . . . . . . . . . . . . . . . 41 Installation in vessel (free space) . . . . . . . . . . . . . . . . . . 41 Installation in stilling well . . . . . . . . . . . . . . . . . . . . . . . 45 Installation in bypass . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Vessels with heat insulation . . . . . . . . . . . . . . . . . . . . . 49 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 50 Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . 50 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Altitude according to IEC61010-1 Ed.3 . . . . . . . . . . . . . . 58 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Cleaning the antenna . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . 59 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 Process temperature, Process pressure . . . . . . . . . . . . . . 60 Dielectric constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62 Mechanical construction . . . . . . . . . . . . . . . . . . . . 63 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 Materials: GT18 housing (stainless steel, corrosion- resistant) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . . 75 Materials: GT20 housing (die-cast aluminum, powder- coated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Materials: Antenna and process connection . . . . . . . . . . . 77 Materials: Weather protection cover . . . . . . . . . . . . . . . . 79 Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Operation with remote display and operating module FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Integration in tank gauging system . . . . . . . . . . . . . . . . . 85 SupplyCare inventory management software . . . . . . . . . . 86 Certificates and approvals . . . . . . . . . . . . . . . . . . . 89 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 RCM-Tick marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . . 89 Functional Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 WHG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . 89 NACE MR 0175 / ISO 15156 . . . . . . . . . . . . . . . . . . . . . 90 NACE MR 0103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Pressure equipment with allowable pressure ≤ 200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Marine certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Radio standard EN302729-1/2 . . . . . . . . . . . . . . . . . . . 90 Radio standard EN302372-1/2 . . . . . . . . . . . . . . . . . . . 91 FCC / Industry Canada . . . . . . . . . . . . . . . . . . . . . . . . . 91 Japanese radio approval . . . . . . . . . . . . . . . . . . . . . . . . 92 CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Track record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 Test, Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Product documentation on paper . . . . . . . . . . . . . . . . . . 94 Other standards and guidelines . . . . . . . . . . . . . . . . . . . 95 Ordering information . . . . . . . . . . . . . . . . . . . . . . . 96 Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 96 3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . . 97 5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . . 98 Customized parametrization . . . . . . . . . . . . . . . . . . . . . 99

Micropilot FMR51, FMR52

Endress+Hauser 3 Tagging (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Application Packages . . . . . . . . . . . . . . . . . . . . . . 100 Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . 100 Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . . 101 Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 102 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103 Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 103 Communication-specific accessories . . . . . . . . . . . . . . . 108 Service-specific accessories . . . . . . . . . . . . . . . . . . . . . 109 System components . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . 110 Standard documentation . . . . . . . . . . . . . . . . . . . . . . . 110 Supplementary documentation . . . . . . . . . . . . . . . . . . . 110 Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . 110 Registered trademarks . . . . . . . . . . . . . . . . . . . . . 113 Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113

Micropilot FMR51, FMR52

4 Endress+Hauser Important document information Symbols Safety symbols Symbol Meaning DANGER DANGER! This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury. CAUTION CAUTION! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury. NOTICE NOTE! This symbol contains information on procedures and other facts which do not result in personal injury. Electrical symbols Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. Protective Earth (PE) A terminal which must be connected to ground prior to establishing any other connections. The ground terminals are situated inside and outside the device: • Inner ground terminal: Connects the protectiv earth to the mains supply. • Outer ground terminal: Connects the device to the plant grounding system. Symbols for certain types of information Symbol Meaning Permitted Procedures, processes or actions that are permitted. Preferred Procedures, processes or actions that are preferred. Forbidden Procedures, processes or actions that are forbidden. Tip Indicates additional information. Reference to documentation. A Reference to page. Reference to graphic. Visual inspection.

Micropilot FMR51, FMR52

Endress+Hauser 5 Symbols in graphics Symbol Meaning 1, 2, 3 ... Item numbers 1., 2. , 3.… Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections - Hazardous area Indicates a hazardous area. . Safe area (non-hazardous area) Indicates the non-hazardous area. Symbols at the device Symbol Meaning Safety instructions Observe the safety instructions contained in the associated Operating Instructions. Temperature resistance of the connection cables Specifies the minimum value of the temperature resistance of the connection cables.

Micropilot FMR51, FMR52

6 Endress+Hauser Function and system design Measuring principle The Micropilot is a "downward-looking" measuring system, operating based on the time-of-flight method (ToF). It measures the distance from the reference point (process connection) to the product surface. Radar impulses are emitted by an antenna, reflected off the product surface and received again by the radar system. F L D E R 0% 100% A0017871  1 Setup parameters of the Micropilot R Reference point of the measurement (lower edge of the flange or threaded connection) E Empty calibration ( = zero) F Full calibration (= span) D Measured distance L Level (L = E - D) Input The reflected radar impulses are received by the antenna and transmitted into the electronics. A microprocessor evaluates the signal and identifies the level echo caused by the reflection of the radar impulse at the product surface. The unambiguous signal identification is accomplished by the PulseMaster® eXact software together with the Multi-echo tracking algorithms, based on many years of experience with time-of-flight technology. The distance D to the product surface is proportional to the time of flight t of the impulse: D = c · t/2, with c being the speed of light. Based on the known empty distance E, the level L is calculated: L = E – D The reference point R of the measurement is located at the process connection. For details see the dimensional drawing: • FMR51: →  64 • FMR52: →  70 The Micropilot is equipped with functions to suppress interference echoes. The user can activate these functions. Together with the multi-echo tracking algorithms they ensure that interference echoes (i.e. from edges and weld seams) are not interpreted as level echo.

Micropilot FMR51, FMR52

Endress+Hauser 7 Output The Micropilot is commissioned by entering an empty distance "E" (=zero), a full distance "F" (=span) and application parameters which automatically adapt the instrument to the process conditions. For models with a current output, the factory adjustment for zero point "E" and span "F" is 4 mA and 20 mA. For digital outputs and the display module, the factory adjustment for zero point "E" and span "F" is 0 % and 100 %. A linearization with max. 32 points, based on a table entered either manually or semi-automatically, can be activated locally or remotely. This function provides a measurement in engineering units and a linear output signal for spheres, horizontal cylindrical tanks and vessels with conical outlet. Life cycle of the product Engineering • Universal measuring principle • Measurement unaffected by medium properties • Hardware and software developed according to SIL IEC 61508 Procurement • Endress+Hauser being the world market leader in level measurement guarantees asset protection • Worldwide support and service Installation • Special tools are not required • Reverse polarity protection • Modern, detachable terminals • Main electronics protected by a separate connection compartment Commissioning • Fast, menu-guided commissioning in only a few steps on site or from the control room • Plain text display in national languages reduces the risk of error or confusion • Direct local access of all parameters • Short instruction manual at the device Operation • Multi-echo tracking: Reliable measurement through self-learning echo-search algorithms taking into account the short-term and long-term history in order to check the found echoes for plausibility and to suppress interference echoes. • Diagnostics in accordance with NAMUR NE107 Maintenance • HistoROM: Data backup for instrument settings and measured values • Exact instrument and process diagnosis to assist fast decisions with clear details concerning remedies • Intuitive, menu-guided operating concept in national languages saves costs for training, maintenance and operation • Cover of the electronics compartment can be opened in hazardous areas Retirement • Order code translation for subsequent models • RoHS-conforming (Restriction of certain Hazardous Substances), unleaded soldering of electronic components • Environmentally sound recycling concept

Micropilot FMR51, FMR52

8 Endress+Hauser Input Measured variable The measured variable is the distance between the reference point and the product surface. The level is calculated from this distance, taking into account the empty distance "E" entered by the user. If required, the level can be converted into other variables (volume, mass) by means of a linearization (up to 32 points). Measuring range Maximum measuring range Device Maximum measuring range FMR51 - standard version 40 m (131 ft) FMR51 - with "Advanced dynamics" application package 70 m (230 ft) FMR52 - standard version 40 m (131 ft) FMR52 - with "Advanced dynamics" application package 60 m (197 ft) Usable measuring range The usable measuring range depends on the size of the antenna, the reflectivity of the medium, the mounting location and eventual interference reflections. The following tables describe the groups of media as well as the achievable measuring range as a function of application and media group. If the dielectric constant of a medium is unknown, it is recommended to assume media group B to ensure a reliable measurement. Media groups Media groups DC (εr) Example A 1.4 to 1.9 non-conducting liquids, e.g. liquefied gas 1) B 1.9 to 4 non-conducting liquids, e.g. benzene, oil, toluene, … C 4 to 10 e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, … D > 10 conducting liquids, e.g. aqueous solutions, dilute acids and alkalis 1) Treat Ammonia NH3 as a medium of group A. For dielectric constants (DC values) of many media commonly used in various industries refer to: • the Endress+Hauser DC manual (CP01076F) • the Endress+Hauser "DC Values App" (available for Android and iOS)

Micropilot FMR51, FMR52

Endress+Hauser 9 Device Storage tank A0018833 Calm product surface (e.g. intermittent filling, filling from bottom, immersion tubes) Antenna size FMR51 40 mm (1½ in) 50 mm (2 in) 80 mm (3 in) 100 mm (4 in) FMR52 - 50 mm (2 in) 80 mm (3 in) - A B 25 (82) 15 (49) C D 3 (9.9) 10 (32) 15 (49) 5 (16) 8 (26) 5 (16) A0018858 A B C D 4 (13) 8 (26) 8 (26) 15 (49) 12 (39) 25 (82) 35 (110) 40 (131) A0018859 A B C D 8 (26) 15 (49) 30 (98) 40 (131) 10 (32) 20 (65) 40 (131) 60 (197) A0018860 A B C D 10 (32) 25 (82) 40 (131) 40 (131) 15 (49) 30 (99) 45 (148) 70 (229) A0018862 Measuring range [m (ft)] Legend Measuring range of the standard version Measuring range for the "Advanced dynamics" application package (product structure: feature 540: "Application Package", Option EM: "Advanced dynamics")

Micropilot FMR51, FMR52

10 Endress+Hauser Device Buffer tank A0018835 Moving surfaces (e.g. continuous filling, from above, mixing jets) Antenna size FMR51 40 mm (1½ in) 50 mm (2 in) 80 mm (3 in) 100 mm (4 in) FMR52 - 50 mm (2 in) 80 mm (3 in) - B C D (6.6) 4 (13) 7.5 (24) 10 (32) (16) 5 2 A0018863 B C D 3 (9.9) 7.5 (24) 10 (32) 5 (16) 10 (32) 15 (49) A0018864 A B C D 2.5 (8) 5 (16) 10 (32) 15 (49) 5 (16) 10 (32) 15 (49) 25 (82) A0018865 A B C D 5 (16) 10 (32) 15 (49) 25 (82) 7.5 (24) 15 (49) 25 (82) 35 (110) A0018866 Measuring range [m (ft)] Legend Measuring range of the standard version Measuring range for the "Advanced dynamics" application package (product structure: feature 540: "Application Package", Option EM: "Advanced dynamics")

Micropilot FMR51, FMR52

Endress+Hauser 11 Device Process tank with agitator A0018837 Turbulent surface (e.g. filling from above, agitators, baffles) Antenna size FMR51 40 mm (1½ in) 50 mm (2 in) 80 mm (3 in) 100 mm (4 in) FMR52 - 50 mm (2 in) 80 mm (3 in) - B C D 1 (3.2) 2 (6.6) 3 (9.8) 5 (16) A0018867 B C D 2 (6.6) 3 (9.8) 5 (16) 7.5 (25) 10 (32) A0018868 B C D 2.5 (8.2) 5 (16) 12 (39) 15 (49) 8 (26) A0018869 B C D 4 (13) (16) 5 8 (26) 15 (49) 20 (65) 10 (32) A0018870 Measuring range [m (ft)] Legend Measuring range of the standard version Measuring range for the "Advanced dynamics" application package (product structure: feature 540: "Application Package", Option EM: "Advanced dynamics")

Micropilot FMR51, FMR52

12 Endress+Hauser Device Stilling well Bypass A0018842 A0018840 Antenna size Antenna size FMR51 40 ... 100 mm ( 1½ ... 4 in) 40 ... 100 mm ( 1½ ... 4 in) FMR52 50 ... 80 mm ( 2 ... 3 in) 50 ... 80 mm ( 2 ... 3 in) 20 (66) A, B, C, D A0018851 20 (66) C, D A0018852 Measuring range [m (ft)]

Micropilot FMR51, FMR52

Endress+Hauser 13 Operating frequency K-band (~ 26 GHz) Up to 8 Micropilot transmitters can be installed in the same tank because the transmitter pulses are statistically coded. Transmitting power Distance Average energy density in beam direction Standard version With "Advanced dynamics" application package 1) 1 m (3.3 ft) < 12 nW/cm2 < 64 nW/cm2 5 m (16 ft) < 0.4 nW/cm2 < 2.5 nW/cm2 1) Product structure, feature 540: "Application package", option EM: "Advanced dynamics"

Micropilot FMR51, FMR52

14 Endress+Hauser Output Output signal HART Signal coding FSK ±0.5 mA over current signal Data transmission rate 1 200 Bit/s Galvanic isolation Yes PROFIBUS PA Signal coding Manchester Bus Powered (MBP) Data transmission rate 31.25 kBit/s, voltage mode Galvanic isolation Yes FOUNDATION Fieldbus Signal coding Manchester Bus Powered (MBP) Data transmission rate 31.25 kBit/s, voltage mode Galvanic isolation Yes Switch output For HART devices, the switch output is available as an option. See product structure, feature 20: "Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output" Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output. Switch output Function Open collector switching output Switching behavior Binary (conductive or non-conductive), switches when the programmable switch point is reached Failure mode non-conductive Electrical connection values U = 16 to 35 VDC, I = 0 to 40 mA Internal resistance RI < 880 Ω The voltage drop at this internal resistance has to be taken into account on planning the configuration. For example, the resulting voltage at a connected relay must be sufficient to switch the relay. Insulation voltage floating, Insulation voltage 1 350 VDC to power supply aund 500 VAC to ground Switch point freely programmable, separately for switch-on and switch-off point Switching delay freely programmable from 0 to 100 s, separately for switch-on and switch-off point Number of switching cycles corresponds to the measuring cycle Signal source device variables • Level linearized • Distance • Terminal voltage • Electronic temperature • Relative echo amplitude • Diagnostic values, Advanced diagnostics Number of switching cycles unlimited

Micropilot FMR51, FMR52

Endress+Hauser 15 Signal on alarm Depending on the interface, failure information is displayed as follows: • Current output (for HART devices) – Failsafe mode selectable (in accordance with NAMUR Recommendation NE 43): Minimum alarm: 3.6 mA Maximum alarm (= factory setting): 22 mA – Failsafe mode with user-selectable value: 3.59 to 22.5 mA • Local display – Status signal (in accordance with NAMUR Recommendation NE 107) – Plain text display • Operating tool via digital communication (HART, PROFIBUS PA, FOUNDATION Fieldbus) or service interface (CDI) – Status signal (in accordance with NAMUR Recommendation NE 107) – Plain text display Linearization The linearization function of the device allows the conversion of the measured value into any unit of length or volume. Linearization tables for calculating the volume in cylindrical tanks are pre- programmed. Other linearization tables of up to 32 value pairs can be entered manually or semi- automatically. Galvanic isolation All circuits for the outputs are galvanically isolated from each other. Protocol-specific data HART Manufacturer ID 17 (0x11) Device type ID 0x1128 HART specification 7.0 Device description files (DTM, DD) Information and files under: • www.endress.com • www.fieldcommgroup.org HART load min. 250 Ω HART device variables The measured values can be freely assigned to the device variables. Measured values for PV (primary variable) • Level linearized • Distance • Electronic temperature • Relative echo amplitude • Area of incoupling • Analog output adv. diagnostics 1 • Analog output adv. diagnostics 2 Measured values for SV, TV, FV (second, third and fourth variable) • Level linearized • Distance • Electronic temperature • Terminal voltage • Relative echo amplitude • Absolute echo amplitude • Area of incoupling • Analog output adv. diagnostics 1 • Analog output adv. diagnostics 2 Supported functions • Burst mode • Additional transmitter status Wireless HART data Minimum start-up voltage 16 V Start-up current 3.6 mA Start-up time 65 s Minimum operating voltage 14.0 V

Micropilot FMR51, FMR52

16 Endress+Hauser Multidrop current 4.0 mA Set-up time 15 s PROFIBUS PA Manufacturer ID 17 (0x11) Ident number 0x1559 Profile version 3.02 GSD file Information and files under: • www.endress.com • www.profibus.org GSD file version Output values Analog Input: • Level linearized • Distance • Terminal voltage • Electronic temperature • Absolute echo amplitude • Relative echo amplitude • Analog output adv. diagnostics 1 • Analog output adv. diagnostics 2 Digital Input: • Digital output AD 1 • Digital output AD 2 • Switch output Input values Analog Output: • Analog value from PLC (for sensor block external pressure to compensate gas phase effects) • Analog value from PLC to be indicated on the display Digital Output: • Extended diagnostic block • Level limiter • Sensor block measurement on • Sensor block save history on • Status output Supported functions • Identification & Maintenance Einfachste Geräteidentifizierung seitens des Leitsystems und des Typenschildes • Automatic Ident Number Adoption GSD compatibility mode with respect to the preceding product Micropilot M FMR2xx • Physical Layer Diagnostics Installation check of the PRFIBUS segment and the Micropilot FMR5x via the terminal voltage and telegram surveillance. • PROFIBUS Up-/Download Up to 10 times faster writing and reading of parameters via PROFIBUS up-/download • Condensed Status Simple and self-explanatory diagnostic information by categorization of occurring diagnostic messages. FOUNDATION Fieldbus Manufacturer ID 0x452B48 Device type 0x1028 Device Revision 0x01 DD Revision Information and files can be found: • www.endress.com • www.fieldcommgroup.org CFF Revision Device Tester Version (ITK Version) 6.0.1 ITK Test Campaign Number IT085300 Link Master (LAS) capable yes

Micropilot FMR51, FMR52

Endress+Hauser 17 Link Master / Basic Device selectable yes; default: Basic Device Node address Default: 247 (0xF7) Features supported Following methods are supported: • Restart • ENP Restart • Setup • Linearization • Self Check Virtual Communication Relationships (VCRs) Number of VCRs 44 Number of Link Objects in VFD 50 Permanent entries 1 Client VCRs 0 Server VCRs 10 Source VCRs 43 Sink VCRs 0 Subscriber VCRs 43 Publisher VCRs 43 Device Link Capabilities Slot time 4 Min. inter PDU delay 8 Max. response delay 20 Transducer Blocks Block Content Output values Setup Transducer Block Contains all parameters for a standard commissioning procedure • Level or volume 1) (Channel 1) • Distance (Channel 2) Advanced Setup Transducer Block Contains all parameters for a more detailed configuration of the device no output values Display Transducer Block Contains all parameters for the configuration of the display module no output values Diagnostic Transducer Block Contains diagnostic information no output values Advanced Diagnostic Transducer Block Contains parameters for the Advanced Diagnostic no output values Expert Configuration Transducer Block Contains parameters which require detailed knowledge of the functionalities of the device no output values Expert Information Transducer Block Contains information about the state of the device no output values Service Sensor Transducer Block Contains parameters which can only be operated by Endress+Hauser service personnel no output values Service Information Transducer Block Contains information on the state of device which is relevant for service operations no output values Data Transfer Transducer Block Contains parameters which allow to backup the device configuration in the display module and to restore it into the device. Access to these parameters is restricted to the Endress+Hauser service. no output values 1) depending on the configuration of the block

Micropilot FMR51, FMR52

18 Endress+Hauser Function Blocks Block Content Number of permanent blocks Number of instantiable blocks Execution time Functionality Resource Block The Resource Block contains all the data that uniquely identifies the field device. It is an electronic version of a nameplate of the device. 1 0 - enhanced Analog Input Block The AI block takes the manufacturer's input data, selected by channel number, and makes it available to other function blocks at its output. 2 3 25 ms enhanced Discrete Input Block The DI block takes a discrete input value (e.g. indication of an level limit), and makes it available to other function blocks at its output. 1 2 20 ms standard Mutiple Analog Output Block This block is used to transfer analog data from the bus into the device 1 0 20 ms standard Mutiple Discrete Output Block This block is used to transfer discrete data from the bus to the device. 1 0 20 ms standard PID Block The PID block serves as proportional-integralderivative controller and is used almost universally to do closed-loop- control in the field including cascade and feedforward. 1 1 25 ms standard Arithmetic Block This block is designed to permit simple use of popular measurement math functions. The user does not have to know how to write equations. The math algorithm is selected by name, chosen by the user for the function to be done. 1 1 25 ms standard Signal Characterizer Block The signal characterizer block has two sections, each with an output that is a non-linear function of the respective input. The non-linear function is determined by a single look-up table with 21 arbitrary x-y pairs. 1 1 25 ms standard Input Selector Block The input selector block provides selection of up to four inputs and generates an output based on the configured action. This block normally receives its inputs from AI blocks. The block performs maximum, minimum, middle, average and ‘first good’ signal selection. 1 1 25 ms standard

Micropilot FMR51, FMR52

Endress+Hauser 19 Block Content Number of permanent blocks Number of instantiable blocks Execution time Functionality Integrator Block The Integrator Function Block integrates a variable as a function of the time or accumulates the counts from a Pulse Input block. The block may be used as a totalizer that counts up until reset or as a batch totalizer that has a setpoint, where the integrated or accumulated value is compared to pre-trip and trip settings, generating discrete signals when these settings are reached. 1 1 25 ms standard Analog Alarm Block 1 1 25 ms standard Up to 20 blocks can be instantiated in the device altogether, including the blocks already instantiated on delivery.

Micropilot FMR51, FMR52

20 Endress+Hauser Power supply Terminal assignment 2-wire: 4-20mA HART 8 9 1 + 2 4...20mA HART 10 mm Spare part 71108xxx 2- wire level 4-20 mA 4-20 mA HART [21] open - 1 + 2 4-20mA 1-channel overvoltage protection - [16] A 7 B � 4...20 mA 5 4 1 2 3 + – + – � + – 4...20 mA 5 4 1 2 3 + – 6 � A0011294  2 Terminal assignment 2-wire; 4-20mA HART A Without integrated overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply (e.g. RN221N): Observe terminal voltage 2 HART communication resistor (≥ 250 Ω): Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) 4 Analog display device: Observe maximum load 5 Cable screen; observe cable specification 6 4-20mA HART (passive): Terminals 1 and 2 7 Overvoltage protection module 8 Terminal for potential equalization line 9 Cable entry

Micropilot FMR51, FMR52

Endress+Hauser 21 2-wire: 4-20mA HART, switch output 1 + 2 4...20 mA HART 10 mm Spare part 71108xxx 2- wire 4-20 mA PFS HART [02/03] open - A 1 + 2 - 3 + 4 - 1 3 + + 2 4 4-20mA/ FIELDBUS 4-20mA/ 2-channel overvoltage protection - - [17] B 10 9 8 7 11 + + - - 2 2 3 3 4 4 6 5 5 1 1 4...20 mA ≥250 Ω 3+ 3+ 4- 4- + + – – � � � 4...20 mA ≥250 Ω A0013759  3 Terminal assignment 2-wire; 4-20mA HART, switch output A Without integrated overvoltage protection B With integrated overvoltage protection 1 Active barrier with power supply (e.g. RN221N): Observe terminal voltage 2 HART communication resistor (≥ 250 Ω): Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) 4 Analog display device: Observe maximum load 5 Cable screen; observe cable specification 6 4-20mA HART (passive): Terminals 1 and 2 7 Switch output (open collector): Terminals 3 and 4 8 Terminal for potential equalization line 9 Cable entry for 4-20mA HART line 10 Cable entry for switch output line 11 Overvoltage protection module

Micropilot FMR51, FMR52

22 Endress+Hauser 2-wire: 4-20mA HART, 4-20mA 1 3 + + 2 4 4...20mA HART 4...20mA 10 mm Spare part 71108xxx 2- wire level 4-20 mA 4-20 mA HART [04/05] open - - 11 A 1 3 + + 2 4 4-20mA/ FIELDBUS 4-20mA/ 2-channel overvoltage protection - - [17] 14 13 12 + + + + - - - - 1 1 2 2 3 3 9 9 5 5 8 8 6 6 7 7 4 4 4 4 + + – – + + – – 4...20 mA 4...20 mA 10 B 4...20 mA 4...20 mA � � � � � A0013923  4 Terminal assignment 2-wire, 4-20 mA HART, 4...20mA A Without integrated overvoltage protection B With integrated overvoltage protection 1 Connection current output 2 2 Connection current output 1 3 Supply voltage for current output 1 (e.g. RN221N); Observe terminal voltage 4 Cable screen; observe cable specification 5 HART communication resistor (≥ 250 Ω): Observe maximum load 6 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) 7 Analog display device ; observe maximum load 8 Analog display device ; observe maximum load 9 Supply voltage for current output 2 (e.g. RN221N); Obeserve terminal voltage 10 Overvoltage protection module 11 Current output 2: Terminals 3 and 4 12 Terminal for the potential equalization line 13 Cable entry for current output 1 14 Cable entry for current output 2 This version is also suited for single-channel operation. In this case, current output 1 (terminals 1 and 2) must be used.

Micropilot FMR51, FMR52

Endress+Hauser 23 4-wire: 4-20mA HART (10.4 to 48 VDC) 3 1 + L+ 4 2 4...20 mA HART 10.4...48 V= 10 mm Spare part 71108xxx 2- wire 4-20 mA HART [08] open - L- A 13 12 11 9 10 2 3 4 7 8 5 1 � � + - 6 � 4...20 mA ≥250 Ω A0011340  5 Terminal assignment 4-wire; 4-20mA HART (10.4 to 48 VDC) 1 Evaluation unit, e.g. PLC 2 HART communication resistor (≥ 250 Ω): Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) 4 Analog display device: Observe maximum load 5 Signal cable including screening (if required), observe cable specification 6 Protective connection; do not disconnect! 7 Protective earth, observe cable specification 8 4...20mA HART (active): Terminals 3 and 4 9 Supply voltage: Terminals 1 and 2 10 Supply voltage: Observe terminal voltage, observe cable specification 11 Terminal for potential equalization 12 Cable entry for signal line 13 Cable entry for power supply

Micropilot FMR51, FMR52

24 Endress+Hauser 4-wire: 4-20mA HART (90 to 253 VAC) 3 1 + L 4 2 4...20 mA HART 90...253 V~ 10 mm Spare part 71108xxx 2- wire 4-20 mA HART [09] open - N A 13 12 11 9 10 6 7 8 � � 2 3 4 5 1 + - 4...20 mA ≥250 Ω A0018965  6 Terminal assignment 4-wire; 4-20mA HART (90 to 253 VAC) 1 Evaluation unit, e.g. PLC 2 HART communication resistor (≥ 250 Ω): Observe maximum load 3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem) 4 Analog display device: Observe maximum load 5 Signal cable including screening (if required), observe cable specification 6 Protective connection; do not disconnect! 7 Protective earth, observe cable specification 8 4...20mA HART (active): Terminals 3 and 4 9 Supply voltage: Terminals 1 and 2 10 Supply voltage: Observe terminal voltage, observe cable specification 11 Terminal for potential equalization 12 Cable entry for signal line 13 Cable entry for power supply LCAUTION To ensure electrical safety: ‣ Do not disconnect the protective connection (6). ‣ Disconnect the supply voltage before disconnecting the protective earth (7). Connect protective earth to the internal ground terminal (7) before connecting the supply voltage. If necessary, connect the potential matching line to the external ground terminal (11). In order to ensure electromagnetic compatibility (EMC): Do not only ground the device via the protective earth conductor of the supply cable. Instead, the functional grounding must also be connected to the process connection (flange or threaded connection) or to the external ground terminal. An easily accessible power switch must be installed in the proximity of the device. The power switch must be marked as a disconnector for the device (IEC/EN61010).

Micropilot FMR51, FMR52

Endress+Hauser 25 PROFIBUS PA / FOUNDATION Fieldbus 1 3 + + 2 4 FIELDBUS 2-channel overvoltage protection - - [17] 4-20mA/ 4-20mA/ B 1 1 + + 2 2 FIELDBUS Spare part 71023457 PA/FF [06/07] FIELDBUS - - 1 3 + + 2 4 PA/FF 10 mm Spare part 71108xxx 2- wire level 4-20 mA PFS FIELDBUS [26/27] open - - A 4 1 1 2 3 6 5 3+ 3+ 4- 4- � A0011341  7 Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus A Without integrated overvoltage protection B With integrated overvoltage protection 1 Cable screen: Observe cable specifications 2 Switch output (open collector): Terminals 3 and 4 3 PROFIBUS PA / FOUNDATION Fieldbus: Terminals 1 and 2 4 Terminal for potential equalization line 5 Cable entries 6 Overvoltage protection module

Micropilot FMR51, FMR52

26 Endress+Hauser Connection examples for the switch output For HART devices, the switch output is available as an option. See product structure, feature 20: "Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output" Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output. 3+ +- 4- Ri A0015909  8 Connection of a relay Suitable relays (examples): • Solid-state relay: Phoenix Contact OV-24DC/480AC/5 with mounting rail connector UMK-1 OM-R/AMS • Electromechanical relay: Phoenix Contact PLC-RSC-12DC/21 3+ 2 1 + 4- Ri A0015910  9 Connection of a digital input 1 Pull-up resistor 2 Digital input For optimum interference immunity we recommend to connect an external resistor (internal resistance of the relay or Pull-up resistor) of < 1 000 Ω.

Micropilot FMR51, FMR52

Endress+Hauser 27 Device plug connectors For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected without opening the housing. Pin assignment of the M12 plug connector 2 1 3 4 A0011175 Pin Meaning 1 Signal + 2 not connected 3 Signal - 4 Ground Pin assignment of the 7/8" plug connector 4 2 3 1 A0011176 Pin Meaning 1 Signal - 2 Signal + 3 Not connected 4 Screen

Micropilot FMR51, FMR52

28 Endress+Hauser Supply voltage An external power supply is required. Various supply units can be ordered from Endress+Hauser: see "Accessories" section→  109 2-wire, 4-20mA HART, passive "Power Supply, Output" 1) "Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the supply unit A: 2-wire; 4-20mA HART • Non-Ex • Ex nA • Ex ic • CSA GP 10.4 to 35 V 3) 4) R [ ] � U0 [V] 10 10.4 21.4 20 30 35 0 500 A0017140 Ex ia / IS 10.4 to 30 V 3) 4) • Ex d(ia) / XP • Ex ic(ia) • Ex nA(ia) • Ex ta / DIP 13 to 35 V 5) U0 [V] 10 13 24 20 30 35 0 500 R [ ] Ω A0034771 Ex ia + Ex d(ia) / IS + XP 13 to 30 V 5) 1) Feature 020 of the product structure 2) Feature 010 of the product structure 3) For ambient temperatures Ta≤ -20 °C (-4 °F) a minimum voltage of 15 V is required for the sartup of the device at the MIN error current (3,6 mA). The startup current can be parametrized. If the device is operated with a fixed current I ≥ 5,5 mA (HART multidrop mode), a voltage of U ≥ 10,4 V is sufficient throughout the entire range of ambient temperatures. 4) In the current simulation mode a voltage U ≥ 12.5 V is required. 5) For ambient temperatures Ta≤ -20 °C (-4 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6 mA). "Power Supply, Output" 1) "Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the supply unit B: 2-wire; 4-20 mA HART, switch output • Non-Ex • Ex nA • Ex nA(ia) • Ex ic • Ex ic(ia) • Ex d(ia) / XP • Ex ta / DIP • CSA GP 13 to 35 V 3) U0 [V] 10 13 24 20 30 35 0 500 R [ ] Ω A0034771 • Ex ia / IS • Ex ia + Ex d(ia) / IS + XP 13 to 30 V 3) 1) Feature 020 of the product structure 2) Feature 010 of the product structure 3) For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6 mA).

Micropilot FMR51, FMR52

Endress+Hauser 29 "Power Supply, Output" 1) "Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the supply unit C: 2-wire; 4-20mA HART, 4-20mA any 13 to 28 V 3) R [ ] � U0 [V] 10 13 24 20 28 0 500 A0034841 1) Feature 020 of the product structure 2) Feature 010 of the product structure 3) For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6 mA). Polarity reversal protection Yes Admissible residual ripple at f = 0 to 100 Hz USS < 1 V Admissible residual ripple at f = 100 to 10000 Hz USS < 10 mV

Micropilot FMR51, FMR52

30 Endress+Hauser 4-wire, 4-20mA HART, active "Power supply; Output" 1) Terminal voltage Maximum load Rmax K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 VAC (50 to 60 Hz), overvoltage category II 500 Ω L: 4-wire 10,4-48VDC; 4-20mA HART 10.4 to 48 VDC 1) Feature 020 of the product structure PROFIBUS PA, FOUNDATION Fieldbus "Power supply; Output" 1) "Approval" 2) Terminal voltage E: 2-wire; FOUNDATION Fieldbus, switch output G: 2-wire; PROFIBUS PA, switch output • Non-Ex • Ex nA • Ex nA(ia) • Ex ic • Ex ic(ia) • Ex d(ia) / XP • Ex ta / DIP • CSA GP 9 to 32 V 3) • Ex ia / IS • Ex ia + Ex d(ia) / IS + XP 9 to 30 V 3) 1) Feature 020 of the product structure 2) Feature 010 of the product structure 3) Input voltages up to 35 V will not spoil the device. Polarity sensitive No FISCO/FNICO compliant according to IEC 60079-27 Yes Power consumption "Power supply; Output" 1) Power consumption A: 2-wire; 4-20mA HART < 0.9 W B: 2-wire; 4-20mA HART, switch output < 0.9 W C: 2-wire; 4-20mA HART, 4-20mA < 2 x 0.7 W K: 4-wire 90-253VAC; 4-20mA HART 6 VA L: 4-wire 10,4-48VDC; 4-20mA HART 1.3 W 1) Feature 020 of the product structure Current consumption HART Nominal current 3.6 to 22 mA, the start-up current for multidrop mode can be parametrized (is set to 3.6 mA on delivery) Breakdown signal (NAMUR NE43) adjustable: 3.59 to 22.5 mA PROFIBUS PA Nominal current 14 mA Error current FDE (Fault Disconnection Electronic) 0 mA

Micropilot FMR51, FMR52

Endress+Hauser 31 FOUNDATION Fieldbus Device basic current 15 mA Error current FDE (Fault Disconnection Electronic) 0 mA FISCO Ui 17.5 V Ii 550 mA Pi 5.5 W Ci 5 nF Li 10 μH Power supply failure • Configuration is retained in the HistoROM (EEPROM). • Error messages (incl. value of operated hours counter) are stored. Potential equalization No special measures for potential equalization are required. If the device is designed for hazardous areas, observe the information in the documentation "Safety Instructions" (XA). Terminals • Without integrated overvoltage protection Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG) • With integrated overvoltage protection Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG) Cable entries Connection of power supply and signal line To be selected in feature 050 "Electrical connection" • Gland M20; Material dependent on the approval: – For Non-Ex, ATEX, IECEx, NEPSI Ex ia/ic: Plastics M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in) – For Dust-Ex, FM IS, CSA IS, CSA GP, Ex nA: – For Ex d: No gland available • Thread – ½" NPT – G ½" – M20 × 1.5 • Plug M12 / Plug 7/8" Only available for Non-Ex, Ex ic, Ex ia Connection of remote display FHX50 Dependent on feature 030: "Display, Operation": • "Prepared for display FHX50 + M12 connection": M12 socket • "Prepared for display FHX50 + custom connection": M16 cable gland Cable specification • Devices without integrated overvoltage protection Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG) • Devices with integrated overvoltage protection Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG) • For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.

Micropilot FMR51, FMR52

32 Endress+Hauser HART • A normal device cable suffices if only the analog signal is used. • A shielded cable is recommended if using the HART protocol. Observe grounding concept of the plant. • For 4-wire devices: Standard device cable is sufficient for the power line. PROFIBUS Use a twisted, screened two-wire cable, preferably cable type A. For further information on the cable specifications, see Operating Instructions BA00034S "PROFIBUS DP/PA: Guidelines for planning and commissioning", PNO Guideline 2.092 "PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP). FOUNDATION Fieldbus Endress+Hauser recommends using twisted, shielded two-wire cables. For further information on the cable specifications, see Operating Instructions BA00013S "FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP). Overvoltage protection If the measuring device is used for level measurement in flammable liquids which requires the use of overvoltage protection according to DIN EN 60079-14, standard for test procedures 60060-1 (10 kA, pulse 8/20 μs), overvoltage protection has to be ensured by an integrated or external overvoltage protection module. Integrated overvoltage protection An integrated overvoltage protection module is available for 2-wire HART as well as PROFIBUS PA and FOUNDATION Fieldbus devices. Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection". Technical data Resistance per channel 2 × 0.5 Ω max. Threshold DC voltage 400 to 700 V Threshold impulse voltage < 800 V Capacitance at 1 MHz < 1.5 pF Nominal arrest impulse voltage (8/20 μs) 10 kA External overvoltage protection HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection. For detailed information please refer to the following documents: • HAW562: TI01012K • HAW569: TI01013K

Micropilot FMR51, FMR52

Endress+Hauser 33 Performance characteristics Reference operating conditions • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F) • Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi) • Humidity = 60 % ±15 % • Reflector: metal plate with a minimum diameter of 1 m (40 in) • No major interference reflections inside the signal beam Maximum measured error Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1; percentage values in relation to the span. Device Value Output digital analog 1) FMR51/FMR52 Standard version Sum of non- linearity, nonrepeatability and hysteresis ± 2 mm (0.08 in) ± 0.02 % Offset/Zero ± 4 mm (0.2 in) ± 0.03 % FMR51/FMR52 Version with application package "Advanced dynamics" 2) Sum of non- linearity, nonrepeatability and hysteresis ± 3 mm (0.12 in) ± 0.02 % Offset/Zero ± 4 mm (0.2 in) ± 0.03 % 1) Only relevant for 4-20mA current output; add error of the analog value to the digital value. 2) Product structure: Feature 540 "Application Package", Option EM "Advanced dynamics" Differing values in near-range applications 2 (0.08) 0 -2 (-0.08) A 1 (3.3) 4 (0.16) -4 (-0.16) R � [mm] ([in]) D [m] ([ft]) A0019035  10 Maximum measured error in near-range applications; values for standard version Δ Maximum measured error A Lower edge of the antenna D Distance from the lower edge A of the antenna R Reference point of the distance measurement

Micropilot FMR51, FMR52

34 Endress+Hauser 3 (0.12) 0 -3 (-12) A 2 (6.6) D [m] ([ft]) 20 (0.79) -20 (-0.79) R � [mm] ([in]) A0019034  11 Maximum measured error in near-range applications; values for version with the "Advanced dynamics" application package Δ Maximum measured error A Lower edge of the antenna D Distance from the lower edge A of the antenna R Reference point of the distance measurement Measured value resolution Dead band according to DIN EN IEC 61298-2 / DIN EN IEC 60770-1: • digital: 1 mm • analog: 1 μA Reaction time The reaction time can be parametrized. The following step response times (as per DIN EN IEC 61298-2 / DIN EN IEC 60770-1) 1)are valid if the damping is switched off: Tank height Sampling rate Step response time < 10 m (33 ft) ≥ 3.6 s–1 < 0.8 s < 70 m (230 ft) ≥ 2.2 s–1 < 1 s Influence of ambient temperature The measurements are carried out in accordance with DIN EN IEC 61298-3 / DIN EN IEC 60770-1 • Digital (HART, PROFIBUS PA, FOUNDATION Fieldbus): – Standard version: average TK = 2 mm/10 K – Version with advanced dynamics 2): average TK = 3 mm/10 K • Analog (current output): – zero point (4 mA): average TK = 0.02 %/10 K – span (20 mA): average TK = 0.05 %/10 K Influence of gas layer High pressures reduce the propagation velocity of the measuring signals in the gas/vapor above the fluid. This effect depends on the kind of gas/vapor and of its temperature. This results in a systematic measuring error that gets bigger as the distance increases between the reference point of the measurement (flange) and the product surface. The following table illustrates this measured 1) According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1 the response time is the time which passes after a sudden change of the input signal until the output signal for the first time assumes 90% of the steady-state value. 2) Feature 540 "Application Package", option EM

Micropilot FMR51, FMR52

Endress+Hauser 35 error for a few typical gases/vapors (with regard to distance; a positive value means that too large a distance is being measured): Gas layer Temperature Pressure °C °F 1 bar (14,5 psi) 10 bar (145 psi) 50 bar (725 psi) 100 bar (1450 psi) 160 bar (2320 psi) Air/ Nitrogen 20 68 0.00 % 0.22 % 1.2 % 2.4 % 3.89 % 200 392 –0.01 % 0.13 % 0.74 % 1.5 % 2.42 % 400 752 –0.02 % 0.08 % 0.52 % 1.1 % 1.70 % Hydrogen 20 68 –0.01 % 0.10 % 0.61 % 1.2 % 2.00 % 200 392 –0.02 % 0.05 % 0.37 % 0.76 % 1.23 % 400 752 –0.02 % 0.03 % 0.25 % 0.53 % 0.86 % Water (saturated steam) 100 212 0.02 % - - - - 180 356 - 2.1 % - - - 263 505,4 - - 8.6 % - - 310 590 - - - 22 % - 364 687 - - - - 41.8 % When the pressure is known and constant, this measured error can, for example, be compensated by means of a linearization. Gas phase compensation with external pressure sensor (PROFIBUS PA, FOUNDATION Fieldbus) PROFIBUS devices can receive the signal of an external pressure sensor through the bus and use it to perform a pressure dependent time-of-flight correction. In the case of saturated steam in the temperature range from 100 to 350 °C (212 to 662 °f), for example, the measuring error of the distance measurement can be reduced by this method from up to 29 % (without compensation) to less than 3 % (with compensation).

Micropilot FMR51, FMR52

36 Endress+Hauser Installation Installation conditions Mounting position A 1 2 3 A0016882 • Recommended distance A from wall to outer edge of nozzle: ~ 1/6 of tank diameter. Nevertheless the device should not be installed closer than 15 cm (5.91 in) to the tank wall. • Not in the center (2), as interference can cause signal loss. • Not above the fill stream (3). • It is recommended to us a weather protection cover (1) in order to protect the device from direct sun or rain. Vessel installations A0018944 Avoid any installations (point level switches, temperature sensors, braces, vacuum rings, heating coils, baffles etc.) inside the signal beam. Take into account the beam angle →  38.

Micropilot FMR51, FMR52

Endress+Hauser 37 Reduction of interference echoes A0016890 Metallic screens mounted at a slope spread the radar signal and can, therefore, reduce interference echoes. Measurement in a plastic vessel If the outer wall of the vessel is made of a non-conductive material (e.g. GRP), microwaves can also be reflected off interfering installations outside the vessel (e.g. metallic pipes (1), ladders (2), grates (3), ...). Therefore, there should be no such interfering installations in the signal beam. Please contact Endress+Hauser for further information. 2 3 1 A0017123

Micropilot FMR51, FMR52

38 Endress+Hauser Optimization options • Antenna size The bigger the antenna, the smaller the beam angle α and the fewer interference echoes →  38. • Mapping The measurement can be optimized by means of electronic suppression of interference echoes. • Antenna alignment Take into account the marker on the flange or threaded connection →  41 →  44. • Stilling well A stilling well can be applied to avoid interferences →  45. • Metallic screens mounted at a slope They spread the radar signals and can, therefore, reduce interference echoes. Beam angle � D W � D _ = 2 2. . tan W A0016891  12 Relationship between beam angle α, distance D and beamwidth diameter W The beam angle is defined as the angle α where the energy density of the radar waves reaches half the value of the maximum energy density (3-dB-width). Microwaves are also emitted outside the signal beam and can be reflected off interfering installations.

Micropilot FMR51, FMR52

Endress+Hauser 39 Beam diameter W as a function of beam angle α and measuring distance D: FMR51 Antenna size 40 mm (1½ in) 50 mm (2 in) 80 mm (3 in) 100 mm (4 in) Beam angle α 23° 18° 10° 8° Measuring distance (D) Beamwidth diameter W 3 m (9.8 ft) 1.22 m (4 ft) 0.95 m (3.1 ft) 0.53 m (1.7 ft) 0.42 m (1.4 ft) 6 m (20 ft) 2.44 m (8 ft) 1.9 m (6.2 ft) 1.05 m (3.4 ft) 0.84 m (2.8 ft) 9 m (30 ft) 3.66 m (12 ft) 2.85 m (9.4 ft) 1.58 m (5.2 ft) 1.26 m (4.1 ft) 12 m (39 ft) 4.88 m (16 ft) 3.80 m (12 ft) 2.1 m (6.9 ft) 1.68 m (5.5 ft) 15 m (49 ft) 6.1 m (20 ft) 4.75 m (16 ft) 2.63 m (8.6 ft) 2.10 m (6.9 ft) 20 m (66 ft) 8.14 m (27 ft) 6.34 m (21 ft) 3.50 m (11 ft) 2.80 m (9.2 ft) 25 m (82 ft) 10.17 m (33 ft) 7.92 m (26 ft) 4.37 m (14 ft) 3.50 m (11 ft) 30 m (98 ft) - 9.50 m (31 ft) 5.25 m (17 ft) 4.20 m (14 ft) 35 m (115 ft) - 11.09 m (36 ft) 6.12 m (20 ft) 4.89 m (16 ft) 40 m (131 ft) - 12.67 m (42 ft) 7.00 m (23 ft) 5.59 m (18 ft) 45 m (148 ft) - - 7.87 m (26 ft) 6.29 m (21 ft) 60 m (197 ft) - - 10.50 m (34 ft) 8.39 m (28 ft) 70 m (230 ft) - - - 9.79 m (32 ft) FMR52 Antenna size 50 mm (2 in) 80 mm (3 in) Beam angle α 18° 10° Measuring distance (D) Beamwidth diameter W 3 m (9.8 ft) 0.95 m (3.1 ft) 0.53 m (1.7 ft) 6 m (20 ft) 1.9 m (6.2 ft) 1.05 m (3.4 ft) 9 m (30 ft) 2.85 m (9.4 ft) 1.58 m (5.2 ft) 12 m (39 ft) 3.80 m (12 ft) 2.1 m (6.9 ft) 15 m (49 ft) 4.75 m (16 ft) 2.63 m (8.6 ft) 20 m (66 ft) 6.34 m (21 ft) 3.50 m (11 ft) 25 m (82 ft) 7.92 m (26 ft) 4.37 m (14 ft) 30 m (98 ft) 9.50 m (31 ft) 5.25 m (17 ft) 35 m (115 ft) 11.09 m (36 ft) 6.12 m (20 ft) 40 m (131 ft) 12.67 m (42 ft) 7.00 m (23 ft) 45 m (148 ft) - 7.87 m (26 ft) 60 m (197 ft) - 10.50 m (34 ft)

Micropilot FMR51, FMR52

40 Endress+Hauser Measuring conditions • In case of boiling surfaces, bubbling or tendency for foaming use FMR53 or FMR54. Depending on its consistence, foam can either absorb microwaves or reflect them off the foam surface. Measurement is possible under certain conditions. For FMR50, FMR51 and FMR52, the additional option "Advanced dynamics" is recommended in these cases (feature 540: "Application Package", option EM). • In case of heavy steam development or condensate, the maximum measuring range of FMR50, FMR51 and FMR52 may decrease depending on density, temperature and composition of the steam → use FMR53 or FMR54. • For the measurement of absorbing gases such as ammonia NH3 or some fluorocarbons 3), please use Levelflex or Micropilot FMR54 in a stilling well. • The measuring range begins, where the beam hits the tank bottom. Particularly with dish bottoms or conical outlets the level cannot be detected below this point. • In stilling well applications, the electromagnetic waves do not propagate completely outside the tube. It must be taken into account that the accuracy may be reduced in the area C. In order to guarantee the required accuracy in these cases, it is recommended to position the zero-point at a distance C above the end of the tube (see figure). • In case of media with a low dielectric constant (εr = 1.5 to 4) 4) the tank bottom can be visible through the medium at low levels (low height C). Reduced accuracy has to be expected in this range. If this is not acceptable, we recommend positioning the zero point at a distance C (see figure) above the tank bottom in these applications. • In principle it is possible to measure up to the tip of the antenna with FMR51, FMR53 and FMR54. However, due to considerations regarding corrosion and build-up, the end of the measuring range should not be chosen any closer than A (see figure) to the tip of the antenna. • When using FMR54 with planar antenna, especially for media with low dielectric constants, the end of the measuring range should not be closer than A: 1 m (3.28 ft) to the flange. • The smallest possible measuring range B depends on the antenna version (see figure). • The tank height should be at least H (see table). 100% 0% B A C H øD A0018872 Device A [mm (in)] B [m (ft)] C [mm (in)] H [m (ft)] FMR51 50(1.97) > 0.2 (0.7) 50 to 250 (1.97 to 9.84) > 0.3 (1.0) FMR52 200(7.87) 3) Affected compounds are e.g. R134a, R227, Dymel 152a. 4) Dielectric constants of important media commonly used in various industries are summarized in the DC manual (CP01076F) and in the Endress +Hauser "DC Values App" (available for Android and iOS).

Micropilot FMR51, FMR52

Endress+Hauser 41 Mounting cladded flanges • Use flange screws according to the number of flange holes. • Tighten the screws with the required torque (see table). • Retighten the screws after 24 hours or after the first temperature cycle. • Depending on process pressure and process temperature check and retighten the screws at regular intervals. Usually, the PTFE flange cladding also serves as a seal between the nozzle and the device flange. Flange size Number of screws Recommended torque [Nm] minimum maximum EN DN50/PN16 4 45 65 DN80/PN16 8 40 55 DN100/PN16 8 40 60 DN150/PN16 8 75 115 ASME 2"/150lbs 4 40 55 3"/150lbs 4 65 95 4"/150lbs 8 45 70 6"/150lbs 8 85 125 JIS 10K 50A 4 40 60 10K 80A 8 25 35 10K 100A 8 35 55 10K 100A 8 75 115 Installation in vessel (free space) Horn antenna (FMR51) Alignment • Align the antenna vertically to the product surface. The maximum range may be reduced if the horn antenna is not vertically aligned. • A marking at the flange (somewhere between the flange holes) or the boss enables alignment of the antenna. This marking must be aligned towards the tank wall as well as possible.

Micropilot FMR51, FMR52

42 Endress+Hauser 90° 90° 90° 90° 90° 90° 90° 90° 90° 90° A B A0018974 Depending on the device version the marking may be a circle or two short parallel lines. Nozzle mounting For optimum measurement, the tip of the antenna should extend below the nozzle. Depending on the antenna size this is achieved by the following maximum nozzle heights: Hmax A0016820  13 Nozzle height for horn antenna (FMR51) Antenna 1) Maximum nozzle height Hmax BA: Horn 40mm/1-1/2" 86 mm (3.39 in) BB: Horn 50mm/2" 115 mm (4.53 in) BC: Horn 80mm/3" 211 mm (8.31 in) BD Horn 100mm/4" 282 mm (11.1 in) 1) Feature 070 of the product structure Conditions for longer nozzles If the medium has good reflective properties, higher nozzles can be accepted. In this case the maximum nozzle height, Hmax, is dependent on the nozzle diameter, D:

Micropilot FMR51, FMR52

Endress+Hauser 43 Hmax øD A0023611 Nozzle diameter D Maximum nozzle height Hmax Recommended antenna 1) 40 mm (1.5 in) 100 mm (3.9 in) BA: Horn 40mm/1-1/2" 50 mm (2 in) 150 mm (5.9 in) BB: Horn 50mm/2" 80 mm (3 in) 250 mm (9.8 in) BC: Horn 80mm/3" 100 mm (4 in) 500 mm (19.7 in) BD: Horn 100mm/4" 150 mm (6 in) 800 mm (31.5 in) BD: Horn 100mm/4" 1) Feature 070 of the product structure If the antenna doesn't extend below the nozzle, observe the following: • The nozzle end must be smooth and free of burrs. If possible its edge should be rounded. • An interference echo suppression must be performed. • Please contact Endress+Hauser for applications with higher nozzles than those indicated in the table. • For mounting in heigh nozzles the device is available in a version with an antenna extension of up to 1 000 mm (39.4 in) 5) • The antenna extension may cause interference echoes in the near range. In this case it may occur that the maximum measurable level is reduced. Threaded connection For devices with a threaded connection it may be necessary - depending on the antenna size - to unmount the horn before fastening the device and to mount it again afterwards. • Tighten with the hexagonal nut only. • Tool : 55 mm hexagonal wrench • Maximum permissible torque: 60 Nm (44 lbf ft) Mesurement from the outside through plastic walls (FMR50/FMR51) • Dielectric constant of the medium: εr > 10 • If possible, use an antenna 100 mm (4 in). • The distance from the lower edge of the antenna to the tank ceiling should be about 100 mm (4 in). • If possible, avoid mounting locations where condensation or build-up might occur. • In case of outdoor mounting, the space between antenna and vessel has to be protected from the elements. • Do not mount any potential reflectors (e.g. pipes) outside the tank in the signal beam. 5) Feature 610 "Accessory mounted" of the product structure.

Micropilot FMR51, FMR52

44 Endress+Hauser Suitable thickness of the tank ceiling: Penetrated material PE PTFE PP Perspex DK / εr 2.3 2.1 2.3 3.1 Optimum thickness 1) 3.8 mm (0.15 in) 4.0 mm (0.16 in) 3.8 mm (0.15 in) 3.3 mm (0.13 in) 1) Other possible values for the thickness are multiples of the values listed (e.g. for PE: 7,6 mm (0.3 in), 11,4 mm (0.45 in) Horn antenna, flush mount (FMR52) Alignment • Align the antenna vertically to the product surface. The maximum range may be reduced if the horn antenna is not vertically aligned. • A marking at the flange (somwhere between the flange holes) or the boss enables alignment of the antenna. This marking must be aligned towards the tank wall as well as possible. 90° 90° 90° 90° 90° 90° 90° 90° 90° 90° A B A0018974 Depending on the device version the marking may be a circle or two short parallel lines. Nozzle mounting Hmax A0016819  14 Nozzle height for horn antenna, flush mount (FMR52)

Micropilot FMR51, FMR52

Endress+Hauser 45 Antenna 1) Maximum nozzle height Hmax BO: Horn 50mm/2" 500 mm (19.7 in) BP: Horn 80mm/3" 500 mm (19.7 in) 1) Feature 070 of the product structure Please contact Endress+Hauser for applications with higher nozzle. • For flanges with PTFE cladding: Observe the notes on the mounting of cladded flanges →  41. • Usually, the PTFE flange cladding also serves as a seal between the nozzle and the device flange. Installation in stilling well 1 90° A0016841  15 Installation in stilling well 1 Marking for antenna alignment • For horn antenna: Align the marking towards the slots of the stilling well. • Measurements can be performed through an open full bore ball valve without any problems. • After mounting, the housing can be turned 350° in order to facilitate access to the display and the terminal compartment. Recommendations for the stilling well • Metal (no enamel coating; plastic on request). • Constant diameter. • Diameter of stilling well not larger than antenna diameter. • Diameter difference between horn antenna and inner diameter of the stilling well as small as possible. • Weld seam as smooth as possible and on the same axis as the slots. • Slots offset 180° (not 90°). • Slot width or diameter of holes max. 1/10 of pipe diameter, de-burred. Length and number do not have any influence on the measurement. • Select horn antenna as big as possible. For intermedaite sizes (e.g. 180 mm (7 in)) select next larger antenna and adapt it mechanically (for horn antennas) • At any transition (i.e. when using a ball valve or mending pipe segments), no gap may be left exceeding 1 mm (0.04 in). • The stilling well must be smooth on the inside (average roughness Rz ≤ 6.3 μm (248 μin)). Use extruded or parallel welded metal pipe. An extension of the pipe is possible with welded flanges or pipe sleeves. Flange and pipe have to be properly aligned at the inside. • Do not weld through the pipe wall. The inside of the stilling well must remain smooth. In case of unintentional welding through the pipe, the weld seam and any unevenness on the inside need to be carefully removed and smoothened. Otherwise, strong interference echoes will be generated and material build-up will be promoted. • In the case of smaller nominal widths flanges must be welded to the pipe such that they allow for a correct orientation (marker aligned toward slots).

Micropilot FMR51, FMR52

46 Endress+Hauser Examples for the construction of stilling wells 100% 0% 100% 0% A B A A A-A 150…500 (5.91...19.7) 150…500 (5.91...19.7) mm (in) 2 1 4 4 C D 5 6 6 7 1 3 1 (0.04) ≤ 1 (0.04) ≤ A0019009 A Micropilot FMR50/FMR51: Horn 40mm(1½") B Micropilot FMR50/FMR51/FMR52/FMR54: Horn 80mm(3") C Stilling well with slots D Full bore ball valve 1 Marking for axial alignment 2 Threaded connection 3 e.g. welding neck flange DIN2633 4  hole max. 1/10  stilling well 5  hole max. 1/10  stilling well; single sided or drilled through 6 Inside of holes deburred 7 Diameter of opening of ball valve must always be equivalent to pipe diameter; avoid edges and constrictions.

Micropilot FMR51, FMR52

Endress+Hauser 47 Installation in bypass 1 2 2 90° 90° A0019446  16 Installation in bypass 1 Marking for antenna alignment 2 Tank connectors • Alighn the marker perpendicular (90°) to the tank connectors. • Measurements can be performed through an open full bore ball valve without any problems. • After mounting, the housing can be turned 350° in order to simplify access to the display and the terminal compartment. Recommendations for the bypass pipe • Metal (no plastic or enamel coating). • Constant diameter. • Select horn antenna as big as possible. For intermediate sizes (e.g. 95 mm (3.5 in)) select next larger antenna and adapt it mechanically (for horn antennas). • Diameter difference between horn antenna and inner diameter of the bypass as small as possible. • At any transition (i.e. when using a ball valve or mending pipe segments), no gap may be created exceeding 1 mm (0.04 in). • In the area of the tank connections (~ ±20 cm (7.87 in)) a reduced accuracy of the measurement has to be expected.

Micropilot FMR51, FMR52

48 Endress+Hauser Example for the construction of a bypass 100% 0% A A A A-A mm (in) 200 (7.87) 200 (7.87) 200 (7.87) 200 (7.87) B 5 4 4 1 2 C 3 3 ≤ 1 (0.04) A0019010 A Micropilot FMR50/FMR51/FMR52/FMR54: Horn 80mm(3") B Full bore ball valve C Minimum distance to upper connection pipe: 400 mm (15,7 in) 1 Marking for axial alignment 2 e.g. welding neck flange DIN2633 3 Diameter of the connection pipes as small as possible 4 Do not weld through the pipe wall; the inside of the bypass must remain smooth. 5 Diameter of opening of ball valve must always be equivalent to pipe diameter. Avoid edges and constrictions.

Micropilot FMR51, FMR52

Endress+Hauser 49 Vessels with heat insulation 1 2 A0019142 If process temperatures are high, the device must be included in normal tank insulation to prevent the electronics heating up as a result of heat radiation or convection. The insulation may not exceed beyond the neck of the housing.

Micropilot FMR51, FMR52

50 Endress+Hauser Environment Ambient temperature range Measuring device –40 to +80 °C (–40 to +176 °F); –50 °C (–58 °F) with manufacturer declaration on request Local display –20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at temperatures outside the temperature range. When operating the device in the open with strong sunlight: • Mount the device in a shady position. • Avoid direct sunlight, especially in warmer regions. • Use a weather protection cover (see accessories). Ambient temperature limits The following diagrams take into account only functional aspects. There may be further restrictions for certified device versions. Please refere to the separate Safety Instructions →  110. With a temperature (Tp) at the process connection the admissible ambient temperature (Ta) is reduced according to the following diagram (temperature derating): Information concerning the derating tables Option Meaning A 2-wire; 4-20 mA HART B 2-wire; 4-20 mA HART, switch output C 2-wire; 4-20 mA HART, 4-20 mA E 2-wire; FF, switch output G 2-wire; PA, switch output K 4-wire 90-253VAC; 4-20 mA HART L 4-wire 10, 4-48VDC; 4-20 mA HART FMR51 Seal: • Viton GLT, –40 to 150 °C (–40 to 302 °F) • Kalrez, –20 to 150 °C (–4 to 302 °F) Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 150 (302) 66 (151) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 150 (302) 67 (153) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 150 (302) 61 (142) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 150 (302) 67 (153) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 150 (302) 63 (145) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 150 (302) 68 (154) 150 (302) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

Endress+Hauser 51 FMR51 Seal: • Viton GLT, –40 to 150 °C (–40 to 302 °F) • Kalrez, –20 to 150 °C (–4 to 302 °F) Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 150 (302) 63 (145) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 150 (302) 62 (144) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - FMR51 Seal: • Viton GLT, –40 to 150 °C (–40 to 302 °F) • Kalrez, –20 to 150 °C (–4 to 302 °F) Housing: GT19 (Plastics PBT) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 80 (176) 80 (176) 80 (176) 150 (302) 56 (133) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 76 (169) 76 (169) 76 (169) 150 (302) 56 (133) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 60 (140) 60 (140) 60 (140) 150 (302) 38 (100) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 150 (302) 56 (133) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 74 (165) 74 (165) 74 (165) 150 (302) 55 (131) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 79 (174) 79 (174) 79 (174) 150 (302) 56 (133) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 63 (145) 63 (145) 63 (145) 150 (302) 41 (106) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - FMR51 Seal: • Viton GLT, –40 to 150 °C (–40 to 302 °F) • Kalrez, –20 to 150 °C (–4 to 302 °F) Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 150 (302) 69 (156) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 150 (302) 70 (158) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 150 (302) 64 (147) 150 (302) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

52 Endress+Hauser FMR51 Seal: • Viton GLT, –40 to 150 °C (–40 to 302 °F) • Kalrez, –20 to 150 °C (–4 to 302 °F) Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 150 (302) 70 (158) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 150 (302) 66 (151) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 150 (302) 71 (160) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 150 (302) 65 (149) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 150 (302) 65 (149) 150 (302) -40 (-40) -40 (-40) -40 (-40) - - FMR51 Seal: Graphite, –40 to 250 °C (–40 to 482 °F) Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 250 (482) 63 (145) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 250 (482) 64 (147) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 250 (482) 58 (136) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 250 (482) 64 (147) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 250 (482) 61 (142) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 250 (482) 65 (149) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 250 (482) 60 (140) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 250 (482) 59 (138) 250 (482) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

Endress+Hauser 53 FMR51 Seal: Graphite, –40 to 250 °C (–40 to 482 °F) Housing: GT19 (Plastics PBT) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 80 (176) 80 (176) 80 (176) 250 (482) 44 (111) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 76 (169) 76 (169) 76 (169) 250 (482) 44 (111) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 60 (140) 60 (140) 60 (140) 250 (482) 32 (90) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 250 (482) 44 (111) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 74 (165) 74 (165) 74 (165) 250 (482) 44 (111) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 79 (174) 79 (174) 79 (174) 250 (482) 44 (111) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 63 (145) 63 (145) 63 (145) 250 (482) 35 (95) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - FMR51 Seal: Graphite, –40 to 250 °C (–40 to 482 °F) Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 250 (482) 67 (153) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 250 (482) 68 (154) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 250 (482) 62 (144) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 250 (482) 68 (154) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 250 (482) 64 (147) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 250 (482) 69 (156) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 250 (482) 64 (147) 250 (482) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 250 (482) 63 (154) 250 (482) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

54 Endress+Hauser FMR51 Seal: Graphite, –196 to 450 °C (–321 to 842 °F) Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P6 P5 P4 P3 P2 P1 Ta A0019344 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -196 (-321) 81 (178) 81 (178) 81 (178) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) B Switch output not used -196 (-321) 82 (180) 82 (180) 82 (180) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) B Switch output used -196 (-321) 77 (171) 77 (171) 77 (171) 450 (842) 25 (77) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) C Channel 2 not used -196 (-321) 82 (180) 82 (180) 82 (180) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) C Channel 2 used -196 (-321) 79 (174) 79 (174) 79 (174) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) E, G Switch output not used -196 (-321) 83 (181) 83 (181) 83 (181) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) E, G Switch output used -196 (-321) 78 (172) 78 (172) 78 (172) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) K, L -196 (-321) 77 (171) 77 (171) 77 (171) 450 (842) 26 (79) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -15 (+5) FMR51 Seal: Graphite, –196 to 450 °C (–321 to 842 °F) Housing: GT19 (Plastics PBT) Temperature unit: °C (°F) Tp P6 P5 P4 P3 P2 P1 Ta A0019344 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -196 (-321) 80 (176) 80 (176) 80 (176) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) B Switch output not used -196 (-321) 76 (169) 76 (169) 76 (169) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) B Switch output used -196 (-321) 60 (140) 60 (140) 60 (140) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) C Channel 2 not used -196 (-321) 82 (180) 82 (180) 82 (180) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) C Channel 2 used -196 (-321) 74 (165) 74 (165) 74 (165) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) E, G Switch output not used -196 (-321) 79 (174) 79 (174) 79 (174) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48) E, G Switch output used -196 (-321) 63 (145) 63 (145) 63 (145) 450 (842) -29 (-20) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) 9 (48)

Micropilot FMR51, FMR52

Endress+Hauser 55 FMR51 Seal: Graphite, –196 to 450 °C (–321 to 842 °F) Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P6 P5 P4 P3 P2 P1 Ta A0019344 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -196 (-321) 81 (178) 81 (178) 81 (178) 450 (842) 39 (102) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) B Switch output not used -196 (-321) 82 (180) 82 (180) 82 (180) 450 (842) 39 (102) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) B Switch output used -196 (-321) 77 (171) 77 (171) 77 (171) 450 (842) 36 (97) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) C Channel 2 not used -196 (-321) 82 (180) 82 (180) 82 (180) 450 (842) 39 (102) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) C Channel 2 used -196 (-321) 79 (174) 79 (174) 79 (174) 450 (842) 38 (100) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) E, G Switch output not used -196 (-321) 83 (181) 83 (181) 83 (181) 450 (842) 39 (102) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) E, G Switch output used -196 (-321) 78 (172) 78 (172) 78 (172) 450 (842) 38 (100) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) K, L -196 (-321) 77 (171 77 (171) 77 (171) 450 (842) 37 (99) 450 (842) -40 (-40) -40 (-40) -40 (-40) -196 (-321) -20 (-4) FMR52 Antenna: Horn 50mm/2" Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 200 (392) 61 (142) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 61 (142) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 55 (131) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 58 (136) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 56 (133) 200 (392) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

56 Endress+Hauser FMR52 Antenna: Horn 50mm/2" Housing: GT19 (Plastics PBT) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 80 (176) 80 (176) 80 (176) 200 (392) 42 (108) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 76 (169) 76 (169) 76 (169) 200 (392) 42 (108) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 60 (140) 60 (140) 60 (140) 200 (392) 31 (88) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 42 (108) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 74 (165) 74 (165) 74 (165) 200 (392) 42 (108) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 42 (108) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 63 (145) 63 (145) 63 (145) 200 (392) 33 (91) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - FMR52 Antenna: Horn 50mm/2" Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 200 (392) 65 (149) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 65 (149 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 59 (138) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 66 (151) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 200 (392) 66 (1512) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 200 (392) 61 (142) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 60 (140) 200 (392) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

Endress+Hauser 57 FMR52 Antenna: Horn 80mm/2" Housing: GT18 (316 L) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 51 (124) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 54 (129) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 200 (392) 53 (127) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 52 (126) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - FMR52 Antenna: Horn 80mm/2" Housing: GT19 (Plastics PBT) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 80 (176) 80 (176) 80 (176) 200 (392) 36 (97) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 76 (169) 76 (169) 76 (169) 200 (392) 36 (97) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 60 (140) 60 (140) 60 (140) 200 (392) 27 (81) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 36 (97) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 74 (165) 74 (165) 74 (165) 200 (392) 36 (97) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 36 (97) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 63 (145) 63 (145) 63 (145) 200 (392) 30 (86) 200 (392) -40 (-40) -40 (-40) -40 (-40) - -

Micropilot FMR51, FMR52

58 Endress+Hauser FMR52 Antenna: Horn 80mm/2" Housing: GT20 (Alu, coated) Temperature unit: °C (°F) Tp P5 P4 P3 P2 P1 Ta A0019351 Power Supply; Output (Pos. 2 of the product structure) P1 P2 P3 P4 P5 P6 Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta A -40 (-40) 81 (178) 81 (178) 81 (178) 200 (392) 61 (142) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - B Switch output used -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 56 (133) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 not used -40 (-40) 82 (180) 82 (180) 82 (180) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - C Channel 2 used -40 (-40) 79 (174) 79 (174) 79 (174) 200 (392) 58 (136) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output not used -40 (-40) 83 (181) 83 (181) 83 (181) 200 (392) 62 (144) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - E, G Switch output used -40 (-40) 78 (172) 78 (172) 78 (172) 200 (392) 58 (136) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - K, L -40 (-40) 77 (171) 77 (171) 77 (171) 200 (392) 57 (135) 200 (392) -40 (-40) -40 (-40) -40 (-40) - - Storage temperature –40 to +80 °C (–40 to +176 °F) –50 °C (–58 °F) with manufacturer declaration on request Climate class DIN EN 60068-2-38 (test Z/AD) Altitude according to IEC61010-1 Ed.3 • Generally up to 2 000 m (6 600 ft) above MSL. • Above 2 000 m (6 600 ft) if the following conditions are met: – Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions) – Supply voltage U < 35 V – Supply voltage of overvoltage category 1 Degree of protection • With closed housing tested according to: – IP68, NEMA6P (24 h at 1.83 m under water surface) – For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water surface) 6) – IP66, NEMA4X • With open housing: IP20, NEMA1 • Display module: IP22, NEMA2 Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the PROFIBUS cable is plugged in and is also rated IP68 NEMA6P. Vibration resistance DIN EN 60068-2-64 / IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s2)2/Hz Cleaning the antenna The antenna can get contaminated, depending on the application. The emission and reception of microwaves can thus eventually be hindered. The degree of contamination leading to an error depends on the medium and the reflectivity, mainly determined by the dielectric constant εr. If the medium tends to cause contamination and deposits, cleaning on a regular basis is recommended. Care has to be taken not to damage the antenna in the process of a mechanical or 6) This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C ("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").

Micropilot FMR51, FMR52

Endress+Hauser 59 hose-down cleaning. The material compatibility has to be considered if cleaning agents are used! The maximum permitted temperature at the flange should not be exceeded. Electromagnetic compatibility (EMC) Electromagnetic compatibility to all relevant requirements of the EN 61326- series and NAMUR recommendation EMC (NE21). For details see declaration of conformity. 7). If only the analogue signal is used, unshielded interconnection lines are sufficient for the installation. In case of using the digital signal (HART/ PA/ FF) use shielded interconnection lines. Max. fluctuations during EMC- tests: < 0.5 % of the span. As an exception to this, the maximum fluctuations may amount to 2 % of the span for devices with plastic housing and see-through lid (integrated display SD02 or SD03) if strong electromagnetic interferences in the frequenency range of 1 to 2 GHz are present. 7) Can be downloaded from www.endress.com.

Micropilot FMR51, FMR52

60 Endress+Hauser Process Process temperature, Process pressure The specified pressure range may be reduced due to the selected process connection. The pressure rating (PN) specified on the flanges refers to a reference temperature of 20 °C, for ASME flanges 100 °F. Pay attention to pressure-temperature dependencies. Please refer to the following standards for the pressure values permitted for higher temperatures: • EN 1092-1: 2001 Tab. 18 With regard to their temperature stability properties, the materials 1.4435 and 1.4404 are grouped under 13E0 in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical. • ASME B 16.5a - 1998 Tab. 2-2.2 F316 • ASME B 16.5a - 1998 Tab. 2.3.8 N10276 • JIS B 2220 FMR51 p p -40 (-40) -40 (-40) -20 (-4) +150 (+302) +250 (+482) 0 (+32) 0 (+32) Tp [°C] ([°F]) Tp [°C] ([°F]) 1 2 3 [bar] ([psi]) [bar] ([psi]) 40 (580) 40 (580) -1 (-14.5) -1 (-14.5) 0 (0) 0 (0) A0022656-EN  17 FMR51: Admissible range of process temperature and process pressure 1 Seal: Graphite (Feature 090 "Seal", Option D3) 2 Seal: Viton (Feature 090 "Seal", Option A5) 3 Seal: Kalrez (Feature 090 "Seal", Option C1)

Micropilot FMR51, FMR52

Endress+Hauser 61 p -196 (-321) -100 (-148) +310 (+590) +350 (+662) +400 (+752) +450 (+842) 0 (+32) Tp [°C] ([°F]) 160 (2320) 154 (2233) 148 (2146) 142 (2059) [bar] ([psi]) A0020555-EN  18 FMR51: Admissible range of process temperature and process pressure of the HT version (Feature 090 "Seal", Option D2) Feature 090 "Seal" Admissible process temperature Admissible process pressure A5: Viton GLT –40 to +150 °C (–40 to 302 °F) prel = –1 to 40 bar (–14.5 to 580 psi) C1: Kalrez –20 to +150 °C (–4 to 302 °F) D3: Graphite, 40...250°C/-40...482°F –40 to +250 °C (–40 to 482 °F) D2: Graphite, 196...450°C/-321...842°F (HT) –196 to +450 °C (–321 to 842 °F) prel = –1 to 160 bar (–14.5 to 2 320 psi) Observe the temperature-dependent restrictions indicated in the diagram above 1). 1) For devices with CRN approval the pressure range may be reduced →  92 FMR52 p -40 (-40) 0 (+32) Tp [°C] ([°F]) +200 (+392) +100 (+212) 2 1 [bar] ([psi]) 25 (362.5) -1 (-14.5) 0 (0) A0023250-EN  19 FMR52: Admissible range of process temperature and process pressure 1 Valid for the following process connections: 6" (ANSI), DN150 (EN), 10K150A (JIS) 2 Valid for all other types of process connection

Micropilot FMR51, FMR52

62 Endress+Hauser Feature 100 "Process Connection" Admissible process temperature Admissible process pressure • AFK: 2" 150lbs (ANSI) • AGK: 3" 150lbs (ANSI) • AHK: 4" 150lbs (ANSI) • CFK: DN50 PN10/16 (EN) • CGK: DN80 PN10/16 (EN) • CHK: DN100 PN10/16 (EN) • KFK: 10K 50A (JIS) • KGK: 10K 80A (JIS) • KHK: 10K 100A (JIS) • MRK: DIN11851 DN50 • MTK: DIN11851 DN80 • TDK: Tri-Clamp 2" • TFK: Tri-Clamp 3" • THK: Tri-Clamp 4" –40 to +200 °C (–40 to 392 °F) prel = –1 to 25 bar (–14.5 to 363 psi) 1) 2) • AJK: 6" 150lbs (ANSI) • CJK: DN150 PN10/16 (EN) • KJK: 10K 150A (JIS) • For T = –40 to +100 °C (–40 to 212 °F): prel = –1 to 25 bar (–14.5 to 363 psi) • For T = 100 to 200 °C (212 to 392 °F): prel = 0 to 25 bar (0 to 363 psi) 1) For devices with CRN approval the pressure range may be reduced →  92 2) Process connections up to 25 bar (363 psi) on request Dielectric constant • For liquids – εr ≥ 1.9 in free-field applications – εr ≥ 1.4 in stilling well • For bulk solids εr ≥ 1.6 For dielectric constants (DC values) of many media commonly used in various industries refer to: • the Endress+Hauser DC manual (CP01076F) • the Endress+Hauser "DC Values App" (available for Android and iOS)

Micropilot FMR51, FMR52

Endress+Hauser 63 Mechanical construction Dimensions Dimensions of the electronics housing 146 (5.75) 142 (5.59) 115.25 (4.54) ø108.5 (4.27) 78 (3.07) 90 (3.54) 98 (3.86)* ø103.5 (4.07) R110 A0011666  20 Housing GT18 (316L); Dimensions in mm (in) *for devices with integrated overvoltage protection. 144 (5.67) 113.5 (4.47) ø106 (4.17) 78 (3.07) 90 (3.54) 99 (3.9)* ø106 (4.17) R110 A0011346  21 Housing GT19 (Plastics PBT); Dimensions in mm (in) *for devices with integrated overvoltage protection. 145.5 (5.73) 141.5 (5.57) ø112 (4.41) ø108.5 (4.27) 78 (3.07) 90 (3.54) 98 (3.86)* ø103.5 (4.07) R110 61 (2.4) A0020751  22 Housing GT20 (Alu coated); Dimensions in mm (in) *for devices with integrated overvoltage protection.

Micropilot FMR51, FMR52

64 Endress+Hauser FMR51: Version T ≤ 150 °C (302 °F); without antenna extension A B C R L L 68 (2.68) 63 (2.48) 5 (0.2) 26 (1.02) b 5 (0.2) ød 55 ød øD øA D L 65 (2.56) b 2 (0.08) ød øD A0023383  23 FMR51: Version T < 150 °C (302 °F); without antenna extension; dimensions: mm (in) A Process connection: Thread B Process connection: Tri-Clamp ISO2852 C Process connection: Flange 316L D Process connection: Flange AlloyC>316L R Reference point of the measurement Valid for the following device versions Feature 090 "Seal": – A5: Viton GLT – C1: Kalrez • Dimensions of the horn antenna (L, ⌀d): →  68 • Dimensions of the process connection (⌀A, ⌀D, b): →  68

Micropilot FMR51, FMR52

Endress+Hauser 65 FMR51: Version T ≤ 150 °C (302 °F); with antenna extension A R L 68 (2.68) 26 (1.02) 5 (0.2) 100(3.94) ød 55 B C L 63 (2.48) b 5 (0.2) 100(3.94) ød øD øA D E 63 (2.48) 68 (2.68) b 47 (1.85) ød øD 101...1000 (3.98...39.3) L2 ø27(1.06) ø42.5(1.7) 55 26 (1.02) A0023384  24 MR51: Version T < 150 °C (302 °F); with antenna extension; dimensions: mm (in) A Process connections: Thread; Accessory mounted: 100mm/4" antenna extension B Prozessanschluss: Tri-Clamp ISO2852; Accessory mounted: 100mm/4" antenna extension C Process connection: Flange 316L; Accessory mounted: 100mm/4" antenna extension D Process connection: Thread; Accessory mounted: ..... mm/inch antenna extension E Process connection: Flange 316L;Accessory mounted: ..... mm/inch antenna extension R Reference point of the measurement Valid for the following device versions • Feature 090 "Seal": – A5: Viton GLT – C1: Kalrez • Feature 610 "Accessory mounted" – OM: 100mm/4" antenna extension (diagrams A, B, C) – OU: ..... mm antenna extension (101 ... 1000 mm) (diagrams D, E) – OV: ..... inch antenna extension (3,98 ... 39,3 in) (diagrams D, E) • Dimensions of the horn antenna (L, L2, ⌀d): →  68 • Dimensions of the process connections (⌀A, ⌀D, b): →  68

Micropilot FMR51, FMR52

66 Endress+Hauser FMR51: Version T ≤ 250 °C (482 °F) A B R L 122 (4.8) 116(4.57) 26 (1.02) b ød 55 øD C D 122 (4.8) 116(4.57) 26 (1.02) b 55 øD 47 (1.85) ød 101...1000 (3.98...39.3) L2 ø27 (1.06) ø42.5(1.7) A0023385  25 FMR51: Version T < 250 °C (482 °F); dimensions: mm (in) A Process connection: Thread B Process connection: Flange 316L C Process connection: Thread; Accessory mounted: ..... mm/inch antenna extension D Process connection: Flange 316L; Accessory mounted: ..... mm/inch antenna extension R Reference point of the measurement Valid for the following device versions Feature 090 "Seal": D3: Graphite, -40...250°C/-40...482°F • Dimensions of the horn antenna (L, ⌀d): →  68 • Dimensions of the process connection (⌀D, b): →  68

Micropilot FMR51, FMR52

Endress+Hauser 67 FMR51: Version T ≤ 450 °C (842 °F) L 141 (5.55) b ød øD R A0023386  26 FMR51: Version T < 450 °C (842 °F); dimensions: mm (in) Valid for the following device versions Feature 090 "Seal": D2: Graphite, -196...450°C/-321...842°F • Dimensions of the horn antenna (L, ⌀d): →  68 • Dimensions of the process connection (⌀D, b): →  68

Micropilot FMR51, FMR52

68 Endress+Hauser FMR51: Dimensions of the horn antenna Feature 070 "Antenna" Dimension BA: Horn 40mm/1-1/2" BB: Horn 50mm/2" BC: Horn 80mm/3" BD: Horn 100mm/4" L 86 mm (3.39 in) 115 mm (4.53 in) 211 mm (8.31 in) 282 mm (11.1 in) L2 51 mm (2.01 in) 80 mm (3.15 in) 176 mm (6.93 in) 247 mm (9.72 in) d 40 mm (1.57 in) 48 mm (1.89 in) 75 mm (2.95 in) 95 mm (3.74 in) FMR51: Dimensions of the process connections Flanges according to EN1092-1 (suitable for DIN2527) 1) Pressure rating Dimension Nominal diameter 2) DN50 DN80 DN100 DN150 PN10/16 b 18 mm (0.71 in) 20 mm (0.79 in) 20 mm (0.79 in) 22 mm (0.87 in) ⌀D 165 mm (6.5 in) 200 mm (7.87 in) 220 mm (8.66 in) 285 mm (11.2 in) PN25/40 b 20 mm (0.79 in) 24 mm (0.94 in) 24 mm (0.94 in) - ⌀D 165 mm (6.5 in) 200 mm (7.87 in) 235 mm (9.25 in) - PN63 b - - 30 mm (1.18 in) - ⌀D - - 250 mm (9.84 in) - PN100 b - - 36 mm (1.42 in) - ⌀D - - 265 mm (10.4 in) - 1) For flanges with a higher pressure rating than those listed in the product structure please contact Endress +Hauser. 2) see feature 100 "Process connection" Flanges according to JIS B2220 1) Nominal diameter 2) Pressure rating 2) Dimension 50A 80A 100A 150A 10K b 16 mm (0.63 in) 18 mm (0.71 in) 18 mm (0.71 in) 22 mm (0.87 in) D 155 mm (6.1 in) 185 mm (7.28 in) 210 mm (8.27 in) 280 mm (11 in) 63K b - - 44 mm (1.73 in) - D - - 270 mm (10.6 in) - 1) For flanges with a higher pressure rating than those listed in the product structure please contact Endress +Hauser. 2) see feature 100 "Process connection"

Micropilot FMR51, FMR52

Endress+Hauser 69 Flanges according to ANSI B16.5 1) Nominal diameter 2) Pressure rating 2) Dimension 2" 3" 4" 6" 150 lbs b 17.5 mm (0.69 in) 22.3 mm (0.88 in) 22.3 mm (0.88 in) 23.9 mm (0.94 in) D 150 mm (6 in) 190 mm (7.5 in) 230 mm (9 in) 280 mm (11 in) 300 lbs b 20.7 mm (0.81 in) 27 mm (1.1 in) 30.2 mm (1.2 in) 35 mm (1.4 in) D 165 mm (6.5 in) 210 mm (8.25 in) 255 mm (10 in) 320 mm (12.5 in) 1) For flanges with a higher pressure rating than those listed in the product structure please contact Endress +Hauser. 2) see feature 100 "Process connection" Tri-Clamp according to ISO 2852 Nominal diameter 1) Dimension DN40-51 (2") DN70-76.1 (3") A 64 mm (2.52 in) 91 mm (3.58 in) 1) see feature 100 "Process connection"

Micropilot FMR51, FMR52

70 Endress+Hauser FMR52 with dairy coupling DIN11851 R A B ø100 (3.94) ø49 (1.93) ø68.5 (2.7) ø81 (3.19) 14 (0.55) 29 (1.14) 5 (0.2) 6 (0.24) 90 (3.54) 122 (4.8) A0023387  27 FMR52 with dairy coupling DIN11851; dimensions: mm (in) A Antenna: Horn 50mm/2"; Process connection DIN11851 DN50 PN25 B Antenna: Horn 80mm/3"; Process connection DIN11851 DN80 PN25 R Reference point of the measurement Valid for the following device versions: Feature 100 "Process connection" – MRK: DIN11851 DN50 PN25 slotted nut, PTFE>316L (diagram A) – MTK: DIN11851 DN80 PN25 slotted nut, PTFE>316L (diagram B)

Micropilot FMR51, FMR52

Endress+Hauser 71 FMR52 with Tri-Clamp ISO2852 R A B øa øa ø81 (3.19) ø49 (1.93) 29 (1.14) 14 (0.55) 2.6 (0.1) 2.6 (0.1) 120 (4.72) 93(3.66) A0023388  28 FMR52 with Tri-Clamp ISO2852; dimensions: mm (in) A Antenna: Horn 50mm/2" B Antenna: Horn 80mm/3" R Reference point of the measurement Antenna 1) BO: "Horn 50mm/2" BP: Horn 80mm/3" Process connection 2) TDK: Tri-Clamp DN40-51 (2") TFK: Tri-Clamp DN70-76.1 (3") THK: Tri-Clamp DN101.6 (4") a 64 mm (2.52 in) 91 mm (3.58 in) 119 mm (4.69 in) 1) Feature 070 of the product structure 2) Feature 100 of the product structure

Micropilot FMR51, FMR52

72 Endress+Hauser FMR52 with flange R A B øb øb øc øc 28.6 (1.13) 13.8 (0.54) 4 (0.16) a a 4 (0.16) 94(3.7) 121(4.76) ø48 (1.89) ø75 (2.95) A0023389  29 Dimensions FMR52 with flange; dimensions: mm (in) A Antenna: Horn 50mm/2" B Antenna: Horn 80mm/3" R Reference point of the measurement Flanges according to EN1092-1 (suitable for DIN2527) Antenna 1) BO: Horn 50mm/2" BP: Horn 80mm/3" Process connection 2) CFK: Flange DN50 CGK: Flange DN80 CHK: Flange DN100 CJK: Flange DN150 a 20 mm (0.79 in) 20 mm (0.79 in) 20 mm (0.79 in) 22 mm (0.87 in) b 102 mm (4.02 in) 138 mm (5.43 in) 158 mm (6.22 in) 212 mm (8.35 in) c 165 mm (6.5 in) 200 mm (7.87 in) 220 mm (8.66 in) 285 mm (11.2 in) 1) Feature 70 of the product structure 2) Feature 100 of the product structure Flanges according to ANSI B16.5 Antenna 1) BO: Horn 50mm/2" BP: Horn 80mm/3" Process connection 2) FK: Flange 2" AGK: Flange 3" AHK: Flange 4" AJK: Flange 6" a 17.5 mm (0.69 in) 22.3 mm (0.88 in) 22.3 mm (0.88 in) 23.9 mm (0.94 in) b 92 mm (3.62 in) 127 mm (5 in) 158 mm (6.22 in) 212 mm (8.35 in) c 150 mm (6 in) 190 mm (7.5 in) 230 mm (9 in) 280 mm (11 in) 1) Feature 70 of the product structure 2) Feature 100 of the product structure Flanges according to JIS B2220 Antenna 1) BO: Horn 50mm/2" BP: Horn 80mm/3" Process connection 2) KFK: Flange 10K 50A KGK: Flange 10K 80A KHK: Flange 10K 100A KJK: Flange 10K 150A a 16 mm (0.63 in) 18 mm (0.71 in) 18 mm (0.71 in) 22 mm (0.87 in) b 96 mm (3.78 in) 127 mm (5 in) 151 mm (5.94 in) 212 mm (8.35 in) c 155 mm (6.1 in) 185 mm (7.28 in) 210 mm (8.27 in) 280 mm (11 in) 1) Feature 70 of the product structure 2) Feature 100 of the product structure

Micropilot FMR51, FMR52

Endress+Hauser 73 Weight Housing Part Weight Housing GT18 - stainless steel approx. 4.5 kg (9.9 lb) Housing GT19 - plastic approx. 1.2 kg (2.7 lb) Housing GT20 - aluminium approx. 1.9 kg (4.2 lb) Antenna and process connection Device Weight of antenna and process connection FMR51 max. 3.0 kg (6.6 lb) + weight of flange 1) FMR52 max. 4 kg (8.8 lb) + weight of flange 1) 1) For the weight of the flange refer to Technical Information TI00426F.

Micropilot FMR51, FMR52

74 Endress+Hauser Materials: GT18 housing (stainless steel, corrosion- resistant) 2.1 1 2.2 5.2 5.1 6 7 8 9 4 3 A0013788 No. Part Material 1 Housing CF3M similar to 316L/1.4404 2.1 Cover of the electronics compartment • Cover: CF3M (similar to 316L/1.4404) • Window: glass • Cover seal: NBR • Seal of the window: NBR • Thread-coating: Graphite-based lubricant varnish 2.2 Cover of the terminal compartment • Cover: CF3M (similar to 316L/1.4404) • Cover seal: NBR • Thread-coating: Graphite-based lubricant varnish 3 Cover lock • Screw: A4 • Clamp: 316L (1.4404) 4 Lock at the housing neck • Screw: A4-70 • Clamp: 316L (1.4404) 5.1 Dummy plug, cable gland, adapter or plug (depending on the device version) • Dummy plug, depending on the device version: – PE – PBT-GF • Cable gland: 316L (1.4404) or nickel-plated brass • Adapter: 316L (1.4404/1.4435) • Seal: EPDM • M12 plug: Nickel-plated brass 1) • 7/8" plug: 316 (1.4401) 2) 5.2 Dummy plug, cable gland or adapter (depending on the device version) • Dummy plug: 316L (1.4404) • Cable gland: 316L (1.4404) or nickel-plated brass • Adapter: 316L (1.4404/1.4435) • Seal: EPDM 6 Dummy plug or M12 socket (depending on the device version) • Dummy plug: 316L (1.4404) • M12 socket: 316L (1.4404) 7 Pressure relief stopper 316L (1.4404) 8 Ground terminal • Screw: A4 • Spring washer: A4 • Clamp: 316L (1.4404) • Holder: 316L (1.4404) 9 Nameplate • Plate: 316L (1.4404) • Groove pin: A4 (1.4571) 1) For the version with M12 plug the sealing material is Viton. 2) For the version with 7/8" plug, the sealing material is NBR.

Micropilot FMR51, FMR52

Endress+Hauser 75 Materials: GT19 housing (plastic) 2.1 1 2.2 5.2 5.1 6 7 8 9 4 A0013788 No. Part Material 1 Housing PBT 2.1 Cover of the electronics compartment • Cover glass: PC • Cover frame: PBT-PC • Cover seal: EPDM • Thread-coating: Graphite-based lubricant varnish 2.2 Cover of the terminal compartment • Cover: PBT • Cover seal: EPDM • Thread-coating: Graphite-based lubricant varnish 4 Lock at the housing neck • Screw: A4-70 • Clamp: 316L (1.4404) 5.1 Dummy plug, cable gland, adapter or plug (depending on the device version) • Dummy plug, depending on the device version: – PE – PBT-GF • Cable gland, depending on the device version: – Nickel-plated brass (CuZn) – PA • Adapter: 316L (1.4404/1.4435) • Seal: EPDM • M12 plug: Nickel-plated brass 1) • 7/8" plug: 316 (1.4401) 2) 5.2 Dummy plug, cable gland or adapter (depending on the device version) • Dummy plug, depending on the device version: – PE – PBT-GF – Nickel-plated steel • Cable gland, depending on the device version: – Nickel-plated brass (CuZn) – PA • Adapter: 316L (1.4404/1.4435) • Seal: EPDM 6 Dummy plug or M12 socket (depending on the device version) • Dummy plug: Nickel-plated brass (CuZn) • M12 socket: Nickel-plated GD-Zn 7 Pressure relief stopper Nickel-plated brass (CuZn) 8 Ground terminal • Screw: A2 • Spring washer: A4 • Clamp: 304 (1.4301) • Holder: 304 (1.4301) 9 Adhesive nameplate Plastic 1) For the version with M12 plug the sealing material is Viton. 2) For the version with 7/8" plug, the sealing material is NBR.

Micropilot FMR51, FMR52

76 Endress+Hauser Materials: GT20 housing (die-cast aluminum, powder- coated) 2.1 1 2.2 5.2 5.1 6 7 8 9 4 3 A0013788 Nr. Part Material 1 Housing, RAL 5012 (blue) • Housing: AlSi10Mg(<0,1% Cu) • Coating: Polyester 2.1 Cover of the electronics compartment; RAL 7035 (gray) • Cover: AlSi10Mg(<0,1% Cu) • Window: Glass • Cover seal: NBR • Seal of the window: NBR • Thread-coating: Graphite-based lubricant varnish 2.2 Cover of the terminal compartment; RAL 7035 (gray) • Cover: AlSi10Mg(<0,1% Cu) • Cover seal: NBR • Thread-coating: Graphite-based lubricant varnish 3 Cover lock • Screw: A4 • Clamp: 316L (1.4404) 4 Lock at the housing neck • Screw: A4-70 • Clamp: 316L (1.4404) 5.1 Dummy plug, cable gland, adapter or plug (depending on the device version) • Dummy plug, depending on the device version: – PE – PBT-GF • Cable gland, depending on the device version: – Nickel-plated brass (CuZn) – PA • Adapter: 316L (1.4404/1.4435) • Seal: EPDM • M12 plug: Nickel-plated brass 1) • 7/8" plug: 316 (1.4401) 2) 5.2 Dummy plug, cable gland or adapter (depending on the device version) • Dummy plug, depending on the device version: – PE – PBT-GF – Nickel-plated steel • Cable gland, depending on the device version: – Nickel-plated brass (CuZn) – PA • Adapter: 316L (1.4404/1.4435) • Seal: EPDM 6 Dummy plug or M12 socket (depending on the device version) • Dummy plug : Nickel-plated brass (CuZn) • M12 socket: Nickel-plated GD-Zn 7 Pressure relief stopper Nickel-plated brass (CuZn)

Micropilot FMR51, FMR52

Endress+Hauser 77 Nr. Part Material 8 Ground terminal • Screw: A2 • Spring washer: A2 • Clamp: 304 (1.4301) • Holder: 304 (1.4301) 9 Adhesive nameplate Plastic 1) For the version with M12 plug the sealing material is Viton. 2) For the version with 7/8" plug, the sealing material is NBR. Materials: Antenna and process connection FMR51 A E F G H D B C 1 7 7 3 2 2 5 2 4 3 6 3 3 6 2 2 7 A0018961 A Version with screw-in adapter B XT version with threaded connection C XT version with flange D HT version with flange E Version with flange and antenna extension > 100 mm (3.94 in) to 500 mm (19.7 in) F Version with threaded connection and antenna extension > 100 mm (3.94 in) to 500 mm (19.7 in) G Version with clamp adapter and antenna extension 100 mm (3.94 in) H Version with flange and antenna extension 100 mm (3.94 in) Pos. Part Material 1 Process adapter 316L (1.4404) Locking washer 2 Horn 316L (1.4404) AlloyC22 (2.4602) Screws A4 AlloyC22 (2.4602) Spring-lock washer (compact version) A4 without spring-lock washer

Micropilot FMR51, FMR52

78 Endress+Hauser Pos. Part Material Nordlock washer (extension >100 mm, XT and HT version) A4 without Nordlock washer Process separation cone Standard: PTFE XT: PEEK HT: Al2O3 Seal Standard: • Viton: FKM • Kalrez: FFKM (K6375) XT: Graphite HT: Graphite 3 Flange 316L (1.4404/1.4435) optionally AlloyC22 (2.4602) clad 4 Tube extension + adapter 316L (1.4404) 5 Extension process adapter 316L (1.4404) Locking washer 6 Process adapter 316L (1.4404) 7 Housing adapter 316L (1.4404) Temperature reduction (XT) 316L (1.4404) Intermediate adapter (HT) 316L (1.4404) Process adapter 316L (1.4404) FMR52 A B C 1 2 4 4 4 2 3 1 1 A0018951 A Dairy coupling B Clamp adapter C Flange version Pos. Part Material 1 Adapter 304 (1.4301) 2 Process connection 316L (1.4435) 3 Flange 316L (1.4404/1.4435) 4 Cladding PTFE (in accordance with FDA 21 CFR 177.1550 and USP <88> Class VI)

Micropilot FMR51, FMR52

Endress+Hauser 79 Materials: Weather protection cover 1 2 3 4 5 7 6 A0015473 No Part: Material 1 Protection cover: 316L (1.4404) 2 Molded rubber part (4x): EPDM 3 Clamping screw: 316L (1.4404) + carbon fibre 4 Bracket: 316L (1.4404) 5 • Cheese head screw: A4-70 • Nut: A4 • Spring washer: A4 6 Ground terminal • Screw: A4 • Spring washer: A4 • Clamp: 316L (1.4404) • Holder: 316L (1.4404) 7 • Washer: A4 • Cheese head screw: A4-70

Micropilot FMR51, FMR52

80 Endress+Hauser Operability Operating concept Operator-oriented menu structure for user-specific tasks • Commissioning • Operation • Diagnostics • Expert level Operating languages • English • Deutsch • Français • Español • Italiano • Nederlands • Portuguesa • Polski • русский язык (Russian) • Svenska • Türkçe • 中文 (Chinese) • 日本語 (Japanese) • 한국어 (Korean) • Bahasa Indonesia • tiếng Việt (Vietnamese) • čeština (Czech) Feature 500 of the product structure determines which of these languages is preset on delivery. Quick and save commissioning • Interactive wizard with graphical interface for easy commissioning via FieldCare/DeviceCare • Menu guidance with brief explanations of the individual parameter functions • Standardized operation at the device and in the operating tools Integrated data storage device (HistoROM) • Enables transfer of configuration when changing electronic modules • Records up to 100 event messages in the device • Records up to 1000 measured values in the device • Saves the signal curve on commissioning which can later be used as a reference. Efficient diagnostics increase measurement reliability • Remedy information is integrated in plain text • Diverse simulation options and line recorder functions

Micropilot FMR51, FMR52

Endress+Hauser 81 Local operation Operation with Pushbuttons Touch Control Order code for "Display; Operation" Option C "SD02" Option E "SD03" A0032219 A0032221 Display elements 4-line display 4-line display white background lighting; switches to red in event of device error Format for displaying measured variables and status variables can be individually configured Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F) The readability of the display may be impaired at temperatures outside the temperature range. Operating elements local operation with 3 push buttons (, , ) external operation via touch control; 3 optical keys: , ,  Operating elements also accessible in various hazardous areas Additional functionality Data backup function The device configuration can be saved in the display module. Data comparison function The device configuration saved in the display module can be compared to the current device configuration. Data transfer function The transmitter configuration can be transmitted to another device using the display module. Operation with remote display and operating module FHX50 1 2 A0032215  30 FHX50 operating options 1 Housing of the remote display and operating module FHX50 2 Display and operating module SD02, push buttons; cover must be removed 3 Display and operating module SD03, optical keys; can be operated through the glass of the cover

Micropilot FMR51, FMR52

82 Endress+Hauser Remote operation Via HART protocol 1 4 5 7 9 6 8 2 3 A0028746  31 Options for remote operation via HART protocol 1 PLC (programmable logic controller) 2 Transmitter power supply unit, e.g. RN221N (with communication resistor) 3 Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475 4 Field Communicator 475 5 Computer with operating tool (e.g. DeviceCare/FieldCare , AMS Device Manager, SIMATIC PDM) 6 Commubox FXA191 (RS232) or FXA195 (USB) 7 Field Xpert SFX350/SFX370 8 VIATOR Bluetooth modem with connecting cable 9 Transmitter Via PROFIBUS PA protocol ENDRESS + HAUSER T PROFIBUS DP PROFIBUS PA 1 2 3 5 4 4 4 A0015775 1 Segment coupler 2 Computer with Profiboard/Proficard and operating tool (e.g. DeviceCare/FieldCare 3 PLC (Progrommable Logic Controller) 4 Transmitter 5 Additional functions (valves etc.)

Micropilot FMR51, FMR52

Endress+Hauser 83 Via FOUNDATION Fieldbus FF-H1 FF-H1 BT BT PS 3 4 PS BT SB BT FF-HSE IN LD ENDRESS + HAUSER E + – % 1 1 2 2 A0017188  32 FOUNDATION Fieldbus system architecture with associated components 1 FFblue Bluetooth modem 2 Field Xpert SFX350/SFX370 3 DeviceCare/FieldCare 4 NI-FF interface card IN Industrial network FF-HSE High Speed Ethernet FF-H1 FOUNDATION Fieldbus-H1 LD Linking Device FF-HSE/FF-H1 PS Bus Power Supply SB Safety Barrier BT Bus Terminator

Micropilot FMR51, FMR52

84 Endress+Hauser Via service interface (CDI) 1 2 3 A0032466 1 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface) 2 Commubox FXA291 3 Computer with DeviceCare/FieldCare operating tool

Micropilot FMR51, FMR52

Endress+Hauser 85 Integration in tank gauging system The Endress+Hauser Tank Side Monitor NRF590 provides integrated communications for sites with multiple tanks, each with one or more sensors on the tank, such as radar, spot or average temperature, capacitive probe for water detection and/or pressure sensors. Multiple protocols out of the Tank Side Monitor guarantee connectivity to nearly any of the existing industry standard tank gauging protocols. Optional connectivity of analog 4...20 mA sensors, digital I/O and analog output simplify full tank sensor integration. Use of the proven concept of the intrinsically safe HART bus for all on-tank sensors yields extremely low wiring costs, while at the same time providing maximum safety, reliability and data availability. NXA820 16...253 VAC 1 2 3 4 6 7 8 5 A0017982  33 The complete measuring system consists of: 1 Tankvision workstation 2 Commubox FXA195 (USB) - optional 3 Computer with operating tool (ControlCare) - optional 4 Level measuring device 5 Temperature measuring device 6 Tank Side Monitor NRF590 7 Pressure measuring device 8 Tankvision Tank Scanner NXA820

Micropilot FMR51, FMR52

86 Endress+Hauser SupplyCare inventory management software SupplyCare is a web-based operating program for coordinating the flow of material and information along the supply chain. SupplyCare provides a comprehensive overview of the levels of geographically distributed tanks and silos, for instance, providing complete transparency over the current inventory situation, regardless of time and location. Based on the measuring and transmission technology installed onsite, the current inventory data are collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide additional security for material requirements planning. The main functions of SupplyCare: Inventory visualization SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current and historical inventory data and calculated forecasts of future demand. The overview page can be configured to suit the user's preferences. Master data management With SupplyCare you can create and manage the master data for locations, companies, tanks, products and users, as well as user authorization. Report Configurator The Report Configurator can be used to create personalized reports quickly and easily. The reports can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be transmitted in many ways, such as by http, ftp or e-mail. Event management Events, such as when levels drop below the safety stock level or plan points, are indicated by the software. In addition, SupplyCare can also notify pre-defined users by e-mail. Alarms If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to the System Administrator and the Local System Administrator. Delivery planning The integrated delivery planning function automatically generates an order proposal if a pre-set minimum inventory level is undershot. Scheduled deliveries and disposals are monitored continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not going to be met as planned. Analysis In the Analysis module, the most important indicators for the inflow and outflow of the individual tanks are calculated and displayed as data and charts. Key indicators of material management are automatically calculated and form the basis for optimizing the delivery and storage process. Geographical visualization All the tanks and the tank inventories are represented graphically on a map (based on Google Maps). The tanks and inventory situations can be filtered by tank group, product, supplier or location. Multi-language support The multi-language user interface supports 9 languages, thereby enabling global collaboration on a single platform. The language and settings are recognized automatically using the browser settings. SupplyCare Enterprise SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server in an Apache Tomcat environment. The operators and administrators operate the application via a Web browser from their workstations.

Micropilot FMR51, FMR52

Endress+Hauser 87 6 . 10 9 1 3 4 6 5 8 7 2 A0034288  34 Example of inventory management platform with SupplyCare Enterprise SCE30B 1 SupplyCare Enterprise (via Web browser) 2 SupplyCare Enterprise installation 3 SupplyCare Enterprise on mobile devices (via Web browser) 4 Ethernet/WLAN/UMTS 5 Fieldgate FXA42 6 Power supply 24 V DC 7 Modbus TCP via Ethernet as server/client 8 Converter from Modbus to HART Multidrop 9 HART Multidrop 10 4 x 4 to 20 mA analog input (2-wire/4-wire)

Micropilot FMR51, FMR52

88 Endress+Hauser Cloud-based application: SupplyCare Hosting SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress +Hauser portal. 6 . 10 9 1 3 4 6 5 8 7 2 WWW A0034289  35 Example of inventory management platform with SupplyCare Hosting SCH30 1 SupplyCare Hosting installation in Endress+Hauser data center 2 PC workstation with Internet connection 3 Warehouse locations with Internet connection via 2G/3G with FXA42 or FXA30 4 Warehouse locations with Internet connection with FXA42 5 Fieldgate FXA42 6 Power supply 24 V DC 7 Modbus TCP via Ethernet as server/client 8 Converter from Modbus to HART Multidrop 9 HART Multidrop 10 4 x 4 to 20 mA analog input (2-wire/4-wire) With SupplyCare Hosting, users do not need to make the initial software purchase or install and run the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance the capability of the software in conjunction with the customer. The hosted version of SupplyCare is thus always up-to-date and can be customized to meet different customer requirements. Other services are also offered in addition to the IT infrastructure and the software that is installed in a secure, redundant Endress+Hauser data center. These services include defined availability of the global Endress+Hauser Service and Support Organization and defined response times in a service event.

Micropilot FMR51, FMR52

Endress+Hauser 89 Certificates and approvals Currently available certificates and approvals can be called up via the product configurator. CE mark The measuring system meets the legal requirements of the applicable EC guidelines. These are listed in the corresponding EC Declaration of Conformity together with the standards applied. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. RoHS The measuring system complies with the substance restrictions of the Restriction on Hazardous Substances Directive 2011/65/EU (RoHS 2). RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media Authority) requirements for network integrity, interoperability, performance characteristics as well as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic compatibility are met. The products are labelled with the RCM- Tick marking on the name plate. A0029561 Ex approval • ATEX • IEC Ex • CSA • FM • NEPSI • KC • INMETRO • TIIS (in preparation) Additional safety instructions must be observed for applications in hazardous areas. They are contained in the separate "Safety Instructions" (XA) document, which is included in the scope of delivery. Reference is made to the XA on the nameplate of the device. Details on the available certificats as well as the associated XAs can be found in the Associated documentation chapter in the Safety instructions section: →  110. Dual seal according to ANSI/ISA 12.27.01 The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the user to waive the use and save the cost of installing external secondary process seals in the conduit as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These instruments comply with the North-American installation practice and provide a very safe and cost- saving installation for pressurized applications with hazardous fluids. Further information can be found in the Safety Instructions (XA) of the relevant devices. Functional Safety Used for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous or inhomogeneous redundancy), independently assessed by TÜV Rhineland as per IEC 61508. Other information see documentation SD01087F: "Functional Safety Manual". WHG WHG approval: Z-65.16-524 Sanitary compatibility FMR52 with flange cladding made of PTFE conforms to FDA 21 CFR 177.1550 and USP <88> Class VI

Micropilot FMR51, FMR52

90 Endress+Hauser 3A and EHEDG approval with Tri-Clamp and DIN11851 process connection. To avoid risk of contamination, install according to the “Hygienic Equipment Design Criteria (HDC)” as stated in the Subgroup Design Principles of the EHEDG, Doc. 8 from April 2004. Suitable fittings and seals must be used to ensure hygiene-compliant design according to the specifications of 3-A SSI and EHEDG. The leak-tight connections can be cleaned with the cleaning methods usually used in this industry without leaving residues. The 316L housing (GT18), option B in the order attribute 40, is not suitable for applications which need a 3A approval. NACE MR 0175 / ISO 15156 For FMR51 and FMR54: • The wetted, metallic materials comply with the requirements of NACE MR 0175 / ISO 15156. • Declaration of conformity: see product structure, feature 580, option JB. NACE MR 0103 For FMR51 and FMR54: • The wetted, metallic materials comply with the requirements of NACE MR 0103. • The certificate of compliance is based on NACE MR 0175. The hardness and the intercrystalline corrision have been tested, as well as the heat treatment (solution annealed) has been carried out. The used materials comply therefore with the requirements of NACE MR 0103. • Declaration of conformity: see product structure, feature 580, option JE. Pressure equipment with allowable pressure ≤ 200 bar (2 900 psi) Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable pressure. Reasons: According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as "devices with an operational function and having pressure-bearing housings". If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber of its own), there is no pressure accessory present within the meaning of the Directive. Marine certificate Device Marine certificate 1) GL ABS LR BV DNV FMR51      FMR52      1) see oredering feature 590 "Additional approval" Radio standard EN302729-1/2 The devices Micropilot FMR50, FMR51, FMR52, FMR56 and FMR57 are conform with the LPR (Level Probing Radar) standard EN302729-1/2. The devices are allowed to be used inside or outside of closed bins or tanks in countries of th EU and the EFTA. Precondition is that the country itself already has implemented the directive. At present, the following countries have already implemented the directive: Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, Great Britain, Ireland, Iceland, Italy, Liechtenstein, Lithuania, Latvia, Malta, Netherland, Norway, Austria, Poland, Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic and Cyprus. All countries not mentioned above are at present in the implementation phase. For the use of the devices outside of closed bins or tanks, the following has to be observed: 1. The installation has to be done by trained personnel. 2. The antenna of the device has to be mounted on a fixed place and vertically to the bottom. 3. The mounting place has to be in a distance of at least 4 km to the mentioned astronomical stations or an appropriate permission has to be issued by the national authority. If the device is mounted at a distance of 4 to 40 km from the stations mentioned, the maximum mounting height is restricted to 15 m (49 ft).

Micropilot FMR51, FMR52

Endress+Hauser 91 Astronomical stations Country Name of the station Geographical latitude Geographical longitude Germany Effelsberg 50°31'32" N 06°53'00" E Finland Metsähovi 60°13'04" N 24°23'37" E Tuorla 60°24'56" N 24°26'31" E France Plateau de Bure 44°38'01" N 05°54'26" E Floirac 44°50'10" N 00°31'37" W United Kingdom Cambridge 52°09'59" N 00°02'20" E Damhall 53°09'22" N 02°32'03" W Jodrell Bank 53°14'10" N 02°18'26" W Knockin 52°47'24" N 02°59'45" W Pickmere 53°17'18" N 02°26'38" W Italy Medicina 44°31'14" N 11°38'49" E Noto 36°52'34" N 14°59'21" E Sardinia 39°29'50" N 09°14'40" E Poland Krakow Fort Skala 50°03'18" N 19°49'36" E Russia Dmitrov 56°26'00" N 37°27'00" E Kalyazin 57°13'22" N 37°54'01" E Pushchino 54°49'00" N 37°40'00" E Zelenchukskaya 43°49'53" N 41°35'32" E Sweden Onsala 57°23'45" N 11°55'35" E Switzerland Bleien 47°20'26" N 08°06'44" E Spain Yebes 40°31'27" N 03°05'22" W Robledo 40°25'38" N 04°14'57" W Hungary Penc 47°47'22" N 19°16'53" E In general the requirements of EN 302729-1/2 have to be taken in consideration. Radio standard EN302372-1/2 The devices Micropilot FMR50, FMR51, FMR52, FMR53, FMR54, FMR56 and FMR57 are conform with the TLPR (Tanks Level Probing Radar) standard EN302372-1/2 and can always be used in closed tanks or bins. For installation, points a to f in Annex B of EN302372-1 have to be taken into account. FCC / Industry Canada This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Canada CNR-Gen Section 7.1.3 This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) This device may not interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement. [Any] changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.

Micropilot FMR51, FMR52

92 Endress+Hauser In addition, the devices FMR50 8), FMR51 9), FMR52 10), FMR56 and FMR57 are compliant with the LPR (Level probe radar) regulation also for free space applicactions according to the FCC Code of Federal Regulations, CFR 47, Part 15, Sections 15.205, 15.207, 15.209, 15.256 for antenna sizes bigger than 50 mm (2.0 in) 11). For these applications the devices must be professionally installed in a downward operating position. In addition, the devices are not allowed to be mounted in a zone of 4 km around RAS stations and within a radius of 40 km around RAS stations the maxium operation height of devices is 15 m (49 ft) above ground. Japanese radio approval The devices FMR50, FMR51, FMR52, FMR54 and FMR57 comply with the Japanese Radio Law, Article 6, Section 1(1). CRN approval Some device versions have a CRN approval. Devices are CRN approved if the folowing two conditions are met: • The device has a CSA approval (Product structure: Feature 010 "Approval") • The device has a CRN approved process connection according to the following table. Feature 100 of the product structure Process connection AFJ NPS 2" Cl.150 RF, 316/316L AFK NPS 2" Cl.150, PTFE >316/316L AFM NPS 2" Cl.150, AlloyC >316/316L AGJ NPS 3" Cl.150 RF, 316/316L AGK NPS 3" Cl.150, PTFE >316/316L AGM NPS 3" Cl.150, AlloyC >316/316L AHJ NPS 4" Cl.150 RF, 316/316L AHK NPS 4" Cl.150, PTFE >316/316L AHM NPS 4" Cl.150, AlloyC >316/316L AJJ NPS 6" Cl.150 RF, 316/316L AJK NPS 6" Cl.150, PTFE >316/316L AJM NPS 6" Cl.150, AlloyC >316/316L ARJ NPS 2" Cl.300 RF, 316/316L ARM NPS 2" Cl.300, AlloyC >316/316L ASJ NPS 3" Cl.300 RF, 316/316L ASM NPS 3" Cl.300, AlloyC >316/316L ATJ NPS 4" Cl.300 RF, 316/316L ATM NPS 4" Cl.300, AlloyC >316/316L MRK DIN11851 DN50 PN25 slotted nut, PTFE>316L MTK DIN11851 DN80 PN25 slotted nut, PTFE>316L RGJ Thread ANSI MNPT1-1/2, 316L RVJ Thread EN10226 R1-1/2, 316L TDJ Tri-Clamp ISO2852 DN40-51 (2"), 316L TDK Tri-Clamp ISO2852 DN40-51 (2"), PTFE>316L, 3A, EHEDG TFJ Tri-Clamp ISO2852 DN70-76.1 (3"), 316L 8) Except of FMR50-######BM* (Horn 40mm/1-1/2", PVDF encapsulated) 9) Except for FMR51-######BA* (Horn 40mm/1-1/2") and FMR51-######BB* (Horn 50mm/2") 10) Except for FMR52-######BO* (Horn 50mm/2", flush mount) 11)

Micropilot FMR51, FMR52

Endress+Hauser 93 Feature 100 of the product structure Process connection TFK Tri-Clamp ISO2852 DN70-76.1 (3"), PTFE>316L, 3A, EHEDG THK Tri-Clamp ISO2852 DN101.6 (4"), PTFE>316L, 3A, EHEDG • Process connections without CRN approval are not included in this table. • Refer to the product structure to see which process connections are available for a specific device type. • For some process connections not listed in the product structure a CRN approval is available on request. • CRN approved devices are marked with the registration number OF15872.5C on the nameplate. For device versions listed in the table below, the maximum allowed pressure is reduced if they have a CRN approval. For device versions not listed in the table below the pressure range as indicated in the "Process" chapter →  60 is not affected by the CRN approval. Product Antenna 1) Process connection 2) Seal 3) max. pressure FMR52 TFK: Tri-Clamp 3" 10 bar (145 psi) THK: Tri-Clamp 4" 10 bar (145 psi) FMR51 D2: Graphite (HT) 100 bar (1 450 psi) 1) Feature 070 of the product structure 2) Feature 100 of the product structure 3) Feature 090 of the product structure Track record FMR5x is the upgrade model of the corresponding FMR2xx series.

Micropilot FMR51, FMR52

94 Endress+Hauser Test, Certificate Feature 580 "Test, Certificate" Designation Available for JA 3.1 Material certificate, wetted metallic parts, EN10204-3.1 inspection certificate FMR51 JB Conformity to NACE MR0175, wetted metallic parts FMR51 JD 3.1 Material certificate, pressurized parts, EN10204-3.1 inspection certificate FMR52 JE Conformity to NACE MR0103, wetted metallic parts FMR51 JF Conformity to AD2000, wetted metallic parts: Material of all wetted/pressurized parts conform to AD2000 (Technical rules W2, W9, W10) • FMR51 • FMR52 KD Helium leak test, internal procedure, inspection certificate FMR51 KE Pressure test, internal procedure, inspection certificate • FMR51 • FMR52 KG 3.1 Material certificate+PMI test (XRF), internal procedure, wetted metallic parts, EN10204-3.1 inspection certificate • FMR51 • FMR52 KP Liquid penetrant test AD2000-HP5-3(PT), wetted/pressurized metallic parts, inspection certificate FMR51 KQ Liquid penetrant test ISO23277-1 (PT), wetted/pressurized metallic parts, inspection certificate FMR51 KR Liquid penetrant test ASME VIII-1 (PT), wetted/pressurized metallic parts, inspection certificate FMR51 KS Welding documentation, wetted/pressurized seams consisting of • Drawing • WPQR (Welding Procedure Qualification Report) • WPS (Welding Process Specification) • Manufacturer Declaration FMR51 KV Conformity to ASME B31.3: The dimensions, materials of construction, pressure / temperature ratings and identification markings of the device comply with the requirements of ASME B31.3 • FMR51 • FMR52 Test reports, declarations and material test certificates can be obtained electronically from the W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer) This applies to the options of the following ordering features: • 550 "Calibration" • 580 "Test, Certificate" • 590 "Additional Approval", option LW: "CoC-ASME BPE" Product documentation on paper Test reports, declarations and material certificates can be ordered as hardcopies via ordering feature 570 "Service", option I7 „Product documentation on paper“. The printouts will be enclosed in the product delivery.

Micropilot FMR51, FMR52

Endress+Hauser 95 Other standards and guidelines • EN 60529 Degrees of protection by housing (IP code) • EN 61010-1 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures. • IEC/EN 61326 "Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC requirements) • NAMUR NE 21 Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment. • NAMUR NE 43 Standardization of the signal level for the breakdown information of digital transmitters with analog output signal. • NAMUR NE 53 Software of field devices and signal-processing devices with digital electronics • NAMUR NE 107 Status classification as per NE107 • NAMUR NE 131 Requirements for field devices for standard applications • IEC61508 Functional safety of electrical/electronic/programmable electronic safety-related systems

Micropilot FMR51, FMR52

96 Endress+Hauser Ordering information Ordering information Detailed ordering information is available from the following sources: • In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate" -> Select your country -> Click "Products" -> Select the product using the filters and search field -> Open product page -> The "Configure" button to the right of the product image opens the Product Configurator. • From your Endress+Hauser Sales Center: www.addresses.endress.com Product Configurator - the tool for individual product configuration • Up-to-the-minute configuration data • Depending on the device: Direct input of measuring point-specific information such as measuring range or operating language • Automatic verification of exclusion criteria • Automatic creation of the order code and its breakdown in PDF or Excel output format • Ability to order directly in the Endress+Hauser Online Shop

Micropilot FMR51, FMR52

Endress+Hauser 97 3-point linearity protocol The following notes must be taken into account if option F3 ("3 point linearity protocol") has been selected in feature 550 ("Calibration"). The 3 points of the linearity protocol are defined as follows: R A 500 (19.7) 6000 (236) 1 3 2 A0023272  36 Points of the 3-point linearity protocol; Dimensions: mm (in) A Distance from reference point R to first measuring point R Reference point of the measurement 1 First measuring point 2 Second measuring point (centrally between first and third measuring point) 3 Third measuring point Measuring point Position 1st measuring point • At the distance A from the reference point • A = antenne length + antenna extension (if present) + 500 mm (19.7 in) • Minimum distance: Amin = 1 000 mm (39.4 in) 2nd measuring point centrally between 1st and 3rd measuring point 3rd measuring point 6 000 mm (236 in) below the reference point, R The position of the measuring points may vary by ±1 cm (±0.04 in). The linearity is checked under reference conditions.

Micropilot FMR51, FMR52

98 Endress+Hauser 5-point linearity protocol The following notes must be taken into account if option F4 ("5 point linearity protocol") has been selected in feature 550 ("Calibration"). The five points of the linearity protocol are evenly distributetd across the measuring range (0% to 100%). In order to define the measuring range, Empty calibration (E) and Full calibration (F) have to be specified 12). The following restrictions have to be taken into account when defining E and F: R A F E A0017983 Device version Minimum distance between reference point (R) and 100% level Minimum span Maximum value for "Empty calibration" FMR51/FMR52 Without antenna extension A ≥ antenna length + 200 mm (8 in) 1) F ≥ 400 mm (16 in) E ≤ 24 m (79 ft) FMR51 with 100 mm/4" antenna extension 2) A ≥ antenna length + antenna extension 100 mm (4 in) + 200 mm (8 in) FMR51 with varaible antenna extension 3) A ≥ antenna length + maximum antenna extension 1 000 mm (40 in) 4) + 200 mm (8 in) 1) Minimum value: A ≥ 400 mm (16 in) 2) Feature 610 "Accessory Mounted", option OM 3) Feature 610 "Accessory Mounted", option OU or OV 4) This value is independent of the actual size of the antenna extension. The linearity is checked under reference conditions. The selected values of Empty calibration and Full calibration are only used to record the linearity protocol and are reset to their probe specific default values thereafter. If values different from the default are required, they must be ordered as a customized parametrization →  99. 12) If E and F are not specified, probe dependent default values will be used instead.

Micropilot FMR51, FMR52

Endress+Hauser 99 Customized parametrization If the option IJ "Customized parametrization HART", IK "Customized parametrization PA" or IL "Customized parametrization FF" has been selected in feature 570 "Service", customer specific presettings can be selected for the following parameters: Parameter Communication Selection list / range of values Setup → Distance unit • HART • PA • FF • in • ft • mm • m Setup → Empty calibration • HART • PA • FF max. 70 m (230 ft) Setup → Full calibration • HART • PA • FF max. < 70 m (230 ft) Setup → Adv. Setup → Current output 1/2 → Damping HART 0 to 999.9 s Setup → Adv. Setup → Current output 1/2 → Failure mode HART • Min • Max • Last valid value Setup → Adv. Setup → Current output 1/2 → Burst mode HART • Off • On Tagging (TAG) Ordering feature 895: Marking Option Z1: Tagging (TAG), see additional spec. Position of the measuring point marking To be selected in the additional specifications: • Tag plate Stainless Steel • Self-adhesive paper label • Supplied label/plate • RFID TAG • RFID TAG + Tag plate Stainless Steel • RFID TAG + Self-adhesive paper label • RFID TAG + Supplied label/plate Definition of the measuring point designation To be defined in the additional specifications: 3 lines containing up to 18 characters each The measuring point designation appears on the selected label and/or the RFID TAG. Designation in the Electronic Name Plate (ENP) The first 32 characters of the measuring point designation Designation on the display module The first 12 characters of the measuring point designation Services The following services can be selected via the product structure in the Product Configurator 13): • PWIS free (PWIS: paint wetting impairment substances) • Customized parametrization HART →  99 • Customized parametrization PA→  99 • Customized parametrization FF →  99 • Without tooling DVD (FieldCare) 13) Feature 570 of the product structure

Micropilot FMR51, FMR52

100 Endress+Hauser Application Packages Heartbeat Diagnostics Availability Available in all device versions. Function • Continuous self-monitoring of the device. • Diagnostic messages output to – the local display. – an asset management system (e.g. FieldCare/DeviceCare). – an automation system (e.g. PLC). Advantages • Device condition information is available immediately and processed in time. • The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation NE 107 and contain information about the cause of the error and remedial action. Detailed description See Operating Instructions of the device (→  110); chapter "Diagnostics and trouble shooting"