Fluke 5320A-LOAD Instruction Sheet

Specifications

Brand: Fluke
Category: Adapter
Model: 5320A-LOAD (Instruction Sheet)
File Type: PDF
File Size: —
Pages: 6

Read Online

Read the Fluke 5320A-LOAD Instruction Sheet online using the built-in PDF viewer.

Text of the manual

PN 3039085 February 2008 ©2008 Fluke Corporation. All rights reserved. Printed in USA. Specifications subject to change without notice.

1 5320A-LOAD High Voltage Load Adapter Instruction Sheet Introduction The 5320A-LOAD High Voltage Load Adapter (hereafter the Load) is designed for loading a hipot tester to create leakage current while performing hipot tester calibration with the 5320A Electrical Safety Tester Calibrator (hereafter the Calibrator). As shown in Figure 1, this Load consists of a series of eight power-rated resistors configured to provide eight resistance taps from 10 kΩ to 5 MΩ. The Load is designed to withstand a maximum voltage of 1.2 kV to 5.5 kV, depending upon the selected resistor. COM 10 k 10 k 20 k 15 k 50 k 150 k 250 k 500 k 4 M 35 k 50 k 100 k 250 k 500 k 1 M 5 M ehq059 Figure 1. Load Schematic XWWarning To avoid possible electric shock or personal injury, use this Load only as specified in this Instruction Sheet or the 5320A Calibrator Users Manual. Preparing for Operation The Load is shipped with a power line adapter (to run the cooling fans), a grounding cable, and this Instruction Sheet. The power line adapter comes with five different mains plug adapters for compatibility with different power plug configurations. Prior to use, select the appropriate plug adapter which matches the local power mains outlet. After attaching the plug to the power line adapter it is ready for use with the Load. After placing the Load on top of the Calibrator, connect the grounding cable between the Load’s rear-panel ground terminal and the Calibrator’s rear-panel ground terminal. XW Warning To avoid electric shock from high- voltage energy, ensure the ground terminal on the rear of the Load is connected to the ground (GND) terminal on the rear of the Calibrator. Any application with non-grounded housing is strictly prohibited. W Caution To prevent damage to the Load, ensure the vents on the bottom of the Load and the fan exhaust vents are clear of obstructions for proper cooling. Connect one end of the power line adapter to the mating power input connector socket on the rear panel of the Load and the other end into the mains power outlet. WCaution Failure to have the fans operating may lead to over heating and component failure. 5320A-LOAD 1200V MAX 140W MAX 2000V MAX 110W MAX 2000V MAX 80W MAX 5500V MAX 300W MAX 5500V MAX 120W MAX 5500V MAX 60W MAX 5500V MAX 30W MAX 5500V MAX 6 W MAX COM 10k 35k 500k 1M 50k 100k 250k 5M 100 mA 40 mA 10 mA 4 mA 1mA POWER I O 20V MAX ehq102.eps Figure 2. 5320A-LOAD Front Panel

High Voltage Load Adapter

2 Turn on the power to the Load’s cooling fans by pressing the front-panel power switch so the “1” side of the switch is depressed (see Figure 2). A light on the power switch illuminates to indicate that power is connected to the fans. The power switch to the fans may be turned off when the Load is no longer being used. Using the Load for Leakage Current Testing The Load resistor should be selected based on the appropriate voltage amplitude used for the test or the value recommended in the hipot tester’s calibration procedure. W Caution To avoid damage, never exceed the maximum rated voltage, power, and current limit of the Load. Usage Limits for Safe Operation Safe operation of the load depends both on the amount of voltage and the length of time it is applied to the Load. For the 10 kΩ, 35 kΩ, 50 kΩ, 100 kΩ, and 250 kΩ resistors, the higher voltage levels have limits to the time that they can be safely applied. The maximum voltages can be applied for up to 3 minutes. Due to self heating, exceeding the 3-minute limit for high voltages may cause both degradation of performance and a permanent shift in the resistor values. However, reduced voltage levels can be safely applied for increasing longer times. At specific levels, sustained voltages can be applied indefinitely. This is graphically shown in Figure 3 for the 10 kΩ, 35 kΩ, 50 kΩ, 100 kΩ, and 250 kΩ resistor sets. Safe Area 0 1000 2000 3000 4000 5000 6000 1 10 100 Time (min) Voltage (V) 250 k 100 k 35k/50k 10k ehq101.eps Figure 3. Safe Area Chart As an example, the 100 kΩ resistor can withstand 5500 V for 3 minutes while it can withstand 4000 V for 60 minutes or longer. Performance and Verification Test With normal operation, the resistance values of the Load should be verified no less frequently than once per year. Also, the Load should be verified if there was a possibility of the resistance values changing due to excessive heating or power dissipation. To verify the electrical performance of the Load, use one of two following test methods to insure the load resistors are within specification. The first test method uses voltage and current levels that insure the resistors dissipate a reasonable power when being verified. The alternative second test method uses a multimeter to measure the load resistance. Both methods check to see if it is within 10 % of the nominal value. The multimeter test uses minimal voltage and current levels, dissipating negligible power when measuring the resistance values of the Load. Either test method can be used, but the voltage method is generally recommended as it verifies the resistance value while the resistor is dissipating actual power – similar to the normal usage of the Load. The voltage method signal levels are based on capabilities of the recommended voltage calibrators. These levels use voltages up to 1 kV. It is possible and acceptable to test using other sources to test at higher voltages, but staying within the limits of the Load’s safe area of operation. Table 1 lists the load resistance nominal value for each test terminal. The voltage test method requires applying a test voltage across each resistance, between the respective input and COM terminals. The resulting current through the resistor is measured, and the value of the resistance is calculated. A calibrator is used as the precision voltage source. A multimeter is recommended to measure the current. Figure 4 shows the test equipment setup. Refer to Table 1 for the respective sourced voltage level settings and the nominal currents to be measured.

High Voltage Load Adapter

Performance and Verification Test 3 Note A Fluke 5520A or 5500A Calibrator is recommended as a source for the voltage method due to its output voltage/current capabilities. A Fluke 8845A Digital Multimeter (or equivalent) is recommended for measurements with either test method. For the voltage test method, test each resistor as follows: 1. Apply the recommended voltage at power mains frequency (either 50 or 60 Hz) between the terminal for the resistor being measured and the COM terminal. 2. Measure the current flowing through the load. 3. Calculate the resistance by dividing the sourced voltage by the measured current (RL=Vs/Im). The calculated resistance should be within 10 % of the nominal resistance value found in Table 1. Repeat steps 1 through 3 for each Load resistor terminal, adjusting the applied voltage per Table 1. For the alternative multimeter test method, measure each resistor as follows: 1. Place the probes of a multimeter between the selected resistor input terminal and the COM terminal. 2. Read the measured resistor using the resistance mode of the multimeter, and note the value. 3. Insure the measured value is within 10 % of the nominal resistance. Repeat steps 1 through 3 for each Load resistor terminal per Table 1. POWER I O 0 • 1 2 3 4 5 6 7 8 9 ENTER M k m V Hz FIELD EDIT / + F OPR EARTH EXGRD SCOPE MENU PREV SHIFT RESET CE SETUP REF NEW TC MEAS ¡F µ n p W dBm sec ¡C A MULT x DIV ÷ MODES MORE STBY HI LO TRIG GUARD TC 20A NORMAL AUX 5520A CALIBRATOR SCOPE OUT V, , ,RTD A, -SENSE, AUX V 20V PK MAX 20V PK MAX CALIBRATION ENABLE SERIAL 1 FROM HOST IEEE-488 IN OUT / 63 Hz 47 Hz 600VA MAX FLUKE CORPORATION MADE IN USA PATENTS PENDING NO INTERNAL USER SERVICEABLE PARTS. REFER SERVICE TO QUALIFIED SERVICE PERSONNEL NORMAL WARNING: TO AVOID ELECTRIC SHOCK GROUNDING CONNECTOR IN POWER CORD MUST BE CONNECTED FUSE CAUTION FOR FIRE PROTECTION REPLACE ONLY WITH A 250V FUSE OF INDICATED RATING MAINS SUPPLY 100V/ 120V 220V/ 240V T5.0A 250V (SB) T2.5A 250V (SB) CHASSIS GROUND SERIAL 2 TO UUT 10 MHz MAX 5V P - P WARNING: TO AVOID PHYSICAL INJURY, INSURE THAT THE FILTER IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRUMENT TO CLEAN THE FILTER: -UNPLUG INSTRUMENT -REMOVE FILTER -FLUSH WITH SOAPY WATER -DRY BEFORE REINSTALLATION INSTALLED OPTIONS - SC300 - SC600 5320A-LOAD 1200V MAX 140W MAX 2000V MAX 110W MAX 2000V MAX 80W MAX 5500V MAX 300W MAX 5500V MAX 120W MAX 5500V MAX 60W MAX 5500V MAX 30W MAX 5500V MAX 6 W MAX COM 10k 35k 500k 1M 50k 100k 250k 5M 100 mA 40 mA 10 mA 4 mA 1mA POWER I O 20V MAX PROTECTIVE EARTH (GROUND) POWER INPUT 12V 10 VA MAX. NO USER SERVICEABLE PARTS CONTAINED WITHIN. HAZARDOUS VOLTAGES PRESENT. DO NOT REMOVE COVERS. SAFETY WARNING DISCONNECT POWER AND SIGNAL LEADS BEFORE SERVICING. READ MANUAL BEFORE USE. REFER SERVICE TO QUALIFIED PERSONNEL. FLUKE CORPORATION www.fluke.com DURING USE, TERMINAL MUST BE CONNECTED TO THE PROTECTIVE EARTH OF THE 5320A 8846A 6-1/2 DIGIT PRECISION MULTIMETER 0.20100 mA~ DCV ACV DCI ACI TRIG FREQ PERIOD TEMP ZERO INSTR SETUP MEAS SETUP RANGE F1 F2 F3 F4 F5 BACK MEMORY ANALYZE INPUT SENSE HI LO FRONT REAR 10A 1V 400 mA 4W HI LO 2W/4W V 300V FUSED 1000V CAT I 600V CAT II Front Rear ehq104.eps Figure 4. Test Equipment Connections for Alternate Test Method

High Voltage Load Adapter

4 Table 1. 5320-LOAD Resistance Verification Values Voltage Method – Verifying the Load Resistances while under Power Alternative DMM Verification Method Nominal Resistance Value Externally Applied Test Voltage at mains Frequency Nominally Required Test Current Measured Current Calculated Resistance RL=Vs/Im (±10 % of nominal) Measured Resistance (±10 % of nominal) 10 kΩ 200 Vac 20 mA 35 kΩ 315 Vac 9 mA 50 kΩ 300 Vac 6 mA 100 kΩ 600 Vac 6 mA 250 kΩ 1000 Vac 4 mA 500 kΩ 1000 Vac 2 mA 1 MΩ 1000 Vac 1 mA 5 MΩ 1000 Vac 0.2 mA Replaceable Parts Table 2 lists the replaceable parts for the Load. To contact Fluke, visit Fluke’s website at www.fluke.com or call one of the following numbers: USA and Canada: 1-888-99-FLUKE (1-888-993-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 China: +86-10-6512-3435-2 Singapore: +65 6799-5588 Anywhere in the world: +1-425-446-5500 Table 2. Replaceable Parts Description Fluke PN POWER ADAPTER 100-240V AC, 12 V DC, with five plug adapters 3132484 GROUNDING CABLE 3132491 General Specifications Power supply voltage........................................... AC adapter 100-240 V, output voltage 12 V @ 0.4 amps min. Warm-up time........................................................ Not applicable Specifications confidence level .......................... 99 % Temperature Operating Temperature ...................................... 5 °C to 40 °C Recommended Calibration Temperature (Tcal) . 23 °C Storage Temperature ......................................... -20 °C to +70 °C Altitude, Maximum Operating............................................................ 3,050 m (10,000 ft) Storage ............................................................... 12,200 m (40,000 ft) Dimensions ........................................................... 430 mm X 462 mm X 95 mm (16.9 in X 18.2 in X 3.7 in)

High Voltage Load Adapter

Electrical Specifications 5 Weight (net)........................................................... 3 kg (8 lb 4.5 oz) Power Consumption............................................. 5 W maximum Safety class........................................................... I according to EN 61010-1 Electrical Specifications Total resistance range.......................................... 10 kΩ to 5 MΩ Number of specific resistance values ................ 8 Tolerance to Nominal Value................................. 10 % (One year, Tcal ±5 °C) Maximum Ratings Nominal Value Max. Voltage Max. Dissipation Power Max. Time at Maximum Power 10 kΩ 1200 V 140 W Limited to 3 minutes (see Figure 3) 35 kΩ 2000 V 110 W Limited to 3 minutes (see Figure 3) 50 kΩ 2000 V 80 W Limited to 3 minutes (see Figure 3) 100 kΩ 5500 V 300 W Limited to 3 minutes (see Figure 3) 250 kΩ 5500 V 120 W Limited to 3 minutes (see Figure 3) 500 kΩ 5500 V 60 W No limit 1 MΩ 5500 V 30 W No limit 5 MΩ 5500 V 5 W No limit

Other Adapter Manuals