ALERT-1 DIGITAL ALARM & SURVEILLANCE SYSTEMS FOR THE HEALTH CARE INDUSTRY SUITABLE FOR: • Hospitals • Laboratories • Nursing Homes • Clinics INSTALLATION & MAINTENANCE MANUAL

Table of Contents

Introduction I Installation 3 1. Power Module 4 2. Pressure Display module 5 3. Sensor Module 5 4. Status Module 6 5. Computer Interface Module 6 Electrical Connections 7 Trouble Shooting 8 Drawings 9 Wiring Diagrams - Area Alarm 10 - Master Alarm/Computer Interface 1 I - Power Supply 12 - Alternate Setpoint Calibration 13 April -1994 Model Number: A-MANUAL-ENG

The Amico medical gas alarm system (ALERT-1) incorporates the latest digital technology in alarm

arid surveillance systems. The system has been designed primarily for user flexibility and reliability. • Digital read-out for every gas service measured. • Bar-graph display for trend indication. • HI/LOW alarm indication with adjustable repeat times. • Interface to building management computer. • Easily expandable for additional gas services. • NO moving parts v e r y reliable. • Modular construction e a s e of maintenance. • Plexiglass front covere a s y to clean. • Adjustable repeat alarm Optional computer interface. • Meets requirements of CSA 2305.1. ALERT-1 is divided into four main modules which plug into the alarm box. 1. POWER SUPPLY MODULE This 'module converts the 115 VAC supply voltage into 24 VDC required for the alarm system and contains the audible alarm and lamp test circuitry. 2. PRESSURE/VACUUM DISPLAY MODULE This module provides a digital display of the actual pressure or vacuum of a gas being monitored. In addition a gas movement trend and HI/LOW alarm section is provided for quick identifica- tion. 3. SENSOR MODULE This module contains the transducer which converts the pressure (vacuum) source into an electric signal that is displayed b y the pressure display module. = I -.I-GAS Features include: MEDICAL CAS ALARM SYSTEM (((co)) P0v,'ER LIGHT AUDIBLE ALARM ...4.1,!2 TEST & I = " I --I-AL;-Rm SILENCE ANDO I N S T R U + 0 0 5 14101070. CANADA POWER MODULE —.1---PRESSURE & TREND INDICATOR LDV I GAS !]—•--VACUUM A L A R M & T R E N D NCIIInAL INDICATOR PRESSURE DISPLAY V A C U U M DiSPL

4. MASTER STATUS MODULE

This module will continuously monitor up to 8 remote signals for operating alarms and emergency alarms such as: pressure switches, low liquid levels, high or low gas pressures, etc. If any of the signals being monitored goes into an alarm condition, a red light will come on and an audible alarm will sound. M C D C & M R t S U P P L Y S T A I L S ( = roculad tow. . . a —ALARM INDICATORS maw. t o , Z O M R . MR. I _ h . * O R M I N O R R O W E MASTER STATUS MODULE ALERT-1 is factory-calibrated for"Normally Closed Circuit" (NC) operation so that the alarm function will be initiated on an open circuit. In this mode the alarm system will signal both alarm conditions and system malfunctions. PIGTAIL T E R M I N A L BLOCK 5. ALARM BOX These modules are pluged into the motherboard of the alarm box once it has been installed in the wall. FUSE TRANSFORMER MOTHERBOARD The box itself (Fig. 1), houses the pigtails that deliver the gases (vacuum) into the alarm and in addition contains: • transformer • main fuse • overload protection unit • terminal block (115VAC) • motherboard 2

A. BOX IN THE WALL

(MASTER STATUS & AREA ALARMS) 1. Remove the alarm box from the shipping carton and install the box securely in the roughed-in wall at the appropriate place. The mounting brack- ets are adjustable to suit the thickness of the wall. Do not remove the Window Frame or any Modules from the ship- ping carton until the box is mounted in the wall and all the plastering, external wiring, and brazing of the copper pig- tails is complete. (AREA ALARMS ONLY) 2. Once the box is installed in the wall remove the dust cover from the front. Using silver-brazing techniques, con- nect each pigtail to its appropriate gas (Vacuum) while ensuring that the bot- tom of the pigtail is wrapped with a damp cloth so as not to damage the check-valve inside the box. (Fig. 2) The system can now be pressure tested. B. MODULES CAUTION Fig. 2 The circuitry on ALERT-1 contains sophisticated integrated circuits and as such can be damaged by static. When installing the modules, hold the boards by the edges and do not touch any of the components or circuitry on the board. Static discharge to a component can cause the module to malfunction. During the installation of the modules, the supply (115 VAC) should not be connected. 3

1. POWER MODULE

(a) Remove the power module from its box by holding the metallic plate that covers the circuit board. (b) Plug the connector which is located at the end of the tail of the power module into the first (leftmost) receptacle on the motherboard. Since the connector is keyed it should slide into the receptacle easily. If resistance is encountered, it indicates the connector is not being inserted the right way. Ensure that there are no pins visible on either end of the motherboard connector. (c) Flip the power module so that it hangs by its cable just below the box. The circuit with all the components should now be visible. (d) Bring the power lines which protrude through the transformer cover up to the female socket on the power module and insert the male plug into the female socket as shown in Figure 3. The con- nector is keyed and hence should slide into the socket easily. Fig. 3 (e) The audible alarm is equipped with a repeater function preset at a 30 minute interval, but shipped disabled. If the repeater function is desired, move black "jumper" from "DIS" (disable) position to "ENABLE" position, then adjust screw dial at the top of module to any desired alarm repeat time between 30 and 2 minutes.

PRESSURE AND SENSOR MODULES

2. PRESSURE DISPLAY AND SENSOR MODULES These two modules are installed concurrently. The sensor module contains the transducer which is connected to the brass quick-disconnect. SENSOR M O D U L E NOTE: A LW AY S HOLD THIS MODULE BY T H E BRASS QUICK-DICONNECT FITTING. (Fig. 4) (a) Plug the brass male connector protruding from the top of the sensor module into the female receptacle in the box in the wall. The connector should click into place. (Fig. 4) (b) Remove the display module from its box and plug the connector atthe end of the tail into the appropriate connector on the motherboard just below the sensor module. (Fig. 5) (c) Flip the display module so that it hangs by its cable just below the box. The circuit with all the components should now be visible. (d) Plug the small cable leading from the end of the sensor module into terminal J2 on the pressure display module. (Fig. 4) (e) The unit has been factory calibrated and requires no adjustments. PRESSURE DISPLAY M O D U L E F i g . 4 Fig. 5 5

3. STATUS MODULE

(a) Plug the connector at the end of the tail of the Status module into the appro- priate place on the motherboard. (b) Flip the status module so that it hangs by its cable just below the box. The circuit with all the components should now be visible. (c) Connect the external switches to the appropriate terminal blocks on strip J4 as per the wiring diagram shown on page 11. 4. COMPUTER INTERFACE MODULE The computer interface module is a piggyback board that fits on top of the status module and plugs into connector (P1) located at the top of the status board. (Fig. 7) Plug the module in (if it is not already on) and attach the board using the five screws provided. Con- nect the module to the computer inter- face as shown in the wiring diagram on page 13. A C O N N E C T O R P 1 T E R M I N A L B L O C K ( J 4 ) M O U N T I N G S C R E W S Fig. 6

NOTE: POWER SHOULD BE BROUGHTTOTHE UNITONLY AFTER ALLTHE MODULES

ARE CONNECTED TO THE MOTHERBOARD. Remove the, transformer cover located at the left hand side of the box by unscrewing the nut at the center of the housing (nearthe fuse). Bring 115 VAC to the appropriately labeled terminal strip. Once the power is applied the display lights up and the buzzer should sound when the "push to test" button is depressed. For problem identification see the section on trouble shooting. When each module has been checked individually for proper operation. insert the screw at the top and bottom of the module to the module in the box. (Fig.8) D. COVER When all the modules have been put in the box, attach the window-frame to the box by means of the two screws on either: side of the frame. The alarm is now ready for use. S C R E W S T O A T T A C H F R A M E Fig. 8 7

D. TROUBLE SHOOTING

Solutions are shown in order of the most probable problem. PROBLEM SOLUTION 1. LIGHTS NOT ILLUMINATED ( a ) Check the fuse by removing the power module. The fuse is directly behind the power module (Fig. 8). Replace the fuse if defective. (b) Check the building electrical breaker to ensure power is ON. (c) Check the voltage at the terminal block above the transformer. (Fig. 1) Ensure 120 VAC is being applied. (d) Replace the power module. 2. N O AUDIBLE ALARM LIGHTS ( a ) Depress the "push to test" button. if the lights come on and NOT ILLUMINATED t h e r e is no audible, replace the power module, if not try solu tions for problem 1. 3. G A S READING INCORRECT ( a ) Ensure the gas going into the pigtails is being measured accu rately. (b) Ensure the sensor board is connected to the display board at terminal J2. (Fig. 4) (c) Ensure the display board is plugged into the motherboard. (Fig. 5) (d) Replace the sensor board. (e) Replace the display board. 4. ALARM CONDITION EXISTS ( a ) Ensure the system was properly ordered. AND THE ALARM UGHTS ARE N o r m a l l y HI pressure 60 psi. NOT GOING ON AT THE L O pressure 40 psi. PROPER SETTINGS L O vacuum 12 in Hg. If any other calibration is required check with your supplier. (b) G o to problem 3. 5. A N Y BURNT OUT LIGHT ( a ) Replace the module. 6. POWER LIGHT IS ON BUT ( a ) LIGHTS ON OTHER MODULES ARE NOT WORKING ( b ) (c) 8 Flat cables of each module are not connected to the mother board. Ensure flat cable on the power board is connected to the mother board. Ensure power lines from the transformer are connected to the power board. (Fig. 3)

(Diagrams in Inches)

J. 6 POSITION BOX -I- 3 . 3 0 0 4 - ;1.- I-"--' — -, 3.30 — • 3 250 --.-. 11,125 A LIMO MEDICAL GAS ALARM STSILLI 0o$ 0110 10010 1•1100•10. 11101110000010 0 N W = = = 1 Q n LDV U N M rr _.. / 14.000 16.400 3 POSITION BOX 6.125 r _ 0.750 -.- 0.750 1.250 02r) C O D 4.000 O 9. 00 9

FOR AREA ALARM

Power Module To Remote Alarm C Low High Back - Box Mother Board - Ribbon Cable Low High Ribbon Cable Pressure Module MP-MOD-L-GAS Sensor Module 10

To Computer Remote

Monitoring inputs WIRING DIAGRAM FOR MASTER ALARM Back - Box Mother Board T Ribbon Cable Optional Computer Interface Module COMMON POWER FAIL. MONITOR 0 0 9 POINT # B NO 0 0 8 POINT # 7 NO 0 0 7 POINT # 6 NO 0 0 6 POINT # 5 NO 0 0 5 POINT # 4 NO 0 0 4 PaNT # N O 0 0 3 POINT # 2 NO 0 0 2 POINT # I NO 0 0 1 C Field Equipment N.C. Common C N.C. C Common Common N.C. C Common Common N.C. C N.C. Common Common N.C. MP-MOD-E-MAS Common Typical examples for closed circuit operation. OXYGEN PRIMARY BULK TANK LIQUID LEVEL LOW OXYGEN RESERVE BULK TANK RESERVE IN USE OXYGEN RESERVE BULK TANK RESERVE LIQUID LEVEL LOW OXYGEN ABNORMAL LINE PRESSURE OXYGEN RESERVE BULK TANK LOW HEAD PRESSURE VACUUM LOW WATER PRESSURE VACUUM OVERLOAD/FAILURE LAG PUMP IN USE VACUUM LOW 11

FOR 120VAC SUPPLY

120VAC Supply 24VDC to Manifold Fuse 1/2 Amp. 0 0 N -0 0 0 cr N Back Box ransformer Power Module 12 24 + Gt4D + 15 MP-MOD-E-POW Power Plug Rectifier Connection to Optional "Remote Alarm"

C F I E L D CALIBRATION

AREA ALARM The Amico Alert-1 Series of Master and Area Alarms are precision electronic monitoring and alarm instruments that are factory calibrated and set according to the most current N.F.P.A. and CSA standards. The N.F.P.A. 1993 edition specifies that all pressure gas alarm points "shall indicate if the pressure in the piping system increases or decreases 20% from the NORMAL OPERATING pressure. NORMAL H I G H ALARM L O W ALARM Oxygen, Nitrous Oxide, Medical Air, Carbon Dioxide 5 0 PSI 6 0 PSI Nitrogen 1 6 5 PSI 1 9 5 PSI Vacuum (Low only) 2 1 Hg 1 2 Hg 40 PSI 140 PSI The Alarm is factory calibrated to the above set-points. The Alarm is also pre-calibrated to ALTERNATE OPERATING set-points for hospitals whose gas pressures are higher than normal. TO CHANGE TO ALTERNATE SETPOINTS: HIGH L O W ALTERNATE O P E R A T I N G A L A R M Oxygen, Nitrous Oxide, 5 5 PSI 6 6 PSI 4 4 PSI Medical Air, Carbon Dioxide Change jumper SW2 on the Area board as shown in the at- tached Figure. The calibrations on the Alarm should not be changed in the field. Should you require any other set-points please call your local Amico distributor or Amico direct at (905) 764-0800. 0 ( ) 7 1 7 , R22 SW4 F I__01 F125 s w 5 I n SW3 S W 1 SW2 6 Normal (513 PSI) SW2 Alternate (55 PSI) 13

Calibration Procedure for the Pressure Display Module

Setup: Connect the display PCB to a well calibrated Sensor module in a working frame and power supply. The calibration of the pressure module requires the following equipment: - Source of known pressure with a reliable gauge or pressure measuring instrument. Procedure: 1 Set the jumper selections SW1, SW2, SW3 and SW5 according to the pressure range to be calibrated. 2 Set the calibrate/operate switch (SW4) to the calibrate position. 3 Apply power to the module. 4 Disconnect the pressure sensor module and short terminals #1 and #4 of connector J2 on the display module. Adjust R22 for zero numerical indication. 5 Reconnect the pressure sensor module and apply a known, measured pressure (preferable at least 90% full scale) to the pressure transducer. 6 Adjust R25 for the appropriate numerical indication for the range and units of display settings of SW1, SW2, SW3 and SW5. Repeat steps 4,5 and 6 until satisfactory. 7 Apply an input for 50% of scale. Adjust R42 to place the bargraph display at center. 8 Confirm that the decimal select jumper SW3 will turn on either the middle or right hand decimal points. 9 Confirm that alarms are generated and that the "HI ALARM" and "LOW ALARM" LED's flash and the "NORMAL" LED extinguishes, by changing the pressure out of the safe area (+/- 20%), or by readjusting R42 to artificially move the LED bar display. 10 Put the calibrate/operate switch (SW4) to the operate position. The Pressure Display Module is factory calibrated for set points at 40 PSI (LOW)/60 PSI (HIGH). If you wish to change the settings, you would adjust "R42" to move the whole range up or down. DO NOT adjust "R22" and "R25", they are factory set at "0" and full scale. Do not adjust the potentiometers on the sensor modules, they are also factory set at "0" and full scale. To obtain a larger or smaller range you must change "SW2" and then adjust "R6" to obtain the proper range. You would then adjust "R42" to obtain the desired values. Calibration Procedure for the Pressure Display Module Setup: Connect the display PCB to a well calibrated Sensor module in a working frame and power supply. The calibration of the pressure module requires the following equipment: - Source of known pressure with a reliable gauge or pressure measuring instrument. Procedure: 1 S e t the jumper selections SW1, SW2, SW3 and SW5 according to the pressure range to be calibrated. 2 S e t the calibrate/operate switch (SW4) to the calibrate position. 3 A p p l y power to the module. 4 D i s c o n n e c t the pressure sensor module and short terminals #1 and #4 of connector J2 on the display module. Adjust R22 for zero numerical indication. 5 R e c o n n e c t the pressure sensor module and apply a known, measured pressure (preferable at least 90% full scale) to the pressure transducer. 6 A d j u s t R25 for the appropriate numerical indication for the range and units of display settings of SW1, SW2, SW3 and SW5. Repeat steps 4,5 and 6 until satisfactory. 7 A p p l y an input for 50% of scale. Adjust R42 to place the bargraph display at center. 8 C o n f i r m that the decimal select jumper SW3 will turn on either the middle or right hand decimal points. 9 C o n f i r m that alarms are generated and that the "HI ALARM" and "LOW ALARM" LED's flash and the "NORMAL" LED extinguishes, by changing the pressure out of the safe area (+/- 20%), or by readjusting R42 to artificially move the LED bar display. 10 P u t the calibrate/operate switch (SW4) to the operate position. The Pressure Display Module is factory calibrated for set points at 40 PSI (LOW)/60 PSI (HIGH). If you wish to change the settings, you would adjust "R42" to move the whole range up or down. DO NOT adjust "R22" and "R25", they are factory set at "0" and full scale. Do not adjust the potentiometers on the sensor modules, they are also factory set at "0" and full scale. To obtain a larger or smaller range you must change "SW2" and then adjust "R6" to obtain the proper range. You would then adjust "R42" to obtain the desired values.

1 2 O N 1

OFF 5 1 III SW2 S W 1 SW3 S W 5 2 R42 4 e 0 0 0 0 0 0 0 0 H o L o C o Pressure display adjustments a n d selections (component side view) Pressure Range a n d Scale O 7 6 0 m m H g 0 1 0 3 2Pa O 1 0 0 p s i 0 6 8 9 k P a 0 2 5 0 p s i O 3 0 in. H g SW1 N u m e r i c a l display range select SW2 B a r display optional g a i n a d j u s t select SW3 N u m e r i c a l display decimal p o i n t select SW4 C a l i b r a t e / operate switch SW5 T h i r d d i g i t o n / o f f select R22 D i s p l a y "Zero" a d j u s t m e n t R25 D i s p l a y "Span" a d j u s t m e n t R6 B a r g r a p h (non—standard) gain a d j u s t m e n t R42 B a r g r a p h " C e n t e r " a d j u s t m e n t Range sel. p o s ' n (SW1) D e c p o i n t sel. p o s ' n (SW3) 1111 1 III 1 III II II III I 1111 E l 2 The b a r display gain s e l e c t o r (SW2) i s n o r m a l l y s e t 6 a n d i s changed t o only when t h e r a n g e o f t h e bargraph i s t o b e different t h a n + / — 2 0 %. The calibrate / operate switch, SW4 i s n o r m a l l y s e t t o t h e lower position b u t i s t o b e m o v e d t o t h e u p p e r position t o reduce t h e time c o n s t a n t during calibration.