Agilent Technologies 1100 Series Reference Manual
Read Online
Read the Agilent Technologies 1100 Series Reference Manual online using the built-in PDF viewer.
Text of the manual
1100 Series AT Reference Manual
Notices
© Agilent Technologies, Inc. 2002
No part of this manual may be reproduced in
any form or by any means (including elec-
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ment and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G1330-90002
Edition
Edition 07/02
Printed in Germany
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn, Germany
Software Revision
This guide is valid for A.01.xx revisions of
the Agilent 1100 Series Autosampler Ther-
mostat software, where xx refers to minor
revisions of the software that do not affect
the technical accuracy of this guide.
Warranty
The material contained in this docu-
ment is provided “as is,” and is sub-
ject to being changed, without notice,
in future editions. Further, to the max-
imum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a par-
ticular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connec-
tion with the furnishing, use, or per-
formance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the sep-
arate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accor-
dance with the terms of such license.
Restricted Rights Legend
Software and technical data rights granted
to federal government customers include
only those rights customarily provided to
end user Customers of Software. Agilent
provides this customary commercial license
in Software and technical data pursuant to
FAR 12.211 (Technical Data) and FAR 12.212
(Computer Software) and, for Department of
Defense purchases, DFARS 252.227-7015
(Technical Data - Commercial Items) and
DFARS 227.7202-3 (Rights in Commercial
Computer Software or Computer Software
Documentation). If a federal government or
other public sector Customer has a need for
rights not conveyed under these terms, it
must negotiate with Agilent to establish
acceptable terms in a written agreement
executed by all relevant parties.
Safety Notices
CAUTION
A CAUTION notice denotes a haz-
ard. It calls attention to an operat-
ing procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated condi-
tions are fully understood and
met.
1100 Series AT Reference Manual
3
In This Guide…
1
Theory of Operation
Theory of operation of mechanical hardware, electronics,
and instrument interfaces
2
Installing the G1330B Thermostat
Site requirements and installation of the G1330B thermostat
3
Troubleshooting and Test Functions
The thermostatted autosampler‘s built-in troubleshooting
and test functions
4
Repairing the ALS Thermostat Module
Instructions on repair procedures of the ALS thermostat
5
ALS Thermostat Parts and Materials
Lists for identification of parts and materials
6
Specifications
Performance specifications of the ALS thermostat
1100 Series AT Reference Manual
5
Contents
1
Theory of Operation
Introduction to the Thermostatted Autosampler
10
ALS Thermostat Operation
12
ALS Thermostat Control and Electronics
14
Temperature Controlled Autosampler Board (TCA)
15
Electrical Connections
17
2
Installing the G1330B Thermostat
Site Requirements
20
Unpacking the Autosampler
23
Optimizing the Stack Configuration
25
Controller Requirements
27
Installing the G1330B Thermostat
28
Stage 1: Preparing the ALS Thermostat and Autosampler
29
Stage 2: Power Cable and Interface Cable Connection
32
Stage 3: Flow Connections
35
Stage 4: Installing the Sample Tray
36
Stage 5: Installing Tray Cover and Front Cover
39
Stage 6: Turning on the Thermostatted Autosampler
40
Stage 7: Update of Control Module Firmware
40
Stage 8: Update of the Agilent ChemStation Software
41
Transporting the Thermostatted Autosampler
43
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Overview of the Thermostat’s Indicators and Test Functions
46
Status Indicators
47
Power Supply Indicator
48
Instrument Status Indicator
48
ALS Thermostat Error Messages
49
Fan Failed
50
Temperature Control Failed (1 - 4)
51
Temperature Sensor Failed (1 - 4)
52
Lost Contact to ALS Thermostat
53
Power Fail for ALS Thermostat Module
54
Temperature out of Range
55
Bad Cooling / Heating Performance
56
4
Repairing the ALS Thermostat Module
Introduction into Repairing the Thermostat
58
Using the ESD Strap
59
Overview of the Repairing the Thermostat
60
Exchanging the Power Supply Fuses
61
Exchanging Internal Parts
62
Removing the Top Cover and Foam
63
Exchanging Heat Exchanger Fan
65
Exchanging the Heatsink Fans
67
Exchanging the Sensors
69
Exchanging the TCA board
72
Assembling the Main Cover
75
1100 Series AT Reference Manual
7
5
ALS Thermostat Parts and Materials
Main Assemblies
78
Accessory Kit G1330-68705
79
Main Assemblies
80
Foam Parts
81
Plastic Parts
82
Heat Exchanger Fan Parts
83
6
Specifications
Performance Specifications
86
A
Safety Information
General Safety Information
88
Radio Interference
91
Sound Emission
92
Agilent Technologies on Internet
93
Index
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
1
Theory of Operation
Introduction to the Thermostatted Autosampler
10
ALS Thermostat Operation
12
ALS Thermostat Control and Electronics
14
Temperature Controlled Autosampler Board (TCA)
15
Electrical Connections
17
1100 Series AT Reference Manual
1
Theory of Operation
Introduction to the Thermostatted Autosampler
The Agilent 1100 Series autosampler is designed for use with other modules of
the Agilent 1100 Series LC system, with the HP 1050 Series, or with other LC
systems if adequate remote control inputs and outputs are available. The
autosampler is controlled from the Agilent 1100 Series control module or from
the Agilent ChemStation for LC systems.
The specially-designed thermostattable sample trays holds either 100 × 1.8 ml
vials or two wellplates and 10 × 1.8 ml vials.
The ALS thermostat contains Peltier-controlled heat exchangers. A fan draws
air from the area above the sample vial tray of the autosampler and is then
blown through the fins of the cooling/heating module. There it is cooled or
heated according to the temperature setting. The thermostatted air enters the
autosampler through a recess underneath the specially-designed sample tray.
The air is then distributed evenly through the sample tray ensuring effective
temperature control, regardless of how many vials are in the tray.
In cooling mode condensation is generated on the cooled side of the Peltier
elements. This condensed water is safely guided into the leak system.
Theory of Operation
1
1100 Series AT Reference Manual
11
Figure 1
Overview of the ALS Thermostat
Power Supply
TCA Board
Heat sink fans
Heat exchanger Fan
Condensation leak drain
Heat exchanger
Heat sink
1100 Series AT Reference Manual
1
Theory of Operation
ALS Thermostat Operation
The thermostatted autosampler is equipped with a cooling/heating module
which uses Peltier elements for efficient air cooling. When turned on the front
side of the Peltier elements is heated/cooled according to the temperature
setting. A fan draws air from the sample tray area and blows it through the
channels of the heating/cooling module. The fan speed is determined
according to the environmental conditions (e.g., ambient temperature,
humidity). In the heating/cooling module the air reaches the temperature of
Figure 2
ALS Thermostat Principle
air channel in
autosampler sampling
unit
special sample tray
air outlet
Peltier elements
heat exchanger
heating/cooling
module
ambient air in
heated air out
thermostatted air
Theory of Operation
1
1100 Series AT Reference Manual
13
the Peltier elements and this thermostatted air is blown underneath the
special sample tray where it is evenly distributed and streams back into the
sample tray area. From there it is again drawn into the ALS thermostat. This
“recycle” mode assures a very efficient cooling/heating of the sample vials.
In cooling mode the opposite side of the Peltier element will become very hot
and to maintain the performance of the elements they have to be cooled down.
This is done with large heat exchangers in the back of the ALS thermostat.
Four fans blow air from left to right through the instrument to remove the
heated air. The fan speed is controlled according to the temperature of the
Peltier elements.
During cooling condensation will appear in the heating/cooling module. The
condensed water will be guided out of the ALS thermostat.
1100 Series AT Reference Manual
1
Theory of Operation
ALS Thermostat Control and Electronics
The ASM board of the autosampler module provides all control signals for the
ALS thermostat (e.g., temperature setting) and receives the sensor signals
from the temperature and fan sensors.
The TCA board controls the internal assemblies in the ALS thermostat. This
chapter describes the electronics of the TCA board.
Figure 3
Block Diagram ALS Thermostat Electronics
TCA Board
ASM Board
Peltier
Assembly A
Peltier
Assembly B
Peltier
Assemblies
C + D
Temperatur Sensor
Fan
1
1
2
2
3
4
5
3
4
Primary Board
Bridge Rectifier
+ 36 V
Line Voltage
Theory of Operation
1
1100 Series AT Reference Manual
15
Temperature Controlled Autosampler Board (TCA)
The board controls all information and activities of all assemblies within the
ALS thermostat. The operator enters parameters, changes modes and controls
the module through the control module or Agilent ChemStation, that is
connected to the autosampler module. Autosampler and ALS thermostat are
connected via a control cable for signal exchange.
Step Down Switcher
The four Peltier assemblies (A to D) are driven by synchronous step down
switching regulators. Switcher 1 and 2 receive their power (+36V) from the
ALS thermostat internal power supply. Switcher 3 receives its power (+24V)
from the autosampler module via the interface cable. The maximum power
consumption for all Peltier assemblies is approximately 180 Watts in cooling
mode and 120 Watts in heating mode.
Regulators
The fans are equipped with hall sensors that allow to determination of the
actual speed of the fans. The speed of the four heat sink fans is variable and
will be set according to the actual temperature at the heat sink side of the
Peltier assemblies. The heat exchanger fan is always running at highest speed.
the speed will be reduced to a minimum when the sample tray is removed from
the autosampler.
Auto Range
When the ALS thermostat power cable is connected to the module the auto
range circuits checks the connected supply voltage. Default setting is a
connection to a 220-240V supply. If the non-regulated 36V is below a certain
limit the auto range circuit activates a switch on the primary board in the
power supply and switches to 100-120V operation.
Multiplexer
The ALS thermostat is equipped with four digital temperature sensors for
temperature control and diagnostics. Four are used to check the temperature
on both sides of the Peltier elements. The temperature sensor signals and the
signals from the hall sensors in the five fans are the input signals for the
multiplexer. The output of the multiplexer is connected to the autosampler
1100 Series AT Reference Manual
1
Theory of Operation
main board via the 26 pin autosampler to ALS thermostat cable. The
autosampler firmware checks the signals for correctness and initiates all
needed activities.
EPROM
The EPROM stores all relevant data for the ALS Thermostat (e.g., serial
number, board revision, etc.). This data are pre-set at the factory.
Figure 4
Block Diagram TCA Board
Step Down
Switcher 1
Step Down
Switcher 1
EPROM
Multiplexer
Auto Range
Regulator 3
Regulator 2
Regulator 1
Step Down
Switcher3
PDC 1
+ 36 V
PDC 2
PDC 3
+ 24 V
+ 36 V
Temp. Sensor 1 to 4
Fan Sensor 1 to 5
to ASM Board
Primary Board
Fan 5
Fan 3 + 4
Fan 1 + 2
Peltier Assembly C + D
Peltier Assembly B
Peltier Assembly A
+ 36 V
Theory of Operation
1
1100 Series AT Reference Manual
17
Electrical Connections
WARNING
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Figure 5
Elctrical Connections
GPIB
RS232C
CAN-bus
Remote
Vial number output
Relay contacts
CAN cable to previous module
CAN cable to next module
Therm-ALS cable
Thermostat -
Autosampler cable
1100 Series AT Reference Manual
1
Theory of Operation
• The GPIB connector is used to connect the autosampler with a computer.
The address and control switch module next to the GPIB connector
determines the GPIB address of your autosampler. The switches are preset
to a default address (see Autosampler Reference Manual) and this is
recognized immediately after power on.
• The CAN bus is a serial bus with high-speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1100 Series
module data transfer and synchronization.
• The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
common shut down, prepare, and so on.
• The RS-232 connector may be used to control the autosampler from a
computer through an RS-232 connection, using appropriate software. This
connector needs to be activated by the configuration switch module next to
the GPIB connector. The software needs the appropriate drivers to support
this communication. See your software documentation for further
information.
• The Thermostat-Autosampler connection is used for control signal transfer
and synchronization of the two modules. The cable must be installed for
operation of the ALS thermostat.
• The power input socket accepts a line voltage of 100–120 or
220–240 volts AC ± 10 % with a line frequency of 50 or 60 Hz. Maximum
power consumption of the autosampler module is 300 Watts (Volt-Amps).
Maximum power consumption of the ALS thermostat module is 260 Watts
(Volt-Amps).There are no voltage selectors on your autosampler because the
power supplies have automatic selection capability. The autosampler
module has no externally accessible fuses, because automatic electronic
fuses are implemented in the power supply. The power supply of the ALS
thermostat has two externally accessible fuses. The security lever at the
power input socket prevents removal of the autosampler cover when line
power is still connected.
• The interface board slot is used for external contacts, BCD output and for
future use.
WARNING
DO NOT disconnect or reconnect the autosampler to ALS thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
2
Installing the G1330B Thermostat
Site Requirements
20
Unpacking the Autosampler
23
Optimizing the Stack Configuration
25
Controller Requirements
27
Installing the G1330B Thermostat
28
Transporting the Thermostatted Autosampler
43
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Site Requirements
A suitable site environment is important to ensure optimum performance of
the autosamplers.
Power Consideration
The autosamplers comprises two modules, the autosampler module (G1329A,
G1389A, G1367A, or G2260A) and the thermostat module (G1330B Therm).
Both modules have a separate power supply and a power plug for the line
connections. The two modules are connected by a control cable and both are
turned on by the autosampler module.
The autosampler power supplies have automatic voltage selectors (see Table 1
on page 22). Consequently there are no voltage selectors in the rear of the two
autosampler modules. The autosampler module has no externally accessible
fuses, because automatic electronic fuses are implemented in its power
supply. The ALS thermostat power supply has two externally accessible fuses.
Power Cords
Your autosampler modules are delivered with power cords which match the
wall socket of your particular country or region. The plug on the power cords
which connect to the rear of the two instrument is identical for all types of
power cord.
WARNING
To disconnect the autosampler from line power, unplug the power cord from the
autosampler and the thermostat. The power supplies still use some power, even if
the power switch on the front panel is turned off. Please make sure that it is always
possible to access the power plug.
WARNING
Shock hazard or damage of your instrumentation can result if the devices are
connected to a line voltage higher than specified.
WARNING
Never operate your instrumentation from a power outlet that has no ground
connection. Never use a power cord other than the power cord designed for your
region.
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
21
Bench Space
The autosampler dimensions and weight (see Table 1 on page 22) allow the
instrument to be placed on almost any laboratory bench. The instrument
requires an additional 25 cm (10 inches) of space on either side for the
circulation of air, and approximately 8 cm (3.1 inches) at the rear for electrical
connections. Ensure the autosampler is installed in a level position.
If a complete Agilent 1100 Series system is to be installed on the bench, make
sure that the bench is designed to carry the weight of all the modules. For a
complete system including the thermostatted autosampler it is recommended
to position the modules in two stacks, see "Optimizing the Stack
Configuration" on page 25. Make sure that in this configuration there is 25 cm
(10 inches) space on either side of the thermostatted autosampler for the
circulation of air.
Environment
Your autosampler modules will work at ambient temperatures and relative
humidity as described in Table 1 on page 22.
CAUTION
Do not store, ship or use your autosampler under conditions where temperature
fluctuations may cause condensation within the autosampler electronics.
Condensation will damage the system electronics. If your autosampler was shipped in
cold weather, leave it in its box, and allow it to warm up slowly to room temperature to
avoid condensation.
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Table 1
Physical Specifications - Thermostatted Autosampler
Type
Specification
Comments
Thermostat
Weight
20.7 kg (45.6 lbs)
Dimensions
(height × width × depth)
140 × 345 × 435 mm
(5.5 × 13.5 × 17 inches)
Line voltage
100 – 120 or 220 – 240 VAC, ± 10 %
Automatic selection
Line frequency
50 or 60 Hz, ± 5 %
Power consumption
Autosampler: 300 VA
ALS Thermostat: 260 VA
Maximum
Maximum
Ambient operating temperature
4 – 40 ∞C (41 – 131 ∞F)
see on page 22
Ambient non-operating temperature
-40 – 70 ∞C (-4 – 158 ∞F)
Humidity
< 95 %, at 25 – 40 ∞C (77 – 104 ∞F)
Non-condensing;
Operating Altitude
Up to 2000 m (6500 ft)
Non-operating altitude
Up to 4600 m (14950 ft)
For storing the
autosampler
Safety standards: IEC, CSA, UL, EN
Installation Category II, Pollution Degree 2
WARNING
Using the autosampler at environmental temperatures higher than 50 ∞C (122 ∞F)
may cause the rear panel to become hot.
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
23
Unpacking the Autosampler
Damaged Packaging
The two modules of the autosampler are shipped in separate boxes. Upon
receipt of your autosampler, inspect the shipping containers for any signs of
damage. If the containers or cushioning material are damaged, save them until
the contents have been checked for completeness and the autosampler has
been mechanically and electrically checked. If the shipping container or
cushioning material is damaged, notify the carrier and save the shipping
material for the carriers inspection.
Delivery Checklist
Unpack the two boxes of the autosampler. Ensure all parts and materials have
been delivered with the autosampler and the ALS thermostat. The delivery
checklist are shown in Table 2. Please report missing or damaged parts to your
local Agilent Technologies sales and service office.
If the thermostatted autosampler was ordered as an upgrade (G1395A) to an
existing autosampler, the shipment will also contain the required software
upgrades for your Agilent ChemStation.
CAUTION
If you intend to ship or transport the autosampler, always park the transport assembly
of the autosampler module before shipment (see “Transporting the Autosampler” in
the corresponding manual).
CAUTION
If there are signs of damage to the autosampler, please do not attempt to install the
autosampler.
Table 2
G1330B Thermostat Checklist
Description
Quantity
Part Number
ALS Thermostat G1330B
1
Power cable
1
as ordered
Accessory kit (Table 3)
1
G1330-68705
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Table 3
ALS Thermostat Accessory Kit Contents G1330-68705
Description
Part Number
Waste Tube
5063-6527
Waste Tube Assembly
G1330-67300
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
25
Optimizing the Stack Configuration
If your autosampler is part of a system, you can ensure optimum performance
by installing the autosampler in the stack in the position shown in Figure 6
and Figure 7. This configuration optimizes the system flow path, ensuring
minimum delay volume.
Figure 6
Recommended Stack Configuration (Front View)
Degasser
Pump
Solvent
Cabinet
Autosampler
Thermostat
Column
Compartment
Detector
Control Module
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Figure 7
Recommended Stack Configuration (Rear View)
GPIB to LC ChemStation
Analog signal
to recorder
AC power
Remote
cable
CAN bus cable
Autosampler -
Thermostat
cable
AC power
Analog signal
to recorder
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
27
Controller Requirements
Agilent 1100 Series control module and Agilent ChemStation must have the
latest revision firmware/software loaded to allow optimum operation with the
thermostattable autosamplers. Older revisions might not recognize or have full
functionality with the thermostattable autosampler.
Control Module Firmware requirements
The control module requires firmware revision A.01.30 or higher to control
the thermostatted autosampler. Previous firmware revisions of the control
module will not run with the thermostatted autosampler. If the control module
was shipped together with the thermostatted autosampler the control module
firmware does not require updating. The firmware update must be done with
a PCMCIA card that has the newest revision loaded. The firmware is not part
of the shipment of the thermostatted autosampler. Contact your local Agilent
Technologies sales and service office for the firmware update of the control
module. For the update procedure see "Stage 7: Update of Control Module
Firmware" on page 40.
Agilent ChemStation Software requirements
To control the thermostatted autosampler from a PC, the Agilent ChemStation
software version A.04.03, or A.05.02beta, or A.05.02 or higher is required.
These software revision might however not support all ALS modules. The
thermostatted autosampler will not run with any previous version of the
Agilent ChemStation software. The software updates are part of the
thermostatted autosampler shipment. For the update procedure see "Stage 8:
Update of the Agilent ChemStation Software" on page 41.
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Installing the G1330B Thermostat
WARNING
To avoid personal injury, keep fingers away from the needle area during autosampler
operation. Do not bend the safety flap away from its position, or attempt to remove
the safety cover (see Figure 8). Do not attempt to insert or remove a vial from the
gripper when the gripper is positioned below the needle.
Figure 8
Safety Flap.
Safety flap
Safety cover
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
29
Stage 1: Preparing the ALS Thermostat and Autosampler
1 Place the ALS thermostat on the bench or in the stack.
2 Remove the front cover. Press the two snap fasteners on the sides of the
cover and move it away.
3 If the ALS thermostat is located on top of another Agilent 1100 Series
Module place the waste tube assembly into the top cover of the ALS
thermostat and locate the other end in the waste funnel of the module
beneath.
Figure 9
Preparation of the ALS Thermostat
Condensation
leak tube
Waste tube
assembly
Module locks
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
4 Connect the condensation leak tube to the main waste exit of the ALS
thermostat and place into an appropriate vessel. It is possible to either let
the condensation leak tubing exit the module at the front or at the left side
of the module. Make sure that the leak tube is fully fixed on the outlet.
5 Install the front cover of the ALS thermostat.
6 Place the autosampler module on top of the ALS thermostat. Make sure that
the autosampler is correctly engaged in the ALS thermostat locks.
WARNING
Make sure that the condensation tube is always above the liquid level in the vessel.
If the tube is located in liquid the condensed water cannot flow out of the tube and
the outlet is blocked. Any further condensation will then remain in the instrument.
This may damage the instruments electronics.
Figure 10
Condensation Leak outlet
Condensation leak tube
Waste bottle
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
31
7 Place the air channel adapter into the autosampler tray base. Make sure the
adapter is fully pressed down. This assures that the cold airstream from the
ALS thermostat is correctly guided to the tray area of the autosampler.
8 If there is no Agilent 1100 Series module located beneath the ALS
thermostat connect the waste tube to the central waste exit of the
autosampler and place in a waste vessel.
Figure 11
Preparation of ALS Thermostat and Autosampler
Air channel adapter
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Stage 2: Power Cable and Interface Cable Connection
1 Ensure the power switch on the front of the autosampler is OFF and the
power cables are disconnected.
2 Connect the cable between the autosampler and the ALS thermostat, see
Figure 5 on page 30.
3 Move the safety lever at the rear of the two modules to the right position,
see Figure 5 on page 30.
4 Connect the power cables to the power connectors.
5 Connect the CAN interface cables to other modules in the system (see
Figure 7 on page 26 and Figure 13 on page 34).
6 If required, connect additional interface and control cables to the
autosampler (see Figure 7 on page 26 and Figure 13 on page 34). Refer to
the documentation of the Agilent 1100 Series control module or
ChemStation for LC for more information.
WARNING
Do not disconnect or reconnect the autosampler to ALS thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
NOTE
In an Agilent 1100 Series system, the individual modules are connected by a CAN cable.
The Agilent 1100 control module can be connected to the CAN bus at any of the modules in
the system. The Agilent Chemstation can be connected to the system by one GPIB cable at
any of the modules, however, it is recommended to connect the GPIB cable to the detector.
For more information about connecting the control module or ChemStation refer to the
respective user manual. For connecting the Agilent 1100 Series equipment to
non-Agilent 1100 Series equipment, see Autosampler manual).
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
33
7 Connect additional cables as required (see Figure 13 on page 34).
Figure 12
Power Connectors and Safety Levers at Rear of thermostatted Autosampler.
Safety lever
Power connector
Autosampler - Thermostat
Cable
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Figure 13
Cable Connections.
GPIB
RS232C
CAN-bus
Remote
Vial number output
Relay contacts
CAN cable to previous module
Therm-ALS cable
Thermostat -
Autosampler cable
CAN cable to next module
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
35
Stage 3: Flow Connections
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
valve.
3 Ensure that the waste tube is positioned inside the leak channel.
WARNING
When opening capillary or tube fittings, solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
Figure 14
Hydraulic Connections
To metering device
To column compartment
From pump
From needle seat
To waste
Plug
Column compartment
Pump
Waste tube in Leak
channel
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Stage 4: Installing the Sample Tray
1 Load the sample tray with sample vials as required.
2 Slide the sample tray into the autosampler so that the rear of the sample
tray is seated firmly against the rear of the sample-tray area.
3 Press the front of the sample tray down to secure the tray in the
autosampler. If the tray pops out of its position the air channel adapter is
not inserted correctly.
Figure 15
Installing the Sample Tray
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
37
Half-Tray Combinations
Half-trays can be installed in any combination enabling both 1.8 ml-and
6 ml-vials to be used simultaneously.
Numbering of Vial Positions
The standard 100-vial tray has vial positions 1 to 100. However, when using
two half-trays, the numbering convention is slightly different. The vial
positions of the right-hand half tray begin at position 101 as follows:
Left-hand 40-position tray: 1–40
Left-hand 15-position tray: 1–15
Right-hand 40-position tray: 101–140
Right-hand 15-position tray: 101–115
NOTE
In the autosampler only the 100 vial tray is supported for temperature control of the vials.
Nevertheless the half trays of the standard autosampler (G1313A) can be used in the
thermostatted autosampler as well. However when these trays are installed cooling or
heating of the vials in the tray will not work.
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Figure 16
Numbering of Tray Positions.
Position 1
Position 101
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
39
Stage 5: Installing Tray Cover and Front Cover
1 Fix the tray cover in the clips of the left autosampler cover side by sliding it
in position. Do not close the tray cover.
2 Position the front cover in the top left corner of the autosampler and turn it
towards the instrument. Press the stop fastener to secure it in the right side
cover of the autosampler.
3 Close the tray cover.
Figure 17
Installation of Tray Cover and Front Cover of the Autosampler
Autosampler front door
Tray cover
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
Stage 6: Turning on the Thermostatted Autosampler
1 Depress the power switch to turn on the two modules of the autosampler.
Stage 7: Update of Control Module Firmware
If the control module has a firmware revision A.01.30 or higher, you do not
need to update the control module firmware.
If you have a control module version A.01.30 or lower, update the firmware as
described.
1 Disconnect the control module before inserting the PC card.
2 Insert the PC card into the card slot of the control module.
3 Reconnect the control module for restarting it.
4 Press “System” (F5) - “Records” (F4). Highlight the LC-System line in the
display using the up-down arrows.
5 Press “FW-Update” (F5).
6 Select the file for the firmware update (LCB202en.BIN).
7 Press “Execute” and select “Yes” to confirm loading of the new firmware.
The control module reboots and loads the firmware indicated by (.) and (*)
on the display. When finished with the update the control module reboots
again.
8 Check that the correct firmware was loaded by pressing “System” (F5) -
“Records” (F4).
9 Disconnect the control module and remove the PC card by pressing the
card-eject button.
NOTE
The power switch stays depressed (1) and a green indicator lamp in the power switch is on
when the autosampler is turned on. When the line power switch stands out (Ø) and the
green light is off, the autosampler is turned off.
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
41
Stage 8: Update of the Agilent ChemStation Software
If you have a Agilent ChemStation software version A.05.02 or higher, you do
not need to update your Agilent ChemStation software.
If you have a Agilent ChemStation software version A.05.01 or lower, please
update your software as described.
Starting from revision A.04.01 or A.04.02 of the Agilent ChemStation
If you have ordered the G1395A Upgrade kit you will be provided with a
A.04.02 and a A.05.01 update CD ROM, which gives you the choice of keeping
your old major software version or updating to A.05.01 release.
1 If you decide to update your A.04.01 Agilent ChemStation to A.05.01, use
the provided A.05.01 CD-ROM and follow the steps described in the
Installing your ChemStation manual, provided as portable document
format (PDF) file on the A.05.01 CD-ROM in the directory MANUALS\
INSTALL\LC. If you do not already have the Adobe Acrobat reader
installed, use the file MANUALS\READER\AR32e30.EXE to install the
reader.
2 After you have updated your system, insert the 3.5” floppy labeled ‘Driver
update Disk (A.05.02 Beta)’ and open a DOS prompt by selecting
Start->RUN and typing command. At the DOS prompt type A: and press
Enter and then HPUPDATE and press Enter only if your
Agilent ChemStation is installed in the directory C:\HPCHEM. If your
Agilent ChemStation is installed in a different directory, e.g. D:\HPCHEM,
you need to type HPUPDATE D:\HPCHEM and press Enter.
CAUTION
If you decide to update to A.05.01, ensure that your PC hard- and software meets the
requirements for A.05.01. A.05.01 requires Windows 95 or Windows NT 4.0 as
operating system and a Pentium PC with a minimum of 24 MB (NT -systems require
also GPIB board Agilent 82341C). Publication 12-5965-6805E gives detailed information
about the PC requirements. The application note can be obtained from the Internet
(http://www.chem.agilent.com/cag/literature/apglit.html) or your local Agilent
Technologies sales office.
1100 Series AT Reference Manual
2
Installing the G1330B Thermostat
3 If you decide to update to A.04.02, insert the A.04.02 CD-ROM and select
SETUP on the CD-ROM using File Manager or Explorer. Select Yes to
continue with the update.
4 After you have updated your system, insert the 3.5 inch floppy labeled
Driver update Disk (A.04.03) and open a DOS prompt by selecting
Start->RUN and typing command. At the DOS prompt type A: and press
Enter and then HPUPDATE and press Enter only if your Agilent
ChemStation is installed in the directory C:\HPCHEM. If your Agilent
ChemStation is installed in a different directory, e.g. D:\HPCHEM, you need
to type HPUPDATE D:\HPCHEM and press Enter.
Starting from revision A.05.01 of the Agilent ChemStation
If you already have the Agilent ChemStation A.05.01 installed, you only need
to install the update to A.05.02 Beta. This update comprises only the driver for
the thermostatted autosampler, no other changes are made.
1 Insert the 3.5” floppy labeled Driver update Disk (A.05.02 Beta) and open a
DOS prompt by selecting Start->RUN and typing command. At the DOS
prompt type A: and press Enter and then HPUPDATE and press Enter only
if your Agilent ChemStation is installed in the directory C:\HPCHEM. If
your Agilent ChemStation is installed in a different directory, e.g. D:\
HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter.
Installing the G1330B Thermostat
2
1100 Series AT Reference Manual
43
Transporting the Thermostatted Autosampler
When moving the autosampler around the laboratory, make sure that any
condensed water inside the thermostat is removed. Tilt the module to the
front, so that the water inside the thermostat can safely flow into the leak
funnel. Otherwise no special precautions are needed for the modules.
If the autosampler needs to be shipped to another location via carrier, ensure:
• The two modules are shipped in separate boxes.
• The transport assembly of the autosampler is parked, see “Park Arm (Park
Gripper)” in your respective Reference Manual.
• The vial tray is secured.
If the autosampler is to be shipped to another location, the transport assembly
of the autosampler must be moved to the park position to prevent mechanical
damage should the shipping container be subjected to excessive shock. Also,
ensure the vial tray is secured in place with suitable packaging, otherwise the
tray may become loose and damage internal components.
NOTE
The ALS thermostat is heavy (20.7 Kg, 45.6 lbs). Carry the module, by putting your hands
under the side covers in a central position of the unit.
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
3
Troubleshooting and Test Functions
Overview of the Thermostat’s Indicators and Test Functions
46
Status Indicators
47
ALS Thermostat Error Messages
49
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Overview of the Thermostat’s Indicators and Test Functions
Status Indicators
The thermostatted autosampler is provided with two status indicators which
indicate the operational state (prerun, run, and error states) of the
instrument. Both are located on the autosampler module. The status
indicators provide a quick visual check of the operation of the thermostatted
autosampler (see "Status Indicators" on page 47).
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the
thermostatted autosampler modules generate an error message in the user
interface. For each message, a short description of the failure, a list of
probable causes of the problem, and a list of suggested actions to fix the
problem are provided. This chapter shows only the error messages that are
generated for the ALS thermostat. For detailed description of the autosampler
error messages see “Autosampler Error Messages” in your corresponding
Reference Manual.
Troubleshooting and Test Functions
3
1100 Series AT Reference Manual
47
Status Indicators
Two status indicators are located on the front of the autosampler. The lower
left indicates the power supply status, the upper right indicates the
thermostatted autosampler status.
Figure 18
Location of Status Indicators
Status indicator
Power supply indicator
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Power Supply Indicator
The power supply indicator is integrated into the main power switch. When
the indicator is illuminated (green) the power is ON.
Instrument Status Indicator
The instrument status indicator indicates one of four possible instrument
conditions:
• When the status indicator is OFF (and power switch light is on), the
instrument is in a prerun condition, and is ready to begin an analysis.
• A green status indicator, indicates the instrument is performing an analysis
(run mode).
• A yellow indicator indicates a not-ready condition. The instrument is in a
not-ready state when it is waiting for a specific condition to be reached or
completed (for example, front cover not installed), or while a self-test
procedure is running.
• An error condition is indicated when the status indicator is red. An error
condition indicates the instrument has detected an internal problem which
affects correct operation of the instrument. Usually, an error condition
requires attention (for example, leak, defective internal components). An
error condition always interrupts the analysis.
Troubleshooting and Test Functions
3
1100 Series AT Reference Manual
49
ALS Thermostat Error Messages
Error messages are displayed in the user interface when an electronic failure
occurs with the ALS thermostat module which requires attention. In the event
of such a failure, the red status indicator at the front of the thermostatted
autosampler is switched on, and an entry is written into the instrument log
book.
This section describes the meaning of ALS thermostat module error messages,
and provides information on probable causes and suggested actions how to
recover from error conditions.
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Fan Failed
Each fan in the ALS thermostat is equipped with a speed sensor, that allows
monitoring and control the speed of the fans. If the sensor shows no signal,
when the fan is activated, it is very likely that the fan is defective. The number
in brackets indicates the fan position. Position numbers of the fans are as
follows (seen from front of the ALS thermostat)
fan failed 1 - left side, fan in front position
fan failed 2 - left side, fan in back position
fan failed 3 - right side, fan in back position
fan failed 4 - right side, fan in front position
fan failed 5 - small cooler fan on top of cooling / heating module
Probable Causes:
• One of the 4 heat-sink fans is defective.
• The cooling heating module fan is defective
Suggested Actions:
✔See Exchange “"Exchanging the Heatsink Fans" on page 67 or “"Exchanging
Heat Exchanger Fan" on page 65.
Troubleshooting and Test Functions
3
1100 Series AT Reference Manual
51
Temperature Control Failed (1 - 4)
There are four Peltier elements build into the ALS thermostat for efficient
cooling / heating. The electronics monitor the current through the Peltier
elements. If the Peltier current is out of a specified limit the Peltier element is
defective.
Probable Causes:
• One of the four peltier elements is defective.
Suggested Action:
✔contact the Technologies service organization
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Temperature Sensor Failed (1 - 4)
The ALS thermostat is equipped with four sensors to monitor the performance
of the instrument. The sensors are connected to both sides of the Peltier
elements for control of the cooling / heating efficiency. As the positioning of
the sensor is critical to meet the performance requirements of the ALS
thermostat the single sensors cannot be replaced separately.
Probable Causes:
• Temperature sensor is defective
Suggested Action:
✔See “"Exchanging the Sensors" on page 69”.
Troubleshooting and Test Functions
3
1100 Series AT Reference Manual
53
Lost Contact to ALS Thermostat
The autosampler control electronics continuously checks whether the ALS
thermostat is active or not.
Probable Causes:
• Bad cable connection between autosampler and ALS thermostat
• Cable between autosampler and thermostat module not connected
• Defective electronic board in autosampler or ALS thermostat
Suggested Actions:
✔Check cable connection between autosampler and ALS thermostat
✔Connect cable between autosampler and ALS thermostat
✔Exchange ASM board in the autosampler
✔Exchange TCA board in the thermostat
✔contact the Agilent Technologies service organization
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Power Fail for ALS Thermostat Module
The +36V that is generated in the ALS thermostat power supply is checked by
the autosampler electronics. If this voltage is missing the error message will be
generated.
Probable Causes:
• ALS thermostat module not connected to line power
• ALS thermostat module fuse(s) defective
• ALS Thermostat power supply defective
Suggested Actions:
✔Check correct power line connection
✔Check fuse(s) and replace if defective
✔contact the Agilent Technologies service organization
Troubleshooting and Test Functions
3
1100 Series AT Reference Manual
55
Temperature out of Range
To protect the Peltier elements for damage their working range is limited to
-3°C to 65°C. The error message is generated when at least one of the three
Peltier elements exceeds this limit.
Probable Causes:
• The four main heat exchanger fans are not able to blow enough air through
the ALS thermostat
• Ambient temperature too high / low
Suggested Actions:
✔Clean the air filters and heat exchanger fins in the ALS thermostat. Switch
off the thermostat and wait 10 min for the peltier elements to equilibrate
✔Make sure that there is enough space on left and right side of the ALS
thermostat and that the air channel inlets and oulets are not blocked
✔Make sure that the ambient air temperature is within its specified limits
(4°C to 55°C)
1100 Series AT Reference Manual
3
Troubleshooting and Test Functions
Bad Cooling / Heating Performance
When turned on the ALS thermostat Peltier elements are activated according
to the given setpoint or to given setpoint change. The electronics check
whether there actual temperature is moving in the correct direction (e.g.,
decreasing when cooled down).
Probable Causes:
• Peltier element(s) defective
• Peltier element(s) not connected or incorrectly connected (e.g., after repair
or maintenance)
• The four main heat exchanger fans are not able to blow enough air through
the ALS thermostat
Suggested Actions:
✔Check connections of the Peltier elements
✔Clean the air filters and heat exchanger fins in the ALS thermostat. Switch
off the thermostat and wait 10 min for the peltier elements to equilibrate
✔Make sure that there is enough space on left and right side of the ALS
thermostat and that the air channel inlets and oulets are not blocked
✔contact the Agilent Technologies service organization
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
4
Repairing the ALS Thermostat Module
Introduction into Repairing the Thermostat
58
Overview of the Repairing the Thermostat
60
Exchanging Internal Parts
62
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
Introduction into Repairing the Thermostat
Simple Repairs
The ALS thermostat is designed for easy repair.
Exchanging Internal Parts
Some repairs may require exchange of defective internal parts. Exchange of
these parts requires removing the ALS thermostat from the stack, removing
the covers, and disassembling the ALS thermostat. The security lever at the
power input socket prevents that the Thermostat cover is taken off when line
power is still connected.
Cleaning the Thermostat
The thermostat covers should be kept clean. Cleaning should be done with a
soft cloth slightly dampened with water or a solution of water and a mild
detergent. Do not use an excessively damp cloth that liquid can drip into the
autosampler.
WARNING
To prevent personal injury, the power cable must be removed from the ALS
thermostat before opening the Thermostat cover. Do not connect the power cable to
the ALS thermostat while the covers are removed.
CAUTION
Electronic boards and components are sensitive to electrostatic discharge (ESD). In
order to prevent damage always use an ESD protection (for example, the ESD wrist
strap from the accessory kit) when handling electronic boards and components.
WARNING
Do not disconnect or reconnect the autosampler to ALS thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
59
Using the ESD Strap
Electronic boards are sensitive to electronic discharge (ESD). In order to
prevent damage, always use an ESD strap supplied in the standard accessory
kit (see "Accessory Kit G1330-68705" on page 79) when handling electronic
boards and components.
1 Unwrap the first two folds of the band and wrap the exposed adhesive side
firmly around your wrist.
2 Unroll the rest of the band and peel the liner from the copper foil at the
opposite end.
3 Attach the copper foil to a convenient and exposed electrical ground.
Figure 19
Using the ESD Strap
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
Overview of the Repairing the Thermostat
Figure 20 shows the main assemblies and their locations which can be
repaired.
Figure 20
Main assemblies
Heat exchanger
fan
TCA
Board
Condensation leak drain
Heat
exchanger
Power Supply
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
61
Exchanging the Power Supply Fuses
The fuse holders are located on the rear panel of the ALS thermostat.
1 Switch off the power switch at the front of the thermostatted autosampler.
2 Remove the power cable from the two modules.
3 Insert the flat head screwdriver in the fuse holder, slightly press and turn
counter clockwise to release the fuse holder from the socket.
4 Pull the fuse holder out of the socket.
5 Remove the fuse from the fuse holder.
6 Insert a new fuse in the fuse holder.
7 Reinsert the fuse holder and fix with the screwdriver.
8 Reinsert the power cables.
9 Switch on the power switch.
Frequency
When defective
Tools required
Flat head screwdriver
Parts required
Fuses T2.5 A/250V (CSA, UL listed), 2110-0015
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
Exchanging Internal Parts
The procedures in this section describe how to exchange internal parts. You
must remove the ALS thermostat from the stack in order to open the main
cover.
WARNING
The following procedures require opening the main cover of the ALS thermostat.
Always ensure the ALS thermostat is disconnected from the line power when the
main cover is removed. The security lever at the power input socket prevents that
the pump cover is taken off when line power is still connected.
WARNING
To disconnect the ALS thermostat from line, unplug the power cord. The power
supply still uses some power, even if the switch on the front panel is turned off.
WARNING
When opening capillary or tube fittings solvents may leak out. Please observe
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
NOTE
The electronics of the ALS thermostat will not allow operation when the top cover and the
top foam are removed. A safety light switch on the main board will inhibit the operation of
the ALS thermostat. Always operate the ALS thermostat with the top foam and top covers
in place.
CAUTION
Internal components may be sensitive to electrostatic discharge (ESD). Always use an
ESD kit when handling internal parts.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
63
Removing the Top Cover and Foam
Tools required
Screwdriver Pozidriv #1
Preparations for this
procedure
Switch off ALS at the main power switch. Disconnect ALS and ALS thermostat
power cords. Remove ALS to ALS thermostat cable, and remove thermostat
from stack.
WARNING
Do not disconnect or reconnect the autosampler to ALS thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.
1 Remove the front cover by pressing the both clip fasteners
on both sides of the cover.
2 Move the lever towards the power socket.
Lever
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
3 Lift the clips on both sides of the top cover (1). Remove the
top cover (2).
4 Unscrew the screws on the top plate and remove the plate
by lifting its back first and then sliding to the front.
5 Unplug all wires at the TCA Board and remove them from
top foam.
6 Remove the top foam.
Screws
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
65
Exchanging Heat Exchanger Fan
When required
When defective
Preparations for this
procedure
Removing the Top Cover and Foam (see "Removing the Top Cover and Foam"
on page 63) item 1 to 4.
Parts required
Fan, PN 3160-1079
1 Unplug the connector at the TCA board and remove the
wire from the Top Foam.
2 Carefully lift the Fan upwards.
3 Remove the Fan from the Fan Gasket.
4 Insert the new Fan in the Fan Gasket. See arrow at the side
of Fan for correct airflow direction.
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
5 Reinstall the fan and route the wire in the appropriate
guide channel.
6 Reconnect the wire to the connector.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
67
Exchanging the Heatsink Fans
When required
Fan not running
Preparations for this
procedure
Removing the Top Cover and Foam (see "Removing the Top Cover and Foam"
on page 63).
Parts required
Fan, part number 3160-0884
1 Carefully lift the pair of Fans upwards.
2 Replace the defective fan, and reinsert the pair of Fans.
Note the orientation of the fans.
3 Reinsert the Top Foam and route the cables in the guide
channels in the Top Foam.
4 Reconnect the connectors in the correct order (See label
on top of Power Supply).
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
5 Replace the metal cover and top cover.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
69
Exchanging the Sensors
When required
Sensors defective
Preparations for this
procedure
Removing the Top Cover and Foam (see "Removing the Top Cover and Foam"
on page 63).
Parts required
Sensor flexboard, PN G1330-66504
Heatsink Compound, PN 6040-0454
1 Unscrew the six holding screws on top of the hardfoam
encapsulation.
2 Carefully lift the upper part of hardfoam encapsulation
away.
Screws
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
3 Remove the sealing adapter by carefully sliding it out of
the seat.
4 Unplug the connectors at the TCA Board, careful lift the
sensors out of their seat and remove the sensor flexboard.
5 Insert some Heatsink Compound into the four sensor
seats, to assure good contact.
6 Install the new sensor flexboard carefully by pressing
each sensor individually into the seats, and then route the
flexboard through the sealing adapter channel.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
71
7 Reinsert the sealing adapter by carefully sliding it in to the
seat.
8 Reinstall the hardfoam encapsulation.
9 Route the cables in the guide channels and reconnect the
connectors in the correct order (See label on Power
Supply).
10 Reinstall the Top Plate and the Top Cover.
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
Exchanging the TCA board
When required
TCA Board defective
Preparations for this
procedure
Removing the Top Cover and Foam (see "Removing the Top Cover and Foam"
on page 63), item 1 to 4.
Parts required
TCA board, part number G1330-66500
1 Remove all connectors on the TCA Board.
2 Unscrew the connector screws for the Autosampler to
Thermostat cable.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
73
3 By carefully pulling back on the upper part of the TCA
board unsnatch it from the holder.
4 Remove the TCA board.
5 Reinsert the new TCA board.
6 Gently push forward on the upper part of the board, until it
snatches into place.
1100 Series AT Reference Manual
4
Repairing the ALS Thermostat Module
7 Refit the connector screws and reconnect all connectors.
8 Replace the top foam section, metal cover and top cover.
Repairing the ALS Thermostat Module
4
1100 Series AT Reference Manual
75
Assembling the Main Cover
When required
If cover is broken
Tools required
None
Parts required
Cover kit G1330-68703 (includes base, top, left and right)
NOTE
The cover kit contains all parts, but it is not assembled.
WARNING
In case you insert the left or right side in the opposite position, you may not be able
to remove the side from the top part.
1 Place the top part on the bench and insert the left and right
side into the top part.
2 Replace the cover.
3 Replace the ALS thermostat into the stack and reconnect the cables and capillaries.
4 Turn on the ALS thermostat.
Front
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
5
ALS Thermostat Parts and Materials
Main Assemblies
78
Accessory Kit G1330-68705
79
Main Assemblies
78
Foam Parts
81
Plastic Parts
82
Heat Exchanger Fan Parts
83
1100 Series AT Reference Manual
5
ALS Thermostat Parts and Materials
Main Assemblies
Figure 21
Main Assemblies
Table 4
Main Assemblies
Item
Description
Part Number
1
Fuse - Power Supply (T2.5A/250V; CSA, UL listed)
2110-0015
2
Fuse TCA - Board (T3A/250V; CSA, UL listed)
2110-0029
3
Front Cover
5062-8582
Cable, autosampler - ALS thermostat
G1330-81600
2
1
3
ALS Thermostat Parts and Materials
5
1100 Series AT Reference Manual
79
Accessory Kit G1330-68705
Table 5
Accessory Kit
Item
Description
Part Number
1
Waste Tube*
*
Reorder Number (5 m)
5062-2463
2
Waste Tube Assembly
G1330-67300
1100 Series AT Reference Manual
5
ALS Thermostat Parts and Materials
Main Assemblies
Figure 22
Main Assemblies
Table 6
Main Assemblies
Item
Description
Part Number
1
Heat sink fans
3160-0884
2
TCA Board
G1330-66500
3
Heat exchanger fan
3160-1079
2
1
3
1
ALS Thermostat Parts and Materials
5
1100 Series AT Reference Manual
81
Foam Parts
Figure 23
Foam Parts
Table 7
Main Assemblies
Item
Description
Part Number
1
Top Foam
G1330-40102
2
Bottom Foam
G1330-40103
1
2
1100 Series AT Reference Manual
5
ALS Thermostat Parts and Materials
Plastic Parts
Figure 24
Plastic Parts
Table 8
Main Assemblies
Item
Description
Part Number
1
Cabinet Kit, includes base, top and sides
G1330-68703
2
Front Cover
5062-8582
3
Leak Pan
5042-1307
2
1
3
ALS Thermostat Parts and Materials
5
1100 Series AT Reference Manual
83
Heat Exchanger Fan Parts
Figure 25
Heat Exchanger Fan Parts
Table 9
Main Assemblies
Item
Description
Part Number
1
Fan Gasket
G1330-44104
2
Heat Exchanger Fan
3160-1079
1
2
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
6
Specifications
Performance Specifications
86
1100 Series AT Reference Manual
6
Specifications
Performance Specifications
Table 10
Performance Specifications Agilent 1100 Series thermostatted autosampler
Type
Specification
Temperature range:
setable from 4°C to 40°C in 1° increments
Agilent 1100 Series Autosampler Thermostat
Reference Manual
Agilent Technologies
A
Safety Information
General Safety Information
88
Radio Interference
91
Sound Emission
92
Agilent Technologies on Internet
93
1100 Series AT Reference Manual
A
Safety Information
General Safety Information
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Agilent
Technologies assumes no liability for the customer’s failure to comply with
these requirements.
General
This is a Safety Class I instrument (provided with terminal for protective
earthing) and has been manufactured and tested according to international
safety standards.
Operation
Before applying power, comply with the installation section. Additionally the
following must be observed.
Do not remove instrument covers when operating. Before the instrument is
switched on, all protective earth terminals, extension cords,
auto-transformers, and devices connected to it must be connected to a
protective earth via a ground socket. Any interruption of the protective earth
grounding will cause a potential shock hazard that could result in serious
personal injury. Whenever it is likely that the protection has been impaired,
the instrument must be made inoperative and be secured against any intended
operation.
Make sure that only fuses with the required rated current and of the specified
type (normal blow, time delay, and so on) are used for replacement. The use of
repaired fuses and the short-circuiting of fuseholders must be avoided.
Some adjustments described in the manual, are made with power supplied to
the instrument, and protective covers removed. Energy available at many
points may, if contacted, result in personal injury.
Any adjustment, maintenance, and repair of the opened instrument under
voltage should be avoided as much as possible. When inevitable, this should be
carried out by a skilled person who is aware of the hazard involved. Do not
1100 Series AT Reference Manual
89
attempt internal service or adjustment unless another person, capable of
rendering first aid and resuscitation, is present. Do not replace components
with power cable connected.
Do not operate the instrument in the presence of flammable gases or fumes.
Operation of any electrical instrument in such an environment constitutes a
definite safety hazard.
Do not install substitute parts or make any unauthorized modification to the
instrument.
Capacitors inside the instrument may still be charged, even though the
instrument has been disconnected from its source of supply. Dangerous
voltages, capable of causing serious personal injury, are present in this
instrument. Use extreme caution when handling, testing and adjusting.
When working with solvents please observe appropriate safety procedures
(e.g. goggles, safety gloves and protective clothing) as described in the material
handling and safety data sheet by the solvent vendor, especially when toxic or
hazardous solvents are used.
Safety Symbols
Table 1 shows safety symbols used on the instrument and in the manuals.
Table 1
Safety Symbols
Symbol
Description
The apparatus is marked with this symbol when the user should refer to the instruction manual in order to
protect the apparatus against damage.
Indicates dangerous voltages.
Indicates a protected ground terminal.
!
1100 Series AT Reference Manual
A
Safety Information
WARNING
A warning alerts you to situations that could cause physical injury or damage to the
equipment. Do not proceed beyond a warning until you have fully understood and
met the indicated conditions.
CAUTION
A caution alerts you to situations that could cause a possible loss of data. Do not
proceed beyond a caution until you have fully understood and met the indicated
conditions.
1100 Series AT Reference Manual
91
Radio Interference
Manufacturer’s Declaration
This is to certify that this equipment is in accordance with the Radio
Interference Requirements of Directive FTZ 1046/1984. The German
Bundespost was notified that this equipment was put into circulation, the
right to check the series for compliance with the requirements was granted.
Test and Measurement
If test and measurement equipment is operated with equipment unscreened
cables and/or used for measurements on open set-ups, the user has to assure
that under operating conditions the radio interference limits are still met
within the premises.
1100 Series AT Reference Manual
A
Safety Information
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German
Sound Emission Directive of 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB.
• Sound Pressure Lp < 70 dB (A)
• At Operator Position
• Normal Operation
• According to ISO 7779:1988/EN 27779/1991 (Type Test)
1100 Series AT Reference Manual
93
Agilent Technologies on Internet
For the latest information on products and services visit our worldwide web
site on the Internet at:
http://www.agilent.com/go/chem
1100 Series AT Reference Manual
95
Index
A
Agilent on internet, 93
air circulation, 21
ALS thermostat accessory kit, 24
ALS thermostat accessory kit contents, 24
ALS thermostat operation, 12
ALS thermostat parts and materials
accessory kit, 79
main assemblies, 78, 79
ALS thermostat repairs
power supply fuses, 61
assembling the main cover, 75
autosampler control, 14
B
bench space, 21
C
cleaning the ALS thermostat, 58
condensation, 10, 21
D
damaged packaging, 23
damper, 65
damping unit, 65
delay volume, 25
delivery checklist, 23
E
electrical connections, 17
CAN, 17
GPIB, 17
Remote, 17
RS-232, 17
thermostat - autosampler, 17
electronics, 14
electrostatic discharge (ESD), 58
environment, 20, 21
error messages, 46
ALS thermostat, 49
bad cooling / heating performance, 56
lost contact to ALS thermostat, 53
power fail ALS thermostat, 54
temperature control failed, 51
temperature out of range, 55
temperature sensor failed, 52
therm fan failed, 50
ESD (electrostatic discharge) strap, 59
exchanging
damper, 65
fan, 67, 69, 72
internal parts, 58, 62
F
failure, 46
fan, 67, 69
exchanging, 72
flow path, 25
fuses, 20, 61
H
half-tray combinations, 37
I
injection valve, 10
installing the thermostatted
autosampler, 28
flow connections, 35
peparation, 29
power cable and interface cable, 32
sample tray, 36
tray cover and front cover, 39
turning on the instrument, 40
update of firmware or software, 40, 41
instrument status indicator, 48
internal parts, 62
internet, 93
introduction to the autosampler, 10
M
missing parts, 23
multi-draw option, 10
O
optimizing the stack configuration, 25
P
park transport assembly, 23, 43
parts and materials, 23
performance specifications, 86
physical specifications, 22
power consideration, 20
power cords, 20
power requirements, 20
power supply indicator, 48
1100 Series AT Reference Manual
Index
R
repairs, 58
using the ESD strap, 59
S
sample tray, 10
sample trays
numbering of vial positions, 37
screwdriver, pozidriv #1, 67, 69, 72
security lever, 58
shipping, 23, 43
simple repairs, 58
site requirements, 20
specifications, 22, 86
status indicator, 46
status indicators, 47
storage, 21
T
temperature, 21
top cover, 75
transport, 43
transport mechanism, 10
U
unpacking, 23
unpacking the autosampler, 23
V
vial racks, 10
vial tray, 43
vials, 10
W
weight, 21
In This Book
This manual contains information
about the Agilent 1100 Series
thermostatted for autosamplers.
Information for the individual
autosamplers can be obtained from
the corresponding autosampler
reference manual. This supplemental
manual describes the following:
• installing the thermostat,
• theory of operation,
• troubleshooting the thermostat,
• repair procedures for the
thermostat,
• parts and materials for the
thermostat.