Agilent Technologies 1100 Series Reference Manual

Specifications

Brand: Agilent Technologies
Category: Thermostat
Model: 1100 Series (Reference Manual)
File Type: PDF
File Size: —
Pages: 98

Read Online

Read the Agilent Technologies 1100 Series Reference Manual online using the built-in PDF viewer.

Text of the manual

1100 Series AT Reference Manual

Notices © Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including elec- tronic storage and retrieval or translation into a foreign language) without prior agree- ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G1330-90002 Edition Edition 07/02 Printed in Germany Agilent Technologies Hewlett-Packard-Strasse 8 76337 Waldbronn, Germany Software Revision This guide is valid for A.01.xx revisions of the Agilent 1100 Series Autosampler Ther- mostat software, where xx refers to minor revisions of the software that do not affect the technical accuracy of this guide. Warranty The material contained in this docu- ment is provided “as is,” and is sub- ject to being changed, without notice, in future editions. Further, to the max- imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par- ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec- tion with the furnishing, use, or per- formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep- arate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor- dance with the terms of such license. Restricted Rights Legend Software and technical data rights granted to federal government customers include only those rights customarily provided to end user Customers of Software. Agilent provides this customary commercial license in Software and technical data pursuant to FAR 12.211 (Technical Data) and FAR 12.212 (Computer Software) and, for Department of Defense purchases, DFARS 252.227-7015 (Technical Data - Commercial Items) and DFARS 227.7202-3 (Rights in Commercial Computer Software or Computer Software Documentation). If a federal government or other public sector Customer has a need for rights not conveyed under these terms, it must negotiate with Agilent to establish acceptable terms in a written agreement executed by all relevant parties. Safety Notices CAUTION A CAUTION notice denotes a haz- ard. It calls attention to an operat- ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per- formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi- tions are fully understood and met.

1100 Series AT Reference Manual

3 In This Guide… 1 Theory of Operation Theory of operation of mechanical hardware, electronics, and instrument interfaces 2 Installing the G1330B Thermostat Site requirements and installation of the G1330B thermostat 3 Troubleshooting and Test Functions The thermostatted autosampler‘s built-in troubleshooting and test functions 4 Repairing the ALS Thermostat Module Instructions on repair procedures of the ALS thermostat 5 ALS Thermostat Parts and Materials Lists for identification of parts and materials 6 Specifications Performance specifications of the ALS thermostat

1100 Series AT Reference Manual

5 Contents 1 Theory of Operation Introduction to the Thermostatted Autosampler 10 ALS Thermostat Operation 12 ALS Thermostat Control and Electronics 14 Temperature Controlled Autosampler Board (TCA) 15 Electrical Connections 17 2 Installing the G1330B Thermostat Site Requirements 20 Unpacking the Autosampler 23 Optimizing the Stack Configuration 25 Controller Requirements 27 Installing the G1330B Thermostat 28 Stage 1: Preparing the ALS Thermostat and Autosampler 29 Stage 2: Power Cable and Interface Cable Connection 32 Stage 3: Flow Connections 35 Stage 4: Installing the Sample Tray 36 Stage 5: Installing Tray Cover and Front Cover 39 Stage 6: Turning on the Thermostatted Autosampler 40 Stage 7: Update of Control Module Firmware 40 Stage 8: Update of the Agilent ChemStation Software 41 Transporting the Thermostatted Autosampler 43

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Overview of the Thermostat’s Indicators and Test Functions 46 Status Indicators 47 Power Supply Indicator 48 Instrument Status Indicator 48 ALS Thermostat Error Messages 49 Fan Failed 50 Temperature Control Failed (1 - 4) 51 Temperature Sensor Failed (1 - 4) 52 Lost Contact to ALS Thermostat 53 Power Fail for ALS Thermostat Module 54 Temperature out of Range 55 Bad Cooling / Heating Performance 56 4 Repairing the ALS Thermostat Module Introduction into Repairing the Thermostat 58 Using the ESD Strap 59 Overview of the Repairing the Thermostat 60 Exchanging the Power Supply Fuses 61 Exchanging Internal Parts 62 Removing the Top Cover and Foam 63 Exchanging Heat Exchanger Fan 65 Exchanging the Heatsink Fans 67 Exchanging the Sensors 69 Exchanging the TCA board 72 Assembling the Main Cover 75

1100 Series AT Reference Manual

7 5 ALS Thermostat Parts and Materials Main Assemblies 78 Accessory Kit G1330-68705 79 Main Assemblies 80 Foam Parts 81 Plastic Parts 82 Heat Exchanger Fan Parts 83 6 Specifications Performance Specifications 86 A Safety Information General Safety Information 88 Radio Interference 91 Sound Emission 92 Agilent Technologies on Internet 93 Index

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 1 Theory of Operation Introduction to the Thermostatted Autosampler 10 ALS Thermostat Operation 12 ALS Thermostat Control and Electronics 14 Temperature Controlled Autosampler Board (TCA) 15 Electrical Connections 17

1100 Series AT Reference Manual

1 Theory of Operation Introduction to the Thermostatted Autosampler The Agilent 1100 Series autosampler is designed for use with other modules of the Agilent 1100 Series LC system, with the HP 1050 Series, or with other LC systems if adequate remote control inputs and outputs are available. The autosampler is controlled from the Agilent 1100 Series control module or from the Agilent ChemStation for LC systems. The specially-designed thermostattable sample trays holds either 100 × 1.8 ml vials or two wellplates and 10 × 1.8 ml vials. The ALS thermostat contains Peltier-controlled heat exchangers. A fan draws air from the area above the sample vial tray of the autosampler and is then blown through the fins of the cooling/heating module. There it is cooled or heated according to the temperature setting. The thermostatted air enters the autosampler through a recess underneath the specially-designed sample tray. The air is then distributed evenly through the sample tray ensuring effective temperature control, regardless of how many vials are in the tray. In cooling mode condensation is generated on the cooled side of the Peltier elements. This condensed water is safely guided into the leak system.

Theory of Operation

1 1100 Series AT Reference Manual 11 Figure 1 Overview of the ALS Thermostat Power Supply TCA Board Heat sink fans Heat exchanger Fan Condensation leak drain Heat exchanger Heat sink

1100 Series AT Reference Manual

1 Theory of Operation ALS Thermostat Operation The thermostatted autosampler is equipped with a cooling/heating module which uses Peltier elements for efficient air cooling. When turned on the front side of the Peltier elements is heated/cooled according to the temperature setting. A fan draws air from the sample tray area and blows it through the channels of the heating/cooling module. The fan speed is determined according to the environmental conditions (e.g., ambient temperature, humidity). In the heating/cooling module the air reaches the temperature of Figure 2 ALS Thermostat Principle air channel in autosampler sampling unit special sample tray air outlet Peltier elements heat exchanger heating/cooling module ambient air in heated air out thermostatted air

Theory of Operation

1 1100 Series AT Reference Manual 13 the Peltier elements and this thermostatted air is blown underneath the special sample tray where it is evenly distributed and streams back into the sample tray area. From there it is again drawn into the ALS thermostat. This “recycle” mode assures a very efficient cooling/heating of the sample vials. In cooling mode the opposite side of the Peltier element will become very hot and to maintain the performance of the elements they have to be cooled down. This is done with large heat exchangers in the back of the ALS thermostat. Four fans blow air from left to right through the instrument to remove the heated air. The fan speed is controlled according to the temperature of the Peltier elements. During cooling condensation will appear in the heating/cooling module. The condensed water will be guided out of the ALS thermostat.

1100 Series AT Reference Manual

1 Theory of Operation ALS Thermostat Control and Electronics The ASM board of the autosampler module provides all control signals for the ALS thermostat (e.g., temperature setting) and receives the sensor signals from the temperature and fan sensors. The TCA board controls the internal assemblies in the ALS thermostat. This chapter describes the electronics of the TCA board. Figure 3 Block Diagram ALS Thermostat Electronics TCA Board ASM Board Peltier Assembly A Peltier Assembly B Peltier Assemblies C + D Temperatur Sensor Fan 1 1 2 2 3 4 5 3 4 Primary Board Bridge Rectifier + 36 V Line Voltage

Theory of Operation

1 1100 Series AT Reference Manual 15 Temperature Controlled Autosampler Board (TCA) The board controls all information and activities of all assemblies within the ALS thermostat. The operator enters parameters, changes modes and controls the module through the control module or Agilent ChemStation, that is connected to the autosampler module. Autosampler and ALS thermostat are connected via a control cable for signal exchange. Step Down Switcher The four Peltier assemblies (A to D) are driven by synchronous step down switching regulators. Switcher 1 and 2 receive their power (+36V) from the ALS thermostat internal power supply. Switcher 3 receives its power (+24V) from the autosampler module via the interface cable. The maximum power consumption for all Peltier assemblies is approximately 180 Watts in cooling mode and 120 Watts in heating mode. Regulators The fans are equipped with hall sensors that allow to determination of the actual speed of the fans. The speed of the four heat sink fans is variable and will be set according to the actual temperature at the heat sink side of the Peltier assemblies. The heat exchanger fan is always running at highest speed. the speed will be reduced to a minimum when the sample tray is removed from the autosampler. Auto Range When the ALS thermostat power cable is connected to the module the auto range circuits checks the connected supply voltage. Default setting is a connection to a 220-240V supply. If the non-regulated 36V is below a certain limit the auto range circuit activates a switch on the primary board in the power supply and switches to 100-120V operation. Multiplexer The ALS thermostat is equipped with four digital temperature sensors for temperature control and diagnostics. Four are used to check the temperature on both sides of the Peltier elements. The temperature sensor signals and the signals from the hall sensors in the five fans are the input signals for the multiplexer. The output of the multiplexer is connected to the autosampler

1100 Series AT Reference Manual

1 Theory of Operation main board via the 26 pin autosampler to ALS thermostat cable. The autosampler firmware checks the signals for correctness and initiates all needed activities. EPROM The EPROM stores all relevant data for the ALS Thermostat (e.g., serial number, board revision, etc.). This data are pre-set at the factory. Figure 4 Block Diagram TCA Board Step Down Switcher 1 Step Down Switcher 1 EPROM Multiplexer Auto Range Regulator 3 Regulator 2 Regulator 1 Step Down Switcher3 PDC 1 + 36 V PDC 2 PDC 3 + 24 V + 36 V Temp. Sensor 1 to 4 Fan Sensor 1 to 5 to ASM Board Primary Board Fan 5 Fan 3 + 4 Fan 1 + 2 Peltier Assembly C + D Peltier Assembly B Peltier Assembly A + 36 V

Theory of Operation

1 1100 Series AT Reference Manual 17 Electrical Connections WARNING Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations. Figure 5 Elctrical Connections GPIB RS232C CAN-bus Remote Vial number output Relay contacts CAN cable to previous module CAN cable to next module Therm-ALS cable Thermostat - Autosampler cable

1100 Series AT Reference Manual

1 Theory of Operation • The GPIB connector is used to connect the autosampler with a computer. The address and control switch module next to the GPIB connector determines the GPIB address of your autosampler. The switches are preset to a default address (see Autosampler Reference Manual) and this is recognized immediately after power on. • The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal Agilent 1100 Series module data transfer and synchronization. • The REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as common shut down, prepare, and so on. • The RS-232 connector may be used to control the autosampler from a computer through an RS-232 connection, using appropriate software. This connector needs to be activated by the configuration switch module next to the GPIB connector. The software needs the appropriate drivers to support this communication. See your software documentation for further information. • The Thermostat-Autosampler connection is used for control signal transfer and synchronization of the two modules. The cable must be installed for operation of the ALS thermostat. • The power input socket accepts a line voltage of 100–120 or 220–240 volts AC ± 10 % with a line frequency of 50 or 60 Hz. Maximum power consumption of the autosampler module is 300 Watts (Volt-Amps). Maximum power consumption of the ALS thermostat module is 260 Watts (Volt-Amps).There are no voltage selectors on your autosampler because the power supplies have automatic selection capability. The autosampler module has no externally accessible fuses, because automatic electronic fuses are implemented in the power supply. The power supply of the ALS thermostat has two externally accessible fuses. The security lever at the power input socket prevents removal of the autosampler cover when line power is still connected. • The interface board slot is used for external contacts, BCD output and for future use. WARNING DO NOT disconnect or reconnect the autosampler to ALS thermostat cable when the power cords are connected to either of the two modules. This will damage the electronics of the modules.

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 2 Installing the G1330B Thermostat Site Requirements 20 Unpacking the Autosampler 23 Optimizing the Stack Configuration 25 Controller Requirements 27 Installing the G1330B Thermostat 28 Transporting the Thermostatted Autosampler 43

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Site Requirements A suitable site environment is important to ensure optimum performance of the autosamplers. Power Consideration The autosamplers comprises two modules, the autosampler module (G1329A, G1389A, G1367A, or G2260A) and the thermostat module (G1330B Therm). Both modules have a separate power supply and a power plug for the line connections. The two modules are connected by a control cable and both are turned on by the autosampler module. The autosampler power supplies have automatic voltage selectors (see Table 1 on page 22). Consequently there are no voltage selectors in the rear of the two autosampler modules. The autosampler module has no externally accessible fuses, because automatic electronic fuses are implemented in its power supply. The ALS thermostat power supply has two externally accessible fuses. Power Cords Your autosampler modules are delivered with power cords which match the wall socket of your particular country or region. The plug on the power cords which connect to the rear of the two instrument is identical for all types of power cord. WARNING To disconnect the autosampler from line power, unplug the power cord from the autosampler and the thermostat. The power supplies still use some power, even if the power switch on the front panel is turned off. Please make sure that it is always possible to access the power plug. WARNING Shock hazard or damage of your instrumentation can result if the devices are connected to a line voltage higher than specified. WARNING Never operate your instrumentation from a power outlet that has no ground connection. Never use a power cord other than the power cord designed for your region.

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 21 Bench Space The autosampler dimensions and weight (see Table 1 on page 22) allow the instrument to be placed on almost any laboratory bench. The instrument requires an additional 25 cm (10 inches) of space on either side for the circulation of air, and approximately 8 cm (3.1 inches) at the rear for electrical connections. Ensure the autosampler is installed in a level position. If a complete Agilent 1100 Series system is to be installed on the bench, make sure that the bench is designed to carry the weight of all the modules. For a complete system including the thermostatted autosampler it is recommended to position the modules in two stacks, see "Optimizing the Stack Configuration" on page 25. Make sure that in this configuration there is 25 cm (10 inches) space on either side of the thermostatted autosampler for the circulation of air. Environment Your autosampler modules will work at ambient temperatures and relative humidity as described in Table 1 on page 22. CAUTION Do not store, ship or use your autosampler under conditions where temperature fluctuations may cause condensation within the autosampler electronics. Condensation will damage the system electronics. If your autosampler was shipped in cold weather, leave it in its box, and allow it to warm up slowly to room temperature to avoid condensation.

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Table 1 Physical Specifications - Thermostatted Autosampler Type Specification Comments Thermostat Weight 20.7 kg (45.6 lbs) Dimensions (height × width × depth) 140 × 345 × 435 mm (5.5 × 13.5 × 17 inches) Line voltage 100 – 120 or 220 – 240 VAC, ± 10 % Automatic selection Line frequency 50 or 60 Hz, ± 5 % Power consumption Autosampler: 300 VA ALS Thermostat: 260 VA Maximum Maximum Ambient operating temperature 4 – 40 ∞C (41 – 131 ∞F) see on page 22 Ambient non-operating temperature -40 – 70 ∞C (-4 – 158 ∞F) Humidity < 95 %, at 25 – 40 ∞C (77 – 104 ∞F) Non-condensing; Operating Altitude Up to 2000 m (6500 ft) Non-operating altitude Up to 4600 m (14950 ft) For storing the autosampler Safety standards: IEC, CSA, UL, EN Installation Category II, Pollution Degree 2 WARNING Using the autosampler at environmental temperatures higher than 50 ∞C (122 ∞F) may cause the rear panel to become hot.

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 23 Unpacking the Autosampler Damaged Packaging The two modules of the autosampler are shipped in separate boxes. Upon receipt of your autosampler, inspect the shipping containers for any signs of damage. If the containers or cushioning material are damaged, save them until the contents have been checked for completeness and the autosampler has been mechanically and electrically checked. If the shipping container or cushioning material is damaged, notify the carrier and save the shipping material for the carriers inspection. Delivery Checklist Unpack the two boxes of the autosampler. Ensure all parts and materials have been delivered with the autosampler and the ALS thermostat. The delivery checklist are shown in Table 2. Please report missing or damaged parts to your local Agilent Technologies sales and service office. If the thermostatted autosampler was ordered as an upgrade (G1395A) to an existing autosampler, the shipment will also contain the required software upgrades for your Agilent ChemStation. CAUTION If you intend to ship or transport the autosampler, always park the transport assembly of the autosampler module before shipment (see “Transporting the Autosampler” in the corresponding manual). CAUTION If there are signs of damage to the autosampler, please do not attempt to install the autosampler. Table 2 G1330B Thermostat Checklist Description Quantity Part Number ALS Thermostat G1330B 1 Power cable 1 as ordered Accessory kit (Table 3) 1 G1330-68705

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Table 3 ALS Thermostat Accessory Kit Contents G1330-68705 Description Part Number Waste Tube 5063-6527 Waste Tube Assembly G1330-67300

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 25 Optimizing the Stack Configuration If your autosampler is part of a system, you can ensure optimum performance by installing the autosampler in the stack in the position shown in Figure 6 and Figure 7. This configuration optimizes the system flow path, ensuring minimum delay volume. Figure 6 Recommended Stack Configuration (Front View) Degasser Pump Solvent Cabinet Autosampler Thermostat Column Compartment Detector Control Module

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Figure 7 Recommended Stack Configuration (Rear View) GPIB to LC ChemStation Analog signal to recorder AC power Remote cable CAN bus cable Autosampler - Thermostat cable AC power Analog signal to recorder

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 27 Controller Requirements Agilent 1100 Series control module and Agilent ChemStation must have the latest revision firmware/software loaded to allow optimum operation with the thermostattable autosamplers. Older revisions might not recognize or have full functionality with the thermostattable autosampler. Control Module Firmware requirements The control module requires firmware revision A.01.30 or higher to control the thermostatted autosampler. Previous firmware revisions of the control module will not run with the thermostatted autosampler. If the control module was shipped together with the thermostatted autosampler the control module firmware does not require updating. The firmware update must be done with a PCMCIA card that has the newest revision loaded. The firmware is not part of the shipment of the thermostatted autosampler. Contact your local Agilent Technologies sales and service office for the firmware update of the control module. For the update procedure see "Stage 7: Update of Control Module Firmware" on page 40. Agilent ChemStation Software requirements To control the thermostatted autosampler from a PC, the Agilent ChemStation software version A.04.03, or A.05.02beta, or A.05.02 or higher is required. These software revision might however not support all ALS modules. The thermostatted autosampler will not run with any previous version of the Agilent ChemStation software. The software updates are part of the thermostatted autosampler shipment. For the update procedure see "Stage 8: Update of the Agilent ChemStation Software" on page 41.

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Installing the G1330B Thermostat WARNING To avoid personal injury, keep fingers away from the needle area during autosampler operation. Do not bend the safety flap away from its position, or attempt to remove the safety cover (see Figure 8). Do not attempt to insert or remove a vial from the gripper when the gripper is positioned below the needle. Figure 8 Safety Flap. Safety flap Safety cover

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 29 Stage 1: Preparing the ALS Thermostat and Autosampler 1 Place the ALS thermostat on the bench or in the stack. 2 Remove the front cover. Press the two snap fasteners on the sides of the cover and move it away. 3 If the ALS thermostat is located on top of another Agilent 1100 Series Module place the waste tube assembly into the top cover of the ALS thermostat and locate the other end in the waste funnel of the module beneath. Figure 9 Preparation of the ALS Thermostat Condensation leak tube Waste tube assembly Module locks

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat 4 Connect the condensation leak tube to the main waste exit of the ALS thermostat and place into an appropriate vessel. It is possible to either let the condensation leak tubing exit the module at the front or at the left side of the module. Make sure that the leak tube is fully fixed on the outlet. 5 Install the front cover of the ALS thermostat. 6 Place the autosampler module on top of the ALS thermostat. Make sure that the autosampler is correctly engaged in the ALS thermostat locks. WARNING Make sure that the condensation tube is always above the liquid level in the vessel. If the tube is located in liquid the condensed water cannot flow out of the tube and the outlet is blocked. Any further condensation will then remain in the instrument. This may damage the instruments electronics. Figure 10 Condensation Leak outlet Condensation leak tube Waste bottle

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 31 7 Place the air channel adapter into the autosampler tray base. Make sure the adapter is fully pressed down. This assures that the cold airstream from the ALS thermostat is correctly guided to the tray area of the autosampler. 8 If there is no Agilent 1100 Series module located beneath the ALS thermostat connect the waste tube to the central waste exit of the autosampler and place in a waste vessel. Figure 11 Preparation of ALS Thermostat and Autosampler Air channel adapter

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Stage 2: Power Cable and Interface Cable Connection 1 Ensure the power switch on the front of the autosampler is OFF and the power cables are disconnected. 2 Connect the cable between the autosampler and the ALS thermostat, see Figure 5 on page 30. 3 Move the safety lever at the rear of the two modules to the right position, see Figure 5 on page 30. 4 Connect the power cables to the power connectors. 5 Connect the CAN interface cables to other modules in the system (see Figure 7 on page 26 and Figure 13 on page 34). 6 If required, connect additional interface and control cables to the autosampler (see Figure 7 on page 26 and Figure 13 on page 34). Refer to the documentation of the Agilent 1100 Series control module or ChemStation for LC for more information. WARNING Do not disconnect or reconnect the autosampler to ALS thermostat cable when the power cords are connected to either of the two modules. This will damage the electronics of the modules. NOTE In an Agilent 1100 Series system, the individual modules are connected by a CAN cable. The Agilent 1100 control module can be connected to the CAN bus at any of the modules in the system. The Agilent Chemstation can be connected to the system by one GPIB cable at any of the modules, however, it is recommended to connect the GPIB cable to the detector. For more information about connecting the control module or ChemStation refer to the respective user manual. For connecting the Agilent 1100 Series equipment to non-Agilent 1100 Series equipment, see Autosampler manual).

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 33 7 Connect additional cables as required (see Figure 13 on page 34). Figure 12 Power Connectors and Safety Levers at Rear of thermostatted Autosampler. Safety lever Power connector Autosampler - Thermostat Cable

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Figure 13 Cable Connections. GPIB RS232C CAN-bus Remote Vial number output Relay contacts CAN cable to previous module Therm-ALS cable Thermostat - Autosampler cable CAN cable to next module

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 35 Stage 3: Flow Connections 1 Connect the pump outlet capillary to port 1 of the injection valve. 2 Connect column-compartment inlet capillary to port 6 of the injection valve. 3 Ensure that the waste tube is positioned inside the leak channel. WARNING When opening capillary or tube fittings, solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used. Figure 14 Hydraulic Connections To metering device To column compartment From pump From needle seat To waste Plug Column compartment Pump Waste tube in Leak channel

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Stage 4: Installing the Sample Tray 1 Load the sample tray with sample vials as required. 2 Slide the sample tray into the autosampler so that the rear of the sample tray is seated firmly against the rear of the sample-tray area. 3 Press the front of the sample tray down to secure the tray in the autosampler. If the tray pops out of its position the air channel adapter is not inserted correctly. Figure 15 Installing the Sample Tray

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 37 Half-Tray Combinations Half-trays can be installed in any combination enabling both 1.8 ml-and 6 ml-vials to be used simultaneously. Numbering of Vial Positions The standard 100-vial tray has vial positions 1 to 100. However, when using two half-trays, the numbering convention is slightly different. The vial positions of the right-hand half tray begin at position 101 as follows: Left-hand 40-position tray: 1–40 Left-hand 15-position tray: 1–15 Right-hand 40-position tray: 101–140 Right-hand 15-position tray: 101–115 NOTE In the autosampler only the 100 vial tray is supported for temperature control of the vials. Nevertheless the half trays of the standard autosampler (G1313A) can be used in the thermostatted autosampler as well. However when these trays are installed cooling or heating of the vials in the tray will not work.

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Figure 16 Numbering of Tray Positions. Position 1 Position 101

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 39 Stage 5: Installing Tray Cover and Front Cover 1 Fix the tray cover in the clips of the left autosampler cover side by sliding it in position. Do not close the tray cover. 2 Position the front cover in the top left corner of the autosampler and turn it towards the instrument. Press the stop fastener to secure it in the right side cover of the autosampler. 3 Close the tray cover. Figure 17 Installation of Tray Cover and Front Cover of the Autosampler Autosampler front door Tray cover

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat Stage 6: Turning on the Thermostatted Autosampler 1 Depress the power switch to turn on the two modules of the autosampler. Stage 7: Update of Control Module Firmware If the control module has a firmware revision A.01.30 or higher, you do not need to update the control module firmware. If you have a control module version A.01.30 or lower, update the firmware as described. 1 Disconnect the control module before inserting the PC card. 2 Insert the PC card into the card slot of the control module. 3 Reconnect the control module for restarting it. 4 Press “System” (F5) - “Records” (F4). Highlight the LC-System line in the display using the up-down arrows. 5 Press “FW-Update” (F5). 6 Select the file for the firmware update (LCB202en.BIN). 7 Press “Execute” and select “Yes” to confirm loading of the new firmware. The control module reboots and loads the firmware indicated by (.) and (*) on the display. When finished with the update the control module reboots again. 8 Check that the correct firmware was loaded by pressing “System” (F5) - “Records” (F4). 9 Disconnect the control module and remove the PC card by pressing the card-eject button. NOTE The power switch stays depressed (1) and a green indicator lamp in the power switch is on when the autosampler is turned on. When the line power switch stands out (Ø) and the green light is off, the autosampler is turned off.

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 41 Stage 8: Update of the Agilent ChemStation Software If you have a Agilent ChemStation software version A.05.02 or higher, you do not need to update your Agilent ChemStation software. If you have a Agilent ChemStation software version A.05.01 or lower, please update your software as described. Starting from revision A.04.01 or A.04.02 of the Agilent ChemStation If you have ordered the G1395A Upgrade kit you will be provided with a A.04.02 and a A.05.01 update CD ROM, which gives you the choice of keeping your old major software version or updating to A.05.01 release. 1 If you decide to update your A.04.01 Agilent ChemStation to A.05.01, use the provided A.05.01 CD-ROM and follow the steps described in the Installing your ChemStation manual, provided as portable document format (PDF) file on the A.05.01 CD-ROM in the directory MANUALS\ INSTALL\LC. If you do not already have the Adobe Acrobat reader installed, use the file MANUALS\READER\AR32e30.EXE to install the reader. 2 After you have updated your system, insert the 3.5” floppy labeled ‘Driver update Disk (A.05.02 Beta)’ and open a DOS prompt by selecting Start->RUN and typing command. At the DOS prompt type A: and press Enter and then HPUPDATE and press Enter only if your Agilent ChemStation is installed in the directory C:\HPCHEM. If your Agilent ChemStation is installed in a different directory, e.g. D:\HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter. CAUTION If you decide to update to A.05.01, ensure that your PC hard- and software meets the requirements for A.05.01. A.05.01 requires Windows 95 or Windows NT 4.0 as operating system and a Pentium PC with a minimum of 24 MB (NT -systems require also GPIB board Agilent 82341C). Publication 12-5965-6805E gives detailed information about the PC requirements. The application note can be obtained from the Internet (http://www.chem.agilent.com/cag/literature/apglit.html) or your local Agilent Technologies sales office.

1100 Series AT Reference Manual

2 Installing the G1330B Thermostat 3 If you decide to update to A.04.02, insert the A.04.02 CD-ROM and select SETUP on the CD-ROM using File Manager or Explorer. Select Yes to continue with the update. 4 After you have updated your system, insert the 3.5 inch floppy labeled Driver update Disk (A.04.03) and open a DOS prompt by selecting Start->RUN and typing command. At the DOS prompt type A: and press Enter and then HPUPDATE and press Enter only if your Agilent ChemStation is installed in the directory C:\HPCHEM. If your Agilent ChemStation is installed in a different directory, e.g. D:\HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter. Starting from revision A.05.01 of the Agilent ChemStation If you already have the Agilent ChemStation A.05.01 installed, you only need to install the update to A.05.02 Beta. This update comprises only the driver for the thermostatted autosampler, no other changes are made. 1 Insert the 3.5” floppy labeled Driver update Disk (A.05.02 Beta) and open a DOS prompt by selecting Start->RUN and typing command. At the DOS prompt type A: and press Enter and then HPUPDATE and press Enter only if your Agilent ChemStation is installed in the directory C:\HPCHEM. If your Agilent ChemStation is installed in a different directory, e.g. D:\ HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter.

Installing the G1330B Thermostat

2 1100 Series AT Reference Manual 43 Transporting the Thermostatted Autosampler When moving the autosampler around the laboratory, make sure that any condensed water inside the thermostat is removed. Tilt the module to the front, so that the water inside the thermostat can safely flow into the leak funnel. Otherwise no special precautions are needed for the modules. If the autosampler needs to be shipped to another location via carrier, ensure: • The two modules are shipped in separate boxes. • The transport assembly of the autosampler is parked, see “Park Arm (Park Gripper)” in your respective Reference Manual. • The vial tray is secured. If the autosampler is to be shipped to another location, the transport assembly of the autosampler must be moved to the park position to prevent mechanical damage should the shipping container be subjected to excessive shock. Also, ensure the vial tray is secured in place with suitable packaging, otherwise the tray may become loose and damage internal components. NOTE The ALS thermostat is heavy (20.7 Kg, 45.6 lbs). Carry the module, by putting your hands under the side covers in a central position of the unit.

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 3 Troubleshooting and Test Functions Overview of the Thermostat’s Indicators and Test Functions 46 Status Indicators 47 ALS Thermostat Error Messages 49

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Overview of the Thermostat’s Indicators and Test Functions Status Indicators The thermostatted autosampler is provided with two status indicators which indicate the operational state (prerun, run, and error states) of the instrument. Both are located on the autosampler module. The status indicators provide a quick visual check of the operation of the thermostatted autosampler (see "Status Indicators" on page 47). Error Messages In the event of an electronic, mechanical or hydraulic failure, the thermostatted autosampler modules generate an error message in the user interface. For each message, a short description of the failure, a list of probable causes of the problem, and a list of suggested actions to fix the problem are provided. This chapter shows only the error messages that are generated for the ALS thermostat. For detailed description of the autosampler error messages see “Autosampler Error Messages” in your corresponding Reference Manual.

Troubleshooting and Test Functions

3 1100 Series AT Reference Manual 47 Status Indicators Two status indicators are located on the front of the autosampler. The lower left indicates the power supply status, the upper right indicates the thermostatted autosampler status. Figure 18 Location of Status Indicators Status indicator Power supply indicator

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Power Supply Indicator The power supply indicator is integrated into the main power switch. When the indicator is illuminated (green) the power is ON. Instrument Status Indicator The instrument status indicator indicates one of four possible instrument conditions: • When the status indicator is OFF (and power switch light is on), the instrument is in a prerun condition, and is ready to begin an analysis. • A green status indicator, indicates the instrument is performing an analysis (run mode). • A yellow indicator indicates a not-ready condition. The instrument is in a not-ready state when it is waiting for a specific condition to be reached or completed (for example, front cover not installed), or while a self-test procedure is running. • An error condition is indicated when the status indicator is red. An error condition indicates the instrument has detected an internal problem which affects correct operation of the instrument. Usually, an error condition requires attention (for example, leak, defective internal components). An error condition always interrupts the analysis.

Troubleshooting and Test Functions

3 1100 Series AT Reference Manual 49 ALS Thermostat Error Messages Error messages are displayed in the user interface when an electronic failure occurs with the ALS thermostat module which requires attention. In the event of such a failure, the red status indicator at the front of the thermostatted autosampler is switched on, and an entry is written into the instrument log book. This section describes the meaning of ALS thermostat module error messages, and provides information on probable causes and suggested actions how to recover from error conditions.

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Fan Failed Each fan in the ALS thermostat is equipped with a speed sensor, that allows monitoring and control the speed of the fans. If the sensor shows no signal, when the fan is activated, it is very likely that the fan is defective. The number in brackets indicates the fan position. Position numbers of the fans are as follows (seen from front of the ALS thermostat) fan failed 1 - left side, fan in front position fan failed 2 - left side, fan in back position fan failed 3 - right side, fan in back position fan failed 4 - right side, fan in front position fan failed 5 - small cooler fan on top of cooling / heating module Probable Causes: • One of the 4 heat-sink fans is defective. • The cooling heating module fan is defective Suggested Actions: ✔See Exchange “"Exchanging the Heatsink Fans" on page 67 or “"Exchanging Heat Exchanger Fan" on page 65.

Troubleshooting and Test Functions

3 1100 Series AT Reference Manual 51 Temperature Control Failed (1 - 4) There are four Peltier elements build into the ALS thermostat for efficient cooling / heating. The electronics monitor the current through the Peltier elements. If the Peltier current is out of a specified limit the Peltier element is defective. Probable Causes: • One of the four peltier elements is defective. Suggested Action: ✔contact the Technologies service organization

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Temperature Sensor Failed (1 - 4) The ALS thermostat is equipped with four sensors to monitor the performance of the instrument. The sensors are connected to both sides of the Peltier elements for control of the cooling / heating efficiency. As the positioning of the sensor is critical to meet the performance requirements of the ALS thermostat the single sensors cannot be replaced separately. Probable Causes: • Temperature sensor is defective Suggested Action: ✔See “"Exchanging the Sensors" on page 69”.

Troubleshooting and Test Functions

3 1100 Series AT Reference Manual 53 Lost Contact to ALS Thermostat The autosampler control electronics continuously checks whether the ALS thermostat is active or not. Probable Causes: • Bad cable connection between autosampler and ALS thermostat • Cable between autosampler and thermostat module not connected • Defective electronic board in autosampler or ALS thermostat Suggested Actions: ✔Check cable connection between autosampler and ALS thermostat ✔Connect cable between autosampler and ALS thermostat ✔Exchange ASM board in the autosampler ✔Exchange TCA board in the thermostat ✔contact the Agilent Technologies service organization

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Power Fail for ALS Thermostat Module The +36V that is generated in the ALS thermostat power supply is checked by the autosampler electronics. If this voltage is missing the error message will be generated. Probable Causes: • ALS thermostat module not connected to line power • ALS thermostat module fuse(s) defective • ALS Thermostat power supply defective Suggested Actions: ✔Check correct power line connection ✔Check fuse(s) and replace if defective ✔contact the Agilent Technologies service organization

Troubleshooting and Test Functions

3 1100 Series AT Reference Manual 55 Temperature out of Range To protect the Peltier elements for damage their working range is limited to -3°C to 65°C. The error message is generated when at least one of the three Peltier elements exceeds this limit. Probable Causes: • The four main heat exchanger fans are not able to blow enough air through the ALS thermostat • Ambient temperature too high / low Suggested Actions: ✔Clean the air filters and heat exchanger fins in the ALS thermostat. Switch off the thermostat and wait 10 min for the peltier elements to equilibrate ✔Make sure that there is enough space on left and right side of the ALS thermostat and that the air channel inlets and oulets are not blocked ✔Make sure that the ambient air temperature is within its specified limits (4°C to 55°C)

1100 Series AT Reference Manual

3 Troubleshooting and Test Functions Bad Cooling / Heating Performance When turned on the ALS thermostat Peltier elements are activated according to the given setpoint or to given setpoint change. The electronics check whether there actual temperature is moving in the correct direction (e.g., decreasing when cooled down). Probable Causes: • Peltier element(s) defective • Peltier element(s) not connected or incorrectly connected (e.g., after repair or maintenance) • The four main heat exchanger fans are not able to blow enough air through the ALS thermostat Suggested Actions: ✔Check connections of the Peltier elements ✔Clean the air filters and heat exchanger fins in the ALS thermostat. Switch off the thermostat and wait 10 min for the peltier elements to equilibrate ✔Make sure that there is enough space on left and right side of the ALS thermostat and that the air channel inlets and oulets are not blocked ✔contact the Agilent Technologies service organization

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 4 Repairing the ALS Thermostat Module Introduction into Repairing the Thermostat 58 Overview of the Repairing the Thermostat 60 Exchanging Internal Parts 62

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module Introduction into Repairing the Thermostat Simple Repairs The ALS thermostat is designed for easy repair. Exchanging Internal Parts Some repairs may require exchange of defective internal parts. Exchange of these parts requires removing the ALS thermostat from the stack, removing the covers, and disassembling the ALS thermostat. The security lever at the power input socket prevents that the Thermostat cover is taken off when line power is still connected. Cleaning the Thermostat The thermostat covers should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the autosampler. WARNING To prevent personal injury, the power cable must be removed from the ALS thermostat before opening the Thermostat cover. Do not connect the power cable to the ALS thermostat while the covers are removed. CAUTION Electronic boards and components are sensitive to electrostatic discharge (ESD). In order to prevent damage always use an ESD protection (for example, the ESD wrist strap from the accessory kit) when handling electronic boards and components. WARNING Do not disconnect or reconnect the autosampler to ALS thermostat cable when the power cords are connected to either of the two modules. This will damage the electronics of the modules.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 59 Using the ESD Strap Electronic boards are sensitive to electronic discharge (ESD). In order to prevent damage, always use an ESD strap supplied in the standard accessory kit (see "Accessory Kit G1330-68705" on page 79) when handling electronic boards and components. 1 Unwrap the first two folds of the band and wrap the exposed adhesive side firmly around your wrist. 2 Unroll the rest of the band and peel the liner from the copper foil at the opposite end. 3 Attach the copper foil to a convenient and exposed electrical ground. Figure 19 Using the ESD Strap

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module Overview of the Repairing the Thermostat Figure 20 shows the main assemblies and their locations which can be repaired. Figure 20 Main assemblies Heat exchanger fan TCA Board Condensation leak drain Heat exchanger Power Supply

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 61 Exchanging the Power Supply Fuses The fuse holders are located on the rear panel of the ALS thermostat. 1 Switch off the power switch at the front of the thermostatted autosampler. 2 Remove the power cable from the two modules. 3 Insert the flat head screwdriver in the fuse holder, slightly press and turn counter clockwise to release the fuse holder from the socket. 4 Pull the fuse holder out of the socket. 5 Remove the fuse from the fuse holder. 6 Insert a new fuse in the fuse holder. 7 Reinsert the fuse holder and fix with the screwdriver. 8 Reinsert the power cables. 9 Switch on the power switch. Frequency When defective Tools required Flat head screwdriver Parts required Fuses T2.5 A/250V (CSA, UL listed), 2110-0015

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module Exchanging Internal Parts The procedures in this section describe how to exchange internal parts. You must remove the ALS thermostat from the stack in order to open the main cover. WARNING The following procedures require opening the main cover of the ALS thermostat. Always ensure the ALS thermostat is disconnected from the line power when the main cover is removed. The security lever at the power input socket prevents that the pump cover is taken off when line power is still connected. WARNING To disconnect the ALS thermostat from line, unplug the power cord. The power supply still uses some power, even if the switch on the front panel is turned off. WARNING When opening capillary or tube fittings solvents may leak out. Please observe appropriate safety procedures (for example, goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet supplied by the solvent vendor, especially when toxic or hazardous solvents are used. NOTE The electronics of the ALS thermostat will not allow operation when the top cover and the top foam are removed. A safety light switch on the main board will inhibit the operation of the ALS thermostat. Always operate the ALS thermostat with the top foam and top covers in place. CAUTION Internal components may be sensitive to electrostatic discharge (ESD). Always use an ESD kit when handling internal parts.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 63 Removing the Top Cover and Foam Tools required Screwdriver Pozidriv #1 Preparations for this procedure Switch off ALS at the main power switch. Disconnect ALS and ALS thermostat power cords. Remove ALS to ALS thermostat cable, and remove thermostat from stack. WARNING Do not disconnect or reconnect the autosampler to ALS thermostat cable when the power cords are connected to either of the two modules. This will damage the electronics of the modules. 1 Remove the front cover by pressing the both clip fasteners on both sides of the cover. 2 Move the lever towards the power socket. Lever

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module 3 Lift the clips on both sides of the top cover (1). Remove the top cover (2). 4 Unscrew the screws on the top plate and remove the plate by lifting its back first and then sliding to the front. 5 Unplug all wires at the TCA Board and remove them from top foam. 6 Remove the top foam. Screws

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 65 Exchanging Heat Exchanger Fan When required When defective Preparations for this procedure Removing the Top Cover and Foam (see "Removing the Top Cover and Foam" on page 63) item 1 to 4. Parts required Fan, PN 3160-1079 1 Unplug the connector at the TCA board and remove the wire from the Top Foam. 2 Carefully lift the Fan upwards. 3 Remove the Fan from the Fan Gasket. 4 Insert the new Fan in the Fan Gasket. See arrow at the side of Fan for correct airflow direction.

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module 5 Reinstall the fan and route the wire in the appropriate guide channel. 6 Reconnect the wire to the connector.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 67 Exchanging the Heatsink Fans When required Fan not running Preparations for this procedure Removing the Top Cover and Foam (see "Removing the Top Cover and Foam" on page 63). Parts required Fan, part number 3160-0884 1 Carefully lift the pair of Fans upwards. 2 Replace the defective fan, and reinsert the pair of Fans. Note the orientation of the fans. 3 Reinsert the Top Foam and route the cables in the guide channels in the Top Foam. 4 Reconnect the connectors in the correct order (See label on top of Power Supply).

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module 5 Replace the metal cover and top cover.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 69 Exchanging the Sensors When required Sensors defective Preparations for this procedure Removing the Top Cover and Foam (see "Removing the Top Cover and Foam" on page 63). Parts required Sensor flexboard, PN G1330-66504 Heatsink Compound, PN 6040-0454 1 Unscrew the six holding screws on top of the hardfoam encapsulation. 2 Carefully lift the upper part of hardfoam encapsulation away. Screws

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module 3 Remove the sealing adapter by carefully sliding it out of the seat. 4 Unplug the connectors at the TCA Board, careful lift the sensors out of their seat and remove the sensor flexboard. 5 Insert some Heatsink Compound into the four sensor seats, to assure good contact. 6 Install the new sensor flexboard carefully by pressing each sensor individually into the seats, and then route the flexboard through the sealing adapter channel.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 71 7 Reinsert the sealing adapter by carefully sliding it in to the seat. 8 Reinstall the hardfoam encapsulation. 9 Route the cables in the guide channels and reconnect the connectors in the correct order (See label on Power Supply). 10 Reinstall the Top Plate and the Top Cover.

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module Exchanging the TCA board When required TCA Board defective Preparations for this procedure Removing the Top Cover and Foam (see "Removing the Top Cover and Foam" on page 63), item 1 to 4. Parts required TCA board, part number G1330-66500 1 Remove all connectors on the TCA Board. 2 Unscrew the connector screws for the Autosampler to Thermostat cable.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 73 3 By carefully pulling back on the upper part of the TCA board unsnatch it from the holder. 4 Remove the TCA board. 5 Reinsert the new TCA board. 6 Gently push forward on the upper part of the board, until it snatches into place.

1100 Series AT Reference Manual

4 Repairing the ALS Thermostat Module 7 Refit the connector screws and reconnect all connectors. 8 Replace the top foam section, metal cover and top cover.

Repairing the ALS Thermostat Module

4 1100 Series AT Reference Manual 75 Assembling the Main Cover When required If cover is broken Tools required None Parts required Cover kit G1330-68703 (includes base, top, left and right) NOTE The cover kit contains all parts, but it is not assembled. WARNING In case you insert the left or right side in the opposite position, you may not be able to remove the side from the top part. 1 Place the top part on the bench and insert the left and right side into the top part. 2 Replace the cover. 3 Replace the ALS thermostat into the stack and reconnect the cables and capillaries. 4 Turn on the ALS thermostat. Front

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 5 ALS Thermostat Parts and Materials Main Assemblies 78 Accessory Kit G1330-68705 79 Main Assemblies 78 Foam Parts 81 Plastic Parts 82 Heat Exchanger Fan Parts 83

1100 Series AT Reference Manual

5 ALS Thermostat Parts and Materials Main Assemblies Figure 21 Main Assemblies Table 4 Main Assemblies Item Description Part Number 1 Fuse - Power Supply (T2.5A/250V; CSA, UL listed) 2110-0015 2 Fuse TCA - Board (T3A/250V; CSA, UL listed) 2110-0029 3 Front Cover 5062-8582 Cable, autosampler - ALS thermostat G1330-81600 2 1 3

ALS Thermostat Parts and Materials

5 1100 Series AT Reference Manual 79 Accessory Kit G1330-68705 Table 5 Accessory Kit Item Description Part Number 1 Waste Tube* * Reorder Number (5 m) 5062-2463 2 Waste Tube Assembly G1330-67300

1100 Series AT Reference Manual

5 ALS Thermostat Parts and Materials Main Assemblies Figure 22 Main Assemblies Table 6 Main Assemblies Item Description Part Number 1 Heat sink fans 3160-0884 2 TCA Board G1330-66500 3 Heat exchanger fan 3160-1079 2 1 3 1

ALS Thermostat Parts and Materials

5 1100 Series AT Reference Manual 81 Foam Parts Figure 23 Foam Parts Table 7 Main Assemblies Item Description Part Number 1 Top Foam G1330-40102 2 Bottom Foam G1330-40103 1 2

1100 Series AT Reference Manual

5 ALS Thermostat Parts and Materials Plastic Parts Figure 24 Plastic Parts Table 8 Main Assemblies Item Description Part Number 1 Cabinet Kit, includes base, top and sides G1330-68703 2 Front Cover 5062-8582 3 Leak Pan 5042-1307 2 1 3

ALS Thermostat Parts and Materials

5 1100 Series AT Reference Manual 83 Heat Exchanger Fan Parts Figure 25 Heat Exchanger Fan Parts Table 9 Main Assemblies Item Description Part Number 1 Fan Gasket G1330-44104 2 Heat Exchanger Fan 3160-1079 1 2

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies 6 Specifications Performance Specifications 86

1100 Series AT Reference Manual

6 Specifications Performance Specifications Table 10 Performance Specifications Agilent 1100 Series thermostatted autosampler Type Specification Temperature range: setable from 4°C to 40°C in 1° increments

Agilent 1100 Series Autosampler Thermostat

Reference Manual Agilent Technologies A Safety Information General Safety Information 88 Radio Interference 91 Sound Emission 92 Agilent Technologies on Internet 93

1100 Series AT Reference Manual

A Safety Information General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. General This is a Safety Class I instrument (provided with terminal for protective earthing) and has been manufactured and tested according to international safety standards. Operation Before applying power, comply with the installation section. Additionally the following must be observed. Do not remove instrument covers when operating. Before the instrument is switched on, all protective earth terminals, extension cords, auto-transformers, and devices connected to it must be connected to a protective earth via a ground socket. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in serious personal injury. Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any intended operation. Make sure that only fuses with the required rated current and of the specified type (normal blow, time delay, and so on) are used for replacement. The use of repaired fuses and the short-circuiting of fuseholders must be avoided. Some adjustments described in the manual, are made with power supplied to the instrument, and protective covers removed. Energy available at many points may, if contacted, result in personal injury. Any adjustment, maintenance, and repair of the opened instrument under voltage should be avoided as much as possible. When inevitable, this should be carried out by a skilled person who is aware of the hazard involved. Do not

1100 Series AT Reference Manual

89 attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present. Do not replace components with power cable connected. Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. Do not install substitute parts or make any unauthorized modification to the instrument. Capacitors inside the instrument may still be charged, even though the instrument has been disconnected from its source of supply. Dangerous voltages, capable of causing serious personal injury, are present in this instrument. Use extreme caution when handling, testing and adjusting. When working with solvents please observe appropriate safety procedures (e.g. goggles, safety gloves and protective clothing) as described in the material handling and safety data sheet by the solvent vendor, especially when toxic or hazardous solvents are used. Safety Symbols Table 1 shows safety symbols used on the instrument and in the manuals. Table 1 Safety Symbols Symbol Description The apparatus is marked with this symbol when the user should refer to the instruction manual in order to protect the apparatus against damage. Indicates dangerous voltages. Indicates a protected ground terminal. !

1100 Series AT Reference Manual

A Safety Information WARNING A warning alerts you to situations that could cause physical injury or damage to the equipment. Do not proceed beyond a warning until you have fully understood and met the indicated conditions. CAUTION A caution alerts you to situations that could cause a possible loss of data. Do not proceed beyond a caution until you have fully understood and met the indicated conditions.

1100 Series AT Reference Manual

91 Radio Interference Manufacturer’s Declaration This is to certify that this equipment is in accordance with the Radio Interference Requirements of Directive FTZ 1046/1984. The German Bundespost was notified that this equipment was put into circulation, the right to check the series for compliance with the requirements was granted. Test and Measurement If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises.

1100 Series AT Reference Manual

A Safety Information Sound Emission Manufacturer’s Declaration This statement is provided to comply with the requirements of the German Sound Emission Directive of 18 January 1991. This product has a sound pressure emission (at the operator position) < 70 dB. • Sound Pressure Lp < 70 dB (A) • At Operator Position • Normal Operation • According to ISO 7779:1988/EN 27779/1991 (Type Test)

1100 Series AT Reference Manual

93 Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/go/chem

1100 Series AT Reference Manual

95 Index A Agilent on internet, 93 air circulation, 21 ALS thermostat accessory kit, 24 ALS thermostat accessory kit contents, 24 ALS thermostat operation, 12 ALS thermostat parts and materials accessory kit, 79 main assemblies, 78, 79 ALS thermostat repairs power supply fuses, 61 assembling the main cover, 75 autosampler control, 14 B bench space, 21 C cleaning the ALS thermostat, 58 condensation, 10, 21 D damaged packaging, 23 damper, 65 damping unit, 65 delay volume, 25 delivery checklist, 23 E electrical connections, 17 CAN, 17 GPIB, 17 Remote, 17 RS-232, 17 thermostat - autosampler, 17 electronics, 14 electrostatic discharge (ESD), 58 environment, 20, 21 error messages, 46 ALS thermostat, 49 bad cooling / heating performance, 56 lost contact to ALS thermostat, 53 power fail ALS thermostat, 54 temperature control failed, 51 temperature out of range, 55 temperature sensor failed, 52 therm fan failed, 50 ESD (electrostatic discharge) strap, 59 exchanging damper, 65 fan, 67, 69, 72 internal parts, 58, 62 F failure, 46 fan, 67, 69 exchanging, 72 flow path, 25 fuses, 20, 61 H half-tray combinations, 37 I injection valve, 10 installing the thermostatted autosampler, 28 flow connections, 35 peparation, 29 power cable and interface cable, 32 sample tray, 36 tray cover and front cover, 39 turning on the instrument, 40 update of firmware or software, 40, 41 instrument status indicator, 48 internal parts, 62 internet, 93 introduction to the autosampler, 10 M missing parts, 23 multi-draw option, 10 O optimizing the stack configuration, 25 P park transport assembly, 23, 43 parts and materials, 23 performance specifications, 86 physical specifications, 22 power consideration, 20 power cords, 20 power requirements, 20 power supply indicator, 48

1100 Series AT Reference Manual

Index R repairs, 58 using the ESD strap, 59 S sample tray, 10 sample trays numbering of vial positions, 37 screwdriver, pozidriv #1, 67, 69, 72 security lever, 58 shipping, 23, 43 simple repairs, 58 site requirements, 20 specifications, 22, 86 status indicator, 46 status indicators, 47 storage, 21 T temperature, 21 top cover, 75 transport, 43 transport mechanism, 10 U unpacking, 23 unpacking the autosampler, 23 V vial racks, 10 vial tray, 43 vials, 10 W weight, 21

In This Book

This manual contains information about the Agilent 1100 Series thermostatted for autosamplers. Information for the individual autosamplers can be obtained from the corresponding autosampler reference manual. This supplemental manual describes the following: • installing the thermostat, • theory of operation, • troubleshooting the thermostat, • repair procedures for the thermostat, • parts and materials for the thermostat.

Other Thermostat Manuals