Keysight U1452A Operation & User’s Manual
Read Online
Read the Keysight U1452A Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
U1451A Insulation Tester
User’s Guide
Distributed by:
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ [email protected] ▪ www.datatec.de
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
Notices
© Keysight Technologies 2014
No part of this manual may be reproduced in
any form or by any means (including elec-
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ment and written consent from Keysight
Technologies as governed by United States
and international copyright laws.
Manual Part Number
U1451-90003
Edition
Edition 3, December 5, 2014
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
The material contained in this docu-
ment is provided “as is,” and is sub-
ject to being changed, without notice,
in future editions. Further, to the max-
imum extent permitted by applicable
law, Keysight disclaims all warran-
ties, either express or implied, with
regard to this manual and any infor-
mation contained herein, including
but not limited to the implied warran-
ties of merchantability and fitness for
a particular purpose. Keysight shall
not be liable for errors or for inciden-
tal or consequential damages in con-
nection with the furnishing, use, or
performance of this document or of
any information contained herein.
Should Keysight and the user have a
separate written agreement with
warranty terms covering the material
in this document that conflict with
these terms, the warranty terms in the
separate agreement shall control.
Technology Licenses
The hardware and or software described in
this document are furnished under a license
and may be used or copied only in accor-
dance with the terms of such license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to
the federal government include only those
rights customarily provided to end user cus-
tomers. Keysight provides this customary
commercial license in Software and techni-
cal data pursuant to FAR 12.211 (Technical
Data) and 12.212 (Computer Software) and,
for the Department of Defense, DFARS
252.227-7015 (Technical Data - Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or Com-
puter Software Documentation).
Safety Notices
CAUTION
A CAUTION notice denotes a haz-
ard. It calls attention to an operat-
ing procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated condi-
tions are fully understood and
met.
U1452A/U1452AT/U1451A User’s Guide
III
Safety Symbols
The following symbols on the instrument and in the documentation indicate precautions
which must be taken to maintain safe operation of the instrument.
Direct current (DC)
Caution, risk of electric shock
Alternating current (AC)
Caution, risk of danger (refer to
this manual for specific Warning
or Caution information)
Both direct and alternating
current
Category III 1000 V overvoltage
protection
Earth (ground) terminal
Category IV 600 V overvoltage
protection
Equipment protected throughout
by double insulation or reinforced
insulation
Do not use in distribution systems
with voltages higher than 600 V
CAT III
1000 V
CAT IV
600 V
U1452A/U1452AT/U1451A User’s Guide
Safety Considerations
Read the information below before using this tester. Model U1452A appears in all
illustrations.
WARNING
•
Do not use the tester if it is damaged. Before you use the tester, inspect the case.
Look for cracks or missing plastic. Pay particular attention to the insulation
surrounding the connectors.
•
Inspect the test leads for damaged insulation or exposed metal. Check the test
leads for continuity. Replace damaged test leads before you use the tester.
•
Do not operate the tester around explosive gas, vapor, or wet environments.
•
Do not apply more than the rated voltage (as marked on the tester) between
terminals, or between terminal and earth ground.
•
Before use, verify the tester's operation by measuring a known voltage.
•
When servicing the tester, use only the specified replacement parts.
•
Use caution when working above 60 VDC, 30 VAC RMS, or 42.4 V peak. Such
voltages pose a shock hazard.
•
When using the probes, keep your fingers behind the finger guards on the probes.
•
Connect the common test lead before you connect the live test lead. When you
disconnect the leads, disconnect the live test lead first.
•
Remove the test leads from the tester before you open the battery cover.
•
Do not operate the tester with the battery cover or portions of the cover removed
or loosened.
•
To avoid false readings, which may lead to possible electric shock or personal
injury, replace the battery as soon as the low battery indicator appears and
flashes.
•
Ensure that you do not perform insulation resistance tests in distribution systems
with voltages higher than 600 V.
•
For insulation resistance tests, ensure that you select a suitable test voltage for
the equipment to be tested.
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
V
CAUTION
•
Disconnect circuit power and discharge all high-voltage capacitors before testing
resistance, continuity, or capacitance.
•
Use the proper terminals, function, and range for your measurements.
•
This device is for use at altitudes of up to 2,000 m.
•
Always use the specified battery type. The power for the tester is supplied with four
1.5 V AA batteries. Observe the correct polarity markings before you insert the
batteries to ensure proper insertion of the batteries in the tester.
•
You are advised to use low leakage batteries when changing to new batteries.
Please remember to remove the batteries when the tester is not in use for a long
period of time. Warning on the risk of battery leakage.
U1452A/U1452AT/U1451A User’s Guide
Environmental Conditions
This instrument is designed for indoor use and in an area with low condensation. The table
below shows the general environmental requirements for this instrument.
Environmental condition
Requirement
Temperature
•
Operating condition
•
–20 °C to 55 °C, 0% to 80% RH (using Alkaline batteries),
20 minutes operating time
•
Storage condition
•
–40 °C to 70 °C, 0% to 80% RH (without batteries)
Humidity
Full accuracy up to 80% RH for temperatures up to 30 °C,
decreasing linearly to 50% RH at 55 °C
Altitude
Up to 2,000 meters
Pollution degree
Pollution degree II
NOTE
The U1452A/U1452AT/U1451A Insulation Tester complies with the following safety and EMC
requirements:
•
Safety compliance
•
Designed in compliance to IEC/EN 61010-1:2010 for Category III 1000 V and Category IV 600 V
•
Designed in compliance to IEC/EN 61557-1, IEC/EN 61557-2, and IEC/EN 61557-4
•
EMC compliance
•
Commercial limits compliance with IEC 61326-1:2005/EN 61326-1:2006
U1452A/U1452AT/U1451A User’s Guide
VII
Regulatory Markings
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
The C-tick mark is a registered
trademark of the Spectrum
Management Agency of Australia. This
signifies compliance with
the Australia EMC Framework
regulations under the terms of the
Radio Communication Act of 1992.
ICES/NMB-001 indicates that this ISM
device complies with the Canadian
ICES-001.
Cet appareil ISM est confomre a la
norme NMB-001 du Canada.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical or electronic product in
domestic household waste.
The CSA mark is a registered
trademark of the Canadian Standards
Association.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life
of the product.
U1452A/U1452AT/U1451A User’s Guide
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement.
This affixed product label indicates that you must not discard this electrical or electronic
product in domestic household waste.
Product Category:
With reference to the equipment types in the WEEE directive Annex 1, this instrument is
classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit
www.keysight.com/environment/product
for more information.
U1452A/U1452AT/U1451A User’s Guide
IX
Declaration of Conformity (DoC)
The Declaration of Conformity (DoC) for this instrument is available on the Keysight
website. You can search the DoC by its product model or description at:
http://regulations.products.keysight.com/DoC/search.htm
NOTE
If you are unable to search for the respective DoC, please contact your
local Keysight representative.
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
XI
Table of Contents
1
Introduction
About This Manual 2
Documentation map 2
Safety notes 2
Preparing Your Tester 3
Check the shipment 3
Install or change the batteries 3
Turn on your tester 6
Select the range 6
Adjust the tilt stand 7
Connect to the Handheld Meter Logger Software 8
Connect the Bluetooth adapter 9
Your Tester in Brief 10
Dimensions 10
Overview 12
Rotary switch 14
Keypad 16
Display screen 20
Input terminals 24
Cleaning Your Tester 25
Additional Features 26
Automatic power-off 26
Hazardous voltage indication 26
Power-on options 26
2
Making Measurements
Insulation Resistance Test 30
Using the Remote Switch Probe 32
U1452A/U1452AT/U1451A User’s Guide
Locking the test 33
Timed insulation resistance/earth-bond resistance test 34
Measuring the Dielectric Absorption Ratio (DAR) 35
Measuring the Polarization Index (PI) 36
Viewing the leakage current 37
Earth-Bond Resistance Test 38
Measuring AC or DC Voltage 41
Auto AC or DC signal identification 43
Measuring Frequency 44
Measuring Resistance 46
Continuity Test 48
Measuring Capacitance 50
3
Tester Features
Making Relative Measurements (Null) 54
Freezing the Display (TrigHold and AutoHold) 55
Performing Limit Comparisons (Limit) 57
Recording Measurement Data (Log) 59
Performing manual logs (HAND) 60
Performing interval logs (AUTO) 60
Performing event logs (TRIG) 61
Reviewing Previously Recorded Data (View) 63
4
Setup Options
Using the Setup Menu 66
Editing numerical values 67
Setup Menu Summary 68
Setup Menu Items 71
U1452A/U1452AT/U1451A User’s Guide
XIII
Changing the variation count 71
Enabling smooth mode 71
Changing the beep frequency 72
Changing the auto power-off (APO) timer 73
Changing the LCD backlight timeout 73
Changing the continuity alert 74
Disabling the power-on melody 75
Changing the battery type 75
Resetting the tester’s Setup options 76
Changing the display count 76
Changing the cable length unit 77
Changing the cable length scale 78
Changing the alert indicators 78
Changing the sample interval duration 79
Changing the recording option 80
Changing the IR and EBR test period 80
Changing the maximum inhibit voltage for IR tests 81
Disabling the lock once feature 82
Changing the button operation on the remote switch probe 82
Changing the Dielectric Absorption Ratio (DAR) for IR
tests 83
5
Characteristics and Specifications
Product Characteristics 86
Specification Considerations 88
Measurement Category 88
Electrical Specifications 89
DC specifications 89
AC specifications 91
Capacitance specifications 92
Frequency specifications 93
Sensitivity for voltage measurements 94
U1452A/U1452AT/U1451A User’s Guide
Insulation resistance specifications 95
Earth-bond resistance specifications 97
EN61557 specifications 97
Display update rate (approximate) 99
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
XV
List of Figures
Figure 1-1 Keysight Handheld Meter Logger Software 8
Figure 1-2 Bluetooth adapter connection 9
Figure 1-3 Width dimension 10
Figure 1-4 Height and depth dimensions 11
Figure 1-5 Display screen allocation example 20
Figure 1-6 Connecting the remote switch probe 24
Figure 2-1 Insulation resistance test example 31
Figure 2-2 T/Time operation 34
Figure 2-3 DAR operation 35
Figure 2-4 PI operation 36
Figure 2-5 Earth-bond resistance test example 39
Figure 2-6 AC or DC voltage measurement example 42
Figure 2-7 Definition of frequency 45
Figure 2-8 Resistance measurement example 47
Figure 2-9 Continuity test example 49
Figure 2-10 Capacitance measurement example 51
U1452A/U1452AT/U1451A User’s Guide
XVII
List of Tables
Table 1-1 Front panel part descriptions 12
Table 1-2 Rear panel parts 13
Table 1-3 U1452A/U1452AT/U1451A rotary switch functions 14
Table 1-4 U1452A/U1452AT/U1451A keypad functions 16
Table 1-5 General annunciators 20
Table 1-6 Measurement units display 22
Table 1-7 Analog bar graph display 23
Table 1-8 Terminal connections for different measuring
functions 24
Table 1-9 Power-on options 26
Table 2-1 Rotary switch position for insulation resistance
tests 30
Table 2-2 Earth-bond resistance test position 38
Table 2-3 AC and DC voltage measurement positions 41
Table 2-4 Resistance measurement position 46
Table 2-5 Continuity test position 48
Table 2-6 Capacitance measurement position 50
Table 3-1 Limit settling default values 57
Table 3-2 Log maximum capacity 59
Table 3-3 Event log trigger conditions 61
Table 4-1 Setup menu key functions 66
Table 4-2 Setup menu item descriptions 68
Table 5-1 DC voltage specifications with accuracy of ± (% of
reading + no. of least significant digit) 89
Table 5-2 Resistance specifications with accuracy of ± (% of
reading + no. of least significant digit) 90
Table 5-3 AC voltage specifications with accuracy of ± (% of
reading + no. of least significant digit) 91
Table 5-4 Capacitance specifications with accuracy of ± (% of
reading + no. of least significant digit)[1][2][3] 92
Table 5-5 Frequency specifications with accuracy of ± (% of
reading + no. of least significant digit)[1][2] 93
Table 5-6 Sensitivity for voltage measurements 94
Table 5-7 Insulation resistance specifications with accuracy of ±
(% of reading + no. of least significant digit) 95
U1452A/U1452AT/U1451A User’s Guide
Table 5-8 Earth-bond resistance specifications with accuracy of ±
(% of reading + no. of least significant digit)[1][2] 97
Table 5-9 Display update rate (approximate) 99
U1452A/U1452AT/U1451A Insulation Tester
User’s Guide
1
Introduction
About This Manual 2
Documentation map 2
Safety notes 2
Preparing Your Tester 3
Check the shipment 3
Install or change the batteries 3
Turn on your tester 6
Select the range 6
Adjust the tilt stand 7
Connect to the Handheld Meter Logger Software 8
Connect the Bluetooth adapter 9
Your Tester in Brief 10
Dimensions 10
Overview 12
Rotary switch 14
Keypad 16
Display screen 20
Input terminals 24
Cleaning Your Tester 25
Additional Features 26
Automatic power-off 26
Hazardous voltage indication 26
Power-on options 26
This chapter teaches you how to set up your tester for the
first time. An introduction to all the features of the tester is
also given.
all-guides.co
m
About This Manual
2
U1452A/U1452AT/U1451A User’s Guide
About This Manual
Documentation map
The following manuals and software related to the
U1452A/U1452AT/U1451A Insulation Tester are available for download.
Please visit our website at http://www.keysight.com/find/hhTechLib for the
latest version.
Check the manual edition on the first page of each manual.
User’s Guide. This manual.
Quick Start Guide. Printed copy for outdoor use, included with shipment.
Service Guide. Downloadable from
http://www.keysight.com/find/hhTechLib
Keysight Handheld Meter Logger Software, Help, and Quick Start
Guide. Downloadable from http://www.keysight.com/hhmeterlogger
Safety notes
The following safety notes are used throughout this manual. More
pertinent safety notes for using this product are located under the “Safety
Symbols” section.
CAUTION
Caution denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, could result in damage to or destruction of the product. Do
not proceed beyond a caution notice until the indicated conditions are fully understood
and met.
WARNING
Warning denotes a hazard. It calls attention to a procedure which, if not correctly
performed or adhered to, could result in injury or loss of life. Do not proceed beyond
a warning note until the indicated conditions are fully understood and met.
Preparing Your Tester
U1452A/U1452AT/U1451A User’s Guide
3
Preparing Your Tester
Check the shipment
When you receive your tester, check the shipment according to the
following procedure.
1 Inspect the shipping container for damage. Signs of damage may include
a dented or torn shipping container or cushioning material that
indicates signs of unusual stress or compacting. Save the packaging
material in case the tester needs to be returned.
2 Carefully remove the contents from the shipping container, and verify
that the standard accessories and your ordered options are included in
the shipment according to the standard shipped items list found in the
printed copy of the U1452A/U1452AT/U1451A Quick Start Guide.
3 For any question or problems, refer to the Keysight contact numbers on
the back of this manual.
Install or change the batteries
Your tester is powered by four 1.5 V AA alkaline batteries (included in the
shipment). When you receive your tester, the batteries are not installed.
Use the following procedure to install or change the batteries.
CAUTION
Before you proceed with the batteries installation, remove all cable connections to the
terminals and ensure that the rotary switch is at the
position. Use only the battery
type specified in the “Product Characteristics” on page 86.
Preparing Your Tester
4
U1452A/U1452AT/U1451A User’s Guide
1 Remove the orange rubber holster. Pull
from a top corner and stretch the orange
rubber holster off the tester.
2 Loosen and remove the two screws with
a suitable Phillips screwdriver as shown
on the right.
3 Lift up and remove the battery cover as
shown on the left.
4 Lift the inner rubber cover to access the
battery compartment.
5 Observe the proper batteries polarity. The
terminal ends of each battery are
indicated inside the battery
compartment. Insert four 1.5 V AA
batteries.
6 Ensure that the inner rubber cover is
positioned properly.
7 Replace the battery cover back in its
original position and tighten the screws.
8 Finally fit the orange rubber holster back
on the tester.
Preparing Your Tester
U1452A/U1452AT/U1451A User’s Guide
5
The battery level indicator in the upper left- hand corner of the display
indicates the relative condition of the batteries.
Replace the batteries as soon as possible when the low battery indicator
(
) is shown.
Battery level indicator
WARNING
To avoid false readings, which could lead to possible electric shock or personal
injury, replace the battery as soon as the low battery indicator appears. Do not
discharge the battery by shorting the battery or reverse the battery polarity in any of
the subjects.
CAUTION
To avoid instruments being damage from battery leakage:
•
Always remove dead batteries immediately.
•
Always remove the batteries and store them separately if the tester is not going to
be used for a long period.
Preparing Your Tester
6
U1452A/U1452AT/U1451A User’s Guide
Turn on your tester
To power ON your tester, turn the rotary switch from the
position to
any other position.
Select the range
The tester’s selected range is always displayed on the right- hand end of
the bar graph.
Selected range
Autoranging indicator
all-guides.co
m
Preparing Your Tester
U1452A/U1452AT/U1451A User’s Guide
7
Pressing
changes the tester range (and disables auto- ranging). Each
additional press of
(in manual ranging) sets the tester to the next
higher range, unless it is already in the highest range, at which point the
range switches to the lowest range.
Press and hold
to switch the tester to auto- ranging. Auto- ranging is
convenient because the tester automatically selects an appropriate range
for sensing and displaying each measurement.
Adjust the tilt stand
To adjust the tester to a 60° standing position, pull the tilt- stand outward
to its maximum reach.
NOTE
•
Changing the tester range (and disabling auto-ranging) is not allowed for earth-bond
resistance tests and insulation resistance tests.
•
In auto-range, the tester selects the lowest range to display the highest available
precision (resolution) for the input signal.
•
If a reading is greater than maximum available range, OL (overload) is shown on the
display — except for earth-bond resistance tests and insulation resistance tests where
to indicate maximum reading, > is shown on the display instead.
To PC (host)
IR-USB cable
Pull until maximum reach
Preparing Your Tester
8
U1452A/U1452AT/U1451A User’s Guide
Connect to the Handheld Meter Logger Software
You can use the IR communication link (IR communication port, located at
the rear panel) and the Keysight Handheld Meter Logger Software to
control your tester remotely, perform data logging operations, and transfer
the contents of your tester’s memory to a PC.
Ensure that the Keysight logo on the U1173A IR- USB cable connected to
the tester is facing up. Firmly push the IR head into the tester’s IR
communication port until it snaps into place.
Refer to the Keysight Handheld Meter Logger Software Help and Quick Start
Guide for more information on the IR communication link and the
Keysight Handheld Meter Logger Software.
The Keysight Handheld Meter Logger Software and its supporting
documents (Quick Start Guide and Help) are available for download from
http://www.keysight.com/hhmeterlogger.
Figure 1-1
Keysight Handheld Meter Logger Software
Preparing Your Tester
U1452A/U1452AT/U1451A User’s Guide
9
Connect the Bluetooth adapter
The U1117A Infrared (IR)- to- Bluetooth® adapter allows you to connect
the tester wirelessly to any Windows PC, Android device, or iOS device.
The U1117A is compatible with the following application or software:
• Keysight Handheld Meter Logger (for Windows PC)
• Keysight Mobile Meter (for Android or iOS devices)
• Keysight Mobile Logger (for Android or iOS devices)
Snap the optic side of the U1117A to the tester’s IR communication port
(see Figure 1- 2).
Refer to the Keysight U1117A IR- to- Bluetooth Adapter Operating
Instructions (download from http://www.keysight.com/find/U1117A) for
more information on how to set up the U1117A to a Windows PC, Android
device, or iOS device.
Figure 1-2
Bluetooth adapter connection
Bluetooth adapter
Your Tester in Brief
10
U1452A/U1452AT/U1451A User’s Guide
Your Tester in Brief
Dimensions
Front view
Figure 1-3
Width dimension
100 mm
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
11
Rear and side view
Figure 1-4
Height and depth dimensions
218 mm
58 mm
all-guides.co
m
Your Tester in Brief
12
U1452A/U1452AT/U1451A User’s Guide
Overview
Front panel
The front panel parts of your tester are described in this section.
Table 1-1 Front panel part descriptions
Legend
Description
Learn more on:
1
Red LED indicator
page 54
2
Display screen
page 20
3
Keypad
page 16
4
Rotary switch
page 14
5
Input terminals
page 24
2
1
3
4
5
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
13
Rear panel
The rear panel parts of your tester are described in this section.
Table 1-2 Rear panel parts
Legend
Description
Learn more on:
1
IR communication port
page 8
2
Test lead/probe holders
-
3
Battery access (under the orange rubber holster)
page 3
4
Tilt stand
page 7
5
Fuse access (under the orange rubber holster)
-
1
2
3
4
5
Your Tester in Brief
14
U1452A/U1452AT/U1451A User’s Guide
Rotary switch
The measurement functions for each rotary switch position are described
in Table 1- 3. Turning the rotary switch changes the measurement function
and resets all other measurement options.
WARNING
Remove the test leads from the measuring source or target before changing the
rotary switch position.
NOTE
Press
to select the alternate measurement function(s) or test methods for insulation
resistance tests. See page 16 for more information on the
key.
U1452A
U1452AT
U1451A
Table 1-3 U1452A/U1452AT/U1451A rotary switch functions
Legend
Measurement function
U1452A
U1452AT
U1451A
Learn more on:
50 V Insulation resistance test
✔
✔
-
page 38
T - Timed test
✔
✔
-
DAR - Dielectric Absorption Ratio test
✔
✔
-
PI - Polarization Index Test
✔
✔
-
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
15
100 V Insulation resistance test
✔
✔
-
page 38
T - Timed test
✔
✔
-
DAR - Dielectric Absorption Ratio test
✔
✔
-
PI - Polarization Index Test
✔
✔
-
250 V Insulation resistance test
✔
-
✔
page 38
T - Timed test
✔
-
✔
DAR - Dielectric Absorption Ratio test
✔
-
-
PI - Polarization Index Test
✔
-
-
500 V Insulation resistance test
✔
-
✔
page 38
T - Timed test
✔
-
✔
DAR - Dielectric Absorption Ratio test
✔
-
-
PI - Polarization Index Test
✔
-
-
1000 V Insulation resistance test
✔
-
✔
page 38
T - Timed test
✔
-
✔
DAR - Dielectric Absorption Ratio test
✔
-
-
PI - Polarization Index Test
✔
-
-
Earth-bond resistance test
✔
✔
✔
page 38
T - Timed test
✔
✔
✔
Resistance measurement
✔
✔
✔
page 46
Continuity test
✔
✔
✔
page 48
Capacitance measurement
✔
✔
✔
page 50
Auto voltage measurement
✔
✔
✔
page 30
DC voltage measurement
✔
✔
✔
AC voltage measurement
✔
✔
✔
Table 1-3 U1452A/U1452AT/U1451A rotary switch functions (continued)
Legend
Measurement function
U1452A
U1452AT
U1451A
Learn more on:
Your Tester in Brief
16
U1452A/U1452AT/U1451A User’s Guide
Keypad
The operation of each key is explained in Table 1- 4 below. Pressing a key
enables a function, displays a related symbol, and emits a beep. Turning
the rotary switch to another position resets the current operation of the
key.
U1452A/U1452AT
U1451A
Table 1-4 U1452A/U1452AT/U1451A keypad functions
Legend
Function when pressed for:
Less than 1 second
More than 1 second
Insulation Resistance (IR) Test: Initiates an IR test (when the rotary switch is in one of the Ω Mega
positions) as long as
is held — the tester sources (outputs) a high voltage and measures
insulation resistance and
is shown on the display.
Earth-Bond Resistance (EBR) Test: Initiates an EBR test (when the rotary switch is in the Ω EB
position) as long as
is held — the tester measures earth-bond resistance and
is
shown on the display.
Trig Hold: Freezes the present reading in the
display (except when the rotary switch is in one
of the Ω Mega positions or in the Ω EB position).
• In TrigHold mode, press
to manually
trigger the holding of the next measured
value.
• Press and hold
again to exit this
mode.
Auto Hold: Automatically freezes the present
reading once the reading is stable (except when
the rotary switch is in one of the Ω Mega
positions or in the Ω EB position).
• In AutoHold mode, the reading is updated
automatically once the reading is stable and
the count setting is exceeded.
• Press and hold
again to exit this
mode.
all-guides.co
m
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
17
Lock: Press
to lock the insulation test or
earth-bond resistance test (when the rotary
switch is in the appropriate position).
Press
>
to initiate an IR or EBR
test. The test will remain active until you press
or
again to release the lock.
View: Press and hold
to enter the Log
Review menu.
• Press
to cycle through the previously
recorded manual (VIEW H), interval (VIEW A),
or event (VIEW E) logging data.
• Press
or
to view first or last
logged data respectively.
• Press
or
to scroll through the logged
data.
• Press
to delete the last logged data.
• Press and hold
to clear all the logged
data for the selected logging mode.
• Press and hold
again to exit this mode.
Hz: Press
to display the frequency for
voltage or current measurements.
Press
again to disable the frequency
display.
Esc: Press
in the Setup menu to discard
your changes.
Press
to switch or cycle between the
default and alternate measurement function(s).
Log: The recording option (HAND, AUTO, or
TRIG) must first be selected in the Setup menu
(see page 80).
• HAND (manual data logging) — Press and
hold
to log the present reading into the
memory. The display will return to normal
after a short while (≈ 1 second). To manually
log another reading, press and hold
again.
• AUTO (automatic data logging) — Press and
hold
to enable the automatic data
logging mode, where data is logged at the
interval defined in the Setup menu (see
page 79). Press and hold
again to exit
this mode.
• TRIG (event data logging) — Press and hold
to enable the event data logging mode,
where data is logged each time a triggering
condition is satisfied (see page 62). Press and
hold
again to exit this mode.
T: Configures the tester for a timed test (when
the rotary switch is in one of the Ω Mega
positions or the Ω EB position).
The test will start when you press
.
DAR: Configures the tester for a dielectric
absorption ratio test (when the rotary switch is
in one of the Ω Mega positions).
The test will start when you press
.
PI: Configures the tester for a polarization index
test (when the rotary switch is in one of the
Ω Mega positions).
The test will start when you press
.
Table 1-4 U1452A/U1452AT/U1451A keypad functions (continued)
Legend
Function when pressed for:
Less than 1 second
More than 1 second
U1452A/U1452AT
U1451A
U1452A/U1452AT
U1451A
Your Tester in Brief
18
U1452A/U1452AT/U1451A User’s Guide
Limit: Press
to enable the comparison for
limit mode.
• Press
again to set the comparison
value. Use the arrow keys to change the value
shown and press
to save your changes.
• Press and hold
to exit this mode.
Smooth: Press and hold
to smoothen the
refresh rate of the readings.
Press and hold
again to exit this mode.
Range: Press
to set a manual range and
disable auto-ranging.
Press
again to cycle through each
available measurement range.
Auto: Press and hold
to enable
auto-ranging.
Leak: Press
to display the leakage current.
Null: Press
to enable the relative function.
• The displayed value is saved as a reference to
be subtracted from subsequent
measurements.
• Press
again to view the stored reference
value that has been saved. The display will
return to normal after a brief period of time
(approx. 3 seconds).
• Pressing
while the stored reference
value is being displayed will cancel the
relative function.
-
Table 1-4 U1452A/U1452AT/U1451A keypad functions (continued)
Legend
Function when pressed for:
Less than 1 second
More than 1 second
U1452A/U1452AT
U1451A
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
19
: Press
to enable or disable the LCD
backlight.
Setup: Press and hold
to enter the Setup
menu.
• In the Setup menu, press
or
to
navigate through the menu pages. Press
or
at each menu page to move the cursor
to a specific menu item.
• Press
to change the value of the
selected menu item. Use the arrow keys to
change the value shown.
• Press
again to save your changes, or
press
to discard your changes.
• Press and hold
again to exit the Setup
menu.
Table 1-4 U1452A/U1452AT/U1451A keypad functions (continued)
Legend
Function when pressed for:
Less than 1 second
More than 1 second
Your Tester in Brief
20
U1452A/U1452AT/U1451A User’s Guide
Display screen
The display annunciators of your tester are described in this section. See
also “Measurement units” on page 22 for a list of available measurement
signs and notations and “Analog bar graph” on page 23 for a tutorial on
the analog bar graph located at the bottom of your display screen.
Display annunciators
The display annunciators of your tester are described in the Table 1- 5.
Figure 1-5
Display screen allocation example
Table 1-5 General annunciators
Legend
Description
Remote control enabled
Battery capacity indication
Analog bar graph
Test indication for insulation resistance and earth-bond resistance tests
Hazardous voltage sign for measuring voltage ≥30 V or OL (overload)
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
21
Auto hold/Trigger hold enabled
Polarization Index test enabled
Dielectric Absorption Ratio test enabled
Timed test enabled
Relative (Null) enabled
Smooth mode enabled
Auto-ranging enabled or Auto signal indicator enabled
Greater than range (for insulation resistance and earth-bond resistance tests)
Test and Test Lock indication for insulation resistance and earth-bond
resistance tests
Primary display
AC or DC indication
Measuring units for primary display
Test time indication for earth-bond resistance and insulation resistance tests
APO (Auto Power-Off) enabled
Limit comparison enabled
Data logging in progress
Table 1-5 General annunciators (continued)
Legend
Description
all-guides.co
m
Your Tester in Brief
22
U1452A/U1452AT/U1451A User’s Guide
Measurement units
The available signs and notations for each measurement function in your
tester are described in Table 1- 6. The units listed below are applicable to
the primary display and secondary display measurements of your tester.
View mode for reviewing previously logged data
Audible continuity test selected
Secondary display
Measuring units and AC+DC indication for secondary display
Tertiary display
Measuring units for tertiary display
Table 1-6 Measurement units display
Sign/Notation
Description
G
Giga
1E+09 (1000000000)
M
Mega
1E+06 (1000000)
k
kilo
1E+03 (1000)
n
nano
1E–09 (0.000000001)
μ
micro
1E–06 (0.000001)
m
milli
1E–03 (0.001)
mV, V
Voltage, units for voltage measurement
nF, μF, mF
Farad, units for capacitance measurement
Table 1-5 General annunciators (continued)
Legend
Description
Your Tester in Brief
U1452A/U1452AT/U1451A User’s Guide
23
Analog bar graph
The analog bar emulates the needle on an analog tester, without displaying
the overshoot.
The “–” sign indicates that the measured or calculated value is negative.
Each segment is presented as a ratio to the range value indicated on the
peak bar- graph. The unit and range will be indicated according to various
measurements (see Table 1- 7).
Ω, kΩ, MΩ, GΩ,
Ohm, units for resistance measurement
kHz, Hz
Hertz, units for frequency measurement
m
Meter, unit for length
ft
Feet, unit for length
Table 1-6 Measurement units display (continued)
Sign/Notation
Description
NOTE
For frequency measurements, the bar graph does not represent the primary display value.
For example, when frequency is displayed on the primary display during voltage
measurement, the bar graph represents the voltage value (not the frequency value).
Table 1-7 Analog bar graph display
Function
Bar graph
V
Ω , Ω EB , Ω Mega
Your Tester in Brief
24
U1452A/U1452AT/U1451A User’s Guide
Input terminals
The terminal connections for the different measurement functions of your
tester are described in the table below.
WARNING
To avoid damaging this device, do not exceed the input limit.
Table 1-8 Terminal connections for different measuring functions
Rotary switch position
Input terminals
Overload protection
1000 Vrms
1000 Vrms for
short circuit <0.3 A
440 mA/1000 V, 30 kA
fast-acting fuse
Figure 1-6
Connecting the remote switch probe
The remote switch probe is
used for insulation resistance
(IR) and earth-bond resistance
(EBR) tests.
Remote switch
probe terminal for
IR and EBR tests
Cleaning Your Tester
U1452A/U1452AT/U1451A User’s Guide
25
Cleaning Your Tester
Dirt or moisture in the terminals can distort readings. Follow the steps
below to clean your tester.
1 Turn the tester off, and remove the test leads.
2 Turn the tester over, and shake out any dirt that may have accumulated
in the terminals.
Wipe the case with a damp cloth and mild detergent — do not use
abrasives or solvents. Wipe the contacts in each terminal with a clean
swab dipped in alcohol.
WARNING
To avoid electrical shock or damage to the tester, ensure that the insides of the
casing stay dry at all times.
U1452A/U1452AT/U1451A User’s Guide
Additional Features
Automatic power-off
Your tester automatically turns off if the rotary switch is not moved or a
key is not pressed for 10 minutes (default). Pressing any key will turn the
tester back on after it is powered off automatically.
To change the timer period or completely disable the automatic power- off,
refer to “Changing the auto power- off (APO) timer” on page 73.
Hazardous voltage indication
The tester will display the hazardous voltage (
) symbol as an early
precaution when the measured voltage is equal to or greater than
±DC 30 V or AC 30 V, or when the measured voltage is over the
measurement range, OL (overload).
Power-on options
Some options can be selected only while you turn the tester on. These
power- on options are listed in the table below. To select a power- on
option, press and hold the specified key while turning the rotary switch
from the
position to any other position. Power- on options remain
selected until the tester is turned off.
Table 1-9 Power-on options
Key
Description
Simulates the Auto Power-Off (APO) mode. Press any key to turn the tester
back on and resume normal operation.
Checks firmware version. The tester’s firmware version will be shown on
the primary display. Press any key to exit this mode.
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
27
Toggles the red LED indicator alert for insulation resistance tests. If
enabled, the red LED indicator will blink every two seconds during an
insulation resistance test.
The red LED indicator alert is disabled when the Limit feature (see page 57)
is enabled.
‘
Change the resolution count to high. To permanently enable change the
resolution count. See “Changing the display count” on page 76.
Tests the LCD. All LCD annunciators are lighted. Press any key to exit this
mode.
Table 1-9 Power-on options (continued)
Key
Description
U1452A/U1452AT
U1451A
U1452A/U1452AT/U1451A User’s Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1452A/U1452AT/U1451A Insulation Tester
User’s Guide
2
Making Measurements
Insulation Resistance Test 30
Using the Remote Switch Probe 32
Locking the test 33
Timed insulation resistance/earth-bond resistance test 34
Measuring the Dielectric Absorption Ratio (DAR) 35
Measuring the Polarization Index (PI) 36
Viewing the leakage current 37
Earth-Bond Resistance Test 38
Measuring AC or DC Voltage 41
Auto AC or DC signal identification 43
Measuring Frequency 44
Measuring Resistance 46
Continuity Test 48
Measuring Capacitance 50
The following sections describe how to take measurements
with your tester.
Insulation Resistance Test
30
U1452A/U1452AT/U1451A User’s Guide
Insulation Resistance Test
Set up your tester as shown in Figure 2- 1. Set the rotary switch to a test
voltage value that does not exceed the maximum voltage limitation of the
circuit under test. Ensure that the device- under- test (DUT) is de- energized
before performing any resistance measurement.
Table 2-1 Rotary switch position for insulation resistance tests
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
50 V insulation
resistance test
AC+DC V
or
DC V (during test)
1 Timed (T) test
2 Dielectric
Absorption
Ratio (DAR)
test
3 Polarization
Index (PI) test
AC+DC V
or
DC V (during test)
100 V insulation
resistance test
250 V insulation
resistance test
500 V insulation
resistance test
1000 V insulation
resistance test
CAUTION
•
DO NOT perform insulation resistance test in distribution systems with voltages
higher than 600 V.
•
The tester automatically detects if the circuit is energized. If the external voltage is
detected to be greater than 30 V (or 50 V or 75 V; depending on selected option in
Setup), the test is inhibited. The symbol is shown on the display when either the
external voltage or the test voltage is greater than 30 V. Disconnect the tester and
remove the power of the circuit before proceeding.
Insulation Resistance Test
U1452A/U1452AT/U1451A User’s Guide
31
Figure 2-1
Insulation resistance test example
3
2
1
4
Remote
Switch Probe
Hold
Test will continue until
is released.
all-guides.co
m
Insulation Resistance Test
32
U1452A/U1452AT/U1451A User’s Guide
Using the Remote Switch Probe
The Remote Switch Probe (included in shipment) is used with insulation
resistance tests and earth- bond resistance tests, enabling the tester to be
controlled remotely from the button on the Remote Switch Probe.
By default the button on the Remote Switch Probe emulates the
button on the tester.
To change the default button operation, see “Changing the button
operation on the remote switch probe” on page 82
CAUTION
The insulation meter will auto-discharge the DUT when the test complete. However,
the DUT will not be auto-discharged when you disconnect the probe before the test is
complete. Avoid touching the DUT when the DUT is not fully discharged as it may lead
to possible electric shock.
NOTE
When an insulation test is in progress, the red LED indicator at the top of the tester will
blink every 2 seconds (if the Limit function is not enabled). To disable this feature see
“Power-on options” on page 26.
=
Insulation Resistance Test
U1452A/U1452AT/U1451A User’s Guide
33
Locking the test
You can lock the insulation resistance tests or earth- bond resistance tests
temporarily.
Press
to enable the lock once feature. The
symbol will be shown
on the display. The test will start when you press
, and it remain
active until
or
is pressed again.
By default, the tester will reset the locked status when the test is stopped
by pressing
or
. See “Changing the Dielectric Absorption
Ratio (DAR) for IR tests” on page 83 to disable this feature.
If you disable this feature, you will need to press
to unlock the
tester, even if the test has already stopped.
2
1
Press once
Press
or
again
to unlock and stop the test.
Lock symbol
Insulation Resistance Test
34
U1452A/U1452AT/U1451A User’s Guide
Timed insulation resistance/earth-bond resistance test
Use the timed test to obtain measurement results with consistent test
times — for later comparisons. Set up your tester as shown in Figure 2- 1,
and follow the steps shown below.
Figure 2-2
T/Time operation
Press
/
and
select the timed test.
After 30 seconds.
After 1 minute, time up.
Press
to view
the resistance value.
Press
to
view the timer.
Press
again to stop the test.
Press
to start the test.
Insulation Resistance Test
U1452A/U1452AT/U1451A User’s Guide
35
Measuring the Dielectric Absorption Ratio (DAR)
Dielectric Absorption Ratio (DAR) is the ratio of the insulation resistance
tested at 60 seconds to the insulation resistance tested at 30 seconds. Set
up your tester as shown in Figure 2- 1, and follow the steps shown below.
Figure 2-3
DAR operation
Press
and
select the DAR test.
Press
to
start the test.
After 30 seconds.
After 1 minute, time up.
Press
to view t0.
Press
to view t00.
DAR > 999.9
Error as IR (t0) > maximum range
Press
to
view the timer.
Press
to view
the resistance value.
Insulation Resistance Test
36
U1452A/U1452AT/U1451A User’s Guide
Measuring the Polarization Index (PI)
Polarization Index (PI) is the ratio of the insulation resistance tested at 10
minutes to the insulation resistance tested at 1 minute. Set up your tester
as shown in Figure 2- 1, and follow the steps shown below.
Figure 2-4
PI operation
Press
and
select the PI test.
Press
to
start the test.
After 1 minute.
After 10 minutes, time up.
Press
to view t0.
Press
to view t00.
PI > 999.9
Press
to
view the timer.
Press
to view
the resistance value.
Error as IR (t0) > maximum range
all-guides.co
m
Insulation Resistance Test
U1452A/U1452AT/U1451A User’s Guide
37
Viewing the leakage current
Press
to view the leakage current display. The leakage current
display is related to the insulation resistance. The higher the resistance
tested, the lower the current is to be measured.
NOTE
•
Because of the time required to perform the T, PI, and DAR tests, the use of alligator test
clips is recommended.
•
For timed (page 34) tests, The length of the timer is 1 minute by default. To change this
value, see “Changing the IR and EBR test period” on page 80 for more information.
•
For DAR (page 35) tests, you can change the DAR from 60:30 to 60:15 in the Setup. See
“Changing the Dielectric Absorption Ratio (DAR) for IR tests” on page 83 for more
information.
•
For DAR (page 35) and PI (page 36) tests, Err is shown on the display if the IR is greater
than the maximum range or less than 0.001 MΩ; if the test is interrupted by the user; or
if the tester’s battery is low.
Earth-Bond Resistance Test
38
U1452A/U1452AT/U1451A User’s Guide
Earth-Bond Resistance Test
Set up your tester to perform earth- bond resistance tests as shown in
Figure 2- 5.
Table 2-2 Earth-bond resistance test position
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
Earth-bond
resistance test
AC+DC V
or
DC V (during test)
Timed (T) test
AC+DC V
or
DC V (during test)
CAUTION
•
To avoid possible damage to your tester or to the equipment under test, disconnect
the circuit power and discharge all high-voltage capacitors before measuring
resistance.
•
The tester automatically detects if the circuit is energized. If the external voltage is
detected to be greater than 2 V, the test will not start. Disconnect the tester and
remove power before proceeding.
NOTE
•
The earth-bond resistance function is used to measure the resistance between earth
conductors, protective earth conductors, and conductors for equipotential bonding;
including their connections and terminals; with an indication of the measured value or
indication of limits.
•
The voltage source is <6.8 V, and the current is >200 mA when the resistance of ≤2 Ω is
to be measured. When the source voltage is <4.7 V, the tester will inhibit the test
automatically. The secondary display indicates the voltage (with auto-ranging enabled).
•
The APO (auto power-off) function is disabled during the test.
•
See also “Timed insulation resistance/earth-bond resistance test” on page 34.
Earth-Bond Resistance Test
U1452A/U1452AT/U1451A User’s Guide
39
Figure 2-5
Earth-bond resistance test example
3
2
1
4
>
Remote
Switch Probe
Hold
Test will continue until
is released.
OR
Earth-Bond Resistance Test
40
U1452A/U1452AT/U1451A User’s Guide
Using the earth-bond resistance test to verify the fuse condition
1 Keep the test leads open, and ensure that no voltage is applied to the
terminals.
2 Press and hold
to verify the fuse condition.
3 If the fuse has been blown, FUSE will be shown on the display. Follow
the instructions in the U1452A/U1452AT/U1451A Service Guide to
replace the fuse.
Fuse blown;
replace fuse.
Measuring AC or DC Voltage
U1452A/U1452AT/U1451A User’s Guide
41
Measuring AC or DC Voltage
Set up your tester to measure AC or DC voltage as shown in Figure 2- 6.
Table 2-3 AC and DC voltage measurement positions
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
Auto (V)
AC+DC V
Cycles between
1 DC V
2 AC V
3 Auto (V)
1 AC+DC V
2 AC+DC V
3 AC+DC V
NOTE
•
This tester displays DC voltage values as well as their polarity. Negative DC voltages will
return a negative sign on the left of the display.
•
Press
to measure the frequency of the voltage source. See “Measuring
Frequency” on page 44 to learn more.
all-guides.co
m
Measuring AC or DC Voltage
42
U1452A/U1452AT/U1451A User’s Guide
Figure 2-6
AC or DC voltage measurement example
AC
3
DC
2
1
Measuring AC or DC Voltage
U1452A/U1452AT/U1451A User’s Guide
43
Auto AC or DC signal identification
The Auto function is able to automatically identify the signal component
(AC or DC) of an electrical source and select a suitable measurement
range according to the AC+DC reading.
The Auto function identifies the signal component using the following
rules:
• It will consider which component value is greater between the AC or
DC.
• The AC value should be greater than a minimum value of 50 counts
(based on 6000 counts) of range to prevent residual value due to range
changing.
• The frequency measured is greater than 10 Hz for the AC mode.
While the signal is being identified, you can press
to lock the (AC or
DC) signal on the primary display.
At any time, you can press
to stop the Auto function and lock the
identified signal component (AC or DC).
The symbol
blinks during the identification.
The AC+DC value is shown
in the secondary display.
U1452A/U1452AT/U1451A User’s Guide
Measuring Frequency
Your tester allows simultaneous monitoring of real- time voltage with
frequency measurements. To measure frequency, rotate the switch to
measure voltage (see Figure 2- 6) and set up the tester accordingly.
Press
. Probe the test points, and read the display.
Frequency measurement techniques
• Measuring the frequency of a signal helps detect the presence of
harmonic currents in neutral conductors and determines whether these
neutral currents are the result of unbalanced phases or non- linear
loads.
• Frequency is the number of cycles a signal completes each second.
Frequency is defined as 1/Period. Period is defined as the time between
the middle threshold crossings of two consecutive, like- polarity edges,
as shown in Figure 2- 7.
• The tester measures the frequency of a voltage signal by counting the
number of times the signal crosses a threshold level within a specified
period of time.
• If a reading shows as 0 Hz or is unstable, the input signal may be
below or near the trigger level. You can usually correct these problems
by manually selecting a lower input range, which increases the
sensitivity of the tester.
WARNING
Never measure the frequency where the voltage level exceeds the specified range.
Manually set the voltage range if you want to measure frequencies below 20 Hz.
NOTE
•
Pressing
controls the input range of the voltage function and not the frequency
range.
•
To obtain the best measuring results for frequency measurements, please use the AC
measuring path.
U1452A/U1452AT/U1451A User’s Guide
45
• If a reading seems to be a multiple of what you expect, the input signal
may be distorted. Distortion can cause multiple triggerings of the
frequency counter. Selecting a higher voltage range might solve this
problem by decreasing the sensitivity of the tester. In general, the
lowest frequency displayed is the correct one.
• The frequency of the input signal is shown in the primary display. The
voltage value of the signal is shown in the secondary display. The bar
graph does not indicate frequency but indicates the voltage value of the
input signal.
Figure 2-7
Definition of frequency
Rise Time
Fall Time
+ Width
– Width
Period
90%
50%
10%
U1452A/U1452AT/U1451A User’s Guide
Measuring Resistance
Set up your tester to measure resistance as shown in Figure 2- 8.
Table 2-4 Resistance measurement position
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
Resistance
-
Continuity
-
CAUTION
To avoid possible damage to your tester or to the equipment under test, disconnect the
circuit power and discharge all high-voltage capacitors before measuring resistance.
NOTE
Resistance (opposition to the current flow) is measured by sending a small current out
through the test leads to the circuit under test. Because this current flows through all
possible paths between the leads, the resistance reading represents the total resistance of
all paths between the leads. Resistance is measured in ohms (Ω).
Keep the following in mind when measuring resistance.
•
The test leads can add 0.1 Ω to 0.2 Ω of error to resistance measurements. To test the
leads, touch the probe tips together and read the resistance of the leads. To remove lead
resistance from the measurement, hold the test lead tips together and press
. Now
the resistance at the probe tips will be subtracted from all future display readings.
•
Because the tester’s test current flows through all possible paths between the probe
tips, the measured value of a resistor in a circuit is often different from the resistor’s
rated value.
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
47
Figure 2-8
Resistance measurement example
3
2
1
U1452A/U1452AT/U1451A User’s Guide
Continuity Test
Set up your tester to perform continuity tests as shown in Figure 2- 9. The
beeper will sound and the red will light LED up as a continuity indication.
Table 2-5 Continuity test position
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
Resistance
-
Continuity
-
CAUTION
To avoid possible damage to your tester or to the equipment under test, disconnect the
circuit power and discharge all high-voltage capacitors before testing for continuity.
NOTE
•
Continuity is the presence of circuit continuities. The beeper sounds as the resistance
falls down to the threshold, and the red LED indicator will be lit (if enabled in the Setup).
The audible and visual alert allows you to perform quick continuity tests without having
to watch the display.
•
The continuity function detects intermittent shorts and opens. A brief short causes the
tester to emit a short beep.
•
You can enable or disable the audible alert via the Setup. See “Changing the continuity
alert” on page 74 for more information on the audible alert option.
U1452A/U1452AT/U1451A User’s Guide
49
Figure 2-9
Continuity test example
2
1
3
ON
(closed)
OFF
(open)
4
U1452A/U1452AT/U1451A User’s Guide
Measuring Capacitance
Set up your tester to measure capacitance as shown in Figure 2- 10. The
cable length of the circuit under test in shown in the secondary display.
• The default cable length scale is 1 km per 40 nF (km/C). To change this
value, see “Changing the cable length scale” on page 78.
• You can also change the cable length unit (Meter or Feet). To change
this value, see “Changing the cable length unit” on page 77.
Table 2-6 Capacitance measurement position
Legend
Default function
Function when
is pressed
Rotary
switch
position
Primary display
Secondary
display
Primary display
Secondary
display
Capacitance
Cable length
-
-
CAUTION
To avoid possible damage to the tester or to the equipment under test, disconnect
circuit power and discharge all high-voltage capacitors before measuring capacitance.
Use the DC voltage function to confirm that the capacitor is fully discharged.
NOTE
•
The tester measures capacitance by using an AC sine wave.
•
The resistance of the test leads will impact the accuracy of the measurement. It is
recommended to use short leads to measure capacitance.
•
The test frequency is 54.5 Hz.
U1452A/U1452AT/U1451A User’s Guide
51
Figure 2-10 Capacitance measurement example
1
3
+
–
2
all-guides.co
m
U1452A/U1452AT/U1451A User’s Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1452A/U1452AT/U1451A Insulation Tester
User’s Guide
3
Tester Features
Making Relative Measurements (Null) 54
Freezing the Display (TrigHold and AutoHold) 55
Performing Limit Comparisons (Limit) 57
Recording Measurement Data (Log) 59
Performing manual logs (HAND) 60
Performing interval logs (AUTO) 60
Performing event logs (TRIG) 61
Reviewing Previously Recorded Data (View) 63
The following sections describe the additional features
available in your tester.
Making Relative Measurements (Null)
54
U1452A/U1452AT/U1451A User’s Guide
Making Relative Measurements (Null)
When making null measurements, also called
relative, each reading is the difference
between a stored (selected or measured) null
value and the input signal.
One possible application is to increase the
accuracy of a resistance measurement by
nulling the test lead resistance.
1 To activate Null, press
. The measurement value at the time that
when Null (
) is enabled, is stored as the reference value.
2 Press
again to view the stored reference value (the
annunciator
blinks). The display will return to normal after 3 seconds.
3 To disable Null, press
while the stored reference value is shown
(step 2).
For any measurement function, you can directly measure and store the
null value by pressing
with the test leads open (nulls the test lead
capacitance), shorted (nulls the test lead resistance), or across a desired
null value circuit.
NOTE
Null can be set for both auto and manual range settings, but not in the case of an overload.
NOTE
•
In resistance measurement, the tester will read a non-zero value even when the two test
leads are in direct contact, because of the resistance of these leads. Press
to
zero-adjust the display.
•
For DC voltage measurements, the thermal effect will influence the accuracy of the
measurements. Short the test leads and press
when the displayed value is stable to
zero-adjust the display.
Freezing the Display (TrigHold and AutoHold)
U1452A/U1452AT/U1451A User’s Guide
55
Freezing the Display (TrigHold and AutoHold)
TrigHold operation
Press
to freeze the display for any
function, except for the Ω Mega or the Ω EB
function.
Press
again to trigger a new
reading. The
annunciator blinks while
waiting for the reading to be stable.
Press and hold
to exit this mode.
AutoHold operation
Press and hold
to activate AutoHold
for any function, except for the Ω Mega or the
Ω EB function.
AutoHold monitors the input signal and
updates the display and, if enabled, emits a
beep, whenever a new stable measurement is
detected. The
annunciator blinks while waiting for the reading to be
stable.
Press and hold
to exit this mode.
NOTE
Pressing
when the rotary switch is in one of the Ω Mega or the Ω EB position will result in a
test being performed instead.
NOTE
Pressing and holding
when the rotary switch is in one of the Ω Mega or the Ω EB position
will result in a test being performed instead.
Freezing the Display (TrigHold and AutoHold)
56
U1452A/U1452AT/U1451A User’s Guide
A trigger point is one that varies more than a selected adjustable
(AutoHold threshold) variation count (default 500 counts). The following
conditions are not included in the update.
To change the default AutoHold threshold count see “Changing the
variation count” on page 71 for more information.
Function
None updated counts
Voltage
50
Resistance
OL or Open
Capacitance
50
NOTE
If the reading value is unable to reach a stable state, the reading value will not be updated.
all-guides.co
m
Performing Limit Comparisons (Limit)
U1452A/U1452AT/U1451A User’s Guide
57
Performing Limit Comparisons (Limit)
Limit is used to compare the test result with
the chosen settling value. The default settling
values are shown in the table below.
You can set pass condition to be > or < the
value measured.
1 Press
to activate Limit.
2 Press
again to set the comparison value. Use the arrow keys to
position the cursor and to change the value shown.
The most significant digit can be set to a value from 1 to 9. The
remaining two digits can be set to 0, 00, or left blank along with any
unit combination (see Table 3- 1).
3 Position the cursor on the > annunciator to change the Limit pass
condition (< or >).
4 Press
to save your changes (or press
to discard your
changes).
Table 3-1 Limit settling default values
Function
Default value
Range
Limit
settling range
Voltage measurement[1]
[1] Absolute value comparison, regardless of polarity.
>30 V
Auto or Manual
1 V < LS < 900 V
Resistance measurement
<10 Ω
Auto or Manual
1 Ω < LS < 90 MΩ
Capacitance measurement[2]
[2] Limit for capacitance is represented without a unit.
>10
Range locked
10 < LS < 900
Earth-bond resistance test
<10 Ω
Auto
1 Ω < LS < 90 kΩ
Insulation resistance test
>10 MΩ
Auto
1 kΩ < LS < 90 GΩ
Performing Limit Comparisons (Limit)
58
U1452A/U1452AT/U1451A User’s Guide
5 If the new value is passed:
• PAS is shown
• A short beep tone is heard
• The red LED blinks once
6 If the new value is failed:
• FAi is shown
• Three short beep tones are heard
• The red LED blinks thrice
NOTE
When the Limit feature is enabled for insulation resistance tests, the red LED indicator
lights up accordingly to the changes in the limit values instead of blinking every 2 seconds.
Recording Measurement Data (Log)
U1452A/U1452AT/U1451A User’s Guide
59
Recording Measurement Data (Log)
Log provides you with the convenience of recording test data for future
review or analysis. Since data is stored in the nonvolatile memory, the
data remains saved even when the tester is turned OFF or if the battery
is replaced.
Log collects measurement information over a user- specified duration.
There are three Log options that can be used to capture measurement
data: manual (HAND), interval (AUTO), or event (TRIG).
• A manual log stores an instance of the measured signal each time you
press and hold
(see page 60).
• An interval log stores a record of the measured signal at a
user- specified interval (see page 60).
• An event log stores a record of the measured signal each time a trigger
condition is satisfied (see page 61).
Before starting a recording session, set up the tester for the measurements
to be recorded.
To change the Log option see “Changing the recording option” on page 80
for more information.
See “Reviewing Previously Recorded Data (View)” on page 63 to review or
erase the recorded entries.
Table 3-2 Log maximum capacity
Log option
Maximum capacity for saving
Manual (HAND)
H00 to H99 (100 entries)
Interval (AUTO)
A00 to A99 (100 entries)
Event (TRIG)
E00 to E99 (100 entries)
NOTE
Each recorded index includes two parameters: the primary display and the secondary
display. Examples include IR-V or V-Hz.
Recording Measurement Data (Log)
60
U1452A/U1452AT/U1451A User’s Guide
Performing manual logs (HAND)
Ensure that HAnd is selected as the Log
option in the Setup.
1 Press and hold
to store the present
input signal value.
and the log entry number are
displayed. The display will return to
normal after a short while (around
1 second).
2 Repeat step 1 again to save the next input signal value.
The maximum number of readings that can be stored for the manual log is
100 entries. When all entries are occupied, HFU will be shown when you
press and hold
.
Performing interval logs (AUTO)
Ensure that AUto is selected as the Log
option in the Setup.
The default recording interval duration is
1 second. To change the recording interval
duration, see “Changing the sample interval
duration” on page 79 for more information.
The duration set in the Setup will determine
how long each recording interval takes. The input signal value at the end
of each interval will be recorded and saved into the tester’s memory.
Start the interval log mode
1 Press and hold
to start interval log mode.
and the log entry number are displayed. Subsequent readings are
automatically recorded into the tester’s memory at the interval specified
in the Setup.
2 Press and hold
again to exit the interval log mode.
Recording Measurement Data (Log)
U1452A/U1452AT/U1451A User’s Guide
61
The maximum number of readings that can be stored for the interval log
is 100 entries. When all entries are occupied, AFU will be shown when you
press and hold
.
Performing event logs (TRIG)
Ensure that triG is selected as the Log option
in the Setup.
Event logs are used only with the following
modes:
• TrigHold and AutoHold (page 55)
• Earth- bond resistance tests (page 38)
• Insulation resistance tests (page 38)
• T/DAR/PI tests (page 30)
Event records are triggered by the measured signal satisfying a trigger
condition set by the measurement function used in the following modes:
NOTE
When the interval log recording session is running, all other keypad operations are
disabled; the exception is
, which, when pressed for more than 1 second, will stop and
exit the recording session. Furthermore, APO (auto power-off) is disabled during the
recording session.
Table 3-3 Event log trigger conditions
Modes
Trigger condition
Primary display recorded
Secondary display
recorded
The input signal value is recorded:
TrigHold
Each time you press
and the reading
update is stable.
Voltage, resistance, or
capacitance
Voltage, frequency,
capacitance cable length,
or output source voltage
AutoHold
When the input signal varies more than the
variation count and the reading update is stable.
Voltage, resistance, or
capacitance
Voltage, frequency,
capacitance cable length,
or output source voltage
all-guides.co
m
Recording Measurement Data (Log)
62
U1452A/U1452AT/U1451A User’s Guide
Start the event log mode
1 Select one of the modes listed in Table 3- 3.
2 Press and hold
to start event log mode.
and the log entry number are displayed. The primary display and
secondary display readings will be recorded into the memory.
Subsequent readings are automatically recorded into the tester’s
memory every time the trigger condition specified in Table 3- 3 is
satisfied.
3 Press and hold
again to exit the event log mode.
The maximum number of readings that can be stored for the event log is
100 entries. When all entries are occupied, EFU will be shown when you
press and hold
.
Earth-bond
resistance test
Each time you press
to stop the test
output source.
Resistance or
leak current value
Test output source voltage
Insulation
resistance test
T/Time
When the time is up (Timer = 00:00), the final
value is recorded before the test output source
is stopped.
Resistance or
leak current value
Test output source voltage
NOTE
The values of DAR t30 (or DAR t15), DAR t60, PI t1, and PI t10 will be recorded in every IR
rotary switch location. For more information on DAR and PI tests, see page 35 and page 36
respectively.
Table 3-3 Event log trigger conditions (continued)
Modes
Trigger condition
Primary display recorded
Secondary display
recorded
The input signal value is recorded:
NOTE
APO (auto power-off) is disabled during the recording session.
Reviewing Previously Recorded Data (View)
U1452A/U1452AT/U1451A User’s Guide
63
Reviewing Previously Recorded Data (View)
Viewing data stored in the tester’s memory is
performed through the
key.
1 Press and hold
to View the
previously recorded data. Press
again
to cycle through the manual (H), interval
(A), or event (E) records.
If nothing has been recorded, H--, A--, or
E-- will be displayed instead.
2 Select the desired recording category to view its entries.
i
Press
to jump to the first stored entry. Press
to jump to
the last stored entry.
ii Press
to view the next stored entry. The index number increases
by one. Press
to view the previous stored entry. The index
number decreases by one.
iii Press
to clear the last stored entry for the selected log type.
Press and hold
to clear all entries for the selected log type.
3 Press and hold
again to exit the View mode.
Sanitizing the Log Memories
You have the option to sanitize the log
memories of your tester. This operation
erases the log memories of your tester
thoroughly. The data stored in the tester’s
memory will not be able to be reconstructed
in any way after the data sanitization
operation.
Prior to sanitizing the log memories, ensure
that all manual (H), interval (A), or event (E) records have been cleared
(see step iii). When all entries are cleared (H--, A--, and E--), press and
hold
to sanitize the log memories.
CAUTION
The data sanitization operation may take up to 30 seconds to complete. Do not press
any keys or turn the rotary switch until the data sanitization operation is completed.
Reviewing Previously Recorded Data (View)
64
U1452A/U1452AT/U1451A User’s Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
U1452A/U1452AT/U1451A Insulation Tester
User’s Guide
4
Setup Options
Using the Setup Menu 66
Editing numerical values 67
Setup Menu Summary 68
Setup Menu Items 71
Editing numerical values 67
Changing the variation count 71
Enabling smooth mode 71
Changing the beep frequency 72
Changing the auto power-off (APO) timer 73
Changing the LCD backlight timeout 73
Changing the continuity alert 74
Disabling the power-on melody 75
Changing the battery type 75
Resetting the tester’s Setup options 76
Changing the display count 76
Changing the cable length unit 77
Changing the cable length scale 78
Changing the alert indicators 78
Changing the sample interval duration 79
Changing the recording option 80
Changing the IR and EBR test period 80
Changing the maximum inhibit voltage for IR tests 81
Disabling the lock once feature 82
Changing the button operation on the remote switch probe 82
Changing the Dielectric Absorption Ratio (DAR) for IR tests 83
The following sections describe how to change the preset
features of your tester.
Using the Setup Menu
66
U1452A/U1452AT/U1451A User’s Guide
Using the Setup Menu
The Setup menu allows you to change a number of nonvolatile preset
features. Modifying these settings affects the general operation of your
tester across several functions. Select a setting to edit in order to perform
one of the following actions:
• Switch between two values, such as on or off.
• Cycle through multiple values from a predefined list.
• Decrease or increase a numerical value within a fixed range.
The contents of the Setup menu are summarized in Table 4- 2 on page 68.
Table 4-1 Setup menu key functions
Legend
Description
• Press
for more than 1 second to access the Setup menu.
• Press and hold
until the tester restarts to exit the Setup menu.
• Press
or
to browse each menu page.
• Press
or
at each menu page to move the cursor to a
specific menu item.
l‘
• Press
to edit the selected menu item. The menu item’s value
will flash to indicate that you can now change the value shown.
• Press
or
again to switch between two values, to cycle
through multiple values from a list, or to decrease or increase a
numerical value.
• Press
to save your changes.
• While the menu item’s value is flashing, press
to discard
your changes.
U1451A
U1452A/U1452AT
NOTE
The tester will automatically exit the Setup menu after 30 seconds of inactivity.
all-guides.co
m
Using the Setup Menu
U1452A/U1452AT/U1451A User’s Guide
67
Editing numerical values
When editing numerical values, use the
and
to position the
cursor on a numerical digit.
• Press
to move the cursor to the left, and
• Press
to move the cursor to the right.
When the cursor is positioned over a digit, use the
and
keys to
change the numerical digit.
• Press
to increment the digit, and
• Press
to decrement the digit.
When you have completed your changes, save the new numerical value by
pressing
. (Or alternatively, if you wish to discard the changes you
made, press
.)
Press and hold
for more than
1 second to enter the Setup menu.
Press
to edit the
selected menu item’s value.
Press
or
to
move the cursor left or right.
Press
or
to increment
or decrement the selected digit.
Press
again to
save the changes made.
Setup Menu Summary
68
U1452A/U1452AT/U1451A User’s Guide
Setup Menu Summary
The Setup menu items are summarized in the table below. Click the
respective “Learn more” pages for more information on each menu item.
Table 4-2 Setup menu item descriptions
Legend
Available settings
Description
Learn more on:
AH
0001 to 9999 counts
Set the AutoHold and Smooth threshold count from 1 to
9999 counts.
Default is 0050 counts.
page 71
SMo
0001 to 9999 (d or E)
Set the settling value from 0001 to 9999. You can also
disable this feature (d).
Default is disabled (0009d).
page 71
bEE
3200 Hz to 4267 Hz,
or oFF
Set the beep frequency from 3200 Hz to
4267 Hz. You can also disable this feature (off).
Default is 3840 Hz.
page 72
APo
01 to 99 minutes (d or E)
Set the auto power-off timer period from 1 to 99 minutes
(1 hour, 39 minutes). You can also disable this feature (d).
Default is 15 minutes (15E).
page 73
bLt
01 to 99 seconds (d or E)
Set the backlight timer period from 1 to 99 seconds
(1 minute, 39 seconds). You can also disable this feature
(d).
Default is 15 seconds (15E).
page 73
tyP
SHor or tonE
Set the tester to sound a single beep (SHor) or a tone
(tone) during continuity alerts for short circuits.
Default is a single beep for short circuits (SHor).
page 74
Snd
MELo, USEr, bEEE, or
oFF
Set the power-on melody to the factory default (MELO),
a beep (bEEE), or disable this feature (off).
Default is a beep (bEEE).
page 75
bAt
PRi or SEC
Change the battery selection from primary (PRi) to
secondary (SEC).
Default is primary (PRi).
page 75
rSt
yES or no
Reset the tester to its factory default settings.
page 76
Setup Menu Summary
U1452A/U1452AT/U1451A User’s Guide
69
Cnt
Hi or Lo
Set the display count to High (Hi) or Low (Lo).
Default is Low (Lo).
page 76
CAb mft
Mete (m) or Foot (ft)
Set the cable length unit to Meter (Mete) or Foot (Foot)
for capacitance measurements.
Default is Mete (m).
page 77
CAb
01 to 99 nF/km
Set the capacitance measurement versus cable length
scale from 1 to 99 nF/km.
Default is 40 nF/km.
page 78
ALE
bE.rL, --.rL,
bE.--, or --.--
Set the tester to beep momentarily and light up the red
LED indicator for limit and continuity alerts. You can also
disable either or both alerts (off).
Default is to beep momentarily and light up the red LED
indicator (bE.rL).
page 78
tiM
0001 to 9999 seconds
Set the logging duration for interval logs from 1 to 9999
seconds (2 hours, 46 minutes, 39 seconds).
Default is 1 second.
page 79
tyP
HAnd, AUto, or triG
Set the data logging option (HAnd: manual log,
AUto: interval log, or triG: event log).
Default is manual log (HAnd).
page 80
tiM
00.05 to 59.59 mm:ss
Set the insulation resistance or earth-bond resistance
test period.
Default is 10 seconds (00.10 mm:ss)
page 80
inH
30 V, 50 V, or 75 V
Set the maximum inhibit voltage for insulation
resistance test.
Default is 30 V.
page 81
onE
yES or no
Enable or disable the lock once feature.
Default is enabled (yES).
page 82
Table 4-2 Setup menu item descriptions (continued)
Legend
Available settings
Description
Learn more on:
Setup Menu Summary
70
U1452A/U1452AT/U1451A User’s Guide
rEM
bt-1 to bt-7 (d or E)
Change or disable the button operation on the remote
probe.
Default is
(bt-7E).
page 82
tiM
60.30 or 60.15 seconds
Set the Dielectric Absorption Ratio in seconds (60:30 or
60:15).
Default is 60:30 (seconds).
page 83
Table 4-2 Setup menu item descriptions (continued)
Legend
Available settings
Description
Learn more on:
Setup Menu Items
U1452A/U1452AT/U1451A User’s Guide
71
Setup Menu Items
Changing the variation count
This setting is used with the AutoHold feature (see page 55). When the
variation of the measured value exceeds the value of the variation count,
the AutoHold feature will be ready to trigger.
To change the variation count:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to AH (
and
are shown on the display), and press
to edit the value.
3 Use the arrow keys to change the variation count.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Enabling smooth mode
Smooth is used to smoothen the refresh rate of the readings in order to
reduce the impact of unexpected noise and to help you achieve a stable
reading.
The smooth refresh rate can be set from 0001 to 9999. The smooth time is
defined as the set value +1. Smooth will be restarted when the variation
count is exceeded, when the range is changed, or after a tester function or
feature is enabled. The variation count is set to the value used for the
AutoHold feature (see “Changing the variation count” on page 71).
Legend
Range
Default setting
AH
(0001 to 9999) counts
0050
all-guides.co
m
Setup Menu Items
72
U1452A/U1452AT/U1451A User’s Guide
To change the smooth refresh rate:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to SMo (
is shown on the display), and press
to edit
the value.
3 Use the arrow keys to change the smooth refresh rate. Select E to
enable the Smooth feature.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the beep frequency
The beeper alerts users to the presence of circuit continuities and newly
sensed values for Max Min recordings.
To change the beep frequency:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to bEE, and press
to edit the value.
3 Use the arrow keys to change the beep frequency. Select oFF to disable
the beeper.
4 Press
to save your changes (or press
to discard your
changes).
Legend
Range
Default setting
SMo
• 0001 to 9999
• d(isabled) or E(nabled)
0009d
Legend
Range
Default setting
bEE
(4267, 4151, 4042, 3938, 3840, 3746, 3675, 3572, 3491,
3413, 3339, 3268, 3200) Hz, or oFF
3840
Setup Menu Items
U1452A/U1452AT/U1451A User’s Guide
73
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the auto power-off (APO) timer
The APO (see page 6) feature uses a timer to determine when to
automatically turn the tester off.
To change the APO timer period:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to APo, and press
to edit the value.
3 Use the arrow keys to change the APO timer period. Select d to disable
the APO feature.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the LCD backlight timeout
The tester’s LCD backlight uses a timer to determine when to turn off the
LCD backlight.
Legend
Range
Default setting
APo
• (01 to 99) minutes
• E(nabled) or d(isabled)
15E
Legend
Range
Default setting
bLT
• (01 to 99) seconds
• E(nabled) or d(isabled)
15E
Setup Menu Items
74
U1452A/U1452AT/U1451A User’s Guide
To change the LCD backlight timer period:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to bLT, and press
to edit the value.
3 Use the arrow keys to change the backlight timer period. Select d to
disable the backlight timer.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the continuity alert
This setting is used with continuity tests (see page 48). The tester will
beep to alert users to the presence of circuit continuities for short
circuits. If tone is selected, the tester will sound different tones according
to the bar graph display.
To change the continuity alert:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to tyP (
is shown on the display), and press
to edit the
value.
3 Use the arrow keys to change the continuity alert.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Legend
Range
Default setting
tyP
SHor(t) or tonE
SHor
Setup Menu Items
U1452A/U1452AT/U1451A User’s Guide
75
Disabling the power-on melody
The tester plays a melody or a beep when it is powered on.
To disable the power- on melody:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to Snd, and press
to edit the value.
3 Use the arrow keys to change the power- on melody. Select oFF to
disable the power- on melody.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the battery type
If you are using rechargeable batteries to power your tester, change the
battery type from PRi to SEC for the tester to accurately reflect the battery
capacity indication.
Legend
Range
Default setting
Snd
MELo(dy), USEr, bEEE, or oFF
bEEE
NOTE
The USEr option is for Keysight internal use.
Legend
Range
Default setting
bAt
PRi(mary) or SEC(ondary)
PRi
Setup Menu Items
76
U1452A/U1452AT/U1451A User’s Guide
To change the battery type:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to bAt, and press
to edit the value.
3 Use the arrow keys to change the battery type.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Resetting the tester’s Setup options
The tester’s Setup options can be reset to its default values through the
Setup menu.
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to rSt, and press
to edit the value.
3 Use the arrow keys to select yES.
4 Press and hold
for more than 1 second to perform the reset. The
tester will beep once and return to the first Setup menu page. Or,
alternatively press
to discard your changes.
Changing the display count
Use these settings to change the display count for the following
measurements/tests. The low resolution count is a tenth of the original
display count. As an example, if the original display count is 6000 counts,
then the low resolution display count is 600 counts.
Legend
Range
Default setting
rSt
yES or no
no
all-guides.co
m
Setup Menu Items
U1452A/U1452AT/U1451A User’s Guide
77
To change the display count:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to Cnt, and press
to edit the value.
3 Use the arrow keys to change the display count.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the cable length unit
This setting is used with capacitance measurements (see page 50). Change
the unit (Meter or Feet) of the cable length display.
To change the cable length unit:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to CAb mft, and press
to edit the value.
3 Use the arrow keys to change the cable length unit.
Legend
Range
Default setting
Cnt
Hi(gh) or Lo(w)
Lo
NOTE
•
Voltage and capacitance measurements are fixed at 6000 counts.
•
DAR and PI tests are fixed at 9999 counts.
Legend
Range
Default setting
CAb mft
Mete(r) or Foot
Mete
Setup Menu Items
78
U1452A/U1452AT/U1451A User’s Guide
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the cable length scale
This setting is used with capacitance measurements (see page 50). Change
the scale from 1 nF to 99 nF per kilometer for capacitance transfers to
cable length.
To change the cable length scale:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to CAb, and press
to edit the value.
3 Use the arrow keys to change the cable length scale.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the alert indicators
The tester’s audible and visual alerts users to the presence of circuit
continuities (see page 48) and values exceeding the Limit values set (see
page 48).
Legend
Range
Default setting
CAb
(01 to 99) nF
40
Legend
Range
Default setting
ALE
bE.rL, --.rL, bE.--, or --.--
bE.rL
Setup Menu Items
U1452A/U1452AT/U1451A User’s Guide
79
To change the alert indicators:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to ALE (
and
is shown on the display), and press
to edit the value.
3 Use the arrow keys to change the alert indicators. Select -- to disable
either the beeper, the red LED, or both the alert indicators.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the sample interval duration
This setting is used with the Interval Data Logging feature (see page 60).
The tester will record a measurement value at the beginning of every
sample interval.
To change the sample interval duration:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to tiM (
is shown on the display), and press
to edit
the value.
3 Use the arrow keys to change the sample interval duration.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Legend
Range
Default setting
tiM
(0001 to 9999) seconds
0001
Setup Menu Items
80
U1452A/U1452AT/U1451A User’s Guide
Changing the recording option
This setting is used with the Data Logging feature (see page 59). There are
three available recording options for the Data Logging feature.
• HAnd: Manual log
• AUto: Interval log
• triG: Event log
To change the recording option:
1 Press
for more than 1 second to enter the Setup menu.
2 Browse to tyP (
is shown on the display), and press
to edit
the value.
3 Use the arrow keys to change the recording option.
4 Press
to save your changes (or press
to discard your
changes).
5 Press and hold
until the tester restarts to return to normal
operation.
Changing the IR and EBR test period
This setting is used with insulation resistance tests (see page 38) or
earth- bond resistance measurements (see page 38). The tester performs
the test over the timed period defined in this setting.
Legend
Range
Default setting
tyP
HAnd, auTO, or triG
HAnd
Legend
Range
Default setting
tiM
(00.05 to 59.59) mm:ss
00.10