O P E R A T I N G M A N U A L
linc84en1-04 (1610)
Catalog No.
155 80
155 81
Adjustable Helium Test Leak TL3-5 / TL4-6
Operating and Handling
1-4
3.1
First Use or Changing the Type of Gas
1-6
3.2
Changing the Reservoir Cylinder without Changing the Type of Gas 1-10
3.3
Storage when unused
1-13
3.4
Maintenance
1-13
4
Applications
1-14
4.1
Use for Vacuum Leak Detection
1-14
4.2
Use for Overpressure Leak Tests (Sniffer Probe)
1-14
5
Using Search Gases other than Helium
1-17
6
Adjustment of leak rate
1-19
7
Spare Parts
1-24
If an appliance is returned to INFICON, indicate whether the appliance is free of
substances damaging to health or whether it is contaminated. If it is contaminated also
indicate the nature of hazard.
For this please use the form provided in the Annex. INFICON must return any appliance
without a declaration of contamination to the sender’s address.
The references to diagrams, e.g. (2/5), consist of the Fig. No. and the Item No. in that
order.
General Note
The right of alterations in the design and the technical data is reserved. The illustrations
are not binding.
Caution
Indicates procedures that must be strictly observed to prevent damage to, or
destruction of, the test leak.
1) with manometer reading 0 bar and p<1 mbar at connecting flange.
Test Leak
Leak rate adjustment range TL3-5
10-3 ... 10-5 mbar l/s
Leak rate adjustment range TL4-6
10-4 ... 10-6 mbar l/s
Nominal leak rate TL3-5
5 · 10-5 mbar l/s
Nominal leak rate TL4-6
5 · 10-6 mbar l/s
Uncertainty of the nominal leak rate1)
± 10%
Temperature coefficient
Negligible
Leak type
Capillary
Calibrated for
Helium
Connecting flange
DN 16 ISO-KF
Manometer reading
-1 ... +15 bar (overpressure)
Helium cylinder
Test gas
Helium 5.0
Purity
99.999 Vol.-%
Bottle capacity
1 l
Filling pressure
12 bar
Filling amount
12 l
Inlet pressure max.
max. 12 bar
Weight
With gas cylinder:
425 g
Without gas cylinder:
300 g
Fig. 1-1 Dimensional drawing (all measures in mm)
1 test leak complete with manometer, gas-reservoir shut-off valve and leak shut-off
valve
•
1 helium bottle (HE 5.0; 12 bar; 1 l)
•
1 holder for setting up the helium bottle with test leak
•
Installation instructions for the holder
Operating and Handling
Chapter 1-3.fm
Operating Manual
(1610)
3
Operating and Handling
Always unscrew valve for a short time only. The leak outlet area on valve should not be
touched, particularly not with your fingers or greasy objects.
Retain the protection cap for the flange and fit it each time the test leak is removed.
Holder
The holder included is used for safe storage of the test leak. It is installed according to
the mounting instructions included, see Fig. 1-2.
Caution
The reservoir cylinder is pressurised.
Protect it against direct sun light, temperatures over 50 °C and damage.
Operating and Handling
1-5
Chapter 1-3.fm
Operating Manual
(1610)
Fig. 1-2 Test leak TL 3-5 / 4-6, complete
Pos.
Description
Pos.
Description
1
overpressure manometer for test gas pressure
5
connecting flange DN 16 ISO-KF
2
test gas supply shut-off valve
6
shut-off valve for leak outlet orifice
3
vent screw
7
test gas bottle
4
body with leak apillary
8
holder
Operating and Handling
Chapter 1-3.fm
Operating Manual
(1610)
3.1
First Use or Changing the Type of Gas
1
Close all valves and the vent screw.
2
Unscrew old test gas bottle.
3
Remove protective cap from new test gas bottle.
4
Screw in the new reservoir (hand tight) until to the stop.
Fig. 1-3
Operating and Handling
1-7
Chapter 1-3.fm
Operating Manual
(1610)
5
Open the test gas supply on the reservoir (the knob turns downwards!)
The pressure gauge must now indicate the pressure of the gas bottle.
6
Close the blocking valve on the reservoir (the knob turns upwards!).
Fig. 1-4
Fig. 1-5
Operating and Handling
Chapter 1-3.fm
Operating Manual
(1610)
7
Open the vent screw so far that the pressure drops to 0 bar (purge).
8
Close the vent screw.
Fig. 1-6
Fig. 1-7
Operating and Handling
1-9
Chapter 1-3.fm
Operating Manual
(1610)
9
Open the test gas supply, so that the pressure reaches the pressure of the test gas
cylinder.
10
Close test gas supply.
Fig. 1-8
Fig. 1-9
Operating and Handling
Chapter 1-3.fm
Operating Manual
(1610)
Repeat process steps 6. to 10. twice so as to ensure that the gas reservoir in front of the
test leak has been exchanged completely.
The test leak will now be ready. The desired pressure can be adjusted with the venting
screw.
3.2
Changing the Reservoir Cylinder without Changing
the Type of Gas
1
Close all valve. (the knob turns upwards!)
2
Unscrew the old test gas bottle.
Fig. 1-10
Operating and Handling
1-11
Chapter 1-3.fm
Operating Manual
(1610)
3
After removing the protective cap screw on the new test gas bottle.
4
Open the test gas supply.
The manometer displays now the test gas bottle pressure.
Fig. 1-11
Fig. 1-12
Operating and Handling
Chapter 1-3.fm
Operating Manual
(1610)
5
Close the test gas supply.
6
Open the vent screw so far, that the pressure drops to the desired pressure.
Fig. 1-13
Fig. 1-14
Operating and Handling
1-13
Chapter 1-3.fm
Operating Manual
(1610)
7
Close the vent screw.
The Helium test leak is now ready for operation.
3.3
Storage when unused
Put the protective cap on the connection flange DN 16 KF and make sure that all valves
and the vent screw are closed. Store the Helium test leak in a lying position or standing
upright in the holder, protected against impact shock and high temperatures
3.4
Maintenance
It is recommended to retighten the union nut of the valve (Fig. 1-2/2) from time to time at
a torque of 2.5 Nm by turning it in clockwise direction (how often this is required depends
on how much the test leak is used). Thereafter it must be easily possible to turn the knob
with two fingers.
Fig. 1-15
Use for Vacuum Leak Detection
The test leak TL3-5 / TL4-6 can be fitted by its DN 16 KF flange connection to any leak
detector or any vacuum apparatus and can then be used for tuning the mass
spectrometer as well as for checking response time and detection sensitivity.
To adjust the test leak rate see chapter 6, Fig. 1-20 and Fig. 1-21.
4.2
Use for Overpressure Leak Tests (Sniffer Probe)
The Helium test leak TL3-5 / TL4-6 is set up in holder (see Fig. 1-1/2). By unscrewing the
valve at the swivel nut the Helium test leak TL3-5 / TL4-6 becomes accessible for a sniffer
probe tip.
Fig. 1-16
The leak outlet orifice itself is on the face of the sealed-in tubulation which is now visible.
Fig. 1-17
Caution
Clogging of the leak capillary.
The size of the leak outlet orifice can change or get choked due to contamination
Caution
Do not touch the leak outlet orifice with the fingers or with greasy objects.
The measured leak rate depends on the distance between sniffer tip and leak, and also
on the speed at which the sniffer tip is moved past the leak.
The size of the leak depends on the pressure as indicated on the pressure gauge and can
be determined from the leak rate diagrams.
Caution
When no longer using the test leak anymore always close the valve.
Using Search Gases other than Helium
1-17
Chapter 1-3.fm
Operating Manual
(1610)
5
Using Search Gases other than Helium
When using test gases other than helium, leak rates can only approximately be
determined, the reason being that the 10-8 to 10-4 mbar·I/s range includes just the
transition-flow region between molecular and laminar gas flow. Moreover, the individual
flow profile of the respective capillary affects the conversion factor. In the molecular flow
region (leak rates < 1 · 10-5 mbar l/s) the leak rate q of different gases at otherwise equal
conditions depends on the gas type as:
M= relative molecular mass
In the laminar flow region the leak rate depends on the gas type as:
= dynamic viscosity
qx
qHe
--------
MHe
Mx
-----------
=
qx
qHe
--------
He
x
---------
=
Using Search Gases other than Helium
Chapter 1-3.fm
Operating Manual
(1610)
Some values of dynamic viscosity in 10-5 Pa•s.
The resulting values must be used according to the predominant type of gas flow, i.e. for
leak rates lower than 10-5 mbar·I/s mainly molecular flow and for leak rates higher than
10-5 mbar·I/s mainly laminar flow can be assumed.
He
R 134a
Air
Ar
H2
1,96
1,36
1,82
2,21
0,88
4
102
28
40
2
Adjustment of leak rate
1-19
Chapter 1-3.fm
Operating Manual
(1610)
6
Adjustment of leak rate
The nominal leak rate was measured during production at a test gas pressure of 1 bar
(abs.) vs. a pressure of lower than 1 mbar at the connecting flange and is shown on the
label. The adjustment diagrams are valid for nominal leak rate of 5 · 10-5 mbar l/s ± 10%
for the test leak TL 3-5 and a nominal leak rate of 5 · 10-6 mbar l/s ± 10% for the test leak
TL 4-6. The leak rate is pressure dependent and can be found in the diagrams Leak Rate/
Pressure (see Fig. 1-18 to Fig. 1-21). Upon request, individual calibration of a single test
leak as a function of the search-gas pressure can be made in our factory at cost price.
If you suspect the nominal leak rate has changed after longer use, also recalibration of
the test leak can be made in our factory against charge. When demanding higher
accuracy we recommend a recalibration once a year.
If a calibration certificate is required, it can be issued by our Cologne factory. Calibration
and calibration certificates are available to order against charge.
The test leak rates depend on the adjusted pressure [bar rel], see the following diagrams.
If a significant deviation from the nominal leak rate of 5 · 10-5 resp. 5 · 10-6 mbar l/s occurs,
a factor can be calculated based on the leak rate diagrams. The nominal leak rate is to be
multiplied by this factor.
Calculation of the factor:
Adjustment of leak rate
Chapter 1-3.fm
Operating Manual
(1610)
Diagram Leak Rate / Pressure
Fig. 1-18 Diagram Test Leak TL3-5
�����������������������������������
Adjustment of leak rate
1-21
Chapter 1-3.fm
Operating Manual
(1610)
Diagram Leak Rate / Pressure
Fig. 1-19 Diagram Test Leak TL4-6
�����������������������������������
Adjustment of leak rate
Chapter 1-3.fm
Operating Manual
(1610)
Diagram Leak Rate / Pressure
Fig. 1-20 Diagram Test Leak TL3-5
����������������������������������
Adjustment of leak rate
1-23
Chapter 1-3.fm
Operating Manual
(1610)
Diagram Leak Rate / Pressure
Fig. 1-21 Diagram Test Leak TL4-6
����������������������������������
Helium (12 bar, 1 l)
200 003 342
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany
Phone: +49 (0)221 347-40
Fax: +49 (0)221 347-41429
E-mail: [email protected]
UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY FRANCE UNITED KINGDOM HONG KONG
Visit our website for contact information and other sales offices worldwide. www.inficon.com
Dokument: linc84en1-04 (1610)