Inficon BPG400 Operating Manual
Read Online
Read the Inficon BPG400 Operating Manual online using the built-in PDF viewer.
Text of the manual
tina03e1 (0206) BPG400 v1.om
In all communications with INFICON, please specify the information on the product
nameplate. For convenient reference copy that information into the space provided
below.
Model:
PN:
SN:
V W
INFICON AG, LI-9496 Balzers
This document applies to products with the following part numbers:
BPG400 (without display)
353-500
(vacuum connection DN 25 ISO-KF)
353-502
(vacuum connection DN 40 CF-R)
BPG400 (with display)
353-501
(vacuum connection DN 25 ISO-KF)
353-503
(vacuum connection DN 40 CF-R)
BPG400-SD (with DeviceNet interface and switching functions)
353-507
(vacuum connection DN 25 ISO-KF)
353-508
(vacuum connection DN 40 CF-R)
BPG400-SP (with Profibus interface and switching functions)
353-505
(vacuum connection DN 25 ISO-KF)
353-506
(vacuum connection DN 40 CF-R)
The part number (PN) can be taken from the product nameplate.
If not indicated otherwise in the legends, the illustrations in this docu-
ment correspond to the KF vacuum connection. They apply to other
vacuum connections by analogy.
In illustrations that apply to all types of the BPG400 gauge family, the
gauge with part number 353-500 is shown.
All BPG400 versions are shipped with an instruction sheet (� � [8]). BPG400-SD
and BPG400-SP come with a supplementary instruction sheet describing the field-
bus interfaces and the switching functions (� � [9]).
We reserve the right to make technical changes without prior notice.
The BPG400 gauges have been designed for vacuum measurement of non-flam-
mable gases and gas mixtures in a pressure range of 5×10-10 … 1000 mbar.
The gauges can be operated in connection with the INFICON Vacuum Gauge
Controller VGC103 or VGC40x or with other control devices.
Product Identification
Validity
Intended Use
tina03e1 (0206) BPG400 v1.om
3
Over the whole measuring range, the gauge has a continuous characteristic curve
and its measuring signal is output as logarithm of the pressure.
The gauge functions with a Bayard-Alpert hot cathode ionization measurement
system (for p < 2.0×10-2 mbar) and a Pirani measurement system (for
p > 5.5×10-3 mbar). In the overlapping pressure range of 2.0×10-2 …
5.5×10-3 mbar, a mixed signal of the two measurement systems is output. The hot
cathode is switched on by the Pirani measurement system only below the switching
threshold of 2.4×10-2 mbar (to prevent filament burn-out). It is switched off when the
pressure exceeds 3.2×10-2 mbar.
DeviceNet™
Open DeviceNet Vendor Association, Inc.
Functional Principle
Trademarks
tina03e1 (0206) BPG400 v1.om
Contents
Product Identification
2
Validity
2
Intended Use
2
Functional Principle
3
Trademarks
3
1 Safety
6
1.1
Symbols Used
6
1.2
Personnel Qualifications
6
1.3
General Safety Instructions
7
1.4
Liability and Warranty
7
2 Technical Data
8
3 Installation
13
3.1
Vacuum Connection
13
3.1.1
Removing and Installing the Electronics Unit
15
3.1.2
Installing the Optional Extension
16
3.1.3
Using the Optional Baffle
17
3.1.3.1
Installing the Baffle
17
3.1.3.2
Replacing the Baffle
19
3.2
Electrical Connection
20
3.2.1
Use With INFICON VGC103 or VGC40x Vacuum Gauge Controller
20
3.2.2
Use With Other Controllers
21
3.2.2.1
Making an Individual Sensor Cable
21
3.2.2.2
Making a DeviceNet Interface Cable (BPG400-SD)
24
3.2.2.3
Making a Profibus Interface Cable (BPG400-SP)
25
3.2.3
Using the Optional Power Supply (With RS232C Line)
26
4 Operation
28
4.1
Measuring Principle, Measuring Behavior
28
4.2
Operational Principle of the Gauge
29
4.3
Putting the Gauge Into Operation
29
4.4
Degas
30
4.5
Display (BPG400)
30
4.6
RS232C Interface
32
4.6.1
Description of the Functions
32
4.6.1.1
Output String (Transmit)
32
4.6.1.2
Input String (Receive)
34
4.7
DeviceNet Interface (BPG400-SD)
35
4.7.1
Description of the Functions
35
4.7.2
Operating Parameters
35
4.7.2.1
Operating Software
35
4.7.2.2
Node Address Setting
35
4.7.2.3
Data Rate Setting
36
4.7.3
Status Lights
36
4.8
Profibus Interface (BPG400-SP)
37
4.8.1
Description of the Functions
37
4.8.2
Operating Parameters
37
4.8.2.1
Operating Software
37
4.8.2.2
Node Address Setting
37
4.9
Switching Functions (BPG400-SD, BPG400-SP)
38
4.9.1
Setting the Switching Functions
38
5 Deinstallation
40
tina03e1 (0206) BPG400 v1.om
5
6 Maintenance, Repair
42
6.1
Maintenance
42
6.1.1
Cleaning the Gauge
42
6.2
Adjusting the Gauge
42
6.2.1
Adjustment at Atmospheric Pressure
42
6.2.2
Zero Point Adjustment
43
6.3
What to Do in Case of Problems
44
6.4
Replacing the Sensor
46
7 Options
47
8 Spare Parts
47
9 Storage
47
10 Returning the Product
48
11 Disposal
48
Appendix
49
A: Relationship Output Signal – Pressure
49
B: Gas Type Dependence
50
C: Literature
52
Declaration of Contamination
53
For cross-references within this document, the symbol (� � XY) is used, for cross-
references to further documents and data sources, the symbol (� � [Z]).
tina03e1 (0206) BPG400 v1.om
1
Safety
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and environmental damage.
Caution
Information on correct handling or use. Disregard can lead to malfunctions or
minor equipment damage.
Notice
Hint, recommendation
The result is O.K.
The result is not as expected
Optical inspection
Waiting time, reaction time
Skilled personnel
All work described in this document may only be carried out by persons who
have suitable technical training and the necessary experience or who have been
instructed by the end-user of the product.
1.1 Symbols Used
1.2 Personnel Qualifications
tina03e1 (0206) BPG400 v1.om
7
��
Adhere to the applicable regulations and take the necessary precautions for the
process media used.
Consider possible reactions between the materials (� � 11) and the process
media.
Consider possible reactions of the process media due to the heat generated by
the product.
��
Adhere to the applicable regulations and take the necessary precautions for all
work you are going to do and consider the safety instructions in this document.
��
Before beginning to work, find out whether any vacuum components are con-
taminated. Adhere to the relevant regulations and take the necessary precau-
tions when handling contaminated parts.
Communicate the safety instructions to all other users.
INFICON assumes no liability and the warranty becomes null and void if the end-
user or third parties
��
disregard the information in this document
��
use the product in a non-conforming manner
��
make any kind of interventions (modifications, alterations etc.) on the product
��
use the product with accessories not listed in the corresponding product
documentation.
The end-user assumes the responsibility in conjunction with the process media
used.
1.3 General Safety
Instructions
1.4 Liability and Warranty
tina03e1 (0206) BPG400 v1.om
2
Technical Data
Measuring range (air, N2, O2)
5×10-10 … 1000 mbar, continuous
Accuracy
15% of reading in the range of
10-8 … 10-2 mbar
(after 5 min stabilization)
Repeatability
5% of reading in the range of
10-8 … 10-2 mbar
(after 5 min stabilization)
Gas type dependence
� Appendix B
Switching on threshold
Switching off threshold
2.4×10-2 mbar
3.2×10-2 mbar
Emission current
p �7.2×10-6 mbar
7.2×10-6 mbar
tina03e1 (0206) BPG400 v1.om
9
Operating voltage at the gauge
+24 VDC (20 … 28 VDC) 1)
ripple max. 2 Vpp
Power consumption
Standard
Degas
Emission start (<200 ms)
�0.5 A
�0.8 A
�1.4 A
Power consumption
BPG400
BPG400-SD, BPG400-SP
�16 W
�18 W
Fuse necessary
1.25 AT
BPG400-SD requires an additional, separate power supply for the
DeviceNet interface (� � 24).
Supply voltage at the DeviceNet con-
nector, (Pin 2 and Pin 3)
+24 VDC (+11 … 25 VDC)
Power consumption
<2 W
The gauge is protected against reversed polarity of the supply voltage.
For reasons of compatibility, the expression "sensor cable" is used for all
BPG400 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BPG400-SD and BPG400-SP) is
normally transmitted via DeviceNet or Profibus.
Electrical connector
BPG400
BPG400-SD, BPG400-SP
D-Sub,15 pins, male
� � 22
� � 23
Cable for BPG400
Analog values only
Without degas function
Analog values
With degas function
Analog values
With degas function
And RS232C interface
Cable for BPG400-SD, BPG400-SP
4 conductors plus shielding
5 conductors plus shielding
7 conductors plus shielding
depending on the functions used,
max. 15 conductors plus shielding
Max. cable length (supply voltage 24 V1))
Analog and fieldbus operation
RS232C operation
�35 m, conductor cross-section 0.25 mm²
�50 m, conductor cross-section 0.34 mm²
�100 m, conductor cross-section 1.0 mm²
�30 m
Gauge identification
42 k� resistor between Pin 10 (sensor
cable) and GND
Switching functions
BPG400
BPG400-SD, BPG400-SP
Adjustment range
none
2 (Setpoints A and B)
1×10-9 mbar … 100 mbar
Setpoints adjustable via potentiometers
(Setpoints A and B), one floating, nor-
mally open relay contact per setpoint
(� � 23, 38)
Relay contact rating
Voltage
Current
�60 V
�0.5 ADC
1)
Measured at sensor cable connector (consider the voltage drop as function of
the sensor cable length).
Sensor cable
tina03e1 (0206) BPG400 v1.om
Data rate
Data format
Connections (sensor cable connector)
TxD (Transmit Data)
RxD (Receive Data)
GND
9600 baud
binary
8 data bits
one stop bit
no parity bit
no handshake
Pin 13
Pin 14
Pin 5
(Function and communication protocol of the RS232C interface � � 32)
Fieldbus name
DeviceNet
Standard applied
� � [6]
Communication protocol, data format
� � [1], [4]
Interface, physical
CAN bus
Data rate (adjustable via "RATE"
switch)
125 kbaud
250 kbaud
500 kbaud
"P" (125 kbaud, 250 kbaud, 500 kbaud
programmable via DeviceNet
(� � [1])
Node address (MAC ID)
(Adjustable via "ADDRESS", "MSD",
"LSD" switches)
0 … 63dec
"P" (0 … 63 programmable via
DeviceNet, � � [1])
DeviceNet connector
Micro-Style, 5 pins, male
Cable
Shielded, special DeviceNet cable,
5 conductors (� � 24 and � [4])
Cable length, system wiring
According to DeviceNet specifications
(� � [6], [4])
Fieldbus name
Profibus
Standard applied
� � [7]
Communication protocol data format
� � [10], [7]
Interface, physical
RS485
Data rate
�12 Mbaud (� � [10])
Node address
(Adjustable via hexadecimal
"ADDRESS", "MSD", "LSD" switches)
00 … 3Fhex (0 … 127dec)
Profibus connection
D-Sub, 9 pins, female
Cable
Shielded, special Profibus cable
(� � 25 and � [5])
Cable length, system wiring
According to Profibus specifications
(� � [7], [5])
RS232C interface
DeviceNet interface
(BPG400-SD)
Profibus interface
(BPG400-SP)
tina03e1 (0206) BPG400 v1.om
11
Materials exposed to vacuum
Housing, supports, screens
Feedthroughs
Insulator
Cathode
Cathode holder
Pirani element
stainless steel
NiFe, nickel plated
glass
iridium, yttrium oxide (Y2O3)
molybdenum
tungsten, copper
Internal volume
353-500, 353-501
353-505, 353-507
353-502, 353-503
353-506, 353-508
�24 cm3
�24 cm3
�34 cm3
�34 cm3
Pressure max.
2 bar (absolute)
Admissible temperatures
Storage
Operation
Bakeout
-20 … 70 °C
0 … 50 °C
+150 °C (without electronics unit or with
bakeout extension � � 16)
Relative humidity
(year's mean / during 60 days)
�65 / 85% (no condensation)
Use
indoors only
altitude up to 2000 m NN
Type of protection
IP 30
Part numbers
353-500
353-501
353-505
(353-507) 1)
4-40UNC 2B
DN 25 ISO-KF
353-500
353-501
Vacuum
Ambiance
Dimensions
tina03e1 (0206) BPG400 v1.om
Part numbers
353-502
353-503
353-506
(353-508) 1)
4-40UNC 2B
DN 40 ISO-KF
353-502
353-503
1)
Gauges with DeviceNet connector are 14 mm longer.
The other dimensions of housing and vacuum connection are identical.
Part numbers
353-507
353-508
Weight
353-500, 353-501
353-502, 353-503
353-505, 353-507
353-506, 353-508
285 g
550 g
430 g
695 g
tina03e1 (0206) BPG400 v1.om
13
3
Installation
DANGER
Caution: overpressure in the vacuum system >1 bar
Injury caused by released parts and harm caused by escaping process
gases can result if clamps are opened while the vacuum system is
pressurized.
Do not open any clamps while the vacuum system is pressurized. Use
the type of clamps which are suited to overpressure.
DANGER
Caution: overpressure in the vacuum system >2.5 bar
KF flange connections with elastomer seals (e.g. O-rings) cannot
withstand such pressures. Process media can thus leak and possibly
damage your health.
Use O-rings provided with an outer centering ring.
DANGER
The gauge must be electrically connected to the grounded vacuum
chamber. This connection must conform to the requirements of a
protective connection according to EN 61010:
��
CF connections fulfill this requirement
��
For gauges with a KF vacuum connection, use a conductive me-
tallic clamping ring.
Caution
Caution: vacuum component
Dirt and damages impair the function of the vacuum component.
When handling vacuum components, take appropriate measures to
ensure cleanliness and prevent damages.
The gauge may be mounted in any orientation. To keep condensates
and particles from getting into the measuring chamber, preferably
choose a horizontal to upright position. See dimensional drawing for
space requirements (� � 11).
��
The gauge is supplied with a built-in grid. For potentially contaminating appli-
cations and to protect the electrodes against light and fast particles, installation
(� � 17) of the optional baffle is recommended (� � 47).
��
For the vacuum connection sealing, use of a metal seal (� � 47) is recom-
mended, since elastomer seals (e.g. FPM) can impair the measurement accu-
racy already in the 10-6 mbar range by outgassing.
��
The sensor can be baked at up to 150 °C. At temperatures exceeding 50 °C,
the electronics unit has to be removed (� � 15) or an extension (Option
� � 47) has to be installed (� � 16).
3.1 Vacuum Connection
tina03e1 (0206) BPG400 v1.om
�
Remove the protective lid.
The protective lid will be
needed for mainte-
nance.
�
Make the flange connection.
When installing the gauge, make sure that the area around the con-
nector is accessible for the tools required for adjustment while the gauge
is mounted (� � 38, 42).
When installing the gauge, allow for installing/deinstalling the connectors
and accommodation of cable loops.
If you are using a gauge with display, make sure easy reading of the
display is possible.
The gauge is now installed.
Procedure
tina03e1 (0206) BPG400 v1.om
15
��
Allen key, size 2.5 mm
�
Unscrew the hexagon socket set screw (1) on the side of the electronics
unit (2).
2
1
�
Remove the electronics unit without twisting it.
Removal of the electronics unit is completed.
�
Place the electronics unit on the sensor (3) (be careful to correctly align the
pins and notch (4)).
4
3
3.1.1 Removing and Installing
the Electronics Unit
Required tools / material
Removing the electronics unit
Installing the electronics unit
tina03e1 (0206) BPG400 v1.om
�
Slide the electronics unit in to the mechanical stop and lock it with the
hexagon socket set screw (1).
The electronics unit is now installed.
With the optional extension (� � 47) the sensor can also be baked during opera-
tion at temperatures up to 150 °C (only at p<10-2 mbar because at high tempera-
tures, the accuracy of the Pirani sensor decreases).
Caution
Caution: rising heat
The electronics unit of gauges that are installed vertically, above the
source of heat can be damaged through rising heat even with an
installed extension.
123
Bakeout area
When installing the extension, make sure that the area around the con-
nector is accessible for the tools required for adjustment while the gauge
is mounted (� � 38, 42).
When installing the gauge, allow for installing/deinstalling the connectors
and accommodation of cable loops.
If you are using a gauge with display, ensure easy reading of the display.
��
Extension (� � 47)
��
Allen key, size 2.5 mm
��
Allen key, size 1.5 mm
�
Remove the electronics unit (2) (� � 15).
�
Slide the sensor (3) into the extension (6) to the mechanical stop (be careful
to correctly position the pins and notch (4)).
3.1.2 Installing the Optional
Extension
Bakeout area
Required tools / material
Procedure
tina03e1 (0206) BPG400 v1.om
17
�
Secure the sensor with the hex socket set screws (7) using an Allen key,
size 1.5 mm.
6
7
3
4
4a
2
1
�
Slide the electronics unit (2) in to the mechanical stop (be careful to cor-
rectly align the pins and notch (4a)).
�
Secure the electronics unit (2) with the hex socket set screw (1) using an
Allen key, size 2.5 mm.
The extension is now installed.
In severely contaminating processes and to protect measurement electrodes opti-
cally against light and fast particles, replacement of the built-in grid by the optional
baffle (� � 47) is recommended.
The optional baffle will be installed at the sensor opening of the deinstalled gauge
(Deinstallation � � 40).
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the
desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
��
Baffle (� � 47)
��
Pointed tweezers
��
Pin (e.g. pencil)
3.1.3 Using the Optional Baffle
3.1.3.1
Installing the Baffle
Required tools / material
tina03e1 (0206) BPG400 v1.om
�
Carefully remove the grid with tweezers.
�
Carefully place the baffle onto the sensor opening.
�
Using a pin, press the baffle down in the center until it catches.
The baffle is now installed (Installation of the gauge � � 13).
Procedure
tina03e1 (0206) BPG400 v1.om
19
In case of severe contamination the baffle can be easily replaced. For this purpose,
the gauge first has to be deinstalled (Deinstallation of the gauge � � 40).
Caution
Caution: dirt sensitive area
Touching the product or parts thereof with bare hands increases the
desorption rate.
Always wear clean, lint-free gloves and use clean tools when working
in this area.
��
New baffle (� � 47)
��
Screwdriver No 1
��
Pin (e.g. pencil)
�
Carefully remove the baffle with the screwdriver.
�
Place new baffle carefully onto the sensor opening.
3.1.3.2
Replacing the Baffle
Required tools / material
Procedure
tina03e1 (0206) BPG400 v1.om
�
Using a pin, press the baffle down in the center until it catches.
The new baffle is now installed (Installation of the gauge � � 13).
If the gauge is used with an INFICON VGC103 or VGC40x controller, a corre-
sponding sensor cable is required (� � 47). The sensor cable permits supplying
the gauge with power, transmitting measurement values and gauge statuses, and
making parameter settings.
Caution
Caution: data transmission errors
If the gauge is operated with the INFICON VGC103 or VGC40x
Vacuum Gauge Controller (RS232C) and a fieldbus interface at the
same time, data transmission errors may occur.
The gauge must not be operated with an INFICON VGC103 or
VGC40x controller and DeviceNet or Profibus at the same time.
��
Sensor cable for INFICON controllers (� � 47)
3.2 Electrical Connection
3.2.1 Use With INFICON
VGC103 or VGC40x
Vacuum Gauge
Controller
Required material
tina03e1 (0206) BPG400 v1.om
21
�
Plug the sensor connector into the gauge and secure it with the locking
screws.
�
Connect the other end of the sensor cable to the INFICON controller and
secure it.
The gauge can now be operated with the VGC103 or VGC40x controller.
The gauge can also be operated with other controllers.
Especially the fieldbus versions BPG400-SD (DeviceNet) and BPG400-SP
(Profibus) are usually operated as part of a network, controlled by a master or bus
controller. In such cases, the control system has to be operated with the
appropriate software and communication protocol (� � [1] or [10]).
For reasons of compatibility, the expression "sensor cable" is used for all
BPG400 versions in this document, although the pressure reading of the
gauges with fieldbus interface (BPG400-SD and BPG400-SP) is
normally transmitted via DeviceNet or Profibus.
The sensor cable is required for supplying all BPG400 types with power.
In connection with the gauges with fieldbus interface (BPG400-SD and
BPG400-SP), it also permits access to the contacts of the switching
functions (� � 23, 38).
The application and length of the sensor cable have to be considered when deter-
mining the number and cross sections of the conductors (� � 9).
�
Open the cable connector (D-Sub, 15 pins, female).
�
Prepare the cable and solder/crimp it to the connector as indicated in the
diagram of the gauge used:
Procedure
3.2.2 Use With Other
Controllers
3.2.2.1
Making an Individual
Sensor Cable
Cable type
Procedure
tina03e1 (0206) BPG400 v1.om
Electrical connection
Pin 2
Signal output (measuring signal)
Pin 5
Supply common, GND
Pin 7
Degas on, active high
Pin 8
Supply
Pin 10 Gauge identification
Pin 12 Signal common, GND
Pin 13 RS232C, TxD
Pin 14 RS232C, RxD
Pin 15 Shielding, housing, GND
8
9
1
15
+24 VDC
+24 VDC
Pins 1, 3, 4, 6, 9 and 11 are
not connected internally.
0 … +10 V
D-Sub, 15 pins
female,
soldering side
TxD
BPG400
42k�
RxD
Measuring
signal
Degas
+Ub
13
14
2
12
7
8
5
15
+
1.25AT
Identification
10
RS232C
+24 V
Sensor cable connection
BPG400
tina03e1 (0206) BPG400 v1.om
23
.
15
8
9
1
D-Sub, 15 pins
female,
soldering side
Electrical connection
Pin 1 Relay Switching function A, common
Pin 2 Signal output (measuring signal)
0 … +10 V
Pin 3 Threshold value (Setpoint) A
0 … +10 V
Pin 4 Relay Switching function A, normally open contact
Pin 5
Supply common, GND
Pin 6 Threshold value (Setpoint) B
0 … +10 V
Pin 7
Degas on, active high
+24 V
Pin 8
Supply voltage
+24 V
Pin 9 Relay Switching function B, common
Pin 10 Gauge identification
Pin 11 Relay Switching function B, normally open contact
Pin 12 Signal common GND
Pin 13 RS232, TxD
Pin 14 RS232, RxD
Pin 15 Shielding, housing GND
SP A
SP B
TxD
RxD
Degas
+Ub
42 k�
BPG400-SD, BPG400-SP
Measuring
signal
Threshold value, SP A
Threshold value, SP B
3
6
1
4
9
11
13
14
7
8
2
12
5
15
1.25 AT
24V
Degas
Identification
RS232
10
WARNING
The supply common (Pin 5) and the shielding (Pin 15) must be
connected at the supply unit with protective ground.
Incorrect connection, incorrect polarity or inadmissible supply
voltages can damage the gauge.
For cable lengths up to 5 m (0.34 mm2 conductor cross-section) the
output signal can be measured directly between the positive signal out-
put (Pin 2) and supply common GND (Pin 5) without loss of accuracy. At
greater cable lengths, differential measurement between signal output
(Pin 2) and signal common (Pin 12) is recommended.
�
Reassemble the cable connector.
�
On the other cable end, terminate the cable according to the requirements
of the gauge controller you are using.
Sensor cable connection
BPG400-SD, BPG400-SP
tina03e1 (0206) BPG400 v1.om
�
Plug the sensor connector into
the gauge and secure it with the
locking screws.
�
Connect the other end of the sensor cable to the connector of the instru-
ment or gauge controller you are using.
The gauge can now be operated via analog and RS232C interface.
For operating BPG400-SD via DeviceNet, an interface cable conforming to the
DeviceNet standard is required.
If no such cable is available, make one according to the following indications.
A shielded special 5 conductor cable conforming to the DeviceNet standard has to
be used (� � [4], [6]).
�
Make the DeviceNet cable according to the following indications.
3
1
4
2
5
Micro-Style, 5 pins,
(DeviceNet)
female, soldering side
Pin
Function (BPG400-SD)
1
Drain
2
Supply
+24 VDC
(DeviceNet) interface only
3
Supply common
GND
(DeviceNet interface only)
4
CAN_H
5
CAN_L
3.2.2.2
Making a DeviceNet
Interface Cable
(BPG400-SD)
Cable type
Procedure
tina03e1 (0206) BPG400 v1.om
25
�
Plug the DeviceNet (and sensor) cable connector into the gauge.
Sensor cable
DeviceNet cable
�
Lock the DeviceNet (and sensor) cable connector.
The gauge can now be operated via DeviceNet interface (� � 35).
For operating BPG400-SP via Profibus, an interface cable conforming to the
Profibus standard is required.
If no such cable is available, make one according to the following indications.
Only a cable that is suited to Profibus operation may be used (� � [5], [7]).
�
Make the Profibus interface cable according to the following indications:
1
5
6
9
D-Sub, 9 pins
male, soldering side
Pin
Function (BPG400-SP)
1
Do not connect
2
Do not connect
3
RxD/TxD-P
4
CNTR-P
1)
5
DGND
2)
6
VP
2)
7
Do not connect
8
RxD/TxD-N
9
Do not connect
1)
Only to be connected if an optical link module is used.
2)
Only required as line termination for devices at both ends of bus cable
(� � [5]).
3.2.2.3
Making a Profibus
Interface Cable
(BPG400-SP)
Cable type
Procedure
tina03e1 (0206) BPG400 v1.om
�
Plug the Profibus (and sensor) cable connector into the gauge.
Sensor cable
Profibus cable
�
Lock the Profibus (and sensor) cable connector.
The gauge can now be operated via Profibus interface (� � 37).
The optional 24 V power supply (� � 47) allows RS232C operation of the BPG400
gauge with any suitable instrument or control device (e.g. PC).
The instrument or control device needs to be equipped with a software that sup-
ports the RS232C protocol of the gauge (� � 32).
Mains connection
Mains voltage
90 … 250 VAC 50 … 60 Hz
Mains cable
1.8 meter (Schuko DIN and U.S. con-
nectors)
Output (operating voltage of gauge)
Voltage
21 … 27 VDC, set to 24 VDC
Current
Max. 1.5 A
Gauge connection
Connector
D-Sub, 15 pins, female
24 V cable
5 m, black
Connection of the instrument or control
device
RS232C connection
D-Sub, 9 pins, female
Cable
5 m, black, 3 conductors, shielded
13
14
8
15
5
2
3
5
4
6
7
8
N
PE
L
DC
AC
GND
+24 V
PE
BPG400
D-Sub, 15 pins
RS232C
D-Sub, 9 pins
Mains
90 ... 250 VAC
50 ... 60 Hz
3.2.3 Using the Optional Power
Supply (With RS232C
Line)
Technical data
Wiring diagram
tina03e1 (0206) BPG400 v1.om
27
�
Connect the gauge to the power supply and lock the connector with the
screws.
�
Connect the RS232C line to the instrument or control device and lock the
connector with the screws.
RS232C
Power supply
Mains
PC
BPG400
�
Connect the power supply to the mains.
�
Turn the power supply on.
The gauge can now be operated via RS232C interface (� � 32).
Connecting the power supply
tina03e1 (0206) BPG400 v1.om
4
Operation
The BPG400 vacuum gauges consist of two separate measuring systems (hot
cathode Bayard-Alpert (BA) and Pirani).
The BA measuring system uses an electrode system according to Bayard-Alpert
which is designed for a low x-ray limit.
The measuring principle of this measuring system is based on gas ionization.
Electrons emitted by the hot cathode (F) ionize a number of molecules proportional
to the pressure in the measuring chamber. The ion collector (IC) collects the thus
generated ion current I+ and feeds it to the electrometer amplifier of the
measurement instrument. The ion current is dependent upon the emission current
Ie, the gas type, and the gas pressure p according to the following relationship:
I+ = Ie × p × C
Factor C represents the sensitivity of the gauge head. It is generally specified for
N2.
The lower measurement limit is 5×10-10 mbar (gauge metal sealed).
To usefully cover the whole range of 5×10-10 mbar … 10-2 mbar, a low emission
current is used in the high pressure range (fine vacuum) and a high emission cur-
rent is used in the low pressure range (high vacuum). The switching of the emis-
sion current takes place at decreasing pressure at approx. 7.2×10-6 mbar, at in-
creasing pressure at approx. 3.2×10-5 mbar. At the switching threshold, the
BPG400 can temporarily (<2 s) deviate from the specified accuracy.
Diagram of the BA measuring system
F
hot cathode (filament)
IC
ion collector
EC
anode (electron collector)
IC
200V
+ 40V
–
+
–
(Degas 2.5V)
(Degas 250V)
EC
F
F
EC
IC
Within certain limits, the thermal conductibility of gases is pressure dependent. This
physical phenomenon is used for pressure measurement in the thermal con-
ductance vacuum meter according to Pirani. A self-adjusting bridge is used as
measuring circuit (� schematic). A thin tungsten wire forms the sensor element.
Wire resistance and thus temperature are kept constant through a suitable control
circuit. The electric power supplied to the wire is a measure for the thermal con-
ductance and thus the gas pressure. The basic principle of the self-adjusting bridge
circuit is shown in the following schematic.
4.1 Measuring Principle,
Measuring Behavior
Bayard-Alpert
Pirani
tina03e1 (0206) BPG400 v1.om
29
UB
Pirani sensor
The bridge voltage UB is a measure for the gas pressure and is further processed
electronically (linearization, conversion).
The BPG400 gauges continuously cover the measuring range
5×10-10 mbar … 1000 mbar.
��
The Pirani constantly monitors the pressure.
��
The hot cathode (controlled by the Pirani) is activated only at pressures
<2.4×10-2 mbar.
If the measured pressure is higher than the switching threshold, the hot cathode is
switched off and the Pirani measurement value is output.
If the Pirani measurement drops below the switching threshold (p = 2.4×10-2 mbar),
the hot cathode is switched on. After heating up, the measured value of the hot
cathode is fed to the output. In the overlapping range of 5.5×10-3 … 2.0×10-2 mbar,
the output signal is generated from both measurements.
Pressure rising over the switching threshold (p = 3.2×10-2 mbar) causes the hot
cathode to be switched off. The Pirani measurement value is output.
The output signal is gas type dependent. The characteristic curves are accurate for
dry air, N2 and O2. They can be mathematically converted for other gases
(� Appendix B).
The measuring currents of the Bayard-Alpert and Pirani sensor are converted into a
frequency. A micro-controller converts this frequency into a digital value repre-
senting the measured total pressure. After further processing this value is available
as analog measurement signal (0 … +10 V) at the output (sensor cable connector
Pin 2 and Pin 12). The maximum output signal is internally limited to +10 V
(�atmosphere). The measured value can be read as digital value through the
RS232C interface (Pins 13, 14, 15) (� � 32). Gauges with a display show the
value as pressure. The default setting of the displayed pressure unit is mbar. It can
be modified via the RS232C interface (� � 32).
In addition to converting the output signal, the micro controller's functions include
monitoring of the emission, calculation of the total pressure based on the meas-
urements of the two sensors, and communication via RS232C interface.
When the operating voltage is supplied (� Technical Data), the output signal is
available between Pin 2 (+) and Pin 12 (–) of the sensor cable connector
(Relationship Output Signal – Pressure � Appendix A).
Allow for a stabilizing time of approx. 10 min. Once the gauge has been switched
on, permanently leave it on irrespective of the pressure.
Communication via the digital interfaces is described in separate sections.
Schematic
Measuring range
Gas type dependence
4.2 Operational Principle of
the Gauge
4.3 Putting the Gauge Into
Operation
tina03e1 (0206) BPG400 v1.om
Gauge failures due to contamination are not covered by the warranty.
Deposits on the electrode system of the BA gauge can lead to unstable measure-
ment readings.
The degas process allows in-situ cleaning of the electrode system by heating the
electron collector grid to approx. 700 °C by electron bombardment.
Depending on the application, this function can be activated manually by a switch
or automatically by the system control via a digital interface. The BPG400
automatically terminates the degas process after 3 minutes, if it has not been
stopped before.
The degas process should be run at pressures below 7.2×10-6 mbar
(emission current 5 mA).
For a repeated degas process, the control signal first has to change from ON
(+24 V) to OFF (0 V), to then start degas again with a new ON (+24 V) command. It
is recommended that the degas signal be set to OFF again by the system control
after 3 minutes of degassing, to achieve an unambiguous operating status.
The gauges with part number
352-501 and
353-503
have a built-in two-line display with an LCD matrix of 32×16 pixels. The first line
shows the pressure, the second line the pressure unit, the function and possible
errors. The background illumination is usually green, in the event of an error, it
changes to red. The pressure is displayed in mbar (default), Torr or Pa. The pres-
sure unit can be changed via RS232C interface (� � 32).
Pressure reading, pressure unit
(none) Pirani operation
Emission 25 �A
Emission 5 mA
Degas
1000 mbar adjustment (Pirani)
4.4 Degas
Contamination
4.5 Display (BPG400)
Pressure display
Function display
tina03e1 (0206) BPG400 v1.om
31
no error
(green background illumination)
Pirani sensor warning
(red background illumination)
Pirani sensor error
(red background illumination)
BA sensor error
(red background illumination)
Internal data connection failure
(red background illumination)
Error display
tina03e1 (0206) BPG400 v1.om
The built-in RS232C interface allows transmission of digital measurement data and
instrument conditions as well as the setting of instrument parameters.
Caution
Caution: data transmission errors
If the gauge is operated with the INFICON VGC103 or VGC40x
Vacuum Gauge Controller (RS232C) and a fieldbus interface at the
same time, data transmission errors may occur.
The gauge must not be operated with the RS232C interface and
DeviceNet or Profibus at the same time.
The interface works in duplex mode. A nine byte string is sent continuously without
a request approx. every 20 ms.
Commands are transmitted to the gauge in a five byte input (receive) string.
��
Data rate
9600 baud
set value, no handshake
��
Byte
8 data bits
1 stop bit
��
TxD
Pin 13
��
RxD
Pin 13
��
GND
Pin 5
(Sensor cable connector)
The complete output string (frame) is nine bytes (byte 0 … 8). The data string is
seven bytes (byte 1 … 7).
Byte No
Function
Value
Comment
0
Length of data string
7
(Set value)
1
Page number
5
(For BPG400)
2
Status
� Status byte
3
Error
� Error byte
4
Measurement high byte
0 … 255
� Calculation of pressure value
5
Measurement low byte
0 … 255
� Calculation of pressure value
6
Software version
0 … 255
� Software version
7
Sensor type
10
(For BPG400)
8
Check sum
0 … 255
� Synchronization
Synchronization of the master is achieved by testing three bytes:
Byte No
Function
Value
Comment
0
Length of data string
7
Set value
1
Page number
5
(For BPG400)
8
Check sum of bytes No 1 … 7
0 … 255
Low byte of check
sum 1)
1)
High order bytes are ignored in the check sum.
4.6 RS232C Interface
4.6.1 Description of the
Functions
Operational parameters
Electrical connections
4.6.1.1
Output String (Transmit)
Format of the output string
Synchronization
tina03e1 (0206) BPG400 v1.om
33
Bit 1
Bit 0
Definition
0
0
Emission off
0
1
Emission 25 �A
1
0
Emission 5 mA
1
1
Degas
Bit 2
Definition
0
1000 mbar adjustment off
1
1000 mbar adjustment on
Bit 3
Definition
0 � 1
Toggle bit, changes with every
string received correctly
Bit 5
Bit 4
Definition
0
0
Current pressure unit mbar
0
1
Current pressure unit Torr
1
0
Current pressure unit Pa
Bit 7
Bit 6
Definition
x
x
Not used
Bit 3
Bit 2
Bit 1
Bit 0
Definition
x
x
x
x
Not used
Bit 7
Bit 6
Bit 5
Bit 4
Definition
0
1
0
1
Pirani adjusted poorly
1
0
0
0
BA error
1
0
0
1
Pirani error
The software version of the gauge can be calculated from the value of byte 6 of the
transmitted string according to the following rule:
Version No = ValueByte 6 / 20
(Example: According to the above formula, ValueByte 6 of 32 means software ver-
sion 1.6)
The pressure can be calculated from bytes 4 and 5 of the transmitted string. De-
pending on the currently selected pressure unit (� byte 2, bits 4 and 5), the ap-
propriate rule must be applied.
As result, the pressure value results in the usual decimal format.
pmbar
= 10((high byte × 256 + low byte) / 4000 - 12.5)
pTorr
= 10((high byte × 256 + low byte) / 4000 - 12.625)
pPa
= 10((high byte × 256 + low byte) / 4000 - 10.5)
Status byte
Error byte
Software version
Calculation of the pressure
value
tina03e1 (0206) BPG400 v1.om
The example is based on the following output string:
Byte No
0
1
2
3
4
5
6
7
8
Value
7
5
0
0
242
48
20
10
69
The instrument or controller (receiver) interprets this string as follows:
Byte No
Function
Value
Comment
0
Length of data
string
7
(Set value)
1
Page number
5
BPG400
2
Status
0
Emission = off
Pressure unit = mbar
3
Error
0
No error
4
5
Measurement
High byte
Low byte
242
48
Calculation of the pressure:
p = 10((242 × 256 + 48) / 4000 - 12.5) = 1000 mbar
6
Software version
20
Software version = 20 / 20 = 1.0
7
Sensor type
10
BPG400
8
Check sum
69
5 + 0 + 0 + 242 + 48 + 20 + 10 =
325dec � 01 45hex
High order byte is ignored �
Check sum = 45hex � 69dec
For transmission of the commands to the gauge, a string (frame) of five bytes is
sent (without ). Byte 1 to byte 3 form the data string.
Byte no
Function
Value
Comment
0
Length of data string
3
(Set value)
1
Data
� admissible input strings
2
Data
� admissible input strings
3
Data
� admissible input strings
4
Check sum
(from bytes No 1 … 3)
0 … 255
(low byte of sum) 1)
1)
High order bytes are ignored in the check sum.
For commands to the gauge, six defined strings are used:
Byte No
Command
0
1
2
3
4 2)
Set the unit mbar in the display
3
16
62
0
78
Set the unit Torr in the display
3
16
62
1
79
Set the unit Pa in the display
3
16
62
2
80
Power-failure-safe storage of current unit
3
32
62
62
156
Switch degas on
(switches itself off after 3 minutes)
3
16
93
148
1
Switch degas off before 3 minutes
3
16
93
105
214
2)
Only low order byte of sum (high order byte is ignored).
Example
4.6.1.2
Input String (Receive)
Format of the input string
Admissible input strings
tina03e1 (0206) BPG400 v1.om
35
This interface allows operation of BPG400-SD with part number
353-507 and
353-508
in connection with other devices that are suited for DeviceNet operation. The
physical interface and communication firmware of BPG400-SD comply with the
DeviceNet standard (� � [4], [6]).
Two adjustable switching functions are integrated in BPG400-SD. The corre-
sponding relay contacts are available at the sensor cable connector (� � 8, 23,
38).
The basic sensor and sensor electronics of all BPG400 gauges are identical.
Caution
Caution: data transmission errors
If the gauge is operated via RS232C interface and DeviceNet interface
at the same time, data transmission errors may occur.
The gauge must not be operated via RS232C interface and DeviceNet
interface at the same time.
Via this interface, the following and further data are exchanged in the standardized
DeviceNet protocol (� � [1]):
��
Pressure reading
��
Pressure unit (Torr, mbar, Pa)
��
Degas function
��
Gauge adjustment
��
Status and error messages
��
Status of the switching functions
As the DeviceNet protocol is highly complex, the parameters and programming of
BPG400-SD are described in detail in the separate Communication Protocol
(� � [1]).
Before the gauge is put into operation, it has to be configured for DeviceNet op-
eration. A configuration tool and the device specific EDS file (Electronic Data
Sheet) are required for this purpose. The EDS file can be downloaded via internet
(� � [3]).
For unambiguous identification of the gauge in a DeviceNet environment, a node
address is required. The node address setting is made on the gauge or pro-
grammed via DeviceNet.
Set the node address (0 … 63dec) via the "ADDRESS" "MSD"
and "LSD" switches. The node address is polled by the firmware
when the gauge is switched on. If the setting deviates from the
stored value, the new value is taken over into the NVRAM. If a
setting higher than 63 is made, the previous node address set-
ting remains valid.
If the „MSD“ switch is in the "P" position, the node address is
programmable via DeviceNet (� � [1]).
4.7 DeviceNet Interface
(BPG400-SD)
4.7.1 Description of the
Functions
4.7.2 Operating Parameters
4.7.2.1
Operating Software
4.7.2.2
Node Address Setting
tina03e1 (0206) BPG400 v1.om
The admissible data rate depends on a number of factors such as system pa-
rameters and cable length � � [4], [6]). It can be set on the gauge or pro-
grammed via DeviceNet.
By means of the "RATE" switch, the data rate can be set to 125
("1"), 250 ("2") or 500 kbaud ("5").
If the switch is in any of the "P" positions, the data rate is pro-
grammable via DeviceNet (� � [1]).
Two lights (LEDs) on the gauge inform on the gauge status and the current
DeviceNet status.
Light status
Description
Dark
No supply
Flashing
red/green
Selftest
Green
Normal operation
Red
Non recoverable error
Light status
Description
Dark
Gauge not online:
�� Selftest not yet concluded
�� No supply, � "STATUS MOD" light
Flashing
green
Gauge online but no communication:
�� Selftest concluded but no communication to other nodes
established
�� Gauge not assigned to any master
Green
Gauge online; necessary connections established
Flashing red
One or several input/output connections in "timed out" status
Red
Communication error. The gauge has detected an error that im-
pedes communication via the network (e.g. two identical node
addresses (MAC IC) or "Bus-off")
The gauge is connected to the DeviceNet system via the 5-pin DeviceNet con-
nector (� � 24).
4.7.2.3
Data Rate Setting
4.7.3 Status Lights
"STATUS MOD"
(gauge status):
"STATUS NET"
(network status):
Electrical connections
tina03e1 (0206) BPG400 v1.om
37
This interface allows operation of BPG400-SP with part number
353-505 and
353-506
in connection with other devices that are suited for Profibus operation. The physical
interface and communication firmware of BPG400-SP comply with the Profibus
standard (� � [7], [5].
Two adjustable switching functions are integrated in the BPG400-SP. The corre-
sponding relay contacts are available at the sensor cable connector (� � 8, 23,
38).
The basic sensor and sensor electronics of all BPG400 gauges are identical.
Caution
Caution: data transmission errors
If the gauge is operated via RS232C interface and Profibus interface at
the same time, data transmission errors may occur.
The gauge must not be operated via RS232C interface and Profibus
interface at the same time.
Via this interface, the following and further data are exchanged in the standardized
Profibus protocol (� � [2]):
��
Pressure reading
��
Pressure unit (Torr, mbar, Pa)
��
Degas function
��
Gauge adjustment
��
Status and error messages
��
Status of the switching functions
As the DeviceNet protocol is highly complex, the parameters and programming of
BPG400-SP are described in detail in the separate Communication Protocol
(� � [2]).
For operating the gauge via Profibus, prior installation of the BPG400 specific GSD
file is required on the bus master side. This file can be downloaded via internet
(� � [3]).
For unambiguous identification of the gauge in a Profibus environment, a node
address is required. The node address setting is made on the gauge.
The node address (0 … 127dec) is set in hexadecimal form
(00 … 3Fhex) via the "ADDRESS", "MSD", and "LSD" switches.
The node address is polled by the firmware when the gauge is
switched on. If the setting deviates from the stored value, the
new value is taken over into the NVRAM. If a value higher than
127dec (3Fhex) is entered, the node address setting currently
stored in the device remains valid.
The gauge is connected to Profibus via the 9-pin Profibus connector (� � 25).
4.8 Profibus Interface
(BPG400-SP)
4.8.1 Description of the
Functions
4.8.2 Operating Parameters
4.8.2.1
Operating Software
4.8.2.2
Node Address Setting
Electrical connections
tina03e1 (0206) BPG400 v1.om
The gauges BPG400-SD and BPG400-SP have two independent, manually
settable switching functions. Each switching function has a floating normally open
relay contact. The relay contacts are accessible at the sensor cable connector
(� � 23).
The threshold values of switching functions A and B can be set within the pressure
range 1×10-9 mbar … 100 mbar via potentiometers "SETPOINT A" and
"SETPOINT B".
The Formula applied to calculate the corresponding threshold voltage
UThreshold depends on the gauge version used (BPG400-SD or
BPG400-SP).
For BPG400-SD:
UThreshold = 0.75 × (log pSetpoint – c) + 7.75
For BPG400-SP:
UThreshold = 0.8129401 × (log pSetpoint – c + 9.30102999)
Constant c is pressure unit dependent (� Appendix A).
Time t
Off
Off
On
Hysteresis
10% U Threshold
Switching function
Threshold
U
Measuring Signal
(Pressure p)
(Setpoint A, B)
Me
ss
ur
ed
val
ue
The hysteresis of the switching functions is 10% of the threshold setting.
The threshold values of the two switching functions "SETPOINT A" and
"SETPOINT B" are set locally on the potentiometers of the gauge that are acces-
sible via the openings on one side of the gauge housing.
��
Voltmeter
��
Ohmmeter or continuity checker
��
Screwdriver, max. ø2.5 mm
4.9 Switching Functions
(BPG400-SD,
BPG400-SP)
4.9.1 Setting the Switching
Functions
Required tools
tina03e1 (0206) BPG400 v1.om
39
The procedure for setting the switching functions is the same for both gauge types.
�
Put the gauge into operation.
�
Connect the �-lead of a voltmeter to the threshold measurement point of
the selected switching function ("Setpoint A" Pin 3, "Setpoint B" Pin 6) and
its �-lead to Pin 5.
max. ø2.5
Setpoint A Pin 3
Setpoint B Pin 6
Pin 5
�
Using a screwdriver (max. ø2.5 mm), set the voltage of the selected
switching function (Setpoint A, B) to the desired value UThreshold.
Setting of the switching functions is now concluded.
There is no local visual indication of the statuses of the switching func-
tions. However, a functional check of the switching functions (On/Off)
can be made with one of the following methods:
��
Reading the status via fieldbus interface � � [1] for BPG400-SD,
� � [2] for BPG400-SP.
��
Measurement of the relay contacts at the sensor cable connector
with a ohmmeter/continuity checker (� � 23).
Procedure
tina03e1 (0206) BPG400 v1.om
5
Deinstallation
DANGER
Caution: contaminated parts
Contaminated parts can be detrimental to health and environment.
Before beginning to work, find out whether any parts are contaminated.
Adhere to the relevant regulations and take the necessary precautions
when handling contaminated parts.
Caution
Caution: vacuum component
Dirt and damages impair the function of the vacuum component.
When handling vacuum components, take appropriate measures to
ensure cleanliness and prevent damages.
�
Vent the vacuum system.
Before taking the gauge out of operation, make sure that this has
no adverse effect on the vacuum system.
Depending on the programming of the superset controller, faults
may occur or error messages may be triggered.
Follow the appropriate shut-down and starting procedures.
�
Take gauge out of operation.
�
Disconnect all cables from the gauge.
�
Remove gauge from the vacuum system.
Procedure
tina03e1 (0206) BPG400 v1.om
41
�
Place the protective lid.
The gauge is now deinstalled.
tina03e1 (0206) BPG400 v1.om
6
Maintenance, Repair
DANGER
Caution: contaminated parts
Contaminated parts can be detrimental to health and environment.
Before beginning to work, find out whether any parts are contaminated.
Adhere to the relevant regulations and take the necessary precautions
when handling contaminated parts.
Small deposits on the electrode system can be removed by baking the anode
(Degas � � 30). In the case of severe contamination, the baffle can be exchanged
easily (� � 19). The sensor itself cannot be cleaned and needs to be replaced in
case of severe contamination (� � 46).
A slightly damp cloth normally suffices for cleaning the outside of the unit. Do not
use any aggressive or scouring cleaning agents.
Make sure that no liquid can penetrate the product. Allow the product to
dry thoroughly before putting it into operation again.
Gauge failures due to contamination are not covered by the warranty.
The gauge is factory-calibrated. Through the use in different climatic conditions,
fitting positions, aging or contamination (� � 30) and after exchanging the sensor
(� � 46) a shifting of the characteristic curve can occur and readjustment can be-
come necessary. Only the Pirani part can be adjusted.
At the push of a button the digital value and thus the analog output are adjusted
electronically to 10 V at atmospheric pressure.
Adjustment is necessary if
��
at atmospheric pressure, the output signal is <10 V
��
the display reads < atmospheric pressure (if the gauge has a display)
��
at atmosphere, the digital value of the RS232C interface is < atmospheric pres-
sure
��
at atmosphere, the digital value received by the bus controller of the fieldbus
gauges (DeviceNet or Profibus) is < atmospheric pressure
��
when the vacuum system is vented, the output voltage reaches 10 V (limited to
10 V by the software) before the measured pressure has reached atmosphere
(gauges with display will show the error "5" at atmospheric pressure (Pirani
sensor warning � � 31))
��
when the vacuum system is vented, the digital value of the RS232C interface
reaches its maximum before the measured pressure has reached atmosphere
��
when the vacuum system is vented, the digital value received by the bus con-
troller of the fieldbus (DeviceNet or Profibus) reaches its maximum before the
measured pressure has reached atmosphere.
6.1 Maintenance
6.1.1 Cleaning the Gauge
6.2 Adjusting the Gauge
6.2.1 Adjustment at
Atmospheric Pressure
tina03e1 (0206) BPG400 v1.om
43
��
Pin approx. ø1.3 × 50 mm (e.g. a bent open paper clip)
Gauges BPG400-SD and -SP are mechanically slightly different from the BPG400.
The adjustment opening of BPG400-SD and -SP is on one side of the gauge
housing. However, the adjustment procedure is the same for all gauge versions.
�
Operate gauge for approx. 10 minutes at atmospheric pressure.
If the gauge was operated before in the BA range, a cooling-
down time of approx. 30 minutes is to be expected (gauge tem-
perature = ambient temperature).
�
Insert the pin through the opening marked and push the
button inside for at least 5 s.
BPG400
BPG400-SD
BPG400-SP
max. ø1.3
Gauges with display will show the
reading "1000 mbar" and the func-
tion "A" when the button has been
pushed for 4 s. Upon completion of
the adjustment, the function
indication "A" disappears.
The gauge is automatically adjusted (�10 s).
The gauge is now adjusted at atmospheric pressure.
A zero point adjustment is recommended
��
after the sensor has been exchanged
��
as part of the usual maintenance work for quality assurance
��
Pin approx. ø1.3 × 50 mm (e.g. a bent open paper clip)
Required tools
Procedure
6.2.2 Zero Point Adjustment
Required tools
tina03e1 (0206) BPG400 v1.om
The push button is also used for the zero point adjustment
(� Illustration in "Adjustment at Atmospheric Pressure").
�
Operate gauge for approx. 10 minutes at a pressure of 1×10-4 mbar.
�
Insert the pin through the opening marked and push the
button inside for at least 2 s.
The adjustment is done automatically and ends after 2 minutes.
The zero point of the gauge is now adjusted.
In the event of a fault or a complete failure of the output signal, the gauge can
easily be checked.
��
Voltmeter / ohmmeter
��
Allen key, size 2.5 mm
��
Spare sensor (if the sensor is faulty)
The output signal is available at the sensor cable connector (Pin 2 and Pin 12).
In case of an error, it may be helpful to just turn off the mains supply and
turn it on again after 5 s.
Problem
Possible cause
Correction
Output signal
permanently �0V
Sensor cable defective or
not correctly connected
Check the sensor cable
No supply voltage
Turn on the power supply
Gauge in an undefined
status
Turn the gauge off and on
again (reset)
Output signal �0.3 V
(Display: error = 8)
Hot cathode error
(sensor faulty)
Replace the sensor
(� � 46)
Output signal �0.5 V
(Display: error = 9)
Pirani error
(sensor defective)
Replace the sensor
(� � 46)
Output signal �0.5 V
Electronics unit not
mounted correctly on
sensor
Check the connection
Display:
Internal data connection
not working
Turn the gauge off and on
again after 5 s
Replace the electronics
unit
Gauge does not switch
over to BA at low pres-
sures
Pirani zero point out of
tolerance
Carry out a zero point
adjustment (� � 43)
Procedure
6.3 What to Do in Case of
Problems
Required tools / material
Troubleshooting (BPG400)
tina03e1 (0206) BPG400 v1.om
45
If the cause of a fault is suspected to be in the sensor, the following checks can be
made with an ohmmeter (the vacuum system need not be vented for this purpose).
Separate the sensor from the electronics unit (� � 15). Using an ohmmeter, make
the following measurements.
Ohmmeter measure-
ment between pins
Possible cause
2 + 4
�37 �
>>37 �
Pirani element 1 broken
4 + 5
�37 �
>>37 �
Pirani element 2 broken
6 + 7
�0.15 �
>>0.15 � Filament of hot cathode broken
4 + 1
�
<<�
Electrode - short circuit to ground
6 + 1
�
<< �
Electrode - short circuit to ground
3 + 1
�
<< �
Electrode - short circuit to ground
9 + 1
�
<< �
Electrode - short circuit to ground
6 + 3
�
<< �
Short circuit between electrodes
9 + 3
�
<< �
Short circuit between electrodes
1
2
3
4
5
6
7
8
9
View on sensor pins
Hot cathode
approx. 0.15 Ohm
Ion collector
Pirani sensor 1
approx. 37 Ohm
Pirani sensor 2
approx. 37 Ohm
6
7
8
2
4
5
3
1
9
Anode
Not connected
GND (connected to sensor housing)
All of the above faults can only be remedied by replacing the sensor (� � 46).
Error diagnosis of fieldbus gauges BPG400-SD and -SP can only be performed as
described above for the basic sensor and sensor electronics. Diagnosis of the
fieldbus interface can only be done via the superset bus controller (� � [1], [2]).
For diagnosis of the BPG400-SD (DeviceNet) gauges, the status lights might pro-
duce some useful information (� � 36).
Troubleshooting (sensor)
Correction
Troubleshooting
(BPG400-SD, BPG400-SP)
tina03e1 (0206) BPG400 v1.om
Replacement is necessary, when
��
the sensor is severely contaminated
��
the sensor is mechanically deformed
��
the sensor is faulty, e.g. filament of hot cathode broken (� � 44)
��
the sensor is faulty, e.g. Pirani element broken (� � 44)
��
Allen key, size 2.5 mm
��
Spare sensor (� � 47)
�
Deinstall the gauge (� � 40).
�
Deinstall the electronics unit from the faulty sensor and mount it to the new
sensor (� � 15).
�
Adjust the gauge (� � 42).
The new sensor is now installed.
6.4 Replacing the Sensor
Required tools / material
Procedure
tina03e1 (0206) BPG400 v1.om
47
7
Options
Part number
24 VDC power supply / RS232C line (� � 26)
353-511
Extension 100 mm (� � 16)
353-510
Baffle DN 25 ISO-KF / DN 40 CF-R (� � 17)
353-512
INFICON Vacuum Gauge Controller VGC103, VGC40x
(� � 20)
1)
Sensor cable BPG400 – VGC103 / VGC40x (� � 20)
1)
Sensor cable with D-Sub female connector for gauge on one
side, no connector on other side
1)
Connection elements, seals, accessories
2)
1)
See INFICON sales literature and data sources for controllers and our range of
sensor cables (� � [10]).
2)
Since there is a vast variety of individual configurations and applications, please
refer to the specific information for your individual configuration (� � [10]).
8
Spare Parts
When ordering spare parts, always indicate:
��
All information on the product nameplate
��
Description and part number
Part number
Sensor BPG400, vacuum connection DN 25 ISO-KF
(including Allen key)
354-490
Sensor BPG400, vacuum connection DN 40 CF-R
(including Allen key)
354-491
9
Storage
Caution
Caution: vacuum component
Inappropriate storage leads to an increase of the desorption rate
and/or may result in mechanical damage of the product.
Cover the vacuum ports of the product with protective lids or grease
free aluminum foil. Do not exceed the admissible storage temperature
range (� � 11).
tina03e1 (0206) BPG400 v1.om
10 Returning the Product
WARNING
Caution: forwarding contaminated products
Contaminated products (e.g. radioactive, toxic, caustic or biological
hazard) can be detrimental to health and environment.
Products returned to INFICON should preferably be free of harmful
substances. Adhere to the forwarding regulations of all involved
countries and forwarding companies and enclose a duly completed
declaration of contamination (� � 53).
Products that are not clearly declared as "free of harmful substances" are decon-
taminated at the expense of the customer.
Products not accompanied by a duly completed declaration of contamination are
returned to the sender at his own expense.
11 Disposal
DANGER
Caution: contaminated parts
Contaminated parts can be detrimental to health and environment.
Before beginning to work, find out whether any parts are contaminated.
Adhere to the relevant regulations and take the necessary precautions
when handling contaminated parts.
N
WARNING
Caution: substances detrimental to the environment
Products or parts thereof (mechanical and electric components, oper-
ating fluids etc.) can be detrimental to the environment.
Dispose of such substances in accordance with the relevant local
regulations.
After disassembling the product, separate its components according to the follow-
ing criteria:
Contaminated components (radioactive, toxic, caustic or biological hazard etc.)
must be decontaminated in accordance with the relevant national regulations,
separated according to their materials, and disposed of.
Such components must be separated according to their materials and recycled.
Separating the components
Contaminated components
Other components
tina03e1 (0206) BPG400 v1.om
49
Appendix
p = 10(U - 7.75) / 0.75 + c
U = 0.75 × (log p - c) + 7.75
where
U
p
c
[V]
[mbar]
0
[V]
[Pa]
2
[V]
[Torr]
-0.125
1E+01
1E–01
1E–03
1E–05
1E–07
1E–09
1E+03
Measuring signal U[V]
0.0
Sensor error
Inadmissible range
Inadmissible range
1E+04
1E+02
1E+00
1E–02
1E–04
1E–06
1E–08
1E–10
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
Pressure p [mbar]
A:
Relationship Output
Signal – Pressure
Conversion formulae
Conversion curve
tina03e1 (0206) BPG400 v1.om
Output signal U
[V]
[mbar]
Pressure p
[Torr]
[Pa]
0.3 / 0.5
Sensor error (� � 44)
0.51 … 0.774
Inadmissible range
0.774
5×10-10
3.75×10-10
5×10-8
1.00
1×10-9
7.5×10-10
1×10-7
1.75
1×10-8
7.5×10-9
1×10-6
2.5
1×10-7
7.5×10-8
1×10-5
3.25
1×10-6
7.5×10-7
1×10-4
4.00
1×10-5
7.5×10-6
1×10-3
4.75
1×10-4
7.5×10-5
1×10-2
5.50
1×10-3
7.5×10-4
1×10-1
6.25
1×10-2
7.5×10-3
1×100
7.00
1×10-1
7.5×10-2
1×101
7.75
1×100
7.5×10-1
1×102
8.50
1×101
7.5×100
1×103
9.25
1×102
7.5×101
1×104
10.00
1×103
7.5×102
1×105
>10.00
Inadmissible range
Pressure indicated (gauge adjusted for air, Pirani-only mode)
102
8
6
4
2
101
8
6
4
2
100
8
6
4
2
10–1
8
6
4
2
10–2
8
6
4
2
10–3
p (mbar)
peff (mbar)
10–3 2
4
6 10–2 2
4
6 10–1 2
4
6 100
2
4
6 101
2
4
6 102
H2 He
Ne
Xe
Ar
Freon 12
H2O
vapor
Air
O2
CO
N2
Kr
Indication range
above 10-2 mbar)
CO2
Conversion table
B:
Gas Type Dependence
Indication range
above 10-2 mbar
tina03e1 (0206) BPG400 v1.om
51
The gas type dependence in the pressure range 10-2 … 1 mbar can be compen-
sated by means of the following formula:
peff = C × indicated pressure
where
Gas type
Calibration factor C
Air, O2, CO
1.0
N2
0.9
CO2
0.5
Water vapor
0.7
Freon 12
1.0
H2
0.5
He
0.8
Ne
1.4
Ar
1.7
Kr
2.4
Xe
3.0
(The above calibration factors are mean values.)
The gas type dependence in the pressure range <10-3 mbar can be compensated
by means of the following formula (gauge adjusted for air):
peff = C × indicated pressure
where
Gas type
Calibration factor C
Air, O2, CO, N2
1.0
N2
1.0
He
5.9
Ne
4.1
H2
2.4
Ar
0.8
Kr
0.5
Xe
0.4
(The above calibration factors are mean values.)
A mixture of gases and vapors is often involved. In this case, accurate
determination is only possible with a partial-pressure measuring instru-
ment.
Calibration in pressure range
10-2 … 1 mbar
Calibration in pressure range
<10-3 mbar
tina03e1 (0206) BPG400 v1.om
� [1]
www.inficon.com
Communication Protocol
DeviceNet™ BPG400-SD
tira03e1
INFICON AG, LI–9496 Balzers, Liechtenstein
� [2]
www.inficon.com
Communication Protocol
Profibus BPG400-SP
tira36e1
INFICON AG, LI–9496 Balzers, Liechtenstein
� [3]
www.inficon.com
Product descriptions and downloads
INFICON AG, LI–9496 Balzers, Liechtenstein
� [4]
www.odva.org
Open DeviceNet Vendor Association, Inc.
DeviceNet™ Specifications
� [5]
www.profibus.com
Profibus user organization
� [6]
European Standard for DeviceNet EN 50325
� [7]
European Standard for Profibus EN 50170
� [8]
www.inficon.com
Instruction Sheet
BPG400 (all versions)
tima03e1 a
INFICON AG, LI–9496 Balzers, Liechtenstein
� [9]
www.inficon.com
Instruction Sheet
BPG400-SD, BPG400-SP
tima36e1
INFICON AG, LI–9496 Balzers, Liechtenstein
� [10]
www.inficon.com
INFICON AG, LI–9496 Balzers, Liechtenstein
C:
Literature
tina03e1 (0206) BPG400 v1.om
53
Declaration of Contamination
Legally binding declaration:
We hereby declare that the information on this form is complete and accurate and that we will assume any further costs that may
arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address
Post code, place
Phone
Fax
E-mail
Name
Date and legally binding signature
Company stamp
1) or not containing any amount
of hazardous residues that
exceed the permissible ex-
posure limits
� Process related contamination of product
toxic
no � 1)
yes �
caustic
no � 1)
yes �
biological hazard
no �
yes � 2)
explosive
no �
yes � 2)
radioactive
no �
yes � 2)
other harmful substances no � 1)
yes �
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has
been submitted. Non-completion will result in delay.
This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies:
Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
��
�
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Manufacturer
Chemical name
(preferably with formula)
Precautions associated
with substance
Action in case of human
contact
Description of product
Type
Part number
Serial number
Reason for return
� Operating fluid(s) used (Must be drained before shipping.)
The product is free of any sub-
stances which are damaging to
health yes �
This form can be downloaded
from our website.
2) Products thus contami-
nated will not be ac-
cepted without written
evidence of decontami-
nation!
+423 / 388 3111
Fax +423 / 388 3700
Original: German tina03d1 (0206)
[email protected]
t i na03e1
www.inficon.com