Text of the manual
Vacuum Gauge Controller
TWO
DISPLAY
THREE
DISPLAY
Instruction Manual GA09037_002_C0
GA09037_002_C0 – 09/2016
3
0.
Table of Contents
0.
Table of Contents
3
0.1
Table of Figures
6
0.2
Table of Tables
6
1.
Description
7
1.1
Validity
7
1.1.1
Part Number
7
1.1.2
Nameplate
7
1.2
Conforming Utilisation
7
1.3
Instrument Versions
8
1.4
Responsibility and Warranty
8
1.5
Shipping Damage
8
2.
Safety
9
2.1
General Information
9
2.2
Key to the Symbols
9
2.3
Basic Safety Information
9
3.
General Equipment Description
10
3.1
Multichannel Gauge
10
3.2
Suitable Sensors
10
4.
Technical Data
11
4.1
General Data
11
4.1.1
Mechanical Data
11
4.1.2
Standard Parameters (factory defaults)
12
4.1.3
Enviroments
12
4.1.4
Standards
12
4.2
Mains Connection
12
4.3
Measurement Channels
13
4.3.1
Sensor Powering
13
4.3.2
Measurement Technology
13
4.4
Switching Functions
14
4.4.1
Switching Function Relay
14
4.4.2
Ready Signal Relay
14
4.5
Outputs and Inputs
14
4.5.1
Analog Output
14
4.5.2
Control Inputs
14
GA09037_002_C0 – 09/2016
5.
Installation
15
5.1
Supplied Equipment
15
5.2
Mechanical Installation
15
5.2.1
Rack Installation
16
5.2.2
Front Panel Installation
16
5.2.3
Benchtop Instrument
17
5.3
Connections
18
5.3.1
Rear Side of the Instrument
18
5.3.2
Mains Connection
18
5.3.3
Earthing
19
5.3.4
Sensor
19
5.3.5
Relay
19
5.3.6
HV Control
20
5.3.7
Analog Output
20
6.
Operation
21
6.1
Front Panel
21
6.1.1
Display
21
6.1.2
Control Pushbuttons
22
6.2
Switching on and off
22
6.2.1
Switching on
22
6.2.2
Switching off
22
6.3
Operating Modes
23
6.4
Measurement Mode
23
6.4.1
Selection
23
6.4.2
Description
23
6.4.3
Pushbutton Functions
24
6.4.3.1
Measurement Channel Selection
24
6.4.3.2
Parameter Mode Selection
24
6.4.3.3
Switching on the High Vacuum Measurement Circuit
24
6.4.3.4
Switching off the High Vacuum Measurement Circuit
25
6.4.3.5
Sensor Identification
25
6.5
Parameter Mode
26
6.5.1
Selection
26
6.5.2
Parameter Groups
26
6.5.3
Operating Concept
27
7.
Parameters
28
7.1
Switching Function Parameters (PArA SP)
28
7.1.1
Basic Terms
28
7.1.2
Configuring the Switching Functions
29
7.1.3
Adjustment Range
29
7.2
Sensor Parameters (PArA SEn)
30
7.2.1
Pirani Range Extension (PrE)
30
7.2.2
Measured Value Filter (FiLt)
30
7.2.3
Ready Function (rAEdY)
31
7.2.4
Type of Gas Correction (Cor)
31
7.2.5
Sensor Switch-on type (S-on)
31
7.2.6
Sensor Switch-on Value (t-on)
32
7.2.7
Sensor Switch-off type (S-oFF)
32
7.2.8
Sensor Switch-off Value (t-off)
33
GA09037_002_C0 – 09/2016
5
7.3
General Parameters (PArA GEn)
33
7.3.1
Unit of Measurement (unit)
33
7.3.2
Display Format (diGit)
33
7.3.3
Display Brightness (bri)
34
7.3.4
Analog Output (AnALoG)
34
8.
Maintenance and Service
35
8.1
Maintenance
35
8.1.1
General Maintenance Information
35
8.2
Troubleshooting
35
8.2.1
Fault Indication
35
8.2.2
Error Messages
35
8.2.3
Help the Case of Faults
36
8.2.4
Fuse Replacement
36
8.2.5
Repair
36
9.
Storing and Waste Disposal
37
9.1
Packaging
37
9.2
Shelving
37
9.3
Waste Disposal
37
Annex 1 - EU Declaration of Conformity
38
Annex 2 - Declaration of Contamination
39
GA09037_002_C0 – 09/2016
0.1
Table of Figures
Figure 1 – Nameplate (Examble)
7
Figure 2 – Multichannel gauge dimensions (in mm)
11
Figure 3 – Rack installation
16
Figure 4 – Front panel cutout dimensions (in mm)
16
Figure 5 – Preparing the top side of the instrument for utilisation as a benchtop unit
17
Figure 6 – Preparing the bottom side of the
17
Figure 7 – Rear panel of the instrument
18
Figure 8 – Connector Sensor (modular socket, 8-way)
19
Figure 9 – Connector Relay (plug-in terminals, screw terminals)
19
Figure 10 – Connector HV-Control (plug-in terminals, screw terminals)
20
Figure 11 – Connector Analog Output (plug-in terminals, screw terminals)
20
Figure 12 – Front panel
21
Figure 13 – Operating the pushbutton CHANNEL
24
Figure 14 – Operating the pushbutton PARA
24
Figure 15 – Operating the pushbutton UP
24
Figure 16 – Operating the pushbutton DOWN
25
Figure 17 – Operating the pushbuttons UP and DOWN
25
Figure 18 – Behaviour of a switching function in response to pressure changes
29
0.2
Table of Tables
Table 1 – Part numbers
7
Table 2 – Suitable sensors
10
Table 3 – Default parameters (factory defaults)
12
Table 4 – Supplied equipment
15
Table 5 – Display components
21
Table 6 – Display in the measurement mode
23
Table 7 – Sensor identification
25
Table 8 – Parameter groups and corresponding parameters
26
Table 9 – Available sensor parameters
30
Table 10 – Values for the parameter PrE
30
Table 11 – Values for the parameter rEAdY
31
Table 12 – Values for the parameter unit
33
Table 13 – Values for the parameter diGit
33
Table 14 – Values for the parameter bri
34
Table 15 – Values for the parameter AnA
34
Table 16 – Error messages
35
GA09037_002_C0 – 09/2016
7
1.
Description
1.1
Validity
1.1.1
Part Number
This document applies to the following products:
Part Number
Product
Version
Serial
Number
230024
DISPLAY TWO
Two-channel measuring instrument for active vacuum sensors
2.1
et seq.
1000
et seq.
230025
DISPLAY THREE
Three-channel measuring instrument for active vacuum sensors
2.1
et seq.
1000
et seq.
Table 1 – Part numbers
1.1.2
Nameplate
A nameplate is located on the bottom side of the instrument. When communicating with the
Leybold GmbH, stating the information on the nameplate is necessary. For this reason
please enter the following information:
Figure 1 – Nameplate (Examble)
1.2
Conforming Utilisation
The DISPLAY TWO and DISPLAY THREE is a display and operating unit for sensors with
an analogue interface of the company Leybold GmbH.
Depending on the version, the unit offers two or three channels, and it is used in combination
with the sensors from the series THERMOVAC and PENNINGVAC as well as DU sensors
for the measurement of pressures above and below atmospheric pressures (vacuum).
The DISPLAY TWO respectively DISPLAY THREE multichannel vacuum gauge is called in
the following "multichannel gauge" for short.
Operate all connected sensors in agreement with the information given in the corresponding
Operating Instructions.
NOTICE:
Based on the technical data please check first whether your measuring
instrument is suited to your kind of application.
GA09037_002_C0 – 09/2016
1.3
Instrument Versions
The multichannel gauge is available in two different versions:
DISPLAY TWO
(two-channel measuring instrument)
DISPLAY THREE (three-channel measuring instrument)
The two versions differ as to the following:
Number of measurement channels
Number of switching functions
Power consumption
Weight
Chapter 4 - Technical Data, page 11
Described in these Operating Instructions are both the DISPLAY TWO as well as the
DISPLAY THREE.
1.4
Responsibility and Warranty
Leybold GmbH will not assume any responsibility or warranty in case the operator or third
persons
do not observe the information given in this document.
do not use the product as intended.
modify the product in any way (conversions, repair work etc).
operate the product with accessories not listed in the corresponding product
documentation.
Subject to technical alterations without prior notice. The figures are not binding.
1.5
Shipping Damage
Examine the shipping package as to any external damage.
In case any damage is determined, file a damage report to the forwarding agent and
the insurer.
Retain the packaging material since damages can only be claimed when returning the
instrument in the original packaging of the manufacturer.
Examine the delivery to ensure that it is complete.
Examine the instrument as to any visually apparent damage.
DANGER: Damaged product.
Commissioning or operating a damaged product is dangerous to life.
GA09037_002_C0 – 09/2016
9
2.
Safety
2.1
General Information
The multichannel gauge is delivered ready for operation. Even so, we recommend that you
carefully read these Operating Instructions so as to ensure optimum operating conditions
right from the start.
This manual contains important information for understanding, installing, commissioning,
operating and troubleshooting the multichannel gauge.
2.2
Key to the Symbols
Important instructions relating to technical safety and safe operation are emphasised by
symbols.
DANGER or WARNING:
Information designed to prevent any kind of injury to persons.
DANGER:
Information designed to prevent injury to persons and damage to equipment
in connection with electricity.
NOTICE:
General information pointing to further information, respectively reference
sections.
2.3
Basic Safety Information
During all work like installation, maintenance and repair activities, comply with the
pertinent safety regulations.
DANGER: Mains Voltage
Coming into contact with components at mains voltage level within the
instrument can be dangerous to life when inserting objects or allowing liquids
to enter the instrument.
WARNING: Improper usage.
Improper usage can damage the instrument. Use the instrument only in
agreement with the specifications issued by the manufacturer.
WARNING: Wrong Connection and Operating Data.
Wrong connection and operating data can damage the instrument. Comply
with all specified connection and operating data.
GA09037_002_C0 – 09/2016
3.
General Equipment Description
3.1
Multichannel Gauge
The DISPLAY TWO and DISPLAY THREE is a display and operating unit for sensors with
an analogue of the company Leybold GmbH.
Depending on the version, the unit offers two or three channels, and it is used in combination
with the sensors from the series THERMOVAC and PENNINGVAC as well as DU sensors
for the measurement of pressures above and below atmospheric pressures (vacuum).
Operate all connected sensors in agreement with the information given in the corresponding
Operating Instructions.
3.2
Suitable Sensors
The following sensors can be operated with the multichannel gauge:
Sensor
Type
Display
THERMOVAC
TTR81N
TTR90 / TTR91 / TTR91N
TTR96S / TTR96SN
TTR211 / TTR216S
TTR911 / TTR911N
TTR916 / TTR916N
()
THERMOVAC
TTR100 / TTR100S2
TTR101 / TTR101N
TTR101S2 / TTR101S2N
()
PENNINGVAC
PTR81N
PTR225 / PTR225N
PTR225S / PTR225SN
PTR237 / PTR237N
PENNINGVAC
PTR82N
PTR90 / PTR90N
DU sensor
DU200 / DU201
DU2000 / DU2001
DU relative pressure sensor
DU2001 rel.
Table 2 – Suitable sensors
GA09037_002_C0 – 09/2016
11
4.
Technical Data
4.1
General Data
4.1.1
Mechanical Data
Dimensions:
Width: 106.4 mm (1/4 19“)
Height: 128.4 mm (3 HU)
Depth: 174.5 mm
Weight:
DISPLAY TWO
≤ 1.3 kg
DISPLAY THREE ≤ 1.4 kg
Installation depth:
≤ 230 mm approx. (including connected plugs)
Usage:
Rack installation
Front panel installation
Benchtop instrument
Figure 2 – Multichannel gauge dimensions (in mm)
GA09037_002_C0 – 09/2016
4.1.2
Standard Parameters (factory defaults)
Parameter
Parameter description
Setting
Pirani range extension
Measured value filter
Function Ready signal
Gas correction
.
Sensor switch-on mode
Sensor switch-off mode
Measurement unit
Display format
Display brightness
Analog output
Table 3 – Default parameters (factory defaults)
4.1.3
Enviroments
Temperature:
Storage:
-20 – +60 °C
Operating: +5 – +50 °C (sea level)
+5 – +30 °C (2000 m above sea level)
Relative atmospheric humidity: 80 % max. (up to 30 °C)
decreasing to 50 % max. (over 40 °C)
Use:
indoors (altitude 2000 m max. above sea level)
Ingress protection type:
IP20
4.1.4
Standards
Conformity with respect to Low Voltage Directive 2014/35/EU
Conformity with respect to EMC Directive 2014/30/EU
Conformity with respect to RoHS Directive 2011/65/EU
International/national standards as well as specifications:
DIN EN 61010-1 (2011)
(Safety requirements for electrical equipment for measurement, control and
laboratory use).
DIN EN 61326-1 (2013)
(Electrical equipment for measurement, control and laboratory use – EMC
requirements. Industrial interference immunity; electromagnetic emissions household
sector Class B).
4.2
Mains Connection
Voltage:
100 – 240 V AC
Frequency:
50/60 Hz
Fuses:
2 x T1.6A H
Power consumption:
DISPLAY TWO
< 10 W
DISPLAY THREE < 15 W
GA09037_002_C0 – 09/2016
13
Current consumption:
0.3 – 0.15 A approx.
Overvoltage category:
II
Protection class:
1
Connection:
Cold-device plug IEC 320 C14
4.3
Measurement Channels
Number:
DISPLAY TWO
2
DISPLAY THREE 3
Connection:
RJ45 (FCC 68)
Suitable sensors:
THERMOVAC
TTR81N
TTR90 / TTR91 / TTR91N
TTR96S / TTR96SN
TTR211 / TTR216S
TTR911 / TTR911N
TTR916 / TTR916N
THERMOVAC
TTR100 / TTR100S2
TTR101 / TTR101N
TTR101S2 / TTR101S2N
PENNINGVAC
PTR81N
PTR225 / PTR225N
PTR225S / PTR225SN
PTR237 / PTR237N
PENNINGVAC
PTR82N
PTR90 / PTR90N
DU sensor
DU200 / DU201
DU2000 / DU2001
DU relative
pressure sensor
DU2001 rel.
4.3.1
Sensor Powering
Voltage:
+24 VDC ±5 %
Current:
100 mA approx.
Fusing:
200 mA, self resetting
Power feeding complies with the requirements of a safety extra-low voltage (SELV-E in
accordance with EN 61010).
4.3.2
Measurement Technology
Measurement ranges:
Sensor dependent
Measurement error:
Gain error
≤ 0.1 % FS
Offset error ≤ 0.02 % FS
Measurement rate:
1000 s-1
Display rate:
4 s-1
Filter time constant:
Slow … Fast
15 … 7 … 3 … 1
Units of measurement:
mbar, Pa, Torr
Correction options:
Correction factor
A/D converter resolution:
> 15 Bit
GA09037_002_C0 – 09/2016
4.4
Switching Functions
Number of switching functions:
DISPLAY TWO
4
DISPLAY THREE 6
Assignment:
2 per channel
Response time:
< 50 ms
Adjustment range:
Sensor dependent
Hysteresis:
adjustable ≥ 10 % of measurement value
4.4.1
Switching Function Relay
Number:
1 per channel
Type of contact:
Changeover contact, floating
Load (resistive):
Switched current:
1 A approx.
Switched voltage: 30 VAC / 30 VDC approx.
Service life:
Mechanical: 5·106 actuations
Electrical:
105 actuations at maximum load
Connection:
Plug-in terminals, screw terminals
4.4.2
Ready Signal Relay
Number:
1 per channel, as the 2. switching function configurable
Type of contact:
Changeover contact, floating
Load (resistive):
Switched current:
1 A approx.
Switched voltage: 30 VAC / 30 VDC approx.
Service life:
Mechanical: 5·106 actuations
Electrical:
105 actuations at maximum load
Connection:
Plug-in terminals, screw terminals
4.5
Outputs and Inputs
4.5.1
Analog Output
Number:
1 per channel
Voltage range:
0 – 10 VDC / 0 – 5 VDC selectable
Deviation from the displ. value:
± 0.1 % FS
Internal resistance:
100 Ω
Relationship between
voltage and pressure:
Sensor dependent
Connections:
Plug-in terminals, screw terminals
4.5.2
Control Inputs
Number:
1 per channel
Voltage range:
0 – 24 VDC
Connections:
Plug-in terminals, screw terminals
GA09037_002_C0 – 09/2016
15
5.
Installation
5.1
Supplied Equipment
Designation
Quantity
Multichannel Gauge
1
Mains cord with shockproof plug (EU)
1
Mains cord with shockproof plug (US)
1
Operating Instructions (each EN and DE)
1
Spare fuse
2
Neck collar screw
4
Plastic sleeve
4
Edge protection rubber
2
Adhesive feet
2
Mating connectors
DISPLAY TWO
4
DISPLAY THREE
5
Table 4 – Supplied equipment
5.2
Mechanical Installation
The multichannel gauge can be used as follows:
Rack installation
Front panel installation
Benchtop instrument
WARNING: Powering down
Install the instrument or place it so that you are in a position to operate the
mains power switch at any time or ensure that the instrument can be
deenergised at any time.
GA09037_002_C0 – 09/2016
5.2.1
Rack Installation
The multichannel gauge has been designed for installation within a sub-rack (19", 3 U) in
accordance with DIN EN 60297 (IEC 60297) ( Figure 3, page 16). For this purpose the
supplied equipment includes four neck collar screws and four plastic sleeves.
Figure 3 – Rack installation
Attach the sub-rack within the rack.
Push the multichannel gauge into the sub-rack.
Affix the instrument to the sub-rack with the neck collar screws and the plastic sleeves
included in the delivery.
5.2.2
Front Panel Installation
For panel mounting of the instrument, the following cutout ( Figure 4, page 16) is
required:
Figure 4 – Front panel cutout dimensions (in mm)
Guide the multichannel gauge into the cut-out.
Affix the instrument with the neck collar screws and the plastic sleeves included in the
delivery.
GA09037_002_C0 – 09/2016
17
5.2.3
Benchtop Instrument
When planning to use the multichannel gauge as a benchtop instrument, proceed as
follows:
Push one of the two edge protection rubber pieces included in the delivery over the top
edge of the front panel ( Figure 5, page 17)
Place the multichannel gauge on its back ( Figure 6, page 17)
Push the second edge protection rubber piece included in the delivery onto the bottom
edge of the front panel
WARNING: Risk of suffering injury.
When using the multichannel gauge as a benchtop instrument fit the two edge
protection rubber pieces onto the top and bottom edge of the front panel so
as to avoid injury by sharp edges.
Stick the two rubber feet included in the delivery onto the bottom of the housing.
Figure 5 – Preparing the top side of the instrument
for utilisation as a benchtop unit
Figure 6 –
Preparing the bottom side of the
instrument for utilisation as a benchtop
unit
Turn the multichannel gauge over again and move it to the desired place.
GA09037_002_C0 – 09/2016
5.3
Connections
5.3.1
Rear Side of the Instrument
The Figure 7, page 18 depicts the rear panel of the multichannel gauge. The way in which
the individual connections have been wired is described in the following sections.
Figure 7 – Rear panel of the instrument
A
Mains switch
B
Ground screw
C
Mains connection
D
Connection Relay
E
Connection for sensor, measurement channel 3
F
Connection for sensor, measurement channel 2
G
Connection for sensor, measurement channel 1
H
Connection Analog Output
I
Connection HV Control
5.3.2
Mains Connection
The mains connection on the rear panel ( Figure 7, C, page 18) has been designed to
accept a mains cord which on the instrument side is equipped with a cold-device plug.
NOTICE: Mains cord
Included in the delivery of the instrument is a mains cord. If the plug on the
mains power side is not compatible with your mains power outlets, you will
need a mains cord which meets the following specifications:
Three-wire cable with protective earthing.
Conductor cross-section: 3 x 0.75 mm² or greater.
Cable length 2.5 m maximum.
DANGER: Mains voltage
Appliances, which have not been professionally connected to Earth, can be
life-threatening in the event of a malfunction. For this reason use three-wire
mains cords, respectively extension cords with protective earthing only. Insert
the mains plug into a mains power socket, which provides an Earth contact.
Insert the plug of the mains cord into the mains socket provided on the instrument.
Insert the mains plug of the mains cord into the mains outlet.
GA09037_002_C0 – 09/2016
19
5.3.3
Earthing
With the aid of the earthing screw ( Figure 7, B, page 18) the multichannel gauge can
be connected to the protective ground of the vacuum chamber.
NOTICE: Earthing
Connect the Earth connection on the vacuum chamber by means of a
protective earth conductor to the earthing screw on the instrument
5.3.4
Sensor
The connection marked Sensor serves the purpose of connecting sensors. For each
measurement channel an 8-way modular socket is available. ( Figure 7, E, F, G, page
18 and Figure 8, page 19).
Figure 8 – Connector Sensor (modular socket, 8-way)
1
+24 V DC
2
Power ground
3
Signal
4
Ident resistant
5
Signal ground
6
Status (for PTR)
7
HV on (for PTR)
8
not available
CAUTION: Impermissible Sensors.
Connecting sensors which have not been designed to be operated in
connection with the multichannel gauge or which do not comply with current
EMC guidelines can impair operation of the instrument or even damage it.
Always operate the multichannel gauge with approved sensors. Chapter
3.2 Suitable Sensors, page 10 .
Connecting:
Measurement channel 1, 2 and/or 3: Connect of the sensor using a shielded straight
through cable to the connection CH1, CH2 and/or CH3.
5.3.5
Relay
The switching functions and the ready monitor influence the switching action of various
relays within the multichannel gauge. Through the connection marked Relay ( Figure
7, D, page 18 and Figure 9, page 19) you may utilise the relay contacts for switching
purposes. The relay contacts are floating.
Figure 9 – Connector Relay (plug-in terminals, screw terminals)
1
SP 1 Normally closed contact (NC)
2
SP 1 Common contact (COM)
3
SP 1 Normaly open contact (NO)
4
SP 2 Normally closed contact (NC)
5
SP 2 Common contact (COM)
6
SP 2 Normaly open contact (NO) or Ready contact
DANGER: Dangerous voltage
Voltages exceeding 60 VDC or 30 VAC are dangerous when touched. You
may only switch at the connector marked Relay Output voltages of 30 VDC or
30 VAC with a maximum current of 1 A. The voltage must comply with the
requirements of a safety extra-low voltage (SELV-E in accordance with EN
61010).
GA09037_002_C0 – 09/2016
5.3.6
HV Control
The connection HV Control ( Figure 7, I, page 18 and Figure 10, page 20) provides the
following connections:
HV ON. Here the high vacuum circuit of the PENNINGVAC sensors PTR81N, PTR225,
PTR225N, PTR225S, PTR225SN, PTR237 and PTR237N can be switched on and off.
The following applies to the signal level: On = +12 – 24 VDC. Off = 0 VDC.
CH2 CH3
CH1
HV Control
1 2 3 4 5 6
Figure 10 – Connector HV-Control (plug-in terminals, screw terminals)
1
Measurement channel 1 -
2
Measurement channel 1 +12 – 24 VDC
3
Measurement channel 2 -
4
Measurement channel 2 +12 – 24 VDC
5
Measurement channel 3 -
6
Measurement channel 3 +12 – 24 VDC
Connecting:
Connect the peripheral components using a shielded connecting cable to the connector
marked HV Control on the rear of the multichannel gauge.
5.3.7
Analog Output
The connection marked Analog Output ( Figure 7, H, page 18 and Figure 11 , page 20)
provides the following connections:
Analog outputs for the signals of the individual measurement channels.
Analog Output
CH2
CH3
CH1
1 2 3 4 5 6
Figure 11 – Connector Analog Output (plug-in terminals, screw terminals)
1
Analog output of measurement channel 1
2
Ground of measurement channel 1
3
Analog output of measurement channel 2
4
Ground of measurement channel 2
5
Analog output of measurement channel 3
6
Ground of measurement channel 3
Connecting:
Connect the peripheral components using a shielded connecting cable to the connector
marked Analog Output on the rear of the multichannel gauge.
GA09037_002_C0 – 09/2016
21
6.
Operation
6.1
Front Panel
Shown in Figure 12, page 21 is the front panel of the DISPLAY THREE. In the case of the
DISPLAY TWO the display for measurement channel 3 (Item C) are not present.
Figure 12 – Front panel
A
Display of channel 1 with two
corresponding switching thresholds
B
Display of channel 2 with two
corresponding switching thresholds
C
Display of channel 3 with two
corresponding switching thresholds
D
Status indication operating pushbuttons
6.1.1
Display
A separate display area is provided for each measurement channel ( Figure 12, A, B,
C, page 21). In the display area, the following information is shown from left to right:
Display
Description
......
Measured value or status message
SP1, SP2
Switching function status When the symbol is lit, the
pressure is then below the lower threshold When the
symbol is not lit, then the pressure is above the
upper threshold
mbar, Pa, Torr
Pressure unit (applies to all channels)
Para
Channel selection, configuration mode for channel
Table 5 – Display components
GA09037_002_C0 – 09/2016
6.1.2
Control Pushbuttons
CH
Through the key marked CH you can select a measurement channel. This is necessary, for
example, when wanting to switch a certain sensor on or off, or when wanting to change the
sensor parameters. The symbol Para for the selected measurement channel flashes for 10
seconds.
PARA
Through the pushbutton PARA you can invoke the parameter mode for switching threshold,
sensor and general parameters. The indicator Para for the selected measurement channel
comes on. You can set up different parameters.
Arrow pushbuttons ( UP)
The arrow pushbuttons are needed so as to be able to enter data in the parameter mode or
switch the PENNINGVAC sensors PTR81N, PTR225, PTR225N, PTR225S, PTR225SN,
PTR237 and PTR237N on or off ( Chapter 6.4.3.3 Switching on the High Vacuum
Measurement Circuit, page 24 and chapter 6.4.3.4 Switching off the High Vacuum
Measurement Circuit, page 25). By operating the pushbuttons, a default value can be
reduced or increased. The corresponding pushbuttons are designated in the following as
DOWN ().
6.2
Switching on and off
6.2.1
Switching on
Switch the instrument on through its mains switch.
After switching on, the multichannel gauge runs the following:
Self test
Display test
Display of the software version used
Re-establishing the parameters set up last
Identification of the connected measuring equipment
Activation of the measurement mode
6.2.2
Switching off
Switch the instrument off through its mains switch.
CAUTION: Waiting Time
Wait for at least five seconds before switching the instrument on again.
GA09037_002_C0 – 09/2016
23
6.3
Operating Modes
The multichannel gauge may be operated in one of the following operating modes:
Measurement Mode
The measurement mode is the standard operating mode. Here the measured values or
status messages are displayed. Chapter 6.4 Measurement Mode, page 23
Parameter Mode
In the parameter mode you have access to different parameters. You can simply view the
parameters all change them with the aid of the arrow keys. In this way you can configure
the multichannel gauge. Chapter 6.5 Parameter Mode, page 26
6.4
Measurement Mode
6.4.1
Selection
After switching on the multichannel gauge it will automatically run the measurement mode.
If running the parameter mode and when not operating a key for 10 seconds the instrument
will then automatically revert back to the measurement mode.
6.4.2
Description
In the measurement mode the measured values of the sensors are displayed. When the
pressure is outside the permissible range, then a status message is displayed.
Measurement channels to which no gauge head has been connected will indicate noSEn.
This status message is erased after 30 seconds and 4 dots are displayed.
Display
Description
or .
Above the permissible range (DU Sensors excluded)
Above the measurement range (DU Sensors only)
.
Within the permissible range
.
Below the permissible range (DU Sensors excluded)
PENNINGVAC sensor PTR81N, PTR225, PTR225N,
PTR225S, PTR225SN, PTR237 or PTR237N has been
switched off
PENNINGVAC sensor PTR81N, PTR225, PTR225N,
PTR225S, PTR225SN, PTR237 or PTR237N is on
(Switching-on procedure, still no valid measured value
available)
Lowly below the measurement range (DU Sensors only)
Middle below the measurement range (DU Sensors only)
Large below the measurement range (DU Sensors only)
Table 6 – Display in the measurement mode
GA09037_002_C0 – 09/2016
6.4.3
Pushbutton Functions
6.4.3.1 Measurement Channel Selection
Press the pushbutton CH.
The instrument selects the next measurement channel. The display of Para for the
selected measurement channel will flash for 10 seconds.
Figure 13 – Operating the pushbutton CHANNEL
6.4.3.2 Parameter Mode Selection
Press the pushbutton PARA and keep it depressed for approximately 2 seconds. The
instrument will change to the parameter mode ( Chapter 6.5 Parameter Mode,
page 26).
When not operating any pushbutton within 10 seconds, the instrument will then
automatically return to the measurement mode.
Figure 14 – Operating the pushbutton PARA
6.4.3.3 Switching on the High Vacuum Measurement Circuit
The high vacuum measurement circuit of the following sensors can be switched on
manually: PENNINGVAC sensors PTR81N, PTR225, PTR225N, PTR225S, PTR225SN,
PTR237 and PTR237N. For this the sensor control must be set to HAnd ( Chapter 7.2.5
Sensor Switch-on Type (S-on), page 31).
Keep the key marked CH pressed so as to select the respective measurement channel.
Keep the key marked UP pressed for approximately two seconds.
The sensor for the selected measurement channel is switched on. Either the
measured value or a status message will be displayed. Table 6, page 23.
Figure 15 – Operating the pushbutton UP
GA09037_002_C0 – 09/2016
25
6.4.3.4 Switching off the High Vacuum Measurement Circuit
The high vacuum measurement circuit of the following sensors can be switched off
manually: PENNINGVAC sensors PTR81N, PTR225, PTR225N, PTR225S, PTR225SN,
PTR237 and PTR237N. For this in the sensor control must be set to HAnd ( Chapter
7.2.7 Sensor Switch-off Type (S-oFF), page 32).
Keep the key marked CH pressed so as to select the respective measurement channel.
Keep the key marked DOWN pressed for approximately two seconds.
The sensor for the selected measurement channel is switched off. The display will
indicate the status oFF.
Figure 16 – Operating the pushbutton DOWN
6.4.3.5 Sensor Identification
Keep the keys UP and DOWN pressed for approximately one second.
On the displays for the individual measurement channels the in each case connected
sensors are indicated. Table 7, page 25.
Figure 17 – Operating the pushbuttons UP and
DOWN
Display
Sensor type
()
THERMOVAC
(TTR81N, TTR90, TTR91, TTR91N,TTR96S, TTR96SN, TTR211, TTR216S,
TTR911, TTR911N, TTR916, TTR916N)
()
THERMOVAC
(TTR100, TTR100S2, TTR101, TTR101N, TTR101S2, TTR101S2N)
PENNINGVAC
(PTR81N, PTR225, PTR225N, PTR225S, PTR225SN, PTR237, PTR237N)
PENNINGVAC
(PTR82N, PTR90, PTR90N)
DU sensor
(DU200, DU201)
DU sensor
(DU2000, DU2001)
DU relative pressure sensor
(DU2001 rel.)
No sensor connected (no sensor). The indication will disappear after 30 seconds.
. . . .
No sensor connected (no sensor).
Table 7 – Sensor identification
GA09037_002_C0 – 09/2016
6.5
Parameter Mode
6.5.1
Selection
By operating the PARA pushbutton for approximately 2 seconds the instrument changes
from the measurement mode to the parameter mode. The Para indicator comes on for the
channel selected in each case. When the instrument is running in the parameter mode and
if no pushbutton is operated for 10 seconds, then the instrument will automatically return
back to the measurement mode. The Para indicator for the channel which was selected in
each case is turned off.
6.5.2
Parameter Groups
In the parameter mode you have access to different parameters. You may view these
parameters or change these with the aid of the arrow pushbuttons. In this way you can
configure the instrument. Table 8, page 26 depicts all available parameters.
Parameter group
Parameter
(only with deselected ready function)
(only with deselected ready function)
Table 8 – Parameter groups and corresponding parameters
The available parameters have been divided into the following parameter sets:
Switching function parameters (PArA SP)
These parameters affect only the sensor of the selected measurement channel. Per
measurement channel, two switching functions are available. Chapter 7.1 Switching
Function Parameters (PArA SP), page 28.
Sensor parameters (PArA SEn)
These parameters affect only the selected measurement channel. For each measurement
channel a separate set of parameters is available which will depend on the type of connected
sensor. Chapter 7.2 Sensor Parameters (PArA SEn), page 29.
General parameters (PArA GEn)
With the aid of these parameters you can generally configure the instrument. The parameter
is applied to all measurement channels. Chapter 7.3 General Parameters (PArA GEn),
Page 33.
GA09037_002_C0 – 09/2016
27
6.5.3
Operating Concept
From the measurement mode, you can select and change a certain parameter as follows:
Press the pushbutton CH, so as to select the desired measurement channel (only
necessary when wanting to change a sensor parameter)
The Para status indicator flashes for the selected channel.
Press the PARA pushbutton for about 2 seconds.
You have now invoked the parameter menu.
Use the arrow pushbuttons to select the desired parameter group.
The name of the parameter group is displayed.
Press the PARA pushbutton to select the desired parameter.
The name and the value of the parameter are displayed.
Use the arrow pushbuttons and to change the value of the parameter.
The value of the parameter is changed.
Repeat the last 2 steps so as to change further parameters of the same parameter
group.
After having accessed the last parameter of a parameter group, the instrument will switch
back to the measurement mode. Changes to the parameters become effective immediately
and are automatically saved to the EEPROM.
When the instrument is running in the parameter mode and if for a period of 10 seconds no
changes are made to the parameters, then the instrument will automatically return to the
measurement mode. Any parameter changes made up to this point are automatically saved
to the EEPROM.
NOTICE:
When the instrument is running in the parameter mode, and when not entering
a change for more than 10 seconds, then the instrument will automatically
return back to the measurement mode.
Any changes, which have been entered and confirmed up to this point of time,
are automatically saved in the EEPROM.
GA09037_002_C0 – 09/2016
7.
Parameters
7.1
Switching Function Parameters (PArA SP)
In this parameter group you can configure the switching functions. The multichannel gauge
provides the following switching function parameters:
SP1-Lo
SP1-Hi
SP2-Lo (only with deselected ready function)
SP2-Hi (only with deselected ready function)
7.1.1
Basic Terms
Switching functions
The DISPLAY TWO contains four relays with change-over contacts, from which two for
switching functions and two are available as ready function. The relays for the ready
functions can be used alternatively also for switching functions.
The DISPLAY THREE contains six relays with change-over contacts, from which three for
switching functions and three are available as ready function. The relays for the ready
functions can be used alternatively also for switching functions.
The contacts of the relays are floating and may be used for switching purposes through the
connection marked Relay Output ( 5.3.5 Relay, page 19).
Threshold values
The switching behaviour of individual relays is defined in each case through two parameters:
the lower threshold and the upper threshold value of the switching function.
Lower threshold value SPx-Lo
The lower threshold value governs switching on of the assigned switching function. When
the pressure drops below the lower threshold value, then the relay is energised. The
common contact of the relay is then connected to the normally open contact.
Upper threshold value SPx-Hi
The upper threshold value governs switching off of the assigned switching function. When
the pressure exceeds the upper switching threshold, then the relay is de-energised. The
common contact of the relay is then connected to the normally closed contact.
Hysteresis
In the pressure range between the two thresholds, the current relay status is maintained.
Within this range the relay will not switch over and the relay status depends on the history
( Figure 18, page 29).
GA09037_002_C0 – 09/2016
29
Figure 18 – Behaviour of a switching function in
response to pressure changes
p
Pressure
t
Time
NO
Normally open contact
COM Common contact
NC
Normally closed contact
The range between the lower and the upper threshold value produces a certain amount of
hysteresis between switching on and switching off of the relay. The hysterisis prevents rapid
switching on and off of the switching function when the pressure is close to a switching
threshold.
7.1.2
Configuring the Switching Functions
Select the desired measurement channel by operating the pushbutton CH several
times.
Keep the pushbutton PARA depressed for approximately 2 seconds.
The instrument is now running the parameter mode in the parameter group SP.
Press the pushbutton PARA to select the desired parameter.
The name and the value of the parameter are displayed.
Use the arrow pushbuttons so as to change the threshold value.
The value of the parameter is changed.
Repeat the steps to change further parameters of the parameter group.
7.1.3
Adjustment Range
The upper and the lower threshold value can be changed depending on the type of sensor
in the range between 1·103 and 1·10-12 mbar.
Hysteresis amounts to at least 10% of the lower threshold value.
CAUTION:
Select the threshold values only within the range of the selected sensor.
CAUTION:
The threshold values have to be adjusted when the sensor type on a
measuring channel has been changed.
GA09037_002_C0 – 09/2016
7.2
Sensor Parameters (PArA SEn)
For each measurement channel there exists a separate set of sensor parameters.
Depending on which sensor is connected to the respective measurement channel, different
parameters are available ( Table 9, page 30).
Sensor
PrE
FiLt
rAEdY
Cor
S-on
S-oFF
()
()
Table 9 – Available sensor parameters
Select the desired measurement channel by operating the pushbutton CH several
times.
Keep the pushbutton PARA depressed for approximately 2 seconds
The instrument is now running the parameter mode.
Use the arrow keys to access the parameter group SEn.
Press the pushbutton PARA to select the desired parameter.
The name and the value of the parameter are displayed.
Use the arrow keys to change the parameter setting.
Repeat the steps to change further parameters of the parameter group.
7.2.1
Pirani Range Extension (PrE)
Pirani range extension for THERMOVAC sensors.
Display
Description
Pirani range extension deactivated
Sensor indicated as or
Pirani range extension activated
Sensor indicated as or
Table 10 – Values for the parameter PrE
7.2.2
Measured Value Filter (FiLt)
The measurement value filter improves processing of unstable signals or signals suffering
from interference. The filter has an effect on the display and on the switching functions.
However, the analog outputs are not influenced.
For the measured value filter you may select between the values 1, 3, 7 and 15. Here 1
stands for fast and 15 for slow. The values 3 and 7 are corresponding intermediate
values. In case of a two digit display the use of filter factor of 3 is recommended, in the case
of a three digit display filter factor 15 is recommended.
GA09037_002_C0 – 09/2016
31
7.2.3
Ready Function (rAEdY)
The second relay per channel makes according to standard a ready signal available.
Alternatively, this relay can be used in addition, as the second switching function.
Display
Description
Ready signal function activeted
Switching function activeted
Table 11 – Values for the parameter rEAdY
7.2.4
Type of Gas Correction (Cor)
Sensors have normally been calibrated for measurements in nitrogen or air. With the aid of
the parameter Cor you can adjust the measurement channel to other type of gases
(PENNINGVAC sensors PTR81N, PTR225, PTR225N, PTR225S, PTR225SN, PTR237 and
PTR237N only).
For this proceed as follows
Select the parameter Cor.
Press the key marked PARA.
The correction factor is displayed.
Use the arrow keys to change the correction factor.
The value of the parameter is changed.
You can adjust the correction factor for a sensor within the range of 0.1 – 1.0 – 9.9. The
setting of 1.0 supplies the uncorrected measured values (for nitrogen or air).
7.2.5
Sensor Switch-on type (S-on)
This parameter defines how the PENNINGVAC sensors PTR81N, PTR225, PTR225N,
PTR225S, PTR225SN, PTR237 and PTR237N are switched on.
You can set up the switch-on type to the following:
HAnd
Manual. The sensor can be switched on by pressing the key marked UP.
( Chapter 6.4.3.3 Switching on the High Vacuum Measurement Circuit, page 24)
ECt
Externally through optocoupler (static signal +12 – 24 V DC)
Hot
Warm start. The sensor is switched on automatically upon switching on the instrument. After
a power failure the measurement is started automatically.
CH 1
Through measurement channel 1. With the aid of the next parameter t-on you can define a
switch-on threshold. As soon as the pressure in measurement channel 1 drops below the
switch-on threshold, the sensor will be switched on.
GA09037_002_C0 – 09/2016
CH 2
Through measurement channel 2. With the aid of the next parameter t-on you can define a
switch-on threshold. As soon as the pressure in measurement channel 2 drops below the
switch-on threshold, the sensor will be switched on.
CH 3
Through measurement channel 3. This setting is only available in the case of the DISPLAY
THREE. With the aid of the next parameter t-on you can define a switch-on threshold. As
soon as the pressure in measurement channel 3 drops below the switch-on threshold, the
sensor will be switched on.
7.2.6
Sensor Switch-on Value (t-on)
This parameter will only appear when the sensor switch-on time has been set to CH 1, CH
2 or CH 3 ( Chapter 7.2.5 Sensor Switch-on type (S-on), page 31).
Through the parameter t-on you can define a switch-on value. If the pressure in the
respective measurement channel drops below the switch-on value, the PENNINGVAC
sensors PTR81N, PTR225, PTR225N, PTR225S, PTR225SN, PTR237 and PTR237N are
switched on.
7.2.7
Sensor Switch-off type (S-oFF)
This parameter defines how the PENNINGVAC sensors PTR81N, PTR225, PTR225N,
PTR225S, PTR225SN, PTR237 and PTR237N are switched off.
You can set the switch-off type to the following:
HAnd
Manual. The sensor can be switched off by pressing the key marked DOWN.
( Chapter 6.4.3.4 Switching off the High Vacuum Measurement Circuit, page 25)
ECt
Externally through optocoupler (static signal +12 – 24 V DC)
SELF
Self monitoring. With the aid of the next parameter t-off you can define a switch-off threshold.
When the pressure at the sensor exceeds the switch off threshold, the sensor will be
switched off.
CH 1
Through measurement channel 1. With the aid of the next parameter t-off you can define a
switch-off threshold. As soon as the pressure in measurement channel 1 exceeds the
switch-off threshold, the sensor will be switched off.
CH 2
Through measurement channel 2. With the aid of the next parameter t-off you can define a
switch-off threshold. As soon as the pressure in measurement channel 2 exceeds the
switch-off threshold, the sensor will be switched off.
CH 3
Through measurement channel 3. This setting is only available in the case of the DISPLAY
THREE. With the aid of the next parameter t-off you can define a switch-off threshold. As
soon as the pressure in measurement channel 3 exceeds the switch-off threshold, the
sensor will be switched off.
GA09037_002_C0 – 09/2016
33
7.2.8
Sensor Switch-off Value (t-off)
This parameter will only appear when the sensor switch-off type has been set to CH 1, CH
2 or CH 3 ( Chapter 7.2.7 Sensor Switch-off type (S-oFF), page 32).
Through the parameter t-off you can define a switch-off value. If the pressure in the
respective measurement channel exceeds the switch-off value, the PENNINGVAC sensors
PTR81N, PTR225, PTR225N, PTR225S, PTR225SN, PTR237 and PTR237N are switched
off.
7.3
General Parameters (PArA GEn)
With the aid of these parameters you can generally configure the instrument. The
parameters apply to all measurement channels.
Keep the pushbutton PARA depressed for approximately 2 seconds.
The instrument is now in the parameter mode.
Use the arrow pushbuttons so as to access the parameter group GEn.
Press the pushbutton PARA to select the desired parameter.
The name and the value of the parameter are displayed.
Use the arrow pushbuttons so as to change the parameter setting.
Repeat the steps so as to change further parameters of the parameter group.
7.3.1
Unit of Measurement (unit)
Unit of measurement for pressure values. This unit of measurement affects the displayed
pressure values, threshold values etc.
Display
Description
Unit of measurement mbar
Unit of measurement Torr
Unit of measurement Pascal
Table 12 – Values for the parameter unit
The unit of measurement is indicated on the display ( Figure 12, page 21).
7.3.2
Display Format (diGit)
Number of digits on the display.
Display
Description
2 digits, for example 2,5-1 or 370
3 digits, for example 2,47-1 or 373
Table 13 – Values for the parameter diGit
GA09037_002_C0 – 09/2016
7.3.3
Display Brightness (bri)
Display brightness.
Display
Description
High brightness
Low brightness
Table 14 – Values for the parameter bri
7.3.4
Analog Output (AnALoG)
Voltage at the analog output.
Display
Description
Analog output voltage same as at the sensor output
Halved Analog output
Table 15 – Values for the parameter AnA
GA09037_002_C0 – 09/2016
35
8.
Maintenance and Service
8.1
Maintenance
8.1.1
General Maintenance Information
For external cleaning please only use a dry piece of cotton cloth. Do not use any aggressive
or abrasive cleaning agents.
DANGER: Mains voltage
The instrument contains voltage carrying components inside. Do not insert
any objects into the openings of the instrument. Protect the instrument against
moisture. Do not open the instrument.
8.2
Troubleshooting
8.2.1
Fault Indication
A malfunction affecting the multichannel gauge is indicated by an error message on the
display. ( Chapter 8.2.2 Error Messages, page 35)
8.2.2
Error Messages
Error (display)
Fault cause and remedy
Significantly below the measurement range.
The Ready relay switches off.
Significantly above the measurement range.
The Ready relay switches off.
Sensor error. Malfunction affecting the connection to
the sensor. The message will only be displayed in
the display field for the affected measurement
channel. Acknowledge by pressing any key.
No sensor connected to the measurement channel.
The status message disappears after 30 seconds.
Error affecting the switch-on or switch-off channel.
No switch-on or switch-off threshold has been
defined for the PENNINGVAC sensors PTR81N,
PTR225, PTR225N, PTR225S, PTR225SN,
PTR237 or PTR237N.
Table 16 – Error messages
GA09037_002_C0 – 09/2016
8.2.3
Help the Case of Faults
If the malfunction persists even after having replaced the sensors, please contact your
nearest Leybold GmbH service office.
8.2.4
Fuse Replacement
To replace blown fuses use only the type of fuse T1.6A H specified on the rear of the
instrument. The two fuses of the instrument are located in the fuse insert at the mains filter
( Figure 7, page 18), which may be levered out using a small screwdriver.
8.2.5
Repair
Send defective products for repair to your nearest Leybold GmbH service office. The compa-
ny Leybold GmbH will not assume any responsibility or honour a warranty if the operator or
third persons have attempted repair work on the multichannel gauge.
GA09037_002_C0 – 09/2016
37
9.
Storing and Waste Disposal
9.1
Packaging
Please retain the original packaging. You will need this packaging material when wanting to
store the multichannel gauge or when returning it to the Leybold GmbH.
9.2
Shelving
The multichannel gauge must only be stored in dry room. During storage, the following
ambient conditions need to be maintained:
Ambient temperature: -20 – +60 °C
Humidity of the air:
As low as possible.
Preferably in a sealed plastic bag with desiccant.
9.3
Waste Disposal
As to waste disposal, the branch-specific and local waste disposal and environment
regulations for equipment containing electronic components apply.
When returning the equipment to the Leybold GmbH, proper and professional separation of
the recyclable fraction and its disposal is ensured.
17200001_002_�� © Leybold
Declaration of Contamination of Compressors, Vacuum Pumps and Components
The repair and / or servicing of compressors, va cuum pumps and components will be carried out only if a correctly completed de claration has
been submitted. Non-completion will result in delay. The manufacturer can refuse to accept any equipment without a declaration.
A separate declaration has to be completed for each single component.
This declaration may be completed and signed only by authorized and qualified staff.
Customer/Dep./Institute :
Reason for return:
applicable please mark
Address :
Repair:
chargeable
warranty
Exchange:
chargeable
warranty
Exchange already arranged / received
Person to contact:
Return only:
rent
loan
for credit
Phone :
Fax:
Calibration:
DKD
Factory-calibr.
End user:
Quality test certificate DIN 55350-18-4.2.1
A. Description of the Leybold product:
Failure description:
Material description :
Catalog number:
Additional parts:
Serial number:
Application-Tool:
Type of oil (ForeVacuum-Pumps) :
Application- Process:
B. Condition of the equipment
No1)
Yes
No
Contamination :
No1)
Yes
11.. Has the equipment been used
toxic
2. Drained (Product/service fluid)
corrosive
3. All openings sealed airtight
flammable
4. Purged
explosive 2)
If yes, which cleaning agent
radioactive 2)
and which method of cleaning
microbiological 2)
1) If answered with “No”, go to D.
other harmful substances
C. Description of processed substances (Please fill in absolutely)
1. What substances have come into contact with the equipment ?
Trade name and / or chemical term of service fluids and substances processed, properties of the substances
According to safety data sheet (e.g. toxic, inflammable, corrosive, radioactive)
X
Tradename:
Chemical name:
a)
b)
c)
d)
No
Yes
2. Are these substances harmful ?
3. Dangerous decomposition products when heated ?
If yes, which ?
2) Components contaminated by microbiological, explosive or radioactive products/substances will not be accepted without written
evidence of decontamination.
D. Legally binding declaration
I / we hereby declare that the information supplied on this form is accurate and sufficient to judge any contamination level.
Name of authorized person (block letters) :
Date
signature of authorized person
firm stamp
Bonner Strasse 498
D-50968 Cologne
T: +49-(0)221-347-0
F: +49-(0)221-347-1250
[email protected]
Germany
Leybold GmbH
Sales, Service, Support Center (3SC)
Bonner Strasse 498
D-50968 Cologne
T:
+49-(0)221-347 1234
F:
+49-(0)221-347 31234
[email protected]
www.leybold.com
Leybold GmbH
Sales Area North
Branch Office Berlin
Industriestrasse 10b
D-12099 Berlin
T:
+49-(0)30-435 609 0
F:
+49-(0)30-435 609 10
[email protected]
Leybold GmbH
Sales Office South
Branch Office Munich
Karl-Hammerschmidt-Strasse 34
D-85609 Aschheim-Dornach
T:
+49-(0)89-357 33 9-10
F:
+49-(0)89-357 33 9-33
[email protected]
[email protected]
Leybold Dresden GmbH
Service Competence Center
Zur Wetterwarte 50, Haus 304
D-01109 Dresden
Service:
T:
+49-(0)351-88 55 00
F:
+49-(0)351-88 55 041
[email protected]
Europe
Belgium
Leybold Nederland B.V.
Belgisch bijkantoor
Leuvensesteenweg 542-9A
B-1930 Zaventem
Sales:
T:
+32-2-711 00 83
F:
+32-2-720 83 38
[email protected]
Service:
T:
+32-2-711 00 82
F:
+32-2-720 83 38
[email protected]
France
Leybold France S.A.S.
Parc du Technopolis, Bâtiment Beta
3, Avenue du Canada
F-91940 Les Ulis cedex
Sales and Service:
T:
+33-1-69 82 48 00
F:
+33-1-69 07 57 38
[email protected]
[email protected]
Leybold France S.A.S.
Valence Factory
640, Rue A. Bergès
B.P. 107
F-26501 Bourg-lès-Valence Cedex
T:
+33-4-75 82 33 00
F:
+33-4-75 82 92 69
[email protected]
Great Britain
Leybold UK LTD.
Unit 9
Silverglade Business Park
Leatherhead Road
Chessington
Surrey (London)
KT9 2QL
Sales:
T:
+44-13-7273 7300
F:
+44-13-7273 7301
[email protected]
Service:
T:
+44-13-7273 7320
F:
+44-13-7273 7303
[email protected]
Italy
Leybold Italia S.r.l.
Via Trasimeno 8
I-20128 Mailand
Sales:
T:
+39-02-27 22 31
F:
+39-02-27 20 96 41
[email protected]
Service:
T:
+39-02-27 22 31
F:
+39-02-27 22 32 17
[email protected]
Netherlands
Leybold Nederland B.V.
Floridadreef 102
NL-3565 AM Utrecht
Sales and Service:
T:
+31-(30) 242 63 30
F:
+31-(30) 242 63 31
[email protected]
[email protected]
Switzerland
Leybold Schweiz AG, Pfäffikon
Churerstrasse 120
CH-8808 Pfäffikon
Warehouse and shipping address:
Riedthofstrasse 214
CH-8105 Regensdorf
Sales:
T:
+41-44-308 40 50
F:
+41-44-302 43 73
[email protected]
Service:
T:
+41-44-308 40 62
F:
+41-44-308 40 60
[email protected]
Spain
Leybold Spain, S.A.
C/. Huelva, 7
E-08940 Cornellà de Llobregat
(Barcelona)
Sales:
T:
+34-93-666 43 11
F:
+34-93-666 43 70
[email protected]
Service:
T:
+34-93-666 46 11
F:
+34-93-685 43 70
[email protected]
America
USA
Leybold USA Inc.
5700 Mellon Road
USA-Export, PA 15632
T:
+1-724-327-5700
F:
+1-724-325-3577
[email protected]
Sales:
T:
+1-724-327-5700
F:
+1-724-333-1217
Service:
T:
+1-724-327-5700
F:
+1-724-325-3577
Brazil
Leybold do Brasil
Rod. Vice-Prefeito Hermenegildo Tonolli,
nº. 4413 - 6B
Distrito Industrial
Jundiaí - SP
CEP 13.213-086
Sales and Service:
T:
+55 11 3395 3180
F:
+55 11 99467 5934
[email protected]
[email protected]
Asia
P. R. China
Leybold (Tianjin)
International Trade Co. Ltd.
Beichen Economic
Development Area (BEDA),
No. 8 Western Shuangchen Road
Tianjin 300400
China
Sales and Service:
T:
+86-22-2697 0808
F:
+86-22-2697 4061
F:
+86-22-2697 2017
[email protected]
[email protected]
India
Leybold India Pvt Ltd.
No. 82(P), 4th Phase
K.I.A.D.B. Plot
Bommasandra Industrial Area
Bangalore - 560 099
Indien
Sales and Service:
T:
+91-80-2783 9925
F:
+91-80-2783 9926
[email protected]
[email protected]
Japan
Leybold Japan Co., Ltd.
Headquarters
Shin-Yokohama A.K.Bldg., 4th floor
3-23-3, Shin-Yokohama
Kohoku-ku, Yokohama-shi
Kanawaga 222-0033
Japan
Sales:
T:
+81-45-471-3330
F:
+81-45-471-3323
[email protected]
Leybold Japan Co., Ltd.
Tsukuba Technical Service Center
1959, Kami-yokoba
Tsukuba-shi, Ibaraki-shi 305-0854
Japan
Service:
T:
+81-29 839 5480
F:
+81-29 839 5485
[email protected]
Malaysia
Leybold Malaysia
Leybold Singapore Pte Ltd.
No. 1 Jalan Hi-Tech 2/6
Kulim Hi-Tech Park
Kulim, Kedah Darul
Aman 09000
Malaysia
Sales and Service:
T:
+604 4020 222
F:
+604 4020 221
[email protected]
[email protected]
South Korea
Leybold Korea Ltd.
3F. Jellzone 2 Tower
Jeongja-dong 159-4
Bundang-gu Sungnam-si
Gyeonggi-do
Bundang 463-384, Korea
Sales:
T:
+82-31 785 1367
F:
+82-31 785 1359
[email protected]
Service:
623-7, Upsung-Dong
Cheonan-Si
Chungcheongnam-Do
Korea 330-290
T:
+82-41 589 3035
F:
+82-41 588 0166
[email protected]
Singapore
Leybold Singapore Pte Ltd.
8 Commonwealth Lane #01-01
Singapore 149555
Singapore
Sales and Service:
T:
+65-6303 7030
F:
+65-6773 0039
[email protected]
[email protected]
Taiwan
Leybold Taiwan Ltd.
No 416-1, Sec. 3
Chunghsin Rd., Chutung
Hsinchu County 310
Taiwan, R.O.C.
Sales and Service:
T:
+886-3-500 1688
F:
+886-3-583 3999
[email protected]
[email protected]
Sales and Service