Medtronic MiniMed 640G Getting Started
Text of the manual
INTRODUCING THE MINIMED 640G SYSTEM
MiniMed Care ................................................................................................................................................3
Welcome
................................................................................................................................................4
The Delivery of Insulin ..................................................................................................................................7
Menu Map
................................................................................................................................................8
GETTING STARTED WITH THE MINIMED 640G INSULIN PUMP
SECTION 1:
Pump Basics..................................................................................................................... 10
SECTION 2: Home Screen................................................................................................................... 12
SECTION 3: Basal Patterns.................................................................................................................. 14
SECTION 4: Giving Boluses................................................................................................................. 18
SECTION 5: ContourTM Next LINK 2.4 Meter................................................................................... 22
SECTION 6: Infusion Set & Reservoir (changing the Mio™ Infusion Set)................................ 25
SECTION 7: Alerts and Alarms............................................................................................................ 32
SECTION 8: Suspend Delivery............................................................................................................ 34
GETTING STARTED WITH CONTINUOUS GLUCOSE MONITORING
SECTION 9: Introduction to Continuous Glucose Monitoring................................................... 35
SECTION 10: Sensor Glucose (SG) and Blood Glucose (BG)........................................................ 36
SECTION11: Trends................................................................................................................................ 37
SECTION 12: Personalised Alerts......................................................................................................... 38
SECTION 13: Reading the Sensor Display......................................................................................... 48
SECTION 14: Connecting your Pump and Transmitter.................................................................. 50
SECTION 15: Inserting and Starting the Sensor............................................................................... 52
SECTION 16: Calibrating......................................................................................................................... 58
SECTION 17: Other Sensor Alerts....................................................................................................... 62
APPENDIX
Charging and storing your transmitter................................................................................................... 63
CareLink™ Personal Therapy Management Software....................................................................... 64
X-Rays and Traveling By Air........................................................................................................................ 65
SmartGuard™ technology Features....................................................................................................... 66
INTRODUCTION I MINIMED CARE
TABLE OF CONTENTS
At Medtronic we aim to develop new and innovative
products to help improve the lives of people with diabetes.
We are also committed to providing our patients with
ongoing personalised support and education to help them
make the most of their diabetes management whilst being
on pump therapy and continuous glucose monitoring.
Register now for My MiniMed Care for personalised support and
education including:
Quarterly newsletters with useful tips and tricks,
testimonials and ongoing support
Access to online personalised practical tools such as ordering
pump clip replacements
Online seminars and education
And come back soon to benefit from our constantly
expanding servicescontent for pump users
REGISTER TODAY
Register for My MiniMed Care today at
www.medtronic-diabetes.co.uk/minimedcare_registration
GETTING STARTED WITH THE MINIMED 640G SYSTEM
Welcome! We are glad that you have chosen insulin pump therapy and are excited
for you to begin using your MiniMed 640G system.
The MiniMed 640G system features innovative technology to more closely mimic
the way a healthy pancreas delivers basal insulin to the body and help you achieve
better glucose control. The MiniMed 640G system has been designed to provide
you with:
Advanced protection from dangerous high and lows
Personalised convenience to help you manage daily
diabetes tasks
The MiniMed 640G insulin pump also has a new, improved
design for consistent ease of use.
Whether you are just starting pump therapy or upgrading from
a previous pump model, this guide provides you step-by-step
instructions on the basic operation and programming of your
MiniMed 640G system, including Continuous Glucose Monitoring.
During your in-person training, your trainer will build on this
information and help ensure you are confident to begin using
your MiniMed 640G insulin pump.
We hope you enjoy learning about your new insulin pump.
Information contained herein does not replace the recommendations of your
healthcare professional.
For a listing of indications, contraindications, precautions, warnings, and potential
adverse events, please refer to the Instructions for Use.
Please refer to the User Guide and your HCP for more information.
When using an insulin pump, check your blood glucose minimum 4 times a day.
INTRODUCTION I WELCOME
INTRODUCTION I MINIMED CARE
INSULIN
DELIVERY
Medtronic offers a wide
range of infusion sets so
that you can choose the
right infusion set for your
comfort and safety.
THERAPY
MANAGEMENT
TOOLS
Upload your pump to CareLink to
conveniently track your glucose
control and remotely share this
information with your
healthcare professional.
www.medtronicdiabetes.com/carelink
CONTINUOUS
GLUCOSE
MONITORING
With our most responsive
sensor system yet, you can
continuously monitor your
glucose levels with superior
comfort* and accuracy**.
BLOOD
GLUCOSE
TESTING
With the accurate1 Contour
Next LINK 2.4 meter you
can wirelessly send blood
glucose results to your
MiniMed 640G insulin pump
and deliver an insulin bolus
discretely.
IMPORTANT: Do NOT attach the insulin pump to your body or attempt to use
insulin in your pump as you use this guide to practice and learn.
Attaching and using must only be done when you receive formal training with
your healthcare professional or a certified product trainer.
DID YOU KNOW?
A complete
explanation of the
technical and
operational aspects
of your pump can be
found in the
MiniMed 640G
system User Guide.
MINIMED 640G SYSTEM
*The size of the new generation Enlite sensor has been reduced by 80%.
**When combined with the MiniMed 640G system and Guardian 2 Link transmitter. Enlite Sensor Performance addendum to user guide.
1. Section 8 clinical study. Data on file. Ascensia Diabetes Care.
THE DELIVERY OF INSULIN
This document does not replace the Instructions for
Use. For a listing of indications, contraindications,
precautions, warnings, and potential adverse events,
please refer to the Instructions for Use.
INTRODUCTION I THE DELIVERY OF INSULIN
*Mio™ infusion set shown in illustration.
**Some infusion sets do not use a cannula but have a small
needle that remains inserted in the body.
INFUSION SET*
1 Tubing: carries insulin from the pump to you
2 Reservoir Connector: end of the tubing that
attaches the reservoir which holds the insulin
3 Insertion Site Section: other end of the
tubing that attaches to you
4 Cannula: tiny flexible tube placed into your
body** by the insertion needle included in
the insertion site section
5 Adhesive: holds the infusion set in place
6 Reservoir Compartment: part of the pump
where the reservoir fits
You should replace both the infusion set and the
reservoir every 2 to 3 days.
3
5
1
2
4
6
INTRODUCTION I MENU MAP
INTRODUCTION I MENU MAP
Home Screen
Press
History
Summary
Daily History
Alarm History
Sensor Glucose Review*
ISIG History*
Reservoir & Tubing
New Reservoir
Fill Cannula
Insulin Settings
Bolus Wizard Setup
Basal Pattern Setup
Preset Temp Setup
Preset Bolus Setup
Dual/Square Wave
Bolus Increment
Max Basal/Bolus
Easy Bolus
Auto Suspend
Bolus Speed
Reminders
Personal
Bolus BG Check
Missed Meal Bolus
Low Reservoir
Set Change
Calibration*
Sensor Settings
Sensor
Sensor Connections*
Alert Silence*
High Settings*
Low Settings*
Calibrate Sensor*
Auto Calibration*
Utilities
Airplane Mode
Display Options
Time & Date
Remote Bolus
Block
Self Test
Carb Unit
Manage Settings
Sensor Demo
Device Options
Language
*Only available when Sensor feature is on.
Event Markers
BG
Injections
Food
Exercise
Other
Audio Options
Audio
Audio & Vibrate
Vibrate
Volume
Suspend Delivery
Press
from any screen to open the Menu.
Press
and
to scroll through the menu items.
Press
on the desired menu item to open.
The scroll bar appears on menus to indicate when additional text is available.
Press
to scroll down to view additional items.
Press
to scroll back up.
Press
to go to previous screen. Hold
to return to the Home screen.
NAVIGATION
MENU MAP
Suspend Delivery
Audio Options
History
Reservoir & Tubing
Insulin Settings
Sensor Settings
Event Markers
Reminders
Utilities
THE FRONT OF YOUR PUMP
Up,
Down,
Left, and
Right
Press to scroll up or down through a
menu or list
Press to move to desired area on the
screen
Press to change the value in an area
Back
Press to return to a previous screen
Press and hold to return to the
starting screen, called the Home
screen
Select
Press to select or confirm a value
or menu option that is highlighted
Press when directions say ‘select ’
Menu
Press to get to the Menu
Press and hold to put pump into sleep
mode
Notification Light
Flashes when an Alert or an Alarm is
occurring
INSULIN PUMP I PUMP BASICS
INSULIN PUMP I PUMP BASICS
SECTION 1: PUMP BASICS
Before inserting the battery or pressing any buttons, let’s take a closer look at your pump.
NOTE: Do not undertighten or try to overtighten
the battery cap. It should be aligned horizontally
with the pump case as shown here.
THE BOTTOM AND BACK OF
YOUR PUMP
Pump Serial
and Model Number
You may need
to provide this
information if you
call for assistance.
INSERTING THE BATTERY
Your insulin pump is powered by a AA battery. A lithium, alkaline, or rechargeable
AA battery can be used. The battery you place into your pump should always be
new or fully charged.
To insert the battery and get started, you will need:
The battery cap found with the pump
The pump clip found with the accessories
The AA battery found with the accessories
Once the battery is inserted, the pump will power on
and the Startup Wizard will begin. You will need to follow it
step-by-step to set up your language, time format, time
and date.
Place the battery
into the battery
compartment with the
negative (flat) end
going in first.
Place the battery cap onto the pump.
Use the edge of the pump clip to turn the cap to
the right (clockwise) and tighten until the slot
is horizontal to the pump. See image below.
Medtronic
Diabetes
HelpLine
Telephone
Number
Back
Down
Menu
Notification
Light
Select
Up
Left
Right
1
2
INSULIN PUMP I HOME SCREEN
INSULIN PUMP I HOME SCREEN
SECTION 2: HOME SCREEN
The Home Screen will be your starting place to access all features in the pump.
Status bar: provides
a quick look at the
pump’s status
Bolus: gives you
access to the bolus
delivery screen and
other bolus insulin
options
Basal: gives you access
to basal insulin options
Current time
BG reading: displays
a BG taken in the last
12 minutes
Active insulin: displays
any insulin still active
from a previous bolus
STATUS BAR
The Status Bar displays the following icons so you can
quickly view important information.
Battery icon: Shows the level of charge your battery has. As the
battery charge decreases, the icon will become less full and change
to yellow and then red.
Reservoir icon: Shows the approximate amount of insulin left in your
reservoir. As insulin is used, the icon will become less full and change
to yellow and then red.
Audio icon: Shows the audio mode you are using: audio
, vibrate
or audio and vibrate
.
BACKLIGHT
When you are not pressing buttons on your pump, you will notice that the Backlight will
soon turn off. The pump is still on; it is just saving battery life. You can simply press any
button to make the screen reappear.
UNLOCKING THE PUMP
After the Backlight has been off for a few minutes,
the pump goes into Sleep mode and the pump is locked.
When you begin using your pump again, you will see a
screen like the one shown here when you leave the
Home screen. You will need to press the arrow key
that is highlighted to unlock the pump. This confirms
you are reading the screen and the button presses are
not accidental.
If the wrong arrow key is pressed, you will be asked to
try again.
You can press and hold
if you wish to put the pump
into Sleep mode and keep it locked when you are not
using it. Doing this can also help save battery life.
STATUS SCREENS
There will be times when you need additional status information such as the number
of insulin units left in your reservoir, the last BG entered or your current basal rate.
To access the Status
Screens, press
to
highlight the Status Bar
and press
.
REMEMBER: You can go back to the previous screen by pressing
.
NOTE: If you are using a 1.8mL insulin reservoir,
the range of reservoir icons will start with the
second green icon.
If starting with a full 1.8mL reservoir and using
the minimum amount of insulin recommended
to fill the shortest tubing (45 cm) and a steel
cannula, the pump will display the second
green (not full) reservoir icon after the
reservoir and infusion set change is complete.
Not displayed with a
1.8mL reservoir
INSULIN PUMP I BASAL PATTERNS
INSULIN PUMP I BASAL PATTERNS
BASAL PATTERNS SET UP - MULTIPLE BASAL RATES
SETTING MULTIPLE BASAL RATES
NOTE: The basal rates shown are for illustration purposes only
– your basal settings will be different.
The pump supplies basal insulin by
delivering small amounts of short-acting
insulin throughout each hour, every
hour of the day and night. This allows for
insulin to be increased and decreased to
adjust for your body’s needs.
Basal insulin amounts must be
programmed into your pump. This is
done by setting a basal pattern. A basal
pattern consists of one or more basal
rates being delivered over the course
of 24 hours.
It is likely when you start on pump therapy, that you will need more than one basal rate
throughout the day and night to meet your body’s insulin needs.
For example, a Basal Pattern may look like this:
00:00
06:00
12:00
18:00
24:00
00:00 – 03:00
03:00 – 08:00
0.700 U/hr
0.800 U/hr
08:00-15:00
0.650 U/hr
15:00-19:00
0.550 U/hr
19:00-24:00
0.600 U/hr
24 Hours
1.000
0.500
U/hr
In this example, the basal pattern includes 5 different basal rates over 24 hours.
5
You can see you are
automatically asked to enter the
end time of the second basal
rate. This basal rate will need to
end at 8:00 and will need to be
changed to 0.800 U/hr.
Press
to 0.700 U/hr and press
.
1
2
Select Basal 1 > Options > Edit.
From the Home screen,
select Basal > Insulin
Settings > Basal Pattern
Setup.
3
4
Press
to change End time
to 03:00 and press
.
Press
on the time
segment. The End time
will be flashing.
Continued on next page
SECTION 3: BASAL PATTERNS
Basal insulin is delivered throughout the day and night to cover insulin
needs between meals and during the night.
NOTE: The Home screen reads Basal (T) since you have a Temp Basal active. Select
Basal (T) to review the details of the active Temp Basal. When the Temp Basal is
complete, the basal will automatically return to the regularly programmed basal rate.
INSULIN PUMP I BASAL PATTERNS
INSULIN PUMP I BASAL PATTERNS
Press
or
to
enter the percent
of current basal rate
desired
.
Select Begin.
Select Next.
Note: If you choose to use Rate, select Type,
and you can then enter the U/hr desired.
Select Percent.
Press
to set
duration and
press
.
From the Home screen,
select Basal > Temp Basal.
Change End time to 08:00 and
basal rate to 0.800 U/hr using
and press
.You can now enter
the next end time.
Repeat steps 3 to 6 to enter the 3 next
time segments and basal rates. For the
last time segment, you will need to enter
24:00 as the end time to complete the
full 24 hours.
Verify that Basal 1 is entered correctly.
Press
to view all basal rates.
Select Done.
If NO changes need to be
made: Select Save.
This basal
pattern
delivers
15.85 U
over 24
hours.
If changes need to be
made: Press
.
Press
and press
.
Repeat steps 7 to 9.
Select
Save.
6
10
11
12
8
7
9
TEMPORARY (TEMP) BASAL RATE
This feature lets you immediately
increase or decrease your basal insulin
for the period of time (duration) that you
set. It is the easiest way to immediately
adapt your basal rate according to your
daily life and is often used for exercise
and sick days.
A Temp Basal can be set in either
Percent (delivers a percent of the
current basal rate) or by Rate (delivers
the amount that you enter).
SET TEMP BASAL RATE
1
From the Home screen,
select Basal (T).
Select Cancel Temp Basal.
Note: Basal rate
has now returned
to the currently
programmed
rate.
If you ever set a Temp Basal and decide you do not need it, it can be canceled.
CANCEL TEMP BASAL RATE
1
2
4
2
5
3
6
SECTION 4: GIVING BOLUSES
A bolus is given to cover food that contains carbohydrate and/or to correct glucose
levels that are above your target range.
TURNING THE BOLUS WIZARD FEATURE ON AND SETUP
From the Menu, select
Insulin Settings > Bolus
Wizard Setup > Bolus
Wizard.
Press
to continue reading text and select Next.
INSULIN PUMP I GIVING BOLUSES
INSULIN PUMP I GIVING BOLUSES
NOTE: Before using the Bolus Wizard calculator, you need to program your individual
settings with the help of your health care professional.
Calculating how much bolus insulin to
give can be challenging. When using the
Bolus Wizard feature, all you will need
to do is enter your current BG reading
along with the amount of carbs you are
about to eat.
Once you do this, the Bolus Wizard
calculator uses the individual
settings provided by your health care
professional to estimate your bolus
amount. Because these settings
are specific to you, you can use it to
calculate the precise amount of bolus
insulin you need for your food and BG.
This can help you better control your
glucose levels.
3
4
BOLUS WIZARD CALCULATOR
USING THE BOLUS WIZARD FEATURE
Here you can see the Bolus Wizard calculation screen and a short description of the
steps below:
You will first test and enter
your current BG.
You will then enter grams of
carbohydrates to be eaten.
The pump displays estimated
amount of insulin to be
delivered.
If you have tested your glucose using your compatible Ascensia
meter, the BG and correction dose will already be showing.
Select Save.
THE BOLUS WIZARD CALCULATOR SETUP IS NOW COMPLETE.
Follow the instructions to program the following settings: Carb Ratio, Insulin
Sensitivity Factor (Sensitivity), BG Target and Active Insulin Time. Each
setting will include a short description: you need to select Next and enter
the requested data.
Carb Ratio
Insulin Sensitivity
Factor
BG Target
Active Insulin Time
NOTE: The boluses shown are for illustration purposes only – your settings and bolus
results will be different.
1
2
GIVING A MANUAL BOLUS
When giving a manual bolus, you simply enter the amount of bolus insulin that you
think you need for the carbohydrates you are eating, or to lower your BG if it is high.
From the Home screen,
select Bolus.
Press
to 1.0 u and press
.
Confirmation that Bolus has
started will appear.
Select Deliver Bolus.
Notice there is Active Insulin now displayed. Active insulin is insulin
from boluses that are still working to lower your blood glucose level
and/or to cover your carbohydrate intake that is still not absorbed.
Each time you give a bolus, it is added to the active insulin amount.
As time passes, the amount will decrease. You will learn more about
active insulin during your training.
Notice that Stop Bolus also
now appears.
The Home
screen will
show the
amount
as it is being
delivered.
INSULIN PUMP I GIVING BOLUSES
INSULIN PUMPI GIVING BOLUSES
2
2
1
3
3
4
Once the bolus
has finished
delivering, the
pump will return
to the normal
Home screen.
STOPPING A BOLUS
To stop a bolus while it is delivering:
From the Home screen,
select Stop Bolus.
CHECKING LAST BOLUS
There may be times when you need to
see the time or amount of the last bolus
that was given. You can see the last bolus
delivered in the Quick Status screen.
From the Home screen, select the Status
Bar > Quick Status (see page 13).
CHECKING BOLUS HISTORY
You may also want to review the
last several boluses that were
delivered. You can see the last
several boluses delivered in Daily
History. Press Menu button >
History > Daily History.
Press
and select Yes.
The (N) behind the Last
bolus amount means
the bolus was delivered
as a normal bolus. There
are additional ways to
give a bolus which you
will learn about later.
The Bolus Stopped
screen will show you
how much of the bolus
insulin was delivered
before it was actually
stopped.
Select Done.
1
CONNECTING YOUR PUMP AND METER
You will connect your pump and meter at your in-person training. For more
information on using your meter, see the User Guide found in the meter box.
5
6
Select Utilities.
Select Device Options.
Place the meter and pump next to each other.
Before Meal
5.2
Bolus
Reminder
Notes
mmol
L
SECTION 5: Contour Next LINK 2.4 METER
Test Strip Port*
and Port Light
Display Screen
Menu and Power Button
USB Connector
Protective USB Cap
Selection/Scrolling Buttons
Contour Next LINK 2.4 METER
CHARGING YOUR METER
Your meter has a permanent rechargeable battery. It is important that the meter be
charged prior to your in-person training. To charge your meter:
1. Plug the USB connector into a computer.**
The computer must be ON and not in sleep,
hibernate or power save mode.
2. The meter will briefly display Do Not Test-
charging and the test strip port light will flash.
You cannot do a blood glucose test while the
battery is charging.
3. When charging is complete, the test strip port
light will turn off. You can then unplug your meter.
* The Contour Next Link 2.4 meter only works with Contour™ Next glucose testing strips.
** If you would prefer not to charge your meter using your computer, you can purchase a
compatible outlet charger by calling Ascensia Customer Service.
10:24 l 11.10
INSULIN PUMP I Contour Next LINK 2.4 METER
INSULIN PUMP I Contour Next LINK 2.4 METER
The Contour Next LINK 2.4 meter from
Ascensia is the only blood glucose meter
able to communicate wirelessly with your
MiniMed 640G insulin pump.
With the Contour Next LINK 2.4 meter,
you can:
Wirelessly send glucose readings to
your pump
Deliver a bolus remotely
Upload your pump data to CareLink
Personal software
Review the parts of your meter here:
1
2
Select "Connect
Device" on your Pump.
OK
Go Back
CONNECT TO PUMP
Connect to a
MiniMed Pump?
OK
Skip
CONNECT TO PUMP
Press OK when asked
Connect to a MiniMed Pump?
Press OK.
3
4
Press Auto Connect.
Press
on your pump.
7
8
Select Connect Device.
Select Auto Connect.
9
10
Press
.
Select Continue.
Put the meter
down and pick up
your pump.
Manual Connect
Auto Connect
CONNECT TO PUMP
SECTION 6: INFUSION SET AND RESERVOIR
The following is a step-by-step guide to changing the Mio™ Infusion set.
START HERE:
Select Rewind.
UPLOADING YOUR PUMP TO CARELINK PERSONAL SOFTWARE
INSULIN PUMP I INFUSION SET AND RESERVOIR
Continued on next page
CareLink Personal software is a web-
based program that is provided free
of charge. This software allows you to
upload the data from your pump and
glucose meter and review it on easy-
to-read reports. This enables you to
conveniently track your glucose control
and remotely share this information with
your healthcare professional.
To upload information from your pump
to CareLink Personal software, you will
use the Contour Next LINK 2.4 meter
as the communication device from the
pump to your computer, through the USB
connector of the meter.
To set up your CareLink Personal software
account, go to
www.medtronicdiabetes.com/carelink
Select Search on your pump.
Select Search on your meter.
The search may take up to 2 minutes.
CONNECT TO PUMP
Select “Search” on
Your Pump and meter.
Search
Cancel
Use Pump to confirm
Device SN:
BGJ133333F
CONNECT TO PUMP
Check to see that the
Device SN (serial number)
on the pump screen
matches the Device SN
on the meter.
If they match, select Confirm
on the pump.
4
5
Remove the used reservoir from the pump.
Remove the infusion set you have been
using by loosening the adhesive and pulling
away from body.
1
Wash your hands.
Press
.
2
Select Reservoir & Tubing.
3
Select New Reservoir.
Check to see that the SN
on the back of the pump
matches the SN now on
the meter screen.
Select Next on the meter.
PUMP CONNECTED
MiniMed Pump
connected:
AA1234567B
Next
Delete
INSULIN PUMP I Contour Next LINK 2.4 METER
Gently push connector
onto reservoir. Turn clock-
wise until locked. You will
hear a click.
IMPORTANT: Do not touch
the top of the reservoir before
connecting it to the infusion
set. Please make sure that the
connector and the top of the
reservoir are dry.
Select Next.
Connector
Select Load Reservoir
and unlock the pump if
necessary.
Plunger
rod
Reservoir
Transfer
guard
Remove from package. Make
sure insulin vial is at room
temperature to reduce the
risk of air bubbles.
Pull plunger down to the
amount that you plan to fill
with insulin.
Plunger
Wipe vial with alcohol.
Place vial on table.
Firmly press the blue
transfer guard onto vial.
Push and hold plunger
down.
Transfer guard
Insulin vial
Plunger
With your thumb still on
the plunger, flip over so
vial is on top. Release
thumb and pull plunger
down to fill with insulin.
Insulin vial
Plunger
Tap the reservoir to move air
bubbles to top of reservoir.
Push plunger up to move air
into vial.
If needed, pull plunger
back down to amount
of insulin needed for
2-3 days.
To avoid getting insulin on
top of the reservoir, turn
vial over so it is upright.
Hold transfer guard and
turn reservoir counter-
clockwise and remove
from transfer guard.
Reservoir
INSULIN PUMP I INFUSION SET AND RESERVOIR
INSULIN PUMP I INFUSION SET AND RESERVOIR
Remove the infusion set from packaging by pulling down the red
plastic tab. With one hand press the three raised markings on
the sides of the lid. Free tubing from slot. Gently unwind tubing in
counter clockwise direction.
IMPORTANT: Do not use the reservoir or infusion set if any liquid gets on the top of the
reservoir or inside the tubing connector. Liquid can temporarily block the vents and may result in
the delivery of too little or too much insulin, which can cause hypoglycaemia or hyperglycaemia.
Twist plunger counter-
clockwise to loosen and
remove.
If you see air bubbles, tap
reservoir to move them to
top. Push plunger just a bit
to move them into tubing.
Continued on next page
CONNECT
RESERVOIR TO
INFUSION SET
You will place the
reservoir connector
onto the end of the
infusion set to the
filled reservoir.
FILL RESERVOIR &
CONNECT TO THE
INFUSION SET
TUBING
Follow the next
steps to fill
reservoir with
insulin and connect
to the infusion set
tubing.
THE BACKLIGHT
MAY HAVE
TURNED OFF
Press any button
to turn the screen
back on
1
3
2
3
4
6
7
8
4
5
2
INSULIN PUMP I INFUSION SET AND RESERVOIR
INSULIN PUMP I INFUSION SET AND RESERVOIR
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over an
by the lined
the sides.
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body.
round indentations on each side of se
insert needle.
PLACE
RESERVOIR
INTO PUMP
Now place the
filled reservoir
into the reservoir
compartment of
the pump.
INSERTING THE
INFUSION SET
Next, follow the
steps to insert the
infusion set into
your body.
LOAD
RESERVOIR AND
FILL TUBING
Follow these
steps to load the
reservoir and fill
the tubing.
Place reservoir into pump.
Turn clockwise, until you feel
reservoir lock into place.
Select Next.
Select Load and keep
holding
.
When you see this screen,
select Next.
Select Fill and keep holding
until you see drops at the
end of tubing, then let go.
Drops at
end of
tubing
After you see drops,
press
and select Next.
Continued on next page
Continue top of next page
With other hand pull up
on centre of serter until it
clicks and locks into place.
Place tubing in slot on
side of serter.
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
Choose an insertion site from
the shaded areas shown here.
Gently peel paper to
expose adhesive.
Turn over and hold
by the lined ridges
on the sides.
Carefully hold needle guard
by tip. To remove it, gently
turn needle guard & pull
away to expose needle.
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6
2
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
2
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
3
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slo
on side of serter.
Next, follow the steps to insert the infusion set into
6.6
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid
serter to
discard n
6.8
Push down on center
of serter to press
adhesive against skin.
4
5
6
2
1
1
1
4
4
3
3
NOTE: Alternative infusion sets offered are Quick-setTM, Silhouette™ and Sure-T™.
Most of the steps described in this section will be different for each infusion set. Please
refer to the instructions for insertion included in the infusion set box.
INSULIN PUMP I INFUSION SET AND RESERVOIR
INSULIN PUMP I INFUSION SET AND RESERVOIR
NOTE: Select
Stop Filling if you need
to stop, for example, if you notice the
Total amount is incorrect. This should
rarely happen if you have verified the Fill
amount on the previous screen.
NOTE: Your pump will remember the Fill
amount that you used last. Always verify that
the Fill amount is correct.
If it is correct, press
to Fill Now
and press
.
If it is incorrect, press
. Change to
correct amount and. Press
.
Fill Now.
Place against prepared site on body.
Press the round indentations on
each side of serter to insert needle.
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
t into
shaded areas
other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
hand pull
er of serter
s and locks
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
n on center
o press
against skin.
Place lid back onto serter to
safely discard needle.
and hold
ed ridges on
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
dy. Press the
f serter to
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
Push down on centre of
serter to press adhesive
against skin.
6.1
Gently peel paper to
expose adhesive.
6.2
Turn over and hold
by the lined ridges on
the sides.
6.3
With other hand pull
up on center of serter
until it clicks and locks
into place.
6.4
Carefully hold needle
guard by tip. To
remove it, gently
turn needle guard &
pull away to expose
needle.
6.5
Place tubing in slot
on side of serter.
w the steps to insert the infusion set into
Choose an insertion site from the shaded areas
shown here. Wipe with alcohol or other antiseptic.
6.7
Place against prepared site on body. Press the
round indentations on each side of serter to
insert needle.
6.9
To remove serter, hold
the center handle and
gently pull it straight
out away from body.
Press adhesive securely
against skin.
6.10
Place lid back onto
serter to safely
discard needle.
6.8
Push down on center
of serter to press
adhesive against skin.
To remove serter, hold the
centre handle and gently
pull it straight out away
from body. Press adhesive
securely against skin.
7
1
4
3
2
8
9
10
The Home screen displays the
insulin as it fills the cannula.
Select Fill Now.
YOUR INFUSION SET CHANGE IS NOW COMPLETE!
Select Fill.
Select Fill amount and enter:
- 0.300 if using 6mm cannula
- 0.500 if using 9mm cannula
Then press
.
FILL CANNULA
You will now fill the
cannula, the little
tube under your
skin, with insulin.
SECTION 7: ALERTS AND ALARMS
ALERTS
An alert makes you aware of a situation that may need your attention. When an alert
occurs, you should check to see what your pump is telling you. Examples of alerts
include Low reservoir or Low battery.
ALARM
When an alarm occurs, something has been detected that is preventing insulin from
being delivered. You are not getting insulin. It is important that you address an alarm
right away. Examples of alarms are Insulin flow blocked and Replace battery now.
WHEN AN ALERT OR ALARM OCCURS:
Notification light
INSULIN PUMP I ALERTS AND ALARMS
INSULIN PUMP I ALERTS AND ALARMS
ALERT
ALARM
Notification
Light
The red light on the pump will blink once
followed by a pause. This sequence
continues until the alert is cleared.
The flashing pattern is shown here:
The red light on the pump will blink twice
followed by a pause. This sequence
continues until the alarm is cleared.
The flashing pattern is shown here:
Audio
Depending on your Audio Option settings, the pump emits a repeated alert tone,
a continuous three-pulse vibration, or both.
Display
The pump will display a notification
with a yellow icon and instructions on
what to do.
The pump will display a notification
with a red icon and instructions on
what to do.
IMPORTANT: It is important that you are able to address an Insulin flow blocked alarm.
This alarm means that insulin is not able to get through the tubing or cannula. If this
alarm occurs, check your blood glucose and check to see if your infusion set has become
dislodged or if your tubing is kinked.
If you don’t detect an issue and are unable to change your reservoir and infusion set right
away, you might choose to select Resume Basal. If an Insulin flow blocked alarm occurs
again, follow the steps on the screen. Select Rewind and change your reservoir and
infusion set.
If you detect an issue or if your reservoir has run out of insulin, follow the steps on the
screen. Select Rewind to change your reservoir and infusion set.
You can call the Product helpline if you have questions about your pump,
alerts or alarms.
TO ADDRESS AND CLEAR THE ALERT OR ALARM:
1. Read the text on the screen to understand the alert or alarm and the steps that
should be taken
2. Press
.
3. Press
on the desired option.
Example of alarm:
Press
Alert: the audio/vibration pattern repeats every 5 minutes or every 15 minutes
(depending on the alert) until the alert is cleared.
Alarm: the audio/vibration pattern repeats every minute for 10 minutes if
the alarm is not cleared. After 10 minutes, the alarm begins to siren.
SECTION 9: INTRODUCTION TO CONTINUOUS
GLUCOSE MONITORING
Continuous glucose monitoring (CGM) gives you a more complete picture of your
glucose control:
Drawings throughout this document are only generic representations of the system components.
1. Conget I., Choudhary P., Olsen B., et al. Prevention of hypoglycaemia by the predictive low glucose management
feature in a user evaluation study. Diabetologia. 2015, 58 (Suppl 1), p 416.
* See Appendix pages 66-67 for further details on how SmartGuard technology works.
** The transmitter must be within 1.8 meters of the insulin pump in order to communicate sensor readings.
The MiniMed 640G insulin pump will not communicate with MiniLink transmitters.
SECTION 8: SUSPEND DELIVERY
Remember your pump is delivering basal insulin throughout every hour of the day.
Although you should never stop this insulin delivery for more than an hour or so, there
will be times when you will want to manually suspend, or stop delivery, and disconnect
from your pump.
CGM I INTRODUCTION TO CONTINUOUS GLUCOSE MONITORING
INSULIN PUMPI SUSPEND DELIVERY
1 GLUCOSE SENSOR
The Enlite™ sensor measures glucose levels in the body.
2 TRANSMITTER**
The Guardian™ 2 Link transmitter connects to the glucose
sensor and sends glucose readings to your insulin pump.
3 INSULIN PUMP
The MiniMed 640G insulin pump displays glucose readings.
Other items include: Enlite serter, Enlite overtape, Guardian 2 Link charger and
watertight tester.
Always use the components that were sent with your MiniMed 640G system.
YOUR CONTINUOUS GLUCOSE MONITORING (CGM)
SYSTEM INCLUDES 3 KEY ITEMS:
1
2
3
This is done using the Suspend Delivery
feature. Using Suspend Delivery stops all
insulin delivery.
The most common reasons to manually
suspend delivery might include bathing
and water activities. Infusion sets are
designed so you can easily disconnect
from your pump and leave it in a safe
place.
When the pump is manually
suspended, all insulin delivery stops. All
insulin delivery will remain stopped until
you resume delivery.
When the delivery is resumed, basal
insulin will begin to deliver again. The
pump will not deliver any of the basal
insulin you missed while the pump was
suspended.
If you manually suspend delivery while
a bolus is delivering, the bolus delivery
will stop. When you resume delivery,
the remainder of the bolus will not be
delivered.
To place the pump in Manual Suspend:
from the Menu, select Suspend
Delivery > Press
and select Yes.
Notice that the Home screen has
changed.
To resume Basal Insulin Delivery, select
Resume from the Home screen.
Using a sensor allows you to receive
up to 288 sensor glucose readings
every 24 hours, filling the gaps
between your BG tests.
Graphs and trend arrows show the
speed and direction your glucose levels
are moving.
CGM alerts notify you of high and low
glucose values.
MiniMed 640G insulin pump also
includes SmartGuard technology,
Medtronic’s exclusive closed loop
technology.
SmartGuard technology mimics some
functions of a healthy pancreas, to
provide you with advanced protection
from hypoglyceamia1. SmartGuard
technology can:
PREDICT when you are approaching
low glucose levels 30 minutes
in advance
Automatically STOP insulin delivery
before you go hypoglycemic
And automatically RESUME it when
your glucose levels recover*.
SECTION 11: TRENDS
Sensor glucose trends give insight into the direction and the speed that your
glucose is changing. The sensor graph and trend arrows are used to show your trend
information.
Example of Sensor information on the Home Screen
Trend Arrows
Sensor Glucose Reading
Sensor Glucose Graph
SECTION10: SENSOR GLUCOSE (SG) AND BLOOD
GLUCOSE (BG)
Skin
Glucose Sensor
Interstitial Fluid
Cell
Glucose
Blood Vessel
Transmitter
CGM I TRENDS
CGM I SENSOR GLUCOSE (SG) AND BLOOD GLUCOSE (BG)
IMPORTANT: When using CGM, focus less on each individual glucose number and more
on the direction and speed that your glucose is changing.
NOTE: You may be likely to notice your glucose trending up or down after eating, giving a
bolus, or when exercising.
IMPORTANT: Sensor glucose is not the same as blood glucose. Your SG and BG
readings will be close to one another, but will rarely match exactly.
Sensor glucose values should not be used to make diabetes treatment decisions.
Always confirm your glucose value with a BG meter first.
Your BG meter measures glucose
levels in your blood. The glucose
sensor measures glucose in the fluid
surrounding the cells of your tissue
called interstitial fluid.
Glucose travels between these two
areas (blood and interstitial fluid). Most
of the time, it travels to your blood
first, and then to your interstitial fluid.
Because of how glucose moves, your
BG meter readings (BG) and sensor
glucose readings (SG) will be close, but
will rarely match exactly. This difference
is normal and should be expected.
When glucose levels are rising or falling
quickly, you should expect to see an
even larger difference between your BG
meter readings and the sensor glucose
readings.
Examples of times when this larger
difference may occur include:
After meals or taking a bolus of insulin
During exercise
When arrows appear on your pump
screen as explained in the next
section
By looking at the sensor information
above, you can see that your current
glucose reading is 5.6 mmol/L. When
you look at the graph, you can see that
you are trending downward.
Furthermore, you see arrows above the
number. These arrows indicate the rate
that your glucose values are moving up
or down:
or - SG has been rising or falling
by about 1-2 mmol/L over the last 20
minutes
or
- SG has been rising or falling by
about 2-3 mmol/L over the last
20 minutes
or
- SG has been rising or falling
by about 3 mmol/L over the last
20 minutes
The Low Settings include alerts, as well as
the SmartGuard™ technology features.
You can choose to be alerted if your
sensor glucose:
is approaching your low limit
(Alert before low)
has reached your low limit
(Alert on low)
The SmartGuard technology features
can automatically suspend insulin if your
sensor glucose:
is approaching your low limit
(Suspend before low)
has reached your low limit
(Suspend on low)
This will keep you from getting additional
insulin that would continue to lower
your sensor glucose level.
Your low (Lo) limit can be set from 2.8 to 5.0 mmol/L. This is the value on which the
other low settings described below are based. You can set up to eight low limits for
different periods of the day or night.
22.2 mmol/L
( Low Snooze)
( High Snooze)
Suspend on Low
Alert on Low
Rise Alert
Alert on High
Suspend before Low
Alert before Low
2.2 mmol/L
High Limit
Alert Before High
Sensor
Glucose
Readings
SECTION 12: PERSONALISED ALERTS
Your CGM alert and suspend settings
are most beneficial if they are
personalised for your needs. Your
healthcare professional will work with
you to determine your initial settings
and help with adjustments that need to
be made, as you learn more from the
information that CGM provides.
The graph below shows an example
of the different settings that can be
personalised for both High and Low
sensor glucose readings.
TURNING SENSOR FEATURE ON
Before setting any of these sensor alerts, you must first turn the sensor feature on.
To turn the sensor feature on, go to the Menu > Sensor Settings and select Sensor.
13.9
4.0
Personalised Alerts
CGM Alert Setting
( Low Snooze)
( High Snooze)
Suspend on Low
Alert on Low
Rise Rate
Alert before High
Alert on High
Suspend before Low
Alert before Low
Sensor Glucose
Readings
CGM I PERSONALISED ALERTS
CGM I PERSONALISED ALERTS
Alert
Reason
Steps to take
Alert before low Alert
before low
If Suspend before low is on, you will
be alerted when insulin is suspended.
If Suspend before low is off, you will
be alerted when the sensor predicts
you will reach your low limit in 30
minutes.
Do not treat your glucose based on
SG. Confirm it using your BG meter.
Treat if necessary based on
instructions from your healthcare
professional and continue to monitor.
Alert on low
Sensor glucose value is equal to or
lower than your low limit.
RESUMING BASAL INSULIN
From the Menu, go to Sensor Settings
> Low Settings and select Low Settings
to turn On.
Press
on the time
segment.
Press
to continue onto
the next screen.
Press
or
to set Lo limit and press
.
SETTING UP YOUR LOW SETTINGS:
In this example, we will set up multiple time segments with different alert and
suspend settings.
CGM I PERSONALISED ALERTS
CGM I PERSONALISED ALERTS
NOTE: If either Suspend on low or Suspend before low is turned on, Alert on low will
automatically be set to on so you know that your glucose is at or below your low limit.
SmartGuard
Suspend features
Impact on insulin
delivery if suspend
feature is turned on
Information displayed by the pump
Suspend before
low
Insulin delivery
is temporarily
stopped if sensor
glucose value is
approaching your
low limit.
You will receive this alert
message and need to
check your BG. Insulin
delivery will remain
suspended after the
alert is cleared.
If the alert is not cleared in
10 minutes, the pump will
begin to siren.
After the alert or alarm
message is cleared and
insulin delivery has stopped,
the Home screen will
display:
Suspended before low
or Suspended on low at
the bottom of the screen
a shaded area to
represent the time
when insulin has been
suspended
a flashing SmartGuard
technology icon.
Suspend on low
Insulin delivery
is temporarily
stopped if sensor
glucose value has
reached or fallen
below your low
limit.
You will receive this alarm
message and need to check
your BG. Insulin will remain
suspended after the alarm
is cleared.
If the alarm is not cleared
after 10 minutes, the pump
will begin to siren and an
emergency message will
appear on the pump screen.
NOTES: Only one suspend feature can be used during each time segment; you cannot
turn both Suspend before low and Suspend on low on.
Insulin delivery will not be suspended if you are more than 3.9 mmol/L above
your low limit.
1
3
4
2
Automatic Basal Resume
In addition to suspending insulin delivery,
the pump can also automatically
resume delivery of basal insulin. If insulin
has been suspended by either Suspend
before low or Suspend on low, insulin
delivery will automatically be resumed:
if SG values are above the low limit and
are trending upward. If you have the
Resume basal alert on, you will be
alerted when this occurs.
after a maximum suspend time of 2
hours. You will always be alerted (even
if the Resume basal alert is off) when
this occurs. It is important that you
check your BG and ensure your
glucose is at a safe level.
Manual Basal Resume
You can choose to resume basal insulin
delivery yourself at any time. You simply
need to select Suspended before/on
low on the Home screen and follow the
instructions on the screen.
Continued on next page
If snooze time needs to be changed,
press
to Snooze and press
.
The low snooze time can be set
from 5 minutes to 1 hour.
Press
or
to the correct time
and press
.
HIGH SETTINGS
The High Settings allow you to be alerted if your sensor glucose:
is rising rapidly (Rise Alert)
is approaching your high limit (Alert before high)
has reached your high limit (Alert on high)
22.2 mmol/L
Sensor
Glucose
Personalized Alerts
CGM Alert Setting
( Low Snooze)
( High Snooze)
Suspend on Low
Alert on Low
Rise Alert
Alert on High
Suspend before Low
Alert before Low
2.2 mmol/L
High Limit
Alert Before High
Sensor Glucose
Readings
Select each feature you wish to
turn on. In this example, Suspend
before low has been turned
on. Notice that Alert on low is
automatically turned on.
Once settings are selected,
select Next.
CGM I PERSONALISED ALERTS
CGM I PERSONALISED ALERTS
REMEMBER: Sensor glucose values must be confirmed with a BG meter reading before
diabetes treatment decisions can be made.
10
5
11
YOUR LOW SETTINGS SETUP IS NOW COMPLETE.
6
Verify that settings are correct
and select Save.
Select Done.
8
9
Press
on the time segment. Repeat steps 3 to 7 to enter the next time segment
and select the features you want to turn on for this segment. In this example,
Alert before low, Suspend on low, and Resume basal alert have been turned on.
7
From the Menu, go to Sensor
Settings > High Settings and
select High Settings to turn On.
If you are changing settings that
are already entered, press
to Setup and press
.
SETTING UP YOUR HIGH SETTINGS:
Press
on the time segment. If you
are setting multiple time segments with
different high limits and alerts, press
to set the first End time and press
.
In this example, only one time segment is
set.
Press
or
to set Hi limit and
press
. In this example, the limit
is set to 13.8 mmol/L.
Press
to continue onto the
next screen and select each
alert you wish to turn on.
*Time before high determines how many minutes before reaching the high limit that you will
receive an Alert before high. This can be set from 5 to 30 minutes.
Your high (Hi) limit can be set from 5.6 to 22.2 mmol/L. This is the value on which
other high settings described below are based.
CGM I PERSONALISED ALERTS
CGM I PERSONALISED ALERTS
REMEMBER: Your high limit is not the same as your glucose target. Your healthcare profes-
sional will help you determine the best setting so that you are alerted when needed while
preventing unnecessary or inconvenient alerts.
REMEMBER: You can set up to 8 different time segments throughout the day and night.
Each time segment can have different high limits and high alerts that work best for you
during that time of day or night.
Alert
Reason
Steps to take
Alert before high
Sensor glucose reading is
expected to reach the high glucose
limit in the length of time that you
set for the Time before high*.
Do not treat your glucose based
on SG. Confirm it using your
BG meter. Treat if necessary
based on instructions from your
healthcare professional and
continue to monitor.
Alert on high
Sensor glucose value is equal to or
higher to the high limit you set.
Rise Alert
Sensor glucose reading is in-
creasing at a rate that is equal to
or faster than the Rate Limit that
you set.
The Rise Alert can be set to alert if
glucose is rising as follows:
- SG is rising at a rate
of 0.056 mmol/L per
minute or more
-
- SG is rising at a rate
of 0.111 mmol/L per
minute or more
- SG is rising at a rate
of 0.167 mmol/L per
minute or more
Custom - SG is rising at the rate
that you set. This can be
set from 0.050 to 0.275
mmol/L per minute
1
3
4
2
Continued on next page
If a sensor alert occurs when Alert Silence is on, a
Sensor alert occurred message is displayed and
the notification light flashes, but there is no beep or
vibration during the set period of time.
To set Alert Silence: from the Menu, go to Sensor
Settings > Alert Silence.
You can select which alerts you would like to silence
and set the time you want these alerts to be silent for.
Alerts will automatically return to audio and/or vibrate
at the end of the duration that you set.
Once settings are selected,
select Next. In this example,
the Alert on high has been
turned on.
Select Done.
Verify that settings are
correct and select Save.
If snooze time needs to be changed,
press
to Snooze and press
.
The high snooze time can be set from
5 min to 3 hours.
CGM I PERSONALISED ALERTS
CGM I PERSONALISED ALERTS
NOTE: If an alert is received during Alert Silence, go to the Menu > History and select Alarm
History to see the alerts that occurred.
CHANGING HIGH AND LOW SETTINGS
To make changes to your existing High or Low settings, go to Sensor Settings
> High Settings or Low Settings > Setup and select Edit.
YOUR HIGH SETTINGS SETUP IS NOW COMPLETE.
5
7
Press
or
to the cor-
rect time and press
.
9
8
6
A graph that shows the last 3 hours of sensor glucose
readings will always display on the Home screen. Your high
and low glucose limits entered in your sensor settings will
be shown in red.
You can also view 6-hour, 12-hour and 24-hour glucose
trend graphs by selecting the sensor graph. Blue squares
at the bottom of the graph represent a bolus.
A gold shaded area represents time when insulin was
suspended by sensor.
SENSOR STATUS
You can go to the Sensor status menu to see, for example,
when your next calibration is due, time left on your sensor,
and battery life remaining on your transmitter.
From the Home screen, select the Status Bar and select
Sensor. You will also see additional sensor status
information in Notifications, Quick Status, and Settings
Review screens.
SMARTGUARD TECHNOLOGY ICON
During any time segment when either Suspend before low or Suspend on low is set
to on, you will see the Suspend by sensor icon on the Home screen :
Suspend before low or Suspend on low is on and ready. If either suspend
becomes active, the icon will flash while insulin delivery is stopped.
Suspend before low or Suspend on low is on but is unavailable. This can be
due to a recent suspend or when no SG values are available.
SECTION 13: READING THE SENSOR DISPLAY
Once the sensor has started giving you sensor glucose readings, the Home screen will
display them similar to what you see here.
STATUS BAR
In addition to the pump icons, you will see additional sensor icons on the Status Bar
when using CGM.
Connection icon: shows radio frequency (RF)
communication between the pump and sensor.
Calibration icon: represents the time left until next
calibration is due. The icon empties as time decreases.
Down arrow means calibration is needed.
Sensor Life icon: represents the number of days before
sensor needs to be changed.
Additional icons: appear when the sensor is in warm up,
pump and transmitter are out of range, system cannot
be calibrated, or calibration or sensor age are unknown.
High Alert Limit
3-hour Sensor
Glucose Graph
Low Alert Limit
Additional Sensor Icons
Trend Arrows
Most recent Sensor
Glucose (SG) Reading
SmartGuard technology
Icon
Bolus Marker
The Sensor Glucose reading is updated every 5 minutes.
CGM I SENSOR DISPLAY
CGM I SENSOR DISPLAY
Immediately select Search
on the pump. The search
can take up to 2 minutes.
Once device is found, confirm that
the serial number (SN) shown on
the pump is the serial number on
the back of your transmitter and
select Confirm. If you receive the No
devices found message, place the
transmitter back onto the charger.
Then remove the transmitter from
the charger and immediately select
Retry on the pump.
Connection is now complete. The
transmitter serial number will be
displayed on the pump screen.
!
Remove transmitter from
charger. If green light on
transmitter does not flash,
reconnect to charger until
fully charged.
SECTION 14: CONNECTING YOUR PUMP AND TRANSMITTER
Before using the sensor for the first time, you will need to wirelessly connect the pump
and transmitter so that they can communicate with each other. This allows the sensor
information to be displayed on the pump screen.
charger
transmitter
Attach your transmitter to
the charger and make sure
it is fully charged.
Press
and select Utilities >
Device Options > Connect Device.
Only one transmitter can be connected
to the pump at one time. When you need
to connect a new transmitter, you must
first select Manage Devices, select the
transmitter number and select Delete.
Select Auto Connect > Press
> Select Continue. Steps for Manual
Connect can be found in the MiniMed 640G
system User Guide.
Make sure the transmitter
is on the charger before
proceeding. Now
start the search processes
on both devices.
TO WIRELESSLY CONNECT YOUR PUMP AND TRANSMITTER:
CGM I CONNECTING YOUR PUMP AND TRANSMITTER
CGM I CONNECTING YOUR PUMP AND TRANSMITTER
NOTE: These steps only need
to be done as a first time set-up.
You will not have to repeat with
each new sensor you start.
NOTE: If you stop using CGM for a period of time and need to store your transmitter, please
make sure to leave it connected to the charger during the storage period. This will help
ensure you get the most life out of your transmitter battery.
1
3
4
4a
5
6
4b
2
FOR GOOD SENSOR GLUCOSE PERFORMANCE, AVOID SITES:
Where clothing may rub or constrict (for example, your beltline)
Where your body naturally bends a great deal which may cause the sensor
to pull out
That are scarred or have hardened tissue or stretch marks
Where there is a great deal of motion or friction
PREPARING YOUR SITE
Wash your hands with soap and water.
Do not use IV prep or the sensor may not work properly.
SELECTING YOUR SITE
Your sensor can be inserted in any of the shaded areas shown here**. The sensor
insertion site should be at least:
5 centimetres from your navel
2.5 centimetres from your insulin
pump infusion site
2.5 centimetres from any manual
insulin injection site
**Clinical trials for glucose sensors were performed on sensors inserted in these areas.
SECTION 15: INSERTING AND STARTING THE SENSOR***
Before you insert your sensor, gather all of your supplies:
One-press serter is required in order to
insert the sensor properly
Enlite sensor is individually packaged
and comes attached to a plastic
pedestal which is necessary for proper
loading into the serter
Sensor overtape is required to keep the
sensor securely in place
Guardian 2 Link transmitter is
connected after the sensor is inserted
and covered with the overtape
*For more details on the system component, consult the User Guides
SYSTEM COMPONENTS*
One-press Serter
A
Sensor Overtape
A – Pre-cut Hole
Enlite sensor & Pedestal
A – Pedestal
B – Needle Housing
C – Sensor
D – Clear Liner
Guardian 2 Link
C
D
A
B
Thumb print
Bump on button
***Please refer to the One-press serter user Guide and your HCP for more information.
CGM I INSERTING AND STARTING THE SENSOR
CGM I INSERTING AND STARTING THE SENSOR
G
TAPING YOUR SENSOR
Before you connect the Guardian 2 Link transmitter to your Enlite sensor it is
very important that you properly secure the sensor against your skin using
the sensor overtape.
Remove large paper backing
from overtape. Do not
remove two smaller paper
tabs on sides of overtape.
Attach the overtape to both
the rounded part of the
sensor and the skin in
front of the sensor.
Important: Make
sure there is
overtape on both
the rounded part
of the sensor and
the skin.
Remove two paper tabs from
sides of overtape.Press
overtape against skin.
Apply rest of overtape, but do
not block sensor connector
with overtape. Press overtape
to your skin for several
seconds.
These images show overtape applied correctly.
Hold sensor by plastic pedestal.
Remove sensor with attached
pedestal by holding pedestal only.
Place sensor/ pedestal on a clean,
flat surface (such as a table).
Tuck adhesive tab.
Make sure that sensor’s
adhesive tab is tucked
under sensor connector
and snaps.
Load sensor into serter.
Grip serter exactly as shown
with thumb on thumbprint marking
on serter. Do not hold side buttons.
Push serter down onto pedestal until
base of serter sits flat on table.
Detach serter from pedestal.
To detach serter from pedestal,
grip serter as shown, with
thumb on thumb print on
serter. With other hand, place
two fingers on pedestal arms,
and slowly pull serter straight
up. Note: Make sure that
pedestal is firmly on table
before pulling serter away.
Warning: Do not detach
pedestal from serter in mid-air
as this may damage sensor.
Correct
Incorrect
Correct
Incorrect
X
Place serter on body. Hold serter steadily
against your cleaned insertion site, without
pushing serter too deeply into skin.
Note: Failing to hold serter securely flat
against body may allow serter to spring
back after pressing buttons and result in
improper insertion of sensor.
Insert sensor. Press and release bump on
both buttons at same time.
Hold serter against body. Continue
holding serter against body to allow
adhesive time to stick to skin.
Remove serter from body. Slowly pull
serter away from skin, making sure buttons
are not pressed.
Remove needle housing. Gently hold base
of sensor against skin with one hand. With
other hand, hold needle housing at top
and slowly pull it straight, away from sensor.
Dispose of needle housing in a sharps
container. Warning: Watch for bleeding at
the insertion site. If bleeding occurs under,
around, or on top of the sensor, apply
steady pressure using sterile gauze or a
clean cloth placed on top of the sensor for
up to three minutes. The use of non-sterile
gauze can cause an infection. If bleeding
does not stop, remove the sensor and
apply steady pressure until the bleeding
stops.
Remove adhesive pad liner.
Hold sensor in place and gently remove liner
from under adhesive pad. Do not remove
the adhesive liner from the rectangular
adhesive tab.
Press entire adhesive pad to skin. Firmly
press adhesive against skin and smooth
entire adhesive pad so it sticks to skin.
NOTE: Enlite sensor adhesive is pressure-
sensitive. Continue applying pressure to
ensure sensor remains inserted in skin for
whole 6 days of wear.
Untuck adhesive tab. Untuck adhesive tab
from under sensor connector.
Straighten adhesive tab.
Straighten adhesive tab so
it lies flat against your skin,
but do not remove adhesive
liner yet.
INSERTING YOUR SENSOR
The instructions below only refer to the One-press Serter. If you have a different
serter model, please refer to your serter User Guide for instructions on how to
insert the sensor.
CGM I INSERTING AND STARTING THE SENSOR
CGM I INSERTING AND STARTING THE SENSOR
NOTE: The sensor remains
inside the serter after removing
the pedestal. The arrow on
both sides of serter indicate
location of the sensor needle.
IMPORTANT: All Enlite tapes and adhesives stick best when you apply pressure for several seconds after putting
them on your skin. Doing so helps the Enlite sensor stay securely placed and fully inserted.
Properly applying the overtape is key to ensuring your success with the Enlite sensor. Due to the sensor’s small
size and flexible nature, the overtape helps to secure it from body motion or physical activity that can cause it to
be pulled out.
1a
1
3
5
4
2
4a
5
4b
4c
4d
6a
6b
7a
7b
1b
2
3
IMPORTANT: Applying an additional
piece of tape over the transmitter once
it is connected to the sensor as shown
here helps to ensure the transmitter
and sensor stay secure and connected
during wear. The Enlite overtape or any
other medical tape can be used.
STARTING THE SENSOR
Once you have inserted the sensor and connected the transmitter, the pump
and transmitter will begin to communicate. Make sure your pump is on the Home
screen so that the message below (in step 1) will be displayed when the sensor is
ready to be started. This typically takes less than a minute, but may take up to
10 minutes.
CONNECTING YOUR TRANSMITTER TO YOUR SENSOR
Before connecting the transmitter, make sure the Sensor feature is On.
See page 38 if you need help with these steps.
With one hand, hold
sensor in place. With
other hand, connect
transmitter
to sensor.
You may hear a faint
“click” when the two
components are
connected. Check for
green light to flash on
transmitter.
It is very helpful to remember the order
of these three steps when changing
your sensor:
1 Insert the sensor.
2 Tape the sensor in place.
3 Connect the transmitter.
CGM I INSERTING AND STARTING THE SENSOR
CGM I INSERTING AND STARTING THE SENSOR
IMPORTANT: If you do not see a
green light flashing on the transmit-
ter after it is connected to the sensor,
then disconnect the transmitter and
place it back on the charger to ensure
that it is fully charged. Then recon-
nect the transmitter to the sensor.
If for any reason you disconnect
the transmitter from the sensor,
wait 5 seconds before reconnecting
it to the sensor.
NOTE: The next time you
connect a transmitter, you will see
these screens. Select Start New
Sensor if you have just inserted
a new sensor. Select Reconnect
Sensor if you have only
disconnected and reconnected
the transmitter.
Enlite overtape
Medical tape*
*A piece of Tegaderm™ cut in half is shown
in this image.
NOTE: When your transmitter is
connected to your sensor they form
a water-tight seal to a depth of 2.4
meters for up to 30 minutes. You can
shower and swim without removing
them.
1
2
Remove the paper on
adhesive tab.
Fold adhesive tab over
transmitter.
Be careful not to pull
the adhesive tab too
tightly or it may cause
the transmitter to
pull from the sensor
connection.
4
3
Press adhesive onto
transmitter.
Apply additional tape over
transmitter as needed.
5
Press
and then
to clear. Warm up… will appear on the Home screen until
sensor is ready for the first calibration.
If 15 minutes have passed and the Warm up bar does not appear or it looks like it
is not progressing, look in the Quick Status screen, if you see the time of Next cal
listed, the sensor is in Warm up.
3
Select Start New Sensor.
The Sensor warm-up started
message will appear.
1
2
When receiving a Calibrate now alert, if you cannot calibrate
right away, you can set the Snooze to remind you to calibrate
in the time that you set.
If you plan to test BG and calibrate right away, simply select
Snooze.
Once you select Snooze, Calibration required will appear on
the Home screen until you enter a BG to calibrate.
EXAMPLE OF OPTIMAL CALIBRATION TIMES
Optimal calibration time
Optimal calibration time
Not optimal for calibration
Optimal calibration time
5.6
0
11.1
16.7
2
0
4
6
8
10
12
14
16
18
20
22
1
2
3
4
Glucose level (mmol/L)
Target range
Blood glucose reading
Low limit
Sensor glucose
Time
SECTION 16: CALIBRATING
Your continuous glucose monitoring
system requires blood glucose meter
readings in order to provide you with
sensor glucose readings. These BG
meter readings are entered into the
pump and are for sensor calibrations.
Calibration is essential for optimal CGM
performance. CGM does not eliminate
the need for BG meter tests.
To calibrate, you must test your blood
glucose on your meter and then enter
that value into your pump. The pump will
accept BG meter readings between 2.2
mmol/L and 22.2 mmol/L.
After inserting a new sensor, a
calibration is needed:
Within 2 hours after you connect the
transmitter to your sensor and start
the Warm up period. Your pump will
notify you with a Calibrate now alert
when it is ready for its first calibration.
Again within 6 hours (first day of
inserting sensor only)
Again every 12 hours (when a
calibration is necessary you will receive
a Calibrate now alert).
CGM I CALIBRATING
CGM I CALIBRATING
1
2
4
3
REMEMBER: Calibrations are necessary in order to continue to receive sensor glucose
readings, alerts and alarms.
IMPORTANT: After the first day, the minimum number of calibrations required is one every
12 hours. However, calibrating 3-4 times a day is optimal and these can be done when it is
convenient for you. To help you remember to calibrate, think “before is best” - typically the
best times to calibrate are before meals, before taking insulin, and before bedtime. Also
check for arrows - calibrating when there are 2 or 3 arrows may decrease sensor accuracy
until the next calibration.
CALIBRATING THE SENSOR
There are 5 different ways that you can enter a BG reading to calibrate the sensor.
CALIBRATING THROUGH HOME SCREEN GRAPH
CALIBRATING BY USING THE Contour™ Next LINK 2.4 METER
When you use the compatible Ascensia meter, you will see the meter value
automatically displayed on the home screen, as shown here.
CALIBRATING THROUGH THE BOLUS WIZARD
Select Calibrate Sensor or if
you plan to give a bolus using
Bolus Wizard, select Bolus.
In the Bolus Wizard:
Select Deliver Bolus.
If you have selected Bolus,
select Yes to Calibrate Sensor?
after bolus is delivered.
Select Yes to calibrate sensor.
Select
to the sensor graph,
press
and hold.
Press
or
to enter BG value,
press
and select Calibrate.
CGM I CALIBRATING
CGM I CALIBRATING
The 2 other ways to calibrate your sensor are through:
Sensor Settings: from the Menu, go to Sensor Settings >
Calibrate Sensor, select BG and press
or
to enter
BG value > press
and select Calibrate.
Event Markers: from the Menu, go to Event Markers >
BG > Enter BG > press
> select Save >
select Yes to calibrate sensor.
Once you have entered a calibration BG, the Home screen
will show you that the system is calibrating.
You will start seeing sensor glucose readings again in 5-15
minutes.
You can use the Calibration Reminder to give you notice before the next calibration
is necessary.
The Calibration Reminder defaults On with a reminder time of 1:00 hour and you can
change it by going to the Reminders menu option.
OTHER WAYS TO CALIBRATE
IMPORTANT: If you notice a large difference between your BG meter reading and sensor
glucose readings, wash your hands and do another BG fingerstick test to help ensure a more
accurate reading. Check the sensor site to ensure the sensor overtape is still holding the
sensor in place. If there is still a large difference in glucose readings, another calibration may
be needed to bring the readings closer together again.
1
1
2
2
1
2
CHARGING AND STORING THE GUARDIAN 2 LINK TRANSMITTER
Charge the transmitter before each use. When
the transmitter is charging, a green light will flash
on the charger. This green light will turn off when
the transmitter is completely charged. You will
need to charge the transmitter after each sensor
use. A fully charged transmitter can be used
for a maximum of six days without recharging.
It can take up to an hour to fully recharge.
When you remove the transmitter from the charger,
a green light should flash on the transmitter.
This indicates that it has enough battery power
to be connected to the sensor. If you do not see
the green flashing light on the transmitter place
it back on the charger until it is fully charged.
Store the transmitter, charger, and test plug in a
clean, dry location at room temperature. Although
not required, you may store the transmitter on
the charger. If the transmitter is not in use, you
must charge it at least once every 60 days.
If you connect transmitter to charger and you see no
lights on the charger:
replace the battery in the charger.
While charging your transmitter you see a flashing red
light on the charger: replace the battery in the charger.
While charging your transmitter you see a mix of short
and long flashing red lights on the charger: replace the
battery in the charger and fully charge the transmitter.
Refer to your Guardian 2 Link transmitter and
charger User Guides for more information.
!
SECTION 17: OTHER SENSOR ALERTS
We discussed personalised alerts earlier in Section 12. There are other sensor alerts
that you will receive as well. The most common alerts that you can expect to receive
when using CGM can be found in the table below.
For a complete list of Alerts and Alarms, refer to the MiniMed 640G system User Guide.
APPENDIX I CHARGING AND STORING YOUR TRANSMITTER
CGM I OTHER SENSOR ALERTS
Alert
Reason
Steps to take
Calibrate now
A calibration is needed in order to
receive sensor glucose readings.
Enter BG value into your pump to
calibrate.
Lost sensor signal
Communication between pump
and transmitter has been lost for
30 minutes during or after warm-
up.
Check that the sensor is still
inserted in the skin and the
transmitter and sensor are still
connected. Move your pump closer
to your transmitter.
Calibration not
accepted
Your system was unable to use
the BG meter readings you
entered to calibrate your sensor.
After 15 minutes, enter a new
BG meter reading for calibration.
If you receive a Calibration not
accepted alert on your second
calibration after 15 minutes, a
Change sensor alert occurs.
BG not received
The transmitter was unable to
receive the calibration BG reading
from the pump.
Move your pump closer to your
transmitter and select OK. The
pump will try sending the BG again.
Sensor expired
Sensor has reached the end of its
useful life.
Remove the sensor and follow
the instructions for inserting and
starting a new sensor.
Change sensor
You have received two Calibration
not accepted alerts in a row.
Remove the sensor and follow
the instructions for inserting and
starting a new sensor.
Cannot find sensor
signal
The pump has not received a signal
from the transmitter.
Disconnect and reconnect your
transmitter and sensor and select
OK.
X-RAYS, MRI, OR CT SCAN
If you are going to have an X-ray, MRI, CT scan, or other
type of diagnostic imaging involving radiation exposure,
remove your insulin pump, transmitter, and glucose
sensor and place them outside of the testing area.
WHAT IS CARELINK SOFTWARE?
CareLink Personal software is a web-based
software that allows you to upload information
from your MiniMed™ 640G system to a secure
online (internet) site for viewing.
CareLink software organises all of your insulin
pump, sensor glucose and blood glucose meter
information into reports (charts, tables and
graphs) that can help you track glucose levels,
insulin usage and carbohydrate intake over time.
With CareLink software, you can grant your
healthcare provider online access, so that your
information can be discussed at your next
appointment.
BENEFITS OF CARELINK SOFTWARE
CareLink Personal software makes it
easier to track your glucose levels and
see how they are affected by your insulin
delivery, meals and exercise routines.
CareLink Personal software provides a
secure place to store your information
and uncover patterns in your glucose
control that meter and logbooks alone
cannot reveal.
Information from CareLink software can
help you and your healthcare provider
make more informed therapy decisions
aimed at improving your glucose control.
For more information on how to upload information to CareLink Personal software using
the Contour™ Next LINK 2.4 glucose meter from Ascensia, see page 24.
CARELINK PERSONAL SOFTWARE
APPENDIX I X-RAYS AND TRAVELING BY AIR
APPENDIX I CARELINK SOFTWARE
CareLink reports can help you and your healthcare provider make decisions that
improve your control and fit your lifestyle.
The combination of insulin pump therapy, continuous glucose monitoring and
CareLink software provides you with the tools and information you may need to
optimise your therapy.
TRAVELING BY AIR
If you wear a CGM device, you may need
to stop the wireless communication
between the transmitter and the pump
during the flight.
To temporarily stop wireless
communication, turn Airplane Mode
on. From the Menu, go to Utilities >
Airplane Mode, select Airplane Mode to
turn On and Save.
The transmitter continues to measure
glucose levels when in Airplane Mode.
To resume wireless communication,
turn Airplane Mode off:
When Airplane Mode is turned off
and communication resumes, the
transmitter will send up to 10 hours of
sensor data to your pump.
If Airplane Mode was on for <6 hours:
1) Wait 15 minutes for sensor data to
appear on pump screen
If Airplane Mode was on for >6 hours:
1) Disconnect transmitter from sensor
and then reconnect it.
2) Select Reconnect Sensor when it
appears on the pump screen to begin
sensor warm-up.
3) The sensor data (up to 10 hours) will
appear on the pump.
4) You will be asked to calibrate in 2
hours to resume sensor readings.
Always remember that it is important
when traveling to be extra attentive to
monitoring your glucose and prepared
to respond if needed.
The images below show additional detail
about using the SmartGuard technology
features of your MiniMed 640G system.
Sensor glucose trend
Estimated sensor glucose trend
Sensor glucose trend during suspend
APPENDIX I SMARTGUARD TECHNOLOGY FEATURES
APPENDIX I SMARTGUARD TECHNOLOGY FEATURES
Suspend on low event:
Low Limit
If sensor glucose (SG) reaches your low limit,
insulin delivery will be stopped.
You will always receive a message and alarm
when this occurs.
You will have 10 minutes to respond before
the pump begins to siren and emergency
message appears.
1
Suspend before low event:
Low Limit
To help keep sensor glucose (SG) from
reaching your low limit, insulin delivery will be
stopped if SG is:
at or within 3.9 mmol/L above the low limit
predicted to be approaching the low limit
in 30 minutes
If Alert before low is on, you will receive an
alert when insulin is suspended.
30 min
2
Alert on low during Suspend before low:
Low Limit
If insulin delivery has stopped due to
Suspend before low, SG may still reach your
low limit.
You will always be alerted when this occurs.
You will have 10 minutes to respond before
the pump begins to siren.
3
Automatic basal resume based on sensor
glucose (SG) value:
During Suspend before low or Suspend
on low, basal insulin will automatically
resume if:
SG is above the low limit and trending
upward
insulin has been suspended by sensor
for at least 30 minutes
If Resume basal alert is on, you will receive
an alert when this occurs. Remember you
can manually resume basal insulin at any
time.
Low Limit
30 min
4
2 hours
Automatic basal resume due to 2 hour
maximum suspend:
During either Suspend before low
or Suspend on low, if basal insulin is
not resumed due to SG values, it will
automatically resume after 2 hours.
You will always receive an alert when you
reach the 2 hour maximum suspend
time, even if the Resume basal alert is
set to off. Remember you can manually
resume basal insulin at any time.
Low Limit
5
Low Limit
Suspend by sensor unavailable:
Once basal insulin resumes following
either a Suspend before low or a
Suspend on low, there will be a period
of time when suspend by sensor is
unavailable.
This will most often be 30 minutes if you
respond to the suspend alarm, but can
be up to 4 hours. See the User Guide
for more specific information about this
unavailable period.
30 min
6
WHO TO CONTACT AND WHEN?
CONTACT MEDTRONIC
Please contact Medtronic for further guidance and technical advice on using
your MiniMed pump.
If you have any concerns that your pump isn’t functioning correctly.
If your pump displays a warning sign or alarm which you cannot switch off.
For more information about a certain pump function.
For guidance when adjusting your basal insulin dose, as instructed by your doctor.
CONTACT YOUR HEALTHCARE PROFESSIONAL
For all other inquiries regarding your health and continuing care please contact
your healthcare professional.
Alternatively call our customer
support helpline 01923 205167
Visit our website at:
www.medtronic-diabetes.co.uk
Europe
Medtronic International Trading Sàrl.
Route du Molliau 31
Case postale
CH-1131 Tolochenaz
www.medtronic.eu
Tel: +41 (0)21 802 70 00
Fax: +41 (0)21 802 79 00
United Kingdom
Medtronic UK Ltd.
Building 9
Croxley Park
Watford, Hertfordshire
WD18 8WW. UK
www.medtronic.co.uk
www.medtronic-diabetes.co.uk
Tel: +44 (0)1923 205 167
Fax: +44 (0)1923 241 004
Medtronic Ireland
Cherrywood Business Park
Block G
Cherrywood Science and
Technology Park
Loughinstown
Co. Dublin
Ireland
www.medtronic.ie
www.medtronic-diabetes.ie
UC201804963 EN REV E Date of preparation: September 2017