Click Gateways and then Configuration Profiles.
Click Add Configuration Profile and type a name (A)
for the configuration. Select the CX logger type. If
the gateway will be used in a storage application
(i.e. a clinic or office), enter these settings (B):
• Select Download and Restart for when to
connect.
• Select “Immediately connect to a logger with
a new sensor alarm.”
• DO NOT select “Immediately connect to a
logger not seen by this gateway.”
• Choose to connect to loggers every week.
Click Save (C).
Insert the appropriate plug for your region into
the AC adapter. Connect the AC adapter to the
gateway and plug it in. Wait a couple of minutes
for the gateway to completely power up. The
LED on the gateway will be yellow-green while it
powers up and then change to deep green once it
is ready to set up.
Upload Data from the Logger
to the Cloud via the Gateway
21853-F MAN-CX5000-QSG
Quick Start for the InTemp
® CX5000 Gateway
4
5
I am setting up a CX5000 gateway for automatic
download of CX loggers to the InTempConnect®
cloud.
Tip: Refer to the full product manual for details on other
configuration settings. Go to www.onsetcomp.com/intemp/
resources/cx5000-manual.
Connect
Connect
Connect
Mobile
Mobile
1
Go to www.intempconnect.com to set up an
account or log into your existing account (A).
A
2
In InTempConnect, click Settings and then Roles.
Click Add Role (A) to create a role for the person
who will be setting up the gateway or click an
existing role to edit it.
A
3
Type the name of the role (A) for the Description.
Select “Login to InTempConnect” plus the gateway-
related privileges (Manage Gateways, Manage
Logger/Gateway Configs, Manage Shipments,
and Gateway Administrator) from the Available
Privileges list on the left and click the right arrow
button (B) to move them to the Assigned Privileges
list on the right. Select a user from the Available
Users list on the left and click the right arrow
button (C) to move the user to the Assigned Users
list on the right. Click Save (D).
Tip: If you need to add a user, click the Users tab and
then Add User, fill in the fields, and assign the user a role
with gateway privileges.
A
B
C
D
C
A
B
Tap Use Current WiFi Network (G) and tap Save (H).
Deploy the gateway. Some guidelines when
selecting a location for the gateway:
• The gateway can support up to 50 loggers
within transmission range, which optimally
is 30.5 m (100 ft) will full line-of-sight.
• Test the range by positioning your mobile
device where you want to deploy the
gateway. If the mobile device can connect
to the logger with the InTemp app from
that location, then the gateway should be
able to connect to the logger as well.
• If you plan to use the gateway for loggers
in different rooms, be sure to test the
connection first.
The gateway is ready for use with loggers. See
www.intempconnect.com/help for details on
configuring and downloading loggers with the
gateway, setting up shipments, and more.
Set up Wi-Fi network
settings. Make sure the
device with the InTemp
app can connect to the
Wi-Fi network. Open
the app and log in as an
InTempConnect user (A).
7
A
B
1-800-LOGGERS (564-4377) • 508-759-9500
www.onsetcomp.com/support/contact
Tip: If the gateway will be using Ethernet with DHCP, plug in the
Ethernet cable. Network settings will be configured automatically.
To configure network settings for Ethernet with static IP addresses
or to set up a gateway for use at another site, refer to the product
manual at www.onsetcomp.com/intemp/resources/cx5000-manual.
Devices
Tap Devices (B) and tap Gateways (C). The gateway
appears in the list. Tap anywhere in the row (D) to
connect to the gateway.
Tap Network Settings (E) and then tap WiFi (F).
C
D
E
F
G
H
Select the gateway configuration (B) you created
in InTempconnect in step 4 (swipe left and right if
there are multiple configurations). Type a name for
the gateway deployment (C) and tap Start (D).
B
C
D
© 2017–2019 Onset Computer Corporation. All rights reserved. Onset, InTemp, and InTempConnect are registered trademarks of
Onset Computer Corporation. Bluetooth is a registered trademark of Bluetooth SIG, Inc. All other trademarks are the property of
their respective companies.
21853-F MAN-CX5000-QSG
Refer to the CX5000 Gateway manual
for full product details, including
specifications and mounting instructions.
Scan the code or go to www.onsetcomp.
com/intemp/resources/cx5000-manual.
Once a gateway has contacted InTempConnect for
the first time, a user is created for the gateway in
your account. Click Settings and then the Users tab
to see the gateway user name, which is listed as
CX5000-.
7
Connect to the gateway (see step 6) and tap
Configure (A).
A
(continued)