Rainin Pipet-Lite Manual
Read Online
Read the Rainin Pipet-Lite Manual online using the built-in PDF viewer.
Text of the manual
10 pipettes for
volume ranges
from
0.1 µL to 20 mL
Pipet-Lite 200 µL shown
Pipet-Lite
Magnetic Assist Pipette
™
Pipet-Lite, LTS, LiteTouch, and Hang-Ups are trademarks of Rainin Instrument,
LLC. Pipet-Lite pipettes are manufactured under U.S. Patent Nos. 5,614,153,
5,700,959, and 5,849,248. LTS Lite Tip Ejection System is protected by U.S.
patents 6,168,761 B1 and 6,171,553 B1. Other national patents pending.
Table of Contents
Introduction..................................................................1
Setting Volume..............................................................2
Filter............................................................................3
Tip Selection and Mounting ............................................3
Tip Immersion Depth .....................................................4
Operation.....................................................................4
Pipetting Guidelines.......................................................5
Pre-Rinsing Recommended............................................5
Reverse-Mode Pipetting .................................................5
Pipetting Liquids of Varying Density.................................6
Temperature Considerations ...........................................6
Autoclaving ..................................................................6
Tip Ejector Arm Removal................................................7
Pipet-Lite Storage..........................................................7
Troubleshooting and Repairs..........................................8
Service, Calibration, and Repair......................................9
Specifications .............................................................10
Replacement Parts ......................................................11
Contacting RAININ............................................Back Page
Figures
Figure 1
Pipet-Lite Pipette........................................1
Figure 2
Filter Orientation.........................................3
Figure 3
Removing the Tip Ejector Arm......................7
Line Drawings
Pipet-Lite
2 µL
................................................12
Pipet-Lite
10 µL
................................................12
Pipet-Lite
20 µL
................................................12
Pipet-Lite
100 µL
................................................12
Pipet-Lite
200 µL
................................................12
Pipet-Lite 1000 µL
................................................13
Pipet-Lite 2000 µL
................................................13
Pipet-Lite 5000 µL
................................................13
Pipet-Lite 10 mL
................................................13
Pipet-Lite 20 mL
................................................13
Pipet-Lite 200 µL
Plunger button /
volume adjustment
Tip ejector
button
Stainless steel tip
ejector arm
Volume lock
Digital
volume
indicator
Body:
high-impact,
low thermal
conductivity
polymer
Shaft
Disposable tip
Introduction
Pipet-Lite is an air-displacement pipette which incorporates
major ergonomic improvements to reduce the risk of repeti-
tive strain injury and pain – low force seals and lower spring
forces contribute to a lighter feel. A magnet is used to help
sense and hold the piston in the zero position, which reduces
static force on the operator’s hand.
In the models with LTS™ shafts, the patented LiteTouch™ Tip
Ejection System reduces total pipetting forces by up to 70%.
Tip ejector arm
1. Turn the volume lock counter-clockwise to the position
shown at left below so the volume setting mechanism is
unlocked and free to turn.
2. With the mechanism unlocked, orient Pipet-Lite so you
are looking at the digital volume indicator, then rotate the
plunger button to change volume – counter-clockwise to
increase, and clockwise to decrease volume.
The volume indicator is read from the top down.
2–20 µL: Black – µL. Red – tenths, hundredths of µL.
100–200 µL: All digits black – whole µL.
1000–5000 µL: Red – mL. Black – tenths, hundredths of mL.
10 mL: Red – mL. Black – tenths of mL
20 mL: Red – mL. Black – tenths of mL.
3. To eliminate errors due to mechanical backlash: when set-
ting the desired volume, first turn the knob 1⁄3 turn above
the desired volume. Then turn the knob slowly clockwise
until the desired volume is displayed. Always dial down to
the desired volume.
4. Turn the volume lock clockwise (see diagram above) to
prevent accidental changes to the volume setting.
Example volumes for the 200 µL model are shown below
(note the intermediate setting at the right).
2
Red digits
Black digits
Volume ranges and increments are shown below:
Pipet-Lite
Range (µL)
Increment
Volume
Adjustable
Recommended
µL
2 µL
0 to 2
0.1 to 2
0.002
10 µL
0 to 10
0.5 to 10
0.02
20 µL
0 to 20
2 to 20
0.02
100 µL
0 to 100
10 to 100
0.2
200 µL
0 to 200
20 to 200
0.2
1000 µL
0 to 1,000
100 to 1,000
2.0
2000 µL
0 to 2,000
200 to 2,000
2.0
5000 µL
0 to 5,000
500 to 5,000
5.0
10 mL
0 to 10 mL
1 mL to 10 mL
20.0
20 mL
0 to 20 mL
2 mL to 20 mL
20.0
Filter
Pipet-Lite 5000 µL, 10 mL, and 20 mL pipettes use a filter in
the end of the shaft to help prevent liquid entering the shaft
and contaminating the piston, should the plunger snap up
during aspiration. Using such a filter is particularly important
when pipetting large volumes. Replace the filter if it gets wet.
Figure 2 Filter Orientation
The 5000 µL and 10 mL pipettes use the same filter oriented
as shown in the diagram above:
5000 µL: insert the small diameter into the shaft.
10 mL: insert the large diameter into the shaft.
Filter part numbers are 6190-164 (pack of 100) and 6190-
165 (pack of 1000).
The filter for 20 mL is a cylinder. Filter part numbers: 6190-
221 (pack of 100) and 6190-222 (pack of 500).
Tip Selection and Mounting
Always use RAININ tips with Pipet-Lite pipettes. Pipet-Lite
pipettes are calibrated with RAININ tips, and performance to
published specifications can only be guaranteed when RAIN-
IN tips are used.
To mount a tip, press the Pipet-Lite shaft into the end of the tip
with light force. The tip will seal properly on the shaft with
minimal force — do not use more force than is required.
3
Tip Immersion Depth
The recommended depth for tip insertion into the sample for
each Pipet-Lite volume is shown below.
Pipet-Lite
Volume Range
Immersion Depth
2 µL
0.1 - 2 µL
1-2 mm
10 µL
0.5 - 10 µL
1-2 mm
20 µL
2 - 20 µL
2 - 3 mm
100 µL
10 - 100 µL
2 - 3 mm
200 µL
20 - 200 µL
2 - 4 mm
1000 µL
100 - 1000 µL
2 - 4 mm
2000 µL
200 - 2000 µL
2 - 6 mm
5000 µL
500 - 5000 µL
6 - 10 mm
10 mL
1 mL - 10 mL
6 - 10 mm
20 mL
2 mL - 20 mL
6 - 10 mm
Tip immersion depth is critical. If these depths are exceeded,
the volume measured may be inaccurate, possibly out of
specification. Tip angle is also important. Hold the pipette
within 20 degrees of vertical.
Operation
Before pipetting valuable samples, it is a good idea to prac-
tice aspirating and dispensing water before pipetting with
actual samples.
1. Set the desired volume as described on page 2.
2. Attach a new RAININ tip. Press the shaft into the tip with
only sufficient force to make a good seal.
3. Press the plunger button to the FIRST STOP, and hold it in
this position. The magnetic latch will help you sense and
hold this position.
4. Holding Pipet-Lite vertically, place the tip into the sample
to the proper depth and relax your thumb pressure on the
plunger. The light piston spring will move the piston
upward, aspirating sample. Do not let go of the plunger
button, or the piston may snap up quickly, resulting in
inaccurate measurement.
5. Pause briefly (longer for macrovolume pipettes) to ensure
that the full volume of sample is drawn into the tip.
6. Withdraw the tip from the sample. If any liquid remains on
the outside of the tip, wipe it carefully with a lint-free tis-
sue, taking care not to touch the tip orifice.
1. Touch the tip end against the side wall of the receiving
vessel and press the plunger slowly, past the FIRST STOP,
to blowout (bottom of stroke.) Wait: 1 second for
2–200 µL volumes, 1-2 seconds for 1000 µL and larger.
(Longer for viscous solutions.)
2. Still holding the plunger, withdraw the tip, sliding it along
the wall of the vessel. Release the plunger.
3. Press the tip ejector button lightly to discard the tip. Use a
new tip for each sample to prevent carry-over. Repeat for
the next pipetting cycle.
Pipetting Guidelines
Pipet-Lite pipettes incorporate several new features which
enhance pipetting consistency. You should also maintain:
1. Consistent pickup and dispense rhythm.
2. Consistent speed and smoothness when pipetting.
3. Consistent pressure on the plunger button at the FIRST
STOP.
4. Consistent immersion depth. See table on page 4.
5. Pipette vertically, or within 20° of vertical.
6. Don’t invert or lay the pipette flat with liquid in
the tip.
Pre-Rinsing Recommended
Some solutions may leave a film on the inside tip wall. This
film remains relatively constant in successive pipettings with
the same tip, so excellent precision can be obtained by re-
filling the tip and using the refilled volume as the sample.
Successive samples from this same tip will exhibit good
reproducibility.
Reverse Mode Pipetting
Another way of reducing error due to film retention, especially
useful for more viscous liquids, is reverse mode pipetting. The
operating sequence is reversed:
1. Mount a disposable tip on the pipette shaft.
2. Press the pushbutton fully to the SECOND STOP.
3. Immerse the tip in liquid and allow the button to return
slowly to the fully UP position. Wait a moment for the liq-
uid column to reach equilibrium in the tip.
5
4. Wipe any excess liquid from the outside of the tip without
touching the orifice.
5. To dispense, rest the end of the tip against the vessel wall
and press the plunger to the first stop. Hold this position a
few seconds, or long enough for the liquid column to
reach equilibrium again.
6. Remove the tip from the receiving vessel without blowing
out the remaining liquid.
7. Return excess sample in the tip to the original sample
container, if desired. Discard the used tip.
Pipetting Liquids of Varying Density
Pipet-Lite lets you compensate for solutions of density much
different from water, by setting the volume slightly higher or
lower than that required. The compensation amount must be
determined empirically.
For example, if pipetting 10 µL of CsCl solution, you deter-
mine that the volume delivered is actually 8.5 µL (≥5 sam-
ples). Change the volume setting to 11.8 µL and repeat the
measurements. If the volumes delivered are still not close
enough to 10 µL, make another slight volume adjustment
until the measurements are as desired.
Temperature Considerations
Warm or cold liquids can be measured with good precision
by using a consistent pipetting rhythm. This will help mini-
mize any differences in heating or cooling effects within the
pipette.
Use a new tip each time for best accuracy and precision
when measuring samples with temperatures greatly different
from ambient, and do not pre-rinse. As with any air-displace-
ment pipette, you will get best results if there is no delay
between picking up the sample and dispensing it.
Autoclaving
Autoclavable parts of Pipet-Lite are the shaft and the tip ejector:
121°C, 1 bar, 15-20 minutes.
Do not autoclave the complete pipette or any parts
other than the shaft and the tip ejector.
Tip Ejector Arm Removal
Three types of tip ejector are used and all types can be
removed with minimum effort - do not use force.
For models up to 2000 µL, press in the quick-release tabs
on the ejector arm and pull the ejector down.
For 5000 µL and 10 mL models, grasp the top of the ejector
arm and pull outward then downward.
For the 20 mL model, pull off the lower part of the ejector
arm; the upper part stays in place. (Replace by aligning the
rod with the hole in the upper part and pressing firmly.)
To replace the ejector arm (except 20 mL), insert the shaft
through the large opening, align the top with the tip ejector
pushrod, and push until the ejector arm snaps in place.
Pipet-Lite Storage
After use, store the pipette in a clean safe place. Pipet-Lite is
a precision instrument and should be treated with the level of
care appropriate for laboratory instrumentation. Three types
of hanger are available to hold your Pipet-Lite conveniently
when not in use.
CR-7:
Free-standing carousel holds seven pipettes.
HU-M3: Set of three individual magnetic Hang-Ups™ for
mounting on ferrous surfaces.
HU-S3: Three Hang-Ups attached to a clamp which fits onto
a shelf.
7
Figure 3 Removing the Tip Ejector Arm
Troubleshooting and Repairs
WARNING:
When removing the shaft from the pipette body, make sure
the spring, seal and o-ring do not fall off the piston, espe-
cially on the smaller models.
Sample Splash (liquid inside the mechanism)
1. Remove the tip ejector arm. Also refer to the appropriate
drawing on page 12 or 13.
2. On pipettes up to 1000 µL, unscrew the shaft coupling (B)
and remove the shaft (C). For 2000 µL, unscrew the shaft.
For 5000 µL, 10 mL and 20 mL unscrew the lower part
of the shaft. Note the way the spring, seal and O-ring fit on
the piston.
3. Inspect the seal assembly and piston for contamination.
The piston should be shiny and free of corrosion. Clean
with distilled water or isopropyl alcohol. Dry with a lint-
free tissue and reassemble after inspecting the interior of
the shaft for any contamination. For 5000 µL, 10 mL
and 20 mL models, apply a small amount of grease to
the seal.
4. If piston corrosion or staining is evident, do not use the
instrument. Call 800-662-7027 for Service.
Leaks, Inaccuracy, Abnormal Stroke
1. Loose shaft. Tighten coupling by hand.
2. Split or cracked shaft. Remove the tip ejector and
inspect the shaft. Replace the shaft if necessary. If the
shaft was dropped, remove it to see if the piston is bent.
If so, return the instrument for service.
3. Worn seal and / or o-ring. All models up to 2000 µL
incorporate a polyethylene seal and o-ring. Examine the
seal and o-ring, replacing them as necessary. Pull off the
old seal and o-ring, position the new seal and o-ring on
the piston assembly as shown in the drawings on pages
12-13, and reassemble the pipette.
Note: It is NOT necessary to recalibrate Pipet-Lite after
changing seals.
4. Do not lubricate any components*. Pipet-Lite uses a dry
sealing system. *Except 5000µL, 10 mL, and L-20ML
models. These pipettes require a small amount of grease
to be applied to the o-ring to make the seal.
8
Acids and Corrosives
After pipetting concentrated acids or highly corrosive solu-
tions, disassemble Pipet-Lite and inspect and clean the pis-
ton assembly, shaft, and seal with distilled water. Dry all
components thoroughly and reassemble.
Extensive contact with corrosive fumes may result in prema-
ture seal wear and damage to the piston. Exposure of inter-
nal components to corrosive fumes can be reduced by using
RAININ tips with aerosol barrier filters.
Service, Calibration and Repair
RAININ Pipette Repair and Calibration facilities:
California: 7500 Edgewater Drive, Oakland CA 94621
Tel. 800-662-7027, Fax 510-564-1683
Massachusetts: Rainin Road, Woburn, MA 01801
Tel. 800-662-7027, Fax 781-935-7631
Japan: 4-1-11, Bunkyo-Ku, Tokyo 113-0033
Tel. (03) 5689-8311, Fax (03) 5689-2670
METTLER TOLEDO Pipette Repair and Calibration facilities:
Belgium: N.V. Mettler-Toledo s.a., B-1932 Zaventem
Tel. (02) 334 02 11, Fax (02) 334 03 34
Germany: Mettler-Toledo GmbH, D-35353 Giessen
Tel. (0641) 50 70, Fax (0641) 507 128
Denmark: Mettler-Toledo A/S, DK-2600 Glostrup
Tel. (43) 270 800, Fax (43) 270 828
Spain: Mettler-Toledo S.A.E., E-08038 Barcelona
Tel. (93) 223 76 00, Fax (93) 223 02 71
France: Mettler-Toledo s.a., F-78222 Viroflay
Tel. (01) 309 717 17, Fax (01) 309 716 16
Italy: Mettler-Toledo S.p.A., I-20026 Novate Milanese
Tel. (02) 333 321, Fax (02) 356 29 73
Netherlands: Mettler-Toledo B.V., NL-4004 JK Tiel
Tel. (0344) 638 363, Fax (0344) 638 390
Sweden: Mettler-Toledo AB, S-12008 Stockholm
Tel. (08) 702 50 00, Fax (08) 642 45 62
Service is also available in many other countries through
authorized RAININ distributors. See www.rainin-global.com.
Note: It is recommended to use only genuine RAININ replace-
ment parts such as seals and shafts. It is NOT necessary to
recalibrate the pipette after changing the seal or shaft. Recali-
bration of the pipette is only necessary when the piston is
replaced, and should be done only by qualified factory-trained
personnel in one of the above-mentioned facilities.
These manufacturer’s specifications should be used as guide-
lines when establishing your own performance specification in
accordance with ISO 8655.
Pipet-Lite Specifications
Volume
Increment
Accuracy
Precision
Model
µL
µL
%
µL (±)
%
µL (≤)
2 µL
0.2
0.002
12.0
0.024
6.0
0.012
1.0
2.7
0.027
1.3
0.013
2.0
1.5
0.030
0.7
0.014
10 µL
1.0
0.02
2.5
0.025
1.2
0.012
5.0
1.5
0.075
0.6
0.03
10.0
1.0
0.1
0.4
0.04
20 µL
2
0.02
7.5
0.15
2.0
0.04
10
1.5
0.15
0.5
0.05
20
1.0
0.2
0.3
0.06
100 µL
10
0.2
3.5
0.35
1.0
0.1
50
0.8
0.4
0.24
0.12
100
0.8
0.8
0.15
0.15
200 µL
20
0.2
2.5
0.5
1.0
0.2
100
0.8
0.8
0.25
0.25
200
0.8
1.6
0.15
0.3
1000 µL
100
2
3.0
3.0
0.6
0.6
500
0.8
4.0
0.2
1.0
1000
0.8
8.0
0.15
1.5
2000 µL
200
2
3.0
6.0
0.6
1.2
1000
0.8
8.0
0.2
2.0
2000
0.8
16.0
0.12
2.4
5000 µL
500
5
2.4
12.0
0.6
3.0
2500
0.6
15.0
0.2
5.0
5000
0.6
30.0
0.16
8.0
10 mL
1 mL
20
5.0
50.0
0.6
6.0
5 mL
1.0
50.0
0.2
10.0
10 mL
0.8
80.0
0.16
16.0
20 mL
2 mL
20
5.0 100.0
0.6
12.0
10 mL
1.0 100.0
0.2
20.0
20 mL
0.8 160.0
0.16
32.0
Specifications are subject to change without notice.
(see pages 12-13)
Legend for Pipet-Lite 2 µL to 2000 µL
A - Plunger Button
B- Shaft Coupling
C - Shaft*
D - Tip Ejector*
E - Piston Assembly
F - Seal
G - O-ring
H - Stroke Spring
J - Seal Retainer
L- Series:
L-2
L-10
L-20
L-100
L-200
L-1000
L-2000
A
6202-055 6202-056 6202-057 6202-058 6202-059 6202-060 6202-192
B
6202-062 6202-062 6202-062 6202-062 6202-062 6202-062 n/a
C
6202-063 6202-064 6202-065 6202-066 6202-229 6202-230 6202-214
D
6202-071 6202-071 6202-071 6202-073 6202-231 6202-232 6200-168
E
6202-076 6202-077 6202-078 6202-079 6202-080 6202-081 6202-082
F
6200-131 6200-138 6200-143 6200-150 6200-154 6200-161 6200-166
G
6200-132 6200-139 6200-170 6200-151 6200-155 6200-162 6200-167
H
6202-083 6202-083 6202-083 6202-084 6202-084
n/a
n/a
J
6200-196 6200-196 6200-198 6200-201 6200-200
n/a
n/a
SL- Series:
SL-2
SL-10
SL-20
SL-100
SL-200
SL-1000
SL-2000
A
6202-109 6202-110 6202-111 6202-112 6202-113 6202-114 6202-115
B
6202-062 6202-062 6202-062 6202-062 6202-062 6202-062 n/a
C
6200-134 6200-140 6200-145 6200-147 6200-382 6200-383 6200-169
D
6200-133 6200-133 6200-144 6200-148 6200-231 6200-232 6200-168
E
6202-183 6202-184 6202-185 6202-079 6202-080 6202-081 6202-082
F
6200-131 6200-138 6200-143 6200-150 6200-154 6200-161 6200-166
G
6200-132 6200-139 6200-170 6200-151 6200-155 6200-162 6200-167
H
6202-083 6202-083 6202-083 6202-084 6202-084 n/a
n/a
J
6200-196 6200-196 6200-198 6200-201 6200-200 n/a
n/a
* These parts are autoclavable (C - shaft, D - tip ejector arm)
Legend for Pipet-Lite 5000 µL , 10 mL and 20 mL
A - Plunger Button
C- Tip Ejector*
D - Piston Assembly
E - Piston O-ring
F - Cylinder
G - Cylinder O-ring H - Shaft*
11
L- Series:
SL- Series:
L-5000
L-10ML
L-20ML
SL-5000
SL-10ML
A
6202-193
6202-194
6202-297
6202-217
6202-218
C
6200-373
6200-374
6202-298
6200-373
6200-374
D
6202-215
6202-216
6202-296
6202-215
6202-216
E
6200-363
6200-369
6202-299
6200-363
6200-369
F
6200-365
6200-371
6202-301
6200-365
6200-371
G
6200-364
6200-370
6202-300
6200-364
6200-370
H
6202-222
6202-223
6202-302
6200-362
6200-368
* These parts are autoclavable (C - tip ejector arm, H - shaft)
Common parts for 5000 µL, 10 mL, and 20 mL:
Tube of grease:
6100-555
Filters
for 5000 µL and 10 mL:
6190-164 (pack of 100)
6190-165 (pack of 1000)
for 20 mL:
6190-221 (pack of 100)
6190-222 (pack of 500)
Pipet-Lite 2 and 10 µL
Pipet-Lite 100 µL
Pipet-Lite 200 µL
A
B
C
D
E
F
G
H
J
Pipet-Lite 20 µL
A
B
C
D
E
F
G
H
J
A
B
C
D
E
F
G
H
J
A
B
C
D
E
F
G
H
J
Legend for Pipet-Lite 2 µL to 2000 µL Models
(this page and top of next page)
A - Plunger Button
B- Shaft Coupling
C - Shaft
D - Tip Ejector
E - Piston Assembly
F - Seal
G - O-ring
H - Stroke Spring
J - Seal Retainer
Pipet-Lite 1000 µL
Pipet-Lite 2000 µL
A
C
D
E
F
G
A
B
C
D
E
F
G
Pipet-Lite 5000 µL & 10 mL
Pipet-Lite 20 mL
A
C
H
H
D
E
F
G
A
C
D
E
F
G
Legend for Pipet-Lite 5000 µL, 10 mL and 20 mL (below)
A - Plunger Button
C- Tip Ejector Arm
D - Piston Assembly
E - Piston O-ring
F- Cylinder
G - Cylinder O-ring
H - Shaft
9920-237 Rev E
Limited Warranty
See the enclosed Limited Warranty and Limitations of Liability
Statement. Please complete and return the Warranty Registra-
tion Card on receipt of your pipette.
RAININ pipettes are calibrated with RAININ tips. To assure
excellent reproducibility and performance, use only RAININ tips
as recommended in this manual. Specified performance is
guaranteed only when RAININ tips are used.
Contacting RAININ
Technical Information:
Phone:
800-543-4030
Fax:
781-938-1152
E-mail:
[email protected]
Pipette Service:
Phone:
800-662-7027
Fax:
781-935-7631
E-mail:
[email protected]
Direct Order Line:
Phone:
800-472-4646
Fax:
781-938-1152
E-mail:
[email protected]
Web:
www.rainin.com
From outside the U.S.:
Massachusetts Office:001-781-935-3050
California Office:
001-510-564-1600
RAININ Japan:
Phone:
++ 81 3 5689-8311
Fax:
++ 81 3 5689-2670
METTLER TOLEDO GmbH, Switzerland
Phone:
++41 1 944 45 45
Fax:
++41 1 944 45 10
Other International Offices and Distributors:
See www.rainin-global.com
E-mail:
[email protected] (from outside the U.S.)
Web:
www.rainin-global.com (from outside the U.S.)
Rainin Instrument, LLC
7500 Edgewater Drive, Box 2160, Oakland, CA 94621-0060
a METTLER TOLEDO Company
Prices and specifications are subject to change without notice.
Copyright 2002, Rainin Instrument, LLC.