Pyxis SP-360 Operation Manual

Text of the manual

1 General Description

1.1 Specification Fluorescein Measurement Range 0 to 600 ppb Excitation Wavelength 470 nm LED Emission Wavelength 520 nm Wavelength Accuracy ±1 nm Resolution 0.1 ppb Accuracy ±1% for > 50 ppb, 0.1 ppb for < 10 ppb Calibration Solution Point 0, 50, 250 and 500 ppb Others Battery 9V alkaline battery Typical Battery Life 3200 reading (480mAh battery) Display 320x240 TFT-LCD, visible under direct sunlight Dimension L160 W74 H33 (mm) Weight 310g (without battery) Temperature Range 40 to 106 ℉ (4 to 41 ℃) Humidity 85% at 106 ℉ (41 ℃) Environmental IP67, dustproof and waterproof ⚫ ⚫ • • •

Insert DI Solution

Click Zero Button Insert 50 ppb Solution √ Click Slope Button Press cycle to choose solution MEASUREMENT Cycle Continue Exit FLUORESCEIN CALIBRATION Insert DI Solution Click Zero Button Insert 50 ppb Solution √ Click Slope Button√ Calibration Succeed Press Continue to Proceed

Insert 50 Solution

Click Slope Button Insert 250 ppb Solution √ Click Slope Button

Insert 50 Solution

Click Slope Button Insert 250 ppb Solution √ Click Slope Button Calibration Succeed Press Continue to Proceed √ MEASUREMENT Cycle Slope Exit FLUORESCEIN CALIBRATION Insert 250 Solution Click Slope Button Insert 500 ppb Solution √ Click Slope Button

Insert 250 Solution

Click Slope Button Insert 500 ppb Solution √ Click Slope Button Calibration Succeed √

Serial Number 170001

Hardware Ver 3.0 Software Ver 1.1.1 Software Rev 500 Battery Satus 54% WiFi SSID SP360-170001 FLRC Calib 500 MEASUREMENT Factory Reset Exit DEVICE INFORMATION [1] 4 [2] 470 [5] 130 [3] 340 [6] 99 [4] 190 [7] 56