Warning!
Use the AED Plus unit only as described in this manual. Improper use of the
device can cause death or injury.
DO NOT use or place the AED Plus unit in service until you have read the AED
Plus Operator’s and Administrator’s Guides.
DO NOT use or place the AED Plus unit in service if the unit’s status indicator
window (located on the left side of the handle) displays a red “X”.
DO NOT use or place the AED Plus unit in service if the unit emits a beeping tone.
Connect the electrode cable to the AED Plus unit after installing batteries.
Keep the electrode cable connected to the AED Plus unit at all times.
This device should only be used by properly trained individuals.
Only use electrodes labeled “Infant/Child” on children less than 8 years old or
weighing less than 55 lbs (25 kg). Use CPR-D padz if patient is older than 8 years
or weighs more than 55 lbs (25 kg).
Always stand clear of patient when delivering treatment. Defibrillation energy
delivered to the patient may be conducted through the patient’s body and cause
a lethal shock to those touching the patient.
DO NOT touch the electrode surfaces, the patient, or any conductive material
touching the patient during ECG analysis or defibrillation.
Move patient away from electrically conductive surfaces prior to use of equipment.
DO NOT use the unit near or within puddles of water.
Keep the patient as motionless as possible during ECG analysis.
DO NOT use the unit near flammable agents, such as gasoline, oxygen-rich
atmospheres, or flammable anesthetics.
Avoid radio frequency interference from high-power sources that might cause the
defibrillator to interpret cardiac rhythms incorrectly by turning off cell phones and
2-way radios.
Disconnect non-defibrillation protected electronic devices or equipment from
patient before defibrillation.
Dry victim’s chest, if wet, before attaching electrodes.
Apply freshly opened and undamaged electrodes, within the electrode expiration
date, to clean and dry skin to minimize burning.
DO NOT place electrodes directly over the patient’s implanted pacemaker.
Pacemaker stimuli may degrade the accuracy of ECG rhythm analyses or the
pacemaker may be damaged by defibrillator discharges.
Check labeling inside the ZOLL AED Plus cover before using the cover as a Passive
Airway Support System (PASS) device to ensure it is intended for this use.
DO NOT use the Passive Airway Support System (PASS) if there is a suspected
head or neck injury. Place the patient on a firm surface before performing CPR.
DO NOT recharge, disassemble, or dispose of batteries in fire. Batteries may
explode if mistreated.
DO NOT use or stack the AED PLUS unit with other equipment. If the unit is used
or stacked with other equipment, verify proper operation prior to use.
Caution!
DO NOT disassemble the unit. A shock hazard exists. Refer all servicing to
qualified personnel.
Use only commercially available type 123A lithium manganese dioxide batteries.
Discard batteries properly after removal from unit. Use only batteries from
recommended manufacturers. See the AED Plus Administrator’s Guide (P/N
9650-0301-01) for a list of recommended battery manufacturers.
If the device is stored outside the recommended environmental conditions, the
electrode pads and/or batteries may be damaged or their useful life reduced.
The CPR-D Padz Electrode can be connected to other ZOLL defibrillators with
Multifunction Cables. Defibrillation can be administered when connected to
other ZOLL defibrillators.The CPR function does not operate with any device
other than the AED Plus defibrillator.
Important!
This symbol indicates that an AED Plus unit is equipped for treating adult and
pediatric patients. An AED Plus unit without this symbol is not equipped to treat
pediatric patients and will NOT work with pedi•padz II ™ pediatric electrodes. To
upgrade an AED Plus unit for use with ZOLL pedi•padz II pediatric electrodes,
contact ZOLL Medical Corporation or an authorized ZOLL distributor for
information on the ZOLL AED Plus Pediatric Upgrade Kit.
Set-up and Check-out Procedure:
1.
Insert 10 new batteries into AED Plus unit.
2.
Connect electrode cable to AED Plus unit and pack sealed electrodes
inside unit cover. Close cover.
3.
Turn unit on and wait for “Unit OK” audio message. Verify that unit
issues appropriate “Adult Pads” or “Pediatric Pads” audio message.
4.
Turn unit off.
5.
Wait 2 minutes. Verify that green check symbol (
) appears in status
indicator window (located on left side of handle) and that unit does not
emit a beeping tone.
6.
Place AED Plus unit in service.
7.
Check AED Plus unit periodically to ensure that green check symbol (
)
appears in status indicator window.
Battery Replacement
Replace batteries before expiration date or if unit prompts. Use only type
123A lithium manganese dioxide batteries from recommended manufacturers.
Remove all batteries from battery compartment and discard before
installing any new batteries.
Insert 10 new batteries into battery well. Do not use old batteries.
Press button in battery well only after installation of new batteries.
Cleaning
Clean and disinfect unit with soft, damp cloth using 90% isopropyl
alcohol or soap and water, or chlorine bleach (30 ml/liter water).
Do not immerse any part of the unit in water.
Do not use ketones (MEK, acetone, etc.).
Avoid using abrasives (e.g., paper towels) on the LCD display, if so
equipped.
Do not sterilize the unit.
Automated External Defibrillator
Operator’s Guide
TROUBLESHOOTING
Problem
Recommended Action
Self-test failed.
Manually test by pressing and
holding the ON/OFF button for
more than 5 seconds. If unit fails
test again, remove from service.
“Change batteries”
Replace all batteries at the
prompt.
same time.
Red “X” in Status
Perform manual test.
Indicator window
Check to see if cable is
OR
attached properly to unit.
beeping noise
Replace batteries.
when unit is OFF.
If unit still does not operate
correctly, remove from service.
Red “X” in Status Indicator
Power cycle the unit. If Red “X”
window when unit is ON.
is still present in Status Indicator
window, remove unit from
service.
© ZOLL Medical Corp. Inc. 2006
REF 9650-0300-01 Rev. H
Federal (U.S.A.) law restricts this device for sale to or on the order of a physician.
Conformité Européene - Complies with the Medical Device Directive 93/42/EEC
AHA/ERC 2005
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA 01824-4105
978-421-9655 • 800-348-9011
www.zoll.com
After one shock...
Person collapses—
suspected
cardiac arrest.
Unit OK.
Turn unit on.
Self-test; unit OK.
Stay calm.
Call for help.
Check breathing.
Give two breaths.
Attach defib pads to patient’s
bare chest.
No shock advised.
Don't touch patient.
Analyzing.
start
here
Indicator light flashes.
Indicator light flashes.
Call or send someone
for help.
Indicator light flashes.
Check responsiveness.
Shout, "Are you OK?”
as you shake the person.
Indicator light flashes.
Prompts you may see in the LCD screen:
Text messages
Shock counter
Elapsed time
ECG waveform
What you see.
What you hear.
What you do.
Start CPR.
Administer CPR:
2 breaths for every
30 compressions.
Indicator light flashes.
Open airway.
Lift chin and tilt head back.
Option: Place Passive Airway
Support System (PASS) under
shoulders to tilt head. DO NOT use
PASS if suspected head or neck injury.
If unresponsive, Open Airway.
Indicator light flashes.
Other prompts you
may hear:
Plug in cable.
Unit failed.
Release shock button.
Data stored in memory.
Change batteries.
(n) shocks delivered.
What you should do:
Attach electrode cable.
Turn unit off and then on.
If unit still fails, perform CPR.
Let go of shock button.
Transmit data.
Replace all 10 batteries.
After 2 minutes of CPR...
Shock advised.
Don't touch patient.
Press flashing shock button.
Stop CPR.
Indicator light flashes.
Give STAND CLEAR
Warning.
DO NOT touch patient.
Press button.
Shock delivered.
Indicator light flashes.
LCD screen
Attach adult or infant/child
electrode pads to patient's
bare chest.
Place electrode pads as
shown on patients under
8 years of age or < 55 lbs (25 kg).
Follow electrode pads
package instructions.
Unit equipped to
treat adult and
pediatric patients.
Check Breathing.
Look for chest rising and falling.
Listen for air flow during exhalation.
Feel for air flow during exhalation.
Look for normal breathing, coughing, or movement.
Give 2 Breaths (AHA Guidelines 2005 only)
If not breathing, administer 2 rescue breaths,
1 second per breath.
.
CONTRAINDICATIONS FOR USE
Do NOT use when a patient:
Is conscious; or
Is breathing; or
Has a detectable pulse or other signs of circulation.
INDICATIONS FOR USE
Use when a patient has an apparent lack of
circulation indicated by:
Unconsciousness; and
Absence of normal breathing; and
Absence of pulse or signs of circulation.
When a victim is less than 8 years of age or weighs less than 55 lbs
(25 kg), the ZOLL AED Plus should be used with ZOLL AED Plus
Pediatric Electrodes. Therapy should not be delayed to determine the
patient’s exact age or weight.
INTENDED USER
The ZOLL AED Plus external defibrillator is intended to be used by
personnel who are qualified by training in the use of the AED Plus device
and basic life support, advanced life support, or other physician-
authorized emergency medical response to defibrillate victims of cardiac
arrest. The CPR monitoring function provides a metronome designed to
encourage rescuers to perform chest compressions at the AHA/ERC
recommended rate of 100 compressions per minute. Voice and visual
prompts encourage a compression depth of 1½ - 2 inches for adult
patients. The CPR monitoring function is not intended for use on patients
under 8 years of age.