Replacing all Electronics

Complete Electronics Kit (#103558) Includes: • Mk2 Circuit Board Assembly (with push-release fittings) – Wiring, Sensitivity Switch and Knob, and Test Button • Red LED Indicator Light with Terminals • Sonalert • Battery Tray, 4 x Screws, 2 x Nuts • 9V Alkaline battery Note: These instructions explain how to replace each electronic component either all at once or separately as required. Tools and Materials Needed 1. Required Repair Kit 2. Phillips Screwdriver 3. Small Flat Screwdriver (3 mm (1/10") 4. 10 mm (3/8") Wrench 5. Pliers 6. Wire Cutters 7. Wire Strippers 8. Solder Wire with Flux 9. Soldering Iron 1. Remove the battery from the meter. Undo the three screws holding the faceplate to the reel. Remove the faceplate. 2. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals of the push-release fittings on the circuit board and pull out to remove the tape leads. 3. From the front of the faceplate, use the wrench to unscrew the nut holding the test button and push the button back out of the faceplate. 4. Cut the red and black wires connected to the light, and push the light out through the front of the faceplate. 5. Use the Phillips screwdriver to undo the two screws from the battery tray. If this is an Mk1 Water Level Meter, cut the red and black wires from the battery tray. If this is an Mk2 Water Level Meter, simply pull the quick- connect fittings off of the terminals on the battery tray. Remove the battery tray from the faceplate. 6. Use the small flat screwdriver to unscrew the small brass screw on the side of the sensitivity knob. Remove the sensitivity knob. Use the pliers to unscrew the nut holding the sensitivity switch and remove the switch from the faceplate. 7. Unscrew the Sonalert retaining ring from the front of the faceplate and remove the old Sonalert and circuit board assembly from the faceplate. Back of Mk1 Water Level Meter Faceplate Back of Mk2 Water Level Meter Faceplate (with push-release fittings) Complete Electronics Kit (#103558) Sonalert Sensitivity Switch Light Battery Tray Test Button Sensitivity Switch Sonalert Light Battery Tray Test Button Molex Connector Push-Release Fittings High Quality Groundwater and Surface Water Monitoring Instrumentation (Page 1 of 6) Water Level Meter Complete Electronics Replacement Model 101/102

8. Insert the new battery tray through front of the faceplate.

Use the two screws (bolts and nuts where required) to secure the tray to the faceplate. 15. Insert the test button through the opening in the faceplate, and secure it in place using the nut. 16. Insert the sensitivity switch through the faceplate with the three connection prongs positioned towards the battery drawer, slightly towards the centre of the reel. Tighten the nut. Fasten the sensitivity knob on the front of the switch using the brass screw. 17. Connect the black wire with quick-connect fitting to the negative terminal on the battery tray. Connect the red wire with quick-connect fitting to the positive terminal. 18. Connect the tape to the new circuit board assembly: a) If this is a Mk2 Water Level Meter, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The lead on the bottom of the tape (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. b) If this is an Mk1 Water Level Meter, with a Molex connector on the tape leads, remove the Molex connector by pushing out the two pins. Cut each pin in half (see image below for correct location). Use pliers to flatten the remainder of the pins against the tape lead, so it fits easily into the terminals on the circuit board. Connect each tape lead to the new circuit board assembly, as described above in step a. Note: To secure the battery tray, the #4 x 1/2” Phillips Screws are for plastic reels, while the #4-40 x 1/2” Phillips Bolts and Nuts are for metal reels. 9. Put the new Sonalert through the faceplate, and secure it using the retaining ring. Use the photo of the fully assembled faceplate below as a reference for its position. The negative terminal should be closest to the battery tray. Sonalert Sensitivity Switch Light Battery Tray Test Button 10. Thread the leads of the LED light assembly through the front of the faceplate and snap the light in place. 11. Unscrew the two screws from the Sonalert. 12. Use the two screws to connect the black wire (-ve terminal) and red wire (+ve terminal) from the light to the outside of the corresponding terminals on the Sonalert. Do not tighten the screws all the way. 13. Slide the circuit board assembly in position on top of the Sonalert, ensuring the sides marked +ve and -ve are lined up with the corresponding terminals on the Sonalert. 14. To connect the circuit board, wrap the two bare wires from the circuit board around the screws, between the light and Sonalert terminals. Start by bringing the wires underneath the screws (see photo below). Tighten the two screws. Remove the Pins from the Tape Molex Connector (Mk1) Molex Connector Wrap bare wires from circuit board around screws High Quality Groundwater and Surface Water Monitoring Instrumentation Water Level Meter Complete Electronics Replacement (Page 2 of 6)

Replacing the Circuit Board (#102977)

Mk2 Circuit Board Assembly (with push-release fittings) includes: • Wiring, Sensitivity Switch and Test Button Release the White Terminals and the Leads Should be Secured. Cut the Pins on Each Tape Lead in Half (as shown) Press Down on the White Terminals on the Circuit Board and Insert the Tape Leads 19. Install the new battery. 20. With the probe in a glass of tap water, turn the Water Level Meter to the ‘ON’ position. If the connections are correct the buzzer and light will activate. If the buzzer or light do not activate, check all connections and the polarity of the battery. 21. Reattach the faceplate to the reel using the three screws. Sonalert Sensitivity Switch Light Battery Tray Test Button 1. Remove the battery. Unscrew the three screws and remove the faceplate. 2. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals of the push-release fittings on the circuit board and pull out to remove the tape leads. 3. If this is an Mk1 Water Level Meter, cut the red and black wires from the battery tray. If this is an Mk2 Water Level Meter, simply pull the quick-connect fittings off of the terminals on the battery tray. 4. From the front of the faceplate, use the wrench to unscrew the nut holding the test button and push the button back out of the faceplate. 5. Use the small flat screwdriver to unscrew the small brass screw on the side of the sensitivity knob. Remove the sensitivity knob. Use the pliers to unscrew the nut holding the sensitivity switch and remove the switch from the faceplate. Save the knob and screw to be used with the new sensitivity switch. 6. Watch for proper connection/polarity on the Sonalert and remove the two screws to disconnect the old circuit board assembly (and the wires connected to the light) from the Sonalert. 7. If this is an Mk1 Water Level Meter, use pliers to bend the terminals on the Sonalert, so they are straight up from the Sonalert (to accommodate the new circuit board in between the terminals). Loosen the Sonalert retaining nut on the front of the faceplate, so you can reposition the Sonalert as shown in the photo below (negative terminal next to the battery tray). 8. Follow Steps 12- 14 on Page 2 to connect the new circuit board, and reconnect the light. Mk2 Circuit Board Showing Wire Connections Square High Quality Groundwater and Surface Water Monitoring Instrumentation Water Level Meter Complete Electronics Replacement (Page 3 of 6)

Replacing the Test Button (#109080)

1. Take out the battery, and remove the faceplate by unfastening the three screws. 2. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals of the push-release fittings on the circuit board and pull out to remove the tape leads. 3. Cut off the wires at the connections to the test button and strip the wires. 4. Unscrew the old test button and remove from the faceplate. 5. Install the new test button in the faceplate and secure with the nut. 6. Attach the wires to the new test button by soldering (for the Mk1 Water Level Metes, ensure the red wire is soldered to the positive terminal on the button, and the black wire to the negative terminal.) 7. Reconnect the tape to the circuit board. For Mk1 Water Level Meters, you will use the Molex connector. If this is an Mk2 Meter, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The lead on the bottom of the tape (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. 8. Replace the faceplate by fastening the three screws. Replace the battery. Test Button Connections – Mk2 Test Button Connections – Mk1 Test Button Test Button 9. Insert the test button through the opening in the faceplate, and secure it in place using the nut. 10. Insert the sensitivity switch through the faceplate with the three connection prongs positioned towards the battery drawer, slightly towards the centre of the reel. Tighten the nut. Fasten the sensitivity knob on the front of the switch using the brass screw. 11. Connect the black wire with quick-connect fitting to the negative terminal on the battery tray. Connect the red wire with quick-connect fitting to the positive terminal. Sonalert Sensitivity Switch Light Battery Tray Test Button 12. Connect the tape to the new circuit board assembly, as described in Step 18, on Page 2. 13. Install the battery. 14. With the probe in a glass of tap water, turn the Water Level Meter to the ‘ON’ position. If the connections are correct the buzzer and light will activate. If the buzzer or light do not activate, check all connections and the polarity of the battery. 15. Reattach the faceplate to the reel using the three screws. Mk2 Circuit Board Showing Wire Connections High Quality Groundwater and Surface Water Monitoring Instrumentation Water Level Meter Complete Electronics Replacement (Page 4 of 6)

Replacing the Light (#106660)

1. Remove the battery. Unscrew and remove the faceplate. 2. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals on the circuit board and pull out to remove the tape leads. 3. Unscrew the two screws holding the wires from the light, to/through the circuit board and the Sonalert (watch for proper polarity). 4. Push the old light out and put the new light through the faceplate. 5. Reconnect the two wires, using screws to the circuit board and the Sonalert. 6. Reconnect the tape to the circuit board. For Mk1 Water Level Meters, you will use the Molex connector. If this is an Mk2 Meter, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The lead on the bottom of the tape (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. 7. Replace the faceplate, with three screws. Replace the battery. Light Connections –Mk2 Light Connections – Mk1 Replacing the Sonalert (#109075) 1. Remove the battery. Unscrew and remove the faceplate. 2. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals of the push-release fittings on the circuit board and pull out to remove the tape leads. 3. Unscrew the Sonalert retaining ring from the front of the faceplate. Push out the old Sonalert. 4. Unscrew the two screws holding the Sonalert to the circuit board (watch for proper connection/polarity). Remove the Sonalert from the circuit board. 5. For Mk1 Water Level Meters, use pliers to bend the connection terminals on the new Sonalert inwards by 90º. Use the old Sonalert you are replacing as a reference. 6. Insert the new Sonalert through the faceplate, and fasten the Sonalert retainer ring until finger tight. Ensure the Sonalert is positioned as shown below. Light/Circuit Board Connections –Mk2 Light/Circuit Board Connections – Mk1 7. Use the two screws to reconnect the wires from the light, and the circuit board to the Sonalert. Make sure the wires from the light and circuit board are connected to the correct terminals. 8. Reconnect the tape to the circuit board. For Mk1 Water Level Meters, you will use the Molex connector. If this is an Mk2 Meter, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The lead on the bottom of the tape (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. 9. Replace the faceplate, with three screws. Replace the battery. Sonalert Sonalert High Quality Groundwater and Surface Water Monitoring Instrumentation Water Level Meter Complete Electronics Replacement (Page 5 of 6)

Note: To gain better access to the sensitivity switch on Mk1 Water

Level Meters, you may want to unscrew the two screws holding the circuit board to the Sonalert, in order to move the circuit board out of the way. Replacing the On/Off Sensitivity Switch Sensitivity Switch and Knob Assembly (#102976), or Sensitivity Switch (#110179) 1. Remove the battery. 2. Unscrew the small brass screw on the side of the sensitivity knob. Remove the sensitivity knob. 3. Unscrew and remove the faceplate. 4. If this is an Mk1 Water Level Meter, disconnect the Molex connector that connects the circuit board to the tape. If this is an Mk2 Water Level Meter, press down on the white terminals of the push-release fittings on the circuit board and pull out to remove the tape leads. 5. Unsolder the wires at the connections to the sensitivity switch; note the order in which they are connected. For the Mk2 Water Level Meters, one of the black wires is soldered to two prongs on the sensitivity switch - ensure to unsolder the wire from both (try to keep the full length of stripped wire). 6. Unscrew the nut holding the sensitivity switch. Remove the switch from the faceplate. 7. Install the new sensitivity switch through the faceplate. For Mk2 Water Level Meters, position the three connection prongs towards the battery drawer, slightly towards the centre of the reel. For Mk1 Water Level Meters, position the three connection prongs opposite the battery tray and slightly towards the centre of the reel. Secure using the nut. Sensitivity Switch Connections –Mk2 Sensitivity Switch Connections – Mk1 8. Strip the wires if required. Attach the wires to the correct positions by soldering: For Mk1 Water Level Meters, the two red wires on the bottom of the circuit board are connected to the first two connections on the sensitivity switch. The black wire on the bottom of the circuit board is connected to the third prong on the sensitivity switch. The black wire closest to the top of the circuit board is attached to the prong on the top of the sensitivity switch. The other black wire connected to the top of the sensitivity switch is connected to the negative terminal on the battery tray (use the diagram on the left for reference). For Mk2 Water Level Meters, the black wire on the bottom left corner of the circuit board is connected to the centre prong on the sensitivity switch. The black wire with the long stripped section goes through the prong on the right of the sensitivity switch and connects to the prong on the top of the switch. The red wire connects to the upper terminal on the top of the switch. 9. Reconnect the tape to the circuit board. For Mk1 Water Level Meters, you will use the Molex connector. If this is an Mk2 Meter, press down on the white terminals on the circuit board and insert the tape leads. Release the terminals and the leads should be secured. The lead on the bottom of the tape (numbers facing up) is inserted into the terminal with a white square below it on the circuit board. 10. Replace the faceplate, with three screws. 11. Fasten the sensitivity knob using the brass screw. 12. Replace the battery. Printed in Canada December 16, 2016 (#108429) (Page 6 of 6) For further information contact: Solinst Canada Ltd. Fax: +1 (905) 873-1992; (800) 516-9081 Tel: +1 (905) 873-2255; (800) 661-2023 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 Web Site: www.solinst.com E-mail: [email protected] ®Solinst is a registered trademark of Solinst Canada Ltd.