RS PRO RS-5500 Instruction Manual

Specifications

Brand: RS PRO
Category: Test Equipment
Model: RS-5500 (Instruction Manual)
File Type: PDF
File Size: —
Pages: 13

Read Online

Read the RS PRO RS-5500 Instruction Manual online using the built-in PDF viewer.

Text of the manual

750V 1000V Insulation Tester 400 125V 250V 500V 1000V TEST LOBAT HV (1) POL1 (2) (1) OHM BZ (2) (2) POL2 4 (1) BAR 1 8 5 DIGITAL INSULATION TESTER RS-5500/5505 INSTRUCTION MANUAL

Read the follow ing safety information carefully before

attempting to operate or service the meter.  To avoid damages to the instrument do not apply the signals which exceed the maximum limits shown in the technical specifications tables.  Do not use the meter or test leads if they look damaged. Use extreme caution when working around bare conductors or bus bars.  Accidental contact with the conductor could result in electric shock.  Use the meter only as specified in this manual; otherw ise, the protection provided by the meter may be impaired.  Read the operating instructions before use and follow all safety Information.  Caution w hen working w ith voltages above 60V DC or 30V AC RMS. Such voltages pose a shock hazard.  Before taking resistance measurements or testing acoustic continuity, disconnect circuit from main pow er supply and all loads from the circuit.

Caution refer to this manual before using the meter.

Dangerous voltages. Meter is protected throughout by double insulation or reinforced insulation. When servicing, use only specified replacement parts. CE Comply w ith EN-61010-1 1. SPECIFICATIONS 1-1 General Information Environment conditions: ① Installation CategoriesⅢ ② Pollution Degree 2 ③ Altitude up to 2000 meters ④ Indoor use only ⑤ Relatively humidity 80% max. ⑥ Operation Ambient 0~40ºC Maintenance & Clearing: ① Repairs or servicing not covered in this manual should only be performed by qualified personnel. ② Periodically wipe the case w ith a dry cloth. Do not use abrasives or solvents on this instruments. Display: Large LCD with dual display Measurement Range: 4000MΩ/125V, 4000MΩ/250V, 4000MΩ/500V, 4000MΩ/1000V, 400Ω/BZ,

Sampling Rate: 2.5 times per second.

Zero Adjustment: Automatic adjustment. Over Range Indicator: “OL” of highest digit is displayed. Low Battery Indication: The is displayed w hen the battery Voltage drop below the operating voltage. Operating Temperature: 0ºC to 40ºC (32ºF to 104ºF) and Humidity below 80% RH Storage Temperature: -10ºC to 60ºC (14ºF to 140ºF) and Humidity below 70% RH Power source: DC9V (6x1.5V Size “AA” battery or Equivalent) Dimensions: 200(L) x 92(W) x 50(H) mm Weight: Approx 700g include battery Accessories: Test leads, 6pcs battery, Carrying case, manual. 1-2 Electrical Specifications Accuracies are specified in the w ay: ±(…% of reading +…digits) at 23ºC±5ºC,below 80% RH. OHMS Range Resolution Accuracy Max. open Circuit Voltage Overload Protection 40.00Ω 0.01Ω +(1.2%+3) 5.8V 250Vrms 400.0Ω 0.1Ω 5.8V Continuity Beeper Range Resolution Operation Resistance Max.open Circuit Voltage Overload Protection

Short circuit current

≧200mA DC Voltage Range Resolution Accuracy Input Impedance Overload Protection 1000V 1V +(0.8%+3) 10MΩ 1000Vrms AC Voltage (40Hz~400Hz) Range Resolution Accuracy Input Impedance Overload Protection 750V 1V +(1.2%+10) 10MΩ 750Vrms Meg OHMS Terminal Voltage Range Resoluti on Accuracy Test Current Short circuit current 125V(0 %~+10 %) 0.125~4.000 MΩ 0.001MΩ +(2%+10) 1mA @load 125kΩ ≤1mA 4.001~40.00 MΩ 0.01MΩ +(2%+10) 40.01~400.0 MΩ 0.1MΩ +(4%+5) 400.1~4000 MΩ 1MΩ +(5%+5) 250V (0%~+1 0%) 0.250~4.000 MΩ 0.001MΩ +(2%+10) 1mA @load 250kΩ ≤1mA 4.001~40.00 MΩ 0.01MΩ +(2%+10) 40.01~400.0 MΩ 0.1MΩ +(3%+5) 400.1~4000 MΩ 1MΩ +(4%+5) 500V(0 %~+10 %) 0.500~4.000 MΩ 0.001MΩ +(2%+10) 1mA @load 500kΩ ≤1mA 4.001~40.00 MΩ 0.01MΩ +(2%+10) 40.01~400.0 MΩ 0.1MΩ +(2%+5)

400.1~4000 MΩ 1MΩ

+(4%+5) 1000V (0%~+1 0%) 1.000~4.000 MΩ 0.001MΩ +(3%+10) 1mA @load 1MΩ ≤1mA 4.001~40.00 MΩ 0.01MΩ +(2%+10) 40.01~400.0 MΩ 0.1MΩ +(2%+5) 400.1~4000 MΩ 1MΩ +(4%+5) 2. PARTS & CONTROLS 1.Digital Display 2.Data Hold Button; MAX/MIN 3.Lock Button 4.Backlight Button; ZERO 5.Test Button 6.Rotary Function sw itch 7.VΩ Jack 8.COM input jack 9. Pothook 10.Battery Cover 2-1 How to connect test leads. On MΩ Range , and 400Ω/BZ ,ACV,.DCV, Connect the red test lead into the “VΩ” terminal and the black lead into the “COM” terminal. 2-2 Battery Check-UP & Replacement a) As battery power is not sufficient. LCD w ill display . Replacement of 6 pcs new batteries, type 1.5V size “AA” is Required. b). Place back the battery cover and four the screws. 2-3 Test leads check Set the range select sw itch to the 400Ω range. With the tip and 750V 1000V Insulation Tester 400 125V 250V 500V 1000V TEST LOBAT HV (1) POL1 (2) (1) OHM BZ (2) (2) POL2 4 (1) BAR 1 8 5 1 2 3 4 6 7 8 9 10 5

alligator clip of the test leads connected. The indicator should read

00.0Ω. When the leads are not connected the display w ill read infinity indicated by “OL”. This w ill ensure that test lead are under w orking condition. 2-4 Rotary Switch positions Turn the Tester on by selecting any measurement Lift < 1000v,500v,250v,125v (4000MΩ) OFF 400Ω/BZ, 1000VDC,750VAC > Right 2-5 Buttons and a display Indicators a). Button HOLD/MAX.MIN : Instant-pressing the “HOLD” button the 1st time, the current values will be hold in the primary display., but it will return in the 2nd pressing ; pressing 2 seconds ,it will enter directly into the “MAX”status, and one another instant-pressing will switch to the “MIN”, if instant-pressing once more, it w ill recycle ,but exit if pressing 2 seconds again. LOCK :In the insulation resistance testing function ,press the “LOCK” button ,and then push down the “ TEST” key, it w ill occur the high-voltage and enter the insulation resistance testing status. Press the “TEST” button once more ,it w ill shutoff the high-voltage and exit from the insulation resistance testing status. TEST : In the insulation resistance testing function ,pressing and holding the “TEST” button, The meter w ill bring high-voltage, and enter into the insulation resistance testing, being free from the “TEST” , it w ill cutoff the high-voltage and exit form the insulation resistance testing. ZERO/LIGHT :Iinstant-pressing the “ZERO/LIGHT” button in the 1st time, the current values in the primary display w ill be set zero,(mainly used for 400Ω, the low resistance testing), ,it will return

if in the 2nd time. pressing for 2 seconds, it will enter directly into the

“LIGHT” status, and the LCD backlight light up. After 15 seconds, the backlight is shut off automatically, the same as pressing for 2 seconds within 15s. b) Display Indicators The Primary Display:Indicate the current function testing values The Secondary Display: It shows the output DCV while you test the insulation resistance, and the battery voltage w hile the ACV The Analog Bar:indicate the current function testing value in synchronous with the primary display. :While testing the insulation resistance, the symbol “ “ flashes frequently if the voltage is over 30V. •))) : While testing the insulation resistance, the symbol” •))) “ flashes frequently and the buzzer warns continually if the outside voltage is over 30V. the symbol “ •))) “ is indicated w hile LOΩ≤35Ω and the BZ w arns continuously. LOCK:Push down the “LOCK” button while you test the insulation resistance and the symbol “ ” is indicated. LOBAT:The display shows “LOBAT” when the voltage drops below 7.5V MAX/MIN:Stand for the maximum or the minimum. ZERO : Digital zero adjusting。 HOLD :The digital holding function for the primary display. AC,DC,:The indicator for the voltage property. V,MΩ,Ω The measured dimension units。

3. INSULATION RESISTANCE MEASUREMENTS

a)Turn the function switch from the “OFF” position to the left(4000 MΩ/1000V---4000/ MΩ500V---4000 MΩ/250V---1000 MΩ/125V),and chose one of the voltage-block( there are 4 ranges namely, 4 MΩ40 MΩ400 MΩ4000 MΩ, can be sw itched automatically for every voltage-block.) b)Connect tw o testing lines to the tested; c)Push down and hold the “TEST” button /or press the “ LOCK” keystoke first and then the “TEST” button, if the tested is electriferous and its voltage ( AC/DC) is over 30V, it w ill refuse work and no high-voltage testing occurs, simultaneity, it shows “>30V” on the LCD, the symbol “ ” flashes, and the buzzer warns frequently. if the tested is diselectriferous or its voltage is low er than 30V, it will enter into the formal testing process and brings the high-voltage. on the primary display, the insulation resistance in MΩ is indicated in-phase w ith analog bar; on the secondary display, the tested insulation voltage in V (DC) is indicated, the symbol “ ” flashes and the buzzer warns frequently d)Being free from the “TEST” button or pushing down the “TEST” button in the “LOCK “ status can exit from the “LOCK” status and shutoff the high-voltage,synchronously, the resistance values is indicated in the primary display w ill be held, and the secondary display still be in the status of monitoring the insulation voltage for the tested. e)Subsequently, discharge the balance insulation voltage of the tested through the inner sw itch of the meter. Turning the function sw itch can exit automatically from testing status during the process。

4.LOW RESISTANCE (CONTINUITY)

MEASUREMENTS a). Set the range sw itch to 400Ω/BZ Position b). Connect the red test lead to the V Ω terminal and black to the COM terminal. c). Connect the tips of the test leads to both ends of the circuit under test. read resistance in Ω on the LCD. The two ranges(40.00/400.0Ω) can be switched automatically; the primary display of the resistance in Ω, flashes in synchronous w ith the analog bar. d). When the impedance on circuit is below approximately ≤35Ω.It w ill indicate by a continuous beeper. e) The current is from 200 to 220mA while the tested resistance is 0Ω f) The high voltage symbol “ “ flashes along w ith a primary display of “>30V” and the buzzer warns frequently if the voltage (AC/DC) is more than 30V. 5.AC/DC VOLTAGE MEASUREMENTS a). Set the range sw itch to ACV or DCV position b). Connect red test lead to “V Ω” terminal and black test lead to terminal “COM”. c). Connect test prods of test leads IN PARALLEL to the circuit being measured. d). Read the voltage value on LCD.

6.BATTERY SAVER (SLEEP MODE)

The meter w ill automatically enter the “sleep mode” if there is no function change or button press for 10minutes, but it w orks as soon as you turn the rotary function switch or push down any button. 7.POWER TOOLS AND AMALL APPLIANCES This test w ould also apply to other similar equipment that has a line cord. For double insulated power tools, the megohmmeter lead shown connected to the housing would be connected to some metal part lf the tool(e..g chuck,blade). Note: The sw itch of the device must be in the “ON” position and the main pow er should be disconnected. MOTORS AC-Disconnect the motor from the line by disconnecting the w ires at the motor terminals or by opening the main sw itch. If the main switch is used and the motor also has a starter then the starter must be held, by some means, in the “ON” position. In the latter case, the measured resistance will include the resistance of the motor, wire and all other components between the motor and the main sw itch. If a w eakness is in dicated, the motor and other components should be checked individually. If the motor is disconnected at the motor terminals, connect one megohmmeter lead to the grounded motor housing and the other lead to One of the motor leads. DC-Disconnect the motor from the line. To test the brush rigging, field coils and armature connect one megohmmeter lead to the grounded motor housing and the other lead to the brush on the commutator. If the resistance measurement indicates a w eakness, raise the brushes off the commutator and separately test the

armature, field coils and brush rigging by connecting one

megohmmeter lead to each of them individually, leaving the other connected to the grounded motor housing. The above also applies to DC Gemerators.

Disconnect the cable from the line. Also disconnect opposite end to

avoid errors due to leakage from other equipment. Check each conductor to ground and /or lead sheath by connecting one megohmmeter lead to a ground and /or lead sheather and the other megohmmeter lead to each of the conductors in turn. Check insulation resistance between conductors by connecting megohmmeter leads to conductors in pairs.