Futek IHH500 Quick Start Manual
Text of the manual
Quick Start Guide
Sensor Solutions Source
Load · Torque · Pressure · Multi Component · Calibration · Instruments · Software
www.futek.com
IHH500
Intelligent Handheld Display
Drawing Number: SP1189
FUTEK reserves the right to modify its design and specifications without notice.
Please visit http://www.futek.com/salesterms for complete terms and conditions.
10 Thomas, Irvine, CA 92618 USA
Tel: (949) 465-0900
Fax: (949) 465-0905
www.futek.com
TECHNICAL SUPPORT
For more IHH500 support, please visit: http://www.futek.com/ihh500/support.aspx
Display Features . . . . . . . . . . . . . . . . . . . . . . 4
Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connector Pin Configuration . . . . . . . . . . . . . . . 6
Wiring Layout . . . . . . . . . . . . . . . . . . . . . . . 7
Profile Quick Setup (Calibration) . . . . . . . . . . . . 10
TEDS and Channel Select . . . . . . . . . . . . . . . . 12
Profile Lockout and LCD Settings . . . . . . . . . . . . 13
Analog Output Quick Setup . . . . . . . . . . . . . . . 14
Alarm and Relay Quick Setup . . . . . . . . . . . . . . 15
Data Logging . . . . . . . . . . . . . . . . . . . . . . 17
SENSIT Test and Measurement Software . . . . . . . . 19
Performance Review . . . . . . . . . . . . . . . . . . . 20
Specifications . . . . . . . . . . . . . . . . . . . . . . 21
Setting Sampling rate and number of digits . . . . . . 22
Troubleshooting . . . . . . . . . . . . . . . . . . . . . 23
Table of Contents
(Compatible with USB 2.0 Only)
Output Connector
Sensor Connector
16×4 Character LCD
Active Channel Number
Peak Value
Tracking
Valley Value
Sampling Rate
Battery Indicator
Larger Font
Display Options
Peak/Valley Reset
Shunt Calibration
Verification
Power Key
1
9
2
10
3
11
4
12
5
6
13
7
14
8
15
IHH500
Battery
Enter
Exit
Shunt
Power
Display
Back
Tare
Gross
Reset
Track
Hold
Unit
Menu
1
5
11
12
13
14
15
2
3
4
+2.00001
Peak
+2.00000
mV/V
+1.99999
Valley
01 Gross
5 SPS
8
9
10
7
6
Displays Battery Life
IN MENU MODE: Enter
Toggles between engineering units (force,
torque, pressure and displacement) and the
standard mV/V.
IN MENU MODE: ▶ Arrow
Toggles between displaying the tare and gross
values. Used to remove any fixture weights.
IN MENU MODE: ▲ Arrow
Alters the current reading by shunting a
resistor across the – Excitation and – Signal
connections.
IN MENU MODE: Exit
Resets peak/valley or angle (in rotary sensors
with an encoder)
IN MENU MODE: ▲ Arrow
ON/OFF
IN MENU MODE: ON/OFF
Start navigation
IN MENU MODE: n/a
Toggles between two font sizes. Selects one
of four states in normal mode, and one of five
states (RPM, torque, angle, power) when a
rotary sensor with encoder is monitored.
IN MENU MODE: Back
Freezes the current reading. HOLD will appear
on the LCD when active. Pressing any key on
the keypad will cancel the hold command.
IN MENU MODE: ▼ Arrow
Connector Pin Configuration
6
7
1
3
4
5
2
8
1
4
2
3
1
2
G
M
F
E
D
C
L
B
A
K
J
H
Power
page 7
USB
page 7
Output
page 7
Sensor
page 8
Note: These pins are not polarity sensitive.
OUTPUT CONNECTIONS
PIN
SYMBOL
DESCRIPTION
1
IDAC
analog current output signal
2
REFERENCE
analog voltage output return
3
+R1
solid state relay 1 (positive)
4
VDAC
analog voltage output signal
5
+R2
solid state relay 2 (positive)
6
GND
analog current output return
7
–R1
solid state relay 1 (negative)
8
–R2
solid state relay 2 (negative)
USB PORT CONNECTIONS
(COMPATIBLE WITH USB 2.0 ONLY)
PIN
SYMBOL
1
VBUS
2
–D
3
+D
4
GND
IHH500 Wiring Connection Examples: http://www.
futek.com/files/Pdf/Manuals_and_Technical_
Documents/IHH500Wiring.pdf
–Excitation, TEDS return
D
–S
–Signal
E
TEDS_IO
TEDS Data
F
24_OUT
24V output
PIN
SYMBOL
DESCRIPTION
G
GND_OUT
Ground/Shield
H
5_OUT
5V Output
J
–V
–V and –mA Amplified Input
Connections
K
+V
+V and +mA Amplified Input
Connections
L
PLEAD
Leading pulse from sensor
M
PLAG
Lagging pulse from sensor
12-PIN SENSOR TO IHH500 BINDER ASSEMBLY
Note: Cable should be connected to pin G, ground, and
should be grounded on one end, either the sensor side or
the IHH500 connection side.
Wiring Layout (continued)
The +/- Excitation and +/- Signal
connections are used with mV/V output
sensors with the +/- Excitation used to
supply power to the sensor.
The 24V output supply can be used to
power amplified sensors such as our
amplified Rotary Torque sensors, for
example TRS605.
The 5V output is typically used to power
items such as the encoder in our rotary
toque sensors.
The IHH500 Elite is callable of
interacting with encoders, such as those
in our Rotary Toque sensor, with the
encoder outputs on the Leading and
Lagging pulse connections.
Typically our rotary torque sensor
encoders will have Angle 1 on the
Leading Pulse connection and Angle 2
on the Lagging pulse connections.
If the Leading Pulse Pin L leads Pin M the
Encoder count will increase and if the
Lagging Pin M leads Pin L the encoder
count will decrease.
Find FUTEK’s wiring codes at
http://www.futek.com/wirecode.aspx
CONNECTION TIPS
The IHH500 can store up to 15 profiles with channel 2 through 15 being user selectable.
1.
Menu
2.
Sensor Profile
3.
New Channel
4.
Set channel using left and right
arrow keys followed by Enter
5.
Sensor Config and choose sensor
output type
6.
Direction
7.
Unit Selection
8.
Sensor Capacity
9.
Sensitivity (+)
10. Sensitivity (–) if applicable
11. Serial Number, (Serial number on
sensor)
12. Back to go back one menu
13. Save Changes
14. Back to exit Main Menu and put the
IHH500 into normal reading display
Profile Quick Setup (Calibration)
The decimal point can be moved by selecting it first and
turning it into a number, and then selecting another
digit to change it into a decimal point.
Sensitivity (+) must be entered. If only one direction is
selected, then sensitivity (+) will be applied to both directions
of output or negative direction output.
Sensitivity is the sensor’s output level, such as mV/V output,
and is found on the calibration certificate.
Online sensor calibration data may be available by sensor
serial number at http://www.futek.com/calibrationData.aspx
to assist in finding your sensor’s output or sensitivity.
Pressing back without saving will prompt to exit without
saving. Enter can be used for Yes, and Back for No.
IHH500 Elite supports Encoders through Pulse profiles.
Current profiles cannot be set up manually and must be set up
using the Live Calibration feature in the profile setup.
Full bridge sensor type is for mV/V output sensors. Voltage
output and current output are for amplified inputs to the
IHH500.
TIPS FOR SETTING UP A PROFILE
A full profile setup guide can be found on our
FUTEK website at: http://www.futek.com/files/
Pdf/QuickStartGuide/IHH500ProfileSetup.pdf
TEDS and Channel Select
TEDS MENU
1.
Menu
2.
TEDS Data
3.
TEDS Device — Will show infor-
mation about TEDS chip.
4.
TEDS Pages — Shows pro-
grammed information stored
in chip.
5.
Load Data — Will load TEDS
information into IHH500.
6.
Auto Detection — Sets the
IHH500 to automatically load
TEDS information upon power
on.
CHANNEL SELECT
(CALIBRATION PROFILE
SELECT)
1.
Menu
2.
System Setting
3.
Channel Select
4.
Use the left and right arrow key
to choose the desired channel.
5.
Enter to load Channel or Profile.
TEDS is a small chip that can store a sen-
sor’s calibration information. The IHH500
can read it upon power up, or by loading
it through the menu, and setup a profile
with the stored information.
TEDS calibration information can be
viewed upon loading TEDS, either during
a power on or while selecting Load Data
under the TEDS menu.
The TEDS chip should be placed within
5ft. of the IHH500 sensor connection
to maintain proper power. Typically the
TEDS chip is placed in the sensor binder
connector.
An auto-calibration may be performed
when loading TEDS information, or a
new channel or profile.
TIPS FOR USING TEDS AND
CHANNEL SELECT
LOCK SETTING MENU
1.
Menu
2.
Lock Setting
a. Lockout Profile
b. Unlock Profile
c. Change Password
LCD SETTINGS
1.
Menu
2.
LCD Setting
a. Contrast
b. Brightness
c. Auto LCD Off
Profile Lockout and LCD Settings
To lockout the ability to create, edit, or delete a calibration
profile in the IHH500, use the Lock Settings menu.
The default lockout password will be the serial number on
the back of the IHH500. The password can be changed to a
new numerical combination.
The LCD contrast and brightness can be set to maximum
and the levels can be saved as default for all future usage
of the IHH500.
Use the left and right arrow keys to increase or decrease
the brightness and contrast of the LCD display in the LCD
display setup menu.
A message stating, “Selected feature is not available at this
time,” will be displayed when profile control is
locked out.
Profile will be locked out when a system calibration at
FUTEK is performed.
TIPS FOR PROFILE LOCKOUT & LCD SETTINGS
IHH500
Battery
Enter
Exit
Shunt
Power
Display
Back
Tare
Gross
Reset
Track
Hold
Unit
Menu
SELECTED FEATURE
IS NOT AVAILABLE
AT THIS TIME
(HIT ENTER/BACK)
Output Type Setup:
1.
Menu
2.
Output Config.
3.
Voltage Config.
4.
Select Bi-polar or Uni-polar
Turn on voltage output:
5.
Back to exit to main menu
6.
Interfaces
7.
Voltage Output
8.
Select Enable
9.
Back to main menu
10. Back again to exit
CURRENT OUTPUT
Output Type Setup:
1.
Menu
2.
Output Config.
3.
Current Value
4.
Select current output range
5.
Current Config.
6.
Select Uni-direction or Bi-direction
Turn on current output:
7.
Back to exit to main menu
8.
Interfaces
9.
Current Output
10. Select Enable
11. Back to exit to main menu
Analog Output Quick Setup
Bi-polar and Uni-polar for voltage
configurations will determine if the
output would be ±5VDC for both
directional outputs from the sensor, or
map both directional outputs to +5VDC
with the zero at 2.5VDC.
Analog Low is set to zero load readings
and Analog High is set to the full
capacity of the sensor profile.
Uni-direction or Bi-direction for current
will determine if the current output range
is for the positive output of the sensor, or
if the zero should be shifted to allow the
selected current output range to adjust
for both directions from the sensor.
A Tare function will affect the reading
on the IHH500 as well as the Analog
Output.
TIPS FOR ANALOG OUTPUT
Alarm Limit set points:
1.
Menu
2.
Sensor Profile
3.
Edit Channel (Choose appropriate channel)
4.
Limit & THD
5.
Alarm Limit Hi and Alarm Limit Lo
(Set the appropriate percentage of the full sensor’s
output for Hi and Lo limits)
6.
Back to Edit Profile Menu
7.
Back to Profile Menu
8.
Save Changes
9.
Back to Main Menu
10. Back to Display Screen
Alarm Configuration and Alarm Activity:
11. Menu
12. System Settings
13. Alarm Config (Latched or Non-Latched)
14. Alarm Activity (Enable or Disable)
15. Back to Main Menu
16. Back to Display Screen
Alarm and Relay Quick Setup
Alarms function off GROSS readings.
TIPS FOR ALARMS AND RELAYS
Alarm Relay 1 and Alarm Relay 2
4.
Choose Normally Open or Normally Closed
5.
Back to Output Config. Menu
6.
Back to Main Menu
7.
Back to Display Screen
Activate Relays:
8.
Menu
9.
Interfaces
10. Relay 1 and Relay 2 Output (Enable or Disable)
11. Back to Main Menu
12. Back to Display Screen
Alarm Relay 1 is mapped to Alarm High
and Alarm Relay 2 is mapped to Alarm
Low.
Threshold limits are set as a percentage
of the capacity of the sensor.
Pressing Reset on the IHH500 will reset
any alarm status.
Peak will flash on the IHH500 for an
active alarm high state, and Valley will
flash for an active low alarm state.
TIPS FOR ALARMS AND RELAYS
Alarm and Relay Quick Setup (continued)
Data logging is the ability to capture readings over a specified time period for later evaluation. The Data logging information
can be uploaded into the FUTEK SENSIT software and further exported into other programs, such as Microsoft Excel.
DATA LOGGING QUICK SETUP
1.
Menu
2.
Data Logging
3.
Logging Rate
4.
Choose Full speed or Half Speed
5.
Duration
6.
Enter duration of data logging test in seconds
7.
Action
8.
Choose to Start or Stop data logging session
Data Logging
The IHH500 can store up to 21,000 data
logging points. Each point is a captured
reading.
The sampling rate used will be the
current sample per second setting in the
IHH500 when Full Speed is selected and
half the current set sampling rate when
Half Speed is selected.
The maximum data logging duration
will be the 21,000 data logging points
divided by the sampling rate used.
The data logging information is lost
when power is turned off on the IHH500.
At slower sampling it is possible to exit
the data logging menu to view a count-
down timer.
A check mark will be present next to
Action to show a data logging session is
active.
When the data logging session is
complete the IHH500 will display a
successful message and return to the
display screen.
Pressing stop before the end of a session
will result in the captured readings
overwriting the prior readings up till the
run time.
TIPS FOR USING DATA LOGGING
The IHH500 can be used with the FUTEK SENSIT software, which can be purchased separately or evaluated through a free trial
version available on the FUTEK website.
FUNCTIONS AVAILABLE IN SENSIT WITH THE
IHH500:
•
Ability to see current reading in Display table
•
Ability to upload, list, plot, and export data logging
information from the IHH500
•
Ability to view, list, and export a live graphing session
•
Ability to remotely control the IHH500 through a virtual
front panel
TO SETUP DATA LOGGING IN SENSIT FOR IHH500:
1.
Data Logging Mode Tab
2.
Settings
3.
File Path
4.
Start Test
In SENSIT the items that do not apply to the IHH500 will be
grayed out in the Data logging Settings.
Start Test will import the data logging information into
SENSIT.
SENSIT
™ Test and Measurement Software
The IHH500 should be turned on and
connected to the computer prior to
starting SENSIT.
USB output, under Interfaces in the
IHH500, should be enabled with a check
mark to confirm its operation.
TIPS FOR USING SENSIT WITH THE IHH500
Input Range 1
–500
500
mV/V nom.
Input Range 2
–12
12
VDC
Input Range 3
0
30
mA
OUTPUT
PARAMETER
MIN
MAX
UNIT
Analog Voltage
0
5
VDC
–5
5
VDC
Analog Current
0
20
mA
4
20
mA
0
25
mA
5
25
mA
POWER ADAPTER
PARAMETER
MIN
MAX
Input
90–264 VAC
0.5 A 50/60 Hz
Output
12 VDC
1A
• Power Supply Adapter: 12 VDC/1 A
• Internal battery: Rechargeable, lithium polymer battery
• Battery size: 7.4V/3000 mAh
• Run Time: up to 30 hours with 350Ω Bridge
• Battery Charger: Wall plug AC powered-unit
SOLID STATE RELAYS
• Part Number: G3VM-2F
• Application: Alarm Relay 1, 2
• Maximum Output Voltage: 250V (AC or DC)
• Maximum Output Current: 120 mA
Note: Internally limited and protected to 110V/100 mA
PASSWORDS
• Default Password: 0000000
• Alternative Password: Device’s Serial Number
RESPONSE TIMES
• Conversion capture rate: every sample
• Peak capture rate: every sample
• Valley capture rate: every sample
• Conversion display update rate: 5 times per second
• Peak display update rate: 5 times per second
• Valley display update rate: 5 times per second
BANDWIDTH
The digital output bandwidth has been specified as SPS/4.
The analog output bandwidth has been specified as SPS/4 for
sampling rates less than 1200 SPS and has been specified as
SPS/5 for sampling rates more than 1200 SPS if:
1. ASCII stream output is deactivated
2. IHH500 is not connected to computer
3. All relays are disabled
4. Display is in menu mode (Peak, Valley, Tracking, and data
display are main display mode)
SETTING SAMPLING RATE
1.
Menu
2.
System Settings
3.
Sampling Rate
CHANGING NUMBER OF DIGITS:
1.
Menu
2.
System Settings
3.
Digits Select
Sampling Rate and Selecting number of Digits
The sampling rate can be set from 5 sps
to 4800 sps at preset intervals.
Increasing the sampling rate may
increase the noise as the IHH500
becomes more sensitive to input
changes.
The decimal point is floating on the
IHH500 which means with 4 decimal
points the format will automatically
adjust from 0000 to 0.000. For example,
a 1,000lb load cell would display from
1000 to 0.001lbs.
The number of digits can be selected
from 3 to 6 and can be saved as default.
TIPS
• An “Overflow” message usually indicates an output
higher than that specified in the profile type and can be
from the load cell or from an open to the IHH500 from the
sensor. The load cell connections should be confirmed
and the load cell for any overload damage.
• If the sensor profile is locked out a message, “Selected
feature is no available at this time,” will appear. See the
lock setting section of this manual for more information.
• A live calibration can be performed using the calibration
option in the IHH500 profie setup and will require known
loads. The load to use should match the capacity listed in
the profile.
• If the readings are not correct for the applied load, review
the sensor profile and confirm the capacity and sensitivity
for your load cell.
• Profile 1 is a system mV/V reference channel and can
be used to confirm the output of a sensor and help in
troubleshooting any questions in a profile setup.
• In the main menu Diagnostic can be used with the
external setting to confirm the resistance on the output
connections of the load cell.
• In the IHH500 Elite the Display button can be pressed to
show encoder information. This screen must be active for
SENSIT Live Graphing to capture encoder information.