Key Pages of Document

When the Multi-Cell does not reset

If the resetting is failed, the photo interrupter switch inside the 5-Pos Variable Pathlength Cell Holder needs replacing. Contact PerkinElmer for replacing the photo interrupter. ...

Configuration of the 5-Pos Variable Pathlength Cell Holder

Figure 2 Lambda 365 5-Pos Variable Pathlength Cell Holder Variable Partition Variable Pathlength Cuvette Holder ① 5 mm cell ② 10 mm cell ③ 20 mm cell ④ 40 mm cell ⑤ 50 mm cell Figure 3 Lambda 365 5-Pos Variable Pathlength Cell Holder ...

4. Pull out the cell holder and base plate by hand.

Figure 7 Pulling out the cell holder and base plate 5. Insert the 5-Pos Variable Pathlength Cell Holder in the sample compartment. Figure 8 Inserting the 5-Pos Variable Pathlength Cell Holder in the sample compartment ...

6. Gently press the 5-Pos Variable Pathlength Cell Holder to connect the communication port (male)

under the bottom of the 5-Pos Variable Pathlength Cell Holder to the port (female) in the sample compartment. Figure 9 Connecting the communication ports 7. Tighten the 5-Pos Variable Pathlength Cell Holder in the sample compartment with the screws (The red circles indicates the location of the screws in the figure). Figure 10 Tighten the 5-PosVariable Pathlength Cell Holder in the sample compartment 8. Turn on the power switch of the instrument. ...

8. ① Click the Accessory tab and ② Select Variable Pathlength Multi-Cell on Accessory

Type, and then click the ③Method Setup button to open the Multicell window. 9. The Multicell window will be shown. Using the button, select cells to be used among the available cells list. Check (√) Multi Baseline, if it is required to use. a. Available Cells: Indicate cells that are available for measurement. b. Using Cells: Shows the cell positions which are selected for measurement. Remove cells using key and switch positions using and keys. c. Multicell Options Multi Zero: If check ...

Dimensions and Specifications

Dimensions Physical Characteristic Specification Comment Outline Height (mm) 139.5 Width (mm) 134.5 Depth (mm) 341 Inner Height (mm) 35 Suitable for the Standard Cells Width (mm) 58 Depth (mm) 12.6 Weight (kg) 2.5 Specification Physical Characteristic Specification Comment Space between cells (mm) 7.4 Moving distance of one cell (mm) 20 Moving distance of five cells (mm) 80 Moving time of one cell (sec) 4 Moving time of five cells (sec) 11 Description Connectable Cells Description Precision cell, 5 mm light path, 1.7 ml, ea Precision cell, 10 mm light ...

PerkinElmer Lambda 365 user guide recommended for: OMNI Bead Ruptor Elite, Lambda XLS, Panthera-Puncher 9, AVIO 200, OPERETTA.

⚠️ Download Safety

The Measuring Instruments Installation instructions manual (12 pages) is completely safe to download (last scan date: 17/12/2025).