8. Prepare the manifold with all the tubing connected in Chap. ‘Manifold and tubing connection’ (p4).

Figure 20. Manifold with all the tubing connected 9. Insert the tubes and cells through the front cover first and hold the manifold in place. Figure 21. Inserting the tubes and cells 10. Tighten the 4 bolts. Figure 22. Tighten the 4 bolts ...

Figure 7. Tubings for cell compartment inside (P/N B2000049) and outside

(P/N B2000515) Figure 8. Flow Cell (N4101052) NOTE: Tubings for cell compartment inside (B2000049) and outside (P/N B2000515) and Flow cell (P/N N4101052) must be purchased separately. Manifold and tubing connection 1. Prepare a manifold 2. Connect a red connector of the inside tube (P/N B2000049) to the In port of the manifold and other white connector of the inside tube (short connector) to the flow cell with the arrow mark (Inlet). Figure 9. Connecting the inlet tube ...

3. Connect a red connector of the inside tube (P/N B2000049) to the Out port of the manifold and

other white connector of the inside tube (long connector) to the flow cell’s outlet port. Figure 10. Connecting the outlet tube 4. Connect a red connector of the outside tube (P/N B2000515) to the port of the manifold. Figure 11. Connecting the outside tube 5. Connect all the connectors in the same way. Figure 12. Manifold and tubing connection ...

1. Prepare the Lambda 365 Spectrophotometer to install this accessory.

2. Connect the power cord and the communication cable. DO NOT turn on the power of the instrument! 3. Remove the two round head screws with washer (M4*12L) to remove the existing cell holder and base plate. Figure 13. Location of the round head screws with washer 4. Pull out the cell holder and base plate by hand. Figure 14. Pulling out the cell holder and base plate CAUTION ATTENTION Make sure the instrument is turned off while installing this accessory. Assurez-vous que l’instrument est éteint ...

Figure 7. Tubings for cell compartment inside (P/N B2000049) and outside

(P/N B2000515) Figure 8. Flow Cell (N4101052) NOTE: Tubings for cell compartment inside (B2000049) and outside (P/N B2000515) and Flow cell (P/N N4101052) must be purchased separately. Manifold and tubing connection 1. Prepare a manifold 2. Connect a red connector of the inside tube (P/N B2000049) to the In port of the manifold and other white connector of the inside tube (short connector) to the flow cell with the arrow mark (Inlet). Figure 9. Connecting the inlet tube ...

7. Tighten the Changer 8-Cell for Dissolution in the sample compartment with the screws. (The red

circles indicate the location of the screws in the figure). Figure 17. Tighten the Changer 8-Cell for Dissolution 8. Attach the plate for calibration on the front plate and tighten the 4 bolts. Figure 18. Tighten the 4 bolts 9. Close the sample compartment cover and turn on the power switch of the instruments. CAUTION ATTENTION Do not start the UV Express software before finishing the initialization of grating. Ne démarrez pas le logiciel UV Express avant de terminer l'initialisation ...

Figure 28. Inserting No.1 and reference cells into the cell holder

8. And then click New to open a new window. 9. Open Method in the main menu. 10. Measure the sample according to the procedure 13-19 of Measurement section. (page 11) Calibration of Multi-Cell Position Calibrate the beam position of the Changer 8-Cell for Dissolution when the Multicell is installed for the first time or beam position is incorrect. 1. Execute Scan Setup mode and ① click the Accessory tab of Method window. ② Select Multi- cell on Accessory Type, and then click the ③Method Setup button to open t ...

PerkinElmer Lambda 365 user guide recommended for: SPECTRUM TWO N, Wallac 1420, FL 8500, AVIO 500, DSC 4000.

⚠️ Download Safety

The Measuring Instruments Installation instructions manual (18 pages) is completely safe to download (last scan date: 11/03/2026).

Related PDF Instructions for PerkinElmer Lambda 365:

Measuring Instruments Equipment: Frequently Requested

Have some Questions about PerkinElmer Lambda 365 Device?

Get valuable and actionable advices from other PerkinElmer Lambda 365 device users. Describe your problem or question, regarding the using of Lambda 365, in as much detail as possible, and other users will be able to answer you.