Keysight Models 6811C, 6812C, and 6813C Programming Guide

Notices Copyright Notice © Keysight Technologies 2017 No part of this manual may be repro- duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol- ogies as governed by United States and international copyright laws. Manual Part Number 06811-90005 Edition Edition 1, October 2, 2017 Printed in: Printed in Malaysia Published by: Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia Technology Licenses The hardware and/or software described in this document are fur- nished under a license and may be used or copied only in accordance with the terms of such license. Declaration of Conformity Declarations of Conformity for this product and for other Keysight prod- ucts may be downloaded from the Web. Go to http://www.keysight.com/ go/conformity. You can then search by product number to find the latest Dec- laration of Conformity. U.S. Government Rights The Software is “commercial computer software,” as defined by Federal Acqui- sition Regulation (“FAR”) 2.101. Pursu- ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple- ment (“DFARS”) 227.7202, the U.S. government acquires commercial com- puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan- dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/ find/sweula. The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft- ware. The EULA and the license set forth therein, does not require or per- mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft- ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern- ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per- form, display, or disclose commercial computer software or commercial com- puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu- ant to the FAR and the DFARS and are set forth specifically in writing else- where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data. Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS- CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA- TION CONTAINED HEREIN, INCLUD- ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER- CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE- QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON- TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR- RANTY TERMS COVERING THE MATE- RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

3 Safety Information CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. La mention ATTENTION signale un danger. Si la manoeuvre ou la procédure correspondante n’est pas exécutée correctement, il peut y avoir un risque d’endommagement de l’appareil ou de perte de données importantes. En présence de la mention ATTENTION, il convient de ne pas poursuivre tant que les conditions indiquées n’ont pas été parfaitement comprises et remplies. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. La mention AVERTISSEMENT signale un danger pour la sécurité de l’opérateur. Si la manoeuvre ou la procédure correspondante n’est pas exécutée correctement, il peut y avoir un risque grave, voire mortel pour les personnes. En présence d’une mention AVERTISSEMENT, il convient de s’interrompre tant que les conditions indiquées n’ont pas été parfaitement comprises et satisfaites.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Safety Considerations Read the information below before using this instrument. The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this guide violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements. WARNING ENERGY HAZARD. AC sources can supply 425 V peak at their output. DEATH on contact may result if the output terminals or circuits connected to the output are touched when power is applied. RISQUE ÉNERGÉTIQUE. Les sources AC peuvent fournir une tension de sortie de 425 V maximum. Tout contact peut entraîner la MORT si les circuits ou bornes de sortie connectés à la sortie sont touchés lorsqu’ils sont sous tension.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

5 Waste Electrical and Electronic Equipment (WEEE) Directive This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste. Product category: With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product. The affixed product label is as shown below. Do not dispose in domestic household waste. To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information. Sales and Technical Support To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/6800 (product-specific information and support, software and documentation updates) – www.keysight.com/find/assist (worldwide contact information for repair and service)

Keysight Models 6811C, 6812C, and 6813C Programming Guide

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

7 Table of Contents Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . .5 Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 1 General Information About this Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24 Documentation summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24 External References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 SCPI references . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 GPIB references . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 Keysight IO Libraries Suite . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26 2 Introduction to Programming Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28 USB Capabilities of the AC Source . . . . . . . . . . . . . . . . . . . . . . . . . . . .29 Vendor Identification (VID) and Product Identification (PID) . . . . . .29 LAN Capabilities of the AC Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 Instrument Web server . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 GPIB Capabilities of the AC Source . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 GPIB address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 RS-232 Capabilities of the AC Source . . . . . . . . . . . . . . . . . . . . . . . . . .32 RS-232 data format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32 Baud rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32 RS-232 Programming Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 RS-232 troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34 Introduction to SCPI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35 Conventions used in this guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35 Types of SCPI commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Types of SCPI messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 The SCPI Command Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 The root level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Active header path . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 The effect of optional headers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Moving among subsystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Including common commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Using queries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 Coupled commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 Structure of a SCPI Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 The message unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 Combining message units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 Headers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Query indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Message unit separator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Root specifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45 Message terminator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 SCPI Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Numerical data formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Suffixes and multipliers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 Character data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 System Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49 Assigning the GPIB address in programs . . . . . . . . . . . . . . . . . . . . . 49 Types of DOS drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 Error handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Keysight BASIC controllers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 3 Language Dictionary Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55 Calibration Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 CALibrate:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 CALibrate:CURRent:AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59

Keysight Models 6811C, 6812C, and 6813C Programming Guide

9 CALibrate:CURRent:EXTernal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 CALibrate:CURRent:MEASure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 CALibrate:DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 CALibrate:DATA? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 CALibrate:DATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61 CALibrate:IMPedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61 CALibrate:LEVel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62 CALibrate:PASSword . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62 CALibrate:PHASe:AUXiliary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63 CALibrate:PWM:FREQuency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63 CALibrate:PWM:RAMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 CALibrate:SAVE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .64 CALibrate:SECure:CODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65 CALibrate:SECure:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66 CALibrate:SHUNt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67 CALibrate:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68 CALibrate:STRing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 CALibrate:VOLTage:AC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69 CALibrate:VOLTage:AUXiliary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 CALibrate:VOLTage:DC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 CALibrate:VOLTage:EXTernal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70 CALibrate:VOLTage:OFFSet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71 CALibrate:VOLTage:PROTection . . . . . . . . . . . . . . . . . . . . . . . . . . . .71 Display Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72 DISPlay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72 DISPlay:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73 DISPlay:TEXT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73 DISPlay:TEXT:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73 LXI Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74 LXI:IDENtify[:STATe] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74 LXI:MDNS[:STATe] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75 LXI:MDNS:ENABle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75 LXI:MDNS:HNAMe[:RESolved]? . . . . . . . . . . . . . . . . . . . . . . . . . . . .76 LXI:MDNS:SNAMe:DESired . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76

Keysight Models 6811C, 6812C, and 6813C Programming Guide

LXI:MDNS:SNAMe[:RESolved]? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 LXI:RESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 LXI:RESTart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Measurement Subsystem (Arrays) Commands . . . . . . . . . . . . . . . . . . . 78 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 MEASure:ARRay:CURRent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 MEASure:ARRay:CURRent:HARMonic? . . . . . . . . . . . . . . . . . . . . . . 80 MEASure:ARRay:CURRent:HARMonic:PHASe? . . . . . . . . . . . . . . . . 81 MEASure:ARRay:VOLTage? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 MEASure:ARRay:VOLTage:HARMonic? . . . . . . . . . . . . . . . . . . . . . . 83 MEASure:ARRay:VOLTage:HARMonic:PHASe? . . . . . . . . . . . . . . . . 84 Measurement Subsystem (Current) Commands . . . . . . . . . . . . . . . . . . 85 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 MEASure:CURRent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 MEASure:CURRent:AC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 MEASure:CURRent:ACDC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 MEASure:CURRent:AMPLitude:MAXimum? . . . . . . . . . . . . . . . . . . . 87 MEASure:CURRent:CREStfactor? . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 MEASure:CURRent:HARMonic? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 MEASure:CURRent:HARMonic:PHASe? . . . . . . . . . . . . . . . . . . . . . . 89 MEASure:CURRent:HARMonic:THD? . . . . . . . . . . . . . . . . . . . . . . . . 89 Measurement Subsystem (Frequency) Commands . . . . . . . . . . . . . . . 90 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 MEASure:FREQuency? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Measurement Subsystem (Power) Commands . . . . . . . . . . . . . . . . . . . 92 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 MEASure:POWer? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 MEASure:POWer:AC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 MEASure:POWer:AC:APParent? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93 MEASure:POWer:AC:REACtive? . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 MEASure:POWer:AC:PFACtor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Measurement Subsystem (TVOLt) Commands . . . . . . . . . . . . . . . . . . . 95 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 MEASure:TVOLt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96

Keysight Models 6811C, 6812C, and 6813C Programming Guide

11 MEASure:TVOLt:ABSolute? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97 Measurement Subsystem (Voltage) Commands . . . . . . . . . . . . . . . . . .98 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98 MEASure:VOLTage? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99 MEASure:VOLTage:AC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99 MEASure:VOLTage:ACDC? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 MEASure:VOLTage:HARMonic? . . . . . . . . . . . . . . . . . . . . . . . . . . .100 MEASure:VOLTage:HARMonic:PHASe? . . . . . . . . . . . . . . . . . . . . .101 MEASure:VOLTage:HARMonic:THD? . . . . . . . . . . . . . . . . . . . . . . .101 MEASure:VOLTage:RANKed? . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102 Output Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103 OUTPut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104 OUTPut:COUPling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104 OUTPut:DFI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105 OUTPut:DFI:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105 OUTPut:IMPedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .106 OUTPut:IMPedance:REAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .106 OUTPut:IMPedance:REACtive . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107 OUTPut:PON:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .107 OUTPut:PROTection:CLEar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108 OUTPut:PROTection:DELay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108 OUTPut:RI:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109 OUTPut:TTLTrg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109 OUTPut:TTLTrg:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110 Sense Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111 SENSe:CURRent:ACDC:RANGe . . . . . . . . . . . . . . . . . . . . . . . . . . .112 SENSe:FREQuency:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113 SENSe:NSELect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114 SENSe:SWEep:OFFSet:POINts . . . . . . . . . . . . . . . . . . . . . . . . . . . .114 SENSe:SWEep:TINTerval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115 SENSe:WINDow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116 Source Subsystem (Current) Commands . . . . . . . . . . . . . . . . . . . . . . .117 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117 all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CURRent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118 CURRent:PEAK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 CURRent:PEAK:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 CURRent:PEAK:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121 CURRent:PROTection:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 Source Subsystem (Frequency) Commands . . . . . . . . . . . . . . . . . . . . 123 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 FREQuency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 FREQuency:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 FREQuency:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 FREQuency:SLEW:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 FREQuency:SLEW:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 FREQuency:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126 Source Subsystem (Frequency Modulation) Commands . . . . . . . . . . 127 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 FM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 FM:DEViation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 FM:FREQuency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 Source Subsystem (Function) Commands . . . . . . . . . . . . . . . . . . . . . 129 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129 FUNCtion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130 FUNCtion:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 FUNCtion:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 FUNCtion:CSINusoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133 Source Subsystem (List) Commands . . . . . . . . . . . . . . . . . . . . . . . . . . 134 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134 LIST:COUNt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 135 LIST:CURRent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136 LIST:CURRent:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136 LIST:DWELl . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137 LIST:DWELl:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137 LIST:FREQuency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 LIST:FREQuency:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138 LIST:FREQuency:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139 LIST:FREQuency:SLEW:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . 139

Keysight Models 6811C, 6812C, and 6813C Programming Guide

13 LIST:PHASe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140 LIST:PHASe:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .140 LIST:SHAPe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141 LIST:SHAPe:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .141 LIST:STEP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142 LIST:TTLTrg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142 LIST:TTLTrg:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143 LIST:VOLTage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143 LIST:VOLTage:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144 LIST:VOLTage:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144 LIST:VOLTage:SLEW:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . . .144 LIST:VOLTage:OFFSet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .145 LIST:VOLTage:OFFSet:POINts? . . . . . . . . . . . . . . . . . . . . . . . . . . . .145 LIST:VOLTage:OFFSet:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .146 LIST:VOLTage:OFFSet:SLEW:POINts? . . . . . . . . . . . . . . . . . . . . . .146 Source Subsystem (Phase) Commands . . . . . . . . . . . . . . . . . . . . . . . .147 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .147 PHASe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .147 PHASe:AUXiliary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .148 PHASe:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .148 PHASe:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .149 Source Subsystem (Pulse) Commands . . . . . . . . . . . . . . . . . . . . . . . .150 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150 PULSe:COUNt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .150 PULSe:DCYCle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151 PULSe:HOLD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .151 PULSe:PERiod . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .153 PULSe:WIDTh . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .153 Source Subsystem (Voltage) Commands . . . . . . . . . . . . . . . . . . . . . . .154 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .154 VOLTage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .155 VOLTage:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .156 VOLTage:AUXiliary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .157 VOLTage:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .158 VOLTage:OFFSet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .159

Keysight Models 6811C, 6812C, and 6813C Programming Guide

VOLTage:OFFSet:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 160 VOLTage:OFFSet:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161 VOLTage:OFFSet:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 161 VOLTage:OFFSet:SLEW:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 162 VOLTage:OFFSet:SLEW:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . 162 VOLTage:PROTection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 VOLTage:PROTection:STATe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163 VOLTage:SENSe:DETector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164 VOLTage:SENSe:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165 VOLTage:SLEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166 VOLTage:SLEW:MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166 VOLTage:SLEW:TRIGgered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167 Status Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168 STATus:PRESet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169 STATus:OPERation? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 169 STATus:OPERation:CONDition? . . . . . . . . . . . . . . . . . . . . . . . . . . . 170 STATus:OPERation:ENABle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170 STATus:OPERation:NTRansition . . . . . . . . . . . . . . . . . . . . . . . . . . . 171 STATus:QUEStionable? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172 STATus:QUEStionable:CONDition? . . . . . . . . . . . . . . . . . . . . . . . . . 173 STATus:QUEStionable:ENABle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173 STATus:QUEStionable:NTRansition . . . . . . . . . . . . . . . . . . . . . . . . 174 System Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176 SECurity[:IMMediate] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177 SYSTem:COMMunicate:LAN:CONTrol? . . . . . . . . . . . . . . . . . . . . . 177 SYSTem:COMMunicate:LAN:DHCP . . . . . . . . . . . . . . . . . . . . . . . . 178 SYSTem:COMMunicate:LAN:DNS[{1|2}] . . . . . . . . . . . . . . . . . . . . . 179 SYSTem:COMMunicate:LAN:DOMain? . . . . . . . . . . . . . . . . . . . . . . 180 SYSTem:COMMunicate:LAN:GATeway . . . . . . . . . . . . . . . . . . . . . . 181 SYSTem:COMMunicate:LAN:HOSTname . . . . . . . . . . . . . . . . . . . . 182 SYSTem:COMMunicate:LAN:IPADdress . . . . . . . . . . . . . . . . . . . . . 183 SYSTem:COMMunicate:LAN:MAC? . . . . . . . . . . . . . . . . . . . . . . . . 184 SYSTem:COMMunicate:LAN:SMASk . . . . . . . . . . . . . . . . . . . . . . . 185 SYSTem:COMMunicate:LAN:TELNet:PROMpt . . . . . . . . . . . . . . . . 186

Keysight Models 6811C, 6812C, and 6813C Programming Guide

15 SYSTem:COMMunicate:LAN:TELNet:WMESsage . . . . . . . . . . . . . .186 SYSTem:COMMunicate:LAN:UPDate . . . . . . . . . . . . . . . . . . . . . . .187 SYSTem:COMMunicate:RLSTate . . . . . . . . . . . . . . . . . . . . . . . . . . .188 SYSTem:COMMunicate:TCPip:CONTrol? . . . . . . . . . . . . . . . . . . . .188 SYSTem:ERRor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .189 SYSTem:LANGuage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .189 SYSTem:PERSona:MANufacturer . . . . . . . . . . . . . . . . . . . . . . . . . .189 SYSTem:PERSona:MANufacturer:DEFault . . . . . . . . . . . . . . . . . . .190 SYSTem:PERSona:MODel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .190 SYSTem:PERSona:MODel:DEFault . . . . . . . . . . . . . . . . . . . . . . . . .190 SYSTem:SECurity:COUNt? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .191 SYSTem:SECurity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .191 SYSTem:SET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .192 SYSTem:VERSion? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .192 SYSTem:LOCal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .192 SYSTem:REMote . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193 SYSTem:RWLock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193 Trace Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .194 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .194 TRACe:CATalog? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .194 TRACe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .195 TRACe:DEFine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196 TRACe:DELete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196 Trigger Subsystem Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .197 Subsystem syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .197 ABORt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .198 INITiate:SEQuence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .199 INITiate:CONTinuous:SEQuence . . . . . . . . . . . . . . . . . . . . . . . . . . .200 TRIGger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .200 TRIGger:DELay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .201 TRIGger:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .201 TRIGger:SEQuence2:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202 TRIGger:SEQuence2:PHASe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203 TRIGger:SEQuence3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203 TRIGger:SEQuence3:SOURce . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204 TRIGger:SEQuence1:DEFine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Common Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206 Common commands syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206 *CLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207 *ESE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207 *ESR? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 208 *IDN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209 *LRN? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 209 *OPC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210 *OPT? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 211 *PSC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 212 *RCL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213 *RST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 214 *SAV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 215 *SRE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 216 *STB? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 217 *TRG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218 *TST? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218 *WAI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219 4 Compatibility of the B and C Versions of the AC Power Solutions Persona Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222 Model number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222 Manufacturer ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222 New Feature Comparison with B and C AC Source . . . . . . . . . . . . . . 223 5 Programming Examples Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228 Programming the Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229 Power-on initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229 Enabling the output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229 AC voltage and frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229 Voltage and frequency slew rates . . . . . . . . . . . . . . . . . . . . . . . . . . 230 Waveform shapes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231 Programming the output phase . . . . . . . . . . . . . . . . . . . . . . . . . . . 233 Current limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233 DC output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234 all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

17 Coupled commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235 Programming Output Transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236 Transient system model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .237 Step and pulse transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .239 List transients . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .241 Triggering Output Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244 SCPI triggering nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244 Output trigger system model . . . . . . . . . . . . . . . . . . . . . . . . . . . . .245 Initiating the output trigger system . . . . . . . . . . . . . . . . . . . . . . . . .246 Selecting the output trigger source . . . . . . . . . . . . . . . . . . . . . . . .246 Specifying a trigger delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247 Synchronizing output changes to a reference phase angle . . . . . .248 Generating output triggers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .248 Specifying a dwell time for each list point . . . . . . . . . . . . . . . . . . .249 Making Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250 Voltage and current measurements . . . . . . . . . . . . . . . . . . . . . . . .251 Power measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .251 Harmonic measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .252 Returning voltage and current data from the data buffer . . . . . . .253 Regulatory-compliant measurement of quasi-stationary harmonics . 253 Triggering Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254 SCPI triggering nomenclature . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254 Measurement trigger system model . . . . . . . . . . . . . . . . . . . . . . . .255 Initiating the measurement trigger system . . . . . . . . . . . . . . . . . . .256 Selecting the measurement trigger source . . . . . . . . . . . . . . . . . . .256 Generating measurement triggers . . . . . . . . . . . . . . . . . . . . . . . . .257 Controlling the instantaneous voltage and current data buffers . .257 Programming the Status Registers . . . . . . . . . . . . . . . . . . . . . . . . . . .259 Power-on conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .259 Operation status group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .259 Questionable Status Group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .263 Standard event status group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .264 Status byte register . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .265

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266 Programming the Trigger In and Trigger Out BNC Connectors . . . . . 269 Remote Inhibit and Discrete Fault Indicator . . . . . . . . . . . . . . . . . . . . 271 Remote inhibit (RI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 271 Discrete fault indicator (DFI) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 272 SCPI Command Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273 A SCPI Command Tree Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276 B SCPI Conformance Information SCPI Confirmed Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 296 Non SCPI Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298 C Error Messages Error Number List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302

Keysight Models 6811C, 6812C, and 6813C Programming Guide

19 List of Figures Figure 2-1 Partial command tree . . . . . . . . . . . . . . . . . . . . . . . . . .37 Figure 2-2 Command message structure . . . . . . . . . . . . . . . . . . .43 Figure 5-1 Model of transient system . . . . . . . . . . . . . . . . . . . . .238 Figure 5-2 Model of output triggers . . . . . . . . . . . . . . . . . . . . . .245 Figure 5-3 Model of measurement triggers . . . . . . . . . . . . . . . .255 Figure 5-4 Pre-event and post-event triggering . . . . . . . . . . . . .258 Figure 5-5 AC source status model . . . . . . . . . . . . . . . . . . . . . . .262 Figure 5-6 BNC connector trigger model . . . . . . . . . . . . . . . . . .270

Keysight Models 6811C, 6812C, and 6813C Programming Guide

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

21 List of Tables Table 5-1 Bit configuration of status registers . . . . . . . . . . . . . .260 Table C-1 Error numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .302 all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Keysight Models 6811C, 6812C, and 6813C AC Power Solutions

Programming Guide 23 1 General Information About this Guide 24 External References 25 Keysight IO Libraries Suite 26 This chapter provides an introduction to this guide.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

About this Guide This guide contains programming information for the Keysight Models 6811C, 6812C, and 6813C AC Power Solutions. These units will be referred to as “ac source” throughout this guide. You will find the following information in the rest of this guide: – Chapter 1, an introduction to this guide. – Chapter 2, an introduction to SCPI messages structure, syntax, and data formats. – Chapter 3, a dictionary of the available SCPI commands. – Chapter 4, information about the compatibility of the B and C versions of the ac source. – Chapter 5, an introduction to programming the ac source with SCPI commands. – Appendix A, the SCPI command tree. – Appendix B, SCPI conformance information. – Appendix C, list of error messages. Documentation summary The following documents that are related to this Programming Guide have additional helpful information for using the ac source. – Quick Start Guide. Information on how to quickly get started using the ac source. – User’s and Service Guide. Includes specifications and supplemental characteristics, how to use the front panel, how to connect to the ac source, and calibration procedures.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

25 External References SCPI references The following documents will assist you with programming in SCPI: – Beginner's Guide to SCPI. Keysight part no. H2325-90001. Highly recommended for anyone who has not had previous experience programming with SCPI. – Tutorial Description of the General Purpose Interface Bus. Keysight part no. 5952-0156. Highly recommended for those not familiar with the IEEE 488.1 and 488.2 standards. To obtain a copy of the above documents, contact your local Keysight Sales and Support Office. GPIB references The most important GPIB documents are your controller programming manuals - Keysight BASIC, GPIB Command Library for MS DOS, etc. Refer to these for all non-SCPI commands (for example: Local Lockout). The following are two formal documents concerning the GPIB interface: – ANSI/IEEE Std. 488.1-1987 IEEE Standard Digital Interface for Programmable Instrumentation. Defines the technical details of the GPIB interface. While much of the information is beyond the need of most programmers, it can serve to clarify terms used in this guide and in related documents. – ANSI/IEEE Std. 488.2-1987 IEEE Standard Codes, Formats, Protocols, and Common Commands. Recommended as a reference only if you intend to do fairly sophisticated programming. Helpful for finding precise definitions of certain types of SCPI message formats, data types, or common commands. The above two documents are available from the IEEE (Institute of Electrical and Electronics Engineers), 345 East 47th Street, New York, NY 10017, USA.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Keysight IO Libraries Suite Keysight IO Libraries Suite is a collection of free instrument control software that automatically discovers instruments and allows you to control instruments over LAN, USB, GPIB, RS-232, and other interfaces. For more information, or to download IO Libraries, go to www.keysight.com/find/iosuite. all-guides.co m

Keysight Models 6811C, 6812C, and 6813C AC Power Solutions

Programming Guide 27 2 Introduction to Programming Introduction 28 USB Capabilities of the AC Source 29 LAN Capabilities of the AC Source 30 GPIB Capabilities of the AC Source 31 RS-232 Capabilities of the AC Source 32 Introduction to SCPI 35 The SCPI Command Tree 37 Structure of a SCPI Message 42 SCPI Data Formats 47 System Considerations 49 This chapter provides an introduction to SCPI messages structure, syntax, and data formats.

Introduction to Programming

28 Keysight Models 6811C, 6812C, and 6813C Programming Guide Introduction The ac source supports the SCPI programming language on all of its remote I/O interfaces. NOTE It is not recommended to send SCPI commands simultaneously using more than one remote I/O interface.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 29 USB Capabilities of the AC Source To use the USB interface for the ac source, no settings are required on the ac source. Vendor Identification (VID) and Product Identification (PID) The vendor identification (VID) and product identification (PID) of the ac sources are specified as below: VID PID Keysight model 0x2A8D 0x3601 6811C 0x3701 6812C 0x3801 6813C

Introduction to Programming

30 Keysight Models 6811C, 6812C, and 6813C Programming Guide LAN Capabilities of the AC Source The ac source conforms to the LAN eXtensions for Instrumentation (LXI) Device Specification 2011 version 1.4. The Telnet, Socket, and VXI-11 interfaces are available to communicate with the ac source that works on either static or dynamic IP address. Instrument Web server The ac source has a built-in Web server that gives you direct control from an Internet browser on your computer. With the Web server, you can access the front panel control functions including the LAN configuration parameters. The Web server is enabled when shipped. Perform the following steps to launch the Web server: 1 Establish a LAN interface connection from your computer to the ac source. 2 To launch the Web server, enter the instrument's IP address or fully-qualified hostname into the browser’s Address field. 3 To begin controlling your instrument, click the Browser Web Control button in the navigation bar on the left. 4 For additional help, click the “Help Page” button on the navigation bar. NOTE The built-in Web server only operates over the LAN interface. It requires Internet Explorer 9 onwards.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 31 GPIB Capabilities of the AC Source All ac source functions except for setting the GPIB address are programmable over the GPIB. GPIB address The ac source operates from a GPIB address that is set from the front panel. To set the GPIB address, press the Address key on the front panel and enter the address using the Entry keys. all-guides.co m

Introduction to Programming

32 Keysight Models 6811C, 6812C, and 6813C Programming Guide RS-232 Capabilities of the AC Source The ac source provides an RS-232 programming interface, which is activated by commands located under the front panel Address key. All SCPI commands are available through RS-232 programming. The EIA RS-232 Standard defines the interconnections between Data Terminal Equipment (DTE) and Data Communications Equipment (DCE). The ac source is designed to be a DTE. It can be connected to another DTE such as a PC COM port through a null modem cable. RS-232 data format The RS-232 data is an 11-bit word with one start bit and two stop bits. The number of start and stop bits is not programmable. The following parity options are selectable using the front panel Address key: Parity options are stored in non-volatile memory. Baud rate The front panel Address key lets you select one of the following baud rates, which is stored in non-volatile memory:1200, 2400, 4800, or 9600. NOTE The RS-232 settings in your program must match the settings specified in the front panel Address menu. Press the front panel Address key if you need to change the settings. EVEN Seven data bits with even parity ODD Seven data bits with odd parity NONE Eight data bits without parity NOTE The ac source must have the same baud rate and parity as the controller.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 33 RS-232 Programming Example The following program illustrates how to program the ac source using RS-232 to set the output voltage and frequency and to read back the model number and output voltage. ' Program to write and read via RS232 ' Configure serial port (e.g. COM1) for: ' Bits per second: select 9600 baud (or desired baudrate) ' Start bits: select 1 ' Data bits: select 8 ' Stop bits: select 2 ' Parity: select None ' Flow control: select None ' Send line feed [Enter] ' Reserve 1000 character buffer for serial I/O ' ' Ensure identical baud and parity settings on ac source (Default - Baudrate: 9600, Parity: None) ' ' Send SCPI commands *RST ' Resets the ac source VOLT 60 ' Set voltage to 60 volts FREQ 50 ' Set frequency to 50 hertz OUTPUT ON ' Turn on the output *IDN? ' Query the ac source identification string [Wait for response] MEAS:VOLT:AC? ' Query the ac source voltage

Introduction to Programming

34 Keysight Models 6811C, 6812C, and 6813C Programming Guide RS-232 troubleshooting If you are having trouble communicating over the RS-232 interface, check the following: – The computer and the ac source must be configured for the same baud rate, parity, and number of data bits. Note that the ac source is configured for 1 start bit and 2 stop bits (these values are fixed). – The correct interface cables or adapters must be used, as described under “RS-232 connector” in the User's Guide. Note that even if the cable has the proper connectors for your system, the internal wiring may be incorrect. – The interface cable must be connected to the correct serial port on your computer (COM1, COM2, etc.). [Wait for response] ' END ' End of main program

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 35 Introduction to SCPI SCPI (Standard Commands for Programmable Instruments) is a programming language for controlling instrument functions over the GPIB. SCPI is layered on top of the hardware-portion of IEEE 488.2. The same SCPI commands and parameters control the same functions in different classes of instruments. For example, you would use the same DISPlay command to control the ac source display and the display of a SCPI-compatible multimeter. Conventions used in this guide Angle brackets < > Items within angle brackets are parameter abbreviations. For example, indicates a specific form of numerical data. Vertical bar | Vertical bars separate alternative parameters. For example, NORM | TEXT indicates that either “TEXT” or “NORM” can be used as a parameter. Square Brackets [ ] Items within square brackets are optional. The representation [SOURce:]LIST means that SOURce: may be omitted. Braces { } Braces indicate parameters that may be repeated zero or more times. It is used especially for showing arrays. The notation {<,B>} shows that parameter “A” must be entered, while parameter “B” may be omitted or may be entered one or more times. Computer font Computer font is used to show program lines in text. TRIGger:DELay .5 shows a program line.

Introduction to Programming

36 Keysight Models 6811C, 6812C, and 6813C Programming Guide Types of SCPI commands SCPI has two types of commands: common and subsystem. – Common commands generally are not related to specific operation but to controlling overall ac source functions, such as reset, status, and synchronization. All common commands consist of a three-letter mnemonic preceded by an asterisk: *RST, *IDN?, *SRE 8 – Subsystem commands perform specific ac source functions. They are organized into an inverted tree structure with the “root” at the top. Some are single commands while others are grouped within specific subsystems. Refer to Appendix A, "SCPI Command Tree” for the ac source SCPI tree structure. Types of SCPI messages There are two types of SCPI messages: program and response. – A program message consists of one or more properly formatted SCPI commands sent from the controller to the ac source. The message, which may be sent at any time, requests the ac source to perform some action. – A response message consists of data in a specific SCPI format sent from the ac source to the controller. The ac source sends the message only when commanded by a program message called a “query”. all-guides.co m

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 37 The SCPI Command Tree As previously explained, the basic SCPI communication method involves sending one or more properly formatted commands from the SCPI command tree to the instrument as program messages. The following figure shows a portion of a subsystem command tree, from which you access the commands located along the various paths (you can refer to the complete tree in Appendix A, "SCPI Command Tree”). Figure 2-1 Partial command tree :OUTPut [:STATe] :DFI :PROTection ROOT :CLEar :DELay :STATus :CONDition? :OPERation [:EVEN] ? [:STATe] :SOURce :COUPling

Introduction to Programming

38 Keysight Models 6811C, 6812C, and 6813C Programming Guide The root level Note the location of the ROOT node at the top of the tree. Commands at the root level are at the top level of the command tree. The SCPI interface is at this location when: – the ac source is powered on – a device clear (DCL) is sent to the ac source – the SCPI interface encounters a message terminator – the SCPI interface encounters a root specifier Active header path In order to properly traverse the command tree, you must understand the concept of the active header path. When the ac source is turned on (or under any of the other conditions listed above), the active path is at the root. That means the SCPI interface is ready to accept any command at the root level, such as OUTPut or STATe. If you enter OUTPut, the active header path moves one colon to the right. The interface is now ready to accept :STATe, :COUPling, :DFI, or :PROTection as the next header. You must include the colon because it is required between headers. If you now enter :PROTection, the active path again moves one colon to the right. The interface is now ready to accept either :CLEar or :DELay as the next header. If you now enter :CLEar, you have reached the end of the command string. The active header path remains at :CLEar. If you wished, you could have entered :CLEar;DELay 20 and it would be accepted as a compound message consisting of: OUTPut:PROTection:CLEAr and OUTPut:PROTection:DELay 20 The entire message would be: OUTPut:PROTection:CLEar;DELay 20 The message terminator after DELay 20 returns the path to the root.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 39 The effect of optional headers If a command includes optional headers, the interface assumes they are there. For example, if you enter OUTPut OFF, the interface recognizes it as OUTPut:STATe OFF. This returns the active path to the root (:OUTPut). But if you enter |OUTPut:STATe OFF,| then the active path remains at :STATe. This allows you to send OUTPut:STATe OFF;PROTection:CLEar in one message. If you tried to send OUTPut OFF;PROTection:CLEar the header path would return to :OUTPut instead of :PROTection. The optional header [SOURce] precedes the current, frequency, function, phase, pulse, list, and voltage subsystems. This effectively makes :CURRent, :FREQuency, :FUNCtion, :PHASe, :PULse, :LIST, and :VOLTage root-level commands. Moving among subsystems In order to combine commands from different subsystems, you need to be able to restore the active path to the root. You do this with the root specifier (:). For example, you could clear the output protection and check the status of the Operation Condition register as follows: OUTPut:PROTection:CLEAr STATus:OPERation:CONDition? Because the root specifier resets the command parser to the root, you can use the root specifier and do the same thing in one message: OUTPut:PROTection:CLEAr;:STATus:OPERation:CONDition? The following message shows how to combine commands from different subsystems as well as within the same subsystem: VOLTage:LEVel 70;PROTection 80;:CURRent:LEVel 3;PROTection:STATe ON Note the use of the optional header LEVel to maintain the correct path within the voltage and current subsystems and the use of the root specifier to move between subsytems.

Introduction to Programming

40 Keysight Models 6811C, 6812C, and 6813C Programming Guide Including common commands You can combine common commands with system commands in the same message. Treat the common command as a message unit by separating it with a semicolon (the message unit separator). Common commands do not affect the active header path; you may insert them anywhere in the message. VOLTage:TRIGger 7.5;INITialize;*TRG OUTPut OFF;*RCL 2;OUTPut ON Using queries Observe the following precautions with queries: – Set up the proper number of variables for the returned data. – Read back all the results of a query before sending another command to the ac source. Otherwise a Query Interrupted error will occur and the unreturned data will be lost. NOTE The “Enhanced Tree Walking Implementation” given in Appendix A of the IEEE 488.2 Standard is not implemented in the ac source.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 41 Coupled commands When commands are coupled it means that the value sent by one command is affected by the settings of the other commands. The following commands are coupled in the ac source: – the voltage, voltage offset, and function shape commands – the step, pulse, and list commands that control output voltages, voltage offsets, and function shapes – the pulse commands that program the width, duty cycle, period, and the hold parameter – the voltage range and current limit commands in some ac source models As explained later in Chapter 5, “Programming Examples”, the order in which data is sent by these coupled commands can be important when more than one parameter is changed. all-guides.co m

Introduction to Programming

42 Keysight Models 6811C, 6812C, and 6813C Programming Guide Structure of a SCPI Message SCPI messages consist of one or more message units ending in a message terminator. The terminator is not part of the syntax, but implicit in the way your programming language indicates the end of a line (such as a newline or end-of-line character). The message unit The simplest SCPI command is a single message unit consisting of a command header (or keyword) followed by a message terminator. ABORt VOLTage? The message unit may include a parameter after the header. The parameter usually is numeric, but it can be a string: VOLTage 20 VOLTage MAX

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 43 Combining message units The following command message is briefly described here, with details in subsequent paragraphs. Figure 2-2 Command message structure The basic parts of the above message are as follows: Message component Example Headers VOLT LEV PROT CURR Header separator The colon in VOLT:LEV Data 8088 Data separator The space in VOLT 80 and PROT 88 Message units VOLT:LEV 80 PROT 88 CURR? Data Headers Header Separator Message Unit Separators Message Unit Query Indicator Message Terminator Root Specifier VOLT:LEV 80 PROT 88 ; ; : CURR?

Introduction to Programming

44 Keysight Models 6811C, 6812C, and 6813C Programming Guide Headers Headers are instructions recognized by the ac source. Headers (which are sometimes known as “keywords”) may be either in the long form or the short form. The SCPI interface is not sensitive to case. It will recognize any case mixture, such as TRIGGER, Trigger, TRIGger. Message unit separator The semicolons in VOLT:LEV 80; and PROT 88; Root specifier The colon in PROT 88;:CURR? Query indicator The question mark in CURR? Message terminator The (newline) indicator. Terminators are not part of the SCPI syntax. Long form The header is completely spelled out, such as VOLTAGE, STATUS, and DELAY. Short form The header has only the first three or four letters, such as VOLT, STAT, and DEL. NOTE Short form headers result in faster program execution. Header convention In the command descriptions in Chapter 3, “Language Dictionary” of this guide, headers are emphasized with boldface type. The proper short form is shown in upper-case letters, such as DELay. Header separator If a command has more than one header, you must separate them with a colon (VOLT:PROT OUTPut:RELay:POLarity). Optional headers The use of some headers is optional. Optional headers are shown in brackets, such as OUTPut[:STATe] ON. As previously explained under “The effect of optional headers” on page 39, if you combine two or more message units into a compound message, you may need to enter the optional header.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 45 Query indicator Following a header with a question mark turns it into a query (VOLTage?, VOLTage:PROTection?). If a query contains a parameter, place the query indicator at the end of the last header (VOLTage:PROTection? MAX). Message unit separator When two or more message units are combined into a compound message, separate the units with a semicolon (STATus:OPERation?;QUEStionable?). Root specifier When it precedes the first header of a message unit, the colon becomes the root specifier. It tells the command parser that this is the root or the top node of the command tree. Note the difference between root specifiers and header separators in the following examples: OUTPut:PROTection:DELay .1 All colons are header separators :OUTPut:PROTection:DELay .1 Only the first colon is a root specifier OUTPut:PROTection:DELay .1;:VOLTage 12.5 Only the third colon is a root specifier NOTE You do not have to precede root-level commands with a colon; there is an implied colon in front of every root-level command.

Introduction to Programming

46 Keysight Models 6811C, 6812C, and 6813C Programming Guide Message terminator A terminator informs SCPI that it has reached the end of a message. Three permitted messages terminators are as follows: – newline (), which is ASCII decimal 10 or hex 0A. – end or identify () – both of the above (). In the examples of this guide, there is an assumed message terminator at the end of each message. If the terminator needs to be shown, it is indicated as regardless of the actual terminator character. all-guides.co m

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 47 SCPI Data Formats All data programmed to or returned from the ac source is ASCII. The data may be numerical or character string. Numerical data formats Symbol Data form Talking formats Digits with an implied decimal point assumed at the right of the least-significant digit. Examples: 273 Digits with an explicit decimal point. Example: .0273 Digits with an explicit decimal point and an exponent. Example: 2.73E+2 Boolean data. Example: 0 | 1 or OFF | ON (0 = OFF; 1 = ON) Listening formats Extended format that includes , and . Examples: 273273. 2.73E2 Expanded decimal format that includes and MINMAX. Examples: 273 73.2 .73E2 MAX. MIN and MAX are the minimum and maximum limit values that are implicit in the range specification for the parameter. Boolean data. Example: 0 | 1

Introduction to Programming

48 Keysight Models 6811C, 6812C, and 6813C Programming Guide Suffixes and multipliers Character data Character strings returned by query statements may take either of the following forms, depending on the length of the returned string: Class Suffix Unit Unit with multiplier Current A ampere mA (milliampere) Amplitude V volt mV (millivolt) Time s second ms (millisecond) Frequency Hz hertz kHz (kilohertz) Common multipliers 1E3 K kilo 1E–3 m milli 1E–6 u micro Character Response Data. Permits the return of character strings. Arbitrary ASCII Response Data. Permits the return of undelimited 7-bit ASCII. This data type has an implied message terminator. String Response Data. Returns string parameters enclosed in double quotes.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 49 System Considerations The remainder of this chapter addresses some system issues concerning programming. These are ac source addressing and the use of the following types of GPIB system interfaces: – HP Vectra PC controller with Keysight 82335A GPIB Interface Command Library – IBM PC controller with National Instruments GPIB-PCII Interface/Handler – Keysight controller with Keysight BASIC Language System Assigning the GPIB address in programs The ac source address cannot be set remotely. It must be set from the front panel. Once the address is set, you can assign it inside programs. The following examples assume that the GPIB select code is 7, and the ac source will be assigned to the variable ACS. For systems using the National Instruments DOS driver, the address is specified in the software configuration program (IBCONFIG.EXE) and assigned a symbolic name. The address then is referenced only by this name within the application program (refer to the National Instruments GPIB documentation). 1070 ACS=706 ! Keysight 82335A Interface 1070 ASSIGN @ACS TO 706 ! Keysight BASIC Interface

Introduction to Programming

50 Keysight Models 6811C, 6812C, and 6813C Programming Guide Types of DOS drivers The Keysight 82335A and National Instruments GPIB are two popular DOS drivers. Each is briefly described here. Refer to the software documentation supplied with the driver for more details. Keysight 82335A driver For GW-BASIC programming, the GPIB library is implemented as a series of subroutine calls. To access these subroutines, your application program must include the header file SETUP.BAS, which is part of the DOS driver software. SETUP.BAS starts at program line 5 and can run up to line 999. Your application programs must begin at line 1000. SETUP.BAS has built-in error checking routines that provide a method to check for GPIB errors during program execution. You can use the error-trapping code in these routines or write your own code using the same variables as used by SETUP.BAS. National Instruments GPIB driver Your program must include the National Instruments header file DECL.BAS. This contains the initialization code for the interface. Prior to running any applications programs, you must set up the interface with the configuration program (IBCONF.EXE). Your application program will not include the ac source symbolic name and GPIB address. These must be specified during configuration (when you run IBCONF.EXE). Note that the primary address range is from 0 to 30 but any secondary address must be specified in the address range of 96 to 126. The instrument expects a message termination on EOI or line feed, so set end of identify (EOI) with last byte of Write. It is also recommended that you set Disable Auto Serial Polling. All function calls return the status word IBSTA%, which contains a bit (ERR) that is set if the call results in an error. When ERR is set, an appropriate code is placed in variable IBERR%. Be sure to check IBSTA% after every function call. If it is not equal to zero, branch to an error handler that reads IBERR% to extract the specific error.

Introduction to Programming

2 Keysight Models 6811C, 6812C, and 6813C Programming Guide 51 Error handling If there is no error-handling code in your program, undetected errors can cause unpredictable results. This includes “hanging up” the controller and forcing you to reset the system. Both of the above DOS drivers have routines for detecting program execution errors. Keysight BASIC controllers The Keysight BASIC Programming Language provides access to GPIB functions at the operating system level. This makes it unnecessary to have the header files required in front of DOS applications programs. Also, you do not have to be concerned about controller “hangups” as long as your program includes a timeout statement. Because the ac source can be programmed to generate Service Request (SRQ) on errors, your program can use an SRQ service routine for decoding detected errors. The detectable errors are listed in Appendix C, "Error Messages”. NOTE Use error detection after every call to a subroutine. all-guides.co m

Introduction to Programming

52 Keysight Models 6811C, 6812C, and 6813C Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Keysight Models 6811C, 6812C, and 6813C AC Power Solutions

Programming Guide 53 3 Language Dictionary Introduction 54 Subsystem Commands 55 Calibration Subsystem Commands 57 Display Subsystem Commands 72 LXI Subsystem Commands 74 Measurement Subsystem (Arrays) Commands 78 Measurement Subsystem (Current) Commands 85 Measurement Subsystem (Frequency) Commands 90 Measurement Subsystem (Power) Commands 92 Measurement Subsystem (TVOLt) Commands 95 Measurement Subsystem (Voltage) Commands 98 Output Subsystem Commands 103 Sense Subsystem Commands 111 Source Subsystem (Current) Commands 117 Source Subsystem (Frequency) Commands 123 Source Subsystem (Frequency Modulation) Commands 127 Source Subsystem (Function) Commands 129 Source Subsystem (List) Commands 134 Source Subsystem (Phase) Commands 147 Source Subsystem (Pulse) Commands 150 Source Subsystem (Voltage) Commands 154 Status Subsystem Commands 168 System Subsystem Commands 176 Trace Subsystem Commands 194 Trigger Subsystem Commands 197 Common Commands 206 This chapter provides a dictionary of the available SCPI commands.

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Introduction This section gives the syntax and parameters for all the IEEE 488.2 SCPI commands and the Common commands used by the ac sources when operating in Normal mode. It is assumed that you are familiar with the material in Chapter 2, "Introduction to Programming". Because the SCPI syntax remains the same for all programming languages, the examples given for each command are generic. Syntax forms Syntax definitions use the long form, but only short form headers (or “keywords”) appear in the examples. Use the long form to help make your program self-documenting. Parameters Most commands require a parameter and all queries will return a parameter.The range for a parameter may vary according to the model of ac source. Parameters for all models are listed in the Specifications table in the User's Guide. Models If a command only applies to specific models, those models are listed in the only entry. If there is no only entry, the command applies to all models. Related commands Where appropriate, related commands or queries are included. These are listed because they are either directly related by function, or because reading about them will clarify or enhance your understanding of the original command or query. Order of presentation The dictionary is organized as follows: – Subsystem commands, arranged by subsystem – IEEE 488.2 common commands

Keysight Models 6811C, 6812C, and 6813C Programming Guide

55 Subsystem Commands Subsystem commands are specific to ac source functions. They can be a single command or a group of commands. The groups comprise commands that extend one or more levels below the root. The description of common commands follows the description of the subsystem commands. The subsystem command groups are listed in alphabetical order and the commands within each subsystem are grouped alphabetically under the subsystem. Commands followed by a question mark (?) take only the query form. When commands take both the command and query form, this is noted in the syntax descriptions. You will find the subsystem command groups discussed on the following pages: Subsystem Page Calibration Subsystem Commands page 57 Display Subsystem Commands page 72 LXI Subsystem Commands page 74 Measurement Subsystem (Arrays) Commands page 78 Measurement Subsystem (Current) Commands page 85 Measurement Subsystem (Frequency) Commands page 90 Measurement Subsystem (Power) Commands page 92 Measurement Subsystem (TVOLt) Commands page 95 Measurement Subsystem (Voltage) Commands page 98 Output Subsystem Commands page 103 Sense Subsystem Commands page 111 Source Subsystem (Current) Commands page 117 Source Subsystem (Frequency) Commands page 123 Source Subsystem (Frequency Modulation) Commands page 127 Source Subsystem (Function) Commands page 129 Source Subsystem (List) Commands page 134 Source Subsystem (Phase) Commands page 147

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Source Subsystem (Pulse) Commands page 150 Source Subsystem (Voltage) Commands page 154 Status Subsystem Commands page 168 System Subsystem Commands page 176 Trace Subsystem Commands page 194 Trigger Subsystem Commands page 197 Common Commands page 206 Subsystem Page all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

57 Calibration Subsystem Commands The commands in this subsystem allow you to do the following: – Enable and disable the calibration mode – Change the calibration password – Calibrate the current and voltage output levels, and store new calibration constants in non-volatile memory. Subsystem syntax CALibrate | CALibration :COUNt? Returns the calibration count of the ac source :CURRent :AC Begin AC current programming calibration sequence :EXTernal Initiates calibration of the external current measurement :MEASure Begin current measurement calibration sequence :DATA Input a calibration measurement :DATA? Returns the present value of the phase calibration :DATE Input a calibration date :IMPedance Begin output impedance calibration sequence :LEVel Advance to next calibration step (P1 | P2 | P3 | P4) :PASSword Set calibration password :PHASe :AUXiliary Initiates calibration of the auxiliary output phase :PWM :FREQuency Trim pulse width modulator frequency :RAMP Trim pulse width modulator ramp :SAVE Save new cal constants in non-volatile memory :SECure :CODE Set calibration password :STATe [, ] Unsecures or secures the instrument for calibration :SHUNt Allows you to enter the external current shunt value :STATE [, ] Enable or disable calibration mode :STRing Stores a message in calibration memory

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:COUNt? CALibration:COUNt? Returns the calibration count of the ac source. Read and record the initial count when you first receive your instrument. The value increments each time you store the calibration constants, as such a complete calibration adds many counts. The calibration count also increments whenever you save a calibration string, change the calibration password, or override the calibration security. You can perform this query regardless of whether the instrument is secured. This setting is non-volatile; it will not be changed by power cycling, a factory reset (*RST), or an instrument preset (SYSTem:PRESet). :VOLTage :AC Begin AC voltage calibration sequence :AUXiliary [:AC] Initiates calibration of the auxiliary output voltage :DC Begin DC voltage calibration sequence :EXTernal Initiates calibration of the external voltage measurement :OFFSet Begin offset voltage calibration sequence :PROTection Begin voltage protection calibration sequence CALibrate | CALibration Query syntax CALibrate:COUNt? CALibration:COUNt? Parameters None Examples CAL:COUN? Returned parameters Related commands SYST:SEC:COUN

Keysight Models 6811C, 6812C, and 6813C Programming Guide

59 CALibrate:CURRent:AC CALibration:CURRent:AC This command can only be used in the calibration mode. It initiates the calibration of the AC current limit and metering circuits. CALibrate:CURRent:EXTernal CALibration:CURRent:EXTernal Keysight 6811C/6812C/6813C Option 020 only. Initiates calibration of the external current measurement. CALibrate:CURRent:MEASure CALibration:CURRent:MEASure This command is used to initiate the calibration of the current metering circuits and the peak current limit circuits. It can only be used in the calibration mode. Command syntax CALibrate:CURRent:AC CALibration:CURRent:AC Parameters None Examples CAL:CURR:AC Related commands CAL:STAT CAL:SAVE CAL:LEV Command syntax CALibrate:CURRent:EXTernal CALibration:CURRent:EXTernal Parameters None Examples CAL:CURR:EXT Command syntax CALibrate:CURRent:MEASure CALibration:CURRent:MEASure Parameters None Examples CAL:CURR:MEAS Related commands CAL:STAT CAL:SAVE CAL:LEV

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:DATA CALibration:DATA This command is only used in calibration mode. It enters a calibration value that you obtain by reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value being entered. These constants are not stored in non-volatile memory until they are saved with CALibrate:SAVE. If CALibrate:STATE OFF is programmed without a CALibrate:SAVE, the previous calibration constants are restored. CALibrate:DATA? CALibration:DATA? Keysight 6812/6813C Option 026 only. Returns the present value of the phase calibration. This command can only be used during the auxiliary phase calibration procedure. It will return an error if it is used at any other time. Command syntax CALibrate:DATA CALibration:DATA Parameters Examples CAL:DATA 3222.3 MA CAL:DATA 5.000 Related commands CAL:STAT CAL:SAVE Query syntax CALibrate:DATA? CALibration:DATA? Returned parameters Examples CAL:DATA?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

61 CALibrate:DATE CALibration:DATE Stores a message in calibration memory with a string input of 11 characters. Storing a calibration message overwrites the previous message. You can perform this query regardless of whether the instrument is secured. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. CALibrate:IMPedance CALibration:IMPedance This command can only be used in calibration mode. It calibrates the output impedance circuits. The command automatically performs the calibration. CALibrate:SAVE is required to store the impedance constant in non-volatile memory. CALibrate:IMPedance is a sequential command that takes several seconds to complete. Command syntax CALibrate:DATE CALibration:DATE Parameters Examples CAL:DATE “14-Jun-2016” Query syntax CALibrate:DATE? CALibration:DATE? Returned parameters Related commands CAL:STR Command syntax CALibrate:IMPedance CALibration:IMPedance Parameters None Examples CAL:IMP Related commands CAL:STAT CAL:SAVE all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:LEVel CALibration:LEVel This command can only be used in calibration mode. It is used to advance to the next state in the calibration sequence. CALibrate:PASSword CALibration:PASSword This command can only be used in calibration mode. It allows you to change the calibration password. A new password is automatically stored in non-volatile memory and does not have to be stored with CALibrate:SAVE. If the password is set to 0, password protection is removed and the ability to enter the calibration mode is unrestricted. Command syntax CALibrate:LEVel CALibration:LEVel Parameters P1 | P2 | P3 | P4 Examples CAL:LEV P2 Related commands CAL:STAT CAL:SAVE Command syntax CALibrate:PASSword CALibration:PASSword Parameters 0 (default) Examples CAL:PASS 6812 CAL:PASS 02.1997 Related commands CAL:STAT CAL:SEC:CODE

Keysight Models 6811C, 6812C, and 6813C Programming Guide

63 CALibrate:PHASe:AUXiliary CALibration:PHASe:AUXiliary Keysight 6812C/6813C Option 026 only. Initiates calibration of the auxiliary output phase. CALibrate:PWM:FREQuency CALibration:PWM:FREQuency This command is only used during manufacture or repair. It trims the switching frequency of the power output stages. The numbers from 0 to 7 are internally mapped to 8 discrete frequencies. Command syntax CALibrate:PHASe:AUXiliary CALibration:PHASe:AUXliary Parameters 0 through 7 Examples CAL:PHAS:AUX 1 Command syntax CALibrate:PWM:FREQuency CALibration:PWM:FREQuency Parameters 0 through 7 Examples CAL:PWM:FREQ 1 Query syntax CALibrate:PWM:FREQuency? CALibration:PWM:FREQuency? Returned parameters Related commands CAL:PWM:RAMP

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:PWM:RAMP CALibration:PWM:RAMP This command modulates the slope of voltage ramp driving the power output stages. Varying the ramp affects the harmonic distortion of the output. The argument is a number from 0 to 255. This command is only used during manufacture or repair of the ac source. CALibrate:SAVE CALibration:SAVE This command can only be used in calibration mode. It saves any new calibration constants (after a current or voltage calibration procedure has been completed) in non-volatile memory. Command syntax CALCulate:PWM:RAMP CALibration:PWM:RAMP Parameters 0 through 255 Examples CAL:PWM:RAMP 100 Query syntax CALibrate:PWM:RAMP? CALibration:PWM:RAMP? Returned parameters Related commands CAL:PWM:FREQ Command syntax CALibrate:SAVE CALibration:SAVE Parameters None Examples CAL:SAVE Related commands CAL:CURR CAL:VOLT CAL:STAT

Keysight Models 6811C, 6812C, and 6813C Programming Guide

65 CALibrate:SECure:CODE CALibration:SECure:CODE This command can only be used in calibration mode. It allows you to change the calibration password. A new password is automatically stored in non-volatile memory and does not have to be stored with CALibrate:SAVE. If the password is set to 0, password protection is removed and the ability to enter the calibration mode is unrestricted. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. Command syntax CALibrate:SECure:CODE CALibration:SECure:CODE Parameters 0 (default) Examples CAL:SEC:CODE 6813 | CAL:SEC:CODE 02.1997 Related commands CAL:PASS

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:SECure:STATe CALibration:SECure:STATe This command enables and disables calibration mode. The calibration mode must be enabled before the ac source will accept any other calibration commands. The first parameter specifies the enabled or disabled state. The second parameter is the password. It is required if the calibration mode is being enabled and the existing password is not 0. If the password is not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query statement returns only the state, not the password. Whenever the calibration state is changed from enabled to disabled, any new calibration constants are lost unless they have been stored with CALibrate:SAVE. Unsecures or secures the ac source for calibration. Command syntax CALibrate:SECure:STATe [, ] CALibration:SECure:STATe [, ] Parameters 0 | 1 | OFF | ON [, ] *RST value OFF Examples CAL:SEC:STAT 1, 6812 | CAL:SEC:STAT OFF Query syntax CALibrate:SECure:STATe? CALibration:SECure:STATe? Returned parameters 0 | 1 Related commands CAL:STAT all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

67 CALibrate:SHUNt CALibration:SHUNt Keysight 6811C/6812C/6813C Option 020 only. Allows you to enter the external current shunt value (in ohms). The external current shunt is used for external current measurement. The programming range is 1E-6 to 1E6. This parameter is non-volatile, so there is no *RST value. As with other CAL commands, calibration must first be enabled with CAL:STATe, and the value must be made permanent with CAL:SAVE. Command syntax CALibrate:SHUNt CALibration:SHUNt Parameters 1E-6 to 1E6 | MINimum | MAXimum Unit Ω (ohms) Examples CAL:SHUN .01

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:STATe CALibration:STATe This command enables and disables calibration mode. The calibration mode must be enabled before the ac source will accept any other calibration commands. The first parameter specifies the enabled or disabled state. The second parameter is the password. It is required if the calibration mode is being enabled and the existing password is not 0. If the password is not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query statement returns only the state, not the password. Whenever the calibration state is changed from enabled to disabled, any new calibration constants are lost unless they have been stored with CALibrate:SAVE. Command syntax CALibrate:STATe [, ] CALibration:STATe [, ] Parameters 0 | 1 | OFF | ON [, ] *RST value OFF Examples CAL:STAT 1, 6812 | CAL:STAT OFF Query syntax CALibrate:STATe? CALibration:STATe? Returned parameters Related commands CAL:PASS CAL:SAVE

Keysight Models 6811C, 6812C, and 6813C Programming Guide

69 CALibrate:STRing CALibration:STRing Stores a message in calibration memory. Common messages include the last calibration date, calibration due date, or calibration department contact information. You can perform this query regardless of whether the ac source is secured. Storing a calibration message overwrites the previous message. This setting is non-volatile; it will not be changed by power cycling or *RST or SYSTem:PRESet. CALibrate:VOLTage:AC CALibration:VOLTage:AC This command can only be used in calibration mode. It initiates the calibration of the AC voltage programming and metering circuits. Command syntax CALibrate:STRing CALibration:STRing Parameters *RST value OFF Examples CAL:STR “CAL BY JOHN [EXT: 1234]” Query syntax CALibrate:STRing? CALibration:STRing? Returned parameters Related commands CAL:DATE Command syntax CALibrate:VOLTage:AC CALibration:VOLTage:AC Parameters None Examples CAL:VOLT:AC Related commands CAL:SAVE CAL:STAT

Keysight Models 6811C, 6812C, and 6813C Programming Guide

CALibrate:VOLTage:AUXiliary CALibration:VOLTage:AUXiliary Keysight 6812C/6813C Option 026 only. Initiates calibration of the auxiliary output voltage. CALibrate:VOLTage:DC CALibration:VOLTage:DC This command can only be used in calibration mode. It initiates the calibration of the DC voltage programming circuits. CALibrate:VOLTage:EXTernal CALibration:VOLTage:EXTernal Keysight 6811C/6812C/6813C Option 020 only. Initiates calibration of the external voltage measurement. Command syntax CALibrate:VOLTage:AUXiliary[:AC] CALibration:VOLTage:AUXiliary[:AC] Parameters None Examples CAL:VOLT:AUX Command syntax CALibrate:VOLTage:DC CALibration:VOLTage:DC Parameters None Examples CAL:VOLT:DC Related commands CAL:SAVE CAL:STAT Command syntax CALibrate:VOLTage:EXTernal CALibration:VOLTage:EXTernal Parameters None Examples CAL:VOLT:EXT

Keysight Models 6811C, 6812C, and 6813C Programming Guide

71 CALibrate:VOLTage:OFFSet CALibration:VOLTage:OFFSet This command can only be used in calibration mode. It initiates the calibration of the offset voltage programming circuits. CALibrate:VOLTage:PROTection CALibration:VOLTage:PROTection This command can only be used in calibration mode. It calibrates the overvoltage protection (OV) circuit. The command automatically performs the calibration. CALibrate:SAVE is required to store the new OV constant in non-volatile memory. CALibrate:VOLTage:PROTection is a sequential command that takes several seconds to complete. Command syntax CALibrate:VOLTage:OFFSet CALibration:VOLTage:OFFSet Parameters None Examples CAL:VOLT:OFFS Related commands CAL:SAVE CAL:STAT CAL:LEV Command syntax CALibrate:VOLTage:PROTection CALibration:VOLTage:PROTection Parameters None Examples CAL:VOLT:PROT Related commands CAL:SAVE CAL:STAT all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Display Subsystem Commands This subsystem programs the front panel display of the ac source. Subsystem syntax DISPlay This command turns the front panel display on and off. It does not affect the annunciators. DISPlay [:WINDow] [:STATe] Enable/disable front panel display :MODE Set display mode (NORM | TEXT) :TEXT [:DATA] Set text displayed in text mode :CLEar Clears text string set on DISP:TEXT Command syntax DISPlay[:WINDow][:STATe] Parameters 0 | 1 | OFF | ON *RST value ON Examples DISP:STAT 1 DISP:STAT OFF Query syntax DISPlay[:WINDow]:STATe? Returned parameters 0 | 1 Related commands DISP:MODE DISP:TEXT

Keysight Models 6811C, 6812C, and 6813C Programming Guide

73 DISPlay:MODE This command sets the display to show either normal instrument functions, or to show a text message. Text messages are defined with DISPlay:TEXT:DATA. DISPlay:TEXT This command sets the character string that is displayed when the display mode is set to TEXT. The argument is a quoted string limited to upper case alpha characters and numbers. The display is capable of showing up to 14 characters. If the string exceeds the display capacity, it will be truncated. DISPlay:TEXT:CLEar Clears text string set on DISP:TEXT. Command syntax DISPlay[:WINDow]:MODE Parameters NORMal | TEXT *RST value NORMal Examples DISP:MODE TEXT Query syntax DISPlay[:WINDow]:MODE? Returned parameters Related commands DISP DISP:TEXT Command syntax DISPlay[:WINDow]:TEXT[:DATA] Parameters *RST value Null string Examples DISP:TEXT “DO TEST1” Query syntax DISPlay[:WINDow]:TEXT? Returned parameters (the last programmed string) Related commands DISP DISP:MODE Command syntax DISPlay[:WINDow]:TEXT:CLEar Parameters None Examples DISP:TEXT:CLE Related commands DISP:MODE DISP:TEXT

Keysight Models 6811C, 6812C, and 6813C Programming Guide

LXI Subsystem Commands This subsystem supports LXI functionality. Subsystem syntax LXI:IDENtify[:STATe] Shows or hides the LXI Web Identify indicator on the display. The indicator helps you identify the device associated with the LAN address. Press the [Local] key or send *RST to turn off the indicator. LXI :IDENtify [:STATe] Shows or hides the LXI Web Identify indicator on the display. :MDNS [:STATe] Disables or enables the mDNS. :ENABle Disables or enables the mDNS. :HNAMe [:RESolved]? Returns the resolved (unique) mDNS hostname. :SNAMe :DESired Sets the desired mDNS service name. [:RESolved]? Returns the desired mDNS service name. :RESet Resets LAN settings to a known operating state. :RESTart Restarts the LAN with the specified settings. Command syntax LXI:IDENtify[:STATe] Parameters 0 | 1 | OFF | ON *RST value OFF Examples LXI:IDEN:STAT ON LXI:IDEN 0 Query syntax LXI:IDENtify[:STATe]? Returned parameters 0 | 1

Keysight Models 6811C, 6812C, and 6813C Programming Guide

75 LXI:MDNS[:STATe] Disables or enables the multicast Domain Name System (mDNS), which provides the capabilities of a DNS server for service discovery in a small network without a DNS server. This setting is non-volatile; it is not changed by power cycling, *RST, or SYSTem:PRESet. This parameter is set to its default value when the ac source is shipped from the factory and after SYSTem:SECurity:IMMediate. LXI:MDNS:ENABle Disables or enables the multicast Domain Name System (mDNS), which provides the capabilities of a DNS server for service discovery in a small network without a DNS server. This setting is non-volatile; it is not changed by power cycling, *RST, or SYSTem:PRESet. This parameter is set to its default value when the ac source is shipped from the factory and after SYSTem:SECurity:IMMediate. Command syntax LXI:MDNS[:STATe] Parameters 0 | 1 | OFF | ON *RST value ON Examples LXI:MDNS ON LXI:MDNS 1 Query syntax LXI:MDNS[:STATe]? Returned parameters 0 | 1 Related commands LXI:MDNS:ENAB Command syntax LXI:MDNS:ENABle Parameters 0 | 1 | OFF | ON *RST value ON Examples LXI:MDNS:ENAB ON LXI:MDNS:ENAB 0 Query syntax LXI:MDNS:ENABle? Returned parameters 0 | 1 Related commands LXI:MDNS

Keysight Models 6811C, 6812C, and 6813C Programming Guide

LXI:MDNS:HNAMe[:RESolved]? Returns the resolved (unique) mDNS hostname in the form K---N, where is the last 5 digits of the ac source's serial number. N is an integer appended if necessary, so that the name is unique. The desired name may be truncated, if necessary, to make room for the appended integer. LXI:MDNS:SNAMe:DESired Sets the desired mDNS service name. This setting is non-volatile; it is not changed by power cycling, a *RST, or SYSTem:PRESet. This parameter is set to its default value when the ac source is shipped from the factory and after SYSTem:SECurity:IMMediate. Query syntax LXI:MDNS:HNAMe[:RESolved]? Parameters None Examples LXI:MDNS:HNAM? Returned parameters Related commands LXI:MDNS:SNAM Command syntax LXI:MDNS:SNAMe:DESired Parameters Examples LXI:MDNS:SNAM:DES “Keysight 6811C ACS” Query syntax LXI:MDNS:SNAMe:DESired? Returned parameters Related commands LXI:MDNS:SNAM:RES all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

77 LXI:MDNS:SNAMe[:RESolved]? The resolved mDNS service name will be the desired service name (LXI:MDNS:SNAMe:DESired), possibly with “()” appended. The N is an integer appended if necessary, to make the name unique. The desired name may be truncated, if necessary, to make room for the appended integer. LXI:RESet Resets LAN settings to a known operating state, beginning with DHCP, and clears the Web Interface password. If DHCP fails, it uses Auto-IP. Depending on your network, the LAN interface may take several seconds to restart after this command is sent. LXI:RESTart Restarts the LAN with the current settings as specified by the SYSTem:COMMunicate:LAN commands. Depending on your network, the LAN interface may take several seconds to restart after this command is sent. Query syntax LXI:MDNS:SNAMe[:RESolved]? Parameters None Examples LXI:MDNS:SNAM? Returned parameters Related commands LXI:MDNS:HNAM Command syntax LXI:RESet Parameters None Examples LXI:RES Related commands LXI:REST Command syntax LXI:RESTart Parameters None Examples LXI:REST Related commands LXI:RES

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Measurement Subsystem (Arrays) Commands This subsystem lets you retrieve arrays containing measurements data. Only current and voltage measurements are stored in an array. Two measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition of new data before returning the readings from the array. FETCh returns previously acquired data from the array. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure | FETCh :ARRay :CURRent [:DC]? Returns the digitized instantaneous current :HARMonic [:AMPLitude]? Returns amplitudes of the first 50 harmonics :PHASe? Returns phase angles of the first 50 harmonics :VOLTage [:DC]? Returns the digitized instantaneous voltage :HARMonic [:AMPLitude]? Returns amplitudes of the first 50 harmonics :PHASe? Returns phase angles of the first 50 harmonics

Keysight Models 6811C, 6812C, and 6813C Programming Guide

79 MEASure:ARRay:CURRent? FETCh:ARRay:CURRent? These queries return an array containing the instantaneous output current in amperes. The output voltage and current are digitized whenever a measure command is given or whenever an acquire trigger occurs. If digitization is caused by a measure command, the time interval between samples is determined by the output frequency. For frequencies greater than 45 Hz, the time interval is 25 microseconds. If digitization is caused by an acquire trigger, the time interval is set by SENSe:SWEep:TINTerval, and the position of the trigger relative to the beginning of the data buffer is determined by SENSe:SWEep:OFFSet:POINts. Query syntax MEASure:ARRay:CURRent[:DC]? FETCh:ARRay:CURRent[:DC]? Parameters None Examples MEAS:ARR:CURR? FETC:ARR:CURR? Returned parameters 4096 NR3 values Related commands MEAS:ARR:VOLT?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:ARRay:CURRent:HARMonic? FETCh:ARRay:CURRent:HARMonic? These queries return an array of harmonic amplitudes of output current in rms amperes. The first value returned is the DC component, the second value is the fundamental frequency, and so on up to the 50th harmonic. Harmonic orders can be measured up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus, the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:ARRay:CURRent:HARMonic[:AMPLitude]? FETCh:ARRay:CURRent:HARMonic[:AMPLitude]? Parameters None Examples MEAS:ARR:CURR:HARM? FETC:ARR:CURR:HARM? Returned parameters 51 NR3 values Related commands MEAS:ARR:VOLT:HARM? MEAS:ARR:CURR:HARM:PHAS?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

81 MEASure:ARRay:CURRent:HARMonic:PHASe? FETCh:ARRay:CURRent:HARMonic:PHASe? These queries return an array of harmonic phases of output current in degrees, referenced to the positive zero crossing of the fundamental component. The first value returned is the DC component (always returned as 0 degrees phase), the second value is the fundamental frequency, and so on up to the 50th harmonic. Harmonic orders can be measured up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:ARRay:CURRent:HARMonic:PHASe? FETCh:ARRay:CURRent:HARMonic:PHASe? Parameters None Examples MEAS:ARR:CURR:HARM:PHAS? FETC:ARR:CURR:HARM:PHAS? Returned parameters 51 NR3 values Related commands MEAS:ARR:VOLT:HARM:PHAS? MEAS:ARR:CURR:HARM? all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:ARRay:VOLTage? FETCh:ARRay:VOLTage? These queries return an array containing the instantaneous output voltage in volts. The output voltage and current are digitized whenever a measure command is given or whenever an acquire trigger occurs. If digitization is caused by a measure command, the time interval between samples is determined by the output frequency. For frequencies greater than 45 Hz, the time interval is 25 microseconds. If digitization is caused by an acquire trigger, the time interval is set by SENSe:SWEep:TINTerval, and the position of the trigger relative to the beginning of the data buffer is determined by SENSe:SWEep:OFFSet:POINts. Query syntax MEASure:ARRay:VOLTage[:DC]? FETCh:ARRay:VOLTage[:DC]? Parameters None Examples MEAS:ARR:VOLT? FETC:ARR:VOLT? Returned parameters 4096 NR3 values Related commands MEAS:ARR:CURR?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

83 MEASure:ARRay:VOLTage:HARMonic? FETCh:ARRay:VOLTage:HARMonic? These queries return an array of harmonic amplitudes of output voltage in rms volts. The first value returned is the DC component, the second value is the fundamental frequency, and so on up to the 50th harmonic. Harmonic orders can be measured up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus, the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:ARRay:VOLTage:HARMonic[:AMPLitude]? FETCh:ARRay:VOLTage:HARMonic[:AMPLitude]? Parameters None Examples MEAS:ARR:VOLT:HARM? FETC:ARR:VOLT:HARM? Returned parameters 51 NR3 values Related commands MEAS:ARR:CURR:HARM? MEAS:ARR:VOLT:HARM:PHAS?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:ARRay:VOLTage:HARMonic:PHASe? FETCh:ARRay:VOLTage:HARMonic:PHASe? These queries return an array of harmonic phases of output voltage in degrees, referenced to the positive zero crossing of the fundamental component. The first value returned is the DC component (always returned as 0 degrees phase), the second value is the fundamental frequency, and so on up to the 50th harmonic. Harmonic orders can be measured up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:ARRay:VOLTage:HARMonic:PHASe? FETCh:ARRay:VOLTage:HARMonic:PHASe? Parameters None Examples MEAS:ARR:VOLT:HARM:PHAS? FETC:ARR:VOLT:HARM:PHAS? Returned parameters 51 NR3 values Related commands MEAS:ARR:CURR:HARM:PHAS? MEAS:ARR:VOLT:HARM?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

85 Measurement Subsystem (Current) Commands This subsystem programs the current measurement capability of the ac source. Two measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition of new measurement data before returning a reading. FETCh returns a reading computed from previously acquired data. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure | FETCh [:SCALar] :CURRent [:DC]? Returns DC component of the current :AC? Returns AC rms current :ACDC? Returns AC+DC rms current :AMPLitude :MAXimum? Returns peak current :CREStfactor? Returns current crest factor :HARMonic [:AMPLitude]? Returns amplitude of the Nth harmonic of current :PHASe? Returns phase of the Nth harmonic of current :THD? Returns % of total harmonic distortion of current

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:CURRent? FETCh:CURRent? These queries return the DC component of the output current being sourced at the output terminals. MEASure:CURRent:AC? FETCh:CURRent:AC? These queries return the AC component rms current being sourced at the output terminals. Query syntax MEASure:[SCALar]:CURRent[:DC]? FETCh:[SCALar]:CURRent[:DC]? Parameters None Examples MEAS:CURR? FETC:CURR? Returned parameters Related commands MEAS:VOLT? MEAS:CURR:AC? Query syntax MEASure:[SCALar]:CURRent:AC? FETCh:[SCALar]:CURRent:AC? Parameters None Examples MEAS:CURR:AC? FETC:CURR:AC? Returned parameters Related commands MEAS:VOLT:AC? MEAS:CURR? all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

87 MEASure:CURRent:ACDC? FETCh:CURRent:ACDC? These queries return the AC+DC rms current being sourced at the output terminals. MEASure:CURRent:AMPLitude:MAXimum? FETCh:CURRent:AMPLitude:MAXimum? These queries return the absolute value of the peak current as sampled over one measurement acquisition of 4096 data points. Query syntax MEASure:[SCALar]:CURRent:ACDC? FETCh:[SCALar]:CURRent:ACDC? Parameters None Examples MEAS:CURR:ACDC? FETC:CURR:ACDC? Returned parameters Related commands MEAS:VOLT:ACDC? MEAS:CURR:AMPL:MAX? Query syntax MEASure:[SCALar]:CURRent:AMPLitude:MAXimum? FETCh:[SCALar]:CURRent:AMPLitude:MAXimum? Parameters None Examples MEAS:CURR:AMPL:MAX? FETC:CURR:AMPL:MAX? Returned parameters Related commands MEAS:CURR:ACDC? MEAS:CURR:CRES?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:CURRent:CREStfactor? FETCh:CURRent:CREStfactor? These queries return the output current crest factor. This is the ratio of peak output current to rms output current. MEASure:CURRent:HARMonic? FETCh:CURRent:HARMonic? These queries return the rms amplitude of the Nth harmonic of output current. The parameter is the desired harmonic number. Queries sent with a value of 0 return the DC component. A value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:[SCALar]:CURRent:CREStfactor? FETCh:[SCALar]:CURRent:CREStfactor? Parameters None Examples MEAS:CURR:CRES? FETC:CURR:CRES? Returned parameters Related commands MEAS:CURR:ACDC? MEAS:CURR:AMPL:MAX? Query syntax MEASure:[SCALar]:CURRent:HARMonic[:AMPLitude]? FETCh:[SCALar]:CURRent:HARMonic[:AMPLitude]? Parameters 0 to 50 Examples MEAS:CURR:HARM? 3 FETC:CURR:HARM? 1 Returned parameters Related commands MEAS:CURR:HARM:PHAS? MEAS:CURR:HARM:THD?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

89 MEASure:CURRent:HARMonic:PHASe? FETCh:CURRent:HARMonic:PHASe? These queries return the phase angle of the Nth harmonic of output current, referenced to the positive zero crossing of the fundamental component. The parameter is the desired harmonic number. Queries sent with a value of 0 return the DC component. A value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. MEASure:CURRent:HARMonic:THD? FETCh:CURRent:HARMonic:THD? These queries return the percentage of total harmonic distortion and noise in the output current. Query syntax MEASure:[SCALar]:CURRent:HARMonic:PHASe? FETCh:[SCALar]:CURRent:HARMonic:PHASe? Parameters 0 to 50 Examples MEAS:CURR:HARM:PHAS? 3 FETC:CURR:HARM:PHAS? 1 Returned parameters Related commands MEAS:CURR:HARM? MEAS:CURR:HARM:THD? Query syntax MEASure:[SCALar]:CURRent:HARMonic:THD? FETCh:[SCALar]:CURRent:HARMonic:THD? Parameters None Examples MEAS:CURR:HARM:THD? FETC:CURR:HARM:THD? Returned parameters Related commands MEAS:CURR:HARM? MEAS:CURR:HARM:PHAS?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Measurement Subsystem (Frequency) Commands This subsystem programs the frequency measurement capability of the ac source. Two measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition of new measurement data before returning a reading. FETCh returns a reading computed from previously acquired data. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure:FREQuency? FETCh:FREQuency? This query returns the output frequency in hertz. For Keysight 6811C/6812C/6813C Option 020 only, when the Power Analyzer input is selected as the measurement source, this command returns the actual frequency measured at the V SENSE input channel. When the Main Output of the ac source is selected as the measurement source, this command returns the programmed output frequency. MEASure | FETCh [:SCALar] :FREQuency Returns the output frequency NOTE With no signal connected to the Power Analyzer input, this command may return a very high frequency value. This is normal, since there is no external frequency for the measurement circuit to detect. Query syntax MEASure:[SCALar]:FREQuency? FETCh:[SCALar]:FREQuency? Parameters None

Keysight Models 6811C, 6812C, and 6813C Programming Guide

91 Examples MEAS:FREQ? FETC:FREQ? Returned parameters all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Measurement Subsystem (Power) Commands This subsystem programs the power measurement capability of the ac source. Two measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition of new measurement data before returning a reading. FETCh returns a reading computed from previously acquired data. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure:POWer? FETCh:POWer? These queries return the DC component of the power being sourced at the output terminals in watts. MEASure | FETCh [:SCALar] :POWer [:DC]? Returns the DC component of power :AC [:REAL]? Returns real power :APParent? Returns VA :REACtive? Returns VAR :PFACtor? Returns power factor Query syntax MEASure:[SCALar]:POWer[:DC]? FETCh:[SCALar]:POWer[:DC]? Parameters None Examples MEAS:POW? FETC:POW? Returned parameters Related commands MEAS:POW:AC?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

93 MEASure:POWer:AC? FETCh:POWer:AC? These queries return the in-phase component of power being sourced at the output terminals in watts. MEASure:POWer:AC:APParent? FETCh:POWer:AC:APParent? These queries return the apparent power being sourced at the output terminals in volt-amperes. Query syntax MEASure:[SCALar]:POWer:AC[:REAL]? FETCh:[SCALar]:POWer:AC[:REAL]? Parameters None Examples MEAS:POW:AC? FETC:POW:AC? Returned parameters Related commands MEAS:POW? Query syntax MEASure:[SCALar]:POWer:AC:APParent? FETCh:[SCALar]:POWer:AC:APParent? Parameters None Examples MEAS:POW:AC:APP? FETC:POW:AC:APP? Returned parameters Related commands MEAS:POW:REAC? MEAS:POW:PFAC?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:POWer:AC:REACtive? FETCh:POWer:AC:REACtive? These queries return the reactive power being sourced at the output terminals in volt-amperes reactive (VAR). Reactive power is computed as: MEASure:POWer:AC:PFACtor? FETCh:POWer:AC:PFACtor? These queries return the output power factor. The power factor is computed as: Query syntax MEASure:[SCALar]:POWer:AC:REACtive? FETCh:[SCALar]:POWer:AC:REACtive? Parameters None Examples MEAS:POW:AC:REAC? FETC:POW:AC:REAC? Returned parameters Related commands MEAS:POW:AC:APP? MEAS:POW:PFAC? VAR (Apparent Power )2 (Real Power )2 – ( ) = Query syntax MEASure:[SCALar]:POWer:AC:PFACtor? FETCh:[SCALar]:POWer:AC:PFACtor? Parameters None Examples MEAS:POW:AC:PFAC? FETC:POW:AC:PFAC? Returned parameters Related commands MEAS:POW:AC:APP? MEAS:POW:REAC? PFactor Real Power Apparent Power -------------------------------- =

Keysight Models 6811C, 6812C, and 6813C Programming Guide

95 Measurement Subsystem (TVOLt) Commands For Keysight 6811C/6812C/6813C Option 020 only, these measurement commands are optimized for use in specific applications such as evaluating uninterruptible power supplies. In this application the new commands can be used to measure transfer time, peak voltage, and phase locked loop performance of the UPS. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure | FETCh [:SCALar] :TVOLt? , Returns the time :TVOLt :ABSolute? , , Returns the time

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:TVOLt? FETCh:TVOLt? Keysight 6811C/6812C/6813C Option 020 only. Returns the time at which the measured voltage crosses a user-specified voltage level. The sign and magnitude of occurrence define the event to be reported. For example, if occurrence is –3, the return value is the third time at which the voltage crosses the level in the negative-going direction. If occurrence is 4, the return value is the fourth time at which the voltage crosses the level in the positive-going direction. The return value is with respect to the start of the measurement buffer. If no point on the waveform satisfies the specified conditions, the return value is 9.91000E+37. Query syntax MEASure:[SCALar]:TVOLt? , FETCh:[SCALar]:TVOLt? , Parameters –1E6 to 1E6 (for level) –4096 to +4096, but not 0 (for occurrence) Examples MEAS:TVOL? 50, 3 FETC:TVOL? 10, -1 Returned parameters Related commands MEAS:TVOL:ABS? all-guides.co m

Keysight Models 6811C, 6812C, and 6813C Programming Guide

97 MEASure:TVOLt:ABSolute? FETCh:TVOLt:ABSolute? Keysight 6811C/6812C/6813C Option 020 only. Returns the time at which the absolute value of the measured voltage first exceeds level, with the following qualifications: The behavior of the signal before start_time is ignored, and the absolute value of the voltage must remain above level for at least the specified min_pulse_width. Both start_time and the return value are with respect to the start of the measurement buffer. If no point on the waveform satisfies the specified conditions, the return value is 9.91000E+37. Query syntax MEASure:[SCALar]:TVOLt:ABSolute? , , FETCh:[SCALar]:TVOLt:ABSolute? , , Parameters 0 to 1E6 (for level) 0 to 1E6 (for start time in seconds) 0 to 1E6 (for minimum pulse width in seconds) Examples MEAS:TVOL:ABS? 50, 0, 0.001 FETC:TVOL:ABS? 100, 0.005, 0 Returned parameters Related commands MEAS:TVOL?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

Measurement Subsystem (Voltage) Commands This subsystem programs the voltage measurement capability of the ac source. Two measurement commands are available: MEASure and FETCh. MEASure triggers the acquisition of new measurement data before returning a reading. FETCh returns a reading computed from previously acquired data. For Keysight 6811C/6812C/6813C Option 020 only, all standard ac source MEASure or FETCh functions are available for the Main Output of the ac source as well as the Power Analyzer input. The SENSe:NSELect command selects the measurement source that will return data. Subsystem syntax MEASure | FETCh [:SCALar] :VOLTage [:DC]? Returns DC component of the voltage :AC? Returns AC rms voltage :ACDC? Returns AC+DC rms voltage :HARMonic [:AMPLitude]? Returns amplitude of the Nth harmonic of voltage :PHASe? Returns phase of the Nth harmonic of voltage :THD? Returns % of total harmonic distortion of voltage :RANKed? Returns the voltage value for the percentile rank given

Keysight Models 6811C, 6812C, and 6813C Programming Guide

99 MEASure:VOLTage? FETCh:VOLTage? These queries return the DC component of the output voltage being sourced at the output terminals. MEASure:VOLTage:AC? FETCh:VOLTage:AC? These queries return the AC rms voltage being sourced at the output terminals. Query syntax MEASure:[SCALar]:VOLTage[:DC]? FETCh:[SCALar]:VOLTage[:DC]? Parameters None Examples MEAS:VOLT? FETC:VOLT? Returned parameters Related commands MEAS:CURR? MEAS:VOLT:AC? Query syntax MEASure:[SCALar]:VOLTage:AC? FETCh:[SCALar]:VOLTage:AC? Parameters None Examples MEAS:VOLT:AC? FETC:VOLT:AC? Returned parameters Related commands MEAS:CURR:AC? MEAS:VOLT?

Keysight Models 6811C, 6812C, and 6813C Programming Guide

MEASure:VOLTage:ACDC? FETCh:VOLTage:ACDC? These queries return the AC+DC rms voltage being sourced at the output terminals. MEASure:VOLTage:HARMonic? FETCh:VOLTage:HARMonic? These queries return the rms amplitude of the Nth harmonic of output voltage. The parameter is the desired harmonic number. Queries sent with a value of 0 return the DC component. A value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the fundamental measurement bandwidth of the measurement system, which is 12.6 kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6 kHz are returned as 0. Query syntax MEASure:[SCALar]:VOLTage:ACDC? FETCh:[SCALar]:VOLTage:ACDC? Parameters None Examples MEAS:VOLT:ACDC? FETC:VOLT:ACDC? Returned parameters Related commands MEAS:CURR:ACDC? MEAS:VOLT? Query syntax MEASure:[SCALar]:VOLTage:HARMonic[:AMPLitude]? FETCh:[SCALar]:VOLTage:HARMonic[:AMPLitude]? Parameters 0 to 50 Examples MEAS:VOLT:HARM? 3 FETC:VOLT:HARM? 1 Returned parameters Related commands MEAS:VOLT:HARM:PHAS? MEAS:VOLT:HARM:THD?