BTM Super Check Pet Quick Start Manual
Read Online
Read the BTM Super Check Pet Quick Start Manual online using the built-in PDF viewer.
Text of the manual
The LCD screen will show as below:
1
SET
2
SET
3
SET
6
SET
4
SET
5
SET
SuperCheck Pet Meter Setup
Before performing a test, setup the time and date on your meter.
With the meter off, long press the M button for 2 seconds to enter the
setting mode. Press C button to adjust the number and then press M
to confirm the number.
Setting the Alarm:
Press C key to toggle between the alarm reminder setting or no-setting
mode. If you select the alarm setting mode, alarm clock icon will stop
flashing and the hour of the time segment will start flashing. Press C
button to adjust the number and then press M to confirm the number.
(1)Measurement Units
(2)Memory
(3)Setting
(4)Blood drop icon
(5)Bettery Symbol
(6)Day Average
(7)Alarm Reminder
(8)Month
(9)Day
(10)Hour
(11)Minute
(14)Control Solution
(13)Cat Blood Testing
(12)Dog Blood Testing
Quick Start Guide
Setting the measurement unit:
Press C key to toggle between mg/dL and mmol/L measurement units.
Press M key to confirm the setting.
To delete the memory:
Press C key to toggle between a flashing “dEL” for memory deletion
or a non-flashing“dEL”. If you want to delete all memory, press M key
while the “dEL” is flashing. If you do not want to delete all memory,
press M key when the “dEL” is not flashing.
How to Perform a Blood Glucose Test
Here are the steps on how to perform a blood glucose test on your
pets with SuperCheck Pet and the Test Strips.
1.Gather all the items you needed before you start the test.
Items including: Meter, Test Strips, Lancing Device, Lancets, Control
Solution.
2.Choose a capillary sampling site.
3.Check the code on the test strip vial. If necessary, change the code
on the meter by pressing C or M after you enter a test strip.
4.Running a Blood Glucose Test
Put the lancet into your lancing device and twists off the protective
cover of the lancet. (Step 1 to 4)
Wash your hands carefully. Be sure to rinse and dry well. Take one
new strip out of the vial and recap the vial immediately. Insert the strip
into the meter. When you insert a strip, choose the code by pressing C
or M to switch from d01~d03 to c01~c03.
The dog or cat icon will be displayed on the screen to remind you
choosing the correct animal code. Use the adjustable lancing device or
disposable lancet to take a blood sample from the paw pad. The meter
will start counting down for 5 seconds and the result will be showed on
the screen.
Remove the strip by pushing the ejector button. The meter will turn off
and result is stored automatically. (Step 5 to 10)