Bacharach Portable Combustion Analyzer 24-9351 Operating & Maintenance Instructions

File Specifications

Brand: Bacharach
Category: Carbon Monoxide Alarm
Model: Portable Combustion Analyzer 24-9351 (Operating & Maintenance Instructions), pr/lw6AD.9NBo
File Info: Adobe Acrobat PDF (DjVu)
File Size: 0.56 Mb
Pages: 100 pages

Download PDF

Download Bacharach Portable Combustion Analyzer 24-9351 Manual (100 pages)

Read Online:

Read the Bacharach Portable Combustion Analyzer 24-9351 Operating & Maintenance Instructions online using the built-in PDF viewer. This document explains how to operate the device, adjust settings and perform regular maintenance.

Key Pages of Document

4.6 Sensor Status Screen

      ! Where: xx = Instrument Model Number z = Sensor(s) in error If there is problem with one or more of the sensors, the Sensor Status Screen will be displayed after the analyzer has gone through its warmup cycle (refer to Section 7.2 for a listing of the error codes). Front Panel Key Functions: – No Action – No Action – No Action – Go to Combustion Test Screen – No Action Operation ...

Do not place a hot probe inside the instrument’s carrying case.

Allow the probe to cool before storage. 2. Loosen the thumbscrew on the probe stop; then remove the probe and probe stop from the stack. 3. If data was saved during the combustion test, you can turn off the analyzer and review or print the stored data at a later time as de- scribed in Sections 4.10 and 4.23. 4.3.5 Turning Off the Analyzer and Purging the CO Sensor Turn off the analyzer by pressing the key. If the key is pressed while the CO reading is 100 ppm or higher, the pump ...

Bacharach, Inc. warrants to Buyer that at the time of delivery this Product will be free from defects

in material and manufacture and will conform substantially to Bacharach Inc.’s applicable specifications. Bacharach’s liability and Buyer’s remedy under this warranty are limited to the repair or replacement, at Bacharach’s option, of this Product or parts thereof returned to Seller at the factory of manufacture and shown to Bacharach Inc.’s reasonable satisfaction to have been defective; provided that written notice of the defect shall have been give ...

4.19 Maintenance Password Screen

    %%% Where: xxx = Password number The Maintenance Password Screen is displayed by repeatedly pressing the MENU key from the Combustion Test Screen. From here a three-digit password must be entered to access the instrument’s Maintenance Screens. The password number is provided on the Portable Combustion Analyzer Calibration Password card that was supplied with the analyzer. To enter the password, first press the key to move the cursor (z) into the first number field, and then press the st key ...

TABLE 4-2. TYPICAL SPREADSHEET FOR A PCA 65

Line 1: Column Headings Line 2: Typical Combustion Readings Line 3: Typical Draft Reading SN ID1 ID2 ID3 TIME DATE FUEL TA TL C/F O2 AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:03:27 19.01.1999 NATGAS 190 20 C 4 AX1020 ID LINE 1 ID LINE 2 ID LINE 3 9:10:35 19.01.1999 C2 CO CU NX NU qA LA DR MB/PA/WC 9.5 12 15 10 12 8.1 1.24 -0.25 MB -0.25 MB TABLE 4-1. COMMA-DELIMITED FIELDS Field Data Name or Value Label in Column Headings 1 Instrument serial number SN 2 ID line 1 (up to 16 characters) ID1 3 ID line 2 (up to ...

5.4 Calibrate TA-Zero

Material Required: • Thermocouple Simulator (K-type) Range: 0 to 300°C Accuracy: ±0.3°C Procedure: 1. With the analyzer turned off, first plug the simulator's K-type connec- tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and wait for its warmup cycle to complete. 2. Enter the Calibration Menu Screen per Section 5.3; then choose TA-ZERO to zero the analyzer’s stack temperature channel. 3. Adjust the simulator to 0° C (32° F).* 4. Wait until the MEASURED reading on the screen stabilizes. Then use the st and keys to enter ...

4.13 Temperature Setup Screen

        The Temperature Setup Screen is displayed by repeatedly pressing the MENU key from the Combustion Test Screen. Use this screen to setup the analyzer to display temperature in either °C or °F. To select the instrument’s temperature unit-of-measure, first use the st keys to move the cursor (z) in front of  or , and then press the key. Front Panel Key Functions: – Move cursor (z) up – Move cursor (z) down – Select temperature unit of measure next to cursor – Go to Combustion ...

Bacharach Portable Combustion Analyzer 24-9351 user guide recommended for: PCA 2 24-9448, MGC200, Snifit 50, ECA450, Monoxor III.

⚠️ Download Safety

The Carbon Monoxide Alarm Operating & maintenance instructions (100 pages) is completely safe to download (last scan date: 29/01/2026).

Related PDF Instructions for Bacharach Portable Combustion Analyzer 24-9351:

Carbon Monoxide Alarm Equipment: Frequently Requested

Have some Questions about Bacharach Portable Combustion Analyzer 24-9351 Device?

Get valuable and actionable advices from other Bacharach Portable Combustion Analyzer 24-9351 device users. Describe your problem or question, regarding the using of Portable Combustion Analyzer 24-9351, in as much detail as possible, and other users will be able to answer you.