TEL AFA4000/1 Installation And Operating Manual
Read Online
Read the TEL AFA4000/1 Installation And Operating Manual online using the built-in PDF viewer.
Text of the manual
AFA4000/1 Airflow Monitor / p.2
Contents
1.
Introduction.......................................................................................................................................................... 5
1.1
Features ...................................................................................................................................................... 6
2.
Safety ...................................................................................................................................................................... 7
2.1
Safety practices ........................................................................................................................................ 7
2.2
Precautions ................................................................................................................................................ 7
2.3
Warnings .................................................................................................................................................... 7
2.4
Operating conditions ............................................................................................................................. 8
2.5
Electrical safety ........................................................................................................................................ 9
2.6
Electrical protection .............................................................................................................................. 10
2.7
EMC compliance .................................................................................................................................... 11
3.
Overview: Airflow Monitor ............................................................................................................................. 12
3.1
Display ...................................................................................................................................................... 12
3.2
Alarms and events................................................................................................................................. 14
3.3
Controls and indicators ....................................................................................................................... 16
3.4
Operator display panel menu navigation ...................................................................................... 17
3.5
External connections ............................................................................................................................ 19
3.6
Functions ................................................................................................................................................. 21
3.7
Components ........................................................................................................................................... 26
4.
Overview: Auto Sash Controller ................................................................................................................... 28
4.1
Features .................................................................................................................................................... 28
4.2
Auto Sash control options .................................................................................................................. 29
4.3
Control system components .............................................................................................................. 32
5.
Installation: Airflow Monitor .......................................................................................................................... 40
5.1
Location .................................................................................................................................................... 40
5.2
Installation ............................................................................................................................................... 42
5.3
Connections ............................................................................................................................................ 45
5.4
Dimensions.............................................................................................................................................. 49
AFA4000/1 Airflow Monitor / p.3
5.5
Start up ..................................................................................................................................................... 51
6.
Installation: Auto Sash Controller ................................................................................................................ 53
6.1
Auto Sash Control Unit ........................................................................................................................ 53
6.2
Auto Sash keypad / alarm panel....................................................................................................... 56
6.3
Auto Sash motor drive unit ................................................................................................................ 57
6.4
Sash light curtain ................................................................................................................................... 65
6.5
Installing and aligning the Under Sash sensor ............................................................................. 72
6.6
Tilt switch ................................................................................................................................................. 74
6.7
Sash Low switch ..................................................................................................................................... 75
6.8
Sash position sensor ............................................................................................................................ 77
6.9
Personnel sensor ................................................................................................................................... 79
6.10 Auto Sash System Control Box - general wiring diagram ........................................................ 87
6.11 Start up ..................................................................................................................................................... 88
7.
Calibration .......................................................................................................................................................... 89
7.1
AFA4000/1 Airflow Monitor ............................................................................................................... 89
7.2
Auto Sash Controller 1, 2 or 3 position. ......................................................................................... 91
7.3
Sash lock function ................................................................................................................................. 93
7.4
Auto Sash Controller – stand-alone ................................................................................................ 94
8.
Configuration ...................................................................................................................................................103
8.1
AFA4000/1 configuration menus ...................................................................................................103
8.2
AFA4000/1/AS configuration menus ............................................................................................111
8.3
Auto Sash Controller system menus .............................................................................................123
9.
Auxiliary features and connections ...........................................................................................................131
9.1
Optional input function - temperature sensor ...........................................................................131
9.2
Optional input function – close sash alarm.................................................................................133
10.
RS485 communications output ..................................................................................................................136
10.1 Overview and connections ...............................................................................................................136
10.2 Configuration settings .......................................................................................................................138
10.3 Config Manager ...................................................................................................................................140
AFA4000/1 Airflow Monitor / p.4
11.
Operation: Airflow Monitor .........................................................................................................................142
11.1 Run screen .............................................................................................................................................142
12.
Operation: Auto Sash Controller ................................................................................................................143
12.1 Control philosophy .............................................................................................................................143
12.2 Modes of operation ............................................................................................................................143
12.3 Operation and alarm indication ......................................................................................................147
13.
Troubleshooting..............................................................................................................................................149
13.1 Fault finding ..........................................................................................................................................149
13.2 Diagnostics menu................................................................................................................................153
13.3 Auto sash alarm indications .............................................................................................................155
14.
Cleaning and maintenance ..........................................................................................................................158
14.1 Cleaning .................................................................................................................................................158
14.2 Contamination .....................................................................................................................................159
14.3 Maintenance .........................................................................................................................................159
15.
Technical specifications ................................................................................................................................160
15.1 AFA4000/1 Airflow Monitor .............................................................................................................160
15.2 Auto Sash Controller ..........................................................................................................................162
16.
Warranty ............................................................................................................................................................169
17.
Index ...................................................................................................................................................................171
AFA4000/1 Airflow Monitor / p.5
1. Introduction
The TEL AFA4000 is a system for controlling and monitoring airflow in industrial and
educational fume cupboards and biological safety cabinets for user safety. The AFA4000
airflow monitor and controller is available with optional Auto Sash integration. Using the
Auto Sash with the AFA4000 controller means you will no longer need a separate Auto
Sash keypad.
The AFA4000 user interface is full colour, intuitive and is easy to navigate. It features a
simple two-step calibration process with on-screen prompts to guide end users. The
unrivalled reliability of the AFA4000 means that once calibrated, you will not need to re-
calibrate in the future. The AFA4000 has in-built communications as standard and it can
be used with either Modbus or BACnet protocol.
Featuring an innovative hot wire sensor, the AFA4000 has no inherent drift meaning that
the sensor will provide stable readings over many years of operation ensuring reliability
and safety.
The AFA4000 has full alarm diagnostics and test functionality, a sixty-minute timeline of
environmental conditions and field-upgradable software, with password protected
menus and diagnostic menus to aid installation and commissioning.
This manual covers the AFA4000/1 and AFA4000/1/AS airflow monitors.
AFA4000/1 Airflow Monitor / p.6
1.1
Features
• Full colour 3.5” LCD Display for continuous velocity reading in m/sec
• Auto Sash integration
• BACnet and Modbus on board
• Velocity bar graph or fault time line over the last 60 minutes
• Visual & audible alarms
• Three digital inputs and three digital outputs
• Diagnostics Menu
• Bespoke overlay designs, with a choice of two, coloured cases
• Bespoke software options tailored to customer requirements
• Touch sensitive buttons
AFA4000/1 Airflow Monitor / p.7
2. Safety
2.1
Safety practices
This guide is a supplement to the safety codes of the user’s country. Not every safety
procedure that should be followed is covered by this guide. Maintaining a safe
laboratory environment is the responsibility of the user.
Review product documentation prior to use and follow instructions carefully.
If the Airflow Monitor or Auto Sash Controller are not used or operated in accordance
with this manual, the protection provided by the devices may be impaired.
2.2
Precautions
• Ensure the equipment voltage corresponds to the voltage available at the installation
position.
• Do not remove equipment panels without shutting down the equipment and
disconnecting the power supply.
2.3
Warnings
2.3.1
Warning notices
WARNINGS must be followed in order to avoid personal injury.
2.3.2
Warning labels
WARNINGS appear as follows:
Switch off the mains voltage and remove the mains cord before
maintenance.
Always follow the instructions given by warning labels on the equipment. For further
information refer to this guide and other documentation provided.
AFA4000/1 Airflow Monitor / p.8
2.4
Operating conditions
2.4.1
General operation
The Airflow Monitor and associated equipment have been designed and tested in
accordance with the International Electrotechnical Commissions (IEC) safety
requirements. The Airflow Monitor conforms to IEC61010-1 (Safety Requirements for
electrical equipment for measurement, control and laboratory use) as it applies to IEC
Class 1 (earthed) appliances, and therefore meets the requirements of EC directive
2014/30/EU.
Adjustment, maintenance or repair of the equipment whilst covers are open or the
equipment is in operation is only to be done by skilled persons who are aware of the
potential hazards.
Make unsafe equipment inoperative.
Examples of unsafe equipment:
• Appears damaged
• Fails to perform correctly
• Has undergone transport stresses
• Has been stored incorrectly
2.4.2
Environmental conditions
Ensure the equipment is used under the following conditions:
• Indoors
• Temperature, 5°C to 40°C
• Relative humidity below 80% for temperatures up to 31°C, decreasing linearly to 50%
at 40°C
• Electrical supply fluctuations not to exceed +10% of nominal voltage
The protection provided by the equipment may be impaired if the
environmental conditions do not lie within these parameters.
AFA4000/1 Airflow Monitor / p.9
2.5
Electrical safety
Lethal voltages are present at certain points within the equipment.
When the equipment is connected to mains power, removing the
equipment's covers is likely to expose live parts.
Even when the power switch is off, high voltage can still be present.
Capacitors within the equipment may still be charged even if the equipment
has been disconnected from all live voltage sources.
Connect the Airflow Monitor and associated equipment correctly to a suitable electrical
supply. The supply must have a correctly installed protective conductor (earth or
ground), which must be installed and checked by a qualified electrician before initial
power up.
Any interruption of the protective conductor inside or outside the Auto-Sash
Controller System, or disconnection of the protective conductor terminal is
likely to make the equipment dangerous.
Never interrupt the protective conductor.
If the mains power cord needs to be replaced, ensure that the replacement
cord is appropriately rated and approved for the intended use.
To prevent potential personal injury and damage to the equipment, switch
OFF all components in the system and disconnect them from the mains
power supply before altering or making any new electrical connections.
When working with the Airflow Monitor System and/or the Auto Sash Controller System:
• Do not operate equipment with any covers or internal parts removed.
• Disconnect the equipment from live voltage sources before adjustments,
replacements, maintenance or repairs are carried out. If the equipment must be
operated during adjustment, maintenance or repair, only a supplier’s Service Engineer
is to carry out the procedure.
AFA4000/1 Airflow Monitor / p.10
• Ensure equipment that is not electrically safe is made inoperative. Secure the
equipment against unauthorised or unintentional operation. Examples of electrically
unsafe equipment conditions are as follows:
• Appears damaged
• Has undergone transport stresses
• Has been stored incorrectly
2.6
Electrical protection
Observe the following electrical protection precautions:
Insulation:
Class I rating for external circuits.
Only connect equipment that meets the requirements of IEC
61010-1, IEC 60950 or equivalent standards.
Installation category:
The equipment can withstand transient over-voltages
typically present on the mains supply. The normal level of
transient over-voltages is impulse withstand (overvoltage)
Category II of IEC 60364-4-443.
Pollution degree 2:
Normally only non-conductive pollution occurs. Occasionally,
however, temporary conductivity caused by condensation
must be expected.
External circuits:
External circuits which are connected to the Auto Sash
controller, except mains connection, shall be insulated by
double or reinforced insulation from the mains.
AFA4000/1 Airflow Monitor / p.11
2.7
EMC compliance
2.7.1
EC Directive
The Airflow Monitor and Auto Sash Controller Systems are designed and tested to meet
the requirements of the EC directive 2014/30/EU and complies with the EMC standard
EN61326 (EMC standard for electrical equipment for measurement, control and
laboratory use) and EN55011 (ISM) Class A (RF emissions).
2.7.2
FCC Rules and Regulations
The Airflow Monitor System and Auto Sash Controller Systems are classified as digital
devices used exclusively as industrial, commercial or medical test equipment. They are
exempt from the technical standards specified in Part 15 of the FCC Rules and
Regulations based on Section 15.103 (c).
AFA4000/1 Airflow Monitor / p.12
3. Overview: Airflow Monitor
3.1
Display
3.1.1
Operator display panel
A
Airflow bargraph or timeline display
B
Airflow velocity display with LED halo
(red/amber/green)
C
Status window, airflow alarm status
Auto Sash status (AFA4000/1/AS only)
D
Menu pushbuttons and alarm mute
Auto Sash Up/Down/Cancel buttons
(AFA4000/1/AS only)
Figure 1: Operator display panel
Note:
Access to the Calibration and Configuration menus is password protected and is
factory set.
To access and or change the password contact the supplier for the engineer’s
password and enter the passwords in the Main Menu, or alternatively use a
laptop connected to the Com port and use the Upload/Download software
provided.
A
B
C
D
AFA4000/1 Airflow Monitor / p.13
3.1.2
Display features
The airflow monitor displays the following features:
• The digital display is a backlit, full colour high resolution graphic unit with a visual
display area of approximately 45 x 34 mm. The display operates through the software
allowing the generation of figures, words and icons. You can choose either blue or
black for the background colour of the display.
• Using the alarm keypad, you can configure the display to show the fume cupboard
velocity in m/sec or fpm. Alternatively, you can configure the display to show AIR
FAIL / AIR SAFE continuously.
The colour of the velocity on the display screen changes when in an alarm condition, to
the following:
• Air Safe = White
• Low/ High Air Alarm = Red
• Warning Air Alarm = Amber
An ‘event time line’ segmented into 60 x 1 minute segments scroll across the display
(when enabled). The event time line is displayed as a graphical scale ranging over 0-1.00
m/sec that progresses across the screen, representing the airflow value at each segment.
The segment colour changes, as above, if the value is in the range of an airflow alarm.
• The alternative to the event time line is a dynamic ‘bar graph’ representing the
airflow velocity.
The display shows an up and down arrow icon,
, in the bottom right hand corner of
the screen when an input function is set to Hi / Lo (2 speed operation). The up arrow
indicates High speed and the down arrow indicates Low speed.
• The display shows a Horn icon (with a line across it) when the audible alarm is muted.
• The display backlight dims to save energy when the following functions are in use:
• Fan Off activated- display backlight dims until Fan On is selected.
Backlight brightens to normal level when you access the Set Up or Diagnostics
menus.
AFA4000/1 Airflow Monitor / p.14
• Setback activated- display backlight dims until the Setback is deactivated.
Backlight brightens to normal level when you access the Set Up or Diagnostics
menus.
3.2
Alarms and events
Message
Displayed when
Sash high
The Sash alarm is activated, and the sash is raised above the
maximum safe working opening. Sash high alternates on/off with
the velocity reading.
External alarm
The external alarm input is activated. External alarm alternates
on/off with the velocity reading.
Alarm fail
The airflow is less than the low air alarm point. Air fail alternates
on/off with the velocity reading.
High alarm
The airflow is more than the high air alarm point. High Air
alternates on/off with the velocity reading.
Setback
The night setback function is activated. Setback alternates on/off
with the velocity reading. The display backlight dims.
Alarm disabled
The alarm disable function is activated. Alarm disable alternates
on/off with the velocity reading.
Close sash
The sash is raised and the operator is not present (when activated).
Close sash alternates on/off with the velocity reading.
Emergency
The emergency input is activated. Emergency alternates on/off with
the velocity reading.
Up / down arrow
The Hi/Lo 2 Speed operation is enabled.
Message
Displayed when
Mains fail
The power fails to connect to the monitor, when activated.
Note: This is an optional extra feature that requires an additional
battery unit.
AFA4000/1 Airflow Monitor / p.15
Message
Displayed when
Low temperature
The cupboard temperature drops below the low temperature alarm
point, when activated. The display alternates on/off with the
velocity reading.
Note: This is an optional extra feature that requires an additional
temperature sensor.
High temperature
The fume cupboard temperature rises above the high temperature
alarm point, when activated.
Note: This is an optional extra feature that requires an additional
temperature sensor.
AFA4000/1 Airflow Monitor / p.16
3.3
Controls and indicators
3.3.1
LED indicators
The alarm unit has three LED indicators:
Red- Alarm
Amber- Caution
Green- Safe
3.3.2
Audible alarm sounder
The AFA4000/1 has an audible sounder with local or remote mute facilities. You can
permanently disable the alarm in the calibration and configuration menus.
You can mute the alarm by activating the Setback or Alarm Disable functions.
When the audible alarm is muted/ disabled, the mute icon displays at the bottom left
side of the screen.
When an input is set to the mute function the keypad Mute button is disabled and you
can only mute the alarm by using the selected mute input.
3.3.3
Pushbuttons
Mute
The alarm has a multifunctional Mute button,
.
Press Mute once, when the alarm is sounding to mute the alarm.
Press Mute for 5 seconds to access the calibration and configuration
menus.
↑ / ↓
Use the ↑ and ↓ buttons to scroll through the calibration and configuration
menus or to select options and values.
AFA4000/1 Airflow Monitor / p.17
3.4
Operator display panel menu navigation
To scroll through menu items, press the ↑ and ↓ arrows, then press Mute to select an
option.
3.4.1
Main menu
To display the Main Menu, from the Run Screen, press and hold the Mute button for 5
seconds until the Main Menu is displayed.
Menu item
Option
Instructions
Run
Run screen
See Operation section
Set up Monitor
Monitor setup menu
See Monitor setup menu
section
Set up Auto Sash
Note: This option is only
available if the Auto Sash is
connected.
Auto Sash Configuration
menu
See Auto Sash
Configuration menu
section
3.4.2
Monitor setup menu
Menu item
Option
Instructions
Configure *
Monitor config menu
See section 8.
Calibration *
Calibration
See section 7.
Passwords *
Monitor passwords menu
Select the password to set:
• Admin
• Calibration
• Configure
Enter and re-enter the
new password, then press
the Mute button.
Select Done to return to
the Main menu.
AFA4000/1 Airflow Monitor / p.18
Menu item
Option
Instructions
Done
Return to Main menu
Note:
Menu items marked * are password protected. To enter the password, use ↑ / ↓
to change each digit of the password, then press Mute.
3.4.3
Auto Sash Configuration menu
Note:
This option is only available if the Auto Sash is connected.
Menu item
Option
Instructions
TEL Sash Closer
Shows the firmware (F/W)
and Hardware (H/W) versions
Read-only information
Setup Menu *
Auto Sash setup menu
See section 8.2.1.
Engineering Menu *
Auto Sash engineering
menu
See section 8.2.2.
Exit
Return to Main menu
Note:
Menu items marked * are password protected. To enter the password, use ↑ / ↓
to change each digit of the password, then press Mute.
AFA4000/1 Airflow Monitor / p.19
3.5
External connections
3.5.1
Inputs
The AFA4000/1 has the following inputs:
3.5.1.1
Input 1, Input2, Input 3
These are volt-free inputs configurable for:
• Normally closed relays
• Normally open relays or
• Analogue 0 - 5 Vdc input
The inputs can be configured as:
Digital input functions (closed or
open volt free contact):
Alarm disabled
Night set-back
External alarm
Emergency
Sash high
High / Low
Sash warning
Mains fail
Fan stop
Mute
Fire Alarm
Analogue input functions:
Temperature
3.5.2
Outputs
The AFA4000/1 has the following outputs:
3.5.2.1
Relay outputs 1, 2, 3
These are volt-free inputs configurable for:
• Normally closed relays or
• Normally open relays
3.5.2.2
Analogue output 1
0 - 10 Vdc re-transmission of face velocity over 0 - 1.00 m/sec, for example 0.50 m/sec =
5 Vdc.
AFA4000/1 Airflow Monitor / p.20
3.5.3
Communications
3.5.3.1
Com port
This enables connection to a laptop or PC for configuration and for communications to
building management or automation systems (BMS/BAS).
For specific information on Modbus RTU options and settings, see section 8.1.1.3.
For specific information on BACnet options and settings, see section 8.1.1.4.
3.5.4
Power supply
Low voltage DC power supply from the mains power adaptor.
3.5.5
Optional inputs
3.5.5.1
Temperature sensor
Bespoke temperature sensor for connecting into inputs 1, 2 or 3 to give temperature
display with high or low temperature alarms.
3.5.5.2
PIR occupancy sensor
Auxiliary PIR used for close sash alarm based on the fume cupboard occupancy.
3.5.5.3
Mains fail battery unit
Auxiliary plug-in battery unit has a mains fail alarm.
AFA4000/1 Airflow Monitor / p.21
3.6
Functions
3.6.1
Airflow functions
The AFA4000/1 airflow display can be set up using the pushbutton menus to display
airflow in units of m/sec or fpm and can also be set to show plain text Air Safe & Air Fail
only.
The AFA4000/1 has 4 programmable airflow alarms:
Safe airflow
• Airflow reading above warning level (for example, > 0.45
m/sec).
• Green Safe LED on.
Warning airflow
• Airflow reads between warning level and air fail level (for
example, > 0.4 m/sec and < 0.45 m/sec).
• Amber Warning LED on.
Low airflow
• Airflow reads below alarm level for longer than the
warning to low air delay time.
• Air Fail toggles on / off with display.
• Red Alarm LED on (Flashing).
• Audible alarm sounds - can be muted by pressing the
Mute pushbutton.
• Low air relay operates (if configured).
• Reset: when airflow rises 0.02 m/sec above Low air level
for longer than the low air to warning air delay time, the
Low air alarm resets automatically.
High airflow
If configured:
• Airflow reading above high level (for example > 1.50
m/sec)
• High Air toggles on / off with display.
• Red Alarm LED on (Solid).
• High air relay operates (when configured).
AFA4000/1 Airflow Monitor / p.22
3.6.2
Audible alarm mute
When the audible alarm is muted via the Mute button, an icon (
) displays.
The audible alarm can be permanently disabled by selecting Not Enabled in the Cal
Config menu (Setup Monitor > Configure > Cal Config Menu > Audible Alarm).
3.6.3
Input functions
The AFA4000/1 has 3 programmable inputs that can be set to:
• Analogue (0 - 5 Vdc)
• Digital open or
• Digital closed operation
3.6.3.1
Analogue input functions
Function
Description
Temperature
• Any Input set to Analogue - Temperature
• Temperature is displayed alongside airflow velocity in °C or °F
• Low Temp toggles on / off with display if temperature is less
than the Low Temp Alarm point
• Red Alarm LED on (Flashing)
• Audible alarm sounds - can be muted by pressing the Mute
pushbutton
• Low Temp relay operates (when configured)
• High Temp toggles on / off with display if temperature is
higher than the High Temp Alarm point
• Red Alarm LED on (Flashing)
• Audible alarm sounds - can be muted by pressing the Mute
pushbutton
• High Temp relay operates (if configured)
AFA4000/1 Airflow Monitor / p.23
3.6.3.2
Digital input functions
Function
Description
Alarm disable
When input configured as Alarm is disabled:
• Alarm disabled is displayed.
• Red LED on (Flashing).
• Audible alarm muted.
• Mute Icon shown on display.
Night set-back
When input configured as Night Set-back is activated:
• Night Set-back Icon displayed.
• Audible alarm muted.
• Mute Icon shown on display.
• Low Air alarm muted if set to Maintain Low Air.
• Reduce Low Air & Warning alarms active if set to Reduce Low
Air.
External alarm
When input configured as External alarm is activated:
• Red LED on (Flashing) – (when configured).
• External Alarm toggles on / off with display - (when
configured).
• Audible alarm sounds. Can be muted by pressing the Mute
pushbutton.
• External Alarm relay operates (when configured).
Sash High
When input configured as Sash High is activated:
• Amber LED on Solid (Flashing when alarm muted)
• Sash High toggles on / off with display if sash height >
calibrated position
• Audible alarm sounds. Can be muted by pressing the Mute
pushbutton
• Sash High relay operates (when configured)
• Alarm re-activates after time delay when repeat time enabled
High / Low
When input configured as High/Low is activated:
• Display Icon shows ↑ (High) or ↓ (Low)
AFA4000/1 Airflow Monitor / p.24
Function
Description
• High / Low relay operates (when configured)
• This function is designed for two speed fan operation or two
position damper operation switched using a micro switch or
proximity switch activated at a given position on the sash.
Sash Warning
When the input configured as Sash Warning is activated:
• Red LED on.
• Sash Warning - toggles on/off with velocity display.
• Audible alarm sounds (after pre-set time).
• Audible can be muted by pressing the Mute pushbutton. This
silences the alarm (when configured).
• Alarm re-activates after time delay when repeat time enabled
• Reset when the sash is lowered to closed position and the
input is de-activated or an operator is present.
Mains Fail
(Optional extra
feature)
When the input configured as Mains Fail is activated:
• Red LED on
• Mains Fail is displayed
• Audible alarm sounds
• Audible can be muted by pressing the Mute pushbutton. This
silences the alarm (if configured)
Fan Stop
When the input configured as Fan Stop is activated:
• Off is displayed.
• Red LED on (Solid).
• Audible alarm muted.
• When the input configured as Fan Stop is de-activated.
• Start Up Timer is displayed.
• Audible alarm and relays inhibited until Start up time has
elapsed.
Note: Used if the AFA4000 is a Slave unit with a common fan.
Remote master On/Off signal is used
Mute
When the input configured as Mute is activated:
AFA4000/1 Airflow Monitor / p.25
Function
Description
• Red LED on (Solid).
• Audible alarm muted.
• Mute Icon shown on display.
Note: The Mute function disables the pushbutton mute so that
any audible alarm can only be muted using the input, for
example remote key switch.
AFA4000/1 Airflow Monitor / p.26
3.7
Components
3.7.1
Airflow monitor components
• 1 – AFA4000/1 Airflow Monitor
• 1 – Airflow Sensor c/w 2 metre RJ45 Sensor Cable
• 1 – Plug in type low voltage power supply with 5 metre Cable
Figure 2: AFA4000/1 airflow monitor
AFA4000/1 Airflow Monitor / p.27
3.7.2
Auxiliary components
The following auxiliary components are available for the AFA4000/1 airflow monitor:
Component
Function
Sash High Proximity Switch Used for Sash High Alarm
Sash High Micro Switch
Used for Sash High Alarm
Personnel Sensor
Passive Infra-Red sensor - used for Close Sash Alarm
Mains Fail Battery Unit
Used for Mains Fail Alarm
Temperature Sensor
Used for Temperature display and alarms
AFA4000/1 Airflow Monitor / p.28
4. Overview: Auto Sash Controller
The Auto Sash Controller automates the opening and closing of the fume hood sash.
Automation of the sash movement provides an increased level of safety to the controlled
environment. A sash cannot be accidentally left open and can respond to safety critical
inputs from external sources, such as fire alarms and ventilation system failures.
The Auto Sash Controller facilitates cost savings. When linked to ventilation controls, the
automatic closing of an unattended fume hood allows extractor fan speeds and
electricity consumption to be reduced.
4.1
Features
The Auto Sash Controller has the following features:
• Auto-Close
The sash automatically closes when the fume hood is
unattended.
• Auto-Open
The sash automatically opens when the user returns.
• Sash Lock
The sash is programmed to lock at a designated
position.
• Manual Sash
Open/Close
To open/close the sash manually.
• Keypad Pushbutton
Open/Close
To open/close the sash using the Keypad Pushbuttons.
• Footswitch
To open / close the sash using an optional, third party
supplied, footswitch.
• Tilt Switch
To disable the Auto Sash Control Unit using a, third
party supplied, tilt switch.
• Building Management
System (BMS) Inputs
The sash automatically closes when a signal is received
from the BMS.
• Building Management
System (BMS) Outputs
The Auto Sash Controller sends signals to the BMS
regarding the current status of the fume hood.
AFA4000/1 Airflow Monitor / p.29
4.2
Auto Sash control options
To open / close the fume hood sash the Auto Sash Controller is fitted with:
• A Tiptronic open/close control via the Sash Position Sensor.
• A Sash Low micro-switch
• A Sash Low proximity switch
4.2.1
Auto Sash Controller with Tiptronic open / close control
Tiptronic Open / Close control is the default control setup. It functions via a Sash
Position Sensor (sprung potentiometer) with a draw wire (1.2m long) attached to either
the sash or the counterweight. The Auto Sash Controller can then be calibrated to give
three methods of operation:
Method A - Close only
• The bottom position is calibrated.
• The sash automatically drives to the bottom position,
when the user is not present.
• Press the ↓ pushbutton to drive the sash to the
bottom position, when the user is present.
• The ↑ pushbutton is disabled.
Method B - 2 Positions
• The bottom and mid-positions are calibrated.
• The sash automatically drives to the bottom position,
when the user is not present.
• Press the ↓ pushbutton to drive the sash to the
bottom position, when the user is present.
• Press the ↑ pushbutton to drive the sash to the mid-
position, when the user is present.
• When the Tiptronic feature is enabled, manually tip
the sash upwards and the sash will drive to the mid-
position.
• Manually tip the sash downwards and the sash will
drive to the bottom position.
AFA4000/1 Airflow Monitor / p.30
• When the Auto Open feature is enabled, the sash
automatically drives to the mid-position when the
user is present.
• There is a time delay following user detection.
• When the Auto Lock feature is enabled and a Sash
Lock position has been calibrated, the Auto Close
feature is disabled and the sash locks in its current
position when the sash is raised to the Sash Lock
position or higher. This facilitates loading the Fume
Cupboard.
Method C – 3 Positions
• The bottom, mid and top positions are calibrated.
• The sash automatically drives to the bottom position,
when the user is not present.
• Press the Down pushbutton to drive the sash to the
bottom position, when the user is present.
• Press the Up pushbutton to drive the sash to the
next calibrated position, when the user is present.
• When the Tiptronic feature is enabled, manually tip
the sash upwards and the sash will drive to the next
calibrated position.
• Manually tip the sash downwards and the sash will
drive to the bottom position.
• When the Auto Open feature is enabled, the user is
present and the sash is lower than the mid position,
the sash automatically drives to the mid-position.
• There is a time delay following user detection.
• When the Auto Lock feature is enabled and a Sash
Lock position has been calibrated, the Auto Close
feature is disabled and the sash locks in its current
position when the sash is raised to the Sash Lock
position or higher. This facilitates loading the Fume
Cupboard.
AFA4000/1 Airflow Monitor / p.31
4.2.2
Auto Sash Controller with Sash Low micro-switch
A Sash Low Micro-Switch can be specified for fitting by TEL or by the fume hood
manufacturer. The switch can be arranged to be open / closed when the sash is closed,
depending on fitting. The wiring arrangement must be modified for installation and the
operational requirements of the fume hood.
The following functionality is available:
• The sash automatically drives to the bottom position, when the user is not present.
• Press the Down pushbutton to drive the sash to the bottom position, when the user is
present.
• The Up pushbutton is disabled.
4.2.3
Auto Sash Controller with Sash Low proximity switch
A Sash Low Proximity Switch can be specified for fitting by TEL or the fume hood
manufacturer. The switch can be arranged to be open/closed when the sash is closed,
depending on fitting. The wiring arrangement must be modified for installation and the
operational requirements of the fume hood.
The following functionality is available:
• The sash automatically drives to the bottom position, when the user is not present.
• Press the Down pushbutton to drive the sash to the bottom position, when the user is
present.
• The Up pushbutton is disabled.
AFA4000/1 Airflow Monitor / p.32
4.3
Control system components
The Auto Sash Control System comprises of a control unit and additional components.
Figure 3: Typical location of main components
Note:
This example installation is fitted with a pulley and wire sash drive.
AFA4000/1 Airflow Monitor / p.33
4.3.1
Auto sash control unit
The Auto Sash Control Unit processes inputs from sensors, switches associated with the
sash control function and output signals relevant to the motor drive(s) and the alarm
panel.
Figure 4: Auto Sash Control Unit - single and dual controller enclosure
AFA4000/1 Airflow Monitor / p.34
4.3.2
Personnel sensor
Figure 5: Detection area clear
Figure 6: Obstruction present
The Personnel Sensor is a Passive Infra-Red (PIR) occupancy detector which detects the
presence / absence of the fume hood user.
Note:
If the user is detected, the sash will not automatically close. The Sash can be
opened / closed manually or using the Up / Down pushbuttons.
On power up, the Personnel Sensor ‘learns’ the reflection characteristics of the
environment within its field of view and stores this information as reference background
data. The sensor then 're-learns' the background every three minutes. A stationary
object, such as a stool, that is left in the field becomes part of the background and is
'ignored' by the sensor. When the user is not present at the fume hood, the sash will
close automatically. This follows a pre-set delay time, if the Sash Light Curtain is not
broken and the sash is open.
Green LED
Red LED
AFA4000/1 Airflow Monitor / p.35
4.3.3
Keypad/alarm panel
Figure 7: UK / EU keypad
Figure 8: US keypad
The Keypad/Alarm Panel provides a visual indication of the status of the Auto Sash, this
includes fault conditions which are displayed on a Liquid Crystal Display (LCD).
4.3.4
Sash light curtain
The Sash Light Curtain is a safety device designed to detect whether the opening below
the sash is clear of obstructions before the sash is automatically closed.
• When an obstruction is detected, the sash drive is disabled and the alarm is activated.
• The Sash Light Curtain comprises an infra-red transmitter mounted on one side of the
fume hood and an infra-red receiver mounted on the opposite side of the fume
hood's inner chamber.
• The infra-red transmitter has multiple, closely spaced, Light Emitting Diodes (LEDs) to
facilitate the detection of small objects.
AFA4000/1 Airflow Monitor / p.36
• Objects are detected when infra-red light from the transmitter is not detected by the
receiver due to an obstruction in its path.
Figure 9: Sash Light Curtain Transmitter and Receiver
A separate Sash Light Curtain Control Box includes a sensitivity potentiometer that
allows the beam sensitivity adjusted to sense glass or other objects that are partially
transparent to infra-red light.
Figure 10: Sash Light Curtain Control Box
4.3.5
Under Sash sensor
The Under Sash Sensor is an alternative to the Sash Light Curtain which detects whether
the opening below the sash is clear from obstructions before the sash is closed.
A sensor on one side of the fume hood detects the reflection of a beam of light from a
self-adhesive strip of retro-reflective tape mounted on the opposite inside wall of the
fume hood.
AFA4000/1 Airflow Monitor / p.37
The sensor incorporates a potentiometer, it allows the beam sensitivity to be adjusted to
sense glass objects.
4.3.6
Sash Low switch
When a Tiptronic control is not in use, a Sash Low switch is used to signal to the Auto
Sash Control Unit that the fume hood sash is closed.
The switch can be either a mechanical roller lever type micro-switch or an inductive
proximity switch.
Roller Lever Micro-Switch
Inductive Proximity Switch
Figure 11: Typical Sash Low switches
4.3.7
Sash position sensor
The Sash Position Sensor comprises of a flexible cable, a spring-loaded spool and a
potentiometer. It is used, in conjunction with the Tiptronic control option, to detect and
measure vertical sash movement and the position of the fume hood sash.
During the initial on-screen calibration set up the sash controller detects the sensor
direction of travel and then provides position information of the sash to the Auto Sash
Control Unit, to enable calibration of the sash open / closed positions.
AFA4000/1 Airflow Monitor / p.38
Sash Position Sensor
Sensor Cable
Sensor Potentiometer
Figure 12: Sash Position Sensor
4.3.8
Tilt switch
The Tilt Switch is an optional, third party supplied component, which is fitted to the fume
hood top panel. It operates to isolate the Auto Sash controls, when the top panel is open
to facilitate the servicing of the fume hood.
Figure 13: Typical third-party tilt switch
AFA4000/1 Airflow Monitor / p.39
4.3.9
Foot switch
The Foot Switch is an optional, third party supplied component, which can be positioned
on the floor in front of the fume hood and used to open the sash.
Figure 14: Typical third party foot switch
4.3.10 Sash drive unit
The Sash Drive Unit is a geared electric motor and clutch assembly used to drive the
sash.
Three types of drive are available, none of which preclude manual opening / closing:
• Rack and pinion drive
• Chain and sprocket drive
• Pulley and wire drive
Chain and sprocket drive
Rack and pinion drive
Pulley and wire drive
Figure 15: Example drives
AFA4000/1 Airflow Monitor / p.40
5. Installation: Airflow Monitor
This section outlines the installation of the various components of the airflow monitor
system. The size and format of individual fume cupboards varies considerably, so specific
instructions are not possible, however the principles outlined are valid for all cases.
5.1
Location
The AFA4000/1 monitor can be mounted on either side of the fume cupboard. However,
when deciding on the location for the airflow monitor and sensor, consider the cable
lengths required:
• The 5 metre low-voltage cable must reach from the terminals on the back of the
AFA4000/1 airflow monitor to the connections on top of the fume cupboard.
• The standard 2 metre ‘telephone style’ sensor cable must reach from the back of the
AFA4000/1 airflow monitor to the airflow sensor. When possible ensure that the
airflow sensor is mounted on the same side of the fume cupboard as the AFA4000/1.
Note:
Longer sensor cables are available on request.
AFA4000/1 Airflow Monitor / p.41
5.1.1
SM7 airflow sensor installation notes
It is very important to position the SM7 airflow sensor in the correct position to give a
long-term stable reading of the face velocity.
Figure 16: Airflow sensor installation diagram
Please read the notes below and if in doubt contact us for further advice.
1. The SM7 sensor must be positioned where it can sense the room pressure of the
laboratory. The back-connection spigot of the sensor is designed to accept a tube
with outside diameter of 25 mm, which should be connected to the inner chamber of
the fume cupboard. This tube and fittings are known as the ‘vent kit’.
The ideal position for the end of the 25 mm diameter tube, for most fume cupboards,
is 100 mm back from the sash glass and 100 mm higher than the normal sash
opening height through the inner side wall.
2. If possible, mount the sensor on the front of the fume cupboard and use a short
length of tube. Tube lengths of more than 1 m or smaller diameter will restrict the
AFA4000/1 Airflow Monitor / p.42
airflow through the sensor. This will lead to too much sensitivity being required to
calibrate the unit which can lead to instability of the readings or incorrect readings at
low velocities.
3. For fume cupboards with a single-skin side-wall, or double-skin, with a small gap
between them, it may not be possible to achieve the ideal sensing position using a
flexible tube.
With a single-skin side-wall it is possible to fix the sensor on the outside of the fume
cupboard and connect directly to the inner chamber in the ideal position. This
method can only be used for up to two fume cupboards when they are positioned
side by side, using the two outer walls.
An alternative method is to mount the sensor on the front of the fume cupboard and
connect it using a short flexible tube to a rigid wall tube attached to the inner side-
wall. Position the open end of this rigid wall tube in the ‘ideal position’ for example,
100 mm back from the sash, and 100 mm higher than the normal sash opening.
Fume cupboards with a large internal height can present a difficulty because the tube
length to reach the ideal position may be longer than 1m. Where possible, use a tube
of 1 metre or less, which may result in a sensing position higher than the ideal.
When fitting a sensor to a ’narrow wall’ fume cupboard for the first time, it may be
necessary to try various combinations of rigid and flexible tube to find the best
combination and position.
4. The sensor should not be mounted in a position where it is subject to draughts from
the laboratory air input or ventilation system.
5.2
Installation
5.2.1
AFA4000/1
1. Use the cut-out details provided with the unit (Figure 17) to cut-out and drill the
holes in the fume cupboard for the AFA4000/1 mounting bracket.
Make sure that the AFA4000 fits into the cut-out before proceeding.
AFA4000/1 Airflow Monitor / p.43
Figure 17: AFA4000 cut-out template
Note: Print to A4. Do not scale or print to fit page.
Note: Separate mounting boxes and adaptor plates are available for retro-fitting to
older fume cupboards.
AFA4000/1 Airflow Monitor / p.44
Figure 18: AFA4000 installation
2. Use 4 fixing screws to secure the mounting bracket to the front panel of the fume
cupboard (A in Figure 18).
Make sure the 4 fixing screws are fully recessed into the counter-sunk fixing holes (B
in Figure 18).
3. Place the AFA4000 onto the top of the mounting bracket and click it into place (C in
Figure 18).
4. Push the bottom of the AFA4000 onto the mounting bracket (D in Figure 18).
5. Secure the AFA4000 to the mounting bracket using the fixing screw in the bottom
face of the monitor (E in Figure 18).
A
B
C
D
E
AFA4000/1 Airflow Monitor / p.45
5.2.2
SM7 airflow sensor
1. Use the cut-out details provided with the sensor (Figure 19) to cut-out and drill the
holes in the fume cupboard for the SM7 sensor.
Make sure that the sensor fits into the cut-out before proceeding.
Figure 19: SM7 sensor cut-out template
Note: Print to A4. Do not scale or print to fit page.
5.3
Connections
1. Connect the power supply cable into the back of the AFA4000/1 (H in Figure 20) and
to the connection on top of the fume cupboard.
2. Connect the ‘telephone style’ airflow sensor plug-in cable to the sensor and the back
of the AFA4000/1 unit (J in Figure 20).
AFA4000/1 Airflow Monitor / p.46
Figure 20: AFA4000/1 monitor connections
A
0-10 V velocity re-transmission
output, 0-10 Vdc over 0-1.00 m/sec /
0-200 fpm
G
Relay output 3 (used if relay interface
is not fitted)
B
Input 1 – digital or analogue
H
15 Vdc power supply (used if relay
interface is not fitted)
C
Input 2 – digital or analogue
I
Auto sash RJ45 connection (only used
on AFA4000/1 – as model)
D
Input 3 – digital or analogue
J
Airflow sensor RJ12 connection
E
Relay output 1 (used if relay interface
is not fitted)
K
14-way ribbon cable connection to
relay interface
F
Relay output 2 (used if relay interface
is not fitted)
L
RS485 comms port connection
A
B
C
D
E
F
G
H
I
J
K
L
AFA4000/1 Airflow Monitor / p.51
5.5
Start up
5.5.1
AFA4000/1
The AFA4000/1 must be field-calibrated once the room air supply and exhaust is
proportionally balanced.
When the unit is powered up, the following sequence of events occurs:
The alarm performs a self-test of its functions and audible alarm which takes
approximately 3 seconds. At the end of the delay, the unit will do one of two things:
When the Airflow monitor has been calibrated, the unit enters normal operating mode.
The monitor displays airflow velocity, LEDs and the audible alarm is enabled.
If the Airflow monitor has not been calibrated, the unit displays the message Requires set
up, press Mute to continue.
Follow the instructions in the Calibration section (section 7.1).
5.5.2
AFA4000/1/AS
The AFA4000/1/AS must be field-calibrated once the room air supply and exhaust is
proportionally balanced.
When the unit is powered up, the following sequence of events occurs:
1. The alarm performs a self-test of its functions and audible alarm which takes
approximately 3 seconds.
2. At the end of the delay, the unit will do one of five things:
• When the Airflow monitor has been calibrated, the unit enters normal operating
mode. The monitor displays airflow velocity, LEDs and the audible alarm is enabled.
• When the Airflow monitor has not been calibrated, the unit displays the message
Requires set up, press Mute to continue.
• When the Auto Sash Controller has been calibrated, the unit enters normal
operating mode. The Controller displays status and alarms in the status window.
• When the Auto Sash Controller is connected and has not been calibrated, the
status window displays the message Auto Sash Not Configured.
AFA4000/1 Airflow Monitor / p.52
• When the Auto Sash Controller is disconnected, the Auto sash controller status and
menus are not displayed.
AFA4000/1 Airflow Monitor / p.53
6. Installation: Auto Sash Controller
As the size and format of individual fume hoods vary, the installation instructions that
follow are general. The principles outlined should remain valid in all cases.
Note:
Before starting installation, ensure that the sash is free to move in its runners.
Use silicon spray on the sash runners or adjust the sash runner's alignment when
the sash does not move freely.
6.1
Auto Sash Control Unit
6.1.1
Mounting the unit
The Auto Sash Control Unit can be mounted as follows:
• Screwed flat to the outside of the fume hood roof - horizontal mounting
Note: For horizontal mounting, the control box should always be mounted with the
lid / front panel facing upwards.
• Screwed to the outside wall of fume hood wall - vertical mounting
Figure 25: Auto Sash Control Unit Lid/Front
Panel
Figure 26: Rear of Auto Sash Control Unit
AFA4000/1 Airflow Monitor / p.54
Figure 27: Auto Sash Control Unit Connectors, Single and Dual Fume Hoods
Once the Auto Sash Control Unit has been positioned, mark its position using the fixing
holes in the rear of the housing as a template. Then, attach the Auto Sash Control Unit
using screws.
6.1.2
Wiring in the Auto Sash Control Unit
For fume hoods with an inner and outer sidewall, route all cables through the sidewall
and into the service void.
For fume hoods with a single wall, route all cables up the outside of the sidewall, using
suitable trunking or fixing pads and ties.
When wiring in the Auto Sash Control Unit, observe the following:
• Keep cables clear of moving parts.
• Use grommets wherever cables pass through sidewalls to prevent abrasion of the
cable insulation and / or sidewall. Where a cable is supplied with a DIN connector, the
grommet should have a minimum internal diameter of 17mm (0.67").
• Do not overstress cables by bending, for example when passing at 90° through a
sidewall.
AFA4000/1 Airflow Monitor / p.55
Figure 28: Sash Control System – general layout, wiring schematic and illustration (inset)
AFA4000/1 Airflow Monitor / p.56
Cable Specifications
Mains cable
3-core, 0.7mm², 2m long
Light Curtain cable
4-core, 0.5mm², 2m long
Keypad cable
RJ45, 3 m long
Sash Position Sensor cable
4-core, 0.2mm², 2.5m long
Personnel Sensor cable
5-core, 0.2mm², 3m long
Motor cable
2-core, 0.5mm², 2m long
6.2
Auto Sash keypad / alarm panel
Alternative Auto Sash Keypad / Alarm Panels can be supplied for US and EU markets.
Both types should be mounted on the front of the fume hood, ensuring there is
adequate clearance behind for the cables connecting the keypad to the Auto Sash
Control Unit.
6.2.1
Fitting the Auto Sash keypad / alarm panel
1. Fit the Auto Sash Keypad to the fume hood using either a single gang socket for UK
applications (Figure 29) or using a cut-out to the (Version 2) dimensions (Figure 30).
Figure 29: Version 1 - UK Single Gang Keypad Fits in Standard Single Gang Socket
AFA4000/1 Airflow Monitor / p.57
Figure 30: Version 2 US Single Gang Keypad – (requires cut out in front panel)
2. Connect the Keypad / Alarm Panel to the Auto Sash Control Unit. See the wiring
schematic in section 6.10 and Figure 31.
Figure 31: Keypad/Alarm Panel Rear View
6.3
Auto Sash motor drive unit
A Motor Drive Unit is used to automatically raise and lower the fume hood sash.
Three types of drives are available:
• Rack and pinion motor drive.
• Chain and sprocket motor drive.
• Wire and pulley motor drive.
The optimal choice of sash drive type depends on whether:
AFA4000/1 Airflow Monitor / p.58
• The installation is part of a new build or retro-fit.
• The size and configuration of the fume hood.
• The fume hoods location.
None of these options preclude manual operation of the sash.
6.3.1
Rack and pinion motor drive
The Rack and Pinion Motor Drive uses a rack fitted directly onto the sash frame and a
motor drive assembly fitted to a framework above the fume hood.
This drive system is ideal for retro-fit installations where there is no access to the sash
wires or counterweights and / or where the sash is fitted to a metal framework, for
example combination type sash fume hoods.
6.3.1.1
Installing the rack and pinion drive
1. Raise the sash to the fully open position.
2. Offer up the motor assembly and rack to ensure that there is a clearance of at least
1.2 m (47.24") above the top of the sash.
Note: The rack is supplied in 1m (39.37") lengths
3. The motor drive assembly should be fitted to a rigid framework and positioned
centrally above the sash. To ensure that the T-bar fixing clears the bottom of the
motor assembly, there should be at least 50 mm between the top of the sash and the
bottom of the motor drive assembly, when the sash is fully opened.
AFA4000/1 Airflow Monitor / p.59
Side view
Front view
Figure 32: Rack and pinion sash drive motor assembly
Note: Ensure that the rack is long enough to protrude through the top of the motor
assembly, when the sash is fully closed.
4. When the motor drive assembly is fixed to the fume hood the T-bar should be fixed
to the rack and offered up with the sash frame.
5. Ensure that the rack is free to move over the full sash opening before marking up and
fixing the T-bar to the sash frame.
Note: The rack must to run vertically at 90º to the horizontal axis of the sash frame.
AFA4000/1 Airflow Monitor / p.60
Figure 33: Front view of rack and pinion sash drive
6. When fitting a sash low switch instead of a sash position sensor, the rack will need to
be cut to the correct length to enable the switch to operate.
Close the sash fully and mark the rack just below the position of the proximity low
switch at the top of the motor assembly to ensure that the rack is clear of the switch
when the sash is closed.
7. Connect the Sash Drive to the Auto Sash Control Unit using the cables provided (with
red and black connectors). See the wiring schematic in section 6.10 and Figure 31.
Note: It is important to observe the correct polarity when making electrical
connections.
8. Connect the clutch assembly to the Auto Sash Control Unit using the cables provided
(with blue and yellow connectors).
AFA4000/1 Airflow Monitor / p.61
Figure 34: Side views of a rack and pinion drive typical installation
6.3.2
Chain and sprocket drives
Two types of chain and sprocket drives are available:
• Shaft Drive
AFA4000/1 Airflow Monitor / p.62
• Counterweight Drive
6.3.2.1
Chain and sprocket shaft drives
In the Shaft Drive arrangement, the motor drive utilises a taper bore sprocket and a roller
chain to drive a shaft running between the counterweight pulleys.
The motor assembly is supplied fitted to an adjustable bracket so that the correct chain
tension can be set.
The Shaft Drive arrangement can accommodate one or two sash wires.
Figure 35: Typical chain and sprocket shaft drive installation
6.3.2.2
Installing the chain and sprocket shaft drive
1. Fit the taper bore sprocket to the shaft and ensure that the sprocket is securely fitted.
Where possible positioned the motor drive and sprocket centrally on the shaft. The
motor assembly should be fitted to a rigid framework to support the motor.
2. Loosely fit the chain before fixing the motor plate to ensure the chain is correctly
tensioned and aligned.
AFA4000/1 Airflow Monitor / p.63
6.3.2.3
Chain and sprocket counterweight drive
In the Counterweight Drive arrangement, the chain and sprocket drive system is used to
directly lift the sash counterweight. This is achieved by fitting a roller chain to the
counterweight and utilising a separate counterweight and freewheeling sprocket to
maintain the correct chain tension.
Figure 36: Typical chain and sprocket counterweight drive installation
6.3.3
Installing the chain and sprocket counterweight drive
1. When the motor, chain and freewheeling sprocket has been fitted, manually open /
close the sash, to ensure that the motor drive is free running and that the chain is not
too slack or too tight.
2. Fit the motor so that the drive sprocket is in line with the counterweight pulley.
3. Fit one end of the chain to the counterweight or sash wire near to the counterweight.
4. Fit the freewheeling sprocket in line with the drive sprocket.
5. Loosely fit the chain over the freewheeling pulley and determine the best position for
the chain counterweight.
AFA4000/1 Airflow Monitor / p.64
6. Fit the chain counterweight, using a counterweight guide if necessary.
7. When the motor, chain and counterweight are fitted manually open / close the sash
to ensure the motor drive, chain and counterweight are free running.
6.3.4
Wire and pulley drive
With the Wire and Pulley Drive, the pulley motor drives the sash wire and can accept
both single and dual sash wire systems. The motor assembly is fitted in line with the sash
wires and uses two guide pulleys and a drive pulley.
The pulley drive is ideal for new fume hoods but can be retro-fitted if a longer
replacement sash wire can be fitted.
Figure 37: Typical wire and pulley drive installation
6.3.4.1
Installing the wire and pulley drive
The motor assembly can be fitted horizontally or vertically on either side of the fume
hood, as follows:
1. Fit the motor assembly so that the guide pulleys are in line with the sash wire.
2. Run the sash wire over the first guide pulley, around the drive pulley and then over
the final guide pulley.
When the motor drive has been installed, open / close the sash to ensure that the sash
can travel freely.
AFA4000/1 Airflow Monitor / p.65
6.4
Sash light curtain
The Sash Light Curtain consists of three main components and associated wiring:
• Sash Light Curtain Control Box
• LED Transmitter
• LED Receiver
The Sash Light Curtain Control Box is connected to the main Auto Sash Control Unit.
6.4.1
Installing the sash light curtain control box
1. Fix the Sash Light Curtain Control Box onto the top of the fume hood, using the tabs
attached to the control box and the screws provided. Ensure that there is access to
the terminals to connect the wiring.
Figure 38: Sash Light Curtain Control Box
2. Use the pre-assembled cables and connectors supplied to connect the Light Curtain
Control Box to the Auto Sash Control Unit. See the wiring schematic in section 6.10
and Figure 31.
AFA4000/1 Airflow Monitor / p.66
6.4.2
Installing the sash light curtain transmitter and receiver
1. Find a suitable position near to the back of the sash track. Ensure that the sash handle
does not protrude into the light beam and that there are no other obstructions on
either side wall.
2. Carefully remove the front plastic covers from the right-hand side of the light curtain
receiver and the left-hand side of the light curtain transmitter cases.
Figure 39: Light curtain receiver plastic cover
3. Position the light curtain transmitter and receiver on opposite sidewalls. Ensure that
the back of the sash handle does not penetrate the light beam path.
The sash light curtain transmitter and receiver should be fitted so that their cable
entry / exit points are at the top.
Ensure that the light beam path is free from other obstructions.
AFA4000/1 Airflow Monitor / p.67
Figure 40: Mounting transmitter / receiver
6.4.3
Aligning the light curtain transmitter and receiver
Mount the light curtain receiver and transmitter at the same height on opposing sides of
the fume hood opening. A small error in alignment is acceptable. The allowable error is
dependent on the spacing between the transmitter and receiver.
Figure 41 illustrates the maximum allowable transmitter to receiver misalignment /
displacement for a 0 to 3 m (0 to 118.11") range of fume hood widths. For a typical fume
hood width of 1 to 2 m (39.37 to 78.740"), the maximum allowable misalignment /
displacement ranges from approximately 20 to 40 mm (0.79 to 1.58").
AFA4000/1 Airflow Monitor / p.68
Figure 41: Maximum allowable displacement for a range of fume hood widths
1. The area protected by the light beams should cover the open area under the sash,
typically 500 mm (20") from its base.
The light curtain transmitter and receiver can be positioned above the bottom of the
sash in the fume hood if they are free from obstructions.
Where possible, make sure the light path is aligned flush to the bottom of the sash
base.
Note: There is a 123 mm (4.84") gap from the top of the light curtain transmitter to
the first LED down and a 22 mm (0.87") gap from the bottom of the transmitter
to the first LED up.
AFA4000/1 Airflow Monitor / p.69
Figure 42: Light curtain vertical range
2. When the light curtain transmitter and receiver have been positioned, mark the
sidewall using the fixing holes in the transmitter and receiver as templates.
3. Drill the holes for the fixing screws. Use the 2 mm (0.08") drill bit provided and fix the
transmitter and receiver.
Figure 43: Transmitter / Receiver fixing holes
AFA4000/1 Airflow Monitor / p.70
4. On a fume hood with an inner and outer sidewall the light curtain transmitter and
receiver cables run through the sidewall into the service void and up to the Sash Light
Curtain Control Box on top of the fume hood.
On a fume hood with a single wall drill the hole in the top soffit panel and run the
cables up the sidewalls using suitable trunking or fixing pads and ties.
Drill a 17mm (0.67") hole to clear the cable DIN connector.
When using a grommet drill a suitable hole in the sidewall to accept the grommet.
Figure 44: Light curtain cable entry
Note: The hole’s centre should be a minimum of 35 mm (1.38") from the top of the
light beam path, so that the cable is not stressed when it bends 90° through the
inner sidewall.
5. Pass the cables through the sidewall and connect to the Sash Light Curtain Control
Box on the top of the fume hood.
6. Connect the light curtain transmitter and receiver to the Sash Light Curtain Control
Box using the plug-in RJ45 cables provided.
AFA4000/1 Airflow Monitor / p.71
Figure 45: Light Curtain Vertical Range Safety Curtain Transmitter and Receiver Wiring
Schematic
7. Use the pre-assembled cables provided to connect the output from the Safety Curtain
Control Box to the Auto Sash Control Unit. See the wiring schematic in section 6.10
and Figure 31.
8. Fix the light curtain transmitter and receiver to the sidewall using their fixing holes
and re-attach the plastic covers.
AFA4000/1 Airflow Monitor / p.72
6.5
Installing and aligning the Under Sash sensor
The Under Sash Sensor consists of two components:
• Sash Sensor
• Retro-reflective Tape Strip
1. Fit the sensor so that its lens is clear of obstructions and below the bottom of the
sash handle.
Figure 46: Under sash sensor mounting
Figure 47: Under sash sensor dimensions
Use a suitable bracket so that the sensor position can be adjusted to the required
position.
2. Fix the 5 m cable, which is attached to the sensor, to the sash wire using cable ties.
Attach the 5 m cable to the sash counterweight to keep the cable taught as the sash
opens or closes.
Encase the cable from the sash counterweight to the sash controller box can in plastic
sheathing to guide the cable slack and stop it from snagging.
AFA4000/1 Airflow Monitor / p.73
When it is not possible to fix the cable to the sash wire, use small diameter trunking
to attach the cable to the sash.
Use a roller bearing or pulley where the cable runs over a metal surface.
Figure 48: Typical under sash sensor retro-installation
3. Fix the retro-reflective tape strip to the inside of the fume hood, on the opposite side
to the sensor.
Position the centre of the retro-reflective tape strip in line with the centre of the
sensor. A small offset between the sensor and retro-reflective tape strip is acceptable.
The maximum offset values are shown in Figure 49.
AFA4000/1 Airflow Monitor / p.74
Figure 49: Maximum under sash sensor offset
For example, on a 1000 mm wide Fume Hood the sensor can detect the retro-
reflective strip up to ± 20 mm either side of the Under Sash Sensor's centreline.
4. Connect to the Auto Sash Control Unit. See the wiring schematic in section 6.10 and
Figure 31.
6.6
Tilt switch
Tilt Switches are provided in a variety of forms, the installation instructions that follow
are therefore general.
• Ensure that the Tilt Switch is installed in the correct orientation – ‘open’ when the
front panel is closed and ‘closed’ when the front panel is open.
• Connect the Tilt Switch to the Auto Sash Control Unit. See the wiring schematic in
section 6.10 and Figure 31.
• Link out / connect the relevant terminals on the Auto Sash Control Unit when not
using a Tilt Switch. See the wiring schematic in section 6.10 and Figure 31.
AFA4000/1 Airflow Monitor / p.75
Figure 50: Tilt switch wiring schematic
6.7
Sash Low switch
Two types of Sash Low switch can be used:
• Micro-Switch
• Proximity Switch
Switches signal the sash as being open when the switch is open or closed, depending on
the installation requirements.
6.7.1
Installing the Sash Low mcro-switch
The Sash Low Micro-Switch is a change-over type switch. When the sash is closed the
switch makes or breaks the circuit depending on which terminals are used on the switch
for each type of activation. Typically, the switch would be fitted so that the lever engages
(pushes in) when the sash is open.
Connect the Sash Low Micro-Switch to the Auto Sash Control Unit (Figure 51 See the
wiring schematic in section 6.10 and Figure 31.
If the switch is installed so that the lever pushes in when the sash is closed, connect the
wiring to the Auto Sash Control Unit to terminals C and NO on the switch.
If the switch is installed so that the lever pushes in when the sash is open, connect the
wiring to the Auto Sash Control Unit to terminals C and NC on the switch.
AFA4000/1 Airflow Monitor / p.76
Figure 51: Sash low micro-switch wiring schematic
6.7.2
Installing the Sash Low proximity switch
The Sash Low Proximity Switch is an inductive switch that operates when it comes within
range of a metal plate attached to the sash, i.e. when the sash is closed.
The type of switch selected makes or breaks the circuit when the sash is closed. Typically,
the switch would be fitted so that the switch is closed when the sash is open.
Connect the Sash Low Proximity Switch to the Auto Sash Control Unit (Figure 52). See
the wiring schematic in section 6.10 and Figure 31.
Figure 52: Sash low proximity switch wiring schedule
AFA4000/1 Airflow Monitor / p.77
Figure 53: Sash Low Proximity Switch Dimensions
Note:
When the Auto Sash control is being retro-fitted, fix the bracket to allow for
over-travel of the sash.
6.8
Sash position sensor
The sash position sensor is fitted with a mounting bracket which can be fitted in four
different orientations to enable the sensor draw wire to be correctly aligned.
1. Offer up the sensor to determine the best fixing position on the fume hood
framework.
Where possible, ensure the sensor draw wire runs straight out of the sensor housing,
in line with the cable housing extension. Where the wire runs at an angle to the
sensor, ensure the angle is as small as possible to prevent the wire from rubbing on
any part of the sensor body, including the cable housing extension.
AFA4000/1 Airflow Monitor / p.78
Figure 54: Sash position sensor
2. Mark the fixing holes on the framework and attach the sensor.
The fixing bracket has two 0.37 mm (0.015") diameter fixing holes.
Note: Ensure that the sensor is in a position that does not over extend the draw wire.
The sensor draw wire full stroke is 1270 mm (50") maximum.
3. Attach the draw wire to either:
• The sash using a suitably sized screw, the draw wire has a 9.53 mm (0.38") diameter
fixing ring with a 4.85 mm (0.18") diameter fixing hole in its centre.
• The sash cable using a cable tie, ensuring that the fixing ring does not run over the
pulleys.
• The counterweight using a cable (Figure 55).
AFA4000/1 Airflow Monitor / p.79
Figure 55: Sash position sensor draw wire attached by cable tie
When the sensor is fitted, manually open / close the sash to ensure that the sensor is
installed correctly.
6.9
Personnel sensor
The Personnel Sensor is supplied as a single transmitter / receiver unit.
6.9.1
Installing the personnel sensor
1. Use the mounting template illustrated in Figure 56 and the fixing screws provided to
mount the Personnel Sensor. Mount the sensor centrally to the front of the fume
hood and above the fume hood sash, use the fixing holes in the back of the sensor.
See the table in Figure 56 for the correct mounting hole dimensions and locations.
AFA4000/1 Airflow Monitor / p.80
A
Cable Entry
10 x 20 mm (5 mm diameter cable provided)
B
Fixing Hole
5 mm diameter maximum
C
Spare Hole
10 x 20 mm
D
Spare Hole
10 x 20 mm
E
Fixing Hole
5 mm diameter maximum
F
Spare Fixing Hole
5 mm diameter maximum
Figure 56: Personnel sensor mounting details
The Personnel Sensor should be positioned so that it points straight down, with the
horizontal axis of the sensor parallel to the horizontal axis of the fume hood (Figure 57).
Note:
Do not mount the sensor higher than 250 0mm (98”) from the laboratory floor.
AFA4000/1 Airflow Monitor / p.81
Figure 57: Vertical positioning of Personnel Sensor
2. Fit the connection cable to the sensor cable inside the sensor housing.
3. Plug the 4-way terminal block into the Auto Sash Control Unit (Figure 58). See the
wiring schematic in section 6.10 and Figure 31.
Figure 58: Personnel sensor 4-way terminal block
AFA4000/1 Airflow Monitor / p.82
6.9.2
Adjusting the personnel sensor beam width
Adjust the width of the Personnel Sensor’s beam to suit different fume hood widths.
To adjust the beam width:
1. Access the two potentiometers under the sensor's front cover (Figure 59).
Figure 59: Personnel sensor beam width adjustment potentiometers
2. Adjust the width of the Personnel Sensor beam using the two potentiometers. This is
to achieve the best field coverage in front of the fume hood for the hood's width
(Figure 60).
AFA4000/1 Airflow Monitor / p.83
Figure 60: Adjusting personnel sensor beam width
6.9.3
Adjusting the beam depth
Adjust the sensor by tilting the bottom of the sensor towards or away from the front wall
of the fume hood (Figure 61). This is to ensure that the sensors beam is at optimum
angle from the fume hood, to achieve the best field of coverage for the height of the
fume hood.
The body of the Personnel Sensor can be tilted from 0° to 5° in three increments to
achieve this.
AFA4000/1 Airflow Monitor / p.84
Figure 61: Adjusting the personnel sensor beam angle away from the fume hood
AFA4000/1 Airflow Monitor / p.85
6.9.4
Personnel sensor dip switch settings
The following Personnel Sensor functions and parameters are controlled by Dip switches,
located within the Sash Light Curtain Control Box (Figure 62) (See section 7.4.7).
Figure 62: Personnel Sensor Dip Switch Settings
Switches
Setting
Description
1 & 2
Presence timer
Select the time delay before the Personnel Sensor
‘re-learns’ the background when a stationary
object is detected.
2, 15, 60 or 180 second delays are selected by
setting associated switches 1 and 2 to
combinations of ON/OFF (Figure 62).
Note: The timer will reset if any movement is
detected during the time period.
3 & 4
Pattern depth
Set the distance of the detection field away from
the sensor.
AFA4000/1 Airflow Monitor / p.86
Switches
Setting
Description
5 & 6
Frequency
Select one of four different sensor frequencies (H;
MH; M and L) if two or more sensor beam paths
overlap.
7
Monitor mode
Select the Monitor Mode. Set the switch to
'Normal'.
Note: The 'Snow' setting is only applicable to
outdoor applications and therefore is not
used.
8
Self-Diagnostics
Setting switch 8 to 'ON' will start a self-diagnostic
test when the sensor is switched on. If the test
detects a fault the LED will continually flash red
and green.
AFA4000/1 Airflow Monitor / p.88
6.11 Start up
Make sure the Auto Sash Controller is installed before attempting to calibrate the
controller.
When the unit is powered up, the following events occur:
The alarm performs a self-test of its functions, LEDs and audible alarm, which takes
approximately 3 seconds. All inputs and outputs are deactivated during the self-test.
At the end of the delay, the unit will do one of two things:
• When the controller has been calibrated, it enters normal operating mode and
displays the current status or alarm condition.
• When the controller has not been calibrated, the monitor displays the messages:
• Sash drive inhibited, if the tilt switch input is open
• Auto Sash Not Configured, if the tilt switch is closed
AFA4000/1 Airflow Monitor / p.89
7. Calibration
7.1
AFA4000/1 Airflow Monitor
The airflow monitor must be calibrated when it is first installed, or when required during
use.
1. From the Run Screen press and hold the Mute button for 5 seconds until the Main
Menu is displayed.
Note: When you go from the main menu to the run screen, a note briefly appears
saying ‘calibration required’.
2. Use the ↑ / ↓ buttons to select Set Up Monitor, then Calibration.
3. Enter the password (the factory default password is 0-0-0-0) and press Mute to
continue.
4. With the sash open to the normal operating height measure the face velocity using a
calibrated anemometer. Use the ↑ / ↓ buttons to enter the measured face velocity
then press Mute. The monitor samples the airflow for 5 seconds.
5. If the sample is accepted, lower the sash by half and measure the face velocity using a
calibrated instrument. Use the ↑ / ↓ buttons to enter the measured face velocity and
press Mute. The monitor samples the airflow for 5 seconds.
6. If the calibration is successful, the monitor returns to the Main Menu. Select Run to
go to normal operating mode and check the velocity reading is accurate and stable.
7. If the calibration is stable and accurate, press and hold the Mute button for 5 seconds
until the Main Menu is displayed.
8. Select Run to go to normal operating mode.
7.1.1
Possible messages
Message
Reason
Actions
Deviations too high
Airflow sample is unstable.
Follow the instructions to
repeat the sample or quit the
calibration.
AFA4000/1 Airflow Monitor / p.90
Message
Reason
Actions
Increase higher airflow
sample
Second sample value is too
close to the first sample
value.
Close the sash slightly and
repeat the higher sample. The
minimum difference between
the samples that the controller
will accept is 0.3m/sec
(60fpm).
Sensor diff too low
The monitor does not detect
any difference in the sensor
output between the 2
airflow samples.
Check that the sensor hose is
connected and repeat the
calibration.
7.1.2
Calibration notes
• Make sure the sensor hose is connected correctly at the side wall and rear of the
sensor.
• Make sure the extraction fan is running and the fume cupboard is balanced before
calibrating the monitor.
• Take extra time, at least 15 to 20 seconds, to wait for the airflow to settle before
capturing the airflow samples. This will minimize the chance of a calibration error due
to turbulence or fluctuations.
• Avoid movement in front of the fume cupboard whilst calibrating the monitor.
• The Low and High air samples must be at least 0.3 m/sec (60 fpm) apart to calibrate
the monitor. This is to avoid inaccuracy in the calibration due to insufficient difference
between the samples. The minimum difference can be changed in the Low High diff
parameter in the Cal Config menu.
• Do not use fully open and fully closed sash positions for the calibration points. The
recommended positions are normal operating height (for example, 500 mm 20”) for
the first sample and approximately half the sash opening (for example, 250 mm 10”)
for the second sample.
AFA4000/1 Airflow Monitor / p.91
7.2
Auto Sash Controller 1, 2 or 3 position.
This section describes the calibration of the Auto Sash Controller from the
AFA4000/1/AS.
Note: Refer to section 7.4 to calibrate the Auto Sash Controller from the Auto Sash keypad
1. From the Run Screen press and hold the Mute button for 5 seconds until the Main
Menu is displayed.
Note: When you go from the main menu to the run screen, a note briefly appears
saying ‘calibration required’.
2. Use the ↑ / ↓ buttons to select Set Up Auto Sash. The display shows the message TEL
SASH CLOSER.
3. Press the Mute button to display the current Hardware and Software version, for
example, 1. HW2.0 FW1.1.
4. Press the ↓ button to scroll down and select Setup Menu and press Mute.
5. Enter the password (the factory default password is 1-0-0-1) and press Mute to
continue.
6. In the Setup menu the controller displays System Profile.
7. Use the ↓ button to scroll through the menu, select Sash 1 Settings and press Mute.
8. In the S1 Settings menu, the display will show 1. S1 Enabled and on-screen
instructions.
9. Set the sash to the centre opening, for example, 250 mm.
10. Press the ↓ button to find which arrow represents the sash closing and press Mute.
This step shows the controller which direction the motor needs to drive to close the
sash.
11. Close the sash and press Mute. This step sets the sash position sensor output, 0 V
position.
12. Fully open the sash and press Mute. This step sets the sash position sensor output, 10
V position.
13. Set the sash to the lower position, for example, close and press Mute.
14. Set the positions as required:
AFA4000/1 Airflow Monitor / p.92
For 1 position operation:
When only the lower position is required (for example,
close only), press ↑ / ↓ together to quit.
For 2 position operation:
Set the sash to the normal working height (for
example, 500 mm) and press Mute.
Press ↑ / ↓ together to quit, when only 2 positions
(lower and middle) are required.
For 3 position operation:
Set the sash to the full opening height (for example,
750 mm) and press Mute.
The lower, middle and top positions are now
calibrated. Press Mute to quit calibration mode.
15. Use the ↓ button and select Back to Set up menu, then press Mute.
16. Use the ↓ button and select Exit and Save Changes, then press Mute.
17. The controller now enters Run mode.
7.2.1
Calibration notes
The sash may slightly over-run past the calibration points, when the sash motor drive is
set to a fast speed. When calibrating the sash positions, allow a margin for the sash to
over-travel past the calibration height.
For example, before calibrating the lower position, fully close the sash and then open it
by 5 mm, or before calibrating the middle position, open the sash to the normal height
and then lower it by 5 mm.
AFA4000/1 Airflow Monitor / p.93
7.3
Sash lock function
This section describes the calibration of the Auto Sash Controller from the
AFA4000/1/AS.
Note: Refer to section 7.4 to calibrate the Auto Sash Controller from the Auto Sash keypad.
1. From the Run Screen press and hold the Mute button for 5 seconds until the Main
Menu is displayed.
Note: When you go from the main menu to the run screen, a note briefly appears
saying ‘calibration required’.
2. Use the ↑ / ↓ buttons to select Set Up Auto Sash. The display shows the message TEL
SASH CLOSER.
3. Press the Mute button to display the current Hardware and Software version, for
example, 1. HW2.0 FW1.1.
4. Press the ↓ button to scroll down and select Setup Menu and press Mute.
5. Enter the password (the factory default password is 1-0-0-1) and press Mute.
6. In the Setup menu the controller will display the message System Profile.
7. Use the ↓ button to scroll through the menu and select Sash 1 Settings, then press
Mute.
8. In the S1 Settings menu the display will show 1. S1 Enabled.
9. Use the ↓ button to scroll through the menu and select 11. S1 Auto Lock. Press
Mute.
10. Set to Enabled and press Mute.
11. Use the ↓ button to scroll through the menu and select 11b. S1 Lock Alarm. Press
Mute.
12. Use the ↑ / ↓ buttons to set the required alarm delay time. Press Mute.
13. Use the ↓ button to scroll through the menu and select 11c. S1 Lock Pos Calibration.
Press Mute.
14. The display will show on-screen instructions.
15. Open the sash to the Lock position and press Mute.
When the selected position is too low, for example, it is lower than the calibrated
Lower, Middle or Top positions, the message Bad position - try again displays.
AFA4000/1 Airflow Monitor / p.94
If the chosen position is OK, the lock position will be set.
16. Use the ↓ button and select Back to Set up menu. Press Mute.
17. Use the ↓ button and select Exit and Save Changes. Press Mute.
The controller now enters Run mode.
7.4
Auto Sash Controller – stand-alone
This section refers to when the Auto Sash Controller is used without the AF4000/1/AS
Airflow Monitor.
Before starting up the Auto Sash Controller, ensure that all Auto Sash Control System
components have been correctly installed and connected.
Note:
Check that the fume hood sash and associated motor drives, pulleys, operating
cables, racks and chains move freely over their entire operating ranges.
Note:
Check that all cables are secured away from moving components and are not
overstressed or damaged.
Note:
Check the security of all electrical connections.
AFA4000/1 Airflow Monitor / p.95
7.4.1
Starting up the Auto Sash Control Unit
Press the Power Up/Menu Entry button to start the Auto Sash Controller (Figure 63 and
Figure 64).
Figure 63: UK / EU keypad
Figure 64: US keypad
1. Switch on the Auto Sash Control Unit.
When powering up, the Auto Sash Control Unit performs a self-test of its functions,
LEDs and the audible alarm. This self-test takes approximately 3 seconds.
During the self-test period, all inputs and outputs are de-activated.
2. At the end of the self-test, the Auto Sash Control Unit will do one of two things:
If the Auto Sash Control Unit has previously been calibrated:
• The Auto Sash Controller enters the Normal Operating Mode and displays its
current status or alarm condition.
If the Auto Sash Control Unit has not been calibrated:
• When the Tilt Switch input is open, the display shows 'Sash Drive Inhibited'.
• When the Tilt Switch input is closed, the display shows ‘Auto Sash Not Configured'.
7.4.2
Calibrating the Auto Sash Control Unit for Tiptronic operation using the
Auto Sash keypad
The Auto Sash Control Unit must have each of its required stationary positions defined
for Tiptronic functionality:
AFA4000/1 Airflow Monitor / p.96
1. Press and hold the Enter button for approximately 5 seconds or until the main menu
appears.
The LCD display will show the current hardware and software versions. For example, 1.
HW2.0 FW1.1.
2. Press ↓ and select 2. Setup.
3. Press Enter.
4. Enter the Password using the ↑ arrow and Enter button, the factory default is 1 0 0
1. In the Set-Up menu, the controller will display Keypad Tones.
5. Press ↓ and select Sash 1 Settings.
6. Press Enter.
In the S1 (Sash 1) Settings menu, the display will show 1. S1 Enabled.
7. Press ↓ and select 8. S1 Calibration.
8. Press Enter.
The display will show on screen instructions. For example:
"Manually move sash to centre and then press the Up/Down keys to find sash down
direction. Press Cancel to abort Calibration"
9. Move the sash to the middle of the sash opening.
10. Press ↑/↓ buttons to determine which arrow activates sash closing.
11. Press the ↑/↓ button which closes the sash, press Enter.
This instructs the controller which direction the motor needs to turn to close the sash.
12. Move the sash to the lowest position required, press Enter.
13. Either:
• Press Mute, if this is the only pre-set position required, or
• Set the sash to the normal working height, for example 500 mm.
14. Press Enter, for two position operation.
15. Either:
• Press Mute, if only two pre-set positions (‘lowest’ and ‘normal’) are required.
• Set the sash to the full opening height.
16. Press Enter, for three position operation.
The lowest, normal and fully open positions are now calibrated.
17. Press Enter, to exit Calibration mode.
AFA4000/1 Airflow Monitor / p.97
18. Press ↓ button, select Back to Set up menu.
19. Press Enter.
20. Press ↓ button, select Exit and Save Changes.
21. Press Enter.
The Controller will now enter Run mode.
Note:
When the Auto Sash motor drive is set to a high speed, allow a margin for the
sash to over-run when calibrating the sash position. See the instructions above.
Allow a 5 mm (0.197") margin at the highest and lowest point.
7.4.3
Calibrating the sash lock function
To define the required Sash Lock position:
1. Press and hold the Enter button for 5 seconds until the Main Menu is displayed.
2. The current Hardware and Software versions are displayed, for example 1. HW2.0
FW1.1
3. Press ↓ button, select 2. Setup and press Enter.
4. Enter the Password using the ↑ arrow and Enter button, the factory default is 1 0 0
1. In the Set-Up menu, the controller will display Keypad Tones.
5. Press ↓ button, select Sash 1 Settings and press Enter. The controller will display 1. S1
Enabled.
6. Press ↓ button, select 11. S1 Auto Lock and press Enter.
7. Set to Enabled and press Enter.
8. Press ↓ button, select 11b. S1 Lock Alarm and press Enter.
9. Press ↑/↓ buttons to set the required alarm delay time and press Enter.
10. Press ↓ button, select 11c. S1 Lock Pos Calibration and press Enter.
Instructions will be displayed, for example: Lock the sash then move the sash to the
lowest lock position. Enter to accept. Cancel to abort
11. Open the sash to the Lock position and press Enter.
If the position chosen is too low, lower than the calibrated Bottom, Middle or Top
positions, Bad position - try again will be displayed.
If the chosen position is acceptable, the lock position will be set.
12. Press ↓ button, select Back to Set up menu and press Enter.
13. Press ↓ button, select Exit and Save Changes and press Enter.
AFA4000/1 Airflow Monitor / p.98
The Auto Sash Controller will now go to RUN mode.
7.4.4
Calibrating the safety light curtain
To adjust the range and sensitivity of the Safety Light Curtain:
1. Ensure the area between the Safety Light Curtain transmitter and receiver is free from
obstructions.
2. Ensure the light beam paths are not broken and the Personnel Sensor is disabled so
that the sash does not Auto-Close whilst the Safety Light Curtain is being calibrated.
Note: If it is not possible to disable the Personnel Sensor, remove the motor / clutch
connector from the Auto Sash Control Unit. This activates the Sash Control
Alarm after the selected time delay and the reset button must be pressed once
the Safety Light Curtain has been calibrated.
3. Remove the cover from the Safety Light Curtain Control Box.
AFA4000/1 Airflow Monitor / p.99
Figure 65: Safety Light Curtain Control Box with lid removed
4. Power up the Auto Sash Control Unit.
5. Check that the Safety Light Curtain Transmitter and Receiver are connected correctly
and are operational.
The red Power On LED will now be visible on the receiver (Figure 66).
AFA4000/1 Airflow Monitor / p.100
Figure 66: Red Power On LED
6. Ensure that Switch 1 (sw1), see Figure 54, is set to the correct range:
Fume Hood Size
Sash Beam Range
sw1 Setting
0 to 10’ (up to 3000mm)
Short
Off
10’ to 16’ (3000 to
5000mm)
Long
On
7. Ensure that Switch 2 (sw2) (Figure 65) is set to Off (Operation Mode).
8. Ensure that Switch 3 (sw3) is set to On (Normal Operation).
9. Set Switch 4 (sw4) to On. A buzzer will then sound when the beam is broken, this aids
the setting up of the beam.
The signal LED indicates beam status:
• Green if the detection area is clear.
• Red if the area is not clear.
10. Turn the gain potentiometer anti-clockwise, to the minimum. The transmitter and
receiver can no longer sense each other.
11. Slowly adjust the gain potentiometer (Figure 65) clockwise (to maximum) until the
transmitter and receiver can sense each other, for example, the receiver can detect
the beam sent from the transmitter.
12. Set sw3 to Off to store the settings.
13. Turn the gain potentiometer approximately 5 degrees clockwise to ensure sufficient
gain.
14. Use a glass object to test whether it breaks the beam. Do this across the entire open
face of the fume hood.