Keysight Technologies 34931A Operation & User’s Manual
Text of the manual
Keysight 34931A-34933A User’s Guide
Notices
Copyright Notice
© Keysight Technologies 2008-2019
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies as governed by
United States and international
copyright laws.
Trademarks
Microsoft® and Windows® are U.S.
registered trademarks of Microsoft
Corporation.
Manual Part Number
34980-90031
Edition
Edition 3, July 1, 2019
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight
products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest
Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal
Acquisition Regulation (“FAR”) 2.101.
Pursuant to FAR 12.212 and 27.405-3
and Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires commercial
computer software under the same
terms by which the software is
customarily provided to the public.
Accordingly, Keysight provides the
Software to U.S. government
customers under its standard
commercial license, which is embodied
in its End User License Agreement
(EULA), a copy of which can be found
at http://www.keysight.com/find/
sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license set
forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation that
is not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall apply,
except to the extent that those terms,
rights, or licenses are explicitly required
from all providers of commercial
computer software pursuant to the FAR
and the DFARS and are set forth
specifically in writing elsewhere in the
EULA. Keysight shall be under no
obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,” AND
IS SUBJECT TO BEING CHANGED,
WITHOUT NOTICE, IN FUTURE
EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY
INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR
CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE FURNISHING,
USE, OR PERFORMANCE OF THIS
DOCUMENT OR OF ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS COVERING
THE MATERIAL IN THIS DOCUMENT
THAT CONFLICT WITH THESE TERMS,
THE WARRANTY TERMS IN THE
SEPARATE AGREEMENT SHALL
CONTROL.
Safety Information
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
Keysight 34931A-34933A User’s Guide
3
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Alternating current (AC)
Caution, risk of electric shock
Frame or chassis (ground) terminal
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
Standby supply. Unit is not completely
disconnected from ac mains when
switch is off
Keysight 34931A-34933A User’s Guide
Additional Safety Notices
The following general safety precautions must be observed during all phases of
operation of this instrument. Failure to comply with these precautions or with
specific warnings or instructions elsewhere in this manual violates safety
standards of design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability of the customer’s failure to comply with the
requirements.
General
Do not use this products in any manner not specified by the manufacturer. The
protective features of this product may be impaired if it is used in a manner not
specified in the operation instructions.
Before Applying Power
Verify that all safety precautions are taken. Make all connections to the unit before
applying power.
Ground the Instrument
This product is provided with protective earth terminals. To minimize shock
hazard, the instrument must be connected to the ac power mains through a
grounded power cable, with the ground wire firmly connected to an electrical
ground (safety ground) at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal will cause
a potential shock hazard that could result in personal injury.
Do Not Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of flammable gases or fumes.
Keysight 34931A-34933A User’s Guide
5
Do Not Remove the Instrument Cover
Only qualified, service-trained personal who are aware of the hazards involved
should remove instrument covers. Always disconnect the power cable and any
external circuits before removing the instrument cover.
Do Not Modify the Instrument
Do not install substitute parts or perform any unauthorized modification to the
product. Return the product to an Keysight Sales and Service Office for service
and repair to ensure that safety features are maintained.
In Case of Damage
Instruments that appear damaged or defective should be made inoperative and
secured against unintended operation until they can be repaired by qualified
service personnel.
Keysight 34931A-34933A User’s Guide
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/
96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
–
www.keysight.com/find/34980a
(product-specific information and support, software and
documentation updates)
–
www.keysight.com/find/assist
(worldwide contact information for repair and service)
all-guides.co
m
Keysight 34931A-34933A User’s Guide
7
Table of Contents
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Additional Safety Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Before Applying Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Ground the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Do Not Operate in an Explosive Atmosphere . . . . . . . . . . . . . . . . . . .4
Do Not Remove the Instrument Cover . . . . . . . . . . . . . . . . . . . . . . . .5
Do Not Modify the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
In Case of Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/
EC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Matrix Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Operating Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
SCPI Programming Examples for the Matrix Modules . . . . . . . . . . . . . .11
Linking Multiple Matrix Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
34931A Dual 4x8 Armature Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
34931A Simplified Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
34931A D-Sub Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
34931T Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
34932A Dual 4x16 Armature Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . .23
34932A Simplified Schematic . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
34932A D-Sub Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
34932T Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
34933A Dual/Quad 4x8 Reed Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . .28
34933A Simplified Schematic for Two-Wire Mode . . . . . . . . . . . . .30
34933A D-Sub Connectors for Two-Wire Mode . . . . . . . . . . . . . . . .31
34933T-001 Terminal Block for Two-Wire Mode . . . . . . . . . . . . . . .33
34933A Simplified Schematic for One-Wire Mode . . . . . . . . . . . . .35
Keysight 34931A-34933A User’s Guide
34933A D-Sub Connectors for One-Wire Mode . . . . . . . . . . . . . . . 36
34933T-002 Terminal Block for One-Wire Mode . . . . . . . . . . . . . . . 38
Keysight 34931A-34933A User’s Guide
9
Matrix Modules
This User’s Guide covers the following three plug-in modules for the Keysight
34980A Multifunction Switch/Measure Unit:
The 34931A has two (dual) matrices of latching armature switches.
Each matrix is organized in a 4-row by 8-column configuration.
The 34932A has two (dual) matrices of latching armature switches.
Each matrix is organized in a 4-row by 16-column configuration.
The 34933A has non-latching reed switches, which you can configure for:
–
differential (2-wire) mode, which has two (dual) matrices.
Each matrix is organized in a 4-row by 8-column configuration.
–
single-ended (1-wire) mode, which has four (quad) matrices.
Each matrix is organized in a 4-row by 8-column configuration.
These matrix switch modules offer a convenient way for you to connect multiple
instruments to multiple points on your device under test. For a lower cost and
better specification alternative, you can connect both matrix and multiplexer
(MUX) modules.
34931A
Dual 4x8 armature matrix
34932A
Dual 4x16 armature matrix
34933A
Dual/Quad 4x8 reed matrix
Keysight 34931A-34933A User’s Guide
Operating Considerations
Connection to Voltage Sources
Although flexible, it is possible to connect more than one source at the same time
with a matrix. Make sure that dangerous or unwanted conditions are not created
by these connections.
Electrical Considerations
See the Introduction to the Plug In Modules chapter of the 34980A Mainframe
User’s Guide for detailed environmental operating conditions for the 34980A
mainframe and its installed modules. That guidance sets maximum per channel
current and power ratings at rated voltage for pollution degree 1 (dry) and
pollution degree 2 (possible condensation) conditions, for each of the matrix
modules.
NOTE
Safety Interlock The Analog Buses of the 34980A are capable of carrying
300V signals. The matrix modules have a hardware Safety Interlock feature that
automatically opens the Analog Bus relays when the associated interlock pins on
the D-sub connectors (faceplate) lose continuity. This prevents signals on the
Analog Buses from being present on the D-sub connector pins. Optional terminal
blocks available from Keysight automatically provide continuity for these
interlock pins. If cables are used, you must provide continuity for the interlock
pins in your DUT assembly. See the pinout information later in this manual for
the location of interlock pins on each module.
The matrix modules have Analog Bus relays on Bank 2 only. Therefore, the
interlock pins are present on only the Bank 2 D-sub connectors.
Normally, if you attempt to connect to the Analog Buses without
a terminal block or cable connected, an error is generated. The
SYSTem:ABUS:INTerlock:SIMulate command allows you to temporarily
disable errors generated by the Safety Interlock feature and enables the
simulation mode. Although Safety Interlock errors are suppressed in this mode,
the actual Analog Bus relays affected by the Safety Interlock are disabled as
long as no terminal block or cable is connected to the module.
Keysight 34931A-34933A User’s Guide
11
SCPI Programming Examples for the Matrix Modules
The programming examples below provide you with SCPI command examples to
use for actions specific to the matrix switch modules.
The slot and channel addressing scheme used in these examples follow the
general form sccc where s is the mainframe slot number (1 through 8) and ccc is
the three-digit channel number. Channel numbers for the matrix modules are
derived as follows:
Two-wire mode: The channel numbers for the 34931A, 34932A, and the
34933A (2-wire mode) are derived from the crosspoint or intersection of rows and
columns, columns having two digits. See the example below.
One-wire mode: The channel numbers for the 34933A (in 1-wire mode) are
derived from a specific matrix number and the crosspoint or intersection of rows
and columns on that matrix. See the example below.
For information on specific configurations, refer to the simplified schematics for
the matrix modules on page 18, page 24, page 30 and page 35.
Displayed Channel
Means This...
5304
A 34931A, 34932A, 34933A (2-wire mode)
matrix module is in slot 5, crosspoint is row
3, column 4. It might be easy to remember
this channel configuration as “srcc” (slot,
row, column, column)
Displayed Channel
Means This...
2437
A 34933A matrix module in 1-wire mode is
in slot 2, matrix of interest is 4, crosspoint is
row 3, column 7. It might be easy to
remember this channel configuration as
“smrc” (slot, matrix, row, column)
all-guides.co
m
Keysight 34931A-34933A User’s Guide
For complete information on the SCPI commands used to program the 34980A,
refer to the Keysight 34980A Programmer’s Reference which can be downloaded
from www.keysight.com/find/34980A.
Opening and Closing Channels
Example: Closing and opening matrix channels (34931A, 34932A, and
34933A in two-wire mode) The following commands close and open channels
311 and 312 through 315 of a 34932A matrix module in 2-wire mode. This module
is in slot 3. The channel number represents the matrix crosspoint of a row (one
digit) and a column (two digits). For example, channel 311 represents crosspoint
at row 3 and column 11 on a 34932A module.
ROUTe:CLOSe (@3311,3312:3315)
ROUTe:OPEN (@3311,3312:3315)
Example: Closing and opening matrix channels (34933A in one-wire
mode) The following commands close and open channels 311 and 312 through
315 of the 34933A module in 1-wire mode. The module is in slot 4. The channel
number represents the matrix and the matrix crosspoint of a row (one digit) and a
column (one digit). For example, channel 311 represents the crosspoint on matrix
3 at row 1, column 1 on a 34933A module in 1-wire mode.
ROUTe:CLOSe (@3311,3312:3315)
ROUTe:OPEN (@3311,3312:3315)
Example: Closing and opening Analog Bus relays The following command
connects the Analog Buses to Matrix 2 for a module (in 2-wire mode) in slot 3.
ROUTe:CLOSe (@3921,3922,3923,3924)
ROUTe:OPEN (@3921,3922,3923,3924)
NOTE
Although the previous two examples show the same channel numbers, the
channels are derived differently as determined by a module’s configuration
mode. See page 11 for channel number derivation.
NOTE
For matrix modules in 2-wire mode, only Matrix 2 connects to the the Analog
Buses. For the 34933A in 1-wire mode, only Matrix 3 and Matrix 4 connect to the
Analog Buses.
Keysight 34931A-34933A User’s Guide
13
The Analog Bus relays (numbered s921, s922, s923, etc.) on the matrix modules
are ignored if they are included in a range of channels. An error will be generated
if an Analog Bus relay is specified as the first or last channel in a range of
channels. For example, the following command closes all valid channels between
channel 304 and channel 615 (slot 2). In addition, this command closes Analog
Bus relay 911 on the module in slot 1 (Bank 1). Note that although the specified
range of channels includes the other Analog Bus relays, they are ignored and are
not closed by this command.
ROUTe:CLOSe (@2304:2615,1911)
Example: Querying channels for open or close state The following
command returns a 1 (true) or 0 (false) state of channel 204 for a module in slot 3.
ROUTe:CLOSe (@3204)
ROUTe:CLOSe? (@3204) !Returns a 1
ROUTe:OPEN? (@3204) !Returns a 0
Configuring a Module
Example: Configuring the 34933A module for 2-wire or 1-wire mode The
following command configures a matrix module in slot 4 for 1-wire measurements.
Because you can configure only the 34933A (and the 34923A and 34925A MUX
modules) for either 2-wire or 1-wire mode, an error is generated if you send this
command to a slot that does not contain one of those three modules. If you are
using terminal blocks with the 34933A module, be sure to use the corresponding
2-wire or 1-wire terminal block.
SYSTem:MODule:WIRE:MODE WIRE1,4
Example: Querying the system for module Identify The following command
returns the identity of the module installed in slot 7.
SYSTem:CTYPe? 7
NOTE
When using a command to configure the system, the new configuration does not
take effect until you cycle power on the 34980A.
Keysight 34931A-34933A User’s Guide
Reading Cycle Count and Resetting Modules to Power-On State
Example: Reading the cycle count for a relay The following command
returns the cycle count on channels 304 and 308 for a matrix module in slot 3.
DIAGnostic:RELay:CYCLes? (@3304,3308)
Example: Resetting module(s) to power-on state The following command
resets a module in slot 4 to its power-on state.
SYSTem:CPON 4
NOTE
For the 34933A matrix module, the query response may include a suffix to
indicate a 1-wire configuration. For example, the response for the 34933A will be
either "34933A" (differential mode) or "34933A-1W" (single-ended mode).
Keysight 34931A-34933A User’s Guide
15
Linking Multiple Matrix Modules
You can link multiple matrix modules to form a larger matrix. The drawings on
page 16 show examples of two-module connections through rows and columns.
Wiring Multiple 34931A or 34932A Modules
With a 34931A you can combine two matrices to form 8x8 (connecting columns)
or 4x16 (connecting rows) configurations. Using two 34932A matrices on a
34932A module, you can create 16x8 (connecting columns) or 4x32 (connecting
rows) configurations.
You can connect rows in separate modules using external wiring. Or, using Bank 2
matrices, you can connect through the mainframe Analog Buses. For a clear idea
of how matrices are arranged and their connections to the Analog Buses, see the
simplified schematics on page 18 (34931A) and page 24 (34932A).
You must use external wiring whenever you connect:
–
Rows in Matrix 1 of separate modules
–
Rows in Matrix 1 to rows in Matrix 2 on the same or separate modules
–
Columns of two matrices on the same or separate modules
You can expand upon these two-module configurations and add up to eight
modules to design your own large matrices. From a programming standpoint,
each matrix module operates as an independent module regardless of the
external connections. When linking modules, the channel numbering scheme
remains the same as for single modules.
Wiring Multiple 34933A Modules
You can connect matrices on the 34933A module in a similar fashion to the
34931A. However, the presence of in-rush resistors on the Analog Buses and
columns require additional consideration, and you must take care when linking
multiple 34933A matrix modules. See the simplified schematics on page 30 and
page 35.
Keysight 34931A-34933A User’s Guide
1
2
3
4
1
2
3
4
1
2
3
n*
n - 1
1
2
3
n*
n - 1
*n can be 8 or 16
8 or 16 Columns
8 Rows
Module 1
Module 2
Increase number of
rows by connecting
through columns
1
2
3
4
1
2
3
4
1
2
3
n*
n - 1
1
2
3
n*
n - 1
Analog
Buses
4 Rows
*n can be 8 or 16
Module 1
Module 2
16 or 32 Columns
Increase number of
columns by connecting
through rows
all-guides.co
m
Keysight 34931A-34933A User’s Guide
17
34931A Dual 4x8 Armature Matrix
The 34931A dual 4x8 armature matrix contains two matrices, each with 32 2-wire
crosspoint latching armature relays organized in a 4-row by 8-column
configuration. Every row and column are made up of two wires each, a high (H)
and a low (L). Each crosspoint relay has a unique channel number representing
the row and column that intersects to create the crosspoint. For example, channel
304 represent the crosspoint connection between row 3 and column 4 (all
columns consisting of two digits; in this case the digits are 04). See the simplified
schematic on page 18.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 2 in this module connects to the Analog Buses. By closing
channels 921 and 922 you can connect rows 5 and 6 respectively to the internal
DMM of the 34980A mainframe for voltage and resistance measurements. You
can connect multiple matrix modules externally and/or through the Analog Buses
for applications that require large matrices. For information on linking multiple
matrices, refer to page 15 of this manual.
When the power is off, matrix relays maintain state, and Analog Bus relays open.
NOTE
When the DMM is scanning, it controls ABus1 and ABus2 relays, which are on
Matrix 2. Therefore, consider this behavior when you are connecting matrices.
Keysight 34931A-34933A User’s Guide
34931A Simplified Schematic
ABus1
DMM
(MEAS)
ABus2
DMM
(SENS)
ABus3
ABus4
922
923
924
921
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
L
L
L
L
H
H
H
H
H
H
H
L
L
L
L
H
H
L
L
H
H
H
H
H
H
H
H
L
L
L
L
L
L
L
L
L
L
L
L
H
H
H
Matrix 1
Matrix 2
Analog Buses
Row 1
Row 2
Row 3
Row 4
Row 5
Row 6
Row 7
Row 8
Col 1
Col 2
Col 3
Col 4
Col 5
Col 6
Col 7
Col 8
Col 1
Col 2
Col 3
Col 4
Col 5
Col 6
Col 7
Col 8
NOTE: Matrix 1 and Matrix 2 are
electrically separate from one another.
NOTE:
Matrix Relays: Armature latching
Analog Bus Relays: Armature non-latching
NOTE: Three-digit channel numbers
are derived from the intersection of
the rows and columns, columns
having two digits. The intersection
shown here represents Channel 304
(Row 3, Column 4)
Keysight 34931A-34933A User’s Guide
19
34931A D-Sub Connectors
Matrix 1
For orientation, the D-sub connector
end of the module is facing you.
Matrix 1
Matrix 2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
NC
C4H
C4L
NC
NC
R4H
R4L
C5H
C5L
NC
NC
NC
NC
C7H
C7L
NC
NC
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
C2L
NC
NC
NC
NC
NC
C2H
NC
NC
R2H
R2L
C8H
C8L
NC
NC
GND
NC
NC
C3H
C3L
C1H
C1L R3H
R3L
C6H
C6L
NC
NC
NC
NC
NC
R1H
R1L
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
NOTE: In this diagram and
the table below, R represents
“row,” and C represents
“column.”
50-Pin D-Sub Male
Connector
Description
Pin
Description
Pin
Description
Pin
R1H
49
C2H
23
C7H
13
R1L
50
C2L
24
C7L
14
R2H
27
C3H
35
C8H
29
R2L
28
C3L
36
C8L
30
R3H
39
C4H
1
GND
33
R3L
40
C4L
2
No Connect pins:
3-4, 9-12, 15-22, 25-26, 31-32, 34,
and 43-48.
R4H
5
C5H
7
R4L
6
C5L
8
C1H
37
C6H
41
C1L
38
C6L
42
Keysight 34931A-34933A User’s Guide
Matrix 2
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
NC
NC
C3H
C3L
C1H
C1L
R7H
R7L
C6H
C6L
NC
NC
NC
NC
NC
R5H
R5L
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Interlock
C4H
C4L
NC
NC
R8H
R8L
C5H
C5L
NC
NC
NC
NC
C7H
C7L
NC
NC
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
C2L
NC
NC
NC
NC
NC
C2H
NC
NC
R6H
R6L
C8H
C8L
NC
NC
Interlock
NOTE: In this diagram and
the table below, R
represents “row,” and C
represents “column.”
50-Pin D-Sub
Male Connector
Description
Pin
Description
Pin
Description
Pin
R5H
49
C2H
23
C7H
13
R5L
50
C2L
24
C7L
14
R6H
27
C3H
35
C8H
29
R6L
28
C3L
36
C8L
30
R7H
39
C4H
1
Interlock
17
R7L
40
C4L
2
Interlock
33
R8H
5
C5H
7
No Connect pins: 3-4, 9-12, 15-16.
18-22, 25-26, 31-32, 34, 43-48
R8L
6
C5L
8
C1H
37
C6H
41
C1L
38
C6L
42
WARNING
As a safety feature, interlock pins (17 and 33) must be shorted to enable the
Analog Bus relays, which are on Matrix 2, to close. The optional 34931T
terminal block shorts these pins for you. This feature protects inadvertent
routing of high voltages from the Analog Buses to the D-sub connector of
the module.
Keysight 34931A-34933A User’s Guide
21
34931T Terminal Block
This terminal block with screw-type connections is labeled with the model number
and the abbreviated module name. In addition, space is available on the label for
you to write the slot number.
NOTE
All modules that connect to the internal DMM are interlock protected. This
means that when an installed module is exposed (no terminal block or cable is
connected), the Analog Bus relays, which are on Matrix 2, are open and
disconnected from the Analog Buses. See page 10 for further information.
all-guides.co
m
Keysight 34931A-34933A User’s Guide
NOTE
On the 34931T terminal block, only two sets of screw terminals are for use with
the 34931A module. See the following drawing.
When using the 34931T terminal block, be sure to
wire your connections to the two sets of screw
terminals closest to the 50-pin D-sub connectors.
Although columns are numbered the same on
Matrix 1 and Matrix 2, they are electrically
separate from one another (e.g., Col C8).
Wire Size:
20 AWG Typical
18 AWG Max
Keysight 34931A-34933A User’s Guide
23
34932A Dual 4x16 Armature Matrix
The 34932A dual 4x16 armature matrix contains two matrices, each with 64
2-wire crosspoint latching armature relays organized in a 4-row by 16-column
configuration. Every row and column are made up of two wires each, a high (H)
and a low (L). Each crosspoint relay has a unique channel number representing
the row and column that intersect to create the crosspoint. For example, channel
315 represents the crosspoint connection between row 3 and column 15 (all
columns consisting of two digits; in this case the digits are 15). See the simplified
schematic on page 24.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 2 in this module connects to the Analog Buses. By closing
channels 921 and 922 you can connect rows 5 and 6 respectively to the internal
DMM of the 34980A mainframe for voltage and resistance measurements. You
can connect multiple matrix modules externally and/or through the Analog Buses
for applications that require large matrices. For information on linking multiple
matrix modules, refer to page 15 of this manual.
When the power is off, matrix relays maintain state, and Analog Bus relays open.
NOTE
When the DMM is scanning, it controls ABus1 and ABus2 relays, which are on
Matrix 2. Therefore, consider this behavior when you are connecting matrices.
Keysight 34931A-34933A User’s Guide
34932A Simplified Schematic
ABus1
DMM
(MEAS)
ABus2
DMM
(SENS)
ABus3
ABus4
922
923
924
921
H
L
H
L
H
L
H
L
H
L
H
L
L
L
L
L
H
H
H
H
H
H
H
L
L
L
L
H
H
H
H
H
H
L
L
L
L
L
L
L
L
L
L
H
H
H
Col 1
Col 2
Col 3
Col 4
Col 15
Col 16
Col 1
Col 2
Col 3
Col 4
Col 15
Col 16
Row 1
Row 2
Row 3
Row 4
Row 5
Row 6
Row 7
Row 8
Matrix 1
Matrix 2
Analog Buses
H
H
L
L
NOTE: Matrix 1 and Matrix 2 are
electrically separate from one another.
NOTE: Three-digit channel numbers are
derived from the intersection of the rows
and columns, columns having two digits.
The intersection shown here represents
Channel 315 (Row 3, Column 15).
NOTE:
Matrix Relays: Armature latching
Analog Bus Relays: Armature non-latching
Keysight 34931A-34933A User’s Guide
25
34932A D-Sub Connectors
Matrix 1
For orientation, the D-sub connector
end of the module is facing you.
Matrix 1
Matrix 2
NOTE: In this diagram
and the table below, R
represents “row,” and C
represents “column.”
50-Pin D-Sub
Male Connector
C12H
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
NC
C4H
C4L
C12L R4H
R4L
C5H
C5L C13H C13L
NC
NC
C7H
C7L C15H C15L
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
C2L
NC
C11H C11L C9H
C9L C2H
C14H C14L R2H
R2L
C8H
C8L
NC
NC
GND
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
NC
NC
C3H
C3L
C1H
C1L
R3H
R3L
C6H
C6L C10H C10L
NC
C16H C16L R1H
R1L
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
R1H
49
C2H
23
C7H
13
C12H
3
GND
33
R1L
50
C2L
24
C7L
14
C12L
4
No Connect pins:
11-12, 17-18, 31-32,
34, and 45-46
R2H
27
C3H
35
C8H
29
C13H
9
R2L
28
C3L
36
C8L
30
C13L
10
R3H
39
C4H
1
C9H
21
C14H
25
R3L
40
C4L
2
C9L
22
C14L
26
R4H
5
C5H
7
C10H
43
C15H
15
R4L
6
C5L
8
C10L
44
C15L
16
C1H
37
C6H
41
C11H
19
C16H
47
C1L
38
C6L
42
C11L
20
C16L
48
Keysight 34931A-34933A User’s Guide
Matrix 2
NOTE: In this diagram
and the table below, R
represents “row,” and C
represents “column.”
50-Pin D-Sub
Male Connector
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Interlock
C4H
C4L C12H C12L R8H
R8L
C5H
C5L C13H C13L
NC
NC
C7H
C7L C15H C15L
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
NC
NC
C3H
C3L
C1H
C1L
R7H
R7L
C6H
C6L C10H C10L
NC
C16H C16L R5H
R5L
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
C2L
NC
C11H C11L C9H
C9L
C2H
C14H C14L R6H
R6L
C8H
C8L
NC
NC Interlock
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
R5H
49
C2H
23
C7H
13
C12H
3
Interlock
17
R5L
50
C2L
24
C7L
14
C12L
4
Interlock
33
R6H
27
C3H
35
C8H
29
C13H
9
No connect pins:
11-12, 18, 31-32, 34,
and 45-46.
R6L
28
C3L
36
C8L
30
C13L
10
R7H
39
C4H
1
C9H
21
C14H
25
R7L
40
C4L
2
C9L
22
C14L
26
R8H
5
C5H
7
C10H
43
C15H
15
R8L
6
C5L
8
C10L
44
C15L
16
C1H
37
C6H
41
C11H
19
C16H
47
C1L
38
C6L
42
C11L
20
C16L
48
WARNING
As a safety feature, interlock pins (17 and 33) must be shorted to enable the
Analog Bus relays, which are on Matrix 2, to close. The optional 34932T
terminal block shorts these pins for you. This feature protects inadvertent
routing of high voltages from the Analog Buses to the D-sub connector of
the module.
all-guides.co
m
Keysight 34931A-34933A User’s Guide
27
34932T Terminal Block
This terminal block with screw-type connections is labeled with the model number
and the abbreviated module name. In addition, space is available on the label for
you to write the slot number.
NOTE
All modules that connect to the internal DMM are interlock protected. This
means that when an installed module is exposed (no terminal block or cable is
connected), the Analog Bus relays, which are on Matrix 2, are open and
disconnected from the Analog Buses. See page 10 for further information.
Although they have
separate
screw-type
connectors, rows
labeled the same
on a matrix are
electrically
connected.
Therefore, you can
wire the
same-matrix rows
in two locations.
Although columns are numbered the same on
Matrix 1 and Matrix 2, they are electrically
separate from one another (e.g., Col C13).
Wire Size:
20 AWG Typical
18 AWG Max
Keysight 34931A-34933A User’s Guide
34933A Dual/Quad 4x8 Reed Matrix
Using program commands or the front panel of the 34980A, you can configure the
34933A dual/quad 4x8 reed matrix module for differential (2-wire) mode or
single-ended (1-wire) mode.
The 34933A module contains 100 Ω in-rush resistors that are used to protect the
reed relays from reactive loads. If you have applications where in-rush resistors
interfere with measurements, connections are provided on the terminal blocks for
you to bypass the in-rush resistors that are located on the columns. See the
simplified schematics on page 30 and page 35. However, if you choose to bypass
the in-rush resistors, the life of the reed relays that you bypass may be degraded.
Two-Wire Mode
To physically configure the module for 2-wire mode, use the 34933T-001 terminal
block, or a compatible standard or custom cable. If using a standard or custom
cable, make sure you connect interlock pins 17 and 33 on the Matrix 2 D-sub
connector. Refer to the pinout drawing and table on page 32.
In 2-wire mode, the 34933A module contains two matrices, each with 32 2-wire
crosspoint non-latching reed relays organized in a 4-row by 8-column
configuration. Every row and column are made up of two wires each, a high (H)
and a low (L). Each crosspoint relay has a unique channel number representing
the row and column that intersect to create the crosspoint. For example, channel
308 represents the crosspoint connection between row 3 and column 08 (all
columns consisting of two digits; in this case the digits are 08). See the simplified
schematic on page 30.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 2 in 2-wire mode of this module connects to the Analog
Buses. By closing channels 921 and 922 you can connect rows 5 and 6
respectively to the internal DMM of the 34980A mainframe for voltage and
resistance measurements.
In 2-wire mode, you can close no more than 20 channels simultaneously due to
power dissipation. However, note that Analog Bus relays count half as much as
channel relays in that total. For example, with one Analog Bus relay closed, you
can close up to a maximum of 19 channel relays. If you try to close more than the
allowed number of channels, you will receive an error message.
Keysight 34931A-34933A User’s Guide
29
One-Wire Mode
To physically configure the module in 1-wire mode, use the 34933T-002 terminal
block, or a compatible standard or custom cable. If using a standard or custom
cable, make sure you connect interlock pins 17 and 33 on the Matrix 2 D-sub
connector. Refer to the pinout drawing and table on page 37.
In 1-wire mode, the 34933A module contains four matrices (1 through 4), each
with 32 1-wire crosspoint non-latching reed relays organized in a 4-row by
8-column configuration. Every row and column has one wire each. Each
crosspoint relay has a unique channel number representing the matrix, and the
single-wire row and column that intersect to make the crosspoint. For example,
channel 218 represents Matrix 2, row 1 and column 8. See the simplified
schematic on page 35.
In 1-wire mode, you can close no more than 40 channels simultaneously due to
power dissipation. For example, with one Analog Bus relay closed you can close
up to a maximum of 39 channel relays. If you try to close more than the allowed
number of channels, you will receive an error message.
You can connect any combination of inputs and outputs at the same time.
However, only Matrix 3 and Matrix 4 in 1-wire mode of this module connect to the
Analog Buses. By closing channels 921 and 922 you can connect rows 1 and rows
2 respectively to the internal DMM of the 34980A mainframe for voltage and
resistance measurements.
You can connect multiple matrix modules externally and/or through the Analog
Buses for applications that require large matrices. For information on linking
multiple matrix modules, refer to page 15 of this manual.
When the power is off, matrix relays and Analog Bus relays open.
Keysight 34931A-34933A User’s Guide
34933A Simplified Schematic for Two-Wire Mode
C2L
C2L
bypass
L
C8L
bypass
L
C8L
C2H
C2H
bypass
H
H
L
L
L
L
H
H
H
C1H
C1H
bypass
H
C1L
C1L
bypass
L
C8H
C8H
bypass
H
C1L
C1L
bypass
L
C8H
C8H
bypass
H
C2L
C2L
bypass
L
C8L
bypass
L
C8L
H
L
L
L
L
H
H
H
C1H
C1H
bypass
H
C2H
C2H
bypass
H
H
H
L
L
Matrix 1
Matrix 2
Col 1H
Col 1L
Col 2H
Col 2L
Col 8H
Col 8L
Col 1H
Col 1L
Col 2H
Col 2L
Col 8H
Col 8L
Row 1
Row 2
Row 3
Row 4
Row 5
Row 6
Row 7
Row 8
ABus1
DMM
(MEAS)
ABus2
DMM
(SENS)
ABus3
ABus4
922
923
924
921
H
H
H
H
L
L
L
L
Analog Buses
NOTE: Three-digit channel numbers are derived from the
intersection of the rows and columns, columns having two
digits. The intersection shown here represents Channel
308 (Row 3, Column 8).
NOTE: Although columns are
numbered the same on Matrix 1
and Matrix 2, they are electrically
separate from one another.
NOTE: All series resistors shown
are 100Ω.
NOTE:
Matrix Relays: Reed non-latching
Analog Bus Relays: Armature
non-latching
Keysight 34931A-34933A User’s Guide
31
34933A D-Sub Connectors for Two-Wire Mode
Matrix 1
For orientation, the D-sub connector
end of the module is facing you.
Matrix 1
Matrix 2
NOTE:
–
In this diagram and the
table below, R represents
“row,” and C represents
“column.”
–
Bypass” means to bypass
the 100Ω in-rush resistor
that protects the reed
relays.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
NC
C4H
C4L
R4H
R4L
C5H
C5L
NC
NC
C7H
C7L
C4H
bypass
C4L
bypass
C5H
bypass
C5L
bypass
C7H
bypass
C7L
bypass
C2L
NC
C2H
R2H
R2L
C8H
C8L
NC
NC
GND
C3H
bypass
C1L
bypass
C1H
bypass
C3L
bypass
C6L
bypass
C6H
bypass
C6L
NC
C3H
C3L
C1H
C1L
R3H
R3L
C6H
NC
R1H
R1L
NC
C2H
bypass
C2L
bypass
C8H
bypass
C8L
bypass
50-Pin D-Sub Male Connector
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
R1H
49
C1H
37
C5H
7
C1H bypass 21
C5H bypass 9
GND
33
R1L
50
C1L
38
C5L
8
C1L bypass
22
C5L bypass
10
No Connect pins:
11-12, 17-18,
31-32, 34, and
45-46
R2H
27
C2H
23
C6H
41
C2H bypass 43
C6H bypass 25
R2L
28
C2L
24
C6L
42
C2L bypass
44
C6L bypass
26
R3H
39
C3H
35
C7H
13
C3H bypass 19
C7H bypass 15
R3L
40
C3L
36
C7L
14
C3L bypass
20
C7L bypass
16
R4H
5
C4H
1
C8H
29
C4H bypass 3
C8H bypass 47
R4L
6
C4L
2
C8L
30
C4L bypass
4
C8L bypass
48
all-guides.co
m
Keysight 34931A-34933A User’s Guide
Matrix 2
NOTE:
–
In this diagram and the
table below, R represents
“row,” and C represents
“column.”
–
“Bypass” means to
bypass the 100Ω in-rush
resistor that protects the
reed relays.
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
C2L
NC
C2H
R6H
R6L
C8H
C8L
NC
NC
Interlock
C3H
bypass
C1L
bypass
C1H
bypass
C3L
bypass
C6L
bypass
C6H
bypass
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
Interlock
C4H
C4L
R8H
R8L
C5H
C5L
NC
NC
C7H
C7L
C4H
bypass
C4L
bypass
C5H
bypass
C5L
bypass
C7H
bypass
C7L
bypass
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
C6L
NC
C3H
C3L
C1H
C1L
R7H
R7L
C6H
NC
R5H
R5L
NC
C2H
bypass
C2L
bypass
C8H
bypass
C8L
bypass
50-Pin D-Sub Male Connector
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
R5H
49
C2H
23
C7H
13
C4H bypass
3
Interlock
17
R5L
50
C2L
24
C7L
14
C4L bypass
4
Interlock
33
R6H
27
C3H
35
C8H
29
C5H bypass
9
No Connect pins:
11-12, 18, 31-32, 34,
and 45-46
R6L
28
C3L
36
C8L
30
C5L bypass
10
R7H
39
C4H
1
C1H bypass
21
C6H bypass
25
R7L
40
C4L
2
C1L bypass
22
C6L bypass
26
R8H
5
C5H
7
C2H bypass
43
C7H bypass
15
R8L
6
C5L
8
C2L bypass
44
C7L bypass
16
C1H
37
C6H
41
C3H bypass
19
C8H bypass
47
C1L
38
C6L
42
C3L bypass
20
C8L bypass
48
WARNING
As a safety feature, interlock pins (17 and 33) must be shorted to enable the
Analog Bus relays, which are on Matrix 2, to close. The optional 34933T-001
(for 2-wire) terminal block shorts these pins for you. This feature protects
inadvertent routing of high voltages from the Analog Bus to the D-sub
connector of the module.
Keysight 34931A-34933A User’s Guide
33
34933T-001 Terminal Block for Two-Wire Mode
This terminal block with screw-type connections is labeled with the model number
and the abbreviated module name. In addition, space is available on the label for
you to write the slot number.
NOTE
All modules that connect to the internal DMM are interlock protected. This
means that when an installed module is exposed (no terminal block or cable is
connected), the Analog Bus relays, which are on Matrix 2, are open and
disconnected from the Analog Buses. See page 10 for further information.
NOTE
If you are using an Keysight terminal block to connect your DUT to this module
be sure to use the 34933T-001 terminal block that corresponds to the 2-wire
configuration mode. Note that an error will not be generated if you have installed
a terminal block that doesn't match the present module configuration.
Keysight 34931A-34933A User’s Guide
When using the 34933T terminal block for 2-wire mode, access is provided to the
bypass columns through the columns labeled C9 through C16. Follow this wiring
convention shown in the table below for both matrices.
Although columns are numbered
the same on Matrix 1 and Matrix 2,
they are electrically separate from
one another (e.g., Col C2).
COLUMN
Wire Size:
20 AWG Typical
18 AWG Max
Terminal marked...
Connects to...
Terminal marked...
Connects to...
C9H
C1Hbypass
C13H
C5H bypass
C9L
C1L bypass
C13L
C5L bypass
C10H
C2H bypass
C14H
C6H bypass
C10L
C2L bypass
C14L
C6L bypass
C11H
C3H bypass
C15H
C7H bypass
C11L
C3L bypass
C15L
CC7L bypass
C12H
C4H bypass
C16H
C8H bypass
C12L
C4L bypass
C16L
C8L bypass
Keysight 34931A-34933A User’s Guide
35
34933A Simplified Schematic for One-Wire Mode
H
H
H
H
ABus1
DMM
(MEAS)
921
ABus2
DMM
(SENS)
922
ABus3
923
ABus4
924
H
H
L
L
H
H
L
L
H
H
H
H
2C2L
2C2
bypass
2C8
2C8
bypass
2C1
2C1
bypass
L
L
H
H
1C2
1C2
bypass
1C8
1C8
bypass
1C1
1C1
bypass
3C2
3C2
bypass
3C8
3C8
bypass
3C1
3C1
bypass
4C2
4C2
bypass
4C8
4C8
bypass
4C1
4C1
bypass
Matrix 1
Matrix 3
Matrix 4
Matrix 2
Row 1
Row 2
Row 3
Row 4
Row 1
Row 2
Row 3
Row 4
Row 1
Row 2
Row 3
Row 4
Row 1
Row 2
Row 3
Row 4
Analog Buses
Row 1
Row 1
Row 2
Row 2
Row 3
Row 3
Row 4
Row 4
L
L
L
L
L
L
L
L
NOTE:
Matrix Relays: Reed non-latching
Analog Bus Relays: Armature non-latching
NOTE: Although rows are numbered the
same across the matrices, they are
electrically separate from one another.
Channel 218
(Matrix 2, Row 1, Column 8)
NOTE: All series resistors shown are 100Ω.
NOTE: Three-digit
channel numbers are
derived from a specific
matrix number and the
intersection of rows and
columns on that matrix.
The channel shown here
is 132 (Matrix 1, Row 3,
Column 2.)
Keysight 34931A-34933A User’s Guide
34933A D-Sub Connectors for One-Wire Mode
Matrices 1 and 2
For orientation, the D-sub connector
end of the module is facing you.
Matrices 3 & 4
Matrices 1 & 2
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
NC
1C4
2C4
1R4
2R4
1C5
2C5
NC
NC
1C7
2C7
1C4
bypass
2C4
bypass
1C5
bypass
2C5
bypass
1C7
bypass
2C7
bypass
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
2C2
NC
1C2
1R2
2R2
1C8
2C8
NC
NC
GND
1C3
bypass
2C1
bypass
1C1
bypass
2C3
bypass
2C6
bypass
1C6
bypass
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
2C6
NC
1C3
2C3
1C1
2C1
1R3
2R3
1C6
NC
1R1
2R1
NC
1C2
bypass
2C2
bypass
1C8
bypass
2C8
bypass
50-Pin D-Sub
Male Connector
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
1R1
49
1C2
23
1C7
13
1C4 bypass
3
GND
33
1R2
27
2C2
24
2C7
14
2C4 bypass
4
No connect pins:
11-12, 17-18, 31-32,
34, and 45-46
1R3
39
1C3
35
1C8
29
1C5 bypass
9
1R4
5
2C3
36
2C8
30
2C5 bypass
10
2R1
50
1C4
1
1C1 bypass
21
1C6 bypass
25
2R2
28
2C4
2
2C1 bypass
22
2C6 bypass
26
2R3
40
1C5
7
1C2 bypass
43
1C7 bypass
15
2R4
6
2C5
8
2C2 bypass
44
2C7 bypass
16
1C1
37
1C6
41
1C3 bypass
19
1C8 bypass
47
2C1
38
2C6
42
2C3 bypass
20
2C8 bypass
48
NOTE
Conventions for these drawings and tables as they relate to pinout information:
–
2R4 means Matrix 2, Row 4.
–
1C5 means Matrix 1, Column 5
–
4C2 bypass means: Matrix 4, Column 2, and the connection bypasses the
100Ω in-rush resistor that protects the reed relays
all-guides.co
m
Keysight 34931A-34933A User’s Guide
37
Matrices 3 and 4
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
Interlock
3C4
4C4
3R4
4R4
3C5
4C5
NC
NC
3C7
4C7
3C4
bypass
4C4
bypass
3C5
bypass
4C5
bypass
3C7
bypass
4C7
bypass
4C2
NC
3C2
3R2
4R2
3C8
4C8
NC
NC
Interlock
3C3
bypass
4C1
bypass
3C1
bypass
4C3
bypass
4C6
bypass
3C6
bypass
4C6
NC
3C3
4C3
3C1
4C1
3R3
4R3
3C6
NC
3R1
4R1
NC
3C2
bypass
4C2
bypass
3C8
bypass
4C8
bypass
50-Pin D-Sub
Male Connector
Description
Pin
Description
Pin
Description
Pin
Description
Pin
Description
Pin
3R1
49
3C2
23
3C7
13
3C4 bypass
3
Interlock
17
3R2
27
4C2
24
4C7
14
4C4 bypass
4
Interlock
33
3R3
39
3C3
35
3C8
29
3C5 bypass
9
No connect pins:
11-12, 18, 31-32, 34,
and 45-46
3R4
5
4C3
36
4C8
30
4C5 bypass
10
4R1
50
3C4
1
3C1 bypass
21
3C6 bypass
25
4R2
28
4C4
2
4C1 bypass
22
4C6 bypass
26
4R3
40
3C5
7
3C2 bypass
43
3C7 bypass
15
4R4
6
4C5
8
4C2 bypass
44
4C7 bypass
16
3C1
37
3C6
41
3C3 bypass
19
3C8 bypass
47
4C1
38
4C6
42
4C3 bypass
20
4C8 bypass
48
WARNING
As a safety feature, interlock pins (17 and 33) must be shorted to enable the
Analog Bus relays, which are on Matrix 2, to close. The optional 34933T-002
(for 1-wire) terminal block shorts these pins for you. This safety feature
protects inadvertent routing of high voltages from the Analog Buses to the
D-sub connector of the module.
Keysight 34931A-34933A User’s Guide
34933T-002 Terminal Block for One-Wire Mode
This terminal block with screw-type connections is labeled with the model number
and the abbreviated module name. In addition, space is available on the label for
you to write the slot number.
NOTE
All modules that connect to the internal DMM are interlock protected. This
means that when an installed module is exposed (no terminal block or cable is
connected), the Analog Bus relays and current channels are open and
disconnected from the Analog Buses. See page 10 for further information.
NOTE
If you are using an Keysight terminal block to connect your DUT to this module
be sure to use the 34933T-002 terminal block that corresponds to the 1-wire
configuration mode. Note that an error will not be generated if you have installed
a terminal block that doesn't match the present module configuration.
NOTE: Analog Bus
connections are
on Matrix 3 and
Matrix 4.
Wire Size:
20 AWG Typical
18 AWG Max
Keysight 34931A-34933A User’s Guide
39
Index
Numerics
34931A
channel numbering,
11
connector pinouts,
19
description,
17
linking multiple modules,
15
programming examples,
11
simplified schematic,
18
terminal block,
21
wiring log,
21
34932A
channel numbering,
11
connector pinouts,
25
description,
23
linking multiple modules,
15
programming examples,
11
simplified schematic,
24
terminal block,
27
wiring log,
27
34933A
channel numbering,
11
connector pinouts,
31, 36
description,
28
linking multiple modules,
15
programming examples,
11
simplified schematic,
30, 35
terminal block,
33, 38
wiring log,
33, 38
C
connector pinouts
34931A,
19
34932A,
25
34933A,
31, 36
D
D-sub pinouts
34931A,
19
34932A,
25
34933A,
31, 36
O
operating considerations,
10
P
pinouts
34931A,
19
34932A,
25
34933A,
31, 36
S
safety notices,
4
This information is subject to change
without notice. Always refer to the
Keysight website for the latest
revision.
© Keysight Technologies 2008-2019
Edition 3, July 1, 2019
Printed in Malaysia
*34980-90031*
34980-90031
www.keysight.com
all-guides.co