CONSTANT CLIMATE CHAMBER HPP COOLED INCUBATOR IPPPLUS 100% ATMOSAFE. MADE IN GERMANY. OPERATING INSTRUCTIONS HPP I PP

Manufacturer and customer service

Memmert GmbH + Co. KG Willi Memmert Straße 90-96 D-91186 Büchenbach Deutschland Phone: +49 (0)9122 925-0 Fax: +49 (0)9122 14585 E-mail: [email protected] Internet: www.memmert.com Customer service: Service hotline: +49 (0)9171 9792 911 Service fax: +49 (0)9171 9792 979 E-mail: [email protected] When contacting customer service, always quote the product serial number on the nameplate (see page 13 ). Shipping address for repairs: Memmert GmbH + Co. KG Kundenservice Willi-Memmert-Str. 90-96 DE-91186 Büchenbach Germany Please contact our customer service before sending appliances for repair or before returning equipment, otherwise, we have to refuse acceptance of the shipment. © 2018 MEMMERT GmbH + Co. KG D24030 | Date 11/2018 We reserve the right to make changes

D24030 | Date 11/2018

3 About this manual About this manual Purpose and target group This manual describes the assembly, function, transport and operation of constant climate chambers HPP and cooled incubators IPPplus. It is intended for use by trained personnel of the owner, who have the task of operating and/or maintaining the respective appliance. If you are asked to work on the appliance, read this manual carefully before starting. Familiar- ise yourself with the safety regulations. Only perform work that is described in this manual. If there is something you do not understand, or certain information is missing, ask your superior or contact the manufacturer. Do not do anything without authorisation. Versions The appliances are available in different configurations and sizes. If specific equipment fea- tures or functions are available only for certain configurations, this is indicated at the relevant points in this manual. The functions described in this manual refer to the latest firmware version. Due to individual configurations and sizes, illustrations in this manual may be slightly different from the actual appearance. Function and operation are identical. Other documents that have to be observed: ►For operation of the appliance with MEMMERT AtmoCONTROL, observe the separate software manual. To open the AtmoCONTROL software manual, click on “Help” in the AtmoCONTROL menu bar. ►for service and repair work (see page 60) – a separate service manual Storage and forwarding This instruction manual belongs with the appliance and should always be stored where persons working on the appliance have access to it. It is the responsibility of the owner to ensure that persons who are working or will work on the appliance are informed as to the whereabouts of this instruction manual. We recommend that it is always stored in a protected location close to the appliance. Make sure that the instruction manual is not damaged by heat or humidity. If the appliance is sold on or transported and then set up again at a different location, the operating instructions must go with it. You will find the current version of our operating manual as pdf file if you go to www.memmert.com/de/service/downloads/bedienungsanleitung/.

D24030 | Date 11/2018

Contents 1. For your Safety 6 1.1 Terms and signs used...........................................................................................................6 1.2 Product safety and dangers ................................................................................................7 1.3 Requirements of the operating personnel ..........................................................................7 1.4 Responsibility of the owner .................................................................................................7 1.5 Intended use ........................................................................................................................8 1.6 Changes and alterations ......................................................................................................8 1.7 Behaviour in case of malfunctions and irregularities ..........................................................8 1.8 Switching off the appliance in an emergency ....................................................................8 2. Construction and description 9 2.1 Construction ........................................................................................................................9 2.2 Description .........................................................................................................................10 2.3 Working range of constant climate chambers HPP ..........................................................10 2.4 Material..............................................................................................................................11 2.5 Electrical equipment ..........................................................................................................11 2.6 Connections and interfaces ...............................................................................................11 2.7 Designation (nameplate) ...................................................................................................13 2.8 Technical data ....................................................................................................................14 2.9 Applied directives and standards ......................................................................................15 2.10 Declaration of conformity .................................................................................................15 2.11 Ambient conditions ...........................................................................................................16 2.12 Scope of delivery ...............................................................................................................16 2.13 Optional accessories ..........................................................................................................16 3. Delivery, transport and setting up 17 3.1 Safety regulations ..............................................................................................................17 3.2 Delivery ..............................................................................................................................18 3.3 Transport ............................................................................................................................18 3.4 Unpacking .........................................................................................................................18 3.5 Storage after delivery ........................................................................................................18 3.6 Setting up ..........................................................................................................................19 4. Putting into operation 23 4.1 Connecting the appliance .................................................................................................23 4.2 Fill up and connect the water tank ...................................................................................23 4.3 Switching on ......................................................................................................................24 5. Operation and control 25 5.1 Operating personnel..........................................................................................................25 5.2 Opening the door ..............................................................................................................25 5.3 Loading the appliance .......................................................................................................26 5.4 Operating the appliance ....................................................................................................26 5.5 Monitoring function ..........................................................................................................33 5.6 Graph .................................................................................................................................38 5.7 Ending operation ...............................................................................................................39 6. Malfunctions, warning and error messages 40 6.1 Warning messages of the monitoring function ................................................................40 6.2 Malfunctions, operating problems and appliance errors ................................................41 6.3 Power failure ......................................................................................................................43 Contents

D24030 | Date 11/2018

5 Contents 7. Menu mode 44 7.1 Overview ............................................................................................................................44 7.2 Basic operation in menu mode using the example of language selection .......................45 7.3 Setup..................................................................................................................................46 7.4 Date and Time ...................................................................................................................50 7.5 Calibration .........................................................................................................................51 7.6 Program .............................................................................................................................56 7.7 Sound ................................................................................................................................57 7.8 Protocol .............................................................................................................................58 7.9 User-ID ...............................................................................................................................59 8. Maintenance and service 60 8.1 Cleaning .............................................................................................................................60 8.2 Regular maintenance.........................................................................................................61 8.3 Repairs and service ............................................................................................................61 9. Storage and disposal 62 9.1 Storage ..............................................................................................................................62 9.2 Disposal .............................................................................................................................62 Index 63

D24030 | Date 11/2018

For your Safety 1. For your Safety 1.1 Terms and signs used In this manual and on the appliance itself, certain common terms and signs are used to warn you of possible dangers or to give you hints that are important in avoiding injury or damage. Observe and follow these hints and regulations to avoid accidents and damage. These terms and signs are explained below. 1.1.1 Terms used "Warning" is used whenever you or somebody else could be injured if you do not observe the accompanying safety regulation. "Caution" is used for information that is important for avoiding damage. 1.1.2 Signs used Warning signs (warning of a danger) Danger of electrocution Danger of explosion Dangerous gases / vapours Danger of top- pling over Hazard area! Observe the opera- ting instructions Prohibition signs (forbidding an action) Do not lift Do not tilt Do not enter Regulation signs (stipulating an action) Disconnect the mains plug Wear gloves Wear safety boots Observe information in separate manual Other icons Important or useful additional information

D24030 | Date 11/2018

7 For your Safety 1.2 Product safety and dangers The appliances described in this manual are technically sophisticated, manufactured using high-quality materials and subject to many hours of testing in the factory. They contain the latest technology and comply with recognised technical safety regulations. However, there are still risks involved, even when the appliances are used as intended. These are described below. Warning! After removing covers, live parts may be exposed. You may receive an electric shock if you touch these parts. Disconnect the mains plug before removing any covers. Only electrical technicians may work on the electrical equipment of the appliances. Warning! When loading the appliance with an unsuitable load, poisonous or explosive vapours or gases may be produced. This could cause the appliance to explode, and people could be severely injured or poi- soned. The appliance may only be loaded with materials/test objects which do not form any toxic or explosive vapours when heated up (see also chapter Intended use on page 8). Warning! If the door is open while the appliance is in operation, the appliance may overheat and pose a fi re hazard. Do not leave the door open during operation. Warning! In case of appliances of a certain size, you can get accidentally locked in, which is life-threatening. Do not climb into the appliance! 1.3 Requirements of the operating personnel The appliance may only be operated and maintained by persons who are of legal age and have been instructed accordingly. Personnel who are to be trained, instructed or who are un- dergoing general training may only work with the appliance under the continuous supervision of an experienced person. Repairs may only be performed by qualified electricians. The regulations in the separate service manual must be observed. 1.4 Responsibility of the owner The owner of the appliance ►is responsible for the flawless condition of the appliance and for it being operated in ac- cordance with its intended use (see page 8); ►is responsible for ensuring that persons who are to operate or service the appliance are qualified to do this, have been instructed accordingly and are familiar with the operating instructions at hand; ►must know about the applicable guidelines, requirements and operational safety regula- tions, and train staff accordingly;

D24030 | Date 11/2018

For your Safety ►is responsible for ensuring that unauthorised persons have no access to the appliance; ►is responsible for ensuring that the maintenance plan is adhered to and that maintenance work is carried out properly (see page 60); ►has to ensure that the appliance and its surroundings are kept clean and tidy, for example through corresponding instructions and inspections; ►is responsible for ensuring that personal protective clothing is worn by operating person- nel, e.g. work clothes, safety shoes and protective gloves. 1.5 Intended use Constant climate chambers HPP and cooled incubators IPPplus may be used exclusively for temperature and climate testing of materials and substances in the context of the procedures and specifications described in this manual. Any other use is improper, and may result in hazards and damage. The appliance is not explosion-proof (does not comply with the German workplace health & safety regulation VBG 24). The appliance may only be loaded with materials and substances which cannot form any toxic or explosive vapours at the set temperature and which cannot explode, burst or ignite. The appliance may not be used for drying, vaporising and branding paints or similar materi- als the solvents of which could form an explosive mixture when combined with air. If there is any doubt as to the composition of materials, they must not be loaded into the appliance. Potentially explosive gas-air mixtures must not form, neither in the working chamber nor in the direct vicinity of the appliance. 1.6 Changes and alterations No unauthorised changes or alterations may be made to the appliance. No parts may be added or inserted which have not been approved by the manufacturer. Unauthorised changes or alterations result in the EC declaration of conformity losing its valid- ity, and the appliance may no longer be operated. The manufacturer is not liable for any damage, danger or injuries that result from unauthor- ised changes or alterations, or from non-observance of the regulations in this manual. 1.7 Behaviour in case of malfunctions and irregularities The appliance may only be used in a flawless condition. If you as the operator notice irregu- larities, malfunctions or damage, immediately take the appliance out of service and inform your superior. You can find information on eliminating malfunctions from page 40. 1.8 Switching off the appliance in an emergency Push the On/Off switch on the control panel (Fig. 1) and disconnect power plug. This disconnects the appliance from the power supply at all poles. ON ONN ON O Fig. 1 Switch off the appliance by pressing the On/ Off switch

D24030 | Date 11/2018

9 Construction and description 2. Construction and description 2.1 Construction 1 2 3 4 5 6 7 8 9 Fig. 2 Construction 1 ControlCOCKPIT with capacitive function keys (see page 27) 2 On/Off switch (see page 24) 3 Working chamber fans 4 Steel grid 5 Working chamber 6 Nameplate (see page 13) 7 Door handle (see page 25) 8 Turn control with confirmation key 9 USB interface (see page 12)

D24030 | Date 11/2018

Construction and description 2.2 Description The appliance can heat the working chamber up to 70 ºC and cool it down to 5 ºC. Low- noise, long-life and energy-saving Peltier cooling and heating technology is used for this. In heating operation, a part of the required energy is extracted from the surroundings (heat pump principle). Additionally, the humidity in the interior of constant climate chambers HPP can be regulated between 10 and 90 % rh (rh = relative humidity). The humidity is increased by water from a tank being evaporated and fed into the interior, and is reduced through condensation on a Peltier module. Optionally, the appliance can be equipped with a light module, making it possible to adjust the interior lighting in steps of 1 %. 2.3 Working range of constant climate chambers HPP The temperature-humidity diagram (Fig. 3) specifies at what range of temperature and humid- ity a permanent, condensation-free operation of the constant climate chamber HPP is possible. Caution: If in operation for long periods at the upper level or outside the working range, puddles of water may form inside the chamber and water may force its way out of the door seal. 0 10 20 30 40 50 60 70 80 90 100 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 Temperature °C Humidity % rH C A B Fig. 3 Temperature-humidity working range of constant climate chambers HPP (at an am- bient temperature of 22 °C ± 3 K, relative humidity < 50 %)

D24030 | Date 11/2018

11 Construction and description Range A: In this range, temperature and humidity can be combined as you please, without resulting in any significant condensation. If there are extreme ambient conditions, the working range may be restricted. Range B: If the specified range is exceeded upwards, e.g. 80 % rh at 60°C, the hot steam fed in will im- mediately condense, due to the dew-point, at the coldest point in the appliance. Range C: At low temperatures and low relative air humidity, the effective range is heavily dependent on the degree of humidity of the chamber load. 2.4 Material For the outer housing, MEMMERT deploys stainless steel (Mat.No. 1.4016 – ASTM 430) and for the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through its high stability, optimal hygienic properties and corrosion-resistance towards many (but not all!) chemical compounds (caution for example with chlorine compounds). The chamber load for the appliance must be carefully checked for chemical compatibility with the materials mentioned. A material resistance table can be requested from the manufacturer. 2.5 Electrical equipment ►Operating voltage and current consumption: See nameplate ►Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010 ►Protection type IP 20 acc. to EN 60 529 ►Interference suppression acc. to EN 55011 class B ►Appliance fuse: Fusible link 250 V/15 A quick-blow ►The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V) 2.6 Connections and interfaces 2.6.1 Electrical connection This appliance is intended for operation on an electrical power system with a system imped- ance Zmax of a maximum of 0.292 ohm at the point of transfer (service line). The operator must ensure that the appliance is operated only on an electrical power system that meets these requirements. If necessary, you can ask your local energy supply company what the system impedance is. Observe the country-specific regulations when connecting (e.g. in Germany DIN VDE 0100 with residual current circuit breaker).

D24030 | Date 11/2018

Construction and description 2.6.2 Communication interfaces The interfaces are intended for appliances which meet the requirements of IEC 60950-1. USB interface The appliance is fitted by default with a USB interface in accordance with the USB specifi- cation. This way, you can ►transfer software stored on a USB stor- age medium to the appliance (see page 56). ►export protocol logs from the appliance to a USB storage medium (see page 58). ►transfer user ID data stored on a USB storage medium to the appliance (see page 59). The USB port can be found on the right side of the ControlCOCKPIT (Fig. 4). Ethernet interface Via Ethernet interface, the appliance can be connected to a network, so that you can transfer programmes created with AtmoCONTROL software to the appliance and read out protocol logs. The Ethernet interface is located on the rear of the appliance (Fig. 5). For identification purposes, each appliance connected must have its own unique IP ad- dress. Setting the IP address is described on page 46 . You will find a description of how to transfer programmes via Ethernet in the en- closed AtmoCONTROL manual. With an optional USB to Ethernet converter, the appliance can be directly connected to a com- puter / laptop (see Optional accessories on page 16). Fig. 4 USB interface Fig. 5 Ethernet interface

D24030 | Date 11/2018

13 Construction and description 2.7 Designation ( nameplate) The nameplate (Fig. 6) provides information about the appliance model, manufacturer and technical data. It is attached to the front of the appliance, on the right behind the door (see page 9). Typ: HPP 260 F.-Nr.: 0109.0088 230 V~ 2.3 A 50/60 Hz 820 W DIN12880-Kl.3.1 Nenntemp.: 70 °C 1 2 3 4 5 6 10 8 9 7 Fig. 6 Nameplate (example) 1 Type designation 2 Operating voltage 3 Applied standard 4 Protection type 5 CE conformity 6 Address of manufacturer 7 Disposal note 8 Temperature range 9 Connection / power ratings 10 Appliance number

D24030 | Date 11/2018

Construction and description 2.8 Technical data Appliance size 30 55 110 260 400 750 1060 Appliance width D1 [mm] 586 586 745 825 825 1225 1225 Appliance height E1 [mm] 703 783 863 1182 1720 1720 1720 Appliance depth F1 (footprint) [mm] 506 586 656 756 756 856 1107 Depth of door lock [mm] 56 Working chamber width A1 [mm] 400 400 560 640 640 1040 1040 Working chamber height B1 [mm] 320 400 480 800 1200 1200 1200 Working chamber depth C1 [mm] 250 330 400 500 500 600 850 Chamber volume [litres] 32 53 108 256 384 749 1060 Weight [kg] 62 74 100 121 213 284 424 Power [W] 140 275 650 820 1300 1500 1600 Current consumption [A] 230 V, 50/60 Hz 0.6 1.2 2.8 3.6 5.7 6.5 7.0 115 V, 50/60 Hz 1.2 2.4 5.7 7.1 11.4 13 13.9 max. number of sliding shelves 3 4 5 9 14 14 14 max. load per sliding shelve [kg] 20 30 20 max. load per appliance [kg] 60 80 150 200 200 200 200 Temperature Adjustment range 0 bis +70 °C2 Adjustment precision 0.1 K Humidity (HPP only) Adjustment range – 10 … 90 % rh 1 see Fig. 7 2 The minimum temperature depends on the outside temperature (see ambient conditions on page 16). If the interior is illuminated, the temperature range is further restricted.

D24030 | Date 11/2018

15 Construction and description )3 9( )3 9( B E 256 D A F 10 319 352 30 C 56 Fig. 7 Dimensions 2.9 Applied directives and standards ►Directive 2004/108/EC amended (Directive of the council on harmonisation of the laws of the member states on electromagnetic compatibility). Fulfilled standards: DIN EN 61326:2004-05, EN 61326:1997, EN 61326/A1:1998, EN 61326/A2:2001 EN 61326/A2:2003 ►Directive 2006/95/EC amended (Directive of the council on harmonisation of the laws of member states relating to electrical equipment designed for use within certain voltage limits). Standards complied with: DIN EN 61 010-1 (VDE 0411 Part 1):2002-08, DIN EN 61 010-2-010 (VDE 0411 Part 2-010):2004-06, EN 61 010-1:2001, EN 61 010-2-010:2003 2.10 Declaration of conformity You can download the EC declaration of conformity of the appliance online: English: http://www.memmert.com/en/service/downloads/ce-statement/ German: http://www.memmert.com/de/service/downloads/eg-konformitaetserklaerung/

D24030 | Date 11/2018

Construction and description 2.11 Ambient conditions ►The appliance may only be used in enclosed rooms and under the following ambient conditions: Ambient temperature 16 ºC to 40 ºC Humidity rh max. 70 %, non-condensing Overvoltage category II Pollution degree 2 Altitude of installation max. 2,000 m above sea level ►The appliance may not be used in areas where there is a risk of explosion. The ambient air must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is not explosion-proof. ►Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to sedimentation in the interior and, as a consequence, could result in short circuits or dam- age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust or aggressive vapours from developing should be taken. 2.12 Scope of delivery ►Power cable ►Tilt protection ►Sliding grid (load capacity 30 kg each) ►USB storage medium with software and AtmoCONTROL manual ►The operating instructions at hand ►Calibration certificate In addition for constant climate chambers HPP ►Water tank with connection hose ►Tank holder (only for appliances of size 400/750/1060, see page 24) 2.13 Optional accessories ►Converter USB to Ethernet (Fig. 8). Makes it possible to connect the appliance's network interface (see page 12) to the USB port of a computer / laptop. ►Reinforced, sliding steel grids with a load capacity of 60 kg each (for appliance size 110 and larger) Fig. 8 Converter USB to Ethernet

D24030 | Date 11/2018

17 Delivery, transport and setting up 3. Delivery, transport and setting up 3.1 Safety regulations Warning! Because of the heavy weight of the appliance, you could injure your- self if you try to lift it. To carry appliances of the sizes 30 and 55, at least two persons, for appliances of the sizes 110 and 260, four persons are needed. Appliances larger than that may not be carried but must be transported with a manual pallet jack or forklift truck. 30 55 110 260 400/750/1060 Warning! You may get your hands or feet squashed when transport- ing and installing the appliance. Wear protective gloves and safety boots. When grasping the bottom of the appliance, grasp it only on the sides: 9 Warning! The appliance could fall over and seriously injure you. Never tilt the appliance and transport it in upright position and without load only (except for standard accessories such as steel grids or shelves). Appliances with castors always have to be moved by two people.

D24030 | Date 11/2018

Delivery, transport and setting up 3.2 Delivery The appliance is packed in cardboard and is delivered on a wooden palette. 3.3 Transport The appliance can be transported in three ways: ►With a forklift truck; move the forks of the truck entirely under the pallet ►On a manual pallet jack ►On its own castors, in case of the corresponding configuration, for which the catch on the (front) castors must be released 3.4 Unpacking To avoid damage, do not unpack the appliance until you reach the installation site. Remove the cardboard packaging by pulling it upwards or carefully cutting along an edge. 3.4.1 Checking for completeness and transport damage ►Check the delivery note to ensure that the delivery is complete. ►Check the appliance for damage. If you notice deviations from the delivery note, damage or irregularities, do not put the appli- ance into operation but inform the haulage company and the manufacturer. 3.4.2 Removing the transport protection Remove the transport protection. It is located between the door hinge, door and frame and has to be removed after opening the door. 3.4.3 Disposing of packaging material Dispose of the packaging material (cardboard, wood, foil) in accordance with the applicable disposal regulations for the respective material in your country. 3.5 Storage after delivery If the appliance is first to be stored after delivery: Read the storage conditions from page 62.

D24030 | Date 11/2018

19 Delivery, transport and setting up 3.6 Setting up Warning! Due to its centre of gravity, the appliance can fall over to the front and injure you or other people. Always attach the appliance to a wall with the tilt protection (see page 21). If this cannot be done due to space problems, do not operate the appliance and do not open the door. Contact the Memmert service team (see page 2). 3.6.1 Prerequisites The installation site must be flat and horizontal and must be able to reliably bear the weight of the appliance (see "Technical data" on page 14). Do not place the appliance on a flam- mable surface. Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail- able at the installation site. The distance between the wall and the rear of the appliance must be at least 15 cm. The clear- ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm (Fig. 9). Sufficient air circulation in the vicinity of the appliance must be guaranteed at all times. For appliances with castors, these need to be positioned in forward direction at all times. •FP •FP •FP •FP Fig. 9 Minimum clearance from walls and ceiling

D24030 | Date 11/2018

Delivery, transport and setting up 3.6.2 Installation options Setting up Comments Suitable for appliance size ... 30 55 110 260 400 750 1060 Floor      Table Check the load capacity first      Stacked two appliances maximum; mount- ing material (feet) provided      Wall mounting Separately packaged fastening material is included in the scope of delivery. Observe the assembly instructions provided.      Base with/without castors      Castor frame      Height adjustable feet     

D24030 | Date 11/2018

21 Delivery, transport and setting up 3.6.3 Tilt protection Attach the appliance to a wall with the tilt protection. The tilt protection is included in the delivery. 1. As illustrated, fasten the tilt protection to the rear side of the appliance. 2. Bend the tilt protec- tion upwards by 90 ° in the desired distance to the wall (consider the minimum distance to the wall, see Fig. 9). 3. Drill a hole, insert a plug and screw the tilt protec- tion to a suitable wall.

D24030 | Date 11/2018

Delivery, transport and setting up 3.6.4 Adjusting the doors (only for model sizes 400/750/1060) For model sizes 400/750/1060, it is possible to adjust doors that warp due to the floor condi- tions. In order to do so, every door has two adjuster screws at the top and at the bottom (Fig. 10). First, adjust the door at the top and then, if further adjustment is necessary, at the bottom as well. 1. Open the door. 2. Undo the screws. 3. Adjust the door. 4. Tighten the screws again. 5. Check door alignment. 6. If necessary, readjust. Fig. 10 Door adjustment screws

D24030 | Date 11/2018

23 Putting into operation 4. Putting into operation Caution: The first time the appliance is operated, it must not be left unattended until it has reached the steady state. 4.1 Connecting the appliance Caution: Observe the country-specific regulations when mak- ing connections (e.g. in Germany DIN VDE 0100 with residual current circuit breaker). Observe the connec- tion and power ratings (see nameplate and Technical data on page 14). Make sure to establish a safe PE conductor connection. Plug the provided power cable into the rear of the appli- ance and connect it to the power supply ( Fig. 11 ). Lay the power cable so that ►it is always accessible and within reach so it can be disconnected quickly in the event of failure or emergencies; ►no one can trip over it ►it does not come into contact with any hot parts. 4.2 Fill up and connect the water tank (for constant climate chambers HPP only) Water specifications For steam creation, use only: demineralised/desalinated water (Aqua Dem) with a conduc- tivity ≤ 10 μS/cm. Otherwise, possible calcification in the steam generators, steam piping and steam generator pumps could affect the correct functioning of the appliance. Only use chlorine-free water with a pH value of > 5 and < 7. Fig. 11 Connect the power cable to the rear of the appliance

D24030 | Date 11/2018

Putting into operation Fill the supplied water tank with water and connect it with the enclosed tube to the „H2O“ connection on the rear of the chamber (Fig. 12). For appliances of sizes 400/750/1060, the tank can be attached to the appliance with the included tank holder (Fig. 13). To do so, hook the tank holder into the slots on the rear of the appliance. For wall mounting, the tank holder also has two drill holes (fastening material not included in the delivery). Fig. 12 Water connection Fig. 13 Tank holder 4.3 Switching on Switch on the appliance by pressing the On/Off switch on the front of the appliance ( Fig. 14 ). The starting process is shown by three animated white dots . If the dots have another colour, an error has occurred (see page 42). After the first start-up, the appliance display is set to English by default. You can change the language as described from page 45. However, to get a basic overview of operating the appliance, you should read the following chapter first. ON ONN ON O Fig. 14 Switching on the ap- pliance

D24030 | Date 11/2018

25 Operation and control 5. Operation and control 5.1 Operating personnel The appliance may only be operated by persons who are of legal age and have been instruct- ed accordingly. Personnel who are to be trained, instructed or who are undergoing general training may only work with the appliance under the continuous supervision of an experi- enced person. 5.2 Opening the door ►To open the door, pull the door handle to the side (to the left or to the right, depending on the door variation, see Fig. 15, A) and open the door completely. ►To close the appliance, push the door closed and the door handle to the side (B). A B Fig. 15 Opening and closing the door Warning! If the door is open while the appliance is in operation, the appliance may overheat and pose a fi re hazard. Do not leave the door open during operation. Warning! In case of appliances of a certain size, you can get accidentally locked in, which is life-threatening. Do not climb into the appliance!

D24030 | Date 11/2018

Operation and control 5.3 Loading the appliance Warning! When loading the appliance with an unsuitable load, poisonous or explosive vapours or gases may be produced. This could cause the appliance to explode, and people could be severely injured or poi- soned. The appliance may only be loaded with materials which do not form any toxic or explosive vapours when heated up and cannot ignite (see also Intended use on page 8). If there is any doubt as to the composition of materials, they must not be loaded into the appliance. Caution: Check the chamber load for chemical compatibility with the materials of the appliance (see page 11). Insert the sliding steel grids or sliding shelves. The maximum number or grids / shelves and the load capacity are specified in the technical data overview from page 14 . The chamber must not be loaded too tightly, so that proper air circulation in the working chamber is guaran- teed. Do not place any of the chamber load on the floor, touching the side walls or right below the ceiling of the working chamber (Fig. 16, see also the "correct loading" sticker on the appliance). In case of improper loading (chamber loaded too tightly), the set temperature may be exceeded or it may take longer until it is reached. To achieve the correct heating capacity, the type of slide-in unit used – Grid or Shelf – must be set in the menu under SETUP (see page 48). 5.4 Operating the appliance 5.4.1 ControlCOCKPIT In manual operation, the desired parameters are entered at the ControlCOCKPIT on the front of the appliance (Fig. 17 and Fig. 18 ). You can also make basic settings here (menu). Addi- tionally, warning messages are displayed, e.g. if the temperature is exceeded. In programme mode, the parameters defined, the programme description, the programme segment cur- rently active and programme duration remaining are displayed (for a more detailed descrip- tion, see page 31). Fig. 16 Correct placement of the chamber load

D24030 | Date 11/2018

27 Operation and control ON ONN ON O TIMER End 14:45 44h:44m Holz trocknen aufheizen 09:12h Fr 20.10.2010 20:31 min 000°C ALARM of °C max 000°C auto off 99K - + 0 12 %rh °C GRAPH off ON LIGHT ON LIG LIG LIGHT HT HT ON LIGHT ON ONN ON O HUMIDITY 30.0%rh Set 30.0%rh 22.4 37.0 TEMP Set °C °C °C 22.4 TEMP °C Set 37.0 LIGHT % 100 ALARM max 38.5°C auto +/- 0.0K min 35.5°C Manual Mode 12.Sept.2012 13:44 3 11 12 13 14 1 2 17 18 9 10 15 16 4 5 6 7 8 TIMER 30m 04h End 23.Nov 13:30 Fig. 17 ControlCOCKPIT for HPP appliances in operating mode (width may differ depending on appliance size) ON ONN ON O TIMER End 14:45 44h:44m Holz trocknen aufheizen 09:12h Fr 20.10.2010 20:31 min 000°C ALARM of °C max 000°C auto off 99K - + 0 12 %rh °C GRAPH off 180.4°C TEMP Set 180.4°C ON LIGHT ON LIG LIG LIGHT HT HT ON LIGHT ON ONN ON O LÜFTER % 0 °C 22.4 TEMP °C Set 37.0 TIMER 30m 04h Ende 23.Nov 13:30 LIGHT % 100 ALARM max 38.5°C +/- 2.0K min 36.5°C auto Manual Mode 12.Sept.2012 13:44 3 11 12 13 14 1 2 17 18 9 10 15 16 4 5 6 Fig. 18 ControlCOCKPIT for IPPplus appliances in operating mode (width may differ de- pending on appliance size) 1 Activation key for temperature setpoint adjustment 2 Setpoint and actual temperature display 3 Switch to menu mode (see page 44) 4 Activation key digital backwards counter with target time setting, adjustable from 1 minute to 99 days 5 On/Off Switch 6 Display digital backwards counter with target time setting, adjustable from 1 min- ute to 99 days 7 Humidity control display 8 Humidity control activation key 9 Turn control for setpoint adjustment 10 Confirmation key (accepts setting made with the turn control) 11 Interior lighting activation key (only for models with light module) 12 Interior lighting display (only for models with light module) 13 Appliance state and programme display 14 Activation key for the appliance state 15 Activation key for temperature monitoring 16 Monitoring display 17 Graphic representation 18 Activation key for graphic representation

D24030 | Date 11/2018

Operation and control 5.4.2 Basic operation In general, all settings are made according to the following pattern: 1. Activate the desired parameter (e.g. tem- perature). To do so, press the correspond- ing activation key on the left or right or the respective display. The activated dis- play is lined in colour, the other displays are dimmed. The set value is highlighted in colour. .5°C 100 TEMP TIMER -- -- m h 22 TEMP Set 2. By turning the turn control to the left or right, adjust the set value (e.g. to 37.0 ºC). 3. Save the set value by pressing the confir- mation key. The display returns to normal and the appliance begins adjusting to the defined set value. °C 22.4 TEMP °C Set 37.0 Additional parameters can be set accordingly. If no new values are entered or confirmed for approx. 30 seconds, the appliance automati- cally returns to the main menu and restores the former values. If you want to cancel the setting procedure, press the activation key on the left or right of the display that you want to exit. The appliance restores the former values. Only the settings that you have confirmed by pressing the confirmation key before cancelling the setting procedure are accepted. 5.4.3 Operating modes The appliance can be operated in different modes: ► Manual mode: The appliance runs in permanent operation at the values set on the Con- trolCOCKPIT. Operation in this mode is described in chapter 5.4.4 . ►Operation with digital backwards counter with target time setting, adjustable from 1 minute to 99 days (Timer): The appliance runs at the values set until the timer has elapsed. Operation in this mode is described in chapter 5.4.5 . ► Programme mode: The appliance automatically runs programme sequences which have been defined using AtmoCONTROL software at a computer / laptop and then transferred to the appliance from a USB stick or via Ethernet. Operation in this mode is described in chapter 5.4.6 . ►via remote control °C 22.4 TEMP °C 37.0 Set T T - T

D24030 | Date 11/2018

29 Operation and control The status display shows you which operating mode or operating state the appliance is currently in. The current operating state is highlighted in colour and indicated by the text display:  Appliance is in programme mode ■ Programme is stopped Appliance is in manual operating mode The example on the right shows the appliance in man- ual mode, identified by the coloured hand symbol. ►When the appliance is in timer mode, Timer active is displayed: ►When the appliance is in remote control mode, the symbol appears in the temperature display: 5.4.4 Manual mode In this operating mode, the appliance runs in permanent operation at the values set on the ControlCOCKPIT. Adjustment options As described in chapter 5.4.2 , you can set the following parameters after pressing the cor- responding activation key (in any sequence): Temperature Adjustment range: model dependent (see nameplate and technical date on page 14) Heating operation is indicated by the symbol. Cooling is indicated by the icon. You can select °C or °F as the temperature unit displayed (see page 47). °C 22.4 TEMP °C Set 37.0 The minimum temperature that can be reached depends on the surrounding conditions. The appliances reach up to to 20 °C below room temperature. In order to do so, the Peltier module needs sufficient ventilation (see Fig. 9 on page 19). Humidity (for constant climate chambers HPP only) Adjustment range: 10 to 90 % rh Humidifying is indicated by the symbol. Dehumidifying is indicated by the icon. HUMIDITY 70 %rh Set34% rh A high level of air humidity in the interior can only be achieved without condensation if the interior is thoroughly heated. For this reason, how fast the humidity is dynamically adjusted to approach the setpoint depends on the interior temperature. Manual Mode 12.Sept.2012 13:44 Timer active 12.Sept.2012 13:44 23.2°C TEMP Set 38 0°C

D24030 | Date 11/2018

Operation and control Interior lighting (only for models with light module) Adjustment range: 0 to 100 % in steps of 1% LIGHT % 40 The interior light is only active at temperatures of up to 40 °C. If this temperature is exceeded, the interior light switches off automatically. The light display will then show “Temp too high”. LIGHT % 0 Temp too high 5.4.5 Operation with digital backwards counter with target time setting, adjustable from 1 minute to 99 days ( Timer) In timer operation, you can adjust the time the appliance runs at the set value. The appliance has to be in manual operating mode for this. 1. Press the activation key to the left of the timer display. The timer display is activated. .5 C 100 Set 1 TIMER - Ende 9:00 23.11. -- h- m 2. Turn the turn control until the desired duration is displayed – in this example 4 hours 30 minutes. The approximate end time is shown beneath, in a smaller font. TIMER End 0 13:30 23.11. 04 m h 3 Up to a duration of 23 hours 59 minutes, the time is displayed in hh:mm (hours:minutes) format. For 24 hours and more, the format dd:hh (days:hours) is used. The maximum duration adjustable is 99 days 00 hours. 3. Press the confirmation key to confirm. The display now shows the remaining time in a large font and the approximate end time in a smaller font beneath. The status display shows „Timer active“. TIMER 30m 04h Ende 13:30 23.11. Timer active 12.Sept.2012 13:44 4. Now, as described under 5.4.2 , set the individual values for temperature, air flap position etc. which you want the appliance to operate at. The set values can be changed at any time during the timer. The changes are effective immediately.

D24030 | Date 11/2018

31 Operation and control In Setup, you can choose if the timer should run setpoint-dependent or not. This deter- mines whether the timer should not start until a tolerance band around the set tempera- ture is reached or if it should start right after activation (see page 48). If the timer runs setpoint-dependent, this is indicated by the symbol in the timer display. When the timer has elapsed, the display shows 00h:00m. All functions (heating etc.) are switched off. In addition, an acoustic alarm sounds, which can be turned off by pressing the confirma- tion key. To deactivate the timer, open the timer display by pressing the activation key again and then turning the turn control to reduce the timer setting until --:-- is displayed. Confirm with the confirma- tion key. 5.4.6 Programme mode In this operating mode, programmes saved in the appliance can be started with different combinations of individual parameters (temperature, humidity, interior lighting) at staggered intervals, which the appliance then automatically processes in sequence. These programmes are not created directly at the appliance but externally at a computer / laptop and using AtmoCONTROL software. Transfer to the appliance is possible using the provided USB storage medium or via Ethernet. A description of how to create and save programmes can be found in the separate AtmoCONTROL software manual. Starting a programme 1. Press the activation key to the right of the status display. The current operating mode is highlighted automatically, in this example manual mode ( ). manueller Betrieb Fr 20.10.2010 20:31 GRAPH GRAPH APH Manual mode 17:44 13.Sept.2012 Activate 2. Turn the turn control until the  start symbol is highlighted. The current programme is displayed, in this example Test 012. Test 012 ready 10:44 12.Sept.2012 Only the programme currently selected in the menu and shown in the display can be used. If you want to process another programme, you need to activate it in the menu first (description from page 56). TIMER 00m 00h End 13:30 23.11. TIMER --m --h End 9:00 23.11.

D24030 | Date 11/2018

Operation and control 3. To start the programme, press the confir- mation key. The programme starts. The display shows: ►the programme description (in this exam- ple Test 012) ►the programme segment description, in this example Ramp 1 ►the current run (in case of loops) Test 012 Ramp 1 10:44 12.Sept.2012 You cannot change any parameters (e.g. the temperature) at the appliance while a programme is running. However, the displays ALARM and GRAPH can still be used. Cancelling a programme You can cancel an active programme at any time. 1. Press the activation key to the right of the status display. The status display is automatically highlighted. 0 % manueller Betrieb % Fr 20.10.2010 20:31 0°C 80 GRAPH 80 GRAPH APH Test 012 Ramp 3 10:44 12.Sept.2012 2. Turn the turn control until the ■ stop symbol is highlighted. Cancel program Test 012 10:48 12.Sept.2012 3. Press the confirmation key to confirm. The programme is cancelled. End Test 012 10:49 12.Sept.2012 A cancelled programme cannot be resumed at the point it was cancelled. It must be restarted from the beginning. End of programme The display shows End when the programme is finished. End Test 012 10:49 12.Sept.2012

D24030 | Date 11/2018

33 Operation and control You can now ►restart the programme as described ►select another programme for processing in menu mode (see page 56) and run it as described. ►return to manual mode. To do so, reactivate it by pressing the activation key next to the status display, then turn the turn control until the hand symbol is highlighted in colour and press the confirmation key. 5.5 Monitoring function 5.5.1 Temperature monitoring The appliance is equipped with a multiple overtemperature protection in accordance with DIN 12 880. This serves to avoid damage to the chamber load and/or appliance in case of a malfunction: ► electronic temperature monitoring (TWW) ►automatic temperature monitor ( ASF) The monitoring temperature of the electronic temperature monitoring is measured via a separate Pt100 temperature sensor in the working chamber. Temperature monitoring settings are made via the ALARM display. The settings made apply to all operating modes. If temperature monitoring has been triggered, this is indicated by the temperature display: the actual tempera- ture is highlighted in red and a warning symbol is shown ( Fig. 19 ). The type of temperature monitoring triggered (TWW in this example) is shown beneath the temperature. If the acoustic alarm has been activated in the menu mode (Sound, see page 57, indicated by the speaker symbol in the alarm display), the alarm is additionally signalled by an intermittent acoustic signal, which can be turned off by pressing the confirmation key. Information on what to do in this case are provided in the chapter Malfunctions, warning and error messages on page 40. Before reading how to adjust temperature monitoring (from page 35), please read the description of the individual monitoring functions here. Manual Mode 12.Sept.2012 13:44 ALARM max 38.5°C 1.0K min 35.5°C auto +- End 14:45 44h:44m min 000 O LI O LI FEUCHTE 30.0%rh Set 30.0%rh °C 22.4 °C Set 37.0 TIMER 30m 04h Ende 23.Nov 13:30 °C TEMP TWW Set 38.5 °C38.9 TIMER FE 22.4 37.0 TEMP Set °C °C °C 22.4 TEMP °C Set 37.0 TIMER °C TEMP TWW Set 38.5 °C38.9 Fig. 19 Temperature monitoring triggered

D24030 | Date 11/2018

Operation and control Electronic temperature monitoring ( TWW) The manually set monitoring temperature min and max of the overtemperature control is monitored by an adjustable over/undertemperature controller (TWW) protection class 3.3 acc. to DIN 12 880. If the manually set monitoring temperature max is exceeded, the TWW takes over temperature control and begins to regulate the monitoring temperature (Fig. 20). t °C Setting MAX Set temperature Controller error Emergency operation Fig. 20 Schematic diagram of how TWW temperature monitoring works

D24030 | Date 11/2018

35 Operation and control Automatic temperature monitor ( ASF) ASF is a monitoring device that automatically follows the set temperature setpoint within an adjustable tolerance band (Fig. 21). The ASF – if switched on – is automatically activated as soon as the actual temperature value reaches 50 % of the set tolerance band of the setpoint (in the example: 50 °C ± 1 K) for the first time (section A). When the temperature violates the set tolerance band around the setpoint (in the example in Fig. 21: 50 °C ± 2 K) – e.g. if the door is opened during operation (section B of illustration) – the alarm is set off. The ASF alarm is automatically triggered as soon as 50 % of the set tolerance band of the setpoint (in the example: 50 °C ± 1 K) are reached again (section C). If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in this example: The setpoint is changed from 50 °C to 25 °C, section D) until the tolerance range of the new temperature setpoint is reached again (section E). 52°C 48°C ASF active 52°C 48°C ASF alarm ASF active ASF active 27°C 23°C °C 50°C 25°C t A B C D E AUTO AUTO AUTO Fig. 21 Schematic diagram of how the ASF temperature monitoring works Adjusting temperature monitoring ►For cooled incubators IPPplus: Press the activation key to the left of the ALARM display. The min setting (under- temperature protection) is automatically activated. Continue with point 1. min 000°C ALARM max 000°C auto off 99K - + ALARM 0.0K min °C auto max 40.0°C 15 0 . +- ►For constant climate chambers HPP: Press the activation key to the left of the ALARM display. The temperature monitoring setting is automatically activated ( ). min 000°C ALARM max 000°C auto off 99K - + ALARM max 40.0°C auto 0.0K min 15.0°C +-

D24030 | Date 11/2018

Operation and control Save the selection by pressing the confirma- tion key. The min setting (undertemperature protection) is automatically activated. ALARM max auto 40 0.0K min 15.0°C .0°C +- 1. By turning the turn control, adjust the desired lower alarm limit value, in the example on the right 35.5 °C. The lower alarm limit value cannot be set higher than the top one. If no undertem- perature protection limit is required, set the lowest temperature. ALARM 0.0K min °C auto max °C 35 5 . 40 0 . +- 2. Press the confirmation key to confirm. The max display (overtemperature protec- tion) is activated. ALARM 0.0K min °C auto max °C 35 5 . 40 0 . +- 3. By turning the turn control, adjust the desired upper alarm limit value, in the example on the right 38.5 °C. The monitoring temperature must be set sufficiently high above the maximum set temperature. We recommend 1 to 3 K. ALARM 0.0K min °C auto max °C 35.5 38 5 . +- 4. Accept the upper alarm limit value by pressing the confirmation key. The setting of the automatic temperature monitor (ASF) is automatically activated (auto). ALARM max 38.5°C 0.0K min 35.5°C auto +- 5. With the turn control, select ON () or OFF (). ALARM max 38.5°C 0.0K min 35.5°C auto +- 6. Press the confirmation key to confirm. The ASF tolerance band setting is acti- vated. ALARM K min °C auto max 38.5°C 35 5 . 0.0 +-

D24030 | Date 11/2018

37 Operation and control 7. With the turn control, adjust the desired tolerance band, e.g. 2.0 K. We recommend a tolerance band of 1 to 3 K. ALARM K min °C auto max 38.5°C 35 5 . 1.0 +- 8. Press the confirmation key to confirm. Temperature monitoring is now active. ALARM max 38.5°C 1.0K min 35.5°C auto +- In the menu, you can set if an acoustic signal should be triggered on alarm (see page 57) 5.5.2 Humidity monitoring (for constant climate chambers HPP only) If humidity monitoring has been triggered, this is indicated by the humidity display: the actual humidity is highlighted in red and a warning symbol is shown ( Fig. 22 ). If the acoustic alarm has been activated in the Sound menu (see page 57), which is indicated by the speaker symbol , the alarm is additionally signalled by an intermittent acoustic signal. Information on what to do in this case are provided in the chapter Malfunctions, warning and error messages on page 40. Adjusting humidity monitoring 1. Press the activation key to the left of the ALARM display. The temperature moni- toring setting is automatically activated. min 000°C ALARM max 000°C auto off 99K - + ALARM max 40.0°C auto 0.0K min 15.0°C +- 2. Turn the turn control until the humidity monitoring entry is highlighted. ALARM min 40.0%rh max 60.0%rh 3. Accept the selection by pressing the con- firmation key. The lower humidity alarm limit is automatically highlighted. ALARM min 40.0%rh max 60.0%rh 4m min 000°C A ON LIGHT ON ON ON ON LIGH IGH IGHT ON LIGHT FEUCHTE 30.0%rh Set 30.0%rh 0 °C °C °C 22.4 EMP °C Set 37.0 L A auto min 35 0m 3:30 HUMIDITY Set 70.0%rh %rh 75.4 m FEUCHTE 30.0%rh Set 30.0%rh C °C Set 37.0 m 0 HUMIDITY Set 70.0%rh %rh 75.4 Fig. 22 Humidity monitoring triggered

D24030 | Date 11/2018

Operation and control 4. By turning the turn control, adjust the desired lower alarm limit, in the example on the right 50 % rh. ALARM min 50.0%rh max 60.0%rh 5. Accept the selection by pressing the con- firmation key. The upper humidity alarm limit is automatically highlighted. ALARM min 50.0%rh max 60.0%rh 6. By turning the turn control, adjust the desired upper alarm limit, in the example on the right 70 % rh. ALARM min 50.0%rh max 70.0%rh 7. Accept the selection by pressing the confirmation key and leave the Alarm display by pressing the activation key on the side. Humidity monitoring is now active. ALARM min 50.0%rh max 70.0%rh 5.6 Graph The GRAPH display provides an overview of the chronological sequence of the set values and the actual values as a curve. 5.6.1 Temperature profi le 1. Press the activation key to the right of the GRAPH display. The display is enlarged and the temperature profile shown. 0 4 8 12 16 20 24 °C 40 20 60 80 100 Fr 20.10.2010 20:34 12.09.2012 14.00 16.00 18.00 38 39 40 ►To change the time frame to be displayed: Press the activa- tion key next to the  ar- row symbols. The time frame to be displayed can now be changed by turning the turn control. 0 4 8 12 16 20 24 °C 40 20 60 80 100 Fr 20.10.2010 20: 12.09.2012 14.00 16.00 18.00 38 39 40

D24030 | Date 11/2018

39 Operation and control ►To zoom the graph in or out: Press the activation key next to the magnifying glass sym- bol. Select whether you want to zoom in or out (+/–) with the turn control and confirm your selection by pressing the confirmation key. .2010 20:34 12 To close the graphical representation, again press the activation key which you have used to activate it. 5.6.2 Humidity profi le (only for constant climate chambers HPP) 1. Activate graphic representa- tion as described above and then press the activation key next to the parameter selec- tion. 0 4 8 12 16 20 24 °C 40 20 60 80 100 Fr 20.10.2010 20 12.09.2012 14.00 16.00 18.00 38 39 40 2. Select humidity with the turn control. °C 40 20 60 80 100 Fr 20.10.2010 20 12.09.2012 38 39 40 3. Press the confirmation key to confirm. The humidity profile is displayed. You can change the display range as described above, as well as extend or reduce it. 0 4 8 12 16 20 24 °C 40 20 60 80 100 Fr 20.10.2010 20:34 12.09.2012 14.00 16.00 18.00 40 60 80 %rh 5.7 Ending operation 1. Switch off active appliance functions (turn back the set values). 2. Remove the chamber load. 3. For constant climate chambers HPP: Check the freshwater tank and fill up if necessary (see page 24). 4. Switch off appliance (Fig. 23). ON ONN ON O Fig. 23 Switching off the appliance.

D24030 | Date 11/2018

Malfunctions, warning and error messages 6. Malfunctions, warning and error messages Warning! After removing covers, live parts may be exposed. You may receive an electric shock if you touch these parts. Malfunctions requiring work inside the appliance may only be rectifi ed by electricians. Ob- serve the separate service manual for this. Do not try to rectify appliance errors yourself but contact the MEMMERT customer service department (see page 2) or an authorised service point. In case of enquiries, please always specify the model and appliance number on the nameplate (see page 13). 6.1 Warning messages of the monitoring function If the acoustic alarm has been activated in the Sound menu (see page 57) which is indicated by the speaker symbol , the alarm is additionally signalled by an intermittent acoustic signal. If the confirmation key is pressed, the acoustic alarm can be temporarily switched off until the next alarm event occurs. 6.1.1 Temperature monitoring Description Cause Action See Temperature alarm and "ASF" are displayed TEMP ASF Set 38.5 °C 40.4°C Automatic tem- perature monitor (ASF) triggered Check if the door is closed. Close the door. Extend the ASF tolerance band If the alarm continues: Contact customer service page 37 page 2 Temperature alarm and "TWW" are displayed TEMP TWW Set 38.5 °C 42.4°C The adjustable undertemperature / overtemperature controller (TWW) has assumed heat- ing control. Increase the difference between the monitoring and setpoint temperature – by either increas- ing the max value of the temper- ature monitoring or decreasing the setpoint temperature. If the alarm continues: Contact customer service page 36 page 2

D24030 | Date 11/2018

41 Malfunctions, warning and error messages 6.1.2 Humidity monitoring (for constant climate chambers HPP only) Description Cause Action See Error symbol HUMIDITY Set 55.0%rh %rh 55.4 Water tank empty Fill the water tank with deminera- lised/distilled water and press the confirmation key page 23 Alarm display (MaxAl) HUMIDITY MaxAl Set 70.0%rh %rh 75.4 Upper humidity limit exceeded Open the door for 30 sec. and wait to see if the appliance reliab- ly adjusts to the setpoint. If the error occurs again, contact customer service. page 2 Alarm display (MinAl) HUMIDITY MinAl Set 60 .0%rh %rh 55.4 Humidity below lower limit Check if the door is closed. Check the water supply and the filling level of the water tank. If required, refill the water tank. If the error occurs again, contact customer service. page 23 page 2 6.2 Malfunctions, operating problems and appliance errors Error description Cause of error Rectifying errors See Displays are dark External power supply was interrupted Check the power supply page 23 Miniature fuse, appliance fuse or power module faulty Contact customer service page 2 Displays cannot be activated Appliance locked by user ID Unlock with user ID page 59 The appliance is in pro- gramme, timer or remote control mode (mode "Write" or "Write + Alarm") Wait until the end of the programme or timer mode or switch off the remote control Displays suddenly look different Appliance is in "wrong" mode Change to operating or menu mode by pressing the MENU key

D24030 | Date 11/2018

Malfunctions, warning and error messages Error description Cause of error Rectifying errors See Display T:E-3 in the temperature display TEMP T:E-3 Set 37.0 °C 37.4°C Temperature operating sensor defective. The moni- toring sensor takes over the measurement function. ►The appliance can be temporarily operated ►Contact customer service as soon as possible page 2 Error message AI E-3 in the temperature display TEMP AI E-3 Set 37.0 °C 37.4°C Temperature monitoring sensor defective. The ope- rating sensor takes over the measurement function. ►The appliance can temporarily be kept in service ►Contact customer service as soon as possible page 2 Error message E-3 in the temperature display TEMP Set 45.0 °C E-3 °C Operating and monitoring sensor defective ►Switch off appli- ance. ►Remove the cham- ber load ►Contact customer service page 2 Error message E-6 in the humidity display HUMIDITY Set 50.0%rh %rh E-6 Humidity sensor defective ►No humidity control possible ►Contact customer service page 2 When switching on the appliance, the start animation is displayed in another colour than white ► Cyan : Not enough storage space on the SD card ►Red : The system files could not be loaded ►Orange : The fonts and images could not be loaded Contact customer service Contact customer service Download the firmware update at memmert. com and install it page 2 page 2

D24030 | Date 11/2018

43 Malfunctions, warning and error messages 6.3 Power failure In case of a power failure, the appliance operates as follows: In manual mode After power supply has been restored, operation is continued with the parameters set. The time and duration of the power failure are documented in the log memory. In timer or programme mode In case of an interruption of the power supply of less than 60 minutes, the current pro- gramme is continued from the point at which it was interrupted. For longer interruptions of the power supply, all appliance functions (heating, fan etc.) are switched off. In remote control mode: The previous values are restored. If a programme has been initiated via remote control, it is continued.

D24030 | Date 11/2018

Menu mode 7. Menu mode In menu mode, you can make basic settings, load programmes and export protocols, as well as adjust appliance parameters. Caution: Before changing menu settings, read the description of the respective functions on the fol- lowing pages to avoid possible damage to the appliance and/or chamber load. To enter menu mode, press the MENU key. To exit the menu mode at any time, press the MENU key again. The appliance then returns to manual mode. Only changes accepted by pressing the confirmation key are saved. 7.1 Overview Press the MENU key to change between the displays in menu mode: SETUP SIGNALTÖNE PROG ZEIT UND DATUM CALIB USER ID PROTOCOL LANGUAGE TIME SOUND ON ONN ON O 1 2 3 4 5 12 14 15 13 6 7 18 19 8 9 10 11 16 17 Fig. 24 ControlCOCKPIT in menu mode 1 Language selection activation key 2 Language selection display 3 Date and time display 4 Date and time setting activation key 5 Exit menu mode and return to operating mode 6 Setup activation key (basic appliance set- tings) 7 Setup display (basic appliance settings) 8 Adjustment display 9 Adjustment activation key 10 Turn control for adjustment 11 Confirmation key (accepts setting made with the turn control) 12 Programme setup activation key 13 Programme setup display 14 Protocol display 15 Protocol activation key 16 Acoustic signal adjustment activation key 17 Acoustic signal adjustment display 18 User ID display 19 User ID activation key

D24030 | Date 11/2018

45 Menu mode 7.2 Basic operation in menu mode using the example of language selection In general, all settings in menu mode are done just like in manual mode: Activate the re- spective display, use the turn control for setting and press the confirmation key to accept the change. A more detailed description is provided in the following, using the example of language selection. 1. Activate the desired parameter (in this example the language). To do so, press the corresponding activation key on the left or right of the respective display. The activated display is enlarged. ENGLISH DEUTSCH ESPANOL POLSKI CESTINA FRANCAIS If you want to exit or cancel your settings, again press the activation key which you have used to activate the display. The appliance returns to the menu overview. Only the settings that you have con- firmed by pressing the confirmation key before cancelling the setting procedure are accepted. SETUP SIGNALTÖNE CALIB LANGUAGE TIME ON ONN ON O 2. With the turn control, select the desired new setting, e.g. Español (Spanish). 3. Save the setting by pressing the confir- mation key. ENGLISH DEUTSCH MAGYAR ITALIANO ESPANOL FRANCAIS POLSKI CESTINA 4. To return to the menu overview, press the activation key again. SETUP SIGNALTÖN CALIB LANGUAGE TIM You can now ►activate another menu function by press- ing the corresponding activation key or ►return to manual mode by pressing the MENU key. Balance Einheit IP Adresse Alarm Temp Timer Mode Unit IP address Subnet mask Alarm Temp Timer mode Slide-in unit ENGLISH DEUTSCH MAGYAR ITALIANO ESPANOL FRANCAIS POLSKI CESTINA

D24030 | Date 11/2018

Menu mode All other settings can be made accordingly. The settings possible are described in the follow- ing sections. If no new values are entered or confirmed for approx. 30 seconds, the appliance automati- cally returns to the main menu and restores the former values. 7.3 Setup 7.3.1 Overview In the SETUP display, you can set the following parameters: ►the IP address and Subnet mask of the appliance's Ethernet interface (for connection to a network) ►the Unit of the temperature display (°C or °F, see page 47) ►how the digital backwards counter with target time setting works (Timer mode, see page 48) ►the type of the slide-in unit (Grid or Shelf, see page 48) ►the heat output distribution (Balance, see page 49) ►Remote control (see page 49) ►Gateway (see page 50) If the SETUP menu contains more entries than can be displayed, this is indicated by the display “1/2”. This means that there is a second „page“ of entries. To display the hidden entries, use the turn control to scroll beyond the lowest entry. The page display changes to “2/2”. 7.3.2 IP address and subnet mask If you want to operate one ore more appliances in a network, each appliance must have its own unique IP address for identification. By default, each appliance is delivered with the IP address 192.168.100.100. LAN 1: 192.168.1.233 192.168.1.216 LAN 2: 192.168.1.215 LAN 3: 192.168.1.241 192.168.1 Editor Programmname Simulation Protokoll - + - + INP 250 Test 01 180.0°C i HPP 250 Labor i 37.0°C 44.4%rh 15 15 STAND BY Programm n e g n u ll e t s n i E ll o k o t o r P t ä r e G e fli H n e k c u r D AtmoCONTROL Fig. 25 Operation of several appliances in a network (schematic example) Setup Unit IP adress 255.145.136.2 Subnet mask 255.255.0.0 °C F 1/2

D24030 | Date 11/2018

47 Menu mode 1. Activate the SETUP display. The entry IP address is automatically highlighted. SETUP Balance Einheit IP Adresse +30% 192.168.100.100 Alarm Temp °C F TWW TWB Timer Mode SETUP Unit IP address 192.168.100.100 Subnet mask 255.255.0.0 Alarm Temp °C F TWW TWB Timer mode Slide-in unit Grid Shelf 2. Accept the selection by pressing the con- firmation key. The first three digits of the IP address are automatically selected. SETUP Unit IP address 192.168.100.100 Subnet mask 255.255.0.0 Alarm temp °C F TWW TWB Timer mode 3. With the turn control, set the new num- ber, e.g. 255. SETUP Unit IP address 255.168.100.100 Subnet mask 255.255.0.0 Alarm temp °C F TWW TWB Timer mode 4. Accept the selection by pressing the confirmation key. The next three digits of the IP address are automatically selected. Setting these is done according to the description above. SETUP Unit IP address 255.168.100.100 Subnet mask 255.255.0.0 Alarm temp °C F TWW TWB Timer mode 5. After setting the last three digits, accept the new IP address by pressing the con- firmation key. The selection returns to the overview. The subnet mask is set accordingly. SETUP Unit IP address 255.145.136.225 Subnet mask 255.255.0.0 Alarm temp °C F TWW TWB Timer mode 7.3.3 Unit Here, you can choose whether the temperature is displayed in °C or °F. SETUP Unit IP address 255.145.136.225 Subnet mask 255.255.0.0 Alarm temp °C °F TWW TWB Timer mode

D24030 | Date 11/2018

Menu mode 7.3.4 Timer mode Here, you can choose whether the digital backwards counter with target time setting (see page 30) should run setpoint-de- pendent or not. This determines whether the timer should not start until a tolerance band of ±3 K around the set temperature is reached (Fig. 26, B) or if it should start right after activation (A). tTimer tTimer A B °C/°F t 6 K ž ž Fig. 26 Timer Mode A Timer independent of setpoint: Timer starts right after activation B Timer setpoint-dependent: Timer does not start until tolerance band is reached 7.3.5 Type of the slide-in unit ( Grid or Shelf) Here, you have to set the type of the slide-in unit (grid or shelf) used. The selection Shelf enables you to adjust the control function to the different air flow characteristics in the interior when using optional sliding shelves instead of the grids that are part of the standard delivery. Unit IP address 255.145.136.225 Subnet mask 255.255.0.0 Alarm temp °C F TWW TWB Timer mode S i i i Unit IP adress 255.145.136.225 Subnet mask 255.255.0.0 °C °F Timer mode Alarm temp TWW TWB Slide-in unit Grid Shelf

D24030 | Date 11/2018

49 Menu mode 7.3.6 Balance (only for model sizes 260, 400 and 750/1060) For appliances of the sizes 260, 400 and 750/1060, application-specific correction of the heat output distribution (balance) between the upper and lower heating groups is possible. The adjustment range is from –50 % to +50 %. upper heat output upper heat output lower heat output -20% lower heat output +30% SETUP Balance -20 % 2/2 SETUP Balance +30 % 2/2 Fig. 27 Heat output distribution (example): The –20 % (left) setting causes the lower heating groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the default heat output distribution. 7.3.7 Remote control In the setup entry Remote control, you can set whether the appliance should be controlled via remote control and if so, in which mode. These settings are available: ►Off ►Read Only ►Write ►Write+Alarm When the appliance is in remote control mode, the symbol appears in the temperature display. In the settings Write and Write+Alarm, the appliance cannot be controlled at the ControlCOCKPIT until the remote control has been switched off (setting Off) or set to Read Only. In order to use the remote control function, program- ming skills and special libraries are required. Setup Balance Remote Control Off Gateway 192.168.5 .1 +30 % 2/2 23.2°C TEMP Set 38.0°C

D24030 | Date 11/2018

Menu mode 7.3.8 Gateway The setup entry Gateway is used to connect two networks with different protocols. The gateway is set the same way as the IP address (see page 46). 7.4 Date and Time In the Time display, you can set date and time, time zone and daylight saving time. Always set the time zone (and summer time yes/no) before you set the date and time. Avoid changing the set time after that since this can lead to gaps or overlapping when recording measured values. If you still need to change the time, you should not run a programme immediately before or after doing so. 1. Activate the time setting. To do so, press the activation key on the right side of the Time display. The display is enlarged and the first adjustment option (Date) automatically highlighted. Date and time Time Time zone Daylight savings 12.05.2012 12:00 GMT 01:00 Date 2. Turn the turn control until Time zone is highlighted. Time Daylight savings 12.05.2012 12:00 GMT 01:00 Date Time zone 3. Accept the selection by pressing the confirmation key. Time Daylight savings 12.05.2012 12:00 GMT Date Time zone 01:00 4. Set the time zone of the installation site with the turn control, e.g. 00:00 for Great Britain, 01:00 for France, Spain or Ger- many. Accept the selection by pressing the confirmation key. Time Daylight savings 12.05.2012 12:00 GMT Date Time zone 00:00 5. With the turn control, select the Daylight savings entry Time 12.05.2012 12:00 GMT 00:00 Date Time zone Daylight savings Balance Remote Control Off Gateway 192.168.5 .1 +30 % 2/2

D24030 | Date 11/2018

51 Menu mode 6. Accept the selection by pressing the confirmation key. The adjustment options are highlighted. Time 12.05.2012 12:00 GMT 00:00 Date Time zone Daylight savings 7. Set daylight savings to off () or on () with the turn control – in this case on (). Save the setting by pressing the confirmation key. Time 12.05.2012 12:00 GMT 00:00 Date Time zone Daylight savings Daylight saving time and standard time are not changed automatically. For this reason, please keep in mind to adjust them at the beginning of each period. 8. Now, set date (day, month year) and time (hours, minutes) in the same way. Accept the selection by pressing the confirma- tion key. Time Time zone Daylight savings . .2012 12:00 GMT 00:00 Date 27 05 7.5 Calibration To guarantee perfect control, we recommend to calibrate the appliance once a year. 7.5.1 Temperature adjustment The appliances are temperature calibrated and adjusted at the factory. In case readjustment should be necessary later on – for example due to influence of the chamber load – the ap- pliance can be calibrated customer-specifically using three calibration temperatures of your choice: ►Cal1 Temperature calibration at low temperature ►Cal2 Temperature calibration at medium temperature ►Cal3 Temperature calibration at high temperature For temperature adjustment, you will need a calibrated reference measuring device.

D24030 | Date 11/2018

Menu mode CAL 1 +0,5°C CAL 3 +1,6°C CAL 2 -0,4°C Factory calibration 20°C 30°C 40°C 0°C 10°C Fig. 28 Schematic example of temperature adjustment Example: Temperature deviation at 30 °C should be corrected. 1. Press the activation key to the right of the CALIB display. The display is enlarged and the temperature adjustment option is automatically highlighted. Cal1 °C -0,2K Cal2 100.0 °C +0,1K Cal3 180.0 °C -0,2K JUSTIEREN Temperatur letzte Justierung 12.10.2012 12:00 40.0C K 0,2 - C K 0,1 + C K 0,2 - Calibration Last updated 12.10.2012 12:00:00 Cal1 Cal2 Cal3 20.0 40.0 5.0 Temperature Humidity 2. Press the confirmation key repeatedly, until the calibration temperature Cal2 is selected. C K 0,2 - C K 0,1 + C K 0,2 - Temperature Cal1 Cal2 Cal3 40.0 5.0 20.0 3. With the turn control, set the calibration temperature Cal2 to 30 °C. C K 0,2 - C K 0,1 + C K 0,2 - Temperature Cal1 Cal2 Cal3 40.0 5.0 30.0 4. Save the setting by pressing the confir- mation key. The corresponding calibra- tion value is automatically highlighted. C K 0,2 - C K C K 0,2 - Temperature Cal1 Cal2 Cal3 40.0 5.0 30.0 0,1 +

D24030 | Date 11/2018

53 Menu mode 5. Set the calibration value to 0.0 K and ac- cept the setting by pressing the confirma- tion key. C K 0,2 - C K C K 0,2 - Temperature Cal1 Cal2 Cal3 40.0 5.0 30.0 0,0 6. Position the sensor of a calibrated refer- ence instrument centrally in the appli- ance's working chamber. 7. Close the door and, in manual mode, adjust the set temperature to 30 °C. 21.4°C TEMP 30.0 °C Set 8. Wait until the appliance reaches the set temperature and displays 30 °C. The reference instrument for example displays 31.6 °C. 30.0°C TEMP Set 30.0°C 31.6 °C 9. In the SETUP, adjust the calibration value Cal2 to +1.6 K (actual value measured minus setpoint temperature) and save the setting by pressing the confirmation key. C K 0,2 - C K C K 0,2 - Temperature Cal1 Cal2 Cal3 40.0 5.0 30.0 1,6 + 10. After the calibration procedure, the temperature measured by the reference instrument should now also be 30 °C. 30.0°C TEMP Set 30.0°C 30,0 °C With Cal1, a calibration temperature below Cal2 can be programmed accordingly, and with Cal3, a temperature above. The minimum difference between the Cal values is 10 K. If all calibration values are set to 0.0 K, the factory calibration settings are restored.

D24030 | Date 11/2018

Menu mode 7.5.2 Humidity adjustment (for constant climate chambers HPP only) Humidity control of the constant climate chamber HPP can be adjusted according to customer requirements by means of three freely selectable balance points. For each selected calibration point, a positive or negative compensation correction value between –10 % and +10 % can be set (Fig. 29). For humidity adjustment, you will need a calibrated reference measuring device. rh 20 +3,0 % rh 90 -1,5 % 60% 40% 0% 20% 80% 100% rh 50 +1,5 % Factory calibration Fig. 29 Humidity calibration (example) Example: Humidity deviation at 90 % should be corrected. 1. Press the activation key to the right of the CALIB display. The display is enlarged and the temperature adjustment option is automatically highlighted. Cal1 °C -0,2K Cal2 100.0 °C +0,1K Cal3 180.0 °C -0,2K JUSTIEREN Temperatur letzte Justierung 12.10.2012 12:00 40.0C K 0,2 - C K 0,1 + C K 0,2 - Calibration Last updated 12.10.2012 12:00:00 Cal1 Cal2 Cal3 20.0 40.0 5.0 Temperature Humidity 2. Turn the turn control until Humidity is highlighted. %rh %rh % 0,5 - % 1,0 + Temperature Cal1 Cal2 40.0 50.0 Humidity %rh % 1,0 + Cal3 80.0 3. Press the confirmation key repeatedly, until the calibration point Cal2 is selected. %rh %rh % 0,5 - % 1,0 + Temperature Cal1 Cal2 40.0 Humidity 50.0 %rh % 1,0 + Cal3 80.0 4. With the turn control, set the calibration point Cal2 to 60 % rh. %rh %rh % 0,5 - % 1,0 + Temperature Cal1 Cal2 40.0 Humidity 60.0 %rh % 1,0 + Cal3 80.0

D24030 | Date 11/2018

55 Menu mode 5. Save the setting by pressing the confir- mation key. The corresponding calibra- tion value is automatically highlighted. %rh %rh % 0,5 - % JUSTIEREN Temperature Cal1 Cal2 40.0 Humidity 60.0 1,0 + %rh % 1,0 + Cal3 80.0 6. Set the calibration value to 0.0 K and ac- cept the setting by pressing the confirma- tion key. %rh %rh % 0,5 - % Temperature Cal1 Cal2 40.0 Humidity 60.0 0,0 %rh % 1,0 + Cal3 80.0 7. Position the sensor of the calibrated refer- ence instrument centrally in the appli- ance's working chamber. 8. Close the door and, in manual mode, adjust the set humidity to 60 % rh. HUMIDITY 29.5%rh Set 60.0%rh 9. Wait until the appliance reaches the set humidity and displays 60 % rh. The reference instrument for example displays 58.5 % rh. HUMIDITY 60.0%rh Set 60.0%rh 58.5 %rh 10. In the SETUP, adjust the calibration value Cal2 to -1.5 % (actual value measured minus setpoint humidity) and save the setting by pressing the confirmation key. %rh %rh % 0,5 - % Temperature Cal1 Cal2 40.0 Humidity 60.0 -1.5 %rh % 1,0 + Cal3 80.0 11. After the calibration procedure, the hu- midity measured by the reference instru- ment should now also be 60 % rh. HUMIDITY 60.0%rh Set 60.0%rh 60.0 %rh

D24030 | Date 11/2018

Menu mode 7.6 Program In the PROG display, programmes created using the AtmoCONTROL software can be trans- ferred to the appliance and saved on a USB storage medium. Here, programme to be used in manual mode can also be selected (see page 31) and programmes can be deleted. To load a programme from a USB stor- age medium: Connect the USB storage medium with the saved programme(s) to the interface on the right side of the control panel. 1. Activate the programme display. To do so, press the activation key on the left side of the PROG display. The display is enlarged and the entry Select automatically highlighted. The programmes available for activation are shown on the right. The programme currently available for use – in this example Test 012 – is highlighted in orange. Program Delete Test 012 Test 013 Test 014 Test 015 Test 017 Test 016 Select 2. Call the Select function by pressing the confirmation key. All programmes available are displayed, including the ones saved on the USB storage medium (identified by the USB symbol). The programme currently available for use is highlighted in orange. Select Delete Test 012 Test 022 Test 013 Test 014 Test 015 Test 022 3. With the turn control, select the pro- gramme you want to make available for use. Select Delete Test 012 Test 022 Test 013 Test 014 Test 015 Test 023 4. Accept the selection by pressing the confirmation key. The programme is now loaded, which is indicated by the transfer symbol. Select Delete Test 012 Test 022 Test 013 Test 014 Test 015 Test 023 5. As soon as the programme is ready, the selection returns to Select. To start the programme: As described on page 31, return to manual mode by pressing the MENU key. Delete Test 012 Test 022 Test 013 Test 014 Test 015 Test 023 Select You can now remove the USB storage medium. To delete a programme, select Delete with the turn control and select the programme to be deleted the same way you can select a programme for activation.

D24030 | Date 11/2018

57 Menu mode 7.7 Sound In the Sound display, it can be define whether or not the appliance should emit acoustic signals and, if yes, on which events: ►on the press of a key ►at the end of a programme ►on alarm ►if the door is open 1. Activate the acoustic signal adjustment. To do so, press the activation key on the left side of the SOUND display. The display is enlarged. The first category (in this case Key sound) is automatically highlighted. On the right, the current set- tings are shown on. Sound Keysound At the end If door open On alarm If you want to edit another list entry: Turn the turn control until the respective entry – e.g. If door open (Special accessories) – is highlighted in colour. Keysound At the end If door open On alarm 2. Save the selection by pressing the confir- mation key. The adjustment options are automatically highlighted. Keysound At the end If door open On alarm 3. With the turn control, select the desired setting – in this example OFF () . Keysound At the end If door open On alarm 4. Save the setting by pressing the confir- mation key. If an acoustic alarm sounds, it can be turned off by pressing the confirmation key. Keysound At the end If door open On alarm

D24030 | Date 11/2018

Menu mode 7.8 Protocol The appliance continually logs all relevant measured values, settings and error messages at 1-minute intervals. The internal log memory is of the continuous memory type. The logging function cannot be switched off and is always active. The measured data are stored in the appliance, safe from manipulation. If the power supply is interrupted, the time of the power failure and voltage recovery are stored in the appliance. You can export the protocol data for different periods to a USB storage medium via the USB interface or, via Ethernet, import them to the AtmoCONTROL software for graphical represen- tation, print-out or storage. The log memory of the appliance is not modified or deleted by reading it out. 1. Connect the USB storage medium to the interface on the right side of the control panel. 2. Activate the protocol. To do so, press the activation key on the right side of the PROTOCOL display. The display is enlarged and the period This Month au- tomatically highlighted. To select another logging period, use the turn control. Protocol 12.10.2012 12:00 control range 3. Save your selection by pressing the confir- mation key. The transfer starts and a status symbol indicates the progress. This Year Complete control range This Month 4. As soon as the transfer is complete, a check mark appears in front of the period selected. The USB storage medium can now be removed. This Year Complete control range This Month For a description of how to import and process protocol data in AtmoCONTROL or read them out via Ethernet, please observe the separate AtmoCONTROL manual.

D24030 | Date 11/2018

59 Menu mode 7.9 User-ID 7.9.1 Description With the User-ID function, you can lock the settings of individual (e.g. temperature) or all parameters, so that they cannot be changed at the appliance by accident or unauthorised persons. You can also lock setting options in menu mode (e.g. adjustment or date and time settings) this way. If adjustment options are locked, this is indicated by the lock symbol in the respective display (Fig. 30). User ID data are entered in the AtmoCONTROL software and saved on the USB storage medium. The USB storage medium is thus acting as a key: Parameters can only be locked or unlocked if it is connected. A description of how to create a user ID in AtmoCONTROL is provided in the separate AtmoCONTROL manual. 7.9.2 User ID activation and deactivation 1. Connect the USB storage medium with the user ID data to the interface on the right side of the control panel. 2. Activate the user ID. To do so, press the activation key on the right side of the User-ID display. The display is enlarged and the entry Activate automatically highlighted. USER-ID Deactivate Activate 3. Confirm the activation by pressing the confirmation key. The new user ID data are transferred from the USB storage medium and activated. As soon as activa- tion is complete, a check mark appears in front of the corresponding entry. USER-ID Deactivate Activate 4. Remove the USB storage medium. Locked parameters are indicated by the lock symbol in the respective display (Fig. 30 ). To unlock the appliance, connect the USB storage medium, activate the User-ID entry and select Deactivate. 22.4 40 4 TEMP Set °C °C Fig. 30 Temperature adjustment at appliance locked (example)

D24030 | Date 11/2018

Maintenance and service 8. Maintenance and service 8.1 Cleaning Warning! Danger due to electric shock. Before doing any maintenance work, pull out the mains plug. Warning! In case of appliances of a certain size, you can get accidentally locked in, which is life-threatening. Do not climb into the appliance! Caution! Danger of cuts due to sharp edges. Always wear gloves when wor- king in the chamber interior. 8.1.1 Working chamber and metal surfaces Regular cleaning of the easy-to-clean working chamber prevents build up of material remains that could impair the appearance and functionality of the stainless steel working chamber over time. The metal surfaces of the appliance can be cleaned with normal stainless steel cleaning agents. Make sure that no rusty objects come into contact with the working chamber or with the stainless steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots should appear on the surface of the working chamber due to impurities, the affected area must be immediately cleaned and polished. 8.1.2 Plastic parts Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic or solvent-based cleaning agents. 8.1.3 Glass surfaces Glass surfaces can be cleaned with a commercially available glass cleaner. 8.1.4 Peltier cooling module In order to guarantee perfect function and long lifetime of the Peltier cooling modules, it is absolutely essential that you remove dust deposits from the heat sink on the rear of the appliance (with a vacuum cleaner, paintbrush or bottle brush, depending on the amount). To make cleaning easier, the cover can be removed after the screws have been loosened (Fig. 31). Fig. 31 Cover of the Peltier cooling modules on the rear of the appliance

D24030 | Date 11/2018

61 Maintenance and service 8.2 Regular maintenance Once a year, grease the moving parts of the doors (hinges and lock) with thin silicone grease and check that the hinge screws are not loose. To guarantee perfect control, we recommend to calibrate the appliance once a year (see page 51). 8.3 Repairs and service Warning! After removing covers, live parts may be exposed. You may receive an electric shock if you touch these parts. Disconnect the mains plug before removing any covers. Any work inside the appliance may only be performed by qualifi ed electri- cians. Repairs and service work are described in a separate service manual.

D24030 | Date 11/2018

Storage and disposal 9. Storage and disposal 9.1 Storage The appliance may only be stored under the following conditions: ►in a dry and enclosed, dust-free room ►frost-free ►disconnected from the power supply Before storage, remove water tube and empty the water tank (see page 24). 9.2 Disposal This product is subject to the Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) of the European Parliament and of the Council. This appliance has been brought to market after August 13th, 2005 in countries which have already integrated this directive into their national laws. It may not be disposed of in normal household waste. For disposal, please contact your dealer or the manufacturer. Any appliances that are infected, infectious or contami- nated with materials hazardous to health are excluded from return. Please also observe all other regulations applicable in this context. Before disposing of the appliance, please render the door locking mechanism unusable, for example, to prevent playing children from being locked inside the appliance. There is a lithium battery in the ControlCOCKPIT of the appliance. Remove it and dispose of it in accordance with the regulations in your country (Fig. 32). Fig. 32 Removing the lithium battery Note for Germany: The appliance may not be left at public or communal recycling or collection points.

D24030 | Date 11/2018

63 Index Index A Accessories 16 Acoustic signals 51 Activation key 28 Adjustment 51 Alarm 35, 37, 40 Ambient conditions 11, 15, 16 Ambient temperature 16 Appliance error 41 ASF 33, 35 AtmoCONTROL 3, 12, 16, 28, 31, 56, 58, 59 B Balance 48 Basic device settings 44 Basic settings 44 C Calibration 51 Carrying 17 Cause of errors 41 Chamber load 26 Changes 8 Cleaning 60 Clock time 50 Communication interfaces 12 Compensation correction value 53 Condensation 11 Connections 11 ControlCOCKPIT 26, 27 Customer service 2 D Date and Time 50 Declaration of conformity 15 Decommissioning 62 Delete programme 56 Delivery 17, 18, 23 Dew point 11 Digital backwards counter 30 Dimensions 15 Disposal 62 Door 25 Dust deposits 60 E Electrical connection 11 Electronic temperature moni- toring 34 Emergency 8 Ending operation 39 End of programme 32 Error description 41 Error messages 40 Ethernet 12 Explosion protection 8 F Forklift truck 18 Fresh water container 23 G Gateway 50 Graph 38 Grid 48 H Hazards 7 Heat output distribution 49 Heat sink 60 Humidity 29 Humidity adjustment 54 Humidity calibration 54 Humidity monitoring 37, 38, 41 Humidity profile 39 I Installation site 19 Intended use 8 Interfaces 11 Interior lighting 30 IP address 46 L Language selection 45 Lighting 30 Loading the appliance 26 Log memory 43, 58 M Maintenance 60 Malfunctions 8, 40, 41 Manufacturer 2 Material 11 Menu 44, 50 Minimum clearances 19 Monitoring function 33 Monitoring temperature 33 N Nameplate 13 Network 12, 46 Normal mode 28, 29 O Operating 25 Operating modes 28 Operating personnel 7, 25 Operating problems 41 Operation 25 P Packaging material 18 Peltier cooling unit 60 Power failure 43 Power supply 58 Press of a key 57 Product safety 7 Programme 56 Programme mode 28, 31 Protocol 58 Pt100 temperature sensor 33 Putting into operation 23 R Rectifying errors 41 Regular maintenance 61 Remote control 49 S Safety regulations 6, 9 Service 61 Servicing 61 Setting parameters 28, 45 Setting up 17, 19 Setting up options 20 Setup 46 Shelf 48 Slide-in unit 48 Speaker symbol 37, 40 Storage after delivery 18 Switching off 39 Switching on 24

D24030 | Date 11/2018

Index T Tank holder 24 Technical data 14 Temperature 29 Temperature adjustment 51 Temperature comparison 51 Temperature deviation 52 Temperature monitor 33, 35 Temperature monitoring 33 Temperature profile 38 Temperature sensor 33 Tilt protection 21 Time 48 Timer 30 Timer mode 48 Transport 17, 18 Transport damage 18 Turn control 28 TWW 34 TWW temperature monitor- ing 34 U Unit 47 Unpacking 18 USB interface 12, 58 User ID 59 W Warning messages 12, 40 Water specifications 23 Water tank 23 Weight 14 Working range 10