INNO V30 Operation & User’s Manual

Specifications

Brand: INNO
Category: Measuring Instruments
Model: V30 (Operation & User’s Manual)
File Type: PDF
File Size: —
Pages: 8

Read Online

Read the INNO V30 Operation & User’s Manual online using the built-in PDF viewer.

Text of the manual

Thank you for choosing V30 Optical Multi Meter produced by

INNO! This manual is only subject to the users who have already purchased V30 Optical Multi Meters. Copying any part of this manual without any written permission by publisher is forbidden.

Chapter 1 Overview

③ ④ ⑥ ⑦ ⑨ ⑧ ⑩ ⑪ ⑤ ① ② 1. Optical fiber connector ( standard FC ) 2. LED lamp 5. Button for wavelength selection (available under“ OPM ” function) 3. VFL interface (2.5mm UPP) 4.LCD 8. Test indicator for 7. Power on/off or function selection button 9. Mini USB charger socket 10. RJ45 network test socket 11. RJ11 phone line test socket 6. Flicker on/off (available under the functions of “ VFL ” or “ LED ” ) Size:124×53×33 (L×W×H) Weight:155g network cable connection system

Chapter 2 Parameters

Parameters Optical Power Meter(OPM) Type of detector InGaAs Type of optical fiber interface 2.5mm of ST/SC/FC universal connectors Calibrated wavelengths 850/1300/1310/1490/1550/1625nm Measuring range Accuracy ± 0.02dB from +6 to -70dBm Linearity ± 2% Units dBm、mW 、uW Visual Fault Locator (VFL/ VLS) Output power 1mW Type of optical fiber interface 5KM of single mode fiber(SMF) Radiation safety class Class 2 Wavelengths 650nm ± 20nm Output modes Continuous /2Hz flicker lasers Battery reverse protection function Available Network cable /phone line test Type of interface RJ45/RJ11 Applicable network / phone lines UTP CAT-5/CAT-5e (2/4/6-core phone lines) Max.test distance 1KM Operating temperature from -10℃ to +60℃ Storage temperature from -20℃ to +70℃ Relative humidity <90% (non-condensing)

Chapter 3 Functions

3.1 power-on and power saving function: Power saving function: 3.2 Function selections: This instrument will shut off automatically after waiting but without receiving any operation instruction for about 10 minutes. If you want to disable this auto shut-off function and keep this instrument working, please press and hold the “MODE” button immediately when power on for about 2 seconds until “PERM” appears on the bottom right screen of the LCD, meaning that the power saving function has been canceled. It defaults to be under “OPM” function when power-on. 4 Function selections: It defaults to be under “OPM” function when power-on.4 functions including “OPM”,“VFL”,“LED”and“network Power-on: Quickly press “MODE” key to turn on the instrument. Power-off: Press and hold “MODE” key for 2 seconds or more until “OFF” displays on the screen. Note: This instrument will shut down automatically without receiving any operation instruction for 10 minutes. PERM cable /phone line tester”will be switched in loop mode. -ons including “OPM”, “VFL”, “LED” and “network cable/phone line tester” will be switched in loop mode. functi

3.4 Visual Fault Locator (VFL)

3.5 LED lamp 3.3 Optical Power Meter (OPM) You can switch the measuring and the units of numeric display include uW and dBm,which can be switched via “U” button. /1300/1310/1490/1550/1625nm “OPM”.The calibrating wavelen- You can select continuous optical wavelengths by pressing “T” button under the function of gths of this instrument are 850 You can select continuous mode or flicker laser mode with 2Hz of flicker frequency under the function of “VFL”. laser light mode of LED or flicker mode of SOS coding by pressing “U” button under the function of LED. 3.6 Null Function Description Tighten the OPM dust cap. “NULL”appears when "T" button and "U" button are pressed at the same time in OPM model.

The “Network cable/ phone line test”(NET) function is used in

network generic wiring and for verifying the link on-off condition of network including the sequence of testing cable connection systems. Test will begin when this NET function is enabled; the 8 indicators are used for indicating the sequence of testing cable connection systems. When the network cable or phone line is connected to this instrument and remote model, 8 indicators will blink in turn. 3.7 Network cable/phone line test Remote module

Company:INNO Instrument Inc.

Address:E-22F, 30, Songdomirae-ro, Yeonsu-gu, Incheon21990, Republic of Korea Tel: +82 32 837 5600 Fax: +82 32 837 5601 Email: [email protected] Please contact with INNO customer engineer or access to the following website for consultation: Any calibration, maintenance or repair to this instrument should be carried out by qualified maintenance staff. www.innoinstrument.com